Microsonic pico+15-F User Manual

Operating Instructions
Ultrasonic proximity switch with one switched output
pico+15/F pico+15/WK/F pico+25/F pico+25/WK/F pico+35/F pico+35/WK/F pico+100/F pico+100/WK/F
Product description
The pico+sensor offers a non-contact measurement of the distance to an object which must be positioned within the sensor’s detection zone. The switched output is set condi­tional upon the adjusted detect dis­tance. Via the Teach-in procedure, the de­tect distance and operating mode can be adjusted. Two LEDs indicate operation and the state of the swit­ched output. The pico+sensors are IO-Link-capable in accordance with IO-Link specifica­tion V1.0.
Safety instructions
Read the operating instructions
prior to start-up.
Sensor adjustment with Teach-in procedure
Set detect point
– method A
Place object at position
Connect Com for about
3 s to +U
B
, until both
LEDs flash simultaneously
Both LEDs:
flash alternately
Set detect point +8 %
– method B
Place object at position
Connect Com for about
3 s to +U
B
, until both
LEDs flash simultaneously
Both LEDs:
flash alternately
Set window mode
Set two way reflective
barrier
Set NOC/NCC
Place object at position
Install reflector at
position
Connect Com for about
3 s to +U
B
, until both
LEDs flash simultaneously
Both LEDs:
flash alternately
Connect Com for about
3 s to +U
B
, until both
LEDs flash simultaneously
Both LEDs:
flash alternately
Connect Com for about
13 s to +U
B
, until both
LEDs flash alternately
Green LED: Yellow LED:
flashes
on: NOC off: NCC
Switch-over
Teach-in / synchronisation
Switch off operating
voltage
Reset to factory setting
Switch off operating
voltage
Connect Com to -U
B
Switch on
operating voltage
Keep Com connected to
-U
B
for about 3 s, until
both LEDs flash
simultaneously
Green LED:
Yellow LED:
flash
on: Teach-in off: Synchro­nisation
Connect Com to -U
B
Switch on
operating voltage
Keep Com connected to
-U
B
for about 13 s, until
both LEDs stop
flashing
Connect Com for about
1 s to +U
B
Normal operating mode
Connect Com for about
3 s to +U
B
, until both
LEDs flash alternately
Place object at position
Both LEDs:
flash mutually
Connect Com for about
1 s to +U
B
Connect Com for about
10 s to +U
B
To change output
characteristic connect
Com for about 1 s to +U
B
Wait for 10 s
To change operation
mode connect
Com for about 1 s to -U
B
Wait for 10 s
Normal operating mode
Disconnect Com from -U
B
before switching off
supply voltage
Connection, installation and ad-
justments may only be carried out by qualified staff.
No safety component in
accordance with the EU Machine Directive
Use for intended purpose only
pico+ultrasonic sensors are used for non-contact detection of objects.
Installation
Mount the sensor at the place of
fitting.
Connect a connection cable to the
M12 device plug.
Fig. 1: Pin assignment with view onto sensor
plug and colour coding of the microsonic connection cables
Start-up
Connect the power supply.Carry out sensor adjustment in
accordance with the diagram.
1
3
colour
+U
B
-U
B
brown
blue
4 2 5
F
-
black white
Com grey
1
5
2
34
Factory setting
Detect point operationSwitched output on NOCDetect distance at operating rangeMulti-function input »Com« set to
»Teach-in«
Filter at F01Filter strength at P01
Operating modes
Three operating modes are available for the switched output:
Operation with one detect point
Window mode
Two-way reflective barrier
object is between sensor and fixed reflector.
Synchronisation
If under multiple sensor operation the assembly distance falls below the values shown in Fig. 2, the internal synchronisation should be used. For this purpose set the switched out­puts of all sensors in accordance with the diagram »Sensor adjustment with the Teach-in procedure«. Then change the multi-function output »Com« to »synchronisation« (see »Further settings«). Finally intercon­nect each pin 5 of the sensors to be synchronised.
Fig. 2: Assembly distances
Maintenance
Notes
The sensors of the pico+ family
have a blind zone, within which a distance measurement is not pos­sible.
