Micro Motion Manual: Micro Motion 1500 Transmitters with Analog Outputs Manuals & Guides

Configuration and Use Manual
MMI-20019023, Rev AC
Micro Motion™ 1500 Transmitters with Analog Outputs
Configuration and Use Manual
February 2022
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step.
Safety and approval information
This Micro Motion product complies with all applicable European directives when properly installed in accordance with the instructions in this manual. Refer to the EU declaration of conformity for directives that apply to this product. The following are available: the EU declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions. In addition the IECEx Installation Instructions for installations outside of the European Union and the CSA Installation Instructions for installations in North America are available on the internet at or through your local Micro Motion support center.
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at . For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion Customer Service department.
Emerson Flow customer service
Email:
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America
United States 800-522-6277 U.K. and Ireland 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 70 413
Mexico +52 55 5809 5010 France +33 (0) 800 917
Argentina +54 11 4809 2700 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy +39 8008 77334 China +86 21 2892 9000
Chile +56 2 2928 4800 Central & Eastern +41 (0) 41 7686
Peru +51 15190130 Russia/CIS +7 495 995 9559 South Korea +82 2 3438 4600
Europe and Middle East Asia Pacific
New Zealand 099 128 804
6666
India 800 440 1468
901
Japan +81 3 5769 6803
111
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684
2
Configuration and Use Manual Contents
MMI-20019023 February 2022

