Millipore Spectroquant Multy Operating Manual

Page 1
F3
Spectroquant
®
Multy
Operating manual
en
Page 2
4
Table of contents
1 Getting started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1.1 Package contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1.2 Inserting the rechargeable batteries/lithium powerpack. . . . . . . . .8
1.2.1 Lithium powerpack – Important information . . . . . . . . . . . . . .9
1.2.2 Charging the battery pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
1.2.3 Using non-rechargeable batteries . . . . . . . . . . . . . . . . . . . . . .10
1.2.4 Using the protective caps . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
1.3 Overview of the key functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
1.4 Starting the colorimeter the first time . . . . . . . . . . . . . . . . . . . . . .12
1.5 Overview of the mode menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
1.6 Setting the language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
1.7 Deleting data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
1.8 Setting the date and time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
1.9 Time and date display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
1.10 Automatic switch-off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
2 Operating mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
2.1 Selecting the method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
2.2 Measuring with test kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
2.3 Differentiation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
2.4 Altering the citation form. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
2.5 Measuring absorbances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
2.6 User countdown (timer function). . . . . . . . . . . . . . . . . . . . . . . . . . .24
2.7 Saving measurement results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
2.8 Retrieving saved measurement results. . . . . . . . . . . . . . . . . . . . . . .26
2.8.1 Retrieving all saved measurement results . . . . . . . . . . . . . . . .26
2.8.2 Retrieving saved measurement results
from a defined date range . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
2.8.3 Retrieving saved measurement results
from a defined code-No. range . . . . . . . . . . . . . . . . . . . . . . . .28
2.8.4 Retrieving saved measurement results
from a defined method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
2.9 Deleting saved measurement results . . . . . . . . . . . . . . . . . . . . 31
en
Page 3
en
5
3 Other functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
3.1 User-specific method list. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
3.1.1 Processing the user-specific method list . . . . . . . . . . . . . . . . .32
3.1.2 User-specific method list: Activate all methods. . . . . . . . . . .34
3.1.3 User-specific method list: Deactivate all methods. . . . . . . . .34
3.2 Profi mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
3.3 Acoustic signals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
3.3.1 Activating/deactivating the key beep . . . . . . . . . . . . . . . . . . .36
3.3.2 Activating/deactivating the signal beep . . . . . . . . . . . . . . . . .36
3.3.3 Activating/deactivating the countdown function
(observance of reaction times) . . . . . . . . . . . . . . . . . . . . . . . . .37
3.4 Setting the display contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
3.5 System info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
4 Trouble-shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
4.1 User messages on the display / Error messages . . . . . . . . . . . . . . .39
4.2 Other potential sources of error. . . . . . . . . . . . . . . . . . . . . . . . . . . .42
4.3 Avoiding errors in photometric measurements . . . . . . . . . . . . . . .43
Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Declaration of CE-Confirmity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Page 4
5 Contents of CD-ROM see enclosed CD
5.1 Overview of preprogrammed methods and analytical procedures
5.2 Standard solutions
5.2.1 Use of Spectroquant® CombiCheck and ready-to-use
standard solutions
5.2.2 Preparation of standard solutions
5.3 Printing measurement results
5.3.1 Setting the print parameters
5.3.2 Printing all measurement results
5.3.3 Printing measurement results from a defined date range
5.3.4 Printing measurement results from a defined code-No. range
5.3.5 Printing measurement results from a defined method
5.4 Transferring data to a PC
5.5 Software update via the internet
5.6 User methods
5.6.1 User concentration method
5.6.2 User polynomials
5.6.3 Deleting a user method
5.6.4 Printing data for a user method
5.6.5 Initializing the user-method system
5.7 User-specific calibration
5.7.1 Saving the user-specific calibration
5.7.2 Deleting the user-specific calibration
5.8 Calculating the Langelier saturation index
5.9 Technical specifications
en
6
Page 5
7
Getting started
h
c
i
k
f
b
g
d
e
a
en
1.1 Package contents
The standard contents of the Spectroquant® Multy Colorimeter package comprise the following items:
1 colorimeter in a plastic carrying case 1 beaker (plastic) 100 ml (a) 1
set of rechargeable batteries (7 nickel-cadmium batteries; type AA) (b
)
1
adapter for 16-mm ø round cells (c
)
3
round cells with lids, ø 16 mm (d
) 1 screwdriver (e) 1
cable for connection with a PC or printer (f
)
1
mains adapter, 100 - 240 V, 50 - 60 Hz with 4 wall-socket adapters (EU, UK, USA, AUS) (g
)
2
protective caps for connection terminals on the back (h)
1
lid for adapter (i
) 1 lithium battery (CR 2032; 3V) 3 round cells with lids, ø 24 mm (k
)
1 operating-instructions
manual
1 certificate of compliance
Page 6
8
1.2 Inserting the rechargeable
batteries/lithium powerpack
Before operating the system for the first time, the rechargeable batteries and the lithium powerpack included in the package must be installed. The batteries enclosed in the package are not charged.
A Battery pack: 7 nickel-cadmium batteries (type AA, 750 mAh) B Battery: lithium battery (type CR 2032, 3V) C Fuse: 1 A, delayed-action, 20 mm
1. Ensure that the Spectroquant® Multy Colorimeter is switched off.
2. Remove, where applicable, the cell from the measurement compartment.
3. Place the unit on its front on a clean, flat surface.
4. Remove the two screws (E) on the battery-compartment cover (F) on the bottom of the unit.
5.
Remove the batteries cover (F).
6. Remove any old batteries (A
) and/or the lithium powerpack (B).
7. Insert 7 new rechargeable batteries and/or the lithium powerpack
.
Make sure that the batteries are inserted correctly!
8.
Attach the battery-compartment cover.
9. Replace the screws and tighten with moderate pressure.
Mains connector
D Serial interface
(printer, PC)
Protective cap
E Screws
A Battery pack C Fuse B Battery Back of unit
F Battery-compartment cover
Dispose of used batteries and/or the lithium powerpack in accordance with the local regulations.
en
Page 7
9
1.2.1 Lithium powerpack –
Important information
The lithium powerpack (B) serves to save all important data (saved measurement results, settings) for approx. ½ year when the unit is not being supplied with power by the battery pack or the mains supply.
As long as the colorimeter is being supplied with mains or battery power, it does not draw any power from the lithium powerpack. Since lithium powerpacks have a very long life, it will in all probability not be necessary to exchange the one supplied. All the same, to ensure proper functioning it is recom mended to replace the lithium powerpack every five years.
Note
When neither the mains adapter nor the battery pack are supplying power, removal of the lithium powerpack will result in a complete loss of data (saved measurement results and settings). For this reason we strongly recommend that the unit is connected to the mains power supply while replacing the lithium powerpack.
1.2.2 Charging the battery pack
The rechargeable batteries (A) remain in place in the photo­ m eter for recharging. Once the mains power supply is established, the batteries start to be charged.
Charge the rechargeable batteries in the instrument for 5 days (working with power from mains is possible). Now use the instrument without mains until the first battery warning comes up. Charge again, 4 days this time. Repeat usage to battery warning and charging four times.
Approximately 10 charging/discharging cycles are required until the batteries reach their full capacity.
The unit can be operated with mains power supply with or without the batteries in place. For this you must attach the country-specific wall-socket adapter to the mains adapter.
en
Page 8
10
1.2.3 Using non-rechargeable batteries
In principle it is also possible to operate the unit using conven­tional batteries – in this case, however, they must never be used in combination with the mains adapter. The charging process would start automatically as soon as the colorimeter is connected to the mains power supply. Conventional batteries are destroyed by the charging current, resulting in severe dam­age to the unit. This also involves the risk of fire and explosion.
1.2.4 Using the protective caps
To protect the connection terminals on the unit from damage (e.g. corrosion) due to environmental influences such as e.g. dust or water splashes, the supplied protective caps should be attached over the connection points (D) when these are not being used.
en
Page 9
en
11
1.3 Overview of the key functions
Switching the unit on and off
Function key: Function explained at the corresponding place in the text
Function key: Function explained at the corresponding place in the text
Function key: Function explained at the corresponding place in the text
Back to method selection / to parent menu
Confirmation of selections (visible on the display as the “ ” symbol)
Menu for settings and other functions
Moves cursor (visible on the display as the “>>” symbol) up or down
Save a displayed result
Zero-calibration function
Run a measurement
Display of date and time / user countdown
On O
F1
F2
F3
tseTedoM
Store
Zero
654
321
987
,0
Spectroquant® Multy
Enter
Esc
Time
Page 10
12
1.5 Overview of the mode menu
Mode No. Mode function Brief description Section
10 Language Setting the language 1.6 11 Key-beep Activating the acoustic key-acknowledgment tone 3.3.1 12 Clock Setting the date and time 1.8 13 Countdown Activating/deactivating the countdown for reaction times 3.3.3 14 Signal-beep Activating/deactivating the acoustic signal at the end of a measurement 3.3.2 20 Print Printing all saved measurement results 5.3.2 21 Print, date Printing measurement results from a defined date range 5.3.3 22 Print, code No. Printing measurement results from a defined code-No. range 5.3.4
1.4 Starting the colorimeter
the first time
Switch on the colorimeter by pressing the [On/Off] button. The unit runs an electronic self-check test.
The display then shows:
Pressing the [Enter] key takes the colorimeter to method selection. Any data already saved in the unit must be deleted (mode 34, see chapter 1.7, "Delete data"), the user-method system must be initialized (mode 69, see section 5.6.5, "Initializing the user-method system (concentration and polynomials)"), and the date and time should be reset. (See below for details.)
The Spectroquant® Multy is supplied with English preset as the standard language setting. Before making the first measurement you should therefore reset the unit to the language of your choice. To do this go from the method list and change to the mode menu by pressing the [Mode] key.
The display shows:
After a short time the selection list appears:
Please initialize the storagesystem with MODE 34
MODE Menu cancel: ESC
>> 10:Language 11:Key-beep 12:Clock
... ...
... ...
en
Page 11
13
Mode No. Mode function Brief description Section
23 Print, method Printing measurement results from a defined method 5.3.5 29 Print params. Setting the printer options 5.3.1
30 Storage Viewing all saved measurement results 2.8.1 31 Stor., date Viewing measurement results from a defined date range 2.8.2 32 Stor., code No. Viewing measurement results from a defined code-No. range 2.8.3 33 Stor., method Viewing measurement results from a defined method 2.8.4 34 Delete data Deleting saved measurement results 2.9
45 User calibration Saving user-specific calibration 5.7.1 46 Clear calibr. Deleting user-specific calibration 5.7.2
50 Profi-Mode Activating the expert-user function (laboratory function) 3.2
60 Method list Processing the method list 3.1.1 61 Mlist all on User-specific method list, activate all methods 3.1.2 62 Mlist all off User-specific method list, deactivate all methods 3.1.3 64 User concentr. User methods, Entry of a concentration method 5.6.1 65 User polynoms User methods, Entry of a user polynomial 5.6.2 66 User m. clear User methods, Deleting a user method 5.6.3 67 User m. print User methods, Printing data for a user method 5.6.4 69 User m. init. User methods, Initializing the user-method system 5.6.5
70 Langelier Calculation of the Langelier saturation index 5.8 71 Temperature Setting the temperature (°C or °F) for Langelier mode 70 5.8
80 LCD contrast Setting the display contrast 3.4
91 System-Info Information on the Spectroquant® Multy, e.g. current unit configuration 3.5
The individual mode functions are selected in the following manner:
Enter the digits for the desired function, e.g.: [1] [0] for setting the language, or
press the [] or [] arrow keys to select the desired function from the display list.
Confirm your selection by pressing [Enter].
Make your settings as described in the respective sections of this manual.
Press the [Esc] key to exit the mode menu.
en
Page 12
14
1.6 Setting the language
Press the keys [Mode] [1] [0] in succession.
Confirm your selection by pressing [Enter].
The display now shows:
Select the desired language using the arrow keys [] or [].
Confirm your selection by pressing [Enter].
(Pressing the [Esc] key takes you back to the method­selection menu.)
1.7 Deleting data
Press the keys [Mode] [3] [4] in succession to delete any stored data.
Confirm by pressing [1] and [Enter].
Press the [0] and [Enter] keys to abort the process.
In the event that you press the [1] key by mistake, you can exit the menu by pressing the [Esc] key if you wish to save the data from deletion.
<Language> Deutsch >> English Français ...
en
Page 13
15
1.8 Setting the date and time
Press the keys [Mode] [1] [2] in succession.
Confirm your selection by pressing [Enter].
The display now shows:
The date and time are entered in the following sequence:
year, month, day, e.g.: July 14, 2006 = [0] [6] [0] [7] [1] [4]
hours, minutes, e.g.: 3.07pm = [1] [5] [0] [7].
Confirm your selection by pressing [Enter].
Note
When you confirm the entry by pressing [Enter] the seconds are automatically set to zero.
Pressing the [Esc] key takes you back to the method­selection mode.
1.9 Time and date display
Press the [Time] key.
The display now shows the current time and date. The unit returns to the previous routine after approx. 15 seconds
or when you press the [Enter] key.
1.10 Automatic switch-off
The Spectroquant® Multy switches off automatically 20 minutes after the last time a key was pressed. In the last 30 seconds before it switches off, the unit emits an acoustic signal. During these 30 seconds you can press a key to prevent the unit from switching off automatically. The automatic switch-off function is inactive while the unit is performing operations (running countdown, printing). After the operation in question has ended, the 20-minute waiting time before the automatic switch-off function starts run­ning anew.
<Clock>
yy-mm-dd hh:mm __-__-__ __:__
en
Page 14
16
Operating mode
2.1 Selecting the method
Switch on the Spectroquant® Multy by pressing the [On/Off] key.
