For online technical support, refer to the EXPERT2™ tool at
www.expert2.com. To speak to a customer service
representative, call the support center nearest you:
• In U.S.A., phone 1-800-522-MASS (1-800-522-6277)
• In Canada and Latin America, phone (303) 530-8400
1.1About this manualThis manual explains how to use the Series 3000 software to configure,
operate, and maintain the Model 3300, 3350, 3500, and 3700
applications platforms.
This manual provides information about the following applications:
•Model 3500 or 3700 transmitter
•Discrete batch control
•Process and inventory totalizers
•Process monitor
•Process comparator
•Security for custody transfer
This manual does not provide information about the Series 3000 Net Oil
Computer or the density application.
•For information about the Net Oil Computer, see the Series 3000 Net Oil Computer Manual.
•For information about the density application, see the Series 3000 Density Application Manual.
This manual does not explain installation or wiring. For information about
installation and wiring, see the Series 3000 Installation Manual.
Person-Process InterfaceConfiguration: InputsConfiguration: System DataBefore You Begin
1.2OrganizationThis manual is organized as follows:
Part 1: Introduction, includes Chapters 1 and 2.
•This Chapter outlines the contents of this manual.
•Chapter 2 explains how to use the Person-Process Interface.
Part 2: Configuration, includes Chapters 3 through 9.
•Chapter 3 explains how to configure system data.
•Chapter 4 explains how to configure inputs.
•Chapter 5 explains how to configure the discrete batch control
application.
•Chapter 6 explains how to configure measurement parameters.
•Chapter 7 explains how to configure outputs.
•Chapter 8 explains how to configure monitoring.
•Chapter 9 explains how to configure digital communication.
Part 3: Security and Language, includes Chapters 10 and 11.
•Chapter 10 explains how to configure security and select the
language for Person-Process Interface displays.
•Chapter 11 explains how to enable security for custody transfer.
Part 4: Operation and Diagnostics, includes Chapters 12 through 14.
•Chapter 12 explains how to use the software in operation mode.
•Chapter 13 explains how to use the diagnostic software.
•Chapter 14 explains how to read inputs and set outputs.
Series 3000 Detailed Setup Manual1
Before You Begincontinued
Part 5: Maintenance, includes Chapters 15 through 18.
•Chapter 15 explains how to use the active alarm log.
•Chapter 16 explains how to monitor and reset totalizers.
•Chapter 17 explains how to perform calibration and trim procedures.
•Chapter 18 explains how to enter meter factors for proving
applications.
1.1AppendixesAppendix A provides software diagrams for all software menus
described in this manual.
Appendix B is the Series 3000 software configuration record. Use it to
record parameters that will be configured as you follow the instructions
in Chapters 3 through 9.
2Series 3000 Detailed Setup Manual
2Person-Process Interface
2.1About this chapterThis chapter explains how to use the security button, function buttons,
and cursor control buttons on the display face.
2.2Person-Process InterfaceFigure 2-1 shows the Person-Process Interface. Use the interface to:
•Configure the application
•Monitor and control the application
•Perform maintenance and diagnostic tasks
Figure 2-1. Person-Process Interface
5←DEVICE 1→2
Mass Flow Rate
Backlit
display
Mass Total
PRINTRESETVIEW
Function buttons
2.33
g/s
485.88
Person-Process InterfaceConfiguration: InputsConfiguration: System DataBefore You Begin
Cursor control
buttons
g
Security button
Security button
Series 3000 Detailed Setup Manual3
The security button is in the lower right of the interface, marked by an
icon of a padlock.
•If security is disabled, press the security button to access the main
menu. See Figure 2-2. When you set up the application for the first
time, security will be disabled.
•If security has been enabled, you will be prompted to enter a
password. See Figure 2-3. To enable security, see pages 63-64.
Person-Process Interfacecontinued
You can use the security button to return to the main menu or password
entry screen. Press the security button once to return to:
•The main menu, shown in Figure 2-2, if security is disabled
•The password entry screen, shown in Figure 2-3, if security is
enabled
At the main menu or password entry screen, press EXIT to return to the
operation screen.
