Micro Motion® Model 1700 Transmitters with
Analog Outputs
Includes the Chinese-Language Display Option
Safety messages
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully
before proceeding to the next step.
Micro Motion customer service
Email
•Worldwide: flow.support@emerson.com
•Asia-Pacific: APflow.support@emerson.com
North and South AmericaEurope and Middle EastAsia Pacific
United States800-522-6277U.K.0870 240 1978Australia800 158 727
Appendix HNE 53 history ..............................................................................................................331
H.1NE 53 history ........................................................................................................................... 331
Index ................................................................................................................................................336
Configuration and Use Manual iii
Contents
iv Micro Motion® Model 1700 Transmitters with Analog Outputs
Part I
Getting Started
Chapters covered in this part:
•Before you begin
•Quick start
Getting Started
Configuration and Use Manual 1
Before you begin
1Before you begin
Topics covered in this chapter:
•About this manual
•Transmitter model code
•Communications tools and protocols
•Additional documentation and resources
1.1About this manual
This manual provides information to help you configure, commission, use, maintain, and
troubleshoot the Micro Motion transmitter.
Important
This manual assumes that the transmitter has been installed correctly and completely, according to
the instructions in the transmitter installation manual, and that the installation complies with all
applicable safety requirements.
1.2Transmitter model code
Your transmitter can be identified by the model number on the transmitter tag.
The transmitter has a model number of the following form:
1700(I/R/C/B)**A******
IIntegral mount
R4-wire remote-mount
C9-wire remote-mount
BRemote core processor with remote transmitter
AAnalog outputs option board
1.3Communications tools and protocols
You can use several different communications tools and protocols to interface with the
transmitter. You may use different tools in different locations or for different tasks.
2 Micro Motion® Model 1700 Transmitters with Analog Outputs
Before you begin
Communications tools, protocols, and related informationTable 1-1:
Communications toolSupported protocolsScopeIn this manualFor more information
Display (standard)
Chinese-language display
ProLink II• HART/RS-485
ProLink III• HART/RS-485
Field Communicator
Not applicableBasic configuration and
commissioning
Not applicableBasic configuration and
commissioning
(1)
• HART/Bell 202
• Modbus/RS-485
• Service port
(1)
• HART/Bell 202
• Modbus/RS-485
• Service port
HART/Bell 202Complete configuration
Complete configuration
and commissioning
Complete configuration
and commissioning
and commissioning
Complete user information. See Appendix A.
Complete user information. See Appendix B.
Basic user information.
See Appendix C.
Basic user information.
See Appendix D.
Basic user information.
See Appendix E.
Not applicable
Not applicable
User manual
• Installed with soft-
ware
• On Micro Motion
user documentation
CD
• On Micro Motion
web site (www.mi-
cromotion.com
User manual
• Installed with soft-
ware
• On Micro Motion
user documentation
CD
• On Micro Motion
web site (www.mi-
cromotion.com
User manual on
Micro Motion web site
(www.micromo-
tion.com
Tip
You may be able to use other communications tools from Emerson Process Management, such as
AMS Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the
Smart Wireless THUM Adapter is not discussed in this manual. The AMS interface is similar to the
ProLink II interface. For more information on the Smart Wireless THUM Adapter, refer to the
documentation available at www.micromotion.com.
1.4Additional documentation and resources
Micro Motion provides additional documentation to support the installation and operation
of the transmitter.
(1) Devices with the Chinese-language display do not support HART/RS-485.
Configuration and Use Manual 3
Before you begin
Additional documentation and resourcesTable 1-2:
TopicDocument
SensorSensor documentation
Transmitter installation
Hazardous area installation See the approval documentation shipped with the transmitter, or
download the appropriate documentation from the Micro Motion
web site at www.micromotion.com.
All documentation resources are available on the Micro Motion web site at
www.micromotion.com or on the Micro Motion user documentation CD.
4 Micro Motion® Model 1700 Transmitters with Analog Outputs
2Quick start
Topics covered in this chapter:
•Power up the transmitter
•Check flowmeter status
•Make a startup connection to the transmitter
•Characterize the flowmeter (if required)
•Verify mass flow measurement
•Verify the zero
2.1Power up the transmitter
The transmitter must be powered up for all configuration and commissioning tasks, or for
process measurement.
1.Ensure that all transmitter and sensor covers and seals are closed.
Quick start
CAUTION!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers
and seals are tightly closed. For hazardous area installations, applying power while
housing covers are removed or loose can cause an explosion.
2.Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. During this period,
Alarm 009 is active. The diagnostic routines should complete in approximately
30 seconds. For transmitters with a display, the status LED will turn green and begin
to flash when the startup diagnostics are complete. If the status LED exhibits
different behavior, an alarm condition is present.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics
can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial
startup, or if power is been off long enough to allow components to reach ambient
temperature, allow the electronics to warm up for approximately 10 minutes before
relying on process measurements. During this warm-up period, you may observe minor
measurement instability or inaccuracy.
2.2Check flowmeter status
Check the flowmeter for any error conditions that require user action or that affect
measurement accuracy.
Configuration and Use Manual 5
Quick start
1.Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions. During the power-up sequence, Alarm A009 is active.
This alarm should clear automatically when the power-up sequence is complete.
2.Check the status LED on the transmitter.
Transmitter status reported by status LEDTable 2-1:
LED stateDescriptionRecommendation
GreenNo alarms are active.Continue with configuration or process meas-
urement.
(3)
(1)
(2)
No alarms are active. One or more previously
active alarms have not been acknowledged.
and have been acknowledged.
One or more low-severity alarms are active
and have not been acknowledged.
and have been acknowledged.
One or more high-severity alarms are active
and have not been acknowledged.
Continue with configuration or process measurement. If you choose, you can acknowledge
the alarms.
A low-severity alarm condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or process measurement. If you choose, you can identify and resolve the alarm condition.
A low-severity alarm condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or process measurement. If you choose, you can identify and resolve the alarm condition. You may
also acknowledge the alarm.
A high-severity alarm condition affects measurement accuracy and output behavior. Resolve the alarm condition before continuing.
A high-severity alarm condition affects measurement accuracy and output behavior. Resolve the alarm condition before continuing.
You may also acknowledge the alarm.
Flashing green
YellowOne or more low-severity alarms are active,
Flashing yellow
RedOne or more high-severity alarms are active,
Flashing red
Postrequisites
For information on viewing the list of active alarms, see Section 8.4.
For information on individual alarms and suggested resolutions, see Section 10.2.
(1) If Status LED Blinking is disabled, the LED will show solid green rather than flashing.
(2) If Status LED Blinking is disabled, the LED will show solid yellow rather than flashing.
(3) If Status LED Blinking is disabled, the LED will show solid red rather than flashing.
6 Micro Motion® Model 1700 Transmitters with Analog Outputs
Quick start
2.3Make a startup connection to the transmitter
For all configuration tools except the display, you must have an active connection to the
transmitter to configure the transmitter. Follow this procedure to make your first
connection to the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in
the appropriate appendix. Use the default communications parameters shown in the
appendix.
Communications toolConnection type to useInstructions
ProLink IIHART/RS-485
ProLink IIIHART/RS-485
Field CommunicatorHARTAppendix E
Postrequisites
(Optional) Change the communications parameters to site-specific values.
(4)
(4)
Appendix C
Appendix D
To change the communications parameters using ProLink II:
•To change the protocol, baud rate, parity, or stop bits, choose ProLink > Configuration >
RS-485.
•To change the address, choose ProLink > Configuration > Device.
To change the communications parameters using ProLink III, choose Device Tools >
Configuration > Communications.
To change the communications parameters using the Field Communicator, choose On-LineMenu > Configure > Manual Setup > Inputs/Outputs > Communications.
Important
If you are changing communications parameters for the connection type that you are using, you will
lose the connection when you write the parameters to the transmitter. Reconnect using the new
parameters.
(4) Devices with the Chinese-language display do not support HART/RS-485. The default connection to use for these devices is Modbus/RS-485.
Configuration and Use Manual 7
Quick start
2.4Characterize the flowmeter (if required)
Display (standard)Not available
Chinese-language
display
ProLink II• ProLink > Configuration > Device > Sensor Type
ProLink IIIDevice Tools > Calibration Data
Field Communicator Configure > Manual Setup > Characterize
Overview
Characterizing the flowmeter adjusts your transmitter to match the unique traits of the
sensor it is paired with. The characterization parameters (also called calibration
parameters) describe the sensor’s sensitivity to flow, density, and temperature.
Depending on your sensor type, different parameters are required. Values for your sensor
are provided by Micro Motion on the sensor tag or the calibration certificate.
If your flowmeter was ordered as a unit, it has already been characterized at the factory. However,
you should still verify the characterization parameters.
Procedure
1.Specify Sensor Type.
• Straight-tube (T-Series)
(5)
• Curved-tube (all sensors except T-Series)
2.Set the flow characterization parameters. Be sure to include all decimal points.
• For straight-tube sensors, set FCF (Flow Cal or Flow Calibration Factor), FTG, and
(5)
FFQ.
• For curved-tube sensors, set Flow Cal (Flow Calibration Factor).
3.Set the density characterization parameters.
• For straight-tube sensors, set D1, D2, DT, DTG, K1, K2, FD, DFQ1, and DFQ2.
(5)
• For curved-tube sensors, set D1, D2, TC, K1, K2, and FD. (TC is sometimes shown
as DT.)
(5) Devices with the Chinese-language display do not support T-Series sensors.
8 Micro Motion® Model 1700 Transmitters with Analog Outputs
2.4.1Sample sensor tags
Tag on older curved-tube sensors (all sensors except T-Series)Figure 2-1:
Quick start
Tag on newer curved-tube sensors (all sensors except T-Series)Figure 2-2:
Configuration and Use Manual 9
Quick start
Tag on older straight-tube sensor (T-Series)Figure 2-3:
Tag on newer straight-tube sensor (T-Series)Figure 2-4:
2.4.2Flow calibration parameters (FCF, FT)
Two separate values are used to describe flow calibration: a 6-character FCF value and a 4character FT value. They are provided on the sensor tag.
Both values contain decimal points. During characterization, these may be entered as two
values or as a single 10-character string. The 10-character string is called either Flowcal or
FCF.
If your sensor tag shows the FCF and the FT values separately and you need to enter a
single value, concatenate the two values to form the single parameter value.
If your sensor tag shows a concatenated Flowcal or FCF value and you need to enter the FCF
and the FT values separately, split the concatenated value:
•FCF = The first 6 characters, including the decimal point
•FT = The last 4 characters, including the decimal point
10 Micro Motion® Model 1700 Transmitters with Analog Outputs
Quick start
Example: Concatenating FCF and FT
FCF = x.xxxx
FT = y.yy
Flow calibration parameter: x.xxxxy.yy
Example: Splitting the concatenated Flowcal or FCF value
Flow calibration parameter: x.xxxxy.yy
FCF = x.xxxx
FT = y.yy
Density calibration parameters are typically on the sensor tag and the calibration
certificate.
If your sensor tag does not show a D1 or D2 value:
•For D1, enter the Dens A or D1 value from the calibration certificate. This value is the
line-condition density of the low-density calibration fluid. Micro Motion uses air. If
you cannot find a Dens A or D1 value, enter 0.001 g/cm3.
•For D2, enter the Dens B or D2 value from the calibration certificate. This value is the
line-condition density of the high-density calibration fluid. Micro Motion uses water.
If you cannot find a Dens B or D2 value, enter 0.998 g/cm3.
If your sensor tag does not show a K1 or K2 value:
•For K1, enter the first 5 digits of the density calibration factor. In the sample tag, this
value is shown as 12500.
•For K2, enter the second 5 digits of the density calibration factor. In the sample tag,
this value is shown as 14286.
If your sensor does not show an FD value, contact Micro Motion customer service.
If your sensor tag does not show a DT or TC value, enter the last 3 digits of the density
calibration factor. In the sample tag, this value is shown as 4.44.
2.5Verify mass flow measurement
Check to see that the mass flow rate reported by the transmitter is accurate. You can use
any available method.
•Read the value for Mass Flow Rate on the transmitter display.
•Connect to the transmitter with ProLink II and read the value for Mass Flow Rate in the
Process Variables window (ProLink > Process Variables).
•Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in
the Process Variables panel.
Configuration and Use Manual 11
Quick start
•Connect to the transmitter with the Field Communicator and read the value for Mass
Flow Rate in the Process Variables menu (On-Line Menu > Overview > Primary Purpose
Variables).
Postrequisites
If the reported mass flow rate is not accurate:
•Check the characterization parameters.
•Review the troubleshooting suggestions for flow measurement issues. See
Section 10.3.
2.6Verify the zero
Verifying the zero helps you determine if the stored zero value is appropriate to your
installation, or if a field zero can improve measurement accuracy.
