The Micro Motion
dual-channel/stream gas applications.
Software Version:
1020 – Gas Applications.
Models Covered
7950MAA0*****
:
®
7950 Signal Converter can be used for
:
Micro Motion
7950 Signal Converte
IMPORTANT NOTICE
Because we are continuously improving our products, some of the menus which appear on
your instrument’s display may not be exactly as illustrated and described in this manual.
However, because the menus are simple and intuitive, this should not cause any major
problems.
This manual is concurrent with embedded software version 501020, issue 2.11.
Static precautions
Some parts of the instrument (such as circuit boards) may be damaged by static electricity. Therefore, when
carrying out any work which involves the risk of static damage to the instrument, the instructions show the
following notice:
CAUTION
At such times you must wear an earthed wrist-strap to protect the instrument.
While carrying out this procedure, you must wear an earthed wrist strap at all times to protect the instrument against static shock.
Safety information
NOTE: This information applies only to those instruments which are mains-powered.
Electricity is dangerous and you risk injury or death if you do not disconnect the power supplies before
carrying out some of the procedures given in this manual. Whenever there is such a hazard, the instructions
show a notice similar to the following:
WARNING
You must heed any such warnings and make sure that, before you go any further:
Electricity is dangerous and can kill.
Disconnect all power supplies before proceeding.
• All power leads are un-powered.
• All power leads are disconnected from the equipment which you are working on unless the
instructions tell you otherwise.
•You obey any other common-sense precautions which may apply to your situation.
If you obey these sensible precautions, you can work on the equipment in complete safety.
Battery-backed Memory notice
• It is essential that the Lithium Cell used for the battery backup is installed at all times (other than during
replacement). The 7950 Micro Motion
If it is necessary to run the units without batteries for Intrinsic Safety reasons, then the battery should be
replaced with a shorting disk inserted in the battery holder. Please consult the factory for further advice.
• Replace the battery when the "Low Battery" system alarm is indicated. The procedure is in Chapter 14.
®
Signal Converter will not power-up correctly if this battery is missing.
CONTENTS
1. About this manual 1.1
1.1 What this man ual tells you 1.1
1.2 Who shou ld use this manual 1.1
1.3 Soft ware versions cover ed by this manual 1.1
2. Getting started 2.1
2.1 What this chapter tells you 2.1
2.2 What the examples show you 2.1
2.3 If you need help… 2.1
2.4 Exam ple 1: 795x with a 7812 Gas density transducer 2.2
2.5 Exam ple 2: 795x with a 3098 Gas specific gravity transducer 2.6
2.6 Exam ple 3: 795x with a mA-type temperature transmitter 2.9
2.7 Exam ple 4: 795x with a mA-type pressure transmitter 2.11
2.8 Example 5: 795x with a RTD-type temperature transmitter 2.14
3. About the Micro Motion® 7950 3.1
3.1 Background 3.1
3.2 What the 7950 Dual Channel Gas Signal Converter does 3.1
3.3 Physical description of the 7950 3.2
3.4 Communications 3.3
3.5 Typical installations 3.3
3.6 Checking your software version 3.4
4. What you can connect to a 7950 4.1
The information in this chapter has been moved to Appendix C
5. Installing the system 5.1
5.1 What this chap ter tells you 5.1
5.2 Hazardous and non-hazardous environments 5.1
5.3 Installation procedure 5.1
5.4 Step 1: Drawing up a wiring schedule 5.1
5.5 Step 2: Unpacking the instrument 5.1
5.6 Step 3: Setting DIP-switches 5.2
5.7 Step 4: Fitting the 7950 5.3
5.8 Step 5: Making the connectio ns 5.3
5.9 Step 6: Earthing the instrument 5.4
5.10 Step 7: Connecting the power supply 5.5
6. The keyboard, display and indicators 6.1
6.1 What this chapter tells you 6.1
6.2 The layout of the front panel 6.1
6.3 What the display shows 6.2
6.4 How the buttons work 6.2
6.5 Using the buttons to move around the menus 6.2
6.6 Using the buttons to view stored data 6.3
6.7 Using the buttons to edit information 6.4
6.7.1 Text editing 6.4
6.7.2 Multiple-choice option selection 6.5
