Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully
before proceeding to the next step.
Emerson Flow customer service
Email:
•Worldwide: flow.support@emerson.com
•Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South AmericaEurope and Middle EastAsia Pacific
United States800-522-6277U.K.0870 240 1978Australia800 158 727
Appendix BUsing the transmitter display ..................................................................................... 139
B.1Components of the transmitter interface ................................................................................ 139
B.2Use the optical switches .......................................................................................................... 139
B.3Access and use the display menu system .................................................................................140
B.3.1Enter a floating-point value using the display ............................................................141
B.4Display codes for process variables ..........................................................................................144
B.5Codes and abbreviations used in display menus ...................................................................... 145
Appendix CUsing ProLink III with the transmitter .........................................................................157
C.1Basic information about ProLink III ...........................................................................................157
C.2Connect with ProLink III ........................................................................................................... 158
C.2.1Connection types supported by ProLink III ................................................................ 158
C.2.2Connect with ProLink III over Modbus/RS-485 ...........................................................159
C.2.3Connect with ProLink III over HART/Bell 202 ............................................................. 162
Appendix DUsing the Field Communicator with the transmitter ................................................... 171
D.1Basic information about the Field Communicator ....................................................................171
D.2Connect with the Field Communicator .................................................................................... 172
ivMicro Motion® Gas Density Meters (GDM)
Part I
Getting Started
Chapters covered in this part:
•Before you begin
•Orientation and planning
•Quick start
Getting Started
Configuration and Use Manual 1
Getting Started
2Micro Motion® Gas Density Meters (GDM)
1Before you begin
Topics covered in this chapter:
•About this manual
•Model codes and device types
•Communications tools and protocols
•Additional documentation and resources
1.1About this manual
This manual provides information to help you configure, commission, use, maintain, and
troubleshoot the Micro Motion Gas Density Meter (GDM).
Important
This manual assumes that the following conditions apply:
•The meter has been installed correctly and completely, according to the instructions in the
installation manual.
•The installation complies with all applicable safety requirements.
•The user is trained in all government and corporate safety standards.
Before you begin
1.2Model codes and device types
Your device can be identified by the model code on the device tag.
Model codes and device typesTable 1-1:
Model codeDevice nicknameI/O
GDM****CGDM mA• Two mA outputs
• RS-485 terminals
GDM****DGDM DO• One mA output
• One discrete output
• RS-485 terminals
GDM****BGDM TPS• One mA output
• One Time Period Sig-
nal output
• RS-485 terminals
GDM****EGDM Fixed• One Time Period Sig-
nal output
Electronics mounting
Integral
Integral
Integral
Integral
Configuration and Use Manual 3
Before you begin
Restriction
The GDM mA and GDM DO support a complete set of application and configuration options. The
GDM TPS and GDM Fixed support a subset of application and configuration options. Refer to the
product data sheet for details.
1.3Communications tools and protocols
You can use several different communications tools and protocols to interface with the
device. You may use different tools in different locations or for different tasks.
Communications tools, protocols, and related informationTable 1-2:
Communications toolSupported protocolsScopeIn this manualFor more information
DisplayNot applicableBasic configuration and
commissioning
ProLink III• Modbus/RS-485
• HART/Bell 202
• Service port
Field Communicator
• HART/Bell 202Complete configuration
Complete configuration
and commissioning
and commissioning
Complete user information. See Appendix B.
Basic user information.
See Appendix C.
Basic user information.
See Appendix D.
Not applicable
User manual
• Installed with soft-
ware
• On Micro Motion
user documentation
CD
• On Micro Motion
web site
(www.micromo‐
tion.com)
User manual on
Micro Motion web site
(www.micromo‐
tion.com )
Tip
You may be able to use other communications tools from Emerson Process Management, such as
AMS Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the
Smart Wireless THUM Adapter is not discussed in this manual. For more information on the Smart
Wireless THUM Adapter, refer to the documentation available at www.micromotion.com.
1.4Additional documentation and resources
Micro Motion provides additional documentation to support the installation and operation
of the device.
4Micro Motion® Gas Density Meters (GDM)
Before you begin
Additional documentation and resourcesTable 1-3:
TopicDocument
Device installationMicro Motion Gas Density Meters (GDM): Installation Manual
Product data sheetMicro Motion Gas Density Meters: Product Data Sheet
All documentation resources are available on the Micro Motion web site at
www.micromotion.com or on the Micro Motion user documentation DVD.
