Micro Motion pursues a policy of continuous development and product improvement. The specification in this document may
therefore be changed without notice. To the best of our knowledge, the information contained in this document is accurate and
Micro Motion cannot be held responsible for any errors, omissions, or other misinformation contained herein. No part of this
document may be photocopied or reproduced without prior written consent of Micro Motion.
Handle the 7835/45/47 liquid density meter with great care.
•Do not drop the meter or subject it to severe mechanical shock.
•Do not expose the meter to excessive vibration.
•Ensure axial loading from pipework does not exceed 1/2 tonne.
•Ensure all electrical safety requirements are applied.
•Ensure the meter and associated pipework have been pressure tested to 1-1/2 times the
maximum operating pressure.
•Do not use liquids incompatible with the construction.
•Do not operate the meter above its rated pressure.
•Do not expose the meter to excessive vibration (> 0.5 g continuous).
•Ensure meter is not transported when it contains hazardous substances. This includes fluids
that may have leaked into, and are still contained, within the case.
•To return a meter, refer to Appendix I for more information on the Micro Motion return policy.
Safety messages are provided throughout this manual to protect personnel and equipment. Read each
safety message carefully before proceeding to the next step.
1.2Product overview
All of the products consist of a mechanical meter and an electronics unit that is normally mounted
inside the meter electronics housing. Together, the electronics and meter provide a system for
continuous on-line measurement of liquid density and temperature.
In applications where the pipeline temperature could exceed 110 °C (230 °F), the electronics should
be mounted in a Remote Amplifier Box (see Chapter 3 and Chapter 4 for more details.)
Liquid density is determined from the resonant frequency of a vibrating tube containing the liquid,
and liquid temperature is determined from a 100-Ω RTD .
1.3Product range
The meters are identical mechanically, except for the material used in the wetted parts and the
flanges/couplings. A fully welded design is utilized to ensure maximum reliability in the most severe
environments. A rupture disc is embodied in the meter end plate furthest from the amplifier housing.
This disc will rupture if a pressure build-up occurs within the case, in the unlikely event of a tube
assembly leak; the operation of the rupture disc is shown in Figure 2-2.
Table 1-1.Meter product range
MeterTube materialFeatures
7835NI-SPAN-C
7845316L Stainless
steel
7847316L Stainless
steel
®
Low temperature coefficient and long term stability, appropriate for fiscal
applications.
Good resistance to corrosion.
Designed for the hygiene requirements of the food processing industry and has
3A authorisation. Please refer to Section 2.7 for special cleaning and installation
requirements in hygienic applications.
1.4Electronics product range
The meters described above may be operated with any of the following electronics options. However,
the electronics options are not all directly interchangeable; this is because the Advanced electronics
boards are physically longer than the Standard electronics boards, and so only fit in the extended
electronics enclosures provided with an Advanced meter.
It should be noted that the Standard Entrained Gas Electronics are NOT suitable for operation in
hazardous areas, whereas all of the other electronics boards are suitable.
For further details of the performance of the different electronics boards, please refer to the
appropriate chapters.
2Micro Motion 7835/45/47 Liquid Density Meters
Introduction
Table 1-2.Electronics product range
Standard Electronics
Density version
Basic amplifier circuit providing a frequency signal (indicating
liquid density) and RTD resistance (indicating liquid temperature).
Interfaces with a Signal Converter.
Entrained gas
version
Advanced Electronics
Density version
Entrained gas
version
Remote Amplifier Box
Advanced
Electronics
Similar amplifier circuit to the above but meter operating at a lower
frequency which is more insensitive to entrained gas. It is NOT
suitable for operation in hazardous areas.
Microprocessor controlled circuit with a choice of analog and
digital outputs providing a direct measurement of density and
temperature. Also providing a selection of calculated parameters
and several diagnostic functions.
As above but meter operating at a lower frequency more
appropriate for liquids with entrained gas. Suitable for operation in
hazardous areas.
Suitable for operation in hazardous areas.
784521A: IIB Approval (ATEX)
784522A: IIC Approval (ATEX)
784523A: CSA Approval
784524A: Non-classified (supplied as spare for customer
installation).
Note that the advanced electronics amplifier cannot be fitted into
the Standard electronics Remote Amplifier Box.
1.5Advanced electronics
In contrast to a meter fitted with Standard electronics (which requires a Signal Converter for
operation), the meter with Advanced electronics will provide a complete measurement system.
An Advanced baseboard, plus meter, provides a complete system for measuring liquid density and
liquid temperature. A Remote Display or one of the optional boards may be required if outputs need
to be configured in the field or if additional functionality is required.
Installation & Configuration Manual3
Introduction
Baseboard
Option board
Remote Display
7965 DISPLAY
Figure 1-2.Advanced electronics baseboard and remote display
For convenience, the Advanced system design and performance are outlined below.
1.5.1Baseboard
The baseboard can be considered the heart of the system. Along with a liquid density meter, it
provides a complete system for measuring liquid density and liquid temperature. The baseboard
performs a range of useful calculations and provides the following outputs:
•Two fully configurable 4–20 mA outputs.
•One pulse output providing either an alarm status signal or the meter tube frequency.
•An RS-485 digital communications link using RTU Modbus protocol.
1.5.2Option board
The option board fits directly onto the baseboard. One option board is presently available.
•HART
®
board – providing an additional 4–20 mA output and full HART communications.
Only one board may be fitted at a time.
1.5.3Remote display
The 7965 remote display is intended for either handheld or wall-mounted use. It provides a
convenient means for displaying calculated data and for configuring or analyzing the system setup. It
communicates via the baseboard RS-485 digital communications link. One remote display can
communicate with a number of meters if they are connected together on the same RS-485 link.
1.62004/22/EC (MID) applications
4Micro Motion 7835/45/47 Liquid Density Meters
Mobrey Limited, a division of Emerson Process Management, has evaluated the 7835 and 7845 liquid
density meters against OIML R117-1:2007 and WELMEC guide 8.8 for use in measuring systems for
the continuous and dynamic measurement of quantities of liquids other than water. This evaluation
was in compliance with the European Measuring Instrument directive (2004/22/EC) annex MI-005.
Introduction
You may use the evaluation certificate for the 7835 and 7845 liquid density meters, with written
permission of Mobrey Limited to assist in obtaining an EC-type examination certificate for the
complete measuring system.
2.2.1General information applicable to the complete system
•These safety instructions are to be used whenever handling or operating this product. Suitably
trained personnel shall carry out the installation both mechanical and electrical in accordance
with the applicable local and national regulations and codes of practice for each discipline.
•Safe working practices for the media and process concerned must be followed during the
installation and maintenance of the equipment. Depressurize and isolate the system before
starting to loosen or remove any connection.
•If the equipment is likely to come into contact with aggressive substances, it is the
responsibility of the user to take suitable precautions that prevent it from being adversely
affected.
•It is the responsibility of the installer/user of this equipment to ensure:
•This product is not used as a support for other equipment or personnel.
•This product is protected from impact.
•It is important that this sensor is handled with care due to its weight and sensitivity to impact;
ensure lifting straps are fitted around flanged ends.
Installation & Configuration Manual7
Installation Procedure
2.2.2Pressure bearing parts
•It is the responsibility of the installer/user of this equipment to ensure:
•The materials of construction are suitable for the application.
•All piping connections conform to the local and national regulations and codes of practice.
•The pressure and temperature limits for this equipment are not exceeded, if necessary by
the use of suitable safety accessories. See Table 2-1.
Class 600 1440.2 psi (99.3 bar)1203.8 psi (83.0 bar)
Class 900 2159.6 psi (148.9 bar)1805.7 psi (124.5 bar)
PN40580.2 psi (40.0 bar)539.5 psi (37.2 bar)
PN1001450.4 psi (100.0 bar) 1348.9 psi (93.0 bar)
Pressure Rating
20°C110°C
•Correct gaskets/seals are fitted and are compatible with the media and process.
•The installed sensor is adequately supported for weight and vibration effects.
•Personnel are protected from hot burns by guards, thermal lagging or limited access.
Allow time to cool prior to carrying out maintenance operations. It is recommended that
“HOT” notices are fitted in the vicinity of the equipment where applicable.
•Regular inspection for corrosion and wear are carried out, both internal and external.
•The sensor must not be fitted until all installation work and final pre commissioning checks are
carried out. Do not remove blanking plugs until the sensor is fitted.
•The sensor must be installed in compliance with this manual, to ensure correct fitting. This
applies to all variants.
•The user should not repair this equipment, but general maintenance can be applied as
described within this manual.
8Micro Motion 7835/45/47 Liquid Density Meters
Installation Procedure
2.3Installation planning
When planning the installation of a meter, it is important to consider the following factors:
Table 2-2.Installation considerations
Safety
Serviceability
When installing in a process line, it is important that the construction material of the wetted
parts (tube) is matched to the non-corrosive performance of the liquid passing through the
instrument. Failure to observe this requirement can cause deterioration of the central tube
(the bellows) and loss in measurement accuracy, or even a failure if leaking occurs. For
advice on which meter in the range is appropriate, please contact Micro Motion.
The NI-SPAN-C material of the central tube is not rated for ‘sour’ service as defined in
NACE specification MR0103-2005. For advice in this application, please contact
Micro Motion.
Installing the meter in a by-pass configuration allows it to be removed for servicing or
calibration without affecting the main pipeline. Possible by-pass configurations are shown in
Figure 2-3.
