Micro Motion Micro Motion 2500 Transmitters with Configurable Input/Outputs Manuals & Guides

Page 1
Configuration and Use Manual
MMI-20019038, Rev AC
Micro Motion™ 2500 Transmitters with Configurable Input/Outputs
Configuration and Use Manual
February 2022
Page 2
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step.
Safety and approval information
This Micro Motion product complies with all applicable European directives when properly installed in accordance with the instructions in this manual. Refer to the EU declaration of conformity for directives that apply to this product. The following are available: the EU declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions. In addition the IECEx Installation Instructions for installations outside of the European Union and the CSA Installation Instructions for installations in North America are available on the internet at or through your local Micro Motion support center.
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at . For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion Customer Service department.
Emerson Flow customer service
Email:
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America
United States 800-522-6277 U.K. and Ireland 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 70 413
Mexico +52 55 5809 5010 France +33 (0) 800 917
Argentina +54 11 4809 2700 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy +39 8008 77334 China +86 21 2892 9000
Chile +56 2 2928 4800 Central & Eastern +41 (0) 41 7686
Peru +51 15190130 Russia/CIS +7 495 995 9559 South Korea +82 2 3438 4600
Europe and Middle East Asia Pacific
New Zealand 099 128 804
6666
India 800 440 1468
901
Japan +81 3 5769 6803
111
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684
2
Page 3
Configuration and Use Manual Contents
MMI-20019038 February 2022

Contents

Part I Getting started
Chapter 1 Before you begin........................................................................................................9
1.1 About this manual....................................................................................................................... 9
1.2 Transmitter model code.............................................................................................................. 9
1.3 Communications tools and protocols.......................................................................................... 9
1.4 Additional documentation and resources.................................................................................. 10
Chapter 2 Quick start............................................................................................................... 11
2.1 Power up the transmitter...........................................................................................................11
2.2 Check meter status....................................................................................................................11
2.3 Make a startup connection to the transmitter............................................................................12
2.4 (Optional) Adjust digital communications settings.................................................................... 12
2.5 Verify mass flow measurement..................................................................................................13
2.6 Verify the zero........................................................................................................................... 13
Part II Configuration and commissioning
Chapter 3 Introduction to configuration and commissioning....................................................17
3.1 Configuration flowchart.............................................................................................................17
3.2 Default values and ranges.......................................................................................................... 18
3.3 Disable write-protection on the transmitter configuration.........................................................18
3.4 Restore the factory configuration.............................................................................................. 18
Chapter 4 Configure process measurement..............................................................................19
4.1 Configure mass flow measurement........................................................................................... 19
4.2 Configure volume flow measurement for liquid applications..................................................... 23
4.3 Configure GSV flow measurement.............................................................................................28
4.4 Configure Flow Direction .......................................................................................................... 33
4.5 Configure density measurement ...............................................................................................38
4.6 Configure temperature measurement....................................................................................... 41
4.7 Configure the petroleum measurement application.................................................................. 43
4.8 Set up concentration measurement ..........................................................................................48
4.9 Configure pressure compensation............................................................................................. 55
Chapter 5 Configure device options and preferences................................................................ 61
5.1 Configure response time parameters.........................................................................................61
5.2 Configure alert handling............................................................................................................ 63
5.3 Configure informational parameters..........................................................................................68
Configuration and Use Manual 3
Page 4
Contents Configuration and Use Manual
February 2022 MMI-20019038
Chapter 6 Integrate the meter with the control system............................................................ 71
6.1 Configure the transmitter channels........................................................................................... 71
6.2 Configure the mA Output.......................................................................................................... 72
6.3 Configure the Frequency Output............................................................................................... 80
6.4 Configure the Discrete Output...................................................................................................85
6.5 Configure the Discrete Input......................................................................................................89
6.6 Configure events....................................................................................................................... 91
6.7 Configure digital communications.............................................................................................94
Chapter 7 Complete the configuration................................................................................... 103
7.1 Test or tune the system using sensor simulation......................................................................103
7.2 Back up transmitter configuration........................................................................................... 105
7.3 Enable write-protection on the transmitter configuration........................................................105
Chapter 8 Set up the Weights & Measures application............................................................ 107
8.1 Weights & Measures application.............................................................................................. 107
8.2 Set up the Weights & Measures application using ProLink III ....................................................108
Part III Operations, maintenance, and troubleshooting
Chapter 9 Transmitter operation............................................................................................113
9.1 Record the process variables....................................................................................................113
9.2 View process variables............................................................................................................. 113
9.3 View transmitter status using the status LED........................................................................... 114
9.4 View and acknowledge status alerts........................................................................................ 114
9.5 Read totalizer and inventory values..........................................................................................116
9.6 Start and stop totalizers and inventories.................................................................................. 116
9.7 Reset totalizers........................................................................................................................ 117
9.8 Reset inventories..................................................................................................................... 117
Chapter 10 Operate the transmitter with the Weights & Measures application.........................119
10.1 Operate the transmitter when the Weights & Measures application is installed..................... 119
10.2 Switch between secured and unsecured mode...................................................................... 123
10.3 Clear Status Alarm A027: Security Breach.............................................................................. 125
10.4 Replacing the core processor in a Weights & Measures installation........................................ 125
Chapter 11 Measurement support............................................................................................127
11.1 Options for measurement support........................................................................................ 127
11.2 Use Smart Meter Verification................................................................................................. 127
11.3 Use Production Volume Reconciliation, Transient Mist Remediation, and Transient Bubble
Remediation............................................................................................................................. 133
11.4 Piecewise linearization (PWL) for calibrating gas meters........................................................ 134
11.5 Use the fuel consumption application....................................................................................134
11.6 Zero the meter...................................................................................................................... 135
4 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 5
Configuration and Use Manual Contents
MMI-20019038 February 2022
11.7 Validate the meter.................................................................................................................135
11.8 Perform a (standard) D1 and D2 density calibration...............................................................137
11.9 Perform a D3 and D4 density calibration (T-Series sensors only)............................................ 140
11.10 Perform temperature calibration......................................................................................... 142
Chapter 12 Troubleshooting.................................................................................................... 145
12.1 Status LED states................................................................................................................... 145
12.2 Check the core processor LED................................................................................................ 145
12.3 Perform a 700 core processor resistance test.........................................................................147
12.4 Check the cutoffs...................................................................................................................148
12.5 Density measurement problems............................................................................................148
12.6 Discrete Input problems........................................................................................................ 149
12.7 Discrete output problems......................................................................................................149
12.8 Check the drive gain.............................................................................................................. 150
12.9 Check for internal electrical problems....................................................................................151
12.10 Check Flow Direction .......................................................................................................... 153
12.11 Flow measurement problems ............................................................................................. 153
12.12 Frequency Output problems................................................................................................156
12.13 Check Frequency Output Mode ...........................................................................................156
12.14 Check Frequency Output Fault Action .................................................................................157
12.15 Check Frequency Output Scaling Method ........................................................................... 157
12.16 Check grounding................................................................................................................. 157
12.17 Check the HART communication loop................................................................................. 158
12.18 Check HART Address and mA Output Action........................................................................158
12.19 Check HART burst mode......................................................................................................159
12.20 Perform loop tests............................................................................................................... 159
12.21 Check Lower Range Value and Upper Range Value ..............................................................162
12.22 Check mA Output Fault Action ............................................................................................162
12.23 Milliamp output problems................................................................................................... 163
12.24 Trim mA Output.................................................................................................................. 165
12.25 Check the pickoff voltage.................................................................................................... 166
12.26 Check power supply wiring.................................................................................................. 167
12.27 Check for radio frequency interference (RFI)........................................................................ 167
12.28 Using sensor simulation for troubleshooting....................................................................... 168
12.29 Check sensor-to-transmitter wiring..................................................................................... 168
12.30 Check for two-phase flow (slug flow)................................................................................... 168
12.31 Status alerts, causes, and recommendations....................................................................... 169
12.32 Temperature measurement problems.................................................................................185
Appendix A Using ProLink III with the transmitter.....................................................................187
A.1 Basic information about ProLink III ..........................................................................................187
A.2 Connect with ProLink III .......................................................................................................... 188
Configuration and Use Manual 5
Page 6
Contents Configuration and Use Manual
February 2022 MMI-20019038
Appendix B Using a field communicator with the transmitter................................................... 197
B.1 Basic information about field communicators..........................................................................197
B.2 Connect with a field communicator ........................................................................................ 198
Appendix C Default values and ranges...................................................................................... 201
C.1 Default values and ranges........................................................................................................201
Appendix D Transmitter components and installation wiring.................................................... 207
D.1 Installation types.....................................................................................................................207
D.2 Power supply terminals .......................................................................................................... 209
D.3 Input/output (I/O) wiring terminals......................................................................................... 210
Appendix E NE 53 history..........................................................................................................211
E.1 NE 53 history............................................................................................................................211
6 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 7
Configuration and Use Manual Getting started
MMI-20019038 February 2022
Part I Getting started
Configuration and Use Manual 7
Page 8
Getting started Configuration and Use Manual
February 2022 MMI-20019038
8 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 9
Configuration and Use Manual Before you begin
MMI-20019038 February 2022

1 Before you begin

1.1 About this manual

This manual helps you configure, commission, use, maintain, and troubleshoot the 2500 transmitter.
Important
This manual assumes that:
The transmitter has been installed correctly and completely according to the instructions in the
transmitter installation manual
Users understand basic transmitter and sensor installation, configuration, and maintenance concepts and
procedures

1.2 Transmitter model code

You can verify that this manual pertains to your transmitter by ensuring the model code on the transmitter tag matches the format.
Example:
The transmitter has a model number of the following form: 2500D**B****** or 2500D**C******
D
4-wire remote DIN rail–mount
B
Configurable input/outputs option board, default configuration (two mA outputs, one frequency output)
C
Configurable input/outputs option board, custom configuration

1.3 Communications tools and protocols

You must have a communications tool to interface with the transmitter. Several different communications tools and protocols are supported. You may use different tools in different locations or for different tasks.
Communications tool
ProLink III HART/Bell 202
Field communicator HART/Bell 202
Tip
You may be able to use other communications tools, such as AMS™ Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this manual. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at .
Supported protocols
Modbus/RS-485
Service port
Configuration and Use Manual 9
Page 10
Before you begin Configuration and Use Manual
February 2022 MMI-20019038

1.4 Additional documentation and resources

The following additional documentation supports the installation and operation of the transmitter.
Topic Document
Fuel consumption
Hazardous area installation See the approval documentation shipped with the
Product Data Sheet
Production Volume Reconciliation (PVR), Transient Bubble Remediation (TBR), and Transient Mist Remediation (TMR) applications
Sensor Sensor documentation
Transmitter installation
Micro Motion Fuel Consumption Application for Transmitters Installation and Operation Guide
transmitter, or download the appropriate documentation at www.emerson.com.
Micro Motion Series 1000 and Series 2000 Transmitters with MVD™ Technology Product Data Sheet (PDS)
Micro Motion Oil and Gas Production Supplement
Micro Motion Model 1500 and Model 2500 Transmitters: Installation Manual
All documentation resources are available at www.emerson.com or on the user documentation DVD.
10 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 11
Configuration and Use Manual Quick start
MMI-20019038 February 2022

2 Quick start

2.1 Power up the transmitter

The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement.
Procedure
1. WARNING
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are tightly closed. For hazardous area installations, applying power while housing covers are removed or loose can cause an explosion resulting in injury or death.
Ensure that all transmitter and sensor covers and seals are closed.
2. Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. The transmitter is self-switching and will automatically detect the supply voltage. When using DC power, a minimum of 1.5 amps of startup current is required. During this period, Alert 009 is active. The diagnostic routines should complete in approximately 30 seconds. The status LED will turn green when the startup diagnostics are complete. If the status LED exhibits different behavior, an alert is active.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power has been off long enough to allow components to reach ambient temperature, allow the electronics to warm up for approximately 10 minutes before relying on process measurements. During this warm-up period, you may observe minor measurement instability or inaccuracy.

2.2 Check meter status

Check the meter for any error conditions that require user action or that affect measurement accuracy.
Procedure
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alert A009 is active. This alert should clear automatically when the power-up sequence is complete.
2. Check the status LED on the transmitter.
Related information
View and acknowledge status alerts Status alerts, causes, and recommendations
Configuration and Use Manual 11
Page 12
Quick start Configuration and Use Manual
February 2022 MMI-20019038
2.2.1 Transmitter status reported by LED
Table 2-1: Status LED states
LED state Alarm condition Description
Solid green No alarm Normal operation
Flashing yellow No alarm Zero calibration procedure is in progress
Loop test is in progress
Solid yellow Low-severity alarm Alarm condition that will not cause measurement error
(outputs continue to report process data)
Solid red High-severity alarm Alarm condition that will cause measurement error
(outputs in fault)

2.3 Make a startup connection to the transmitter

To configure the transmitter, you must have an active connection from a communications tool. Follow this procedure to make your first connection to the transmitter.
Procedure
Identify the connection type to use, and follow the instructions for that connection type in the appropriate appendix. Use the default communications parameters shown in the appendix.
Communications tool
ProLink III Modbus/RS-485 Using ProLink III with the transmitter
Field Communicator HART Using a field communicator with the
Connection type to use Instructions
transmitter

2.4 (Optional) Adjust digital communications settings

Change the communications parameters to site-specific values.
Important
If you are changing communications parameters for the connection type that you are using, you will lose the connection when you write the parameters to the transmitter. Reconnect using the new parameters.
Procedure
1. To change the communications parameters using ProLink III, choose Device Tools Configuration
Communications.
2. To change the communications parameters using the Field Communicator, choose On-Line Menu
Configure Manual Setup Inputs/Outputs Communications.
12 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 13
Configuration and Use Manual Quick start
MMI-20019038 February 2022

2.5 Verify mass flow measurement

Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available method.
Procedure
Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the Process Variables
panel.
Connect to the transmitter with a field communicator and read the value for Mass Flow Rate.
Online Overview Mass Flow Rate
Postrequisites
If the reported mass flow rate is not accurate:
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues.

2.6 Verify the zero

Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
The zero verification procedure analyzes the Live Zero value under conditions of zero flow, and compares it to the Zero Stability range for the sensor. If the average Live Zero value is within a reasonable range, the zero value stored in the transmitter is valid. Performing a field calibration will not improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one of the following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Do not verify the zero or zero the meter if a high-severity alert is active. Correct the problem, then verify the zero or zero the meter. You may verify the zero or zero the meter if a low-severity alert is active.
Procedure
1. Allow the flowmeter to warm up for at least 20 minutes after applying power.
2. Run the process fluid through the sensor until the sensor temperature reaches the normal process
operating temperature.
3. Stop flow through the sensor by shutting the downstream valve, and then the upstream valve if
available.
4. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is completely full of
process fluid.
5. From ProLink III, choose Device Tools Calibration Zero Verification and Calibration Verify
Zero and wait until the procedure completes.
6. Observe the drive gain, temperature, and density readings. If they are stable, check the Live Zero or
Field Verification Zero value. If the average value is close to 0, you should not need to zero the meter.
Configuration and Use Manual 13
Page 14
Quick start Configuration and Use Manual
February 2022 MMI-20019038
7. If the zero verification procedure fails:
a) Confirm that the sensor is completely blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
b) Verify that the process fluid is not flashing or condensing, and that it does not contain particles
that can settle out.
c) Remove or reduce sources of electromechanical noise if appropriate.
d) Repeat the zero verification procedure.
e) If it fails again, zero the meter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Related information
Zero the meter
2.6.1 Terminology used with zero verification and zero calibration
Term Definition
Zero In general, the offset required to synchronize the left pickoff and the right pickoff under
conditions of zero flow. Unit = microseconds.
Factory Zero The zero value obtained at the factory, under laboratory conditions.
Field Zero The zero value obtained by performing a zero calibration outside the factory.
Prior Zero The zero value stored in the transmitter at the time a field zero calibration is begun. May
be the factory zero or a previous field zero.
Manual Zero The zero value stored in the transmitter, typically obtained from a zero calibration
procedure. It may also be configured manually. Also called “mechanical zero” or “stored zero”.
Live Zero The real-time bidirectional mass flow rate with no flow damping or mass flow cutoff
applied. An adaptive damping value is applied only when the mass flow rate changes dramatically over a very short interval. Unit = configured mass flow measurement unit.
Zero Stability A laboratory-derived value used to calculate the expected accuracy for a sensor. Under
laboratory conditions at zero flow, the average flow rate is expected to fall within the range defined by the Zero Stability value (0 ± Zero Stability). Each sensor size and model has a unique Zero Stability value.
Zero Calibration The procedure used to determine the zero value.
Zero Time The time period over which the Zero Calibration procedure is performed. Unit = seconds.
Field Verification Zero A 3-minute running average of the Live Zero value, calculated by the transmitter. Unit =
configured mass flow measurement unit.
Zero Verification A procedure used to evaluate the stored zero and determine whether or not a field zero
can improve measurement accuracy.
14 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 15
Configuration and Use Manual Configuration and commissioning
MMI-20019038 February 2022
Part II Configuration and commissioning
Configuration and Use Manual 15
Page 16
Configuration and commissioning Configuration and Use Manual
February 2022 MMI-20019038
16 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 17
Configuration and Use Manual Introduction to configuration and commissioning
MMI-20019038 February 2022

3 Introduction to configuration and commissioning

3.1 Configuration flowchart

Use the following flowchart as a general guide to the configuration and commissioning process.
Some options may not apply to your installation. Detailed information is provided in the remainder of this manual. If you are using the Weights & Measures application, additional configuration and setup are required.
Configure process measurement
Configure mass flow
measurement
Configure device options and preferences
Configure fault handling
parameters
Test and move to production
Test or tune transmitter using sensor simulation
Configure volume flow
meaurement
Volume flow type
Liquid
Configure flow direction
Configure density
measurement
Configure temperature
measurement
Configure petroleum
measurement (API)
application (if available)
Configure concentration
measurement application
(if available)
Gas
Define gas properties
Configure sensor
parameters
Configure device
parameters
Integrate device with control system
Configure the channel(s)
Configure the mA
output(s)
Configure the frequency
output(s)
Configure the discrete
output(s)
Back up transmitter
configuration
Enable write-protection on
transmitter configuration
Done
Configure pressure
compensation (optional)
Configure PVR, TMR,
TBR, or fuel consumption
(if available)
Configure events
Configure digital communications
Configuration and Use Manual 17
Page 18
Introduction to configuration and commissioning Configuration and Use Manual
February 2022 MMI-20019038

3.2 Default values and ranges

See Default values and ranges to view the default values and ranges for the most commonly used parameters.
3.3 Disable write-protection on the transmitter
configuration
Display OFF-LINE MAINT CONFG LOCK
ProLink III Device Tools Configuration Write-Protection
If the transmitter is write-protected, the configuration is locked and you must unlock it before you can change any configuration parameters. By default, the transmitter is not write-protected.
Tip
Write-protecting the transmitter prevents accidental changes to configuration. It does not prevent normal operational use. You can always disable write-protection, perform any required configuration changes, then re-enable write-protection.