Configuration and Use Manual 5
Before you begin
6Micro Motion® Gas Density Meters (GDM)
Orientation and planning
2Orientation and planning
Topics covered in this chapter:
•Terms and definitions
•Core process variable and available gas process variables
•GDM equations
2.1Terms and definitions
Terms used in meter setup and measurementTable 2-1:
TermDefinition or usage
Gas
Calibration gasThe gas used during factory calibration, typically nitrogen or argon.
Sample gasThe gas stream to be measured by the meter.
Pressure
Line pressureThe pressure in the main pipeline, independent of the meter.
Sample pressureThe pressure of the sample gas.
Measurement
Base density (standard density, normal density)
Calorific valueThe amount of heat released during the combustion of a specified amount of a
Compressibility factor “z”The correction factor for interactive molecular behavior of non-ideal gas mix-
Concentration (gas purity)In a gas mixture, the quantity of the primary gas in comparison to the quantity
Energy flowThe energy content of the process gas flowing through the pipe per unit of
Molecular weightThe ratio of the mass of a gas to its volume. Typically measured in g/mol.
Net mass flow rateThe flow rate as measured in mass flow units and multiplied by the current con-
Net volume flow rateThe flow rate as measured in volume flow units, corrected to base temperature
Relative densityThe ratio of the weight of a volume of gas (or gas mixture) to the weight of an
The absolute density of a gas at reference conditions (base temperature and
base pressure). Can be used to calculate standard volume flow from mass flow.
Measured in user-specified units.
gas. Measured in units of energy per units of the gas. Energy = calorific value.
tures.
of the secondary gas (contaminant). Measured in user-specified units.
time. Measured in units of energy per units of time.
centration value.
and base pressure, and multiplied by the current concentration value.
equal volume of dry air, where the weights of both the gas and air are taken
under identical conditions of temperature and pressure. Unitless.
Configuration and Use Manual 7
Orientation and planning
Terms used in meter setup and measurement (continued)Table 2-1:
TermDefinition or usage
Specific gravityThe ratio of the molecular weight of a gas (or gas mixture) to the molecular
weight of dry air. The molecular weight of dry air is normally assumed to be
28.96469. Unitless.
Wobbe indexThe ratio of the calorific value of a gas to its specific gravity. Measured in volu-
metric units (BTU/SCF, and MJ/SCM).
2.2Core process variable and available gas process
variables
The gas process variables that the GDM can report are determined by the core process
variable that you select during calibration.
Core process variable and available process variablesTable 2-2:
Available process variablesDefault measurement unit
Specific gravityUnitless
Molecular weightg/mol
Relative densityUnitless
Base densityg/cc
Line densityg/cc
Base compressibilityUnitless
Calorific valueMJ/Nm3
Wobbe indexMJ/Nm3
Energy flowMJ/hr
Concentration (gas purity)Concentration (% mass)
Net mass flow rateg/cc
Net volume flow rateSCFM
2.3GDM equations
Line density
All of the pressures in the following line density calculations are calculated in absolute
pressure. If gauge pressure is used for the polling and polled device, a new parameter is set
to convert the input pressure to absolute.
8Micro Motion® Gas Density Meters (GDM)
Orientation and planning
Uncorrected density Equation 2-1:
Du = K0+ K1 × τ + K2 × τ
2
Du
K0, K1, and K
τ
Uncorrected density (kg/m3)
Instrument calibration factors
2
Instrument time period (µs). Normally the upper frequency is 3dB point, but it can
also be the time period of the resonance peak.
Temperature-corrected density Equation 2-2:
Dт= Du × 1 + K18×(т –20)+ K19×(т –20
Dт
K18 and K
19
т
Note
Depending on your calibration option, you may have selectable K18 and K19 calibrations for multiple
temperature ranges.
Temperature-corrected density (kg/m3)
Instrument calibration factors
Instrument temperature (°C)
)
Velocity-of -sound-corrected density Equation 2-3:
D
V
Dт
VOS
cal
VOS
meas
K3 and K4
DV= Dт × {1 +(K3 /(DT+ K4)) × (VOS
Velocity of sound-corrected density (kg/m3)
Temperature-corrected density (kg/m3)
Sound velocity of the calibration gas (m/s) / 293°K
Sound velocity of the measured gas (m/s) / temperature for the VOS of the measured
gas (°K)
VOS calibration coefficients
cal
– VOS
meas
) }
Note
VOScal and VOSmeas are defined as SG / γ0 of the fluid, and are entered as the VOS. γ0 = Low
pressure ratio of specific heats.
