Safety First: The primary concern of Midmark Corporation is that this sterilizer is maintained with the
safety of the patient and staff in mind. To assure that
services and repairs are completed safely and correctly,
proceed as follows:
(1) Read this entire manual before performing any
services or repairs on this sterilizer.
(2) Be sure you understand the instructions con-
tained in this manual before attempting to service or repair this sterilizer.
Safety Alert Symbols
Throughout this manual are safety alert symbols that
call attention to particular procedures. These items are
used as follows:
DANGER
A DANGER is used for an imminently
hazardous operating procedure, practice, or condition which, if not correctly followed,
will result in loss of life or serious personal
injury.
NOTE
A NOTE is used to amplify an operating procedure,
practice or condition.
Warranty Instructions
Refer to the Midmark “Limited Warranty” printed in the
Installation and Operation Manual for warranty information. Failure to follow the guidelines listed below will
void the warranty and/or render the M4•9 Easyclave™
Sterilizer unsafe for operation.
•In the event of a malfunction, do not attempt to use
the sterilizer until necessary repairs have been
made.
•Do not attempt to disassemble sterilizer, replace
malfunctioning or damaged components, or perform adjustments unless you are one of Midmark’s
authorized service technicians.
•Do not substitute parts of another manufacturer
when replacing inoperative or damaged components. Use only Midmark replacement parts.
WARNING
A WARNING is used for a potentially
hazardous operating procedure, practice, or condition which, if not correctly followed,
could result in loss of life or serious personal
injury.
CAUTION
A CAUTION is used for a potentially haz-
ardous operating procedure, practice, or
condition which, if not correctly followed, could result
in minor or moderate injury. It may also be used to
alert against unsafe practices.
EQUIPMENT ALERT
An EQUIPMENT ALERT is used for an
imminently or potentially hazardous operating procedure, practice, or condition which, if not
correctly followed, will or could result in serious, moderate, or minor damage to unit.
The M 4•9 EasyClave™ Sterilizer is to be used in medical and dental offices, hospitals, clinics, nursing homes,
laboratories, and other facilities to sterilize heat and
moisture stable reusable items (including dental handpieces) that are compatible with steam sterilization at
270°F (132°C). Follow state and federal guidelines for
sterilization of single use devices.
The major serviceable components of the sterilizer
are (refer to Figure 1-1, Major Components ).
Door Pulse Solenoid (1)
Door Pulse Solenoid is held in closed (latched) position
during sterilizer operation and automatically unlatches
after vent mode. Controlled by main control p.c. board.
Door Latch Interlock Switch (2)
Monitors position of door assembly. A normally open
switch that closes when door is closed. Prevents Initiating a Cycle unless door is closed.
Axial Flow Fan (3)
Thermostatically controlled fan for circulating air, in area
between chamber and inside of panels, to decrease
heat and condensation.
Fan Thermostat (4)
Controls Off / On of Axial Flow Fan. Normally open,
bimetal, auto-reset, thermostat that automatically
closes contacts at 122°F (50°C) to 138°F (59°C) to
power Fan. Thermostat opens at 95°F (35°C) to 106°F
(41°C), removing power, stopping Fan.
Air Solenoid Valve (5)
Removes air from chamber during operation. Controlled by main control p.c. board.
Steam (Side) Temperature Probe (6)
Temperature probe inside chamber monitors internal
chamber temperature during operation and transmits it
back to main p.c. board. Steam Probe is located in
lower front right side of chamber.
Auxiliary (Lower, Trough) Temperature Probe (7)
Located inside chamber in water trough. Senses temperature near lower heating element and transmits it
back to main control p.c. board.
Contains the program logic, EPROM, and various connections to send and receive data to operate the various
programs. SW2 dip switch assembly is located on top,
right corner of board. Service Diagnostics, calibrations,
and various display options can be accessed by changing switch positions.
Fuses F1 and F2 protect the main control p.c. board
should a problem exist in the transformer or related
components.
Fuse size is 1.6 Amp, 250 VAC, Slowblo, 3/16" D. x
13/16" L.
Auxiliary (Upper) Hi-Limit Switch (9)
Controls power to heating elements should overheat
conditions exist.
Normally closed, bimetal, thermostat that automatically
opens contacts at 450°F (232°C) to 500°F (260°C)
removing power to heating elements.
Automatically resets to close position at 370°F (188°C)
to 420°F (216°C).
Located on upper
Lower Hi-Limit Switch (10)
Controls power to heating elements should overheat
conditions exist.
Normally closed, bimetal, thermostat that automatically
opens contacts at 450°F (232°C) to 500°F (260°C)
removing power to heating elements.
Automatically resets to close position at 370°F (188°C)
to 420°F (216°C).
Located on lower
Transform e r(11)
Steps line voltage down to 18.5 VAC to power the main
control p.c. board.
Can be used with 115 VAC or 230 VAC supply voltage.
Main Fuse (12)
Protects sterilizer electrical circuit. 115 VAC Sterilizer
fuse is a 15 Amp, 600 VAC, 3AB Fast Acting, 13/32” D.
x 1 1/2" L.
The 230 VAC Sterilizer fuse is the same style but rated
at 8 amp.
Solid State Relay (13)
Solid state switch that closes when 12 VDC is supplied
from the main control P.C. Board to relay terminals 3
and 4.
With the relay operating, line current can pass thru (terminals 1 and 2) to common terminal of Main Relay.
Depending if Main Relay is operated, current will be
supplied to both heaters (auxiliary heat) or just the lower
(steam) heater.
Solid State Relay opens and closes, turning on and off
the heater(s), to control temperature in chamber.
front, outside surface of chamber.
front, outside surface of chamber.
SECTION I
Return To Table Of Contents
GENERAL INFORMATION
Main Relay (14)
Main Relay contains a set of Normally Open (NO) and
Normally Closed (NC) contacts that allow one (Steam,
lower) or both (Auxiliary, upper and lower) heaters to
operate depending if its coil is being supplied with 12
VDC from main control P.C. Board.
When coil is energized
(12 VDC) the contacts change
position, supplying line voltage thru the SSR and the
NO contacts to the Steam (lower) heater.
closed
If the coil is not energized
, the contacts remain in their
normally closed positions. Line voltage is now supplied
thru the SSR and the NC contacts to both the Upper
and Lower heaters.
Water Level Sensor (15)
Located in back of chamber, it monitors the water level
during the FILL cycle. When the water level reaches the
sensor, a circuit is completed which signals the main
control P.C. board to remove power from the Fill Valve
solenoid, closing the valve.