The display shows the selection list of the stored methods:
There are two ways to select the desired method:
a) by entering the method number directly, e.g.: [1] [6] [3] for COD 14541, or
b) by pressing the [] or [] arrow keys to select the desired method from the displayed list.
Confirm your selection by pressing [Enter].
Note
Pressing the [F1] key switches between the compact and the detailed method-selection list. The method-selection list must be shown on the display for this option.
Example for the detailed method-selection list
Line 1: Method number, method name, item number Line 2: Measuring range Line 3: Type of test (cell test or test) Line 4: Cells used (16 mm/24 mm)
>> 10 Acid cap. 01758 20 Aluminium 14825 21 Aluminium 00594
... ...
... ...
163 COD 14541 25-1500 mg/l Cell Test 16 mm
en
Page 15
17
Note
The five-digit item number (e.g. 14541) gives the five digits in the middle of the catalogue/ordering number 1.XXXXX.0001, in this case 1.14541.0001. In some cases in which the assign­ment is self-evident (e.g. monochloramine) or else in which all Spectroquant® tests available for one parameter can be used (e.g. chlorine), this number is not shown.
For an overview of all programmed methods please refer to the included CD, section 5.1, “Overview of preprogrammed methods and analytical procedures”.
2.2 Measuring with test kits
A detailed description of the procedure for the selected method is given on the provided CD in section 5.1, “Overview of prepro­grammed methods and analytical procedures”. The procedures may differ slightly from those described in the respective pack­age inserts.
After selecting the method, prepare the blank and sample for measurement.
In the case of analysis specifications in which reaction times must be observed, a timer (countdown) is integrated in the method programme. (In such cases the cells must not be inserted into the measurement compartment.)
After the method has been selected the display shows: Example: Method 90, Bromine 00605
If you wish to exit the menu at this stage, simply press the [Enter] key twice (= abort countdown) and then the [Esc] key once.
After the method has been selected, the countdown function is started by pressing the [Enter] key. The remaining time is shown continuously. An acoustic signal is emitted in the last 10 seconds before the countdown expires. After the countdown has run out, proceed as described in the analysis procedures.
In some methods there are several reaction times that have to be considered; these are shown and processed in the proper sequence.
en
90 Bromine 00605
0.10-5.00 mg/l Br2 Count-down 1 1:00 Start:
Page 16
18
Note
The running countdown can be skipped by pressing the [Enter] key once. In this case the user must observe the necessary reaction time him-/herself. (Failure to observe the specified reaction time can lead to erroneous results.) Further options to deactivate the countdown procedure: mode No. 13 or Profi mode (mode No. 50).
After the countdown has expired the display shows:
Place the prepared blank in the measurement compartment with the mark on the cell pointing towards the mark on the unit case.
Positioning the cell (ø 24 mm)
Align the triangular mark on the cell with that on the Spectroquant® Multy: To afford better protection against sunlight, press the o-ring firmly into place.
90 Bromine 00605
0.10-5.00 mg/l Br2
prepare Zero press ZERO
en
Page 17
19
en
Positioning the cell (ø 16 mm)
Attach the 16-mm cell adapter. Align the line mark above the item number of the cell with the triangular mark on the Spectroquant® Multy. Subsequently attach the lid to cover the adapter.
Press the [Zero] key.
The display shows:
Insert the prepared sample into the measurement compartment with the cell mark aligned with the mark on the unit case.
Press the [Test] key.
The result is displayed in the following manner: Example: Method 90 (Bromine 00605)
Line 1: Method number, method name, item number Line 2: Measuring range Line 3: Result (expressed as the concentration)
90 Bromine 00605
0.10-5.00 mg/l Br2 Zero accepted prepare Test press TEST
90 Bromine 00605
0.10 - 5.00mg/l Br2
2.11mg/l Br2
Page 18
90 Bromine 00605
0.10-5.00 mg/l Br2 Zero accepted prepare Test press TEST
20
In the event the result lies outside the respective measuring range, the following message appears on the display:
the concentration of the sample lies below the measuring range
or, respectively,
the concentration of the sample lies above the measuring range.
After the result has been displayed
• it can be saved (see section 2.7, “Saving measurement results” and section 2.8, “Retrieving saved measurement results” for further details)
• it can be printed out (see section 5.3)
• a new method can be selected: pressing the [Esc] key takes the colorimeter back to the method-selection mode; you can also enter a new method number directly (confirm by pressing [Enter])
• further measurements can be made using the same or a new zero setting:
• If you wish to measure other samples using the same method:
Press the [Test] key.
The display shows:
Confirm by pressing [Test].
90 Bromine 00605
0.10 - 5.00mg/l Br2
Underrange Br2
90 Bromine 00605
0.10 - 5.00mg/l Br2
Overrange Br2
en
Page 19
21
en
• If you wish to measure other samples with a new zero setting:
Press the [Zero] key to reset to zero.
The display shows:
2.3 Differentiation
Some methods permit further differentiation (e.g. chlorine). After selecting the method, e.g. 131 Chlorine Test, you are prompted to state the type of measurement (e.g. differentiated, free, or total).
Use the [] or [] arrow keys to select the desired measurement type.
Confirm your selection by pressing [Enter].
90 Bromine 00605
0.10-5.00 mg/l Br2 Count-down 1 1:00 Start:
Chlorine Test >> diff free total
Page 20
22
2.4 Altering the citation form
Wherever this is appropriate, it is possible to alter the citation form (see section 5.1, “Overview of preprogrammed methods and analytical procedures” for possible citation-form alterna­tives).
After the first sample has been measured using a specific method and the result is shown on the display, you can alter the citation form in the following manner:
Result shown on display using Method 380 (Phosphate 14543) as an example:
Pressing the [] arrow key gives you the option to select a citation form. The result shown on the display changes to this:
Pressing the [] arrow key again shows the next citation form:
Pressing the [] arrow key takes you back to the previous citation form.
The citation form last shown on the display remains valid for all consequent measurements until changed again.
380 Phosphate 14543
0.05-4.00mg/l PO4-P
0.33mg/l PO4-P
380 Phosphate 14543
0.15-12.26mg/l PO4
1.01mg/l PO4
380 Phosphate 14543
0.11-9.17mg/l P205
0.76mg/l P205
en
Page 21
23
en
2.5 Measuring absorbances
Besides measuring concentrations using a selected method, the unit is also capable of measuring absorbances. For this you call up the desired wavelength by entering the corresponding method number or by choosing from the method-selection list.
Measuring range: -2600 mAbs to +2600 mAbs
Method No. Designation
600 A 430 nm 610 A 530 nm 620 A 560 nm 630 A 580 nm 640 A 610 nm 650 A 660 nm
The display shows e.g.:
Always zero the colorimeter using a filled cell (e.g. with DI water). The display shows e.g.:
Then measure the sample.
The display shows e.g.:
100 mAbs = 0.100 A (absorbance units)
Tip Reaction times for your own measurements in the absorbance mode can be more easily observed by using the user-countdown function (see the following section 2.6, “User countdown”).
600 A 430 nm
-2600 - +2600 mAbs
prepare Zero press ZERO
600 A 430 nm
-2600 - +2600 mAbs Zero accepted prepare Test press TEST
600 A 430 nm
-2600 - +2600 mAbs
100 mAbs
Page 22
24
2.6 User countdown
(timer function)
This function enables the user to employ a self-defined countdown time.
Press the [Time] key.
The display shows the current time and date.
Press the [Time] key again.
The display shows:
Enter the time in double digits, in the sequence minutes and seconds, e.g.: 2 minutes, 0 seconds = [0][2][0][0].
Confirm your selection by pressing [Enter].
The display shows:
Press the [Enter] key to start the countdown.
After the countdown has expired, the unit returns to the previous routine.
Note
The user-countdown function is available even when the preset countdown function is deactivated.
15.58.40 2004-10-21
count-down mm:ss
count-down 2:00
en
Page 23
25
en
2.7 Saving measurement results
Press the [Store] key while the result is shown on the display.
The display shows (Example: Method 31, Ammonium 14558):
The user is able to enter a six-digit code at this stage. (The code No. can be used to show e.g. information regarding the user or the sampling site.)
Confirm the code No. by pressing [Enter].
If you do not wish to enter a code No., simply confirm by pressing [Enter]. (This results in the automatic assigment of a code No. starting with 0.)
The entire data set is then stored together with the date, time, code No., method, and result.
The display shows:
Note
The number of available memory records is also shown on the display. Subsequently the measurement result is shown again.
When fewer than 30 free memory records are available, the display shows:
It is advisable to delete the data memory as soon as possible when no longer required (see section 2.9, “Deleting saved mea­surement results”). When all memory records are occupied, it is not possible to save any further results.
31 Ammonium 14588
0.20-8.00 mg/l NH4-N
Code-No.: _ _ _ _ _ _
31 Ammonium 14588
0.20-8.00 mg/l NH4-N Stored! storage: 997 free records left
storage: only 29 free records left
Page 24
26
2.8 Retrieving saved
measurement results
2.8.1 Retrieving all saved measurement results
Press the keys [Mode] [3] [0] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Confirm by pressing [Enter]. The data sets are then shown in reverse chronological sequence, starting with the most recently saved measurement result.
Pressing the [] key takes you to the next data set.
Pressing the [] key takes you back to the previous data set.
Pressing the [F3] key prints out the result shown on the display.
Exit by pressing the [Esc] key.
If there are no data saved in the memory, the display shows:
<storage> display all data Start: cancel: ESC print: F3 print all: F2
<storage> display all data
no data
en
Page 25
27
en
2.8.2 Retrieving saved measurement results
from a defined date range
Press the keys [Mode] [3] [1] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Enter the starting date in the sequence year, month, day:
e.g.: July 14, 2006 = [0] [6] [0] [7] [1] [4].
Confirm by pressing [Enter].
The display shows:
Enter the end date in the sequence year, month, day:
e.g.: July 19, 2006 = [0] [6] [0] [7] [1] [9].
Confirm by pressing [Enter].
The display shows:
<storage> sorted: date from yy-mm-dd __-__-__
<storage> sorted: date to yy-mm-dd __-__-__
<storage> sorted: date from 2006-07-14 to 2006-07-19 Start: cancel:ESC print:F3 print all:F2
Page 26
28
Pressing the [Enter] key shows the saved test results for the defined period of time.
Pressing the [F3] key prints out the result shown on the display.
Pressing the [F2] key prints out all selected results.
Exit by pressing the [Esc] key.
Note
To show test results obtained on just one day, enter the same date for both start and end date.
2.8.3 Retrieving saved measurement results
from a defined code-No. range
Press the keys [Mode] [3] [2] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Enter the start code No. (max. 6 digits), e.g.: [1].
Confirm by pressing [Enter].
The display shows:
<storage> sorted: Code-No. from _ _ _ _ _ _
<storage> sorted: Code-No. from 1 _ _ _ _ _ to _ _ _ _ _ _
en
Page 27
29
en
Enter the end code No. (max. 6 digits), e.g.: [1] [0].
Confirm by pressing [Enter].
The display shows:
Pressing the [Enter] key shows all saved test results for the selected code-No. range.
Pressing the [F3] key prints out the result shown on the display.
Pressing the [F2] key prints out all selected results.
Exit by pressing the [Esc] key.
Note
To show test results with one and the same code No., enter the same number for both the start and the end code No. To show all test results without the code No. (code No. = 0), enter a zero [0] for both the start and the end code No.
<storage> sorted: Code-No. from 000001 to 000010 Start: cancel:ESC print:F3 print all:F2
Page 28
30
2.8.4 Retrieving saved measurement results
from a defined method
Press the keys [Mode] [3] [3] in succession.
Confirm your selection by pressing [Enter].
The display shows e.g.:
Select the desired method from the list or otherwise enter the method number directly, e.g. 21 (aluminium 00594).
Confirm by pressing [Enter].
In the case of differentiated methods make the corresponding new selection and confirm by pressing the [Enter] key.
The display shows:
Pressing the [Enter] key shows all saved test results for the selected method.
Pressing the [F3] key prints out the result shown on the display.
Pressing the [F2] key prints out all selected results.
Exit by pressing the [Esc] key.
<storage> >> 10 Acid cap. 01758 20 Aluminium 14825 21 Aluminium 00594 ...
<storage> method 21 Aluminium 00594 Start: cancel:ESC print:F3 print all:F2
en
Page 29
31
en
2.9 Deleting saved
measurement results
Press the keys [Mode] [3] [4] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Pressing the [0] key saves the data for further use.
Pressing the [1] key deletes the data from the memory records.
The display shows:
or, respectively,
Confirm by pressing [Enter].
In the event that you press the [1] key by mistake, you can exit the menu by pressing the [Esc] key if you wish to save the data from deletion.
Note
All saved measurement results are deleted by this operation, irrespective of the method.
<Delete data> Delete all Data?
YES:1 NO:0
<Delete data> Delete all Data?
YES:1 NO:0 do not delete
<Delete data> Delete all Data?
YES:1 NO:0 Delete Data
Page 30
32
Other functions
3.1 User-specific method list
In the unit's delivery configuration the method-selection list displays all available methods. Additionally the user can configure the method-selection list to suit his/her specific requirements.
After an update all new methods are automatically added to the user list.
For software-technical reasons at least one method must be activated in the user-specific method list. The unit automatically activates the first method stored in the sorting list. For this reason another method must be activated before the automatically activated method can be deactivated.
3.1.1 Processing the user-specific method list
Press the keys [Mode] [6] [0] in succession.
Confirm your selection by pressing [Enter].
The display shows:
<methods list> selected: • toggle: F2 save: cancel: ESC
en
Page 31
33
en
Start by pressing the [Enter] key.
