Micro Motion™ 5700 Transmitters with
FOUNDATION™ Fieldbus
Configuration and Use Manual
June 2021
Safety messages
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully
before proceeding to the next step.
Safety and approval information
This Micro Motion product complies with all applicable European directives when properly installed in accordance with the
instructions in this manual. Refer to the EU declaration of conformity for directives that apply to this product. The following are
available: the EU declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings
and Instructions. In addition the IECEx Installation Instructions for installations outside of the European Union and the CSA
Installation Instructions for installations in North America are available on the internet at www.emerson.com or through your local
Micro Motion support center.
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at
www.emerson.com. For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration
manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government
transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept
your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion
Customer Service department.
Emerson Flow customer service
Email:
• Worldwide: flow.support@emerson.com
• Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America
United States800-522-6277U.K. and Ireland0870 240 1978Australia800 158 727
G.1 RoHS and WEEE.......................................................................................................................365
6Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualBefore you begin
MMI-20029970June 2021
1 Before you begin
1.1 About this manual
This manual helps you configure, commission, use, maintain, and troubleshoot Micro Motion 5700
transmitters with FOUNDATION Fieldbus.
Important
This manual assumes that:
• The transmitter has been installed correctly and completely according to the instructions in the
transmitter installation manual
• Users understand basic transmitter and sensor installation, configuration, and maintenance concepts and
procedures
1.2 Hazard messages
This document uses the following criteria for hazard messages based on ANSI standards Z535.6-2011
(R2017).
DANGER
Serious injury or death will occur if a hazardous situation is not avoided.
WARNING
Serious injury or death could occur if a hazardous situation is not avoided.
CAUTION
Minor or moderate injury will or could occur if a hazardous situation is not avoided.
NOTICE
Data loss, property damage, hardware damage, or software damage can occur if a situation is not avoided.
There is no credible risk of physical injury.
Physical access
NOTICE
Unauthorized personnel can potentially cause significant damage and/or misconfiguration of end users'
equipment. Protect against all intentional or unintentional unauthorized use.
Physical security is an important part of any security program and fundamental to protecting your system.
Restrict physical access to protect users' assets. This is true for all systems used within the facility.
Configuration and Use Manual7
Before you beginConfiguration and Use Manual
June 2021 MMI-20029970
1.3 Related documents
You can find all product documentation on the product documentation DVD shipped with the product or at
www.emerson.com.
See any of the following documents for more information:
• Micro Motion 5700 Product Data Sheet
• Micro Motion 5700 Transmitters for FOUNDATION™ Fieldbus: Installation Manual
• Replacing the Junction Box for the 4200 Transmitter and the 5700 Transmitter
• Replacing the Sensor Cable for the 4200 Transmitter and the 5700 Transmitter
• Sensor installation manual
1.4 Communication methods
You can use several different communications methods to interface with the transmitter. You may use
different methods in different locations or for different tasks.
InterfaceTool
DisplayInfrared-sensitive buttons
Universal Service PortProLink™ III
FOUNDATION Fieldbus channel• Field communicator
• FOUNDATION Fieldbus (FF) host
— On an enhanced FF host, the transmitter parameters are displayed either
in the form of a menu tree (for example, the 475 Field Communicator)
or in the form of UIRD (for example, the AMS Intelligent Device Manager
with DeltaV™ System). Both the menu tree and UIRD are provided as
part of the Device Description.
— A basic FF host displays the transmitter parameters in the form of a list
under the Resource block and transducer blocks.
— The configuration sections contain information for both types of host.
For information about how to use the communication tools, see the appendices in this manual.
Tip
You may be able to use other communications tools, such as AMS™ Suite: Intelligent Device Manager.
8Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualQuick start
MMI-20029970June 2021
2 Quick start
2.1 Power up the transmitter
The transmitter must be powered up for all configuration and commissioning tasks, or for process
measurement.
Procedure
1. Verify that all transmitter and sensor covers and seals are closed.
WARNING
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are
tightly closed. For hazardous area installations, applying power while housing covers are removed or
loose can cause an explosion resulting in injury or death.
2. Turn on the electrical power at the power supply.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to
10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power has been off long
enough to allow components to reach ambient temperature, allow the electronics to warm up for
approximately 10 minutes before relying on process measurements. During this warm-up period, you may
observe minor measurement instability or inaccuracy.
2.2 Check meter status
Check the meter for any error conditions that require user action or that affect measurement accuracy.
Procedure
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error
conditions. During the power-up sequence, the Transmitter Initializing alert is active. This
alert should clear automatically when the power-up sequence is complete.
2. Check the status LED on the transmitter.
Table 2-1: Status LED and device status
Status LED conditionDevice status
Solid greenNo alerts are active.
Solid yellowOne or more alerts are active with Alert Severity = Out of Specification,
Maintenance Required, or Function Check.
Solid redOne or more alerts are active with Alert Severity = Failure.
Flashing yellow (1 Hz)The Function Check in Progress alert is active.
Configuration and Use Manual9
Quick startConfiguration and Use Manual
June 2021 MMI-20029970
2.3 Determine the FOUNDATION Fieldbus unique device ID
using the display
Every FOUNDATION Fieldbus device has a unique 24-digit number that the fieldbus segment uses to identify it.
You can determine the number using the display.
Procedure
Choose Menu→About→Device Information.
The number is located under Device Unique ID.
2.4 Commissioning wizards
The transmitter menu includes a Guided Setup to help you move quickly through the most common
configuration parameters. ProLink III also provides a commissioning wizard.
By default, when the transmitter starts up, the Guided Setup menu is offered. You can choose to use it or not.
You can also choose whether or not Guided Setup is displayed automatically.
• To enter Guided Setup upon transmitter startup, choose Yes at the prompt.
• To enter Guided Setup after transmitter startup, choose Menu→Startup Tasks.
• To control the automatic display of Guided Setup, choose Menu→Configuration→Guided Setup.
For information on the ProLink III commissioning wizard, see the Micro Motion ProLink III with ProcessVizSoftware User Manual.
As the commissioning wizards are self guided, they are not documented in detail.
2.5 Make a startup connection to the transmitter
For all configuration tools except the display, you must have an active connection to the transmitter to
configure the transmitter.
Procedure
Identify the connection type to use, and follow the instructions for that connection type in the appropriate
appendix.
Basic FF hostDevice TB→Set Clock Date-Time (OD Index 136)
Menu → Configuration → Time/Date/Tag
10Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualQuick start
MMI-20029970June 2021
The transmitter clock provides timestamp data for alerts, service logs, history logs, and all other timers and
dates in the system. You can set the clock for your local time or for any standard time you want to use.
Tip
You may find it convenient to set all of your transmitter clocks to the same time, even if the transmitters are
in different time zones.
Procedure
1. Select the time zone that you want to use.
2. If you need a custom time zone, select Special Time Zone and enter your time zone as a difference
from UTC (Coordinated Universal Time).
3. Set the time appropriately for the selected time zone.
Tip
The transmitter does not adjust for Daylight Savings Time. If you observe Daylight Savings Time, you
must reset the transmitter clock manually.
4. Set the month, day, and year.
The transmitter tracks the year and automatically adds a day for leap years.
2.7 View the licensed features
Display
ProLink IIIDevice Tools→Device Information→Licensed Features
Field communicatorOverview→Device Information→Licenses
Basic FF hostDevice TB→Permanent Feature (OD Index 142)
You can view the licensed features to ensure that the transmitter was ordered with the required features.
Licensed features are purchased and available for permanent use. The options model code represents the
licensed features.
A trial license allows you to explore features before purchasing. The trial license enables the specified features
for a limited number of days. This number is displayed for reference. At the end of this period, the feature will
no longer be available.
To purchase additional features or request a trial license, document the Unique ID Number and current
license key from your transmitter and contact customer service. To enable the additional features or trial
license, you will need to install the new license on the transmitter.
Basic FF hostDevice TB→Transmitter Information (OD Index 14–21)
Device TB → Core Processor Information (OD Index 22–25)
Device TB → Sensor Information (OD Index 28–33)
You can set several parameters that identify or describe the transmitter and sensor. These parameters are not
used in processing and are not required.
Procedure
1. Set informational parameters for the transmitter.
a) Set Transmitter Serial Number to the serial number of your transmitter.
The transmitter serial number is provided on the metal tag that is attached to the transmitter
housing.
b) Set Descriptor to any desired description of this transmitter or measurement point.
c) Set Message to any desired message.
d) Verify that Model Code (Base) is set to the base model code of the transmitter.
The base model code completely describes your transmitter, except for the features that can be
licensed independently. The base model code is set at the factory.
e) Set Model Code (Options) to the options model code of the transmitter.
The options model code describes the independent features that have been licensed for this
transmitter. The original options model code is set at the factory. If you license additional
options for this transmitter, Micro Motion will supply an updated options model code.
For a field communicator, configuring model code options is not available for this release.
2. Set informational parameters for the sensor.
a) Set Sensor Serial Number to the serial number of the sensor connected to this transmitter.
The sensor serial number is provided on the metal tag that is attached to the sensor case.
b) Set Sensor Material to the material used for the sensor.
c) Set Sensor Liner to the material used for the sensor liner, if any.
d) Set Flange Type to the type of flange that was used to install the sensor.
Do not set Sensor Type. Sensor Type is set or derived during characterization.
2.9 Characterize the meter (if required)
Display
ProLink IIIDevice Tools→Calibration Data
Field communicatorConfigure→Manual Setup→Characterization
12Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Basic FF hostMeasurement TB→Device Calibration (OD Index 95–113)
Characterizing the meter adjusts your transmitter to match the unique traits of the sensor it is paired with.
The characterization parameters (also called calibration parameters) describe the sensor’s sensitivity to flow,
density, and temperature. Depending on your sensor type, different parameters are required.
Values for your sensor are provided on the sensor tag or the calibration certificate.
• If your transmitter was ordered with a sensor, it was characterized at the factory. However, you should still
verify the characterization parameters.
• Perform a characterization whenever you replace a core processor.
Procedure
1. Optional: Specify Sensor Type.
• Straight Tube (T-Series sensors)
• Curved Tube (all sensors except T-Series)
Note
Unlike earlier transmitters, the 5700 derives Sensor Type from the user-specified values for FCF and K1
in combination with an internal ID.
2. Set the flow calibration factor: FCF (also called Flow Cal or Flow Calibration Factor). Be sure to include
all decimal points.
3. Set the density characterization parameters: D1, D2, TC, K1, K2, and FD. (TC is sometimes shown as
DT.)
4. Apply the changes as required by the tool you are using.
The transmitter identifies your sensor type, and characterization parameters are adjusted as required:
• If Sensor Type changed from Curved Tube to Straight Tube, five characterization parameters are
added to the list.
• If Sensor Type changed from Straight Tube to Curved Tube, five characterization parameters are
removed from the list.
• If Sensor Type did not change, the list of characterization parameters does not change.
5. T-Series sensors only: Set the additional characterization parameters listed below.
Characterization parameter type
FlowFTG, FFQ
DensityDTG, DFQ1, DFQ2
Parameters
Configuration and Use Manual13
Quick startConfiguration and Use Manual
June 2021 MMI-20029970
2.9.1 Sample sensor tags
Figure 2-1: Tag on newer curved-tube sensors (all sensors except T-Series)
Figure 2-2: Tag on older straight-tube sensor (T-Series)
Figure 2-3: Tag on newer straight-tube sensor (T-Series)
2.9.2 Flow calibration parameters (FCF, FT)
Two separate values are used to describe flow calibration: a 6-character FCF value and a 4-character FT value.
They are provided on the sensor tag.
Both values contain decimal points. During characterization, these are entered as a single 10-character string.
The 10-character string is called either Flowcal or FCF.
14Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualQuick start
MMI-20029970June 2021
If your sensor tag shows the FCF and the FT values separately and you need to enter a single value,
concatenate the two values to form the single parameter value, retaining both decimal points.
Concatenating FCF and FT
FCF = x.xxxx FT = y.yy Flow calibration parameter: x.xxxxy.yy
Density calibration parameters are typically on the sensor tag and the calibration certificate.
If your sensor tag does not show a D1 or D2 value:
• For D1, enter the Dens A or D1 value from the calibration certificate. This value is the line-condition
density of the low-density calibration fluid. Micro Motion uses air. If you cannot find a Dens A or D1 value,
enter 0.001 g/cm3.
• For D2, enter the Dens B or D2 value from the calibration certificate. This value is the line-condition density
of the high-density calibration fluid. Micro Motion uses water. If you cannot find a Dens B or D2 value,
enter 0.998 g/cm3 .
If your sensor tag does not show a K1 or K2 value:
• For K1, enter the first five digits of the density calibration factor. In this sample tag, this value is shown as
12500.
• For K2, enter the second five digits of the density calibration factor. In this sample tag, this value is shown
as 14286.
Figure 2-4: K1, K2, and TC values in the density calibration factor
If your sensor does not show an FD value, contact customer service.
If your sensor tag does not show a DT or TC value, enter the last four characters of the density calibration
factor. In the sample tag shown above, the value is shown as 4.44.
Do not confuse the Meter Factor line on the pictured sensor tag with any meter factor settings discussed in
this manual.
Configuration and Use Manual15
Quick startConfiguration and Use Manual
June 2021 MMI-20029970
2.10 Verify mass flow measurement
Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available
method.
Procedure
• Read the value for Mass Flow Rate on the transmitter display.
• Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the Process Variables
panel.
• Connect to the transmitter with a field communicator and read the value for Mass Flow Rate.
Online → Overview → Mass Flow Rate
Postrequisites
If the reported mass flow rate is not accurate:
• Check the characterization parameters.
• Review the troubleshooting suggestions for flow measurement issues.
ProLink IIIDevice Tools→Calibration→Smart Zero Verification and Calibration→Verify Zero
Field communicatorService Tools→Maintenance→Calibration→Zero Calibration→Perform Zero Verify
Enhanced FF hostService Tools→Maintenance→Calibration→Zero Calibration→Perform Zero Verify
Basic FF hostMeasurement TB→Perform Zero Verify (OD Index 124)
Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field
zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one of the
following is true:
• The zero is required by site procedures.
• The stored zero value fails the zero verification procedure.
Do not verify the zero or zero the meter if a high-severity alert is active. Correct the problem, then verify the
zero or zero the meter. You may verify the zero or zero the meter if a low-severity alert is active.
Procedure
1. Prepare the meter:
a) Allow the meter to warm up for at least 20 minutes after applying power.
b) Run the process fluid through the sensor until the sensor temperature reaches the normal
process operating temperature.
c) Stop flow through the sensor by shutting the downstream valve, and then the upstream valve if
available.
16Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualQuick start
MMI-20029970June 2021
d) Verify that the sensor is blocked in, that flow has stopped, and that the sensor is completely full
of process fluid.
2. Start the zero verification procedure, and wait until it completes.
3. If the zero verification procedure fails:
a) Confirm that the sensor is completely blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
b) Verify that the process fluid is not flashing or condensing, and that it does not contain particles
that can settle out.
c) Repeat the zero verification procedure.
d) If it fails again, zero the meter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Related information
Zero the meter
Configuration and Use Manual17
Quick startConfiguration and Use Manual
June 2021 MMI-20029970
18Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20029970June 2021
3 Introduction to configuration and
commissioning
3.1 Security and write protection
The transmitter has several features that can help to protect it against intentional or unintentional access and
configuration changes.
• When locked, the mechanical lock switch on the front of the display prevents any configuration changes
to the transmitter from any local or remote configuration tool. A transmitter without a display does not
have a lock switch.