The zero verification procedure analyzes the Live Zero value under conditions of zero flow,
and compares it to the Zero Stability range for the sensor. If the average Live Zero value is
within a reasonable range, the zero value stored in the transmitter is valid. Performing a
field calibration will not improve measurement accuracy.
2.6.1Verify the zero using ProLink II
Verifying the zero helps you determine if the stored zero value is appropriate to your
installation, or if a field zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the flowmeter unless
one of the following is true:
•The zero is required by site procedures.
•The stored zero value fails the zero verification procedure.
Prerequisites
ProLink II v2.94 or later
Important
Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem,
then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a lowseverity alarm is active.
Procedure
1.Prepare the flowmeter:
a. Allow the flowmeter to warm up for at least 20 minutes after applying power.
12 Micro Motion® Model 1700 Transmitters with Analog Outputs
Quick start
b. Run the process fluid through the sensor until the sensor temperature reaches
the normal process operating temperature.
c. Stop flow through the sensor by shutting the downstream valve, and then the
upstream valve if available.
d. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
2.Choose ProLink > Calibration > Zero Verification and Calibration > Verify Zero and wait until
the procedure completes.
3.If the zero verification procedure fails:
a. Confirm that the sensor is completely blocked in, that flow has stopped, and that
the sensor is completely full of process fluid.
b. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
c. Repeat the zero verification procedure.
d. If it fails again, zero the flowmeter.
For instructions on zeroing the flowmeter, see Zero the flowmeter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
2.6.2Verify the zero using ProLink III
Verifying the zero helps you determine if the stored zero value is appropriate to your
installation, or if a field zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the flowmeter unless
one of the following is true:
•The zero is required by site procedures.
•The stored zero value fails the zero verification procedure.
Prerequisites
ProLink III v1.0 with Patch Build 31, or a later release
Important
Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem,
then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a lowseverity alarm is active.
Procedure
1.Prepare the flowmeter:
a. Allow the flowmeter to warm up for at least 20 minutes after applying power.
Configuration and Use Manual 13
Quick start
b. Run the process fluid through the sensor until the sensor temperature reaches
the normal process operating temperature.
c. Stop flow through the sensor by shutting the downstream valve, and then the
upstream valve if available.
d. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
2.Choose Device Tools > Device Calibration > Zero Verification and Calibration > Verify Zero and
wait until the procedure completes.
3.If the zero verification procedure fails:
a. Confirm that the sensor is completely blocked in, that flow has stopped, and that
the sensor is completely full of process fluid.
b. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
c. Repeat the zero verification procedure.
d. If it fails again, zero the flowmeter.
For instructions on zeroing the flowmeter, see Zero the flowmeter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
2.6.3Terminology used with zero verification and zero
calibration
Terminology used with zero verification and zero calibrationTable 2-2:
TermDefinition
ZeroIn general, the offset required to synchronize the left pickoff and the right pickoff under
conditions of zero flow. Unit = microseconds.
Factory ZeroThe zero value obtained at the factory, under laboratory conditions.
Field ZeroThe zero value obtained by performing a zero calibration outside the factory.
Prior ZeroThe zero value stored in the transmitter at the time a field zero calibration is begun. May
be the factory zero or a previous field zero.
Manual ZeroThe zero value stored in the transmitter, typically obtained from a zero calibration proce-
dure. It may also be configured manually. Also called “mechanical zero” or “stored zero.”
Live ZeroThe real-time bidirectional mass flow rate with no flow damping or mass flow cutoff ap-
plied. An adaptive damping value is applied only when the mass flow rate changes dramatically over a very short interval. Unit = configured mass flow measurement unit.
Zero StabilityA laboratory-derived value used to calculate the expected accuracy for a sensor. Under
laboratory conditions at zero flow, the average flow rate is expected to fall within the
range defined by the Zero Stability value (0 ± Zero Stability). Each sensor size and model
has a unique Zero Stability value. Statistically, 95% of all data points should fall within the
range defined by the Zero Stability value.
Zero CalibrationThe procedure used to determine the zero value.
14 Micro Motion® Model 1700 Transmitters with Analog Outputs
Quick start
Terminology used with zero verification and zero calibration (continued)Table 2-2:
TermDefinition
Zero TimeThe time period over which the Zero Calibration procedure is performed. Unit = seconds.
Field Verification ZeroA 3-minute running average of the Live Zero value, calculated by the transmitter. Unit =
configured mass flow measurement unit.
Zero VerificationA procedure used to evaluate the stored zero and determine whether or not a field zero
can improve measurement accuracy.
Configuration and Use Manual 15
Configuration and commissioning
Part II
Configuration and commissioning
Chapters covered in this part:
•Introduction to configuration and commissioning
•Configure process measurement
•Configure device options and preferences
•Integrate the meter with the control system
•Completing the configuration
16 Micro Motion® Model 1700 Transmitters with Analog Outputs
Introduction to configuration and commissioning
3Introduction to configuration and
commissioning
Topics covered in this chapter:
•Configuration flowchart
•Default values and ranges
•Enable access to the off-line menu of the display
•Disable write-protection on the transmitter configuration
•Restore the factory configuration
3.1Configuration flowchart
Use the following flowchart as a general guide to the configuration and commissioning
process.
Some options may not apply to your installation. Detailed information is provided in the
remainder of this manual. If you are using the Weights & Measures application, additional
configuration and setup are required.
Configuration and Use Manual 17
Introduction to configuration and commissioning
Configuration flowchartFigure 3-1:
Configure process measurement
Configure device options and
preferences
Test and move to production
Configure mass flow
measurement
Configure volume flow
meaurement
Volume flow type
Liquid
Configure flow direction
Configure density
measurement
Configure temperature
measurement
Gas
Define gas properties
Configure display
parameters
Configure fault handling
parameters
Configure sensor
parameters
Configure device
parameters
Integrate device with control system
Configure the channels
Configure the mA
output(s)
Test or tune transmitter
using sensor simulation
Back up transmitter
configuration
Enable write-protection on
transmitter configuration
Done
Configure pressure
compensation (optional)
Configure the frequency
output(s)
Configure the discrete
output(s)
Configure events
Configure digital
communications
18 Micro Motion® Model 1700 Transmitters with Analog Outputs
Introduction to configuration and commissioning
3.2Default values and ranges
See Section F.1 to view the default values and ranges for the most commonly used
parameters.
3.3Enable access to the off-line menu of the
display
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features
Offline Maintain > Configuration > Display
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must
enable it if you want to use the display to configure the transmitter.
Restriction
You cannot use the display to enable access to the off-line menu. You must make a connection from
another tool.
3.4Disable write-protection on the transmitter
configuration
Field Communicator Configure > Manual Setup > Info Parameters > Transmitter Info > Write Protect
Offline Maintain > Configuration > Lock
Overview
If the transmitter is write-protected, the configuration is locked and you must unlock it
before you can change any configuration parameters. By default, the transmitter is not
write-protected.
Configuration and Use Manual 19
Introduction to configuration and commissioning
Tip
Write-protecting the transmitter prevents accidental changes to configuration. It does not prevent
normal operational use. You can always disable write-protection, perform any required configuration
changes, then re-enable write-protection.
ProLink IIIDevice Tools > Configuration Transfer > Restore Factory Configuration
Field Communicator Not available
Overview
Not available
Restoring the factory configuration returns the transmitter to a known operational
configuration. This may be useful if you experience problems during configuration.
Tip
Restoring the factory configuration is not a common action. You may want to contact Micro Motion
to see if there is a preferred method to resolve any issues.
20 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
4Configure process measurement
Topics covered in this chapter:
•Configure mass flow measurement
•Configure volume flow measurement for liquid applications
•Configure gas standard volume (GSV) flow measurement
•Configure Flow Direction
•Configure density measurement
•Configure temperature measurement
•Configure pressure compensation
4.1Configure mass flow measurement
The mass flow measurement parameters control how mass flow is measured and reported.
The mass flow measurement parameters include:
•Mass Flow Measurement Unit
•Flow Damping
•Mass Flow Cutoff
4.1.1Configure Mass Flow Measurement Unit
Display (standard)OFF-LINE MAINT > OFF-LINE CONFG > UNITS > MASS
Chinese-language
display
ProLink IIProLink > Configuration > Flow > Mass Flow Unit
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Unit
Overview
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow
rate. The unit used for mass total and mass inventory is derived from this unit.
Procedure
Offline Maintain > Configuration > Units > Mass Flow Rate
Set Mass Flow Measurement Unit to the unit you want to use.
The default setting for Mass Flow Measurement Unit is g/sec (grams per second).
Configuration and Use Manual 21
Configure process measurement
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement
Unit, plus one user-defined special measurement unit. Different communications tools may
use different labels for the units.
Options for Mass Flow Measurement UnitTable 4-1:
Label
Display (stand-
Unit description
Grams per secondG/Sg/secg/secg/secg/s
Grams per minuteG/MINg/ming/ming/ming/min
Grams per hourG/Hg/hrg/hrg/hrg/h
Kilograms per secondKG/Skg/seckg/seckg/seckg/s
Kilograms per minuteKG/MINkg/minkg/minkg/minkg/min
Kilograms per hourKG/Hkg/hrkg/hrkg/hrkg/h
Kilograms per dayKG/Dkg/daykg/daykg/daykg/d
Metric tons per minuteT/MINmTon/minmTon/minmTon/minMetTon/min
Metric tons per hourT/HmTon/hrmTon/hrmTon/hrMetTon/h
Metric tons per dayT/DmTon/daymTon/daymTon/dayMetTon/d
Pounds per secondLB/Slbs/seclbs/seclbs/seclb/s
Pounds per minuteLB/MINlbs/minlbs/minlbs/minlb/min
Pounds per hourLB/Hlbs/hrlbs/hrlbs/hrlb/h
Pounds per dayLB/Dlbs/daylbs/daylbs/daylb/d
Short tons (2000 pounds)
per minute
Short tons (2000 pounds)
per hour
Short tons (2000 pounds)
per day
Long tons (2240 pounds)
per hour
Long tons (2240 pounds)
per day
Special unitSPECLSpecialspecialspecialSpcl
ard)
ST/MINsTon/minsTon/minsTon/minSTon/min
ST/HsTon/hrsTon/hrsTon/hrSTon/h
ST/DsTon/daysTon/daysTon/daySTon/d
LT/HlTon/hrlTon/hrlTon/hrLTon/h
LT/DlTon/daylTon/daylTon/dayLTon/d
Chinese-language display
ProLink IIProLink IIIField Commu-
nicator
22 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
Define a special measurement unit for mass flow
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Special Units
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Mass Special Units
Overview
A special measurement unit is a user-defined unit of measure that allows you to report
process data, totalizer data, and inventory data in a unit that is not available in the
transmitter. A special measurement unit is calculated from an existing measurement unit
using a conversion factor.
Note
Although you cannot define a special measurement unit using the display (standard option), you can
use the standard display to select an existing special measurement unit, and to view process data
using the special measurement unit.
Offline Maintain > Configuration > Units > Special Mass Flow
Procedure
1.Specify Base Mass Unit.
Base Mass Unit is the existing mass unit that the special unit will be based on.
2.Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3.Calculate Mass Flow Conversion Factor as follows:
a. x base units = y special units
b. Mass Flow Conversion Factor = x/y
4.Enter Mass Flow Conversion Factor.
5.Set Mass Flow Label to the name you want to use for the mass flow unit.
6.Set Mass Total Label to the name you want to use for the mass total and mass
inventory unit.
The special measurement unit is stored in the transmitter. You can configure the
transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for mass flow
You want to measure mass flow in ounces per second (oz/sec).
Offline Maintain > Configuration > Damping > Mass Flow Damping
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping
Value specifies the time period (in seconds) over which the transmitter will spread changes
in the reported process variable. At the end of the interval, the reported process variable
will reflect 63% of the change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
The default value is 0.8 seconds. The range depends on the core processor type and the
setting of Update Rate, as shown in the following table.
Core processor typeUpdate Rate settingFlow Damping range
StandardNormal0 to 51.2 seconds
Special0 to 10.24 seconds
EnhancedNot applicable0 to 51.2 seconds
Tips
• A high damping value makes the process variable appear smoother because the reported value
changes slowly.
• A low damping value makes the process variable appear more erratic because the reported value
changes more quickly.
• The combination of a high damping value and rapid, large changes in flow rate can result in
increased measurement error.
24 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
• Whenever the damping value is non-zero, the reported measurement will lag the actual
measurement because the reported value is being averaged over time.
• In general, lower damping values are preferable because there is less chance of data loss, and less
lag time between the actual measurement and the reported value.
• For gas applications, Micro Motion recommends setting Flow Damping to 2.56 or higher.
The value you enter is automatically rounded down to the nearest valid value. Valid
damping values are shown in the following table.
Flow Damping affects volume measurement for liquid volume data. Flow Damping also affects
volume measurement for gas standard volume data. The transmitter calculates volume
data from the damped mass flow data.