Pressure Transducer (16)
Monitors pressure inside chamber and transmits signal
back to main control p.c. board so conditions can be
controlled inside chamber. If conditions are not within
tolerances, specific Codes will be displayed to inform
Operator.
Fill Valve (17)
Fill Valve is a normally closed valve. When START key
is depressed K3 contacts on main control P.C. board
close supplying line voltage to the Fill Valve solenoid.
Normally closed valve opens, allowing water to enter
chamber. When water level reaches sensor a circuit is
complete. P.C. Board opens K3 contacts, removing
power from Fill Valve solenoid coil, closing the valve.
Vent Valve (18)
Vent Valve is a Normally Open
valve.Controlled by main
control P.C. board. When K2 contacts close Vent Valve
solenoid energizes, closing
the valve. Valve vents back
thru condensing coil in reservoir.
Lower (Steam) Heating Element (19)
Submersible, tubular, 1550 watt, heating element used
to generate steam for unit.
During the Heatup and Sterilization cycles, the Steam
temperature probe, monitors internal chamber conditions, sending signals back to main control P.C. board
which controls On / Off cycling of heater thru the Solid
State Relay (SSR).
Upper Heating Element (20)
Upper element is wired in series with lower element to
maintain conditions inside chamber during DRY
With the Main Relay not
plied thru the SSR and NC contacts of the Main Relay
to both
Lower and Upper temperature probes monitor internal
chamber conditions sending signal back to main P.C.
board which controls On / Off cycling of both heaters by
opening and closing the SSR.
Pressure Relief Valve (21)
Safety valve vents chamber should pressure inside
chamber reach 40 PSIG (275.8 kPa). Steam is vented
out below base of sterilizer.
Reservoir Tank w/ Condensing Coil (22)
Retains supply of distilled water for operation. During
Vent cycle steam is discharged from chamber back thru
condensing coil into reservoir.
Vent Muffler (23)
Located inside reservoir at end of Air Valve line. Used to
supress exhaust noise during operation. Periodically
inspect muffler for restriction to assure proper operation.
Filter Screen (24)
Fine stainless steel mesh, screen for filtering particals
from system preventing migration to Vent Valve. Filter
should be cleaned monthly.
Internal Filter Screen (25)
Fine stainless steel, mesh, tubular screen for filtering
particals from system preventing migration to Air Valve
and Pressure Transducer. Filter should be cleaned
monthly.
Pressure Vessel (26)
Includes chamber and door assembly which are ASME
certified.
Door Interlock Plunger (27)
Plunger device, installed in center of door assembly that
prevents unlatching and opening door. Operated by
internal chamber pressure. Locks at 2 PSIG (13.8 kPa).
Door Gasket (28)
Seals door when in latched position to prevent pressure
from escaping around door during operation.
Display / P.C. Board (29)
Operation and Service Diagnostic messages are
viewed though use of a liquid crystal display (LCD) on
front of top cover.
heaters.
energized, line voltage is sup-
1.2Specifications
Factual data for the M 4•9 EasyClave™ Steam Sterilizer
is provided in Table 1-2.
Table 1-2. Specifications
DescriptionData
Dimensions (overall):
Width.................................................15.3 in. (38.9 cm)
Depth ................................................21.2 in. (53.8 cm)
Height (w/o printer) ...........................15.9 in. (40.4 cm)
Height (w/ printer ..............................16.3 in. (41.4 cm)
Chamber Size: ..........4.5 in. “H” x 9 in. “W” x 12 in. “D”
(11.4 cm x 22.9 cm x 30.5 cm)
Shipping Carton: 24.25 in. "L" x 20.5 in. "W" x 21 in. "H"
(61.6 cm x 52.1 cm x 53.3 cm)
Weight:
With Reservoir Empty .......................82.0 lbs (37.2 kg)
Shipping weight w/ skid ..................115.0 lbs (52.2 kg)
Water Reservoir Capacity ....................Apprx. 1.0 gallon
(3.8 Liters) to full mark
Electrical Requirements:
115 VAC Unit ................................. 115 VAC, 50/60 HZ
A separate (dedicated) circuit is recommended for this
sterilizer. Sterilizer should not
trical circuit with other appliances or equipment unless
circuit is rated for the additional load.
In order to effectively diagnose a malfunction of sterilizer, it may be necessary to perform an operational test
as follows:
WARNING
Refer to Operator Manual for complete
instructions on operating sterilizer.
Failure to do so could result in personal injury.
NOTE
Operational Test, for most part, only describes what
should happen when sterilizer is operated. If sterilizer does something other than described, a problem
has been discovered. Refer to Troubleshooting
Guide to determine cause of problem and its correction.
WARNING
Use 220 - 240 VAC, 50 / 60 HZ alternat-
ing current only for 230 VAC models
and 110 - 120 VAC, 50 / 60 HZ alternating current
only for 115 VAC models. Failure to do so could
result in electrical shock to personnel and will result in damage to sterilizer.
Do not use this sterilizer in an explosive or oxygen-enriched atmosphere. Failure to do so could
result in serious personal injury or death.
(1) Plug sterilizer power cord into a grounded, non-
isolated, correctly polarized outlet, that has
proper voltage for sterilizer.
NOTE
Grounding reliability can only be achieved if unit is
connected to a matching three-pronged, grounded,
non-isolated, correctly polarized receptacle.
(2) Electrical rating for 115 VAC Sterilizer:
115 VAC, 50 / 60 Hz, 12 amps
Connect to a dedicated branch circuit rated for
115 VAC, 50 / 60 Hz, 15 amps minimum.
Electrical rating for 230 VAC Sterilizer:
230 VAC, 50 / 60 Hz, 6.5 amps
Connect to a dedicated branch circuit rated for
220 VAC, 50 / 60 Hz, 10 amps.
Check label on back panel to verify electrical
specifications required for unit.
SELECT
CYCLE
U
nw
rapped
W
rapped
Figure 2-1. Display and Operator Control Pad
(3) Table 2-1, Operational Chart, describes a com-
plete cycle during Normal
operation, identifying
audible signals, various display prompts and
component operation.
Note:
At Apprx. 271°F (133°C) and
28 Psi (193 kPa) countdown
begins.
Audible beeps (Occurs 5 seconds
before Door opens.)
