Midmark 9A408H1, 9A408F1, 9A408T1 User Manual

The wireless control’s “association” is preset at the factory. This procedure is only required if there is a malfunction, or if you are attempting to use the controller with a chair other than the one it was originally associated with.
Association Procedure:
A. Using the wired hand or foot control, press & hold
the Stop & Home buttons simultaneously.
(after two seconds you will hear a single “beep”)
B*. Using the wireless hand control, press & hold
the Stop & Home buttons until you hear three “beeps”.
(one long “beep” indicates the association failed. Repeat Steps A & B)
C. Check for proper operation of wireless hand control.
D. Repeat procedure for wireless foot control.
* Note: Step B must be performed within 3 seconds of Step A,
failure to do so will cause the procedure to fail.
Specifications
Range of Operation ................. 20 ft (6 m) radius
Batteries:
Hand Control ......................... AA (qty: 2)
Foot Control .......................... AA (qty: 2)
(Note: Battery access cover is located on back of controller)
MA771100i
Troubleshooting
• When any control button is pressed, that controller sounds a single “beep”. (All functions operate)
Solution: Replace batteries
• When any control button is pressed, nothing happens. (No “beeps”, no movement, etc.)
Causes: 1. Batteries completely dead
Solution: 1. Replace batteries
* The wireless controllers will only function with the
2. Wireless controller has lost “association” to chair.*
2. Perform Association Procedure*.
chair they have been “associated” with. This prevents unwanted movement in locations with more than one chair.
If further service is required,
contact your authorized Midmark dealer.
Compliance Information
Limited Warranty
General Information
FCC Compliance Statement
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interfer­ence when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Any modifications to this product may violate the rules of the Federal Communications Commission and make operation of the product unlawful.
Declaration of Conformity
Midmark Corporation declares that 9A408001 thru 9A408008 conforms to the applicable requirements of the Radio Equipment & Telecommunications Equipment Directive 1999/5/EC, Annex II.
Serial Number: ALL to present
Quality Assurance Manager
Gene Harshbarger Midmark Corp. 60 Vista Dr. P.O. Box 286 Versailles, OH 45380 U.S.A.
077-0520-00
SCOPE OF WARRANTY
Midmark Corporation (“Midmark”) warrants to the original purchaser its new Alternate Care products and components (except for components not warranted under “Exclusions”) manufactured by Midmark to be free from defects in material and workmanship under normal use and service. Midmark’s obligation under this warranty is limited to the repair or replacement, at Midmark’s option, of the parts or the products the defects of which are reported to Midmark within the applicable warranty period and which, upon examination by Midmark, prove to be defective.
APPLICABLE WARRANTY PERIOD
The applicable warranty period, measured from the date of delivery to the original user, shall be one (1) year for all warranted products and components.
EXCLUSIONS
This warranty does not cover and Midmark shall not be liable for the following: (1) repairs and replacements because of misuse, abuse, negligence, alteration, accident, freight damage, or tampering; (2) products which are not installed, used, and properly cleaned as required in the Midmark “Installation” and or “Installation / Operation Manual for this applicable product. (3) products considered to be of a consumable nature; (4) accessories or parts not manufactured by Midmark; (5) charges by anyone for adjustments, repairs, replacement parts, installation, or other work performed upon or in connection with such products which is not expressly authorized in writing in advance by Midmark.
EXCLUSIVE REMEDY
Midmark’s only obligation under this warranty is the repair or replace­ment of defective parts. Midmark shall not be liable for any direct, special, indirect, incidental, exemplary, or consequential damages or delay, including, but not limited to, damages for loss of profits or loss of use.
NO AUTHORIZATION
No person or firm is authorized to create for Midmark any other
obligation or liability in connection with the products.
THIS WARRANTY IS MIDMARK’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. MIDMARK MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS.
SF-1487 REV. A1
Attention: Wired Hand & Foot Controls
Wired hand & foot controls are shipped standard with all models. A wired controller (either hand or foot) is required to perform the Association Procedure. This procedure may be required in the event of a malfunction, do not discard the wired controllers.
(Refer to: Association Procedure in this User Guide)
Intended Use
This product is intended to control patient positioning while providing better patient access by eliminating the cord from the hand & foot controls.
Authorized CE Representative
Countries in the EEC should direct all questions, incidents, and complaints to Midmark’s authorized CE Representative listed below:
Promotal 22, Rue de St. Denis 53500 Ernee, FRANCE Phone: 33-2-430-517-76
Fax: 33-2-430-569-10
Calling for Service
If service is required, contact your Midmark dealer or to locate your nearest service provider, log onto:
http://www.midmark.com
To contact Midmark directly:
1-800-Midmark (1-800-643-6275) 8:00 am until 5:00 pm Mon-Fri (EST)
User’s Guide
Wireless Hand / Foot Controls
9A408
This guide outlines basic operation of the wireless hand / foot control accessory. If service is required, contact your authorized Midmark dealer.
Dealer (name / phone):
___________________________________ ___________________________________ ___________________________________ ___________________________________
Midmark Corporation 60 Vista Drive Versailles, OH 45380-0286 Phone: 937-526-3662 Fax: 937-526-5542 midmark.com
Date of Purchase:
© Midmark Corporation 2007
_____________________
003-1735-00 Rev.B
Loading...