Performance
Pipe stresses and
vibration
Gas bubblesThe presence of gas bubbles can seriously affect the meter performance and so the
Meter orientation• For low flow rates, for example 750 liters/hour (2.7 gal/min.), the meter should preferably
Flow rateA fast flow rate, for example 3000 liters/hour (11 gal/min.), will help to achieve good
Temperature stabilityThermally lag the meter and the inlet and slipstream/bypass-loop pipework to ensure good
Axial load should not exceed ½ tonne, so pipe-work should have a degree of flexibility.
Excessive pipe vibration should be avoided. Figure 2-2 for preferable mounting positions.
following points should be considered:
• The liquid must always be at a pressure substantially above its vapor pressure.
• All pipe-work couplings and joints must be airtight.
• No vortex should be present at the inlet to the meter.
• Cavitations, caused by pumping, should not generate bubbles from dissolved gases.
• If a pump is used it should ‘push’ rather than ‘pull’ the product through the meter.
be mounted vertically or at an incline, with the flow in an upwards direction.
• If the liquid contains solid particles, the direction of flow should be upwards unless the
particles are large enough not to be carried with the flow, in which case the direction of
flow should be reversed.
• The meter should be mounted with the electric cable running downwards thereby
minimizing the ingress of water should a cable gland become defective.
temperature equilibrium and have a self-cleaning action.
A low flow rate, for example 1000 liters/hour (3.7 gal/min.), is recommended if the product
contains particles which may cause erosion.
The meters exhibit a small flow dependent density reading. For flow rates up to 15000
liters/hour (55 gal/min) and assuming no consequent line pressure or product changes, the
maximum density offset will be less than 0.2 kg/m
This section considers in more detail the mounting of the meters and the design of the associated
pipework, including the calculation of pressure drop in the meter.
The preferred methods of supporting the meter are shown in Figure 2-1.
Installation & Configuration Manual9
Installation Procedure
1st
Meter supported
in two positions around
1” schedule 80 pipe.
Meter supported
in two positions around
outer case (4” diameter).
3rd
Meter supported
in two positions around
the immediate pipework.
2nd
Figure 2-1Preferred methods of mounting meters (support)
For continuously high flow rates, the mounting position can be selected to simplify the associated
pipework and help minimize the pressure and temperature losses (see Figure 2-2).
10Micro Motion 7835/45/47 Liquid Density Meters
Installation Procedure
1st
FLOW
90
o
Rupture plate blow out
Electric Cable
Direction of flow should
be reversed for slurries.
2nd
>60
o
Rupture plate blow out
Electric Cable
Direction of flow should
be reversed for slurries.
FLOW
Flow rate should be kept high
to prevent gas bubbles & sediment
from forming on the resonant tube.
3rd
Rupture plate blow out
Electric Cable
FLOW
Figure 2-2Preferred methods of mounting meter (angles)
(1) Indicates laminar flow (fluid density 1.0 g/cc)
υ = 2 cSu = 10 cS
2.6Calculation of pressure drop in the meter
The meter should be considered as a straight pipe of 23.6 mm (0.929”) internal diameter and 1.03 m
(40.551”) in length. The following formula has been proven to apply to the meter by measurements at
12000 liters/hour (44 gal/min).
In addition to the pressure drop caused by the liquid flow through the instrument, it will be necessary
to calculate the pressure drop in any associated sample pipework before concluding the system design
requirements.
Installation & Configuration Manual13
Installation Procedure
2.7Special considerations for hygienic applications (7847 meter only)
The 7847 meter is specially designed for use in hygienic applications. The following points should be
considered when planning an installation for a hygienic application.
Table 2-4.Considersations for hygienic applications
Meter orientation
Meter mounting
Steam cleaning
Electrical installation
Post installation
The 7847 shall be installed in the vertical plane to prevent the accumulation of product
residue in the convolutions of the bellows, causing contamination, especially during the
final rinse when cleaning. The installation shall include a means of draining the meter.
The method of mounting the meter shall comply with 3-A recommendations. Process seals
suitable for the media and complying with 3-A recommendations shall be used.
Where it is necessary to sterilize the meter using the steam cleaning process, ensure the
temperature and duration of cleaning does not exceed 250 °F (121 °C) for a period of
30 minutes. Exceeding this limit may permanently damage the meter’s amplifier circuit.
Cable glands and blanking plugs with a minimum rating of NEMA 4/IP66, shall be used to
ensure the environmental rating of the enclosure is not reduced.
Ensure the cable glands, blanking plugs, lid, and seal are in place and tightened to prevent
moisture and dust ingress.
2.87845/47 and 7845/47 Entrained Gas Meters with Remote Amplifier
For operating in product temperatures greater than 230°F (110 °C), it is necessary to remove the
amplifier unit from direct contact with the meter to a remote position. For this purpose, a flexible
PTFE conduit with over-braiding of stainless steel mesh is introduced between the amplifier housing
and the new amplifier housing. Three twisted pair cable looms are fed through the conduit to
complete the extended meter/amplifier connections.
This high temperature arrangement allows the amplifier unit to be mounted in a more temperate
environment and in no way impairs the operational accuracy of the meter. The maintenance and
calibration procedures remain relevant to the re-configured meter assembly.
Figure 2-4 shows the installation for the 78452 version, which is for Advanced Electronics only.
14Micro Motion 7835/45/47 Liquid Density Meters
Installation Procedure
Figure 2-4.7845 with Advanced electronics remote amplifier unit installation
After installation, the meter should be pressure tested to 1.5 times the maximum working pressure of
the system but not to a value exceeding the meter test figure shown on the meter label.
If the pressure test figure is exceeded, the meter may be irrevocably damaged.
16Micro Motion 7835/45/47 Liquid Density Meters
Installation Procedure
2.10Installation Drawings
Figure 2-5Installation for the 7835/45/47 with Standard electronics
This chapter describes the electrical installation of the 7835/45/47 liquid density meters with Standard
Electronics fitted. The units are identical, except where the 7845 and 7847 meters are to be used at
temperatures above 110
3.2MID (2004/22/EC) Requirements
To comply with the MID (2004/22/EC) directive:
•The digital communication (Modbus and HART) must be disconnected during operational use.
It can only be used for commissioning and maintenance.
•Unused cable ports must be sealed with suitably rated blanking plugs.
•After commissioning or maintenance of the meter, you must seal the enclosure cover to secure
legally relevant parameters from unauthorized modification.
See Section 3.2.1 for more information on securing the meter from unauthorized access to the
meter controls.
°C (230 °F) when the remote amplifier version is recommended.
To seal the meter from unauthorized access after commissioning or maintenance, Micro Motion has
provided additional holes on the electronics housing cover to attach a locking wire to the transmitter
cover. The securing component must bear the mark as laid down by the national inspection authority.
Figure 3-1 illustrates the suggested method for sealing the meter.
Note: When installing the meter in a MID measuring system, you must consider the method in which
the system will be verified to meet MID requirements. This method may impact the design of the
measurement system, and we recommend you involve the national inspection authority early in the
design process.
Installation & Configuration Manual21
Electrical Connections (Standard)
Stainless steel locking wire (supplied by MID inspector)
Bonding crimp (supplied by MID inspector)
Use hole in screw(s) to attach locking wire
Figure 3-1MID seal method
3.3Ground connections
The earthing pads on the mounting face of the amplifier unit MUST make good contact with the
meter case by the M3 cage nuts. The meter should be grounded via the pipework.
The external earth bonding point of the meter is located inside the maintaining amplifier housing.
The 0V power supply lead should be earthed at the supply end, or at the safety barriers if applicable.
3.4Use with Micro Motion signal converters
3.4.1System Connections (Hazardous Area only)
When the meter is used in a hazardous area, a safety barrier MUST be interposed between the meter
and the signal processing equipment. (See Chapter 3 for information on selecting a safety barrier.)
•For installation of the CSA certified unit in a hazardous area, refer to Appendix H.
•For installation of the ATEX certified unit in a hazardous area, refer to the appropriate safety
instructions booklet shipped with your unit (also available at www.micromotion.com).
3.4.2System connections (Safe Area only)
The density system connections are illustrated in Figure 3-2.
22Micro Motion 7835/45/47 Liquid Density Meters
Electrical Connections (Standard)
Figure 3-2Electric connection diagram to signal converters
•Power supply to density meter: 15.5 V to 33 V dc, 25 mA minimum
•Power supply to RTD: 5 mA maximum
The frequency at which the meter is operating can be detected by using a series resistor in the +VE
power line. The value of resistance to be used for a given supply voltage must not exceed the value
obtained from the LOAD NOMOGRAM (Figure 3-4). The electrical connections to be made are
shown in Figure 3-3.
Installation & Configuration Manual23
Electrical Connections (Standard)
7835/45/46/47 with
Standard Electronics
PRT
1
2
3
4
5
6
See Note
POWER +VE
SIGNAL +VE
POWER −VE
SIGNAL −VE
PRT SUPPLY +VE
PRT SIGNAL
PRT SUPPLY −VE
Figure 3.2- Electrical Connection Diagram
7835/45/46/47 with StandardElectronicsto Customer’s OwnEquipment
R
POS +
NEG -
SIG
Note: See Load Nomogram (Figure 3.3) to determine R value.
1μF
1μF
Meter
RTD
Figure 3.3 Load Resistance Nomogram
Supply Voltage (volts d.c.)
