To comply with changes in Canadian regulatory requirements you must now provide documentation, a Manufacturer’s Data Report (MDR), to your appropriate provincial regulatory agency when purchasing and using a steam sterilizer.
To obtain the MDR from Midmark, record the sterilizer Serial Number and the ASME National Board Num-ber of your sterilizer. See the illustration below for the location of these numbers. Then, contact Midmark customer service at 1.800.MIDMARK and provide the recorded numbers.
Midmark will then complete the MDR and send it to you. Once you have the MDR, you must submit it to the appropriate provincial regulatory agency.
We apologize for any inconvenience and appreciate your assistance in helping maintain compliance to Ca-nadian regulations for sterilizers. If you have any questions or concerns, please contact 1.800.MIDMARK.
The primary concern of Midmark is that this equipment is operated and maintained with the safety of
the patient and sta in mind. To assure safe and reliable operation:
• Read and understand this manual before attempting to install or operate the sterilizer.
• Assure that the appropriate personnel are informed on the contents of this manual.
(This is the responsibility of the purchaser).
• Assure that this manual is located near the sterilizer, or if possible, permanently axed to the sterilizer.
Intended Use
The M3 Steam Sterilizer can be used in medical, dental and veterinary oces, hospitals,
clinics, nursing homes, laboratories and other facilities to sterilize heat and moisture stable reusable items
(including dental handpieces) that are compatible with steam sterilization.
Refer to ‘Loading the Tray’ and ‘Cycle Parameters’ in this manual for detailed information.
Sterilization Technology
The Sterilization Technology used in the M3 utilizes a dynamic air removal system called Steam Flush Pressure Pulse to remove air from the chamber.
Electromagnetic Interference
The M3 is designed and built to minimize electromagnetic interference with other devices.
However, if interference is noticed between another device and this sterilizer:
• Remove interfering device from room
• Plug sterilizer into a dedicated circuit
• Increase separation between sterilizer and interfering device
The information below is provided for reference only. Contact appropriate state / local agencies for
specific sterilization guidelines for your office. Additional information on infection control is available
from the Centers for Disease Control and Prevention (CDC), Organization for Safety and Asepsis
Procedures (OSAP) and the American Dental Association (ADA).
Physical Monitors
Temperature and pressure measuring devices can help detect sterilizer malfunctions.
The sterilizer’s control system aborts the cycle and displays a message if physical conditions go
outside established limits. The optional printer can be used to create a record of each load’s actual
cycle time, temperature and pressure.
Note
Use only FDA cleared chemical and biological indicators designed for steam sterilization that are
compatible with the particular sterilization cycle temperature and exposure time being monitored.
Use sterility monitors with each sterilization load. If a sterilization cycle is terminated prematurely,
reprocess instruments to ensure sterility of the load. Process the load according to your regular
practice, placing indicators near the handle side of tray. Follow manufacturer’s instructions for
proper disposal of used indicators.
Chemical Indicators
Chemical indicators are designed to verify that conditions in the sterilizer chamber were adequate to
achieve sterilization. They do not validate that a processed item is sterile. If a chemical indicator shows
a failure, items in that load are considered non-sterile. Potential causes for sterilization failures include:
improper cleaning, packing, loading, or a sterilizer malfunction. Determine the cause of any sterilization
failure and remedy the situation before running the next cycle. Only FDA cleared chemical indicators
labeled for use with the nontraditional steam sterilization cycle parameters, e.g. temperature and exposure
time, of the M3 Sterilizer should be used for monitoring the three M3 cycles. Follow the chemical
indicator’s instructions for proper storage, use, interpretation and disposal.