Contents

Part I Getting started
Chapter 1 Before you begin........................................................................................................9
1.1 About this manual....................................................................................................................... 9
1.2 Transmitter model code.............................................................................................................. 9
1.3 Communications tools and protocols.......................................................................................... 9
1.4 Additional documentation and resources.................................................................................. 10
Chapter 2 Quick start............................................................................................................... 11
2.1 Power up the transmitter...........................................................................................................11
2.2 Check meter status....................................................................................................................11
2.3 Make a startup connection to the transmitter............................................................................12
2.4 (Optional) Adjust digital communications settings.................................................................... 12
2.5 Verify mass flow measurement..................................................................................................13
2.6 Verify the zero........................................................................................................................... 13
Part II Configuration and commissioning
Chapter 3 Introduction to configuration and commissioning....................................................17
3.1 Configuration flowchart.............................................................................................................17
3.2 Default values and ranges.......................................................................................................... 18
3.3 Disable write-protection on the transmitter configuration.........................................................18
3.4 Restore the factory configuration.............................................................................................. 18
Chapter 4 Configure process measurement..............................................................................19
4.1 Configure mass flow measurement........................................................................................... 19
4.2 Configure volume flow measurement for liquid applications..................................................... 23
4.3 Configure GSV flow measurement.............................................................................................27
4.4 Configure Flow Direction .......................................................................................................... 33
4.5 Configure density measurement ...............................................................................................38
4.6 Configure temperature measurement....................................................................................... 42
4.7 Configure pressure compensation............................................................................................. 43
Chapter 5 Configure device options and preferences................................................................ 49
5.1 Configure response time parameters.........................................................................................49
5.2 Configure alert handling............................................................................................................ 51
5.3 Configure informational parameters..........................................................................................55
Chapter 6 Integrate the meter with the control system............................................................ 59
6.1 Configure the transmitter channels........................................................................................... 59
Configuration and Use Manual 3
Contents Configuration and Use Manual
February 2022 MMI-20019023
6.2 Configure the mA Output.......................................................................................................... 59
6.3 Configure the Frequency Output............................................................................................... 65
6.4 Configure the Discrete Output...................................................................................................68
6.5 Configure events....................................................................................................................... 72
6.6 Configure digital communications.............................................................................................74
Chapter 7 Complete the configuration..................................................................................... 83
7.1 Test or tune the system using sensor simulation........................................................................83
7.2 Back up transmitter configuration............................................................................................. 85
7.3 Enable write-protection on the transmitter configuration..........................................................85
Part III Operations, maintenance, and troubleshooting
Chapter 8 Transmitter operation..............................................................................................89
8.1 Record the process variables......................................................................................................89
8.2 View process variables............................................................................................................... 89
8.3 View transmitter status using the status LED............................................................................. 90
8.4 View and acknowledge status alerts.......................................................................................... 90
8.5 Read totalizer and inventory values............................................................................................92
8.6 Start and stop totalizers and inventories.................................................................................... 92
8.7 Reset totalizers.......................................................................................................................... 93
8.8 Reset inventories....................................................................................................................... 93
Chapter 9 Measurement support..............................................................................................95
9.1 Options for measurement support.............................................................................................95
9.2 Use Smart Meter Verification..................................................................................................... 95
9.3 Use Production Volume Reconciliation, Transient Mist Remediation, and Transient Bubble
Remediation............................................................................................................................. 101
9.4 Piecewise linearization (PWL) for calibrating gas meters.......................................................... 102
9.5 Zero the meter........................................................................................................................ 102
9.6 Validate the meter...................................................................................................................103
9.7 Perform a (standard) D1 and D2 density calibration.................................................................105
9.8 Perform a D3 and D4 density calibration (T-Series sensors only).............................................. 107
9.9 Perform temperature calibration............................................................................................. 110
Chapter 10 Troubleshooting.................................................................................................... 113
10.1 Status LED states................................................................................................................... 113
10.2 Check the core processor LED................................................................................................ 113
10.3 Perform a 700 core processor resistance test.........................................................................115
10.4 Check the cutoffs...................................................................................................................116
10.5 Density measurement problems............................................................................................116
10.6 Check the drive gain.............................................................................................................. 117
10.7 Check for internal electrical problems....................................................................................118
10.8 Check Flow Direction ............................................................................................................ 120
4 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Contents
MMI-20019023 February 2022
10.9 Flow measurement problems ................................................................................................121
10.10 Frequency Output problems................................................................................................123
10.11 Check Frequency Output Fault Action .................................................................................123
10.12 Check Frequency Output Scaling Method ........................................................................... 124
10.13 Check grounding................................................................................................................. 124
10.14 Check the HART communication loop................................................................................. 124
10.15 Check HART Address and mA Output Action........................................................................125
10.16 Check HART burst mode......................................................................................................125
10.17 Perform loop tests............................................................................................................... 125
10.18 Check Lower Range Value and Upper Range Value ..............................................................128
10.19 Milliamp output problems................................................................................................... 128
10.20 Check mA Output Fault Action ............................................................................................130
10.21 Check the trimming of the mA Output.................................................................................130
10.22 Check the pickoff voltage.................................................................................................... 130
10.23 Check power supply wiring.................................................................................................. 131
10.24 Check for radio frequency interference (RFI)........................................................................ 132
10.25 Using sensor simulation for troubleshooting....................................................................... 132
10.26 Check sensor-to-transmitter wiring..................................................................................... 132
10.27 Check for two-phase flow (slug flow)................................................................................... 133
10.28 Status alerts, causes, and recommendations....................................................................... 133
10.29 Temperature measurement problems.................................................................................148
Appendix A Using ProLink III with the transmitter.....................................................................151
A.1 Basic information about ProLink III ..........................................................................................151
A.2 Connect with ProLink III .......................................................................................................... 152
Appendix B Using a field communicator with the transmitter................................................... 161
B.1 Basic information about field communicators..........................................................................161
B.2 Connect with a field communicator ........................................................................................ 162
Appendix C Default values and ranges...................................................................................... 165
Appendix D Transmitter components and installation wiring.................................................... 171
D.1 Installation types.....................................................................................................................171
D.2 Power supply terminals .......................................................................................................... 173
D.3 Input/output (I/O) wiring terminals......................................................................................... 174
Appendix E NE 53 history..........................................................................................................175
Configuration and Use Manual 5
Contents Configuration and Use Manual
February 2022 MMI-20019023
6 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Getting started
MMI-20019023 February 2022
Part I Getting started
Configuration and Use Manual 7
Getting started Configuration and Use Manual
February 2022 MMI-20019023
8 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Before you begin
MMI-20019023 February 2022

1 Before you begin

1.1 About this manual

This manual helps you configure, commission, use, maintain, and troubleshoot the 1500 transmitter.
Important
This manual assumes that:
The transmitter has been installed correctly and completely according to the instructions in the
transmitter installation manual
Users understand basic transmitter and sensor installation, configuration, and maintenance concepts and
procedures

1.2 Transmitter model code

You can verify that this manual pertains to your transmitter by ensuring the model code on the transmitter tag matches the format.
Example:
The transmitter has a model number of the following form: 1500DEB**A******
D
4-wire remote DIN rail–mount
E
4-wire remote DIN rail transmitter with 9-wire remote enhanced core processor
B
4-wire remote DIN rail transmitter with 9-wire remote core processor
A
Analog outputs option board

