Please, read and follow strictly all the instructions in the manual.
Due to continuous improvement in the device, information contained in this
manual is subject to change without notice. Contact the company-producer
for revisions and corrections
4, Narodni Buditeli Str.
8900 Nova Zagora
BULGARIA
Phone/Fax: + 359 457 67082
e-mail: office@lactoscan.com
www.lactoscan.com
www.milkotronic.com
Operation manual 11.04.19 2/110
Ultrasonic milk analyser
SAFETY INSTRUCTIONS
1. Read this manual carefully and make sure that you understand all the
instructions.
2. For safety purposes the device is equipped with grounded power
cable. If there is no grounded electrical outlet where the device will
be used, please, install suc h be f ore using the device.
3. Place the device on leveled and stable plate. In case it falls or is
severely shocked it may be damaged.
4. Connect to the electri cal network in such a wa y that the power cable
to stay away from the side for accessing the device and not to be
stepped on.
5. Every time before cleani ng the device swit ch it off and unplug it from
the electrical outlet. The device has to remain unplugged till the
cleaning completion.
6. Do not disassemble the unit in order to avoid possible electrical
shock. In case of malfunct ion contact your local dealer.
7. Handle the liquids the device works with carefully, following all the
instructions for their preparation.
8. Place the switching adaptor in such a way as to be protected from
overflow and spillage of liquids.
Operation manual 11.04.19 3/110
Ultrasonic milk analyser
№
Description
Item №
pcs
1.
Ultrasonic portable milk analyzer
LSSA001
1
50 sec
30 sec
Operation
manual
LSSA00
2.
Plastic sample holder
LSSA003
1
3.
Spare Pipes
LSSA004
2
4.
12 V DC Power Supply Cable
LSSA005
2
5.
Alkaline cleaning solution Lactodaily
100 g
1
6.
Acidic cleaning solution Lactoweekly
100 g
1
№
Description
not included in the set (may be
Item №
pcs
7.
RS232 Interface Cable - Analyser-IBM PC
LSSA006
8.
Service Pack - CD
LSSA007
9.
pH measuring system
LSSA009
1
10.
pH probe with cable and holder
1
11.
Buffer solution Ph 60 ml
(pH7.00±0.01/20˚C)
LSSA011
1
12.
Buffer solution pH 60 ml
(pH4.00±0.01/20˚C)
LSSA012
1
13.
Milk conductivity measuring system
LSSA013
1
14.
Buffer solution conductivity 50 ml (5.02
(±5%) mS/cm (18±0.1˚C)
LSSA014
1
15.
Real time clock
LSSA015
1
16.
High-fat measuring function
LSSA016
1
17.
ECS POS Serial Printer
LSSA017
1
PARTS AND ACCESSORIES
In the table below the standard delivery configuration of the milk analyzer is
listed:
1 sample measurement time
2
In the table below the milk analyzer spares and accessories, which are
delivered on customers request are listed:
a) included in the set:
b)
additionally bought):
LSSA010
/
Operation manual 11.04.19 4/110
Ultrasonic milk analyser
18.
12 V Serial Printer Power Supply Cable
LSSA018
1
19.
RS232 Interface Cable - Milk Analyser –
Serial Printer
LSSA019
1
20.
Carrying case
LSSA020
1
21.
Plug type
1
1
22.
Spare O-ring for the pH probe
1
23.
USB flash drive (slot)
1
Operation manual 11.04.19 5/110
Ultrasonic milk analyser
1. FUNCTION
The function of the milk a na lyzer is to make quick analysis of mil k on f a t
(FAT), non-fat solids (SNF), proteins, lactose and water content
percentages, temperature (
conductivity as well as density of one and the same sample directly
after milking, at collecting and during processing.
o
С), pH, freezing point, salts, total solids,
Operation manual 11.04.19 6/110
Ultrasonic milk analyser
2. TECHNICAL PARAMETERS
2.1. Working modes characteristics:
The program of the milk analyzer has four working modes.
2.1.1. Measurement mode milk / dairy product – first type
2.1.2. Measurement mode milk / dairy product – second type
2.1.3. Measurement mode UHT milk / dairy product – third type
These modes have been calibrated on customers’ request for 3 milk types
from the following: cow, sheep, UHT, buffalo, goat, camel milk, cream, ice
cream mixtures, whey, recovered milk, etc. before leaving the production
facilities and the text on the display will be for the corresponding types, as is
indicated on page 2 Measurement modes.
