Micro Motion Modbus Interface Tool - SmartMeter Verification Manuals & Guides

Operation and Use Guide
MMI-20017073, Rev AD
Micro Motion® Modbus Interface Tool
Smart Meter Verification
October 2019
Safety and approval information
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at
www.emerson.com.
For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion Customer Service department.
Emerson Flow customer service
Email:
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America
United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 704 136
Mexico +41 (0) 41 7686
111
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy 8008 77334 China +86 21 2892 9000
Europe and Middle East Asia Pacific
666
France 0800 917 901 India 800 440 1468
Central & Eastern +41 (0) 41 7686
111
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684
New Zealand 099 128 804
Japan +81 3 5769 6803
2
Operation and Use Guide Contents
MMI-20017073 October 2019

Contents

Chapter 1 Smart Meter Verification................................................................................................5
Chapter 2 Execute an SMV test from the Modbus host....................................................................7
Chapter 3 Configure and view scheduled test executions............................................................... 9
Chapter 4 Read stored test results................................................................................................11
Operation and Use Guide iii
Contents Operation and Use Guide
October 2019 MMI-20017073
iv Micro Motion Modbus Interface Tool for SMV
Operation and Use Guide Smart Meter Verification
MMI-20017073 October 2019

1 Smart Meter Verification

Smart Meter Verification (SMV) is a Micro Motion diagnostic tool that allows you to compare the sensor’s current performance to benchmark data established at the factory.
Using the Modbus interface, you can:
Execute an SMV test from the Modbus host
Set up a scheduled test execution
Read result data for the twenty test results stored on the transmitter
Note
You can view more than twenty previous test results and see detailed test reports only if SMV is licensed.
For more information on requirements and features, see your transmitter configuration and use manual.
Operation and Use Guide 5
Smart Meter Verification Operation and Use Guide
October 2019 MMI-20017073
6 Micro Motion Modbus Interface Tool for SMV

Operation and Use Guide Execute an SMV test from the Modbus host

MMI-20017073 October 2019
2 Execute an SMV test from the
Modbus host
Prerequisites
The Modbus host must be connected to the transmitter.
Procedure
1. If you are fixing outputs during the test, set the output state from Register 3093.
Option Description
0 Last Measured Value
1 Fault Action
2. From Register 3000, start or abort the test. 0 = Abort
1 = Start with outputs fixed (apply Register 3093) 6 = Start in Continue Measurement Mode
3. Read Register 3001. Is the test running?
Option
> 0 = Yes Read the percent complete from Register 3020.
0 = No Continue to the next step.
4. Read Register 3003. Did the test run to completion?
Option
16 = Yes Continue to the next step.
< 16 = No Read the abort code on Register 3002.
Description
Range: 0 to 100
Description
1 = User Initiated Abort 3 = Frequency Drift 5 = High Drive Gain 8 = Unstable Flow 12 = Fault Condition 13 = No Air Reference 14 = No Water Reference 15 = Missing Configuration Data
Operation and Use Guide 7
Execute an SMV test from the Modbus host Operation and Use Guide
October 2019 MMI-20017073
5. Read Register 3004. Is the test passed?
0 = Yes The test passed.
> 0 = No The test result is caution. Continue to the next step.
6. For Model 5700 transmitters, read Register 6348. For all other transmitters, read Register 6364.
128 Possible coating or corrosion/erosion; coating detection is available only on
the Model 5700
256 Possible extreme temperature or corrosion
512 Possible extreme temperature or damage
1024 Possible coating or damage; coating detection is available only on the Model
5700
2048 Possible corrosion/erosion
4096 Possible coating; coating detection is available only on the Model 5700
8 Micro Motion Modbus Interface Tool for SMV

Operation and Use Guide Configure and view scheduled test executions

MMI-20017073 October 2019
3 Configure and view scheduled test
executions
Use the Modbus host to set up the schedule.
Prerequisites
To configure a test execution and view a scheduled test execution, the Modbus host must be connected to the transmitter. Once the schedule is loaded into the transmitter, the Modbus host no longer has to be connected to the transmitter.
Procedure
1. Choose any of the following options:
Option Description
Specify the hours until test execution for a single execution
Specify the hours for a recurring execution
View the number of hours until the next execution
Cancel a scheduled execution
2. Disconnect as desired.
Write a floating-point value to Register 2993.
To specify the number of hours until the first
test is executed, write a floating-point value to Register 2993.
To specify the number of hours between test
executions, write a floating-point value to Register 2995.
Read Register 2997.
Write 0 to Register 2993.
Write 0 to Register 2995.
Operation and Use Guide 9
Configure and view scheduled test executions Operation and Use Guide
October 2019 MMI-20017073
10 Micro Motion Modbus Interface Tool for SMV
Operation and Use Guide Read stored test results
MMI-20017073 October 2019

4 Read stored test results

The results of the twenty most recent tests are stored on the transmitter.
Note
You can view more than twenty previous test results and see detailed test reports only if SMV is licensed.
Prerequisites
To read test results, the Modbus host must be connected to the transmitter.
Procedure
1. Specify the test record you want to read by writing a value between 0 and 19 to the appropriate register.
Register for Model 5700 transmitters
5779 2984 For Model 5700 transmitters:
Register for all other transmitters
Description
Register 5825 = the most recent test For all other meters: 0 = the most recent test
2. Read the Modbus register values using the descriptions in the following table.
Registers for Model 5700 transmitters
5697 2985 Test number
5820 and 5821 2986 Bit #7 Test result
5819 2986 Bits #6–4 State
5818 2986 Bits #3–0 Abort code (compressed)
5780 2987 Time initiated
5782 and 5783 2989 Inlet normalized data
5784 and 5785 2991 Outlet normalized data
Registers for all other transmitters
Description
0 = Pass 1 = Caution
3. Disconnect as desired.
Operation and Use Guide 11
*MMI-20017073*
MMI-20017073
Rev. AD
2019
Micro Motion Inc. USA
Worldwide Headquarters 7070 Winchester Circle Boulder, Colorado USA 80301 T +1 303-527-5200 T +1 800-522-6277 F +1 303-530-8459
www.emerson.com
Micro Motion Asia
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Republic of Singapore T +65 6363-7766 F +65 6770-8003
©
2019 Micro Motion, Inc. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Micro Motion, ELITE, ProLink, MVD and MVD Direct Connect marks are marks of one of the Emerson Automation Solutions family of companies. All other marks are property of their respective owners.
Micro Motion Europe
Emerson Automation Solutions Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 T +31 (0) 70 413 6666 F +31 (0) 318 495 556
www.emerson.com/nl-nl
Micro Motion United Kingdom
Emerson Automation Solutions Emerson Process Management Limited Horsfield Way Bredbury Industrial Estate Stockport SK6 2SU U.K. T +44 0870 240 1978 F +44 0800 966 181
Loading...