3.4 Restore the factory configuration

ProLink III Device Tools Configuration Transfer Restore Factory Configuration
Field communicator Service Tools Maintenance Reset/Restore Restore Factory Configuration
Restoring the factory configuration returns the transmitter to the same configuration it had when it left the factory. This may be useful if you experience problems during configuration.
Important
You cannot restore factory configurations with a 700 core.
Tip
Restoring the factory configuration is not a common action. You may want to contact customer support to see if there is a preferred method to resolve any issues.
18 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 19
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022

4 Configure process measurement

4.1 Configure mass flow measurement

The mass flow measurement parameters control how mass flow is measured and reported.
4.1.1 Configure Mass Flow Measurement Unit
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements Flow Mass Flow Unit
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow rate. The unit used for mass total and mass inventory is derived from this unit.
Any selected measurement unit, (mass, volume or gas standard volume), is automatically applied to both the mA and Frequency Outputs.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
The default setting for Mass Flow Measurement Unit is g/sec (grams per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Label
Unit description
Grams per second g/sec g/s
Grams per minute g/min g/min
Grams per hour g/hr g/h
Kilograms per second kg/sec kg/s
Kilograms per minute kg/min kg/min
Kilograms per hour kg/hr kg/h
ProLink III Field Communicator
Kilograms per day kg/day kg/d
Metric tons per minute mTon/min MetTon/min
Metric tons per hour mTon/hr MetTon/h
Metric tons per day mTon/day MetTon/d
Pounds per second lbs/sec lb/s
Configuration and Use Manual 19
Page 20
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
Label
Unit description
Pounds per minute lbs/min lb/min
Pounds per hour lbs/hr lb/h
Pounds per day lbs/day lb/d
Short tons (2000 pounds) per minute sTon/min STon/min
Short tons (2000 pounds) per hour sTon/hr STon/h
Short tons (2000 pounds) per day sTon/day STon/d
Long tons (2240 pounds) per hour lTon/hr LTon/h
Long tons (2240 pounds) per day lTon/day LTon/d
Special unit special Spcl
ProLink III Field Communicator
Define a special measurement unit for mass flow
ProLink III Device Tools Configuration Process Measurement Flow Special Units
Field Communicator Configure Manual Setup Measurements Special Units Mass Special Units
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Procedure
1. Specify Base Mass Unit.
Base Mass Unit is the existing mass unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Mass Flow Conversion Factor as follows:
a) x base units = y special units
b) Mass Flow Conversion Factor = x ÷ y
The original mass flow rate value is divided by this value.
4. Enter Mass Flow Conversion Factor.
5. Set Mass Flow Label to the name you want to use for the mass flow unit.
6. Set Mass Total Label to the name you want to use for the mass total and mass inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for mass flow
You want to measure mass flow in ounces per second (oz/sec).
1. Set Base Mass Unit to Pounds (lb).
20 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 21
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
2. Set Base Time Unit to Seconds (sec).
3. Calculate Mass Flow Conversion Factor:
a. 1 lb/sec = 16 oz/sec
b. Mass Flow Conversion Factor = 1 ÷ 16 = 0.0625
4. Set Mass Flow Conversion Factor to 0.0625.
5. Set Mass Flow Label to oz/sec.
6. Set Mass Total Label to oz.
4.1.2 Configure Flow Damping
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements Flow Flow Damping
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
The default value is 0.8 seconds. The range depends on the core processor type and the setting of Update Rate, as shown in the following table.
Update Rate setting
Normal 0 to 51.2 seconds
Special 0 to 40.96 seconds
Damping range
The value you enter is automatically rounded off to the nearest valid value. For example, if the damping is currently set to 0.8 seconds, any value entered up to 1.2 seconds will be rounded down to 0.8 seconds, and any value entered from 1.21 to 1.59 seconds will be rounded up to 1.6 seconds.
Update Rate setting
Normal 0.0, 0.2, 0.4, 0.8, 1.6, 3.2, 6.4, 12.8, 25.6, 51.2
Special 0.0, 0.04, 0.08, 0.16, 0.32, 0.64, 1.28, 2.56, 5.12, 10.24,
Valid damping values
20.48, 40.96
Configuration and Use Manual 21
Page 22
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
Effect of flow damping on volume measurement
Flow damping affects volume measurement for liquid volume data. Flow damping also affects volume measurement for gas standard volume data. The transmitter calculates volume data from the damped mass flow data.
Interaction between Flow Damping and mA Output Damping
In some circumstances, both Flow Damping and mA Output Damping are applied to the reported mass flow value.
Flow Damping controls the rate of change in flow process variables. mA Output Damping controls the rate of change reported through mA Output. If mA Output Process Variable is set to Mass Flow Rate, and both Flow Damping and mA Output Damping are set to non-zero values, flow damping is applied first, and the added damping calculation is applied to the result of the first calculation.
4.1.3 Configure Mass Flow Cutoff
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements Flow Mass Flow Cutoff
Mass Flow Cutoff specifies the lowest mass flow rate that will be reported as measured. All mass flow rates below this cutoff will be reported as 0.
Procedure
Set Mass Flow Cutoff to the value you want to use.
The default value for Mass Flow Cutoff is 0.0 g/sec or a sensor-specific value set at the factory. The recommended value is 0.5% of the nominal flow rate of the attached sensor. See the sensor specifications. Do not leave Mass Flow Cutoff at 0.0 g/sec.
Effect of Mass Flow Cutoff on volume measurement
Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the actual mass data rather than the reported value.
Volume flow has a separate Volume Flow Cutoff that is not affected by the Mass Flow Cutoff value.
Interaction between Mass Flow Cutoff and mA Output Cutoff
Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported through mA Output. If mA Output Process Variable is set to Mass Flow Rate, the mass flow rate reported through mA Output is controlled by the higher
of the two cutoff values.
Mass Flow Cutoff affects all reported values and values used in other transmitter behavior (e.g., events defined on mass flow).
mA Output Cutoff affects only mass flow values reported through mA Output.
22 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 23
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Example: Cutoff interaction with mA Output Cutoff lower than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
mA Output Cutoff: 10 g/sec
Mass Flow Cutoff: 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mA Output Cutoff higher than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
mA Output Cutoff: 15 g/sec
Mass Flow Cutoff: 10 g/sec
Result:
If the mass flow rate drops below 15 g/sec but not below 10 g/sec:The mA Output will report zero flow.
The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
4.2 Configure volume flow measurement for liquid
applications
The volume flow measurement parameters control how liquid volume flow is measured and reported.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.
Note
If you need to switch from gas standard volume to liquid volume, polling for base density will automatically be disabled.
4.2.1 Configure Volume Flow Type for liquid applications
ProLink III
Field Communicator Configure Manual Setup Measurements GSV Volume Flow Type Liquid
Configuration and Use Manual 23
Device Tools Configuration Process Measurement Flow
Page 24
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type to Liquid if you are using any of the following applications:
Petroleum measurement
Concentration measurement
Fuel consumption
Production Volume Reconciliation (PVR)
Procedure
Set Volume Flow Type to Liquid.
4.2.2 Configure Volume Flow Measurement Unit for liquid
applications
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements Flow Volume Flow Unit
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the volume flow rate. The unit used for the volume total and volume inventory is based on this unit.
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
To read US gallons, select that unit from this menu. G/MIN stands for grams per minute (USGPM), not gallons per minute. The default setting for Volume Flow Measurement Unit is l/sec (liters per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement Unit, plus one user-defined measurement unit. Different communications tools may use different labels for the units.
Label
Unit description
Cubic feet per second ft3/sec Cuft/s
Cubic feet per minute ft3/min Cuft/min
Cubic feet per hour ft3/hr Cuft/h
Cubic feet per day ft3/day Cuft/d
24 Micro Motion 2500 Transmitters with Configurable Input/Outputs
ProLink III Field Communicator
Page 25
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Label
Unit description
ProLink III Field Communicator
Cubic meters per second m3/sec Cum/s
Cubic meters per minute m3/min Cum/min
Cubic meters per hour m3/hr Cum/h
Cubic meters per day m3/day Cum/d
U.S. gallons per second US gal/sec gal/s
U.S. gallons per minute US gal/min gal/min
U.S. gallons per hour US gal/hr gal/h
U.S. gallons per day US gal/day gal/d
Million U.S. gallons per day mil US gal/day MMgal/d
Liters per second l/sec L/s
Liters per minute l/min L/min
Liters per hour l/hr L/h
Million liters per day mil l/day ML/d
Imperial gallons per second Imp gal/sec Impgal/s
Imperial gallons per minute Imp gal/min Impgal/min
Imperial gallons per hour Imp gal/hr Impgal/h
Imperial gallons per day Imp gal/day Impgal/d
(1)
(1)
(1)
(1)
(2)
(2)
(2)
(2)
barrels/sec bbl/s
barrels/min bbl/min
barrels/hr bbl/h
barrels/day bbl/d
Beer barrels/sec bbbl/s
Beer barrels/min bbbl/min
Beer barrels/hr bbbl/h
Beer barrels/day bbbl/d
Barrels per second
Barrels per minute
Barrels per hour
Barrels per day
Beer barrels per second
Beer barrels per minute
Beer barrels per hour
Beer barrels per day
Special unit special Spcl
(1) Unit based on oil barrels (42 U.S. gallons). (2) Unit based on U.S. beer barrels (31 U.S. gallons).
Define a special measurement unit for volume flow
ProLink III
Field Communicator Configure Manual Setup Measurements Special Units Volume Special Units
Device Tools Configuration Process Measurement Flow Special Units
Configuration and Use Manual 25
Page 26
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Note
Procedure
1. Specify Base Volume Unit.
Base Volume Unit is the existing volume unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Volume Flow Conversion Factor as follows:
a) x base units = y special units
b) Volume Flow Conversion Factor = x ÷ y
4. Enter Volume Flow Conversion Factor.
The original volume flow rate value is divided by this conversion factor.
5. Set Volume Flow Label to the name you want to use for the volume flow unit.
6. Set Volume Total Label to the name you want to use for the volume total and volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Defining a special measurement unit for volume flow
You want to measure volume flow in pints per second (pints/sec).
1. Set Base Volume Unit to Gallons (gal).
2. Set Base Time Unit to Seconds (sec).
3. Calculate the conversion factor:
a. 1 gal/sec = 8 pints/sec
b. Volume Flow Conversion Factor = 1 ÷ 8 = 0.1250
4. Set Volume Flow Conversion Factor to 0.1250.
5. Set Volume Flow Label to pints/sec.
6. Set Volume Total Label to pints.
4.2.3 Configure Volume Flow Cutoff
ProLink III
Field Communicator Configure Manual Setup Measurements Flow Volume Flow Cutoff
26 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Device Tools Configuration Process Measurement Flow
Page 27
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Volume Flow Cutoff specifies the lowest volume flow rate that will be reported as measured. All volume flow rates below this cutoff are reported as 0.
Procedure
Set Volume Flow Cutoff to the value you want to use.
The default value for Volume Flow Cutoff is 0.0 l/sec (liters per second). The lower limit is 0. Leaving the volume flow cutoff at 0 is not recommended.
Interaction between Volume Flow Cutoff and mAO Cutoff
Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will report as measured. mAO Cutoff defines the lowest flow rate that will be reported through mA Output. If mA Output Process Variable is set to Volume Flow Rate, the volume flow rate reported through mA Output is controlled by the
higher of the two cutoff values.
Volume Flow Cutoff affects both the volume flow values reported via the outputs and the volume flow values used in other transmitter behavior (e.g., events defined on the volume flow).
mAO Cutoff affects only flow values reported through mA Output.
Example: Cutoff interaction with mAO Cutoff lower than Volume Flow Cutoff
Configuration:
mA Output Process Variable: Volume Flow Rate
Frequency Output Process Variable: Volume Flow Rate
AO Cutoff: 10 l/sec
Volume Flow Cutoff: 15 l/sec
Result: If the volume flow rate drops below 15 l/sec, volume flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mAO Cutoff higher than Volume Flow Cutoff
Configuration:
mA Output Process Variable: Volume Flow Rate
Frequency Output Process Variable: Volume Flow Rate
AO Cutoff: 15 l/sec
Volume Flow Cutoff: 10 l/sec
Result:
If the volume flow rate drops below 15 l/sec but not below 10 l/sec:The mA Output will report zero flow.
The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
If the volume flow rate drops below 10 l/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
Configuration and Use Manual 27
Page 28
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038

4.3 Configure GSV flow measurement

The gas standard volume (GSV) flow measurement parameters control how volume flow is measured and reported in a gas application.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.
4.3.1 Configure Volume Flow Type for gas applications
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements GSV Volume Flow Type Standard Gas Volume
Volume Flow Type controls whether liquid or gas standard volume flow measurement is used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type to Liquid if you are using any of the following applications:
Petroleum measurement
Concentration measurement
Fuel consumption
Production Volume Reconciliation (PVR)
Procedure
Set Volume Flow Type to Gas Standard Volume.
4.3.2 Configure Standard Density of Gas
ProLink III
Field Communicator Configure Manual Setup Measurements GSV Gas Ref Density
The Standard Density of Gas value is the gas density at standard reference conditions. Use it to convert the measured mass flow data to volume flow at reference conditions.
Prerequisites
Ensure that Density Measurement Unit is set to the measurement unit you want to use for Standard Density of Gas.
Procedure
From the Source field, choose the method to supply gas base density data and perform the required setup.
Device Tools Configuration Process Measurement Flow
Option
Fixed Value or Digital Communications
28 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Description
A host writes gas base density data to the meter at appropriate intervals. Continue to Configure fixed value or digital communications.
Page 29
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Option Description
Poll for external value The meter polls an external HART device for gas base density data in order
to then compute gas standard volume from the mass flow and gas base density.
Continue to Poll for external value.
Configure fixed value or digital communications
Prerequisites
Configure Standard Density of Gas
Procedure
1. Set Standard Density of Gas to the standard reference density of the gas you are measuring.
Note
ProLink III provides a guided method that you can use to calculate your gas base density, if you do not know it.
2. Continue to Configure Gas Standard Volume Flow Unit .
Poll for external value
Prerequisites
Configure Standard Density of Gas
Procedure
1. Set Polling Slot to an available slot.
2. Set Polling Control n as one of the following options:
The n is the value you selected in the Polling Slot field.
If there is another master, and if that master is primary, then set this field to secondary. If the other master is secondary, then set this field to primary.
Option
Poll as Primary No other HART masters will be on the network.
Poll as Secondary Other HART masters will be on the network.
3. Set External Device Tag n to the HART tag of the device being polled.
The n is the value you selected in the Polling Slot field.
Description
The device being polled (slave) cannot have special units set for density. Otherwise, the master will
reject the base density and report the following alarm:
A115: No External Input or Polled Data Alert
On the slave side, setup the HART Primary Variable for Base Density. The master will reject anything
other than Base Density for the HART Primary Variable and trigger an A115 alarm.
Configuration and Use Manual 29
Page 30
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
The density units on the transmitter and the polled device can be different as long as they can be
classified as density units; for example, kg/m3 and g/cm3. The transmitter converts the polled units into compatible specified units.
For wiring and setup instructions for a polled device, refer to the Micro Motion Gas Density Meters (GDM) Installation manual or the Micro Motion Specific Gravity Meters (SGM) Installation manual.
4. Continue to Configure Gas Standard Volume Flow Unit .
4.3.3 Configure Gas Standard Volume Flow Unit
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements GSV GSV Flow Unit
Gas Standard Volume Flow Unit specifies the unit of measure that will be displayed for the gas standard volume flow. The measurement unit used for the gas volume total and the gas volume inventory is derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Unit, be sure that Volume Flow Type is set to Gas Standard Volume.
For polling, the first transmitter (master) requests density from a second transmitter (slave) via HART communications. Special units for GSV are allowed on the master side, but the device being polled (slave) cannot have special units set for density, otherwise the master will reject the base density and report an A115: No External Input or Polled Data Alert.
Procedure
Set Gas Standard Volume Flow Unit to the unit you want to use.
The default setting for Gas Standard Volume Flow Unit is SCFM (Standard Cubic Feet per Minute).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Gas Standard Volume Flow Unit
The transmitter provides a standard set of measurement units for Gas Standard Volume Flow Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Label
Unit description
Normal cubic meters per second Nm3/sec Nm3/sec
ProLink III Field Communicator
Normal cubic meters per minute Nm3/sec Nm3/min
Normal cubic meters per hour Nm3/hr Nm3/hr
Normal cubic meters per day Nm3/day Nm3/day
Normal liters per second NLPS NLPS
Normal liters per minute NLPM NLPM
30 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 31
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Label
Unit description
Normal liters per hour NLPH NLPH
Normal liters per day NLPD NLPD
Standard cubic feet per second SCFS SCFS
Standard cubic feet per minute SCFM SCFM
Standard cubic feet per hour SCFH SCFH
Standard cubic feet per day SCFD SCFD
Standard cubic meters per second Sm3/sec Sm3/sec
Standard cubic meters per minute Sm3/min Sm3/min
Standard cubic meters per hour Sm3/hr Sm3/hr
Standard cubic meters per day Sm3/day Sm3/day
Standard liters per second SLPS SLPS
Standard liters per minute SLPM SLPM
Standard liters per hour SLPH SLPH
Standard liters per day SLPD SLPD
Special measurement unit special Special
ProLink III Field Communicator
Define a special measurement unit for gas standard volume flow
ProLink III
Field Communicator Configure Manual Setup Measurements Special Units Special GSV Units
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Procedure
1. Specify Base Gas Standard Volume Unit.
Base Gas Standard Volume Unit is the existing gas standard volume unit that the special unit will be
based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Gas Standard Volume Flow Conversion Factor as follows:
a) x base units = y special units
b) Gas Standard Volume Flow Conversion Factor = x ÷ y
4. Enter the Gas Standard Volume Flow Conversion Factor.
The original gas standard volume flow value is divided by this conversion factor.
Device Tools Configuration Process Measurement Flow Special Units
Configuration and Use Manual 31
Page 32
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
5. Set Gas Standard Volume Flow Label to the name you want to use for the gas standard volume flow unit.
6. Set Gas Standard Volume Total Label to the name you want to use for the gas standard volume total and gas standard volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for gas standard volume flow
You want to measure gas standard volume flow in thousands of standard cubic feet per minute.
1. Set Base Gas Standard Volume Unit to SCF.
2. Set Base Time Unit to minutes (min).
3. Calculate the conversion factor:
a. 1 thousands of standard cubic feet per minute = 1000 cubic feet per minute
b. Gas Standard Volume Flow Conversion Factor = 1 ÷ 1000 = 0.001 standard
4. Set Gas Standard Volume Flow Conversion Factor to 0.001.
5. Set Gas Standard Volume Flow Label to MSCFM.
6. Set Gas Standard Volume Total Label to MSCF.
4.3.4 Configure Gas Standard Volume Flow Cutoff
ProLink III
Field Communicator Configure Manual Setup Measurements GSV GSV Cutoff
Gas Standard Volume Flow Cutoff specifies the lowest gas standard volume flow rate that will reported as measured. All gas standard volume flow rates below this cutoff will be reported as 0.
Procedure
Set Gas Standard Volume Flow Cutoff to the value you want to use.
The default value for Gas Standard Volume Flow Cutoff is 0.0. The lower limit is 0.0. There is no upper limit. The recommended value is 0.5% of the nominal flow rate of the attached sensor. See the sensor specifications.
Interaction between Gas Standard Volume Flow Cutoff and mA Output Cutoff
Gas Standard Volume Flow Cutoff defines the lowest Gas Standard Volume flow value that the transmitter will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported through mA Output. If mA Output Process Variable is set to Gas Standard Volume Flow Rate, the volume flow rate reported through mA Output is controlled by the higher of the two cutoff values.
Device Tools Configuration Process Measurement Flow
Gas Standard Volume Flow Cutoff affects both the gas standard volume flow values reported through
outputs and the gas standard volume flow values used in other transmitter behavior (for example, events defined on gas standard volume flow).
32 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 33
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
mA Output Cutoff affects only flow values reported through mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Gas Standard Volume Flow Cutoff
Configuration:
mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow Rate
Frequency Output Process Variable: Gas Standard Volume Flow Rate
mA Output Cutoff for the primary mA Output: 10 SLPM (standard liters per minute)
Gas Standard Volume Flow Cutoff: 15 SLPM
Result: If the gas standard volume flow rate drops below 15 SLPM, the volume flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mA Output Cutoff higher than Gas Standard Volume Flow Cutoff
Configuration:
mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow Rate
Frequency Output Process Variable: Gas Standard Volume Flow Rate
mA Output Cutoff for the primary mA Output: 15 SLPM (standard liters per minute)
Gas Standard Volume Flow Cutoff: 10 SLPM
Result:
If the gas standard volume flow rate drops below 15 SLPM but not below 10 SLPM:The primary mA Output will report zero flow.
The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
If the gas standard volume flow rate drops below 10 SLPM, both outputs will report zero flow, and 0 will be
used in all internal processing.