Biological Indicators
Biological indicators are microbiological devices designed to accompany items being sterilized to monitor
adequacy of the sterilization process. If a biological indicator shows a failure, items in that load are
considered non-sterile. Potential causes for sterilization failures include: improper cleaning, packing,
loading, or a sterilizer malfunction. Determine the cause of any sterilization failure and remedy the
situation before running the next cycle. Only FDA cleared biological indicators labeled for use with the
nontraditional steam sterilization cycle parameters, e.g. temperature and exposure time, of the M3 Sterilizer
should be used for monitoring the three M3 cycles. Follow the biological indicator’s instructions
for proper storage, use, interpretation and disposal.
• Surface mustbe level to ensure proper operation.
• Surface should meet minimum dimensions listed below:
Dimensions
Depth (front to back).................24 in. (61 cm)
Width (side to side)...................22 in. (56 cm)
Clearance Requirements
To ensure proper air circulation and to allow access to the reservoir ll port and drain coupling,adhere to the minimum clearance requirements listed below.
Clearance Requirements
Back of Unit - Back Wall.................................................4 in. (10 cm)
Front Sterilizer Feet - Front of Support Surface.............4 in. (10 cm)
Side of Unit - Side Wall...................................................2 in. (5 cm) each side
Distance Above Unit*......................................................2 in (5 cm)*
* The minimum clearance for proper air circulation is listed.
However, be sure to allow access to the reservoir ll port located on top of the sterilizer.
Page 10
Operating Environment
Equipment Alert
Unit should be allowed to reach room temperature before operating.
Failure to do so could result in damage.
Ambient Temperature Range: ..........+68°F to 104°F (+20°C to 40°C)
Relative Humidity:............................ less than 80% (non-condensing)
(Pollution Degree 2, in accordance to IEC664)
Normal Operating Altitude: .............. less than 6000 ft. (1829 m) above sea level
• Approved for indoor use only
• Environment should be relatively dust-free
Electrical Ratings / Requirements
Note
To ensure unit is properly grounded, it must be connected to a
matching grounded, dedicated, correctly polarized receptacle.
WARNING
Use 104-127 VAC, 50/60 HZ alternating current only for 115 VAC rated models and
207-253, 50/60 HZ alternating current only for 230 VAC rated models. Failure to do
so could result in electrical shock to personnel and will result in damage to sterilizer.
Check the serial number label on back panel of sterilizer to verify voltage rating for the unit. Failure to connect sterilizer to an appropriate power supply could result in damage to the unit and electrical shock to personnel.
Voltage
Rating
WARNING
Equipment is not suitable for use in the presence of a flammable anesthetic mixture with oxygen, air, or nitrous oxide.
Clarification: Equipment issuitable for use in the presence of oxygen, air, or nitrous oxide.
Equipment Alert
For optimal performance, allow sterilizer to reach room temperature before operating.
To connect the power cord...
A)Plug power cord into receptacle on back of sterilizer.B)Plug power cord into a properly polarized and grounded receptacle rated for a minimum of 15 amps. Adedicated circuit only used for the sterilizer is recommended.
The tank hose mustbe properly connected(and not kinked). Improper connection willcause water / steam leaks or a sterilizer
malfunction if water flow to the tank is restricted.
Note
External Condensing Tank is not needed if M3 is connectedto the VistaCool™Direct-To-Drain Thermal Reduction System.
To connect the condensing tank...
A)If connected, disconnect sterilizer power cord.B)Push tank hose into compression fitting on lid. Tighten nut. (If step B is difficult - see NOTE)C)Connect tank hose to back of sterilizer.D)Fill condensing tank with tap water to minimumwater level indicator line.E)Secure the lid on the condensing tank by aligningthe tabs and rotate lid clockwise to secure.
Note
If tubing is difficult to install...
A) Remove compression nut and two ferrules.B) Install nut and ferrules onto tank hose as shown.(Large and small ferrules mustbe positioned as shown)C) Insert hose into tank fitting, then tighten nut.
Power Cord
Compression
Fitting
Tank Hose
Min. Water Level
Indicator Line
Ferrule(small)
Ferrule
(large)
Nut
Page 13
0
External Condensing Tank Connections - continued
Power Cord
Sensor Plug
0
To connect the condensing tank - continued...