The complete method list appears on the display.
Methods showing a dot (•) following the method number appear in the method-selection list, while methods without the dot are not shown.
Press the [] or [] keys to position the cursor next to the method to be processed.
Use the [F2] key to switch between “activated” and “deactivated”. Deactivated methods are then shown without the dot.
Select the next method and follow the above procedure to adjust the list to match your requirements until all methods show the desired settings.
Confirm your selection for saving by pressing [Enter].
Pressing the [Esc] key
and then the [Enter] key enables you to exit this mode at any time without adopting the alterations.
Tip
In the event that you wish for only a few methods to be shown in the method-selection list, it is advisable to first execute mode 62 “Mlist all off” (deactivate all methods) and then to process the method-selection list using mode 60 “Method list”. All you then need do is select the methods that you would like to include in the method-selection list for later use by marking them with the dot (•). The names of the user polynomials (1-25) and user concentra­tions (1-10) all appear in the method list, even when they are not programmed. Unprogrammed methods cannot be activated!
<methods list> >>10•Acid cap. 01758 20•Aluminium 14825 21•Aluminium 00594 ...
<methods list> >>10•Acid cap. 01758 20 Aluminium 14825 21•Aluminium 00594 ...
Page 32
34
3.1.2 User-specific method list:
Activate all methods
This mode function activates all methods and the complete method-selection list appears when the unit is switched on.
Press the keys [Mode] [6] [1] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Pressing the [1] key shows all methods in the method­selection list.
Pressing the [0] key saves the current method-selection list for later use.
The unit then returns to the mode menu.
3.1.3 User-specific method list:
Deactivate all methods
For software-technical reasons at least one method must be activated in the user-specific method list. The unit auto­matically activates the first method stored in the sorting list.
Press the keys [Mode] [6] [2] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Pressing the [1] key shows just one method in the method list.
Pressing the [0] key saves the current method list for later use.
The unit then returns to the mode menu.
<Mlist all on> switch on all methods YES:1 NO:0
<Mlist all off> switch off all methods YES:1 NO:0
en
Page 33
35
en
3.2 Profi mode
As a rule the methods include the following information:
a) Method b) Measuring range c) Date and time d) Differentiation of measurement results e) Observance of reaction times (countdown)
When the Profi mode is activated, the colorimeter restricts itself to a minimum of user guidance. Items d and e are omitted.
Press the keys [Mode] [5] [0] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Pressing the [0] key deactivates the Profi mode.
Pressing the [1] key activates the Profi mode.
The display shows:
or
Confirm by pressing [Enter].
Note
It is also possible to save results in the Profi mode. When results are saved here, the display also shows the message: “Profi mode”. This selected setting remains activated even when the unit is switched off until a new setting is made.
<Profi-Mode> actual: switched off ON:1 OFF:0
<Profi-Mode> actual: switched off ON:1 OFF:0 switched on
<Profi-Mode> actual: switched off ON:1 OFF:0 switched off
Page 34
36
3.3 Acoustic signals
3.3.1 Activating/deactivating the key beep
Press the keys [Mode] [1] [1] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Pressing the [0] key deactivates the key beep function.
Press the [1] key activates the key beep function.
Confirm by pressing [Enter].
Note
In connection with methods that include a reaction time, the unit emits an acoustic signal in the last 10 seconds before the countdown expires even when the key beep function is inactive.
3.3.2 Activating/deactivating the signal beep
It takes the colorimeter approx. 8 seconds to perform a zero calibration and measurements. It emits a brief signal beep at the end of a measurement.
Press the keys [Mode] [1] [4] in succession.
Confirm your selection by pressing [Enter].
The display shows:
<Key-Beep>
ON:1 OFF:0
<signal-beep>
ON:1 OFF:0
en
Page 35
37
en
Pressing the [0] key deactivates the signal beep function.
Press the [1] key activates the signal beep function.
Confirm by pressing [Enter].
Note
In connection with methods that include a reaction time, the unit emits an acoustic signal in the last 10 seconds before the countdown expires even when the signal beep function is inactive.
3.3.3 Activating/deactivating the countdown
function (observance of reaction times)
Certain methods require the observance of reaction times. These waiting times are stored as standard settings in the respective methods in the form of a timer (countdown) function. The countdown can be deactivated for all methods involved in the following manner:
Press the keys [Mode] [1] [3] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Pressing the [0] key deactivates the countdown function.
Pressing the [1] key activates the countdown function.
Confirm by pressing [Enter].
Note
When the countdown function is inactive, the necessary reaction time must be observed by the user him-/herself. Failure to observe the specified reaction time can lead to erroneous results.
<countdown> actual: switched on ON:1 OFF:0
Page 36
38
3.4 Setting the display contrast
Press the keys [Mode] [8] [0] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Pressing the [] key enhances the contrast of the LC display.
Pressing the [] key reduces the contrast of the LC display.
Confirm by pressing [Enter].
3.5 System info
Press the keys [Mode] [9] [1] in succession.
Confirm your selection by pressing [Enter].
This mode provides details on the current software version, the number of measurements that have already been made, and the number of free memory records.
Pressing the [] key takes you on to further information.
Press the [Esc] key to return to the mode menu.
< LCD contrast>
“LCD contrast”
<System-Info> Software: VO12.004.3.003.002 mains power: yes more:cancel: ESC
<System-Info> Number of Tests: 6 free records left: 999 cancel: ESC
en
Page 37
39
4.1 User messages on the display
/ Error messages
Display message: ADU Err 01: xx (xx = 01, 02, 10, or 20)
Measure: Initialize LED parameters
Please carry out the following steps to set and save the LED brightness parameters: Test the software version of the unit by pressing the keys [Mode] [9] [1]. The version should be V012.00x.3.003.zzz, with zzz = 003 or higher. If you have a lower software version you should per­form a software update (see chapter 5.5, "Software update via Internet"). To do this download the current version for the Spectroquant® Multy from our website (www.service-test-kits.com).
The cell compartment must be clean and dry.
Fill one cell with distilled water and place it in the cell com­partment. It is irrelevant whether you use a 16-mm or 24-mm round cell.
Take care that the unit is NOT exposed to direct sunlight or a bright spotlight.
Keep the [Mode] key pressed, switch off the unit
and switch it back on while pressing the [Mode] key.
Trouble-shooting
en
Page 38
40
The display shows:
Enter the password: [1] [7] [0] [9] [1] [9] [9] [1].
Confirm by pressing [Enter].
The display shows:
Press the [1] key.
The display then shows:
Press the [0] key. The LED parameters are now determined. The unit displays the wavelength in operation, and shortly thereafter "LED-Opt. o.k." if the determined parameter is correct.
When all the LEDs have been processed the display shows:
Confirm by pressing [Enter].
The display then shows:
Press the [Esc] key.
The unit continues to run normally. Now switch off by pressing the [On/Off] key. Thereafter the unit can be used as usual.
<Service Menu 01> Password:
<Service Menu 01> 1: LED-Optim. cancel: ESC
<LED-Optim.> 0: alle LEDs 1: LED wählen cancel: ESC
<LED-Optim.> LED Opt. o.k. weiter mit
<Service Menu 01> 1: LED-Optim. cancel: ESC
en
Page 39
41
en
Display message Possible causes Measures
Battery capacity
Full capacity
Warning signal every 3 minutes The battery capacity will be out soon. Recharge battery pack; Warning signal every 12 seconds operate the unit with mains adapter
Warning signal, the colorimeter switches itself off
E40 If the test result appears with Overrange/Underrange, Use a test with a standard User cal. a user calibration is not possible of lower/higher concentration here not possible Check sources of error, e.g.: user error (correct procedure, observance of the reaction time, ...) Jus Overrange E4, When the user makes calibrations, Check error sources, e.g.: Jus Underrange E4 the setting of the specified value user error (correct procedure, is possible only within defined limits. observance of reaction time,...) These were exceeded or, respectively, not reached. standard (sample weight, dilution, age, pH,...) Repeat adjustment Overrange Measuring range exceeded Where possible dilute sample or select another measuring range
Turbidities in the sample Heed possible interferences
Light entering measurement compartment Seal ring attached to cell lid? Repeat measurement with seal ring attached Overrange During the user calibration the upper measuring-range limit Carry out test with a standard E1 was exceeded while setting to the specified value of lower concentration Underrange Result below measuring range State result with lower than x mg/l x = Lower limit of measuring range; if necessary use a different analytical method Underrange E1 During the user calibration the lower measuring-range limit Carry out test with a standard was not met while setting to the specified value. of a higher concentration Zero Too much, too little incident light Zero cell in place? not accepted Insert zero cell, repeat measurement Clean measurement compartment Repeat zero calibration Printer Timeout Printer inactive, no connection Connect printer, check contacts, switch on printer Storage-system Power supply for storage Insert or replace the lithium error system interrupted or not available. powerpack. Then execute mode 34 Use Mode 34 to delete the data ADU Err 02: xx Light entering measurement compartment Seal ring attached to cell lid? (xx = 01, 02, 10, Repeat measurement with seal or 20) ring attached
Page 40
42
Display message Possible causes Measures
Err 51 LED used is too bright Perform a software update Initialize LED parameters see chapter 4.1 Err 53 Light path in measurement compartment is completely blocked Clean measurement compartment Optics unit is mechanically defe Initialize LED parameters Electronic defect on mainboard see chapter 4.1 It is not possible to calculate a value Correctly measured? (e.g.: bound chlorine). If not, repeat
Example 1 Example: 1 While the values displayed differ in terms of magnitude, in consideration of the tolerances they are identical. In this case there is no bound chlorine present in the sample.
Example 2 Example: 2 The result for free chlorine lies outside the measuring range, which is why the value for free chlorine cannot be calculated. Since no detectable free chlorine is present, the proportion of bound chlorine can be assumed to be the total chlorine content.
Example 3 Example: 3 The result for total chlorine lies outside the measuring range, which is why the unit is not able to calculate the value for bound chlorine. In this case the sample must be diluted to obtain the total chlorine content.
4.2 Other potential sources of error
Problem Possible cause Measure
The colorimeter runs off mains power, Fuse (type A, delayed action, 20 mm) Replace fuse but not with the otherwise fully is defective functionable battery pack.
...
??? ...
130 Chlorine CT
0.05-5.00 mg/l Cl2
0,60 mg/l free Cl
??? comb. Cl
0,59 mg/l total Cl
130 Chlorine CT
0.05-5.00 mg/l Cl2
Underrange free Cl
??? comb. Cl
0,59 mg/l total Cl
130 Chlorine CT
0.05-5.00 mg/l Cl2
0,60 mg/l free Cl
??? comb. Cl
Overrange total Cl
en
Page 41
43
en
4.3 Avoiding errors in
photometric measurements
1. The cells and lid must be cleaned thoroughly after each analy­ sis run to prevent errors due to cross-contamination. Even the smallest residues of reagents will lead to erroneous results.
2. The outer walls of the cells must be clean and dry before the analysis is carried out. Fingerprints or water droplets on the light-path surfaces of the cells will lead to erroneous results.
3. The cells for the zero calibration and the test itself must always be inserted into the measurement compartment in such a way that the white triangle or, respectively, the line of the graduation is correctly aligned with the corresponding mark on the case (see page 16 or 17).
4. The zero calibration and the test itself must both be made with the cell lid in place. The cell lid of the 24-mm cell must be fitted with a seal ring. The 16-mm cell must be covered with the lid for the adapter.
5. The formation of air bubbles on the inner walls of the cell will lead to erroneous results. In this case attach the cell lid to the cell and swirl the cell to eliminate any air bubbles before carrying out the test.
6. Care must be taken to prevent any water from entering the measurement compartment. Any entry of water into the case of the colorimeter may result in the destruction of electronic components and in damage due to corrosion.
7. Any contamination of the optical components in the measure­ ment compartment will lead to erroneous results. The light-path surfaces of the measurement compartment must be checked at regular intervals and cleaned wherever necessary. Use moist wipes and cotton-wool buds for these cleaning operations.
8. Major differences in temperature between the colorimeter and the local environment can lead to erroneous results, e.g. due to condensation on the optical components and on the cell.
9. When operating the colorimeter make sure that it is protected from direct sunlight.
Page 42
Important information
m CAUTION m
The colorimeter was developed for use in the laboratory for water analysis.
The accuracy of the instrument is only valid if the instrument is used in an
environment with controlled electromagnetic disturbances according to DIN 61326.
Wireless devices, e. g. wireless phones, must not be used near the instrument.
Important disposal instructions for batteries and accumulators
EC Guideline 2006/66/EG requires users to return all used and worn-out batteries and accumulators.
They must not be disposed of in normal domestic waste. Because our
products include batteries and accumulators in the delivery package our advice is as
follows:
Used batteries and accumulators are not items of domestic waste. They must be disposed
of in a proper manner. Your local authority may have a disposal facility; alternatively
you can hand them in at any shop selling batteries and accumulators.
You can also return them to the company which supplied them to you; the company is obliged
to accept them.
Important information
To Preserve, Protect and Improve the Quality of the Environment
Disposal of Electrical Equipment in the European Union
Because of the European Directive 2002/96/EC your electrical instrument must not be
disposed of with normal household waste!
For more information please visit the following website:
www.millipore.com/company/flx4/eu_regulatory_compliance
44
en
Page 43
45
en
Declaration of CE-Conformity
Declaration of EC-Conformity according to
DIRECTIVE 2004/108/EG OF THE EUROPEAN PARLIAMENT
AND OF THE COUNCIL of 2004, December the 15
th
Name of manufacturer: Merck KGaA
64271 Darmstadt Germany
declares that this product
Product name: Spectroquant® Multy
meets the requirements of the following product family standard:
DIN EN 61326-1:2006
Immunity test requirements for equipment intended for use in industrial locations (Table 1)
Emission according to the requirements for class B equipment
Darmstadt, 10th January 2013
Merck KGaA
i. V. i. A.