The pico+ sensors are equipped
with an internal temperature com­pensation. Due to the sensors self heating, the temperature compen­sation reaches its optimum work­ing-point after approx. 20 minutes of operation.
In the normal operating mode, an il-
luminated yellow LED signals that the switched output is switched through.
The pico+sensors have a push-pull
switched output.
In the »Two-way reflective barrier«
operating mode, the object has to be within the range of 0-85 % of the set distance.
0,25 m 1,30 m
0,35 m
0,40 m
2,50 m
2,50 m
0,70 m
4,00 m
A
B
Ultrasonic Sensors
Set switched output
Further Settings
Technical data
blind zone
operating range
maximum range
angle of beam spread
20 mm
150 mm
30 mm
250 mm
250 mm
see detection zone
350 mm
see detection zone
transducer frequency
resolution
reproducibility
380 kHz
0.069 mm
320 kHz
0.069 mm
± 0.15 %
± 0.15 %
accuracy
operating voltage U
B
voltage ripple
no-load current consumption
±1 % (temperature drift internally compensated)
10 - 30 V DC, reverse polarity protection
±1 % (temperature drift internally compensated)
10 - 30 V DC, reverse polarity protection
±10 %
< 40 mA
±10 %
< 40 mA
housing
max. tightening torque of nuts
brass sleeve, nickel-plated, plastic parts: PBT;
ultrasonic transducer: polyurethane foam,
brass sleeve, nickel-plated, plastic parts: PBT;
ultrasonic transducer: polyurethane foam,
class of protection per EN 60 529
type of connection
controls
indicators
IP 67
IP 67
Teach-in via pin 5 (Com)
LED green (operation)
Teach-in via pin 5 (Com)
LED green (operation)
programmable
synchronisation
operating temperature
LED yellow (state of output)
Teach-in, LinkControl
LED yellow (state of output)
Teach-in, LinkControl
internal synchronisation up to 10 sensors
-25°C to +70°C
internal synchronisation up to 10 sensors
-25°C to +70°C
storage temperature
switched output
switching hysteresis
1)
-40°C to +85°C
Push-Pull, I
max
= 100 mA
-40°C to +85°C
Push-Pull, I
max
= 100 mA
switchable NOC/NCC, short-circuit-proof
2 mm
switchable NOC/NCC, short-circuit-proof
3 mm
switching frequency
1)
response time
1)
time delay before availability
1)
norm conformity
25 Hz
32 ms
25 Hz
32 ms
< 300 ms
EN 60947-5-2
< 300 ms
EN 60947-5-2
order no. directly radiating
weight
pico+15/F
pico+25/F
30 g
30 g
order no. angular head
weight
1) Can be programmed with LinkControl
pico+15/WK/F
35 g
pico+25/WK/F
35 g
65 mm
350 mm
120 mm
1,000 mm
600 mm
see detection zone
1,300 mm
see detection zone
400 kHz
0.069 mm
200 kHz
0.069 mm
± 0.15 %
± 0.15 %
±1 % (temperature drift internally compensated)
10 - 30 V DC, reverse polarity protection
±1 % (temperature drift internally compensated))
10 - 30 V DC, reverse polarity protection
±10 %
< 40 mA
±10 %
< 40 mA
brass sleeve, nickel-plated, plastic parts: PBT;
ultrasonic transducer: polyurethane foam,
brass sleeve, nickel-plated, plastic parts: PBT;
ultrasonic transducer: polyurethane foam,
IP 67
IP 67
Teach-in via pin 5 (Com)
LED green (operation)
Teach-in via pin 5 (Com)
LED green (operation)
LED yellow (state of output)
Teach-in, LinkControl
LED yellow (state of output)
Teach-in, LinkControl
internal synchronisation up to 10 sensors
-25°C to +70°C
internal synchronisation up to 10 sensors
-25°C to +70°C
-40°C to +85°C
Push-Pull, I
max
= 100 mA
-40°C to +85°C
Push-Pull, I
max
= 100 mA
switchable NOC/NCC, short-circuit-proof
5 mm
switchable NOC/NCC, short-circuit-proof
20 mm
12 Hz
64 ms
10 Hz
80 ms
< 300 ms
EN 60947-5-2
< 300 ms
EN 60947-5-2
pico+35/F
pico+100/F
30 g
30 g
pico+35/WK/F
35 g
pico+100/WK/F
35 g
+U
B
-U
B
F Com
1 2 4 5 3
Push-Pull output in pnp circuit
U
+U
B
-U
B
F Com
1 2 4 5 3
Push-Pull output in npn circuit
U
pico+15...
pico+25...
pico+35...
pico+100...