F) Place tank on a level surface, preferably below
the sterilizer but in no case should it be higher
than the sterilizer support surface.
G) Connect sensor plug.
H) Connect sterilizer power cord.
003-1658-99
Note
Clearance...
Maintain a minimum of 6 inch clearance above the condensing tank for
proper steam ventilation. If enclosed in a cabinet the support surface
and surrounding surfaces, should be protected with a water resistant
material (e.g. plastic, laminate, stainless steel, etc.). If enclosed in a
cabinet, it is recommended that the door be vented to avoid heat,
moisture build up and potential damage to the inside of cabinet.
Water that is discharged to external condensing tank can be VERY HOT;
person emptying pitcher should allow the temperature to cool.
Always use carrying handle and use caution when emptying.
To remove lid...
A) Rotate lid counterclockwise.
B) Lift lid assembly out of
condensing tank pitcher.
Lid
Assembly
To replace lid...
Note: Lid can be reinstalled in one of four locations
to facilitate safe water disposal.
A) Insert lid assembly into pitcher.
B) Align the four tabs for correct positioning.
C) Rotate lid clockwise to secure.
003-1658-99
English - 14
Tab
To drain pitcher...
A) Grasp pitcher by handle.
B) Pour cool water into an
approved drain.
(Detailed instructions for each step are outlined in the following pages of the Operation section).
Basic Operation...
A)Turn power switch ON (I).B)Fill reservoir.C)Load tray.D)Press desired cycle button.E)Press <Start> button.
Note
There is a 10 minute (dry) heat cycle that can be used to pre-heat the chamber, or for additional drying time.Refer to: ‘Additional Heat’Cyclein this section.
Pour one (1) gallon of distilled water into fill port.Do notfill above lower lip of fill port.
The power cord must be connectedand the power switch must be ON (I)for the sterilizer to operate.
Filling the Reservoir
Lower
Lip
Equipment Alert
Use distilled water or water that meets thereferenced water purity specifications.Failure to comply may result in sterilizermalfunction and/or premature failure due
to excessive corrosion.
OFF (O)
ON (I)
Page 17
Qualication Testing
Your sterilizer should be tested after sterilizer installation, malfunctions, relocation, major repairs and
after sterilization process failure. Qualication testing should be performed prior to placing the sterilizer in
service. If multiple cycles types are used, e.g. “Pouches” and “Low Temp” each cycle type should be
qualied. Qualication testing should include at least one Biological Indicator (BI) (sometimes referred to as
Spore Tests) and one Chemical Indicator (CI). The test pack should be placed near the front of the tray and
performed with items routinely processed and considered to be the most dicult to sterilize. Additional items
should be placed in the chamber along with the Biological Indicator and Chemical Indicator so that cham-
ber is fully loaded (don’t exceed the maximum capacities listed in the tables under “Guidelines for Loading”
in this manual). Three consecutive test runs, for each cycle type tested, with negative results from the BIs
and the appropriate readings from all physical monitors and chemical indicators demonstrating complete
sterilization, provide verication that the sterilizer has been properly installed (or reinstalled after relocation)
or repaired to the manufacturer’s specications and that it will function eectively in the facility in which it
is installed. All items processed during qualication testing should be quarantined until the results of the
biological testing for all three test runs are available.
Loading the Tray
Types of Items
Before placing any instrument in the M3,
check with the instrument manufacturer to be sure the
materials are compatible with steam sterilization and
to verify the acceptability of sterilization parameters.
The M3 is capable of immediate use sterilization - sterilizing unwrapped instruments for immediate use.
Please consider the following when choosing whether or not to sterilize your instruments unwrapped:
• The sterility of unwrapped instruments is compromised upon exposure to a non-sterile environment. Follow CDC guidelines for using unwrapped, sterilized instruments.
