Micro Motion Model 5700 Transmitter Net Oil
Calculations
Multiwell Supplement
Safety messages
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully
before proceeding to the next step.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration
manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government
transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept
your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion
Customer Service department.
Emerson Flow customer service
Email:
• Worldwide: flow.support@emerson.com
• Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South AmericaEurope and Middle EastAsia Pacific
United States800-522-6277U.K.0870 240 1978Australia800 158 727
Appendix ADifferences between Series 3000 and Model 5700 NOC Modbus registers.....................39
4Micro Motion® Model 5700 Transmitter
Multiwell Supplement Before you begin
MMI-20052554March 2019
1 Before you begin
1.1 About this manual
This manual explains how to understand, plan, configure, and use the Net Oil Calculations (NOC) application
on a Model 5700 transmitter.
The information in this document assumes that users understand all corporate, local government, and
national government safety standards and requirements that guard against injuries and death.
1.2 Related documentation
You can find all product documentation on the product documentation DVD shipped with the product or at
www.emerson.com.
See any of the following documents for more information:
• Micro Motion Model 5700 Transmitters with Configurable Outputs: Configuration and Use Manual
•
Micro Motion Model 5700 Transmitters with Configurable Outputs: Installation Manual
• Modbus Interface Tool
• Micro Motion ProLink III User Manual
• Sensor installation manual
1.3 Overview of Model 5700 NOC tasks
Step
1Install the Model 5700✓
2Set up digital commmunications✓
3Start up the system✓
3Configure security and language✓
4Configure system data✓
5Configure inputs✓
6Configure digital communications✓
Task
Model 5700
NOC (multiwell)
supplement
(this manual)
Model 5700
installation
manual
Model 5700
configuration
and use
manual
7Configure the NOC application✓
8Configure outputs✓✓
9Configure NOC status alert severity✓✓
10Perform optional configuration✓
11Run an NOC well test✓
Multiwell Supplement 5
Before you beginMultiwell Supplement
March 2019 MMI-20052554
StepTask
12Perform NOC continuous measurement✓
13Perform calibrations and meter verification✓
14Troubleshoot alerts✓✓
Model 5700
NOC (multiwell)
supplement
(this manual)
Model 5700
installation
manual
Model 5700
configuration
and use
manual
6Micro Motion® Model 5700 Transmitter
Multiwell Supplement Planning
MMI-20052554March 2019
2 Planning
2.1 NOC overview
NOC is a software option that customers can purchase for the Model 5700 transmitter.
The NOC application can provide real-time measurements of water cut when paired with a Micro Motion
sensor installed on either of the following options:
• The oil or water leg of a two-phase separator
• The oil leg of a three-phase separator
Alternatively, the NOC application can receive water cut data from an external water cut monitor. When the
water cut value is known, net oil volume flow and net water volume flow can be calculated.
Net oil volume flow and net water volume flow can also be calculated by the Advanced Phase Measurement
software from current density data.
All volume rates and totals are at line conditions, uncorrected, unless stated otherwise.
2.2 NOC terminology
TermDefinition
ActualThe flow rate as measured at the time of viewing.
AverageThe flow-weighted average, calculated from the beginning of the applicable
time period.
Back flowUncorrected volume flow that is moving backward through the sensor.
Density or density @ lineThe density of the mixture, with no corrections applied.
Gross or gross @ refThe sum of the oil volume and the water volume, as measured by the NOC
application.
Net oil or net oil @ refOil that is measured by volume, corrected to reference temperature, with the
oil shrinkage factor applied.
Net water or net water @ refWater that is corrected to reference temperature, with the water shrinkage
factor applied.
PVProcess Variable
TotalThe rolling total, calculated from the beginning of the applicable time period.
Uncorrected gross or gross @ lineThe raw volume flow measurement from the oil or oil and water leg.
Uncorrected oil or net oil @ lineOil moving through the oil or oil and water leg that is measured by volume,
with no corrections applied. Temperature correction and shrinkage factors are
not applied.
Uncorrected water cut or water cut @
line
Uncorrected water or net water @ lineWater moving through the oil or oil and water leg, measured by volume, with
Multiwell Supplement 7
The percentage of water in a production stream that is at operating conditions.
no corrections applied. Temperature correction and shrinkage factors are not
applied.
Planning
March 2019 MMI-20052554
TermDefinition
Water cut or water cut @ refThe percentage of water in a production stream, corrected to reference
temperature, with the water shrinkage factor applied.
Multiwell Supplement
2.3 NOC system components
The Model 5700 NOC system requires:
• One Model 5700 transmitter with the NOC software option
• One Micro Motion sensor installed on the oil and water, or oil leg
• (Optional) A water cut monitor on the oil and water, or oil leg
2.4 Installation architecture
You can install the Model 5700 NOC system with a two-phase separator, or a three-phase separator.
Note
The following examples do not illustrate all possible combinations.
NOC system with a two-phase separator
With a two-phase separator, the NOC system uses density-based water cut data.
A. Model 5700 transmitter
B. Separator
C. Oil/water leg
D. Sensor
NOC system with a three-phase separator
With a three-phase separator:
• A meter is installed on the water leg, but the Model 5700 transmitter does not monitor or record flow data
from this source.
• A water cut monitor is installed on the oil leg. The NOC system can be configured to use either density-
based water cut data, or data from the water cut monitor.