Configuration and Use Manual 9
Orientation and planning
Specific gravity
Specific gravity calculated from molecular weight Equation 2-4:
MW
SG =
MW
Gas
Air
SG
MW
MW
Specific gravity of process gas
Molecular weight of process gas (g/mol)
Gas
Molecular weight of air (user specified; default = 28.96469 g/mol)
Air
Molecular weight
Molecular weight calculated from line density Equation 2-5:
MW
ρ
Line
UGC
T
Line
Z
Line
P
Line
Molecular weight of process gas
Density of the process gas at line conditions
Universal gas constant
Temperature of the process gas at line conditions
Compressibility of the process gas at line conditions
Line pressure
Base density
MW
Gas
=
ρ
Line
× UGC × T
P
Line
Line
× Z
Line
Base density from line density Equation 2-6:
ρ
P
ρ
Z
T
Z
T
P
B
B
Line
Line
Line
B
B
Line
PB× ρ
ρB=
Base density
Base pressure
Density of the process gas at line conditions
Compressibility of the process gas at line conditions
Temperature of the process gas at line conditions
Compressibility of the process gas at base conditions
Temperature of the process gas at base conditions
Line pressure
× Z
Line
ZB× TB× P
Line
Line
× T
Line
10Micro Motion® Gas Density Meters (GDM)
Relative density
Orientation and planning
Relative density from base density Equation 2-7:
ρ
RD =
B
ρ
Air
RD
P
B
ρ
Air
Relative density
Base pressure
Relative density of air
Configuration and Use Manual 11
Orientation and planning
12Micro Motion® Gas Density Meters (GDM)
3Quick start
Topics covered in this chapter:
•Power up the transmitter
•Check meter status
•Make a startup connection to the transmitter
3.1Power up the transmitter
The transmitter must be powered up for all configuration and commissioning tasks, or for
process measurement.
1.Ensure that all transmitter and sensor covers and seals are closed.
WARNING!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers
and seals are tightly closed. For hazardous area installations, applying power while
housing covers are removed or loose can cause an explosion.
Quick start
2.Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. During this period,
Alert 009 is active. The diagnostic routines should complete in approximately
30 seconds.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics
can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial
startup, or if power has been off long enough to allow components to reach ambient
temperature, allow the electronics to warm up for approximately 10 minutes before
relying on process measurements. During this warm-up period, you may observe minor
measurement instability or inaccuracy.
3.2Check meter status
Check the meter for any error conditions that require user action or that affect
measurement accuracy.
1.Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions. During the power-up sequence, Alert A009 is active.
This alert should clear automatically when the power-up sequence is complete.
2.Check the status LED on the transmitter.
Configuration and Use Manual 13
Quick start
Transmitter status reported by status LEDTable 3-1:
LED stateDescriptionRecommendation
GreenNo alerts are active.Continue with configuration or process meas-
urement.
YellowOne or more low-severity alerts are active.A low-severity alert condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or process measurement. If you choose, you can identify and resolve the alert condition.
Flashing yellowCalibration in progress, or Known Density Veri-
fication in progress.
RedOne or more high-severity alerts are active.A high-severity alert condition affects meas-
The measurement can fluctuate during the
calibration process or change as a result of the
calibration process. The alert will clear when
the calibration is complete. Check the calibration results before continuing.
urement accuracy and output behavior. Resolve the alert condition before continuing.
•View and acknowledge status alerts (Section 9.3)
•Status alerts, causes, and recommendations (Section 11.6)
3.3Make a startup connection to the transmitter
For all configuration tools except the display, you must have an active connection to the
transmitter to configure the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in
the appropriate appendix. Use the default communications parameters shown in the
appendix.
Communications toolConnection type to useInstructions
ProLink IIIModbus/RS-485
HART/Bell 202
Field CommunicatorHART/Bell 202Appendix D
Postrequisites
(Optional) Change the communications parameters to site-specific values.
Appendix C
•To change the communications parameters using ProLink III, choose Device Tools >
Configuration > Communications.
•To change the communications parameters using the Field Communicator, choose
Configure > Manual Setup > HART > Communications.
14Micro Motion® Gas Density Meters (GDM)
Quick start
Important
If you are changing communications parameters for the connection type that you are using, you will
lose the connection when you write the parameters to the transmitter. Reconnect using the new
parameters.