• Due to the sensitive nature of some types of surgery (including, but not limited to ophthalmological), instruments used in such procedures must be wrapped or pouched
in order to reduce their exposure to sterilization process residues. The water reservoir
should also be drained and relled with fresh distilled water on a daily basis when
processing instruments for these procedures on a routine basis.
Pouching and Wrapping Items
The M3 is capable of sterilizing pouched or wrapped items.
• When pouching or wrapping items, use only sterilizer pouches and wraps that have
been cleared by the FDA and labeled for use with the nontraditional steam sterilization
cycle parameters, e.g. temperature and exposure time, of the M3 sterilizer. Follow the
manufacturer’s instructions for use.
• When using cassettes in the M3 follow the manufacturer’s instructions for use.
• Pouched items to be sterilized should be placed lengthwise with plastic side of
pouch facing up in the M3 tray. The pouches may overlap slightly, but items must not be layered.
Refer to diagram below.
POUCH
POUCH
POUCH
POUCH
POUCH
POUCH
POUCH
POUCH
WARNING
Failure to comply with these guideline may adversely affect sterilization and/or drying.
Load Size
The M3 Steam Sterilizer can accommodate loads weighing up to 2.4 lbs (1.1 kg).
[Note: This is the weight of the contents in the tray (ex. instruments, cassettes, pouches, etc.).
The weight of the tray itself has already been accounted for].
WARNING
Do not overload the sterilizer tray. Failure to allow adequate space
around items for circulation will compromise sterilization and drying.
Use the table below as a general guideline for weights of commonly used items.
Consult manufacturer’s specications for the exact weight of any particular instrument.
Item Description Weight*
lbs. kg
Scissors0.0660.030
Dental Scalers0.0440.020
Forceps0.0330.015
Dental Handpiece0.1210.055
Suction Cannula0.0220.010
Plastic Mouth Mirror0.0180.008
Impression Tray0.0330.015
Plastic X-Ray Positioning Ring0.0440.020
Hu-Friedy Signa-Stat Cassette1.5000.680
Packing the Tray
Equipment Alert
Do not use towels or packaging which contains chlorine bleach residue. Chamber and/or tray
rusting or discoloration may occur. The life of the sterilizer may be shortened significantly.
(*actual weights may vary)
WARNING
Clean and dry instruments thoroughly before placing them into tray. Improper cleaning
may result in non-sterile instruments or damage to the unit. Follow instrument
manufacturer’s guidelines and CDC recommendations for handling and cleaning instruments
prior to sterilization.
In addition to total load weight outlined above, all items must be processed in accordance with Centers
for Disease Control and Prevention (CDC), ‘Guidelines for Infection Control in Dental Healthcare
Settings’ - 2003, MMWR 2003; 52 (no. RR-17), which states:
“Items to be sterilized should be arranged to permit free circulation of the sterilizing agent (e.g., steam,
chemical vapor, or dry heat); manufacturer’s instructions for loading the sterilizer should be followed.”
• All items must t in M3 tray.
• Loaded tray must slide into chamber opening without scraping.
• Items must not touch one another.
• Pouched items should be loosely packed.
• Pouches may overlap slightly, but items must not be layered.
WARNING
Failure to comply with these guideline may adversely affect sterilization and/or drying.
After sterilization is complete, all items must be handled in accordance with accepted and documented
standards, such as the Centers for Disease Control and Prevention (CDC) document, ‘Guidelines for
Infection Control in Dental Healthcare Settings’ - 2003, MMWR 2003; 52 (no. RR-17), as well as any local
requirements that may apply.
Qualied personnel responsible for infection control should prepare a protocol for handling sterilized items.
This protocol should be followed by all personnel responsible for handling sterilized items.
‘Additional Heat’ Cycle
The Additional Heat Cycle activates the
dry heaters for ten minutes.
This cycle can be used to pre-heat the
chamber at the beginning of the workday,
or for extended drying time at the
end of a cycle.
To pre-heat chamber prior to running a cycle...