8Micro Motion® Model 5700 Transmitter
Multiwell Supplement Planning
MMI-20052554March 2019
A. Model 5700 transmitter
B. Separator
C. Oil leg
D. Sensor
E. (Optional) Water cut monitor
F. Water leg
G. Flowmeter for water
2.5 Operation modes
The NOC system operates in either well test mode or continuous mode.
You can change modes after initial configuration. However, changing modes affects current measurement
and data collection.
Well test mode
Well tests can be performed on up to 48 wells. A manifold system is used to ensure that output from a single
well is routed through the test separator and the NOC system. The system can save data for a total of three
well tests. If more than three tests are run, older tests are overwritten as required.
Multiwell Supplement 9
PlanningMultiwell Supplement
March 2019 MMI-20052554
A. Model 5700 transmitter
B. Manifold
C. One well to test separator
D. Separator
E. Liquid leg
F. Sensor
G. To production separator
H. Other wells to production separator
Continuous mode
One well is measured continuously. The NOC system supplies current flow data, plus running averages and
totals.
A. Model 5700 transmitter
B. From a single well
C. Separator
D. Liquid leg
E. Sensor
F. To production separator
10Micro Motion® Model 5700 Transmitter
Multiwell Supplement Planning
MMI-20052554March 2019
2.6 Required well data
The following information is required for each well tested or measured by the NOC system:
• Density of dry oil from this well at reference temperature and reference pressure. To ensure the most
accurate net oil data, the density should be based on live oil rather than dead oil. “Live oil” refers to crude
oil at line pressure.
• Density of the water from this well, at reference temperature and reference pressure.
If you do not know the density values, do any of the following tasks:
• Take samples of produced oil and produced water, perform laboratory analysis, and enter the results into
the well configuration.
• Perform an in-line density determination for oil, water, or both. During in-line density determination, the
appropriate process fluid (water or live oil) is routed through the sensor, density values are averaged over a
user-specified time period, and these values are converted to reference temperature. The water cut
calculation uses these average values for DO and DW.
• Enter approximate values at initial configuration, begin measurement, and recalculate NOC data at a later
time when well-specific density values are known.
2.7 Water cut determination
There are two options for determining water cut, density-based and water cut monitoring.
Density-based
The NOC application derives the water cut by applying the following equation:
D
− D
Watercut=
D
Mix
D
O
D
W
Water cut monitoring (external water cut)
A water cut monitor is used to measure the process stream directly. The Model 5700 transmitter retrieves the
water cut data via a HART connection. A HART connection between the primary mA output on the NOC
platform and the water cut monitor is required.
The Model 5700 can also use external water cut using an mA Input.
Density of the oil, water, and gas mixture
Density of produced oil (user-supplied value)
Density of produced water (user-supplied value)
Mix
DW− D
O
O
Multiwell Supplement 11
PlanningMultiwell Supplement
March 2019 MMI-20052554
2.8 NOC features and options
This section describes several features and options of the NOC application.
2.8.1 Temperature correction
Temperature correction refers to the conversion of data collected at the observed process temperature to the
equivalent values at reference temperature. The NOC application automatically applies temperature
correction to NOC data, using the temperature data from the RTD built into the sensor.
2.8.2 Shrinkage factors
Shrinkage is a reduction in mass and volume of crude oil due to the vaporization and evaporation of the
volatile components in the oil. By estimating the shrinkage during oil storage or transport, you can estimate
sellable oil based on upstream volume measurement.
Note
The contraction of the volume due to cooling is not a mass loss. It is the thermal contraction of the oil
accounted for by “temperature correction” via the Volume Correction Factors in API Ch 11.1.
The NOC application includes one shrinkage factor for oil. The net oil flow rate measured by the NOC
application is automatically multiplied by the corresponding shrinkage factor and output on a separate PV. By
default, the shrinkage factors are set to 1.0, resulting in no compensation for shrinkage.
Use your standard methods to determine the appropriate shrinkage factors, taking into consideration the
location of the sensor in your process.
2.8.3 Liquid with gas
This measurement option improves flow measurement in liquid processes with intermittent entrained gas.
Note
The liquid with gas measurement option can also be combined with the net oil measurement option or
concentration measurement. See the Micro Motion Enhanced Density Application Manual to configure
concentration measurement.
Liquid with gas measurement process
The presence of entrained gas (or bubbles) can cause significant errors when measuring the volume flow of
liquid through a Coriolis meter. Because bubbles displace some of the liquid in a flow stream, the measured
volume of the mixture may differ from the actual amount of liquid that emerges from the pipe downstream.
So how can you tell when a liquid contains gas? When bubbles are present in a liquid stream, Coriolis meters
will report an increase in drive gain coinciding with a decrease in both fluid density
due to the lower amount of mass contained in the liquid-gas mixture. Therefore, in order to measure only the
liquid portion of the stream, the volume of the bubbles must be ignored or subtracted from the mixture total.
(1)
and mass flow rate
(2)
(1) High frequency sensors may erroneously report a higher fluid density when entrained gas is present, and therefore are not
recommended for use on liquids with entrained gas. High frequency sensors include the F300/H300 compact, and all TSeries sensors.
(2) The accuracy and repeatability of the mass flow and density measurements for liquids with entrained gas is dependent on
the sensor-fluid decoupling ratio, which is a complex function of fluid velocity, fluid viscosity, fluid density, the difference
between the liquid and gas densities, the operating sensor frequency, and the Gas Volume Fraction (GVF) of gas. For best
measurement performance, GVF should be kept below 15%.
12Micro Motion® Model 5700 Transmitter
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.