Configuration and Use Manual 15
Quick start
16Micro Motion® Gas Density Meters (GDM)
Configuration and commissioning
Part II
Configuration and commissioning
Chapters covered in this part:
•Introduction to configuration and commissioning
•Configure process measurement
•Configure device options and preferences
•Integrate the meter with the control system
•Complete the configuration
Configuration and Use Manual 17
Configuration and commissioning
18Micro Motion® Gas Density Meters (GDM)
Introduction to configuration and commissioning
4Introduction to configuration and
commissioning
Topics covered in this chapter:
•Default values
•Enable access to the off‐line menu of the display
•Disable HART security
•Set the HART lock
•Restore the factory configuration
4.1Default values
Default values for your meter are configured at the factory.
Important
Default values are based on your purchase order options. Therefore, the default values described in
the following tables may not be the factory default values configured for your system. For absolute
accuracy, refer to the configuration sheet that was shipped with your meter.
4.1.1GDM default values
GDM default mA scaling valuesTable 4-1:
VariableDefault 4 mADefault 20 mA
Line density0.0 g/cc0.400 g/cc
Line temperature-50.000°C
Drive gain0.000 %100.000 %
External temperature-50.000°C
External pressure0.000 PSIg1450.377 PSIg
Base density0.000 g/cc0.400 g/cc
Relative density0.01.0
Molecular weight0.0 g/mol28.96469 g/mol
Specific gravity03
Calorific value20 MJ/Nm360 MJ/Nm3
Wobbe index20 MJ/Nm360 MJ/Nm3
Sensor time period400 us1200 us
-58°F
-58.00000°F
200.000°C
392°F
200.000°C
392.0000°F
Configuration and Use Manual 19
Introduction to configuration and commissioning
GDM default mA scaling values (continued)Table 4-1:
Field Communicator Configure > Manual Setup > Display > Display Menus > Offline Menu
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must
enable it if you want to use the display to configure the transmitter.
20Micro Motion® Gas Density Meters (GDM)
Restriction
A
You cannot use the display to enable access to the off-line menu. You must make a connection from
another tool.
4.3Disable HART security
If you plan to use HART protocol to configure the device, HART security must be disabled.
HART security is disabled by default, so you may not need to do this.
Prerequisites
•Strap wrench
•3 mm hex key
Procedure
1.Power down the meter.
2.Using the strap wrench, loosen the grub screws and remove the transmitter end-
cap.
Introduction to configuration and commissioning
Transmitter with end-cap removedFigure 4-1:
A. Transmitter end‐cap
3.Using the hex key, remove the safety spacer.
Configuration and Use Manual 21
A
B
A
B
Introduction to configuration and commissioning
Transmitter with end-cap and safety spacer removedFigure 4-2:
A. Transmitter end‐cap
B. Safety spacer
4.Move the HART security switch to the OFF position (up).
The HART security switch is the switch on the left.
HART security switchFigure 4-3:
A. HART security switch
B. Unused
22Micro Motion® Gas Density Meters (GDM)
5.Replace the safety spacer and end-cap.
6.Power up the meter.
4.4Set the HART lock
If you plan to use a HART connection to configure the device, you can lock out all other
HART masters. If you do this, other HART masters will be able to read data from the device
but will not be able to write data to the device.
Restrictions
•This feature is available only when you are using the Field Communicator or AMS.
•This feature is available only with a HART 7 host.
2.If you are locking the meter, set Lock Option as desired.
Introduction to configuration and commissioning
OptionDescription
Permanent Only the current HART master can make changes to the device. The device will
remain locked until manually unlocked by a HART master. The HART master can
also change Lock Option to Temporary.
Temporary Only the current HART master can make changes to the device. The device will
remain locked until manually unlocked by a HART master, or a power-cycle or
device reset is performed. The HART master can also change Lock Option to Perma-
nent.
Lock AllNo HART masters are allowed to make changes to the configuration. Before
changing Lock Option to Permanent or Temporary, the device must be unlocked. Any
HART master can be used to unlock the device.
Postrequisites
To avoid future confusion or difficulties, ensure that the device is unlocked after you have
completed your tasks.
4.5Restore the factory configuration
DisplayNot available
ProLink IIIDevice Tools > Configuration Transfer > Restore Factory Configuration
Field Communicator Service Tools > Maintenance > Reset/Restore > Restore Factory Configuration
Configuration and Use Manual 23
Introduction to configuration and commissioning
Overview
Restoring the factory configuration returns the transmitter to a known operational
configuration. This may be useful if you experience problems during configuration.
Tip
Restoring the factory configuration is not a common action. You may want to contact Micro Motion
to see if there is a preferred method to resolve any issues.
24Micro Motion® Gas Density Meters (GDM)