Micro Motion Manual: Series 3000 | Micro Motion Manuals & Guides

Instruction Manual
P/N 3300992, Rev. C November 2003
Micro Motion
®
Series 3000
Detailed Setup Manual
Micro Motion
TM
Micro Motion
®
Series 3000
Detailed Setup Manual
For online technical support, refer to the EXPERT2™ tool at www.expert2.com. To speak to a customer service representative, call the support center nearest you:
• In U.S.A., phone 1-800-522-MASS (1-800-522-6277)
• In Canada and Latin America, phone (303) 530-8400
• In Asia, phone (65) 6770-8155
• In the U.K., phone 0800 - 966 180 (toll-free)
• Outside the U.K., phone +31 (0) 318 495 670
©2003, Micro Motion, Inc. All rights reserved. Micro Motion is a registered trademark of Micro Motion, Inc. The Micro Motion and Emerson logos are trademarks of Emerson Electric Co. All other trademarks are property of their respective owners.
Contents
1 Before You Begin . . . . . . . . . . . . . . . . . . . . . . 1
1.1 About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Organization. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Appendixes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 Person-Process Interface . . . . . . . . . . . . . . . . . 3
2.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Person-Process Interface. . . . . . . . . . . . . . . . . . . . . . . . . 3
2.3 Scientific notation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3 System Data. . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Recording system data . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.3 System data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4 Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2 Recording inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.3 Disabling Coriolis inputs, Coriolis alarms, and
sensor alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.4 Configure process variables. . . . . . . . . . . . . . . . . . . . . . 14
4.5 Sensor calibration data . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.6 Sensor information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.7 Frequency input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5 Discrete Batch . . . . . . . . . . . . . . . . . . . . . . . 29
5.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.2 Recording discrete batch parameters . . . . . . . . . . . . . . 29
5.3 Flow source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.4 Control options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.5 Configure presets. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.6 Discrete inputs or discrete events . . . . . . . . . . . . . . . . . 37
6 Measurements . . . . . . . . . . . . . . . . . . . . . . . 39
6.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.2 Recording measurement parameters . . . . . . . . . . . . . . 39
6.3 Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.4 Process comparator . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Series 3000 Detailed Setup Manual i
Contents continued
7 Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.2 Recording outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.3 Discrete outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.4 Milliamp outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.5 Frequency output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8 Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.2 Recording monitoring data. . . . . . . . . . . . . . . . . . . . . . . 55
8.3 Process monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
9 Digital Communication . . . . . . . . . . . . . . . . . 57
9.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
9.2 Recording printer settings . . . . . . . . . . . . . . . . . . . . . . . 57
9.3 Configuring the printer setup . . . . . . . . . . . . . . . . . . . . . 58
9.4 Weights and measures ticket. . . . . . . . . . . . . . . . . . . . . 61
9.5 Printer test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
10 Passwords and Language . . . . . . . . . . . . . . . 63
10.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
10.2 Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
10.3 Security for weights and measures . . . . . . . . . . . . . . . . 65
10.4 Language. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
11 Custody Transfer. . . . . . . . . . . . . . . . . . . . . . 67
11.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.2 Custody transfer configuration procedure . . . . . . . . . . . 67
11.3 Security breach . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
12 Operation Mode . . . . . . . . . . . . . . . . . . . . . . 81
12.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
12.2 Startup and display test . . . . . . . . . . . . . . . . . . . . . . . . . 81
12.3 Sensor zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
12.4 Default operation mode . . . . . . . . . . . . . . . . . . . . . . . . . 84
12.5 Operation mode for discrete batch control. . . . . . . . . . . 85
12.6 Using the view menu . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
13 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
13.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
13.2 Alarm messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
13.3 Active alarm log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
13.4 Customer service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
ii Series 3000 Detailed Setup Manual
Contents continued
14 Diagnostics Menu . . . . . . . . . . . . . . . . . . . . 111
14.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
14.2 Reading inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
14.3 Setting outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
15 Active Alarm Log . . . . . . . . . . . . . . . . . . . . 115
15.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
15.2 Active alarm log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
15.3 For more information about alarms . . . . . . . . . . . . . . . 116
16 Totalizers. . . . . . . . . . . . . . . . . . . . . . . . . . 117
16.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
16.2 Configuring totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . 117
16.3 Batch inventory totalizers . . . . . . . . . . . . . . . . . . . . . . . 118
16.4 Process inventory totalizers . . . . . . . . . . . . . . . . . . . . . 118
17 Calibration and Trim . . . . . . . . . . . . . . . . . . 121
17.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
17.2 Necessary versus optional calibration and
trim procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
17.3 Sensor zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
17.4 Density calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
17.5 Milliamp output trim . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
17.6 Batch AOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
17.7 Temperature calibration . . . . . . . . . . . . . . . . . . . . . . . . 133
17.8 Viewing current data for calibrations . . . . . . . . . . . . . . 135
18 Meter Factors. . . . . . . . . . . . . . . . . . . . . . . 137
18.1 About this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
18.2 Meter factors and measurements . . . . . . . . . . . . . . . . 137
18.3 Proving factors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
18.4 Volume method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
18.5 Multivariable method . . . . . . . . . . . . . . . . . . . . . . . . . . 140
18.6 Viewing current data for meter factors . . . . . . . . . . . . . 150
18.7 Resetting meter factors and proving factors. . . . . . . . . 150
Appendixes
Software Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Series 3000 Software Configuration Record. . . . . . . . . . . . . . . 159
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
Series 3000 Detailed Setup Manual iii
Series 3000 Detailed Setup Manual iv

