Configuration and Use Manual
MMI-20020949, Rev AC
Micro Motion® Gas Density Meters (GDM)
Configuration and Use Manual
April 2016

Safety messages

Emerson Flow customer service

Email:
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 704 136 666 New Zealand 099 128 804
Mexico +41 (0) 41 7686 111 France 0800 917 901 India 800 440 1468
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy 8008 77334 China +86 21 2892 9000
Venezuela +58 26 1731 3446 Central & Eastern +41 (0) 41 7686 111 Japan +81 3 5769 6803
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684

Contents

Contents
Part I Getting Started
Chapter 1 Before you begin ............................................................................................................3
1.1 About this manual ....................................................................................................................... 3
1.2 Model codes and device types ..................................................................................................... 3
1.3 Communications tools and protocols .......................................................................................... 4
1.4 Additional documentation and resources .................................................................................... 4
Chapter 2 Orientation and planning ...............................................................................................7
2.1 Terms and definitions .................................................................................................................. 7
2.2 Core process variable and available gas process variables ............................................................ 8
2.3 GDM equations ........................................................................................................................... 8
Chapter 3 Quick start ................................................................................................................... 13
3.1 Power up the transmitter ...........................................................................................................13
3.2 Check meter status ................................................................................................................... 13
3.3 Make a startup connection to the transmitter ............................................................................14
Part II Configuration and commissioning
Chapter 4 Introduction to configuration and commissioning ....................................................... 19
4.1 Default values ............................................................................................................................19
4.1.1 GDM default values .....................................................................................................19
4.2 Enable access to the off-line menu of the display ....................................................................... 20
4.3 Disable HART security ................................................................................................................21
4.4 Set the HART lock ...................................................................................................................... 23
4.5 Restore the factory configuration .............................................................................................. 23
Chapter 5 Configure process measurement ..................................................................................25
5.1 Verify the calibration factors ......................................................................................................25
5.1.1 Calibration factors ...................................................................................................... 26
5.2 Configure line density measurement ........................................................................................ 26
5.2.1 Configure Density Measurement Unit ................................................................................26
5.2.2 Configure Density Damping ........................................................................................... 28
5.2.3 Configure Density Cutoff ................................................................................................ 29
5.3 Configure temperature measurement .......................................................................................29
5.3.1 Configure Temperature Measurement Unit ........................................................................ 30
5.3.2 Configure Temperature Damping .....................................................................................30
5.3.3 Configure Temperature Input .......................................................................................... 31
5.4 Configure the pressure input ..................................................................................................... 35
5.4.1 Configure the pressure input using ProLink III ............................................................. 35
5.4.2 Configure the pressure input using the Field Communicator .......................................36
5.5 Configure energy content management ................................................................................... 37
5.5.1 Configure energy content management using ProLink III ............................................38
5.5.2 Configure energy content management using the Field Communicator .....................41
Configuration and Use Manual i
Contents
5.6 Configure base density measurement ....................................................................................... 42
5.7 Set up flow rate measurement ...................................................................................................43
5.7.1 Set up flow rate measurement using ProLink III ...........................................................43
5.7.2 Set up flow rate measurement using the Field Communicator .................................... 45
Chapter 6 Configure device options and preferences ....................................................................49
6.1 Configure the transmitter display .............................................................................................. 49
6.1.1 Configure the language used for the display ............................................................... 49
6.1.2 Configure the process variables and diagnostic variables shown on the display ...........50
6.1.3 Configure the number of decimal places (precision) shown on the display ..................50
6.1.4 Configure the refresh rate of data shown on the display ..............................................51
6.1.5 Enable or disable automatic scrolling through the display variables ............................ 51
6.2 Enable or disable the Acknowledge All Alerts display command ......................................................52
6.3 Configure security for the display menus .................................................................................. 53
6.4 Configure alert handling ............................................................................................................54