6.7.3 Numerical editing 6.5
6.7.4 Date and time editing 6.6
6.8 The 795x character set 6.7
6.9 LED indicators 6.7
6.10 Summary of key functions 6.8
7. The menu system 7.1
7.1 What this chap ter tells you 7.1
7.2 What the menu system does 7.1
7.3 How the menu system works 7.1
8. Alarms 8.1
8.1 Alarms 8.1
8.1.1 Alarm types 8.1
8.1.2 Alarm indicators 8.1
8.1.3 How alarms are received and stored 8.2
8.1.4 Examining the Alarm Status Display and Historical Alarm Log 8.2
8.1.5 What the Status Display tells you 8.3
8.1.6 What the entries in the Historical Log tell you 8.3
8.1.7 Clearing all alarms in the Historical Alarm Log 8.4
8.1.8 Alarm messages 8.4
9. Additional facilities 9.1
9.1 What this chap ter tells you 9.1
9.2 Selecting u nits and data forma ts 9.1
9.3 Limits 9.1
9.4 Fallback values and modes 9.2
9.5 Analogue 0% and 100% values 9.2
9.6 Live and set da ta 9.2
9.7 Units which the 795x can display 9.3
10. Configuring the instrument using wizards 10.1
10.1 What this chapter tells you 10.1
10.2 Wizards : Configuring the easy way 10.1
10.3 Wizard Maps : Conventions used 10.1
10.4 Quick-start Guide ( Set-up Wizards ) 10.3
10.5 Set-up wizard selection map 10.4
10.6 Units wizard selection map 10.5
10.7 Density 1 application wizard 10.7
10.8 Density 2 application wizard 10.8
10.9 SG-1 application wizard 10.9
10.10 SG-2 application wizard 10.11
10.11 SG-1&2 application wizard 10.12
10.12 Line density wizard 10.13
10.13 Base density wizard 10.16
10.14 Specific gravity wizard 10.18
10.15 Temperature wizard 10.20
10.16 Pressure wizard 10.22
10.17 Transmitter wizard 10.24
10.18 Special calculation wizard 10.27
10.19 Analogue outputs wizard 10.28
10.20 Alarms wizard 10.29
10.21 Multi-view wizard 10.30
10.22 Full setup wizard 10.31
11. Configuring by using the menus 11.1
11.1 What does configuration involve? 11.1
11.2 Before you start 11.1
11.3 Recommended sequence for configuration 11.2
11.4 What Sections 11.5 - 11.18 tell you 11.3
11.5 Configuring analogue inputs 11.5
11.6 Configuring transducer details 11.6
11.7 Configuring transmitter detail 11.7
11.8 Configuring line density
11.11
11.9 Configuring base density 11.16
11.10 Configuring specific gravity 11.19
11.11 Configuring energy 11.22
11.12 Configuring custom applications 11.23
11.13 Configuring mA outputs 11.24
11.14 Configuring other parameters 11.25
11.15 Configuring Multiview 11.29
12. Routine operation 12.1
12.1 What this chapter tells you 12.1
12.2 The menu diagrams 12.1
12.3 Security and passwords 12.8
12.4 How you can edit displayed information 12.8
12.5 Making data values Live or Set 12.10
12.6 Changing the units which are displayed 12.11
12.7 Changing fallback values 12.11
12.8 Changing the time and date 12.11
12.9 Checking the performance of the 795x 12.11
12.10 Giving your 795x a unique identifier 12.12
12.11 Printed reports 12.12
13. Routine maintenance and fault-finding 13.1
13.1 Cleaning the instrument 13.1
13.2 Fault-finding 13.1
14. Removal and replacement of parts 14.1
14.1 Front panel assembly 14.1
14.2 Display 14.3
14.3 Connector Board 14.4
14.4 Microprocessor board 14.5
14.5 Screen and RFI conductive strips 14.5
14.6 Terminal cover seal 14.6
14.7 Gland plate seal 14.8
14.8 Fuses 14.9
14.9 Back-up battery 14.10
15. Assembly drawing and parts list 15.1
15.1 What the drawing and parts list tells you 15.1
15.2 How to obtain spare parts 15.1
Appendices
Appendix A Glossary A.1
Appendix B Blank wiring schedule B.1
Appendix C Technical data for the 7950 C.1
Appendix D Calculations and theory D.1
1. About this manual
1.1 What this manual tells you
This manual tells you how to install, configure, operate, and service the instrument. In addition, some information
is given to help you identify and correct some of the more common faults which may occur. However, since
repairs are done by changing suspected faulty assemblies, fault-finding to board component level is not covered.
This manual assumes that all devices or peripherals to be connected to the 795x have their own documentatio n
which tells you how to install and configure them. For this reason it is assumed that anything which you want to
link to the 795x is already installed and working correctly in accordance with the manufacturer’s instructions.