2.1.4. Cleaning
2.1.4.1. Current
2.1.4.2. Final
Note:
For devices with a 5 button keypad on the front panel by pressing the extra
buttons, you can run the following commands:
Pressing the button labeled A - start re-printing the results
Pressing the button labeled B - start of flushing (2.1.4.)
These commands are possible when the unit is in Idle mode, i.e. expecting
command for measuring sample.
Operation manual 11.04.19 7/110
Ultrasonic milk analyser
Fat ………………........... .. ........
…………..from 0.01% to 25% (45%*)
SNF ………………………………
...…………………....from 3% to 40%
Density ** ……………………….
……...…….from 1000 to 1150 kg/m3
Proteins ………………………….
………………………..from 2% to 7%
Lactose ………………………...
…………………from 0.01 % to 20 %
Water content …………………
……………...……..from 0 % to 70 %
Temperature of milk ……………
from 1oC to 40oC (if measurement is
30 sec, then t° is from 15 to 40°)
Freezing point ***.………………
…………..from – 0,400 to – 0,700oC
Salts …………………………….
…...…………...………from 0,4 to 4%
PH*..……………………………...
…………………..………from 0 to 14
Conductivity* ** …………………
……………...…from 3 to 14 [mS/cm]
Total Solids* …………………….
………………………...from 0 to 50 %
Fat …..………………………………..
…...………………….…...± 0.06%
SNF …………………………………..
…………...………….…...± 0.15%
Density ………………………………
………………………..± 0.3 kg/m3
Proteins ……………………………...
……...……………….…...± 0.15%
Lactose ………………………………
.…………………………...± 0.20%
Water content ……………………….
.…………………………….± 3.0%
Temperature of milk ………………..
.………………………………± 1oC
Freezing point………………………..
..……...…………………± 0.005oC
Salts ………………………………...
…..…………...…………...± 0.05%
2.2. Measuring range:
* Option, on customers’ request
** Density data are shown in an abbreviated form. For example 27.3 have to
be understood as 1027.3 kg/m
3
. To determine the milk density, write down
the result from the display and add 1000.
Example: result 21,20; density = 1000 + 21,20 = 1021,2 kg/m
3
The abbreviated form of the density is used also when entering data for
samples in working mode Recalibrate, for example:
If the measured sample density is 1034.5 kg/m3, then in the menu for
entering the samples parameters used for calibration, across the parameter
Den = , you have to enter 34.5.
*** pH and conductivity measurements are optional and are embedded in the
device on customers' request.
extent of contamination at normal
environmental conditions
…………………….…………..2
2.4 Correct ambient condit ions:
Accuracy is guaranteed in case of normal ambient conditions:
Accuracy values in point 2.3 are in dependence on the correctness of the
corresponding chemical method, used for component content determination.
In point 2.3. are used the following reference methods: Gerber – for fat,
gravimetric – for SNF, Kjeldahl – for protein. The boundary for maximum
variation of repeatability when the power supply voltage is from +10 to – 15%
from the nominal voltage values (220V) have to be no more than 0.8
accuracy according point 2.3. The analyser is used in conditions free of outer
electrical and magnetic fields (except the magnetic field of the Earth) and
vibrations.
2.5. Dimensions: ……………… ……………..240/220/100 mm, mass 3,0 kg
2.6. Continuous working ti m e : ………………………………………non-stop
2.7 Milk sample volume per one measurement: ………..15 cm3 (= 25 ml)
Operation manual 11.04.19 9/110
Ultrasonic milk analyser
1. pH probe(option)
10. Sample holder
2. printer (option)
11 Input pipe
3. keypad (option)
12. Hanger
4. Display
13. USB f lash drive (slot)
5-9. Buttons
Fig.1 Front panel
Operation manual 11.04.19 10/110
Ultrasonic milk analyser
1. keypad (option)
8. Serial printer interface/RS232
2. Printer (option)
9. Waste solution output pipe orifice
3. pH-probe (option)
10. Cleaning solution input pipe orifice
4. Weigh Scale input (option)
11. Switching adapter
5. AC adaptor input
12. Power switch
6. DC Power Supply output
13. pH-meter input (option)
7. USB (option)
fig. 2 Back panel
Operation manual 11.04.19 11/110
Ultrasonic milk analyser
Fig 3 Principle working schem e
Operation manual 11.04.19 12/110
Ultrasonic milk analyser
Fig. 4 Connecting peripheral de vi ce s
Operation manual 11.04.19 13/110
Ultrasonic milk analyser
Fig. 5 Cables description
90-1801-0008
RS232 Interface Cable - Milk Analyser – Serial Printer/ IBM PC
The connection of the analyser with other devices via RS232 Interface is
made through the coupling DB9 on the back panel of the device. In it can be
plugged only the cable, supplied by the producer with the analyser (or
produced by the customer as per above scheme). The coupling from the
cable, marked on the scheme with „TO LACTOSCAN” must be plugged in the
computer. If due to some reason is used an external cable USB to RS232
Converter (usually when working with a laptop the customers buy such
cables from the market), it must be plugged in the coupling “TO PC” of the
original cable, not directly in the device.