1 Before You Begin

1.1 About this manual This manual explains how to use the Series 3000 software to configure,

operate, and maintain the Model 3300, 3350, 3500, and 3700 applications platforms.
This manual provides information about the following applications:
Model 3500 or 3700 transmitter
Discrete batch control
Process and inventory totalizers
Process monitor
Process comparator
Security for custody transfer
This manual does not provide information about the Series 3000 Net Oil Computer or the density application.
For information about the Net Oil Computer, see the Series 3000 Net Oil Computer Manual.
For information about the density application, see the Series 3000 Density Application Manual.
This manual does not explain installation or wiring. For information about installation and wiring, see the Series 3000 Installation Manual.
Person-Process Interface Configuration: InputsConfiguration: System DataBefore You Begin

1.2 Organization This manual is organized as follows:

Part 1: Introduction, includes Chapters 1 and 2.
This Chapter outlines the contents of this manual.
Chapter 2 explains how to use the Person-Process Interface.
Part 2: Configuration, includes Chapters 3 through 9.
Chapter 3 explains how to configure system data.
Chapter 4 explains how to configure inputs.
Chapter 5 explains how to configure the discrete batch control application.
Chapter 6 explains how to configure measurement parameters.
Chapter 7 explains how to configure outputs.
Chapter 8 explains how to configure monitoring.
Chapter 9 explains how to configure digital communication.
Part 3: Security and Language, includes Chapters 10 and 11.
Chapter 10 explains how to configure security and select the language for Person-Process Interface displays.
Chapter 11 explains how to enable security for custody transfer.
Part 4: Operation and Diagnostics, includes Chapters 12 through 14.
Chapter 12 explains how to use the software in operation mode.
Chapter 13 explains how to use the diagnostic software.
Chapter 14 explains how to read inputs and set outputs.
Series 3000 Detailed Setup Manual 1
Before You Begin continued
Part 5: Maintenance, includes Chapters 15 through 18.
Chapter 15 explains how to use the active alarm log.
Chapter 16 explains how to monitor and reset totalizers.
Chapter 17 explains how to perform calibration and trim procedures.
Chapter 18 explains how to enter meter factors for proving applications.

1.1 Appendixes Appendix A provides software diagrams for all software menus

described in this manual.
Appendix B is the Series 3000 software configuration record. Use it to record parameters that will be configured as you follow the instructions in Chapters 3 through 9.
2 Series 3000 Detailed Setup Manual

2 Person-Process Interface

2.1 About this chapter This chapter explains how to use the security button, function buttons,

and cursor control buttons on the display face.