A) Press <Start> button when ‘SELECT CYCLE’ appears on display.
B) During the ten minute pre-heat mode, ‘ADDITIONAL HEAT’ will flash on the display.
C) When ‘ADDITIONAL HEAT’ stops flashing, press desired cycle button, then press <Start>.
For extended drying time at the end of a cycle...
A) Press <Start> button when ‘SELECT CYCLE’ appears on display.
B) During the ten minute drying mode, ‘ADDITIONAL HEAT’ will flash on the display.
The M3 allows the operator to adjust
the drying time from 20 - 60 minutes
using 1 minute increments for the three
pre-programmed cycles.
To adjust the drying time for a pre-programmed cycle...
A) After pressing desired cycle button, press <P> button.
[Display will show current setting. (ex. DRY TIME: 30 MINUTES)]
B) Press the <+> / <_> buttons to increase / decrease the drying time.
C) Press the <P> button to save your changes.
(Pressing the <Stop> button cancels the changes and returns to last saved setting.
Maintenance
Maintenance Messages
To assure correct operation and maximum sterilizer life, the M3 provides the operator with reminders
when it’s time to perform operator maintenance. After the M3 is powered ON for 7, 14 and 21 days,
a message “Perform Periodic Maintenance” will be displayed. After 28 days, a “Perform Monthly
Maintenance” message will be displayed. Refer to the appropriate maintenance instructions in this manual.
The maintenance reminders are removed from the display when a cycle is started.
If power is turned OFF, the timer will reset, initiating a new cycle of messages.
A. Wash unit according to your facility’s procedure for clinical contact surfaces noting thefollowing:(Use only quaternary disinfectants to disinfect unit. Staining, pitting, discoloration, orsoftening could occur if phenolic, iodophor, or glutaraldehyde-based disinfectant is usedon plastic surfaces of the unit. Also, use of alcohol or aerosol spray cleaner / disinfectantcontaining substantial amounts of alcohol in the formula can damage the faceplate.)B. Wring excess solution from cloth.C. Using soft cloth, wipe all external surfaces.D. Follow the instructions provided with the cleaner / disinfectant used regarding rinsing and dryingof external surfaces.
WARNING
Check to assure thechamber filter is not blocked bydebris and that it is properly seated;in the bottom hole in the back of thechamber. Failure to make sure thefilter is in place and clear of debriscould result in serious personal
injury and or equipment damage.
• Clean Chamber Filter
Check that the chamber lter is free of debris andproperly positioned in the hole at the back of thesterilizer chamber. If the lter is blocked by debrisfollow the procedures under “Monthly Maintenance”
to remove and clean the lter.
• Clean Door Gasket / Mating Surface
A. Wash with a damp cloth.B. Inspect gasket for damage.C. Replace gasket if necessary.
A. Remove drain hose from storage location.B. Place open-end of drain hose into container or sink.C. Connect adapter-end of drain hose to coupling as shown.D. Once water has drained, press release lever and remove hose.E. Return drain hose to storage location.
F. Rell reservoir with distilled water.
Distilled
Water
Note
If coupling leaks, reinsert hoseseveral times to clean seal.
Release Lever
Drain Hose (storage location)
Periodic Maintenance
Equipment Alert
Use distilled water or water that meets thereferenced water purity specifications.Failure to comply may result in sterilizermalfunction and/or premature failure due
• Empty / Clean External Condensing Tank (As Needed)
A. Empty water from tank. (Do notreuse water!)B. Clean tank with diluted bleach solution (1/4 cup bleach : 1 gallon water) and a brush.C. Rinse tank thoroughly.
D. Rell tank to minimum water level indicator line.
Note
This process is not necessary if the sterilizer is connected to a direct-to-drain thermal reduction system.
Periodic Maintenance - continued
Caution
Water may be HOT!Allow water to cool before emptying tank.
A) Slide notch in filter tool (provided) over filter.B) Pull out to remove.
To install filter...