• When enabled, the software setting Write Protection prevents any configuration changes. The setting
can only be enabled if the transmitter does not have a display.
• If the Universal Service Port (USP) is disabled, the port cannot be used by any service tool to communicate
with or make changes to the transmitter.
• When enabled, Security prevents any configuration changes being made from the display unless the
appropriate password is entered.
• When enabled, the fieldbus write lock prevents any configuration changes being written from the fieldbus
segment.
3.1.1 Universal Service Port security
This transmitter is equipped with a Universal Service Port that works with USB type A connections, including
compatible flash drives. There are multiple levels of security built into the transmitter's service port that you
can configure according to your needs and security standards.
The service port offers the following features that enhance interface security:
• The service port is inaccessible without physical access to the transmitter and requires removal of the
terminal cover
• The service port can be disabled from the transmitter through software
• The transmitter has a non-traditional operating system that is not designed to execute programs or run
scripts
• The display can be password protected to limit access to the USB file menu
• Overall transmitter security switches such as the lock switch or write-protection disallows configuration
changes from all interfaces including the Universal Service Port
This transmitter:
• Was designed to be implemented in an industrial automation control system (Level 1 and Level 2 of the
Purdue Reference Architecture Model), with defense in depth security controls
• Is not intended to be directly connected to an enterprise or to an internet-facing network without a
compensating control in place
Configuration and Use Manual19
Introduction to configuration and commissioningConfiguration and Use Manual
June 2021 MMI-20029970
3.1.2 Lock or unlock the transmitter
If the transmitter has a display, a mechanical switch on the display can be used to lock or unlock the
transmitter. When locked, no configuration changes can be made using any configuration tool.
Figure 3-1: Lock switch on transmitter display
You can determine whether you need to lock or unlock the transmitter by looking at the switch.
• If the switch is in the right position, the transmitter is locked.
• If the switch is in the left position, the transmitter is unlocked.
Note
The top switch is reserved for future use.
Procedure
1.
2. Remove the transmitter housing cover.
Figure 3-2: Removing the transmitter housing cover
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is
powered up. Failure to follow these instructions can cause an explosion resulting in injury or death.
If you are in a hazardous area, power down the transmitter.
20Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20029970June 2021
3. Using a fine-pointed tool, move the switch to the desired position.
4. Replace the transmitter housing cover.
5. If necessary, power up the transmitter.
3.1.3 Enable or disable the service port
DisplayMenu→Configuration→Security→Service Port
ProLink IIINot available
Field communicatorConfigure→Manual Setup→Security→Enable/Disable Service Port
Enhanced FF hostConfigure→Manual Setup→Security→Enable/Disable Service Port
Basic FF hostDevice TB→Enable Service Port (OD Index 146)
The service port is enabled by default, so you can use it for transferring files or connect to it with ProLink III. If
you want to completely prevent it from being used, you can disable it.
Note
Enabling or disabling the service port will not take effect until power has been cycled to the transmitter.
WARNING
Do not use the service port if the transmitter is in a hazardous area because using the service port means
that you must open the transmitter wiring compartment. Opening the wiring compartment in a hazardous
area while the transmitter is powered up can cause an explosion resulting in injury or death.
Basic FF hostResource Block→Write Lock (OD Index 34)
When enabled, Write-Protection prevents changes to the transmitter configuration. You can perform all
other functions, and you can view the transmitter configuration parameters.
Note
The write protection setting via software methods (such as ProLink III) is available only on transmitters
without a display.
For transmitters with a display, write protection is available only using the lock switch on the display. See Lock
or unlock the transmitter.
Use the mechanical switch on the display.
Configuration and Use Manual21
Introduction to configuration and commissioningConfiguration and Use Manual
June 2021 MMI-20029970
Write-protecting the transmitter primarily prevents accidental changes to configuration, not intentional
changes. Any user who can make changes to the configuration can disable write protection.
Basic FF hostDevice TB→Offline Menu Passcode Required (OD Index 67)
Device TB → Passcode (4 Digits alphanumeric) (OD Index 68)
Device TB → Alert Passcode (OD Index 89)
When using the display, you can require users to enter a password to do any of the following tasks:
• Enter the main menu
• Change a parameter
• Access alert data through the display
• Start, stop, or reset totalizers or inventories via the context menu
The display password can be the same or different from the totalizer/inventory context menu control
password. If different, the display password is used to reset, start, and stop totalizers or inventories using
Menu → Operations → Totalizers.
22Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20029970June 2021
Procedure
1. Configure Password Required as desired.
OptionDescription
At WriteWhen an user chooses an action that leads to a configuration change, they are
prompted to enter the display password.
Enter MenuWhen the menu is selected from the process variable screen, the display password
will be immediately required if Password Required is set.
Never (default) When a user chooses an action that leads to a configuration change, they are
prompted to activate ⇦⇧⇩⇨. This is designed to protect against accidental changes
to configuration. It is not a security measure.
2. If the At Write or Enter Menu option was selected, enable or disable alert security as desired.
OptionDescription
Enabled If an alert is active, the alert symbol ⓘ is shown in the upper right corner of the display but
the alert banner is not displayed. If the operator attempts to enter the alert menu, they are
prompted to enter the display password.
Disabled If an alert is active, the alert symbol ⓘ is shown in the upper right corner of the display and
the alert banner is displayed automatically. No password or confirmation is required to
enter the alert menu.
Restriction
You cannot set Password Required to Never and enable alert security.
• If you did not enable Password Required, alert security is disabled and cannot be enabled.
• Alert security is disabled automatically if you set Password Required to Never after:
— Password Required is initially set to either At Write or Enter Menu
— Alert security is enabled
3. If Password Required has been set to At Write or Enter Menu, you will be prompted to enter the
desired password.
• Default: AAAA
• Range: Any four alphanumeric characters
• Password Required must be set to At Write or Enter Menu to enable the totalizer/inventory control
context menu password option.
Important
If you enable Password Required but you do not change the display password, the transmitter will post
a configuration alert.
Configuration and Use Manual23
Introduction to configuration and commissioningConfiguration and Use Manual
June 2021 MMI-20029970
4. Configure Main Menu Available as desired.
OptionDescription
EnabledThe local display Menu option from the process variable screen will be accessible.
Disabled The local display Menu option from the process variable screen will not be accessible.
Important
Once Main Menu Available has been disabled, you cannot enable it from the local display. Use another
configuration tool, such as ProLink III, to re-enable main menu access from the local display.
3.1.6 Enable or disable fieldbus write lock
When locked, the fieldbus write lock prevents any configuration changes being written from the fieldbus
segment.
Procedure
Set the Write Lock parameter (OD index 34) of the Resource block to Locked (1) or Unlocked (0).
3.2 Work with configuration files
You can save the current transmitter configuration in two forms: a backup file and a replication file. You can
save the configuration to the SD card on your transmitter or to a USB drive.
Tip
You can use a saved configuration file to change the nature of the transmitter quickly. This might be
convenient if the transmitter is used for different applications or different process fluids.
You can load a configuration file to the transmitter's working memory or to the transmitter's SD card. You can
load either a backup file or a replication file.
Backup files
Replication files
3.2.1 Save a configuration file using the display
Prerequisites
If you are planning to use the USB drive, the service port must be enabled. It is enabled by default. However, if
you need to enable it, choose Menu→Configuration→Security and set Service Port to On.
Contain all parameters. They are used to restore the current device if required.
The .spare extension is used to identify backup files.
Contain all parameters except the device-specific parameters, e.g., calibration factors or
meter factors. They are used to replicate the transmitter configuration to other devices.
The .xfer extension is used to identify replication files.
24Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20029970June 2021
Procedure
• To save the current configuration to the transmitter's SD card as a backup file:
a) Choose Menu→Configuration→Save/Restore Config→Save Config to Memory.
b) Enter the name for this configuration file.
The configuration file is saved to the transmitter's SD card as yourname.spare.
• To save the current configuration to a USB drive, as either a backup file or a replication file:
a)
b) Choose Menu→USB Options→Transmitter --> USB Drive→Save Active Config to USB Drive.
c) Choose Backup or Replicate.
d) Enter the name for this configuration file.
The configuration file is saved to the USB drive as yourname.spare or yourname.xfer.
• To copy a configuration file from the transmitter's SD card to the USB drive:
a)WARNING
b) Choose Menu→USB Options→Transmitter --> USB Drive→Transfer Config File to USB Drive.
c) Choose Backup or Replicate.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is
powered up. Failure to follow these instructions can cause an explosion resulting in injury or
death.
Open the wiring compartment on the transmitter and insert a USB drive into the service port.
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is
powered up. Failure to follow these instructions can cause an explosion resulting in injury or
death.
Open the wiring compartment on the transmitter and insert a USB drive into the service port.
d) Select the file that you want to transfer.
The configuration file is copied to the USB drive, using its existing name.
3.2.2 Save a configuration file using ProLink III
Note
When you use ProLink III format for configuration files, you can specify configuration parameters individually
or by groups. Therefore, you can use this format for both backup and replication.
Procedure
• To save the current configuration to the transmitter's SD card:
a) Choose Device Tools→Configuration Transfer→Save Configuration.
b) Select On my 5700 Device Internal Memory and select Next.
c) Select Save.
Configuration and Use Manual25
Introduction to configuration and commissioningConfiguration and Use Manual
June 2021 MMI-20029970
d) Enter the name for this configuration file.
e) Set the file type.
— To save a backup file, set the file type to Backup.
— To save a replication file, set the file type to Transfer.
f) Select Save.
The configuration file is saved to the transmitter's SD card as yourname.spare or yourname.xfer.
• To save the current configuration to your PC, in 5700 format:
a) Choose Device Tools→Configuration Transfer→Save Configuration.
b) Select On my computer in 5700 device file format and select Next.
c) Select Save.
d) Browse to the desired location, then enter the name for this configuration file.
e) Set the file type.
— To save a backup file, set the file type to Backup.
— To save a replication file, set the file type to Transfer.
f) Select Save.
The configuration file is saved to the specified location as yourname.spare or yourname.xfer.
• To save the current configuration to your PC, in ProLink III format:
a) Choose Device Tools→Configuration Transfer→Save Configuration.
b) Select On my computer in ProLink III file format and click Next.
c) Select Save.
d) Select the configuration parameters to be included in this file.
— To save a backup file, select all parameters.
— To save a replication file, select all parameters except device-specific parameters.
e) Select Save.
f) Browse to the desired location, then enter the name for this configuration file.
g) Set the file type to ProLink configuration file.
h) Select Start Save.
The configuration file is saved to the specified location as yourname.pcfg.
3.2.3 Save a configuration file using a basic FF host
Procedure
• To save the current configuration to the transmitter's SD card as a backup or replication file:
26Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20029970June 2021
a) Verify or write the appropriate value to the Config file type parameter of the Device TB for the type
of file you want to save.
— 1 for a backup (spare) file.
— 3 for a replication file.
b) Enter the name for the configuration file in the File Name parameter of the Device TB.
c) Write a 1 to the Save Config File parameter of the Device TB.
The configuration file is saved to the transmitter's SD card as yourname.spare or yourname.xfer,
depending on the type.
3.2.4 Load a configuration file using the display
Prerequisites
You must have a backup file or a replication file available for use.
If you are planning to use the USB drive, the service port must be enabled. It is enabled by default. However, if
you need to enable it, choose Menu→Configuration→Security and set Service Port to On.
Procedure
• To load either a backup file or a replication file from the transmitter's SD card:
a) Choose Menu→Configuration→Save/Restore Config→Restore Config from Memory.
b) Select Backup or Replicate.
c) Select the file that you want to load.
The file is loaded to working memory and becomes active immediately.
• To load a either a backup file or a replication file from a USB drive:
a)
b) Choose Menu→USB Options→USB Drive --> Transmitter→Upload Configuration File.
c) Select Backup or Replicate.
d) Select the file that you want to load.
e) Choose Yes or No when prompted to apply the settings.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is
powered up. Failure to follow these instructions can cause an explosion resulting in injury or
death.
Open the wiring compartment on the transmitter and insert the USB drive containing the backup
file or replication file into the service port.
— Yes: The file is loaded to working memory and becomes active immediately.
— No: The file is loaded to the transmitter's SD card but not to working memory. You can load it
from the SD card to working memory at a later time.
Configuration and Use Manual27
Introduction to configuration and commissioningConfiguration and Use Manual
June 2021 MMI-20029970
3.2.5 Load a configuration file using ProLink III
You can load a configuration file to the transmitter's working memory. You can load a backup file or a
replication file. Two PC file formats are supported: the 5700 format and the ProLink III format.
Note
When you use ProLink III format for configuration files, you can specify configuration parameters individually
or by groups. Therefore, you can use this format for both backup and replication.
Procedure
• To load a backup file or replication file from the transmitter's SD card:
a) Choose Device Tools→Configuration Transfer→Load Configuration.
b) Select On my 5700 Device Internal Memory and select Next.
c) Select Restore.
d) Set the file type.
— To load a backup file, set the file type to Backup.
— To load a replication file, set the file type to Transfer.
e) Select the file that you want to load and select Load.
The parameters are written to working memory, and the new settings become effectively immediately.
• To load a backup file or replication file in 5700 format from the PC:
a) Choose Device Tools→Configuration Transfer→Load Configuration.
b) Select On my computer in 5700 device file format and select Next.
c) Select Restore.
d) Set the file type.
— To load a backup file, set the file type to Backup.
— To load a replication file, set the file type to Transfer.
e) Navigate to the file you want to load, and select it.
The parameters are written to working memory, and the new settings become effectively immediately.
• To load a file in ProLink III format from the PC:
a) Choose Device Tools→Configuration Transfer→Load Configuration.
b) Select On my computer in ProLink III file format and select Next.
c) Select the parameters that you want to load.
d) Select Load.
e) Set the file type to Configuration file.
f) Navigate to the file you want to load, and select it.
g) Select Start Load.
28Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20029970June 2021
The parameters are written to working memory, and the new settings become effectively immediately.
3.2.6 Load a configuration file using a basic FF host
You can load a backup or replication configuration file to the transmitter's working memory from the SD card
using a basic FF host. If you need to load a file from a USB drive, you must use ProLink III or the display.
Prerequisites
You must have a backup file or a replication file available for use.
Procedure
• To load either a backup file or a replication file from the transmitter's SD card:
a) Verify or write the appropriate value to the Config file type parameter of the Device TB for the type
of file you want to load.
— 1 for a backup (spare) file.
— 3 for a replication file.
b) Enter the name of the file you want to restore in the File Name parameter of the Device TB.
c) Write a 1 to the Restore Config File parameter of the Device TB.
The file is loaded to working memory and becomes active immediately.
Basic FF hostMeasurement TB→Restore Factory Configuration (OD Index 122)
A file containing the factory configuration is always saved in the transmitter's internal memory, and is
available for use.
This action is typically used for error recovery or for repurposing a transmitter.
If you restore the factory configuration, the real-time clock, the audit trail, the historian, and other logs are
not reset.
Menu → Configuration → Save/Restore Configuration → Restore Config from Memory
3.2.8 Replicate a transmitter configuration
Replicating a transmitter configuration is a fast method to set up similar or identical measurement points.
Procedure
1. Configure a transmitter and verify its operation and performance.
2. Use any available method to save a replication file from that transmitter.
3. Use any available method to load the replication file to another transmitter.
Configuration and Use Manual29
Introduction to configuration and commissioningConfiguration and Use Manual
June 2021 MMI-20029970
4. At the replicated transmitter, set device-specific parameters and perform device-specific procedures:
a) Set the clock.
b) Set the tag and related parameters.
c) Characterize the transmitter.
d) Perform zero validation and take any recommended actions.
e) Perform loop tests and take any recommended actions, including mA Output trim.
f) Use sensor simulation to verify transmitter response.
