Micronics 204 User Manual

PORTAFLOW 204 Plus!
Flowmeter
Operating Manual
Loudwater, High Wycombe, Bucks. HP10 9QR
TEL: +44 (0) 1628 810456 FAX: +44 (0) 1628 531540
e-mail
sales@micronicsltd. co.uk
Web site www.micronicsltd.co.uk
CONTENTS Page No
System Component Pack
1
Introduction 1
Transducer Mounting
2 Programming – Main Menu 4 Main Menu – Quick Start 4 Main Menu – View/Edit site data 6 Main Menu – Select sensor set 7 Main Menu – Set up Instrument 7 Keypad Options – Pulse, 4-20mA, RS232, Logger, Options 8 Status/Error/Warning messages 10 Application and Performance 12 Flow Range 12 Specification 13 Warranty 14 CE Marking 14 Liquid Sound Speeds 15 Battery Charge Circuit Operation 19 Product Care and Maintenance 21
1. Portaflow Electronics 855-0016A
2. Sensor Assembly complete 855-0015A
3. Interconnecting Cables 740-0007
4. Ultrasonic Couplant 292-0020
5. Carrying Case 855-0013
6. Operating Manual 855-1000
7. Rechargeable Battery Pack 740-0005
System Component Pack
204 Plus!!
PORTAFLOW 204 Plus!!
INTRODUCTION
The PORTAFLOW 204 Plus!!! non-invasive Ultrasonic portable flow meter has been designed to help Service, Maintenance and Commissioning Engineers make quick, accurate flow readings of any liquids in pipes from 13mm internal diameter to 115mm outside Diameter.
These compact, rugged instruments will measure flow rate in litres/second, litres/minute, gallons/minute, US gallons/minute, m as litres, gallons, US gallons and m
Simple to set up, the PORTAFLOW is able to accurately measure the flow of clear or cloudy liquids from 0.3 metres per second up to 8 metres per second, in any pipe material and over a temperature range of -20°C to +125°C.
Programming is menu driven with the user supplying OD, wall thickness, pipe material and pipe temperature information when measuring water. Liquid sound speed data is required when measuring other fluids. It is possible to program the instrument and mount the transducers in under 2 minutes, with stable flow data becoming available immediately. The unit is powered by a rechargeable battery pack included in the instrument, which gives an operating life of 8-10 hours from full charge. For continuous operation the meter can be operated from the mains, via the battery charger.
3
/hour, metres/second or feet/sec. The total is displayed
3
.
1
TRANSDUCER MOUNTING
1. OPERATION COMMENT
1.1 Remove oxidization, bitumen or rubber from pipe surface where the transducer assembly is to be mounted.
Mounting transducer assembly.
1.2 Turn the two knurled lock nuts on the rail assembly (figure 3a), CLOCKWISE to withdraw the transducers into the guide rail and then apply ultrasonic couplant as shown in figure 2, page 3.
1.3 Position transducer assembly
Very thick asbestos, concrete, old porous cast iron tubes and steel pipes with scaled or badly corroded internal surfaces can weaken the signal and prevent the unit from operating correctly.
Straight pipe lengths either side of the flow sensors should be 20 diameters UPSTREAM and 10 diameters DOWNSTREAM. See figures 1a and 1b, page 3.
The guide rail assembly must be in onto prepared section of pipe as recommended at 1.1. The fixed transducer (Blue) should be DOWNSTREAM from the floating transducer (Red). Strap to pipe securely using the ball chain attached to the rail.
1.4 Turn the knurled lock nut for the fixed transducer ANTI­CLOCKWISE to contact pipe surface.
1.5 Connect transducer assembly to the handset with the cables
alignment with the pipe axis and positioned as shown in figure 3b when attached to a horizontal pipe. If the sensor cables are connected in reverse the instrument will display a negative flow rate, but this will not affect the accuracy of the reading.
Do not over tighten. Do not lock the floating transducer until the ‘separation distance’ has been determined after programming the unit.
The Portaflow is now ready to program.
provided.
2
Fig 1a. Fig 1b.
0d
CORRECT INCORRECT
DN
UP
DN
UP
>10d
DN
FLOW
>20d
>20d
UP
FLOW
<10d
FLOW
<20d
<2
UP
DN
FLOW
Fig 2. Inverted Plan View of Transducer Assembly
Ultrasonic
couplant
Ultrasonic couplant
2mm
20mm
Fig 3a. Guide Rail Assembly
Connectors
Knurled Knobs
Fig 3b
Mount Transducer Assembly as shown
45°
Transducer Assembly
Separation
Distance
Sensor blocks
Pipe
3
PROGRAMING-MAIN MENU
Switch 0n
Micronics Ltd.

PORTAFLOW

204+3 V1.04
Press Enter to start

Main Menu

Press SCROLL up or down to move cursor to the required option and press ENTER to select.
MAIN MENU
Quick start View/Edit Site Data Sensor set Data Logger Set up RS232 Set up Instrument Read flow

Main Menu - Quick Start

Selecting quick start allows the user to start entering application information. If the instrument has been used previously, it stores the last application data entered. If the unit is to be used on the same application the user can scroll immediately to Read flow on the menu, and press enter without spending time entering new data.
If the unit is to be used on a new application select QUICK START and proceed with the following routine. Use the scroll keys to select, and then press
ENTER.
QUICK START
Dimension units? mm Inches
The instrument now asks for the Pipe outside diameter? After entering the outside diameter in millimetres press
ENTER.
QUICK START
Dimension units mm Pipe O.D.? 58.0
Pipe wall thickness now appears on the display. Enter the pipe wall thickness in millimetres, then press ENTER.
QUICK START
Dimension units MILLIMETRES Pipe O.D.? 58.0 Wall thick? 4.0
Pipe lining thickness now appears on the display. If the pipe you are measuring has a lining, enter the Pipe lining thickness. If nothing is entered the instrument automatically assumes there is no lining. Press ENTER to move on.
QUICK START
Dimension units MILLIMETRES Pipe outside diameter? 58.0 Wall thick? 4.0 Lining? 0.0
The instrument now displays Select pipe wall material. Using the scroll keys select from the options available and press
ENTER.
QUICK START
Select pipe wall material: Mild Steel S’ less Steel 316 S’ less Steel 303 Plastic Cast Iron Ductile Iron Copper Brass Concrete Glass Other (m/s)
Note: The following will only be displayed if a lining thickness had been entered previously. Use the scroll keys to select the required material, and then press ENTER. If Other is selected, enter the sound speed of the lining in metres/sec.
QUICK START
Select pipe lining material: Steel Rubber Glass Epoxy Concrete Other (mps)
4
Select fluid type now appears on the display. Use the scroll keys to select the fluid type and press ENTER.
If the liquid you are measuring is not listed select Other and enter a liquid sound speed in metres/second. The sound speed information can be found in the back of the manual under Liquid Sound Speeds.
QUICK START
Select fluid type: Water Glycol/water 50/50 Lubricating oil Diesel oil Freon Other (m/sec)
The instrument will now display the screen below and provide the user with details of the mode of operation and the maximum flow rate that can be achieved from the information entered. At this stage use the keypad to check maximum volumetric flows. If the display reads double bounce or treble bounce the unit has calculated a larger separation distance, but the transducers should still be set up in reflex mode.
Attach sensor set in REFLEX mode
Approx. max. flow:
X.XX m/s
ENTER to continue
SCROLL changes mode
Select ENTER and the display will now ask you to enter a temperature.
Enter the application temperature and press ENTER. The display will now display a sensor separation distance.
Follow the steps below to attach the guide rail and transducers to the pipe.
1. Take the guide rail and apply
2. Turn thumbscrews on the top of
ENTER to continue
grease to the sensors as shown on page 3 fig 2.
the guide rail clockwise to retract the sensors back up into the guide
Set sensor
Separation to
XXX
rail. This will keep the grease away from the pipe until the guide rail is attached.
3. Now strap the guide rail to the pipe.
4. Screw down the fixed transducer and slide the moveable transducer to the required separation distance (front edge of block), and screw down on to the pipe.
5. Connect the RED and BLUE sensor cables, between the guide rail and the electronics.
6. Press ENTER to read flow.
The flow reading now appears on the display.
Batt CHRG
Sig 48%
(ERROR MESSAGES APPEAR HERE)
m/s
When reading volumetric flow the instrument will display a positive and negative total. Selecting OPTIONS from the keypad can reset these totals (See page 8).
The instrument will continually display the battery and signal levels. Signal levels should be above 30%, to obtain an accurate reading.
If there is an error with the site data entered or the application, the instrument will display an Error or warning message (See page 10), which will appear above the flow reading. If there is more than one message it will continually scroll between them.
To stop reading flow, press ENTER ONCE. The display will read the following.
This will stop all
outputs
Press ENTER to EXIT
SCROLL to return
to READ FLOW
Pressing ENTER a second time will stop outputs and return the instrument to the MAIN MENU.
Press the scroll key to return the instrument to READ FLOW.
5