A) Position filter in notch of tool as shown.B) Align filter with hole in back of chamber* and press in gently.C) Flip tool over, then gentlypress filter in to secure.
*Tip:Slide tool along bottom of chamber to align filter with hole.
Filter Tool
Filter
Extended Use Maintenance
The M3 is designed and tested to provide exceptional reliability throughout its service life. However, likeall electro-mechanical devices it is subject to wear and degradation with use.
To ensure the integrity, performance and safety of all major components it is the responsibility of the userto have the sterilizer performance / operation veried by a Midmark Authorized Service Provider after8 years or 20,000 cycles of use, whichever comes rst. After 8 years or 20,000 cycles of use an annual
inspection by a Midmark Authorized Service Provider is recommended.
A)Separate lid / printer from housing.B)Press Paper Feed buttonfor three seconds.(This advances the ribbon inside the cartridge).C)Repeat step B as necessary*.
*Note:If pressing the Paper Feed button fails to correct the problem, replace the printer cartridge.
A) Loosen thumbscrew, thenseparate lid / printer from housing.B) Slide paper into slot as shown*.C) Press paper feed button until approx.2 in. (5 cm) of paper is fed thru printer.D) Pull paper thru slot in lid.E) Position paper roll / spindle in slots.F) Reassemble lid / printer / housing.
A)Turn sterilizer power switch OFF(O).B)Separate lid from printer.C)Press down on side of printer cartridge labeled “EJECT”.
To install new printer ribbon cartridge...
A)Install new cartridge as shown* (it will “snap” into place). *Note: The printer paper mustbe between the cartridge ribbon andthe metal plate.B)Turn Adjustment Knob clockwise until ribbon is tight.
Metal Plate
(Printer Paper)
Ribbon
Adjustment
Knob
Printer Paper
Cartridge
Ribbon
Page 34
Troubleshooting
Error Codes
If a malfunction is detected during a cycle, a numeric error code will appear on the display panel.
Use the chart below to diagnose and correct the most common, maintenance-related error codes.
If you encounter an error code not identied below, follow the instructions on the display panel.
If error code persists, contact your authorized service provider.
Example:
003-1658-99
Error CodeProbable CauseCorrective Action
C010
C100-series (all
(C101, C102, etc.)
C231, C232
C441, C442
C533, C633Water pump needs primed.
Sterilizer lost power
during cycle.
<STOP> button was
pressed during the cycle.
Not enough water in
reservoir to complete
the cycle.
External condensing
tank is full.
Press <STOP> button,
then restart cycle.
Press <STOP> button,
then restart cycle.
Fill reservoir with distilled water,
or water that meets the
referenced water purity specications.
Empty external condensing tank.
Put the sterilizer in the User
Diagnostic Mode and prime
• To set “English” or “Metric” units on the display panel,
• To retrieve the last ve (5) error codes stored in the unit memory.
• To prime the water pump if needed.
To activate User Diagnostic Mode...
A) Turn Power Switch OFF (O).
B) Press and hold the <START> button.
C) Turn Power Switch ON ( I ).
D) Press the <START> button when the display shows “USER DIAGNOSTIC”.
To change Display Units...
A) Put the unit in User Diagnostic Mode.
B) Press the <P> button to select units.
C) Press the <+> button to change the temperature and pressure display from
English-to-Metric or Metric-to-English units. (Factory default setting is English).
D) Press the <START> button to continue.
E) Turn the power switch OFF (O) to exit User Diagnostic Mode.
To retrieve the five (5) most recent error codes...
A) Put the unit in User Diagnostic Mode.
B) Press the <STOP> button to recall errors.
C) The last five (5) error codes will be displayed.
D) Press the <START> button to return to the User Diagnostic Mode display.
E) Turn the power switch OFF (O) to exit User Diagnostic Mode.
To prime the sterilizer pump...
A) Put the unit in User Diagnostic Mode.
B) Press the <START> button to start the pump priming progress.