1.3 Communications tools and protocols

You must have a communications tool to interface with the transmitter. Several different communications tools and protocols are supported. You may use different tools in different locations or for different tasks.
Communications tool
ProLink III HART/Bell 202
Field communicator HART/Bell 202
Tip
You may be able to use other communications tools, such as AMS™ Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this manual. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at .
Supported protocols
Modbus/RS-485
Service port
Configuration and Use Manual 9
Before you begin Configuration and Use Manual
February 2022 MMI-20019023

1.4 Additional documentation and resources

The following additional documentation supports the installation and operation of the transmitter.
Topic Document
Hazardous area installation See the approval documentation shipped with the
transmitter, or download the appropriate documentation at www.emerson.com.
Product Data Sheet
Production Volume Reconciliation (PVR), Transient Bubble Remediation (TBR), and Transient Mist Remediation (TMR) applications
Sensor Sensor documentation
Transmitter installation
Micro Motion Series 1000 and Series 2000 Transmitters with MVD™ Technology Product Data Sheet (PDS)
Micro Motion Oil and Gas Production Supplement
Micro Motion Model 1500 and Model 2500 Transmitters: Installation Manual
All documentation resources are available at www.emerson.com or on the user documentation DVD.
10 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Quick start
MMI-20019023 February 2022

2 Quick start

2.1 Power up the transmitter

The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement.
The 1500:
Is DC powered only
Has a minimum 19.2 to 28.8 VDC, 6.3 watts
At startup, the transmitter power source must provide a minimum of 1.0 amperes of short-term current
per transmitter
The length and conductor diameter of the power cable must be sized to provide 19.2 VDC minimum at the
power terminals, at a load current of 330 mA
Procedure
1. Ensure that all transmitter and sensor covers and seals are closed.
WARNING
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are tightly closed. For hazardous area installations, applying power while housing covers are removed or loose can cause an explosion.
2. Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. During this period, Alert 009 is active. The diagnostic routines should complete in approximately 30 seconds. The status LED will turn green when the startup diagnostics are complete. If the status LED exhibits different behavior, an alert is active.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power has been off long enough to allow components to reach ambient temperature, allow the electronics to warm up for approximately 10 minutes before relying on process measurements. During this warm-up period, you may observe minor measurement instability or inaccuracy.

2.2 Check meter status

Check the meter for any error conditions that require user action or that affect measurement accuracy.
Procedure
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alert A009 is active. This alert should clear automatically when the power-up sequence is complete.
Configuration and Use Manual 11
Quick start Configuration and Use Manual
February 2022 MMI-20019023
2. Check the status LED on the transmitter.
Related information
View and acknowledge status alerts
2.2.1 Transmitter status reported by LED
Table 2-1: Status LED states
LED state Alarm condition Description
Solid green No alarm Normal operation
Flashing yellow No alarm Zero calibration procedure is in progress
Loop test is in progress
Solid yellow Low-severity alarm Alarm condition that will not cause measurement error
(outputs continue to report process data)
Solid red High-severity alarm Alarm condition that will cause measurement error
(outputs in fault)

2.3 Make a startup connection to the transmitter

To configure the transmitter, you must have an active connection from a communications tool. Follow this procedure to make your first connection to the transmitter.
Procedure
Identify the connection type to use, and follow the instructions for that connection type in the appropriate appendix. Use the default communications parameters shown in the appendix.
Communications tool
ProLink III Modbus/RS-485 Using ProLink III with the transmitter
Field Communicator HART Using a field communicator with the
Connection type to use Instructions
transmitter

2.4 (Optional) Adjust digital communications settings

Change the communications parameters to site-specific values.
Important
If you are changing communications parameters for the connection type that you are using, you will lose the connection when you write the parameters to the transmitter. Reconnect using the new parameters.
Procedure
1. To change the communications parameters using ProLink III, choose Device Tools Configuration
Communications.
2. To change the communications parameters using the Field Communicator, choose On-Line Menu
Configure Manual Setup Inputs/Outputs Communications.
12 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Quick start
MMI-20019023 February 2022

2.5 Verify mass flow measurement

Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available method.
Procedure
Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the Process Variables
panel.
Connect to the transmitter with a field communicator and read the value for Mass Flow Rate.
Online Overview Mass Flow Rate
Postrequisites
If the reported mass flow rate is not accurate:
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues.