Main Menu - View/Edit Site Data

The VIEW/EDIT SITE DATA mode can be accessed from the main menu. It allows the user to enter application details for up to 20 different sites. This facility is useful if a number of sites are being monitored on a regular basis. Application data can be programmed into each site before getting there! This also allows you to view the last data entered and edit if required.
When scrolling up/down the menu press
ENTER to select at each prompt.
VIEW/EDIT SITE DATA
List sites Site number 0 Name QUICK START Units MM Pipe O.D. 58.0 Wall thick 4.0 Lining 0.0 Wall MILD STEEL Lining --------­Fluid WATER Read flow Exit
Note:
Site Zero is always the QUICK START data and cannot be changed.
Changing the data in any site is automatically saved when leaving this menu. Data will have to be re-entered to over ride the old data.
View/Edit Site Data - List Sites
Selecting LIST SITES allows the user to view the names of up to 20 sites, numbers 1-5 appear first. Pressing ENTER will display sites from 6-
10. Pressing again will display sites 11-15,
and again to display 15-20.
1 site not named 2 site not named 3 site not named 4 site not named 5 site not named
Press ENTER to continue
View/Edit Site Data - Site Number Site number allows the user to enter the
number of the site data that you wish to be displayed. If the site has not been used then no data would be stored. You can now enter new application data.
View/Edit Site Data - Site Name Site name allows the user to edit or enter
a site name. Use the scroll keys to move the cursor to the letter/figure required and
press ENTER to select. Press zero, to return the instrument to VIEW/EDIT SITE DATA. The new site name will appear on the display.
View/Edit Site Data - Units Dimension units allow the user to switch
between millimetres and inches. The software converts all the application data in a particular site.
View/Edit Site Data - Pipe wall/lining thickness and Pipe wall/lining material
can now be changed as required. Lining material is ignored if a lining thickness has not been entered. A selection of pipe wall/lining materials will be displayed when these options are selected.
View/Edit Site Data - Fluid type Fluid type allows the user to scroll
through a selection of fluid types. Select OTHER in the menu if a liquid is not mentioned. When Other (m/s) is selected the user must enter the liquid sound speed in m/s. This can be supplied by Micronics or found in the back of the manual under Liquid Sound Speeds.

View/Edit Site Data - Read Flow

Selecting Read flow informs the user of the mode of operation and the approximate maximum flow rate. Press the appropriate key can change the units required.
Pressing ENTER asks the user to enter a temperature in °C.
Fluid temp? 20.0 (°C)
SCROLL & ENTER
Select for space, 0 to end
abcdefghijklmnopqr
stuvwxyz01234567890
>……...............<
Attach sensor set in REFLEX mode
Approx. max. flow:
X.XX m/s
ENTER to continue
SCROLL changes mode
6
Enter the temperature and press ENTER. The instrument will display the separation distance before displaying flow. Follow the instructions on the screen and the unit will read flow.
Main Menu – Sensor set
When the application information is programmed into the instrument it calculates the optimum separation distance for the transducers in reflex mode, double bounce reflex or triple bounce reflex.
Sensor Mode
All methods displayed are reflex mode, but on smaller pipes the unit will calculate the separation distance in Double or Triple Bounce Reflex. Should the actual flow be higher than the one specified on the instrument, another mode of operation can be selected, but is very unlikely.
Read Flow
Moving the cursor to Read flow and pressing ENTER informs the user of the mode of operation and the maximum flow capable. Selecting EXIT will take you back to MAIN MENU.

Main Menu – Data Logger

Not available on this model.

Main Menu – Setup RS232
Not available on this model.
Main Menu – Set up instrument Set up Instrument - Calibrate 4-
20mA
(Note: A meter is required to
measure the output.) The 4-20mA Output is calibrated before it leaves the factory and also allows the user to adjust the calibration to match a specific display. The DAC value is a number between 0 and 40,000 and is a number internal to the Portaflow that will change when calibrating the 4-20mA.
The first stage is to adjust the output current to 4mA. When connected to any device that accepts 4-20mA, it may require adjustment to exactly 4mA or 20mA and this is possible by using the scroll keys or keys 5 and 6. The scroll keys move the DAC value in larger steps of 25 and keys 5 & 6 move the value one at a time.
The DAC value will be approximately 8000 for 4mA and 40000 for 20mA. By watching the actual current value displayed on the meter, it is possible to scroll up and down or use keys 5 and 6 to calibrate the 4­20mA to the exact value.
When the 4mA is adjusted press ENTER. Now adjust the 20mA in the same way. Press ENTER when complete and the display will return to the SETUP Instrument menu. If the 4-20mA is not connected then the instrument will still display the DAC number but display Error instead of OK.
yy-mm-dd hh:mm:ss
Adjust to 4mA
Use UP/DOWN to set,
5/6 to trim
DAC value: 8590
mA OK/ERROR
Press ENTER when done
Set up instrument - Display backlight
Use the scroll key to select backlight and press ENTER. This allows the user to enable or disable the backlight. Enable, means the backlight will stay on for 15secs with every key press. It will stay on permanently with the mains plugged in. Use the scroll key to select and press ENTER. The backlight will draw power from the batteries and reduce the operating life of the battery cell. (follow
the application note at the back of the manual)
Set up instrument - Application Options
Contact Micronics.
Set up instrument - Sensor Parameters
Micronics use only. Set up instrument - Factory Settings
Micronics use only.

Set up instrument - Exit

Means EXIT and will take you back to the Main Menu.

Main Menu - Read Flow

When choosing the Read flow option from the MAIN MENU the instrument defaults directly to the data that was last entered. The instrument will have to be reprogrammed if it is to be used on a new application.
7

KEYPAD OPTIONS

The output options can only be adjusted and operated in flow mode.

