NOTE - It is highly recommended that a three
valve manifold be installed in the
instrumentation loop to minimize
possible damage during pressurization.
1. CONNECTIONS
Model 130 P is provided with 1/4" compression
tube fittings, one low and one high, on the top
and bottom of gauge body.
Model 130 A, B, H, S, is provided with dual 1/4"
FNPT connections top and bottom as standard.
One pair of high and low pressure ports is for
process connections. The other pair is plugged or
used as drain/bleed connections.
Ports are identified for high and low pressure. Be
sure to plumb to proper connections on system.
Improper connection will not damage instrument
but it will not function properly.
Flexible tubing is recommended to minimize effect
of any vibration that may exist.
When attaching connections to the 130 P, care
must be taken to prev ent o v er-tightening which ma y
result in stripping threads and/or cracking ports. In
order to prevent this, use a suitable wrench to hold
the factory supplied fitting in place while tightening,
sometimes called “double wrenching”.
2. INSTRUMENT LOCATION
On liquid service, the instrument should be
mounted below the process connections to
facilitate self-bleeding.On gas service, it should be
located above the process connections to promote
self-draining. If the process contains particulates, a
“pigtail” loop or drop leg (manometer “U-tube”
configuration) in the tubing will minimize the
possibility of it migrating into the instrument.
NOTE - On liquid applications, unequal liquid
heads on high or low side will result in
an inaccurate differential pressure
indication.
3. PANEL MOUNTING
The Model 130 is designed for mounting through
the front of the instrument panel and is provided
with a panel mount kit, consisting of (4) panel
mounting studs/nuts.
Make cutouts as indicated in (Fig. 1). Insert (4) panel
mounting studs, finger tight, into metal inserts located
in the rear of the bezel.
Insert gauge through the panel, aligning panel
mounting studs with holes in the panel. Install #832 nuts onto studs and tighten securely.
4. PIPE MOUNTING
If specified, your Model 130 will have a pipe mount
kit installed. This provides for mounting on a 2"
vertical or horizontal pipe. See (Fig. 1).
5. ELECTRICAL - See Mid-West Bulletin
IMELEC130/latest
TR
OUBLE SHOOTING
1. Gauge does not indicate differential.
A. Check for proper hook up, high to “hi” and
low to “lo”.
B. Make certain block valves are open and, if
using a 3-valve manifold, that the equalizer
(balance) valve is closed.
C. If A & B check out, loosen high pressure line
to determine if there is pressure to
instrument.
D. If there is pressure to instrument, check to
determine that there is differential across the
unit being monitored.If so, contact factory for
assistance and/or an “RGA” (return goods
authorization) number to return instrument
for repair or replacement.