4.4 Configure Flow Direction

ProLink III
Field Communicator Configure Manual Setup Measurements Flow Flow Direction
Flow Direction controls how forward flow and reverse flow affect flow measurement and reporting.
Flow Direction is defined with respect to the flow arrow on the sensor:
Forward flow (positive flow) moves in the direction of the flow arrow on the sensor.
Reverse flow (negative flow) moves in the direction opposite to the flow arrow on the sensor.
Device Tools Configuration Process Measurement Flow
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow direction or the setting of the Flow Direction parameter.
Procedure
Set Flow Direction to the value you want to use.
Configuration and Use Manual 33
Page 34
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
The default setting is Forward.
4.4.1 Options for Flow Direction
Flow Direction setting
Relationship to Flow Direction arrow on sensorProLink III Field Communicator
Forward Forward Appropriate when the Flow Direction arrow is in
the same direction as the majority of flow.
Reverse Reverse Appropriate when the Flow Direction arrow is in
the opposite direction from the majority of flow.
Absolute Value Absolute Value Flow Direction arrow is not relevant.
Bidirectional Bi directional Appropriate when both forward and reverse flow
are expected, and forward flow will dominate, but the amount of reverse flow will be significant.
Negate Forward Negate/Forward Only Appropriate when the Flow Direction arrow is in
the opposite direction from the majority of flow.
Negate Bidirectional Negate/Bi-directional Appropriate when both forward and reverse flow
are expected, and reverse flow will dominate, but the amount of forward flow will be significant.
Effect of Flow Direction on mA Outputs
Flow Direction affects how the transmitter reports flow values via the mA Outputs. The mA Outputs are affected by Flow Direction only if mA Output Process Variable is set to a flow variable.
Flow Direction and mA Outputs
The effect of Flow Direction on the mA Outputs depends on Lower Range Value configured for the mA Output:
If Lower Range Value is set to 0, see Figure 4-1.
If Lower Range Value is set to a negative value, see Figure 4-2.
34 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 35
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Figure 4-1: Effect of Flow Direction on the mA Output: Lower Range Value = 0
Flow Direction = Forward
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
Flow Direction = Reverse, Negate Forward
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
Flow Direction = Absolute Value, Bidirectional, Negate Bidirectional
20
12
mA output
4
Lower Range Value = 0
Upper Range Value = x
Figure 4-2: Effect of Flow Direction on the mA Output: Lower Range Value < 0
Flow Direction = Forward
20
Flow Direction = Reverse, Negate Forward
20
Flow Direction = Absolute Value, Bidirectional, Negate Bidirectional
20
-x 0 x
Reverse flow Forward flow
12
mA output
4
-x 0 x
Reverse flow Forward flow
12
mA output
4
-x 0 x
Reverse flow Forward flow
12
mA output
4
-x 0 x
Reverse flow Forward flow
Lower Range Value = x
Upper Range Value = x
Flow Direction = Forward and Lower Range Value = 0
Configuration:
Flow Direction = Forward
Lower Range Value = 0 g/sec
Upper Range Value = 100 g/sec
Configuration and Use Manual 35
Page 36
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
Result:
Under conditions of zero flow, the mA Output is 4 mA.
Under conditions of forward flow, up to a flow rate of 100 g/sec, the mA Output varies between 4 mA and
20 mA in proportion to the flow rate.
Under conditions of forward flow, if the flow rate equals or exceeds 100 g/sec, the mA Output will be
proportional to the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher flow rates.
Flow Direction = Forward and Lower Range Value < 0
Configuration:
Flow Direction = Forward
Lower Range Value = 100 g/sec
Upper Range Value = +100 g/sec
Result:
Under conditions of zero flow, the mA Output is 12 mA.
Under conditions of forward flow, for flow rates between 0 and +100 g/sec, the mA Output varies between
12 mA and 20 mA in proportion to (the absolute value of) the flow rate.
Under conditions of forward flow, if (the absolute value of) the flow rate equals or exceeds 100 g/sec, the
mA Output is proportional to the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher flow rates.
Under conditions of reverse flow, for flow rates between 0 and 100 g/sec, the mA Output varies between
4 mA and 12 mA in inverse proportion to the absolute value of the flow rate.
Under conditions of reverse flow, if the absolute value of the flow rate equals or exceeds 100 g/sec, the mA
Output is inversely proportional to the flow rate down to 3.8 mA, and will be level at 3.8 mA at higher absolute values.
Flow Direction = Reverse
Configuration:
Flow Direction = Reverse
Lower Range Value = 0 g/sec
Upper Range Value = 100 g/sec
Result:
Under conditions of zero flow, the mA Output is 4 mA.
Under conditions of reverse flow, for flow rates between 0 and +100 g/sec, the mA Output level varies
between 4 mA and 20 mA in proportion to the absolute value of the flow rate.
Under conditions of reverse flow, if the absolute value of the flow rate equals or exceeds 100 g/sec, the mA
Output will be proportional to the absolute value of the flow rate up to 20.5 mA, and will be level at
20.5 mA at higher absolute values.
36 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 37
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Effect of flow direction on Frequency Outputs
Flow direction affects how the transmitter reports flow values via the Frequency Outputs. The Frequency Outputs are affected by flow direction only if Frequency Output Process Variable is set to a flow variable.
Table 4-1: Effect of the flow direction parameter and actual flow direction on Frequency Outputs
Actual flow direction
Flow Direction setting
Forward Hz > 0 0 Hz 0 Hz
Reverse 0 Hz 0 Hz Hz > 0
Bidirectional Hz > 0 0 Hz Hz > 0
Absolute Value Hz > 0 0 Hz Hz > 0
Negate Forward 0 Hz 0 Hz Hz > 0
Negate Bidirectional Hz > 0 0 Hz Hz > 0
Forward Zero flow Reverse
Effect of flow direction on Discrete Outputs
The flow direction parameter affects the Discrete Output behavior only if Discrete Output Source is set to Flow Direction.
Table 4-2: Effect of the flow direction parameter and actual flow direction on Discrete Outputs
Actual flow direction
Flow Direction setting
Forward OFF OFF ON
Reverse OFF OFF ON
Bidirectional OFF OFF ON
Absolute Value OFF OFF ON
Negate Forward ON OFF OFF
Negate Bidirectional ON OFF OFF
Forward Zero flow Reverse
Effect of flow direction on digital communications
Flow direction affects how flow values are reported via digital communications. The following table describes the effect of the flow direction parameter and actual flow direction on flow values reported via digital communications.
Table 4-3: Effect of the flow direction on flow values
Actual flow direction
Flow Direction setting
Forward Positive 0 Negative
Reverse Positive 0 Negative
Bidirectional Positive 0 Negative
Configuration and Use Manual 37
Forward Zero flow Reverse
Page 38
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
Table 4-3: Effect of the flow direction on flow values (continued)
Actual flow direction
Flow Direction setting
Absolute Value Positive
Negate Forward Negative 0 Positive
Negate Bidirectional Negative 0 Positive
(1) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
Forward Zero flow Reverse
(1)
0 Positive
(1)
Effect of flow direction on flow totals
Flow direction affects how flow totals and inventories are calculated.
Actual flow direction
Flow Direction setting
Forward Totals increase Totals do not change Totals do not change
Reverse Totals do not change Totals do not change Totals increase
Bidirectional Totals increase Totals do not change Totals decrease
Absolute Value Totals increase Totals do not change Totals increase
Negate Forward Totals do not change Totals do not change Totals increase
Negate Bidirectional Totals decrease Totals do not change Totals increase
Forward Zero flow Reverse

4.5 Configure density measurement

The density measurement parameters control how density is measured and reported.
4.5.1 Configure Density Measurement Unit
ProLink III
Field Communicator Configure Manual Setup Measurements Density Density Unit
Density Measurement Unit controls the measurement units that will be used in density calculations and reporting.
Procedure
Set Density Measurement Unit to the option you want to use.
The default setting for Density Measurement Unit is g/cm3 (grams per cubic centimeter).
Device Tools Configuration Process Measurement Density
38 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 39
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement Unit. Different communications tools may use different labels.
Label
Unit description
Specific gravity unit
Grams per cubic centimeter g/cm3 g/Cucm
Grams per liter g/l g/L
Grams per milliliter g/ml g/mL
Kilograms per liter kg/l kg/L
Kilograms per cubic meter kg/m3 kg/Cum
Pounds per U.S. gallon lbs/Usgal lb/gal
Pounds per cubic foot lbs/ft3 lb/Cuft
Pounds per cubic inch lbs/in3 lb/CuIn
Degrees API degAPI degAPI
Short ton per cubic yard sT/yd3 STon/Cuyd
(1) Non-standard calculation. This value represents line density divided by the density of water at 60 °F.
(1)
ProLink III Field Communicator
SGU SGU
4.5.2 Configure two-phase flow parameters
ProLink III
Field Communicator Configure Manual Setup Measurements Density Slug Low Limit
Device Tools Configuration Process Measurement Density
Configure Manual Setup Measurements Density Slug High Limit
Configure Manual Setup Measurements Density Slug Duration
The two-phase flow parameters control how the transmitter detects and reports two-phase flow (gas in a liquid process or liquid in a gas process).
Note
Two-phase flow is also referred to as slug flow.
Procedure
1. Set Two-Phase Flow Low Limit to the lowest density value that is considered normal in your process.
Values below this will cause the transmitter to post Alert A105 (Two-Phase Flow).
Tip
Gas entrainment can cause your process density to drop temporarily. To reduce the occurrence of two­phase flow alerts that are not significant to your process, set Two-Phase Flow Low Limit slightly below your expected lowest process density.
You must enter Two-Phase Flow Low Limit in g/cm³, even if you configured another unit for density measurement.
Configuration and Use Manual 39
Page 40
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
The default value for Two-Phase Flow Low Limit is 0.0 g/cm³. The range is 0.0 to 10.0 g/cm³.
2. Set Two-Phase Flow High Limit to the highest density value that is considered normal in your process.
Micro Motion recommends leaving Two-Phase Flow High Limit at the default value.
Values above this will cause the transmitter to post Alert A105 (Two-Phase Flow).
You must enter Two-Phase Flow High Limit in g/cm³, even if you configured another unit for density measurement.
The default value for Two-Phase Flow High Limit is 5.0 g/cm³. The range is 0.0 to 10.0 g/cm³.
3. Set Two-Phase Flow Timeout to the number of seconds that the transmitter will wait for a two-phase
flow condition to clear before posting the alert.
The default value for Two-Phase Flow Timeout is 0.0 seconds, meaning that the alert will be posted immediately. The range is 0.0 to 60.0 seconds.
The Two-Phase Flow alert is set immediately. The flow rate will hold the last measured value for the Timeout time. Then the flow rate will report zero flow. If the density goes back in range, the error clears immediately.
Detecting and reporting two-phase flow
Two-phase flow (gas in a liquid process or liquid in a gas process) can cause a variety of process control issues. By configuring the two-phase flow parameters appropriately for your application, you can detect process conditions that require correction.
Micro Motion recommends leaving Two-Phase Flow High Limit at the default value.
A two-phase flow condition occurs whenever the measured density goes below Two-Phase Flow Low Limit or above Two-Phase Flow High Limit. If this occurs:
A two-phase flow alert is posted to the active alert log.
All outputs that are configured to represent flow rate hold their last pre-alert value for the number of
seconds configured in Two-Phase Flow Timeout.
If the two-phase flow condition clears before Two-Phase Flow Timeout expires:
Outputs that represent flow rate revert to reporting actual flow.
The two-phase flow alert is deactivated, but remains in the active alert log until it is acknowledged.
If the two-phase flow condition does not clear before Two-Phase Flow Timeout expires, the outputs that represent flow rate report a flow rate of 0.
If Two-Phase Flow Timeout is set to 0.0 seconds, the outputs that represent flow rate will report a flow rate of 0 as soon as two-phase flow is detected.
4.5.3 Configure Density Damping
ProLink III
Field Communicator Configure Manual Setup Measurements Density Density Damping
40 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Device Tools Configuration Process Measurement Density
Page 41
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Density Damping controls the amount of damping that will be applied to the line density value.
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Procedure
Effect of Density Damping on volume measurement
Density Damping affects liquid volume measurement. Liquid volume values are calculated from the damped density value rather than the measured density value. Density Damping does not affect gas standard volume measurement.
Interaction between Density Damping and Added Damping
When the mA Output is configured to report density, both Density Damping and Added Damping are applied to the reported density value.
Density Damping controls the rate of change in the value of the process variable in transmitter memory. Added Damping controls the rate of change reported via the mA Output.
If mA Output Process Variable is set to Density, and both Density Damping and Added Damping are set to non-zero values, density damping is applied first, and the added damping calculation is applied to the result of the first calculation. This value is reported over the mA Output.
4.5.4 Configure Density Cutoff
Density Cutoff specifies the lowest density value that will be reported as measured. All density values below this cutoff will be reported as 0.
Procedure
Set Density Cutoff to the value you want to use.
For most applications, the default setting (0.2 g/cm³) is sufficient. The range is 0.0 g/cm³ to 0.5 g/cm³.
Effect of Density Cutoff on volume measurement
Density Cutoff affects liquid volume measurement. If the density value goes below Density Cutoff, the volume flow rate is reported as 0. Density Cutoff does not affect gas standard volume measurement. Gas standard volume values are always calculated from the value configured for Standard Gas Density or polled value if configured for polled base density.