Interaction between Flow Damping and Added Damping
In some circumstances, both Flow Damping and Added Damping are applied to the reported
mass flow value.
Flow Damping controls the rate of change in flow process variables. Added Damping controls
the rate of change reported via the mA output. If mA Output Process Variable is set to MassFlow Rate, and both Flow Damping and Added Damping are set to non-zero values, flow
damping is applied first, and the added damping calculation is applied to the result of the
first calculation.
4.1.3Configure Mass Flow Cutoff
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Flow > Mass Flow Cutoff
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Cutoff
Mass Flow Cutoff specifies the lowest mass flow rate that will be reported as measured. All
mass flow rates below this cutoff will be reported as 0.
Procedure
Set Mass Flow Cutoff to the value you want to use.
The default value for Mass Flow Cutoff is 0.0 g/sec or a sensor-specific value set at the
factory. The recommended setting is 0.05% of the sensor's rated maximum flow rate or a
value below the highest expected flow rate. Do not set Mass Flow Cutoff to 0.0 g/sec.
Effect of Mass Flow Cutoff on volume measurement
Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the
actual mass data rather than the reported value.
Interaction between Mass Flow Cutoff and AO Cutoff
Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report asmeasured. AO Cutoff defines the lowest flow rate that will be reported via the mA output. If
mA Output Process Variable is set to Mass Flow Rate, the mass flow rate reported via the mA
output is controlled by the higher of the two cutoff values.
Mass Flow Cutoff affects all reported values and values used in other transmitter behavior
(e.g., events defined on mass flow).
AO Cutoff affects only mass flow values reported via the mA output.
Example: Cutoff interaction with AO Cutoff lower than Mass Flow Cutoff
Configuration:
•mA Output Process Variable: Mass Flow Rate
•Frequency Output Process Variable: Mass Flow Rate
•AO Cutoff: 10 g/sec
•Mass Flow Cutoff: 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0
will be used in all internal processing.
Example: Cutoff interaction with AO Cutoff higher than Mass Flow Cutoff
Configuration:
•mA Output Process Variable: Mass Flow Rate
•Frequency Output Process Variable: Mass Flow Rate
•AO Cutoff: 15 g/sec
•Mass Flow Cutoff: 10 g/sec
Result:
26 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
•If the mass flow rate drops below 15 g/sec but not below 10 g/sec:
-The mA output will report zero flow.
-The frequency output will report the actual flow rate, and the actual flow rate will
be used in all internal processing.
•If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0
will be used in all internal processing.
4.2Configure volume flow measurement for liquid
applications
The volume flow measurement parameters control how liquid volume flow is measured
and reported.
The volume flow measurement parameters include:
•Volume Flow Type
•Volume Flow Measurement Unit
•Volume Flow Cutoff
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. You
must choose one or the other.
4.2.1Configure Volume Flow Type for liquid applications
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Liquid
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be
used.
Procedure
Set Volume Flow Type to Liquid.
Not available
Configuration and Use Manual 27
Configure process measurement
4.2.2Configure Volume Flow Measurement Unit for liquid
applications
Display (standard)OFF-LINE MAINT > OFF-LINE CONFG > UNITS > VOL
Chinese-language
display
ProLink IIProLink > Configuration > Flow > Vol Flow Unit
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Volume Flow Unit
Overview
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the
volume flow rate. The unit used for the volume total and volume inventory is based on this
unit.
Prerequisites
Offline Maintain > Configuration > Units > Volume Flow Rate
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to
Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
The default setting for Volume Flow Measurement Unit is l/sec (liters per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement
Unit, plus one user-defined measurement unit. Different communications tools may use
different labels for the units.
Options for Volume Flow Measurement Unit for liquid applicationsTable 4-3:
Label
Display (stand-
Unit description
Cubic feet per secondCUFT/Sft3/secft3/secft3/secCuft/s
Cubic feet per minuteCUF/MNft3/minft3/minft3/minCuft/min
Cubic feet per hourCUFT/Hft3/hrft3/hrft3/hrCuft/h
Cubic feet per dayCUFT/Dft3/dayft3/dayft3/dayCuft/d
ard)
Chinese-language display
ProLink IIProLink IIIField Commu-
nicator
28 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
Options for Volume Flow Measurement Unit for liquid applications (continued)Table 4-3:
Label
Unit description
Display (standard)
Chinese-language display
ProLink IIProLink IIIField Commu-
nicator
Cubic meters per secondM3/Sm3/secm3/secm3/secCum/s
Cubic meters per minuteM3/MINm3/minm3/minm3/minCum/min
Cubic meters per hourM3/Hm3/hrm3/hrm3/hrCum/h
Cubic meters per dayM3/Dm3/daym3/daym3/dayCum/d
U.S. gallons per secondUSGPSUS gal/secUS gal/secUS gal/secgal/s
U.S. gallons per minuteUSGPMUS gal/minUS gal/minUS gal/mingal/min
U.S. gallons per hourUSGPHUS gal/hrUS gal/hrUS gal/hrgal/h
U.S. gallons per dayUSGPDUS gal/dayUS gal/dayUS gal/daygal/d
Million U.S. gallons per
MILG/Dmil US gal/daymil US gal/daymil US gal/dayMMgal/d
day
Liters per secondL/Sl/secl/secl/secL/s
Liters per minuteL/MINl/minl/minl/minL/min
Liters per hourL/Hl/hrl/hrl/hrL/h
Million liters per dayMILL/Dmil l/daymil l/daymil l/dayML/d
Imperial gallons per sec-
UKGPSImp gal/secImp gal/secImp gal/secImpgal/s
ond
Imperial gallons per mi-
UKGPMImp gal/minImp gal/minImp gal/minImpgal/min
nute
Imperial gallons per hourUKGPHImp gal/hrImp gal/hrImp gal/hrImpgal/h
Imperial gallons per dayUKGPDImp gal/dayImp gal/dayImp gal/dayImpgal/d
Barrels per second
(1)
BBL/Sbarrels/secbarrels/secbarrels/secbbl/s
Barrels per minuteBBL/MNbarrels/minbarrels/minbarrels/minbbl/min
Barrels per hourBBL/Hbarrels/hrbarrels/hrbarrels/hrbbl/h
Barrels per dayBBL/Dbarrels/daybarrels/daybarrels/daybbl/d
Beer barrels per minuteBBBL/MNBeer barrels/minBeer barrels/minBeer barrels/minbbbl/min
Beer barrels per hourBBBL/HBeer barrels/hrBeer barrels/hrBeer barrels/hrbbbl/h
Beer barrels per dayBBBL/DBeer barrels/dayBeer barrels/dayBeer barrels/daybbbl/d
Special unitSPECLSpecialspecialspecialSpcl
(1) Unit based on oil barrels (42 U.S. gallons).
(2) Unit based on U.S. beer barrels (31 U.S. gallons).
Configuration and Use Manual 29
Configure process measurement
Define a special measurement unit for volume flow
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Special Units
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Volume Special Units
Overview
A special measurement unit is a user-defined unit of measure that allows you to report
process data, totalizer data, and inventory data in a unit that is not available in the
transmitter. A special measurement unit is calculated from an existing measurement unit
using a conversion factor.
Note
Although you cannot define a special measurement unit using the display, you can use the display to
select an existing special measurement unit, and to view process data using the special
measurement unit.
Not available
Procedure
1.Specify Base Volume Unit.
Base Volume Unit is the existing volume unit that the special unit will be based on.
2.Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3.Calculate Volume Flow Conversion Factor as follows:
a. x base units = y special units
b. Volume Flow Conversion Factor = x/y
4.Enter Volume Flow Conversion Factor.
5.Set Volume Flow Label to the name you want to use for the volume flow unit.
6.Set Volume Total Label to the name you want to use for the volume total and volume
inventory unit.
The special measurement unit is stored in the transmitter. You can configure the
transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for volume flow
You want to measure volume flow in pints per second (pints/sec).
1.Set Base Volume Unit to Gallons (gal).
2.Set Base Time Unit to Seconds (sec).
30 Micro Motion® Model 1700 Transmitters with Analog Outputs
Volume Flow Cutoff specifies the lowest volume flow rate that will be reported as measured.
All volume flow rates below this cutoff are reported as 0.
Procedure
Set Volume Flow Cutoff to the value you want to use.
The default value for Volume Flow Cutoff is 0.0 l/sec (liters per second). The lower limit is 0.
The upper limit is the sensor’s flow calibration factor, in units of l/sec, multiplied by 0.2.
Interaction between Volume Flow Cutoff and AO Cutoff
Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will
report as measured. AO Cutoff defines the lowest flow rate that will be reported via the mA
output. If mA Output Process Variable is set to Volume Flow Rate, the volume flow rate reported
via the mA output is controlled by the higher of the two cutoff values.
Volume Flow Cutoff affects both the volume flow values reported via the outputs and the
volume flow values used in other transmitter behavior (e.g., events defined on the volume
flow).
AO Cutoff affects only flow values reported via the mA output.
Example: Cutoff interaction with AO Cutoff lower than Volume Flow Cutoff
Configuration:
•mA Output Process Variable: Volume Flow Rate
•Frequency Output Process Variable: Volume Flow Rate
Configuration and Use Manual 31
Configure process measurement
•AO Cutoff: 10 l/sec
•Volume Flow Cutoff: 15 l/sec
Result: If the volume flow rate drops below 15 l/sec, volume flow will be reported as 0, and
0 will be used in all internal processing.
Example: Cutoff interaction with AO Cutoff higher than Volume Flow Cutoff
Configuration:
•mA Output Process Variable: Volume Flow Rate
•Frequency Output Process Variable: Volume Flow Rate
•AO Cutoff: 15 l/sec
•Volume Flow Cutoff: 10 l/sec
Result:
•If the volume flow rate drops below 15 l/sec but not below 10 l/sec:
•If the volume flow rate drops below 10 l/sec, both outputs will report zero flow, and
-The mA output will report zero flow.
-The frequency output will report the actual flow rate, and the actual flow rate will
be used in all internal processing.
0 will be used in all internal processing.
4.3Configure gas standard volume (GSV) flow
measurement
The gas standard volume (GSV) flow measurement parameters control how gas standard
volume flow is measured and reported.
The GSV flow measurement parameters include:
•Volume Flow Type
•Standard Gas Density
•Gas Standard Volume Flow Measurement Unit
•Gas Standard Volume Flow Cutoff
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. You
must choose one or the other.
32 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
4.3.1Configure Volume Flow Type for gas applications
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Flow > Vol Flow Type
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Standard Gas Volume
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement is
used.
Procedure
Set Volume Flow Type to Gas Standard Volume.
Not available
4.3.2Configure Standard Gas Density
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Flow > Std Gas Density
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Gas Ref Density
Overview
The Standard Gas Density value is used to convert the measured flow data to the standard
reference values.
Prerequisites
Ensure that Density Measurement Unit is set to the measurement unit you want to use for
Standard Gas Density.
Procedure
Set Standard Gas Density to the standard reference density of the gas you are measuring.
Not available
Note
ProLink II and ProLink III provide a guided method that you can use to calculate the standard density
of your gas, if you do not know it.
Configuration and Use Manual 33
Configure process measurement
4.3.3Configure Gas Standard Volume Flow Measurement Unit
Display (standard)OFF-LINE MAINT > OFF-LINE CONFG > UNITS > GSV
Chinese-language
display
ProLink IIProLink > Configuration > Flow > Std Gas Vol Flow Unit
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > GSV Flow Unit
Overview
Gas Standard Volume Flow Measurement Unit specifies the unit of measure that will be displayed
for the gas standard volume flow rate. The measurement unit used for the gas standard
volume total and the gas standard volume inventory is derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Measurement Unit, be sure that Volume Flow Type
is set to Gas Standard Volume.
Offline Maintain > Configuration > Units > Gas Std Volume Flow
Procedure
Set Gas Standard Volume Flow Measurement Unit to the unit you want to use.
The default setting for Gas Standard Volume Flow Measurement Unit is SCFM (Standard Cubic
Feet per Minute).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Gas Standard Volume Flow Measurement Unit
The transmitter provides a standard set of measurement units for Gas Standard Volume Flow
Measurement Unit, plus one user-defined special measurement unit. Different
communications tools may use different labels for the units.