2.2 Person-Process Interface Figure 2-1 shows the Person-Process Interface. Use the interface to:

Configure the application
Monitor and control the application
Perform maintenance and diagnostic tasks
Figure 2-1. Person-Process Interface
5 DEVICE 1 2
Mass Flow Rate
Backlit
display
Mass Total
PRINT RESET VIEW
Function buttons
2.33
g/s
485.88
Person-Process Interface Configuration: InputsConfiguration: System DataBefore You Begin
Cursor control buttons
g

Security button

Security button
Series 3000 Detailed Setup Manual 3
The security button is in the lower right of the interface, marked by an icon of a padlock.
If security is disabled, press the security button to access the main menu. See Figure 2-2. When you set up the application for the first time, security will be disabled.
If security has been enabled, you will be prompted to enter a password. See Figure 2-3. To enable security, see pages 63-64.
Person-Process Interface continued
You can use the security button to return to the main menu or password entry screen. Press the security button once to return to:
The main menu, shown in Figure 2-2, if security is disabled
The password entry screen, shown in Figure 2-3, if security is enabled
At the main menu or password entry screen, press EXIT to return to the operation screen.
Figure 2-2. Pressing security button, security disabled
5 DEVICE 1 2
Mass Flow Rate
2.33
g/s
Mass Total
485.88 g
PRINT RESET VIEW
Figure 2-3. Pressing security button, security enabled
5 DEVICE 1 2
Mass Flow Rate
2.33
g/s
Mass Total
485.88 g
PRINT RESET VIEW
DEVICE 1
Configuration
Maintenance Security Language
SEL EXIT
Enter Password
SEL HELP EXIT
4 Series 3000 Detailed Setup Manual
Person-Process Interface continued

Function buttons The pushbuttons below the display are the function buttons. The action

each button performs appears above the button. See Figure 2-4.
Figure 2-4. Function buttons
ALARMS
DEVICE 1
Configuration
Maintenance Security Language
SEL HELP EXIT
VIEW Access the view menu
ACK Acknowledge an alarm message
EXIT Exit to previous menu or cancel a change
NO Cancel action
PREV Return to the previous screen
ABORT • Abort sensor zero
• Abort calibration
HELP Show a help screen
RESUME Resume a batch that has been stopped
RESET Reset total
PRINT Print a ticket
NEXT Advance to the next screen
START Start batch
STOP • Stop batch before target is achieved
• Batch can be resumed
END • End batch before target is achieved
• Batch cannot be resumed
RESET Reset total
PAUSE Pause counting of all displayed totals
RESUME Resume counting of all displayed totals
SEL Select the highlighted menu item
CHG Make a change to the highlighted menu item
SAVE Save a change
ENTER Enter a password
YES Proceed with action
PRINT Print a ticket
Configuration: InputsConfiguration: System DataPerson-Process InterfaceBefore You Begin
Series 3000 Detailed Setup Manual 5
Person-Process Interface continued

Using cursor control buttons

The actions performed by the function buttons apply to the item at the cursor.
Figure 2-5, page 7, shows a typical configuration sequence involving both a menu item and a variable. Pressing HELP produces a screen that has help for the item at the cursor.
Menus
A menu is a list of items.
The cursor is a reverse-video highlight bar.
Use the up or down arrow buttons to locate the cursor at the menu item you want to select or change.
After locating the cursor at the desired menu item, press CHG or the right arrow button to select the item.
Vari a bles
After a menu item has been selected, the cursor enables you to enter or change a variable:
The cursor appears as a line under a character.
If the variable has a value of Yes or No, all arrows toggle between the two choices. Otherwise, press the up and down arrow buttons to increase or decrease the value of the character at the cursor.
If the variable has more than one digit or character (like the slug low limit in the example), press the left and right arrow buttons to move the cursor to the next or previous character.
When the variable is correct, press SAVE.
If you wish to cancel the change, press EXIT before pressing SAVE. The interface will return to the previous screen without saving the changes.
Process monitor
In the process monitor, use the left and right arrows to scroll from one screen to the next or previous screen.
Press the right arrow (
Press the left arrow (
There are five screens.
To assign variables to each process monitor screen, see page 56.
) to scroll to the next screen.
) to scroll to the previous screen.

2.3 Scientific notation Scientific notation is used for displaying values that include 10 or more

digits. For example, the value 123,400,000 would be displayed as
1.234+8.
6 Series 3000 Detailed Setup Manual
Person-Process Interface continued
Figure 2-5. Cursor control buttons
Menu item
Indicates items
available to scroll
Cursor is a
highlight bar
Variab le
Cursor is an
underscore
Density
Density Units
Density Damping
Slug Low Limit
0.005000 g/cc
Slug High Limit
0.100000 g/cc
CHG HELP EXIT
Density
Density Units
Density Damping
Slug Low Limit
0.005
Slug High Limit
0.100000 g/cc
SAVE HELP EXIT
g/cc
1.7 sec
g/cc
1.7 sec
000 g/cc
EXIT
Move cursor to left
Move cursor up/Scroll up
SELECT
Move cursor down/Scroll down
Person-Process Interface Configuration: InputsConfiguration: System DataBefore You Begin
Increase value at cursor or toggle YES/NO
Move cursor to right
Decrease value at cursor or toggle YES/NO
Process monitor
5 DEVICE 1 2
Scroll to previous screen
Mass Flow Rate
2.33
g/s
Mass Total
485.88 g
PRINT RESET VIEW
Series 3000 Detailed Setup Manual 7
Scroll to next screen
8 Series 3000 Detailed Setup Manual