90-1801-0009
DC 12V Power Supply Milk Analyzer Cable
1. GND
2. No connection
3. No connection
4. 12V DC
There’s an option in the analyser – embedded USB interface (USB to
RS232 Serial Converter) for connection with computer type IBM PC. It is
Operation manual 11.04.19 14/110
Ultrasonic milk analyser
intended for connecting computers from laptop type, which have no RS 232
interface (COM Ports – DB9 Connector). This option may be available
together with the standard RS232 – connector DB9, which always exists in
the analyzers. At one and the same time, connection analyzer – computer,
can be established only through one of the couplings – either only DB9, or
only USB, i.e. analyzer could not be connected at one and the same time,
using one of the interfaces with one and using the other interface to another,
second computer. The coupling for this option (Printer Type) is at the back of
the analyzer, next to the standard RS232 DB9 connector. Ot her USB devices
like
printers, keypads etc COULD NOT be connected to this coupling.
USB interface is based on the element MCP2200 of the company
Microchip Technology Inc. – site:
http://www.microchip.com. For establishing
a connection through this interface a driver has to be installed for MCP2200
in the correspondent computer, which will be connected with the analyzer.
Please, follow the procedure, from the Internet site of the producer:
by choosing the suitable for your operation system driver.
After installing the driver, choose the СОМ port, which will be used for
the real communication. For Win XP it is done by executing the following
commands: Start -> Settings -> Control panel -> System -> Hardware ->
Device Manager -> Ports (COM and LPT) – Right Click -> Properties -> Port
Settings -> Advanced -> COM Port Number. Choose Number 1 or if another
number is chosen, for example 3, then, when working with the software tools,
in the field COM Port (upper right corner), you have to set the new number of
the port, in this case 3.
Operation manual 11.04.19 15/110
Ultrasonic milk analyser
3. QUALIFICATION OF RAW MILK, THERMALLY TREATED
MILK, OTHER DAIRY PRODUCTS AND DERIVATIVES
3.1. Taking samples and preparation for analysis
In order to receive reliable results in qualification of milk, dairy products and
derivatives are needed: precise samples taking; correct samples storing (in
need to be preserved); correct preparation before making measurement. The
rules and requirements for this are described in details in Appendix Milk
sampling.
3.2. Making the measurement.
3.2.1. Preparing the a na lyzer for working mode
3.2.1.1. Put the analyzer on the working place, providing good ventilation and
not in the vicinity of heat providing devices or sources. The temperature in the
premises has to be in the boundaries 10-30
3.2.1.2. Check if the power switch is in "0" position and that the outlet voltage
complies with the voltage indicated on the rating plate of the analyzer.
Connect the switching adapter.
3.2.1.3. Switch on the “POWER” button, which starts the identification
procedure. For a short time the display shows the number of the software
versions, for example:
o
C.
Milkanalyzer xxx
LCD vers xx
MA vers yy
MA ser. N. xxxx
where:
Milkanalyzer xxx is the time for measurement.
LCD vers XX is display control software version.
MA vers YY is the motherboard software version.
MA ser. N. xxxx is the serial number – written on the rear panel of the
analyzer.
Operation manual 11.04.19 16/110
Ultrasonic milk analyser
If in the process of exploitation there is a need to ask a question the
company-producer, you have to send the data, written on the display during
the above described initialization procedure.
3.2.1.4.Till the analyzer is prepared for work (at about 5 minutes) the
following message is written on the display: “Getting ready”. Above pointed
time is in dependence of the environmental temperature and increases with
decreasing the temperature.
3.2.1.5. When the device is ready for work the display shows: “Ready to
start”.