15
20
25
30
(33)
35
Maximum Supply Voltage ‘E’
0
100
200
300
400500600
700
Loa
d R
e
s
is
t
a
n
c
e
L
i
n
e
Supply Voltage
Maximum Load Resistance
for Given Supply Voltage
Maximum Load Resistance (ohms)
Note: It is recommended that the actual load resistor should
be 50 ohms less than that given by the Nomogram.
Figure 3-3Electrical connection diagram to customer’s own equipment
24Micro Motion 7835/45/47 Liquid Density Meters
Figure 3-4Load resistance
Electrical Connections (Standard)
3.5.2System Connections (Hazardous Area only)
Installation of the meter, the safety barriers, and the customer's signal processing equipment is shown
in Figure 4.4.
•For installation of the CSA certified unit in a hazardous area, refer to Appendix H.
•For installation of the ATEX certified unit in a hazardous area, refer to the appropriate safety
instructions booklet shipped with your unit (also available at www.micromotion.com).
Note: Whenever there is disagreement between connection details on the figures dealing with
Hazardous Areas and Certified System Diagrams, the Certified System Diagrams are the authoritative
documents.
Figure 3-5Electrical Connection Diagram with Standard Electronics to Customer’s Own Equipment
After installation, the following procedure will indicate to a high degree of confidence that the meter
is operating correctly.
Measure the current consumption and the supply voltage at the meter amplifier. This should be within
the limits:
•15.5 to 33Vdc, 17mA ±1mA (Safe Areas)
•15.5 to 21.5Vdc, 17mA ±1mA (Hazardous Areas)
With the meter empty, clean and dry, measure the periodic time of the output signal and check that it
is as specified on the meter calibration certificate (air check), to within the limits given in the table
below.
Installation & Configuration Manual25
Electrical Connections (Standard)
Meter typeAir check limit at 20°CAdded temperature effect
7835±60 ns±10 ns/°C
7845/47±60 ns–300 ns/°C
26Micro Motion 7835/45/47 Liquid Density Meters
Chapter 4
Base board
Option board
Remo te Display
7965 DISPLAY
Electrical Connections (Advanced)
4.1General
This chapter describes the electrical installation of the 7835 and the 7845/7847 Liquid Density
Meters when fitted with the Advanced electronics option.
The first sections of this chapter address the installation and configuration of the Advanced
Baseboard, and the later sections concern the Advanced option boards. Further details about the
installation of the Remote Display are given in the Micro Motion 7835/45/47 liquid density meter
Configuration and Use Manual.
4.2MID (2004/22/EC) Requirements (7835/7845 Only)
To comply with the MID (2004/22/EC) directive:
•The digital communication (Modbus and HART) must be disconnected during operational use.
It can only be used for commissioning and maintenance.
•Unused cable ports must be sealed with suitably rated blanking plugs.
•After commissioning or maintenance of the meter, you must seal the enclosure cover to secure
legally relevant parameters from unauthorized modification.
See Section 4.2.1 for more information on securing the meter from unauthorized access to the
meter controls.
Installation & Configuration Manual27
Electrical Connections (Advanced)
Stainless steel locking wire (supplied by MID inspector)
Bonding crimp (supplied by MID inspector)
Use hole in screw(s) to attach locking wire
4.2.1Securing the meter for MID
To seal the meter from unauthorized access after commissioning or maintenance, Micro Motion has
provided additional holes on the electronics housing cover to attach a locking wire to the transmitter
cover. The securing component must bear the mark as laid down by the national inspection authority.
Figure 4-1 illustrates the suggested method for sealing the meter.
Note: When installing the meter in a MID measuring system, you must consider the method in which
the system will be verified to meet MID requirements. This method may impact the design of the
measurement system, and we recommend you involve the national inspection authority early in the
design process.
Figure 4-1MID seal method
28Micro Motion 7835/45/47 Liquid Density Meters
Electrical Connections (Advanced)
4.3Planning an Electrical Installation
When planning the electrical installation of an Advanced unit, it is important to consider the points
given below.
Safety
Power supply
Ground
connections
Cable parameters
EMC
• Electrical installation in hazardous areas requires strict adherence to local codes of practice.
• For installation of the CSA certified unit in a hazardous area, refer to Appendix H.
• For installation of the ATEX certified unit in a hazardous area, refer to the appropriate safety
instructions booklet (available at www.micromotion.com).
• The Advanced electronics operate from a nominal 24V supply, but will operate from any
supply in the range 9.5V to 28V, measured at the supply terminals on the baseboard.
• The output circuits on the baseboard are all loop-powered and are isolated from the main
circuit. If required, the main circuit and the output circuits can be powered from a common
power supply.
• When selecting a suitable power supply voltage, you must take into account voltage drops
caused by the connecting cable (see below) and in hazardous areas, across zener barriers or
galvanic isolators.
• The earthing pads on the baseboard must make good contact with the meter case via the M3
bolts.
•If a HART® option board is used, the indicated earthing point must make good contact with
the baseboard earthing points.
• The 0V power supply lead should be earthed at the supply end, or at the safety barriers if
applicable.
• Where long cable lengths are required the cable resistance may be significant. When
operating from a 24V supply in safe areas the following limits apply:
- Maximum line resistance (Ω)
- Power supply260
- Remote Display60
- Outputs500
For further details relating to the maximum line resistance, please refer to Appendix D.
• When calculating the maximum cable lengths please note that the current loop is 2 times the
cable length, and so the cable resistance is given by: 2 x (dc resistance per unit length) x
(cable length).
• Typical cables would comply with BS5308 Type 1 or 2.
• To meet the EC Directive for EMC (Electromagnetic Compatibility), it is recommended that the
meter be connected using a suitable instrumentation cable. The instrumentation cable should
have individual screen(s), foil or braid over each twisted pair and an overall screen to cover all
cores. Where permissible, the overall screen should be connected to earth at both ends
(360° bonded at both ends). The inner individual screen(s) should be connected at only one end, the controller (e.g. signal converter) end.
• Note that for intrinsic safety, termination of the inner individual screen(s) to earth in the
hazardous area is NOT generally permitted.
• Metal cable glands should be used where the cables enter the meter amplifier box. Unused
cable ports should be fitted with metal blanking plugs.
• When the 78452 (Advanced Remote Amplifier) is used, the ferrite ring, which is supplied,
must be fitted around the connecting cable. For installation in hazardous areas, refer to the
certified system drawings in Appendix H, and the safety instruction booklet that came with the
meter (also available at www.micromotion.com) .
All connections to the baseboard are made to terminals 1 to 12 of the terminal block PL2 as shown in
Figure 4-2.
Installation & Configuration Manual29
Electrical Connections (Advanced)
Ana lo g 1
Anal og 2
Alarm
22mA
2mA
Alar m
22m A
2mA
OPTIONAL
BOAR D
PL1
PL 2
P/U 1
P/U 2
0V
PRT -
SIG SIG +
PRT +
DRIVE
PUL SE 1
+
-
ANA LO G 1
+
-
SUP P LY +
0V
+
A
B
DISPLAY
Eart hing Point
Ea r t h i ng Po i n t
ANA LO G 2
+
-
12 34 5
678910
11
12
12345
6
789
Jumper Links
Foam support strip
Figure 4-2Baseboard Layout
30Micro Motion 7835/45/47 Liquid Density Meters
Note: There is no reason to remove the baseboard under normal circumstances. If it is removed,
however, care should be taken, when replacing the board, to push it firmly against the foam strip, as
this helps to prevent the board from excessive flexing.
Figure 4-3 shows a schematic representation of the electrical connection diagram for the Advanced
Baseboard and Remote Display in safe areas. In this example, the pulse output is shown driving an
alarm device such as a relay.
4.4.1Electrical Installation with Signal Converter
Figure 4-4 shows a typical electrical connection diagram for use in safe areas using signal converters.
Electrical Connections (Advanced)
1
1000Ω
+ Densit y Pwr
2
+ Density Input
- Density Pw r
- Density Input
+ Analog Input
SEE FLOW
COMPUTER/
SIGNAL
CONVERTER
HANDBOOK
FOR
CONNECTION
DETAILS
3
4
(5)
(6)
+ Density Pwr
- Density Pwr
- Analog Input
+Density Pwr
- Density Pwr
Power Supp ly
Analog 1 (Analog 2)
Fre q/A larm
+VE
-VE
+VE
-V E
+VE
-VE
7
8
7835/45/47 with advanced
electronics
7950/51
Figure 4-4Electrical connection diagram when using a signal converter (Safe area)
4.5Electrical installation in hazardous
When used in hazardous areas, safety barriers MUST be interposed between the meter and the signal
processing equipment. Some of the safety barriers are unsuitable for certain installations, as discussed
below.
4.5.1Safety Barrier and Galvanic Isolator Selection
Power Supply (PL2 terminals 7 and 8)
As a general rule, the IIB safety barrier should be used where possible as this allows the maximum
power to the meter, facilitating a wide range of installations and system configurations.
Where the installation requires the IIC safety barrier, it is important to check that sufficient power is
available to power the meter and all of the options. The table below summarizes the maximum line
resistances allowable for the main system configurations assuming a 24V supply and a minimum of
9.5V available at the supply terminals on the baseboard.
areas
•For installation of the CSA certified unit in a hazardous area, refer to Appendix H.
•For installation of the ATEX certified unit in a hazardous area, refer to the appropriate
safety instructions booklet shipped with the unit (also available at www.micromotion.com).