2004/108/EC
detection zones
The dark grey areas represent the zone
where it is easy to recognise the normal
reflector (round bar). This indicates the typical
operating range of the sensors. The light
grey areas represent the zone where a
very large reflector – for instance a plate – can still be regognized. The
requirement here is for an optimum
alignment to the sensor. It is not
for different objects:
possible to evaluate ultrasonic
reflections outside this area.
In the »Set detect point – method A« Teach-in procedure the actual distance to the object is taught to the sensor as the detect point. If the object moves towards the sensor (e.g. with level control) then the taught distance is the level at which the sensor has to switch the output.
If the object to be scanned moves
into the detection area from the side, the »Set detect point +8 % – method B« Teach-in procedure should be used. In this way the switching distance is set 8 % fur­ther than the actual measured dis­tance to the object. This ensures a
4 cm
0 cm
0 cm
0 cm
0 cm
4 cm
4 cm
4 cm
Rohr ø 10 mm
Rohr ø 10 mm
Rohr ø 10 mm
Round bar ø 10 mm
ausgerichtete Platte
ausgerichtete Platte
Plate
8 cm
8 cm
8 cm
8 cm
0 cm
0 cm
0 cm
4 cm
4 cm
4 cm
8 cm
8 cm
8 cm
12 cm
12 cm
12 cm
16 cm
16 cm
16 cm
20 cm
20 cm
20 cm
24 cm
24 cm
24 cm
10 cm
5 cm
0 cm
5 cm
Plate
Round bar ø 10 mm
10 cm
0 cm
5 cm
10 cm
15 cm
20 cm
25 cm
30 cm
35 cm
20 cm
10 cm
0 cm
10 cm
Plate
Round bar ø 27 mm
20 cm
0 cm
10 cm
20 cm
30 cm
35 cm
40 cm
50 cm
60 cm
0,4 m
0,2 m
0 m
0 m
0,2m
Plate
0,4 m
0,6 m
Round bar ø 27 mm
0,8 m
1,0 m
1,0 m
1,3 m
1,4 m
0,4 m
0,2 m
reliable switching distance even if the height of the objects varies slightly.
Fig. 4: Setting the detect point for different
directions of movement of the object
4 cm
4 cm
8 cm
8 cm
4 cm
4 cm 8 cm
8 cm
epoxy resin with glass content 15 Nm
5-pin M12 circular plug
epoxy resin with glass content 15 Nm
5-pin M12 circular plug
epoxy resin with glass content 15 Nm
5-pin M12 circular plug
epoxy resin with glass content 15 Nm
5-pin M12 circular plug
If synchronization is activated the
Teach-in is disabled (see »Further settings«).
The sensor can be reset to its fac-
tory setting (see »Further set­tings«).
Using the LinkControl adapter (op-
tional accessory) and the LinkCon­trol software for Windows, all Teach-in and additional sensor pa­rameter settings can be optionally undertaken.
IO-Link mode
The pico+ sensors are IO-Link-capa­ble in accordance with IO-Link speci­fication V1.0.
Pointer
In IO-Link mode Teach-in, Link-
Control and synchronization via pin 5 are not available.
In IO-Link mode pin 5 must not be
connected to any potential.
For current information about IO-
Link please contact the microsonic sales department.
Synchronisation in IO-Link mode
Process data
The pico+ cyclically transmits the measured distance value with a reso­lution of 0,1 mm and the state of
the switched output.
Service data
The following sensor parameters may be set via IO-Link interface using the IO-Link device description (IODD).
Detect point 1
The switched output is activated when the distance to an object is un­der that of the present detect point.
Return detect point 1
The switched output is reactivated when the distance to an object is greater than the present return de­tect point (detect point + hysteresis).