The unit will automatically cycle through a pre programmed priming cycle...
• Closing the sterilizer door.
• Heating the boiler.
• Cycling the pump ON and OFF until the pump is primed.
When finished the 2nd line of the display will show “PRIMING COMPLETE”.
C) Press the <START> button to return to the User Diagnostic Menu.
D) Turn the power switch OFF (O) to exit User Diagnostic Mode.
Please mark down any displayed Code(s) and be sure to relay this information to the service technician.
Contact your Midmark Authorized Dealer, or log on to www.midmark.com/technical-library.
Model and serial number information will be required when calling for service.
To contact Midmark directly:
1.800.MIDMARK (1.800.643.6275) or 937.526.3662
8 am to 5 pm EST (Monday thru Friday)
[excluding standard US holidays]
F1 .........................................................................15 Amp, 250 V, Fast Acting, 1/4” x 1 1/4”
F2 .........................................................................0.25 Amp, 250 V, Slo-blo, 1/4” x 1 1/4”
230 VAC Unit
F1 .........................................................................8 Amp, 250 V, Fast Acting, 5 x 20 mm
F2 .........................................................................0.125 Amp, 250 V, Slo-blo, 5 x 20 mm
Certications:
ASME Boiler and Pressure Vessel Code, Section VIII, Division 1
Canadian Registration Number Available
UL 61010-1, 2nd Edition
IEC 61010-2-040
CAN/CSA-C22.2 No. 61010-1, 2nd Edition
FCC Part 15, Sub-part B
Physical Dimensions:
Overall Length: .....................................................21 in. (53.3 cm)
Overall Width: ....................................................... 17.8 in. (45.2 cm)
Overall Height: ......................................................7.1 in. (18 cm)
Shipping Carton Length: .......................................25 in. (63.5 cm)
Shipping Carton Width:.........................................22 in. (55.9 cm)
Shipping Carton Height: .......................................16.6 in. (42.2 cm)
Counter Area: .......................................................24 in. (61 cm) deep x 22 in. (55.9 cm) wide
Chamber Volume: .................................................0.49 gal (1.8 liter)
Weight:
Empty Reservoir: .................................................. 71 lbs. (32.2 kg)
Full Reservoir: ......................................................80 lbs. (36.3 kg)
With Shipping Carton:...........................................80 lbs. (36.3 kg)
Water Reservoir Capacity: ....................................1.20 gal (4.5 liter)
SCOPE OF WARRANTY Midmark Corporation (“Midmark”) warrants to the original retail purchaser that it will, at Midmark’s option, repair or
under “Exclusions”) that are defective in material or workmanship under normal use and service. The sole remedy under this limited warran-
and/or “Installation/Operation Manual” for the applicable product, including the speci ed structural and operational environmental conditions
THAT THE SOFTWARE: (1) IS ERROR FREE; (2) CAN BE USED WITHOUT PROBLEMS OR INTERRUPTIONS; OR (3) IS FREE FROM
EXCLUSIVE REMEDY; CONSEQUENTIAL DAMAGES DISCLAIMER MIDMARK’S ONLY OBLIGATION UNDER THIS LIMITED WARRAN-
TIES, EXPRESS OR IMPLIED. MIDMARK MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACE-
STATUTE OF LIMITATIONS No action may be brought against Midmark for breach of this limited warranty, an implied warranty, if any, or for
Warranty Information
Scope of Warranty
replace components of the domestic and international medical products manufactured by Midmark (except for components not warranted
ty is the repair or replacement, at Midmark’s option, of the applicable components. This limited warranty shall only apply to defects that are
reported to Midmark within the applicable warranty period and which are determined to exist upon examination by Midmark,. This warranty
extends only to the original retail purchaser of a product and is not transferable or assignable. Replacement components or products may
be used and/or refurbished components or products, provided they are of like quality and speci cations as new components or products.