32Micro Motion 7835/45/47 Liquid Density Meters
Electrical Connections (Advanced)
()
1150
95.02.828
+
×−
Table 4-1Maximum line resistances for main system configurations
Advanced System combination
Baseboard340
Baseboard + remote display260
Baseboard + locally powered remote display280
®
Baseboard + HART
Baseboard + HART
board270
®
+ display250
Zener Safety Barriers
Power to the main circuit may be obtained through a simple 164Ω barrier or two 300Ω 28V barriers
in parallel for IIB applications, or through a single 234Ω 28V barrier in IIC applications.
The main characteristics of the safety barriers are given in Table 4-2. Using this information and the
information given in Table 4-1, the most suitable barriers for a particular application can be
ascertained.
Example TypeGroupSafety Description (Ω)Max. resistance (Ω)
MTL 729PIIB164184
MTL 728PIIC234253
P&F Z728HIIC240250
Ω
Note: The power supply input is protected internally by an 8.2V±5% clamp diode and a 1
resistor.
This limits the maximum current that can flow into the device:
I
=
max
Maximum voltage from barrier
Minimum resistance of barrier/input combination
- Minimum voltage at input
For example, if two 28V, 300Ω barriers are used in parallel, the effective resistance is 150Ω.
The maximum current is:
I
== 134mA
max
…and not 185mA, as might be expected if the input protection diode was not present.
Galvanic Isolators
Galvanic isolators are suitable for powering the main board in IIB applications, but are NOT suitable
for powering the main board in IIC applications. Also, IIC isolators are not suitable for use when
Modbus communications are required.
The main characteristics of the galvanic isolators are given here. Using this information and the
information given in the table above, the most suitable galvanic isolators for a particular application
can be ascertained.
Installation & Configuration Manual33
Electrical Connections (Advanced)
Table 4-3Galvanic isolator characteristics
Example TypeGroupMax output impedance (Ω)
MTL 3022IIB165
P&F KFD2-SD-Ex1.36IIB160
P&F KFD2-SL-Ex1.36IIB160
MTL 5022IIB143
Analog (4 to 20mA) outputs including HART
®
(PL2 terminals 3 & 4, 5 & 6 and HART PL3 3 & 4)
Any of the zener safety barriers listed on the system certificates are suitable for operation with the
Advanced Density analog outputs. Some galvanic isolators may not be capable of driving the 2mA
and 22mA out-of-range alarm states available on the Advanced Density system; for details please
check with the barrier supplier.
Pulse output
(PL2 terminals 1 & 2)
The pulse output can be configured to output either a status / alarm signal or the resonant frequency of
the density meter; the latter requires the safety barrier to have a bandwidth of at least 1 kHz.
The table below indicates which galvanic isolators may be used for frequency-configured pulse
output; zener barriers may be used for either configuration.
Table 4-4Pulse output configurations
Example TypeOutput descriptionPulse output configuration
MTL3011RelayStatus / alarm
MTL3012Solid state (dc to 2kHz)All
MTL4013Solid state (dc to 5kHz)All
MTL4014RelayStatus / alarm
MTL5011RelayStatus / alarm
MTL5016RelayStatus / alarm
MTL5017RelayStatus / alarm
4.5.2Electrical connections in a hazardous area
Table 4-5Electrical connections in a hazardous area
Drawing ReferenceDescription
Advanced Electronics, HART, MODBUS, and Zener Barrier
(Gas Groups A, B, C and D)
Advanced Electronics, HART, MODBUS, Remote Display, Zener Barrier
(Gas Groups A, B, C, and D)
34Micro Motion 7835/45/47 Liquid Density Meters
Electrical Connections (Advanced)
Table 4-5Electrical connections in a hazardous area
Figure 4-5Electrical connection diagram when using Advanced Electronics, HART, MODBUS, and
Zener Barrier in a Hazardous Area (Gas Groups A, B, C and D)
Advanced Electronics with HART Multi-drop, and Zener Barrier
(Gas Groups C and D)
Advanced Electronics with HART Multi-drop, Remote Display, Zener Barrier
(Gas Groups C and D)
Electronics with HART Multi-drop, Remote Display, Galvanic Isolator
(Gas Groups C and D)
The baseboard is supplied with one of the following two software versions:
•General software version - normally used in the food and process industries.
•Fiscal software version - normally used in Crude oil or refined petroleum applications.
The only difference between the two software versions is the available calculations.
Installation & Configuration Manual39
Electrical Connections (Advanced)
Irrespective of which software version is running, when the unit is received from the factory, it is
pre-configured to output the following signals:
OutputOutput Parameter
Analog 1 (4-20mA)Line density (700 to 1000 kg/m³)
Analog 2 (4-20mA)Temperature (0 to 100°C)
Pulse outputFrequency (default)
For many applications, the factory default configuration described above will be quite acceptable.
However, if any of the additional calculated parameters or different output ranges are required, then a
simple Baseboard re-configuration can be performed using a Remote Display, or PC.
Alarm: Untriggered – High (default)
4.7Baseboard plus HART
The HART
®
option board connects with the Baseboard using the 40-way connector provided and
supported by two plastic posts. A tag is provided for connecting to the chassis earth point on the
Baseboard.
®
The HART
option board is a loop-powered 4-20mA output which can support HART®
communications or can be used to provide a third analog output. The unit is always designated as a
®
HART
For HART
slave unit, i.e. it only communicates when it receives a message asking it to do so.
®
communications, the option board output is regarded as the primary output, and the
baseboard outputs as the secondary and tertiary outputs.
4.7.1Electrical Installation for HART
For safe area installations, electrical connections to the HART
Figure 4-10.
•For installation of the CSA certified unit in a hazardous area, refer to Appendix H.
•For installation of the ATEX certified unit in a hazardous area, refer to the appropriate safety
instructions booklet shipped with your unit (also available at www.micromotion.com).
In safe areas, up to 15 HART
should be attached in parallel across the two points indicated as X and Y in Figure 4-10. If more than
one HART
®
Advanced unit is installed on a single HART® communication link, each unit must be
given a unique HART
non-zero value, the output current is automatically set at 4 mA.
®
Option Board
®
Communications
®
option board are shown in
®
slave units may be installed on one HART® communication link. They
®
slave address in the range 1 to 15. Whenever the HART® address is set to a
After installation, the following procedure will indicate to a high degree of confidence that the meter
and Advanced system is operating correctly.
1. Measure the supply voltage at the meter amplifier (PL2, pins 7 and 8). This voltage should be
within the limits of 9.5 to 24Vdc in safe areas and 9.5 to 20Vdc in hazardous areas. In safe
areas, the current to the baseboard should not exceed 80mA.
2. With the meter empty, clean and dry, measure the periodic time of the output signal and check
that it is as specified on the meter calibration certificate (air check), to within the limits given
in the table below.
Meter typeAir check limit at 20 °CAdded temperature effect
7835±60 ns±10 ns / °C
7845/47±60 ns– 300 ns / °C
Installation & Configuration Manual41
Electrical Connections (Advanced)
42Micro Motion 7835/45/47 Liquid Density Meters
Chapter 5
Electrical Connections (Entrained Gas Option)
5.1General
This chapter concerns the operation of 7845/47 liquid density meters when fitted with the Standard
Entrained Gas amplifier.
7845/47 Entrained Gas Liquid Density Meters with Standard Electronics are NOT intrinsically
safe.
5.2Ground Connections
The earthing pads on the mounting face of the amplifier unit MUST make good contact with the
meter case by the M3 cage nuts. The external earth bonding point of the meter is located inside the
maintaining amplifier housing. The meter should be grounded via the pipework.
The 0V power supply lead should be earthed at the supply end, or at the safety barriers if applicable.
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
5.3Use with Signal Converters
The 7845/47 E.G. Liquid Density Meter (with Standard Electronics)/signal converter system can only
be operated in SAFE AREAS.
5.3.1System Connections
The density system connections are illustrated in Figure 5-1 below.
Installation & Configuration Manual43
Electrical Connections (Entrained Gas Option)
Figure 5-1Electrical connection diagram (7945/47 E.G. to Signal Converters)
5.4Use with Customer’s Own Equipment
5.4.1System Connections for Safe Areas
7845/47 Entrained Gas Liquid Density Meters with Standard Electronics are NOT intrinsically
safe.
Power supply to Density Meter:15.5 V to 33 V d.c., 25 mA min.
Power supply to RTD:5 mA max.
The frequency at which the meter is operating can be detected by using a series resistor in the +VE
power line. The value of resistance to be used for a given supply voltage must not exceed the value
obtained from the LOAD NOMOGRAM (Figure 5-2). The electrical connections to be made are
shown in Figure 5-3.
44Micro Motion 7835/45/47 Liquid Density Meters
Electrical Connections (Entrained Gas Option)
7845/47 E.G.
PRT
1
2
3
4
5
6
See Note
POWER +VE
SIGNAL +VE
POWER -VE
SIGNAL -VE
PRT SUPPLY +VE
PRT SIGNAL
PRT SUPPLY -VE
R
POS +
NEG -
SIG
Note: See Load Nomogram (Figure 5.2) to determine R value.
1μF
1μF
Figure 5-2Load Resistance Nomogram
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
Figure 5-3Electrical Connection Diagram – 7845/47 E.G. to Customer’s Own Equipment (SAFE
AREAS)
Installation & Configuration Manual45
Electrical Connections (Entrained Gas Option)
5.5Post-Installation Checks
After installation, the following procedure will indicate that, to a high degree of confidence, the meter
is operating correctly.