Audible beepsDRY CYCLE COMPLETE
STERILIZING
: XX XXX°F XX . X PSI
XX
(time) (temp.) (pressure)
READY TO VENT
. XX XXX °F XX . X PSI
XX
(time) (temp.) (pressure)
VENTING CHAMBER
°F XX.X PSI
XXX
(temp.) (pressure)
DOOR TO OPEN
°F XX . X PSI
XXX
(temp.) (pressure)
ITEMS STERILIZED
DRY CYCLE XX:XX
(time counts down from 20 to 40
min.)
SELECT CYCLE
CHAMBER PREHEATED
or greater.
Display shows total amount of cycles that have been run.
(VER is the software version number).
*If a cycle has previously been run display will also indicate CHAMBER PREHEATED.
NOTE: Pressing WRAPPED button, display time for sterilization cycle will be 9 MINUTES.
Press START Key
valve. Water from reservoir enters chamber.
opening
K2 (Vent) contacts close
open valve.
At same time, K4 contacts close
mally closed valve. Main Relay coil energizes, changing the positions of the contacts. Normally open contacts close
Solid State Relay (SSR) is not energized, contacts are open
operate.
Water reaches level of Water Level Sensor sending signal back to main P.C. board.
K3 contacts on Main Control P.C. Board open
Fill Valve closes
Solid State Relay (SSR) energizes closing
With Main Relay, N/C contacts opened
Steam Heater. Air Valve, being open
temperature reaches 210°F (99°C) K4 contacts open
pressure to build up. Air Valve will
may
also open at 10, 14, & 18 Psi ( 69, 97, and 124 kPa).
Sterilization Time (3 1/2 min. Unwrapped or 9 min. Wrapped) counts down.
RTD Temperature probes and Pressure Transducer monitor temperature and pressure.
K4 contacts and Air Valve may open and close
off by Solid State Relay (SSR) to maintain temp. / pressure.
10 seconds before end of STERILIZING cycle “READY TO VENT” is displayed.
Main Control P.C. board K2 relay contacts Open
solenoid, returning Vent Valve to normally open
Chamber vents.
Main Control P.C. board K5 relay contacts close
just 1/2 second
Door opens
Main Relay is OFF
SSR cycles
and Off to maintain temp. in chamber during dr ying cycle.
Drying Cycle time is varible between 20 and 40 Minutes.
Open door completely and remove contents. Unit is ready for next sterilization cycle.
For 90 minutes, if no other cycle is initiated, SSR will open and close. With Main
Relay Off
and Steam (Lower) heaters to maintain preheat temperature at 212°F (100°C). After
90 minutes, if no cycle has been initiated, power is removed.
to partially open position.
open & closed turning Auxiliary (Upper) and Steam (Lower) heaters On
, current flows thru Main Relay N/C contacts energizing Auxiliary (Upper)
run if the temperature inside cabinet is 130°F (+/- 8°F) (54°C).
on Main Control P.C. Board, Fill Valve solenoid coil energizes,
, energizing Vent Valve solenoid coil, closing the normally
, energizing Air Valve solenoid coil, opening nor-
and normally closed contacts open.
and heaters do not
,
stopping water flow into chamber.
its normally open contacts.
and N/O contacts closed, power is applied to
, allows air to be expelled from chamber. When
and Air Valve closes allowing
open at 6 Psi (41 kPa) and 22 Psi (152 kPa) and
while steam heater is cycled on and
, removing power from Vent Valve
position.
. Door Pulse solenoid energizes for
, releasing door latch.
, contacts are in their normally open and normally closed position.
Always complete Service Diagnostics
(Section 2.2) before replacing any major
components especially the Main P.C. Board.
Service Diagnostics allows testing various components
of sterilizers through use of operator buttons, display
unit and LED’s on Main P.C. board. It will be necessary
to remove the panels of the sterilizer in order to view the
various LED’s.
WARNING
Unplug unit from electrical outlet to prevent possibility of personal injury or
death from electrical shock.
(1) Unplug unit from electrical outlet.
(2) Remove panels (Refer to para 4-2).
(3) Place SW2 Switch 1 (Figure 2-3) in ON
position.
(left)
WARNING
Use extreme caution when unit is
plugged into outlet to prevent possibil-
ity of personal injury from electrical shock.
(4) Plug unit into outlet.
(5) Use Table 2-2 to perform the Service Diagnos-
tics.
NOTE
Refer to Section 5 for Electrical Schematics / Wiring
Diagrams.
(6) Unplug unit from electrical outlet.
(7) Place dip switch 1 ( Figure 2-3) located on SW2
on Main P.C. board in OFF
normal operation.
(right) position for
NOTE
SW2, Dip Switch Assembly is located on upper, right
corner of Main P.C. Board.
Audible beeps whenever
START or STOP buttons are
depressed.
Valve will emit a “clicking”
sound when operated.
Audible beeps whenever
START or STOP buttons are
depressed.
Audible beeps whenever
START or STOP buttons are
depressed.
Audible beeps whenever
START or STOP buttons are
depressed.
TESTING AND TROUBLESHOOTING
Table 2-2. Service Diagnostics (Refer to Schematic Fig. 2-2).
FILL VALVE TEST
(START) TO TOGGLE
(STOP) FOR NEXT TEST
DOOR OPEN TEST
(START) TO OPERATE
(STOP) FOR NEXT TEST
MAIN RELAY TEST
(START) TO TOGGLE
(STOP) FOR NEXT TEST
STEAM HEATER TEST
(START) TO TOGGLE
(STOP) FOR NEXT TEST
(Checks Lower
Note: Amperage draw when
Steam heater is powered is
approximately 12 Amps.
Heater)
WARNING: Pressing START button energizes (Opens) FILL valve allowing
water to flow into chamber. Water Level Probe does not
test. Water can overflow
again or STOP button is pressed terminating test.
This function checks operation of Fill Valve and related L1 Fuse, and K3 relay contacts on P.C. board.
Press START
K3 relay contacts close and K3 LED illuminates.
Power is applied to solenoid of FILL valve, opening
ber.
Press START
K3 relay opens and K3 LED goes out.
Power is removed from Fill Valve solenoid returning valve to Normally Closed position.
Press STOP
WAR NING: Door will immediately go to
partially open position from closed position
when START button is depressed.
Checks operation of Door Pulse Solenoid and related L1 Fuse, and K5 relay contacts on P.C. board.
Press START
K5 relay contacts close and K5 LED illuminates .
Power is briefly ( 1/2 second ) applied to Door Pulse Solenoid, door springs to partially open position.
K5 relay opens, removing power from K5 LED and Door Pulse Solenoid.
Press STOP
Checks operation of Main Relay coil, N/O, N/C contacts and L1 Fuse.