Function buttonsThe pushbuttons below the display are the function buttons. The action
each button performs appears above the button. See Figure 2-4.
Figure 2-4. Function buttons
ALARMS
DEVICE 1
Configuration
Maintenance
Security
Language
SELHELPEXIT
VIEWAccess the view menu
ACKAcknowledge an alarm message
EXITExit to previous menu or cancel a change
NOCancel action
PREVReturn to the previous screen
ABORT• Abort sensor zero
• Abort calibration
HELPShow a help screen
RESUMEResume a batch that has been stopped
RESETReset total
PRINTPrint a ticket
NEXTAdvance to the next screen
STARTStart batch
STOP• Stop batch before target is achieved
• Batch can be resumed
END• End batch before target is achieved
• Batch cannot be resumed
RESETReset total
PAUSEPause counting of all displayed totals
RESUMEResume counting of all displayed totals
SELSelect the highlighted menu item
CHGMake a change to the highlighted menu item
SAVESave a change
ENTEREnter a password
YESProceed with action
PRINTPrint a ticket
Configuration: InputsConfiguration: System DataPerson-Process InterfaceBefore You Begin
Series 3000 Detailed Setup Manual5
Person-Process Interface continued
Using cursor control
buttons
The actions performed by the function buttons apply to the item at the
cursor.
Figure 2-5, page 7, shows a typical configuration sequence involving
both a menu item and a variable. Pressing HELP produces a screen that
has help for the item at the cursor.
Menus
A menu is a list of items.
•The cursor is a reverse-video highlight bar.
•Use the up or down arrow buttons to locate the cursor at the menu
item you want to select or change.
•After locating the cursor at the desired menu item, press CHG or the
right arrow button to select the item.
Vari a bles
After a menu item has been selected, the cursor enables you to enter or
change a variable:
•The cursor appears as a line under a character.
•If the variable has a value of Yes or No, all arrows toggle between the
two choices. Otherwise, press the up and down arrow buttons to
increase or decrease the value of the character at the cursor.
•If the variable has more than one digit or character (like the slug low
limit in the example), press the left and right arrow buttons to move
the cursor to the next or previous character.
•When the variable is correct, press SAVE.
•If you wish to cancel the change, press EXIT before pressing SAVE.
The interface will return to the previous screen without saving the
changes.
Process monitor
In the process monitor, use the left and right arrows to scroll from one
screen to the next or previous screen.
•Press the right arrow (
•Press the left arrow (
•There are five screens.
To assign variables to each process monitor screen, see page 56.
→ ) to scroll to the next screen.
←) to scroll to the previous screen.
2.3Scientific notationScientific notation is used for displaying values that include 10 or more
digits. For example, the value 123,400,000 would be displayed as
1.234+8.
6Series 3000 Detailed Setup Manual
Person-Process Interface continued
Figure 2-5. Cursor control buttons
Menu item
Indicates items
available to scroll
Cursor is a
highlight bar
Variab le
Cursor is an
underscore
Density
Density Units
Density Damping
Slug Low Limit
0.005000 g/cc
Slug High Limit
0.100000 g/cc
CHGHELPEXIT
Density
Density Units
Density Damping
Slug Low Limit
0.005
Slug High Limit
0.100000 g/cc
SAVEHELPEXIT
g/cc
1.7 sec
g/cc
1.7 sec
000 g/cc
EXIT
↓
Move cursor to left
↓
Move cursor up/Scroll up
SELECT
Move cursor down/Scroll down
Person-Process InterfaceConfiguration: InputsConfiguration: System DataBefore You Begin
Increase value at cursor
or toggle YES/NO
Move cursor to right
Decrease value at cursor
or toggle YES/NO
Process monitor
5←DEVICE 1→2
Scroll to previous screen
Mass Flow Rate
2.33
g/s
Mass Total
485.88
g
PRINTRESETVIEW
Series 3000 Detailed Setup Manual7
Scroll to next screen
8Series 3000 Detailed Setup Manual
3System Data
3.1About this chapterThis chapter explains how to configure system data. System data
include all the software parameters listed in Figure 3-1.