B. Grau C. Klein Director MM WFA Product Manager Photometry
Page 44
Table of contents
5 Contents of CD-ROM
5.1 Overview of preprogrammed methods and analytical procedures
5.2 Standard solutions
5.2.1 Use of Spectroquant® CombiCheck and ready-to-use
standard solutions
5.2.2 Preparation of standard solutions
5.3 Printing measurement results
5.3.1 Setting the print parameters
5.3.2 Printing all measurement results
5.3.3 Printing measurement results from a defined date range
5.3.4 Printing measurement results from a defined code-No. range
5.3.5 Printing measurement results from a defined method
5.4 Transferring data to a PC
5.5 Software update via the internet
5.6 User methods
5.6.1 User concentration method
5.6.2 User polynomials
5.6.3 Deleting a user method
5.6.4 Printing data for a user method
5.6.5 Initializing the user-method system
5.7 User-specific calibration
5.7.1 Saving the user-specific calibration
5.7.2 Deleting the user-specific calibration
5.8 Calculating the Langelier saturation index
5.9 Technical specifications
Page 45
CD-ROM
5.1 Overview of preprogrammed methods and analytical procedures
Meth. No. Parameter Cat. No. Measuring range Blank Type of Type of test cell
10 Acid cap. 01758 1.01758.0001 0.40 - 8.00 mmol/l OH RB Cell test 16 mm 20 Aluminium 14825 1.14825.0001* 20 - 700 μg/l Al RB Cell test 16 mm 21 Aluminium 00594 1.00594.0001 0.05 - 0.50 mg/l Al RB Test 24 mm 30 Ammonium 14739 1.14739.0001 10 - 2000 μg/l NH
4
-N RB Cell test 16 mm
31 Ammonium 14558 1.14558.0001 0.20 - 8.00 mg/l NH
4
-N RB Cell test 16 mm
32 Ammonium 14559 1.14559.0001 4.0 - 80.0 mg/l NH
4
-N RB Cell test 16 mm
33 Ammonium 14752 1.14752.0001* 0.02 - 1.30 mg/l NH
4
-N RB Test 24 mm
34 Ammonium 00683 1.00683.0001 1.0 - 50.0 mg/l NH
4
-N RB Test 16 mm 40 AOX 00675 1.00675.0001 0.05 - 2.50 mg/l AOX RB Cell test 16 mm 50 Arsenic 01747 1.01747.0001 5 - 100 μg/l As RB Test 16 mm 70 BOD 00687 1.00687.0001 0.5 - 3000 mg/l BOD H
2
O Cell test 16 mm 80 Boron 00826 1.00826.0001 0.05 - 2.00 mg/l B RB Cell test 16 mm 90 Bromine 00605 1.00605.0001 0.10 - 5.00 mg/l Br
2
H
2
O Test 24 mm 100 Cadmium 14834 1.14834.0001 25 - 1000 μg/l Cd RB Cell test 16 mm 101 Cadmium 01745 1.01745.0001 5 - 500 μg/l Cd RB Test 24 mm 110 Calcium 00858 1.00858.0001 10 - 250 mg/l Ca RB Cell test 16 mm 111 Calcium 14815 1.14815.0001 5 - 160 mg/l Ca RB Test 16 mm 120 Chloride 14730 1.14730.0001 5 - 125 mg/l Cl RB Cell test 16 mm 121 Chloride 14897 1.14897.0001 10 - 250 mg/l Cl RB Test 16 mm 122 Chloride 01804 1.01804.0001 0.5 - 15.0 mg/l Cl RB Cell Test 16 mm 123 Chloride 01807 1.01807.0001 0.50 - 5.00 mg/l Cl RB Test 24 mm 130 Cl2 CT 0.05-5.00 1.00595.0001
(free) 0.05 - 5.00 mg/l Cl
2
H
2
O Cell test 16 mm
1.00597.0001
(free + total)
131 Cl2 0.02-4.50 1.00598.0002 (free) 0.02 - 4.50 mg/l Cl2 H
2
O Test 24 mm
1.00598.0001
(free)
1.00602.0001 (total)
1.00602.0002 (total)
1.00599.0001 (free + total)
* in contrast to the instructions given in the package insert the sample volume as well as the reagent volume must be doubled RB = own reagent blank value
Page 46
Meth. No. Parameter Cat. No. Measuring range Blank Type of Type of test cell
134 Cl2 0.10-6.00 1.00598.0002
(free) 0.10 - 6.00 mg/l Cl
2
H
2
O Test 16 mm
1.00598.0001
(free)
1.00602.0001 (total)
1.00602.0002 (total)
1.00599.0001 (free + total) 132 Cl2 LR 0.10-6.00 1.00086.0001 + 1.00087.0001
(free)
1.00086.0001 + 1.00087.0001 + 1.00088.0001 (total)
0.10 - 6.00 mg/l Cl
2
H
2
O Cell test 16 mm 133 Cl2 LR 0.02-4.50 1.00086.0001 + 1.00087.0001
(free)
1.00086.0001 + 1.00087.0001 + 1.00088.0001 (total)
0.02 - 4.50 mg/l Cl
2
H
2
O Test 24 mm 140 ClO2 0.05-8.50 1.00608.0001 0.05 - 8.50 mg/l ClO
2
H
2
O Test 24 mm 141 ClO2 0.20-10.00 1.00608.0001 0.20 - 10.00mg/l ClO
2
H
2
O Test 16 mm 150 Chromate 14552 1.14552.0001 0.05 - 2.00 mg/l Cr H
2
O Cell test 16 mm 151 Chromate 14758 1.14758.0001* 10 - 1400 μg/l Cr H
2
O Test 24 mm 168 COD 01796 1.01796.0001 5.0 - 80.0 mg/l COD RB Cell test 16 mm 160 COD 14540 1.14540.0001 10 - 150 mg/l COD RB Cell test 16 mm 161 COD 14895 1.14895.0001 15 - 300 mg/l COD RB Cell test 16 mm 162 COD 14690 1.14690.0001 50 - 500 mg/l COD RB Cell test 16 mm 163 COD 14541 1.14541.0001 25 - 1500 mg/l COD RB Cell test 16 mm 164 COD 14691 1.14691.0001 300 - 3500 mg/l COD RB Cell test 16 mm 165 COD 14555 1.14555.0001 0.50 -10.00 g/l COD RB Cell test 16 mm 169 COD 01797 1.01797.0001 5.00 - 90.00 g/l COD RB Cell test 16 mm 166 COD 09772 1.09772.0001 10 - 150 mg/l COD RB Cell test 16 mm 167 COD 09773 1.09773.0001 100 - 1500 mg/l COD RB Cell test 16 mm 570 COD 17058 1.17058.0001 5.0 - 60.0 mg/l COD RB Cell test 16 mm 571 COD 17059 1.17059.0001 50 - 3000 mg/l COD RB Cell test 16 mm 170 Color - 0 - 1000 mg/l Pt/Co (Hazen) H
2
O Method 24 mm 180 Copper 14553 1.14553.0001 0.05 - 8.00 mg/l Cu H
2
O Cell test 16 mm 181 Copper 14767 1.14767.0001 0.10 - 6.00 mg/l Cu H
2
O Test 16 mm 192 Cyanide 02531 1.02531.0001 10 - 350 μg/l CN H
2
O Cell test 16 mm 190 Cyanide 14561 1.14561.0001 10 - 350 μg/l CN H
2
O Cell test 16 mm 191 Cyanide 09701 1.09701.0001* 5 - 200 μg/l CN H
2
O Test 24 mm 200 Cyan. acid19250 119250.0001 2 - 160 mg/l Cys SB Test 24 mm 201 Cyan. acid19253 119253.0001 2 - 160 mg/l CyA SB Test 24 mm 220 Fluoride 14557 1.14557.0001 0.10 - 1.50 mg/l F RB Cell test 16 mm 222 Fluoride 00809 1.00809.0001 0.10 - 1.80 mg/l F RB Cell test 16 mm 221 Fluoride 14598 1.14598.0001 0.10 - 2.00 mg/l F RB Test 16 mm 223 Fluoride 00822 1.00822.0001 0.08 - 2.00 mg/l F RB Test 24 mm 230 Hydrazine 09711 1.09711.0001* 10 - 1200 μg/l N
2H4
RB Test 24 mm
560 HydroPerox 18789 1.18789.0001 0.02 - 5.50 mg/l H
2O2
RB Test 16 mm
240 Iodine 00606 1.00606.0001 0.10 - 5.00 mg/l I
2
H
2
O Test 24 mm 250 Iron 14549 1.14549.0001 0.05 - 4.00 mg/l Fe H
2
O Cell test 16 mm 251 Iron 14761 1.14761.0001* 0.01 - 2.00 mg/l Fe H
2
O Test 24 mm
1.14761.0002* 252 Iron 00796 1.00796.0001 0.10 - 5.00 mg/l Fe H
2
O Test 16 mm 260 Lead 14833 1.14833.0001 0.10 - 5.00 mg/l Pb RB Cell test 16 mm 261 Lead 09717 1.09717.0001 0.05 - 5.00 mg/l Pb RB Test 24 mm 270 Magnesium 00815 1.00815.0001 5.0 - 75.0 mg/l Mg RB Cell test 16 mm 280 Manganese 00816 1.00816.0001 0.10 -5.00 mg/l Mn H
2
O Cell test 16 mm 281 Manganese 01739 1.01739.0001 0.05 - 1.80 mg/l Mn RB Test 24 mm 282 Manganese 14770 1.14770.0001* 0.05 - 6.00 mg/l Mn H
2
O Test 24 mm 283 Manganese 01846 1.01846.0001 0.05 - 1.80 mg/l Mn RB Test 24 mm 290 Molybdenum 00860 1.00860.0001 0.02 - 1.00 mg/l Mo H
2
O Cell test 16 mm 291 Molybdenum 19252 119252.0001 0.5 - 45.0 mg/l Mo H
2
O Test 24 mm 300 Monochloramine 1.01632.0001 0.10 - 5.00 mg/l Cl
2
H
2
O Test 24 mm 310 Nickel 14554 1.14554.0001 0.10 - 6.00 mg/l Ni RB Cell test 16 mm 311 Nickel 14785 1.14785.0001* 0.05 - 5.00 mg/l Ni RB Test 24 mm
* in contrast to the instructions given in the package insert the sample volume as well as the reagent volume must be doubled
RB = own reagent blank value SB = sample blank value
Page 47
Meth. No. Parameter Cat. No. Measuring range Blank Type of Type of test cell
320 Nitrate 14542 1.14542.0001 0.5 - 15.0 mg/l NO
3
-N RB Cell test 16 mm
321 Nitrate 14773 1.14773.0001 0.5 - 15.0 mg/l NO
3
-N RB Test 16 mm
322 Nitrate 14556 1.14556.0001 0.10 - 3.00 mg/l NO
3
-N H2O Cell test 16 mm
323 Nitrate 01842 1.01842.0001 0.3 - 30.0 mg/l NO
3
-N RB Test 24 mm
330 Nitrite 14547 1.14547.0001 10 - 700 μg/l NO
2
-N H2O Cell test 16 mm
331 Nitrite 14776 1.14776.0001* 5 - 400 μg/l NO
2
-N H2O Test 24 mm
1.14776.0002* 332 Nitrite 00609 1.00609.0001 1.0 - 90.0 mg/l NO
2
-N H2O Cell test 16 mm 340 Nitrogen 14537 1.14537.0001 0.5 - 15.0 mg/l N RB Cell test 16 mm 550 Oxygen 14694 1.14694.0001 0.5 - 12.0 mg/l O
2
H2O Cell test 16 mm 555 Oxyg. scavengers 119251.0001 20 - 500 μg/l DEHA RB Test 24 mm 350 Ozone 0.02-3.00 1.00607.0001 0.02 - 3.00 mg/l O
3
H
2
O Test 24 mm
1.00607.0002 351 Ozone 0.10-4.00 1.00607.0001 0.10 - 4.00 mg/l O
3
H
2
O Test 16 mm
1.00607.0002 360 pH 01744 1.01744.0001 6.4 - 8.8 H
2
O Cell test 16 mm
370 Phenol 14551 1.14551.0001 0.10 - 2.50 mg/l C
6H5
OH RB Cell test 16 mm
371 Phenol 00856 1.00856.0001 0.10 - 5.00 mg/l C
6H5
OH RB Test 24 mm
387 Phosphate 00474 1.00474.0001 0.05 - 4.00 mg/l PO
4
-P H2O Cell test 16 mm
380 Phosphate 14543 1.14543.0001 0.05 - 4.00 mg/l PO
4
-P H2O Cell test 16 mm
388 Phosphate 00475 1.00475.0001 0.5 - 20.0 mg/l PO
4
-P H2O Cell test 16 mm
381 Phosphate 14729 1.14729.0001 0.5 - 20.0 mg/l PO
4
-P H2O Cell test 16 mm
382 Phosphate 00616 1.00616.0001 3.0 - 100.0 mg/l PO
4
-P H2O Cell test 16 mm
389 Phosphate 00673 1.00673.0001 3.0 - 100.0 mg/l PO
4
-P H2O Cell test 16 mm
383 Phosphate 14848 1.14848.0001* 0.01 - 2.50 mg/l PO
4
-P H2O Test 24 mm
384 Phosphate 00798 1.00798.0001 1.0 - 60.0 mg/l PO
4
-P H2O Test 16 mm
385 Phosphate 14842 1.14842.0001 0.5 - 30.0 mg/l PO
4
-P RB Test 16 mm
386 Phosphate 14546 1.14546.0001 0.5 - 25.0 mg/l PO
4
-P RB Cell test 16 mm
400 Potassium 14562 1.14562.0001 5.0 - 50.0 mg/l K H
2
O Cell test 16 mm
401 Potassium 00615 1.00615.0001 30 - 300 mg/l K H
2
O Cell test 16 mm 410 Residual hardness 14683 1.14683.0001 0.50 - 5.00 mg/l Ca RB Cell test 16 mm 420 Silicate 14794 1.14794.0001* 0.11 - 8.56 mg/l SiO
2
H
2
O Test 24 mm 421 Silicate 00857 1.00857.0001 11 - 1070 mg/l SiO
2
H
2
O Test 16 mm 422 Silicate 01813 1.01813.0001 0.004 - 0.500 mg/l SiO
2
RB Test 24 mm 430 Sodium 00885 1.00885.0001 10 - 300 mg/l Na RB Cell test 16 mm 444 Sulfate 02532 1.02532.0001 2.0 - 50.0 mg/l SO
4
RB Cell test 16 mm
440 Sulfate 14548 1.14548.0001 5 - 250 mg/l SO
4
H
2
O Cell test 16 mm
441 Sulfate 00617 1.00617.0001 50 - 500 mg/l SO
4
H
2
O Cell test 16 mm
442 Sulfate 14564 1.14564.0001 100 - 1000 mg/l SO
4
H
2
O Cell test 16 mm
443 Sulfate 01812 1.01812.0001 1.0 - 25.0 mg/l SO
4
H
2
O Test 24 mm
445 Sulfate 02537 1.02537.0001 5 - 300 mg/l SO
4
RB Test 16 mm
1.02537.0002 450 Sulfide 14779 1.14779.0001 0.10 - 1.50 mg/l S H
2
O Test 16 mm
460 Sulfite 14394 1.14394.0001 1.0 - 20.0 mg/l SO
3
RB Cell test 16 mm
461 Sulfite 01746 1.01746.0001 1.0 - 60.0 mg/l SO
3
RB Test 16 mm 470 Surfact-a 14697 1.14697.0001 0.05 - 2.00 mg/l MBAS RB Cell test 16 mm 473 Surfact-a 02552 1.02552.0001 0.05 - 2.00 mg/l MBAS RB Cell test 16 mm 471 Surfact-c 01764 1.01764.0001 0.05 - 1.50 mg/l RB Cell test 16 mm 472 Surfact-n 01787 1.01787.0001 0.10 - 7.50 mg/l RB Cell test 16 mm 480 Susp. solids - 50 - 750 mg/l H
2
O Method 24 mm
490 Tin 14622 1.14622.0001 0.10 - 2.50 mg/l Sn H
2
O Cell test 16 mm 500 TOC 14878 1.14878.0001 5.0 - 80.0 mg/l TOC RB Cell test 16 mm 501 TOC 14879 1.14879.0001 50 - 800 mg/l TOC RB Cell test 16 mm 510 Total hardness 00961 1.00961.0001 5 - 215 mg/l Ca RB Cell test 16 mm 520 Turbidity - 1 - 100 FAU H