Press START
Power is supplied thru P.C. board thru Plug J7 to coil of Main Relay energizing it.
Normally Open and Normally Closed contacts change condition, Open contacts
close and Close contacts open.
MAIN LED on P.C. board at Plug J7 illuminates.
Press START
Press STOP
WARNING: There is a 15 second maximum timing period that the
heater will be energized before it shuts off. To restart press START.
Do not restart cycle more than twice before allowing unit to cool for
apprx. 5 minutes.
Checks operation of Steam (Lower) Heater and related L1 Fuse, Upper and Lower
Thermostat O/L’s, Solid State Relay (SSR), and Main Relay.
Press START
12 VDC Power is supplied thru P.C. board thru Plug J7 to coil of Main Relay, energizing it, N/O contacts close.
12 VDC is also supplied thru Plug J7 to Solid State Relay (SSR) (3 & 4) closing it.
Line power is supplied thru the L1 fuse, N/C upper and lower thermostat O/L’s thru
the SSR closed contacts (1 & 2) to the Common terminal of Main Relay.
With Main Relay N/O contacts closed
Main Relay and SSR Steam Relay LED’s on P.C. board at Plug J7 illuminate.
SSR Steam LED on P.C. board at Plug J7 illuminates.
Heater remains energized for a 15 second time period before it automatically
off.
To restart test depress START
Press STOP
once.
again.
to go to DOOR OPEN test.
.
to go to MAIN RELAY test.
.
again to change back to normal unenergized condition
Both heaters are powered is
approximately 6.5 Amps.
Return To Table Of Contents
Audible beeps whenever
START or STOP buttons are
depressed.
TESTING AND TROUBLESHOOTING
Table 2-2. Service Diagnostics (Refer to Schematic Fig. 2-2).
AUXILIARY HEATER TEST
(START) TO TOGGLE
(STOP) FOR NEXT TEST
(Checks Both
STEAM AUX PRESS
XX.X F XX.X F XX.X P
END DIAGNOSTIC TEST
UNPLUG/RE-PLUG UNIT
Heaters)
WAR NING: There is a 15 second maximum timing period that the
heater will be energized before it shuts off. To restart press START.
Do not restart cycle more than twice before allowing unit to cool for
apprx. 5 minutes.
Checks operation of both Auxiliary and Steam Heaters, L1 Fuse, Upper and Lower
Thermostat O/L’s, Solid State Relay (SSR), and Main Relay.
Press START
12 VDC is supplied thru Plug J7 to Solid State Relay (SSR) (3 & 4) closing it.
Line power is supplied thru L1 fuse and N/C Upper and Lower Thermostat O/L’s thru
the SSR closed contacts to the Common terminal of Main Relay.
With Main Relay N/C contacts closed current flows thru the Relay contacts to the
Auxiliary (Upper) Heater and then to the Steam (Lower) Heater.
SSR Surface LED on P.C. board at Plug J7 illuminates.
Heaters remain energized for 15 seconds before automatically
To restart test depress START
Press STOP
DIAGNOSTIC TEST.
Checks Steam (Side) and Auxiliary (Trough) Temperature Sensors and Pressure.
Displays temperature in °F or °C depending on setting of SW2, switch #7.
Should the particular sensor be open, shorted, or have a disconnected lead, display
will indicate a 32°F or 0.0°C reading.
Sensor plugs are located on P.C. Board at J3 for Auxiliary (Trough) Sensor and J5 for
Steam (Side) Sensor.
NOTE: When Service Diagnostics has been completed:
Unplug unit
Position SW2, Switch #1 to OFF
Install covers
Plug in unit.
.
.
to go to TEMPERATURE / PRESSURE DISPLAY & END
position
SECTION II
turning off.
2.3Steam and Pressure Calibration.
When replacing a Steam temperature sensor (RTD),
pressure transducer or main P.C. board it will be necessary to assure the components and system are calibrated. Use the following steps and tables to calibrate
the system. Table 2-3 is for Temperature Calibration and
Table 2-4 is for Pressure Calibration.
NOTE
When a Lower Temperature Probe is replaced, calibration is not required.
When changing a P.C. Board:
You must have the Pressure Span
Steam Offset
replaced. This information can be obtained two ways:
(1) Documentation is supplied with sterilizer
(2) Use SW2 dipswitches #2 and #3 to recover
a.) Dipswitch #2: Pressure Span and Offset.
b.) Dipswitch #3: Steam Offset.
(3) After P.C. Board has been replaced re-enter old
IMPORTANT: Place SW2 dip switch #2 back to OFF position.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
2.5Troubleshooting Procedures
Table 2-5 is a Troubleshooting Guide which is used to
help to determine the cause of a malfunction.
Refer to Section V Schematics and Diagrams for tracing
circuits and referencing points in Troubleshooting
Guide.
Error Categories (Severity)
There are 4 categories of errors in the M4•9 software:
Automatic Errors:
Minor problem causing Operator to do something such
as close the door, etc. Once cleared the software continues on where it left off.
Fast Restart Errors:
• Possible problems with M4•9 control system or with
producing an effective sterilization cycle.
•May be due to something the Operator did such as
pressing the STOP button during a sterilization mode.
•Could also be due to temperature values being outside
the normal range required as sensed by the Lower
Temperature probe.
• Operator is prompted by display to press STOP
button to Restart
Software will abandon the sterilization cycle and return
to SELECT CYCLE / PREHEAT MODE.
.
Fatal Errors
• This would indicate possible serious problems
M4•9 control system.
• Pressing the STOP button before 60 seconds
(when it turns off automatically) will silence the audible
signal but the Power must be disconnected
error code display.
Unit should not be ran until a trained Service Technician
has repaired the malfunction.
with the
to clear the
NOTE
In the Troubleshooting Guide, Table 2-5, the Type of
Error Category of the specific Code will be included
on the chart.
EQUIPMENT ALERT
Always complete Service Diagnostics
(Section 2.2) before replacing any major
components, especially the Main P.C. Control Board.
WARNING
Use extreme caution when testing com-
ponents with Sterilizer plugged into
outlet. Line voltage is present. Failure to comply
could result in personal injury.
Restart Errors:
• Like Fast Restart, indicates possible problems with
M4•9 control system or with producing an effective sterilization cycle.
• Operator is prompted to press the STOP button,
abandoning the sterilization cycle.
Pressing the STOP button will stop the audible signal
but 60+ seconds must
indicated and the pressure inside the chamber must
equal to or below 5 Psig (34.5 Kpa) before it returns to
SELECT CYCLE / PREHEAT MODE to allow starting
another cycle.