Pointer
The return detect point 1 must al-
ways be greater than the detect point 1.
Detect point 2, return detect point 2 By programming these two detect distances the window mode is acti­vated.
Pointer
The return detect point 2 must al-
ways be smaller than the detect point 2.
NOC/NCC operation
The NCC or NOC output function can be present for the switched output.
Measurement filter
F00
No filter, each ultrasonic measure­ment acts in an unfiltered manner on the output.
F01
Standard filter, on the object con­tinuously approaching the sensor, the ongoing interval is immediate­ly taken on and the output correspondingly activated. The ef­fect of the object abruptly moving away from the sensor is for the existing distance to be saved for a retaining time dependent on the filter strength and for the switched
output state to be maintained.
F02
Average value filter, forms the arithmetic mean across a number of measurements. The output is activated in keeping with the ave­rage value. The number of measu­rements, from which the average value is formed, depends on the selected filter strength.
Filter strength
A filter strength between 0 – weak filter effect – and 9 – pronounced fil­ter effect – can be selected for each measurement filter.
Foreground suppression
Spurious reflections, caused by ob­jects in the foreground of the sensor may be blocked out by the fore­ground suppression.
Pointer
Check that the object in the fore-
ground does not cause multiple reflections.
The object in the foreground must
not cover the sensor in a way that the detection zone is influenced.
System commands
With 4 system commands the follo­wing settings may be carried out:
Teach-in detect point – method A.Teach-in detect point – method B.Teach-in two way reflective barrier.Reset sensor to factory settings.
Pointer
To achieve the maximum resolution the Master Cycle Time has to comply with the following requirements:
Min Cycle Time ≤ Master Cycle Time ≤ Min Cycle Time + 1.2 ms.
If this condition can not be full-
filled, sporadic discontinuities of the measurement value can occur. In this case the Master Cycle Time has to be increased in 400 µs steps until the discontinuities of the measurement disappear.
Pointer
If the pico+ sensor was set using
Teach-in or LinkControl it is re­commended to reset the sensor to the factory setting prior to using it in IO-Link mode (s. »Further settings«).
IODD file
The latest IODD file you will find on
the internet under
www.microsonic.de/en/IODD.
For further informations on IO-Link see www.io-link.com.
index
index
format
UINT16
UINT16
UINT16
UINT16
index
format
UINT16
UINT16
UINT16
UINT16
format
UINT16
466-65,512 (32 - 350 mm) 1)
> 5,094: window mode deactivated
451-65,512 (31 - 349 mm) 1)
UINT16
UINT16
UINT16
format
UINT16
975-65,512 (67 - 600 mm) 1)
> 8,733: window mode deactivated
961-65,512 (66 - 599 mm) 1)
UINT16
UINT16
UINT16
> 5,094: window mode deactivated
> 8,733: window mode deactivated
1,776-65,512 (122 - 1,300 mm) 1)
> 18,922: window mode deactivated
1,761-65,512 (121 - 1,299 mm) 1)
> 18,922: window mode deactivated
pico+15...
pico+35...
pico+100...
IO-Link mode
MV-DO-108322-372394
microsonic GmbH | Hauert 16 | 44227 Dortmund | Germany | telefone +49 2 31 / 97 51 51-0 | telefax +49 2 31 / 97 51 51-51 | e-mail: info@microsonic.de | www.microsonic.eu The content of this document is subject to technical changes. Specifications in this document are presented in a descriptive way only. They do not confirm any product features.
IO-Link data
physical layer
SIO mode support
min cycle time
format of process data
content of process data
service data IO-Link specific
service data sensor specific
return detect point 1
return detect point 2
foreground suppression
Teach-in via Pin 5 in SIO mode
Teach-in detect point – method A
Teach-in detect point – method B
Teach-in two way reflective barrier
reset to factory settings
baud rate
Vendor name
Vendor text
Product name
Product ID
Product text
detect point 1
detect point 2
switching mode
filter strength
system commands
yes 8 ms COM 2 (38.400 Bd) 16 Bit, R, UNI16 Bit 0: state of switched output; Bit 1-15: distance value with 0,1 mm resolution
0x10 0x11 0x12 0X13 0x14
0x40 0x41 0x47
0x48
0x42 UINT8 0x43
filter
0x44 0x49 0x4A
0x02 0x02 0x02 0x02
1)
Distance values, e.g. detect points, are given as multiple of the internal resolution of the measurement value = 0,069 mm (example: 320 22 mm). The values in the table are decimal.