4.6 Configure temperature measurement

The temperature measurement parameters control how temperature data from the sensor is reported.
4.6.1 Configure Temperature Measurement Unit
ProLink III
Field Communicator Configure Manual Setup Measurements Temperature Temperature Unit
Configuration and Use Manual 41
Device Tools Configuration Process Measurement Temperature
Page 42
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
Temperature Measurement Unit specifies the unit that will be used for temperature measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
The default setting is Degrees Celsius.
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit. Different communications tools may use different labels for the units.
Label
Unit description
Degrees Celsius °C degC
Degrees Fahrenheit °F degF
Degrees Rankine °R degR
Kelvin °K Kelvin
ProLink III Field Communicator
4.6.2 Configure Temperature Damping
ProLink III Device Tools Configuration Temperature
Field Communicator Configure Manual Setup Measurements Temperature Temp Damping
Temperature Damping controls the amount of damping that will be applied to the line temperature value, when the on-board temperature data is used (RTD).
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Tip
Temperature Damping affects all process variables, compensations, and corrections that use temperature data from the sensor.
Procedure
Enter the value you want to use for Temperature Damping.
The default value is 4.8 seconds. For most applications, the default temperature damping setting is sufficient. The range is 0.0 to 38.4 seconds.
Tip
A high damping value makes the process variable appear smoother because the reported value changes
slowly.
A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
Whenever the damping value is non-zero, the reported measurement will lag the actual measurement
because the reported value is being averaged over time.
42 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 43
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the reported value.
The value you enter is automatically rounded off to the nearest valid value. Valid values for Temperature Damping are 0, 0.6, 1.2, 2.4, 4.8, 9.6, 19.2, and 38.4.
4.6.3 Effect of Temperature Damping on process measurement
Temperature Damping affects all processes and algorithms that use temperature data from the internal sensor RTD.
Temperature compensation
Temperature compensation adjusts process measurement to compensate for the effect of temperature on the sensor tubes.
Petroleum measurement
Temperature Damping affects petroleum measurement process variables only if the transmitter is configured to use temperature data from the sensor. If an external temperature value is used for petroleum measurement, Temperature Damping does not affect petroleum measurement process variables.
Concentration measurement
Temperature Damping affects concentration measurement process variables only if the transmitter is configured to use temperature data from the sensor. If an external temperature value is used for concentration measurement, Temperature Damping does not affect concentration measurement process variables.

4.7 Configure the petroleum measurement application

The petroleum measurement application corrects line density to reference temperature according to American Petroleum Institute (API) standards. The resulting process variable is referred density.
4.7.1 Configure petroleum measurement using ProLink III
The petroleum measurement parameters specify the API table, measurement units, and reference values to be used in referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process fluid.
You must know the reference temperature that you want to use.
Procedure
1. Choose Device Tools Configuration Process Measurement Petroleum Measurement.
2. Specify the API table to use to calculate referred density.
Each API table is associated with a specific set of equations.
Configuration and Use Manual 43
Page 44
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
a) Set Process Fluid to the API table group that your process fluid belongs to.
API table group Process fluids
A tables Generalized crude and JP4
B tables Generalized products: Gasoline, jet fuel, aviation fuel, kerosene, heating oils,
fuel oils, diesel, gas oil
C tables Liquids with a constant base density or known thermal expansion coefficient
(TEC). You will be required to enter the TEC for your process fluid.
D tables Lubricating oils
b) Set Referred Density Measurement Unit to the measurement units that you want to use for
referred density.
c) Click Apply.
These parameters uniquely identify the API table to be used to calculate referred density. The selected API table is displayed, and the meter automatically changes the density unit, temperature unit, pressure unit, and reference pressure to match the API table.
Your choice also determines the API table that will be used to calculate the correction factor for volume (CTL).
Restriction
Not all combinations are supported by the petroleum measurement application. See the list of API tables in this manual.
3. Refer to the API documentation and confirm your table selection.
a) Verify that your process fluid falls within range for line density, line temperature, and line
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
5. Set Reference Temperature to the temperature to which density will be corrected in referred density
calculations. If you choose Other, select the temperature measurement unit and enter the reference temperature.
4.7.2 Set up temperature data for petroleum measurement using
ProLink III
The petroleum measurement application uses temperature data in its calculations. You must decide how to provide this data, then perform the required configuration and setup.
Tip
Fixed values for temperature are not recommended. Using a fixed temperature value may produce inaccurate process data.
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART communications.
44 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 45
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
If you are using an external temperature device, it must use the temperature unit that is configured in the transmitter.
Procedure
1. Choose Device Tools Configuration Process Measurement Petroleum Measurement.
2. Choose the method to be used to supply temperature data, and perform the required setup.
Option Description Setup
Internal RTD temperature data
Polling The meter polls an external
Digital communications
Temperature data from the on­board temperature sensor (RTD) is used.
device for temperature data. This data will be available in addition to the internal RTD temperature data.
A host writes temperature data to the meter at appropriate intervals. This data will be available in addition to the internal RTD temperature data.
a. Set Line Temperature Source to Internal RTD.
b. Click Apply.
a. Set Line Temperature Source to Poll for External Value.
b. Set Polling Slot to an available slot.
c. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
network. The Field Communicator is not a HART master.
Poll as Secondary Other HART masters will be on the
network. The Field Communicator is not a HART master.
d. Set External Device Tag to the HART tag of the
temperature device.
e. Click Apply.
a. Set Line Temperature Source to Fixed Value or Digital
Communications.
b. Click Apply.
c. Perform the necessary host programming and
communications setup to write temperature data to the meter at appropriate intervals.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs group on the ProLink III main window.
Need help?
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:Verify the wiring between the meter and the external device.
Verify the HART tag of the external device.
Configuration and Use Manual 45
Page 46
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
For digital communications:Verify that the host has access to the required data.
Verify that the host is writing to the correct register in memory, using the correct data type.
4.7.3 Configure petroleum measurement using the Field
Communicator
Procedure
1. Choose Online Configure Manual Setup Measurements Set Up Petroleum.
2. Specify the API table to use.
a) Open the Petroleum Measurement Source menu and select the API table number.
Depending on your choice, you may be prompted to enter a reference temperature or a thermal expansion coefficient.
b) Enter the API table letter.
These two parameters uniquely specify the API table.
3. Determine how the transmitter will obtain temperature data for the petroleum measurement
calculations, and perform the required setup.
Option Setup
Temperature data from the sensor
A user-configured static temperature value
Polling for temperature a. Ensure that the primary mA output has been wired to support HART
a. Choose Online Configure Manual Setup Measurements
External Pressure/Temperature Temperature.
b. Set External Temperature to Disabled.
a. Choose Online Configure Manual Setup Measurements
External Pressure/Temperature Temperature.
b. Set External Temperature to Enabled.
c. Set Correction Temperature to the value to be used.
polling.
b. Choose Online Configure Manual Setup Measurements
External Pressure/Temperature Temperature.
c. Set External Temperature to Enabled.
d. Choose External Polling.
e. Set Poll Control to Poll As Primary or Poll as Secondary.
f. Determine whether you will use Polling Slot 1 or Polling Slot 2.
g. For the chosen slot, set Ext Dev Tag to the HART tag of the external
temperature device.
h. For the chosen slot, set Polled Variable to Temperature.
46 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 47
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Option Setup
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network. The Field
Communicator is not a HART master.
A value written by digital communications
a. Choose Online Configure Manual Setup Measurements
External Pressure/Temperature Temperature.
b. Set External Temperature to Enabled.
c. Perform the necessary host programming and communications setup
to write temperature data to the transmitter at appropriate intervals.
Note
If the Weights & Measures application is implemented and the transmitter is secured, digital communications cannot be used to write temperature or pressure data to the transmitter.
4.7.4 API tables supported by the petroleum measurement application
The API tables listed here are supported by the petroleum measurement application.
Table name Process fluid CTL source data Reference temperature Density unit
5A Generalized crude and
JP4
5B Generalized products Observed density and
Observed density and observed temperature
observed temperature
60 °F (configurable) Degrees API
Range: 0 to 100
60 °F (configurable) Degrees API
Range: 0 to 85
5D Lubricating oils Observed density and
observed temperature
6C Liquids with a constant
density base or known thermal expansion coefficient
23A Generalized crude and
JP4
23B Generalized products Observed density and
23D Lubricating oils Observed density and
Configuration and Use Manual 47
User-supplied reference density (or thermal expansion coefficient) and observed temperature
Observed density and observed temperature
observed temperature
observed temperature
60 °F (configurable) Degrees API
Range: 10 to +45
60 °F (configurable) Degrees API
60 °F (configurable) Relative density
Range: 0.6110 to 1.0760
60 °F (configurable) Relative density
Range: 0.6535 to 1.0760
60 °F (configurable) Relative density
Range: 0.8520 to 1.1640
Page 48
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
Table name Process fluid CTL source data Reference temperature Density unit
24C Liquids with a constant
density base or known thermal expansion coefficient
53A Generalized crude and
JP4
53B Generalized products Observed density and
53D Lubricating oils Observed density and
54C Liquids with a constant
density base or known thermal expansion coefficient
User-supplied reference density (or thermal expansion coefficient) and observed temperature
Observed density and observed temperature
observed temperature
observed temperature
User-supplied reference density (or thermal expansion coefficient) and observed temperature
60 °F (configurable) Relative density
15 °C (configurable) Base density
Range: 610 to 1075 kg/m
15 °C (configurable) Base density
Range: 653 to 1075 kg/m
15 °C (configurable) Base density
Range: 825 to 1164 kg/m
15 °C (configurable) Base density in kg/m
3
3
3
3
Restriction
These tables are not appropriate for the following process fluids: propane and propane mixes, butane and butane mixes, butadiene and butadiene mixes, isopentane, LNG, LPG, NGL, ethylene, propylene, cyclohexane, aeromatics, asphalts, and road tars.

4.8 Set up concentration measurement

This section guides you through loading and setting up a concentration matrix used for measurement. It does not cover building a concentration matrix.
The concentration measurement application calculates concentration data from process temperature and density. Micro Motion provides a set of concentration matrices that provide the reference data for several standard industry applications and process fluids. If desired, you can build a custom matrix for your process fluid, or purchase a custom matrix from Micro Motion.
Note
Concentration matrices can be made available on your transmitter either by loading an existing matrix from a file or by building a new matrix. Up to 6 matrices can be available on your transmitter, but only 1 can be used for measurement at any given time. For detailed information on building a matrix, see the .
Prerequisites
Before you can configure concentration measurement:
The concentration measurement application must be purchased on your transmitter.
The concentration matrix you want to use must be available on your transmitter, or it must be available as
a file on your computer.
48 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 49
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
You must know the derived variable that your matrix is designed for.
You must know the density unit used by your matrix.
You must know the temperature unit used by your matrix.
The concentration measurement application must be unlocked.
4.8.1 Configure concentration measurement using ProLink III
Procedure
1. Choose Device Tools Configuration Process Measurement Density and set Density Unit to
the density unit used by your matrix.
2. Choose Device Tools Configuration Process Measurement Temperature and set
Temperature Unit to the temperature unit used by your matrix.
3. Choose Device Tools Configuration Process Measurement Concentration Measurement.
4. Set Derived Variable to the derived variable that your matrix is designed for, and click Apply.
Important
All concentration matrices on your transmitter must use the same derived variable. If you are using
one of the standard matrices from Micro Motion, set Derived Variable to Mass Concentration (Density). If you are using a custom matrix, see the reference information for your matrix.
If you change the setting of Derived Variable, all existing concentration matrices will be deleted
from transmitter memory. Set Derived Variable before loading concentration matrices.
5. Load one or more matrices. a) Set Matrix Being Configured to the location to which the matrix will be loaded.
b) Click Load Matrix from a File, navigate to the matrix file on your computer, and load it.
c) Repeat until all required matrices are loaded.
6. Configure or review matrix data. a) If necessary, set Matrix Being Configured to the matrix you want to configure or review, and
click Change Matrix.
b) Set Concentration Unit to the label that will be used for the concentration unit.
c) If you set Concentration Unit to Special, enter the custom label.
d) If desired, change the matrix name.
e) Review the data points for this matrix.
f) Do not change Reference Temperature or Curve Fit Maximum Order.
g) If you changed any matrix data, click Apply.
7. Set up extrapolation alarms.
Each concentration matrix is built for a specific density range and a specific temperature range. If process density or process temperature goes outside the range, the transmitter will extrapolate
Configuration and Use Manual 49
Page 50
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
concentration values. However, extrapolation may affect accuracy. Extrapolation alarms are used to notify the operator that extrapolation is occurring.
a) If necessary, set Matrix Being Configured to the matrix you want to view, and select Change
Matrix.
b) Set Extrapolation Alarm Limit to the point, in percent, at which an extrapolation alarm will be
posted.
c) Enable or disable the high and low limit alarms for temperature and density, as desired, and
select Apply.
Restriction
The high and low limit alarms require the enhanced core processor.
Example
If Extrapolation Alarm Limit is set to 5%, High Extrapolation Limit (Temperature) is enabled, and the matrix is built for a temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), an extrapolation alarm will be posted if process temperature goes above 82 °F (27.8 °C).
8. Set Temperature Source to the method that the transmitter will use to obtain temperature data.
Option Description
Poll for external value
(1)
The transmitter will poll an external temperature device, using HART protocol over the primary mA Output.
RTD The transmitter will use the temperature data from the sensor.
Static or Digital Communications
The transmitter will use the temperature value that it reads from memory.
Static: The configured value is used. (Not recommended.)
Digital Communications: A host writes transmitter data to
transmitter memory.
Note
If the Weights & Measures application is implemented and the transmitter is secured, digital communications cannot be used to write temperature or pressure data to the transmitter.
9. If you chose RTD, no more configuration is required. Select Apply and exit.
10. If you chose to poll for temperature data: a) Select the Polling Slot to use.
b) Set Polling Control to Poll as Primary or Poll as Secondary, and click Apply.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network. The Field Communicator is
not a HART master.
50 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 51
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
c) Set External Device Tag to the HART tag of the external temperature device, and select Apply.
11. If you chose to use a static temperature value, set External Temperature to the value to use, and select Apply.
12. If you want to use digital communications, select Apply, then perform the necessary host programming and communications setup to write temperature data to the transmitter at appropriate intervals.
13. Set Active Matrix to the matrix to be used for measurement.
Concentration process variables are now available on the transmitter. You can view and report them in the same way that you view and report other process variables.
4.8.2 Configure concentration measurement using the Field Communicator
Procedure
1. Choose Online Configure Manual Setup Measurements Density and set Density Unit to match the density unit used by your matrix.
2. Choose Online Configure Manual Setup Measurements Temperature and set Temperature Unit to match the temperature unit used by your matrix.
3. Choose Online Configure Manual Setup Measurements and click Concentration Measurement.
4. Enable or disable matrix switching, as desired.
5. Set up extrapolation alerts. Each concentration matrix is built for a specific density range and a specific temperature range. If
process density or process temperature goes outside the range, the transmitter will extrapolate concentration values. However, extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is occurring.
a) Click Next.
b) On the Matrix Configuration page, set Matrix Being Configured to the matrix that you want to
configure.
c) Modify the matrix name if desired.
d) Set Extrapolation Alert Limit to the point, in percent, at which an extrapolation alert will be
posted.
e) Choose Online Configure Alert Setup CM Alerts.
f) Enable or disable the high and low limit alarms for temperature and density, as desired.
Restriction
The high and low limit alarms require the enhanced core processor.
Example
If Alarm Limit is set to 5%, the high-temperature extrapolation alert is enabled, and the matrix is built for a temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), an extrapolation alarm will be posted if process temperature goes above 82 °F (27.8 °C).
Configuration and Use Manual 51
Page 52
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
6. Select the label that will be used for the concentration unit.
a) Click Next.
b) On the Concentration Measurement page, set Concentration Units to the desired label.
c) Set Concentration Units to the desired label.
d) If you set Units to Special, enter the custom label.
e) Click Finish.
7. Determine how the transmitter will obtain temperature data for the concentration measurement calculations, and perform the required setup.
Option Setup
Temperature data from the sensor
A user-configured static temperature value
Polling for temperature
(1)
a. Choose Online Configure Manual Setup Measurements.
b. Click External Inputs.
c. Click Next.
d. Disable External Temperature.
a. Choose Online Configure Manual Setup Measurements.
b. Click External Inputs.
c. Click Next.
d. Enable External Temperature.
e. Set Correction Temperature to the value to be used.
a. Ensure that the primary mA output has been wired to support HART
polling.
b. Choose Online Configure Manual Setup Measurements.
c. Click External Inputs.
d. Click Next.
e. Enable External Temperature.
f. Click Next.
g. Choose an unused polling slot.
h. Set Poll Control to Poll As Primary Host or Poll as Secondary Host.
i. Set External Tag to the HART tag of the external temperature device.
j. Set Polled Variable to Temperature.
Tip
Poll as Primary: No other HART masters will be on the network.
52 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 53
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Option Setup
Poll as Secondary: Other HART masters will be on the network. The Field
Communicator is not a HART master.
A value written by digital communications
a. Choose Online Configure Manual Setup Measurements
External Pressure/Temperature Temperature.
b. Enable External Temperature.
c. Perform the necessary host programming and communications setup
to write temperature data to the transmitter at appropriate intervals.
Note
If the Weights & Measures application is implemented and the transmitter is secured, digital communications cannot be used to write temperature or pressure data to the transmitter.
8. Choose Online Configure Manual Setup Measurements Conc Measurement (CM) CM Configuration and set Active Matrix to the matrix to be used for measurement.
Concentration process variables are now available on the transmitter. You can view and report them in the same way that you view and report other process variables.
4.8.3 Standard matrices for the concentration measurement application
The standard concentration matrices available from Micro Motion are applicable for a variety of process fluids. These matrices are included in the ProLink III installation folder.
Tip
If the standard matrices are not appropriate for your application, you can build a custom matrix or purchase a custom matrix from Micro Motion.
Matrix name
Deg Balling Matrix represents percent extract, by
Deg Brix Matrix represents a hydrometer scale
Configuration and Use Manual 53
Description Density unit
3
g/cm mass, in solution, based on °Balling. For example, if a wort is 10 °Balling and the extract in solution is 100% sucrose, the extract is 10% of the total mass.
3
g/cm for sucrose solutions that indicates the percent by mass of sucrose in solution at a given temperature. For example, 40 kg of sucrose mixed with 60 kg of water results in a 40 °Brix solution.
Temperature unit
°F Mass
°C Mass
Derived variable
Concentration (Density)
Concentration (Density)
Page 54
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
Matrix name Description Density unit
Deg Plato Matrix represents percent extract, by
mass, in solution, based on °Plato. For example, if a wort is 10 °Plato and the extract in solution is 100% sucrose, the extract is 10% of the total mass.
HFCS 42 Matrix represents a hydrometer scale
for HFCS 42 (high-fructose corn syrup) solutions that indicates the percent by mass of HFCS in solution.
HFCS 55 Matrix represents a hydrometer scale
for HFCS 55 (high-fructose corn syrup) solutions that indicates the percent by mass of HFCS in solution.
HFCS 90 Matrix represents a hydrometer scale
for HFCS 90 (high-fructose corn syrup) solutions that indicates the percent by mass of HFCS in solution.
g/cm
g/cm
g/cm
g/cm
3
3
3
3
Temperature unit
°F Mass
°C Mass
°C Mass
°C Mass
Derived variable
Concentration (Density)
Concentration (Density)
Concentration (Density)
Concentration (Density)
4.8.4 Derived variables and calculated process variables
The concentration measurement application calculates a different set of process variables from each derived variable. The process variables are then available for viewing or reporting.
Derived variable Description
Density at Reference
Specific Gravity The ratio of the density
Mass Concentration (Density)
Mass/unit volume, corrected to a given reference temperature
of a process fluid at a given temperature to the density of water at a given temperature
Note
The two given temperature conditions do not need to be the same.
The percent mass of solute or of material in suspension in the total solution, derived from reference density
Calculated process variables
Density at reference temp
Standard volume flow rate
Specific gravity
ConcentrationNet mass
flow rate
Net volume flow rate
54 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 55
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
Calculated process variables
Derived variable Description
Mass Concentration (Specific Gravity)
Volume Concentration (Density)
Volume Concentration (Specific Gravity)
Concentration (Density)
The percent mass of solute or of material in suspension in the total solution, derived from specific gravity
The percent volume of solute or of material in suspension in the total solution, derived from reference density
The percent volume of solute or of material in suspension in the total solution, derived from specific gravity
The mass, volume, weight, or number of moles of solute or of material in suspension in proportion to the total solution, derived from reference density
Density at reference temp
Standard volume flow rate
Specific gravity
ConcentrationNet mass
flow rate
Net volume flow rate
Concentration (Specific Gravity)
The mass, volume, weight, or number of moles of solute or of material in suspension in proportion to the total solution, derived from specific gravity