5. At the replicated transmitter, make any other configuration changes.
6. Follow your standard procedures to ensure that the replicated transmitter is performing as desired.
30Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement
MMI-20029970June 2021
4 Configure process measurement
4.1 Configure Sensor Flow Direction Arrow
DisplayMenu→Configuration→Process Measurement→Flow Variables→Flow Direction
ProLink IIIDevice Tools→Configuration→Process Measurement→Flow→Sensor Direction
Field communicatorConfigure→Manual Setup→Measurements→Flow→Sensor Direction
Enhanced FF hostConfigure→Manual Setup→Measurements→Flow→Sensor Direction
Basic FF hostMeasurement TB→Flow Direction (OD Index 30)
Sensor Flow Direction Arrow is used to accommodate installations in which the Flow arrow on the sensor
does not match the majority of the process flow. This typically happens when the sensor is accidentally
installed backwards.
Sensor Flow Direction Arrow interacts with mA Output Direction, Frequency Output Direction, and
Totalizer Direction to control how flow is reported by the outputs and accumulated by the totalizers and
inventories.
The Sensor Flow Direction Arrow also affects how flow is reported on the transmitter display and via digital
communications. This includes ProLink III, a field communicator, the FF host, and all other user interfaces.
Figure 4-1: Flow arrow on sensor
A. Flow arrow
B. Actual flow direction
Procedure
Set Sensor Flow Direction Arrow as appropriate.
Option
Description
With ArrowThe majority of flow through the sensor matches the Flow arrow on the sensor. Actual
forward flow is processed as forward flow.
Against Arrow The majority of flow through the sensor is opposite to the Flow arrow on the sensor. Actual
forward flow is processed as reverse flow.
Configuration and Use Manual31
Configure process measurementConfiguration and Use Manual
June 2021 MMI-20029970
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow direction or the
setting of Sensor Flow Direction Arrow. Sensor Flow Direction Arrow controls only whether actual flow is
processed as forward flow or reverse flow.
Related information
Configure mA Output Direction
Configure Frequency Output Direction
Configure Discrete Output Source
Configure totalizers and inventories
Effect of Sensor Flow Direction Arrow on digital communications
4.2 Configure mass flow measurement
The mass flow measurement parameters control how mass flow is measured and reported. The mass total
and mass inventory are derived from the mass flow data.
ProLink IIIDevice Tools→Configuration→Process Measurement→Flow→Mass Flow Rate Unit
Field communicatorConfigure→Manual Setup→Measurements→Flow→Mass Flow Unit
Enhanced FF hostConfigure→Manual Setup→Measurements→Flow→Mass Flow Unit
Basic FF hostMeasurement TB→Mass Flow Unit (OD Index 19)
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow rate. The
default unit used for mass total and mass inventory is derived from this unit.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
Default: g/sec (grams per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one
user-defined special measurement unit. Different communications tools may use different labels for the
units.
Label
Unit description
DisplayProLink III
Grams per secondgram/sg/secg/sg/s1318
Grams per minutegram/ming/ming/ming/min1319
32Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Field
communicator
Enhanced FF
host
Basic FF host
Configuration and Use ManualConfigure process measurement
MMI-20029970June 2021
Label
Unit description
DisplayProLink III
Grams per hourgram/hg/hrg/hg/h1320
Kilograms per secondkg/skg/seckg/skg/s1322
Kilograms per minutekg/minkg/minkg/minkg/min1323
Kilograms per hourkg/hkg/hrkg/hkg/h1324
Kilograms per daykg/dkg/daykg/dkg/d1325
Metric tons per minuteMetTon/minmTon/minMetTon/mint/min1327
Metric tons per hourMetTon/hmTon/hrMetTon/ht/h1328
Metric tons per dayMetTon/dmTon/dayMetTon/dt/d1329
Basic FF hostMeasurement TB→Flow Damping (OD Index 29)
Flow Damping controls the amount of damping that will be applied to the measured mass flow rate. It affects
flow rate process variables that are based on the measured mass flow rate. This includes volume flow rate and
gas standard volume flow rate.
Menu → Configuration → Process Measurement → Flow Variables → Flow Damping
Flow Damping also affects specialized flow rate variables such as temperature-corrected volume flow rate
(API Referral) and net mass flow rate (concentration measurement).
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value
specifies the time period, in seconds, over which the transmitter will spread changes in the process variable.
34Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement
MMI-20029970June 2021
At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the
change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
• Default: 0.64 seconds
• Range: 0 seconds to 60 seconds
Note
If a number greater than 60 is entered, it is automatically changed to 60.
Tip
• A high damping value makes the process variable appear smoother because the reported value changes
slowly.
• A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
• The combination of a high damping value and rapid, large changes in flow rate can result in increased
measurement error.
• Whenever the damping value is non-zero, the reported measurement will lag the actual measurement
because the reported value is being averaged over time.
• In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the reported value.
• The transmitter automatically rounds off any entered damping value to the nearest valid value. Therefore,
the recommended damping value for gas applications should be 3.2 seconds. If you enter 2.56, the
transmitter will round it off to 3.2.
• For filling applications, Micro Motion recommends using the default value of 0.04 seconds.
Effect of flow damping on volume measurement
Flow damping affects volume measurement for liquid volume data. Flow damping also affects volume
measurement for gas standard volume data. The transmitter calculates volume data from the damped mass
flow data.
Interaction between Flow Damping and mA Output Damping
In some circumstances, both Flow Damping and mA Output Damping are applied to the reported mass flow
value.
Flow Damping controls the rate of change in flow process variables. mA Output Damping controls the rate
of change reported via the mA Output. If mA Output Process Variable is set to Mass Flow Rate, and both
Flow Damping and mA Output Damping are set to non-zero values, flow damping is applied first, and the
added damping calculation is applied to the result of the first calculation.
Configuration and Use Manual35
Configure process measurementConfiguration and Use Manual
Basic FF hostMeasurement TB→Mass Flow Cutoff (OD Index 26)
Mass Flow Cutoff specifies the lowest mass flow rate that will be reported as measured. All mass flow rates
below this cutoff will be reported as 0.
Procedure
Set Mass Flow Cutoff to the value you want to use.
• Default: A sensor-specific value set at the factory. If your transmitter was ordered without a sensor, the
default may be 0.0.
• Recommendation: 0.5% of maximum flow rate of the attached sensor. See the sensor specifications.
Important
Do not use your meter for measurement with Mass Flow Cutoff set to 0.0 g/sec. Ensure that Mass Flow
Cutoff is set to the value that is appropriate for your sensor.
Cutoff
Effect of Mass Flow Cutoff on volume measurement
Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the actual mass data
rather than the reported value.
Volume flow has a separate Volume Flow Cutoff that is not affected by the Mass Flow Cutoff value.
Interaction between Mass Flow Cutoff and mA Output Cutoff
Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured. mA
Output Cutoff defines the lowest flow rate that will be reported via the mA Output. If mA Output Process
Variable is set to Mass Flow Rate, the mass flow rate reported via the mA Output is controlled by the higher of
the two cutoff values.
Mass Flow Cutoff affects all reported values and values used in other transmitter behavior (e.g., events
defined on mass flow).
mA Output Cutoff affects only mass flow values reported via the mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Mass Flow Cutoff
Configuration:
• mA Output Process Variable: Mass Flow Rate
• Frequency Output Process Variable: Mass Flow Rate
• mA Output Cutoff: 10 g/sec
• Mass Flow Cutoff: 15 g/sec
36Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement
MMI-20029970June 2021
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0 will be used in all
internal processing.
Example: Cutoff interaction with mA Output Cutoff higher than Mass Flow Cutoff
Configuration:
• mA Output Process Variable: Mass Flow Rate
• Frequency Output Process Variable: Mass Flow Rate
• mA Output Cutoff: 15 g/sec
• Mass Flow Cutoff: 10 g/sec
Result:
• If the mass flow rate drops below 15 g/sec but not below 10 g/sec:
— The mA Output will report zero flow.
— The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
• If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
4.3 Configure volume flow measurement for liquid
applications
The volume flow measurement parameters control how liquid volume flow is measured and reported. The
volume total and volume inventory are derived from volume flow data.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one
or the other.
4.3.1 Configure Volume Flow Type for liquid applications
Display
ProLink III
Field communicator
Enhanced FF host
Basic FF host
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
Gas standard volume measurement is incompatible with the following applications:
• API Referral
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings → Flow
Type → Liquid
Device Tools → Configuration → Process Measurement → Flow → Volume Flow Type → Liquid Volume
ProLink IIIDevice Tools→Configuration→Process Measurement→Flow→Volume Flow Rate Unit
Field communicatorConfigure→Manual Setup→Measurements→Flow→Volume Flow Unit
Enhanced FF hostConfigure→Manual Setup→Measurements→Flow→Volume Flow Unit
Basic FF hostMeasurement TB→Volume Flow Unit (OD Index 31)
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the volume
flow rate. The unit used for the volume total and volume inventory is based on this unit.
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
Default: l/sec (liters per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid applications
Label
Unit description
DisplayProLink III
Cubic feet per secondft3/sft3/secCuft/sCFS1356
Cubic feet per minuteft3/minft3/minCuft/minCFM1357
Cubic feet per hourft3/hft3/hrCuft/hCFH1358
Cubic feet per dayft3/dft3/dayCuft/dft³/d1359
Cubic meters per secondm3/sm3/secCum/sm³/s1347
Cubic meters per minutem3/minm3/minCum/minm³/min1348
Field
communicator
Enhanced FF
host
Basic FF host
code
Cubic meters per hourm3/hm3/hrCum/hm³/h1349
Cubic meters per daym3/dm3/dayCum/dm³/d1350
U.S. gallons per secondgal/sUS gal/secgal/sgal/s1362
38Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement
MMI-20029970June 2021
Label
Unit description
DisplayProLink III
U.S. gallons per minutegal/mUS gal/mingal/minGPM1363
U.S. gallons per hourgal/hUS gal/hrgal/hgal/h1364
U.S. gallons per daygal/dUS gal/daygal/dgal/d1365
Field
communicator
Enhanced FF
host
Basic FF host
code
Million U.S. gallons per
MMgal/dmil US gal/dayMMgal/dMgal/d1366
day
Liters per secondL/sl/secL/sL/s1351
Liters per minuteL/minl/minL/inL/min1352
Liters per hourL/hl/hrL/hL/h1353
Million liters per dayMML/dmil l/dayML/dML/d1355
Imperial gallons per
Impgal/sImp gal/secImpgal/sImpGal/s1367
second
Imperial gallons per
Impgal/mImp gal/minImpgal/minImpGal/min1368
minute
Imperial gallons per hourImpgal/hImp gal/hrImpgal/hImpGal/h1369
Imperial gallons per dayImpgal/dImp gal/dayImpgal/dImpGal/d1370
Basic FF hostMeasurement TB→Volume Flow Cutoff (OD Index 38)
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings → Low Flow
Cutoff
40Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement
MMI-20029970June 2021
Volume Flow Cutoff specifies the lowest volume flow rate that will be reported as measured. All volume flow
rates below this cutoff are reported as 0.
Procedure
Set Volume Flow Cutoff to the value you want to use.
• Default: 0.0 l/sec (liters per second)
• Range: 0 l/sec to x l/sec, where x is the sensor’s flow calibration factor, multiplied by 0.0002.
Interaction between Volume Flow Cutoff and mAO Cutoff
Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will report as measured.
mAO Cutoff defines the lowest flow rate that will be reported via the mA Output. If mA Output Process
Variable is set to Volume Flow Rate, the volume flow rate reported via the mA Output is controlled by the
higher of the two cutoff values.
Volume Flow Cutoff affects both the volume flow values reported via the outputs and the volume flow values
used in other transmitter behavior (e.g., events defined on the volume flow).
mAO Cutoff affects only flow values reported via the mA Output.
Example: Cutoff interaction with mAO Cutoff lower than Volume Flow Cutoff
Configuration:
• mA Output Process Variable: Volume Flow Rate
• Frequency Output Process Variable: Volume Flow Rate
• AO Cutoff: 10 l/sec
• Volume Flow Cutoff: 15 l/sec
Result: If the volume flow rate drops below 15 l/sec, volume flow will be reported as 0, and 0 will be used in all
internal processing.
Example: Cutoff interaction with mAO Cutoff higher than Volume Flow Cutoff
Configuration:
• mA Output Process Variable: Volume Flow Rate
• Frequency Output Process Variable: Volume Flow Rate
• AO Cutoff: 15 l/sec
• Volume Flow Cutoff: 10 l/sec
Result:
• If the volume flow rate drops below 15 l/sec but not below 10 l/sec:
— The mA Output will report zero flow.
— The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
• If the volume flow rate drops below 10 l/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
Configuration and Use Manual41
Configure process measurementConfiguration and Use Manual
June 2021 MMI-20029970
4.4 Configure Gas Standard Volume (GSV) flow
measurement
The gas standard volume (GSV) flow measurement parameters control how gas standard volume flow is
measured and reported.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one
or the other.
4.4.1 Configure Volume Flow Type for gas applications
Display
ProLink III
Field communicator
Enhanced FF host
Basic FF host
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings → Flow
Type → Gas
Device Tools → Configuration → Process Measurement → Flow → Volume Flow Type → Gas Standard
Volume
Configure → Manual Setup → Measurements → Optional Setup → GSV → Volume Flow Type →
Standard Gas Volume
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
Gas standard volume measurement is incompatible with the following applications:
• API Referral
• Concentration measurement
• Advanced Phase Measurement — liquid with gas
For these applications, set Volume Flow Type to Liquid.
Procedure
Set Volume Flow Type to Gas.
4.4.2 Configure Standard Gas Density
Display
ProLink IIIDevice Tools→Configuration→Process Measurement→Flow→Standard Density of Gas
Field communicatorConfigure→Manual Setup→Measurements→Optional Setup→GSV→Gas Ref Density
Enhanced FF hostConfigure→Manual Setup→Measurements→Optional Setup→GSV→Gas Ref Density
Basic FF hostMeasurement TB→Gas Reference Density (OD Index 53)
42Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings → Standard
Gas Density
Configuration and Use ManualConfigure process measurement
MMI-20029970June 2021
Standard Gas Density is the density of your gas at reference temperature and reference pressure. This is
often called standard density or base density. It is used to calculate the GSV flow rate from the mass flow rate.
Procedure
Set Standard Gas Density to the density of your gas at reference temperature and reference pressure.
You can use any reference temperature and reference pressure that you choose. It is not necessary to
configure these values in the transmitter.
Tip
ProLink III provides a guided method that you can use to calculate the standard density of your gas if you do
not know it.
4.4.3 Configure Gas Standard Volume Flow Measurement Unit
ProLink IIIDevice Tools→Configuration→Process Measurement→Flow→Gas Standard Volume Flow Unit
Field communicatorConfigure→Manual Setup→Measurements→Flow→GSV Flow Unit
Enhanced FF hostConfigure→Manual Setup→Measurement→Gas Standard Volume Flow→Unit
Basic FF hostMeasurement TB→Gas Standard Volume Flow Unit (OD Index 55)
Gas Standard Volume Flow Measurement Unit specifies the unit of measure that will be used for the gas
standard volume (GSV) flow rate. The unit used for gas standard volume total and gas standard volume
inventory is derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Measurement Unit, be sure that Volume Flow Type is set
to Gas Standard Volume.