access value
RRmicrosonic GmbH
RRpico+
R Ultraschall-Sensor
access
R/W R/W R/W
R/W
R/W
UINT8
R/W
UINT8
R/W
UINT8
R/W R/W
UINT8
access value
WW161
WW164
www.microsonic.de
15/F;15/WK/F
range (dez)
306-3,609 (21-248 mm) 1) 320-3,624 (22-249 mm) 1) 335-65,512 (23 - 250 mm) 1) > 3,638: window mode deactivated 320-65,512 (22 - 250 mm) 1) > 3.638: window mode deactivated 00: NCC, 02: NOC 00-02: F00 - F02 00-09: P00 - P09 0-1,878 (0-129 mm) 1) 00: deactivated, 16: activated
162
168
yes 8 ms COM 2 (38.400 Bd) 16 Bit, R, UNI16 Bit 0: state of switched output; Bit 1-15: distance value with 0,1 mm resolution
index
0x10 0x11 0x12 0X13 0x14
index
0x40 0x41 0x47
0x48
0x42 UINT8 0x43 0x44 0x49 0x4A
index
0x02 0x02 0x02 0x02
access value
RRmicrosonic GmbH
RRpico+
R Ultraschall-Sensor
access
R/W R/W R/W
R/W
R/W
UINT8
R/W
UINT8
R/W
UINT8
R/W R/W
UINT8
access value
WW161
WW164
pico+25...
www.microsonic.de
25/F;25/WK/F
range (dez)
436-5,065 (30 - 348 mm) 1) 451-5,080 (31 - 349 mm) 1)
00: NCC, 02: NOC 00-02: F00 - F02 00-09: P00 - P09 0-3,246 (0-223 mm) 1) 00: deactivated, 16: activated
162
168
yes 16 ms COM 2 (38.400 Bd) 16 Bit, R, UNI16 Bit 0: state of switched output; Bit 1-15: distance value with 0,1 mm resolution
index
0x10 0x11 0x12 0X13 0x14
index
0x40 0x41 0x47
0x48
0x42 UINT8 0x43 0x44 0x49 0x4A
index
0x02 0x02 0x02 0x02
access value
RRmicrosonic GmbH
RRpico+
R Ultraschall-Sensor
access
R/W R/W R/W
R/W
R/W
UINT8
R/W
UINT8
R/W
UINT8
R/W R/W
UINT8
access value
WW161
WW164
www.microsonic.de
35/F;35/WK/F
range (dez)
946-8,704 (65 - 598 mm) 1) 961-8,718 (66 - 599 mm) 1)
00: NCC, 02: NOC 00-02: F00 - F02 00-09: P00 - P09 0-4,236 (0-291 mm) 1) 00: deactivated, 16: activated
162
168
yes 20 ms COM 2 (38.400 Bd) 16 Bit, R, UNI16 Bit 0: state of switched output; Bit 1-15: distance value with 0,1 mm resolution
index
0x10 0x11 0x12 0X13 0x14
index
0x40 0x41 0x47
0x48
0x42 UINT8 0x43 0x44 0x49 0x4A
index
0x02 0x02 0x02 0x02
access value
RRmicrosonic GmbH
RRpico+
R Ultraschall-Sensor
access
R/W R/W R/W
R/W
R/W
UINT8
R/W
UINT8
R/W
UINT8
R/W R/W
UINT8
access value
WW161
WW164
www.microsonic.de
100/F;100/WK/F
range (dez)
1,747-18,892 (120 - 1,298 mm) 1) 1,761-18,907 (121 - 1,299 mm) 1)
00: NCC, 02: NOC 00-02: F00 - F02 00-09: P00 - P09 0-12,969 (0-891 mm) 1) 00: deactivated, 16: activated
162
168
Loading...