4.9 Configure pressure compensation

Pressure compensation adjusts process measurement to compensate for the pressure effect on the sensor. The pressure effect is the change in the sensor’s sensitivity to flow and density caused by the difference between the calibration pressure and the process pressure.
Tip
Not all sensors or applications require pressure compensation. The pressure effect for a specific sensor model can be found in the product data sheet located at . If you are uncertain about implementing pressure compensation, contact customer service.
Prerequisites
You will need the flow factor, density factor, and calibration pressure values for your sensor.
For the flow factor and density factor, see the product data sheet for your sensor.
For the calibration pressure, see the calibration sheet for your sensor. If the data is unavailable, use 20 PSI.
Configuration and Use Manual 55
Page 56
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
4.9.1 Configure pressure compensation using ProLink III
Procedure
1. Choose Device Tools Configuration Process Measurement Pressure Compensation.
2. Set Pressure Compensation Status to Enabled.
3. Set Pressure Unit to the appropriate unit.
If you will use an external pressure value, set Pressure Unit to match the pressure unit used by the external pressure device.
4. Enter Flow Calibration Pressure for your sensor.
The calibration pressure is the pressure at which your sensor was calibrated, and defines the pressure at which there is no pressure effect. If the data is unavailable, enter 20 PSI.
5. Enter Flow Factor for your sensor.
The flow factor is the percent change in the flow rate per PSI. When entering the value, reverse the sign.
Example
If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI.
6. Enter Density Factor for your sensor.
The density factor is the change in fluid density, in g/cm3/PSI. When entering the value, reverse the sign.
Example
If the density factor is 0.000006 g/cm3/PSI, enter +0.000006 g/cm3/PSI.
7. Set Pressure Source to the method that the transmitter will use to obtain pressure data.
Option
Poll for external value The transmitter will poll an external pressure device, using HART
Fixed Value or Digital Communications
Description
protocol over the primary mA Output.
The transmitter will use the pressure value that it reads from memory.
Fixed Value: The configured value is used.
Digital Communications: A host writes transmitter data to
transmitter memory.
Note
If the Weights & Measures application is implemented and the transmitter is secured, digital communications cannot be used to write temperature or pressure data to the transmitter.
8. If you chose to poll for pressure data: a) Select the Polling Slot to use.
56 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 57
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
b) Set Polling Control to Poll as Primary or Poll as Secondary, and click Apply.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network. The Field Communicator is
not a HART master.
c) Set External Device Tag to the HART tag of the external pressure device, and click Apply.
d) Ensure that the primary mA Output is wired to support HART communications with the external
pressure device.
9. If you chose to use a fixed pressure value: a) Set Fixed Value to the value to use, and click Apply
10. If you want to use digital communications, click Apply, then perform the necessary host programming and communications setup to write pressure data to the transmitter at appropriate intervals.
Postrequisites
If you are using an external pressure value, verify the setup by checking the External Pressure value displayed in the Inputs area of the main window.
4.9.2 Configure pressure compensation using the Field Communicator
Procedure
1. Choose OnlineConfigureManual SetupMeasurementsExternal Pressure/TemperaturePressure.
2. Set Pressure Compensation to Enabled.
3. Enter Flow Cal Pressure for your sensor.
The calibration pressure is the pressure at which your sensor was calibrated, and defines the pressure at which there is no pressure effect. If the data is unavailable, enter 20 PSI.
4. Enter Flow Press Factor for your sensor.
The flow factor is the percent change in the flow rate per PSI. When entering the value, reverse the sign.
Example
If the flow factor is 0.0002 % per PSI, enter +0.0002 % per PSI.
5. Enter Dens Press Factor for your sensor.
The density factor is the change in fluid density, in g/cm3/PSI. When entering the value, reverse the sign.
Configuration and Use Manual 57
Page 58
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
Example
If the density factor is 0.000006 g/cm3/PSI, enter +0.000006 g/cm3/PSI.
6. Determine how the transmitter will obtain pressure data, and perform the required setup.
Option Setup
A user-configured static pressure value
a. Set Pressure Unit to the desired unit.
b. Set Compensation Pressure to the desired value.
Polling for pressure a. Ensure that the primary mA Output has been wired to support HART
polling.
b. Choose Online Configure Manual Setup Measurements
External Pressure/Temperature External Polling.
c. Set Poll Control to Poll As Primary Host or Poll as Secondary Host.
d. Choose an unused polling slot.
e. Set External Tag to the HART tag of the external pressure device.
f. Set Polled Variable to Pressure.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network. The Field
Communicator is not a HART master.
A value written by digital communications
a. Set Pressure Unit to the desired unit.
b. Perform the necessary host programming and communications setup
to write pressure data to the transmitter at appropriate intervals.
Note
If the Weights & Measures application is implemented and the transmitter is secured, digital communications cannot be used to write temperature or pressure data to the transmitter.
Postrequisites
If you are using an external pressure value, verify the setup by choosing Service Tools Variables External Variables and checking the value displayed for External Pressure.
58 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 59
Configuration and Use Manual Configure process measurement
MMI-20019038 February 2022
4.9.3 Options for Pressure Measurement Unit
The transmitter provides a standard set of measurement units for Pressure Measurement Unit. Different communications tools may use different labels for the units. In most applications, Pressure Measurement Unit should be set to match the pressure measurement unit used by the remote device.
Label
Unit description
Feet water @ 68 °F Ft Water @ 68°F ftH2O
Inches water @ 4 °C In Water @ 4°C inH2O @4DegC
Inches water @ 60 °F In Water @ 60°F inH2O @60DegF
Inches water @ 68 °F In Water @ 68°F inH2O
Millimeters water @ 4 °C mm Water @ 4°C mmH2O @4DegC
Millimeters water @ 68 °F mm Water @ 68°F mmH2O
Millimeters mercury @ 0 °C mm Mercury @ 0°C mmHg
Inches mercury @ 0 °C In Mercury @ 0°C inHG
Pounds per square inch PSI psi
Bar bar bar
Millibar millibar mbar
Grams per square centimeter g/cm2 g/Sqcm
Kilograms per square centimeter kg/cm2 kg/Sqcm
Pascals pascals Pa
Kilopascals Kilopascals kPa
Megapascals Megapascals MPa
ProLink III Field Communicator
Torr @ 0 °C Torr @ 0°C torr
Atmospheres atms atms
Configuration and Use Manual 59
Page 60
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019038
60 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 61
Configuration and Use Manual Configure device options and preferences
MMI-20019038 February 2022

5 Configure device options and preferences

5.1 Configure response time parameters

You can configure the rate at which process data is polled and process variables are calculated.
5.1.1 Configure Update Rate
ProLink III Device Tools Configuration Process Measurement Response Update Rate
Field Communicator Configure Manual Setup Measurements Update Rate
Update Rate controls the rate at which process data is polled and process variables are calculated. Update Rate = Special produces faster and “noisier” response to changes in the process. Do not use Special mode
unless required by your application.
Prerequisites
Before setting Update Rate to Special:
Check the effects of Special mode on specific process variables.
Contact customer support.
Tip
For systems with a standard core processor, Special mode can improve performance for applications with entrained air or Empty-Full-Empty conditions. This does not apply to systems with an enhanced core processor.
Procedure
1. Set Update Rate as desired.
Option
Normal All process data is polled at the rate of 20 times per second (20 Hz).
Special A single, user-specified process variable is polled at the rate of 100 times per second
Description
All process variables are calculated at 20 Hz. This option is appropriate for most applications.
(100 Hz). Other process data is polled at 6.25 Hz. Some process, diagnostic, and calibration data is not polled.
All available process variables are calculated at 100 Hz. Use this option only if required by your application.
If you change Update Rate, the settings for Flow Damping, and Density Damping are automatically adjusted.
2. If you set Update Rate to Special, select the process variable to be polled at 100 Hz.
Configuration and Use Manual 61
Page 62
Configure device options and preferences Configuration and Use Manual
February 2022 MMI-20019038
Effects of Update Rate = Special
Incompatible features and functions
Special mode is not compatible with the following features and functions:
Enhanced events. Use basic events instead.
All calibration procedures.
Zero verification.
Restoring the factory zero or the prior zero.
If required, you can switch to Normal mode, perform the desired procedures, and then return to Special mode.
Process variable updates
Some process variables are not updated when Special mode is enabled.
Table 5-1: Special mode and process variable updates
Always polled and updated Updated only when the petroleum
measurement application is disabled
Mass flow
Volume flow
Gas standard volume flow
Density
Temperature
Drive gain
LPO amplitude
Status [contains Event 1 and Event
2 (basic events)]
Mass total
Volume total
Live zero
Gas standard volume total
Temperature-corrected volume
total
Temperature-corrected density
Temperature-corrected volume
flow
Batch-weighted average
temperature
Batch-weighted average density
RPO amplitude
Core input voltage
Mass inventory
Volume inventory
Gas standard volume inventory
Never updated
All other process variables and calibration data. They retain the values held at the time you enabled Special mode.
62 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 63
Configuration and Use Manual Configure device options and preferences
MMI-20019038 February 2022
5.1.2 Configure Response Time
ProLink III Device Tools Configuration Process Measurement Response Response Time
Field Communicator Not available
Response Time is used to apply a different algorithm to the calculation of process variables from the raw process data.
Restriction
Response Time is available only on systems with the enhanced core processor.
Procedure
Set Response Time as desired.
Option Description
Normal (Legacy) Transmitter calculates process variables at the standard speed. This option is
selected if this parameter was configured on an earlier version of ProLink III software.
Special (Legacy) Transmitter calculates process variables at a faster speed. This option is
selected if this parameter was configured on an earlier version of ProLink III software.
Normal - Optimal Filtering Transmitter calculates process variables at standard filtering and speed.
Low Filtering - Fastest Response
High Filtering - Smoothest Output
Service For factory use only.
Transmitter calculates process variables at the fastest speed.
Transmitter calculates process variables at the smoothest (least noisy) response to changes in the process.

5.2 Configure alert handling

The alert handling parameters control the transmitter’s response to process and device conditions.
5.2.1 Configure Fault Timeout
ProLink III
Field Communicator Configure Alert Setup Alert Severity Fault Timeout
Fault Timeout controls the delay before fault actions are performed.
Device Tools Configuration Fault Processing
Restriction
Fault Timeout is applied only to the following alerts (listed by Status Alert Code): A003, A004, A005, A008, A016, A017, A033. For all other alerts, fault actions are performed as soon as the alert is detected.
Configuration and Use Manual 63
Page 64
Configure device options and preferences Configuration and Use Manual
February 2022 MMI-20019038
Procedure
Set Fault Timeout as desired.
The default value is 0 seconds. The range is 0 to 60 seconds.
If you set Fault Timeout to 0, fault actions are performed as soon as the alert condition is detected.
The fault timeout period begins when the transmitter detects an alert condition. During the fault timeout period, the transmitter continues to report its last valid measurements.
If the fault timeout period expires while the alert is still active, the fault actions are performed. If the alert condition clears before the fault timeout expires, no fault actions are performed.
5.2.2 Configure Status Alert Severity
ProLink III Device Tools Configuration Alert Severity
Field Communicator Configure Alert Setup Alert Severity Set Alert Severity
Use Status Alert Severity to control the fault actions that the transmitter performs when it detects an alert condition.
Restriction
For some alerts, Status Alert Severity is not configurable.
For some alerts, Status Alert Severity can be set only to two of the three options.
Tip
Use the default settings for Status Alert Severity unless you have a specific requirement to change them.
Procedure
1. Select a status alert.
2. For the selected status alert, set Status Alert Severity as desired.
Option
Fault Actions when fault is detected:
Description
The alert is posted to the Alert List.
Outputs go to the configured fault action (after Fault Timeout has expired, if
applicable).
Digital communications go to the configured fault action (after Fault Timeout has
expired, if applicable).
The status LED (if available) changes to red or yellow (depending on alert severity).
Actions when alert clears:
Outputs return to normal behavior.
Digital communications return to normal behavior.
The status LED (if available) returns to green and may or may not flash.
Informational Actions when fault is detected:
64 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 65
Configuration and Use Manual Configure device options and preferences
MMI-20019038 February 2022
Option Description
The alert is posted to the Alert List.
The status LED (if available) changes to red or yellow (depending on alert severity).
Actions when alert clears:
The status LED (if available) returns to green and may or may not flash.
Ignore No action
Status alerts and options for Status Alert Severity
Table 5-2: Status alerts and Status Alert Severity
Alert code Status message Default severity Notes Configurable?
A001 EEPROM Error Fault No
A002 RAM Error Fault No
A003 No Sensor Response Fault Yes
A004 Temperature Overrange Fault No
A005 Mass Flow Rate Overrange Fault Yes
A006 Characterization Required Fault Yes
A008 Density Overrange Fault Yes
A009 Transmitter Initializing/
Warming Up
A010 Calibration Failure Fault No
A011 Zero Calibration Failed:
Low
A012 Zero Calibration Failed:
High
A013 Zero Calibration Failed:
Unstable
A014 Transmitter Failure Fault No
A016 Sensor RTD Failure Fault Yes
A017 T-Series RTD Failure Fault Yes
A018 EEPROM Error
(Transmitter)
A019 RAM Error (Transmitter) Fault No
Fault Yes
Fault Yes
Fault Yes
Fault Yes
Fault No
A020 No Flow Cal Value Fault Yes
A021 Incorrect Sensor Type (K1) Fault No
A022 Configuration Database
Corrupt (Core Processor)
Configuration and Use Manual 65
Fault Applies only to flowmeters with the
standard core processor.
No
Page 66
Configure device options and preferences Configuration and Use Manual
February 2022 MMI-20019038
Table 5-2: Status alerts and Status Alert Severity (continued)
Alert code Status message Default severity Notes Configurable?
A023 Internal Totals Corrupt
(Core Processor)
A024 Program Corrupt (Core
Processor)
A025 Boot Sector Fault (Core
Processor)
A026 Sensor/Transmitter
Communications Failure
A027 Security Breach Fault No
A028 Core Processor Write
Failure
A031 Low Power Fault Applies only to flowmeters with the
A032 Meter Verification in
Progress: Outputs to Fault
A033 Insufficient Right/Left
Pickoff Signal
A034 Meter Verification Failed Fault Applies only to transmitters with
Fault Applies only to flowmeters with the
standard core processor.
Fault Applies only to flowmeters with the
standard core processor.
Fault Applies only to flowmeters with the
standard core processor.
Fault No
Fault No
enhanced core processor.
Varies Applies only to transmitters with
Smart Meter Verification. If outputs are set to Last Measured
Value, severity is Info. If outputs are set to Fault, severity is Fault.
Fault Applies only to flowmeters with the
enhanced core processor.
Smart Meter Verification.
No
No
No
No
No
Yes
Yes
A035 Meter Verification Aborted Fault Applies only to transmitters with
Smart Meter Verification.
A100 mA Output 1 Saturated Informational Can be set to either Informational
or Ignore, but cannot be set to Fault.
A101 mA Output 1 Fixed Informational Can be set to either Informational
or Ignore, but cannot be set to Fault.
A102 Drive Overrange Informational Yes
A103 Data Loss Possible (Totals
and Inventories)
A104 Calibration in Progress Informational Can be set to either Informational
A105 Slug Flow Informational Yes
Informational Applies only to flowmeters with the
enhanced core processor. Can be set to either Informational
or Ignore, but cannot be set to Fault.
or Ignore, but cannot be set to Fault.
Yes
Yes
Yes
Yes
Yes
66 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 67
Configuration and Use Manual Configure device options and preferences
MMI-20019038 February 2022
Table 5-2: Status alerts and Status Alert Severity (continued)
Alert code Status message Default severity Notes Configurable?
A106 Burst Mode Enabled Informational Can be set to either Informational
or Ignore, but cannot be set to Fault.
A107 Power Reset Occurred Informational Normal transmitter behavior;
occurs after every power cycle.
A108 Basic Event 1 On Informational Applies only to basic events. Yes
A109 Basic Event 2 On Informational Applies only to basic events. Yes
A110 Frequency Output
Saturated
A111 Frequency Output Fixed Informational Can be set to either Informational
A112 Upgrade Transmitter
Software
A113 mA Output 2 Saturated Informational Can be set to either Informational
A114 mA Output 2 Fixed Informational Can be set to either Informational
Informational Can be set to either Informational
or Ignore, but cannot be set to Fault.
or Ignore, but cannot be set to Fault.
Informational Applies only to systems with
transmitter software earlier than v5.0.
or Ignore, but cannot be set to Fault.
or Ignore, but cannot be set to Fault.
Yes
Yes
Yes
Yes
Yes
Yes
Yes
A115 No External Input or Polled
Data
A116 Temperature Overrange
(Petroleum)
A117 Density Overrange
(Petroleum)
A118 Discrete Output 1 Fixed Informational Can be set to either Informational
A119 Discrete Output 2 Fixed Informational Can be set to either Informational
A120 Curve Fit Failure
(Concentration)
A121 Extrapolation Alarm
(Concentration)
Informational Yes
Informational Applies only to transmitters with
the petroleum measurement application.
Informational Applies only to transmitters with
the petroleum measurement application.
or Ignore, but cannot be set to Fault.
or Ignore, but cannot be set to Fault.
Informational Applies only to transmitters with
the concentration measurement application.
Informational Applies only to transmitters with
the concentration measurement application.
Yes
Yes
Yes
Yes
No
Yes
Configuration and Use Manual 67
Page 68
Configure device options and preferences Configuration and Use Manual
February 2022 MMI-20019038
Table 5-2: Status alerts and Status Alert Severity (continued)
Alert code Status message Default severity Notes Configurable?
A131 Meter Verification in
Progress: Outputs to Last Measured Value
A132 Sensor Simulation Active Informational Applies only to flowmeters with the
A141 DDC trigger(s) have
completed
Informational Applies only to transmitters with
Smart Meter Verification.
enhanced core processor. Can be set to either Informational
or Ignore, but cannot be set to Fault.
Informational Applies only to flowmeters with the
enhanced core processor. Can be set to either Informational
or Ignore, but cannot be set to Fault.
Yes
To Informational or Ignore only
Yes