2.6 Verify the zero

Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
The zero verification procedure analyzes the Live Zero value under conditions of zero flow, and compares it to the Zero Stability range for the sensor. If the average Live Zero value is within a reasonable range, the zero value stored in the transmitter is valid. Performing a field calibration will not improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one of the following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Do not verify the zero or zero the meter if a high-severity alert is active. Correct the problem, then verify the zero or zero the meter. You may verify the zero or zero the meter if a low-severity alert is active.
Procedure
1. Allow the flowmeter to warm up for at least 20 minutes after applying power.
2. Run the process fluid through the sensor until the sensor temperature reaches the normal process
operating temperature.
3. Stop flow through the sensor by shutting the downstream valve, and then the upstream valve if
available.
4. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is completely full of
process fluid.
5. From ProLink III, choose Device Tools Calibration Zero Verification and Calibration Verify
Zero and wait until the procedure completes.
6. Observe the drive gain, temperature, and density readings. If they are stable, check the Live Zero or
Field Verification Zero value. If the average value is close to 0, you should not need to zero the meter.
Configuration and Use Manual 13
Quick start Configuration and Use Manual
February 2022 MMI-20019023
7. If the zero verification procedure fails:
a) Confirm that the sensor is completely blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
b) Verify that the process fluid is not flashing or condensing, and that it does not contain particles
that can settle out.
c) Remove or reduce sources of electromechanical noise if appropriate.
d) Repeat the zero verification procedure.
e) If it fails again, zero the meter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Related information
Zero the meter
2.6.1 Terminology used with zero verification and zero calibration
Term Definition
Zero In general, the offset required to synchronize the left pickoff and the right pickoff under
conditions of zero flow. Unit = microseconds.
Factory Zero The zero value obtained at the factory, under laboratory conditions.
Field Zero The zero value obtained by performing a zero calibration outside the factory.
Prior Zero The zero value stored in the transmitter at the time a field zero calibration is begun. May
be the factory zero or a previous field zero.
Manual Zero The zero value stored in the transmitter, typically obtained from a zero calibration
procedure. It may also be configured manually. Also called “mechanical zero” or “stored zero”.
Live Zero The real-time bidirectional mass flow rate with no flow damping or mass flow cutoff
applied. An adaptive damping value is applied only when the mass flow rate changes dramatically over a very short interval. Unit = configured mass flow measurement unit.
Zero Stability A laboratory-derived value used to calculate the expected accuracy for a sensor. Under
laboratory conditions at zero flow, the average flow rate is expected to fall within the range defined by the Zero Stability value (0 ± Zero Stability). Each sensor size and model has a unique Zero Stability value.
Zero Calibration The procedure used to determine the zero value.
Zero Time The time period over which the Zero Calibration procedure is performed. Unit = seconds.
Field Verification Zero A 3-minute running average of the Live Zero value, calculated by the transmitter. Unit =
configured mass flow measurement unit.
Zero Verification A procedure used to evaluate the stored zero and determine whether or not a field zero
can improve measurement accuracy.
14 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Configuration and commissioning
MMI-20019023 February 2022
Part II Configuration and commissioning
Configuration and Use Manual 15
Configuration and commissioning Configuration and Use Manual
February 2022 MMI-20019023
16 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Introduction to configuration and commissioning
MMI-20019023 February 2022

3 Introduction to configuration and commissioning

3.1 Configuration flowchart

Use the following flowchart as a general guide to the configuration and commissioning process.
Some options may not apply to your installation. Detailed information is provided in the remainder of this manual.
Configure process measurement
Configure mass flow
measurement
Configure device options and preferences
Configure fault handling
parameters
Test and move to production
Test or tune transmitter using sensor simulation
Configure volume flow
meaurement
Volume flow type
Liquid
Configure flow direction
Configure density
measurement
Configure temperature
measurement
Configure pressure
compensation (optional)
Configure PVR, TMR,
or TBR (if available)
Gas
Define gas properties
Configure sensor
parameters
Configure device
parameters
Integrate device with control system
Configure the channel(s)
Configure the mA
output(s)
Configure the frequency
output(s)
Configure the discrete
output(s)
Back up transmitter
configuration
Enable write-protection on
transmitter configuration
Done
Configure events
Configure digital communications
Configuration and Use Manual 17
Introduction to configuration and commissioning Configuration and Use Manual
February 2022 MMI-20019023

3.2 Default values and ranges

See Default values and ranges to view the default values and ranges for the most commonly used parameters.
3.3 Disable write-protection on the transmitter
configuration
Display OFF-LINE MAINT CONFG LOCK
ProLink III Device Tools Configuration Write-Protection
If the transmitter is write-protected, the configuration is locked and you must unlock it before you can change any configuration parameters. By default, the transmitter is not write-protected.
Tip
Write-protecting the transmitter prevents accidental changes to configuration. It does not prevent normal operational use. You can always disable write-protection, perform any required configuration changes, then re-enable write-protection.