Procedure
Set Gas Standard Volume Flow Measurement Unit to the unit you want to use.
Default: SCFM (Standard Cubic Feet per Minute)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Configuration and Use Manual43
Configure process measurementConfiguration and Use Manual
June 2021 MMI-20029970
Options for Gas Standard Volume Flow Measurement Unit
The transmitter provides a standard set of measurement units for Gas Standard Volume Flow Measurement
Unit, plus one user-defined special measurement unit. Different communications tools may use different
labels for the units.
Label
Unit description
DisplayProLink III
Field
communicator
Enhanced FF
host
Basic FF host
Normal cubic meters per
second
Normal cubic meters per
minute
Normal cubic meters per
hour
Normal cubic meters per
day
Normal liter per secondNLPSNLPSNLPSNL/s1532
Normal liter per minuteNLPMNLPMNLPMNL/min1533
Normal liter per hourNLPHNLPHNLPHNL/h1534
Normal liter per dayNLPDNLPDNLPDNL/d1535
Standard cubic feet per
second
Standard cubic feet per
minute
Standard cubic feet per
hour
Standard cubic feet per
day
NCMSNm3/secNm3/secNm³/s1522
NCMMNm3/minNm³/minNm³/min1523
NCMHNm3/hrNm3/hrNm³/h1524
NCMDNm3/dayNm3/dayNm³/d1525
SCFSSCFSSCFSSCFS33000
SCFMSCFMSCFMSCFM1360
SCFHSCFHSCFHSCFH1361
SCFDSCFDSCFDSCFD33001
Standard cubic meters per
second
Standard cubic meters per
minute
Standard cubic meters per
hour
Standard cubic meters per
day
Standard liter per secondSLPSSLPSSLPSSL/s1537
Standard liter per minuteSLPMSLPMSLPMSL/min1538
Standard liter per hourSLPHSLPHSLPHSL/h1539
Standard liter per daySLPDSLPDSLPDSL/d1540
Special measurement unit SPECIALSpecialSpecialSpecial253
44Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
SCMSSm3/secSm3/secSm³/s1527
SCMMSm3/minSm3/minSm³/min1528
SCMHSm3/hrSm3/hrSm³/h1529
SCMDSm3/daySm3/daySm³/d1530
Configuration and Use ManualConfigure process measurement
MMI-20029970June 2021
Define a special measurement unit for gas standard volume flow
Display
ProLink III
Field communicator
Enhanced FF host
Basic FF host
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings → Units →
SPECIAL
Device Tools → Configuration → Process Measurement → Flow → Gas Standard Volume Flow Unit →
Special
Configure → Manual Setup → Measurements → Optional Setup → Special Units → Special Gas Standard
Volume Units
Configure → Manual Setup → Measurements → Optional Setup → Special Units → Special Gas Standard
Volume Units
Measurement TB → Gas Process Variables (OD Index 56–59, 61)
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer
data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is
calculated from an existing measurement unit using a conversion factor.
Procedure
1. Specify Base Gas Standard Volume Unit.
Base Gas Standard Volume Unit is the existing gas standard volume unit that the special unit will be
based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Gas Standard Volume Flow Conversion Factor as follows:
a) x base units = y special units
b) Gas Standard Volume Flow Conversion Factor = x ÷ y
4. Enter the Gas Standard Volume Flow Conversion Factor.
The original gas standard volume flow value is divided by this conversion factor.
5. Set Gas Standard Volume Flow Label to the name you want to use for the gas standard volume flow
unit.
6. Set Gas Standard Volume Total Label to the name you want to use for the gas standard volume total
and gas standard volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the
special measurement unit at any time.
Example: Defining a special measurement unit for gas standard volume flow
You want to measure gas standard volume flow in thousands of standard cubic feet per minute.
1. Set Base Gas Standard Volume Unit to SCFM.
2. Set Base Time Unit to minutes (min).
3. Calculate the conversion factor:
a. One thousands of standard cubic feet per minute = 1000 cubic feet per minute
b. Gas Standard Volume Flow Conversion Factor = 1 ÷ 1000 = 0.001
Configuration and Use Manual45
Configure process measurementConfiguration and Use Manual
June 2021 MMI-20029970
4. Set Gas Standard Volume Flow Conversion Factor to 0.001.
5. Set Gas Standard Volume Flow Label to KSCFM.
6. Set Gas Standard Volume Total Label to KSCF.
4.4.4 Configure Gas Standard Volume Flow Cutoff
Display
ProLink III
Field communicator
Enhanced FF host
Basic FF host
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings → Low Flow
Cutoff
Device Tools → Configuration → Process Measurement → Flow → Gas Standard Volume Flow Cutoff
Configure → Manual Setup → Measurements → Optional Setup → Gas Standard Volume Flow → Cutoff
Measurement TB → Gas Standard Volume Cutoff (OD Index 60)
Gas Standard Volume Flow Cutoff specifies the lowest gas standard volume flow rate that will reported as
measured. All gas standard volume flow rates below this cutoff will be reported as 0.
Procedure
Set Gas Standard Volume Flow Cutoff to the value you want to use.
• Default: 0.0
• Range: 0.0 to any positive value
Interaction between Gas Standard Volume Flow Cutoff and mA Output
Cutoff
Gas Standard Volume Flow Cutoff defines the lowest Gas Standard Volume flow value that the transmitter
will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported via the mA
Output. If mA Output Process Variable is set to Gas Standard Volume Flow Rate, the volume flow rate
reported via the mA Output is controlled by the higher of the two cutoff values.
Gas Standard Volume Flow Cutoff affects both the gas standard volume flow values reported through
outputs and the gas standard volume flow values used in other transmitter behavior (for example, events
defined on gas standard volume flow).
mA Output Cutoff affects only flow values reported via the mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Gas Standard Volume Flow Cutoff
Configuration:
• mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow Rate
• Frequency Output Process Variable: Gas Standard Volume Flow Rate
• mA Output Cutoff for the primary mA Output: 10 SLPM (standard liters per minute)
• Gas Standard Volume Flow Cutoff: 15 SLPM
Result: If the gas standard volume flow rate drops below 15 SLPM, the volume flow will be reported as 0, and 0
will be used in all internal processing.
46Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement
MMI-20029970June 2021
Example: Cutoff interaction with mA Output Cutoff higher than Gas Standard Volume Flow Cutoff
Configuration:
• mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow Rate
• Frequency Output Process Variable: Gas Standard Volume Flow Rate
• mA Output Cutoff for the primary mA Output: 15 SLPM (standard liters per minute)
• Gas Standard Volume Flow Cutoff: 10 SLPM
Result:
• If the gas standard volume flow rate drops below 15 SLPM but not below 10 SLPM:
— The primary mA Output will report zero flow.
— The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
• If the gas standard volume flow rate drops below 10 SLPM, both outputs will report zero flow, and 0 will be
used in all internal processing.
4.5 Configure density measurement
The density measurement parameters control how density is measured and reported. Density measurement
is used with mass flow rate measurement to determine liquid volume flow rate.
4.5.1 Configure Density Measurement Unit
Display
ProLink IIIDevice Tools→Configuration→Process Measurement→Density→Density Unit
Field communicatorConfigure→Manual Setup→Measurements→Density→Density Unit
Enhanced FF hostConfigure→Manual Setup→Measurements→Density→Density Unit
Basic FF hostMeasurement TB→Density Unit (OD Index 45)
Density Measurement Unit controls the measurement units that will be used in density calculations and
reporting.
Restriction
If the API Referral application is enabled, you cannot change the density measurement unit here. The density
measurement unit is controlled by the API table selection.
Procedure
Set Density Measurement Unit to the option you want to use.
Menu → Configuration → Process Measurement → Density → Units
Default: g/cm3 (grams per cubic centimeter)
Configuration and Use Manual47
Configure process measurementConfiguration and Use Manual
June 2021 MMI-20029970
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement Unit. Different
communications tools may use different labels.
Label
Unit description
Specific gravity
(1)
DisplayProLink III
SGUSGUSGUSGU1114
Field
communicator
Enhanced FF
host
Basic FF host
Grams per cubic
centimeter
Grams per literg/Lg/lg/Lg/L1105
Grams per milliliterg/mLg/mlg/mLg/ml1104
Kilograms per literkg/Lkg/lkg/Lkg/L1103
Kilograms per cubic meter kg/m3kg/m3kg/Cumkg/m³1097
Pounds per U.S. gallonlb/gallbs/USgallb/gallb/gal1108
Pounds per cubic footlb/ft3lbs/ft3lb/Cuftlb/ft³1107
Pounds per cubic inchlb/in3lbs/in3lb/CuInlb/in³1106
Degrees APIAPIAPIdegAPIdegAPI1113
Short ton per cubic yardSTon/yd3sT/yd3STon/yd³STon/yd³1109
(1) Non-standard calculation. This value represents line density divided by the density of water at 60 °F (15.6 °C).
Basic FF hostMeasurement TB→Density Damping (OD Index 49)
Menu → Configuration → Process Measurement → Density → Damping
Density Damping controls the amount of damping that will be applied to density data.
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value
specifies the time period, in seconds, over which the transmitter will spread changes in the process variable.
At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the
change in the actual measured value.
Procedure
Set Density Damping to the desired value.
• Default: 1.28 seconds
• Range: 0.0 to 60 seconds
48Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement
MMI-20029970June 2021
Tip
• A high damping value makes the process variable appear smoother because the reported value changes
slowly.
• A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
• The combination of a high damping value and rapid, large changes in density can result in increased
measurement error.
• Whenever the damping value is non-zero, the damped value will lag the actual measurement because the
damped value is being averaged over time.
• In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the damped value.
• If a number greater than 60 is entered, it is automatically changed to 60.
Effect of Density Damping on volume measurement
Density Damping affects liquid volume measurement. Liquid volume values are calculated from the damped
density value rather than the measured density value. Density Damping does not affect gas standard volume
measurement.
Interaction between Density Damping and mA Output Damping
When the mA Output is configured to report density, both Density Damping and mA Output Damping are
applied to the reported density value.
Density Damping controls the rate of change in the value of the process variable in transmitter memory. mA
Output Damping controls the rate of change reported via the mA Output.
If mA Output Source is set to Density, and both Density Damping and mA Output Damping are set to nonzero values, density damping is applied first, and the mA Output damping calculation is applied to the result
of the first calculation. This value is reported over the mA Output.
Basic FF hostMeasurement TB→Density Cutoff (OD Index 50)
Density Cutoff specifies the lowest density value that will be reported as measured. All density values below
this cutoff will be reported as 0.
Procedure
Menu → Configuration → Process Measurement → Density → Cutoff
Set Density Cutoff to the value you want to use.
Configuration and Use Manual49
Configure process measurementConfiguration and Use Manual
June 2021 MMI-20029970
• Default: 0.2 g/cm
• Range: 0.0 g/cm3 to 0.5 g/cm
3
3
Effect of Density Cutoff on volume measurement
Density Cutoff affects liquid volume measurement. If the density value goes below Density Cutoff, the
volume flow rate is reported as 0.
4.6 Configure temperature measurement
The temperature measurement parameters control how temperature data is processed. Temperature data is
used in several different ways, including temperature compensation, API Referral, and concentration
measurement.
Basic FF hostMeasurement TB→Temperature Unit (OD Index 41)
Temperature Measurement Unit specifies the unit that will be used for temperature measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
Default: °C (Celsius)
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit. Different
communications tools may use different labels for the units.
Basic FF hostMeasurement TB→Temperature Damping (OD Index 44)
Temperature Damping controls the amount of damping that will be applied to temperature data from the
sensor. Temperature Damping is not applied to external temperature data.
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value
specifies the time period, in seconds, over which the transmitter will spread changes in the process variable.
At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the
change in the actual measured value.
Procedure
Set Temperature Damping to the desired value.
• Default: 4.8 seconds
• Range: 0.0 to 80 seconds
Note
If a number greater than 80 is entered, it is automatically changed to 80.
Tip
• A high damping value makes the process variable appear smoother because the reported value changes
slowly.
• A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
• The combination of a high damping value and rapid, large changes in temperature can result in increased
measurement error.
• Whenever the damping value is non-zero, the damped value will lag the actual measurement because the
damped value is being averaged over time.
• In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the damped value.
Effect of Temperature Damping on process measurement
Temperature Damping affects all processes and algorithms that use temperature data from the internal
sensor RTD.
Temperature compensation
Temperature compensation adjusts process measurement to compensate for the effect of temperature on
the sensor tubes.
API Referral
Temperature Damping affects API Referral process variables only if the transmitter is configured to use
temperature data from the sensor. If an external temperature value is used for API Referral, TemperatureDamping does not affect API Referral process variables.
Configuration and Use Manual51
Configure process measurementConfiguration and Use Manual
June 2021 MMI-20029970
Concentration measurement
Temperature Damping affects concentration measurement process variables only if the transmitter is
configured to use temperature data from the sensor. If an external temperature value is used for
concentration measurement, Temperature Damping does not affect concentration measurement process
variables.
4.7 Configure Pressure Measurement Unit
Display
ProLink III
Field communicator
Enhanced FF host
Basic FF host
Menu → Configuration → Process Measurement → Pressure → Units
Device Tools → Configuration → Process Measurement → Pressure Compensation → Pressure Unit
Pressure Measurement Unit controls the measurement unit used for pressure. This unit must match the unit
used by the external pressure device.
Pressure data is used for pressure compensation and for API Referral. The device does not measure pressure
directly. You must set up a pressure input.
Procedure
Set Pressure Measurement Unit to the desired unit.
Default: psi
4.7.1 Options for Pressure Measurement Unit
The transmitter provides a standard set of measurement units for Pressure Measurement Unit. Different
communications tools may use different labels for the units. In most applications, set Pressure MeasurementUnit to match the pressure measurement unit used by the remote device.
Label
Unit description
Feet water @ 68 °FftH2O @68°FFt Water @ 68°FftH₂OftH2O (68°F)1154
Inches water @ 4 °CinH2O @4°CIn Water @ 4°CinH2O @4DegCinH2O (4°C)1147
Inches water @ 60 °FinH2O @60°FIn Water @ 60°FinH2O @60DegF inH2O (60°F)33003
Inches water @ 68 °FinH2O @68°FIn Water @ 68°FinH2OinH2O (68°F)1148
Millimeters water @ 4 °CmmH2O @4°Cmm Water @
Millimeters water @ 68 °FmmH2O @68°Fmm Water @
Millimeters mercury @
0 °C
52Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
DisplayProLink III
4°C
68°F
mmHg @0°Cmm Mercury @
0°C
Field
communicator
mmH2O
@4DegC
mmH2OmmH2O (68°F)1151
mmHgmmHg (0°F)1158
Enhanced FF
host
mmH2O (4°C)1150
Basic FF host
code
Configuration and Use ManualConfigure process measurement
Field communicatorConfigure→Manual Setup→Measurements→Approximate Velocity→Velocity Unit
Enhanced FF hostConfigure→Manual Setup→Measurements→Approximate Velocity→Velocity Unit
Basic FF hostMeasurement TB→Velocity Unit (OD Index 51)
Velocity Measurement Unit controls the measurement unit used to report velocity.
Menu → Configuration → Process Measurement → Velocity → Units
Procedure
Set Velocity Measurement Unit to the desired unit.
