Microsonic crm+25-IU-TC-E User Manual

T1 + T2
Press T1 and T2 simultaneously for about 3 s until welcome message has passed
T1 + T2
T2
T1
T2
T1
T2
T1
T1 + T2
Set sensor-close window margin in mm or cm
Choose rising ( ) / falling ( ) output characteristic curve
T1
T2
T1 + T2
T2
T1
T1 + T2
T1 + T2
Set analogue output
Set sensor-distant window margin in mm or cm
Ready
Start here
Set sensor parameters alternatively numerically using LED-display...
Product description
The crm+sensor with one analogue out-
put measures the distance to an object within the detection zone contactless. A signal proportional to distance is created according to the adjusted window margings of the analogue characteristic curve.
The ultrasonic transducer surface of the
crm+sensors is laminated with a PEEK film. The transducer itself is sealed against the housing by a PTFE joint ring. This composition ensures a high resitance against many aggressive substances.
The sensor automatically detects the load
put to the analogue output and switches to current output or voltage output re­spectively.
All settings are done with two push-but-
tons and a three-digit LED-display (TouchControl).
Light emitting diodes (three-colour LEDs)
indicate all operation conditions.
Choosing between rising and falling out-
put characteristic is possible.
The sensors are adjustable manually using
the numerical LED-display or may be trai­ned using Teach-in processes.
Useful additional functions are set in the
Add-on-menu.
Using the LinkControl adapter (optional
accessory) all TouchControl and additio­nal sensor parameter settings may be made by a Windows-Software.
Safety Notes
Read the operating instructions prior to
start-up.
Connection, installation and adjustment
works may only be carried out by expert personnel
.
No safety component in accordance with
the EU Machine Directive.
The crm+sensors indicate a blind zone, in which the distance cannot be measured. The operating range indicates the distance of the sensor that can be applied with normal re­flectors with sufficient function reserve. When using good reflectors, such as a calm water surface, the sensor can also be used up to its maximum range. Objects that strongly absorb (e.g. plastic foam) or diffusely reflect sound (e.g. pebble stones) can also reduce the defined operating range.
Synchronisation
If the assembly distances shown in Fig.1 for two or more sensors are exceeded the inte­grated synchronisation should be used. Con­nect Sync/Com-channels (pin 5 at the units re­ceptable) of all sensors (10 maximum).
Fig. 1: Assembly distances, indicating syn-
chronisation/multiplex
Multiplex mode
The Add-on-menu allows to assign an indivi­dual address »01« to »10« to each sensor connected via the Sync/Com-channel (Pin5). The sensors perform the ultrasonic measure­ment sequentially from low to high address. Therefore any influence between the sensors is rejected. The address »00« is reserved to synchronisa­tion mode and deactivates the multiplex
0.35 m
0.40 m
2.50 m
2.50 m
1.10 m
2.00 m
8.00 m
18.00 m
4.00 m 30.00 m
A
B
mode. (To use synchronised mode all sensors must be set to address »00«.)
Assembly instructions
Assemble the sensor at the installation lo-
cation.
Plug in the connector cable to the M 12
connector
Fig. 2: Pin assignment with view onto sen-
sor plug and colour coding of the microsonic connection cable
Start-up
crm+ sensors are delivered factory made with the following settings: Rising analogue characteristic
1 3
colour
+U
B
-U
B
brown
blue 4 2 5
-
I/U
black
white
Sync/Com. grey
1
5
2
34
Window margins for the analogue out-
put set to blind zone and operating range
Measurement range set to maximum ran-
ge
Set the parameters of the sensor manually or use the Teach-in procedure to adjust the de­tect points.
Fig. 3: TouchControl
Operation
crm+sensors work maintenance free. Small amounts of dirt on the surface do not influ­ence function. Thick layers of dirt and caked­on dirt affect sensor function and therefore must be removed.
Note
As a result of the design the assembly of
PEEK film and PTFE joint ring is not gas­proof.
The chemical resistance has to be tested
experimentally if necessary.
crm+sensors have internal temperature
compensation. Because the sensors heat up on their own, the temperature com­pensation reaches its optimum working point after approx. 30 minutes of opera­tion.
If an object is within the set window mar-
gins of the analogue output, then LED D1 lights up green, if the object is outside the window margins, then LED D1 lights up red.
The load put to the analogue output is
detected automatically when turning supply voltage on.
During normal mode operation, the mea-
sured distance value is displayed on the LED-indicator in mm (up to 999 mm) or cm (from 100 cm). Scale switches auto­matically and is indicated by a point on top of the digits. Alternatively a percent­age scale may be set in the add-on menu. In this connection 0% and 100% corre­spond to the set window margins of the analogue output.
If no objects are placed within the detec-
tion zone the LED-indicator shows »- - -«.