6.4.1 Configure Fault Timeout .................................................................................................54
6.4.2 Configure Alert Severity ................................................................................................. 55
6.5 Configure informational parameters ......................................................................................... 57
Chapter 7 Integrate the meter with the control system ................................................................59
7.1 Configure Channel B ..................................................................................................................59
7.2 Configure the mA output .......................................................................................................... 60
7.2.1 Configure mA Output Process Variable ............................................................................. 60
7.2.2 Configure Lower Range Value (LRV) and Upper Range Value (URV) ....................................62
7.2.3 Configure Added Damping ............................................................................................. 64
7.2.4 Configure mA Output Fault Action and mA Output Fault Level ...............................................65
7.3 Configure the discrete output ................................................................................................... 66
7.3.1 Configure Discrete Output Source ....................................................................................66
7.3.2 Configure Discrete Output Polarity ................................................................................... 67
7.3.3 Configure Discrete Output Fault Action ............................................................................. 68
7.4 Configure an enhanced event ....................................................................................................69
7.5 Configure HART/Bell 202 communications ............................................................................... 70
7.5.1 Configure basic HART parameters ...............................................................................70
7.5.2 Configure HART variables (PV, SV, TV, QV) ..................................................................71
7.5.3 Configure burst communications ............................................................................... 73
7.6 Configure Modbus communications ..........................................................................................78
7.7 Configure Digital Communications Fault Action ................................................................................. 79
7.7.1 Options for Digital Communications Fault Action .................................................................80
Chapter 8 Complete the configuration ......................................................................................... 81
8.1 Test or tune the system using sensor simulation ........................................................................81
8.2 Back up transmitter configuration ............................................................................................. 81
8.3 Enable HART security .................................................................................................................82
Part III Operations, maintenance, and troubleshooting
Chapter 9 Transmitter operation ................................................................................................. 87
9.1 Record the process variables ..................................................................................................... 87
9.2 View process variables and diagnostic variables .........................................................................87
9.2.1 View process variables using the display ..................................................................... 88
ii Micro Motion® Gas Density Meters (GDM)
Contents
9.2.2 View process variables and other data using ProLink III ............................................... 88
9.2.3 View process variables using the Field Communicator ................................................ 89
9.3 View and acknowledge status alerts .......................................................................................... 89
9.3.1 View and acknowledge alerts using the display .......................................................... 89
9.3.2 View and acknowledge alerts using ProLink III .............................................................91
9.3.3 View alerts using the Field Communicator .................................................................. 92
9.3.4 Alert data in transmitter memory ............................................................................... 92
Chapter 10 Measurement support ................................................................................................. 95
10.1 Perform the Known Density Verification procedure ................................................................... 95
10.1.1 Perform the Known Density Verification procedure using the display ......................... 96
10.1.2 Perform the Known Density Verification procedure using ProLink III ........................... 96
10.1.3 Perform the Known Density Verification procedure using the
Field Communicator ................................................................................................... 97
10.2 Adjust density measurement with Density Offset or Density Meter Factor ......................................... 97
10.3 Adjust temperature measurement with Temperature Offset or Temperature Slope ........................... 99
10.4 Perform temperature calibration .............................................................................................100
10.4.1 Perform temperature calibration using the display ................................................... 101
10.4.2 Perform temperature calibration using ProLink III ..................................................... 102
10.4.3 Perform temperature calibration using the Field Communicator .............................. 103
10.5 Configure VOS compensation ................................................................................................. 103
10.5.1 VOS User G ............................................................................................................... 104
10.6 Set up user-defined calculations .............................................................................................. 105
10.6.1 Equations used in user-defined calculations ..............................................................106
10.6.2 Measurement units used in user-defined calculations ............................................... 107
10.7 Diagnostic density ...................................................................................................................107
10.7.1 Enable diagnostic input density ................................................................................ 108
10.7.2 Enable calculated diagnostic density .........................................................................108