Options for Gas Standard Volume Measurement UnitTable 4-4:
Unit description
Normal cubic meters per
second
Normal cubic meters per
minute
Normal cubic meters per
hour
Label
Display (standard)
NM3/SNm3/secNm3/secNm3/secNm3/sec
NM3/MNNm3/minNm3/minNm3/secNm3/min
NM3/HNm3/hrNm3/hrNm3/hrNm3/hr
Chinese-language display
ProLink IIProLink IIIField Commu-
nicator
34 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
Options for Gas Standard Volume Measurement Unit (continued)Table 4-4:
Label
Display (stand-
Unit description
Normal cubic meters per
day
Normal liter per secondNLPSNLPSNLPSNLPSNLPS
Normal liter per minuteNLPMNLPMNLPMNLPMNLPM
Normal liter per hourNLPHNLPHNLPHNLPHNLPH
Normal liter per dayNLPDNLPDNLPDNLPDNLPD
Standard cubic feet per
second
Standard cubic feet per
minute
Standard cubic feet per
hour
Standard cubic feet per
day
Standard cubic meters
per second
Standard cubic meters
per minute
Standard cubic meters
per hour
Standard cubic meters
per day
Standard liter per secondSLPSSLPSSLPSSLPSSLPS
Standard liter per minuteSLPMSLPMSLPMSLPMSLPM
Standard liter per hourSLPHSLPHSLPHSLPHSLPH
Standard liter per daySLPDSLPDSLPDSLPDSLPD
Special measurement
unit
ard)
NM3/DNm3/dayNm3/dayNm3/dayNm3/day
SCFSSCFSSCFSSCFSSCFS
SCFMSCFMSCFMSCFMSCFM
SCFHSCFHSCFHSCFHSCFH
SCFDSCFDSCFDSCFDSCFD
SM3/SSm3/secSm3/SSm3/secSm3/sec
SM3/MNSm3/minSm3/minSm3/minSm3/min
SM3/HSm3/hrSm3/hrSm3/hrSm3/hr
SM3/DSm3/daySm3/daySm3/daySm3/day
SPECLSpecialspecialspecialSpecial
Chinese-language display
ProLink IIProLink IIIField Commu-
nicator
Configuration and Use Manual 35
Configure process measurement
Define a special measurement unit for gas standard volume
flow
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Special Units
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Special GSV Units
Overview
A special measurement unit is a user-defined unit of measure that allows you to report
process data, totalizer data, and inventory data in a unit that is not available in the
transmitter. A special measurement unit is calculated from an existing measurement unit
using a conversion factor.
Not available
Note
Although you cannot define a special measurement unit using the display, you can use the display to
select an existing special measurement unit, and to view process data using the special
measurement unit.
Procedure
1.Specify Base Gas Standard Volume Unit.
Base Gas Standard Volume Unit is the existing gas standard volume unit that the special
unit will be based on.
2.Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3.Calculate Gas Standard Volume Flow Conversion Factor as follows:
a. x base units = y special units
b. Gas Standard Volume Flow Conversion Factor = x/y
4.Enter the Gas Standard Volume Flow Conversion Factor.
5.Set Gas Standard Volume Flow Label to the name you want to use for the gas standard
volume flow unit.
6.Set Gas Standard Volume Total Label to the name you want to use for the gas standard
volume total and gas standard volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the
transmitter to use the special measurement unit at any time.
36 Micro Motion® Model 1700 Transmitters with Analog Outputs
Example: Defining a special measurement unit for gas standard volume flow
You want to measure gas standard volume flow in thousands of standard cubic feet per
minute.
1.Set Base Gas Standard Volume Unit to SCFM.
2.Set Base Time Unit to minutes (min).
3.Calculate the conversion factor:
a. 1 thousands of standard cubic feet per minute = 1000 cubic feet per minute
b. Gas Standard Volume Flow Conversion Factor = 1/1000 = 0.001
4.Set Gas Standard Volume Flow Conversion Factor to 0.001.
Gas Standard Volume Flow Cutoff specifies the lowest gas standard volume flow rate that will
reported as measured. All gas standard volume flow rates below this cutoff will be
reported as 0.
Procedure
Set Gas Standard Volume Flow Cutoff to the value you want to use.
The default value for Gas Standard Volume Flow Cutoff is 0.0. The lower limit is 0.0. There is no
upper limit.
Interaction between Gas Standard Volume Flow Cutoff and AO
Cutoff
Gas Standard Volume Flow Cutoff defines the lowest Gas Standard Volume flow value that the
transmitter will report as measured. AO Cutoff defines the lowest flow rate that will be
reported via the mA output. If mA Output Process Variable is set to Gas Standard Volume FlowRate, the volume flow rate reported via the mA output is controlled by the higher of the
two cutoff values.
Configuration and Use Manual 37
Configure process measurement
Gas Standard Volume Flow Cutoff affects both the gas standard volume flow values reported
via outputs and the gas standard volume flow values used in other transmitter behavior
(e.g., events defined on gas standard volume flow).
AO Cutoff affects only flow values reported via the mA output.
Example: Cutoff interaction with AO Cutoff lower than Gas Standard Volume Flow Cutoff
Configuration:
•mA Output Process Variable for the primary mA output: Gas Standard Volume Flow Rate
•Frequency Output Process Variable: Gas Standard Volume Flow Rate
•AO Cutoff for the primary mA output: 10 SLPM (standard liters per minute)
•Gas Standard Volume Flow Cutoff: 15 SLPM
Result: If the gas standard volume flow rate drops below 15 SLPM, the volume flow will be
reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with AO Cutoff higher than Gas Standard Volume Flow Cutoff
Configuration:
•mA Output Process Variable for the primary mA output: Gas Standard Volume Flow Rate
•Frequency Output Process Variable: Gas Standard Volume Flow Rate
•AO Cutoff for the primary mA output: 15 SLPM (standard liters per minute)
•Gas Standard Volume Flow Cutoff: 10 SLPM
Result:
•If the gas standard volume flow rate drops below 15 SLPM but not below 10 SLPM:
-The primary mA output will report zero flow.
-The frequency output will report the actual flow rate, and the actual flow rate will
be used in all internal processing.
•If the gas standard volume flow rate drops below 10 SLPM, both outputs will report
zero flow, and 0 will be used in all internal processing.
4.4Configure Flow Direction
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Flow > Flow Direction
ProLink IIIDevice Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Flow Direction
Not available
38 Micro Motion® Model 1700 Transmitters with Analog Outputs
Overview
Flow Direction controls how forward flow and reverse flow affect flow measurement and
reporting.
Flow Direction is defined with respect to the flow arrow on the sensor:
•Forward flow (positive flow) moves in the direction of the flow arrow on the sensor.
•Reverse flow (negative flow) moves in the direction opposite to the flow arrow on
the sensor.
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow
direction or the setting of the Flow Direction parameter.
Procedure
Set Flow Direction to the value you want to use.
4.4.1Options for Flow Direction
Configure process measurement
Options for Flow DirectionTable 4-5:
Flow Direction settingRelationship to Flow Direction ar-
ProLink IIProLink IIIField Communicator
ForwardForwardForwardAppropriate when the Flow Direction
ReverseReverseReverseAppropriate when the Flow Direction
Absolute ValueAbsolute ValueAbsolute ValueFlow Direction arrow is not relevant.
BidirectionalBidirectionalBi directionalAppropriate when both forward and
Negate ForwardNegate ForwardNegate/Forward OnlyAppropriate when the Flow Direction
Negate BidirectionalNegate BidirectionalNegate/Bi-directionalAppropriate when both forward and
row on sensor
arrow is in the same direction as the
majority of flow.
arrow is in the same direction as the
majority of flow.
reverse flow are expected, and forward flow will dominate, but the
amount of reverse flow will be significant.
arrow is in the opposite direction from
the majority of flow.
reverse flow are expected, and reverse
flow will dominate, but the amount of
forward flow will be significant.
Configuration and Use Manual 39
Configure process measurement
Effect of Flow Direction on mA outputs
Flow Direction affects how the transmitter reports flow values via the mA outputs. The mA
outputs are affected by Flow Direction only if mA Output Process Variable is set to a flow
variable.
Flow Direction and mA outputs
The effect of Flow Direction on the mA outputs depend on Lower Range Value configured for
the mA output:
•If Lower Range Value is set to 0, see Figure 1.
•If Lower Range Value is set to a negative value, see Figure 2.
Effect of Flow Direction on the mA output: Lower Range Value = 0Figure 4-1:
Flow Direction = Forward
20
12
mA output
4
-x0x
Reverse flowForward flow
•Lower Range Value = 0
•Upper Range Value = x
Flow Direction = Reverse, Negate Forward
20
12
mA output
4
-x0x
Reverse flowForward flow
Flow Direction = Absolute Value, Bidirectional,
Negate Bidirectional
20
12
mA output
4
-x0x
Reverse flowForward flow
40 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
Effect of Flow Direction on the mA output: Lower Range Value < 0Figure 4-2:
Flow Direction = Forward
20
12
mA output
4
-x0x
Reverse flowForward flow
•Lower Range Value = −x
•Upper Range Value = x
Example: Flow Direction = Forward and Lower Range Value = 0
Configuration:
•Flow Direction = Forward
•Lower Range Value = 0 g/sec
•Upper Range Value = 100 g/sec
Flow Direction = Reverse, Negate Forward
20
12
mA output
4
-x0x
Reverse flowForward flow
Flow Direction = Absolute Value, Bidirectional,
Negate Bidirectional
20
12
mA output
4
-x0x
Reverse flowForward flow
Result:
•Under conditions of reverse flow or zero flow, the mA output is 4 mA.
•Under conditions of forward flow, up to a flow rate of 100 g/sec, the mA output
varies between 4 mA and 20 mA in proportion to the flow rate.
•Under conditions of forward flow, if the flow rate equals or exceeds 100 g/sec, the
mA output will be proportional to the flow rate up to 20.5 mA, and will be level at
20.5 mA at higher flow rates.
Example: Flow Direction = Forward and Lower Range Value < 0
Configuration:
•Flow Direction = Forward
•Lower Range Value = −100 g/sec
•Upper Range Value = +100 g/sec
Result:
•Under conditions of zero flow, the mA output is 12 mA.
•Under conditions of forward flow, for flow rates between 0 and +100 g/sec, the mA
output varies between 12 mA and 20 mA in proportion to (the absolute value of) the
flow rate.
Configuration and Use Manual 41
Configure process measurement
•Under conditions of forward flow, if (the absolute value of) the flow rate equals or
•Under conditions of reverse flow, for flow rates between 0 and −100 g/sec, the mA
•Under conditions of reverse flow, if the absolute value of the flow rate equals or
Example: Flow Direction = Reverse
Configuration:
•Flow Direction = Reverse
•Lower Range Value = 0 g/sec
•Upper Range Value = 100 g/sec
Result:
•Under conditions of forward flow or zero flow, the mA output is 4 mA.
•Under conditions of reverse flow, for flow rates between 0 and +100 g/sec, the mA
•Under conditions of reverse flow, if the absolute value of the flow rate equals or
exceeds 100 g/sec, the mA output is proportional to the flow rate up to 20.5 mA,
and will be level at 20.5 mA at higher flow rates.
output varies between 4 mA and 12 mA in inverse proportion to the absolute value
of the flow rate.
exceeds 100 g/sec, the mA output is inversely proportional to the flow rate down to
3.8 mA, and will be level at 3.8 mA at higher absolute values.
output level varies between 4 mA and 20 mA in proportion to the absolute value of
the flow rate.
exceeds 100 g/sec, the mA output will be proportional to the absolute value of the
flow rate up to 20.5 mA, and will be level at 20.5 mA at higher absolute values.
Effect of Flow Direction on frequency outputs
Flow Direction affects how the transmitter reports flow values via the frequency outputs. The
frequency outputs are affected by Flow Direction only if Frequency Output Process Variable is set
to a flow variable.
Table 4-6:
Flow Direction setting
ForwardHz > 00 Hz0 Hz
Reverse0 Hz0 HzHz > 0
BidirectionalHz > 00 HzHz > 0
Absolute ValueHz > 00 HzHz > 0
Negate Forward0 Hz0 HzHz > 0
Negate BidirectionalHz > 00 HzHz > 0
Effect of the Flow Direction parameter and actual flow direction on frequency
outputs
Actual flow direction
ForwardZero flowReverse
42 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
Effect of Flow Direction on discrete outputs
The Flow Direction parameter affects the discrete output behavior only if Discrete Output
Source is set to Flow Direction.
Table 4-7:
Effect of the Flow Direction parameter and actual flow direction on discrete
outputs
Actual flow direction
Flow Direction setting
ForwardOFFOFFON
ReverseOFFOFFON
BidirectionalOFFOFFON
Absolute ValueOFFOFFOFF
Negate ForwardONOFFOFF
Negate BidirectionalONOFFOFF
ForwardZero flowReverse
Effect of Flow Direction on digital communications
Flow Direction affects how flow values are reported via digital communications.
Table 4-8:
Effect of the Flow Direction parameter and actual flow direction on flow
values reported via digital communications
Actual flow direction
Flow Direction setting
ForwardPositive0Negative
ReversePositive0Negative
BidirectionalPositive0Negative
Absolute ValuePositive
Negate ForwardNegative0Positive
Negate BidirectionalNegative0Positive
ForwardZero flowReverse
(3)
0Positive
Effect of Flow Direction on flow totals
Flow Direction affects how flow totals and inventories are calculated.