Since the instrument can be used for a wide variety of purposes, it is driven by software specially for your application.
This manual gives information about the so ftw are which applie s to y our machine only .
Throughout this manual the term '795x' is used to refe r to al l members of the 795x family (7950 and 7951).
Chapter 1 About this manual
1.2 Who should use this manual
This manual is for anyone who installs, uses, services or repairs the 795x.
1.3 Software version covered by this manual
The software version dealt with in this manual is given on the title page. Chapter 3 tells you about the software is
installed in your instrument.
Page 1.1
Chapter 1 About this manual
Page 1.2
2. Getting started
2.1 What this chapter tells you
If you are new to the Micro Motion® 7950 Signal Converter, the worked examples in this chapter can help you to
become familiar with the installation and configuration procedures. The examples are:
• Example 1: 7950 with a 12 (See page 2.2)
• Example 2: 7950 with a 3098 Gas Specific Gravity (See page 2.6)
• Example 3: 7950 with a mA-type temperature transmitter (See page 2.9)
• Example 4: 7950 with a mA-type pressure transmitter (See page 2.12)
• Example 5: 7950 with a RTD-type temperature transmitter (See page 2.14)
Work through whichever one is most like your installation.
2.2 What the examples show you
Chapter 2 Getting started
Each example shows you how to:
• wire up a simple system
• set the DIP switches inside the 7950
• find the menu from which you start configuration
• clear the memory of details of any existing configuration (OPTIONAL)
• select the appropriate wizard to configure the simple system
• work through the wizard and button in information
• view the results of your configuration
The examples do not give full instructions on how to fit and configure installations. They are intended purely to
give you confidence to install and configure your own equipment. Chapter 5 tells you how to make permanent
installations.
2.3 If you need help...
If you get into difficulties...
If you get into difficulties when using the wizards, you can abandon the configuration and start again as follows:
1. From the menu, keep selecting NO (usually by pressing the c-button) or, if that option is not available:
2. Press ENTER until you can start selecting NO.
3. Carry on with (1) and (2) until you return to the wizards menu where you started.
4. Start the worked example again. The configuration you abandoned is clear ed from the instrument’s
memory when you begin again.
If you don’t know where the buttons are...
Chapter 6 shows how to find all the buttons referred to in the worked examples.
Page 2.1
Chapter 2 Getting started
2.4 Example 1: 7950 with a 7812 gas density meter
About this example
†
example shows you how to connect either a 7812
7950
This NON-HAZARDOUS (SAFE) AREA ONLY INSTALLATION
gas density meter to the 7950, and then uses the “Density 1” wizard to configure the system.
In this example, the “Density 1” wizard is used to configure a connection as follows:
• A single densitometer is connected to Density Input 1.
Work through the example by following the instructions below. If you are not sure where the buttons are, refer to
Chapter 6.
Connect the
meter
1. Wire the meter to the 7950, as in:
Figure 2.1 for a 7812 gas density meter.
Note: Refer to meter documentation for other wiring arrangements.
2. Earth the 7950 to a suitable earth point.
Notes:
(a) You must comply with EMC regulations.
(b) Use a screened multi-core cable that has twisted pairs.
7812
1
SIG A
2
3
SIG B
4
+24V Power
Signal +
0V
Power
Signal -
PL9/1
PL9/2
PL9/4
PL9/3
Figure 2.1: Safe area wiring for a 7812 (3-wire arrangement)
†
Hazardous area considerations: Refer to meter documentation for details of Intrinsically Safe Barrier/Isolator requirements.
Page 2.2
Chapter 2 Getting started
Turn on the
power
Go to the
wizards menu
Clear existing
configuration
(This is optional)
Select the wizard 12. Press the b-button then the UP-ARROW or DOWN-ARROW button to scroll through
Start of wizard 14. Press the d-button to answer YES to “Edit Gas density A?”
Enter
densitometer
calibration
factors
3. Turn on the power to the system. The system goes through a Power On Self Test
(POST) routine which takes less than 30 seconds. When it is finished, ignore any
flashing alarm lights which may appear.
4. Press the MENU button to go to Page 1 of the Main Menu (if you aren’t there already).
5. Press the DOWN-ARROW button twice (to go to page 3 of the menu).
6. Press the b-button to select “Configure”.
7. Press the a-button twice to go to the wizards menu.
8. Press the b-button then the UP-ARROW or DOWN-ARROW button to scroll through the
option list until “Initialise” is shown.