Chapter 11 Troubleshooting ........................................................................................................ 111
11.1 Quick guide to troubleshooting ...............................................................................................111
11.2 Check power supply wiring ......................................................................................................112
11.3 Check grounding .....................................................................................................................113
11.4 Perform loop tests ...................................................................................................................113
11.4.1 Perform loop tests using the display ......................................................................... 113
11.4.2 Perform loop tests using ProLink III ........................................................................... 115
11.4.3 Perform loop tests using the Field Communicator .................................................... 116
11.5 Status LED states ..................................................................................................................... 117
11.6 Status alerts, causes, and recommendations ........................................................................... 118
11.7 Density measurement problems ............................................................................................. 122
11.8 Temperature measurement problems .....................................................................................123
11.8.1 Thermal insulation ....................................................................................................124
11.9 Gas measurement problems ................................................................................................... 124
11.10 Concentration measurement problems ...................................................................................125
11.11 Milliamp output problems ....................................................................................................... 125
11.12 Discrete output problems ........................................................................................................127
11.13 Time Period Signal (TPS) output problems ...............................................................................127
11.14 Using sensor simulation for troubleshooting ........................................................................... 128
11.15 Trim mA outputs ..................................................................................................................... 128
11.15.1 Trim mA outputs using ProLink III ..............................................................................128
11.15.2 Trim mA outputs using the Field Communicator .......................................................129
11.16 Check HART communications ................................................................................................. 130
11.17 Check Lower Range Value and Upper Range Value ......................................................................... 131
Configuration and Use Manual iii
Contents
11.18 Check mA Output Fault Action ...................................................................................................... 132
11.19 Check for radio frequency interference (RFI) ............................................................................132
11.20 Check the cutoffs .................................................................................................................... 132
11.21 Check the drive gain ................................................................................................................ 133
11.21.1 Excessive or erratic drive gain ....................................................................................133
11.21.2 Collect drive gain data .............................................................................................. 134
11.22 Check the pickoff voltage ........................................................................................................ 134
11.22.1 Collect pickoff voltage data ...................................................................................... 135
11.23 Check for internal electrical problems ..................................................................................... 135
11.24 Locate a device using the HART 7 Squawk feature ................................................................... 135
Appendices and reference
Appendix A Calibration certificate ................................................................................................ 137
A.1 Sample calibration certificate ................................................................................................. 137
Appendix B Using the transmitter display ..................................................................................... 139
B.1 Components of the transmitter interface ................................................................................ 139
B.2 Use the optical switches .......................................................................................................... 139
B.3 Access and use the display menu system .................................................................................140
B.3.1 Enter a floating-point value using the display ............................................................141
B.4 Display codes for process variables ..........................................................................................144
B.5 Codes and abbreviations used in display menus ...................................................................... 145
Appendix C Using ProLink III with the transmitter .........................................................................157
C.1 Basic information about ProLink III ...........................................................................................157
C.2 Connect with ProLink III ........................................................................................................... 158
C.2.1 Connection types supported by ProLink III ................................................................ 158
C.2.2 Connect with ProLink III over Modbus/RS-485 ...........................................................159
C.2.3 Connect with ProLink III over HART/Bell 202 ............................................................. 162
Appendix D Using the Field Communicator with the transmitter ................................................... 171
D.1 Basic information about the Field Communicator ....................................................................171
D.2 Connect with the Field Communicator .................................................................................... 172
iv Micro Motion® Gas Density Meters (GDM)
Part I
Getting Started
Chapters covered in this part:
Before you begin
Orientation and planning
Quick start
Getting Started
Configuration and Use Manual 1
Getting Started
2 Micro Motion® Gas Density Meters (GDM)
1 Before you begin
Topics covered in this chapter:

