MILLENNIUM
Operation Manual
SF822B Blood Pressure Monitor
• Welcome to use MILLENNIUM Microcomputer Blood Pressure
Monitor Series.
• Please read the Manual carefully before use so as to operate the
Blood Pressure Monitor correctly.
• Please retain this manual for future reference.
.

- 1 of 35 -
Foreword
Hypertension (High Blood Pressure) is one of the three major
diseases that are of great danger to mankind. It is known as
the INVISIBLE KILLER. The higher the blood pressure, the
higher the incidence of the heart disease or incidental
cerebrovascular (apoplexy) disease. Therefore, early
diagnosis as well as prompt treatment of hypertension is of
great significance.
Attention :
Hypertension could cause the narrow of arterial blood vessels
or the obstacle in twig blood circulation, which shall cause the
difference of blood pressure values between wrist blood
pressure and upper arm blood pressure. Even for healthy
persons, the difference between wrist and arm blood pressure
values might be about 20mmHg (2.5Kpa). So, please do not
judge your blood pressure by yourself and consult your
doctors.

- 2 of 35 -
Blood pressure values change is the same by measuring wrist
or upper arm. Therefore, one can monitor the change of his
blood pressure values by measuring wrist blood pressure.
Blood pressure changes throughout the day and even without
feeling, body blood pressure will change ranging from
30mmHg-50mmHg according to body tension or relaxation
state.
About Blood Pressure
Blood pressure is the pressure exerted on the arterial walls
when blood flows through the arteries. The pressure exerted
when the heart contracts and sends out blood is called systolic
(highest) pressure. The pressure exerted when the heart dilates
and blood flows into the heart is called diastolic (lowest)
pressure. Blood pressure is a physiological index subject to
wide fluctuations, which might be influenced by talk, movement,
posture, mood, etc., and possesses the property of periodical
change throughout the day. And therefore, using different body

- 3 of 35 -
parts for measurement (such as wrist or upper arm), different
blood vessel wall thickness as well as different measurement
time will probably generate different measuring results.
SF822B
It uses the Oscillometric method of measurement and
automatically pressurizes and measures blood pressure and
is capable of displaying minor changes of the blood pressure
measured. It also has saving function and can save pulse,
blood pressure values and measured time.

- 4 of 35 -
Special Features of SF822A Microcomputer
Wrist Blood Pressure Monitor
z All in one compact unit and easy to be portable.
z Memory function for storing 30 measurements for 3
persons including measured time, total 90 memories.
z Query function to view saved pressure data available,
helping you constantly monitor your blood pressure.
z Display time and date (year, month, day, hour and minute).
z Clear function to erase saved data available.
z Super large LCD display and easy to read.
z User friendly operation and automatically pressurize
cyclically.
z Automatically shut off function after two minutes upon
finishing measurement.

- 6 of 35 -
Operation Guide
1. Remove the battery compartment cover as illustrated by
sliding the cover in the direction of the arrow.
2. Insert the batteries as per the polarity indicated inside the
compartment.
3. Replace the battery cover.
Batteries Installation

- 7 of 35 -
Caution:
1. Please remove batteries when unit is not in use for
extended periods.
2. Replace promptly with new batteries in case low battery
indicator is displayed on the LCD (must be two NEW AAA
size batteries of the same brand).
3. Please use only alkaline batteries instead of
rechargeable ones.
Please replace batteries in case
of the following circumstances:
1. A battery symbol appears on the
Display while in use as illustrated.
2. Nothing appears on LCD screen when unit is turned on.

- 8 of 35 -
Time Display and Set
1. Time Display
When the unit is on, the system
can display time and date
except for query and save
mode. Day/Month, Year is in
the left lower corner of the
screen such us 18/01 2004
09:28. In QUE mode, displayed
time is measured time of these measurements.
2. Time Set
Time can be set after the
batteries are replaced. At this
time, MODE key is time set key
“+” and can not be used for
other functions.

