Manual Supplement
P/N MMI-20007746, Rev. BA
February 2011
Micro Motion
®
Model 2400S Transmitters
for DeviceNet
™
Device Profile
©2011, Micro Motion, Inc. All rights reserved. ELITE and ProLink are registered trademarks, and MVD and MVD Direct Connect
are trademarks of Micro Motion, Inc., Boulder, Colorado. Micro Motion is a registered trade name of Micro Motion, Inc., Boulder,
Colorado. The Micro Motion and Emerson logos are trademarks and service marks of Emerson Electric Co. All other trademarks
are property of their respective owners.
Contents
Chapter 1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 About device profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2.1 Object models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2.2 I/O data format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2.3 Configuration data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Customer service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Chapter 2 Object Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Object classes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.3 Object diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Chapter 3 I/O Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.2 Summary of input assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.3 Summary of output assemblies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.4 Input assembly details. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4.1 Detail of Input Assembly 1 – Status and mass flow . . . . . . . . . . . . . . . . . . 9
3.4.2 Detail of Input Assembly 2 – Status and volume flow . . . . . . . . . . . . . . . . 9
3.4.3 Detail of Input Assembly 3 – Status, mass flow, and mass total . . . . . . . 10
3.4.4 Detail of Input Assembly 4 – Status, volume flow, and volume total . . . . 10
3.4.5 Detail of Input Assembly 5 – Status and basic process variables . . . . . . 10
3.4.6 Detail of Input Assembly 6 – Status, mass flow, mass totals, and
other process variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.4.7 Detail of Input Assembly 7 – Status, volume flow, volume totals, and
other process variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.4.8 Detail of Input Assembly 8 – Status, gas standard volume process
variables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.4.9 Detail of Input Assembly 9 – Status, gas standard volume process
variables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.10 Detail of Input Assembly 10 – Status, gas standard volume process
variables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.11 Detail of Input Assembly 11 – Status, gas standard volume process
variables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.12 Detail of Input Assembly 12 – Status, gas standard volume process
variables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4.13 Detail of Input Assembly 13 – Status, petroleum measurement
process variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4.14 Detail of Input Assembly 14 – Status, petroleum measurement
process variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4.15 Detail of Input Assembly 15 – Status, petroleum measurement
process variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.4.16 Detail of Input Assembly 16 – Status, petroleum measurement
process variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Device Profile i
Contents
3.4.17 Detail of Input Assembly 17 – Concentration measurement
densities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.4.18 Detail of Input Assembly 18 – Concentration measurement
concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.4.19 Detail of Input Assembly 19 – Concentration measurement Baume . . . . 15
3.4.20 Detail of Input Assembly 20 – Concentration measurement net mass
flow and totals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.4.21 Detail of Input Assembly 21 – Concentration measurement net
volume flow and totals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.22 Detail of Input Assembly 22 – Concentration measurement reference
density and net mass flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.23 Detail of Input Assembly 23 – Concentration measurement reference
density and net volume flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.24 Detail of Input Assembly 24 – Concentration measurement reference
density and standard volume flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4.25 Detail of Input Assembly 25 – Status, concentration measurement
reference density and concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4.26 Detail of Input Assembly 26 – Status, user-selected process
variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.5 Output assembly details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5.1 Detail of Output Assembly 0 – No output data. . . . . . . . . . . . . . . . . . . . . 18
3.5.2 Detail of Output Assembly 50 – External pressure . . . . . . . . . . . . . . . . . 18
3.5.3 Detail of Output Assembly 51 – External temperature. . . . . . . . . . . . . . . 18
3.5.4 Detail of Output Assembly 52 – External pressure and temperature. . . . 18
3.5.5 Detail of Output Assembly 53 – Start/stop all totalizers and
inventories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5.6 Detail of Output Assembly 54 – Reset all totalizers . . . . . . . . . . . . . . . . . 19
3.5.7 Detail of Output Assembly 55 – Reset all inventories . . . . . . . . . . . . . . . 19
3.5.8 Detail of Output Assembly 56 – Start/stop all totalizers and
inventories and reset all totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.5.9 Detail of Output Assembly 57 – Start/stop all totalizers and
inventories, reset all totalizers, and reset all inventories . . . . . . . . . . . . . 19
3.5.10 Detail of Output Assembly 58 – Start meter verification . . . . . . . . . . . . . 20
3.5.11 Detail of Output Assembly 59 – External pressure, start meter
verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.12 Detail of Output Assembly 60 – External temperature and start
meter verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.13 Detail of Output Assembly 61 – Start/stop all totalizers and
inventories, start meter verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.14 Detail of Output Assembly 62 – Reset all totalizers and start meter
verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.5.15 Detail of Output Assembly 63 – Reset all inventories and start meter
verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.5.16 Detail of Output Assembly 64 – Start/stop all totalizers, reset all
totalizers, start meter verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.5.17 Detail of Output Assembly 65 – Start/stop all totalizers, reset all
totalizers, reset all inventories, start meter verification . . . . . . . . . . . . . . 21