Pulse Output Key

Use the scroll key to move the cursor up or down the display. To change the flow units press the key required. This will also change the flow units when returning to the flow mode. Changing the flow units will also re-scale the litres per pulse.
PULSE OUTPUT
Flow units Output OFF Max. pulse rate1 per sec Litres per pulse XXXXX Exit
Outputs allow the user to select from the following.
1. Selecting Off switches the pulse
off and returns to the PULSE OUTPUT display.
2. Selecting the Forward total
counts the pulses of the forward flow only.
3. Selecting Net total counts the
pulses of the sum of the forward total less the reverse total.
OUTPUT
Off Forward total Net total
Max. Pulse Rate This option allows the user to select between fast/slow pulses or large/small pulse width. Select 1 per second for slow pulses and 100 for a fast pulse. The pulse width for 1 per second is 100ms and 5ms for 100 per second.
XXXX per pulse This will change when the flow units are changed above. When the correct flow units are selected this allows the user to scale the pulses to there own requirements or it can be left in the default setting.
4 - 20mA Output KEY
The 4-20mA Output can be scaled to the minimum and maximum flow rate. It is also possible to enter a negative figure for the minimum output and would enable a reverse flow to be monitored. For example the 4mA would be the maximum reverse
flow (e.g. –100 lpm) and the 20mA would be maximum positive flow (e.g. +100 lpm).
mA Out This displays the current output at any particular time.
yy-mm-dd hh:mm:ss 4-20MA Units l/m Flow at max. xxx Flow at min. xxx mA for error 22.0 Exit
Output This option allows the user to select between three different outputs or switching the output off. The display will read as follows. Scroll down the options to select required output, and press ENTER. The display will then revert back to the 4- 20mA menu and Flow at max. output.
OUTPUT OFF 4 - 20mA 0 - 20mA 0 - 16mA
The Units The flow units can be changed at this stage by selecting them from the keypad. When selected, scroll down to move onto the next option.
Flow at Max. Output This sets the output at the top end of the scale so that the maximum flow gives 20mA (or 16mA).
The instrument automatically defaults to the maximum flow rate. The user can press ENTER and set the output to a level required. When selected press ENTER to continue.
If the flow was over the maximum range set, the instrument will go to a maximum of
24.4 mA and stay there until either the flow reduces or the output is re-scaled. The instrument will also display a warning message- mA out over range-if the output is greater than 20mA or 16mA.
Flow at Min. Output This sets the output at the bottom end of the scale so that the minimum flow gives 4mA or 0mA. The instrument automatically defaults to zero, but the user is able to enter any
yy-mm-dd hh:mm:ss
8
figure they wish including a minus figure for reverse flow conditions.
Output mA For Error This gives an error output to inform the user of loss of signal. The figure set to between zero and 24mA, but defaults to 22mA.
Exit
RS232 Key
Not available on this instrument.
Logger Key
Not available on this instrument.

Delete Key

If anything is entered in error, press the DELETE key and re-enter the information required.
Options Key
This can only be used in flow mode. Scroll down the options then press ENTER to select.
OPTIONS
Cutoff (m/s) 0.05 Set zero Total RUN Reset + total Reset – total Damping (sec) 5 Cal Factor 1.00 Corr Factor 1.00 Diagnostics Exit
Options key - Cut Off (m/s)
The instrument has an automatic CUTOFF which defaults to 0.05 m/s. Micronics cannot guarantee measuring flows below this range due to the nature of the applications, installation, and instabilities in the measuring system.
Adjusting the cutoff allows the user not to see or record any flow below that figure. For example it may be that the user may not want to measure flows below 50 LPM in a 50mm pipe which is equivalent to 0.42 m/sec. In this case 0.42 m/sec would be entered into the instrument and nothing would be recorded below that level. The maximum cut off is 1 m/sec.