Touch pad and display do not work.
Sterilizer has
power.
No Display.
Touch pads work.
Sterilizer has
power.
Table 2-5. Troubleshooting Guide
When depressed Touchpad or Display do not work.
There is no display and
touch pad does not work.
Displayed data is a line of
undefined characters and
intermitent beep continues.
SECTION II
TESTING AND TROUBLESHOOTING
Sterilizer is not plugged in. Check that power cord is
No power at outletCheck for voltage at outlet. Reset circuit breaker.
Main (L1) Fuse blown on
back of unit.
Transformer has Open
windings.
Plug on pin connection
J12 on Main P.C. board
has come loose.
Wire harness from Transformer and J12 pin connection has broken
wire(s).
Transformer has open or
shorted winding.
Fuses F1 and F2 on Main
P.C. board are blown or
open.
Ribbon cable from J1 on
Main P.C. board to J3 on
Display P.C. board is disconnected.
J1/ J3 ribbon cable has
open lead(s).
Main P.C. Board is malfunctioning.
Display P.C. Board is malfunctioning.
Ribbon cable from J2 to
touch pad is disconnected
J2 ribbon cable to touch
pad has open lead(s).
plugged into sterilizer and/
or outlet.
Check continuity of fuse.Replace fuse with same style
Check for line
pins 1 & 6 on primary side
of transformer.
Check condition of connector on J12 of Main P.C.
board.
Check continuity of wire
harness from transformer
to J12 pin connection.
Check for 18.5 VAC output
on terminals 7 & 12 at
transformer.
Check continuity of fuses
F1 and F2 with Ohm
meter.
Check plug connectors of
J1 on Main P.C. board and
J2 on Display P.C. board.
Check continuity of J1/J2
ribbon cable.
Check for 18.5 VAC on
outlet side of F1/F2 fuses
and 5 VDC on pin connector J1 across pins 1 & 2,
then across pins 15 & 16
(see Main P.C. Board diagram).
Check for 5 VDC
betweeen R1and Ground
on Display P.C. Board (see
Main P.C. Board diagram).
Check J2 Plug connection. Assure J2 Plug pins are aligned
Check continuity of J2 ribbon cable and touch pad.
voltage at
Plug power cord into sterilizer
and/or outlet.
Assure outlet is on dedicated
circuit capable of handling sterilizer power requirements.
and rating.
If 18.5 VAC is not present on
secondary side when line voltage is present on primary side,
replace Transformer (Refer to
para 4.17).
Plug connector into J12 on Main
P.C. board.
Repair or replace wire harness.
Replace transformer
(Refer to para 4.17).
Replace fuses with same style
and rating.
Assure printer ribbon cable connectors are plugged in securely
Replace ribbon cable.
If voltage is present at F1/F2 but
no voltage is present at J1 and
leads between transformer secondary and J12 have no broken
connections replace Main P.C
Board (Refer to para 4.22).
If no 5 VDC is present replace
Display P.C. Board (Refer to
para 4.4).
and connected securely.
Replace touch pad and ribbon
cable assembly (Refer to
para 4.5).
during Fill Mode,
prompted by TEMP REG
WARNING then it would be
RESTART.)
C103 HEATUP
Code
Code
C104 STERILIZE
Code
C105 VENT
C106 DOOR
Code
(Type: RESTART)
Table 2-5. Troubleshooting Guide
Display:
CHAMBER VENTING
BEFORE FILL
C010: POWER UP MODE
SYSTEM PWR LOSS
ITEMS NOT STERILE
PUSH STOP TO RESTART
Unit lost power for a few
seconds during Fill, Heat
Up, or Sterilization Mode.
Also, if power was lost during the Vent Mode for a
Temperature Regulation
Warning.
C060: POWER UP MODE
SYSTEM HARDWARE
ITEMS NOT STERILE *
UNPLUG/RE-PLUG UNIT
*No message on this line
power loss occurs
unless
during Fill, Heatup, Sterilization or Vent (only in
case of a Temp Reg Warning).
Also, if an error that
occurred in one of these
modes was not properly
cleared before unit was
Unplugged and Replugged
into power source.
C102: FILL MODE
STOP PRESSED
ITEMS NOT STERILE *
PUSH STOP TO RESTART
*No message on this line
unless it occurs during
Venting because of a Temp
Reg Warning.
(Refer to column on left for
codes and description)
C10X: XXXXXX MODE
STOP PRESSED
ITEMS NOT STERILE*
PUSH STOP TO RESTART
*No message on this line
unless it occurs during
Venting because of a Temp
Reg Warning.
TESTING AND TROUBLESHOOTING
Atmospheric pressure setting out of calibration.
Power interruption.Check voltage supply to
May be an internal glitch
due to a brief interruption
of power with components
on P.C. Board (RAM,
EEPROM, A/DC.
Operator pressed STOP
button, terminating FILL
Mode.
assure outlet is on a dedicated circuit with correct
circuit breaker or fuse.
Check voltage supply to
assure outlet is on a dedicated circuit with correct
circuit breaker or fuse.
Check with person operating unit.
Check with person operating unit.
SECTION II
• Unplug Unit & Open Door.
• Place SW2 dip switch #4 ON
• Plug in Unit.
• Displays: START TO ZERO.
• Press START
• Wait 1 - 2 minutes
• Unplug unit.
• Place SW2 dip switch #4 OFF
• Plug in unit.
Plug unit into a dedicated circuit
with correct circuit breaker or
fuse. Refer to Specifications
para 1.5).
Power interruption could be due
to an electrical storm, brown
out, etc.
Plug unit into a dedicated circuit
with correct circuit breaker or
fuse. Refer to Specifications
para 1.5).
Power interruption could be due
to an electrical storm, etc.
Auxiliary temperature
(trough) has not risen to at
least 212°F (100°C) in 10
minutes while unit is in
SELECT/PREHEAT
Note: An initial cycle must
be ran and unit must be
warm for this code to display.
mode.
TESTING AND TROUBLESHOOTING
Door Switch contacts
opened during Heatup
Mode.
Door Switch contacts
opened during Sterilize
Mode.
Door Switch contacts
opened during Vent Mode.
Door left wide open and
tray rack not in position.
Water level of reservoir is
covering outlet of condensing coil. Water
siphons back into chamber
trough resulting in low
Auxiliary (trough) temperature.
Solid State Relay (SSR)
malfunctioning.
Auxiliary (upper) or Steam
(lower) heater has open
winding.