Failure to perform configuration tasks in the proper sequence could
result in an incomplete configuration. Perform configuration tasks in the
following sequence:
1. Configure system data.
2. Configure inputs (see Chapter 4).
3. Configure the discrete batch control application, if it is present (see
Chapter 5).
4. Configure measurements (see Chapter 6).
5. Configure outputs (see Chapter 7).
6. Configure monitoring (see Chapter 8).
7. Configure digital communication (see Chapter 9).
CAUTION
Selecting configuration will interrupt measurement
and control functions. All outputs will go to their
configured fault settings.
Person-Process InterfaceConfiguration: InputsConfiguration: System DataBefore You Begin
Set control devices for manual operation before accessing
configuration menus.
3.2Recording system dataWhile you are configuring system data, record the data in the Series
3000 Series 3000 configuration record (Appendix B).
Figure 3-1. System menu
SystemTag
TimeHour
Minute
Second
DateDay
Master resetMonth
Yea r
Series 3000 Detailed Setup Manual9
System Data continued
3.3System dataTo configure system data:
Configuration
System
System
Tag
Time
Date
Master Reset
SELEXIT
1. Press the security button on the display face.
2. Select Configuration.
3. Select System.
4. Use the function buttons and cursor control
buttons to configure the parameters that are listed
in Tabl e 3-1.
Table 3-1. System parameters
VariableDefaultDescription
TagDevice 1• Enter up to 8 digits and/or characters that uniquely identify this platform
TimeCurrent timeEnter 2 digits for hours, 2 digits for minutes, and 2 digits for seconds
DateCurrent dateEnter 4 digits for the year, a character code for the month, and 2 digits for the day
• The tag will appear on operation screens
10Series 3000 Detailed Setup Manual
4Inputs
4.1About this chapterThis chapter explains how to configure inputs. Inputs include all the
software parameters listed in Figure 4-1, page 12.
Failure to perform configuration tasks in the proper sequence could
result in an incomplete configuration. Perform configuration tasks in the
following sequence:
1. Configure system data (see Chapter 3).
2. Configure inputs.
3. Configure the discrete batch control application, if it is present (see
Chapter 5).
4. Configure measurements (see Chapter 6).
5. Configure outputs (see Chapter 7).
6. Configuring monitoring (see Chapter 8).
7. Configure digital communication (see Chapter 9).
CAUTION
Selecting configuration will interrupt
measurement and control functions. All outputs
will go to their configured fault settings.
Person-Process InterfaceConfiguration: InputsConfiguration: System DataBefore You Begin
Set control devices for manual operation before
accessing configuration menus.
4.2Recording inputsWhile you are configuring inputs, record them in the Series 3000
configuration record (Appendix B).
Series 3000 Detailed Setup Manual11
Inputs continued
Figure 4-1. Inputs menu
InputsCoriolisEnable/disableEnable Coriolis
Enable sensor alarms
Alarm timeout
Configure process varFlow variablesFlow damping
DensityDensity units
TemperatureTemperature units
Sensor cal dataT-Series setup
Flow factor
Flowcal temp coef
4
FCF
4
FT
4
FTG
4
FFQ
D1
D2
K1
K2
FD
4
DT
4
DTG
DFQ1
DFQ2
Dens temp coeff
Temperature slope
Temperature offset
Sensor informationSensor model no.
Sensor serial no.
Sensor material
Sensor end connection
Sensor liner
Frequency inputFlow rate units
Scaling methodFrequency = flow
Frequency
Flow
Pulses/unit
Units/pulse
5
5
6
7
Pulses/unit
Units/pulse
K-factor
1
If enable sensor alarms is set to NO.
2
If a sensor is not connected.
3
If an ELITE, BASIS, Model D, Model DL, or Model DT sensor is connected, or if T-Series setup is set to NO.
4
If a T-Series sensor is connected or if T-Series setup is set to YES.
5
If frequency = flow is selected.
6
If pulses/unit is selected.
7
If units/pulse is selected.