Options key – Set zero flow

On some applications and in some conditions it may be possible that although there is no flow the instrument may show a small offset due to “noise” or “ringing”. The offset can be cancelled out and will increase the accuracy of the instrument.
Selecting this option and pressing ENTER the display will show the following.
Stop the flow
COMPLETELY and
press ENTER or
SCROLL to cancel
Pressing ENTER before the flow has stopped will result in a message asking if you are you sure the flow has stopped. This occurs when the flow is still above
0.25m/sec. When this facility has already been selected, press ENTER to cancel the previous instruction, it is then possible to re-set the Zero balance. The option is not available when error messages E1 and E2 (See 10) are being displayed.
Set up instrument - Total
This option allows the user to disable the positive and negative totalizers. When you select either of these options the totaliser will start or stop functioning. It does not zero the total, this is a separate function described below.
Set up instrument - Reset + Total and - Total
The Portaflow 204 Plus! has a forward and reverse totaliser that can be reset. Use the scroll keys to select and press ENTER to reset. The Total is stored when the unit is switched off or battery goes flat.
Set up instrument - Damping (sec)
Damping maybe used when the flow readings are unstable due to turbulence caused by obstructions or bends etc. Damping or averaging can be used to make the readings more stable. It can be set to up-date the display, anything between 3 and 100 seconds.
Set up instrument - Calibration Factor
This allows the user to calibrate the unit for a specific application. If for example the flow reading is 4% higher than expected entering 0.96 will reduce the reading by 4%. If the reading is 4% lower than expected then entering 1.04 would increase the reading by 4%.
When the instrument is supplied it will always default to 1.00 and when this is changed it will stay in the memory, until such time as it needs to be changed again.
9
Set up instrument - Correction Factor
This is a facility that can be used when errors occur due to lack of straight pipe or the sensors have been placed too close to a bend, this could give an incorrect reading to what is expected. The user can set this as a % in the same way as the calibration factor, but it will not be stored in the memory, when the unit is switched off.
Set up instrumentDiagnostics
The following list allows Micronics to see if the instrument is functioning correctly, and if the signals are being sent and received in the correct way.
Calculated μs
This is a value the instrument predicts will be the time in μsecs that it should take for the transmitted signal to go across a particular pipe size. This value is ascertained from the data entered by the user. i.e. Pipe size, material, sensor set etc.
Up μs, Dn μs
This is the actual transit time measured by the instrument and will be slightly (5-10μs depending on the pipe size and signal condition) less than the calculated value above.
Measurement μs
A point in the signal transmitted, where the flow measurement is taken from. It is used to see if the signal is being taken from the burst at the correct time to get the strongest signal. It is normally used on smaller pipes when the instrument is being used in double or triple bounce as signals can sometimes interfere with each other. This value is normally a few μs below the
Up μs, Dn μs
Phase up/dn μs
Only valid if Calculated μs and Up μs, Dn μs are correct. If the reading is zero then
there is no signal, which could mean the pipe is empty, or the liquid is contaminated with particles or air.
Phase offset
This value will be between 0 and 15. The exact value is not important and will vary between applications. It should however, be stable when the flow condition is good and velocity is within the range of the transducers being used. As the flow rate increases towards and beyond the maximum, this figure will continuously change. In flow mode the instrument will read unstable or high flow.
value.
Flow (m/s)
This displays flow velocity in m/sec to 3 decimal places.
Signal
This is the average value of Signal up/dn and is a value between 800 and 2400 which Display’s the signal strength as a percentage (800=0%, 2400=86%).
Signal up/dn
This value is internal to the electronics and must be greater than 800. There is an option in the SET UP INSTRUMENT menu to allow this value to be taken down to 400 in extreme circumstances and is useful on some applications when the signal levels are poor.
Sensor separation
This is a reminder for the user to check for correct sensor separation and sensor mode.
STATUS/ERROR/WARNING MESSAGES
There are three types of message that will appear, Status, Error and Warning. These messages appear under the time and date on the display when in flow mode.
Status Messages
S1: Initialising
Appears when first entering flow mode to show instrument is starting up.
Error Messages
E1: unstable or high flow
This error message occurs when either the sensors have been positioned too near to an obstruction or bend causing turbulence, or the instrument is being used outside its normal flow range. When the instrument is programmed it calculates the maximum flow rate and if this is exceeded then the high flow message occurs.
E2: No Flow Signal
This message appears when the two transducers cannot send or receive signals, which could happen for various reasons. Firstly check that all cables are connected, transducers are on the pipe correctly with grease between the sensor face and the pipe.
No flow signal will happen if:
The pipe is empty or partially filled.
The liquid is aerated.
The particulate content of the liquid is
too high.
10
The grease has not been applied to
the transducers
The condition of the pipe being
measured is poor.
Warning Messages
W1: Check site data
This message occurs when the application information has been entered incorrectly or the wrong sensors have been attached to the wrong pipe size causing the system timing to be in error. The site data needs to be checked and the instrument reprogrammed.
W2: Signal timing poor
Unstable signal timing or differing up/down stream times, indicate that the liquid is aerated or pipe surface is of poor quality.
W3: Flow signals poor
This warning appears when there is a signal lower than 25%. This could be due to the application, a poor quality pipe, amongst others.
W4: mA OUT Overange
The mA output is over-range when the flow is higher than the maximum mA range. Once the 4-20mA has been set up and the flow goes above the range set then this message will appear. It is possible to re-scale the 4-20mA to be able to cope with the higher flow.