3 System Data

3.1 About this chapter This chapter explains how to configure system data. System data

include all the software parameters listed in Figure 3-1.
Failure to perform configuration tasks in the proper sequence could result in an incomplete configuration. Perform configuration tasks in the following sequence:
1. Configure system data.
2. Configure inputs (see Chapter 4).
3. Configure the discrete batch control application, if it is present (see Chapter 5).
4. Configure measurements (see Chapter 6).
5. Configure outputs (see Chapter 7).
6. Configure monitoring (see Chapter 8).
7. Configure digital communication (see Chapter 9).
CAUTION
Selecting configuration will interrupt measurement and control functions. All outputs will go to their configured fault settings.
Person-Process Interface Configuration: InputsConfiguration: System DataBefore You Begin
Set control devices for manual operation before accessing configuration menus.

3.2 Recording system data While you are configuring system data, record the data in the Series

3000 Series 3000 configuration record (Appendix B).
Figure 3-1. System menu
System Tag
Time Hour
Minute
Second
Date Day
Master reset Month
Yea r
Series 3000 Detailed Setup Manual 9

System Data continued

3.3 System data To configure system data:
Configuration
System
System
Tag
Time Date Master Reset
SEL EXIT
1. Press the security button on the display face.
2. Select Configuration.
3. Select System.
4. Use the function buttons and cursor control buttons to configure the parameters that are listed
in Tabl e 3-1.
Table 3-1. System parameters
Variable Default Description
Tag Device 1 • Enter up to 8 digits and/or characters that uniquely identify this platform
Time Current time Enter 2 digits for hours, 2 digits for minutes, and 2 digits for seconds
Date Current date Enter 4 digits for the year, a character code for the month, and 2 digits for the day
• The tag will appear on operation screens
10 Series 3000 Detailed Setup Manual

4 Inputs

4.1 About this chapter This chapter explains how to configure inputs. Inputs include all the

software parameters listed in Figure 4-1, page 12.
Failure to perform configuration tasks in the proper sequence could result in an incomplete configuration. Perform configuration tasks in the following sequence:
1. Configure system data (see Chapter 3).
2. Configure inputs.
3. Configure the discrete batch control application, if it is present (see Chapter 5).
4. Configure measurements (see Chapter 6).
5. Configure outputs (see Chapter 7).
6. Configuring monitoring (see Chapter 8).
7. Configure digital communication (see Chapter 9).
CAUTION
Selecting configuration will interrupt measurement and control functions. All outputs will go to their configured fault settings.
Person-Process Interface Configuration: InputsConfiguration: System DataBefore You Begin
Set control devices for manual operation before accessing configuration menus.

4.2 Recording inputs While you are configuring inputs, record them in the Series 3000

configuration record (Appendix B).
Series 3000 Detailed Setup Manual 11
Inputs continued
Figure 4-1. Inputs menu
Inputs Coriolis Enable/disable Enable Coriolis
Enable sensor alarms
Alarm timeout
Configure process var Flow variables Flow damping
Density Density units
Temperature Temperature units
Sensor cal data T-Series setup
Flow factor
Flowcal temp coef
4
FCF
4
FT
4
FTG
4
FFQ
D1
D2
K1
K2
FD
4
DT
4
DTG
DFQ1
DFQ2
Dens temp coeff
Temperature slope
Temperature offset
Sensor information Sensor model no.
Sensor serial no.
Sensor material
Sensor end connection
Sensor liner
Frequency input Flow rate units
Scaling method Frequency = flow
Frequency
Flow
Pulses/unit
Units/pulse
5
5
6
7
Pulses/unit
Units/pulse
K-factor
1
If enable sensor alarms is set to NO.
2
If a sensor is not connected.
3
If an ELITE, BASIS, Model D, Model DL, or Model DT sensor is connected, or if T-Series setup is set to NO.
4
If a T-Series sensor is connected or if T-Series setup is set to YES.
5
If frequency = flow is selected.
6
If pulses/unit is selected.
7
If units/pulse is selected.
1
Meter direction Forward
Mass units Backward
Mass low flow cutoff
Volume units
Vol low flow cutoff
Density damping
Slug low limit
Slug high limit
Slug time
Temperature damping
2
3
3
4
4
3
12 Series 3000 Detailed Setup Manual
Inputs continued