The analyzer is ready to make analysis in mode 1 (normally Cow)
3.2.1.6. If you want to pass to another mode press the button Enter and hold
it pressed. The following message appears on the display:
Release button to
start menu
Release the button Enter. The display shows the possible working modes:
Milk selector
Cal1 – Cow
Cal2 – Sheep
Cal3 – UHT
-----------------------------Cleaning
Final cleaning
Using “up”▲and ”down”▼buttons, choose the working mode and press Enter
in order to start it.
3.2.2. Making analysis
To start measurement:
- pour the preliminary prepared sample in the sample holder of the
analyzer;
- put the sample holder in the recess of the analyzer;
- press the button Enter.
Operation manual 11.04.19 17/110
Ultrasonic milk analyser
The analyzer sucks the milk, makes the measurement and returns the milk in
the waste liquid reservoir. During the measurement the temperature of the
sample is shown on the display.
Ignore the results received immediately after switching on the analyzer and
after measuring distilled water. Make a second measurement with new
portion of the same sample.
3.2.3. Displaying the r e s ult s
3.2.3.1. When the measurement is finished, the sample returns in the
sample-holder and the display shows the results. For example:
F= ff.ff - measured FAT in percentage;
S= ss.ss - measured SNF in percentage;
D= dd.dd - measured density in percentage;
P= pp.pp - measured protein in percentage;
L= ll.ll - measured lactose in percentage;
W= ww.ww - measured sample’s added water in percentage;
By pressing the button” Down” ▼ the display shows the second page,
containing the results:
tt.tC - sample’s temperature;
pp.pp - sample’s pH result – if there is a pH probe connected;
-0.fff - measured sample’s freezing point;
0.sss - measured salts values;
aa.aa - measured total solids
Operation manual 11.04.19 18/110
Ultrasonic milk analyser
By pressing the button”Up” ▲ display shows the third page with results:
Page 3 Results:
L=l.ll
Where:
L= ll.ll- measured Lactose in %;
By pressing the buttons “up”▲ and ”down”▼, the operator has the possibility
to pass from one page result to another.
If the device has an embedded option “Conductivity” and "conductivity
measurement" started, the result is shown on the display, showing the basic
results replacing lactose results in the following way:
C=xx.xx
In this case the Lactose result is shown on a new page - Page 3 Results.
xx.xx is the measured milk sample’s conductivity in [mS/cm]. If the results are
outside the limits for this type of sample (see table from the Appendix
Conductivity measurement), the cursor flashes after the letter С, reminding
that the sample is not correct. On the printout it is printed as !!!.
If the conductiity value is outside measuring range (2-14 mS/cm), the
following message appears on the display:
C=OutRg (Out of Range), and on the printout there isn’t any line with
conductivity value.
3.2.3.2. Write down the results in the form. The results remain on the display
till a new measurement is started. If the analyzer is connected to a computer
or a printer, it sends the data to the computer or prints them.
Operation manual 11.04.19 19/110
Ultrasonic milk analyser
Time: xx:xx:xx
Conduct.(mS/sm)…………..xx.xx
Fig. 6 Printing the results
For more information for the printer see 8.2. Working with Printer.
This procedure prevents gathering milk fat residues and milk stone on the
sensor. The milk stone consists of milk solids, calcium, iron, sulphates,
magnesium, etc. All these substances form layer on the pipe and sensor’s
walls, which leads to deviations in the measurement results and blocking up
the piping.
The company-producer recommends usage of the chemicals, supplied with
the analyser – alkaline and acidic (Lactodaily and Lactoweekly). You may
order them separately or together with the analyser. Try to use only these
chemicals for cleaning the analyser.
In case you missed to order these chemicals, the alternative is to use alkaline
and acidic cleaning solutions for dairy equipment by one the companies,
producing such chemi ca l s , as for example:
Analyzer's contamination as a result of the irregular cleaning is the basic
reason for inexactness during measurement. In order to be avoid this, in the
milk analyzers with peristaltic pump is embedded automatic cleaning. For this
purpose the analyzer has to be prepared on the following way:
Fig.7 Additional pipe to the input pipe for
automatic cleaning
94 - additional pipe,
input pipe
Please, pay attention to the additional pipe, connected towards the input pipe.
It has to be long enough to suck all the content in the glass, but to be bent in
order not to impede the pump function.