W5: BATTERY LOW
The battery low warning occurs when battery indication is on 40%. The instrument has approximately 30 minute’s usage before it needs recharging. (See application note at the back of this manual)
W6: mA load too high
The 4-20mA Output is designed to work with a load up to 750Ω. When the load is too high or not connected, the above warning message will be displayed.
Other Messages
The messages below appear mainly when data has been incorrectly entered or the Portaflow 204 Plus! is trying to be used on an application that it is not capable of working on.
Pipe OD out of range
The outside diameter of the pipe has been entered and is out of range of the instrument.
Wall thickness out of range
The wall thickness that has been entered is out of range of the instrument.
Lining thickness out of range
The pipe lining thickness has been incorrectly entered.
11
APPLICATION & PERFORMANCE
WARNING - Users should or note that: a) The PORTAFLOW is not certified for use in Hazardous areas. b) The local site safety regulations must be complied with. c) Work is carried out in accordance with The Health & Safety at Work Act 1974.
The operational temperature of the fluid to be measured is -20°C to +125°C The velocity of the liquid marginally accelerates the speed at which the ultrasound is
transmitted from transducer ‘A’ to ‘B’. However, when the direction of the transmission is reversed, i.e. from transducer ‘B’ to ‘A’, a deceleration will occur because the sound is being transmitted against the liquid flow. Therefore the difference in time taken over the same distance is directly proportional to the flow velocity - see fig. 4.
A B
FLOW
Fig. 4 As it is not possible to determine from the outside what flow conditions prevail inside the pipe,
it must be assumed that the liquid is flowing uniformly in either a turbulent or laminar flow condition, and that the flow velocity profile is also uniform for 360 degrees around the pipe axis. Distortion of the flow velocity profile caused by bends, tees or any other type of obstruction may create measurement errors.
LIQUID CONDITIONS
Transit time ultrasonic meters perform best on liquids that are totally free from entrained air and solids. With sufficient air in the system, the ultrasound beam can be attenuated totally and therefore prevent the instrument from working. Often it is impossible to tell whether there is air in the system or not.
If a flow signal cannot be obtained, a simple test to determine whether the flow is aerated involves cutting off the flow for a period of 10-15 minutes.
During this time, the air bubbles will rise to the top of the pipe and the flow signal should return. When the flow signal has returned, “switch on” the flow and if entrained air is still present then signal loss
Pipe NB
PORTAFLOW 204 Plus! OPERATING RANGE
mm
100
75
50
25 13
0.07 0.10 0.13 0.22 0.29 0.32 0.42 5.9 8.2 11 14.76 23.94
0
0.02 0.03 0.04 0.07 0.09 0.10 0.13 1.8 2.5 4.5 7.3
.40
3.5
12
PORTAFLOW 204 Plus! SPECIFICATION
Carrying Case
Outside Dimensions Protection Class Material Total Weight Complete Electronics Housing Assembly Approximate Dimensions Material Protection class Approx. weight Operating Temp Storage Temp Battery Pack Data input Via 16 Key Tactile Membrane Keypad Display Graphics LCD Display Low Batt’ Indication Signal Level Indication Output data Flow Rate Indication
Totaliser (To 12 Digits) Pulse Output
Analogue Display Resolution
0.1% of the reading or better Repeatability Changed Transducer Position Response Time Less than 2 seconds
Transducers Clamp on with Ball Chain strapping Guide Rail Length Operating Range Cable Length Pipe range 13mm to 115mm Outside Diameter Accuracy
115mm 13mm
: +/-1-3% within the velocity range. Operating Range. : 0.3 metres/sec to 4 metres/sec : 0.5 metres/sec to 8 metres/sec : Maximum velocity dependent on the pipe size
: 210mm x 36mm x 27mm : -20°C to +125°C : 2 metres
: 350mmx330mmx170mm : IP54 : Plastic : 6 Kilos
: 235mmx125mmx42mm : Pebble Grey Abs : IP 54 : 400grms : 0°C to +45°C : -20°C to +55°C : 5 AA Nickel Metal Hydride
: Litres/sec, litres/min, gallons/min, US gallons/min, m3/hr : Feet/sec, metres/sec
: Gallons, litres, US gallons, m : 5 Volts. Maximum 1 pulse per second
: 4-20mA in 500 ohms
:+/- 10%
204 Plus!
3
MICRONICS reserve the right to alter any specification without notification.
13
WARRANTY
The material and workmanship of the PORTAFLOW 204 Plus! is guaranteed by MICRONICS LTD for one year from the date of purchase provided the equipment has been used for the purpose for which it has been designed, and has been operated in accordance with the operating manual supplied.
Misuse by the purchaser, or any other person, will immediately revoke any warranty given or implied.
Repair or replacement will be at MICRONICS LTD discretion and will be made without charge at MICRONICS LTD plant during the warranty period. MICRONICS LTD reserves the right, without prior notices, to discontinue manufacture, redesign or modify any of its products. Your statutory rights are not affected by this warranty.
If any problems develop, customers are requested to take the following steps: Notify MICRONICS LTD or the Distributor/Agent from whom the flow meter was purchased
giving details of the problem. Be sure to include the Model & Serial Number of the instrument. When returning a product to the factory, carefully package and ship freight prepaid. Be sure to include a complete description of the application and problem and identify any hazardous material used with the product. The Warranty of the PORTAFLOW is strictly in accordance with that stated above, and cannot in any way be extended.
CE MARKING
The PORTAFLOW 204 Plus! has been tested and found to conform to EN50081 - 1 Emission Standards and EN50082 - 1 Immunity Standards.
The tests were conducted by AQL - EMC Ltd, of 16 Cobham Road, Fern Down Industrial Estate, Windborne, U.K. BH21 7PG.
The unit was tested with all cables as supplied of a maximum length of 3m. While the operation of the unit may not be affected by the use of longer cables, MICRONICS LTD can make no statement about conformance to the above standards when these cables are in use.
The PORTAFLOW 204 Plus! is supplied with an external battery-charging unit. This unit is manufactured by Frieman & Wolf, Gerätebau GmbH. P.O. Box 1164 D-48342 Ostbevan, Germany who have CE marked the equipment. MICRONICS LTD have purchased this equipment on the understanding that the manufacturers have tested the unit to the relevant standards prior to CE marking the product. MICRONICS LTD have not tested the charger unit and cannot accept responsibility for any non-conformance from the relevant standards.