1. Measure the current consumption and the supply voltage at the meter amplifier. This should be
within the limits:
•15.5V to 33V d.c.
•75mA
±10mA
With the meter empty, clean and dry, measure the periodic time of the output signal and check that it
is as specified on the meter calibration certificate (air check), to within acceptable limits (e.g. 500ns),
after making allowances for different ambient conditions.
46Micro Motion 7835/45/47 Liquid Density Meters
Chapter 6
Calibration and Performance
6.1General
The 7835/45/47 liquid density meters (including Entrained Gas versions) are calibrated at the factory,
and are supplied with their own test and calibration certificates.
The calibration certificate specifies various calibration constants that allow the user to convert the
output periodic time signal from the meter into a density value. (See Appendix E for specimen
calibration certificates.)
6.1.1For Standard Electronics Units
For units with Standard electronics, the calibration constants will need to be programmed into a signal
processing instrument such as a signal converter. Density calculations are performed on the signal
processing instrument.
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
6.1.2For Advanced Electronics Units
For units with Advanced electronics, the calibration constants are pre-programmed into the
electronics and normally require no further consideration. The calculations in this chapter are
performed by the electronics on the meter.
Important Information
If you have obtained a replacement calibration certificate for an Advanced Electronics unit, the set of
pressure coefficient constants K20A, K20B, K21A and K21B that fall within your operating pressure
range can be programmed into the Advanced Electronics using ADView or ProLink II software
(downloadable from web sites listed on the back page).
(If your operating pressure range falls between two of the sets of operating pressure ranges on the new
certificate, contact the factory for a new calibration certificate.)
The Advanced Electronics keeps a write-protected copy and a working copy of all coefficients. The
integrity of the working coefficients is safe, and so for simplicity it is recommended that the working
coefficients only be changed. This is achieved by writing to registers 131 and 132. (See Chapter 8 for
a guide to using ADView or ProLink II.)
Alternatively, a new FRAM memory chip can be issued which holds the calibration coefficients.
However, in replacing the FRAM device some user configured data may be lost (e.g. upper and lower
limits on the analog outputs, matrix referral points, special function, user defined line pressure etc.).
Installation & Configuration Manual47
Calibration and Performance
The general density equation is: D = KK K01 2
2
++
ττ
Where :
D = The un cor rected densi ty (kg/m
3
) of liqui d
τ
= Periodic time (μs) of v ibration
= 1/f where ‘f’’ is the frequency of vibration
2&1,0KKK = Constants fr om the Calibration Certificate
The e qua t ion used for thi s cor recti on is :
D
t
=
()
[]
()
DKtKt11820 1920+−+−
Where:
D
t
= Te mper ature co rrec ted d ensi ty (kg/ m3)
D = Densi ty ca lcul at ed using eq uat ion 1
t = Temperature (degrees C)
Kand K1819 = Constants from the Calibration Certificate
6.2Interpretation of calibration certificate
6.2.1General density equation
The basic meter constants, K0, K1, and K2 are computed from the factory calibration on three fluids.
Using these constants and the general density equation, the density of the liquid within the meter can
be calculated.
It is stated on the calibration certificate that the basic constants are determined from a calibration at a
temperature of 20 °C (68 °F) and at a pressure of 1 bar (14.5 psi). If the operating conditions of the
meter differ from that of the calibration conditions, a correction to the density calculated using the
general equation is required.
6.2.2Temperature correction
If the meter operates at temperatures other than 20 °C (68 °F), a correction to the density calculated
using equation (1) must be made using the temperature coefficient data given on the calibration
certificate.
6.2.3Pressure correction
The meter design has a unique facility to reduce the influence of the line pressure on the density
measurement but there is a residual effect for which correction may be required. This residual
pressure effect before a pressure correction is illustrated for the 7835/45/47 meter in the following
figures.
48Micro Motion 7835/45/47 Liquid Density Meters
Calibration and Performance
-10
0
10
20
30
40
50
020406080100120140
Pressure (Bar Absolute)
Density Offset (kg/m
3
)
Uncorrected pressure effects
on the meter fall within these
bands
0
5
10
15
20
25
30
35
40
45
05101520
Pressure (Bar Absolute)
Density Offset (kg/m
3
)
Figure 6-1Pressure effect on the7835/45/47 meter before pressure correction (at 20°C)
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
Figure 6-2Pressure effect on 7845/47 Entrained Gas before pressure correction (at 20°C)
Installation & Configuration Manual49
Calibration and Performance
The equation used to apply pressure correction is:
D
P
=
()
[]
()
DKPKP
t
120 1 21 1+−+−
Where:
D
P
= Temper ature an d pressure cor rec te d d ensity (kg/m3)
D
t
= Tem per at ure correct ed d ensity (kg/ m3)
P = Pressure in bar absolute
K 20 =
()
KAKBP20201+−
K 21=
()
KAKBP21211+−
During the calibration of the meter, which is normally performed at a pressure of 1 bar (14.5 psi), the
pressure influence is also measured. This data is also shown on the calibration certificate (see
Appendix E).
Note: K20A, K20B, K21A, and K21B are the pressure coefficient constants on the calibration
certificate.
Note: The pressure correction is further enhanced on units that operate above 41 bar (595 psi) by
having sets of pressure coefficient constants covering subsets of the full operating pressure range.
Only one set of pressure coefficient constants is selected from your calibration certificate according to
your operating pressure range. If your operating pressure range falls within the range of two sets of
pressure coefficient constants, contact Micro Motion for a new calibration certificate. See Appendix E
for an example calibration certificate.
Note: If it is required to apply temperature and pressure corrections, the temperature correction is
applied first.
Figure 6-3 shows the typical residual error curves after pressure corrections for 7835 (100Bar) units
using three sets of pressure coefficient constants. Each set covers a sub-set of the 100Bar range. The
uncertainty specification for a 7835 is indicated by the upper and lower limit lines. The uncertainty
for the 7835 pressure coefficients is
uncertainty of
Note: Only one set of pressure coefficient constants is selected from your calibration certificate
according to your operating pressure range. For specimen calibration certificates, see Appendix E.
±0.15 kg/m
3
.
±0.003 kg/m
3
. This is in addition to the instrument calibration
Figure 6-3Residual pressure effect after pressure correction – 7835 (100Bar) units (at 20°C)
Figure 6-4 shows the typical residual error curves after pressure correction for 7835 150Bar units
using four sets of pressure coefficient constants. Each set covers a sub-set of the 150Bar range. The
uncertainty specification for a 7835 is indicated by the upper and lower limit lines. The uncertainty
for the 7835 pressure coefficients is ±0.003 kg/m
uncertainty of +/-0.15kg/m
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
3
3
.
. This is in addition to the instrument calibration
Note: Only one set of pressure coefficient constants is selected from your calibration certificate
according to your operating pressure range. For specimen calibration certificates, see Appendix E.
Figure 6-4Residual pressure effect after pressure correction – 7835 (150Bar) units (at 20°C)
Figure 6-5 shows the typical residual error curves after pressure correction for 7845K (100Bar) units
using three sets of pressure coefficient constants. Each set covers a sub-set of the 100Bar range. The
7845 uncertainty specification is indicated by the upper and lower limit lines. The uncertainty for the
7845K pressure coefficients is ±0.006kg/m
uncertainty of +/-0.035kg/m3.
Installation & Configuration Manual51
3
. This is in addition to our stated instrument calibration
Calibration and Performance
-1.50
-1.00
-0.50
0.00
0.50
1.00
1.50
0 102030405060708090100110
Pressure (BarG)
Magnitude (kg/m3)
New K20 K21
Upp er limit
Low er limit
-1.00
-0.80
-0.60
-0.40
-0.20
0.00
0.20
0.40
0.60
0.80
1.00
0 102030405060
Pressure (BarG)
Magnitude (kg/m3)
New K20 K21
Upper limit
Low e r limit
Note: Only one set of pressure coefficient constants is selected from your calibration certificate
according to your operating pressure range. For specimen calibration certificates, see Appendix E.
Figure 6-5Residual pressure effect after pressure correction – 7845K (100Bar) units (at 20°C)
Figure 6-6 shows the typical residual error curves after pressure correction for 7845 (50Bar) units
using one set of pressure coefficient constants. The set covers the full 50 Bar range. The 7845
uncertainty specification is indicated by the upper and lower limit lines. The uncertainty for the
7845K pressure coefficients is ±0.006kg/m
uncertainty of +/-0.035kg/m3.
Note: Only one set of pressure coefficient constants is selected from your calibration certificate
according to your operating pressure range. For specimen calibration certificates, see Appendix E.
Figure 6-6Residual pressure effect after pressure correction – 7845 (50Bar) units (at 20°C)
52Micro Motion 7835/45/47 Liquid Density Meters
3
. This is in addition to our stated instrument calibration
Calibration and Performance
Alternatively, the following equations may be used:
D
VOS
=
Dp 1
14 06
1400
11
22
+
+
×−
⎛
⎝
⎜
⎞
⎠
⎟
⎡
⎣
⎢
⎤
⎦
⎥
. E
DVV
PCA
Where:
D
VOS
=
(kg/m3)
D
P
= Temperature and pressure corrected density (kg/m3)
V
C
= Calibration VOS (m/s)
V
A
= Liquid VOS (m/s)
V
C
can be obtained direct from Figure 5.2 or may b e calculated as follows:
V
C
= 1001455+ .D
P
for a DP of 300kg/m3 to 1100k g/m3
V
C
= 269009− . D
P
for a DP of 1100kg/m3 to 1600kg/m3
Optimization for pressure-temperature coupling effect (7835 meters only)
For the calibration of 7835 meters, a new generic constant is being applied to calculate the K21A
pressure coefficient that is valid for use over a limited operating temperature and pressure range. The
revised K21A pressure coefficient is selected from a table in a new format calibration certificate and
is unique to the 7835 meter. The application of this K21A coefficient does not change the density
calibration coefficient format or the density calibration equations previously used in the flow
computer software.
