Midmark Casework Maintance Manual

Casework Pre-Installation Guide
C
H
Attention!
For OSHPD Regulated Installations, please go to
www.midmark.com to download the appropriate
OSHPD Pre-Approval and Installation documents.
MA707900i
NOTES:
• The wall layout shows a LH unit. If unit being installed is a RH unit, layout and dimensions are mirror image.
© Midmark Corporation 2006
Page 1
003-1620-00 Rev. D (3/11)
Wall Layout Preparation for Midmark (w/insert) & Ritter
NOTE:
Use the layout below for electrical / plumbing installation.
47 3/4"
CORNER OR BEGINNING
OF UNOBSTRUCTED WALL
(121.3 cm)
ATTENTION:
Recommended: 2 x 6 backing prior to installation of drywall.
73"
(185.4 cm)
DESIRED HEIGHT
OF CABINET
(MINIMUM 3" FROM CEILING)
24"
(60.9 cm)
18"
[SUGGESTED]
(45.7)
COUNTER TOP
34 1/2"
(87.6 cm)
36"
(94.1 cm)
39 1/4"
(99.7 cm)
28 3/8"
(72.1 cm)
(6.3 cm)
22"
(55.9 cm)
6 3/8"
(16.2 cm)
2 1/2"
21 1/2"
(54.6 cm)
19"
(48.3 cm)
PLUMBING
CUT_OUT
48 3/4"
(123.8 cm)
33 1/4"
(76.2 cm)
23 1/4"
(59 cm)
5/8"
(1.6 cm)
24"
(60.9 cm)
72"
(182.9 cm)
48"
(121.9 cm)
MA731200i
© Midmark Corporation 2006
Page 2
Wall Layout Preparation for Midmark Woodgrain with Storage End Panels
NOTE:
Use the layout below for electrical / plumbing installation.
CORNER OR BEGINNING
OF UNOBSTRUCTED WALL
ATTENTION:
Recommended: 2 x 6 backing prior to installation of drywall.
73 3/4"
(187.3 cm)
48 1/2"
(123.2 cm)
DESIRED HEIGHT
OF CABINET
(MINIMUM 3" FROM CEILING)
24"
(60.9 cm)
18"
[SUGGESTED]
(45.7)
COUNTER TOP
34 1/2"
(87.6 cm)
36"
(94.1 cm)
39 1/4"
(99.7 cm)
28 3/8"
(72.1 cm)
(6.3 cm)
22"
(55.9 cm)
6 3/8"
(16.2 cm)
2 1/2"
21 1/2"
(54.6 cm)
19"
(48.3 cm)
PLUMBING
CUT-OUT
49 1/2"
(125.7 cm)
48 3/4"
(123.8 cm)
24"
(60.9 cm)
72 3/4"
(184.8 cm)
33 1/4"
(76.2 cm)
23 1/4" (59 cm)
5/8"
(1.6 cm)
MA731300i
© Midmark Corporation 2006
Page 3
Wall Layout Preparation for Midmark Woodgrain with Storage & Sink End Panels
NOTE:
Use the layout below for electrical / plumbing installation.
CORNER OR BEGINNING
OF UNOBSTRUCTED WALL
ATTENTION:
Recommended: 2 x 6 backing prior to installation of drywall.
74 1/2"
(189.2 cm)
49 1/4"
(125.1 cm)
(MINIMUM 3" FROM CEILING)
18"
[SUGGESTED]
(45.7)
36"
(94.1 cm)
39 1/4"
(99.7 cm)
DESIRED HEIGHT
OF CABINET
24"
(60.9 cm)
COUNTER TOP
34 1/2"
(87.6 cm)
28 3/8"
(72.1 cm)
22"
(55.9 cm)
6 3/8"
(16.2 cm)
22 1/4"
(56.5 cm)
19"
(48.3 cm)
PLUMBING
CUT-OUT
50 1/4"
(127.6 cm)
49 1/2"
(125.7 cm)
24"
(60.9 cm)
72 1/2"
(186.7 cm)
33 1/4"
(76.2 cm)
23 1/4"
(59 cm)
5/8" (1.6 cm)
MA731400i
3 1/4"
(8.3 cm)
© Midmark Corporation 2006
Page 4
8' Steri-Center Casework
NOTE:
Use the layout below for electrical / plumbing installation.
Refer to
"36" High Sink Base Cabinet"
"84" High Sterilizer Cabinet"
and Cabinet cut-outs.
HEIGHT
DESIRED
OF CABINET
(MINIMUM 3" FROM CEILING)
)
24"
(61.0 cm
)
30"
(76.2 cm
and
for details of Sink
7' - 11 3/4" (243.2 cm
(61.0 cm
)
29 3/4"
(75.6 cm
24"
ATTENTION:
Recommended: 2 x 6 backing prior to installation of drywall.
)
24"
(61.0 cm
)
)
18"
(45.7 cm
)
84"
(205.7 cm
39 1/4"
(99.7 cm
)
COUNTER TO
)
(91.4 cm
36"
18"
(45.7 cm
P
34 1/2"
(87.6 cm
)
)
)
24"
(61.0 cm
)
)
3/4"
(1.9 cm
)
15"
(38.1 cm
15"
(38.1 cm
)
24"
(61.0 cm
)
8' - 3/4"
(245.7 cm
18"
(45.7 cm
)
)
MA731500i
© Midmark Corporation 2006
Page 5