About this manual

Model codes and device types
Communications tools and protocols
Additional documentation and resources
1.1 About this manual
This manual provides information to help you configure, commission, use, maintain, and troubleshoot the Micro Motion Gas Density Meter (GDM).
Important
This manual assumes that the following conditions apply:
The meter has been installed correctly and completely, according to the instructions in the
installation manual.
The installation complies with all applicable safety requirements.
The user is trained in all government and corporate safety standards.

Before you begin

1.2 Model codes and device types
Your device can be identified by the model code on the device tag.

Model codes and device typesTable 1-1:

Model code Device nickname I/O
GDM****C GDM mA Two mA outputs
RS-485 terminals
GDM****D GDM DO One mA output
One discrete output
RS-485 terminals
GDM****B GDM TPS One mA output
One Time Period Sig-
nal output
RS-485 terminals
GDM****E GDM Fixed One Time Period Sig-
nal output
Electronics mount­ing
Integral
Integral
Integral
Integral
Configuration and Use Manual 3
Before you begin
Restriction
The GDM mA and GDM DO support a complete set of application and configuration options. The GDM TPS and GDM Fixed support a subset of application and configuration options. Refer to the product data sheet for details.

1.3 Communications tools and protocols

You can use several different communications tools and protocols to interface with the device. You may use different tools in different locations or for different tasks.
Communications tools, protocols, and related informationTable 1-2:
Communica­tions tool Supported protocols Scope In this manual For more information
Display Not applicable Basic configuration and
commissioning
ProLink III Modbus/RS-485
HART/Bell 202
Service port
Field Commu­nicator
HART/Bell 202 Complete configuration
Complete configuration and commissioning
and commissioning
Complete user informa­tion. See Appendix B.
Basic user information. See Appendix C.
Basic user information. See Appendix D.
Not applicable
User manual
Installed with soft-
ware
On Micro Motion
user documentation CD
On Micro Motion
web site (www.micromo‐
tion.com)
User manual on Micro Motion web site (www.micromo‐
tion.com )
Tip
You may be able to use other communications tools from Emerson Process Management, such as AMS Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this manual. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at www.micromotion.com.

1.4 Additional documentation and resources

Micro Motion provides additional documentation to support the installation and operation of the device.
4 Micro Motion® Gas Density Meters (GDM)
Before you begin
Additional documentation and resourcesTable 1-3:
Topic Document
Device installation Micro Motion Gas Density Meters (GDM): Installation Manual
Product data sheet Micro Motion Gas Density Meters: Product Data Sheet
All documentation resources are available on the Micro Motion web site at
www.micromotion.com or on the Micro Motion user documentation DVD.
Configuration and Use Manual 5
Before you begin
6 Micro Motion® Gas Density Meters (GDM)

Orientation and planning

2 Orientation and planning
Topics covered in this chapter:

Terms and definitions

Core process variable and available gas process variables
GDM equations
2.1 Terms and definitions
Terms used in meter setup and measurementTable 2-1:
Term Definition or usage
Gas
Calibration gas The gas used during factory calibration, typically nitrogen or argon.
Sample gas The gas stream to be measured by the meter.
Pressure
Line pressure The pressure in the main pipeline, independent of the meter.
Sample pressure The pressure of the sample gas.
Measurement
Base density (standard density, nor­mal density)
Calorific value The amount of heat released during the combustion of a specified amount of a
Compressibility factor “z” The correction factor for interactive molecular behavior of non-ideal gas mix-
Concentration (gas purity) In a gas mixture, the quantity of the primary gas in comparison to the quantity
Energy flow The energy content of the process gas flowing through the pipe per unit of
Molecular weight The ratio of the mass of a gas to its volume. Typically measured in g/mol.
Net mass flow rate The flow rate as measured in mass flow units and multiplied by the current con-
Net volume flow rate The flow rate as measured in volume flow units, corrected to base temperature
Relative density The ratio of the weight of a volume of gas (or gas mixture) to the weight of an
The absolute density of a gas at reference conditions (base temperature and base pressure). Can be used to calculate standard volume flow from mass flow. Measured in user-specified units.
gas. Measured in units of energy per units of the gas. Energy = calorific value.
tures.
of the secondary gas (contaminant). Measured in user-specified units.
time. Measured in units of energy per units of time.
centration value.
and base pressure, and multiplied by the current concentration value.
equal volume of dry air, where the weights of both the gas and air are taken under identical conditions of temperature and pressure. Unitless.
Configuration and Use Manual 7
Orientation and planning
Terms used in meter setup and measurement (continued)Table 2-1:
Term Definition or usage
Specific gravity The ratio of the molecular weight of a gas (or gas mixture) to the molecular
weight of dry air. The molecular weight of dry air is normally assumed to be
28.96469. Unitless.
Wobbe index The ratio of the calorific value of a gas to its specific gravity. Measured in volu-
metric units (BTU/SCF, and MJ/SCM).