Check with person operating unit to see if door was
opened as unit was heating up.
Check Door Switch. Contact should be closed
when door is in closed
position.
Check Door Switch. Contact should be closed
when door is in closed
position.
Check with person operating unit to see if door was
opened during Vent.
Check Door Switch. Contact should be closed
when door is in closed
position.
Door is left wide open for
an extended length of
time* and tray rack is not in
place in chamber.
*Heaters will cycle on/off
for 90 minutes to maintain
preheat temp. 212°F
(100°C).
Check if condensing coil
outlet is beneath water
level in reservoir.
• Disconnect power from
unit.
• Remove line voltage
leads from terminals 1 & 2
of SSR, place terminals
out of way to prevent electrical shock.
• Place meter probes on
terminals 1 & 2 and check
ohms.
• Run Service Diagnostics,
go to Steam and Auxiliary
Heater Test. When Heaters are on (switch is
closed) apprx. 700 ohms
should register on meter.
Check continuity of both
heaters.
115 VAC
ohms, cold reading.
230 VAC
ohms, cold reading.
heaters are 10
heaters are 39
SECTION II
Explain proper operation.
Push STOP to restar t back to
SELECT CYCLE Mode.
If contact of Door Switch shows
open when door is closed adjust
or replace switch (Refer to para
4.20).
If contact of Door Switch shows
open when door is closed adjust
or replace switch (Refer to para
4.20).
Explain proper operation.
Push STOP to restar t back to
SELECT CYCLE Mode.
If contact of Door Switch shows
open when door is closed adjust
or replace switch (Refer to para
4.20).
Inform Operator that door
should be placed in the partially
open or closed position after
each cycle.
• Inform operator not to over fill
reservoir.
• If reservoir is not over filled but
coil outlet is submerged, bend
coil so outlet is a maximum of 1
7/8” below top of lid opening.
Do not allow coil to touch top or
sides of reservoir.
If ohms value did not register on
meter when heaters were energized replace Solid State Relay
(SSR) (Refer to para 4.15).
If a heater shows no or very different continuity, replace the
heater (Refer to para(s) 4.31 or
Auxiliary temperature
(trough) has not risen to at
least 212°F (100°C) in 10
minutes while unit is in
SELECT/PREHEAT
Note: An initial cycle must
be ran and unit must be
warm for this code to dis-
C437: DRY MODE
AUX TEMP LOW
PUSH STOP TO RESTART
Chamber Temperature
sensed by Auxiliary
(trough) Temp. Sensor
(RTD) is lower than 158° F
(70°C) after 5 minutes into
DRY mode.
mode.
play.
TESTING AND TROUBLESHOOTING
Auxiliary Temperature
Sensor malfunctioning.
Main P.C. Board malfunctioning.
Door left wide open and
tray rack not in position.
Solid State Relay (SSR)
malfunctioning.
Auxiliary (upper) or Steam
(lower) heater has open
winding.
Auxiliary Temperature
Sensor malfunctioning.
Run Service Diagnostics
and check STEAM and
AUX temp readings.
Run Service Diagnostics
and check each function.
Also, make the various
voltage checks as indicated on the P.C. Board
Schematic in Section 5.
Door is left wide open for
an extended length of
time* and tray rack is not in
place in chamber.
*Heaters will cycle on/off
for 90 minutes to maintain
preheat temp. 212°F
(100°C).
• Disconnect power from
unit.
• Remove line voltage
leads from terminals 1 & 2
of SSR, place terminals
out of way to prevent electrical shock.
• Place meter probes on
terminals 1 & 2 and check
ohms.
• Run Service Diagnostics,
go to Steam and Auxiliary
Heater Test. When Heaters are on (switch is
closed) apprx. 700 ohms
should register on meter.
Check continuity of both
heaters.
115 VAC
ohms, cold reading.
230 VAC
ohms, cold reading.
Run Service Diagnostics
and check STEAM and
AUX temp readings.
Remove plug connector at
J3 on P.C. Board and
check resistance value
(ohms) of RTD.
Resistance values should
be between the following:
75°F (23°C) = 1090 ohms.
545°F (285°C) = 2066
ohms.
Above values +/- 10 Ohms
heaters are 10
heaters are 39
SECTION II
If AUX temperature reading is
not in line with STEAM replace
Auxiliary Temperature Sensor
(Refer to para 4.30).
Replace the P.C. Board (Refer
to para 4.22).
run Service Diagnostics
Always
(Para 2.2) before replacing P.C.
Board.
Inform Operator that door
should be placed in the partially
open or closed position after
each cycle.
If ohms value did not register on
meter when heaters were energized replace Solid State Relay
(SSR) (Refer to para 4.15).
If a heater shows no or very different continuity, replace the
heater (Refer to para(s) 4.31 or
4.32).
If AUX temperature reading is
not in line with STEAM replace
Auxiliary Temperature Sensor
(Refer to para 4.30).
If Auxiliary Temperature probe
shows open or values outside
resistance ranges replace
probe (Refer to para 4.30).
(trough) Temp. Sensor
(RTD) is lower than 158° F
(70°C) after 5 minutes into
DRY mode.
C461: SELECT MODE
AUX TEMP HARDWARE
PUSH STOP TO RESTART
Auxiliary A / D converter
reports an average value
outside the limits for normal operation during the
SELECT mode.
Note: This will only occur
during a warm start (back
to back cycles).
C467: DRY MODE
AUX TEMP HARDWARE
PUSH STOP TO RESTART
Auxiliary A / D converter
reports an average value
outside the limits for normal operation during the
DRY mode.
TESTING AND TROUBLESHOOTING
Main P.C. Board malfunctioning.
Auxiliary Temperature
Sensor (RTD) unplugged
from P.C. Board.
Auxiliary Temperature
Sensor (RTD) malfunctioning.
Main P.C. Board Malfunctioning.
Auxiliary Temperature
Sensor (RTD) unplugged
from P.C. Board.
Auxiliary Temperature
Sensor (RTD) malfunctioning.
Main P.C. Board Malfunctioning.
Run Service Diagnostics
and check each function.
Also, make the various
voltage checks as indicated on the P.C. Board
Schematic in Section 5.
Check condition of Auxiliiary RTD plug on P.C.
Board. (Refer to P.C.
Board Schematic in Section 5).
Remove plug connector at
J3 on P.C. Board and
check resistance value
(ohms) of RTD.
Resistance values should
be between the following:
75°F (23°C) = 1090 ohms.
545°F (285°C) = 2066
ohms.