2
O Method 24 mm 530 Volatile org. acids 1.01763.0001 50 - 3000 mg/l RB Cell test 16 mm 531 Volatile org. acids 1.01749.0001 50 - 3000 mg/l RB Cell test 16 mm
1.01809.0001 50 - 3000 mg/l RB Test 16 mm 540 Zinc 00861 1.00861.0001 25 - 1000 μg/l Zn RB Cell test 16 mm 541 Zinc 14566 1.14566.0001 0.20 - 5.00 mg/l Zn RB Cell test 16 mm
* in contrast to the instructions given in the package insert the sample volume as well as the reagent volume must be doubled
RB = own reagent blank value
Page 48
Meth. No. Parameter Cat. No. Measuring range Blank Type of Type of test cell
600 A 430 nm Absorbance 610 A 530 nm Absorbance 620 A 560 nm Absorbance 630 A 580 nm Absorbance 640 A 610 nm Absorbance 650 A 660 nm Absorbance
Page 49
Release 08/2014 - Spectroquant® Multy
Acid Capacity to pH 4.3
(Total Alkalinity)
101758
Cell Test
Select method 1 0 .
Measuring range: 0.40 – 8.00 mmol/l OH 16-mm cell 20 – 400 mg/l CaCO
3
16-mm cell
Pipette 4.0 ml each of AC-1 into two round cells.
Add to one cell 1.0 ml of the sample with pipette, close with the screw cap, and mix.
Add to each cell 0.50 ml of AC-2 with pipette, close with the screw cap, and mix.
Add to the second cell
1.0 ml of distilled water with pipette, close with the screw cap, and mix. (Blank cell)
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a sodium h
ydroxide solu­tion 1.0 mol/l, Cat.No. 109141, can be used after diluting accordingly (see section “Standard solutions”).
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Page 50
Release 08/2014 - Spectroquant® Multy
Measuring range: 20– 700 µg/l Al 24-mm cell
Aluminium
114825
Test
Attention: In contrast to the instructions given in the package insert the sample volume as well as the reagent volume must be doubled.
Transfer each solution into a separate 24-mm cell, close with the screw caps.
Select method 2 0 .
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 2 minutes Press Enter to start the countdown.
Check the pH of the sample, specified range
: pH 3 – 10. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Add to each test tube 2 level blue microspoons of Al-1 and dissolve the solid substance.
Add to each test tube
2.4 ml of Al-2 with pi­pette and mix.
Add to each test tube
0.5 ml of Al-3 with pi­pette and mix.
Pipette 10 ml of the sample into a test tube.
Pipette 10 ml of distilled water into a second test tube. (Blank)
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use aluminium
standard solution Certipur
®
, Cat.No. 119770, concentra­tion 1000 mg/l Al can be used after diluting accordingly. The measurement results are expressed in µg/l.
Page 51
Add to each cell 1 level blue microspoon of Al-1K, close with the screw cap.
Measuring range: 0.05 –0.50 mg/l Al 16-mm cell
Add to each test tube
0.25 ml of Al-2K with pipette, close with the screw cap, and mix.
100594
Cell Test
Aluminium
Pipette 6.0 ml of the sample into a reaction cell, close with the screw cap, and mix.
Pipette 6.0 ml of distilled water into a second reaction cell, close with the screw cap, and mix. (Blank cell)
Shake both cells vigor­ously to dissolve the solid substance.
Select method 2 1 .
Reaction time: 5 minutes Press Enter to start the countdown.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use aluminium
standard solution Certipur
®
, Cat.No. 119770, concentra­tion 1000 mg/l Al can be used after diluting accordingly.
Check the pH of the sample, specified range
: pH 3 – 10. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Release 08/2014 - Spectroquant
®
Multy
Page 52
Release 08/2014 - Spectroquant® Multy
114739
Cell Test
Measuring range: 10 –2000 µg/l NH4-N 16-mm cell 13 –2576 µg/l NH
4
16-mm cell
10 –2000 µg/l NH
3
-N 16-mm cell
12 –2432 µg/l NH
3
16-mm cell
Ammonium
Pipette 5.0 ml of the sample into a reaction cell, close with the screw cap, and mix.
Add to each cell 1 dose of NH
4
-1K using the blue
dose-metering cap, close with the screw cap.
Shake both cells vigor­ously to dissolve the solid substance.
Pipette 5.0 ml of distilled water into a second reaction cell, close with the screw cap, and mix. (Blank cell)
Select method 3 0 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 15 minutes Press Enter to start the countdown.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 50, Cat.No. 114695, or the Standard solution for photometric applications, CRM, Cat.No. 125022 and 125023. The measurement results are expressed in µg/l.
Ready-for-use ammonium standard solution
Certipur®,
Cat.No. 119812,
concentration 1000 mg/l NH
4
+
, can also
be used after diluting accordingly. To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 50) is highly recommended.
Important:
Very high ammonium concentrations in the sample produce turquoise-colored solutions (measurement solu­tion should be yellow-g
reen to green) and false-low read­ings are yielded. In such cases the sample must be di­luted (plausibility check).
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 13. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Page 53
Release 08/2014 - Spectroquant® Multy
114558
Cell Test
Measuring range: 0.20 – 8.00 mg/l NH4-N 16-mm-cell
0.26 –10.30 mg/l NH
4
16-mm-cell
0.20 – 8.00 mg/l NH
3
-N 16-mm-cell
0.24 – 9.73 mg/l NH
3
16-mm-cell
Ammonium
Pipette 1.0 ml of the sample into a reaction cell, close with the screw cap, and mix.
Add to each cell 1 dose of NH
4
-1K using the blue
dose-metering cap, close with the screw cap.
Shake both cells vigor­ously to dissolve the solid substance.
Pipette 1.0 ml of distilled water into a second reaction cell, close with the screw cap, and mix. (Blank cell)
Select method 3 1 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 15 minutes Press Enter to start the countdown.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 10, Cat.No. 114676, or the Standard solution for photometric applications, CRM, Cat.No. 125022, 125023, 125024, and 125025.
Ready-for-use ammonium standard solution
Certipur®,
Cat.No. 119812,
concentration 1000 mg/l NH
4
+
, can also
be used after diluting accordingly. To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 10) is highly recommended.
Important:
Very high ammonium concentrations in the sample produce turquoise-colored solutions (measurement solu­tion should be yellow-g
reen to green) and false-low read­ings are yielded. In such cases the sample must be di­luted (plausibility check).
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 13. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Page 54
Release 08/2014 - Spectroquant® Multy
114559
Cell Test
Measuring range: 4.0– 80.0 mg/l NH4-N 16-mm-cell
5.2 – 103.0 mg/l NH
4
16-mm-cell
4.0– 80.0 mg/l NH
3
-N 16-mm-cell
4.9– 97.3 mg/l NH
3
16-mm-cell
Ammonium
Pipette 0.10 ml of the sample into a reaction cell, close with the screw cap, and mix.
Add to each cell 1 dose of NH
4
-1K using the blue
dose-metering cap, close with the screw cap.
Shake both cells vigor­ously to dissolve the solid substance.
Pipette 0.10 ml of distilled water into a second reaction cell, close with the screw cap, and mix. (Blank cell)
Select method 3 2 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 15 minutes Press Enter to start the countdown.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 70, Cat.No. 114689, or the Standard solution for photometric applications, CRM, Cat.No. 125025, 125026, and 125027.
Ready-for-use ammonium standard solution
Certipur®,
Cat.No. 119812,
concentration 1000 mg/l NH
4
+
, can also
be used after diluting accordingly. To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 70) is highly recommended.
Important:
Very high ammonium concentrations in the sample produce turquoise-colored solutions (measurement solu­tion should be yellow-g
reen to green) and false-low read­ings are yielded. In such cases the sample must be di­luted (plausibility check).
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 13. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Page 55
Release 08/2014 - Spectroquant® Multy
Measuring range: 0.02 –1.30 mg/l NH4-N 24-mm cell
0.03 –1.67 mg/l NH
4
24-mm cell
0.02 –1.30 mg/l NH
3
-N 24-mm cell
0.02 –1.64 mg/l NH
3
24-mm cell
Ammonium
114752
Test
Add to each test tube 8 drops of NH
4
-3 and
mix.
Pipette 10 ml of the sample into a test tube.
Add to each test tube
1.2 ml of NH
4
-1 with pi-
pette and mix.
Add to each test tube 2 level blue microspoons of NH
4
-2.
Shake both test tubes vigorously to dissolve the solid substance.
Attention: In contrast to the instructions given in the package insert the sample volume as well as the reagent volume must be doubled.
Pipette 10 ml of distilled water into a second test tube. (Blank)
Transfer each solution into a separate 24-mm cell, close with the screw caps.
Select method 3 3 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 5 minutes Press Enter to start the countdown.
Reaction time: 5 minutes Press Enter to start the countdown.
Quality assurance:
To check the measurement system (test re­agents, measurement device
, and handling)
we recommended to use Spectroquant
®
CombiCheck 50, Cat.No. 114695, or the Stan­dard solution for photometric applications, CRM, Cat.No. 125022 and 125023. Use 10 ml R-1 instead of the sample.
Ready-for-use ammonium standard solution Certipur
®
, Cat.No. 119812,
concentration
1000 mg/l NH
4
+
, can also be used after diluting
accordingly. To check for sample-dependent effects the use
of addition
solutions (e.g. in CombiCheck 50) is highly recommended. Use 10 ml sample + 0.1 ml R-2.
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 13. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Important:
Very high ammonium concentrations in the sample produce turquoise-colored solutions (measurement solu­tion should be yellow-g
reen to green) and false-low read­ings are yielded. In such cases the sample must be di­luted (plausibility check).
Page 56
Release 08/2014 - Spectroquant® Multy
Measuring range: 1.0 –50.0 mg/l NH4-N 16-mm cell
1.3 –64.4 mg/l NH
4
16-mm cell
1.0 –50.0 mg/l NH
3
-N 16-mm cell
1.2 –60.8 mg/l NH
3
16-mm cell
Ammonium
100683
Test
Pipette 5.0 ml each of
NH
4
-1 into two 16-mm
cells.
Add
to the second cell
0.20 ml of distilled water with pipette, close with the screw cap, and mix. (Blank cell)
Add to each cell 1 level blue microspoon of
NH
4
-2, close with the
screw cap.
Shake both cells vigor­ously to dissolve the solid substance.
Add
to one cell 0.20 ml of the sample with pipet­te, close with the screw cap, and mix.
Select method 3 4 .