9. Press the b-button to select “Initialise”.
10. Press the d-button to confirm that you want to lose the current configuration.
11. Wait a few seconds until “initialise” on line 2 of the display changes back to
“Choose option”.
the option list until “Density 1” is shown.
13. Press the b-button to select “Density 1”.
15. Press the b-button, and then input the factor K0 from the Calibration Certificate that
was shipped with the meter.
16. Press the b-button then ENTER to confirm the K0 value.
17. Enter values for factors K1 and K2 in the same way as for K0.
Note: Figure 2.3 on page 2.4 shows where to find the K0, K1 and K2 factors on a
calibration certificate. Always
shipped with the connected meter.
use values from the calibration certificate that was
Enter
temperature
correction
factors
Skip over the
next few
questions
View how you
have configured
Line density
18. Press b-button to start the correction selection process.
19. Use the UP-ARROW button to scroll through the options until “Temp” appears on line 2.
20. Press the b-button and then the ENTER button to confirm that temperature correction is
to be applied.
21. Enter factors K18 and K19 in the same way as for K0, K1 and K2.
22. Press the ENTER button to skip past the “Density offset” prompt
23. Keep pressing the c-button (to answer NO to all questions) until the wizard is exited.
24. Press the MENU button.
25. Press the a-button twice. The display looks similar to that shown in Figure 2.2,
although values and titles shown may vary.
Figure 2.2: Prime Line density display
Page 2.3
Chapter 2 Getting started
C A L I B R A T I O N C E R T I F I C A T E
7812xx GAS DENSITY METER
PRESSURE TESTED TO 375 BAR
DENSITY CALIBRATION FOR NITRIGEN AT 20 DEG C
Based on Pressure-Temperature-Density Data in IUPAC Tables
DENSITY
[ kg/m2]
0
1
2
5
10
15
20
30
40
50
60
70
80
90
TEMPERATURE COEFFICIENT DATA
Dt = DI ( 1 + K18 ( t - 20 ) ) + K19 ( t - 20 ) K18 = -1.36E-05
USER GAS OFFSET DATA
Nitrogen/Methane Gas Mixture Over Density Range 10 to 60 KG/M3
Figure 2.3: Circled areas on an example calibration certificate showing where to find
values for K0, K1, K2, K18 and K19.
DATE14JUL97
View the Multiview display
Chapter 2 Getting started
26. Press the MULTI-VIEW DISPLAY button. The display looks similar to that in Figure
2.4, although values shown may vary.
27. Pressing the DOWN-ARROW button results in the message:
“Invalid Multiview Page”
This appears because it is possible to have more than one Multiview page and it is
simply saying that no more pages exist. In this case, only four items are defined and
they fit on one page.
Pressing the UP-ARROW button makes the previous page to re-appear.
Note that it may be necessary to press the UP-ARROW button several times before the
first Multi-view page appears.
From Multi-view key
End of Worked Example 1
a
b
c
d
a
b
c
d
V
V
Density0.000
Invalid Multiview
Page
Figure 2.4: Multi-view display
Page 2.5
Chapter 2 Getting started
r
2.5 Example 2: 7950 with a 3098 Gas Specific Gravity Meter
About this example
This example shows you how to connect a 3098 to the 7950 and then use the “SG 1” wizard to configure the
system.
In this example, the “SG 1” wizard is used to configure a connection as follows:
• A single 3098 is connected to Density Input 3.
Work through the example by following the instructions below. If you are not sure where the buttons are, refer to
Chapter 6.
Connect the
meter
1. Wire the meter to the 7950, as in :
Figure 2.5 for a NON-HAZARDOUS (SAFE) AREA OR
Figure 2.6 for a HAZARDOUS AREA
2. Earth the 7950 to a suitable earth point.
Notes:
(a) You must comply with EMC regulations.
(b) Use a screened multi-core cable that has twisted pairs.
30987950
40ohms
+
1
Sig
3
Neg
2
4
(24V)Power +
Signal +
Signal -
(0V) Power -
PL 10/1
PL 10/2
PL 10/3
PL 10/4
Figure 2.5: Non-hazardous (Safe) area wiring for a 3098
Barrier as defined in
3098
the latest
1
2
3
+
Sig
Neg
3
4
3098
specification
1
2
(24v) Power +
Signal +
Signal -
(0v) Powe
7950
PL 10/ 1
PL 10 / 2
PL 10 / 3
-
PL 10 / 4
Intrinsically Safe Earth
Figure 2.6: Hazardous area wiring for a 3098
Page 2.6
Turn on the
power
Chapter 2 Getting started
3. Turn on the power to the system. The system goes through a Power On Self Test
(POST) routine which takes less than 30 seconds. When it is finished, ignore any
flashing alarm lights which may appear.