3.4 Restore the factory configuration

ProLink III Device Tools Configuration Transfer Restore Factory Configuration
Field communicator Service Tools Maintenance Reset/Restore Restore Factory Configuration
Restoring the factory configuration returns the transmitter to the same configuration it had when it left the factory. This may be useful if you experience problems during configuration.
Important
You cannot restore factory configurations with a 700 core.
Tip
Restoring the factory configuration is not a common action. You may want to contact customer support to see if there is a preferred method to resolve any issues.
18 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Configure process measurement
MMI-20019023 February 2022

4 Configure process measurement

4.1 Configure mass flow measurement

The mass flow measurement parameters control how mass flow is measured and reported.
4.1.1 Configure Mass Flow Measurement Unit
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements Flow Mass Flow Unit
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow rate. The unit used for mass total and mass inventory is derived from this unit.
Any selected measurement unit, (mass, volume or gas standard volume), is automatically applied to both the mA and Frequency Outputs.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
The default setting for Mass Flow Measurement Unit is g/sec (grams per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Label
Unit description
Grams per second g/sec g/s
Grams per minute g/min g/min
Grams per hour g/hr g/h
Kilograms per second kg/sec kg/s
Kilograms per minute kg/min kg/min
Kilograms per hour kg/hr kg/h
ProLink III Field Communicator
Kilograms per day kg/day kg/d
Metric tons per minute mTon/min MetTon/min
Metric tons per hour mTon/hr MetTon/h
Metric tons per day mTon/day MetTon/d
Pounds per second lbs/sec lb/s
Configuration and Use Manual 19
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019023
Label
Unit description
Pounds per minute lbs/min lb/min
Pounds per hour lbs/hr lb/h
Pounds per day lbs/day lb/d
Short tons (2000 pounds) per minute sTon/min STon/min
Short tons (2000 pounds) per hour sTon/hr STon/h
Short tons (2000 pounds) per day sTon/day STon/d
Long tons (2240 pounds) per hour lTon/hr LTon/h
Long tons (2240 pounds) per day lTon/day LTon/d
Special unit special Spcl
ProLink III Field Communicator
Define a special measurement unit for mass flow
ProLink III Device Tools Configuration Process Measurement Flow Special Units
Field Communicator Configure Manual Setup Measurements Special Units Mass Special Units
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Procedure
1. Specify Base Mass Unit.
Base Mass Unit is the existing mass unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Mass Flow Conversion Factor as follows:
a) x base units = y special units
b) Mass Flow Conversion Factor = x ÷ y
The original mass flow rate value is divided by this value.
4. Enter Mass Flow Conversion Factor.
5. Set Mass Flow Label to the name you want to use for the mass flow unit.
6. Set Mass Total Label to the name you want to use for the mass total and mass inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for mass flow
You want to measure mass flow in ounces per second (oz/sec).
1. Set Base Mass Unit to Pounds (lb).
20 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Configure process measurement
MMI-20019023 February 2022
2. Set Base Time Unit to Seconds (sec).
3. Calculate Mass Flow Conversion Factor:
a. 1 lb/sec = 16 oz/sec
b. Mass Flow Conversion Factor = 1 ÷ 16 = 0.0625
4. Set Mass Flow Conversion Factor to 0.0625.
5. Set Mass Flow Label to oz/sec.
6. Set Mass Total Label to oz.
4.1.2 Configure Flow Damping
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements Flow Flow Damping
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
The default value is 0.8 seconds. The range depends on the core processor type and the setting of Update Rate, as shown in the following table.
Update Rate setting
Normal 0 to 51.2 seconds
Special 0 to 40.96 seconds
Damping range
The value you enter is automatically rounded off to the nearest valid value. For example, if the damping is currently set to 0.8 seconds, any value entered up to 1.2 seconds will be rounded down to 0.8 seconds, and any value entered from 1.21 to 1.59 seconds will be rounded up to 1.6 seconds.
Update Rate setting
Normal 0.0, 0.2, 0.4, 0.8, 1.6, 3.2, 6.4, 12.8, 25.6, 51.2
Special 0.0, 0.04, 0.08, 0.16, 0.32, 0.64, 1.28, 2.56, 5.12, 10.24,
Valid damping values
20.48, 40.96
Configuration and Use Manual 21
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019023