Insert the blank cell into the cell compartment. Press Zero .
Insert the cell containing the sample into the cell compartment. Press Test .
Reaction time: 15 minutes Press Enter to start the countdown.
Check the pH of the sample, specified range
: pH 4 – 13. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Important:
Very high ammonium concentrations in the sample produce turquoise-colored solutions (measurement solu­tion should be yellow-g
reen to green) and false-low read­ings are yielded. In such cases the sample must be di­luted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 70, Cat.No. 114689, or the Standard solution for photometric applications, CRM, Cat.No. 125025 and 125026.
Ready-for-use ammonium standard solution
Certipur®,
Cat.No. 119812,
concentration 1000 mg/l NH
4
+
, can also
be used after diluting accordingly. To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 70) is highly recommended.
Page 57
Release 08/2014 - Spectroquant® Multy
100675
Cell Test
Measuring range: 0.05– 2.50 mg/l AOX 16-mm cell
AOX
Adsorbable Organic Halogens (x)
Place the column in an empty cell. Fill 1 level blue microspoon of AOX-1 into the column using the glass funnel.
Run 3 separate 1-ml portions of AOX-2 through the column. Discard the wash solution.
Run 3 separate 1-ml portions of AOX-3 through the column. Discard the wash solution.
Close the bottom end of the column with the stopper. Apply to the column 1 ml of AOX-3. Close the top end of the column with the stopper and swirl to eliminate air bubbles. Remove the stopper on the top end and fill the column to the brim with AOX-3.
Attach the glass reservoir to the prepared column (closed at the bottom end).
Fill 100 ml of the sample and 6 drops of AOX-4 into the reservoir.
Remove the stopper from the column outlet and run the sample through completely.
Detach the column from the reservoir. Apply 3 separate 1-ml portions of AOX-3. Discard the wash solution.
Preparation of the adsorption column:
Sample enrichment:
Check the pH of the sample, specified range
: pH 6 – 7. If required, add dilute sodium hydroxide solution or nitric acid drop by drop to adjust the pH.
Page 58
Release 08/2014 - Spectroquant® Multy
100675
Cell Test
AOX
Adsorbable Organic Halogens (x)
Pipette 0.20 ml each of AOX-1K into two re- action cells, close with the screw cap, and mix.
Add to the second cell
7.0 ml distilled water with pipette, close with the screw cap, and mix. (Blank cell)
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) Spectroquant
®
AOX Stan­dard, Cat.No. 100680, concentration 0.2 – 2.0 mg/l can be used.
Add to one cell 7.0 ml of pretreated sample (without charcoal) with glass pipette, close with the srew cap, and mix.
Select method 4 0 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 15 minutes Press Enter to start the countdown.
Fill the 10-ml syringe with 10 ml of reagent AOX-5 and attach the syringe with the column outlet using the connector. Place the top end of the column on an empty cell and rinse the charcoal filling of the column into an empty 16-mm cell.
Add 2 level green microspoons of AOX-6, close with the screw cap, and mix.
Heat the cell at 120 °C in the thermoreactor for 30 minutes.
Add 5 drops of AOX-4, close the cell, and mix; clear supernatant: pretreated sample.
Remove the cell from the thermoreactor and place in a test-tube rack to cool to room temperature.
Digestion:
Determination:
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Page 59
Release 08/2014 - Spectroquant® Multy
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use arsenic stan-
dard solution Certipur
®
, Cat.No. 119773, concentration 1000 mg/l As can be used after diluting accordingly. The measurement results are expressed in µg/l.
Measuring range: 5 –100 µg/l As 16-mm cell
Arsenic
101747
Test
Fill approx. 10 ml of distilled water into a 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 5 0 . Insert the cell containing
the sample into the cell compartment. Press Test .
Place 350 ml of the sample into an Erlen­meyer flask with ground joint.
Add 5 drops of As-1 and mix.
Add 20 ml of As-2 with pipette and mix.
Add 1 level green dosing spoon of As-3 and dissolve.
Add 1.0 ml of As-4 with pipette and mix.
Pipette 5.0 ml of As-5 into the absorption tube.
Add 1.0 ml of As-6 with pipette to the solution in the Erlenmeyer flask and mix.
Add 3 level red dosing spoons of As-7. Imme- diately attach the ab­sorption tube to the Erlenmeyer flask.
Leave to stand for 2 hours. During this time carefully swirl the flask several times or stir slowly with a magnetic stirrer.
Transfer the solution from the absorption tube into a 16-mm cell, close with the screw cap.
Insert the blank cell into the cell compartment. Press Zero .
Check the pH of the sample, specified range
:
pH 0 – 13.
Page 60
Release 08/2014 - Spectroquant® Multy
100687
Cell Test
Measuring range: 0.5 – 3000
1)
mg/l O
2
16-mm cell
1)
after corresponding dilution (details see package insert)
BOD
Biochemical Oxygen Demand
Fill approx. 10 ml of distilled water into a 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Insert the blank cell (
nutrient-salt solution
) into the cell compart­ment. Align the mark on the cell with that on the photometer. Press Test .
Fill 2 oxygen reaction bottles each with
pretreated sample
and 2 glass beads to overflowing. Close bubble-free with the slanted ground-glass stoppers.
Fill 2 oxygen reaction bottles each with inocu-
lated nutrient-salt solu­tion and 2 glass beads
to overflowing. Close bubble-free with the slanted ground-glass stoppers.
Incubate one bottle of pretreated sample and one of inoculated
nutrient-salt solution
closed in a thermostatic incubation cabinet at 20 + 1°C for 5 days.
Use one bottle of
pretreated sample and one of inoculated nutrient-salt solution
for the measurement of the initial oxygen concentration.
After incubation, use one bottle of pretreated
sample and one of inoculated nutrient­salt solution for the
measurement of the final oxygen concentration.
Measurement of inital
oxygen concentration
= Result 1
(measurement sample)
= Result 1
(blank)
Measurement of final
oxygen concentration
= Result 2
(measurement sample)
= Result 2
(blank)
Preparation and incubation:
Determination:
Select method 7 0 . Add to each
oxygen re-
action bottle
10 drops of BOD-3K, reclose, and mix.
Add to each
oxygen
reaction bottle
5 drops of BOD-1K and then 10 drops of BOD-2K, close bubble-free, and mix for approx. 10 seconds.
Transfer each solution into a separate 16-mm cell, close with the screw caps.
Calculation:
BOD of measurement sample: Result 1 – Result 2 (measurement sample) = A in mg/l
BOD of blank: Result 1 – Result 2 (b
lank) = B in mg/l
BOD of original sample in mg/l = A • dilution factor – B
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Insert the blank cell into the cell compartment. Press Zero .
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) Spectroquant BOD Stan­dard (analogous to EN 1899), Cat.No
. 100718, can be
used.
Check the pH of the sample, specified range
: pH 6 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Reaction time: 1 minute
Page 61
Release 08/2014 - Spectroquant® Multy
100826
Cell Test
Measuring range: 0.05– 2.00 mg/l B 16-mm cell
Boron
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use boron stan-
dard solution Certipur
®
, Cat.No. 119500, concentration
1000 mg/l B can also be used after diluting accordingly.
Pipette 1.0 ml each of B-1K into two reaction cells, close with the screw cap, and mix.
Add to one cell 4.0 ml of the sample with pipette, close with the screw cap.
Shake both cells vigor­ously to dissolve the solid substance.
Add to the second cell
4.0 ml of distilled water with pipette, close with the screw cap. (Blank cell)
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 60 minutes
Check the pH of the sample, specified range
: pH 2 – 12. If required, add dilute sodium hydroxide solution or nitric acid drop by drop to adjust the pH.
Select method 8 0 .
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Page 62
Release 08/2014 - Spectroquant® Multy
Pipette 10 ml of the sample into a 24-mm cell.
Add 1 level blue micro­spoon of Br
2
-1, close
with the screw cap.
Measuring range: 0.10– 5.00 mg/l Br2 24-mm cell
Bromine
100605
Test
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Select method 9 0 .
Important:
Very high bromine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution
can be used (see section “Standard solutions”).
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Page 63
Release 08/2014 - Spectroquant® Multy
114834
Cell Test
Measuring range: 25– 1000 µg/l Cd 16-mm cell
Cadmium
Pipette 5.0 ml of the sample into a reaction cell, close with the screw cap, and mix.
Add 0.20 ml each of Cd-1K with pipette, close with the screw cap, and mix.
Add to each cell 1 level green microspoon of Cd-2K, close with the screw cap.
Shake both cells vigor­ously to dissolve the solid substance.
Pipette 5.0 ml of distilled water into a second reaction cell, close with the screw cap, and mix. (Blank cell)
Select method 1 0 0 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 2 minutes Press Enter to start the countdown.
Check the pH of the sample, specified range
: pH 3 – 11. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 30, Cat.No. 114677.
The measurement results are expressed in µg/l. Ready-for-use cadmium standard solution Certipur
®
, Cat.No. 119777, concentration 1000 mg/l Cd, can also be used after diluting accordingly.
To check for sample-dependent effects the use of addition solutions (e.g. in CombiCheck 30) is highly recommended.
Important:
For the determination of total cadmium a pretreatment with Crack Set 10C
, Cat.No. 114688 or Crack Set 10,
Cat.No. 114687, and thermoreactor is necessary.
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Page 64
Release 08/2014 - Spectroquant® Multy
Pipette 1.0 ml each of Cd-1 into two 24-mm
cells.
Add to one cell 10 ml of the sample with pipette, close with the screw cap, and mix.
Add to each cell 0.20 ml of Cd-2 with pipette, close with the screw cap, and mix.
Measuring range: 5– 500 µg/l Cd 24-mm cell
Add to each cell 1 level green microspoon of Cd-3, close with the screw cap, and dissolve the solid substance.
Cadmium
101745
Test
Add to the second cell 10 ml of distilled water with pipette, close with the screw cap, and mix. (Blank cell)
Select method 1 0 1 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 2 minutes Press Enter to start the countdown.
Check the pH of the sample, specified range
: pH 3 – 11. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use cadmium
standard solution Certipur
®
, Cat.No. 119777, concentra­tion 1000 mg/l Cd, can be used after diluting accordingly. The measurement results are expressed in µg/l.
Important:
For the determination of total cadmium a pretreatment with Crack Set 10C
, Cat.No. 114688 or Crack Set 10,
Cat.No. 114687, and thermoreactor is necessary.
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Page 65
Release 08/2014 - Spectroquant® Multy
Measuring range: 10 –250 mg/l Ca 16-mm cell 14 – 350 mg/l CaO 16-mm cell 25 – 625 mg/l CaCO
3
16-mm cell
Pipette 1.0 ml of the sample into a reaction cell, close with the screw cap, and mix.
Pipette 1.0 ml of distilled water into a second re­action cell, close with the screw cap, and mix. (Blank cell)
Add to each cell 0.50 ml of Ca-2K with pipette, close with the screw cap, and mix.
100858
Cell Test
Calcium
Add to each cell 1.0 ml of Ca-1K with pipette, close with the screw cap, and mix.
Select method 1 1 0 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Check the pH of the sample, specified range
: pH 3 – 9. If required, add dilute sodium hydroxide solution or hydrochloric acid drop by drop to adjust the pH.
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Reaction time: exactly 3 minutes Press Enter to start the countdown.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Page 66
Release 08/2014 - Spectroquant® Multy
Pipette 0.10 ml of the sample into a 16-mm cell.
Add to each cell 4 drops of Ca-2, close with the screw cap, and mix.
Measuring range: 5 –160 mg/l Ca 16-mm cell 7 – 224 mg/l CaO 16-mm cell 13 – 400 mg/l CaCO
3
16-mm cell
Calcium
114815
Test
Add to each cell 4 drops of Ca-3, close with the screw cap, and mix.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use calcium stan-
dard solution Certipur
®
, Cat.No. 119778, concentration
1000 mg/l Ca, can be used after diluting accordingly.
Add to each cell 5.0 ml of Ca-1 with pipette, close with the screw cap, and mix.
Pipette 0.10 ml of distilled water into a second 16-mm cell. (Blank cell)
Select method 1 1 1 .
Insert the cell containing the sample into the cell compartment. Press Test .
Check the pH of the sample, specified range
: pH 4 – 10. If required, add dilute sodium hydroxide solution or hydrochloric acid drop by drop to adjust the pH.
Reaction time: 8 minutes, measure immediately. Press Enter to start the countdown.
Insert the blank cell into the cell compartment. Press Zero .
Page 67
Release 08/2014 - Spectroquant® Multy
114730
Cell Test
Measuring range: 5–125 mg/l Cl 16-mm cell
Chloride
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 10 and 20, Cat.No. 114676
and 114675. Ready-for-use chloride standard solution Certipur
®
,
Cat.No. 119897, concentration 1000 mg/l Cl
–
, can also
be used after diluting accordingly. To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck) is highly recommended.
Pipette 0.50 ml each of
Cl-1K
into two reaction cells, close with the screw cap, and mix
.
Add to one cell 1.0 ml of the sample with pipette, close with the screw cap, and mix.
Add to the second cell
1.0 ml of distilled water, close with the screw cap, and mix. (Blank cell)
Select method 1 2 0 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Check the pH of the sample, specified range
: pH 1 – 12. If required, add dilute ammonia solution or nitric acid drop by drop to adjust the pH.
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Page 68
Release 08/2014 - Spectroquant® Multy
Pipette 1.0 ml of the sample into a 16-mm cell.
Add to each cell 2.5 ml of Cl-1 with pipette, close with the screw cap, and mix.
Measuring range: 10– 250 mg/l Cl 16-mm cell
Chloride
114897
Test
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 60, Cat.No. 114696.