2.2 Core process variable and available gas process variables

The gas process variables that the GDM can report are determined by the core process variable that you select during calibration.
Core process variable and available process variablesTable 2-2:
Available process variables Default measurement unit
Specific gravity Unitless
Molecular weight g/mol
Relative density Unitless
Base density g/cc
Line density g/cc
Base compressibility Unitless
Calorific value MJ/Nm3
Wobbe index MJ/Nm3
Energy flow MJ/hr
Concentration (gas purity) Concentration (% mass)
Net mass flow rate g/cc
Net volume flow rate SCFM

2.3 GDM equations

Line density
All of the pressures in the following line density calculations are calculated in absolute pressure. If gauge pressure is used for the polling and polled device, a new parameter is set to convert the input pressure to absolute.
8 Micro Motion® Gas Density Meters (GDM)
Orientation and planning
Uncorrected density Equation 2-1:
Du = K0+ K1 × τ + K2 × τ
2
Du
K0, K1, and K
τ
Uncorrected density (kg/m3)
Instrument calibration factors
2
Instrument time period (µs). Normally the upper frequency is 3dB point, but it can also be the time period of the resonance peak.
Temperature-corrected density Equation 2-2:
Dт= Du × 1 + K18×(т –20)+ K19×(т –20
K18 and K
19
т
Note
Depending on your calibration option, you may have selectable K18 and K19 calibrations for multiple temperature ranges.
Temperature-corrected density (kg/m3)
Instrument calibration factors
Instrument temperature (°C)
)
Velocity-of -sound-corrected density Equation 2-3:
D
V
VOS
cal
VOS
meas
K3 and K4
DV= Dт × {1 +(K3 /(DT+ K4)) × (VOS
Velocity of sound-corrected density (kg/m3)
Temperature-corrected density (kg/m3)
Sound velocity of the calibration gas (m/s) / 293°K
Sound velocity of the measured gas (m/s) / temperature for the VOS of the measured gas (°K)
VOS calibration coefficients
cal
– VOS
meas
) }
Note
VOScal and VOSmeas are defined as SG / γ0 of the fluid, and are entered as the VOS. γ0 = Low pressure ratio of specific heats.
Configuration and Use Manual 9
Orientation and planning
Specific gravity
Specific gravity calculated from molecular weight Equation 2-4:
MW
SG =
MW
Gas
Air
SG
MW
MW
Specific gravity of process gas
Molecular weight of process gas (g/mol)
Gas
Molecular weight of air (user specified; default = 28.96469 g/mol)
Air
Molecular weight
Molecular weight calculated from line density Equation 2-5:
MW
ρ
Line
UGC
T
Line
Z
Line
P
Line
Molecular weight of process gas
Density of the process gas at line conditions
Universal gas constant
Temperature of the process gas at line conditions
Compressibility of the process gas at line conditions
Line pressure
Base density
MW
Gas
=
ρ
Line
× UGC × T
P
Line
Line
× Z
Line
Base density from line density Equation 2-6:
ρ
P
ρ
Z
T
Z
T
P
B
B
Line
Line
Line
B
B
Line
PB× ρ
ρB=
Base density
Base pressure
Density of the process gas at line conditions
Compressibility of the process gas at line conditions
Temperature of the process gas at line conditions
Compressibility of the process gas at base conditions
Temperature of the process gas at base conditions
Line pressure
× Z
Line
ZB× TB× P
Line
Line
× T
Line
10 Micro Motion® Gas Density Meters (GDM)
Relative density
Orientation and planning
Relative density from base density Equation 2-7:
ρ
RD =
B
ρ
Air
RD
P
B
ρ
Air
Relative density
Base pressure
Relative density of air
Configuration and Use Manual 11
Orientation and planning
12 Micro Motion® Gas Density Meters (GDM)
3 Quick start
Topics covered in this chapter:

Power up the transmitter

Check meter status

Make a startup connection to the transmitter
3.1 Power up the transmitter
The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement.
1. Ensure that all transmitter and sensor covers and seals are closed.
WARNING!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are tightly closed. For hazardous area installations, applying power while housing covers are removed or loose can cause an explosion.