Go to the
wizards menu
Clear existing
configuration
(This is optional)
Select the
wizard
Start of wizard 14. Press the d-button to answer YES to the question “Edit Gravitometer A?”.
Enter
gravitometer
calibration
factors
4. Press the MENU button to go to Page 1 of the Main Menu (if you aren’t there already).
5. Press the DOWN-ARROW button twice (to go to page 3 of the menu).
6. Press the b-button to select “Configure”.
7. Press the a-button twice to go to the wizards menu.
8. Press the b-button then the UP-ARROW or DOWN-ARROW button to scroll through the
option list until “Initialise” is shown.
9. Press the b-button to select “Initialise”.
10. Press the d-button to confirm that you want to lose the current configuration.
11. Wait a few seconds until “initialise” on line 2 of the display changes back to
“Choose option”.
12. Press the b-button then the UP-ARROW or DOWN-ARROW button to scroll through
the option list until “SG 1” is shown.
13. Press the b-button to select “SG 1”.
15. Press the b-button, then input the factor K2 from the Calibration Certificate that was
shipped with the meter.
16. Press the b-button then ENTER to confirm the K2 value.
17. Press the b-button, then input the factor K0 from the Calibration Certificate that was
shipped with the meter.
18. Press the b-button then ENTER to confirm the K0 value.
Skip over other
questions
View how you
have configured
Specific gravity
View the
Multiview
display
19. Press the c-button several times (to answer NO to all questions) until the wizard is
exited.
20. Press the MENU button.
21. Press the c-button and then press the a-button. The display looks similar to that
shown in Figure 2.7 although values shown may vary.
Figure 2.7: Prime Specific Gravity display
22. Press the MULTI-VIEW DISPLAY button. The display looks similar to that in
Figure 2.8, although values and titles shown may vary.
Page 2.7
Chapter 2 Getting started
From Multi-view key
a
b
c
d
a
b
c
d
V
V
Density0.000
Invalid Multiview
Page
Figure 2.8: Multi-View display
End of Worked Example 2
Page 2.8
Chapter 2 Getting started
2.6 Example 3: 7950 with a mA-type temperature transmitter
About this example
This example shows you how to connect a mA-type temperature transmitter to the 7950, and then use the
“Temperature” wizard to configure the system.
In this example, the “Temperature” wizard is used to configure connections as follows:
• A single temperature transmitter is connected to Analogue Input 3.
Now work through the example by following the instructions below. If you are not sure where the buttons are,
refer to the diagram at the start of this chapter.
Connect the
meter
Set DIP switch 3. Ensure that the DIP-switch, inside the 7950, is set as shown below.
1. Wire the temperature transmitter to the 7950, as in Figure 2.9
2. Earth the 7950 to a suitable earth point.
Notes:
(a) You must comply with EMC regulations.
(b) Use a screened multi-core cable that has twisted pairs.
Loop-powered
4-20mA
temperature
transmitter
7950
+
-
Power +
Signal +
Signal -
Power -
4-20mA
4
3
2
1
PL14/5
PL13/2
PL13/3
PL14/8
D
C
PRT
B
A
Notes:
1. Specified 7950 pins are for Analogue Input 3.
2. DIP switch position 3 must be set to 4-20mA.
Figure 2.9: DIP-switch and safe area wiring for a
mA-type temperature transmitter
Page 2.9
Chapter 2 Getting started
Turn on the
power
4. Turn on the power to the system. The system goes through a Power On Self Test (POST)
routine which takes less than 30 seconds. When it is finished, ignore any flashing al arm
lights which may appear.
Go to the
wizards menu
5. Press the MENU button to go to Page 1 of the Main Menu (if you aren’t there already).
6. Press the DOWN-ARROW button twice to go to Page 3 of the menu.
7. Press the b-button to select “Configure”.
8. Press the a-button twice to go to the wizards menu.
Clear existing
configuration
(This is optional)
9. Press the b-button then the UP-ARROW or DOWN-ARROW button to scroll through the
option list until “Initialise” is shown.
10. Press the b-button to select “Initialise”.
11. Press the d-button to confirm that you want to lose the current configuration.
12. Wait a few seconds until “initialise” on line 2 of the display changes back to
“Choose option”.