Operation manual 11.04.19 21/110
Ultrasonic milk analyser
Pay attention to the
ith the
cleaning solution has to be
well dipped in the cleaning
solution while the pipe in the
reservoir with the already
used samples and cleaning
solution MUST NOT be
Pay attention to the level of
the cleaning solution in the
reservoir. Add solution when
used samples and used
The reservoir with the cleaning solution have to be connected towards the
analyser and outgoing pipe towards reservoir for collecting the measured
samples and used cleaning solution, as is shown on the scheme:
Fig.8 Connecting the re s er voi rs towards the milk analyzer
pipes in both reservoirs. The
pipe in the reservoir w
dipped in the liquid.
the level is below 2/3!
1 - cleaning solution input
2 cleaning solution output
Description of the cleaning procedures
There are two ways of cleaning the analyzer embedded in it – current and
end (final) cleaning.
Operation manual 11.04.19 22/110
:
Ultrasonic milk analyser
4.1.1. Current cleaning.
4.1.1.1. Auto matic start
The current cleaning is made with 3% alkaline solution (for working solution
preparation see 4.2.1.), with which the reservoir for the cleaning solution is
filled in.
It is automatically started, without operator's interference after the set time
intevals elapse
1. 55 min. after switching on the power supply of the analyser,
but idle* work;
2. 15 min. after the last measurement of real milk sample.
*Idle Mode is that part of the standard working mode, when the analyser is
not making measurements. There’s embedded in the analyser system for
measurement of the idle time. The idle time is measured starting from the last
action of the operator. In dependence of it (what the operator last did), are
taken decisions regarding the cleaning.
There are 2 options:
Option A: If the analyser:
1. Was only switched on but was not started in measurement mode,
2. Or the last action was cleaning,
3. Or the last action was measuring sample with very low Fat (similar to
water)
Then the automatic cleaning is started after 55 min.
Option B: If the last thing done with the analyser was measurement of normal
milk sample, the automatic cleaning is started after 15 min.
The display shows the following message and a sound signal is emitted:
Auto clean
started!
Put empty glass
After this the cleaning i s started. The display shows:
Operation manual 11.04.19 23/110
Ultrasonic milk analyser
Cow
Cleaning
Please wait
If there is a glass with sample it is completely emptied and then the analyser
automatically fulfills procedure for sucking the cleaning solution and c rinsing
the analyzer's inner system. In order the input metal pipe to be cleaned out
for a short time a cleaning solution is pumped out for a short time in the
already empty glass. It is filled to the middle and then is sucked back and
poured in the reservoir with the used liquids. If there wasn't glass before the
start of the automatic cleaning and if you do not replace it or put an empty
glass then the cleaning solution will be poured in front of the device.
After the cleaning is finished the displays shows the following:
End of cleaning
After 2 seconds the display shows:
Analyzer ready
Then the analyser is ready for normal measurement.
4.1.1.2. Manual start
The current cleaning may be completed by manual start of the menu
Cleaning. It is used before starting the menu Final Clean. It serves for
cleaning the fats from the measuring tract with alkaline cleaning solution. It is
started using the menu for choosing the working mode of the analyzer. After
the measurement is completes, by continuous pressing the button Enter, the
possible analyser operation modes are shown on the display.
Sheep
UHT
Cleaning
Final Clean
Operation manual 11.04.19 24/110
Ultrasonic milk analyser
By choosing Cleaning the current cleaning is started. The display shows the
following:
Auto clean
started!
Put empty glass
In this way the current cleaning is started. The procedure and operator’s
actions are described above.
Do not switch off the device at the end of the working day before the
automatic cleaning procedure is completed. If it is not automatically cleaned
and there is not a possibility to wait starting the automatic cleaning, then start
manually the cleaning procedure with alkaline cleaning solution and if it is
necessary with acidic cleaning solution as it is described below.
4.1.2. End (final) cleaning.
4.1.2.1. Final cleaning.
It is done with 3% acidic cleaning detergent (for working solution preparation
see 4.2.2.). It is done daily. Serves for cleaning the protein deposits from the
measuring system of the analyser, which were not removed during work with
the alkaline cleaning solution and removing milk stone layers.
Do not use chemicals, which are not intended for milking equipment and
vessels in dairy industry. Pay particular attention to the acidic cleaning
solution concentration. The higher concentration may damage the measuring
sensor.
Always before the final acidic cleaning rinse the device with alkaline cleaning
solution by manual starting if it was not already automatically cleaned. It is
necessary to be done in order to remove the milk residues which could react
with the acidic cleaning solution.
Do not forget to rinse with clean water!