(3) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
Configuration and Use Manual 43
Configure process measurement
Table 4-9:
Effect of the Flow Direction parameter and actual flow direction on flow
totals and inventories
Actual flow direction
Flow Direction setting
ForwardTotals increaseTotals do not changeTotals do not change
ReverseTotals do not changeTotals do not changeTotals increase
BidirectionalTotals increaseTotals do not changeTotals decrease
Absolute ValueTotals increaseTotals do not changeTotals increase
Negate ForwardTotals do not changeTotals do not changeTotals increase
Negate BidirectionalTotals decreaseTotals do not changeTotals increase
ForwardZero flowReverse
4.5Configure density measurement
The density measurement parameters control how density is measured and reported.
Density measurement (along with mass measurement) is used to determine liquid volume
flow.
The density measurement parameters include:
•Density Measurement Unit
•Slug Flow Parameters
•Density Damping
•Density Cutoff
4.5.1Configure Density Measurement Unit
Display (standard)OFF-LINE MAINT > OFF-LINE CONFG > UNITS > DENS
Chinese-language
display
ProLink IIProLink > Configuration > Density > Dens Unit
ProLink IIIDevice Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Unit
Overview
Density Measurement Unit specifies the units of measure that will be displayed for density
measurement.
Procedure
Set Density Measurement Unit to the option you want to use.
44 Micro Motion® Model 1700 Transmitters with Analog Outputs
Offline Maintain > Configuration > Units > Density
Configure process measurement
The default setting for Density Measurement Unit is g/cm3 (grams per cubic centimeter).
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement Unit.
Different communications tools may use different labels.
Options for Density Measurement UnitTable 4-10:
Label
Display (stand-
Unit description
Specific gravity unit (not
temperature-corrected)
Grams per cubic centimeter
Grams per literG/Lg/Lg/lg/lg/L
Grams per milliliterG/mLg/mLg/mlg/mlg/mL
Kilograms per literKG/Lkg/Lkg/lkg/lkg/L
Kilograms per cubic meter
Pounds per U.S. gallonLB/GALlbs/US gallbs/Usgallbs/Usgallb/gal
Pounds per cubic footLB/CUFlbs/ft3lbs/ft3lbs/ft3lb/Cuft
Pounds per cubic inchLB/CUIlbs/in3lbs/in3lbs/in3lb/CuIn
API gravityD APIdeg APIdegAPIdegAPIdegAPI
Short ton per cubic yardST/CUYsTon/yd3sT/yd3sT/yd3STon/Cuyd
ard)
SGUSGUSGUSGUSGU
G/CM3g/cm3g/cm3g/cm3g/Cucm
KG/M3kg/m3kg/m3kg/m3kg/Cum
Chinese-language display
ProLink IIProLink IIIField Commu-
nicator
4.5.2Configure slug flow parameters
Display (standard)Not available
Chinese-language
display
ProLink II• ProLink > Configuration > Density > Slug High Limit
ProLink IIIDevice Tools > Configuration > Process Measurement > Density
Field Communicator • Configure > Manual Setup > Measurements > Density > Slug Low Limit
Not available
• ProLink > Configuration > Density > Slug Low Limit
• ProLink > Configuration > Density > Slug Duration
• Configure > Manual Setup > Measurements > Density > Slug High Limit
The slug flow parameters control how the transmitter detects and reports two-phase flow
(gas in a liquid process or liquid in a gas process).
Procedure
1.Set Slug Low Limit to the lowest density value that is considered normal in your
process.
Values below this will cause the transmitter to perform the configured slug flow
action. Typically, this value is the lowest density value in the normal range of your
process.
Tip
Gas entrainment can cause your process density to drop temporarily. To reduce the
occurrence of slug flow alarms that are not significant to your process, set Slug Low Limit
slightly below your expected lowest process density.
You must enter Slug Low Limit in g/cm3, even if you configured another unit for
density measurement.
The default value for Slug Low Limit is 0.0 g/cm3. The range is 0.0 to 10.0 g/cm3.
2.Set Slug High Limit to the highest density value that is considered normal in your
process.
Values above this will cause the transmitter to perform the configured slug flow
action. Typically, this value is the highest density value in the normal range of your
process.
Tip
To reduce the occurrence of slug flow alarms that are not significant to your process, set Slug
High Limit slightly above your expected highest process density.
You must enter Slug High Limit in g/cm3, even if you configured another unit for
density measurement.
The default value for Slug High Limit is 5.0 g/cm3. The range is 0.0 to 10.0 g/cm3.
3.Set Slug Duration to the number of seconds that the transmitter will wait for a slug
flow condition to clear before performing the configured slug flow action.
The default value for Slug Duration is 0.0 seconds. The range is 0.0 to 60.0 seconds.
Slug flow detection and reporting
Slug flow is typically used as an indicator of two-phase flow (gas in a liquid process or liquid
in a gas process). Two-phase flow can cause a variety of process control issues. By
configuring the slug flow parameters appropriately for your application, you can detect
process conditions that require correction.
46 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
Tip
To decrease the occurrence of slug flow alarms, lower Slug Low Limit or raise Slug High Limit.
A slug flow condition occurs whenever the measured density goes below Slug Low Limit or
above Slug High Limit. If this occurs:
•A slug flow alarm is posted to the active alarm log.
•All outputs that are configured to represent flow rate hold their last “pre-slug flow”
value for the configured Slug Duration.
If the slug flow condition clears before Slug Duration expires:
•Outputs that represent flow rate revert to reporting actual flow.
•The slug flow alarm is deactivated, but remains in the active alarm log until it is
acknowledged.
If the slug flow condition does not clear before Slug Duration expires, the outputs that
represent flow rate report a flow rate of 0.
If Slug Duration is set to 0.0 seconds, the outputs that represent flow rate will report a flow
rate of 0 as soon as slug flow is detected.
4.5.3Configure Density Damping
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Density > Density Damping
ProLink IIIDevice Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Damping
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping
Value specifies the time period (in seconds) over which the transmitter will spread changes
in the reported process variable. At the end of the interval, the reported process variable
will reflect 63% of the change in the actual measured value.
Procedure
Set Density Damping to the value you want to use.
The default value is 1.6 seconds. The range depends on the core processor type and the
setting of Update Rate, as shown in the following table:
Offline Maintain > Configuration > Damping > Density Damping
Core processor typeUpdate Rate settingDensity Damping range
StandardNormal0 to 51.2 seconds
Special0 to 10.24 seconds
Configuration and Use Manual 47
Configure process measurement
Core processor typeUpdate Rate settingDensity Damping range
EnhancedNot applicable0 to 40.96 seconds
Tips
• A high damping value makes the process variable appear smoother because the reported value
changes slowly.
• A low damping value makes the process variable appear more erratic because the reported value
changes more quickly.
• Whenever the damping value is non-zero, the reported measurement will lag the actual
measurement because the reported value is being averaged over time.
• In general, lower damping values are preferable because there is less chance of data loss, and less
lag time between the actual measurement and the reported value.
The value you enter is automaticaly rounded down to the nearest valid value. The valid
values for Density Damping depend on the setting of Update Rate.
Density Damping affects liquid volume measurement. Liquid volume values are calculated
from the damped density value rather than the measured density value. Density Damping
does not affect gas standard volume measurement.
Interaction between Density Damping and Added Damping
In some circumstances, both Density Damping and Added Damping are applied to the reported
density value.
Density Damping controls the rate of change in the density process variable. Added Damping
controls the rate of change reported via the mA output. If mA Output Process Variable is set to
Density, and both Density Damping and Added Damping are set to non-zero values, density
damping is applied first, and the added damping calculation is applied to the result of the
first calculation.
48 Micro Motion® Model 1700 Transmitters with Analog Outputs
4.5.4Configure Density Cutoff
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Density > Low Density Cutoff
ProLink IIIDevice Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Cutoff
Overview
Density Cutoff specifies the lowest density value that will be reported as measured. All
density values below this cutoff will be reported as 0.
The default value for Density Cutoff is 0.2 g/cm3. The range is 0.0 g/cm3 to 0.5 g/cm3.
Effect of Density Cutoff on volume measurement
Density Cutoff affects liquid volume measurement. If the density value goes below Density
Cutoff, the volume flow rate is reported as 0. Density Cutoff does not affect gas standard
volume measurement. Gas standard volume values are always calculated from the value
configured for Standard Gas Density.
4.6Configure temperature measurement
The temperature measurement parameters control how temperature data from the
sensor is reported. Temperature data is used to compensate for the effect of temperature
on the sensor tubes during flow measurement .
The temperature measurement parameters include:
•Temperature Measurement Unit
•Temperature Damping
Configuration and Use Manual 49
Configure process measurement
4.6.1Configure Temperature Measurement Unit
Display (standard)OFF-LINE MAINT > OFF-LINE CONFG > UNITS > TEMP
Chinese-language
display
ProLink IIProLink > Configuration > Temperature > Temp Unit
ProLink IIIDevice Tools > Configuration > Process Measurement > Temperature
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temperature Unit
Overview
Temperature Measurement Unit specifies the unit that will be used for temperature
measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
Offline Maintain > Configuration > Units > Temperature
The default setting is Degrees Celsius.
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit. Different
communications tools may use different labels for the units.
Options for Temperature Measurement UnitTable 4-12:
Label
Chinese-lan-
Display
Unit description
Degrees Celsius°C°CdegC°CdegC
Degrees Fahrenheit°F°FdegF°FdegF
Degrees Rankine°R°RdegR°RdegR
Kelvin°K°KdegK°KKelvin
(standard)
guage displayProLink IIProLink III
Field Communicator
50 Micro Motion® Model 1700 Transmitters with Analog Outputs
4.6.2Configure Temperature Damping
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Temperature > Temp Damping
ProLink IIIDevice Tools > Configuration > Temperature
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temp Damping
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping
Value specifies the time period (in seconds) over which the transmitter will spread changes
in the reported process variable. At the end of the interval, the reported process variable
will reflect 63% of the change in the actual measured value.
Procedure
Offline Maintain > Configuration > Damping > Temperature Damping
Configure process measurement
Enter the value you want to use for Temperature Damping.
The default value is 4.8 seconds. The range is 0.0 to 76.8 seconds.
Tips
• A high damping value makes the process variable appear smoother because the reported value
changes slowly.
• A low damping value makes the process variable appear more erratic because the reported value
changes more quickly.
• Whenever the damping value is non-zero, the reported measurement will lag the actual
measurement because the reported value is being averaged over time.
• In general, lower damping values are preferable because there is less chance of data loss, and less
lag time between the actual measurement and the reported value.
The value you enter is automaticaly rounded down to the nearest valid value. Valid values
for Temperature Damping are 0, 0.6, 1.2, 2.4, 4.8, … 76.8.
Effect of Temperature Damping on process measurement
Temperature Damping affects the response speed for temperature compensation with
fluctuating temperatures. Temperature compensation adjusts the process measurement
to compensate for the effect of temperature on the sensor tube.
Configuration and Use Manual 51
Configure process measurement
4.7Configure pressure compensation
Pressure compensation adjusts process measurement to compensate for the pressure
effect on the sensor. The pressure effect is the change in the sensor’s sensitivity to flow
and density caused by the difference between the calibration pressure and the process
pressure.
Tip
Not all sensors or applications require pressure compensation. The pressure effect for a specific
sensor model can be found in the product data sheet located at www.micromotion.com. If you are
uncertain about implementing pressure compensation, contact Micro Motion customer service.
4.7.1Configure pressure compensation using ProLink II
Prerequisites
You will need the flow factor, density factor, and calibration pressure values for your
sensor.
•For the flow factor and density factor, see the product data sheet for your sensor.
•For the calibration pressure, see the calibration sheet for your sensor. If the data is
unavailable, use 20 PSI.
Procedure
1.Choose View > Preferences and ensure that Enable External Pressure Compensation is
checked.
2.Choose ProLink > Configuration > Pressure.
3.Enter Flow Factor for your sensor.
The flow factor is the percent change in the flow rate per PSI. When entering the
value, reverse the sign.
Example:
If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI.
4.Enter Density Factor for your sensor.
The density factor is the change in fluid density, in g/cm3/PSI. When entering the
value, reverse the sign.
Example:
If the density factor is 0.000006 g/cm3/PSI, enter −0.000006 g/cm3/PSI.
5.Enter Cal Pressure for your sensor.
The calibration pressure is the pressure at which your sensor was calibrated, and
defines the pressure at which there is no pressure effect. If the data is unavailable,
enter 20 PSI.
52 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
6.Determine how the transmitter will obtain pressure data, and perform the required
setup.
