Miele E 901/2 Operating instructions

0 (0)
Miele E 901/2 Operating instructions

Mobile unit E 901/2 for surgical instruments

Items supplied:

Mobile unit E 901/2 with four levels, H 590, W 640, D 790 mm,

fitted with a magnetic strip for automatic mobile unit recognition and spray arm magnets for spray arm monitoring (PG 8527/28 only).

Supplied as separate items:

2 x E 447 Luer lock adapters for Luer locks

2 x E 449 Luer lock adapters for Luer locks

3 x E 362 blanking screws

3 x E 980 angle adapters

4 x Ø 4.0 mm injector jets, 110 mm long

3 x Ø 4.0 mm injector jets with spring clips, 120 mm long

Optional accessories

E 142 DIN mesh tray

E 476 holders for micro instruments

E 448 hose connectors for Luer lock

Insert E 473/1 for small instruments

Half insert E 492 for kidney dishes

Please read the Warning and Safety instructions in the operating Instructions for your G 7827/28 or PG 8527/28 Washer-disinfector carefully. You should only process instruments/medical products in this mobile unit if they have been declared as suitable for machine processing by the manufacturer. Please also observe the manufacturer's care instructions. The programme selected must be suitable for the materials from which the items to be processed are made. Where possible the programme should include thermal disinfection. The final rinse should, if possible, always be carried out with fully demineralised water.

If using automatic mobile unit recognition, please make sure that the coding is correctly set.

Areas of application:

DIN or ISO mesh trays containing surgical instruments can be processed on the lower three levels.

Up to 8 hollow instruments/suction tubes can be processed in addition to the mesh trays.

Loading dimensions

[H x W x D in mm]

(Levels are numbered from the bottom upwards)

 

 

 

Level 1

80 x 593 x 780

 

 

Level 2

80 x 593 x 780

 

 

Level 3

80 x 593 x 780

 

 

Before loading the machine and before starting a programme, please check:

that the spray arms rotate freely,

that the spray arm jets are clean and free of any blockages,

that the mobile unit is correctly fitted into the cabinet and that it has connected correctly with the water intake.

For validation purposes please make sure that you follow the loading instructions given on the template.

To ensure a suitable standardised spray pressure for all connections, all screw connectors must be fitted with jets, adapters, sleeves or blanking screws.

Damaged jets, adapters or sleeves etc. must not be used.

Empty jets, etc. do not have to be replaced with blanking screws.

Level 4

 

210 x 520 x 780

 

 

 

 

 

 

Hinged instruments must be opened up before being placed in the mesh trays to minimise the risk of surfaces not being accessed during processing.

To prevent the risk of light instruments such as bulldog clips and other small items moving about during processing, closable inserts (e.g. E 473) and silicone holders (e.g. E 476) can be used to keep them secure.

Containers e.g. beakers and kidney dishes, should be placed in the mobile unit with the open side facing downwards, ensuring that all surfaces can be accessed easily and that cleaning solution can flow freely off them. They must also be secure so that they cannot be flipped over by the force of water during processing.

If hollow containers are placed such that cleaning solution can collect in recesses etc., they will not get properly clean and rinsing results will be impaired.

Miele Company Ltd., Fairacres, Marcham Road, Abingdon, Oxon, OX14 1TW, Tel 0330 160 6600, Internet: http://www.miele.co.uk / 0297 Alteration rights reserved / Publication date 2015-11-01

If necessary, a cover net or suitable insert, e. g. an E 492 for kidney dishes, can be used to secure these containers.

M.-Nr. 10 335 950 / 00 en