Default: m/sec
Configuration and Use Manual53
Configure process measurementConfiguration and Use Manual
June 2021 MMI-20029970
4.8.1 Options for Velocity Measurement Unit
The transmitter provides a standard set of measurement units for Velocity Measurement Unit. Different
communications tools may use different labels.
Label
Unit description
DisplayProLink III
Feet per minuteft/minft/minft/min1070
Feet per secondft/sft/secft/s1067
Inches per minutein/minin/minin/min1069
Inches per secondin/sin/secin/s1066
Meters per hourm/hm/hrm/h1063
Meters per secondm/sm/secm/s1061
Field
communicator or
enhanced FF host
Basic FF host code
54Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
5 Configure process measurement
applications
5.1 Set up the API Referral application
The API Referral application corrects line density to reference temperature and reference pressure according
to American Petroleum Institute (API) standards. The resulting process variable is referred density.
Restriction
The API Referral application is not compatible with the following applications:
• Gas Standard Volume Measurement (GSV)
• Advanced Phase Measurement
• Concentration measurement
5.1.1 Set up the API Referral application using the display
Enable the API Referral application using the display
The API Referral application must be enabled before you can perform any setup. If the API Referral application
was enabled at the factory, you do not need to enable it now.
Prerequisites
The API Referral application must be licensed on your transmitter.
Procedure
1. Choose Menu→Configuration→Process Measurement.
2. Choose Flow Variables→Volume Flow Settings and ensure that Flow Type is set to Liquid.
3. Return to the Process Measurement menu.
4. If the concentration measurement application is displayed in the list, choose Concentration
Measurement and ensure that Enabled/Disabled is set to Disabled.
The concentration measurement application and the API Referral application cannot be enabled
simultaneously.
5. Enable API Referral.
a) Choose Menu→Configuration→Process Measurement→API Referral.
b) Set Enabled/Disabled to Enabled.
Configuration and Use Manual55
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
Configure API Referral using the display
The API Referral parameters specify the API table, measurement units, and reference values to be used in
referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process
fluid.
You must know the reference temperature and reference pressure that you want to use.
2. Set API Table to the API table that you want to use to calculate referred density.
Each API table is associated with a specific set of equations. Choose your API table based on your
process fluid and the measurement unit that you want to use for referred density.
Your choice also determines the API table that will be used to calculate the correction factor for volume
(CTPL or CTL).
3. Refer to the API documentation and confirm your table selection.
a) Verify that your process fluid falls within range for line density, line temperature, and line
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
Acceptable limits:
• 230.0 x 10-6 to 930.0 x 10-6 per °F
• 414.0 x 10-6 to 1674.0 x 10-6 per °C
5. If required, set Reference Temperature to the temperature to which density will be corrected in
referred density calculations.
The default reference temperature is determined by the selected API table.
6. If required, set Reference Pressure to the pressure to which density will be corrected in referred
density calculations.
The default reference pressure is determined by the selected API table.
Set up temperature and pressure data for API Referral using the display
The API Referral application uses temperature and, optionally, pressure data in its calculations. You must
decide how to provide this data, then perform the required configuration and setup.
Note
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value
may produce inaccurate process data.
56Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
Prerequisites
The pressure measurement must be gauge pressure, not atmospheric pressure.
The pressure device must use the pressure unit that is configured in the transmitter.
If you are using an external temperature device, it must use the temperature unit that is configured in the
transmitter.
Procedure
1. Choose the method to be used to supply temperature data, and perform the required setup.
MethodDescriptionSetup
Internal
temperature
Digital
communications
Temperature data from the onboard temperature sensor
(RTD) will be used for all
measurements and
calculations. No external
temperature data will be
available.
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal temperature data.
a. Choose Menu→Configuration→Process Measurement
→Temperature.
b. Set External Temperature to Off.
a. Choose Menu→Configuration→Process Measurement
→Temperature.
b. Set External Temperature to On.
c. Perform the necessary host programming and
communications setup to write temperature data to the
transmitter at appropriate intervals.
2. Choose the method to be used to supply pressure data, and perform the required setup.
Method
Digital
communications
DescriptionSetup
A host writes pressure data to
the meter at appropriate
intervals.
a. Choose Menu→Configuration→Process Measurement
b. Set External Pressure to On.
c. Perform the necessary host programming and
→ Pressure → External Pressure.
communications setup to write pressure data to the
transmitter at appropriate intervals.
Postrequisites
Choose Menu→Service Tools→Service Data→View Process Variables and verify the values for External
Temperature and External Pressure.
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For digital communications:
— Verify that the host has access to the required data.
Configuration and Use Manual57
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
— Verify that the output variable is being correctly received and processed by the transmitter.
5.1.2 Set up the API Referral application using ProLink III
Enable the API Referral application using ProLink III
Prerequisites
The API Referral application must be licensed and enabled on your transmitter before you can perform any
setup. If the API Referral application was enabled at the factory, you do not need to enable it now.
Procedure
1. Choose Device Tools → Configuration → Process Measurement → Flow and ensure that Volume
C tablesLiquids with a constant base density or known thermal expansion coefficient
(TEC). You will be required to enter the TEC for your process fluid.
58Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
API table groupProcess fluids
D tablesLubricating oils
E tablesNGL (Natural Gas Liquids) and LPG (Liquid Petroleum Gas)
b) Set Referred Density Measurement Unit to the measurement units that you want to use for
referred density.
c) Select Apply.
These parameters uniquely identify the API table to be used to calculate referred density. The selected
API table is displayed, and the meter automatically changes the density unit, temperature unit,
pressure unit, and reference pressure to match the API table.
Your choice also determines the API table that will be used to calculate the correction factor for volume
(CTPL or CTL).
Restriction
Not all combinations are supported by the API Referral application. See the list of API tables in this
manual.
3. Refer to the API documentation and confirm your table selection.
a) Verify that your process fluid falls within range for line density, line temperature, and line
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
Acceptable limits:
• 230.0 x 10-6 to 930.0 x 10-6 per °F
• 414.0 x 10-6 to 1674.0 x 10-6 per °C
5. Set Reference Temperature to the temperature to which density will be corrected in referred density
calculations. If you choose Other, select the temperature measurement unit and enter the reference
temperature.
6. Set Reference Pressure to the pressure to which density will be corrected in referred density
calculations.
Set up temperature and pressure data for API Referral using ProLink III
The API Referral application uses temperature and, optionally, pressure data in its calculations. You must
decide how to provide this data, then perform the required configuration and setup.
Note
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value
may produce inaccurate process data.
Prerequisites
The pressure measurement must be gauge pressure, not atmospheric pressure.
The pressure device must use the pressure unit that is configured in the transmitter.
Configuration and Use Manual59
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
If you are using an external temperature device, it must use the temperature unit that is configured in the
transmitter.
2. Choose the method to be used to supply temperature data, and perform the required setup.
OptionDescriptionSetup
Internal RTD
temperature
data
Digital
communications
Temperature data from the onboard temperature sensor
(RTD) is used.
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal RTD temperature data.
a. Set Line Temperature Source to Internal RTD.
b. Select Apply.
a. Set Line Temperature Source to Fixed Value or Digital
Communications.
b. Select Apply.
c. Perform the necessary host programming and
communications setup to write temperature data to the
meter at appropriate intervals.
3. Choose the method you will use to supply pressure data, and perform the required setup.
OptionDescriptionSetup
Digital
communications
A host writes pressure data to
the meter at appropriate
intervals.
a. Set Pressure Source to Fixed Value or Digital
Communications.
b. Perform the necessary host programming and
communications setup to write pressure data to the meter
at appropriate intervals.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs group
on the ProLink III main window.
The current pressure value is displayed in the External Pressure field. Verify that the value is correct.
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For digital communications:
— Verify that the host has access to the required data.
— Verify that the output variable is being correctly received and processed by the transmitter.
60Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
5.1.3 Set up the API Referral application using a basic FF host
Enable the API Referral application using a basic FF host
The API Referral application must be enabled before you can perform any setup. If the API Referral application
was enabled at the factory, you do not need to enable it now.
Procedure
1. If necessary, disable the concentration measurement application: Write 0 to Device TB →
Concentration Measurement.
The concentration measurement application and the API Referral application cannot be enabled
simultaneously.
2. Enable the API Referral application: Write 1 to Device TB→API Referral.
Configure API Referral using a basic FF host
The API Referral parameters specify the API table, measurement units, and reference values to be used in
referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process
fluid.
You must know the reference temperature and reference pressure that you want to use.
Procedure
1. Specify the API table to use: API Referral TB→2540 CTL Table Type.
Each API table is associated with a specific set of equations. Your choice also determines the
measurement unit to be used for temperature and pressure, and the default values for reference
temperature and reference pressure.
The meter automatically changes the density unit, temperature unit, pressure unit, and reference
pressure to match the API table.
2. Refer to the API documentation and confirm your table selection.
a) Verify that your process fluid falls within range for line density, line temperature, and line
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
3. If you chose a C table, enter the Thermal Expansion Coefficient (TEC) for your process fluid: API Referral
TB→Thermal Expansion Coefficient.
Acceptable limits:
• 230.0 x 10-6 to 930.0 x 10-6 per °F
• 414.0 x 10-6 to 1674.0 x 10-6 per °C
Configuration and Use Manual61
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
4. If required, set the temperature to which density will be corrected in referred density calculations: API
Referral TB→Reference Temp.
The default reference temperature is determined by the selected API table.
5. If required, set the reference pressure to the pressure to which density will be corrected in referred
density calculations: API Referral TB→Reference Pressure.
The default reference pressure is determined by the selected API table. API Referral requires gauge
pressure.
Set up temperature and pressure data for API Referral using a basic FF
host
The API Referral application uses line temperature and line pressure data in its calculations. You must decide
how to provide this data, then perform the required configuration and setup.
Tip
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value
may produce inaccurate process data.
Important
• Line temperature data is used in several different measurements and calculations. It is possible to use the
internal RTD temperature in some areas and an external temperature in others. The transmitter stores the
internal RTD temperature and the external temperature separately. However, the transmitter stores only
one alternate temperature value, which may be either the external temperature or the configured fixed
value. Accordingly, if you choose a fixed temperature for some uses, and an external temperature for
others, the external temperature will overwrite the fixed value.
• Line pressure data is used in several different measurements and calculations. The transmitter stores only
one pressure value, which may be either the external pressure or the configured fixed value. Accordingly, if
you choose a fixed pressure for some uses, and an external pressure for others, the external pressure will
overwrite the fixed value.
Prerequisites
• The pressure measurement must be gauge pressure, not atmospheric pressure.
• The pressure device must use the pressure unit that is configured in the transmitter.
• If you are using an external temperature device, it must use the temperature unit that is configured in the
transmitter.
Procedure
1. Choose the method to be used to supply temperature data, and perform the required setup.
62Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
OptionDescriptionSetup
Internal RTD
temperature data
Fieldbus AO
function block
Digital
communications
Temperature data from the onboard temperature sensor
(RTD) is used.
Temperature from an external
device is used, supplied via the
AO function block.
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal RTD temperature data.
a. Write 0 to Measurement TB→Temperature
Compensation.
a. Write 1 to Measurement TB→Temperature
Compensation .
b. Ensure that the AO function block is set up as a
temperature source.
c. Connect the AO function block of the transmitter to the AI
function block of the external temperature device.
a. Perform the necessary host programming and
communications setup to write temperature data to the
meter at appropriate intervals.
2. Set up the pressure input.
a) Ensure that the AO function block is set up as a pressure source.
b) Connect the AO function block of the transmitter to the AI function block of the external
pressure device.
5.1.4 Set up the API Referral application using a field communicator
or an enhanced FF host
This section guides you through the tasks required to set up and implement the API Referral application using
a field communicator or an enhanced FF host.
Enable the API Referral application using a field communicator or an
enhanced FF host
Prerequisites
The API Referral application must be licensed and enabled on your transmitter. If the API Referral application
was enabled at the factory, you do not need to enable it now.
Volume Flow Type must be set to Liquid.
Procedure
1. Choose Configure → Manual Setup → Measurements → Optional Setup → GSV and ensure that
Volume Flow Type is set to Liquid.
This parameter is available only if API Referral or concentration measurement is not enabled. If you do
not see this parameter, it is already set correctly.
2. If the concentration measurement application is enabled, disable it.
The concentration measurement application and the API Referral application cannot be enabled
simultaneously.
Configuration and Use Manual63
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
3. Enable the API Referral application.
4. If Advance Phase Measurement→Output Type is other than Disabled, disable it.
The Advance Phase Measurement application and the API Referral application cannot be enabled
simultaneously.
Configure API Referral using a field communicator or an enhanced FF host
The API Referral parameters specify the API table, measurement units, and reference values to be used in
referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process
fluid.
You must know the reference temperature and reference pressure that you want to use.
3. Specify the API table that you want to use to calculate referred density.
Each API table is associated with a specific set of equations.
a) Set API Table Number to the number that matches the API table units that you want to use for
referred density.
Your choice also determines the measurement unit to be used for temperature and pressure,
and the default values for reference temperature and reference pressure.
Measurement
API table number
5°API°Fpsi (g)60 °F0 psi (g)
(1)
6
23SGU°Fpsi (g)60 °F0 psi (g)
(1)
24
53kg/m³°CkPa (g)15 °C0 kPa (g)
(1)
54
(2)
59
(2)
60
(1) Used only with API Table Letter = C.
(2) Used only with API Table Letter = E.
unit for referred
density
°API°Fpsi (g)60 °F0 psi (g)
SGU°Fpsi (g)60 °F0 psi (g)
kg/m³°CkPa (g)15 °C0 kPa (g)
kg/m³°CkPa (g)20 °C0 kPa (g)
kg/m³°CkPa (g)20 °C0 kPa (g)
Temperature
measurement
unit
Pressure
measurement
unit
Default reference
temperature
Default reference
pressure
b) Set API Table Letter to the letter of the API table group that is appropriate for your process fluid.
64Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
(1) Used only with API Table Number= 6, 24, or 54.
(2) Used only with API Table Number = 23, 24, 53, 54, 59, or 60.
Liquids with a constant base density or known thermal expansion coefficient
(TEC). You will be required to enter the TEC for your process fluid.
NGL (Natural Gas Liquids) and LPG (Liquid Petroleum Gas)
API Table Number and API Table Letter uniquely identify the API table. The selected API table is
displayed, and the meter automatically changes the density unit, temperature unit, pressure unit,
reference temperature, and reference pressure to match the API table.
Your choice also determines the API table that will be used to calculate the correction factor for volume
(CTPL or CTL).
Restriction
Not all combinations are supported by the API Referral application. See the list of API tables in this
manual.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
Acceptable limits:
• 230.0 x 10-6 to 930.0 x 10-6 per °F
• 414.0 x 10-6 to 1674.0 x 10-6 per °C
5. Refer to the API documentation and confirm your table selection.
a) Verify that your process fluid falls within range for line density, line temperature, and line
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
6. If required, set Reference Temperature to the temperature to which density will be corrected in
referred density calculations.
The default reference temperature is determined by the selected API table.
7. If required, set Reference Pressure to the pressure to which density will be corrected in referred
density calculations.
The default reference pressure is determined by the selected API table. API Referral requires gauge
pressure.
Configuration and Use Manual65
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
Set up temperature and pressure data for API Referral using a field
communicator
The API Referral application uses temperature and, optionally, pressure data in its calculations. You must
decide how to provide this data, then perform the required configuration and setup.
Note
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value
may produce inaccurate process data.
Procedure
1. Choose the method to be used to supply temperature data, and perform the required setup.
MethodDescriptionSetup
Internal RTD
temperature
data
Digital
communications
Temperature data from the onboard temperature sensor
(RTD) is used.