14
LIQUID SOUND SPEEDS
Note: All the following sound speeds are calculated at 25°C.
The speed of sound in liquids at temperatures other than 25°C is calculated as follows.
Example:
Glycol C2H6O
Substance Form
Index
2
Specific
Gravity
1.113 1658 2.1
Sound
Speed
Δv/ºC ­m/s/ºC
Water, distilled (49,50) H2O 0.996 1498 -2.4
For every 1°C higher than 25°C take off the value in the Δv/°C- m/s/°C column. Glycol at 50°C = 1658 - (2.1 x 25) = 1605.5
For every 1°C less than 25°C add on the value in the Δv/°C- m/s/°C column. Glycol at 5°C = 1658 + (2.1 x 20) = 1700
If the value has a minus sign in front of it then do the opposite of above. Distilled Water at 50°C = 1498 - (-2.4 x 25) =1558 Distilled Water at 10°C = 1498 + (-2.4 x 15) = 1462
Acetic anhydride (22) (CH3CO)2O 1.082 (20ºC) 1180 2.5 Acetic acid, anhydride (22) (CH3CO)2O 1.082 (20ºC) 1180 2.5 Acetic acid, nitrile C2H3N 0.783 1290 4.1 Acetic acid, ethyl ester (33) C4H8O Acetic acid, methyl ester C3H6O Acetone C3H6O 0.791 1174 4.5 Acetonitrile C2H3N 0.783 1290 4.1 Acetonylacetone C6H10O Acetylene dichloride C2H2Cl2 1.26 1015 3.8 Acetylene tetrabromide (47) C2H2Br Acetylene tetrachloride (47) C2H2Cl Alcohol C2H6O 0.789 1207 4.0 Alkazene-13 C15H Alkazene-25 C10H12Cl 2-Amino-ethanol C2H7NO 1.018 1724 3.4 2-Aminotolidine (46) C7H9N 0.999 (20ºC) 1618 4-Aminotolidine (46) C7H9N 0.966 (45ºC) 1480 Ammonia (35) NH Amorphous Polyolefin 0.98 962.6 t-Amyl alcohol C5H12O 0.81 1204 Aminobenzene (41) C6H5NO Aniline (41) C6H5NO Argon (45) Ar 1.400 (-188ºC) 853 Azine C6H5N 0.982 1415 4.1 Benzene (29,40,41) C6H Benzol (29,40,41) C6H Bromine (21) Br Bromo-benzene (46) C6H5Br 1.522 1170 1-Bromo-butane (46) C4H9Br 1.276 (20ºC) 1019 Bromo-ethane (46) C2H5Br 1.460 (20ºC) 900 Bromoform (46,47) CHBr n-Butane (2) C4H 2-Butanol C4H10O 0.81 1240 3.3 sec-Butylalcohol C4H10O 0.81 1240 3.3 n-Butyl bromide (46) C4H9Br 1.276 (20ºC) 1019 n-Butyl chloride (22,46) C4H9Cl 0.887 1140 4.57 tert Butyl chloride C4H9Cl 0.84 984 4.2 Butyl oleate C22H42O 2,3 Butylene glycol C4H10O Cadmium (7) Cd 2237.7 Carbinol (40,41) CH4O 0.791 (20ºC) 1076 2.92 Carbitol C6H14O Carbon dioxide (26) CO Carbon disulphide CS Carbon tetrachloride(33,35,47) CCl Carbon tetrafluoride (14) CF Cetane (23) C16H Chloro-benezene C6H5Cl 1.106 1273 3.6 1-Chloro-butane (22,46) C4H9Cl 0.887 1140 4.57 Chloro-diFluoromethane (3) (Freon 22) CHClF Chloroform (47) CHCl 1-Chloro-propane (47) C3H7Cl 0.892 1058 Chlorotrifluoromethane (5) CClF
Substance Form Index Specific Gravity Sound Speed
2 2
2
4 4
24
2
3
2 2
6 6
2
3
10
2
2
3 2 2
4
4
34
2
3
3
0.901 1085 4.4
0.934 1211
0.729 1399 3.6
2.966 1027
1.595 1147
0.86 1317 3.9
1.20 1307 3.4
0.771 1729 6.68
1.022 1639 4.0
1.022 1639 4.0
0.879 1306 4.65
0.879 1306 4.65
2.928 889 3.0
2.89 (20ºC) 918 3.1
0.601 (0ºC) 1085 5.8
1404 3.0
1.019 1484 1.51
0.988 1458
1.101 (-37ºC) 839 7.71
1.261 (22ºC) 1149
1.595 (20ºC) 926 2.48
1.75 (-150ºC) 875.2 6.61
0.773 (20ºC) 1338 3.71
1.491 (-69ºC) 893.9 4.79
1.489 979 3.4 724 5.26
Δv/ºC -m/s/ºC
15
Substance Form Index Specific Gravity Sound Speed
Δv/ºC -m/s/ºC
Cinnamaldehyde C9H8O 1.112 1554 3.2 Cinnamic aldehyde C9H8O 1.112 1554 3.2 Colamine C2H7NO 1.018 1724 3.4 o-Cresol (46) C7H8O 1.047 (20ºC) 1541 m-Cresol (46) C7H8O 1.034 (20ºC) 1500 Cyanomethane C2H3N 0.783 1290 4.1 Cyclohexane (15) C6H Cyclohexanol C6H12O 0.962 1454 3.6
12
0.779 (20ºC) 1248 5.41
Cyclohexanone C6H10O 0.948 1423 4.0 Decane (46) C10H 1-Decene (27) C10H n-Decylene (27) C10 H Diacetyl C4H6O Diamylamine C10H23N 1256 3.9 1,2 Dibromo-ethane (47) C2H4Br trans-1,2-Dibromoethene(47) C2H2Br Dibutyl phthalate C8H22O4 1408
22 20
20
2
2 2
0.730 1252
0.746 1235 4.0
0.746 1235 4.0
0.99 1236 4.6
2.18 995
2.231 935
Dichloro-t-butyl alcohol C4H8Cl2O 1304 3.8 2,3 Dichlorodioxane C2H6Cl2O Dichlorodifluoromethane (3) (Freon 12) CCl2F 1,2 Dichloro ethane (47) C2H4Cl cis 1,2-Dichloro-Ethene(3,47) C2H2Cl trans 1,2-Dichloro-ethene(3,47) C2H2Cl Dichloro-fluoromethane (3) (Freon 21) CHCl2F 1.426 (0ºC) 891 3.97 1-2-Dichlorohexafluoro cyclobutane (47) C4Cl2F 1-3-Dichloro-isobutane C4H8Cl Dichloro methane (3) CH2Cl 1,1-Dichloro-1,2,2,2 tetra fluoroethane CClF2-CClF Diethyl ether C4H10O 0.713 985 4.87 Diethylene glycol, monoethyl ether C6H14O Diethylenimide oxide C4H9NO 1.00 1442 3.8 1,2-bis(DiFluoramino) butane (43) C4H8(NF2) 1,2bis(DiFluoramino)- 2-methylpropane (43) C4H9(NF2) 1,2bis(DiFluoramino) propane (43) C3H6(NF2) 2,2bis(DiFluoramino) propane (43) C3H6(NF2) 2,2-Dihydroxydiethyl ether C4H10O Dihydroxyethane C2H6O 1,3-Dimethyl-benzene (46) C8H 1,2-Dimethyl-benzene(29,46) C8H 1,4-Dimethyl-benzene (46) C8H 2,2-Dimethyl-butane (29,33) C6H Dimethyl ketone C3H6O 0.791 1174 4.5 Dimethyl pentane (47) C7H Dimethyl phthalate C8H10O Diiodo-methane CH2I Dioxane C4H8O Dodecane (23) C12H 1,2-Ethanediol C2H6O Ethanenitrile C2H3N 0.783 1290
2
2
2 2 2
6
2
2
2
3
2 2 2 2
3
2 10 10 10 14
16
4
2
2
26
2
1391 3.7
1.516(-40ºC) 774.1 4.24
1.253 1193
1.284 1061
1.257 1010
1.654 669
1.14 1220 3.4
1.327 1070 3.94
1.455 665.3 3.73
0.988 1458
1.216 1000
1.213 900
1.265 960
1.254 890
1.116 1586 2.4
1.113 1658 2.1
0.868 (15ºC) 1343
0.897 (20ºC) 1331.5 4.1 1334
0.649 (20ºC) 1079
0.674 1063
1.2 1463
3.235 980
1.033 1376
0.749 1279 3.85
1.113 1658 2.1
Ethanoic anhydride (22) (CH3CO)2O 1.082 1180 Ethanol C2H6O 0.789 1207 4.0 Ethanol amide C2H7NO 1.018 1724 3.4 Ethoxyethane C4H10O 0.713 985 4.87 Ethyl acetate (33) C4H8O Ethyl alcohol C2H6O 0.789 1207 4.0 Ethyl benzene (46) C8H Ethyl bromide (46) C2H5Br 1.461 (20ºC) 900
2
10
0.901 1085 4.4
0.867(20ºC) 1338
Ethyliodide (46) C2H5I 1.950 (20ºC) 876 Ether C4H10O 0.713 985 4.87 Ethyl ether C4H10O 0.713 985 4.87 Ethylene bromide (47) C2H4Br Ethylene chloride (47) C2H4Cl Ethylene glycol C2H6O 50% Glycol/ 50% H2O 1578
2 2
2
2.18 995
1.253 1193
1.113 1658 2.1
d-Fenochone C10H16O 0.947 1320 d-2-Fenechanone C10H16O 0.947 1320 Fluorine F 0.545 (-143ºC) 403 11.31 Fluoro-benzene (46) C6H5F 1.024 (20ºC) 1189 Formaldehyde, methyl ester C2H4O Formamide CH3NO 1.134 (20ºC) 1622 2.2
2
0.974 1127 4.02
Formic acid, amide CH3NO 1.134 (20ºC) 1622 Freon R12 774.2 Furfural C5H4O Furfuryl alcohol C5H6O Fural C5H4O 2-Furaldehyde C5H4O 2-Furancarboxaldehyde C5H4O 2-Furyl-Methanol C5H6O Gallium Ga 6.095 2870 (30ºC) Glycerin C3H8O Glycerol C3H8O Glycol C2H6O Helium (45) He Heptane (22,23) C7H n-Heptane (29,33) C7H Hexachloro-Cyclopentadiene(47) C5Cl Hexadecane (23) C16H Hexalin C6H12O 0.962 1454 3.6 Hexane (16,22,23) C6H n-Hexane (29,33) C6H 2,5-Hexanedione C6H10O
2 2 2 2 2 2
3 3 2
4
16 16
6
34
14 14
2
1.157 1444