- 9 of 35 -
The detailed time set process is as following procedures.
Turn on the unit after replacing batteries.
The screen time display is 01/01 (day/month), 2000(year) and
00:00(hour: minute). Moreover, the first character of “day” is
flashing and by then “day” can be set by pressing MODE/+ key.
After pressing set key to confirm, the second character of
“day” can be set Day/month set can be set cyclically. Upon
finishing set “year” and pressing SET key, if day/month and
year set is correct, the system
will automatically transfer to
“hour” set (the first character of
“hour” is flashing.) At this time,
hour and minute can be set
cyclically. The time set will finish
after pressing SET key and the
system will automatically enter
waiting command status.
If day/month/year date set is incorrect, for example, 31/02

- 10 of 35 -
2003, the screen will display “EE”. Please power on the unit
and re-set time.
Remark:
If time set is not finished, Measure, Save, Query functions
shall not be operated; when the device is switched on again,
time set is still needed.
Fastening the Wrist Holder
1. Turn left hand so that palm is facing
upward. Wrap the Velcro wrist holder
over the left wrist with the LCD display
facing you. The wrist holder should be in
contact with your skin directly with no
shirt sleeve or other clothing between the
Wrist Holder and your skin as this may
Measuring Blood Pressure

- 11 of 35 -
lead to inaccurate readings.
2. The upper end of the unit should be located about 5mm to
10mm away from the palm base line and the centerline of the
unit should be in line with the middle finger as shown in the
figure. Avoid capitulum ulnae (wrist
tuberositas ulnae bone) wrapped in
the wrist cuff.
3. Pull up the flap of the Wrist Holder
and roll to fasten it as illustrated, neither
leaving any space between the Wrist
Holder and your wrist nor tightening it
too much so as to feel discomfort.
4. If the wrist is thick, do not fasten the
wrist holder too hard avoid the distortion of the wrist holder.
Otherwise, it will affect the measurement.

- 12 of 35 -
5. If the left hand wrist cannot be used due to reasons such as
injury, sprain etc, you may use your right hand wrist following
the above procedure.

- 13 of 35 -
Posture for Measuring
Be seated as shown while measuring
blood pressure.
1.Place your left arm with the unit
wrapped around your wrist gently
on a table .
2. Place your forearm on the plastic
case (the packing case of the unit ) as illustrated, making the
center of the Wrist Holder in the same horizontal level as that
of your heart.
3. The supporting plastic case should not press against the
Wrist Holder, and the wrist should be held straight as shown.
Neither incurve nor outcurve your wrist.
4. Keep yourself dispassionate and relax your entire body.
Remarks:
Please try to maintain correct posture while measuring. Avoid
leaning forward or backward during measurements to get
correct readings.

- 14 of 35 -
Do not speak and do not touch the monitor when measuring is
taking place.
Your blood pressure readings displayed may be higher
than normal under the following circumstances:
a) if you measure immediately after drinking alcohol or
smoking, having meals or after a strenuous physical
workout
b) if you are stressed out, have mood swings or have had
inadequate sleep
c) if you feel like going to the lavatory
Measuring Method
1. Press the On/Off button. All
characters and symbols will flash
on the screen, confirming normal
display.

- 15 of 35 -
2. After about one second , the
air relief symbol Ð begins to
flash, indicating the Blood
Pressure Monitor is deflating the
residual air in the Wrist Holder.
3. Upon the completion of air
deflating, you will hear the buzzer
beep and meanwhile the air relief symbol Ð will disappear.
This completes the initialization process and the unit is ready
for measurement.
Indicator ♥ will appear, and the digit
0 (zero) will show up on the Display.
The system is set to measure mode
by default and the Mode Indicator
MEA is displayed.
4. The unit is in command waiting
status at this moment and other functions can be navigated by
pressing the MODE key. In the measurement mode, in case

- 16 of 35 -
setting of the upper limit of pressurization is not activated by
the user for 3 seconds, the unit will automatically commence
to pressurize. Pressurizing symbol Ï will be displayed and the
air pump will automatically inflate and pressurize the Wrist
Holder and the system enters the measurement mode. The
MODE key and SET key will not respond to operation before
the completion of measurement.
5. Gradually ascending
pressure value can be seen on
the Display Window and you
can feel the Wrist Holder
tightening your wrist gradually.
The air pump will stop
automatically as the pressure reaches the set value, and the
pressurization symbol Ï will disappear, indicating the end of
pressurization. When air pressure detected by the system is
inadequate, the air pump will be started again to continue to
pressurize to reach a higher air pressure level. The air