If no push-buttons are pressed for 20 se-
conds during parameter setting mode the made changes are stored and the sensor returns to normal mode operation.
Show parameters
Tapping push-button T1 shortly during nor­mal mode operation shows »PAr« on the LED-display. Each time you tap push-button T1 the actual settings of the analogue output are shown.
Instruction manual
crm+ Ultrasonic Sensors with one analogue output
c
m m m %
T 1 D 1 D 2 T
2
measuring range
LED D1 and D2
3-digit LED-display
Push-buttons T1 and T2
crm+25/IU/TC/E crm+35/IU/TC/E crm+130/IU/TC/E crm+340/IU/TC/E crm+600/IU/TC/E
2004/108/EEC
Ultrasonic Sensors
...or with the Teach-in procedure
Set window margins
Place object at position
Press T1 until »IU« is shown
Place object at position
Current measuring value
Press T1 until »End« is shown
Current measuring value
Set rising/falling output
characteristic curve
Press T1 until countdown passed
from »- 8 -« to »- 0 -« and symbol
for rising/falling characteristic
is displayed
Sybol rising/falling characteristis
To change output characteristic
press T1
Press T1 and T2 simultaneously until
»End« is displayed
Sybol rising/falling characteristis
Normal mode operation
T1 + T2
T1 + T2
T1 + T2
T1 + T2
T1 + T2
T1 + T2
T1 + T2
T1 + T2
T1 + T2
T1 + T2
T1 + T2
T1 + T2
Teach-in analogue output
Key lock and factory setting
Activate/deactivate
TouchControl
Reset to factory
setting
Turn supply voltage
OFF
While pressing T1 turn
supply voltage ON
until »on« or »off« is
displayed
Turn supply voltage
OFF
While pressing T1 turn
supply voltage ON for
ca. 15 s until «rESET» has past through the
display
To activate or
deactivate press T1
»on« or »off«
To activate or
deactivate press T1
»on« or »off«
Normal mode operation
Usefull additional functions in Add-on menu (for experienced users only, settings not required for standard applications)
Start here
T1 + T2
T1
T1
T1 + T2
T2
T2
T1 + T2
Ready
T1
»C01«: Display bright
»C02«: Display dim­med
»C03«: Display off
Low power mode Display mode Choose current/
»- - -«: Display in mm or. cm
» «: Display in %, 100% at mini-
mum measured va­lue
» «: Display in %, 100% at maxi-
mum measured va­lue
T2
Note
Changes in the Add-on menu may impair the sensor function. A6, A7, A8 , A10 , A11, A12 have influence on the response time of the sensor.
Press T1 and T2 simultaneously for about 13 s until »Add« is shown in the LED-display
T1
T2
T1
T2
T1
T1 + T2
T1
»Aut«: automatic detection of the load
»U«: voltage output
»I«: current output
voltage output
T1 + T2
T2
T1
T1 + T2
T1
T2
»F00«: no filter
»F01«: standard fil­ter
»F02«: averaging fil­ter
»F03«: foreground filter
»F04«: background filter
Measurement filter Filter strength Foreground
T2
T2
T1
T1
T1 + T2
T2
Defines the strength of the chosen filter.
»P00«: weak filter up to
»P09«: strong filter
T1
T1
T2
Delay in seconds between the detec­tion of an object and the output of the measured dis­tance in case of ob­ject approach (beha­ves as on-delay).
"00": 0 s (no delay) up to
"20": 20 s response time
Response time
T1 + T2
T2
T2
T1
T1
T1 + T2
T2
Minimum value: blind zone
Maximum value: ne­arwindow limit - 1
suppression
T1
T1
T2
»00«: synchronisati­on
»01« to »10«: sensor address for multi­plex mode
»oFF«: synchronisati­on deactivated
Multiplex mode
device addressing
T1 + T2
T2
T2
T1
T1 + T2
T1
To optimize multi­plex speed the hig­hest sensor address may be set.
Setting range »01« to »10«
Multiplex mode highest address
T1
T2
T1
T2
T1 + T2
T2
T2
Minimum value: sen­sor-distant window margin
Maximum value: 999 mm for crm+25/... and crm+35/... and 999 cm for all other typs
Measurement range Calibration
T1
T1 + T2
T1
Put plane reflector vertically disposed in front of sensor: in an exact distance of 250 mm for crm+ 25/... and crm+35/... and 900 mm for all other typs. Adjust display to 250 mm or 900 mm. Confirm calibration with T1 + T2.
display
T2
T2
T1
T1 + T2
T1
Affects the size of the detection zone.
»E01«: high
»E02«: standard
»E03«: slight
Detection zone
sensitivity
T2
T2
Loading...
+ 1 hidden pages