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal RTD temperature data.
communications setup to write pressure data to the
transmitter at appropriate intervals.
Postrequisites
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For digital communications:
— Verify that the host has access to the required data.
— Verify that the host is writing to the correct register in memory, using the correct data type.
66Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
Set up temperature and pressure data for API Referral using an enhanced
FF host
The API Referral application uses temperature and, optionally, pressure data in its calculations. You must
decide how to provide this data, then perform the required configuration and setup.
Note
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value
may produce inaccurate process data.
Procedure
1. Choose the method to be used to supply temperature data, and perform the required setup.
MethodDescriptionSetup
Internal RTD
temperature
data
Digital
communications
Temperature data from the onboard temperature sensor
(RTD) is used.
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal RTD temperature data.
Liquids with a
constant density
base or known
thermal expansion
coefficient
(5)
API tables
(calculations)
Referred
density
(2)
5A6AUnit: °API
(1)
CTL or
CTPL
Referred density
(API): unit and
range
(3) (4)
Default
reference
temp
Default
reference
pressure
API standard
60 °F0 psi (g)API MPMS 11.1
Range: 0 to 100 °API
23A24AUnit: SGU
60 °F0 psi (g)
Range: 0.6110 to
1.0760 SGU
53A54AUnit: kg/m
3
15 °C0 kPa (g)
Range: 610 to
1075 kg/m³
5B6BUnit: °API
60 °F0 psi (g)API MPMS 11.1
Range: 0 to 85 °API
23B24BUnit: SGU
60 °F0 psi (g)
Range: 0.6535 to
1.0760 SGU
53B54BUnit: kg/m
3
15 °C0 kPa (g)
Range: 653 to
1075 kg/m³
N/A6CUnit: °API60 °F0 psi (g)API MPMS 11.1
N/A24CUnit: SGU60 °F0 psi (g)
N/A54CUnit: kg/m³15 °C0 kPa (g)
5D6DUnit: °API
60 °F0 psi (g)API MPMS 11.1
Range: −10 to
+40 °API
Lubricating oils
23D24DUnit: SGU
Range: 0.8520 to
60 °F0 psi (g)
1.1640 SGU
53D54DUnit: kg/m³
15 °C0 kPa (g)
Range: 825 to
1164 kg/m³
NGL (natural gas
liquids) and LPG
(liquid petroleum
gas)
23E24EUnit: SGU60 °F0 psi (g)API MPMS 11.2.4
53E54EUnit: kg/m³15 °C0 psi (g)
59E60EUnit: kg/m³20 °C0 psi (g)
(1) Each API table represents a specialized equation defined by the American Petroleum Institute for a specific combination
of process fluid, line conditions, and output.
68Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
(2) Referred density is calculated from line density. You must specify this table, either directly or by selecting the process
fluid and base density measurement unit.
(3) You do not need to specify this table. It is invoked automatically as a result of the previous table selection.
(4) CTL is a correction factor based on online temperature. CTPL is a correction factor based on both line pressure and line
temperature. Calculation of CTL and CTPL for A, B, C, and D table products is in accordance with API MPMS Chapter
11.1. Calculation of CTL and CTPL for E table products is in accordance with API MPMS Chapters 11.2.2, 11.2.4, and
11.2.5.
(5) The Thermal Expansion Coefficient (TEC) replaces the referred density calculation. Use the CTL/CTPL table instead.
5.1.6 Process variables from the API Referral application
The API Referral application calculates several different process variables according to API standards.
CTPL
CTL
Referred density
API volume flow
Batch-weighted
average density
Batch-weighted
average
temperature
API volume total
API volume
inventory
Correction factor based on line temperature and line pressure.
Correction factor based on line temperature at saturation conditions.
The measured density after CTL or CTPL has been applied.
The measured volume flow rate after CTL or CTPL has been applied. Also called
corrected volume flow.
One density value is recorded for each unit of flow (e.g., barrel, liter). The average is
calculated from these values. The average is reset when the API totalizer is reset. Not
available unless a totalizer has been configured with Source set to Corrected Volume
Flow.
One temperature value is recorded for each unit of flow (e.g., barrel, liter). The
average is calculated from these values. The average is reset when the API totalizer is
reset. Not available unless a totalizer has been configured with Source set to
Temperature-Corrected Volume Flow.
The total API volume measured by the transmitter since the last API totalizer reset.
Also called corrected volume total. Not available unless a totalizer has been
configured with Source set to Corrected Volume Flow.
The total API volume measured by the transmitter since the last API inventory reset.
Also called corrected volume inventory. Not available unless an inventory has been
configured with Source set to Corrected Volume Flow.
5.2 Set up concentration measurement
The concentration measurement application calculates concentration from line density and line temperature.
5.2.1 Preparing to set up concentration measurement
The procedure for setting up concentration measurement application depends on how your device was
ordered and how you want to use the application. Review this information before you begin.
Requirements for concentration measurement
To use the concentration measurement application, the following conditions must be met:
• The concentration measurement application must be enabled.
• The API Referral application must be disabled.
Configuration and Use Manual69
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
• The Advanced Phase Measurement application must be disabled or set for the Liquid with Gas application.
• A concentration matrix must be loaded into one of the six slots on the transmitter.
Tip
In most cases, the concentration matrix that you ordered was loaded at the factory. If it was not, you have
several options for loading a matrix. You can also build a matrix.
• Temperature Source must be configured and set up.
• One matrix must be selected as the active matrix (the matrix used for measurement).
Requirements for matrices
A matrix is the set of coefficients used to convert process data to concentration, plus related parameters. The
matrix can be saved as a file.
The transmitter requires all matrices to be in .matrix format. You can use ProLink III to load matrices in other
formats:
• .edf (used by ProLink II)
• .xml (used by ProLink III)
The transmitter can store matrices in two locations:
• One of the six slots in memory
• The transmitter's SD card
Any matrix in a slot is available for use. In other words, it can be selected as the active matrix and used for
measurement. Matrices on the SD card are not available for use. Matrices must be loaded into a slot before
they can be used for measurement.
All matrices in slots must use the same derived variable. Matrices on the SD card have no requirement for
their derived variables to match.
Requirements for derived variables
A derived variable is the process variable that a concentration matrix measures. All other process variables are
calculated from the derived variable. There are eight possible derived variables. Each matrix is designed for
one specific derived variable.
The transmitter can store up to six matrices in six slots. There are additional matrices on the transmitter's SD
card. All matrices in the six slots must use the same derived variable. If you change the setting of DerivedVariable, all matrices are deleted from the six slots. Any matrices on the transmitter's SD card are not
affected.
Tip
Always ensure that Derived Variable is set correctly before loading matrices into slots.
Derived variables and net flow rate
If you want the transmitter to calculate Net Mass Flow Rate, the derived variable must be set to Mass
Concentration (Density). If your matrix is not designed for Mass Concentration (Density), contact customer
support for assistance.
If you want the transmitter to calculate Net Volume Flow Rate, the derived variable must be set to Volume
Concentration (Density). If your matrix is not designed for Volume Concentration (Density), contact customer
support for assistance.
70Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
Derived variables based on specific gravity
The following derived variables are based on specific gravity:
• Specific Gravity
• Concentration (Specific Gravity)
• Mass Concentration (Specific Gravity)
• Volume Concentration (Specific Gravity)
If you are using one of these derived variables, two additional parameters can be configured:
• Reference Temperature of Water (default setting: 4 °C)
• Water Density at Reference Temperature (default setting: 999.99988 kg/m³)
These two parameters are used to calculate specific gravity.
You cannot set these parameters from the display. If the default values are not appropriate, you must use
another method to set them.
Optional tasks in setting up concentration measurement
The following tasks are optional:
• Modifying names and labels
• Configuring extrapolation alerts
5.2.2 Set up concentration measurement using the display
This section guides you through most of the tasks related to setting up and implementing the concentration
measurement application.
Restriction
This section does not cover building a concentration matrix. For detailed information on building a matrix, see
the Micro Motion Enhanced Density Application Manual.
Enable concentration measurement using the display
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Disable the following applications before enabling concentration measurement as concentration
measurement cannot be enabled at the same time:
• Advanced Phase Measurement — gas with liquid
• API Referral
• Gas Standard Volume
Procedure
1. Choose Menu→Configuration→Process Measurement.
Configuration and Use Manual71
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
2. Choose Flow Variables→Volume Flow Settings and ensure that Flow Type is set to Liquid.
3. Return to the Process Measurement menu.
4. If the API Referral application is displayed in the menu, choose API Referral and ensure that Enabled/Disabled is set to Disabled.
The concentration measurement application and the API Referral application cannot be enabled
simultaneously.
5. If the Advanced Phase Measurement application is displayed in the menu, choose Advanced PhaseMeasurement→Application Setup and ensure that Enabled/Disabled is set to Disabled.
6. Enable concentration measurement.
a) Choose Menu→Configuration→Process Measurement→Concentration Measurement.
b) Set Enabled/Disabled to Enabled.
Load a concentration matrix from a USB drive using the display
At least one concentration matrix must be loaded into one of the six slots on your transmitter. You can load
up to six matrices into slots. You can also copy matrices to the transmitter's SD card, and load them into slots
at a later time.
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not
need to load any matrices.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is powered
up. Failure to follow these instructions can cause an explosion resulting in injury or death.
Prerequisites
The concentration measurement application must be enabled on your device.
For each concentration matrix that you want to load, you need a file containing the matrix data. The
transmitter's SD card and the ProLink III installation include a set of standard concentration matrices. Other
matrices are available from Micro Motion.
Each concentration matrix file must be in .matrix format.
Tip
• If you have a custom matrix on another device, you can save it to a file, then load it to the current device.
• If you have a matrix file in a different format, you can load it using ProLink III.
The .matrix files must be copied to the root directory of a USB drive.
You must know the derived variable that the matrix is designed to calculate.
Important
• All concentration matrices on your transmitter must use the same derived variable.
72Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
• If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the
six slots on the transmitter, but not from the SD card. Set Derived Variable before loading concentration
matrices.
Procedure
1. Choose Menu → Configuration → Process Measurement → Concentration Measurement →
Configure Application and ensure that the setting of Derived Variable matches the derived variableused by your matrix. If it does not, change it as required and click Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from
the six slots, but not from the transmitter's SD card. Verify the setting of Derived Variable before
continuing.
2. Load the matrix.
a) Remove the cover from the transmitter's wiring compartment, open the snap flap to access the
service port, and insert the USB drive into the service port.
b) Choose Menu→USB Options→USB Drive --> Transmitter→Upload Configuration File.
c) Set Config File Type to Concentration Measurement Matrix.
d) Select the .matrix file that you want to load, and wait for the transfer to complete.
3. Choose Yes or No when you are asked if you want to apply the settings.
The transmitter has six slots that are used to store concentration matrices. Any one of these can be
used for measurement. The transmitter also has the capability to store multiple concentration matrices
on its SD card. These cannot be used for measurement until they are moved to a slot.
Option
Description
YesThe matrix is saved to the SD card, and the loading process continues with loading the
matrix into one of the slots.
NoThe matrix is saved to the SD card, and the loading process ends. You must load a matrix into
a slot before you can use it for measurement.
4. If you chose Yes, select the slot to load this matrix into, and wait until the load is complete.
You can load the matrix into any empty slot, or you can overwrite an existing matrix.
Postrequisites
If you loaded the matrix into a slot, choose Menu → Configuration → Process Measurement →
Concentration Measurement → Configure Application → Active Matrix and ensure that the matrix is listed.
If you loaded the matrix onto the SD card only, choose Menu → Configuration → Process Measurement →
Concentration Measurement → Load Matrix and ensure that the matrix is listed.
Configuration and Use Manual73
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
Load a concentration matrix from the SD card using the display
If you have a concentration matrix on the transmitter's SD card, you can load it into one of the six slots on your
transmitter. You cannot use the matrix for measurement until it has been loaded into a slot. You can load up
to six matrices into slots.
Prerequisites
You must have one or more concentration matrices stored on the transmitter's SD card. The standard
matrices are loaded to the SD card at the factory.
You must know the derived variable that the matrix is designed to calculate.
Procedure
1. Choose Menu→Configuration→Process Measurement→Concentration Measurement and ensure
that the setting of Derived Variable matches the derived variable used by your matrix. If it does not,
change it as required and click Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from
the six slots, but not from the transmitter's SD card. Verify the setting of Derived Variable before
continuing.
Matrix.
The transmitter displays a list of all matrices that are on the SD card.
3. Select the matrix that you want to load.
4. Select the slot that you want to load it into.
You can load the matrix into any empty slot, or you can overwrite an existing matrix.
Postrequisites
Choose Menu → Configuration → Process Measurement → Concentration Measurement → Configure
Application → Active Matrix and ensure that the matrix is listed.
Set up temperature data using the display
The concentration measurement application uses line temperature data in its calculations. You must decide
how to provide this data, then perform the required configuration and setup. Temperature data from the onboard temperature sensor (RTD) is always available. Optionally, you can set up an external temperature
device and use external temperature data.
The temperature setup that you establish here will be used for all concentration measurement matrices on
this meter.
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the
internal RTD temperature in some areas and an external temperature in others. The transmitter stores the
internal RTD temperature and the external temperature separately. However, the transmitter stores only one
alternate temperature value, which may be either the external temperature or the configured fixed value.
Accordingly, if you choose a fixed temperature for some uses, and an external temperature for others, the
external temperature will overwrite the fixed value.
74Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
Procedure
Choose the method to be used to supply temperature data, and perform the required setup.
MethodDescriptionSetup
Internal
temperature
Digital
communications
Temperature data from the onboard temperature sensor
(RTD) will be used for all
measurements and
calculations. No external
temperature data will be
available.
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal temperature data.
a. Choose Menu→Configuration→Process Measurement
→Temperature.
b. Set External Temperature to Off.
a. Choose Menu→Configuration→Process Measurement
→Temperature.
b. Set External Temperature to On.
c. Perform the necessary host programming and
communications setup to write temperature data to the
transmitter at appropriate intervals.
Postrequisites
Choose Menu→Service Tools→Service Data→View Process Variables and verify the value for External
Temperature.
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For digital communications:
— Verify that the host has access to the required data.
— Verify that the output variable is being correctly received and processed by the transmitter.
Modify matrix names and labels using the display
For convenience, you can change the name of a concentration matrix and the label used for its measurement
unit. This does not affect measurement.
Procedure
1. Choose Menu → Configuration → Process Measurement → Concentration Measurement →
Configure Matrix.
2. Select the matrix that you want to modify.
3. Set Matrix Name to the name that will be used for this matrix.
4. Set Concentration Unit to the label that will be used for the concentration unit.
If you want to use a custom label, you can use the display to select Special. However, you cannot use
the display to configure the custom label. You must use another tool to change the label from Special
to a user-defined string.
Configuration and Use Manual75
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
Modify extrapolation alerts using the display
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control
the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density
or line temperature goes outside the range, the transmitter will extrapolate concentration values. However,
extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is
occurring.
Each concentration matrix has its own extrapolation alert limits.
Procedure
1. Choose Menu → Configuration → Process Measurement → Concentration Measurement →
Configure Matrix.
2. Select the matrix that you want to modify.
3. Set Extrapolation Limit to the point, in percent, at which an extrapolation alert will be posted.
5. Enable or disable the high and low limit alerts for temperature and density as desired.
Example: Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a
temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if
line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using the display
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose Menu → Configuration → Process Measurement → Concentration Measurement →
Configure Application.