Ready-for-use chloride standard solution Certipur
®
,
Cat.No. 119897, concentration 1000 mg/l Cl
–
, can also
be used after diluting accordingly. To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 60) is highly recommended.
Add to each cell 0.50 ml of Cl-2 with pipette, close with the screw cap, and mix.
Pipette 1.0 ml of distilled water into a second 16-mm cell. (Blank cell)
Select method 1 2 1 .
Insert the cell containing the sample into the cell compartment. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Check the pH of the sample, specified range
: pH 1 – 12. If required, add dilute ammonia solution or nitric acid drop by drop to adjust the pH.
Insert the blank cell into the cell compartment. Press Zero .
Page 69
Release 08/2014 - Spectroquant® Multy
101804
Cell Test
Measuring range: 0.5–15.0 mg/l Cl 16-mm cell
Chloride
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Check the pH of the sample, specified range
: pH 3 – 11. If required, add dilute ammonia solution or nitric acid drop by drop to adjust the pH.
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Pipette 10 ml of the sample into a reaction cell, close with the screw cap, and mix.
Add 0.25 ml each of Cl-1K with pipette, close with the screw cap, and mix.
Pipette 10 ml of distilled water into a second reaction cell, close with the screw cap, and mix. (Blank cell)
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use chloride
standard solution Certipur
®
, Cat.No. 119897, concentra-
tion 1000 mg/l Cl
-
, can be used after diluting accordingly.
Select method 1 2 2 .
Reaction time: 10 minutes Press Enter to start the countdown.
Page 70
Release 08/2014 - Spectroquant® Multy
Pipette 0.20 ml each of Cl-1 into two 24-mm cells.
Add
to the second cell 10 ml of distilled water (Water for analysis EMSURE
®
, Cat.No. 116754, is recommen­ded) with pipette, close with the screw cap, and mix.(Blank cell)
Measuring range: 0.50– 5.00 mg/l Cl 24-mm cell
Chloride
101807
Test
Add to each cell 0.20 ml of Cl-2 with pipette, close with the screw cap, and mix.
Add 10 ml of the sam­ple with pipette, close with the screw cap, and mix.
Check the pH of the sample, specified range
: pH 3 – 11. If required, add dilute ammonia solution or nitric acid drop by drop to adjust the pH.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use chloride
standard solution Certipur
®
, Cat.No. 119897, concentra-
tion 1000 mg/l Cl
-
, can be used after diluting accordingly.
Select method 1 2 3 .
Reaction time: 10 minutes Press Enter to start the countdown.
Page 71
Release 08/2014 - Spectroquant® Multy
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Pipette 5.0 ml of the sample into a round cell.
100595
Cell Test
Measuring range: 0.05– 5.00 mg/l Cl
2
16-mm cell
Chlorine
Determination of free chlorine
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 0 , select subitem >>free.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Insert the blank cell into the cell compartment. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Page 72
Release 08/2014 - Spectroquant® Multy
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Pipette 5.0 ml of the sample into a round cell.
100597
Cell Test
Measuring range: 0.05– 5.00 mg/l Cl
2
16-mm cell
Chlorine
Determination of free chlorine
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 0 , select subitem >>free.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Insert the blank cell into the cell compartment. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Page 73
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Pipette 5.0 ml of the sample into a round cell.
100597
Cell Test
Measuring range: 0.05– 5.00 mg/l Cl
2
16-mm cell
Chlorine
Determination of total chlorine
Add 2 drops of Cl2-2, close with the screw cap, and mix.
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 0 , select subitem >>total.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard from Chloramine T GR, Cat.No. 102426, can be used (see section “Standard solutions”).
Insert the blank cell into the cell compartment. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Release 08/2014 - Spectroquant
®
Multy
Page 74
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Pipette 5.0 ml of the sample into a round cell.
100597
Cell Test
Measuring range: 0.05– 5.00 mg/l Cl
2
16-mm cell
Chlorine
Determination of free chlorine, total chlorine,
and combined chlorine
Remove sample cell from the photometer, open, add 2 drops of
Cl
2
-2, close with the
screw cap, and mix.
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 0 , select subitem >>diff.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test . (= T1)
Reaction time: 1 minute Press Enter to start the countdown.
Insert anew the cell con­taining the sample into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Test .
(= T2)
Insert the blank cell into the cell compartment. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solu­tions (measurement solution should be red) and false-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water. For on-the-spot determinations where there are no suitable facilities for rinsing, the cell contents can be transferred to a new 16-mm cell before the addition of reagent Cl
2
-2. Use this second cell only for the determination of total chlorine!
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a freshly prepared standard solution can be used (see section “Stan­dard solutions”).
Release 08/2014 - Spectroquant® Multy
Page 75
Release 08/2014 - Spectroquant® Multy
Pipette 10 ml of the sample into a 24-mm cell.
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Measuring range: 0.02– 4.50 mg/l Cl2 24-mm cell
Chlorine
Determination of free chlorine
100598
Test
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 1 , select subitem >>free.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Page 76
Insert the cell containing the sample into the cell compartment. Press Test .
Insert the blank cell into the cell compartment. Press Zero .
Release 08/2014 - Spectroquant® Multy
Pipette 10 ml of the sample into a 16-mm cell.
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Measuring range: 0.10– 6.00 mg/l Cl2 16-mm cell
Chlorine
Determination of free chlorine
100598
Test
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 4 , select subitem >>free.
Reaction time: 1 minute Press Enter to start the countdown.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Page 77
Release 08/2014 - Spectroquant® Multy
Pipette 10 ml of the sample into a 24-mm cell.
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Measuring range: 0.02– 4.50 mg/l Cl2 24-mm cell
Chlorine
Determinatin of total chlorine
100602
Test
Add 2 drops of Cl2-2, close with the screw cap, and mix.
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 1 , select subitem >>total.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard from Chloramine T GR, Cat.No. 102426, can be used (see section “Standard solutions”).
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water.
Page 78
Insert the cell containing the sample into the cell compartment. Press Test .
Insert the blank cell into the cell compartment. Press Zero .
Release 08/2014 - Spectroquant® Multy
Pipette 10 ml of the sample into a 16-mm cell.
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Measuring range: 0.10– 6.00 mg/l Cl2 16-mm cell
Chlorine
Determinatin of total chlorine
100602
Test
Add 2 drops of Cl2-2, close with the screw cap, and mix.
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 4 , select subitem >>total.
Reaction time: 1 minute Press Enter to start the countdown.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard from Chloramine T GR, Cat.No. 102426, can be used (see section “Standard solutions”).
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water.
Page 79
Release 08/2014 - Spectroquant® Multy
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Measuring range: 0.02– 4.50 mg/l Cl2 24-mm cell
Chlorine
Detemination of free chlorine
100599
Test
Shake the cell vigor­ously to dissolve the solid substance.
Pipette 10 ml of the sample into a 24-mm cell.
Select method 1 3 1 , select subitem >>free.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Page 80
Insert the cell containing the sample into the cell compartment. Press Test .
Insert the blank cell into the cell compartment. Press Zero .
Release 08/2014 - Spectroquant® Multy
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Measuring range: 0.10– 6.00 mg/l Cl2 16-mm cell
Chlorine
Detemination of free chlorine
100599
Test
Shake the cell vigor­ously to dissolve the solid substance.
Pipette 10 ml of the sample into a 16-mm cell.
Select method 1 3 4 , select subitem >>free.
Reaction time: 1 minute Press Enter to start the countdown.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Page 81
Pipette 10 ml of the sample into a 24-mm cell.
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Measuring range: 0.02– 4.50 mg/l Cl2 24-mm cell
Chlorine
Determination of total chlorine
100599
Test
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Add 2 drops of Cl2-2, close with the screw cap, and mix.
Select method 1 3 1 , select subitem >>total.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard from Chloramine T GR, Cat.No. 102426, can be used (see section “Standard solutions”).
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Release 08/2014 - Spectroquant
®
Multy
Page 82
Insert the cell containing the sample into the cell compartment. Press Test .
Insert the blank cell into the cell compartment. Press Zero .
Pipette 10 ml of the sample into a 16-mm cell.
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Measuring range: 0.10– 6.00 mg/l Cl2 16-mm cell
Chlorine
Determination of total chlorine
100599
Test
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Add 2 drops of Cl2-2, close with the screw cap, and mix.
Select method 1 3 4 , select subitem >>total.
Reaction time: 1 minute Press Enter to start the countdown.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard from Chloramine T GR, Cat.No. 102426, can be used (see section “Standard solutions”).
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Release 08/2014 - Spectroquant
®
Multy
Page 83
Pipette 10 ml of the sample into a 24-mm cell.
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Measuring range: 0.02– 4.50 mg/l Cl2 24-mm cell
Chlorine
Determination of free chlorine, total chlorine,
and combined chlorine
100599
Test
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Remove the sample cell from the photometer, open, add 2 drops of
Cl
2
-2, close with the
screw cap, and mix.
Select method 1 3 1 , select subitem >>diff.
Reaction time: 1 minute Press Enter to start the countdown.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test . (= T1)
Insert anew the cell con­taining the sample into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Test .
(= T2)
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solu­tions (measurement solution should be red) and false-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water. For on-the-spot determinations where there are no suitable facilities for rinsing, the cell contents can be transferred to a new 24-mm cell before the addition of reagent Cl
2
-2. Use this second cell only for the determination of total chlorine!
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a freshly prepared standard solution can be used (see section “Stan­dard solutions”).
Release 08/2014 - Spectroquant® Multy
Page 84
Pipette 10 ml of the sample into a 16-mm cell.
Add 1 level blue micro­spoon of Cl
2
-1, close
with the screw cap.
Measuring range: 0.10– 6.00 mg/l Cl2 16-mm cell
Chlorine
Determination of free chlorine, total chlorine,
and combined chlorine
100599
Test
Shake the cell vigor­ously to dissolve the solid substance.
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Remove the sample cell from the photometer, open, add 2 drops of
Cl
2
-2, close with the
screw cap, and mix.
Select method 1 3 4 , select subitem >>diff.
Reaction time: 1 minute Press Enter to start the countdown.
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solu­tions (measurement solution should be red) and false-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water. For on-the-spot determinations where there are no suitable facilities for rinsing, the cell contents can be transferred to a new 24-mm cell before the addition of reagent Cl
2
-2. Use this second cell only for the determination of total chlorine!
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a freshly prepared standard solution can be used (see section “Stan­dard solutions”).
Release 08/2014 - Spectroquant® Multy
Insert the cell containing the sample into the cell compartment. Press Test . (= T1)
Insert the blank cell into the cell compartment. Press Zero .
Insert anew the cell con­taining the sample into the cell compartment. Press Test . (= T2)
Page 85
Add 3 drops of Cl2-2, close with the screw cap, and mix.
Place
6 drops of Cl2-1
into a round cell
.
Measuring range: 0.10– 6.00 mg/l Cl2 16-mm cell
Add 10 ml of the sample with pipette
, close with
the screw cap, and mix.
Chlorine (with liquid reagents)
Detemination of free chlorine
100086/100087
Cell Test
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 2 , select subitem >>free.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Insert the blank cell into the cell compartment. Press Zero .
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Release 08/2014 - Spectroquant
®
Multy
Reaction time: 1 minute Press Enter to start the countdown.
Page 86
Add 3 drops of Cl2-2, close with the screw cap, and mix.
Place
6 drops of Cl2-1
into a 24-mm cell
.
Measuring range: 0.02– 4.50 mg/l Cl2 24-mm cell
Add 10 ml of the sample with pipette
, close with
the screw cap, and mix.
Chlorine (with liquid reagents)
Detemination of free chlorine
100086/100087
Test
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 3 , select subitem >>free.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Release 08/2014 - Spectroquant
®
Multy
Reaction time: 1 minute Press Enter to start the countdown.
Page 87
Add 3 drops of Cl2-2, close with the screw cap, and mix.
Place
6 drops of Cl2-1
into a round cell
.
Measuring range: 0.10– 6.00 mg/l Cl2 16-mm cell
Add 10 ml of the sample with pipette
, close with
the screw cap, and mix.
Chlorine (with liquid reagents)
Detemination of total chlorine
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 2 , select subitem >>total.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Insert the blank cell into the cell compartment. Press Zero .
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard from Chloramine T GR, Cat.No. 102426, can be used (see section “Standard solutions”).
100086/100087/
100088
Cell Test
Add 2 drops of Cl2-3, close with the screw cap, and mix.
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Release 08/2014 - Spectroquant
®
Multy
Reaction time: 1 minute Press Enter to start the countdown.
Page 88
Add 3 drops of Cl2-2, close with the screw cap, and mix.
Place
6 drops of Cl2-1
into a 24-mm cell
.
Measuring range: 0.02– 4.50 mg/l Cl2 24-mm cell
Add 10 ml of the sample with pipette
, close with
the screw cap, and mix.
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 3 , select subitem >>total.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard from Chloramine T GR, Cat.No. 102426, can be used (see section “Standard solutions”).
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Chlorine (with liquid reagents)
Detemination of total chlorine
100086/100087/
100088
Test
Add 2 drops of Cl2-3, close with the screw cap, and mix.
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water.
Release 08/2014 - Spectroquant® Multy
Reaction time: 1 minute Press Enter to start the countdown.
Page 89
Add 3 drops of Cl2-2, close with the screw cap, and mix.
Place
6 drops of Cl2-1
into a round cell
.
Measuring range: 0.10– 6.00 mg/l Cl2 16-mm cell
Add 10 ml of the sample with pipette
, close with
the screw cap, and mix.
Chlorine (with liquid reagents)
Determination of free chlorine,
total chlorine, and combined chlorine
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 3 2 , select subitem >>dif.