Note: This constant can only be applied to 7835 meters that have been calibrated at the factory
beginning in January 2011. Additionally, it is not possible to recalculate a revised K21A for units that
have been recertified at external calibration facilities.
The new constant is being applied as an intermediary measure to meet the requirements of the United
Kingdom Department of Energy and Climate Change (DECC) directive regarding the calibration of
liquid density meters. The DECC directive recommended that by July 2011 all density meters be
calibrated at the anticipated operating conditions (such as simultaneously at temperature and
pressure). Micro Motion is in process of redesigning the calibration stands so that they can operate at
a combined elevated temperature and pressure. These stands are planned to be completed and
operational by July 2011.
For an example of the calibration certificate that includes the new K21A pressure coefficients, see
Appendix E.
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
6.2.4Velocity of sound correction
The Velocity of Sound (VOS) in the process liquid may have an effect on the accuracy of the
indicated density. The calibration of the 7835 meter has been optimized to a density/VOS relationship
as indicated in Figure 6-7. If the VOS of the process fluid deviates substantially from the relationship
in Figure 6-7, it may be desirable to apply a correction. This may be achieved by the simple
introduction of a calibration offset using the data in Figure 6-7. Adjustment of the value
basic equation will achieve this.
Vel ocit y of s ou nd and tem perat ure correct ed d ensity
K
0
in the
Installation & Configuration Manual53
Calibration and Performance
0
200
400
600
800
1000
1200
1400
1600
1800
2000
2200
2400
2600
2
00
4
00
6
00
800
1
000
120
0
1400
160
0
Indicated Density (kg/m3)
Velocity of Sound (m/s)
+2kg/m3
+2kg/m3
Nominal
-2kg/m3
Values shown are the required corrections
True density = Indicated density + Corrections
Figure 6-7Optimized velocity of sound relationship for the 7835/45/47 meter
6.3Calibration
6.3.1Factory calibration
The 7835/45/47 liquid density meters are calibrated prior to leaving the factory against Transfer
Standard instruments, traceable to National Standards. Three fluids are used in the calibration –
ambient air whose density is derived from look-up tables, hydrocarbon oil of about 815 kg/m
3
and a high-density fluid in the range 1400 to 1500 kg/m
density. Several of the instruments-under-test
are connected in parallel between two Transfer Standard Instruments on the Micro Motion special
flow rig. During a calibration, and as the liquid flows through the instruments, readings are only taken
when the indicated densities on the two Transfer Standard Instruments agree. In this way, a high
integrity of calibration is achieved.
Measurements are also made under conditions of changing temperature and pressure to establish the
magnitude of these effects on the instrument. From all this data, a calibration certificate is generated
for each instrument.
Samples of the instruments are further tested by the Micro Motion Quality Assurance Department to
54Micro Motion 7835/45/47 Liquid Density Meters
verify the calibration.
6.3.2Calibration of transfer standards
The Transfer Standard instruments used in the calibration are selected instruments that are calibrated
and certified by the ISO/IEC17025-certified calibration laboratory.
Transfer Standard calibration uses a number of ‘density certified’ liquids. The densities of these
certified liquids are obtained using the Primary measurement system, whereby glass sinkers of
defined volumes are weighed in samples of the liquids.
3
density
Calibration and Performance
Calibration is performed by pumping each certified liquid through the Transfer Standard in a closely
controlled manner and recording the output signal in each case. A calibration certificate is issued for
each Transfer Standard.
Calibrations are repeated, typically every six months, producing a well-documented density standard.
6.3.3Instrument calibration certificate
Each instrument is issued with its own calibration certificate (see Appendix E for samples),
containing four important pieces of data:
•The instrument serial number.
•The output signal/density relationship. This is based on three calibration points – air, medium
density and high-density fluids. The air and high density fluid points are offset to achieve the
product velocity of sound/density profile described earlier. However, the signal value at Air
Density is also given for check purposes.
•Temperature coefficient data, describing the correction which should be applied to achieve the
best accuracy if the instrument is operating at product temperatures other than 20 °C (68 °F).
•Pressure coefficient data, describing the correction that should be applied to achieve the best
accuracy if the instrument is operating at elevated pressures.
A second page of the calibration certificate is retained by Micro Motion and contains all the
calibration measurements.
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
6.3.4Pressure test
A hydrostatic pressure test is carried out to a pressure value specified on the instrument label and on
the instrument calibration certificate. This test loads the instrument structure to a pressure that
exceeds the maximum permitted operating pressure of the instrument.
Note: During manufacture, the welded structure is pressure tested to conform to the requirements of
EN50018:1997. The outer case is able to withstand 100 bar of internal pressure in the event of
tube/bellows failure.
6.3.5Insulation test
To comply with Intrinsic Safety requirements, a 500 Vac insulation test is carried out between the
electrical terminals and the instrument case.
6.3.6Calibration check methods
There are two methods employed in calibration checks:
•Air checkpoint, which is simple and convenient and highlights long term drift, corrosion and
deposition.
•Liquid calibration verification comprising two choices:
-Drawing off a sample of the liquid being measured and obtaining its density, using a
hydrometer (for stable liquids) or pyknometer (for unstable liquids).
-Using a second density meter.
Ambient air check
1. Isolate, drain and if necessary, disconnect the meter from the pipeline.
Installation & Configuration Manual55
Calibration and Performance
2. Clean and dry the wetted parts of the meter and leave them open to the ambient air.
3. Apply power to the instrument and check that the time period of the output signal agrees with
the 'Air Check' figure shown in the calibration certificate, to within acceptable limits.
Some variation between the two figures is to be expected due to changes in ambient air
conditions. The density indication if using the K0, K1 and K2 factors will be about –0.9 kg/m
because the basic density equation has been optimized for best performance over the normal
operating density range.
This test will indicate whether there has been a calibration offsets due to corrosion, deposition
or long term drift. When this test is applied to the the meters, their temperature coefficient has
a significant effect and must be considered. The following table shows the temperature
correction for the 7835 and 7845 Standard and Entrained Gas versions.
3
Temperature
Meter
7835±10 ns/°C±60 ns
7845–300 ns/°C±60 ns
7845 Entrained
Gas
correction
–700 ns/°C±500 ns
Air check limit
at 20 °C
4. Reconnect the meter to the pipeline if serviceable, or remove it for further servicing.
Liquid density check – sample method
If it is necessary to verify the calibration using liquid at operating conditions, then the following
sample methods are recommended:
For Stable Liquids:
1. Draw off a sample of the liquid into a suitable container, at the same time noting the indicated
density, temperature and pressure of the liquid.
2. Measure the density of the sample under defined laboratory conditions, using a hydrometer or
other suitable instrument.
3. Refer the density measurement under laboratory conditions to that under the line operating
conditions of temperature and pressure.
4. Compare the referred density figure with that indicated by the density meter.
Note: It is essential that a good understanding of the physical properties (temperature coefficient,
etc.) of the liquid is acquired when using this method.
For Unstable Liquids:
1. Couple a pressure pyknometer and its associated pipework to the pipeline so that a sample of
the liquid flows through it.
2. When equilibrium conditions are reached, the meter density reading is noted as the
pyknometer is isolated from the sample flow.
3. Remove the pyknometer for weighing to establish the product density.
4. Compare the pyknometer registered density with that obtained from the meter.
Sampling Techniques
Sampling should comply with the international sampling standards (ISO 3171, ASTM D 4177, API
8.2 and IP 6.2).
56Micro Motion 7835/45/47 Liquid Density Meters
Calibration and Performance
For further details of these procedures, reference should be made to:
Institute of Petroleum:Petroleum Measurement Manual
Institute of Petroleum:Petroleum Measurement Manual
American Petroleum Institute:Manual of Petroleum Measurement Standards
Liquid density check – second density meter
It is often the practice, especially in fiscal metering applications, to use two or more density meters in
a continuous measurement mode as a means of improving the integrity of the measurement system.
Any unacceptable discrepancies between the measurements can immediately raise the necessary
alarm signals.
1. Connect the second density meter to the pipeline adjacent to meter being checked so that it
receives the same sample of fluid under the same conditions of temperature and pressure as the
meter under test.
Part VII Section 1 – Method IP 160 (Hydrometer Method)
(BS2000–160, ISO3675, ASTM 1298)
Part VII Section 2 – Continuous Density Measurement
Chapter 14 – Natural Gas Fluids – Section 6:
Installing and proving density meters used to measure
hydrocarbon liquid with densities between 0.3 and 0.7 g/cc at
15.56 °C (60 °F) and saturation vapor pressure, 1991.
2. Connect the second meter to its readout equipment, switch on and allow both systems to reach
equilibrium conditions.