Chapter 4 Configuration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2 Object summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.3 How to use the object tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
ii Micro Motion ® Model 2400S Transmitters for DeviceNet
™
Contents
4.4 Identity Object (0x01) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.4.1 Identity Object attributes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.4.2 Identity Object services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.4.3 Reset service details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.5 Message Router Object (0x02) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.5.1 Message Router attributes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.5.2 Message Router services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.6 DeviceNet Object (0x03) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.6.1 DeviceNet Object attributes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.6.2 DeviceNet Object services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4.6.3 Service details: Allocate Master/Slave Connection service . . . . . . . . . . . 31
4.6.4 Service details: Release Master/Slave Connection Service . . . . . . . . . . 32
4.6.5 Rules governing Master/Slave connections . . . . . . . . . . . . . . . . . . . . . . . 32
4.7 Assembly Object (0x04) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.7.1 Assembly Object attributes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.7.2 Assembly Object services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.8 Connection Object (0x05) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.8.1 Connection Object attributes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.8.2 Connection Object services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.9 Acknowledge Handler Object (0x2B) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.9.1 Acknowledge Handler Object attributes . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.9.2 Acknowledge Handler Object services . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.10 Analog Input Point Object (0x0A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.10.1 Instances and attributes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.10.2 Analog Input Point Object Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
4.11 Gas Standard Volume Object (0x64). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4.11.1 Instances and attributes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4.11.2 Gas Standard Volume Object services . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.12 Calibration Object (0x65) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
4.12.1 Instances and attributes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
4.12.2 Calibration Object services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
4.12.3 Zero service details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.13 Diagnostics Object (0x66) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.13.1 Instances and attributes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.13.2 Diagnostics Object services. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
4.13.3 Service Details: Meter Verification service. . . . . . . . . . . . . . . . . . . . . . . . 65
4.14 Sensor Information Object (0x67) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
4.14.1 Instances and attributes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
4.14.2 Sensor Information Object services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
4.15 Local Display Object (0x68) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
4.15.1 Instances and attributes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
4.15.2 Local Display Object Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
4.16 Petroleum Measurement Object (0x69) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
4.16.1 Instances and attributes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
4.16.2 Petroleum Measurement Object services . . . . . . . . . . . . . . . . . . . . . . . . 71
4.17 Concentration Measurement Object (0x6A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.17.1 Instances and attributes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.17.2 Concentration Measurement Object services . . . . . . . . . . . . . . . . . . . . . 75
4.18 Get and Set services. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.18.1 Get service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.18.2 Set service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.19 DeviceNet data types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
4.20 Measurement unit codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
4.21 Process variable codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
4.22 Alarm index codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Device Profile iii
iv Micro Motion ® Model 2400S Transmitters for DeviceNet
™
Chapter 1
Introduction
1.1 About this manual
This manual describes the device profile for the Micro Motion
DeviceNet
™
.
®
Model 2400S transmitter for
1.2 About device profiles
A device profile must contain:
• An object model for the device type
• The I/O data format for the device type
• Configuration data and the public interface(s) to that data
1.2.1 Object models
This section of the device profile defines the grouping of objects used in a device. Each device has a
single object model.
An object model specification:
• Identifies all object classes present in the device (required and optional).
• Indicates the number of instances present in each object class. If the device supports the
dynamic creation and deletion of instances, then the object model states the maximum number
of instances that can exist within the object class.
• States whether or not the object affects behavior of the device. If it does affect behavior, the
object model states how.
• Defines the interface to each object. This defines how objects and object classes are linked.
The object model for the Model 2400S DN transmitter is described in detail in Chapter 2.
Object Model Configuration Data I/O Data FormatIntroduction
1.2.2 I/O data format
This section of the device profile defines how a device communicates on the DeviceNet network. This
includes an exact specification of the device's I/O data format.
Communicating multiple pieces of data (attributes) across a single I/O connection requires that the
attributes be grouped or assembled together into a single block. Instances of the Assembly Object
Class perform this grouping. Thus, the definition of a device's I/O data format is equivalent to the
definition of the assembly object instances used to group the device's I/O data.