5.3 Configure informational parameters

The informational parameters can be used to identify or describe your meter. They are not used in process measurement and they are not required.
5.3.1 Configure Sensor Serial Number
ProLink III
Field Communicator Configure Manual Setup Info Parameters Sensor Information Sensor Serial Number
Device Tools Configuration Informational Parameters Sensor
Sensor Serial Number lets you store the serial number of the sensor component of your flowmeter in
transmitter memory. This parameter is not used in processing and is not required.
Procedure
1. Obtain the sensor serial number from your sensor tag.
2. Enter the serial number in the Sensor Serial Number field.
5.3.2 Configure Sensor Material
ProLink III
Field Communicator Configure Manual Setup Info Parameters Sensor Information Tube Wetted Material
Sensor Material lets you store the type of material used for your sensor’s wetted parts in transmitter memory. This parameter is not used in processing and is not required.
Procedure
1. Obtain the material used for your sensor’s wetted parts from the documents shipped with your sensor, or from a code in the sensor model number.
To interpret the model number, refer to the product data sheet for your sensor.
Device Tools Configuration Informational Parameters Sensor
68 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 69
Configuration and Use Manual Configure device options and preferences
MMI-20019038 February 2022
2. Set Sensor Material to the appropriate option.
5.3.3 Configure Sensor Liner Material
ProLink III Device Tools Configuration Informational Parameters Sensor
Field Communicator Configure Manual Setup Info Parameters Sensor Information Tube Lining
Sensor Liner Material lets you store the type of material used for your sensor liner in transmitter memory. This parameter is not used in processing and is not required.
Procedure
1. Obtain your sensor’s liner material from the documents shipped with your sensor, or from a code in the sensor model number.
To interpret the model number, refer to the product data sheet for your sensor.
2. Set Sensor Liner Material to the appropriate option.
5.3.4 Configure Sensor Flange Type
ProLink III Device Tools Configuration Informational Parameters Sensor
Field Communicator Configure Manual Setup Info Parameters Sensor Information Sensor Flange
Sensor Flange Type lets you store your sensor’s flange type in transmitter memory. This parameter is not used in processing and is not required.
Procedure
1. Obtain your sensor’s flange type from the documents shipped with your sensor, or from a code in the sensor model number.
To interpret the model number, refer to the product data sheet for your sensor.
2. Set Sensor Flange Type to the appropriate option.
5.3.5 Configure Descriptor
ProLink III
Field Communicator Configure Manual Setup Info Parameters Transmitter Info Descriptor
Descriptor lets you store a description in transmitter memory. The description is not used in processing and is not required.
Procedure
Enter a description for the transmitter or device You can use up to 16 characters for the description.
Device Tools Configuration Informational Parameters Transmitter
5.3.6 Configure Message
ProLink III
Configuration and Use Manual 69
Device Tools Configuration Informational Parameters Transmitter
Page 70
Configure device options and preferences Configuration and Use Manual
February 2022 MMI-20019038
Field Communicator Configure Manual Setup Info Parameters Transmitter Info Message
Message lets you store a short message in transmitter memory. This parameter is not used in processing and is not required.
Procedure
Enter a short message for the transmitter or device. Your message can be up to 32 characters long.
5.3.7 Configure Date
ProLink III Device Tools Configuration Informational Parameters Transmitter
Field Communicator Configure Manual Setup Info Parameters Transmitter Info Date
Date lets you store a static date (not updated by the transmitter) in transmitter memory. This parameter is not used in processing and is not required.
Procedure
Enter the date you want to use, in the form mm/dd/yyyy.
Tip
ProLink III provides a calendar tool to help you select the date.
70 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 71
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022

6 Integrate the meter with the control system

6.1 Configure the transmitter channels

ProLink III Device Tools Configuration I/O Channels
Field communicator Configure Manual Setup Inputs/Outputs Channels Channel B
Configure Manual Setup Inputs/Outputs Channels Channel C
You can configure the channels on your transmitter to operate in several ways. The channel configuration must match the wiring at the transmitter terminals.
Prerequisites
To avoid causing process errors:
Configure the channels before configuring the outputs.
Before changing the channel configuration, ensure that all control loops affected by the channel are under
manual control.
Important
Before configuring a channel to operate as a Discrete Input, check the status of the remote input device and the actions assigned to the Discrete Input. If the Discrete Input is ON, all actions assigned to the Discrete Input will be performed when the new channel configuration is implemented. If this is not acceptable, change the state of the remote device or wait to configure the channel as a Discrete Input until an appropriate time.
Note
Channel A always operates as an internally-powered mA Output. If Channel B is configured as an mA Output, it is internally powered.
Important
If you need both a Frequency Output and a Discrete Output, you must first configure Channel B as the Frequency Output, then configure Channel C as the Discrete Output. Other combinations are invalid and will be rejected by the transmitter.
Procedure
1. Set Channel B as desired.
Option
Secondary mA Output Channel B will operate as an mA Output.
Frequency Output Channel B will operate as a Frequency Output.
Description
Discrete Output Channel B will operate as a Discrete Output.
2. If you set Channel B to operate as a Frequency Output or Discrete Output, configure the power source
for the channel.
Configuration and Use Manual 71
Page 72
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
Option Description
Internal (Active) The channel is powered by the transmitter.
External (Passive) The channel is powered by an external power source.
3. Set Channel C as desired.
Option Description
Frequency Output Channel C will operate as a Frequency Output.
Discrete Output Channel C will operate as a Discrete Output.
Discrete Input Channel C will operate as a Discrete Input.
4. Configure the power source for Channel C.
Option Description
Internal (Active) The channel is powered by the transmitter.
External (Passive) The channel is powered by an external power source.
Postrequisites
For each channel that you configured, perform or verify the corresponding input or output configuration. When the configuration of a channel is changed, the channel’s behavior will be controlled by the configuration that is stored for the selected input or output type, and the stored configuration may not be appropriate for your process.
After verifying channel and output configuration, return the control loop to automatic control.

6.2 Configure the mA Output

The mA Output is used to report the configured process variable. The mA Output parameters control how the process variable is reported.
Your transmitter may have one or two mA Outputs:
Channel A is always an mA Output (the primary mA Output).
Channel B can be configured as an mA Output (the secondary mA Output).
Important
Whenever you change an mA Output parameter, verify all other mA Output parameters before returning the meter to service. In some situations, the transmitter automatically loads a set of stored values, and these values may not be appropriate for your application.
6.2.1 Configure mA Output Process Variable
ProLink III
72 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Device Tools Configuration I/O Outputs mA Output
Page 73
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
Field Communicator Configure Manual Setup Inputs/Outputs mA Output 1 Primary Variable
Configure Manual Setup Inputs/Outputs mA Output 2 Secondary Variable
Use mA Output Process Variable to select the variable that is reported over the mA Output.
Prerequisites
If you plan to configure the output to report volume flow, ensure that you have set Volume Flow Type as
desired: Liquid or Gas Standard Volume.
If you plan to configure an output to report a concentration measurement process variable, ensure that
the concentration measurement application is configured so that the desired variable is available.
If you are using the HART variables, be aware that changing the configuration of mA Output Process
Variable will change the configuration of the HART Primary Variable (PV) and/or the HART Secondary Variable (SV).
Procedure
Set mA Output Process Variable as desired.
Default settings are as follows:
Primary mA Output: Mass Flow Rate
Secondary mA Output: Density
Postrequisites
If you changed the setting of mA Output Process Variable, verify the settings of Lower Range Value (LRV) and Upper Range Value (URV).
Options for mA Output Process Variable
The transmitter provides a basic set of options for mA Output Process Variable, plus several application­specific options. Different communications tools may use different labels for the options.
Table 6-1: Standard mA Output process variables
Process variable Label
ProLink III Field Communicator
Density Density Dens
Drive gain Drive Gain Driv signl
External pressure External Pressure External pres
External temperature External Temperature External temp
Gas standard volume flow rate Gas Standard Volume Flow Rate Gas vol flo
Mass flow rate Mass Flow Rate Mass flo
Temperature Temperature Temp
Volume flow rate Volume Flow Rate Vol flo
Configuration and Use Manual 73
Page 74
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
Table 6-2: Petroleum measurement mA Output process variables
Process variable Label
ProLink III Field Communicator
Average corrected density Average Density TC Avg Dens
Average temperature Average Temperature TC Avg Temp
Temperature-corrected (standard) volume flow rate
Temperature-corrected density Density at Reference Temperature TC Dens
Volume Flow Rate at Reference Temperature
TC Vol
Table 6-3: Concentration measurement mA Output process variables
Process variable Label
ProLink III Field Communicator
Baume Baume ED Dens (Baume)
Concentration Concentration ED Concentration
Density at reference Density at Reference Temperature ED Dens at Ref
Net mass flow Net Mass Flow Rate ED Net Mass flo
Net volume flow rate Net Volume Flow Rate ED Net Vol flo
Specific gravity Density (Fixed SG Units) ED Dens (SGU)
Standard volume flow rate Volume Flow Rate at Reference
Temperature
ED Std Vol flo
Table 6-4: Fuel consumption mA Output process variables
Process variable Label
ProLink III Field Communicator
Differential mass flow Differential Mass Flow Rate Differential Mass Flow Rate
Table 6-5: PVR mA Output process variables
Process variable Label
ProLink III Field Communicator
Uncorrected oil flow Oil Flow Rate At Line Oil Flow Rate at Line
Uncorrected water cut Water Cut At Line Water Cut at Line
Uuncorrected water flow Water Flow Rate At Line Water Flow Rate at Line
Corrected oil flow Oil Flow Rate At Reference Oil Flow Rate at Reference
Corrected water cut Water Cut At Reference Water Cut at Reference
Corrected water flow Water Flow Rate At Reference Water Flow Rate at Reference
Shrinkage factor corrected net oil at line
74 Micro Motion 2500 Transmitters with Configurable Input/Outputs
SF Oil Flow Rate At Line Shrinkage Factor Oil Flow Rate at Line
Page 75
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
Table 6-5: PVR mA Output process variables (continued)
Process variable Label
ProLink III Field Communicator
Shrinkage factor corrected net oil at 60F
Shrinkage factor corrected volume of mix at 60F
SF Oil Flow Rate At Reference Shrinkage Factor Oil Flow Rate at
Reference
SF Volume Flow Rate At Reference Shrinkage Factor Volume Flow Rate at
Reference
6.2.2 Configure Lower Range Value (LRV) and Upper Range Value
(URV)
ProLink III Device Tools Configuration I/O Outputs mA Output
Field Communicator
The Lower Range Value (LRV) and Upper Range Value (URV) are used to scale the mA Output, that is, to define the relationship between mA Output Process Variable and the mA Output level.
Prerequisites
Ensure that mA Output Process Variable is set to the desired process variable. Each process variable has its own set of LRV and URV values. When you change the values of LRV and URV, you are configuring values for the currently assigned mA Output process variable.
Configure Manual Setup Inputs/Outputs mA Output 1 mA Output Settings PV LRV
Configure Manual Setup Inputs/Outputs mA Output 1 mA Output Settings PV URV
Configure Manual Setup Inputs/Outputs mA Output 2 mA Output Settings SV LRV
Configure Manual Setup Inputs/Outputs mA Output 2 mA Output Settings SV URV
Ensure that the measurement unit for the configured process variable has been set as desired.
Procedure
Set LRV and URV as desired.
LRVis the value of mA Output Process Variable represented by an output of 4 mA. The default value for
LRV depends on the setting of mA Output Process Variable. Enter LRV in the measurement units that are configured for mA Output Process Variable.
URV is the value of mA Output Process Variable represented by an output of 20 mA. The default value for
URV depends on the setting of mA Output Process Variable. Enter URV in the measurement units that are configured for mA Output Process Variable.
The mA Output uses a range of 4–20 mA to represent mA Output Process Variable. Between LRV and URV, the mA Output is linear with the process variable. If the process variable drops below LRV or rises above URV, the transmitter posts an output saturation alert.
Configuration and Use Manual 75
Page 76
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
Default values for Lower Range Value (LRV) and Upper Range Value (URV)
Each option for mA Output Process Variable has its own LRV and URV. If you change the configuration of mA Output Process Variable, the corresponding LRV and URV are loaded and used.
Table 6-6: Default values for Lower Range Value (LRV) and Upper Range Value (URV)
Process variable LRV URV
All mass flow variables –200.000 g/sec 200.000 g/sec
All liquid volume flow variables –0.200 l/sec 0.200 l/sec
All density variables 0.000 g/cm
All temperature variables –240.000 °C 450.000 °C
Drive gain 0.00% 100.00%
Gas standard volume flow –423.78 SCFM 423.78 SCFM
External temperature –240.000 °C 450.000 °C
External pressure 0.000 bar 100.000 bar
3
10.000 g/cm
3
Concentration 0% 100%
Baume 0 10
Specific gravity 0 10
6.2.3 Configure AO Cutoff
ProLink III
Field Communicator Configure Manual Setup Inputs/Outputs mA Output 1 mA Output Settings MAO
AO Cutoff (Analog Output Cutoff) specifies the lowest mass flow rate, volume flow rate, or gas standard volume flow rate that will be reported through the mA Output. Any flow rates below AO Cutoff will be reported as 0.
Restriction
AO Cutoff is applied only if mA Output Process Variable is set to Mass Flow Rate, Volume Flow Rate, or Gas Standard Volume Flow Rate. If mA Output Process Variable is set to a different process variable, AO Cutoff is not configurable, and the transmitter does not implement the AO cutoff function.
Procedure
Device Tools Configuration I/O Outputs mA Output
Cutoff
Configure Manual Setup Inputs/Outputs mA Output 2 mA Output Settings MAO
Cutoff
Set AO Cutoff as desired.
The default values for AO Cutoff are as follows:
Primary mA Output: 0.0 g/sec
Secondary mA Output: Not-A-Number
76 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 77
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
Tip
For most applications, the default value of AO Cutoff should be used. Contact customer service before changing AO Cutoff.
Interaction between AO Cutoff and process variable cutoffs
When mA Output Process Variable is set to a flow variable (for example, mass flow rate or volume flow rate), AO Cutoff interacts with Mass Flow Cutoff or Volume Flow Cutoff. The transmitter puts the cutoff into effect
at the highest flow rate at which a cutoff is applicable.
Example: Cutoff interaction
Configuration:
mA Output Process Variable = Mass Flow Rate
Frequency Output Process Variable = Mass Flow Rate
AO Cutoff = 10 g/sec
Mass Flow Cutoff = 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, all outputs representing mass flow will report zero flow.
Example: Cutoff interaction
Configuration:
mA Output Process Variable = Mass Flow Rate
Frequency Output Process Variable = Mass Flow Rate
AO Cutoff = 15 g/sec
Mass Flow Cutoff = 10 g/sec
Result:
If the mass flow rate drops below 15 g/sec but not below 10 g/sec:The mA Output will report zero flow.
The Frequency Output will report the actual flow rate.
If the mass flow rate drops below 10 g/sec, both outputs will report zero flow.
6.2.4 Configure Added Damping
ProLink III
Field Communicator Configure Manual Setup Inputs/Outputs mA Output 1 mA Output Settings PV Added
Device Tools Configuration I/O Outputs mA Output
Damping
Configure Manual Setup Inputs/Outputs mA Output 2 mA Output Settings SV Added
Damping
Added Damping
controls the amount of damping that will be applied to the mA Output.
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Configuration and Use Manual 77
Page 78
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
Added Damping affects the reporting of mA Output Process Variable through the mA Output only. It does not affect the reporting of that process variable via any other method (e.g., a Frequency Output or digital communications), or the value of the process variable used in calculations.
Note
Added Damping is not applied if the mA Output is fixed (for example, during loop testing) or if the mA Output is reporting a fault. Added Damping is applied while sensor simulation is active.
Procedure
Set Added Damping to the desired value.
The default value is 0.0 seconds. The range is 0.0 to 440 seconds.
When you specify a value for Added Damping, the transmitter automatically rounds the value down to the nearest valid value.
Note
Added Damping values are affected by the setting of Update Rate and 100 Hz Variable.
Table 6-7: Valid values for Added Damping
Update rate
Setting of Update Rate
Normal N/A 20 Hz 0.0, 0.1, 0.3, 0.75, 1.6, 3.3, 6.5, 13.5, 27.5,
Special 100 Hz variable (if assigned
Process variable
to the mA Output)
100 Hz variable (if not assigned to the mA Output)
All other process variables
in effect
100 Hz 0.0, 0.04, 0.12, 0.30, 0.64, 1.32, 2.6, 5.4,
6.25 Hz 0.0, 0.32, 0.96, 2.40, 5.12, 10.56, 20.8,
Valid values for Added Damping
55, 110, 220, 440
11, 22, 44, 88, 176, 350
43.2, 88, 176, 352
Interaction between mA Output Damping and process variable damping
When mA Output Source is set to a flow rate variable, density, or temperature, mA Output Damping interacts with Flow Damping, Density Damping, or Temperature Damping. If multiple damping parameters are applicable, the effect of damping the process variable is calculated first, and the mA Output damping calculation is applied to the result of that calculation.
Example: Damping interaction
Configuration:
Flow Damping = 1 second
mA Output Source = Mass Flow Rate
mA Output Damping = 2 seconds
Result: A change in the mass flow rate will be reflected in the mA Output over a time period that is greater than 3 seconds. The exact time period is calculated by the transmitter according to internal algorithms which are not configurable.
78 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 79
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
6.2.5 Configure mA Output Fault Action and mA Output Fault Level
ProLink III Device Tools Configuration Fault Processing
Field Communicator Configure Manual Setup Inputs/Outputs mA Output 1 MA01 Fault Settings
Configure Manual Setup Inputs/Outputs mA Output 2 MA02 Fault Settings
mA Output Fault Action controls the behavior of the mA Output if the transmitter encounters an internal fault condition.
Note
For some faults only: If Fault Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed.
Procedure
1. Set mA Output Fault Action to the desired value.
The default setting is Downscale.
Restriction
If Digital Communications Fault Action is set to NAN (not a number), you cannot set mA Output Fault Action or Frequency Output Fault Action to None. If you try to do this, the transmitter will not accept
the configuration.
2. If you set mA Output Fault Action to Upscale or Downscale, set mA Output Fault Level as desired.
Postrequisites
NOTICE
If you set mA Output Fault Action or Frequency Output Fault Action to None, be sure to set Digital Communications Fault Action to None. If you do not, the output will not report actual process data, and this
may result in measurement errors or unintended consequences for your process.
Options for mA Output Fault Action and mA Output Fault Level
Option
Upscale Goes to the configured fault level Default: 22.0 mA
Downscale (default) Goes to the configured fault level Default: 2.0 mA
Internal Zero Goes to the mA Output level associated with a
None Tracks data for the assigned process variable; no
mA Output behavior
process variable value of 0 (zero), as determined by Lower Range Value and Upper Range Value settings
fault action
mA Output Fault Level
Range: 21.0 to 24.0 mA
Range: 1.0 to 3.6 mA
Not applicable
Not applicable
Configuration and Use Manual 79
Page 80
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038