Quick start

2. Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. During this period, Alert 009 is active. The diagnostic routines should complete in approximately 30 seconds.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power has been off long enough to allow components to reach ambient temperature, allow the electronics to warm up for approximately 10 minutes before relying on process measurements. During this warm-up period, you may observe minor measurement instability or inaccuracy.
3.2 Check meter status
Check the meter for any error conditions that require user action or that affect measurement accuracy.
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alert A009 is active. This alert should clear automatically when the power-up sequence is complete.
2. Check the status LED on the transmitter.
Configuration and Use Manual 13
Quick start
Transmitter status reported by status LEDTable 3-1:
LED state Description Recommendation
Green No alerts are active. Continue with configuration or process meas-
urement.
Yellow One or more low-severity alerts are active. A low-severity alert condition does not affect
measurement accuracy or output behavior. You can continue with configuration or proc­ess measurement. If you choose, you can iden­tify and resolve the alert condition.
Flashing yellow Calibration in progress, or Known Density Veri-
fication in progress.
Red One or more high-severity alerts are active. A high-severity alert condition affects meas-
The measurement can fluctuate during the calibration process or change as a result of the calibration process. The alert will clear when the calibration is complete. Check the calibra­tion results before continuing.
urement accuracy and output behavior. Re­solve the alert condition before continuing.
View and acknowledge status alerts (Section 9.3)
Status alerts, causes, and recommendations (Section 11.6)

3.3 Make a startup connection to the transmitter

For all configuration tools except the display, you must have an active connection to the transmitter to configure the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in the appropriate appendix. Use the default communications parameters shown in the appendix.
Communications tool Connection type to use Instructions
ProLink III Modbus/RS-485
HART/Bell 202
Field Communicator HART/Bell 202 Appendix D
Postrequisites
(Optional) Change the communications parameters to site-specific values.
Appendix C
To change the communications parameters using ProLink III, choose Device Tools >
Configuration > Communications.
To change the communications parameters using the Field Communicator, choose
Configure > Manual Setup > HART > Communications.
14 Micro Motion® Gas Density Meters (GDM)
Quick start
Important
If you are changing communications parameters for the connection type that you are using, you will lose the connection when you write the parameters to the transmitter. Reconnect using the new parameters.
Configuration and Use Manual 15
Quick start
16 Micro Motion® Gas Density Meters (GDM)
Configuration and commissioning
Part II
Configuration and commissioning
Chapters covered in this part:
Introduction to configuration and commissioning
Configure process measurement
Configure device options and preferences
Integrate the meter with the control system
Complete the configuration
Configuration and Use Manual 17
Configuration and commissioning
18 Micro Motion® Gas Density Meters (GDM)

Introduction to configuration and commissioning

4 Introduction to configuration and
commissioning
Topics covered in this chapter:

Default values

Enable access to the off‐line menu of the display
Disable HART security
Set the HART lock
Restore the factory configuration
4.1 Default values
Default values for your meter are configured at the factory.
Important
Default values are based on your purchase order options. Therefore, the default values described in the following tables may not be the factory default values configured for your system. For absolute accuracy, refer to the configuration sheet that was shipped with your meter.
4.1.1 GDM default values
GDM default mA scaling valuesTable 4-1:
Variable Default 4 mA Default 20 mA
Line density 0.0 g/cc 0.400 g/cc
Line temperature -50.000°C
Drive gain 0.000 % 100.000 %
External temperature -50.000°C
External pressure 0.000 PSIg 1450.377 PSIg
Base density 0.000 g/cc 0.400 g/cc
Relative density 0.0 1.0
Molecular weight 0.0 g/mol 28.96469 g/mol
Specific gravity 0 3
Calorific value 20 MJ/Nm3 60 MJ/Nm3
Wobbe index 20 MJ/Nm3 60 MJ/Nm3
Sensor time period 400 us 1200 us
-58°F
-58.00000°F
200.000°C
392°F
200.000°C
392.0000°F
Configuration and Use Manual 19
Introduction to configuration and commissioning
GDM default mA scaling values (continued)Table 4-1:
Variable Default 4 mA Default 20 mA
User-defined calculation out­put
%CO
2
%N
2
%H
2
%CO 0 % 100 %
Concentration Measurement enabled
Gas purity concentration 0.000 % 100.000 %
Flow input enabled
Energy flow 40.0 MJ/hr 120.0 MJ/hr
Mass flow rate (calculated) -200 g/cc 200 g/cc
Mass flow rate (external) -200 g/cc 200 g/cc
Volume flow rate (calculated) -0.42378 SCFM 0.42378 SCFM
Volume flow rate (external) -0.20000 0.20000
0 100
0 % 100 %
0 % 100 %
0 % 100 %
GDM default variablesTable 4-2:
Default variable Output option A Output options B and C
Primary Variable (PV), mA1 Sample Temperature Density
Secondary Variable (SV), mA2
Tertiary Variable (TV) Sensor Time Period Sensor Time Period
Quaternary Variable (QV) Drive Gain Drive Gain
Sensor Time Period Sample Temperature