Device Profile 1
Introduction
Assemblies are grouped into two categories: Input and Output. From a device perspective:
The definition of a device's I/O data format:
The I/O data format for the Model 2400S DN transmitter is described in detail in Chapter 3.
• An Input Assembly is the data PRODUCED from the device ONTO the network (or can also
be viewed as data INPUT to the Master FROM the device).
• An Output Assembly is the data CONSUMED from the network INTO the device (or can also
be viewed as the data OUTPUT from the Master TO the device).
• Identifies the I/O assembly by instance number, type, and name. For a summary of input
assemblies, see Table 3-1. For a summary of output assemblies, see Table 3-2.
• Specifies the I/O assembly data attribute format. Refer to the assembly detail tables in
Sections 3.4 and 3.5.
• Maps the I/O assembly data attribute components to other attributes. Refer to the assembly
detail tables in Sections 3.4 and 3.5. In these tables, EPATHs are denoted as follows:
[Class_ID -Instance_ID -Attribute_ID ]
For example, the EPATH [0x0A-1-3] specifies the location of the mass flow process variable.
1.2.3 Configuration data
A device profile includes specification of the device's configurable parameters and the public interface
to those parameters.
Definition of a device's configuration includes the following information for each configurable
feature:
• Configuration parameter data, also known as attributes
• Description of the services available for use with the attribute
• Public interface to the device's configuration (e.g., bulk configuration via a configuration
assembly, full/stub instances of the Parameter Object Class, etc.)
• The configuration data for the Model 2400S DN transmitter is described in detail in Chapter 4.
1.3 Customer service
For customer service, phone the support center nearest you:
• In the U.S.A., phone
• In Canada and Latin America, phone +1 303-527-5200
•I n A s i a :
- In Japan, phone 3 5769-6803
- In other locations, phone +65 6777-8211 (Singapore)
800-522-MASS (800-522-6277) (toll-free)
•I n E u r o p e :
- In the U.K., phone 0870 240 1978 (toll-free)
- In other locations, phone +31 (0) 318 495 555 (The Netherlands)
Customers outside the U.S.A. can also email Micro Motion customer service at
FlowSupport@Emerson.com .
2 Micro Motion® Model 2400S Transmitters for DeviceNet
™
Chapter 2
Object Model
2.1 About this chapter
This chapter describes the object model of the Model 2400S DN transmitter. The following
information is provided:
• Object classes – see Section 2.2
• Object diagram – see Section 2.3
2.2 Object classes
Table 2-1 lists and describes all object classes present in the Model 2400S DN transmitter.
Table 2-1 Object classes and descriptions
Optional/
Object Class ID
Identity 0x01 Required 1 Supports the Reset service
Message Router 0x02 Required 1 No effect
DeviceNet 0x03 Required 1 Configures port attributes (baud rate, MAC ID, BOI
Connection 0x05 Required 2 One instance for I/O and one for explicit messaging.
Assembly 0x04 Required 33 Defines I/O data format
Acknowledge
Handler
Analog Input Point 0x0A Optional 4 Defines AI values: mass flow, volume flow, density, and
Gas Standard
Volume
Calibration 0x65 Optional 1 Affects response of process variable values
Diagnostics 0x66 Optional 1 No effect
Sensor Information 0x67 Optional 1 No effect
Local Display 0x68 Optional 1 Affects behavior of display
Petroleum
Measurement
Concentration
Measurement
0x2B Required 1 Affects retries and timeouts for acknowledged Cyclic
0x64 Optional 1 Affects response of gas standard volume calculation
0x69 Optional 1 Affects calculations for the petroleum measurement
0x6A Optional 1 Affects calculations for the concentration measurement
required
# of
instances Effect on behavior of device
processing)
Contains the number of logical ports into or out of the
device
I/O connection types
temperature. Affects response of these values
application
application
Object Model Configuration Data I/O Data Format Introduction
Device Profile 3
Object Model
Explicit
Msg
I/O
Connection
Object Class
DeviceNet Network
DeviceNet
Obje ct
Message
Router
Identity
Obje ct
Assembly Object
Class
Diagnostics
Obje ct
AI Point
Obje ct
Application Objects
Instance
#n
Calibrat ion
Obje ct
Local
Displ ay
Objec t
Sensor
Info
Object
Petro -
leum
Msmt
Obje ct
Conce n-
tration
Msmt.
Object
Gas
Standard
Volume
Object
2.3 Object diagram
The object diagram for the Model 2400S DN transmitter is shown in Figure 2-1.