Effect of flow damping on volume measurement
Flow damping affects volume measurement for liquid volume data. Flow damping also affects volume measurement for gas standard volume data. The transmitter calculates volume data from the damped mass flow data.
Interaction between Flow Damping and mA Output Damping
In some circumstances, both Flow Damping and mA Output Damping are applied to the reported mass flow value.
Flow Damping controls the rate of change in flow process variables. mA Output Damping controls the rate of change reported through mA Output. If mA Output Process Variable is set to Mass Flow Rate, and both Flow Damping and mA Output Damping are set to non-zero values, flow damping is applied first, and the added damping calculation is applied to the result of the first calculation.
4.1.3 Configure Mass Flow Cutoff
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements Flow Mass Flow Cutoff
Mass Flow Cutoff specifies the lowest mass flow rate that will be reported as measured. All mass flow rates below this cutoff will be reported as 0.
Procedure
Set Mass Flow Cutoff to the value you want to use.
The default value for Mass Flow Cutoff is 0.0 g/sec or a sensor-specific value set at the factory. The recommended value is 0.5% of the nominal flow rate of the attached sensor. See the sensor specifications. Do not leave Mass Flow Cutoff at 0.0 g/sec.
Effect of Mass Flow Cutoff on volume measurement
Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the actual mass data rather than the reported value.
Volume flow has a separate Volume Flow Cutoff that is not affected by the Mass Flow Cutoff value.
Interaction between Mass Flow Cutoff and mA Output Cutoff
Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported through mA Output. If mA Output Process Variable is set to Mass Flow Rate, the mass flow rate reported through mA Output is controlled by the higher
of the two cutoff values.
Mass Flow Cutoff affects all reported values and values used in other transmitter behavior (e.g., events defined on mass flow).
mA Output Cutoff affects only mass flow values reported through mA Output.
22 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Configure process measurement
MMI-20019023 February 2022
Example: Cutoff interaction with mA Output Cutoff lower than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
mA Output Cutoff: 10 g/sec
Mass Flow Cutoff: 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mA Output Cutoff higher than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
mA Output Cutoff: 15 g/sec
Mass Flow Cutoff: 10 g/sec
Result:
If the mass flow rate drops below 15 g/sec but not below 10 g/sec:The mA Output will report zero flow.
The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
4.2 Configure volume flow measurement for liquid
applications
The volume flow measurement parameters control how liquid volume flow is measured and reported.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.
Note
If you need to switch from gas standard volume to liquid volume, polling for base density will automatically be disabled.
4.2.1 Configure Volume Flow Type for liquid applications
ProLink III
Field Communicator Configure Manual Setup Measurements GSV Volume Flow Type Liquid
Configuration and Use Manual 23
Device Tools Configuration Process Measurement Flow
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019023
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type to Liquid if you are using any of the following applications:
Production Volume Reconciliation (PVR)
Procedure
Set Volume Flow Type to Liquid.
4.2.2 Configure Volume Flow Measurement Unit for liquid
applications
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements Flow Volume Flow Unit
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the volume flow rate. The unit used for the volume total and volume inventory is based on this unit.
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
To read US gallons, select that unit from this menu. G/MIN stands for grams per minute (USGPM), not gallons per minute. The default setting for Volume Flow Measurement Unit is l/sec (liters per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement Unit, plus one user-defined measurement unit. Different communications tools may use different labels for the units.
Label
Unit description
Cubic feet per second ft3/sec Cuft/s
Cubic feet per minute ft3/min Cuft/min
Cubic feet per hour ft3/hr Cuft/h
Cubic feet per day ft3/day Cuft/d
ProLink III Field Communicator
Cubic meters per second m3/sec Cum/s
Cubic meters per minute m3/min Cum/min
Cubic meters per hour m3/hr Cum/h