OptionSetup
A user-configured
static pressure value
Polling for pres-
(4)
sure
a. Set Pressure Units to the desired unit.
b. Set External Pressure to the desired value.
a. Ensure that the primary mA output has been wired to support
HART polling.
b. Choose ProLink > Configuration > Polled Variables.
c. Choose an unused polling slot.
d. Set Polling Control to Poll As Primary or Poll as Secondary, and click
Apply.
e. Set External Tag to the HART tag of the external pressure device.
f. Set Variable Type to Pressure.
Tip
• Poll as Primary: No other HART masters will be on the network.
• Poll as Secondary: Other HART masters will be on the network.
The Field Communicator is not a HART master.
A value written by
digital communications
a. Set Pressure Units to the desired unit.
b. Perform the necessary host programming and communications
setup to write pressure data to the transmitter at appropriate in-
tervals.
Postrequisites
If you are using an external pressure value, verify the setup by choosing ProLink > Process
Variables and checking the value displayed in External Pressure.
4.7.2Configure pressure compensation using ProLink III
Prerequisites
You will need the flow factor, density factor, and calibration pressure values for your
sensor.
•For the flow factor and density factor, see the product data sheet for your sensor.
•For the calibration pressure, see the calibration sheet for your sensor. If the data is
3.Enter Flow Calibration Pressure for your sensor.
(4) Not available on all transmitters.
Configuration and Use Manual 53
Configure process measurement
4.Enter Flow Factor for your sensor.
5.Enter Density Factor for your sensor.
6.Set Pressure Source to the method that the transmitter will use to obtain pressure
The calibration pressure is the pressure at which your sensor was calibrated, and
defines the pressure at which there is no pressure effect. If the data is unavailable,
enter 20 PSI.
The flow factor is the percent change in the flow rate per PSI. When entering the
value, reverse the sign.
Example:
If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI.
The density factor is the change in fluid density, in g/cm3/PSI. When entering the
value, reverse the sign.
Example:
If the density factor is 0.000006 g/cm3/PSI, enter −0.000006 g/cm3/PSI.
data.
OptionDescription
Poll for external value
Static or Digital CommunicationsThe transmitter will use the pressure value that it reads from
(5)
The transmitter will poll an external pressure device, using
HART protocol over the primary mA output.
memory.
• Static: The configured value is used.
• Digital Communications: A host writes transmitter data
to transmitter memory.
7.If you chose to poll for pressure data:
a. Select the Polling Slot to use.
b. Set Polling Control to Poll as Primary or Poll as Secondary, and click Apply.
Tip
• Poll as Primary: No other HART masters will be on the network.
• Poll as Secondary: Other HART masters will be on the network. The Field Communicator
is not a HART master.
c. Set External Device Tag to the HART tag of the external pressure device, and click
Apply.
8.If you chose to use a static pressure value:
a. Set Pressure Unit to the desired unit.
(5) Not available on all transmitters.
54 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
b. Set Static or Current Pressure to the value to use, and click Apply
9.If you want to use digital communications, click Apply, then perform the necessary
host programming and communications setup to write pressure data to the
transmitter at appropriate intervals.
Postrequisites
If you are using an external pressure value, verify the setup by checking the External Pressure
value displayed in the Inputs area of the main window.
4.7.3Configure pressure compensation using the
Field Communicator
Prerequisites
You will need the flow factor, density factor, and calibration pressure values for your
sensor.
•For the flow factor and density factor, see the product data sheet for your sensor.
•For the calibration pressure, see the calibration sheet for your sensor. If the data is
The calibration pressure is the pressure at which your sensor was calibrated, and
defines the pressure at which there is no pressure effect. If the data is unavailable,
enter 20 PSI.
4.Enter Flow Press Factor for your sensor.
The flow factor is the percent change in the flow rate per PSI. When entering the
value, reverse the sign.
Example:
If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI.
5.Enter Dens Press Factor for your sensor.
The density factor is the change in fluid density, in g/cm3/PSI. When entering the
value, reverse the sign.
Example:
If the density factor is 0.000006 g/cm3/PSI, enter −0.000006 g/cm3/PSI.
6.Determine how the transmitter will obtain pressure data, and perform the required
setup.
Configuration and Use Manual 55
Configure process measurement
OptionSetup
A user-configured
static pressure value
Polling for pres-
(6)
sure
a. Set Pressure Unit to the desired unit.
b. Set Compensation Pressure to the desired value.
a. Ensure that the primary mA output has been wired to support
Pressure/Temperature > External Polling.
c. Set Poll Control to Poll As Primary Host or Poll as Secondary Host.
d. Choose an unused polling slot.
e. Set External Tag to the HART tag of the external pressure device.
f. Set Polled Variable to Pressure.
Tip
• Poll as Primary: No other HART masters will be on the network.
• Poll as Secondary: Other HART masters will be on the network.
The Field Communicator is not a HART master.
A value written by
digital communications
a. Set Pressure Unit to the desired unit.
b. Perform the necessary host programming and communications
setup to write pressure data to the transmitter at appropriate in-
tervals.
Postrequisites
If you are using an external pressure value, verify the setup by choosing Service Tools >
Variables > External Variables and checking the value displayed for External Pressure.
4.7.4Options for Pressure Measurement Unit
The transmitter provides a standard set of measurement units for Pressure Measurement Unit.
Different communications tools may use different labels for the units. In most
applications, Pressure Measurement Unit should be set to match the pressure measurement
unit used by the remote device.
Options for Pressure Measurement UnitTable 4-13:
Label
Display (stand-
Unit description
Feet water @ 68 °FFTH2OFt Water @ 68°FFt Water @ 68°FFt Water @ 68°FftH2O
Inches water @ 4 °CINW4CIn Water @ 4°CIn Water @ 4°CIn Water @ 4°CinH2O @4DegC
Inches water @ 60 °FINW60In Water @ 60°FIn Water @ 60°FIn Water @ 60°FinH2O @60DegF
Inches water @ 68 °FINH2OIn Water @ 68°FIn Water @ 68°FIn Water @ 68°FinH2O
ard)
Chinese-language display
ProLink IIProLink IIIField Commu-
nicator
(6) Not available on all transmitters.
56 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure process measurement
Options for Pressure Measurement Unit (continued)Table 4-13:
•Enable or disable operator actions from the display
•Configure security for the display menus
•Configure response time parameters
•Configure alarm handling
•Configure informational parameters
5.1Configure the transmitter display
You can control the process variables shown on the display and a variety of display
behaviors.
The transmitter display parameters include:
•Display Language
•Display Variables
•Display Precision
•Update Period
•Auto Scroll and Auto Scroll Rate
•Backlight
•Status LED Blinking
5.1.1Configure the language used for the display
Display (standard)OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > LANG
Chinese-language
display
ProLink IIProLink > Configuration > Display > Display Language
ProLink IIIDevice Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Language
Offline Maintain > Configuration > Display > Language
Overview
Display Language controls the language used for process data and menus on the display.
58 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure device options and preferences
Procedure
Select the language you want to use.
Tip
For devices with the Chinese-language display, you can use a shortcut key, or an optical switch
combination, to change the language without having to access the display menu. The optical switch
combination is shown on the front of the display.
The languages available depend on your transmitter model and version.
5.1.2Configure the process variables shown on the display
You can control the process variables shown on the display and the order in which they
appear. The display can scroll through up to 15 process variables in any order you choose.
In addition, you can repeat variables or leave slots unassigned.
Restrictions
•You cannot set Display Variable 1 to None. Display Variable 1 must be set to a process variable.
•If you have configured Display Variable 1 to track the primary mA output, you cannot change
the setting of Display Variable 1 using this procedure. To change the setting of Display Variable 1,
you must change the configuration of mA Output Process Variable for the primary mA output.
Note
If you configure a display variable as a volume process variable and then change Volume Flow Type, the
display variable is automatically changed to the equivalent process variable. For example, Volume FlowRate would be changed to Gas Standard Volume Flow Rate.
Procedure
For each display variable you want to change, assign the process variable you want to use.
Configuration and Use Manual 59
Configure device options and preferences
Example: Default display variable configuration
Display variableProcess variable assignment
Display Variable 1Mass flow
Display Variable 2Mass total
Display Variable 3Volume flow
Display Variable 4Volume total
Display Variable 5Density
Display Variable 6Temperature
Display Variable 7External pressure
Display Variable 8Mass flow
Display Variable 9None
Display Variable 10None
Display Variable 11None
Display Variable 12None
Display Variable 13None
Display Variable 14None
Display Variable 15None
Configure Display Variable 1 to track the primary mA output
You can configure Display Variable 1 to track mA Output Process Variable for the primary mA
output. When tracking is enabled, you can control Display Variable 1 from the display menu.
Tip
This feature is the only way to configure a display variable from the display menus (standard display
option only), and it applies only to Display Variable 1.
Procedure
Configure Display Variable 1 to track the primary mA output.
Offline Maintain > Configuration > Display > Configure 1st Var
60 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure device options and preferences
Display Variable 1 will automatically be set to match mA Output Process Variable for the primary
mA output. If you change the configuration of mA Output Process Variable, Display Variable 1
will be updated automatically.
5.1.3Configure the precision of variables shown on the
display
Setting Display Precision determines the precision (number of decimal places) shown on the
display. You can set Display Precision independently for each variable.
Setting Display Precision does not affect the actual value of the process variable.
Offline Maintain > Configuration > Display > Num of Decimals
Procedure
1.Select a process variable.
2.Set Display Precision to the number of decimal places you want shown when the
process variable appears on the display.
For temperature and density process variables, the default value is 2 decimal places.
For all other process variables, the default value is 4 decimal places. The range is
0 to 5.
Tip
The lower the selected precision, the greater the process change must be for it to be reflected
on the display. Do not set Display Precision value too low or too high to be useful.
5.1.4Configure the refresh rate of data shown on the display
ProLink IIIDevice Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Auto Scroll
Offline Maintain > Configuration > Display > Auto Scroll
Overview
You can configure the display to automatically scroll through the configured display
variables or to show a single display variable until the operator activates Scroll. When you
set automatic scrolling, you can also configure the length of time each display variable is
displayed.
Procedure
1.Enable or disable Auto Scroll as desired.
OptionDescription
EnabledThe display automatically scrolls through each display variable as specified
by Scroll Rate. The operator can move to the next display variable at any
time using Scroll.
Disabled (default)
The display shows Display Variable 1 and does not scroll automatically. The
operator can move to the next display variable at any time using Scroll.
2.If you enabled Auto Scroll, set Scroll Rate as desired.
The default value is 10 seconds.
Tip
Scroll Rate may not be available until you apply Auto Scroll.
62 Micro Motion® Model 1700 Transmitters with Analog Outputs
ProLink IIProLink > Configuration > Display > Display Options > Display Status LED Blinking
ProLink IIIDevice Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Status LED Blinking
Overview
By default, the status LED blinks (flashes) to indicate unacknowledged alarms. If you
disable Status LED Blinking, the status LED does not blink, whether alarms are acknowledged
or not. It still changes color to indicate active alarms.
Procedure
Enable or disable Status LED Blinking.
The default setting is Enabled.
Not available
Configuration and Use Manual 63
Configure device options and preferences
5.2Enable or disable operator actions from the
display
You can configure the transmitter to let the operator perform specific actions using the
display.
You can configure the following:
•Totalizer Start/Stop
•Totalizer Reset
•Acknowledge All Alarms
5.2.1Enable or disable Totalizer Start/Stop from the display
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features
Offline Maintain > Configuration > Display
66 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure device options and preferences
Overview
You can control operator access to different sections of the display off-line menu. You can
also configure a password to control access.
Procedure
1.To control operator access to the maintenance section of the off-line menu, enable
or disable Off-Line Menu.
OptionDescription
Enabled (default)
DisabledOperator cannot access the maintenance section of the off-line menu.
Operator can access the maintenance section of the off-line menu. This
access is required for configuration and calibration, but is not required to
view alarms or to access Smart Meter Verification (if applicable).
2.To control operator access to the alarm menu, enable or disable Alarm Menu.
OptionDescription
Enabled (default) Operator can access the alarm menu. This access is required to view and
acknowledge alarms, but is not required for Smart Meter Verification (if
applicable), configuration, or calibration.
DisabledOperator cannot access the alarm menu.
Note
The transmitter status LED changes color to indicate that there are active alarms, but does
not show specific alarms.
3.To require a password for access to the maintenance section of the off-line menu
and the Smart Meter Verification menu, enable or disable Off-Line Password.
OptionDescription
EnabledOperator is prompted for the off-line password at entry to the Smart Meter
Verification menu (if applicable) or entry to the maintenance section of the
off-line menu.
Disabled (default)
No password is required for entry to the Smart Meter Verification menu (if
applicable) or entry to the maintenance section of the off-line menu.
4.To require a password to access the alarm menu, enable or disable Alarm Password.
OptionDescription
EnabledOperator is prompted for the off-line password at entry to the alarm
menu.