Figure 2-1 Object diagram
4 Micro Motion® Model 2400S Transmitters for DeviceNet
™
Chapter 3
I/O Data Format
3.1 About this chapter
This chapter describes the I/O data format of the Model 2400S DN transmitter. The following
information is provided:
• Summary of the input assemblies and output assemblies – see Section 3.2
• Summary of the output assemblies – see Section 3.3
• Detailed descriptions of the input assemblies – see Section 3.4
• Detailed descriptions of the output assemblies – see Section 3.5
3.2 Summary of input assemblies
The Model 2400S DN transmitter input assemblies are summarized in Table 3-1. In this table:
• The Status field is a single byte, reflecting the status of the data from the referenced object
(e.g., the Analog Input Point Object, Instance 2). Status values are as follows:
-0 = n o f a u l t
-1 = f a u l t
• The listed data type describes the data type of all other fields in the input assembly.
Table 3-1 Summary of input assemblies
Instance ID Data description Size (bytes) Data type Notes
1 Status
Mass flow
2 Status
Volume flow
3 Status
Mass flow
Mass total
4 Status
Volume flow
Volume total
5 Status
Mass flow
Temperature
Density
Volume flow
Drive gain
5 BOOL
REAL
5 BOOL
REAL
9 BOOL
REAL
REAL
9 BOOL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
Mass flow
Volume flow
Mass flow and total
Volume flow and
total
Basic process
variables
Object Model Configuration Data I/O Data FormatIntroduction
Device Profile 5
I/O Data Format
Table 3-1 Summary of input assemblies continued
Instance ID Data description Size (bytes) Data type Notes
6 Status
Mass flow
Mass total
Mass inventory
Temperature
Density
7 Status
Volume flow
Volume total
Volume inventory
Temperature
Density
8 Status
Mass flow
Mass total
Temperature
Gas standard volume flow
Gas standard volume total
9 Status
Mass flow
Temperature
Gas standard volume flow
Gas standard volume total
Gas standard volume inventory
10 Status
Temperature
Drive gain
Gas standard volume flow
Gas standard volume total
Gas standard volume inventory
11 Status
Gas standard volume flow
12 Status
Gas standard volume flow
Gas standard volume total
Gas standard volume inventory
13 Status
Volume flow
Volume total
Volume inventory
Temperature-corrected volume flow
Temperature-corrected volume total
14 Status
Volume flow
Volume total
Temperature-corrected density
Temperature-corrected volume flow
Temperature-corrected volume inventory
15 Status
Mass flow
Mass total
Volume flow
Volume total
Temperature-corrected density
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
5 BOOL
REAL
13 BOOL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
Mass flow, mass
totals and other
process variables
Volume flow, volume
totals and other
process variables
Gas standard
volume flow
Gas standard
volume flow
Gas standard
volume flow
Gas standard
volume flow
Gas standard
volume flow
Petroleum
measurement
application
Petroleum
measurement
application
Petroleum
measurement
application
6 Micro Motion ® Model 2400S Transmitters for DeviceNet
™
I/O Data Format
Table 3-1 Summary of input assemblies continued
Instance ID Data description Size (bytes) Data type Notes
16 Status
Temperature-corrected density
Temperature-corrected volume flow
Temperature-corrected volume inventory
Average temperature-corrected density
Average temperature
17 Status
Mass flow
Volume flow
Temperature
Reference density
Specific gravity
18 Status
Mass flow
Volume flow
Temperature
Density
Concentration
19 Status
Mass flow
Volume flow
Temperature
Density
Baume
20 Status
Temperature
Density
Net mass flow
Net mass total
Net mass inventory
21 Status
Temperature
Density
Net volume flow
Net volume total
Net volume inventory
22 Status
Mass flow
Temperature
Density
Reference density
Net mass flow
23 Status
Volume flow
Temperature
Density
Reference density
Net volume flow
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
Petroleum
measurement
application
Concentration
measurement
application
Concentration
measurement
application
Concentration
measurement
application
Concentration
measurement
application
Concentration
measurement
application
Concentration
measurement
application
Concentration
measurement
application
Object Model Configuration Data I/O Data FormatIntroduction
Device Profile 7
I/O Data Format
Table 3-1 Summary of input assemblies continued
Instance ID Data description Size (bytes) Data type Notes
24 Status
Mass flow
Volume flow
Density
Reference density
Standard volume flow
25 Status
Mass flow
Temperature
Density
Reference density
Concentration
26 Status
User-specified variable 1
User-specified variable 2
User-specified variable 3
User-specified variable 4
User-specified variable 5
(1) Default = Mass flow.
(2) Default = Temperature.
(3) Default = Density.
(4) Default = Volume flow.