24 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Configure process measurement
MMI-20019023 February 2022
Label
Unit description
ProLink III Field Communicator
Cubic meters per day m3/day Cum/d
U.S. gallons per second US gal/sec gal/s
U.S. gallons per minute US gal/min gal/min
U.S. gallons per hour US gal/hr gal/h
U.S. gallons per day US gal/day gal/d
Million U.S. gallons per day mil US gal/day MMgal/d
Liters per second l/sec L/s
Liters per minute l/min L/min
Liters per hour l/hr L/h
Million liters per day mil l/day ML/d
Imperial gallons per second Imp gal/sec Impgal/s
Imperial gallons per minute Imp gal/min Impgal/min
Imperial gallons per hour Imp gal/hr Impgal/h
Imperial gallons per day Imp gal/day Impgal/d
(1)
(1)
(1)
(1)
(2)
(2)
(2)
(2)
barrels/sec bbl/s
barrels/min bbl/min
barrels/hr bbl/h
barrels/day bbl/d
Beer barrels/sec bbbl/s
Beer barrels/min bbbl/min
Beer barrels/hr bbbl/h
Beer barrels/day bbbl/d
Barrels per second
Barrels per minute
Barrels per hour
Barrels per day
Beer barrels per second
Beer barrels per minute
Beer barrels per hour
Beer barrels per day
Special unit special Spcl
(1) Unit based on oil barrels (42 U.S. gallons). (2) Unit based on U.S. beer barrels (31 U.S. gallons).
Define a special measurement unit for volume flow
ProLink III
Field Communicator Configure Manual Setup Measurements Special Units Volume Special Units
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Note
Configuration and Use Manual 25
Device Tools Configuration Process Measurement Flow Special Units
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019023
Procedure
1. Specify Base Volume Unit.
Base Volume Unit is the existing volume unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Volume Flow Conversion Factor as follows:
a) x base units = y special units
b) Volume Flow Conversion Factor = x ÷ y
4. Enter Volume Flow Conversion Factor.
The original volume flow rate value is divided by this conversion factor.
5. Set Volume Flow Label to the name you want to use for the volume flow unit.
6. Set Volume Total Label to the name you want to use for the volume total and volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Defining a special measurement unit for volume flow
You want to measure volume flow in pints per second (pints/sec).
1. Set Base Volume Unit to Gallons (gal).
2. Set Base Time Unit to Seconds (sec).
3. Calculate the conversion factor:
a. 1 gal/sec = 8 pints/sec
b. Volume Flow Conversion Factor = 1 ÷ 8 = 0.1250
4. Set Volume Flow Conversion Factor to 0.1250.
5. Set Volume Flow Label to pints/sec.
6. Set Volume Total Label to pints.
4.2.3 Configure Volume Flow Cutoff
ProLink III
Field Communicator Configure Manual Setup Measurements Flow Volume Flow Cutoff
Volume Flow Cutoff specifies the lowest volume flow rate that will be reported as measured. All volume flow rates below this cutoff are reported as 0.
Device Tools Configuration Process Measurement Flow
Procedure
Set Volume Flow Cutoff to the value you want to use.
The default value for Volume Flow Cutoff is 0.0 l/sec (liters per second). The lower limit is 0. Leaving the volume flow cutoff at 0 is not recommended.
26 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Configure process measurement
MMI-20019023 February 2022
Interaction between Volume Flow Cutoff and mAO Cutoff
Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will report as measured. mAO Cutoff defines the lowest flow rate that will be reported through mA Output. If mA Output Process Variable is set to Volume Flow Rate, the volume flow rate reported through mA Output is controlled by the
higher of the two cutoff values.
Volume Flow Cutoff affects both the volume flow values reported via the outputs and the volume flow values used in other transmitter behavior (e.g., events defined on the volume flow).
mAO Cutoff affects only flow values reported through mA Output.
Example: Cutoff interaction with mAO Cutoff lower than Volume Flow Cutoff
Configuration:
mA Output Process Variable: Volume Flow Rate
Frequency Output Process Variable: Volume Flow Rate
AO Cutoff: 10 l/sec
Volume Flow Cutoff: 15 l/sec
Result: If the volume flow rate drops below 15 l/sec, volume flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mAO Cutoff higher than Volume Flow Cutoff
Configuration:
mA Output Process Variable: Volume Flow Rate
Frequency Output Process Variable: Volume Flow Rate
AO Cutoff: 15 l/sec
Volume Flow Cutoff: 10 l/sec
Result:
If the volume flow rate drops below 15 l/sec but not below 10 l/sec:The mA Output will report zero flow.
The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
If the volume flow rate drops below 10 l/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.