- 17 of 35 -
pressure levels set as follows, in
mmHg unit:
180, 240 and 290, totaling 3 grades.
6. Upon the completion of
pressurization, the air pressure inside the Wrist Holder will
decrease automatically. A buzzer
accompanied by flashing of the
symbol ♥ on the Display Window
indicates the beginning of the
measurement of your blood pressure
and pulse.
Please wait patiently for the readings.
7. Upon the completion of
measurement, the air relief
symbol Ð will flash, and the residual air inside the Wrist
Holder will be discharged quickly. The buzzer will beep when

- 18 of 35 -
the air pressure inside the Wrist Holder
descends below 5mmHg, and
meanwhile, the symbol ♥ will
appear on the Display Window. And
then, your blood pressure values as
well as pulse will show upon the
screen alternately (blood pressure and
pulse displayed about 3.5 seconds and
2 seconds respectively).
8. Upon the completion of your measurement after you turn
the machine on, the system will not pressurize automatically
in case MEA still displays. Next operation is available by
pressing the SET key with one touch.
9. Just press On/Off button to turn the machine off.
Blood Pressure Commonsense
Blood pressure fluctuates all the time and possesses the
property of periodical change within one day.

- 20 of 35 -
The basic blood pressure is your blood pressure taken when
you are at your calmest, both mentally and physically. Since it
is the basic index used when assessing the current state of
your health, due importance should be accorded.
This basic blood pressure should be measured ideally under
the following conditions:
30 minutes to one hour after waking up (before breakfast)
At room temperature of 15 degrees to 25 degrees Celsius
(60 F to 80 F)
After resting your body for more than 30 minutes
After using the lavatory
When you feel physically and mentally calm
Please try to keep the correct posture while measuring.
Do not be too anxious or too affirmative about your temporary
blood pressure readings. Please record the variations of your
daily blood pressure, and then consult with your doctor. The
measuring results of continuous examinations might be

- 21 of 35 -
dramatically different, and it is caused by changed volume of
blood flow resulted from deformed blood vessels upon
pressure (particularly common for the old because of the poor
flexibility of their blood vessel, their high blood viscosity as
well as low recovery speed).
Generally, your blood pressure value taken in a hospital is
higher 20mmHg-30mmHg than that taken at home. This is
caused by a sudden rise of your blood pressure on account of
your nervousness resulting from your body’s alert reactions to
movement, crowd, surroundings as well as the presence of
medical staff.
Your pulse, just as your blood pressure value, will also change
along with the difference of your body condition as well as
physical movement.

- 22 of 35 -
Save, Inquiry and Clear
The SAVE function can record the high blood pressure
(systolic) value, low blood pressure (diastolic) value, pulse
and measured time upon the completion of measurement by
pressing MODE key to select SAVE function followed by
pressing SET key to confirm SAVE function. The system will
not respond to SAVE function other than in the above situation
even if the MODE key is pressed down to select the SAVE
function followed by pressing SET key to confirm. On the
contrary, the system will automatically transfer to the next
operation mode---QUE mode.
After finishing measurement, press Mode key to choose save
function. With the display of indication SAVE, it will appear
NO.1, which indicates to save the first measurements. You
Saving Measurements

- 23 of 35 -
can continue to press MODE key to save the second (NO.2)
or the third (NO.3) measurement. It can save measured pulse,
systolic, diastolic values and
measured time only after pressing
set key again. The system
automatically chooses saving order
and can save 30 measurements for
each one at most.
Upon the number of sets saved exceeding 30, the system will
erase the oldest values and keep the latest values. Saving
order keeps unchanged.
After switching the unit on, the
symbols 0, ♥ and the measuring mode
indicating SAVE will display on the
Query Function of Blood Pressure Data

- 24 of 35 -
screen. The system is in command waiting status and press
MODE key to select query measurements function.
When the mode indication QUE
displays, NO.1 will appear on the
screen and indicate to query the first
person’s measurements. You can
choose to query the second or the third
person’s measurements by pressing
MODE key continuously. Press SET
key again to confirm.
A set of saved measurement values
numbered 1 will appear on the screen.
(The first screen display is pulse
symbol P, saving NO., pulse value and
measuring time. The second screen
display is systolic and diastolic pressure values. ) Upon
pressing SET key once again, another set of saved values
numbered 2 will appear on the screen and so on. After the last