Disabled (default) No password is required for entry to the alarm menu.
Configuration and Use Manual 67
Configure device options and preferences
If both Off-Line Password and Alarm Password are enabled, the operator is prompted for
the off-line password to access the off-line menu, but is not prompted thereafter.
5.(Optional) Set Off-Line Password to the desired value.
The same value is used for both the off-line password and the alarm password. The
default value is 1234. The range is 0000 to 9999.
Tip
Record your password for future reference.
5.4Configure response time parameters
You can configure the rate at which process data is polled and process variables are
calculated.
Update Rate controls the rate at which process data is polled and process variables are
calculated. Update Rate = Special produces faster and “noisier” response to changes in the
process. Do not use Special mode unless required by your application.
Tip
For systems with a standard core processor, Special mode can improve performance for applications
with entrained air or Empty-Full-Empty conditions. This does not apply to systems with an enhanced
core processor.
Prerequisites
Not available
Before setting Update Rate to Special:
•Check the effects of Special mode on specific process variables.
68 Micro Motion® Model 1700 Transmitters with Analog Outputs
Configure device options and preferences
•Contact Micro Motion.
Procedure
1.Set Update Rate as desired.
Option Description
NormalAll process data is polled at the rate of 20 times per second (20 Hz).
All process variables are calculated at 20 Hz.
This option is appropriate for most applications.
SpecialA single, user-specified process variable is polled at the rate of 100 times per sec-
ond (100 Hz). Other process data is polled at 6.25 Hz). Some process, diagnostic,
and calibration data is not polled.
All available process variables are calculated at 100 Hz.
Use this option only if required by your application.
If you change Update Rate, the settings for Flow Damping, Density Damping, and
Temperature Damping are automatically adjusted.
2.If you set Update Rate to Special, select the process variable to be polled at 100 Hz.
Effects of Update Rate = Special
Incompatible features and functions
Special mode is not compatible with the following features and functions:
•Enhanced events. Use basic events instead.
•All calibration procedures.
•Zero verification.
•Restoring the factory zero or the prior zero.
If required, you can switch to Normal mode, perform the desired procedures, and then
return to Special mode.
Process variable updates
Some process variables are not updated when Special mode is enabled.
Configuration and Use Manual 69
Configure device options and preferences
Special mode and process variable updatesTable 5-1:
Always polled and updated
• Mass flow
• Volume flow
• Gas standard volume flow
• Density
• Temperature
• Drive gain
• LPO amplitude
• Status [contains Event 1 and Event
2 (basic events)]
• Raw tube frequency
• Mass total
• Volume total
• Gas standard volume total
Updated only when the petroleum
measurement application is disabledNever updated
• RPO amplitude
• Board temperature
• Core input voltage
• Mass inventory
• Volume inventory
• Gas standard volume inventory
All other process variables and calibration data. They retain the values held
at the time you enabled Special mode.
5.4.2Configure Calculation Speed (Response Time)
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Device > Response Time
Calculation Speed is used to apply a different algorithm to the calculation of process variables
from the raw process data. Calculation Speed = Special produces faster and “noisier” response
to changes in the process.
In ProLink II, Calculation Speed is called Response Time.
Restriction
Calculation Speed is available only on systems with the enhanced core processor.
Tip
You can use Calculation Speed = Special with either setting of Update Rate. The parameters control
different aspects of flowmeter processing.
Not available
Procedure
Set Calculation Speed as desired.
70 Micro Motion® Model 1700 Transmitters with Analog Outputs
OptionDescription
NormalTransmitter calculates process variables at the standard speed.
SpecialTransmitter calculates process variables at a faster speed.
5.5Configure alarm handling
The alarm handling parameters control the transmitter’s response to process and device
conditions.
Alarm handling parameters include:
•Fault Timeout
•Status Alarm Severity
5.5.1Configure Fault Timeout
Configure device options and preferences
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Analog Output > Last Measured Value Timeout
ProLink > Configuration > Frequency/Discrete Output > Frequency > Last Measured Value Timeout
Overview
Fault Timeout controls the delay before fault actions are performed.
Restriction
Fault Timeout is applied only to the following alarms (listed by Status Alarm Code): A003, A004, A005,
A008, A016, A017, A033. For all other alarms, fault actions are performed as soon as the alarm is
detected.
Procedure
Set Fault Timeout as desired.
The default value is 0 seconds. The range is 0 to 60 seconds.
If you set Fault Timeout to 0, fault actions are performed as soon as the alarm condition is
detected.
The fault timeout period begins when the transmitter detects an alarm condition. During
the fault timeout period, the transmitter continues to report its last valid measurements.
Configuration and Use Manual 71
Configure device options and preferences
If the fault timeout period expires while the alarm is still active, the fault actions are
performed. If the alarm condition clears before the fault timeout expires, no fault actions
are performed.
Tip
ProLink II allows you to set Fault Timeout in two locations. However, there is only one parameter, and
the same setting is applied to all outputs.
Field Communicator Configure > Manual Setup > Info Parameters > Transmitter Info > Date
Overview
Date lets you store a static date (not updated by the transmitter) in transmitter memory.
This parameter is not used in processing and is not required.
Procedure
Enter the date you want to use, in the form mm/dd/yyyy.
Tip
ProLink II and ProLink III provide a calendar tool to help you select the date.
Field Communicator Configure > Manual Setup > Info Parameters > Sensor Information > Sensor Serial Number
Overview
Sensor Serial Number lets you store the serial number of the sensor component of your
flowmeter in transmitter memory. This parameter is not used in processing and is not
required.
Procedure
1.Obtain the sensor serial number from your sensor tag.
2.Enter the serial number in the Sensor Serial Number field.
Field Communicator Configure > Manual Setup > Info Parameters > Sensor Information > Tube Wetted Material
Overview
Sensor Material lets you store the type of material used for your sensor’s wetted parts in
transmitter memory. This parameter is not used in processing and is not required.
Procedure
1.Obtain the material used for your sensor’s wetted parts from the documents
shipped with your sensor, or from a code in the sensor model number.
To interpret the model number, refer to the product data sheet for your sensor.
2.Set Sensor Material to the appropriate option.
Not available
78 Micro Motion® Model 1700 Transmitters with Analog Outputs
Field Communicator Configure > Manual Setup > Info Parameters > Sensor Information > Tube Lining
Overview
Sensor Liner Material lets you store the type of material used for your sensor liner in
transmitter memory. This parameter is not used in processing and is not required.
Procedure
1.Obtain your sensor’s liner material from the documents shipped with your sensor, or
from a code in the sensor model number.
Not available
Configure device options and preferences
To interpret the model number, refer to the product data sheet for your sensor.
2.Set Sensor Liner Material to the appropriate option.
Field Communicator Configure > Manual Setup > Inputs/Outputs > Channels > Channel B
Overview
You can configure Channel B on your transmitter to operate as a frequency output or a
discrete output. The channel configuration must match the wiring at the transmitter
terminals.
Prerequisites
To avoid causing process errors:
•Configure the channels before configuring the outputs.
•Before changing the channel configuration, ensure that all control loops affected by
the channel are under manual control.
Procedure
Set Channel B as desired.
Offline Maintain > Configuration > Input/Output > Channel B Setup
OptionDescription
Frequency outputChannel B will operate as a frequency output.
Discrete outputChannel B will operate as a discrete output.
80 Micro Motion® Model 1700 Transmitters with Analog Outputs
Postrequisites
For each channel that you configured, perform or verify the corresponding input or output
configuration. When the configuration of a channel is changed, the channel’s behavior will
be controlled by the configuration that is stored for the selected input or output type, and
the stored configuration may not be appropriate for your process.
After verifying channel and output configuration, return the control loop to automatic
control.
6.2Configure the mA output
The mA output is used to report the configured process variable. The mA output
parameters control how the process variable is reported. Your transmitter has one mA
output: Channel A.
The mA output parameters include:
•mA Output Process Variable
•Lower Range Value (LRV) and Upper Range Value (URV)
•AO Cutoff
•Added Damping
•AO Fault Action and AO Fault Value
Integrate the meter with the control system
Important
Whenever you change an mA output parameter, verify all other mA output parameters before
returning the flowmeter to service. In some situations, the transmitter automatically loads a set of
stored values, and these values may not be appropriate for your application.
6.2.1Configure mA Output Process Variable
Display (standard)OFF-LINE MAINT > OFF-LINE CONFG > IO > CH A > AO
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output
Overview
Use mA Output Process Variable to select the variable that is reported over the mA output.
Prerequisites
•If you plan to configure the output to report volume flow, ensure that you have set
Volume Flow Type as desired: Liquid or Gas Standard Volume.
Offline Maintain > Configuration > Input/Output > Channel A Setup > MAO Source
Configuration and Use Manual 81
Integrate the meter with the control system
•If you are using the HART variables, be aware that changing the configuration of mA
Output Process Variable will change the configuration of the HART Primary Variable
(PV) and the HART Tertiary Variable (TV).
•If you have configured Display Variable 1 to track mA Output Process Variable, be aware
that changing the configuration of mA Output Process Variable will change the contents
of Display Variable 1.
Procedure
Set mA Output Process Variable as desired.
The default setting is Mass Flow Rate.
Options for mA Output Process Variable
The transmitter provides a basic set of options for mA Output Process Variable, plus several
application-specific options. Different communications tools may use different labels for
the options.
Options for mA Output Process VariableTable 6-1:
Label
Display (stand-
Process variable
Mass flow rateMFLOWMass Flow RateMass Flow RateMass Flow RateMass flo
Field Communicator • Configure > Manual Setup > Inputs/Outputs > mA Output > mA Output Settings > PV LRV
• Offline Maintain > Configuration > Input/Output > Channel A Setup > Variable at 4 mA
• Offline Maintain > Configuration > Input/Output > Channel A Setup > Variable at 20 mA
• ProLink > Configuration > Analog Output > Primary Output > Upper Range Value
• Configure > Manual Setup > Inputs/Outputs > mA Output > mA Output Settings > PV URV
82 Micro Motion® Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
Overview
The Lower Range Value (LRV) and Upper Range Value (URV) are used to scale the mA output,
that is, to define the relationship between mA Output Process Variable and the mA output
level.
Note
For transmitter software v5.0 and later, if you change LRV and URV from the factory default values,
and you later change mA Output Process Variable, LRV and URV will not reset to the default values. For
example, if you set mA Output Process Variable to Mass Flow Rate and change the LRV and URV, then you
set mA Output Process Variable to Density, and finally you change mA Output Process Variable back to MassFlow Rate, LRV and URV for Mass Flow Rate reset to the values that you configured. In earlier versions of
the transmitter software, LRV and URV reset to the factory default values.
Procedure
Set LRV and URV as desired.
• LRV is the value of mA Output Process Variable represented by an output of 4 mA. The
default value for LRV depends on the setting of mA Output Process Variable. Enter LRV in
the measurement units that are configured for mA Output Process Variable.
• URV is the value of mA Output Process Variable represented by an output of 20 mA. The
default value for URV depends on the setting of mA Output Process Variable. Enter URV in
the measurement units that are configured for mA Output Process Variable.
Tips
For best performance:
• Set LRV ≥ LSL (lower sensor limit).
• Set URV ≤ USL (upper sensor limit).
• Set these values so that the difference between URV and LRV is ≥ Min Span (minimum span).
Defining URV and LRV within the recommended values for Min Span, LSL, and USL ensures that the
resolution of the mA output signal is within range of the bit precision of the D/A converter.
Note
You can set URV below LRV. For example, you can set URV to 50 and LRV to 100.
The mA output uses a range of 4–20 mA to represent mA Output Process Variable. Between
LRV and URV, the mA output is linear with the process variable. If the process variable
drops below LRV or rises above URV, the transmitter posts an output saturation alarm.
Default values for Lower Range Value (LRV) and Upper Range Value
(URV)
Each option for mA Output Process Variable has its own LRV and URV. If you change the
configuration of mA Output Process Variable, the corresponding LRV and URV are loaded and
used.
Configuration and Use Manual 83
Integrate the meter with the control system
Default values for Lower Range Value (LRV) and Upper Range Value (URV)Table 6-2:
Process variableLRVURV
All mass flow variables–200.000 g/sec200.000 g/sec
All liquid volume flow variables–0.200 l/sec0.200 l/sec
Gas standard volume flow–423.78 SCFM423.78 SCFM
6.2.3Configure AO Cutoff
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Analog Output > Primary Output > AO Cutoff
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output > mA Output Settings > PV MAO Cutoff
Not available
Overview
AO Cutoff (Analog Output Cutoff) specifies the lowest mass flow rate, volume flow rate, or
gas standard volume flow rate that will be reported through the mA output. Any flow rates
below AO Cutoff will be reported as 0.