2. Set Active Matrix to the matrix you want to use.
5.2.3 Set up concentration measurement using ProLink III
This section guides you through the tasks required to set up, configure, and implement concentration
measurement.
Enable concentration measurement using ProLink III
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
76Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
Disable the following applications before enabling concentration measurement as concentration
measurement cannot be enabled at the same time:
• Advanced Phase Measurement — gas with liquid
• API Referral
• Gas Standard Volume
Procedure
1. Choose Device Tools → Configuration → Process Measurement → Flow and ensure that Volume
4. Disable API Referral and set the Advance Phase Measurement application to Disabled or Single Liquid.
5. Set Concentration Measurement to Enabled and select Apply.
Load a concentration matrix using ProLink III
At least one concentration matrix must be loaded onto your transmitter. You can load up to six.
Prerequisites
The concentration measurement application must be enabled on your device.
For each concentration matrix that you want to load, you need a file containing the matrix data. The ProLink III
installation includes a set of standard concentration matrices. Other matrices are available from Micro
Motion. The file can be on your computer or in the transmitter's internal memory.
The file must be in one of the formats that ProLink III supports. This includes:
• .xml (ProLink III)
• .matrix (5700)
If you are loading an .xml file, you must know the following information for your matrix:
• The derived variable that the matrix is designed to calculate
• The density unit that the matrix was built with
• The temperature unit that the matrix was built with
If you are loading a .matrix file, you must know the derived variable that the matrix is designed to calculate.
Important
• All concentration matrices on your transmitter must use the same derived variable.
• If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the
six slots on the transmitter, but not from the transmitter's SD card. Set Derived Variable before loading
concentration matrices.
• ProLink III loads matrices directly to one of the transmitter's six slots.
Configuration and Use Manual77
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not
need to load any matrices.
Restriction
You cannot use ProLink III to load a matrix to the transmitter's SD card.
Procedure
1. If you are loading an .xml file, choose Device Tools → Configuration → Process Measurement → Line
Density and set Density Unit to the density unit used by your matrix.
Important
When you load a matrix in one of these formats, if the density unit is not correct, concentration data
will be incorrect. The density units must match at the time of loading. You can change the density unit
after the matrix is loaded.
2. If you are loading an .xml file, choose Device Tools→Configuration→Process Measurement→Line
Temperatureand set Temperature Unit to the temperature unit used by your matrix.
Important
When you load a matrix in one of these formats, if the temperature unit is not correct, concentration
data will be incorrect. The temperature units must match at the time of loading. You can change the
temperature unit after the matrix is loaded.
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
4. In Step 1, ensure that the setting of Derived Variable matches the derived variable used by your
matrix. If it does not, change it as required and select Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from
the six slots. Verify the setting of Derived Variable before continuing.
5. Load one or more matrices.
a) In Step 2, set Matrix Being Configured to the location (slot) to which the matrix will be loaded.
b) To load a .xml file from your computer, select Load Matrix from File, navigate to the file, and
load it.
c) To load a .matrix file from your computer, select Load Matrix from My Computer, navigate to
the file, and load it.
d) To load a .matrix file from the transmitter's internal memory, select Load Matrix from 5700
Device Memory, navigate to the file on the transmitter, and load it.
e) Repeat until all required matrices are loaded.
78Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
Set reference temperature values for specific gravity using ProLink III
When Derived Variable is set to any option based on specific gravity, you must set the reference temperature
for water, then verify the density of water at the configured reference temperature. These values affect
specific gravity measurement.
This requirement applies to the following derived variables:
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and select Change
Matrix.
3. Scroll to Step 3, then perform the following actions:
a) Set Reference Temperature for Referred Density to the temperature to which line density will
be corrected for use in the specific gravity calculation.
b) Set Reference Temperature for Water to the water temperature that will be used in the specific
gravity calculation.
c) Set Water Density at Reference Temperature to the density of water at the specified reference
temperature.
The transmitter automatically calculates the density of water at the specified temperature. The
new value will be displayed the next time that transmitter memory is read. You can enter a
different value if you prefer.
4. Select Apply at the bottom of Step 3.
Set up temperature data using ProLink III
The concentration measurement application uses line temperature data in its calculations. You must decide
how to provide this data, then perform the required configuration and setup. Temperature data from the onboard temperature sensor (RTD) is always available. Optionally, you can set up an external temperature
device and use external temperature data.
The temperature setup that you establish here will be used for all concentration measurement matrices on
this meter.
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the
internal RTD temperature in some areas and an external temperature in others. The transmitter stores the
internal RTD temperature and the external temperature separately. However, the transmitter stores only one
alternate temperature value, which may be either the external temperature or the configured fixed value.
Configuration and Use Manual79
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
Accordingly, if you choose a fixed temperature for some uses, and an external temperature for others, the
external temperature will overwrite the fixed value.
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 4.
3. Choose the method to be used to supply temperature data, and perform the required setup.
OptionDescriptionSetup
Internal
temperature
Digital
communications
Temperature data from the onboard temperature sensor
(RTD) will be used for all
measurements and
calculations. No external
temperature data will be
available.
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal RTD temperature data.
a. Set Line Temperature Source to Internal.
b. Click Apply.
a. Set Line Temperature Source to Fixed Value or Digital
Communications.
b. Click Apply.
c. Perform the necessary host programming and
communications setup to write temperature data to the
meter at appropriate intervals.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs group
on the ProLink III main window.
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For digital communications:
— Verify that the host has access to the required data.
— Verify that the output variable is being correctly received and processed by the transmitter.
Modify matrix names and labels using ProLink III
For convenience, you can change the name of a concentration matrix and the label used for its measurement
unit. This does not affect measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
80Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and click Change
Matrix.
3. Scroll to Step 3, then perform the following actions:
a) Set Concentration Units Label to the label that will be used for the concentration unit.
b) If you set Concentration Units Label to Special, enter the custom label in User-Defined Label.
c) In Matrix Name, enter the name to be used for the matrix.
4. Select Apply at the bottom of Step 3.
Modify extrapolation alerts using ProLink III
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control
the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density
or line temperature goes outside the range, the transmitter will extrapolate concentration values. However,
extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is
occurring.
Each concentration matrix has its own extrapolation alert limits.
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and click Change
Matrix.
3. Scroll to Step 4.
4. Set Extrapolation Alert Limit to the point, in percent, at which an extrapolation alert will be posted.
5. Enable or disable the high and low limit alerts for temperature and density, as desired, and click Apply.
Example: Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a
temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if
line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using ProLink III
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
2. Scroll to Step 2, set Active Matrix to the matrix you want to use and select Change Matrix.
Configuration and Use Manual81
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
5.2.4 Set up concentration measurement using a field
communicator or an enhanced FF host
This section guides you through most of the tasks related to setting up and implementing the concentration
measurement application.
Enable concentration measurement using a field communicator or an
enhanced FF host
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Disable the following applications before enabling concentration measurement as concentration
measurement cannot be enabled at the same time:
• Advanced Phase Measurement — gas with liquid
• API Referral
• Gas Standard Volume
Procedure
1. Choose Overview → Device Information → Licenses → Enable/Disable Applications and ensure that
3. Enable the concentration measurement application.
Load a concentration matrix from the transmitter's SD card using a field
communicator
If you have a concentration matrix on the transmitter's SD card, you can move it into one of the six slots on
your transmitter. You cannot use the matrix for measurement until it has been loaded into a slot. You can
load up to six matrices into slots.
Prerequisites
You must have one or more concentration matrices loaded onto the transmitter's SD card.
You must know the derived variable that the matrix is designed to calculate.
CM Configuration and ensure that the setting of Derived Variable matches the derived variable usedby your matrix. If it does not, change it as required and click Apply.
82Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from
the six slots, but not from the transmitter's SD card. Verify the setting of Derived Variable before
continuing.
You can load the matrix into any empty slot, or you can overwrite an existing matrix.
4. Enter the name of the matrix file on the SD card, without the .matrix extension.
Example
If the matrix file name is test.matrix, enter test.
Postrequisites
Choose Configure → Manual Setup → Measurements → Optional Setup → Conc Measurement → CM
Configuration → Active Matrix and ensure that the selected slot contains the matrix that you loaded.
Set reference temperature values for specific gravity using a field
communicator or an enhanced FF host
When Derived Variable is set to any option based on specific gravity, you must set the reference temperature
for water, then verify the density of water at the configured reference temperature. These values affect
specific gravity measurement.
To check the setting of Derived Variable, choose:
• Field communicator: Configure→Manual Setup→Measurements→Optional Setup→Conc
Do not change the setting of Derived Variable. If you change the setting of Derived Variable, all existing
concentration matrices will be deleted from transmitter memory.
1. Set Matrix Being Configured to the matrix you want to modify.
2. Choose Reference Conditions, then perform the following actions:
a) Set Reference Temperature to the temperature to which line density will be corrected for use in
the specific gravity calculation.
Configuration and Use Manual83
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
b) Set Water Reference Temperature to the water temperature that will be used in the specific
gravity calculation.
c) Set Water Reference Density to the density of water at the specified reference temperature.
The transmitter automatically calculates the density of water at the specified temperature. The
new value will be displayed the next time that transmitter memory is read. Optionally, you can
enter a different value.
Provide temperature data using a field communicator or an enhanced FF
host
The concentration measurement application uses line temperature data in its calculations. You must decide
how to provide this data, then perform the required configuration and setup. Temperature data from the onboard temperature sensor (RTD) is always available. Optionally, you can set up an external temperature
device and use external temperature data.
The temperature setup that you establish here will be used for all concentration measurement matrices on
this meter.
Procedure
Choose the method to be used to supply temperature data, and perform the required setup.
MethodDescriptionSetup
Internal RTD
temperature data
Digital
communications
Temperature data from the onboard temperature sensor
(RTD) is used.
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal RTD temperature data.
a. Choose Configure→Manual Setup→Measurements→
Optional Setup→External Variables
b. Set Temperature Compensation to Disable.
a. Choose Configure→Manual Setup→Measurements→
Optional Setup→External Variables.
b. Set Temperature Compensation to Enable.
c. Perform the necessary host programming and
communications setup to write temperature data to the
meter at appropriate intervals.
Postrequisites
To verify the External Temperature, choose:
Field communicator
Enhanced FF hostService Tools→Variables→Variable Summary→External Temperature
Service Tools → Variables → Process → External Temperature
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For digital communications:
— Verify that the host has access to the required data.
84Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
— Verify that the output variable is being correctly received and processed by the transmitter.
Modify matrix names and labels using a field communicator or an
enhanced FF host
Field communicatorConfigure→Manual Setup→Measurements→Optional Setup→Conc Measurement→Configure
4. Set Concentration Unit to the label that will be used for the concentration unit.
5. If you set Concentration Unit to Special, choose Label and enter the custom label.
Matrix
Matrix Configuration → Matrix Selection
Modify extrapolation alerts using a field communicator
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control
the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density
or line temperature goes outside the range, the transmitter will extrapolate concentration values. However,
extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is
occurring.
Each concentration matrix has its own extrapolation alert limits.
2. Set Matrix Being Configured to the matrix you want to modify.
3. Set Extrapolation Alert Limit to the point, in percent, at which an extrapolation alert will be posted.
4. Choose Configure→Alert Setup→CM Alerts.
5. Enable or disable the high and low alerts for temperature and density, as desired.
Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a
temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if
line temperature goes above 82 °F (27.8 °C).
Configuration and Use Manual85
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
Modify extrapolation alerts for concentration measurement using an
enhanced FF host
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control
the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density
or line temperature goes outside the range, the transmitter will extrapolate concentration values. However,
extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is
occurring.
Each concentration matrix has its own extrapolation alert limits.
6. Enable or disable the high and low alerts for temperature and density, as desired.
Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a
temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if
line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using a field communicator or an
enhanced FF host
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
86Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
5.2.5 Set up concentration measurement using a basic FF host
This section guides you through most of the tasks related to setting up and implementing the concentration
measurement application.
Restriction
This section does not cover building a concentration matrix. See Micro Motion Enhanced Density Application
Manual for detailed information on building a matrix.
Enable concentration measurement using a basic FF host
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Procedure
1. Set the GSV Volume Flow Type to liquid: write a 0 to the Volume Flow Type parameter on the
Measurement TB.
2. Enable the concentration measurement application: write 1 to the Concentration Measurement
parameter on the Device TB (OD Index 144).
Set reference temperature values for specific gravity using a basic FF host
When Derived Variable is set to any option based on specific gravity, you must set the reference temperature
for water, then verify the density of water at the configured reference temperature. These values affect
specific gravity measurement.
To check the setting of Derived Variable, read the value of the Derived Variable parameter in the
Concentration Measurement TB.
Do not change the setting of Derived Variable. If you change the setting of Derived Variable, all existing
concentration matrices will be deleted from transmitter memory.
Configuration and Use Manual87
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
Procedure
Write the desired values into the appropriate parameters in the Concentration Measurement TB for
Reference Temperature, Water Reference Temperature, and Water Reference Density.
The transmitter automatically calculates the density of water at the specified temperature. The new value will
be displayed the next time that transmitter memory is read. You can enter a different value if you want to.
Modify matrix names and labels using a basic FF host
For convenience, you can change the name of a concentration matrix and the label used for its measurement
unit. This does not affect measurement.
Procedure
1. Choose the matrix you want to modify by writing to the Matrix Being Configured parameter in the
Concentration Measurement TB. Each saved matrix has a unique value of 0 through 5.
2. Write the desired values into the Matrix Name and Concentration Unit parameters in the
Concentration Measurement TB.
Table 5-3: Concentration unit codes
Fieldbus codeUnit
1110degTwad
1426degBrix
1111degBaum hv
1112degBaum It
1343% sol/wt
1344% sol/vol
1427degBall
1428proof/vol
1429proof/mass
33004deg plato
253special
3. Write a value into the Special Concentration Unit Label parameter if Concentration Unit is set to code
253 (special).
Modify extrapolation alerts for concentration measurement using a basic
FF host
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control
the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density
or line temperature goes outside the range, the transmitter will extrapolate concentration values. However,
extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is
occurring.
88Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure process measurement applications
MMI-20029970June 2021
Each concentration matrix has its own extrapolation alert limits.
Procedure
1. Choose the matrix you want to configure using the Matrix Being Configured parameter in the
Concentration Measurement TB. Each saved matrix has a unique value of 0 through 5.
2. Write the desired values into the appropriate parameters in the Concentration Measurement TB.
Parameter nameDescription
Extrapolation LimitExtrapolation Alert Limit The point, in percent, at which an
extrapolation alert will be posted.
Density LowEnable low density extrapolation alarm (write 1 to enable; 0 to
disable).
Density HighEnable high density extrapolation alarm (write 1 to enable; 0 to
disable).
Temperature LowEnable low temperature extrapolation alarm (write 1 to enable; 0
to disable).
Temperature HighEnable high temperature extrapolation alarm (write 1 to enable; 0
to disable).
Extrapolation alert in action
If the following conditions exist, the high temperature extrapolation alert will be posted when the line
temperature exceeds 82 °F (27.8 °C):
• The Extrapolation Alert Limit is set to 5%
• The high temperature alarm is enabled
• The active matrix is built for a temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C)
Select the active concentration matrix using a basic FF host
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
Choose the matrix you want to use by writing to the Matrix Being Configured parameter in the
Concentration Measurement TB. Each saved matrix has a unique value of 0 through 5.