(5) Default = Drive gain.
(1)
(2)
(3)
(4)
(5)
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
21 BOOL
REAL
REAL
REAL
REAL
REAL
Concentration
measurement
application
Concentration
measurement
application
Configurable
assembly
3.3 Summary of output assemblies
The Model 2400S DN transmitter output assemblies are summarized in Table 3-2.
Table 3-2 Summary of output assemblies
Instance ID Data Description Size (bytes) Data Type Notes
0 No output data 0 NULL No data
50 External pressure 4 REAL Used for pressure
compensation
51 External temperature 4 REAL Used for
temperature
correction
52 External pressure
External temperature
53 Start/stop totalizers and inventories 1 BOOL Totalizer control
54 Reset all totalizers 1 BOOL Totalizer control
55 Reset all inventories 1 BOOL Totalizer control
56 Start/stop totalizers and inventories
Reset all totalizers
57 Start/stop totalizers and inventories
Reset all totalizers
Reset all inventories
58 Start meter verification 1 BOOL Meter verification
8R E A L
REAL
Used for pressure
compensation and
temperature
correction
2 BOOL
Totalizer control
BOOL
3 BOOL
Totalizer control
BOOL
BOOL
8 Micro Motion ® Model 2400S Transmitters for DeviceNet
™
I/O Data Format
Table 3-2 Summary of output assemblies continued
Instance ID Data Description Size (bytes) Data Type Notes
59 External pressure
Start meter verification
60 External temperature
Start meter verification
61 Start/stop totalizers
Start meter verification
62 Reset all totalizers
Start meter verification
63 Reset all inventories
Start meter verification
64 Start/stop totalizers
Reset all totalizers
Start meter verification
65 Start/stop totalizers
Reset all totalizers
Reset all inventories
Start meter verification
5R E A L
BOOL
5R E A L
BOOL
2 BOOL
BOOL
2 BOOL
BOOL
2 BOOL
BOOL
3 BOOL
BOOL
BOOL
4 BOOL
BOOL
BOOL
BOOL
Pressure
compensation and
meter verification
Temperature
correction and meter
verification
Totalizer control and
meter verification
Totalizer control and
meter verification
Totalizer control and
meter verification
Totalizer control and
meter verification
Totalizer control and
meter verification
Object Model Configuration Data I/O Data FormatIntroduction
3.4 Input assembly details
3.4.1 Detail of Input Assembly 1 – Status and mass flow
• Assembly instance: 1
• Data size: 5 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
3.4.2 Detail of Input Assembly 2 – Status and volume flow
• Assembly instance: 2
• Data size: 5 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-2-4]
Bytes 1–4 Volume flow [0x0A-2-3]
Device Profile 9
I/O Data Format
3.4.3 Detail of Input Assembly 3 – Status, mass flow, and mass total
• Assembly instance: 3
• Data size: 9 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Mass total [0x0A-1-100]
3.4.4 Detail of Input Assembly 4 – Status, volume flow, and volume total
• Assembly instance: 4
• Data size: 9 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-2-4]
Bytes 1–4 Volume flow [0x0A-2-3]
Bytes 5–8 Volume total [0x0A-2-100]
3.4.5 Detail of Input Assembly 5 – Status and basic process variables
• Assembly instance: 5
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Temperature [0x0A-4-3]
Bytes 9–12 Density [0x0A-3-3]
Bytes 13–16 Volume flow [0x0A-2-3]
Bytes 17–20 Drive gain [0x66-1-20]
10 Micro Motion ® Model 2400S Transmitters for DeviceNet
™
I/O Data Format
3.4.6 Detail of Input Assembly 6 – Status, mass flow, mass totals, and other process variables
• Assembly instance: 6
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Mass total [0x0A-1-100]
Bytes 9–12 Mass inventory [0x0A-1-101]
Bytes 13–16 Temperature [0x0A-4-3]
Bytes 17–20 Density [0x0A-3-3]
3.4.7 Detail of Input Assembly 7 – Status, volume flow, volume totals, and other process
variables
• Assembly instance: 7
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-2-4]
Bytes 1–4 Volume flow [0x0A-2-3]
Bytes 5–8 Volume total [0x0A-2-100]
Bytes 9–12 Volume inventory [0x0A-2-101]
Bytes 13–16 Temperature [0x0A-4-3]