Operation manual 11.04.19 25/110
Ultrasonic milk analyser
It is started using the working modes menu. After the measurement is
completed, by pressing and holding pressed the button Enter, the display
shows the possible working modes of the analyser.
Cow
Sheep
UHT
Cleaning
Final Clean
1. Choose Final Clean. After choosing the final cleaning, the following
message appears on the display:
Put filled with
Water glass
and press Enter
to Continue
2. Put a glass filled with water.
3. The hose from the Waste pipe has to be placed in the vessel for waste
liquids.
4. Press the button Enter in order to continue the procedure. Then the
following message appears on the display:
Cleaning
Please wait
Do not miss to clean with water because the residues from the alkaline
cleaning solution could react with the acidic solution, which will lead to gas
and sediment formation.
After rinsing with water
, the display shows the following:
Operation manual 11.04.19 26/110
Ultrasonic milk analyser
Cleaning: X
Glass Alcal Detrg
Waste pipe
Glass
press Enter
5. Place glass filled with alkaline cleaning solution.
6. The pipe, coming from the orifice with inscription Waste Pipe has to be
placed in the glass.
Fig 9. Cleaning
7. Pressing button Enter starts the procedure for cleaning with alkaline
solution.
The following appears on the display:
Operation manual 11.04.19 27/110
Ultrasonic milk analyser
Glass Acid Detrg
End of Cleaning
Cleaning: X
Where, X is changing number showing number of cycles done by the pump.
8. After the cleaning procedure is finished, the following message appears,
reminding the analyzer needs rinsing.
9. Rinsing procedure is executed, described in p. 2. After it is finished, the
display shows:
Waste pipe Glass
press Enter
10. Place glass filled with acidic cleaning solution.
11. The pipe, coming from the orifice with inscriptionс Waste Pipe has to be
placed in the glass.
12. Pressing button Enter starts the procedure for cleaning with acidic
solution.
The following appears on the display:
Where, X is changing number showing number of cycles done by the pump.
13. After the c leaning procedure is finished, the following message appears,
reminding the analyzer needs rinsing.
14. Rinsing procedure is executed, described in p. 2. After it is finished, the
display shows:
Ready to Start
15. Procedure Final Clean is finished.
3. Pour warm alkaline solution in the glass and press the button Enter, in
order to continue the cleaning.
Then the following message appears on the display:
Operation manual 11.04.19 28/110
Ultrasonic milk analyser
Cleaning
Please wait
Follow repeatedly sucking/pouring out the cleaning solution and heating it up.
After the cleaning is finished, the cleaning solution is emitted in the reservoir
with the waste liquids.
For maximum cleaning effect it is recommended the cleaning solution to be
preliminary heated up to 40-50 degrees centigrade.
Follows rinsing with water:
4. Put a glass with water for rinsing the device
.
Put filled with
Water glass
and press Enter
to Continue
After rinsing the display shows the following:
End of cleaning
After 2 seconds the following appears on the display:
Analyser ready
After this the analyser is ready for normal measurement or to be switched off
from the power supply.
Operation manual 11.04.19 29/110
Ultrasonic milk analyser
4.2. Preparation of cleaning s olution
For a single cleaning cycle you need only 25 ml cleaning solution. We
recommend you to prepare working solutions of cleaning chemicals, enough
for normal work for 1 week, because, during their stay unused, the working
solutions loose their strength and also is difficult to store them.
Preparation of 3 % alkaline solution of Lactodaily for circulation cleaning in
the milk analyser:
1. Take the package concentrated chemical Lactodaily
2. In appropriate vessel (for example bucket) pour 1 l water.
3. Add the chemical and then again water up to 3 l.
Then follow the instruction for milk analyser cleaning.
4.2.2. Preparation of acidic cleaning solution
Preparation of 3 % acidic solution of Lactoweekly for circulation cleaning in
the milk analyser:
1. Take the package concentrated chemical Lactoweekly
2. In appropriate vessel (for example bucket) pour 1 l water.
3. Add the chemical and then again water up to 3 l.
Then follow the instruction for milk analyser cleaning.
Fig. 10 Labels for acidic cleaning chemical Lactoweekly and alkaline
cleaning chemical Lactodaily
The following procedure is executed:
1. Rinsing the milk residues:
Fill in the glass with water. Put it in the recess of the analyser and start
command Cleaning from the main menu. After finishing it pour out the
contaminated water.
Operation manual 11.04.19 30/110
Loading...
+ 80 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.