4.3 Configure GSV flow measurement

The gas standard volume (GSV) flow measurement parameters control how volume flow is measured and reported in a gas application.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.
Configuration and Use Manual 27
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019023
4.3.1 Configure Volume Flow Type for gas applications
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements GSV Volume Flow Type
Volume Flow Type controls whether liquid or gas standard volume flow measurement is used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type to Liquid if you are using any of the following applications:
Production Volume Reconciliation (PVR)
Procedure
Set Volume Flow Type to Gas Standard Volume.
4.3.2 Configure Standard Density of Gas
ProLink III Device Tools Configuration Process Measurement Flow
Field Communicator Configure Manual Setup Measurements GSV Gas Ref Density
The Standard Density of Gas value is the gas density at standard reference conditions. Use it to convert the measured mass flow data to volume flow at reference conditions.
Prerequisites
Ensure that Density Measurement Unit is set to the measurement unit you want to use for Standard Density of Gas.
Procedure
From the Source field, choose the method to supply gas base density data and perform the required setup.
Option
Fixed Value or Digital Communications
Poll for external value The meter polls an external HART device for gas base density data in order
Description
A host writes gas base density data to the meter at appropriate intervals. Continue to Configure fixed value or digital communications.
to then compute gas standard volume from the mass flow and gas base density.
Continue to Poll for external value.
Configure fixed value or digital communications
Prerequisites
Configure Standard Density of Gas
Procedure
1. Set Standard Density of Gas to the standard reference density of the gas you are measuring.
28 Micro Motion 1500 Transmitters with Analog Outputs
Configuration and Use Manual Configure process measurement
MMI-20019023 February 2022
Note
ProLink III provides a guided method that you can use to calculate your gas base density, if you do not know it.
2. Continue to Configure Gas Standard Volume Flow Unit .
Poll for external value
Prerequisites
Configure Standard Density of Gas
Procedure
1. Set Polling Slot to an available slot.
2. Set Polling Control n as one of the following options:
The n is the value you selected in the Polling Slot field.
If there is another master, and if that master is primary, then set this field to secondary. If the other master is secondary, then set this field to primary.
Option Description
Poll as Primary No other HART masters will be on the network.
Poll as Secondary Other HART masters will be on the network.
3. Set External Device Tag n to the HART tag of the device being polled.
The n is the value you selected in the Polling Slot field.
The device being polled (slave) cannot have special units set for density. Otherwise, the master will
reject the base density and report the following alarm:
A115: No External Input or Polled Data Alert
On the slave side, setup the HART Primary Variable for Base Density. The master will reject anything
other than Base Density for the HART Primary Variable and trigger an A115 alarm.
The density units on the transmitter and the polled device can be different as long as they can be
classified as density units; for example, kg/m3 and g/cm3. The transmitter converts the polled units into compatible specified units.
For wiring and setup instructions for a polled device, refer to the Micro Motion Gas Density Meters (GDM) Installation manual or the Micro Motion Specific Gravity Meters (SGM) Installation manual.
4. Continue to Configure Gas Standard Volume Flow Unit .
4.3.3 Configure Gas Standard Volume Flow Unit
ProLink III
Field Communicator Configure Manual Setup Measurements GSV GSV Flow Unit
Configuration and Use Manual 29
Device Tools Configuration Process Measurement Flow
Configure process measurement Configuration and Use Manual
February 2022 MMI-20019023
Gas Standard Volume Flow Unit specifies the unit of measure that will be displayed for the gas standard volume flow. The measurement unit used for the gas volume total and the gas volume inventory is derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Unit, be sure that Volume Flow Type is set to Gas Standard Volume.
For polling, the first transmitter (master) requests density from a second transmitter (slave) via HART communications. Special units for GSV are allowed on the master side, but the device being polled (slave) cannot have special units set for density, otherwise the master will reject the base density and report an A115: No External Input or Polled Data Alert.
Procedure
Set Gas Standard Volume Flow Unit to the unit you want to use.
The default setting for Gas Standard Volume Flow Unit is SCFM (Standard Cubic Feet per Minute).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Gas Standard Volume Flow Unit
The transmitter provides a standard set of measurement units for Gas Standard Volume Flow Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Label
Unit description
Normal cubic meters per second Nm3/sec Nm3/sec
Normal cubic meters per minute Nm3/sec Nm3/min
Normal cubic meters per hour Nm3/hr Nm3/hr
Normal cubic meters per day Nm3/day Nm3/day
Normal liters per second NLPS NLPS
Normal liters per minute NLPM NLPM
Normal liters per hour NLPH NLPH
Normal liters per day NLPD NLPD
Standard cubic feet per second SCFS SCFS
Standard cubic feet per minute SCFM SCFM
Standard cubic feet per hour SCFH SCFH
Standard cubic feet per day SCFD SCFD
ProLink III Field Communicator
Standard cubic meters per second Sm3/sec Sm3/sec
Standard cubic meters per minute Sm3/min Sm3/min
Standard cubic meters per hour Sm3/hr Sm3/hr
Standard cubic meters per day Sm3/day Sm3/day
30 Micro Motion 1500 Transmitters with Analog Outputs
Loading...
+ 152 hidden pages