1
Meter directionForward
Mass unitsBackward
Mass low flow cutoff
Volume units
Vol low flow cutoff
Density damping
Slug low limit
Slug high limit
Slug time
Temperature damping
2
3
3
4
4
3
12Series 3000 Detailed Setup Manual
Inputs continued
4.3Disabling Coriolis inputs, Coriolis alarms,
and sensor alarms
Configuration
Inputs
Coriolis
Enable/disable
Enable/Disable
Enable Coriolis
YES
Enable Sensor Alarms
YES
Alarm Timeout
1
Coriolis and sensor alarms are enabled as the
default. By disabling Coriolis, you disable all Coriolis
input signals and alarms. Disabling them might be
desirable while you are connecting the sensor, or if
you are using only the frequency input to measure
flow. Disabling sensor alarms disables a subset of
Coriolis alarms to prevent them from driving outputs
to fault levels, stopping internal totalizers, and
stopping a running batch.
To disable Coriolis inputs and Coriolis alarms, or to
disable sensor alarms:
1. Press the security button on the display face.
2. Select Configuration.
3. Select Inputs.
4. Select Coriolis.
5. Select Enable/Disable.
6. Use the function buttons and cursor control
buttons to configure the parameters that are listed
in Tabl e 4-1.
Person-Process InterfaceConfiguration: InputsConfiguration: System DataBefore You Begin
CHGEXIT
Table 4-1. Enabling or disabling inputs and alarms
Note
For more information about alarms, see Chapter 13.
VariableDefaultDescription
Enable CoriolisYesIf set to NO:
Enable sensor alarmsYesIf set to NO, warming up, transmitter failure, density failure, and sensor failure
Alarm timeout1 minute• If enable sensor alarms is set to NO, enter the number of minutes, from 1 to 20, for
• The platform will not use input signals from the sensor to measure flow, density, or
temperature
• The platform will not produce the following alarms: warming up, cal in progress,
drive overrange, temperature overrange, temperature failure, sensor failure,
transmitter failure, density overrange, density failure, mass flow overrange, volume
overrange, calibration failure, calibration complete, calibration aborted, RTD
failure, charize required, slug flow, slug timeout
alarms will be downgraded to informational alarms for the amount of time
configured for the alarm timeout:
• During alarm timeout, outputs will not go to fault levels
• During alarm timeout, sensor alarms will not require acknowledgment
• During alarm timeout, internal totalizers will not stop counting
• During alarm timeout, batches in progress will not stop
which sensor alarms will be disabled
• Sensor alarms will revert to fault alarms after the alarm timeout has ended
Series 3000 Detailed Setup Manual13
Inputs continued
4.4Configure process variablesProcess variables include flow variables, density,
temperature, sensor calibration data, and sensor
information.
Flow variablesTo configure flow variables:
Configuration
Inputs
Coriolis
Config process var
Flow variables
Flow Variables
↓
Flow Damping
0.8 sec
Meter Direction
Forward
Mass Units
g/s
Mass Low Flow Cutoff
0.00000 g/s
CHGEXIT
1. Press the security button on the display face.
2. Select Configuration.
3. Select Inputs.
4. Select Coriolis.
5. Select Config Process Var.
6. Select Flow Variables.
7. Use the function buttons and cursor control
buttons to configure the parameters that are listed
in Tabl e 4-2.