Midmark warrants to the original retail purchaser that during the applicable warranty period it will repair or replace software contained
within the products manufactured by Midmark (except for those not warranted under “Exclusions”) if: (1) the media on which the software is
furnished exhibits defects in material or workmanship under normal use; or (2) the software does not substantially conform to its published
speci cations.
APPLICABLE WARRANTY PERIOD The applicable warranty period, measured from the date of invoice to the original retail purchaser of
the product and shall be one (1) year for all warranted products and components.
OBTAINING WARRANTY SERVICE Warranty service must be obtained through either Midmark or an authorized dealer in the Midmark
product line for which warranty service is requested. Midmark may be contacted for warranty service inquiries or issues via email at www.
midmark.com; by phone at 1.800.MIDMARK or by mail to Midmark Corporation, 60 Vista Drive, Versailles, Ohio 45380. It is the retail purchaser’s obligation to arrange for delivery of a product to Midmark or one of its authorized dealers for warranty service, which delivery shall
be at retail purchaser’s expense. It is also the retail purchaser’s obligation to comply with the warranty service instructions provided either
by Midmark or its authorized dealer. The retail purchaser must provide Midmark with completed warranty registration information within
thirty (30) days after purchase in order to obtain the bene ts of this limited warranty.
EXCLUSIONS: This limited warranty does not cover and Midmark shall not be liable for the following:
(1) Defects, damage or other conditions caused, in whole or in part, by misuse, abuse, negligence, alteration, accident, freight damage,
negligent storage, tampering or failure to seek and obtain repair or replacement in a timely manner;
(2) Products which are not installed, used, and properly cleaned and maintained as required or recommended in the Midmark “Installation”
and electrical requirements;
(3) Products considered to be of a consumable or sterile nature;
(4) Accessories or parts not manufactured by Midmark;
(5) Charges by anyone for adjustments, repairs, replacement parts, installation or other work performed upon or in connection with such
products which are not expressly authorized in writing in advance by Midmark;
(6) Costs and expenses of routine maintenance and cleaning; and
(7) Representations and warranties made by any person or entity other than Midmark.
(8) Matching of color, grain or texture except to commercially acceptable standards;
(9) Changes in color caused by natural or arti cial light;
(10) Custom manufactured products;
(11) Alterations or modi cations to the product by any person or entity other than Midmark; and
(12) Products that would otherwise by covered under this limited warranty, but are acquired: (i) from a person or entity that is not Midmark
or one of its authorized dealers; or (ii) from a Midmark dealer that is not authorized to sell the product at issue in the geographic territory
where the purchaser is located, or is not authorized to sell the product at issue within the medical market.
SOFTWARE; WITH RESPECT TO SOFTWARE THAT IS A PRODUCT OR COMPONENT THEREOF, MIDMARK DOES NOT WARRANT
VULNERABILITY TO INTRUSION OR ATTACK BY VIRUSES OR OTHER METHODS.
TY IS THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS. MIDMARK SHALL NOT BE LIABLE FOR AND HEREBY DISCLAIMS
ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, EXEMPLARY OR CONSEQUENTIAL DAMAGES OR DELAYS, INCLUDING, BUT
NOT LIMITED TO, DAMAGES FOR LOSS OF PROFITS OR INCOME, LOSS OF USE, LOSS OF DATA, DOWNTIME, COVER AND
EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES, PAYMENTS AND BENEFITS. THIS DISCLAIMER SHALL SURVIVE ANY
FAILURE OR ASSERTED FAILURE OF THE ESSENTIAL PURPOSE OF THIS LIMITED WARRANTY OR ITS REMEDIES SPECIFIED
HEREIN. WARRANTY DISCLAIMER THIS WARRANTY IS MIDMARK’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRAN-
MENT OF DEFECTIVE PARTS.
any other claim arising out of or relating to the products, more than ninety (90) days following expiration of the limited warranty period.
NO AUTHORIZATION No person or rm is authorized to create or approve for Midmark any other obligation or liability in connection with