Configuration and Use Manual89
Configure process measurement applicationsConfiguration and Use Manual
June 2021 MMI-20029970
90Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure advanced options for process measurement
MMI-20029970June 2021
6 Configure advanced options for process
measurement
6.1 Configure Response Time
DisplayMenu→Configuration→Process Measurement→Response Time
ProLink IIIDevice Tools→Configuration→Process Measurement→Response Time
Field communicatorNot available
Enhanced FF hostNot available
Basic FF hostNot available
Response Time controls the speed of various internal processes that are involved in retrieving electronic data
from the sensor and converting it to process data.
Response Time affects all process and diagnostic variables.
Restriction
Response Time is configurable only if you are using the enhanced core processor. If you are using the
standard core processor, Response Time is set to Low Filtering and cannot be changed.
Procedure
Set Response Time as desired.
Option
NormalAppropriate for typical applications.
High Filtering Slower response. Appropriate for applications with significant amount of entrained gas or
Low Filtering Fastest response. Appropriate for proving or filling applications.
ServiceDo not select unless directed by Micro Motion personnel.
Description
process noise.
Configuration and Use Manual91
Configure advanced options for process measurementConfiguration and Use Manual
June 2021 MMI-20029970
6.2 Detect and report two-phase flow
Two-phase flow (gas in a liquid process or liquid in a gas process) can cause a variety of process control issues.
The transmitter provides two methods to detect and report or respond to two-phase flow.
6.2.1 Detect two-phase flow using density
Display
ProLink III
Field communicator
Enhanced FF host
Basic FF host
Menu → Configuration → Process Measurement → Density
Device Tools → Configuration → Process Measurement → Density
Measurement TB → Two Phase Flow Setup (OD Index 91–94)
The transmitter can use line density data to detect two-phase flow (gas in a liquid process or liquid in a gas
process). The density limits are user-specified. When two-phase flow is detected, an alert is posted.
Procedure
1. Set Two-Phase Flow Low Limit to the lowest density value that is considered normal in your process.
Values below this will cause the transmitter to post a Process Aberration alert.
Tip
Gas entrainment can cause your process density to drop temporarily. To reduce the occurrence of twophase flow alerts that are not significant to your process, set Two-Phase Flow Low Limit slightly below
your expected lowest process density.
You must enter Two-Phase Flow Low Limit in g/cm³, even if you configured another unit for density
measurement.
• Default: 0 g/cm³
• Range: 0 g/cm³ to the sensor limit
2. Set Two-Phase Flow High Limit to the highest density value that is considered normal in your process.
Values above this will cause the transmitter to post a Process Aberration alert.
Tip
To reduce the occurrence of two-phase flow alerts that are not significant to your process, set TwoPhase Flow High Limit slightly above your expected highest process density.
You must enter Two-Phase Flow High Limit in g/cm³, even if you configured another unit for density
measurement.
• Default: 5 g/cm³
• Range: 5 g/cm³ to the sensor limit
92Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure advanced options for process measurement
MMI-20029970June 2021
3. Set Two-Phase Flow Timeout to the number of seconds that the transmitter will wait for a two-phase
flow condition to clear before posting the alert.
• Default: 0 seconds, meaning that the alert will be posted immediately
• Range: 0 to 60 seconds
6.2.2 Detect two-phase flow using sensor diagnostics
Display
ProLink III
Field communicator
Enhanced FF host
Basic FF host
Menu → Configuration → Inputs/Outputs → Channel B → I/O Settings → Source
Configure → Manual Setup → Inputs/Outputs → Channel B
Device TB → mAO Source Variable (OD Index 94)
The transmitter always monitors sensor diagnostics and applies a two-phase flow algorithm. You can assign
an mA Output to report the results of this calculation: single-phase flow, moderate two-phase flow, or severe
two-phase flow. Severe two-phase flow can cause the meter to stop functioning.
Procedure
Set mA Output Source to Two-Phase Flow Detection.
The signal from the mA Output indicates the current state of the process:
Flow Rate Switch is used to indicate that the flow rate has moved past a user-specified setpoint, in either
direction. The flow rate switch is implemented with a user-configurable hysteresis.
Typically, a Discrete Output is assigned as the flow rate switch indicator. The Discrete Output can be wired to
an external device such as a light or a horn.
Prerequisites
A channel must be configured as a Discrete Output, and the Discrete Output must be available for this use.
Configuration and Use Manual93
Configure advanced options for process measurementConfiguration and Use Manual
June 2021 MMI-20029970
Procedure
1. Set Discrete Output Source to Flow Switch, if you have not already done so.
2. Set Flow Switch Variable to the flow variable that you want to use to control the flow rate switch.
3. Set Flow Switch Setpoint to the value at which the flow switch will be triggered (after Hysteresis is
applied).
Depending on the polarity of the Discrete Output:
• If the flow rate is below this value, the Discrete Output is ON.
• If the flow rate is above this value, the Discrete Output is OFF.
4. Set Hysteresis to the percentage of variation above and below the setpoint that will operate as a
deadband.
Hysteresis defines a range around the setpoint within which the flow rate switch will not change.
• Default: 5%
• Range: 0.1% to 10%
Example
If Flow Switch Setpoint = 100 g/sec and Hysteresis = 5%, and the first measured flow rate is above
100 g/sec, the discrete output is OFF. It will remain OFF unless the flow rate drops below 95 g/sec. If
this happens, the discrete output will turn ON, and remain ON until the flow rate rises above 105 g/sec.
At this point it turns OFF and will remain OFF until the flow rate drops below 95 g/sec.
Related information
Configure a Discrete Output
6.4 Configure events
An event occurs when the real-time value of a user-specified process variable moves past a user-defined
setpoint. Events are used to provide notification of process changes or to perform specific transmitter actions
if a process change occurs.
Basic FF hostDevice TB→Discrete Events (OD Index 153–159)
Menu → Configuration → Alert Setup → Enhanced Events
An enhanced event is used to provide notification of process changes and, optionally, to perform specific
transmitter actions if the event occurs. An enhanced event occurs (is ON) if the real-time value of a user-
94Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure advanced options for process measurement
MMI-20029970June 2021
specified process variable moves above (HI) or below (LO) a user-defined setpoint, or in range (IN) or out of
range (OUT) with respect to two user-defined setpoints.
You can define up to five enhanced events. For each enhanced event, you can assign one or more actions that
the transmitter will perform if the enhanced event occurs.
Procedure
1. Select the event that you want to configure.
2. Assign a process variable to the event.
3. Specify Event Type.
OptionDescription
HIx > A
The event occurs when the value of the assigned process variable (x) is greater
than the setpoint (Setpoint A), endpoint not included.
LOx < A
The event occurs when the value of the assigned process variable (x) is less than
the setpoint (Setpoint A), endpoint not included.
INA ≤ x ≤ B
The event occurs when the value of the assigned process variable (x) is in range,
that is, between Setpoint A and Setpoint B, endpoints included.
OUTx ≤ A or x ≥ B
The event occurs when the value of the assigned process variable (x) is out ofrange, that is, less than Setpoint A or greater than Setpoint B, endpoints
included.
4. Set values for the required setpoints.
• For HI and LO events, set Setpoint A.
• For IN and OUT events, set Setpoint A and Setpoint B.
5. Optional: Configure a Discrete Output to switch states in response to the event status.
6. Optional: Specify the action or actions that the transmitter will perform when the event occurs.
Option
Description
DisplayMenu → Configuration → Alert Setup →
Enhanced Events, select any enhanced event,
and choose Assign Actions
ProLink IIIDevice Tools→Configuration→I/O→Inputs→
Action Assignment
Field communicatorConfigure → Alert Setup → Enhanced Events
Related information
Configure Discrete Output Source
Configuration and Use Manual95
Configure advanced options for process measurementConfiguration and Use Manual
June 2021 MMI-20029970
Options for Enhanced Event Action
Label
Action
DisplayProLink III
Standard
Start sensor zeroStart Zero CalibrationStart Sensor ZeroStart Sensor Zero
Totalizers
Field communicator
or enhanced FF host
Basic FF host code
Start Sensor Zero
Start/stop all
totalizers and
inventories
Reset totalizer XReset Total XTotalizer XReset Total XReset Total X
Reset all totalizers and
inventories
Concentration measurement
Increment CM matrixIncrement MatrixIncrement ED CurveIncrement Curve
Meter verification
Start meter
verification test
Start/stop all
totalizers
Reset All TotalsReset All TotalsReset All TotalsReset All Totals
Menu → Configuration → Process Measurement → Totalizers & Inventories
Totalizers and Inventories TB
The transmitter provides seven configurable totalizers and seven configurable inventories. Each totalizer and
each inventory can be configured independently.
Totalizers track the process since the last totalizer reset. Inventories track the process since the last inventory
reset. Inventories are typically used to track the process across totalizer resets.
Tip
The default configurations cover the most typical uses of totalizers and inventories. You may not need to
change any configurations.
Prerequisites
Before configuring the totalizers and inventories, ensure that the process variables you plan to track are
available on the transmitter.
Procedure
1. Select the totalizer or inventory that you want to configure.
96Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure advanced options for process measurement
MMI-20029970June 2021
2. Set Totalizer Source or Inventory Source to the process variable that the totalizer or inventory will
track.
OptionDescription
Mass flowThe totalizer or inventory will track Mass Flow Rate and calculate total
mass since the last reset.
Volume flowThe totalizer or inventory will track Volume Flow Rate and calculate total
volume since the last reset.
Gas standard volume
flow
Temperature-corrected
volume flow
The totalizer or inventory will track Gas Standard Volume Flow Rate and
calculate total volume since the last reset.
The totalizer or inventory will track Temperature-Corrected Volume Flow
Rate and calculate total volume since the last reset.
Standard volume flowThe totalizer or inventory will track Standard Volume Flow Rate and
calculate total volume since the last reset.
Net mass flowThe totalizer or inventory will track Net Mass Flow Rate and calculate total
mass since the last reset.
Net volume flowThe totalizer or inventory will track Net Volume Flow Rate and calculate
total volume since the last reset.
Note
The totalizer/inventory value will not automatically be reset when the source is changed. The user must
manually reset the totalizer/inventory.
Tip
If you are using the API Referral application and you want to measure batch-weighted average density
or batch-weighted average temperature, you must have a totalizer configured to measure
temperature-corrected volume flow.
3. Set Totalizer Direction to specify how the totalizer or inventory will respond to forward or reverse flow.
Option
Flow directionTotalizer and inventory behavior
Forward OnlyForwardTotals increment
ReverseTotals do not change
Reverse OnlyForwardTotals do not change
ReverseTotals increment
BidirectionalForwardTotals increment
ReverseTotals decrement
Absolute ValueForwardTotals increment
ReverseTotals increment
Important
Actual flow direction interacts with Sensor Flow Direction Arrow to determine the flow direction that
the transmitter uses in processing. See the following table.
Configuration and Use Manual97
Configure advanced options for process measurementConfiguration and Use Manual
June 2021 MMI-20029970
Table 6-1: Interaction between actual flow direction and Sensor Flow Direction Arrow
Actual flow direction
Forward (same direction as Flow
arrow on sensor)
Reverse (opposite from Flow arrow
on sensor)
Setting of Sensor Flow Direction
Arrow
With ArrowForward
Against ArrowReverse
With ArrowReverse
Against ArrowForward
Flow direction sent to outputs
and totalizers
4. Optional: Set User Name to the name you want to use for the inventory or totalizer.
User Name can have a maximum of 16 characters.
The transmitter automatically generates a name for each totalizer and inventory, based on its source,
direction, and type.
Example
• Totalizer Source=Mass Flow
• Totalizer Direction=Forward Only
• Totalizer name=Mass Fwd Total
Example
• Inventory Source=Gas Standard Volume Flow
• Inventory Direction=Bidirectional
• Inventory name = GSV Bidir Inv
The specified name is used on the transmitter display and on all interfaces that support it. If User Name
contains only spaces, the transmitter-generated name is used. Not all interfaces support totalizer and
inventory names.
Example: Checking for backflow
You suspect that there is a significant amount of backflow through the sensor. To collect data, configure two
totalizers as follows:
• Source=Mass Flow, Direction=Forward Only
• Source=Mass Flow, Direction=Reverse Only
Reset both totalizers, allow them to run for an appropriate period, then look at the amount of reverse flow as
a percentage of forward flow.
Example: Tracking three different process fluids
Three tanks are connected to a loading dock through a single meter. Each tank contains a different process
fluid. You want to track each process fluid separately.
1. Set up three totalizers, one for each tank.
2. Name the totalizers Tank 1, Tank 2, and Tank 3.
98Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Configuration and Use ManualConfigure advanced options for process measurement
MMI-20029970June 2021
3. Configure each totalizer as required for the corresponding process fluid.
4. Stop and reset all three totalizers to ensure that the beginning values are 0.
5. When loading from a tank, start the corresponding totalizer, and stop it when the load is finished.
6.5.1 Default settings for totalizers and inventories
Totalizer or
inventory
1Mass flowForward OnlyMass Fwd Total
2Volume flowForward OnlyVolume Fwd Total
3Temperature-corrected volume
4Gas standard volume flowForward OnlyGSV Fwd Total
5Standard volume flowForward OnlyStandard Vol Fwd Total
6Net mass flowForward OnlyNet Mass Fwd Total
7Net volume flowForward OnlyNet Vol Fwd Total
Source (process variable
assignment
flow
DirectionName of totalizer
Name of inventory
Mass Fwd Inv
Volume Fwd Inv
Forward OnlyAPI Volume Fwd Total
API Volume Fwd Inv
GSV Fwd Inv
Standard Vol Fwd Inv
Net Mass Fwd Inv
Net Vol Fwd Inv
6.6 Configure logging for totalizers and inventories
The transmitter can write the current value of four totalizers or inventories to a log, at user-specified intervals.
You can generate a log file from this data for viewing and analysis.
Procedure
1. Specify the date on which totalizer logging will begin.
You must specify a future date. If you try to specify the current date, the transmitter will reject the
setting.
2. Specify the time at which totalizer logging will begin.
3. Specify the number of hours between records.
4. Select up to four totalizers or inventories to be logged.
Configuration and Use Manual99
Configure advanced options for process measurementConfiguration and Use Manual
Process Variable Fault Action specifies the values that will be reported via the display and digital
communications if the device encounters a fault condition. The values are also sent to the outputs for
processing against their configured fault actions.
Procedure
Set Process Variable Fault Action as desired.
Default: None
Restriction
If you set Process Variable Fault Action to NAN, you cannot set mA Output Fault Action or Frequency
Output Fault Action to None. If you try to do this, the transmitter will not accept the configuration.
Important
• If you want the mA Output to continue reporting process data during fault conditions, you must set both
Process Variable Fault Action and mA Output Fault Action to None. If mA Output Fault Action is set to
None and Process Variable Fault Action is set to any other option, the mA Output will produce the signal
associated with the selection.
• If you want the Frequency Output to continue reporting process data during fault conditions, you must set
both Process Variable Fault Action and Frequency Output Fault Action to None. If Frequency OutputFault Action is set to None and Process Variable Fault Action is set to any other option, the Frequency
Output will produce the signal associated with the selection.
6.7.1 Options for Process Variable Fault Action
Label
DisplayProLink III
UpscaleUpscaleUpscaleUpscale
DownscaleDownscaleDownscaleDownscale
Field
communicator
Fieldbus host
Description
• Process variable values indicate
that the value is greater than the
upper sensor limit.
• Totalizers stop incrementing.
• Process variable values indicate
that the value is lower than the
lower sensor limit.
• Totalizers stop incrementing.
100Micro Motion 5700 Transmitters with FOUNDATION Fieldbus
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.