Table 4-2. Flow variables
VariableDefaultDescription
Flow damping0.8 sec• Damping filters out noise or the effects of rapid changes in the flow rate without
Meter directionForward• Select the direction in which process fluid will flow through the sensor relative to
Mass unitsg/s• Select the desired unit of mass flow (see Table 4-4, page 15)
Mass low flow cutoff0.00000 g/s• Enter the mass flow rate below which mass flow outputs and displays will indicate
Volume unitsl/s• Select the desired unit of volume flow (see Tabl e 4- 4, page 15)
Volume low flow cutoff0.00000 l/s• Enter the volume flow rate below which volume flow outputs and displays will
affecting measurement accuracy
• If the platform will operate with a Micro Motion T-Series sensor, the recommended
flow damping value is 0.3 seconds
• Milliamp outputs have their own damping
the flow direction arrow on the sensor
• The sensor can measure forward or backward flow
• For the effect of flow direction on outputs and totalizers, see Ta ble 4-3, page 15
• Mass flow outputs and displays will indicate mass flow in the selected unit
zero flow
• Milliamp outputs have their own mass low flow cutoffs
• Volume flow outputs and displays will indicate volume flow in the selected unit
indicate zero flow
• Milliamp outputs have their own volume low flow cutoffs
14Series 3000 Detailed Setup Manual
Inputs continued
Table 4-3. Effect of flow direction on outputs and totalizers
Fluid flow
directionOutput or totalizer
Fluid flowing in
same direction
as flow arrow
on sensor
Fluid flowing in
opposite direction
from flow arrow
on sensor
4-20 mA outputOutput increases as flow
Frequency outputOutput increases as flow
Totalizer configured for forward flowTotals increaseTotals remain constant
Totalizer configured for reverse flowTotals remain constantTotals increase
Totalizer configured for absolute value
forward/reverse
Totalizer configured for subtractive forward/reverseTotals increaseTotals decrease
4-20 mA outputOutput goes to 2 mAOutput increases as flow
Frequency outputOutput remains at 0 HzOutput increases as flow
Totalizer configured for forward flowTotals remain constantTotals increase
Totalizer configured for reverse flowTotals increaseTotals remain constant
Totalizer configured for absolute value
forward/reverse
Totalizer configured for subtractive forward/reverseTotals decreaseTotals increase
Short tons (2000 lb)/minuteSTon/minImperial gallons/dayUKgpd
Short tons (2000 lb)/hourSTon/hrMillion gallons/dayMilGal/day
Short tons (2000 lb)/daySTon/dayLiters/secondl/sec
Long tons (2240 lb)/minuteLTon/minLiters/minutel/min
Long tons (2240 lb)/hourLTon/hrLiters/hourl/hr
Long tons (2240 lb)/dayLTon/dayMilliliters/dayMilL/day
Ounces/secondoz/sBarrels/secondbbl/s
Ounces/minuteoz/minBarrels/minutebbl/min
Ounces/houroz/hrBarrels/hourbbl/hr
Barrels/daybbl/day
Fluid ounces/secondFloz/s
Fluid ounces/minuteFloz/min
Fluid ounces/hourFloz/hr
Person-Process InterfaceConfiguration: InputsConfiguration: System DataBefore You Begin
Series 3000 Detailed Setup Manual15
Inputs continued
Density inputsTo configure density inputs:
Configuration
Inputs
Coriolis
Config process var
Density
Density
↓
Density Units
g/cc
Density Damping
1.7 sec
Slug Low Limit
0.000000 g/cc
Slug High Limit
5.000000 g/cc
CHGEXIT
1. Press the security button on the display face.
2. Select Configuration.
3. Select Inputs.
4. Select Coriolis.
5. Select Config Process Var.
6. Select Density.
7. Use the function buttons and cursor control
buttons to configure the parameters that are listed
in Tabl e 4-5.
Table 4-5.Density inputs
VariableDefaultDescription
Density unitsg/cc• Select the desired unit of density (see Tab le 4-6 )
Density damping1.7 sec• Damping filters out noise or the effects of rapid changes in density without affecting
Slug low limit0.000000 g/cc• Enter the desired low limit, in g/cc, for the process density
Slug high limit5.000000 g/cc• Enter the desired high limit, in g/cc, for the process density
Slug time1.0 sec• Enter the number of seconds for which flow outputs will hold their last measured flow
• Density outputs and displays will indicate density in the selected unit
measurement accuracy
• If the platform will operate with a Micro Motion T-Series sensor, the recommended
density damping value is 0.3 seconds
• Milliamp outputs have their own damping
• The entered value is the density below which a slug flow alarm will be generated
• For more information about slug flow, see page 98
• The entered value is the density above which a slug flow alarm will be generated
• For more information about slug flow, see page 98
rate while density is outside the range specified by the slug low limit and slug high limit
• The maximum slug time is 1200 seconds
• If a value of 0.0 is entered, as soon as slug flow is detected, flow outputs will go to the
level that indicates zero flow
• For more information about slug time, see page 98
16Series 3000 Detailed Setup Manual
Inputs continued
Table 4-6. Density units
UnitSoftware label
Grams/cubic centimeterg/cc
Kilograms/cubic meterkg/cum
Pounds/gallonlb/gal
Pounds/cubic footlb/cuft
Grams/milliliterg/mL
Kilograms/literkg/L
Grams/literg/L
Pounds/cubic inchlb/CuIn
Short tons (2000 lb)/cubic yardSTon/CuYd
TemperatureTo configure temperature inputs:
Configuration
Inputs
Coriolis
Config process var
Temperature
Temperature
Temperature Units
degC
Temperature Damping
3.5 sec
CHGEXIT
1. Press the security button on the display face.
2. Select Configuration.
3. Select Inputs.
4. Select Coriolis.
5. Select Config Process Var.
6. Select Temperature.
7. Use the function buttons and cursor control
buttons to configure the parameters that are listed
in Tabl e 4-7.
Table 4-7. Temperature inputs
Person-Process InterfaceConfiguration: InputsConfiguration: System DataBefore You Begin
VariableDefaultDescription
Temperature unitsdegC• Select °Celsius, °Fahrenheit, °Rankine, or Kelvin
Temperature damping3.5 sec• Damping filters out noise or the effects of rapid changes in temperature without
Series 3000 Detailed Setup Manual17
• Temperature outputs and displays will indicate temperature in the selected unit
affecting measurement accuracy
• If the platform will operate with a Micro Motion T-Series sensor, the recommended
temperature damping value is 0.0 seconds
• Milliamp outputs have their own damping
Inputs continued
4.5Sensor calibration dataSensor calibration data describe the sensor’s
Configuration
Inputs
Coriolis
Sensor cal data
Sensor Cal Data
↓
Flow Factor
1.00000
Flocal Temp Coef
5.130
D1
0.000000
D2
1.000000
CHGEXIT
Sensor Cal Data
↓
T-Series Setup
NO
Flow Factor
1.00000
Flowcal Temp Coef
5.130
D1
0.000000
CHGEXIT
sensitivity to flow, density, and temperature.
To configure sensor calibration data:
1. Press the security button on the display face.
2. Select Configuration.
3. Select Inputs.
4. Select Coriolis.
5. Select Sensor Cal Data.
6. If the applications platform is connected to a
sensor, skip to step 7. If the applications platform
is not connected to a sensor, select T-Series
Setup, then:
• Select Yes to enter calibration data for a Micro
Motion T-Series sensor (see page 19), or
• Select No to enter calibration data for an
ELITE, BASIS, Model D, Model DL, or Model
DT sensor (see pages 20-26).
7. Use the function buttons and cursor control
buttons to configure sensor calibration data.
• Sensor cal data should be entered from the
sensor serial number tag or factory calibration
certificate.
• Tags and certificates vary in appearance,
depending on the sensor model number and
manufacturing date.
18Series 3000 Detailed Setup Manual
Inputs continued
Calibration data for Micro Motion T-Series
sensors
If the applications platform is connected to a Micro
Motion T-Series sensor, or if YES was selected at
step 6, page 18, the Person-Process Interface
enables configuration of calibration data for a
T-Series sensor.
Sensor Cal Data
T-Series Setup
FCF
FT
FTG
0.000000
CHGEXIT
YES
1.00000
5.130
↓
Flow calibration values include the FCF, FT, FTG,
and FFQ. Enter the flow calibration values that
appear on the sensor tag. See Figure 4-2.
Density calibration values include D1, D2, K1, K2,
FD, DT, DTG, DFQ1, and DFQ2. Enter the density
calibration values that appear on the sensor tag. See
Figure 4-2.
Temperature calibration values include the
temperature slope and the temperature offset. To
enter temperature calibration values, see page 26.