4.3 Disabling Coriolis inputs, Coriolis alarms, and sensor alarms

Configuration
Inputs
Coriolis
Enable/disable
Enable/Disable
Enable Coriolis
YES
Enable Sensor Alarms
YES
Alarm Timeout
1
Coriolis and sensor alarms are enabled as the default. By disabling Coriolis, you disable all Coriolis input signals and alarms. Disabling them might be desirable while you are connecting the sensor, or if you are using only the frequency input to measure flow. Disabling sensor alarms disables a subset of Coriolis alarms to prevent them from driving outputs to fault levels, stopping internal totalizers, and stopping a running batch.
To disable Coriolis inputs and Coriolis alarms, or to disable sensor alarms:
1. Press the security button on the display face.
2. Select Configuration.
3. Select Inputs.
4. Select Coriolis.
5. Select Enable/Disable.
6. Use the function buttons and cursor control buttons to configure the parameters that are listed
in Tabl e 4-1.
Person-Process Interface Configuration: InputsConfiguration: System DataBefore You Begin
CHG EXIT
Table 4-1. Enabling or disabling inputs and alarms
Note
For more information about alarms, see Chapter 13.
Variable Default Description
Enable Coriolis Yes If set to NO:
Enable sensor alarms Yes If set to NO, warming up, transmitter failure, density failure, and sensor failure
Alarm timeout 1 minute • If enable sensor alarms is set to NO, enter the number of minutes, from 1 to 20, for
• The platform will not use input signals from the sensor to measure flow, density, or temperature
• The platform will not produce the following alarms: warming up, cal in progress, drive overrange, temperature overrange, temperature failure, sensor failure, transmitter failure, density overrange, density failure, mass flow overrange, volume overrange, calibration failure, calibration complete, calibration aborted, RTD failure, charize required, slug flow, slug timeout
alarms will be downgraded to informational alarms for the amount of time configured for the alarm timeout:
• During alarm timeout, outputs will not go to fault levels
• During alarm timeout, sensor alarms will not require acknowledgment
• During alarm timeout, internal totalizers will not stop counting
• During alarm timeout, batches in progress will not stop
which sensor alarms will be disabled
• Sensor alarms will revert to fault alarms after the alarm timeout has ended
Series 3000 Detailed Setup Manual 13
Inputs continued

4.4 Configure process variables Process variables include flow variables, density,

temperature, sensor calibration data, and sensor information.
Flow variables To configure flow variables:
Configuration
Inputs
Coriolis
Config process var