4.2 Enable access to the off-line menu of the display

Display Not available
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Display Menus > Offline Menu
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must enable it if you want to use the display to configure the transmitter.
20 Micro Motion® Gas Density Meters (GDM)
Restriction
A
You cannot use the display to enable access to the off-line menu. You must make a connection from another tool.

4.3 Disable HART security

If you plan to use HART protocol to configure the device, HART security must be disabled. HART security is disabled by default, so you may not need to do this.
Prerequisites
Strap wrench
3 mm hex key
Procedure
1. Power down the meter.
2. Using the strap wrench, loosen the grub screws and remove the transmitter end-
cap.
Introduction to configuration and commissioning
Transmitter with end-cap removedFigure 4-1:
A. Transmitter end‐cap
3. Using the hex key, remove the safety spacer.
Configuration and Use Manual 21
A
B
A
B
Introduction to configuration and commissioning
Transmitter with end-cap and safety spacer removedFigure 4-2:
A. Transmitter end‐cap B. Safety spacer
4. Move the HART security switch to the OFF position (up).
The HART security switch is the switch on the left.
HART security switchFigure 4-3:
A. HART security switch B. Unused
22 Micro Motion® Gas Density Meters (GDM)
5. Replace the safety spacer and end-cap.
6. Power up the meter.

4.4 Set the HART lock

If you plan to use a HART connection to configure the device, you can lock out all other HART masters. If you do this, other HART masters will be able to read data from the device but will not be able to write data to the device.
Restrictions
This feature is available only when you are using the Field Communicator or AMS.
This feature is available only with a HART 7 host.
Procedure
1. Choose Configure > Manual Setup > Security > Lock/Unlock Device.
2. If you are locking the meter, set Lock Option as desired.
Introduction to configuration and commissioning
Option Description
Permanent Only the current HART master can make changes to the device. The device will
remain locked until manually unlocked by a HART master. The HART master can also change Lock Option to Temporary.
Temporary Only the current HART master can make changes to the device. The device will
remain locked until manually unlocked by a HART master, or a power-cycle or device reset is performed. The HART master can also change Lock Option to Perma-
nent.
Lock All No HART masters are allowed to make changes to the configuration. Before
changing Lock Option to Permanent or Temporary, the device must be unlocked. Any HART master can be used to unlock the device.
Postrequisites
To avoid future confusion or difficulties, ensure that the device is unlocked after you have completed your tasks.

4.5 Restore the factory configuration

Display Not available
ProLink III Device Tools > Configuration Transfer > Restore Factory Configuration
Field Communicator Service Tools > Maintenance > Reset/Restore > Restore Factory Configuration
Configuration and Use Manual 23
Introduction to configuration and commissioning
Overview
Restoring the factory configuration returns the transmitter to a known operational configuration. This may be useful if you experience problems during configuration.
Tip
Restoring the factory configuration is not a common action. You may want to contact Micro Motion to see if there is a preferred method to resolve any issues.
24 Micro Motion® Gas Density Meters (GDM)