Figure 4-2. Sensor calibration data on Micro Motion T-Series sensor tag
Flow calibration values
Person-Process InterfaceConfiguration: InputsConfiguration: System DataBefore You Begin
FCF
FTG
Density calibration values
D1
D2
DT
DTG
FT
FFQ
K1
K2
FD
DFQ1DFQ2
Series 3000 Detailed Setup Manual19
Inputs continued
Calibration data for ELITE®, BASIS®,
Model D, Model DL, or Model DT sensors
Sensor Cal Data
↓
T-Series Setup
NO
Flow Factor
1.00000
Flowcal Temp Coef
5.130
D1
0.000000
CHGEXIT
If the applications platform is connected to an ELITE,
BASIS, Model D, Model DL, or Model DT sensor, or if
NO was selected at step 6, page 18, the PersonProcess Interface enables configuration of calibration
data for the appropriate sensor.
Flow calibration values include the flow factor and
the flow calibration temperature coefficient. To
configure flow calibration values, see page 20.
Density calibration values include D1 and D2
density values, K1 and K2 tube periods, the flowing
density correction factor, and the density calibration
temperature coefficient. To configure density
calibration values, see pages 21-25.
Temperature calibration values include the
temperature slope and the temperature offset. To
configure temperature calibration values, see
page 26.
Flow calibration values
Flow calibration values include the flow factor and the
flow calibration temperature coefficient (flowcal temp
coef). To configure flow calibration values, see
Table 4-8 and Figure 4-3.
Table 4-8. Flow calibration values
VariableDefaultDescription
Flow factor1.00000 g/sec• Enter the first 5 digits of the flow cal factor (see Figure 4-3)
• The entered value is the flow rate, in g/sec, that generates 1
velocity signals from the sensor
Flowcal temp coef5.130• Enter the last 3 digits of the flow cal factor (see Figure 4-3)
• The entered value represents the percent change in the measured flow rate per 100°C
change in temperature
µsec of time shift between
20Series 3000 Detailed Setup Manual
Inputs continued
Figure 4-3. Flow calibration values on sensor serial number tag
Flow factor on newer tagFlow factor on older tag
19.0005.13
Flocal temp coef on newer tagFlocal temp coef on older tag
19.0005.13
Person-Process InterfaceConfiguration: InputsConfiguration: System DataBefore You Begin
19.0005.13
19.0005.13
Density calibration values
Density calibration values include D1 and D2 density values, K1 and K2
tube periods, the flowing density correction factor (FD), and the density
calibration temperature coefficient (dens temp coeff).
•To configure D1 and D2, see Ta ble 4-9 and Figure 4-4, page 22.
•To configure K1 and K2, see Ta ble 4-10 and Figure 4-5, page 23.
•To configure FD and the dens temp coeff, see Table 4-1 1 and
Figure 4-6, page 24.
Table 4-9. D1 and D2 density values
VariableDefaultDescription
D10.000000 g/cc• If the sensor tag shows a D1 value, enter the D1 value (see Figure 4-4)
D21.000000 g/cc• If the sensor tag shows a D2 value, enter the D2 value (see Figure 4-4)
• If the sensor tag does not show a D1 value, enter the Dens A or D1 value from the calibration
certificate
• The entered value is the line-condition density of the low-density calibration fluid (Micro
Motion uses air)
• If the sensor tag does not show a D2 value, enter the Dens B or D2 value from the calibration
certificate
• The entered value is the line-condition density of the high-density calibration fluid (Micro
Motion uses water)
Series 3000 Detailed Setup Manual21
Inputs continued
Figure 4-4. D1 and D2 on sensor serial number tag
D1 on newer tagD2 on newer tag
0.0010
0.9980
Table 4-10. K1 and K2 tube period values
VariableDefaultDescription
K1500.000• If the sensor tag shows a K1 value, enter the K1 value (see Figure 4-5)
K250000.000• If the sensor tag shows a K2 value, enter the K2 value (see Figure 4-5)
• If the sensor tag does not show a K1 value, enter the first 5 digits of the density calibration factor
(see Figure 4-5)
• The entered value represents the sensor flow tube period associated with D1, adjusted to 0°C
• If the sensor tag does not show a K2 value, enter the second 5 digits of the density calibration factor
(see Figure 4-5)
• The entered value represents the sensor flow tube period associated with D2, adjusted to 0°C
22Series 3000 Detailed Setup Manual
Loading...
+ 154 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.