3. Compare the two readings, making any necessary corrections.
This method of automatic checking has proved to be a very successful technique and where there is a
facility for two instruments, the practice of exchanging one for a newly calibrated instrument is
proving successful. This is sometimes referred to as the "Substitution Method".
It is very important when using one instrument to verify the performance of a second and similar
instrument, to ensure there are no unaccounted for systematic errors which would are not highlighted.
6.4Performance
Micro Motion meters are generally calibrated using specified fluids at 20 °C and 1 bar absolute. When
operating at other conditions, it is necessary to increase the uncertainty of measurement by the
magnitude of the offsets if no corrections are applied or by a fraction of the offsets if corrections are
applied.
The following table lists the sources and magnitudes of the offsets affecting the meters covered in this
manual (including an example below).
Table 6-1.Source and magnitude of measurement offsets
Error source78357845/477845/47 Entrained Gas
A Primary Standard0.05 kg/m
B Transfer Standard0.1 kg/m
C Instrument Accuracy (at
calibration conditions)
0.15 kg/m
3
3
3
0.05 kg/m
0.1 kg/m
30.15 kg/m
3
3
3
0.05 kg/m
0.1 kg/m
1.0 kg/m
3
3
3
57Micro Motion 7835/45/47 Liquid Density Meters
Calibration and Performance
22222
GFEDC++++
Table 6-1.Source and magnitude of measurement offsets
D Temperature (uncorrected)
Temperature (corrected)
E Pressure (uncorr’d at 50 bar)
Pressure (uncorr’d at 100 bar)
Pressure (corrected)
F Velocity of Sound (uncorr’d)
Velocity of Sound (corrected)
G Long term stability0.15 kg/m
0.02 kg/m3/deg C
0.005 kg/m
–1 to +2 kg/m
+7 to +15 kg/m
3
/deg C
3
3
0.003 kg/m3/bar
See Section 6.2
20% of offset
3
/year0.35 kg/m3/year1.0 kg/m3/year
For total operational accuracy, the square root of the sum of the squares of each error source (C to G)
is recommended, such as:
•Effective Total =
For example, if we consider instruments operating at 50 °C (122 °F) and 50 bar, six months after
calibration and with no VOS offset, the total operational accuracy after corrections have been applied
is derived as follows:
Table 6-2.Total operational accuracy for example quoted
0.9 kg/m
0.05 kg/m
0 to +8 kg/m
---
0.006 kg/m
3
/deg C
3
/deg C
3
3
/bar
See Section 6.2
20% of offset
1.2 kg/m
0.5 kg/m
3
/deg C
3
/deg C
–40 to +40 kg/m
---
---
---
---
3
Error Source78357845/47
C
D
E
F
G
0.150.35
0.151.5
0.150.30
––
0.070.175
Effective Total0.271.58
For better accuracy, it would be necessary to carry out an on-line calibration at the operating
conditions. Higher accuracy can be obtained, by request, for all instruments by the use of water
calibration or by UKAS certified laboratory calibration of selected fluids.
Note: The tables above relate to the effect of uncertainties on the time period output of the meter, and
do not take into account any uncertainty in the measurement of the time period itself.
58Micro Motion 7835/45/47 Liquid Density Meters
Chapter 7
S
7965 DISPLAY
BaseD
LineD
S.G.
Tem
p
987.7 K /m
g
3
123.400 °
777.0 K /m
g
3
21.3 °
or
Remote Display and Digital Communications
7.1Introduction
The Advanced Baseboard provides an RS-485 digital communications link which can be used to
communicate, using RTU Modbus communications protocol, with either a 7965 Remote Display or
computer device up to 3280 ft (1000 m) from the meter. (Only one controlling device is permitted on
the RS-485 link at any one time.) It provides a convenient means for configuring the Baseboard and
for displaying or logging measurement data.
Figure 7-1Advanced baseboard with RS-485 digital communications link
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
The RS-485 link will support multi-drop installations where, for example, one Remote Display can
communicate with several Advanced units at the same time.
7.2Mechanical Installation of the 7965 Remote Display
The 7965 Remote Display is suitable for handheld or wall-mounted operation, and is designed for use
in both safe and hazardous areas.
Take care not to scratch the transparent front screen.
To install the 7965 Remote Display, the front display section must be separated from its backplate.
First, gently prise off both cover plates by using a small screwdriver inserted into a slot at the corner
of the cover. This reveals the four clamping screws that hold the front section to the backplate; undo
these and separate the two parts.
Installation & Configuration Manual59
Remote Display and Digital Communications
Cover plates
Clamping
screws
Slots for
removing
cover
plates
7965 DISPLAY
Figure 7-2Installation of the 7965 Remote Display
The display may be wall mounted by first screwing the backplate to the wall through the holes spaced
as shown below, and then screwing the main enclosure to the back plate with the clamping screws.
Figure 7-37965 Remote Display Dimensions
60Micro Motion 7835/45/47 Liquid Density Meters
Remote Display and Digital Communications
7965 DISPLAY
BaseD 777.0 Kg/m
LineD 987.7 Kg/m
S.G. 0.778 °
Temp 21.3 °
3
3
7.3Safe Area Electrical installation
Electrical installation of the Remote Display in a safe area is shown in Chapter 3. The maximum
resistance on the power lines between the Baseboard and the Remote Display is 60Ω, which equates
to a maximum cable length of 750 meters (2500 ft) for a cable of dc resistance 40Ω/km (12Ω/
1000ft). If the Remote Display is to be operated at distances of up to 1000 meters (3280 ft) from the
meter, it should be powered locally with a supply in the range 8 to 28 volts.
7.4Hazardous Area Electrical installation
When installing in hazardous areas, contact the factory for wiring information. Please also
refer to ATEX safety instruction booklet 79655010/SI for information on safety matters.
7.5Configuring the Baseboard using the Remote Display
On either side of the liquid crystal display there are four press-keys, each of which corresponds to one
of the four lines of text. These keys are used for entering data and for navigating the menu.
Figure 7-47965 Remote Display Press Keys
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
For security purposes, a connector, located within the enclosure, can be set to disable the keypad so
that the Remote Display simply displays measurement data. For additional security, password control
may be used to restrict access to certain areas of the menu structure.
7.5.1Power-up
Once the electrical installation is completed, the meter and display can be powered-up. The Remote
Display undergoes a sequence of self-checks and should then give a display similar to the one above.
If not, it will produce the screen shown below which indicates that communication between the meter
and display has not been established.
Installation & Configuration Manual61
Remote Display and Digital Communications
7965 DISPLAY
No devices replying
< Enter demo mode >
< Try again >
Failure to establish communications is normally due to incorrect electrical installation or incorrect
meter slave address (see below).
7.5.2Slave address
Each meter is given a Modbus slave address. The factory default slave address is 1 (one). By
entering the [Change setup] section of the display menu, the Remote Display can be configured
to communicate with any valid slave address (i.e. in the range 1 to 247), or to poll all slave addresses
in any given range.
If more than one meter is connected to a single display then each meter should be given a different
slave address (see Section 7.5.2). If the slave address has to be changed, it is recommended that the
lowest unused slave address be used.
7.5.3Demo mode
The demo or demonstration mode allows the user to explore the Display menu structure without a
meter attached.
7.5.4Navigating the menu structure
Access to the menu structure is obtained by pressing any of the 8 keys either side of the display,
giving the following menu-screen:
62Micro Motion 7835/45/47 Liquid Density Meters
Remote Display and Digital Communications
7965 DISPLAY
Process variables
Outputs
Transducer setup
↓
ade
,
solartro
solartron
arnborough, Hamps
Y
Madein theU.K.bySolartron,
ough, Hamps
d
n
Farnborough,Hampshire,England.
solartron
L
3
B
3
S
T
C
P
s
O
s
T
V
^
P re s su re
Communications
V
^
S
x
L
p
V
e
Figure 7-67965 Remote Display Menu Screen
Corresponding to each line of text, there are two keys, one either side of the display. Unless indicated
otherwise, the keys on either side of the display perform the same function.
Over 100 separate menu screens are provided. However, after a few minutes of familiarization, the
user should find it simple to navigate through the entire menu.
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
Figure 7-7Remote Display Menu Screens Sequence
n
7965 D ISPLAY
ineD 987.7 kg/m
aseD 967.4 kg/m
. G . 98 .0 0
e m p 21 .3 °
M
intheU.K.bySolartron
Farnborough, Hampshire, England
796 5 DI SP LAY
^
Diagnostics
Servic
Finish
Made inthe U.K. bySolartron
The menu structure can be thought of as being cyclic, so that, that by repeatedly pressing the
downward arrows (↓), the user will eventually end up back at the first screen.
solartro
79 65 D ISP L A
et Referral Matri
o c al s e tu
Farnbor
hire, Englan
solartron
Made inthe U.K.bySolartron
F
79 65 DI SP LAY
rocess variable
ut pu t
ransducer setup
Made inthe U.K. bySolartron
Farnborough, Hampshire,England.
79 65 DI SPLA Y
hire,England.
Some menu items lead to another level of menus, which are also cyclic. In some cases, there are
several levels of menus. The simplest way of getting to know these levels is to experiment.
Installation & Configuration Manual63
Remote Display and Digital Communications
If required, the Remote Display can be reset by pressing the four corner keys together and holding
them down for about 2 seconds. When this is done, the Remote Display will re-establish
communication with the meter and will bring back the front menu page displaying measurement data,
from which the menu structure can be accessed once again.