Above values +/- 10 Ohms
Remove plug connector at
J3 on P.C. Board and
check voltage. Voltage
should be apprx. 4.5 VDC
at all times with unit
plugged in.
Also run Service Diagnostics.
Check condition of Auxiliiary RTD plug on P.C.
Board. (Refer to P.C.
Board Schematic in Section 5).
Remove plug connector at
J3 on P.C. Board and
check resistance value
(ohms) of RTD.
Resistance values should
be between the following:
75°F (23°C) = 1090 ohms.
545°F (285°C) = 2066
ohms.
Above values +/- 10 Ohms
Remove plug connector at
J3 on P.C. Board and
check voltage. Voltage
should be apprx. 4.5 VDC
at all times with unit
plugged in.
Also run Service Diagnostics.
SECTION II
Replace the P.C. Board (Refer
to para 4.22).
Always
run Service Diagnostics
(Para 2.2) before replacing P.C.
Board.
Assure Auxiliary RTD plug connector is secure.
If Auxiliary Temperature probe
shows open or values outside
resistance ranges replace
probe (Refer to para 4.30.
If voltage is not present or Service Diagnostics reveals a problem replace P.C. Board (Refer to
para 4.22).
Always
run Service Diagnostics
(Para 2.2) before replacing P.C.
Board.
Assure Auxiliary RTD plug connector is secure.
If Auxiliary Temperature probe
shows open or values outside
resistance ranges replace
probe (Refer to para 4.30).
If voltage is not present or Service Diagnostics reveals a problem replace P.C. Board (Refer to
para 4.22).
Note: When a Code ends
in “0” it means unit had a
supply power interruption.
The operator could have
unplug the unit when a previous Code was displayed.
Always go to Service Diagnostics and check previous
5 codes that are displayed.
Code
C471
(Type: FA TAL)
Table 2-5. Troubleshooting Guide
C470: POWER UP MODE
AUX TEMP OVERLIM
UNPLUG / RE-PLUG UNIT
Auxiliary temperature
exceeds 568.4°F (298°C)
during POWER UP mode.
C471: SELECT MODE
AUX TEMP OVERLIM
UNPLUG / RE-PLUG UNIT
Auxiliary temperature
exceeds 568.4°F (298°C)
during SELECT mode.
TESTING AND TROUBLESHOOTING
Auxiliary Temperature
Sensor (RTD) malfunctioning.
Resistance value of RTD is
above 2113 ohms, indicating temperature is over
568.4°F (298°C)
May be due to a malfunction that occurred previously to Code displayed.
Auxiliary (upper) high limit
switch is weak and opened
prematurely.
P.C. Board malfunctioning. Run Service Diagnostics
Auxiliary Temperature
Sensor (RTD) malfunctioning.
Resistance value of RTD is
above 2113 ohms, indicating temperature is over
568.4°F (298°C)
Solid State Relay (SSR)
shorted causing power
being supplied to heaters
constantly.
P.C. Board malfunctioning. Run Service Diagnostics
Remove plug connector at
J3 on P.C. Board and
check resistance value
(ohms) of RTD.
Resistance values should
be between the following:
75°F (23°C) = 1090 ohms.
545°F (285°C) = 2066
ohms.
Above values +/- 10 Ohms
Go to Service Diagnostics
and check previous 5 displayed codes.
Disconnect power and
check continuity across
high limit switch.
Limit opens at 475° F
(246° C) +/- 25° F (-4° C)
Close at 395° F (202° C)
+/- 25° F (-4° C)
and make various voltage
checks as indicated on
P.C. Schematic.
Remove plug connector at
J3 on P.C. Board and
check resistance value
(ohms) of RTD.
Resistance values should
be between the following:
75°F (23°C) = 1090 ohms.
545°F (285°C) = 2066
ohms.
Above values +/- 10 Ohms
• Disconnect power from
unit.
• Remove line voltage
leads from terminals 1 & 2
of SSR, place terminals
out of way to prevent electrical shock.
• Place meter probes on
terminals 1 & 2 and check
ohms.
• Run Service Diagnostics,
go to Steam and Auxiliary
Heater Test. When Heaters are on (switch is
closed) apprx. 700 ohms
should register on meter.
and make various voltage
checks as indicated on
P.C. Schematic.
SECTION II
If Auxiliary Temperature probe
shows values outside resistance ranges replace probe
(Refer to para 4.30).
Refer to Code Numbers in Troubleshooting Guide to determine
correction.
If continuity is not present and
temperature at limit is below the
Limit’s “Open” temperature
replace Limit Switch
(Refer to para 4.28).
Replace P.C. Board (Refer to
para 4.22).
run Service Diagnostics
Always
(Para 2.2) before replacing P.C.
Board.
If Auxiliary Temperature probe
shows values outside resistance ranges replace probe
(Refer to para 4.30).
If ohms value did not register on
meter when heaters were energized replace Solid State Relay
(SSR) (Refer to
para 4.15).
Auxiliary temperature
exceeds 568.4°F (298°C)
during FILL mode.
C473: HEATUP MODE
AUX TEMP OVERLIM
UNPLUG / RE-PLUG UNIT
Auxiliary temperature
exceeds 568.4°F (298°C)
during HEATUP mode.
C474: STERILIZE MODE
AUX TEMP OVERLIM
UNPLUG / RE-PLUG UNIT
Auxiliary temperature
exceeds 568.4°F (298°C)
during STERILIZE mode.
TESTING AND TROUBLESHOOTING
Auxiliary Temperature
Sensor (RTD) malfunctioning.
Resistance value of RTD is
above 2113 ohms, indicating temperature is over
568.4°F (298°C)
P.C. Board malfunctioning. Run Service Diagnostics
Sterilizer not level, water
fill is not adequate.
Water is not filling trough
adequately due to dirty or
corroded Level Sensor
Probe.
Vent Valve malfunctioning,
staying open.
Air Valve doesn’t close.Check operation of Air
Auxiliary Temperature
Sensor (RTD) malfunctioning.
Resistance value of RTD is
above 2113 ohms, indicating temperature is over
568.4°F (298°C)
P.C. Board malfunctioning. Run Service Diagnostics
Sterilizer not level, water
fill is not adequate.
Remove plug connector at
J3 on P.C. Board and
check resistance value
(ohms) of RTD.
Resistance values should
be between the following:
75°F (23°C) = 1090 ohms.
545°F (285°C) = 2066
ohms.
Above values +/- 10 Ohms
and make various voltage
checks as indicated on
P.C. Schematic.