Insert the blank cell into the cell compartment. Press Zero .
100086/100087/
100088
Cell Test
Remove the sample cell from the photometer, open, add 2 drops of
Cl
2
-3, close with the
screw cap, and mix.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test . (= T1)
Insert anew the cell con­taining the sample into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Test .
(= T2)
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solu­tions (measurement solution should be red) and false-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water. For on-the-spot determinations where there are no suitable facilities for rinsing, the cell contents can be transferred to a new 16-mm cell before the addition of reagent Cl
2
-3. Use this second cell only for the determination of total chlorine!
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a freshly prepared standard solution can be used (see section “Stan­dard solutions”).
Release 08/2014 - Spectroquant® Multy
Reaction time: 1 minute Press Enter to start the countdown.
Page 90
Add 3 drops of Cl2-2, close with the screw cap, and mix.
Measuring range: 0.02– 4.50 mg/l Cl2 24-mm cell
Add 10 ml of the sample with pipette
, close with
the screw cap, and mix.
Chlorine (with liquid reagents)
Determination of free chlorine,
total chlorine, and combined chlorine
Select method 1 3 3 , select subitem >>diff.
100086/100087/
100088
Test
Remove the sample cell from the photometer, open, add 2 drops of
Cl
2
-3, close with the
screw cap, and mix.
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test . (= T1)
Insert anew the cell con­taining the sample into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Test .
(= T2)
Place
6 drops of Cl2-1
into a 24-mm cell
.
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Important:
Very high chlorine concentrations in the sample produce yellow-colored solu­tions (measurement solution should be red) and false-low readings are yielded. In such cases the sample must be diluted (plausibility check). After each determination of total chlorine rinse the cell with sulfuric acid 25 % and subsequently several times with distilled water. For on-the-spot determinations where there are no suitable facilities for rinsing, the cell contents can be transferred to a new 24-mm cell before the addition of reagent Cl
2
-3. Use this second cell only for the determination of total chlorine!
Quality assurance:
To check the measurement system (test reagents, measurement device, and handling) a freshly prepared standard solution can be used (see section “Stan­dard solutions”).
Release 08/2014 - Spectroquant® Multy
Reaction time: 1 minute Press Enter to start the countdown.
Page 91
Release 08/2014 - Spectroquant® Multy
Pipette 10 ml of the sample into a 24-mm cell.
Add 2 drops of ClO
2
-1,
close with the screw cap, and mix.
Chlorine Dioxide
100608
Test
Add 1 level blue micro­spoon of ClO
2
-2, close
with the screw cap.
Shake the cell vigor­ously to dissolve the solid substance.
Measuring range: 0.05– 8.50 mg/l ClO2 24-mm cell
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 4 0 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Important:
Very high chlorine dioxide concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such
cases the sample must be diluted (plausibility check).
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Reaction time: 1 minute Press Enter to start the countdown.
Reaction time: 2 minutes Press Enter to start the countdown.
Page 92
Insert the cell containing the sample into the cell compartment. Press Test .
Insert the blank cell into the cell compartment. Press Zero .
Release 08/2014 - Spectroquant® Multy
Pipette 10 ml of the sample into a 16-mm cell.
Add 2 drops of ClO
2
-1,
close with the screw cap, and mix.
Chlorine Dioxide
100608
Test
Add 1 level blue micro­spoon of ClO
2
-2, close
with the screw cap.
Shake the cell vigor­ously to dissolve the solid substance.
Measuring range: 0.20– 10.00 mg/l ClO2 16-mm cell
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) a freshly prepared standard solution can be used (see section “Standard solutions”).
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 4 1 .
Important:
Very high chlorine dioxide concentrations in the sample produce yellow-colored solutions (measurement solution should be red) and f
alse-low readings are yielded. In such
cases the sample must be diluted (plausibility check).
Check the pH of the sample, specified range
: pH 4 – 8. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Reaction time: 1 minute Press Enter to start the countdown.
Reaction time: 2 minutes Press Enter to start the countdown.
Page 93
Release 08/2014 - Spectroquant® Multy
114552
Cell Test
Measuring range: 0.05– 2.00 mg/l Cr 16-mm cell
0.11– 4.46 mg/l CrO
4
16-mm cell
Chromate
Determination of chromium(VI)
Add 6 drops of Cr-3K into a reaction cell, close with the screw cap.
Shake the cell vigor­ously to dissolve the solid substance and leave to stand for
1 minute.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use chromate
standard solution Certipur
®
, Cat.No. 119780, concen -
tration 1000 mg/l CrO
4
2–
, can be used after diluting
accordingly.
Add 5.0 ml of the sample with pipette
, close with
the screw cap, and mix.
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 5 0 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Check the pH of the sample, specified range
: pH 1 – 9. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Insert the blank cell into the cell compartment. Press Zero .
Page 94
Release 08/2014 - Spectroquant® Multy
114552
Cell Test
Chromate
Determination of total chromium
= sum of chromium(VI) and chromium(III)
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use chromate
standard solution Certipur
®
, Cat.No. 119780, concen-
tration 1000 mg/l CrO
4
2–
, can be used after diluting
accordingly.
Pipette 10 ml of the sample into an empty 16-mm cell.
Measuring range: 0.05– 2.00 mg/l Cr 16-mm cell
0.11– 4.46 mg/l CrO
4
16-mm cell
Fill approx. 10 ml of distilled water into an empty 16-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 5 0 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Check the pH of the sample, specified range
: pH 1 – 9. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Add 6 drops of Cr-3K into a reaction cell, close with the screw cap.
Shake the cell vigorous­ly to dissolve the solid substance and leave to stand for 1 minute.
Add 5.0 ml of the pre- treated sample
with
pipette
, close with the
screw cap, and mix.
Add 1 drop of Cr-1K, close with the screw cap, and mix.
Add 1 dose of Cr-2K using the blue dose­metering cap, close the reaction cell with the screw cap.
Heat the cell in the thermoreactor at 120°C (100°C) for 1 hour.
Remove the cell from the thermoreactor and place in a test-tube rack to cool to room tempera­ture:
pretreated sample
.
Insert the blank cell into the cell compartment. Press Zero .
Page 95
Release 08/2014 - Spectroquant® Multy
114758
Test
Shake the cell vigorous­ly to dissolve the solid substance.
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) ready-f
or-use chromate
standard solution Certipur
®
, Cat.No. 119780, concen-
tration 1000 mg/l CrO
4
2–
, can be used after diluting accordingly. The measurement results are expressed in µg/l.
Add 10 ml of the sample with pipette
, close with
the screw cap, and mix.
Place 2 level grey micro­spoons of Cr-1 into a dry 24-mm cell.
Add 12 drops of Cr-2, close with the screw cap.
Chromate
Determination of chromium(VI)
Important:
For the determination of total chromium = sum of c
hromium(VI) and chromium(III) a pretreatment
with Crack Set 10C, Cat.No. 114688, or Crack Set 10, Cat.No. 114687 and thermoreactor is necessary.
Measuring range: 10 –1400 µg/l Cr 24-mm cell 22 –3123 µg/l CrO
4
24-mm cell
Attention: In contrast to the instructions given in the package insert the sample volume as well as the reagent volume must be doubled.
Fill approx. 10 ml of distilled water into a 24-mm cell (do not add any reagents!), close with the screw cap
.
(Blank cell)
Select method 1 5 1 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Reaction time: 1 minute Press Enter to start the countdown.
Check the pH of the sample, specified range
: pH 1 – 9. If required, add dilute sodium hydroxide solution or sulfuric acid drop by drop to adjust the pH.
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Page 96
Release 08/2014 - Spectroquant® Multy
101796
Cell Test
COD
Chemical Oxygen Demand
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 50, Cat.No. 114695, or the Standard solution for photometric applications, CRM, Cat.No. 125028.
To check for sample-dependent effects the use of addition solutions (e.g. in CombiCheck 50) is highly recommended.
Measuring range: 5.0–80.0 mg/l COD or O2 16-mm cell
Select method 1 6 8 . Insert the cell containing
the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Carefully pipette 2.0 ml of the sample into a reaction cell, close tight­ly with the screw cap, and mix vigorously.
Caution, the cell becomes very hot!
Suspend the bottom sediment in two cells by swirling.
Heat both cells in the thermoreactor at 148°C for 2 hours.
Remove both cells from the thermoreactor and place in a test-tube rack to cool.
Swirl both cells after 10 minutes.
Replace both cells in the rack for complete cool­ing to room temperature. (Very important!)
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Carefully pipette 2.0 ml of distilled water (Water for chromatography LiChrosolv
®
, Cat.No. 115333, is recommen­ded) into a second reac­tion cell, close tightly with the screw cap, and mix vigorously.
Caution, the cell becomes very hot!
(Blank cell)
Page 97
Release 08/2014 - Spectroquant® Multy
114540
Cell Test
COD
Chemical Oxygen Demand
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 10, Cat.No. 114676, or the Standard solution for photometric applications, CRM, Cat.No. 125029.
To check for sample-dependent effects the use of addition solutions (e.g. in CombiCheck 10) is highly recommended.
Measuring range: 10–150 mg/l COD or O2 16-mm cell
Select method 1 6 0 . Insert the cell containing
the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Carefully pipette 3.0 ml of the sample into a reaction cell, close tight­ly with the screw cap, and mix vigorously.
Caution, the cell becomes very hot!
Suspend the bottom sediment in two cells by swirling.
Heat both cells in the thermoreactor at 148°C for 2 hours.
Remove both cells from the thermoreactor and place in a test-tube rack to cool.
Swirl both cells after 10 minutes.
Replace both cells in the rack for complete cool­ing to room temperature. (Very important!)
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Carefully pipette 3.0 ml of distilled water (Water for chromatography LiChrosolv
®
, Cat.No. 115333, is recommen­ded) into a second reac­tion cell, close tightly with the screw cap, and mix vigorously.
Caution, the cell becomes very hot!
(Blank cell)
Page 98
114895
Cell Test
COD
Chemical Oxygen Demand
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 60, Cat.No. 114696, or the Standard solution for photometric applications, CRM, Cat.No. 125029 and 125030.
To check for sample-dependent effects the use of addition solutions (e.g. in CombiCheck 60) is highly recommended.
Measuring range: 15–300 mg/l COD or O2 16-mm cell
Select method 1 6 1 . Insert the cell containing
the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Carefully pipette 2.0 ml of the sample into a reaction cell, close tight­ly with the screw cap, and mix vigorously.
Caution, the cell becomes very hot!
Suspend the bottom sediment in two cells by swirling.
Heat both cells in the thermoreactor at 148°C for 2 hours.
Remove both cells from the thermoreactor and place in a test-tube rack to cool.
Swirl both cells after 10 minutes.
Replace both cells in the rack for complete cool­ing to room temperature. (Very important!)
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Carefully pipette 2.0 ml of distilled water (Water for chromatography LiChrosolv
®
, Cat.No. 115333, is recommen­ded) into a second reac­tion cell, close tightly with the screw cap, and mix vigorously.
Caution, the cell becomes very hot!
(Blank cell)
Release 08/2014 - Spectroquant
®
Multy
Page 99
114690
Cell Test
COD
Chemical Oxygen Demand
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 60, Cat.No. 114696, or the Standard solution for photometric applications, CRM, Cat.No. 125029, 125030, and 125031.
To check for sample-dependent effects the use of addition solutions (e.g. in CombiCheck 60) is highly recommended.
Measuring range: 50–500 mg/l COD or O2 16-mm cell
Select method 1 6 2 . Insert the cell containing
the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Carefully pipette 2.0 ml of the sample into a reaction cell, close tight­ly with the screw cap, and mix vigorously.
Caution, the cell becomes very hot!
Suspend the bottom sediment in two cells by swirling.
Heat both cells in the thermoreactor at 148°C for 2 hours.
Remove both cells from the thermoreactor and place in a test-tube rack to cool.
Swirl both cells after 10 minutes.
Replace both cells in the rack for complete cool­ing to room temperature. (Very important!)
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Carefully pipette 2.0 ml of distilled water (Water for chromatography LiChrosolv
®
, Cat.No. 115333, is recommen­ded) into a second reac­tion cell, close tightly with the screw cap, and mix vigorously.
Caution, the cell becomes very hot!
(Blank cell)
Release 08/2014 - Spectroquant
®
Multy
Page 100
114541
Cell Test
COD
Chemical Oxygen Demand
Quality assurance:
To check the measurement system (test reagents, meas­urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 20, Cat.No. 114675, or the Standard solution for photometric applications, CRM, Cat.No. 125029, 125030, 125031, and 125032.
To check for sample-dependent effects the use of addition solutions (e.g. in CombiCheck 20) is highly recommended.
Measuring range: 25–1500 mg/l COD or O2 16-mm cell
Carefully pipette 3.0 ml of distilled water into a second reaction cell, close tightly with the screw cap, and mix vigor­ously.
Caution, the cell becomes very hot!
(Blank cell)
Select method 1 6 3 .
Insert the cell containing the sample into the cell compartment. Align the mark on the cell with that on the photometer. Press Test .
Carefully pipette 3.0 ml of the sample into a reaction cell, close tight­ly with the screw cap, and mix vigorously.
Caution, the cell becomes very hot!
Suspend the bottom sediment in two cells by swirling.
Heat both cells in the thermoreactor at 148°C for 2 hours.
Remove both cells from the thermoreactor and place in a test-tube rack to cool.
Swirl both cells after 10 minutes.
Replace both cells in the rack for complete cool­ing to room temperature. (Very important!)
Insert the blank cell into the cell compartment. Align the mark on the cell with that on the pho­tometer. Press Zero .
Release 08/2014 - Spectroquant® Multy
Loading...