1.135 1450 3.4
1.157 1444 3.7
1.157 1444 3.7
1.157 1444 3.7
1.135 1450 3.4
1.26 1904 2.2
1.26 1904 2.2
1.113 1658 2.1
0.125(-268.8ºC) 183
0.684 (20ºC) 1131 4.25
0.684 (20ºC) 1180 4.0
1.7180 1150
0.773 (20ºC) 1338 3.71
0.659 1112 2.71
0.649 (20ºC) 1079 4.53
0.729 1399 3.6
16
Substance Form Index Specific Gravity Sound Speed
Δv/ºC -m/s/ºC
n-Hexanol C6H14O 0.819 1300 3.8 Hexahydrobenzene (15) C6H Hexahydrophenol C6H12O 0.962 1454 3.6 Hexamethylene (15) C6H Hydrogen (45) H 2-Hydroxy-toluene (46) C7H8O 1.047 (20ºC) 1541
12
12
2
0.779 1248 5.41
0.779 1248 5.41
0.071 (-256ºC) 1187
3-Hydroxy-tolune (46) C7H8O 1.034 (20ºC) 1500 Iodo-benzene (46) C6H5I 1.823 1114 Iodo-ethane (46) C2H5I 1.950 (20ºC) 876 Iodo-methane CH3I 2.28 (20ºC) 978 Isobutyl acetate (22) C6H12O 1180 4.85 Isobutanol C4H10O 0.81 (20ºC) 1212 Iso-Butane 1219.8 Isopentane (36) C5H Isopropanol (46) C3H8O 0.785 (20ºC) 1170
12
0.62 (20ºC) 980 4.8
Isopropyl alcohol (46) C3H8O 0.785 (20ºC) 1170 Kerosene 0.81 1324 3.6 Ketohexamethylene C6H10O 0.948 1423 4.0 Lithium fluoride (42) LiF 2485 1.29 Mercury (45) Hg 13.594 1449 Mesityloxide C6H16O 0.85 1310 Methane (25,28,38,39) CH Methanol (40,41) CH4O 0.791 (20ºC) 1076 2.92 Methyl acetate C3H6O o-Methylaniline (46) C7H9N 0.999 (20ºC) 1618
4
2
0.162 405(-89.15ºC) 17.5
0.934 1211
4-Methylaniline (46) C7H9N 0.966 (45ºC) 1480 Methyl alcohol (40,44) CH4O 0.791 (20ºC) 1076 2.92 Methyl benzene (16,52) C7H 2-Methyl-butane (36) C5H Methyl carbinol C2H6O 0.789 1207 4.0 Methyl-chloroform (47) C2H3Cl Methyl-cyanide C2H3N 0.783 1290
8
12
3
0.867 1328 4.27
0.62 (20ºC) 980
1.33 985
3-Methyl cyclohexanol C7H14O 0.92 1400 Methylene chloride (3) CH2Cl Methylene iodide CH2I Methyl formate (22) C2H4O Methyl iodide CH3I 2.28 (20ºC) 978 α-Methyl naphthalene 2-Methylphenol (46) C7H8O 1.047 (20ºC) 1541
C
2
2
2
11H10
1.327 1070 3.94
3.235 980
0.974 (20ºC) 1127 4.02
1.090 1510 3.7
3-Methylphenol (46) C7H8O 1.034 (20ºC) 1500 Milk, homogenized 1548 Morpholine C4H9NO 1.00 1442 3.8 Naphtha 0.76 1225 Natural Gas (37) 0.316 (-103ºC) 753 Neon (45) Ne 1.207 (-246ºC) 595 Nitrobenzene (46) C6H5NO Nitrogen (45) N Nitromethane (43) CH3NO2 1.135 1300 4.0
2
2
1.204 (20ºC) 1415
0.808 (-199ºC) 962
Nonane (23) C9H2O 0.718 (20ºC) 1207 4.04 1-Nonene (27) C9H Octane (23) C8H n-Octane (29) C8H 1-Octene (27) C8H Oil of Camphor Sassafrassy 1390 3.8
18 18 18 16
0.736 (20ºC) 1207 4.0
0.703 1172 4.14
0.704 (20ºC) 1212.5 3.50
0.723 (20ºC) 1175.5 4.10
Oil, Car (SAE 20a.30) 1.74 870 Oil, Castor C11H10O Oil, Diesel 0.80 1250
10
0.969 1477 3.6
Oil, Fuel AA gravity 0.99 1485 3.7 Oil (Lubricating X200) 1530 Oil (Olive) 0.912 1431 2.75 Oil (Peanut) 0.936 1458 Oil (Sperm) 0.88 1440 Oil, 6 1509 2,2-Oxydiethanol C4H10O Oxygen (45) O Pentachloro-ethane (47) C2HCl Pentalin (47) C2HCl Pentane (36) C5H n-Pentane (47) C5H Perchlorocyclopentadiene(47) C5Cl Perchloro-ethylene (47) C2Cl Perfluoro-1-Hepten (47) C7F Perfluoro-n-Hexane (47) C6F Phene (29,40,41) C6H β-Phenyl acrolein Phenylamine (41) C6H5NO Phenyl bromide (46) C6H5Br 1.522 1170
3
2
5
5 12 12
6
4 14 14
6
C
O 1.112 1554 3.2
9H8
2
1.116 1586 2.4
1.155 (-186ºC) 952
1.687 1082
1.687 1082
0.626 (20ºC) 1020
0.557 1006
1.718 1150
1.632 1036
1.67 583
1.672 508
0.879 1306 4.65
1.022 1639 4.0
Phenyl chloride C6H5Cl 1.106 1273 3.6 Phenyl iodide (46) C6H5I 1.823 1114 Phenyl methane (16,52) C7H 3-Phenyl propenal C9H8O 1.112 1554 3.2
8
0.867 (20ºC) 1328 4.27
Phthalardione C8H4O3 1125 Phthalic acid, anhydride C8H4O Phthalic anhydride C8H4O Pimelic ketone C6H10O 0.948 1423 4.0
3 3
1125 1125
Plexiglas, Lucite, Acrylic 2651 Polyterpene Resin 0.77 1099.8 Potassium bromide (42) Kbr 1169 0.71 Potassium fluoride (42) KF 1792 1.03 Potassium iodide (42) KI 985 0.64 Potassium nitrate (48) KNO
3
1.859 (352ºC) 1740.1 1.1
17
Propane (2,13)(-45 to -130ºC) C3H 1,2,3-Propanetriol C3H8O 1-Propanol (46) C3H8O 0.78 (20ºC) 1222 2-Propanol (46) C3H8O 0.785 (20ºC) 1170 2-Propanone C3H6O 0.791 1174 4.5 Propene (17,18,35) C3H n-Propyl acetate (22) C5H10O n-Propyl alcohol C3H8O 0.78 (20ºC) 1222 Propylchloride (47) C3H7Cl 0.892 1058 Propylene (17,18,35) C3H Pyridine C6H5N 0.982 1415 4.1 Refrigerant 11 (3,4) CCl3F 1.49 828.3 3.56 Refrigerant 12 (3) CCl2F Refrigerant 14 (14) CF Refrigerant 21 (3) CHCl2F 1.426 (0ºC) 891 3.97 Refrigerant 22 (3) CHClF Refrigerant 113 (3) CCl2F-CClF Refrigerant 114 (3) CClF2-CClF Refrigerant 115 (3) C2ClF Refrigerant C318 (3) C4F Selenium (8) Se 1072 0.68 Silicone (30 cp) 0.993 990 Sodium fluoride (42) NaF 0.877 2082 1.32 Sodium nitrate (48) NaNO Sodium nitrite (48) NaNO Solvesso 3 0.877 1370 3.7 Spirit of wine C2H6O 0.789 1207 4.0 Sulphur (7,8,10) S 1177 -1.13 Sulphuric acid (1) H2SO Tellurium (7) Te 991 0.73 1,1,2,2-Tetrabromo-ethane(47) C2H2Br4 2.966 1027 1,1,2,2-Tetrachloro-ethane(67) C2H2Cl Tetrachloroethane (46) C2H2Cl Tetrachloro-ethene (47) C2Cl Tetrachloro-methane (33,47) CCl Tetradecane (46) C14H3O 0.763 (20ºC) 1331 Tetraethylene glycol C8H18O Tetrafluoro-methane (14) (Freon 14) CF Tetrahydro-1,4-isoxazine C4H9NO 1442 3.8 Toluene (16,52) C7H o-Toluidine (46) C7H9N 0.999 (20ºC) 1618 p-Toluidine (46) C7H9N 0.966 (45ºC) 1480 Toluol C7H Tribromo-methane (46,47) CHBr 1,1,1-Trichloro-ethane (47) C2H3Cl Trichloro-ethene (47) C2HCl Trichloro-fluoromethane (3) (Freon 11) CCl3F 1.49 828.3 3.56 Trichloro-methane (47) CHCl 1,1,2-Trichloro-1,2,2-Trifluoro-Ethane CCl2F-CClF Triethyl-amine (33) C6H15N 0.726 1123 4.47 Triethylene glycol C6H14O 1,1,1-Trifluoro-2-Chloro-2-Bromo-Ethane C2HClBrF 1,2,2-Trifluorotrichloro- ethane (Freon 113) CCl2F-CClF d-1,3,3-Trimethylnor- camphor C10H16O 0.947 1320 Trinitrotoluene (43) C7H5(NO2) Turpentine 0.88 1255 Unisis 800 0.87 1346 Water, distilled (49,50) H2O 0.996 1498 -2.4 Water, heavy D²O 1400 Water, sea 1.025 1531 -2.4 Wood Alcohol (40,41) CH4O 0.791 (20ºC) 1076 2.92 Xenon (45) Xe 630 m-Xylene (46) C8H o-Xylene (29,46) C8H p-Xylene (46) C8H Xylene hexafluoride C8H4F Zinc (7) Zn 3298
Substance Form Index Specific Gravity Sound Speed
8
3
6
2
6
2
4
2
5
8
3 2
4
4 4
4
4
5
4
8
8
3
3
3
3
4
3
10 10 10
6
0.585 (-45ºC) 1003 5.7
1.26 1904 2.2
0.563 (-13ºC) 963 6.32 1280 (2ºC) 4.63
0.563 (-13ºC) 963 6.32
1.516 (-40ºC) 774.1 4.24
1.75 (-150ºC) 875.24 6.61
1.491 (-69ºC) 893.9 4.79
2 2
2
2
3
1.563 783.7 3.44
1.455 665.3 3.73
656.4 4.42
1.62 (-20ºC) 574 3.88
1.884 (336ºC) 1763.3 0.74
1.805 (292ºC) 1876.8
1.841 1257.6 1.43
1.595 1147
1.553 (20ºC) 1170
1.632 1036
1.595 (20ºC) 926
1.123 1586/5203.4 3.0
1.75 (-150ºC) 875.24 6.61
0.867 (20ºC) 1328 4.27
0.866 1308 4.2
2.89 (20ºC) 918
1.33 985
1.464 1028
1.489 979 3.4
1.563 783.7
1.123 1608 3.8
1.869 693
1.563 783.7 3.44
1.64 1610
0.868 (15ºC) 1343
0.897 (20ºC) 1331.5 4.1 1334
1.37 879
Δv/ºC -m/s/ºC