Bytes 17–20 Density [0x0A-3-3]
3.4.8 Detail of Input Assembly 8 – Status, gas standard volume process variables
• Assembly instance: 8
• Data size: 21 bytes
•T y p e : I n p u t
Object Model Configuration Data I/O Data FormatIntroduction
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Mass total [0x0A-1-100]
Bytes 9–12 Temperature [0x0A-4-3]
Bytes 13–16 Gas standard volume flow [0x64-1-1]
Bytes 17–20 Gas standard volume total [0x64-1-2]
Device Profile 11
I/O Data Format
3.4.9 Detail of Input Assembly 9 – Status, gas standard volume process variables
• Assembly instance: 9
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Temperature [0x0A-4-3]
Bytes 9–12 Gas standard volume flow [0x64-1-1]
Bytes 13–16 Gas standard volume total [0x64-1-2]
Bytes 17–20 Gas standard volume inventory [0x64-1-3]
3.4.10 Detail of Input Assembly 10 – Status, gas standard volume process variables
• Assembly instance: 10
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-4-4]
Bytes 1–4 Temperature [0x0A-4-3]
Bytes 5–8 Drive gain [0x66-1-20]
Bytes 9–12 Gas standard volume flow [0x64-1-1]
Bytes 13–16 Gas standard volume total [0x64-1-2]
Bytes 17–20 Gas standard volume inventory [0x64-1-3]
3.4.11 Detail of Input Assembly 11 – Status, gas standard volume process variables
• Assembly instance: 11
• Data size: 5 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-2-4]
Bytes 1–4 Gas standard volume flow [0x64-1-1]
12 Micro Motion ® Model 2400S Transmitters for DeviceNet
™
I/O Data Format
3.4.12 Detail of Input Assembly 12 – Status, gas standard volume process variables
• Assembly instance: 12
• Data size: 13 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-2-4]
Bytes 1–4 Gas standard volume flow [0x64-1-1]
Bytes 5–8 Gas standard volume total [0x64-1-2]
Bytes 9–12 Gas standard volume inventory [0x64-1-3]
3.4.13 Detail of Input Assembly 13 – Status, petroleum measurement process variables
• Assembly instance: 13
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-2-4]
Bytes 1–4 Volume flow [0x0A-2-3]
Bytes 5–8 Volume total [0x0A-2-100]
Bytes 9–12 Volume inventory [0x0A-2-101]
Bytes 13–16 Temperature-corrected volume flow [0x69-1-2]
Bytes 17–20 Temperature-corrected volume total [0x69-1-3]
3.4.14 Detail of Input Assembly 14 – Status, petroleum measurement process variables
• Assembly instance: 14
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-2-4]
Bytes 1–4 Volume flow [0x0A-2-3]
Bytes 5–8 Volume total [0x0A-2-100]
Bytes 9–12 Temperature-corrected density [0x69-1-1]
Bytes 13–16 Temperature-corrected volume flow [0x69-1-2]
Bytes 17–20 Temperature-corrected volume inventory [0x69-1-4]
Object Model Configuration Data I/O Data FormatIntroduction
Device Profile 13
I/O Data Format
3.4.15 Detail of Input Assembly 15 – Status, petroleum measurement process variables
• Assembly instance: 15
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Mass total [0x0A-1-100]
Bytes 9–12 Volume flow [0x0A-2-3]
Bytes 13–16 Volume total [0x0A-2-100]
Bytes 17–20 Temperature-corrected density [0x69-1-1]
3.4.16 Detail of Input Assembly 16 – Status, petroleum measurement process variables
• Assembly instance: 16
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-3-4]
Bytes 1–4 Temperature-corrected density [0x69-1-1]
Bytes 5–8 Temperature-corrected volume flow [0x69-1-2]
Bytes 9–12 Temperature-corrected volume inventory [0x69-1-4]
Bytes 13–16 Average temperature-corrected density [0x69-1-5]
Bytes 17–20 Average temperature [0x69-1-6]
3.4.17 Detail of Input Assembly 17 – Concentration measurement densities
• Assembly instance: 17
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Volume flow [0x0A-2-3]
Bytes 9–12 Temperature [0x0A-4-3]
Bytes 13–16 Reference density [0x6A-1-1]
Bytes 17–20 Specific gravity [0x6A-1-2]
14 Micro Motion ® Model 2400S Transmitters for DeviceNet
™
I/O Data Format
3.4.18 Detail of Input Assembly 18 – Concentration measurement concentration
• Assembly instance: 18