Select the
wizard
13. Press the b-button then the UP-ARROW or DOWN-ARROW button to scroll through the
option list until “Temperature” is shown.
14. Press the b-button to select “Temperature”.
Start of wizard 15. Press the d-button to answer YES to the question “Edit Line Temperature?”.
Choose the
Analogue Input
16. Press the b-button
17. Press the UP-ARROW button until “Analogue input 3” appears.
18. Press the b-button and then the ENTER button to confirm selection of
“Analogue input 3”.
Select the type
19. Press the ENTER button to keep the default selection of a 4-20mA type input.
of Analogue
Input
Set Analogue
0% and 100%
20. Press the b-button.
21. Type in a suitable maximum temperature value and then press the ENTER button.
range
22. Press the ENTER button to move on to the next prompt.
23. Press the b-button.
24. Type in a suitable minimum temperature value and then press the ENTER button.
25 Press the ENTER button to move on to the next prompt.
Make the
Analogue Input
“live”
Skip over the
next few
26. Press the d-button.
27. Press the UP-ARROW button so that “Set” changed “Live”.
28. Press the ENTER button once.
29. Press the c-button several times to answer NO to all questions until the wizard is exited.
questions
View how you
have configured
Line temperature
30. Press the MENU button.
31. Press the d-button and then press the a-button. The display looks similar to that shown
in Figure 2.10 although values shown may vary.
Page 2.10
Chapter 2 Getting started
Figure 2.10: Line temperature data display
End of Worked Example 3
Page 2.11
Chapter 2 Getting started
2.7 Example 4: 7950 with a mA-type pressure transmitter
About this example
This example shows you how to connect a mA-type pressure transmitter to the 7950, and then use the
“Pressure” wizard to configure the system.
In this example, the “Pressure” wizard is used to configure a connection as follows:
• A single pressure transmitter is connected to Analogue Input 3.
Work through the example by following the instructions below. If you are not sure where the buttons are, refer to
Chapter 6.
Connect the
meter
Set DIP switch 3. Ensure that the DIP-switch, inside the 7950, is set as shown in Figure 2.11 .
1. Wire the transmitter to the 7950, as in Figure 2.11.
2. Earth the 7950 to a suitable earth point.
Notes:
(a) You must comply with EMC regulations.
(b) Use a screened multi-core cable that has twisted pairs.
Loop-powered
4-20mA pressure
transmitter
7950
+
-
Power +
Signal +
Signal -
Power -
4-20mA
4
3
2
1
PL14/5
PL13/2
PL13/3
PL14/8
D
C
PRT
B
A
Notes:
1. Specified 7950 pins are for Analogue Input 3.
2. DIP-switch position 3 must be set to 4-20mA.
Figure 2.11: DIP-switch and safe area wiring for a
mA-type pressure transmitter
Turn on the
power
4. Turn on the power to the system. The system goes through a Power On Self Test (POST)
routine which takes less than 30 seconds. When it is finished, ignore any flashing al arm
lights which may appear.
Page 2.12
Go to the
wizards menu
5. Press the MENU button to go to Page 1 of the Main Menu (if you aren’t there already).
6. Press the DOWN-ARROW button twice to go to Page 3 of the menu.
7. Press the b-button to select “Configure”.
8. Press the a-button twice to go to the wizards menu.
Chapter 2 Getting started
Clear existing
configuration
(This is optional)
Select the wizard 13. Press the b-button then the UP-ARROW or DOWN-ARROW button to scroll through
Start of wizard 15. Press the d-button to answer YES to the question “Edit Line Pressure?”.
9. Press the b-button then the UP-ARROW or DOWN-ARROW button to scroll through the
option list until “Initialise” is shown.
10. Press the b-button to select “Initialise”.
11. Press the d-button to confirm that you want to lose the current configuration.
12. Wait a few seconds until “initialise” on line 2 of the display changes back to “Choose
option”.
the option list until “Pressure” is shown.
14. Press the b-button to select “Pressure”.
Choose the
Analogue Input
Set Analogue 0%
and 100% range
23. Type in a suitable minimum pressure value and then press the ENTER button.
Select the type
of Analog Input
Make the
Analogue Input
“live”
Skip over other
questions
View how you
have configured
Line pressure
16. Press the b-button
17. Press the UP-ARROW button until “mA input 3” appears.
18. Press the b-button and then the ENTER button to confirm selection of
“mA input 3”.