6.3 Configure the Frequency Output

The Frequency Output is used to report a process variable. The Frequency Output parameters control how the process variable is reported. Your transmitter may have zero, one, or two Frequency Outputs, depending on the configuration of Channels B and C. If both Channels B and C are configured as Frequency Outputs, they are electrically isolated but not independent. You cannot configure them separately.
Important
Whenever you change a Frequency Output parameter, verify all other Frequency Output parameters before returning the flowmeter to service. In some situations, the transmitter automatically loads a set of stored values, and these values may not be appropriate for your application.
6.3.1 Configure Frequency Output Process Variable
ProLink III Device Tools Configuration I/O Outputs Frequency Output
Field Communicator Configure Manual Setup Inputs/Outputs Frequency Output FO Settings Third Variable
Frequency Output Process Variable controls the variable that is reported over the Frequency Output.
Prerequisites
If you plan to configure the output to report volume flow, ensure that you have set Volume Flow Type as desired: Liquid or Gas Standard Volume.
If you plan to configure an output to report a concentration measurement process variable, ensure that the concentration measurement application is configured so that the desired variable is available.
If you are using the HART variables, be aware that changing the configuration of Frequency Output Process Variable will change the configuration of the HART Tertiary Variable (TV).
Procedure
Set Frequency Output Process Variable as desired.
The default setting is Mass Flow Rate.
Options for Frequency Output Process Variable
The transmitter provides a basic set of options for Frequency Output Process Variable, plus several application-specific options. Different communications tools may use different labels for the options.
Table 6-8: Standard FO process variables
Process variable Label
ProLink III Field Communicator
Gas standard volume flow rate Gas Standard Volume Flow Rate Gas vol flo
Mass flow rate Mass Flow Rate Mass flo
Volume flow rate Volume Flow Rate Vol flo
80 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 81
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
Table 6-9: Petroleum measurement FO process variables
Process variable Label
ProLink III Field Communicator
Temperature-corrected (standard) volume flow rate
Volume Flow Rate at Reference Temperature
TC Vol
Table 6-10: Concentration measurement FO process variables
Process variable Label
ProLink III Field Communicator
Net mass flow Net Mass Flow Rate ED Net Mass flo
Net volume flow rate Net Volume Flow Rate ED Net Vol flo
Standard volume flow rate Volume Flow Rate at Reference
Temperature
ED Std Vol flo
Table 6-11: Fuel consumption FO process variables
Process variable Label
ProLink III Field Communicator
Differential mass flow Differential Mass Flow Rate Differential Mass Flow Rate
6.3.2 Configure Frequency Output Polarity
ProLink III
Field Communicator Configure Manual Setup Inputs/Outputs Frequency Output FO Settings FO Polarity
Device Tools Configuration I/O Outputs Frequency Output
Frequency Output Polarity controls how the output indicates the ON (active) state. The default value, Active
High, is appropriate for most applications. Your receiving device might require an Active Low setting.
Procedure
Set Frequency Output Polarity as desired.
The default setting is Active High.
Options for Frequency Output Polarity
Polarity option
Active High 0 As determined by power supply,
Active Low As determined by power supply,
Configuration and Use Manual 81
Reference voltage (OFF) Pulse voltage (ON)
pull-up resistor, and load. See the installation manual for your transmitter.
0 pull-up resistor, and load. See the installation manual for your transmitter.
Page 82
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
6.3.3 Configure Frequency Output Scaling Method
ProLink III Device Tools Configuration I/O Outputs Frequency Output
Field Communicator Configure Manual Setup Inputs/Outputs Frequency Output FO Scaling
Frequency Output Scaling Method defines the relationship between output pulse and flow units. Set Frequency Output Scaling Method as required by your frequency receiving device.
Procedure
1. Set Frequency Output Scaling Method.
Option Description
Frequency=Flow (default) Frequency calculated from flow rate
Pulses/Unit A user-specified number of pulses represents one flow unit
Units/Pulse A pulse represents a user-specified number of flow units
2. Set additional required parameters.
If you set Frequency Output Scaling Method to Frequency=Flow, set Rate Factor and Frequency
Factor.
If you set Frequency Output Scaling Method to Pulses/Unit, define the number of pulses that will
represent one flow unit.
If you set Frequency Output Scaling Method to Units/Pulse, define the number of units that each
pulse will indicate.
For all scaling methods, the transmitter puts out a fixed number of pulses per unit, and at the same time, the Frequency Output signal varies in proportion to flowrate.
Calculate frequency from flow rate
The Frequency=Flow option is used to customize the Frequency Output for your application when you do not know appropriate values for Units/Pulse or Pulses/Unit.
If you specify Frequency=Flow, you must provide values for Rate Factor and Frequency Factor:
Rate Factor
Frequency Factor
The maximum flow rate that you want the Frequency Output to report.
A value calculated as follows:
FrequencyFactor
where:
RateFactor
T
 × N
T
Factor to convert selected time base to seconds
N
Number of pulses per flow unit, as configured in the receiving device
The resulting Frequency Factor must be within the range of the Frequency Output 0 to 10,000 Hz:
If Frequency Factor is less than 1 Hz, reconfigure the receiving device for a higher pulses/unit setting.
82 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 83
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
If Frequency Factor is greater than 10,000 Hz, reconfigure the receiving device for a lower pulses/unit
setting.
6.3.4 Configure Frequency Output Mode
ProLink III Device Tools Configuration I/O Outputs Frequency Output
Field communicator Not available
Frequency Output Mode defines the relationship between the two Frequency Outputs (dual-pulse mode).
Prerequisites
Before configuring Frequency Output Mode, ensure that both Channel B and Channel C are configured to operate as Frequency Outputs. If you do not have two Frequency Outputs on your transmitter, Frequency Output Mode is set to Single and cannot be changed.
Procedure
Set Frequency Output Mode as desired.
The default value is Quadrature.
Options for Frequency Output Mode
Option Channel behavior Process condition
In-Phase 50% duty cycle
90° Phase Shift 50% duty cycle
–90° Phase Shift 50% duty cycle
180° Phase Shift 50% duty cycle
Quadrature 50% duty cycle
(1)
Channel B
Channel C
Channel B
Channel C
Channel B
Channel C
Channel B
Channel C
Channel B Forward flow
Channel C
Channel B Reverse flow
Channel C
Channel B Fault condition
Channel C
Channel C lags Channel B by 90°
Channel C leads Channel B by 90°
Channel C is driven to 0
(1) Quadrature mode is used only for specific Weights & Measures applications where required by law.
Configuration and Use Manual 83
Page 84
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
6.3.5 Configure Frequency Output Fault Action and Frequency
Output Fault Level
ProLink III Device Tools Configuration Fault Processing
Field Communicator Configure Manual Setup Inputs/Outputs Frequency Output FO Fault Parameters FO
Frequency Output Fault Action controls the behavior of the Frequency Output if the transmitter encounters an internal fault condition.
Note
For some faults only: If Fault Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed.
Procedure
1. Set Frequency Output Fault Action as desired.
The default value is Downscale (0 Hz).
Fault Action
Configure Manual Setup Inputs/Outputs Frequency Output FO Fault Parameters FO
Fault Level
2. If you set Frequency Output Fault Action to Upscale, set Frequency Fault Level to the desired value.
The default value is 15000 Hz. The range is 10 to 15000 Hz.
Options for Frequency Output Fault Action
Table 6-12: Options for Frequency Output Fault Action
Frequency output behavior
Label
Upscale Goes to configured Upscale value:
Downscale 0 Hz Channel B: goes to configured Upscale
Internal Zero 0 Hz Channel B: Goes to configured Upscale
None (default) Tracks data for the assigned process
All modes except Quadrature Quadrature mode
Channel B: Goes to configured Upscale
Range: 10 Hz to 15000 Hz
Default: 15000 Hz
variable; no fault action
value Channel C: 0 Hz
value Channel C: 0 Hz
value Channel C: 0 Hz
Channel B: Tracks data for the assigned process variable
Channel C: Tracks data for the assigned process variable
If your transmitter has two Frequency Outputs, their fault behavior is the same for all modes except Quadrature.
84 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 85
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
NOTICE
If you set mA Output Fault Action or Frequency Output Fault Action to None, be sure to set Digital Communications Fault Action to None. If you do not, the output will not report actual process data, and this
may result in measurement errors or unintended consequences for your process.
Restriction
If Digital Communications Fault Action is set to NAN (not a number), you cannot set mA Output Fault Action or Frequency Output Fault Action to None. If you try to do this, the transmitter will not accept the
configuration.

6.4 Configure the Discrete Output

The Discrete Output is used to report specific meter or process conditions. The Discrete Output parameters control which condition is reported and how it is reported. Your transmitter may have zero, one, or two Discrete Outputs, depending on the configuration of Channels B and C. If both Channels B and C are configured as Discrete Output, they operate independently and you can configure them separately.
Restriction
Before you can configure the Discrete Output, you must configure a channel to operate as a Discrete Output.
Important
Whenever you change a Discrete Output parameter, verify all other Discrete Output parameters before returning the meter to service. In some situations, the transmitter automatically loads a set of stored values, and these values may not be appropriate for your application.
6.4.1 Configure Discrete Output Source
ProLink III
Field Communicator Configure Manual Setup Inputs/Outputs Discrete Output DO 1 Assignment
Discrete Output Source controls which device condition or process condition is reported via the Discrete Output.
Procedure
Set Discrete Output Source to the desired option.
Default settings for Discrete Output Source are as follows:
Discrete Output 1: Flow Direction
Discrete Output 2: Flow Switch, with Flow Switch Variable set to Mass Flow Rate, Flow Switch Setpoint
set to 0.0 g/s, and Flow Switch Hysteresis set to 0.05 (5%).
Device Tools Configuration I/O Outputs Discrete Output
Configure Manual Setup Inputs/Outputs Discrete Output DO 2 Assignment
Options for Discrete Output Source
Option
ProLink III Field Communicator
Discrete Event 1–5
Configuration and Use Manual 85
(1)
Enhanced Event 1 Discrete Event x ON Site-specific
Label State Discrete Output
voltage
Page 86
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
Option Label State Discrete Output
ProLink III Field Communicator
voltage
Enhanced Event 2 Enhanced Event 3 Enhanced Event 4 Enhanced Event 5
Event 1–2
Flow Switch Flow Switch Indicator Flow Switch ON Site-specific
Flow Direction Forward Reverse
Calibration in Progress Calibration in Progress Calibration in Progress ON Site-specific
Fault Fault Indication Fault ON Site-specific
(1) Events configured using the enhanced event model. (2) Events configured using the basic event model.
(2)
Event 1 Event 2
Event 1 or Event 2 Status
Indicator
Event 1 Event 2
Event 1 or Event 2
Forward/Reverse Forward flow 0 V
OFF 0 V
ON Site-specific
OFF 0 V
OFF 0 V
Reverse flow Site-specific
OFF 0 V
OFF 0 V
Important
If you assign Flow Switch to the Discrete Output, you should also configure Flow Switch Variable, Flow Switch Setpoint, and Hysteresis.
Note
If your transmitter has two Discrete Outputs:
You can configure them independently. For example, you can assign one to Flow Switch and one to Fault.
If you assign both to Flow Switch, the same settings for Flow Switch Variable, Flow Switch Setpoint, and
Flow Switch Hysteresis will be implemented for both Discrete Outputs.
Related information
Configure an enhanced event Fault indication with a Discrete Output
Configure Flow Switch parameters
ProLink III
Field Communicator Configure Manual Setup Inputs/Outputs Discrete Output Flow Switch Source
86 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Device Tools Configuration I/O Outputs Discrete Output
Configure Manual Setup Inputs/Outputs Discrete Output Flow Switch Setpoint
Configure Manual Setup Inputs/Outputs Discrete Output Hysteresis
Page 87
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
Flow Switch is used to indicate that the flow rate (measured by the configured flow variable) has moved past the configured setpoint, in either direction. The flow switch is implemented with a user-configurable hysteresis.
Procedure
1. Set Discrete Output Source to Flow Switch, if you have not already done so.
2. Set Flow Switch Variable to the flow variable that you want to use to control the flow switch.
3. Set Flow Switch Setpoint to the value at which the flow switch will be triggered (after Hysteresis is
applied).
If the flow rate is below this value, the Discrete Output is ON.
If the flow rate is above this value, the Discrete Output is OFF.
4. Set Hysteresis to the percentage of variation above and below the setpoint that will operate as a
deadband. Hysteresis defines a range around the setpoint within which the flow switch will not change. The
default is 5%. The valid range is 0.1% to 10%.
Example
If Flow Switch Setpoint = 100 g/sec and Hysteresis = 5%, and the first measured flow rate is above 100 g/sec, the Discrete Output is OFF. It will remain OFF unless the flow rate drops below 95 g/sec. If
this happens, the Discrete Output will turn ON, and remain ON until the flow rate rises above 105 g/sec. At this point it turns OFF and will remain OFF until the flow rate drops below 95 g/sec.
6.4.2 Configure Discrete Output Polarity
ProLink III
Field Communicator Configure Manual Setup Inputs/Outputs Discrete Output DO 1 Polarity
Discrete Outputs have two states: ON (active) and OFF (inactive). Two different voltage levels are used to represent these states. Discrete Output Polarity controls which voltage level represents which state.
Procedure
Set Discrete Output Polarity as desired.
The default setting is Active High.
Options for Discrete Output Polarity
Polarity option
Active High Internal When asserted (condition
Device Tools Configuration I/O Outputs Discrete Output
Configure Manual Setup Inputs/Outputs Discrete Output DO 2 Polarity
Discrete output power supply Description
tied to DO is true), the circuit provides a pull-up to 15 V.
When not asserted
(condition tied to DO is false), the circuit provides 0 V.
Configuration and Use Manual 87
Page 88
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
Polarity option Discrete output power supply Description
External When asserted (condition
tied to DO is true), the circuit provides a pull-up to a site­specific voltage, maximum 30 V.
When not asserted
(condition tied to DO is false), the circuit provides 0 V.
Active Low Internal When asserted (condition
tied to DO is true), the circuit provides 0 V.
When not asserted
(condition tied to DO is false), the circuit provides a pull-up to 15 V.
External When asserted (condition
tied to DO is true), the circuit provides 0 V.
When not asserted
(condition tied to DO is false), the circuit provides a pull-up to a site-specific voltage, to a maximum of 30 V.
6.4.3 Configure Discrete Output Fault Action
ProLink III
Field Communicator Configure Manual Setup Inputs/Outputs Discrete Output DO 1 Fault Action
Discrete Output Fault Action controls the behavior of the Discrete Output if the transmitter encounters an internal fault condition.
Note
For some faults only: If Fault Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed.
NOTICE
Do not use Discrete Output Fault Action as a fault indicator. If you do, you may not be able to distinguish a fault condition from a normal operating condition. If you want to use the Discrete Output as a fault indicator, set Discrete Output Source to Fault and set Discrete Output Fault Action to None.
Procedure
Set Discrete Output Fault Action as desired.
Device Tools Configuration Fault Processing
Configure Manual Setup Inputs/Outputs Discrete Output DO 2 Fault Action
88 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 89
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
The default setting is None.
Related information
Fault indication with a Discrete Output
Options for Discrete Output Fault Action
Label Discrete Output behavior
Upscale Fault: Discrete Output is ON (site-specific voltage)
No fault: Discrete Output is controlled by its assignment
Downscale Fault: Discrete Output is OFF (0 V)
No fault: Discrete Output is controlled by its assignment
None (default) Discrete Output is controlled by its assignment
Fault indication with a Discrete Output
To indicate faults via a Discrete Output, set Discrete Output Source to Fault. Then, if a fault occurs, the Discrete Output is always ON and the setting of Discrete Output Fault Action is ignored.