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Volume flow [0x0A-2-3]
Bytes 9–12 Temperature [0x0A-4-3]
Bytes 13–16 Density [0x0A-3-3]
Bytes 17–20 Concentration [0x6A-1-12]
3.4.19 Detail of Input Assembly 19 – Concentration measurement Baume
• Assembly instance: 19
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Volume flow [0x0A-2-3]
Bytes 9–12 Temperature [0x0A-4-3]
Bytes 13–16 Density [0x0A-3-3]
Bytes 17–20 Baume [0x6A-1-13]
3.4.20 Detail of Input Assembly 20 – Concentration measurement net mass flow and totals
• Assembly instance: 20
• Data size: 21 bytes
•T y p e : I n p u t
Object Model Configuration Data I/O Data FormatIntroduction
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Temperature [0x0A-4-3]
Bytes 5–8 Density [0x0A-3-3]
Bytes 9–12 Net mass flow [0x6A-1-6]
Bytes 13–16 Net mass total [0x6A-1-7]
Bytes 17–20 Net mass inventory [0x6A-1-8]
Device Profile 15
I/O Data Format
3.4.21 Detail of Input Assembly 21 – Concentration measurement net volume flow and totals
• Assembly instance: 21
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-2-4]
Bytes 1–4 Temperature [0x0A-4-3]
Bytes 5–8 Density [0x0A-3-3]
Bytes 9–12 Net volume flow [0x6A-1-9]
Bytes 13–16 Net volume total [0x6A-1-10]
Bytes 17–20 Net volume inventory [0x6A-1-11]
3.4.22 Detail of Input Assembly 22 – Concentration measurement reference density and net
• Assembly instance: 22
• Data size: 21 bytes
mass flow
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Temperature [0x0A-4-3]
Bytes 9–12 Density [0x0A-3-3]
Bytes 13–16 Reference density [0x6A-1-1]
Bytes 17–20 Net mass flow [0x6A-1-6]
3.4.23 Detail of Input Assembly 23 – Concentration measurement reference density and net
volume flow
• Assembly instance: 23
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-2-4]
Bytes 1–4 Volume flow [0x0A-2-3]
Bytes 5–8 Temperature [0x0A-4-3]
Bytes 9–12 Density [0x0A-3-3]
Bytes 13–16 Reference density [0x6A-1-1]
Bytes 17–20 Net volume flow [0x6A-1-9]
16 Micro Motion ® Model 2400S Transmitters for DeviceNet
™
I/O Data Format
3.4.24 Detail of Input Assembly 24 – Concentration measurement reference density and
• Assembly instance: 24
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Volume flow [0x0A-2-3]
Bytes 9–12 Density [0x0A-3-3]
Bytes 13–16 Reference density [0x6A-1-1]
Bytes 17–20 Standard volume flow [0x6A-1-3]
3.4.25 Detail of Input Assembly 25 – Status, concentration measurement reference density
standard volume flow
Object Model Configuration Data I/O Data FormatIntroduction
and concentration
• Assembly instance: 25
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Mass flow [0x0A-1-3]
Bytes 5–8 Temperature [0x0A-4-3]
Bytes 9–12 Density [0x0A-3-3]
Bytes 13–16 Reference density [0x6A-1-1]
Bytes 17–20 Concentration [0x6A-1-12]
3.4.26 Detail of Input Assembly 26 – Status, user-selected process variables
• Assembly instance: 26
• Data size: 21 bytes
•T y p e : I n p u t
Bytes Description EPATH
Byte 0 Status [0x0A-1-4]
Bytes 1–4 Process variable 1 (Default = Mass flow) [0x04-26-100]
Bytes 5–8 Process variable 2 (Default = Temperature) [0x04-26-101]
Bytes 9–12 Process variable 3 (Default = Density) [0x04-26-102]
Bytes 13–16 Process variable 4 (Default = Volume flow) [0x04-26-103]
Bytes 17–20 Process variable 5 (Default = Drive gain) [0x04-26-104]
Device Profile 17
I/O Data Format
3.5 Output assembly details
3.5.1 Detail of Output Assembly 0 – No output data
• Assembly instance: 0
• Data size: 0 bytes
• Type: Output
3.5.2 Detail of Output Assembly 50 – External pressure
• Assembly instance: 50
• Data size: 4 bytes
• Type: Output
Bytes Description EPATH
Bytes 0–3 External pressure [0x65-1-28]
3.5.3 Detail of Output Assembly 51 – External temperature
• Assembly instance: 51
• Data size: 4 bytes
• Type: Output
Bytes Description EPATH
Bytes 0–3 External temperature [0x65-1-26]
3.5.4 Detail of Output Assembly 52 – External pressure and temperature
• Assembly instance: 52
• Data size: 8 bytes
• Type: Output
Bytes Description EPATH