Flow variables

Flow Variables
Flow Damping
0.8 sec
Meter Direction
Forward
Mass Units
g/s
Mass Low Flow Cutoff
0.00000 g/s
CHG EXIT
1. Press the security button on the display face.
2. Select Configuration.
3. Select Inputs.
4. Select Coriolis.
5. Select Config Process Var.
6. Select Flow Variables.
7. Use the function buttons and cursor control buttons to configure the parameters that are listed in Tabl e 4-2.
Table 4-2. Flow variables
Variable Default Description
Flow damping 0.8 sec • Damping filters out noise or the effects of rapid changes in the flow rate without
Meter direction Forward • Select the direction in which process fluid will flow through the sensor relative to
Mass units g/s • Select the desired unit of mass flow (see Table 4-4, page 15)
Mass low flow cutoff 0.00000 g/s • Enter the mass flow rate below which mass flow outputs and displays will indicate
Volume units l/s • Select the desired unit of volume flow (see Tabl e 4- 4, page 15)
Volume low flow cutoff 0.00000 l/s • Enter the volume flow rate below which volume flow outputs and displays will
affecting measurement accuracy
• If the platform will operate with a Micro Motion T-Series sensor, the recommended flow damping value is 0.3 seconds
• Milliamp outputs have their own damping
the flow direction arrow on the sensor
• The sensor can measure forward or backward flow
• For the effect of flow direction on outputs and totalizers, see Ta ble 4-3, page 15
• Mass flow outputs and displays will indicate mass flow in the selected unit
zero flow
• Milliamp outputs have their own mass low flow cutoffs
• Volume flow outputs and displays will indicate volume flow in the selected unit
indicate zero flow
• Milliamp outputs have their own volume low flow cutoffs
14 Series 3000 Detailed Setup Manual
Inputs continued
Table 4-3. Effect of flow direction on outputs and totalizers
Fluid flow direction Output or totalizer
Fluid flowing in same direction as flow arrow on sensor
Fluid flowing in opposite direction from flow arrow on sensor
4-20 mA output Output increases as flow
Frequency output Output increases as flow
Totalizer configured for forward flow Totals increase Totals remain constant
Totalizer configured for reverse flow Totals remain constant Totals increase
Totalizer configured for absolute value forward/reverse
Totalizer configured for subtractive forward/reverse Totals increase Totals decrease
4-20 mA output Output goes to 2 mA Output increases as flow
Frequency output Output remains at 0 Hz Output increases as flow
Totalizer configured for forward flow Totals remain constant Totals increase
Totalizer configured for reverse flow Totals increase Totals remain constant
Totalizer configured for absolute value forward/reverse
Totalizer configured for subtractive forward/reverse Totals decrease Totals increase
Platform configuration for meter direction
Forward Backward
Output goes to 2 mA
rate increases
Output remains at 0 Hz
rate increases
Totals increase Totals increase
rate increases
rate increases
Totals increase Totals increase
Table 4-4. Mass and volume flow units
Mass flow units Volume flow units
Unit Software label Unit Software label
Grams/second g/s Cubic feet/second cuft/s
Grams/minute g/min Cubic feet/minute cuft/min
Grams/hour g/hr Cubic feet/hour cuft/hr
Kilograms/second kg/s Cubic feet/day cuft/day
Kilograms/minute kg/min Cubic meters/second cu m/s
Kilograms/hour kg/hr Cubic meters/minute cu m/min
Kilograms/day kg/day Cubic meters/hour cu m/hr
Metric tons (1000 kg)/minute t/min Cubic meters/day cu m/day
Metric tons (1000 kg)/hour t/hr U.S. gallons/second USgps
Metric tons (1000 kg)/day t/day U.S. gallons/minute USgpm
Pounds/second lb/s U.S. gallons/hour USgph
Pounds/minute lb/min Imperial gallons/second UKgps
Pounds/hour lb/hr Imperial gallons/minute UKgpm
Pounds/day lb/day Imperial gallons/hour UKgph
Short tons (2000 lb)/minute STon/min Imperial gallons/day UKgpd
Short tons (2000 lb)/hour STon/hr Million gallons/day MilGal/day
Short tons (2000 lb)/day STon/day Liters/second l/sec
Long tons (2240 lb)/minute LTon/min Liters/minute l/min
Long tons (2240 lb)/hour LTon/hr Liters/hour l/hr
Long tons (2240 lb)/day LTon/day Milliliters/day MilL/day
Ounces/second oz/s Barrels/second bbl/s
Ounces/minute oz/min Barrels/minute bbl/min
Ounces/hour oz/hr Barrels/hour bbl/hr
Barrels/day bbl/day
Fluid ounces/second Floz/s
Fluid ounces/minute Floz/min
Fluid ounces/hour Floz/hr
Person-Process Interface Configuration: InputsConfiguration: System DataBefore You Begin
Series 3000 Detailed Setup Manual 15
Inputs continued

Density inputs To configure density inputs:

Configuration
Inputs
Coriolis
Config process var
Density
Density
Density Units
g/cc
Density Damping
1.7 sec
Slug Low Limit
0.000000 g/cc
Slug High Limit
5.000000 g/cc
CHG EXIT
1. Press the security button on the display face.
2. Select Configuration.
3. Select Inputs.
4. Select Coriolis.
5. Select Config Process Var.
6. Select Density.
7. Use the function buttons and cursor control buttons to configure the parameters that are listed
in Tabl e 4-5.
Table 4-5. Density inputs
Variable Default Description
Density units g/cc • Select the desired unit of density (see Tab le 4-6 )
Density damping 1.7 sec • Damping filters out noise or the effects of rapid changes in density without affecting
Slug low limit 0.000000 g/cc • Enter the desired low limit, in g/cc, for the process density
Slug high limit 5.000000 g/cc • Enter the desired high limit, in g/cc, for the process density
Slug time 1.0 sec • Enter the number of seconds for which flow outputs will hold their last measured flow
• Density outputs and displays will indicate density in the selected unit
measurement accuracy
• If the platform will operate with a Micro Motion T-Series sensor, the recommended density damping value is 0.3 seconds
• Milliamp outputs have their own damping
• The entered value is the density below which a slug flow alarm will be generated
• For more information about slug flow, see page 98
• The entered value is the density above which a slug flow alarm will be generated
• For more information about slug flow, see page 98
rate while density is outside the range specified by the slug low limit and slug high limit
• The maximum slug time is 1200 seconds
• If a value of 0.0 is entered, as soon as slug flow is detected, flow outputs will go to the level that indicates zero flow
• For more information about slug time, see page 98
16 Series 3000 Detailed Setup Manual
Inputs continued
Table 4-6. Density units
Unit Software label
Grams/cubic centimeter g/cc
Kilograms/cubic meter kg/cum
Pounds/gallon lb/gal
Pounds/cubic foot lb/cuft
Grams/milliliter g/mL
Kilograms/liter kg/L
Grams/liter g/L
Pounds/cubic inch lb/CuIn
Short tons (2000 lb)/cubic yard STon/CuYd
Temperature To configure temperature inputs:
Configuration
Inputs
Coriolis
Config process var

Temperature

Temperature
Temperature Units
degC
Temperature Damping
3.5 sec
CHG EXIT
1. Press the security button on the display face.
2. Select Configuration.
3. Select Inputs.
4. Select Coriolis.
5. Select Config Process Var.
6. Select Temperature.
7. Use the function buttons and cursor control buttons to configure the parameters that are listed
in Tabl e 4-7.
Table 4-7. Temperature inputs
Person-Process Interface Configuration: InputsConfiguration: System DataBefore You Begin
Variable Default Description
Temperature units degC • Select °Celsius, °Fahrenheit, °Rankine, or Kelvin
Temperature damping 3.5 sec • Damping filters out noise or the effects of rapid changes in temperature without
Series 3000 Detailed Setup Manual 17
• Temperature outputs and displays will indicate temperature in the selected unit
affecting measurement accuracy
• If the platform will operate with a Micro Motion T-Series sensor, the recommended temperature damping value is 0.0 seconds
• Milliamp outputs have their own damping
Inputs continued

4.5 Sensor calibration data Sensor calibration data describe the sensor’s

Configuration
Inputs
Coriolis
Sensor cal data
Sensor Cal Data
Flow Factor
1.00000
Flocal Temp Coef
5.130
D1
0.000000
D2
1.000000
CHG EXIT
Sensor Cal Data
T-Series Setup
NO
Flow Factor
1.00000
Flowcal Temp Coef
5.130
D1
0.000000
CHG EXIT
sensitivity to flow, density, and temperature.
To configure sensor calibration data:
1. Press the security button on the display face.
2. Select Configuration.
3. Select Inputs.
4. Select Coriolis.
5. Select Sensor Cal Data.
6. If the applications platform is connected to a sensor, skip to step 7. If the applications platform is not connected to a sensor, select T-Series Setup, then:
• Select Yes to enter calibration data for a Micro
Motion T-Series sensor (see page 19), or
• Select No to enter calibration data for an
ELITE, BASIS, Model D, Model DL, or Model DT sensor (see pages 20-26).
7. Use the function buttons and cursor control buttons to configure sensor calibration data.
• Sensor cal data should be entered from the
sensor serial number tag or factory calibration certificate.
• Tags and certificates vary in appearance,
depending on the sensor model number and manufacturing date.
18 Series 3000 Detailed Setup Manual
Inputs continued

Calibration data for Micro Motion T-Series sensors

If the applications platform is connected to a Micro Motion T-Series sensor, or if YES was selected at step 6, page 18, the Person-Process Interface enables configuration of calibration data for a T-Series sensor.
Sensor Cal Data
T-Series Setup
FCF
FT
FTG
0.000000
CHG EXIT
YES
1.00000
5.130
Flow calibration values include the FCF, FT, FTG, and FFQ. Enter the flow calibration values that appear on the sensor tag. See Figure 4-2.
Density calibration values include D1, D2, K1, K2, FD, DT, DTG, DFQ1, and DFQ2. Enter the density calibration values that appear on the sensor tag. See
Figure 4-2.
Temperature calibration values include the
temperature slope and the temperature offset. To enter temperature calibration values, see page 26.
Figure 4-2. Sensor calibration data on Micro Motion T-Series sensor tag
Flow calibration values
Person-Process Interface Configuration: InputsConfiguration: System DataBefore You Begin
FCF FTG
Density calibration values
D1
D2 DT
DTG
FT
FFQ
K1 K2 FD
DFQ1 DFQ2
Series 3000 Detailed Setup Manual 19
Inputs continued