Restriction
AO Cutoff is applied only if mA Output Process Variable is set to Mass Flow Rate, Volume Flow Rate, or Gas
Standard Volume Flow Rate. If mA Output Process Variable is set to a different process variable, AO Cutoff is
not configurable, and the transmitter does not implement the AO cutoff function.
Procedure
Set AO Cutoff as desired.
The default value for AO Cutoff is 0.0 g/sec.
Tip
For most applications, the default value of AO Cutoff should be used. Contact Micro Motion customer
service before changing AO Cutoff.
Interaction between AO Cutoff and process variable cutoffs
When mA Output Process Variable is set to a flow variable (for example, mass flow rate or
volume flow rate), AO Cutoff interacts with Mass Flow Cutoff or Volume Flow Cutoff. The
transmitter puts the cutoff into effect at the highest flow rate at which a cutoff is
applicable.
84 Micro Motion® Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
Example: Cutoff interaction
Configuration:
•mA Output Process Variable = Mass Flow Rate
•Frequency Output Process Variable = Mass Flow Rate
•AO Cutoff = 10 g/sec
•Mass Flow Cutoff = 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, all outputs representing mass flow will
report zero flow.
Example: Cutoff interaction
Configuration:
•mA Output Process Variable = Mass Flow Rate
•Frequency Output Process Variable = Mass Flow Rate
•AO Cutoff = 15 g/sec
•Mass Flow Cutoff = 10 g/sec
Result:
•If the mass flow rate drops below 15 g/sec but not below 10 g/sec:
-The mA output will report zero flow.
-The frequency output will report the actual flow rate.
•If the mass flow rate drops below 10 g/sec, both outputs will report zero flow.
6.2.4Configure Added Damping
Display (standard)Not available
Chinese-language
display
ProLink IIProLink > Configuration > Analog Output > Primary Output > AO Added Damp
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output > mA Output Settings > PV Added Damping
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping
Value specifies the time period (in seconds) over which the transmitter will spread changes
in the reported process variable. At the end of the interval, the reported process variable
will reflect 63% of the change in the actual measured value. Added Damping controls the
amount of damping that will be applied to the mA output. It affects the reporting of mAOutput Process Variable through the mA output only. It does not affect the reporting of that
process variable via any other method (e.g., the frequency output or digital
communications), or the value of the process variable used in calculations.
Not available
Configuration and Use Manual 85
Integrate the meter with the control system
Note
Added Damping is not applied if the mA output is fixed (for example, during loop testing) or if the mA
output is reporting a fault. Added Damping is applied while sensor simulation is active.
Procedure
Set Added Damping to the desired value.
The default value is 0.0 seconds.
When you specify a value for Added Damping, the transmitter automatically rounds the value
down to the nearest valid value.
Note
Added Damping values are affected by the setting of Update Rate and 100 Hz Variable.
Interaction between Added Damping and process variable
damping
When mA Output Process Variable is set to a flow variable, density, or temperature, Added
Damping interacts with Flow Damping, Density Damping, or Temperature Damping. If multiple
damping parameters are applicable, the effect of damping the process variable is
calculated first, and the added damping calculation is applied to the result of that
calculation.
Example: Damping interaction
Configuration:
•Flow Damping = 1 second
•mA Output Process Variable = Mass Flow Rate
•Added Damping = 2 seconds
Result: A change in the mass flow rate will be reflected in the mA output over a time period
that is greater than 3 seconds. The exact time period is calculated by the transmitter
according to internal algorithms which are not configurable.
86 Micro Motion® Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
6.2.5Configure mA Output Fault Action and mA Output Fault Level
Display (standard)Not available
Chinese-language
display
ProLink II• ProLink > Configuration > Analog Output > Primary Output > AO Fault Action
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output > MA0 Fault Settings
Overview
mA Output Fault Action controls the behavior of the mA output if the transmitter encounters
an internal fault condition.
Note
For some faults only: If Last Measured Value Timeout is set to a non-zero value, the transmitter will not
implement the fault action until the timeout has elapsed.
Not available
• ProLink > Configuration > Analog Output > Primary Output > AO Fault Level
Procedure
1.Set mA Output Fault Action to the desired value.
The default setting is Downscale.
2.If you set mA Output Fault Action to Upscale or Downscale, set mA Output Fault Level as
desired.
Options for mA Output Fault Action and mA Output Fault Level
Options for mA Output Fault Action and mA Output Fault LevelTable 6-4:
OptionmA output behaviormA Output Fault Level
UpscaleGoes to the configured fault levelDefault: 22.0 mA
Range: 21 to 24 mA
Downscale (default)Goes to the configured fault level
Internal ZeroGoes to the mA output level associated
with a process variable value of 0 (zero),
as determined by Lower Range Value and
Upper Range Value settings
NoneTracks data for the assigned process vari-
able; no fault action
Default: 2.0 mA
Range: 1.0 to 3.6 mA
Not applicable
Not applicable
Configuration and Use Manual 87
Integrate the meter with the control system
CAUTION!
If you set mA Output Fault Action or Frequency Output Fault Action to None, be sure to set DigitalCommunications Fault Action to None. If you do not, the output will not report actual process data,
and this may result in measurement errors or unintended consequences for your process.
Restriction
If you set Digital Communications Fault Action to NAN, you cannot set mA Output Fault Action or Frequency
Output Fault Action to None. If you try to do this, the transmitter will not accept the configuration.
6.3Configure the frequency output
The frequency output is used to report a process variable. The frequency output
parameters control how the process variable is reported. Your transmitter may have zero
or one frequency output: Channel B can be configured as a frequency output or a discrete
output.
The frequency output parameters include:
•Frequency Output Polarity
•Frequency Output Scaling Method
•Frequency Output Maximum Pulse Width
•Frequency Output Fault Action and Frequency Output Fault Value
Restriction
The process variable assigned to the primary mA output is automatically assigned to the frequency
output. You cannot assign a different process variable.
Important
Whenever you change a frequency output parameter, verify all other frequency output parameters
before returning the flowmeter to service. In some situations, the transmitter automatically loads a
set of stored values, and these values may not be appropriate for your application.
6.3.1Configure Frequency Output Polarity
Display (standard)OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET FO > FO POLAR
Field Communicator Configure > Manual Setup > Inputs/Outputs > Frequency Output > FO Settings > FO Polarity
Offline Maintain > Configuration > Input/Output > Channel B Setup > Frequency Output > Polarity
88 Micro Motion® Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
Overview
Frequency Output Polarity controls how the output indicates the ON (active) state. The default
value, Active High, is appropriate for most applications. Active Low may be required by
applications that use low-frequency signals.
Procedure
Set Frequency Output Polarity as desired.
The default setting is Active High.
Options for Frequency Output Polarity
Options for Frequency Output PolarityTable 6-5:
PolarityReference voltage (OFF)Pulse voltage (ON)
Active High0As determined by power sup-
ply, pull-up resistor, and load
(see the installation manual for
your transmitter)
Active LowAs determined by power sup-
ply, pull-up resistor, and load
(see the installation manual for
your transmitter)
0
6.3.2Configure Frequency Output Scaling Method
Display (standard)OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET FO > FO SCALE
Field Communicator Configure > Manual Setup > Inputs/Outputs > Frequency Output > FO Scaling
Overview
Frequency Output Scaling Method defines the relationship between output pulse and flow
units. Set Frequency Output Scaling Method as required by your frequency receiving device.
Procedure
1.Set Frequency Output Scaling Method.
OptionDescription
Frequency=Flow (default)Frequency calculated from flow rate
Offline Maintain > Configuration > Input/Output > Channel B Setup > Frequency Output > Scaling Meth-
od
Configuration and Use Manual 89
Integrate the meter with the control system
OptionDescription
Pulses/UnitA user-specified number of pulses represents one flow unit
Units/PulseA pulse represents a user-specified number of flow units
2.Set additional required parameters.
• If you set Frequency Output Scaling Method to Frequency=Flow, set Rate Factor and
Frequency Factor.
• If you set Frequency Output Scaling Method to Pulses/Unit, define the number of pulses
that will represent one flow unit.
• If you set Frequency Output Scaling Method to Units/Pulse, define the number of units
that each pulse will indicate.
Calculate frequency from flow rate
The Frequency=Flow option is used to customize the frequency output for your application
when you do not know appropriate values for Units/Pulse or Pulses/Unit.
If you specify Frequency=Flow, you must provide values for Rate Factor and Frequency Factor:
Rate FactorThe maximum flow rate that you want the frequency output to report.
Above this rate, the transmitter will report A110: Frequency Output Saturated.
Frequency
Factor
A value calculated as follows:
FrequencyFactor =x N
RateFactor
T
where:
TFactor to convert selected time base to seconds
NNumber of pulses per flow unit, as configured in the receiving device
The resulting Frequency Factor must be within the range of the frequency output (0 to
10,000 Hz):
•If Frequency Factor is less than1 Hz,reconfigure the receiving device for a higher
pulses/unit setting.
•If Frequency Factor is greater than 10,000 Hz, reconfigure the receiving device for a
lower pulses/unit setting.
Tip
If Frequency Output Scale Method is set to Frequency=Flow, and Frequency Output Maximum Pulse Width is set
to a non-zero value, Micro Motion recommends setting Frequency Factor to a value below 200 Hz.
Example: Configure Frequency=Flow
You want the frequency output to report all flow rates up to 2000 kg/min.
The frequency receiving device is configured for 10 pulses/kg.
Solution:
90 Micro Motion® Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
FrequencyFactor =x N
FrequencyFactor =x 10
RateFactor
T
2000
60
333.33FrequencyFactor =
Set parameters as follows:
•Rate Factor: 2000
•Frequency Factor: 333.33
6.3.3Configure Frequency Output Maximum Pulse Width
Field Communicator Configure > Manual Setup > Inputs/Outputs > Frequency Output > FO Settings > Max Pulse Width
Offline Maintain > Configuration > Input/Output > Channel B Setup > Frequency Output
Overview
Frequency Output Maximum Pulse Width is used to ensure that the duration of the ON signal is
great enough for your frequency receiving device to detect.
The ON signal may be the high voltage or 0.0 V, depending on Frequency Output Polarity.
Interaction of Frequency Output Maximum Pulse Width and Frequency Output PolarityTable 6-6:
PolarityPulse width
Active High
Active Low
Procedure
Set Frequency Output Maximum Pulse Width as desired.
Configuration and Use Manual 91
Integrate the meter with the control system
The default value is 277 milliseconds. You can set Frequency Output Maximum Pulse Width to
0 milliseconds or to a value between 0.5 milliseconds and 277.5 milliseconds. The
transmitter automatically adjusts the value to the nearest valid value.
Tip
Micro Motion recommends leaving Frequency Output Maximum Pulse Width at the default value. Contact
Micro Motion customer service before changing Frequency Output Maximum Pulse Width.
6.3.4Configure Frequency Output Fault Action and Frequency
Frequency Output Fault Action controls the behavior of the frequency output if the transmitter
encounters an internal fault condition.
Note
For some faults only: If Last Measured Value Timeout is set to a non-zero value, the transmitter will not
implement the fault action until the timeout has elapsed.
Procedure
1.Set Frequency Output Fault Action as desired.
The default value is Downscale (0 Hz).
2.If you set Frequency Output Fault Action to Upscale, set Frequency Fault Level to the desired
value.
The default value is 15000 Hz. The range is 10 to 15000 Hz.
92 Micro Motion® Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
Options for Frequency Output Fault Action
Options for Frequency Output Fault ActionTable 6-7:
LabelFrequency output behavior
UpscaleGoes to configured Upscale value:
• Range: 10 Hz to 15000 Hz
• Default: 15000 Hz
Downscale0 Hz
Internal Zero0 Hz
None (default)Tracks data for the assigned process variable; no fault action
CAUTION!
If you set mA Output Fault Action or Frequency Output Fault Action to None, be sure to set DigitalCommunications Fault Action to None. If you do not, the output will not report actual process data,
and this may result in measurement errors or unintended consequences for your process.
Restriction
If you set Digital Communications Fault Action to NAN, you cannot set mA Output Fault Action or Frequency
Output Fault Action to None. If you try to do this, the transmitter will not accept the configuration.
6.4Configure the discrete output
The discrete output is used to report specific flowmeter or process conditions. The discrete
output parameters control which condition is reported and how it is reported. Your
transmitter may have zero or one discrete output: Channel B can be configured as a
frequency output or a discrete output.
The discrete output parameters include:
•Discrete Output Source
•Discrete Output Polarity
•Discrete Output Fault Action
Restriction
Before you can configure the discrete output, you must configure a channel to operate as a discrete
output.
Important
Whenever you change a discrete output parameter, verify all other discrete output parameters
before returning the flowmeter to service. In some situations, the transmitter automatically loads a
set of stored values, and these values may not be appropriate for your application.
Configuration and Use Manual 93
Integrate the meter with the control system
6.4.1Configure Discrete Output Source
Display (standard)OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET DO > DO SRC