19. Press the b-button.
20. Type in a suitable maximum pressure value and then press the ENTER button.
21. Press the ENTER button to move on to the next prompt
22. Press the b-button
24 Press the ENTER button to move on to the next prompt
25. Press the ENTER button to keep the default selection of a 4-20mA type input.
26. Press the d-button
27. Press the UP-ARROW button so that “Set” changed “Live”
28. Press the ENTER button twice
29. Press the c-button several times to answer NO to all questions until the wizard is
exited
30. Press the MENU button.
31. Press the DOWN-ARROW button.
32. Press the a-button twice. The display looks similar to that shown in Figure 2.12
although values shown may vary.
Figure 2.12: Line pressure data display
End of Worked Example 4
Page 2.13
Chapter 2 Getting started
2.8 Example 5: 7950 with a RTD-type temperature transmitter
About this example
This example shows you how to connect a PT100 transmitter to the 7950, and then use the “Temperature”
wizard to configure the system.
In this example, the “Temperature” wizard is used to configure a connection as follows:
• A single PT100 transmitter is connected to Analogue Input 1.
Work through the example by following the instructions below. If you are not sure where the buttons are, refer to
Chapter 6.
Connect the
meter
Set DIP switch 3. Set DIP-switch position 1 to “RTD” (for Analogue input 1).
1. Wire the transmitter to the 7950, as in Figure 2.13.
2. Earth the 7950 to a suitable earth point.
Notes:
(a) You must comply with EMC regulations.
(b) Use a screened multi-core cable that has twisted pairs.
Turn on the
power
PT100 on
Analogue
Input 1
PRT
Power +
Signal +
Signal Power -
4-20mA
4
3
2
1
7950
PL12/1
PL12/2
PL12/3
PL12/4
D
C
PRT
B
A
Notes:
1. Specified 795x pins are for Analogue Input 1.
2. DIP switch position 1 must be set to PRT.
Figure 2.13: DIP switch and safe area wiring for a
RTD-type temperature transmitter
4. Turn on the power to the system. The system goes through a Power On Self Test
(POST) routine which takes less than 30 seconds. When it is finished, ignore any
flashing alarm lights which may appear.
Page 2.14
Go to the
wizards menu
5. Press the MENU button to go to Page 1 of the Main Menu (if you aren’t there already).
6. Press the DOWN-ARROW button twice to go to Page 3 of the menu.
7. Press the b-button to select “Configure”.
8. Press the a-button twice to go to the wizards menu.
Chapter 2 Getting started
Clear existing
configuration
(This is optional)
Select the wizard 13. Press the b-button then the UP-ARROW or DOWN-ARROW button to scroll through
Start of wizard 15. Press the d-button to answer YES to the question “Edit Line Temperature?”.
9. Press the b-button then the UP-ARROW or DOWN-ARROW button to scroll through
the option list until “Initialise” is shown.
10. Press the b-button to select “Initialise”.
11. Press the d-button to confirm that you want to lose the current configuration.
12. Wait a few seconds until “initialise” on line 2 of the display changes back to
“Choose option”.
the option list until “Temperature” is shown.
14. Press the b-button to select “Temperature”.
Choose the
Analogue Input
Choose the type
of Analogue Input
Make the
Analogue Input
Channel “live”
Skip over other
questions
View how you
have configured
Line temperature
16. Press the b-button.
17. Press the UP-ARROW button until “Analogue input 1” appears.
18. Press the b-button and then the ENTER button to confirm selection of
“Analogue input 1”.
19. Press the b-button.
20. Press the UP-ARROW button until “PT100 input” appears on line 2.
21. Press the ENTER button twice to select this RTD-type input.
22. Press the d-button.
23. Press the UP-ARROW button so that “Set” has changed to “Live”.
24. Press the ENTER button twice.
25. Press the c-button several times (to answer NO to all further questions) until the
wizard is exited.
26. Press the MENU button.
27. Press the d-button and then press the a-button. The display looks similar to that
shown in Figure 2.14 although values shown may vary.
Figure 2.14: Line temperature display
End of Worked Example 5
Page 2.15
Chapter 2 Getting started
Page 2.16
3. About the Micro Motion® 7950
3.1 Background
The Micro Motion® 7950 was developed to meet the demand for a reliable, versatile, user-friendl y and costeffective instrument for liquid and gas metering. It has a Motorola 68332 32-bit microprocessor and surfacemounted circuit board components so that it is powerful, reliable and compact.
Features of the 7950 include:
• Simple access to information.
• Comprehensive interrogation facilities.
• Alarm and alarm history facilities.
• A menu-driven, user-friendly interface.