6.5 Configure the Discrete Input

The Discrete Input is used to initiate one or more transmitter actions from a remote input device. Your transmitter may have zero or one Discrete Input, depending on the configuration of Channel C.
The Discrete Input parameters include:
Discrete Input Action
Discrete Input Polarity
Important
Whenever you change a Discrete Input parameter, verify all other Discrete Input parameters before returning the flowmeter to service. In some situations, the transmitter automatically loads a set of stored values, and these values may not be appropriate for your application.
6.5.1 Configure Discrete Input Action
ProLink III
Field Communicator Configure Manual Setup Inputs/Outputs Discrete Input DI Assignment
Discrete Input Action controls the action or actions that the transmitter will perform when the Discrete Input transitions from OFF to ON.
Important
Before assigning actions to an enhanced event or Discrete Input, check the status of the event or the remote input device. If it is ON, all assigned actions will be performed when the new configuration is implemented. If this is not acceptable, wait until an appropriate time to assign actions to the event or Discrete Input.
Device Tools Configuration I/O Action Assignment
Configuration and Use Manual 89
Page 90
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
Procedure
1. Select an action.
2. Select the Discrete Input that will perform the selected action.
3. Repeat until you have assigned all the actions to be performed by the Discrete Input.
Options for Discrete Input Action
Table 6-13: Standard options
Action Label
ProLink III Field Communicator
None (default) None None
Start sensor zero Start Sensor Zero Perform auto zero
Start/stop all totalizers Start/Stop All Totalization Start/stop totals
Reset mass total Reset Mass Total Reset mass total
Reset volume total Reset Volume Total Reset volume total
Reset gas standard volume total Reset Gas Std Volume Total Reset gas standard volume total
Reset all totals Reset All Totals Reset totals
Table 6-14: Petroleum measurement options
Action Label
ProLink III Field Communicator
Reset temperature-corrected volume total
Reset Volume Total at Reference Temperature
Reset corrected volume total
Table 6-15: Concentration measurement options
Action Label
ProLink III Field Communicator
Reset CM reference volume total Reset Volume Total at Reference
Temperature
Reset CM net mass total Reset Net Mass Total Not available
Reset CM net volume total Reset Net Volume Total Not available
Increment CM matrix Increment Concentration Matrix Not available
Not available
Table 6-16: Meter verification options
Action Label
ProLink III Field Communicator
Start meter verification test Start Meter Verification Not available
90 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 91
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
6.5.2 Configure Discrete Input Polarity
ProLink III Device Tools Configuration I/O Inputs Discrete Input
Field Communicator Configure Manual Setup Inputs/Outputs Discrete Input DI Polarity
The Discrete Input has two states: ON and OFF. Discrete Input Polarity controls how the transmitter maps the incoming voltage level to the ON and OFF states.
Procedure
Set Discrete Input Polarity as desired.
The default setting is Active Low.
Options for Discrete Input Polarity
Polarity option Discrete Input power
supply
Active High Internal Voltage across terminals
External Voltage applied across
Active Low Internal Voltage across terminals
External Voltage applied across
Voltage Status of discrete input
is high
Voltage across terminals is 0 VDC
terminals is 3–30 VDC
Voltage applied across terminals is <0.8 VDC
is 0 VDC
Voltage across terminals is high
terminals is <0.8 VDC
Voltage applied across terminals is 3–30 VDC

6.6 Configure events

at transmitter
ON
OFF
ON
OFF
ON
OFF
ON
OFF
An event occurs when the real-time value of a user-specified process variable moves past a user-defined setpoint. Events are used to provide notification of process changes or to perform specific transmitter actions if a process change occurs.
Your transmitter supports two event models:
Basic event model
Enhanced event model
Configuration and Use Manual 91
Page 92
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
6.6.1 Configure a basic event
ProLink III Device Tools Configuration Events Basic Events
Field Communicator Not available
A basic event is used to provide notification of process changes. A basic event occurs (is ON) if the real-time value of a user-specified process variable moves above (HI) or below (LO) a user-defined setpoint. You can define up to two basic events. Event status can be queried via digital communications, and a Discrete Output can be configured to report event status.
Procedure
1. Select the event that you want to configure.
2. Specify Event Type.
Option Description
HI x > A
The event occurs when the value of the assigned process variable (x) is greater than the setpoint (Setpoint A), endpoint not included.
LO x < A
The event occurs when the value of the assigned process variable (x) is less than the setpoint (Setpoint A), endpoint not included.
3. Assign a process variable to the event.
4. Set a value for Setpoint A.
5. Optional: Configure a Discrete Output to switch states in response to the event status.
6.6.2 Configure an enhanced event
ProLink III
Field Communicator Configure Alert Setup Discrete Events
An enhanced event is used to provide notification of process changes and, optionally, to perform specific transmitter actions if the event occurs. An enhanced event occurs (is ON) if the real-time value of a user­specified process variable moves above (HI) or below (LO) a user-defined setpoint, or in range (IN) or out of range (OUT) with respect to two user-defined setpoints. You can define up to five enhanced events.
Procedure
1. Select the event that you want to configure.
2. Specify Event Type.
Device Tools Configuration Events Enhanced Events
Option
HI x > A
92 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Description
The event occurs when the value of the assigned process variable (x) is greater than the setpoint (Setpoint A), endpoint not included.
Page 93
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
Option Description
LO x < A
The event occurs when the value of the assigned process variable (x) is less than the setpoint (Setpoint A), endpoint not included.
IN A ≤ x ≤ B
The event occurs when the value of the assigned process variable (x) is in range, that is, between Setpoint A and Setpoint B, endpoints included.
OUT x ≤ A or x ≥ B
The event occurs when the value of the assigned process variable (x) is out of range, that is, less than Setpoint A or greater than Setpoint B, endpoints included.
3. Assign a process variable to the event.
4. Set values for the required setpoints.
For HI and LO events, set Setpoint A.
For IN and OUT events, set Setpoint A and Setpoint B.
5. Optional: Configure a Discrete Output to switch states in response to the event status.
6. Optional: Specify the action or actions that the transmitter will perform when the event occurs.
With ProLink III: Device Tools Configuration I/O Action Assignment
With a field communicator: Configure Alert Setup Discrete Events Assign Discrete Action
Options for Enhanced Event Action
Table 6-17: Standard options
Action Label
ProLink III Field Communicator
None (default) None None
Start sensor zero Start Sensor Zero Perform auto zero
Start/stop all totalizers Start/Stop All Totalization Start/stop totals
Reset mass total Reset Mass Total Reset mass total
Reset volume total Reset Volume Total Reset volume total
Reset gas standard volume total Reset Gas Std Volume Total Reset gas standard volume total
Reset all totals Reset All Totals Reset totals
Table 6-18: Petroleum measurement options
Action Label
ProLink III Field Communicator
Reset temperature-corrected volume total
Configuration and Use Manual 93
Reset Volume Total at Reference Temperature
Reset corrected volume total
Page 94
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
Table 6-19: Concentration measurement options
Action Label
ProLink III Field Communicator
Reset CM reference volume total Reset Volume Total at Reference
Temperature
Reset CM net mass total Reset Net Mass Total Not available
Reset CM net volume total Reset Net Volume Total Not available
Increment CM matrix Increment Concentration Matrix Not available
Table 6-20: Meter verification options
Action Label
ProLink III Field Communicator
Start meter verification test Start Meter Verification Not available
Not available

6.7 Configure digital communications

The digital communications parameters control how the transmitter will communicate using digital communications.
Your transmitter supports the following types of digital communications:
HART/Bell 202 over the primary mA terminals
Modbus/RS-485 over the RS-485 terminals
Modbus RTU via the service port
Note
The service port responds automatically to a wide range of connection requests. It is not configurable.
6.7.1 Configure HART/Bell 202 communications
HART/Bell 202 communications parameters support HART communications with the transmitter's primary mA terminals over a HART/Bell 202 network.
Configure basic HART parameters
Basic HART parameters include the HART address, HART tags, and the operation of the primary mA Output.
HART/Bell 202 communications parameters support HART communication with the transmitter's primary mA terminals over a HART/Bell 202 network. The HART/Bell 202 communications parameters include:
HART Address (Polling Address)
mA Output Action
Burst Parameters (optional)
HART Variables (optional)
94 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 95
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
Procedure
1. Set HART Address to a value that is unique on your network.
Default: 0
Range: 0 to 15
Tip
The default address is typically used unless you are in a multidrop environment.
Devices using HART protocol to communicate with the transmitter may use either HART Address or
HART Tag (Software Tag) to identify the transmitter. Configure either or both, as required by your other HART devices.
2. Ensure that mA Output Action is configured appropriately.
Option Description
Enabled (Live) The primary mA Output reports process data as configured.
Disabled (Fixed) The primary mA Output is fixed at 4 mA and does not report process data.
Important
If you use ProLink III to set HART Address to 0, the program automatically enables mA Output Action. If you use ProLink III to set HART Address to any other value, the program automatically disables mA Output Action. This is designed to make it easier to configure the transmitter for legacy behavior. Always verify mA Output Action after setting HART Address.
Configure burst parameters
ProLink III
Field Communicator Configure Manual Setup Inputs/Outputs Communications Set Up Burst Mode
Burst mode is a mode of communication during which the transmitter regularly broadcasts HART digital information over the primary mA Output. The burst parameters control the information that is broadcast when burst mode is enabled.
Tip
In typical installations, burst mode is disabled. Enable burst mode only if you are using a HART Triloop.
Procedure
1. Enable Burst Mode.
2. Set Burst Mode Output as desired.
Device Tools Configuration Communications Communications (HART)
Label
Description
ProLink III Field Communicator
Source (Primary Variable) PV The transmitter sends the primary variable (PV) in the
configured measurement units in each burst (e.g.,
14.0 g/sec, 13.5 g/sec, 12.0 g/sec.
Configuration and Use Manual 95
Page 96
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
Label Description
ProLink III Field Communicator
Primary Variable (Percent Range/Current)
Process Variables/Current Process variables/current The transmitter sends PV, SV, TV, and QV values in
Transmitter variables Fld dev var The transmitter sends four user-specified process variables
% range/current The transmitter sends the PV’s percent of range and the
PV’s actual mA level in each burst (e.g., 25%, 11.0 mA.
measurement units and the PV’s actual milliamp reading in each burst (e.g., 50 g/sec, 23 °C, 50 g/sec, 0.0023 g/cm3,
11.8 mA.
in each burst.
3. Ensure that the burst output variables are set appropriately.
If you set Burst Mode Output to send four user-specified variables, set the four process variables to
be sent in each burst.
If you set Burst Mode Output to any other option, ensure that the HART variables are set as desired.
Configure HART variables (PV, SV, TV, QV)
ProLink III Device Tools Configuration Communications Communications (HART)
Field Communicator Configure Manual Setup Inputs/Outputs Variable Mapping
The HART variables are a set of four variables predefined for HART use. The HART variables include the Primary Variable (PV), Secondary Variable (SV), Tertiary Variable (TV), and Quaternary Variable (QV). You can assign specific process variables to the HART variables, and then use standard HART methods to read or broadcast the assigned process data.
Tip
The Tertiary Variable and Quaternary Variable are also called the Third Variable (TV) and Fourth Variable (FV).
Options for HART variables
Table 6-21: Standard HART process variables
Process variable Primary
Variable (PV)
Board Temperature
Density
Drive Gain
External Pressure
External Temperature
Gas Standard Volume Flow Rate
Gas Standard Volume Inventory
Gas Standard Volume Total
Line (Gross) Volume Flow Rate
Secondary Variable (SV)
Third Variable (TV)
Fourth Variable (QV )
96 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 97
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
Table 6-21: Standard HART process variables (continued)
Process variable Primary
Variable (PV)
Line (Gross) Volume Inventory
Line (Gross) Volume Total
Live Zero
LPO Amplitude
Mass Flow Rate
Mass Inventory
Mass Total
Meter Temperature (T-Series)
RPO Amplitude
Temperature
Tube Frequency
Secondary Variable (SV)
Third Variable (TV)
Table 6-22: Petroleum measurement HART process variables
Process variable Primary
Variable (PV)
API Average Density
Secondary Variable (SV)
Third Variable (TV)
Fourth Variable (QV )
Fourth Variable (QV )
API Average Temperature
API Corrected Volume Flow
API Corrected Volume Inventory
API Corrected Volume Total
API CTL
API Density At Reference
Table 6-23: Concentration measurement HART process variables
Process variable Primary
Variable (PV)
CM Baume (700 core processor only)
CM Concentration
CM Density at Reference
CM Net Mass Flow
CM Net Mass Inventory
CM Net Mass Total
CM Net Volume Flow
CM Net Volume Inventory
Secondary Variable (SV)
Third Variable (TV)
Fourth Variable (QV )
Configuration and Use Manual 97
Page 98
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
Table 6-23: Concentration measurement HART process variables (continued)
Process variable Primary
Variable (PV)
CM Net Volume Total
CM Specific Gravity
CM Standard Volume Flow
CM Standard Volume Inventory
CM Standard Volume Total
Secondary Variable (SV)
Third Variable (TV)
Table 6-24: Fuel consumption HART process variables
Process variable Primary
Variable (PV)
Differential mass flow
Differential mass inventory
Differential mass total
Secondary Variable (SV)
Third Variable (TV)
Table 6-25: PVR-only HART process variables
Process variable Primary
Variable (PV)
Corrected Oil Flow
Secondary Variable (SV)
Third Variable (TV)
Fourth Variable (QV )
Fourth Variable (QV )
Fourth Variable (QV )
Corrected Oil Total
Corrected Water Cut
Corrected Water Flow
Corrected Water Total
Density of Oil @ Line Fixd degAPI
Density of Oil @ Line Fixd SGU
Oil Total @ Line
Shrinkage Factor Corrected Oil Flow @ 60F
Shrinkage Factor Corrected Oil Flow @ Line
Shrinkage Factor Corrected Oil Total @ 60F
Shrinkage Factor Corrected Oil Total @ Line
Shrinkage Factor Corrected Total of Mix @ 60F
Shrinkage Factor Corrected Volume Of Mix @ 60F
Uncorrected Oil Flow
Uncorrected Water Cut
Uncorrected Water Flow
Volume Flow of Mix at Line
98 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Page 99
Configuration and Use Manual Integrate the meter with the control system
MMI-20019038 February 2022
Table 6-25: PVR-only HART process variables (continued)
Process variable Primary
Variable (PV)
Volume Total Of Mix @ Line
Water Total @ Line
Secondary Variable (SV)
Third Variable (TV)
Table 6-26: TMR-only HART process variables
Process variable Primary
Variable (PV)
Remediated Mass Flow
Remediated Mass Total
Remediated Mass Inventory
Secondary Variable (SV)
Third Variable (TV)
Table 6-27: PVR- and TBR-only HART process variables
Process variable Primary
Variable (PV)
Unremediated Density
Secondary Variable (SV)
Third Variable (TV)
Table 6-28: PVR, TBR, and TMR HART process variables
Process variable Primary
Variable (PV)
Secondary Variable (SV)
Third Variable (TV)
Fourth Variable (QV )
Fourth Variable (QV )
Fourth Variable (QV )
Fourth Variable (QV )
Total Remediated Time
Interaction of HART variables and transmitter outputs
The HART variables are automatically reported through specific transmitter outputs. They may also be reported through HART burst mode, if enabled on your transmitter.
Table 6-29: HART variables and transmitter outputs
HART variable Reported via Comments
Primary Variable (PV) Primary mA output If one assignment is changed, the other is changed
automatically, and vice versa.
Secondary Variable (SV) Secondary mA Output (if
present on your transmitter)
Tertiary Variable (TV) Frequency Output (if
present on your transmitter)
Quaternary Variable (QV) Not associated with an
output
If one assignment is changed, the other is changed automatically, and vice versa. If your transmitter is not configured for a secondary mA Output, the SV must be configured directly, and the value of the SV is available only via digital communications.
If one assignment is changed, the other is changed automatically, and vice versa. If your transmitter does not have a Frequency Output, the TV must be configured directly, and the value of the TV is available only via digital communications.
The QV must be configured directly, and the value of the QV is available only via digital communications.
Configuration and Use Manual 99
Page 100
Integrate the meter with the control system Configuration and Use Manual
February 2022 MMI-20019038
6.7.2 Configure Modbus/RS-485 communications
ProLink III Device Tools Configuration Communications RS-485 Terminals
Field Communicator Configure Manual Setup Inputs/Outputs Communications Set Up RS-485 Port
Modbus/RS-485 communications parameters control Modbus communication with the transmitter's RS-485 terminals.
Procedure
1. Set Disable Modbus ASCII as desired.
Support for Modbus ASCII limits the set of addresses that are available for the transmitter's Modbus address.
Modbus ASCII support Available Modbus addresses
Disabled 1–127, excluding 111 (111 is reserved to the service port)
Enabled 1–15, 32–47, 64–79, and 96–110
2. Set Protocol to match the protocol used by your Modbus/RS-485 host.
Option Description
Modbus RTU (default) 8–bit communications
Modbus ASCII 7–bit communications
If support for Modbus ASCII is disabled, you must use Modbus RTU.
3. Set Modbus Address to a unique value on the network.
4. Set Parity, Stop Bits, and Baud Rate as appropriate for your network.
5. Set Floating-Point Byte Order to match the byte order used by your Modbus host.
Code
Byte order
0 1–2 3–4
1 3–4 1–2
2 2–1 4–3
3 4–3 2–1
See the following table for the bit structure of bytes 1 through 4.
Table 6-30: Bit structure of floating-point bytes
Byte Bits Definition
1 SEEEEEEE S=Sign
E=Exponent
100 Micro Motion 2500 Transmitters with Configurable Input/Outputs
Loading...