Calibration data for ELITE®, BASIS®, Model D, Model DL, or Model DT sensors

Sensor Cal Data
T-Series Setup
NO
Flow Factor
1.00000
Flowcal Temp Coef
5.130
D1
0.000000
CHG EXIT
If the applications platform is connected to an ELITE, BASIS, Model D, Model DL, or Model DT sensor, or if NO was selected at step 6, page 18, the Person­Process Interface enables configuration of calibration data for the appropriate sensor.
Flow calibration values include the flow factor and the flow calibration temperature coefficient. To configure flow calibration values, see page 20.
Density calibration values include D1 and D2 density values, K1 and K2 tube periods, the flowing density correction factor, and the density calibration temperature coefficient. To configure density calibration values, see pages 21-25.
Temperature calibration values include the temperature slope and the temperature offset. To configure temperature calibration values, see page 26.
Flow calibration values
Flow calibration values include the flow factor and the flow calibration temperature coefficient (flowcal temp coef). To configure flow calibration values, see Table 4-8 and Figure 4-3.
Table 4-8. Flow calibration values
Variable Default Description
Flow factor 1.00000 g/sec • Enter the first 5 digits of the flow cal factor (see Figure 4-3)
• The entered value is the flow rate, in g/sec, that generates 1 velocity signals from the sensor
Flowcal temp coef 5.130 • Enter the last 3 digits of the flow cal factor (see Figure 4-3)
• The entered value represents the percent change in the measured flow rate per 100°C change in temperature
µsec of time shift between
20 Series 3000 Detailed Setup Manual
Inputs continued
Figure 4-3. Flow calibration values on sensor serial number tag
Flow factor on newer tag Flow factor on older tag
19.0005.13
Flocal temp coef on newer tag Flocal temp coef on older tag
19.0005.13
Person-Process Interface Configuration: InputsConfiguration: System DataBefore You Begin
19.0005.13
19.0005.13
Density calibration values
Density calibration values include D1 and D2 density values, K1 and K2 tube periods, the flowing density correction factor (FD), and the density calibration temperature coefficient (dens temp coeff).
To configure D1 and D2, see Ta ble 4-9 and Figure 4-4, page 22.
To configure K1 and K2, see Ta ble 4-10 and Figure 4-5, page 23.
To configure FD and the dens temp coeff, see Table 4-1 1 and Figure 4-6, page 24.
Table 4-9. D1 and D2 density values
Variable Default Description
D1 0.000000 g/cc • If the sensor tag shows a D1 value, enter the D1 value (see Figure 4-4)
D2 1.000000 g/cc • If the sensor tag shows a D2 value, enter the D2 value (see Figure 4-4)
• If the sensor tag does not show a D1 value, enter the Dens A or D1 value from the calibration certificate
• The entered value is the line-condition density of the low-density calibration fluid (Micro Motion uses air)
• If the sensor tag does not show a D2 value, enter the Dens B or D2 value from the calibration certificate
• The entered value is the line-condition density of the high-density calibration fluid (Micro Motion uses water)
Series 3000 Detailed Setup Manual 21
Inputs continued
Figure 4-4. D1 and D2 on sensor serial number tag
D1 on newer tag D2 on newer tag
0.0010
0.9980
Table 4-10. K1 and K2 tube period values
Variable Default Description
K1 500.000 • If the sensor tag shows a K1 value, enter the K1 value (see Figure 4-5)
K2 50000.000 • If the sensor tag shows a K2 value, enter the K2 value (see Figure 4-5)
• If the sensor tag does not show a K1 value, enter the first 5 digits of the density calibration factor (see Figure 4-5)
• The entered value represents the sensor flow tube period associated with D1, adjusted to 0°C
• If the sensor tag does not show a K2 value, enter the second 5 digits of the density calibration factor (see Figure 4-5)
• The entered value represents the sensor flow tube period associated with D2, adjusted to 0°C
22 Series 3000 Detailed Setup Manual
Loading...
+ 154 hidden pages