Bytes 0–3 External pressure [0x65-1-28]
Bytes 4–7 External temperature [0x65-1-26]
3.5.5 Detail of Output Assembly 53 – Start/stop all totalizers and inventories
• Assembly instance: 53
• Data size: 1 byte
• Type: Output
Bytes Description EPATH
Byte 0 Start/stop all totalizers and inventories [0x0A-0-100]
18 Micro Motion ® Model 2400S Transmitters for DeviceNet
™
I/O Data Format
3.5.6 Detail of Output Assembly 54 – Reset all totalizers
• Assembly instance: 54
• Data size: 1 byte
• Type: Output
Bytes Description EPATH
Byte 0 Reset all totalizers [0x0A-0-101]
3.5.7 Detail of Output Assembly 55 – Reset all inventories
• Assembly instance: 55
• Data size: 1 byte
• Type: Output
Bytes Description EPATH
Byte 0 Reset all inventories [0x0A-0-102]
Object Model Configuration Data I/O Data FormatIntroduction
3.5.8 Detail of Output Assembly 56 – Start/stop all totalizers and inventories and reset
all totalizers
• Assembly instance: 56
• Data size: 2 bytes
• Type: Output
Bytes Description EPATH
Byte 0 Start/stop all totalizers and inventories [0x0A-0-100]
Byte 1 Reset all totalizers [0x0A-0-101]
3.5.9 Detail of Output Assembly 57 – Start/stop all totalizers and inventories, reset all
totalizers, and reset all inventories
• Assembly instance: 57
• Data size: 3 bytes
• Type: Output
8
Bytes Description EPATH
Byte 0 Start/stop all totalizers and inventories [0x0A-0-100]
Byte 1 Reset all totalizers [0x0A-0-101]
Byte 2 Reset all inventories [0x0A-0-102]
Device Profile 19
I/O Data Format
3.5.10 Detail of Output Assembly 58 – Start meter verification
• Assembly instance: 58
• Data size: 1 byte
• Type: Output
Bytes Description EPATH
Byte 0 Start meter verification [0x66-1-86]
3.5.11 Detail of Output Assembly 59 – External pressure, start meter verification
• Assembly instance: 59
• Data size: 5 bytes
• Type: Output
Bytes Description EPATH
Bytes 0–3 External pressure [0x65-1-28]
Byte 4 Start meter verification [0x66-1-86]
3.5.12 Detail of Output Assembly 60 – External temperature and start meter verification
• Assembly instance: 60
• Data size: 5 bytes
• Type: Output
Bytes Description EPATH
Bytes 0–3 External temperature [0x65-1-26]
Byte 4 Start meter verification [0x66-1-86]
3.5.13 Detail of Output Assembly 61 – Start/stop all totalizers and inventories, start meter
verification
• Assembly instance: 61
• Data size: 2 bytes
• Type: Output
Bytes Description EPATH
Byte 0 Start/stop all totalizers and inventories [0x0A-0-100]
Byte 1 Start meter verification [0x66-1-86]
20 Micro Motion ® Model 2400S Transmitters for DeviceNet
™
I/O Data Format
3.5.14 Detail of Output Assembly 62 – Reset all totalizers and start meter verification
• Assembly instance: 62
• Data size: 2 bytes
• Type: Output
Bytes Description EPATH
Byte 0 Reset all totalizers [0x0A-0-101]
Byte 1 Start meter verification [0x66-1-86]
3.5.15 Detail of Output Assembly 63 – Reset all inventories and start meter verification
• Assembly instance: 63
• Data size: 2 bytes
• Type: Output
Bytes Description EPATH
Byte 0 Reset all inventories [0x0A-0-102]
Byte 1 Start meter verification [0x66-1-86]
3.5.16 Detail of Output Assembly 64 – Start/stop all totalizers, reset all totalizers, start meter
verification
• Assembly instance: 64
• Data size: 3 bytes
• Type: Output
Bytes Description EPATH
Byte 0 Start/stop all totalizers and inventories [0x0A-0-100]
Byte 1 Reset all totalizers [0x0A-0-101]
Byte 2 Start meter verification [0x66-1-86]
3.5.17 Detail of Output Assembly 65 – Start/stop all totalizers, reset all totalizers, reset all
inventories, start meter verification
Object Model Configuration Data I/O Data FormatIntroduction
• Assembly instance: 65
• Data size: 4 bytes
• Type: Output
Bytes Description EPATH
Byte 0 Start/stop all totalizers and inventories [0x0A-0-100]
Byte 1 Reset all totalizers [0x0A-0-101]
Byte 2 Reset all inventories [0x0A-0-102]
Byte 3 Start meter verification [0x66-1-86]
Device Profile 21
22 Micro Motion ® Model 2400S Transmitters for DeviceNet
™