• NEMA 4X, IP65 enclosure.
• ac and dc powered.
• Three serial communication ports (using RS232 or RS485) for MODBUS communications and printing.
These facilities are described in more detail in the rest of this chapter.
Chapter 3 About the Micro Motion® 7950
3.2 What the 7950 Dual Channel Gas Signal Converter does
The 7950 Gas Signal Converter is primarily used to convert signal s from one fo rma t to another. There is a need for
this conversion when a system is unable to accept a raw signal from a transducer or, perhaps, some intermediate
signal processing is required.
A common conversion is where a frequency input from a 7812 gas density transducer can be accepted by a 795x
and then transmitted (by the same 795x) through an analogue output as a 4-20mA signal.
In this application, the 7950 can calcu late :
• Line density
• Specific gravity
• Base density
• Energy (Cv/m) (from AGA-5 method, or mA-type input)
Note: Dual channel measurements are available for Items marked with a
It can also obtain:
• Line temperature.
• Densitometer temperature.
• Density pressure.
• Atmospheric pressure.
• Percentage of CO
• Percentage of N
• Compressibility (from S-GERG, NX19, NX19mod or NX19 3h)
• Special equations 1 and 2.
√
(from Transducer, PTZ method, or mA-type input)
√
(from Transducer, mA-type input, or Base density)
√
(from Specific gravity method, PTZ method, or mA-type input)
√
(from a mA-type input)
2
(from a mA-type input)
2
Page 3.1
Chapter 3 About the Micro Motion® 7950
3.3 Physical description of the 7950
The 7950 is a wall-mounted instrument housed in a one-piece case. The upper part of the instrument has a
panel on which are mounted the keyboard and display. Below this, and stepped back slightly, is a terminal cover
which, when removed allows access to the electrical connectors on the connector board inside the instrument.
All wiring enters the case from underneath, through the gland plate which has to be drille d for the purpose.
The connector board is mounted vertically inside the back of the case. The microprocessor board is attached, by
six screws and stand-offs, to the back of the keyboard and display.
The upper and lower parts of the instrument are separated by a horizontal metal plate (the screen) which helps
to protect the instrument against electro-magnetic interference.
Page 3.2
Figure 3.1 The 7950 and its major assemblies
p
3.4 Communications
The 7950 can operate as a MODBUS slave. It can:
• Download a configuration from a PC, DCS, etc.
• Upload a configuration.
• Monitor random locations in the 7950.
• Interrogate the alarm and data logger buffers.
• Manipulate the alarm and data logger buffers.
• Set random locations with new data.
• Instigate printed reports.
3.5 Typical installations
The diagram below illustrates a typical installation utilising the 7950.
DT/1
Chapter 3 About the Micro Motion® 7950
Part of pipeline
Transducers and transmitters
PT
DT/2
TE
4-wire PRT
PT
DT/n
TE
Static pressure transmitter
Gas density transducer 1 and 2
Temperature element (PRT)
7950
solartron
2 * Frequency
S
instruments
V
V
1
2
89
7
a
6
45
b
12
3
c
+/-
0
d
CLREXP
Alarm
Analogue outputs
4-20mA
Printer
MODBUS communications
to and from host com
Figure 3.2: Typical installation for a Gas Signal Converter system
uter
Page 3.3
Chapter 3 About the Micro Motion® 7950
3.6 Checking your software version
The 7950 is driven by pre-loaded software which differs according to the application for which the instrument is to
be used. To check hardware configuration, see Ordering Information in Appendix C.
PREFIX:
HARDWARE
PLATFORM
50 7950
51 7951
DIGIT 1:
METERED
PRODUCT
1 GAS
2 LIQUID
3 BOTH
4 OTHER
PREFIXDIGIT 1DIGIT 2DIGIT 3DIGIT 4
DIGIT 2:
FLOW
METER
0 NONE
1 OR IF ICE
2 TURBINE/PD
3 VENTURI
4 MASS
5 MULTI
DIGIT 3:
STREAMS/
CHANNELS
1 SINGLE
2 DUAL
3 TRIPLE
4 QUAD
5 1, 2, 3 or 4
DIGIT 4:
SPECIAL
0 – 9
SOFTWARE VERSION NUMBER
Figure 3.3: Software version number
For example, for a 7950 Dual-Channel Gas Signal Converter, the software version number is 501020.
You can find the software version number in two ways:
1. It is printed on a label at the rear panel of the 7950.
2. It is written into the menu structure – see Chapter 12.
Page 3.4
Loading...
+ 148 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.