18
PORTAFLOW 204 Plus Battery Charge circuit Operation.
Charging Controller IC:
A Maxim IC MAX712 or MAX713 controls the Ni-Cd and Ni-Mh battery charger. It has two modes, fast charge and trickle charge; an output indicates the fast-charge status. In both modes it supplies, via a PNP power transistor, a constant current to the battery, by keeping a constant voltage across a current sensing resistor. In fast charge m ode it is 250mV, in trickle charge mode 31mV, so the trickle charge current is 1/8 of the fast charge current.
By wiring up input pins on the IC, the number of cells is set to 5, the voltage sampling interval to 168 sec, and the fast-charge time limit to 264 minutes (the maximum). The battery temperature limits are not used.
The IC starts the fast-charge timer when a battery is connected or when power is applied. It terminates the fast charge and returns to trickle charge, either after the 264 min (~4.5 hrs) time limit, or when it senses that the battery voltage remains constant or begins to decrease, meaning that the battery is fully charged.
Charging Voltage:
The voltage available to charge the 6V battery is restricted by the 9V charger input and the two diodes in the input. The S2D silicon diodes had a fwd drop of 0.75V, limiting the available charge voltage to 7.5V, which caused the MAX712 to sense that the battery voltage had stopped rising, and therefore prematurely end the fast charge. With several days of trickle charging the battery could however still reach its full capacity. In Dec.2000 the S2D diodes were replaced by SS14 Schottky diodes with a fwd drop of 0.35V, thus raising the available charge voltage to 8.3V. At the same time the current was increased.
Instrument differences:
The current sensing resistor consists of either 2 or 4 parallel 1.2Ω resisto rs, giving about 0.4A or 0.8A fast-charge current. PF-300 and UFM610P: Battery Capacity 3.5Ah, or 4.0Ah after Oct.2000 Current 0.4A before, 0.8A after Dec.2000 PF 204, PF-SE and 216: Battery Capacity 1.2Ah Current 0.4A
Software:
The fast-charge status output is not used by the present software (ver.3.06); in a future software update a message will be added, indicating charging status.
Quicker full charge:
The fastest way to fully charge the battery is to charge for 4.5 hrs, then switch the power supply off and on again, thus re-starting the fast charge for another 4.5 hr period, followed by trickle charge.
19
Warning:
If the battery is getting warm, that would indicate that it is full, and the power supply should not be connected again - overcharging reduces the life of the battery. Note: After a recently fully charged battery is connected to the charger, it seems that it takes the MAX712 about 30 min to sense that the battery voltage stops changing, and go to trickle charge.
Examples:
Older PF-300:-
10.5 hrs of trickle charge (50mA): 4.5*0.4+10.5*0.05=2.325Ah=3.5Ah*0.66, which fills the battery to 66% of capacity (3.5Ah). To fill the remaining 34% at 50mA takes 3.5*0.34/0.05=23.8hrs, +15hrs = 39hrs to 100%. Assuming 20% losses: (3.5Ah*20%)/50mA=0.7Ah/0.05A=14hrs of trickle charge to cover losses, +39hrs=53hrs total. In fact it needs ~9hrs * 0.4A = 3.6Ah to fill the battery from empty to 103% full capacity. Assuming 20% losses: (3.5Ah*20%-0.1)/50mA=0.6Ah/0.05A=12hrs of trickle charge to cover losses, +9hrs=21hrs total. A third session of fast charge would fill the last 17% in 3.5Ah*17%/0.4A=1.5hrs, = 10.5hrs total. Newer PF-300:-
4.5hrs fast: 0.8A*4.5h=3.6Ah = 90% of 4.0Ah Slow: 10%= 0.4Ah/0.1A = 4h, total 8.5h to 100%
A 15 hour charge consists of 4.5 hrs of fast charge (400mA), followed by
with 20% losses: 0.8Ah/0.1A = 8h Total time fast and slow: 16.5hrs to 120%. Fast only: 4.0Ah/0.8A=5hrs, +20%=6hrs, that needs 2 sessions: 4.5hrs + 1.5hrs to 120%. PF204, PF-SE & 216:­3h+20%=3.6hrs total. This is well within the first 4.5hrs.
1.2Ah/0.4A=3hrs to 100% capacity; with 20% losses
20
PRODUCT CARE AND MAINTENANCE
WARNING:-USE ONLY BATTERIES, CHARGER, ACCESSORIES, CABLES APPROVED FOR THIS PARTICULAR MODEL.
THE USE OF ANY OTHER TYPES MAY INVALIDATE ANY APPROVAL OR WARRANTY APPLYING TO THE INSTRUMENT AND MAY BE DANGEROUS. IF IN DOUBT CONTACT A MICRONICS SALES OR SERVICE REPRESENTATIVE.
Do not disassemble this unit unless advised by Micronics. Return the unit to the place of purchase for further advice.
Do not drop.
Wipe the exterior of the instrument with a clean damp cloth or paper towel-the use of
a solvent may damage the paint surface. Ensure the unit is switched off and disconnected from the mains.
Do not place the instrument electronics near to naked flames or sources of intense heat such as an electric fire or hot pipes in excess 50°C.
Dispose of any batteries safely and in accordance with any regulations in force in the country of operation.
Ensure all connectors are kept clean and free from grease. They may be cleaned with a general purpose switch cleaner.
Avoid the use of excessive grease/ultrasonic couplant on the sensors. This may impair the performance of the equipment. Read the instructions in the manual on how to apply the couplant. Any excessive grease/couplant can be removed from the sensors and guide rails using an absorbent paper towel and a gene ral purpose solvent cleaner.
Regularly check all cables/parts for damage. Replacement parts are available from Micronics.
On fixed instruments, it is recommended the ultrasonic couplant is replaced on the sensors every 6 months especially on pipes where the application is too hot to touch. If signal level drops below 30% this is also an indication that the sensors need re­greasing.
Ensure the person who services your instrument is qualified to do so. If in doubt, return the instrument to Micronics with a detailed report on the nature of the problem.
Ensure that suitable protective precautions are taken when using any materials to clean the instrument/sensors.
Calibration of the instrument and sensors is recommended to be done at least once every 12 months.
If the instrument was supplied with dust or dirt caps make sure they are re-fitted when the instrument is not in use.
It is the users responsibility to tidy cables so they will not cause harm to other people.
When returning product to Micronics make sure it is clean. Notify Micronics if the
instrument has been in contact with any hazardous substances.
MICRONICS Ltd
Knaves Beech Business Centre, Davies Way, Loudwater,
High Wycombe, Bucks. HP10 9QR U.K.
TELEPHONE: +44 (0)1628 810456
FAX: +44 (0)1628 531540 e-mail: Website: www.micronicsltd.co.uk
January 2006
sales@micronicsltd.co.uk
PUBLICATION Dec 2005
MICRONICS LTD
Software Version: v1.04
21
Loading...