- 25 of 35 -
numbered 30 set of saved values has thus been displayed,
pressing SET key once again, the system will display saved
values numbered 1 again. Further pressing of SET key will
again display the values at number 1 and so on cyclically.
When the system is in command waiting status, press MODE
key to select clear function. When the mode indication CLR
displays in the LCD, CLr no.1
displays and CLr is flashing,
which indicates to clear the first
person’s measurements. You can
choose to clear the second (NO.2)
or the third (NO.3) person’s
measurements by pressing MODE key continuously. Press
SET key again to confirm and the system will clear all saved
Clear Saved Measurements

- 26 of 35 -
measurements.
Remarks:
1. The saved data will still be kept in the system when the
machine is turned off and then on or after the batteries are
replaced.
2. The system will appear 0 in case no data has been stored
and QUE function is selected.

- 27 of 35 -
Error Display
Display
Cause Treatment
With Noise
Release the Wrist
Holder and Re-wear
it for another measurement.
Your wrist, finger
or body moved
or you talked
while measuring
Keep calm & quiet for 5 minutes
and measure again.
Wrist Holder
worn over shirt
sleeve
Unbutton sleeve and wrap unit
directly over the wrist.
E
Pressure exceedin
300mmHg
Measure again.
Battery Used
Out
Replace the old batteries.

- 28 of 35 -
EE
Upper Limit for
Pressurization
not
High Enough
Set a higher upper limit and
measure again. For e.g. if the
upper limit was earlier set to
180, reset this to the next higher
limit, 240.
Faults & Causes
Fault Treatment
Systolic Blood Pressure Value
too High while Diastolic Blood
Pressure Value too Low
Wrong Wrist Holder Location,
or Wrist Holder incorrectly
fastened, or you talked or
moved your wrist.
Systolic Blood Pressure Value
too Low while Diastolic Blood
Pressure Value too High
Wrist Holder incorrectly
fastened, or shirt sleeve
wrapped by the Wrist Holder.

- 29 of 35 -
Difference in readings when
measured at the doctor’s clinic
People tend to have subtle
psychosomatic changes in
the presence of a doctor, and
identical results are rare.
No display when turning the
machine on by pressing
On/Off button
Battery used out or wrong
installation of the batteries.
The wrist blood pressure measurements are the different with
that of upper arm blood pressure because of different
measuring parts. Normally, the systolic and diastolic blood
pressure difference between wrist and upper arm blood
pressure is within ± 10mmHg for healthy adults.

- 30 of 35 -
Maintenance & Operation Precautions
DO NOT drop or give a
strong shock to the unit.
DO NOT inflate the unit without
strapping it on your wrist.

- 31 of 35 -
DO NOT attempt to
disassemble the unit.
DO NOT store in extremely low(less
than 0°C) or high (more than 60°C)
temperature. Do not expose to
moisture.

- 32 of 35 -
DO NOT bend the Wrist
Holder in the direction
opposite to its shape.
Warranty will be void if any of the above occurs.

- 33 of 35 -
Specifications
Name
SHENGFU Microcomputer Wrist
Blood Pressure Monitor
Model □SF822B
Display Mode
Large Digital Liquid Crystal Display
(LCD)
Measurement Range
Pressure: 0 - 37.3kPa
( 0 - 280mmHg )
Pulse: 40-200 Beats/Min
Accuracy
Pressure: ± 0.4kPa (± 3mmHg )
Pulse: ± 5%
Pressurization Mode Automatic Pressurization
Power Supply Two AAA Size Batteries
Gross Weight Around 275g
Operating
Temperature &
Relative Humidity (RH)
airpressure
Temperature: 10°C-40°C
Relative Humidity (RH)≤80%
Airpressure:86kPa--106kPa
645mmHg—795mmHg

- 34 of 35 -
Maintenance
Temperature &
Relative Humidity (RH)
airpressure
Temperature: 0°C-40°C
Relative Humidity (RH)≤80%
Airpressure:50kPa--106kPa
375mmHg—795mmHg
Packing Dimension
100mm×83.3mm×88mm(L×W×H)
Guarantee
MILLENNIUM blood pressure monitor is guaranteed for
2 years from date of purchase.