Check level as described
in Installation of I/O manual.
Check condition of Level
Sensor Probe.
Run Service Diagnostics
and check Vent Valve
operation and for leakage
when closed. Check Vent
Valve Solenoid coil:
115 VAC = 92 Ohms
230 VAC = 376 Ohms
(+ / - 10 %)
Valve by running Service
Diagnostics and check coil
resistance. Refer to Schematic of Main P.C. Board.
Remove plug connector at
J3 on P.C. Board and
check resistance value
(ohms) of RTD.
Resistance values should
be between the following:
75°F (23°C) = 1090 ohms.
545°F (285°C) = 2066
ohms.
Above values +/- 10 Ohms
and make various voltage
checks as indicated on
P.C. Schematic.
Check level as described
in Installation of I/O manual.
SECTION II
If Auxiliary Temperature probe
shows values outside resistance ranges replace probe
(Refer to para 4.30).
Replace P.C. Board (Refer to
para 4.22).
run Service Diagnostics
Always
(Para 2.2) before replacing P.C.
Board.
Level unit.
Clean or replace Level Sensor
Probe (Refer to para 4.33).
If coil shows an “open” or leakage is determined, replace Vent
Valve (Refer to para 4.27).
Replace Air Valve Assembly
(Refer to para 4.24).
If Auxiliary Temperature probe
shows values outside resistance ranges replace probe
(Refer to para 4.30).
Water is not filling trough
adequately due to dirty or
corroded Level Sensor
Probe.
Vent Valve malfunctioning,
staying open.
Air Valve doesn’t close.Check operation of Air
Auxiliary Temperature
Sensor (RTD) malfunctioning.
Resistance value of RTD is
above 2113 ohms, indicating temperature is over
568.4°F (298°C)
Solid State Relay (SSR)
shorted causing power
being supplied to heaters
constantly.
P.C. Board malfunctioning. Run Service Diagnostics
Check condition of Level
Sensor Probe.
Run Service Diagnostics
and check Vent Valve
operation and for leakage
when closed. Check Vent
Valve Solenoid coil:
115 VAC
230 VAC
(+ / - 10 %)
Valve by running Service
Diagnostics and check coil
resistance. Refer to Schematic of Main P.C. Board.
Remove plug connector at
J3 on P.C. Board and
check resistance value
(ohms) of RTD.
Resistance values should
be between the following:
75°F (23°C) = 1090 ohms.
545°F (285°C) = 2066
ohms.
Above values +/- 10 Ohms
• Disconnect power from
unit.
• Remove line voltage
leads from terminals 1 & 2
of SSR, place terminals
out of way to prevent electrical shock.
• Place meter probes on
terminals 1 & 2 and check
ohms.
• Run Service Diagnostics,
go to Steam and Auxiliary
Heater Test. When Heaters are on (switch is
closed) apprx. 700 ohms
should register on meter.
and make various voltage
checks as indicated on
P.C. Schematic.
= 92 Ohms
= 376 Ohms
SECTION II
Clean or replace Level Sensor
Probe (Refer to para 4.33).
If coil shows and “open” or leakage is determined, replace Vent
Valve (Refer to para 4.27).
Replace Air Valve Assembly
(Refer to para 4.24).
If Auxiliary Temperature probe
shows values outside resistance ranges replace probe
(Refer to para 4.30).
If ohms value did not register on
meter when heaters were energized replace Solid State Relay
(SSR) (Refer to para 4.15).
*No message on this line
unless it occurs during
Venting because of a Temp
Reg Warning.
Auxiliary temperature
exceeds 568.4°F (298°C)
during VENT mode.
C477: DRY MODE
AUX TEMP OVERLIM
UNPLUG / RE-PLUG UNIT
Auxiliary temperature
exceeds 568.4°F (298°C)
during DRY mode.
TESTING AND TROUBLESHOOTING
Auxiliary Temperature
Sensor (RTD) malfunctioning.
Resistance value of RTD is
above 2113 ohms, indicating temperature is over
568.4°F (298°C)
Solid State Relay (SSR)
shorted causing power
being supplied to heaters
constantly.
P.C. Board malfunctioning. Run Service Diagnostics
Auxiliary Temperature
Sensor (RTD) malfunctioning.
Resistance value of RTD is
above 2113 ohms, indicating temperature is over
568.4°F (298°C)
Main Relay malfunctioning
keeping Steam Heater
energized.
Solid State Relay (SSR)
shorted causing power
being supplied to heaters
constantly.
Remove plug connector at
J3 on P.C. Board and
check resistance value
(ohms) of RTD.
Resistance values should
be between the following:
75°F (23°C) = 1090 ohms.
545°F (285°C) = 2066
ohms.
Above values +/- 10 Ohms
• Disconnect power from
unit.
• Remove line voltage
leads from terminals 1 & 2
of SSR, place terminals
out of way to prevent electrical shock.
• Place meter probes on
terminals 1 & 2 and check
ohms.
• Run Service Diagnostics,
go to Steam and Auxiliary
Heater Test. When Heaters are on (switch is
closed) apprx. 700 ohms
should register on meter.
and make various voltage
checks as indicated on
P.C. Schematic.
Remove plug connector at
J3 on P.C. Board and
check resistance value
(ohms) of RTD.
Resistance values should
be between the following:
75°F (23°C) = 1090 ohms.
545°F (285°C) = 2066
ohms.
Above values +/- 10%
Run Service Diagnostics,
Main Relay Test, Steam &
Auxiliary Heater Test.
• Disconnect power from
unit.
• Remove line voltage
leads from terminals 1 & 2
of SSR, place terminals
out of way to prevent electrical shock.
• Place meter probes on
terminals 1 & 2 and check
ohms.
• Run Service Diagnostics,
go to Steam and Auxiliary
Heater Test. When Heaters are on (switch is
closed) apprx. 700 ohms
should register on meter.
SECTION II
If Auxiliary Temperature probe
shows values outside resistance ranges replace probe
(Refer to para 4.30).
If ohms value did not register on
meter when heaters were energized replace Solid State Relay
(SSR) (Refer to para 4.15).
Replace P.C. Board (Refer to
para 4.22).
run Service Diagnostics
Always
(Para 2.2) before replacing P.C.
Board.
If Auxiliary Temperature probe
shows values outside resistance ranges replace probe
(Refer to para 4.30).
If Main Relay did not function
correctly replace Relay (Refer
to para 4.16).
If ohms value did not register on
meter when heaters were energized replace Solid State Relay
(SSR) (Refer to
para 4.15).