7.5.5Menu Structure
The tables below represent some of the top-level menu screens.
Menu-screen 1 at Level 1
Process variables
Outputs
Meter setup
↓
Menu-screen 2 at Level 1
↑
Pressure
Communications
↓
Menu-screen 3 at Level 1
↑
Set referral matrix
Local setup
↓
Menu-screen 4 at Level 1
↑
Diagnostics
Service
Finish
(See Level 2 “Process Variables” menu below)
→
(See Level 2 “Outputs” menu below)
→
Set slave address, view meter serial no. etc.
→
Line pressure, atmospheric pressure, pressure units
→
Set device addresses, poll network again
→
Temperature points, referral temperature, referral points
→
(See “Local Setup” menu below)
→
(See “Diagnostics” menu below)
→
For service engineers only
→
Return to live display
→
Process Variables (PV) Menu (Level 2)
•PV Menu-screen 1
Line density units
Base density units
Temperature units
↓
Set line density units to kg/m³, lb/gal etc.
→
Set base density units to kg/m³, lb/gal etc.
→
Set temperature units to °C or °F.
→
•PV Menu-screen 2
↑
P.V. averaging
Set special function
↓
64Micro Motion 7835/45/47 Liquid Density Meters
Select special function
→
Select process variable averaging (1s, 2s, 5s etc.)
→
Remote Display and Digital Communications
•PV Menu-screen 3
↑
Density offset
Temperature offset
< Exit variables >
Set pulse output to be alarm status or tube frequency
→
Set alarm state, coverage and hysteresis
→
Select menu language (currently only English)
→
Set screen contrast
→
Turn system warnings on and off
→
View display unit details
→
Set device addresses, poll network again
→
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
Diagnostics Menu (Level 2)
•Diagnostics Menu-screen 1
Pickup level
Q of resonance
↓
•Diagnostics Menu-screen 2
↑
Tube period
RTD resistance
↓
•Diagnostics Menu-screen 3
↑
Change fixed values
Fix meter readings
< Exit diagnostics >
Enter values for density etc. to use when fixing readings
→
Fix density and temperature readings to set values
→
Installation & Configuration Manual65
Remote Display and Digital Communications
OR
+ Pwr
-Pwr
A
B
A
A
A
B
B
B
7965 DISPLAY
7.6Multi-drop installation
For a multi-drop installation each meter is allocated a different slave address and linked together in
parallel as shown below. Each meter must be individually programmed with its unique slave address
using the PC or remote display before they are linked together. Up to 24 meters can be connected in
such an arrangement, depending on the cable parameters.
Please note that the arrangement shown below is NOT
Figure 7-8RS-485 Multidrop Arrangement
suitable for hazardous area installations.
Note: The meters must be given different slave addresses before
they are connected together in a
multidrop arrangement, otherwise there will be communications conflicts between the meters.
The Remote Display communicates with one meter at a time; to obtain information from several
meters the Remote Display must be set, each time, to interrogate the individual meters.
7.7Electrical installation of Computer Devices
Electrical installation of computer devices having an RS-485 serial port is the same as for a Remote
Display, except that the power lines are not required.
When installing in hazardous areas, a safety barrier or galvanic isolator must be
interposed between the RS-232/RS-485 converter and the meter – contact the factory
for wiring information. Please also refer to ATEX safety instruction booklet 79655010/SI
for information on safety matters.
Personal Computers (PCs) with an RS-232 serial port can readily communicate with the Baseboard
using a line powered RS-232-to-RS-485 converter, as shown below.
66Micro Motion 7835/45/47 Liquid Density Meters
Remote Display and Digital Communications
RS232 to RS485
CONVERTER
A
B
RS232
RS485
Figure 7-9Advanced System Linked to a Personal Computer
In order for a computer device to communicate with an Advanced unit, a software program is required
to generate and interpret Modbus messages. The information provided in Appendix H is intended to
enable users to write their own Modbus software. For information on using ADView software, please
see Chapter 8.
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
7.7.1Connections using an RS-232/485 Converter
Terminals 11 and 12 on the Advanced unit’s Baseboard are for RS-485 (Modbus) connections to the
converter, as shown in Figure 7.8. Note: The PC and converter are always located in a non-hazardous
(safe) area.
Converters are available from a number of sources, and can range from simple in-line devices that
simply plug into a PC’s RS-232 port, to programmable devices with full isolation between the two
networks.
The Advanced unit uses a half-duplex implementation of RS-485, such that the A and B signals are
used for data transmission in both directions. This requires that the RTS line is toggled to indicate the
transmission direction; it can be done by the host computer, or automatically by an RS-485/232
converter which has the facility to do so. If you are using Windows NT, 2000 or XP on your PC, you
should use a converter which automatically changes RTS (as detailed below) otherwise the link may
not work correctly.
The optional ADView software kit includes a K3 RS-485/RS-232 converter that is manufactured by
KK Systems Ltd.
The K3 converter derives its power from the PC’s RS-232 port RTS or DTR line, which must be held
permanently in the high state. This is normally adequate for short distances where there are only a
few devices on the network. However, the ability of the port to supply sufficient power is not
guaranteed, especially for laptop PCs, and it may be necessary to connect an external power supply.
This may also be necessary if using Windows NT, 2000 or XP.
To check the voltage levels, measure the voltages on the RTS input (pin 7) and the DTR input (pin 4)
while the converter is connected to the PC (or other RS-232 device). This procedure needs a
break-out box (not supplied).
Whichever input is powering the converter must have at least +6V during communications. Where
the power is found to be insufficient, a 9V dc supply can be plugged into the option DC Input socket
(Figure 7-9). See also the manufacturer’s technical information for details.
Installation & Configuration Manual67
Remote Display and Digital Communications
PC (+Adview)
1 (A)
2 (B)
K3-ADE
RS232/RS485
Converter
1
2
34
56
ON
R
S
2
3
2
R
S
4
8
5
1
2
78 9101112
AB
12
345
6
PL2
If you encounter communication
difficulties with RS-485, swap
over the ‘A’ and ‘B’ signal
connections at one end of the
network.
PC (+Adview)
K2-ADE
RS232/RS485
Converter
R
S
2
3
2
9 (+9V)
9V dc
Power
Supply
-
+
RS232
5 (GND)
+6V to +16V dc
Figure 7-10 RS-485 connections < 50 meters
Figure 7-11 Powering the Converter with an External 9V dc Supply (Non-hazardous Area)
68Micro Motion 7835/45/47 Liquid Density Meters
For permanent installations, and where the network length is more than 100 meters or so, use the
DIN-rail mounted device KD485-ADE from KK Systems Ltd.
The KD485-ADE is three-way isolated, providing isolation between the two ports and the power
supply. It requires a +7 to +35V power supply and typically takes 1 to 2W; (power consumption is
largely independent of supply voltage). It is capable of working with Windows 98, NT, 2000 and XP.
For a PC running Windows NT/2000/XP, the RTS connection can be omitted.
Remote Display and Digital Communications
Port 1
RS232
Port 2
RS485
KD485-ADE
RS232 to RS485 Interface Converter/Isolator
Switch
Tx
Rx
Port 1 GND
RTS In
1
2
3
4
5
6
Power Input
+
-
7
8
6
5
4
3
2
1
Port 1 GND
TxB
TxA
RxB
RxA
Figure 7-12 Modbus connections > 100 meters
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
The default configuration of the KD485-ADE has Port 2 configured for 9600 baud. The meter uses
the following parameter settings, which are not selectable:
•Baud rate: 9600
•Bits: 8
•Parity: None
•Stop bits: 2
Figure 7-13 Switches on KD485-ADE
The switch on the KD485-ADE should be set with SW1 On (to enable half-duplex operation on Port
2), with the other three switches (SW2, SW3, SW4) set to Off.
Note: In most systems, the ground (GND) connection on pin 6 of port 2 will be unnecessary.
Installation & Configuration Manual69
Remote Display and Digital Communications
70Micro Motion 7835/45/47 Liquid Density Meters
Chapter 8
Using ADView and ProLink II
8.1Using ADView Software
8.1.1What is ADView?
ADView is a software package provided by Micro Motion to enable you to:
•Configure our density and viscosity transmitters.
•View and save data from them.
•Check that they are functioning correctly.
ADView is installed on a PC and interacts with the density/viscosity transmitter through one of the
PC’s standard serial (RS-232) ports.
ADView requires Microsoft’s Windows operating system: Windows 3.1, 95, 98, NT, 2000 or XP.
Note: To connect to an RS-485/Modbus device, such as the 7835, you will need an adapter between
the PC and the meter (see Chapter 4).
ADView provides many useful facilities, such as:
•Setting up serial link to communicate with the meter.
•Configuring the transmitter.
•Displaying data in real time, or as a graph.
•Logging data to a file.
•Verifying correct operation of the system, and diagnosing faults.
•Loading or storing Modbus register values.
•Read/write to individual Modbus registers.
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
8.1.2Installing ADView
ADView software is available for the PC on a variety of media (for example, CD-ROM) and is freely
available to download from the Micro Motion web site (at www.micromotion.com).
1. Identify the media containing the installation files for ADView.
2. Insert the media into an appropriate drive on your PC.
3. If the installation program does not begin automatically, run the set-up ‘.exe’ file that is on the
media. This does vary between different PC operating systems. In general, open the File
Manager or Windows Explorer, browse the drive containing the media and double-click on the
set-up ‘.exe’.)
4. When the installation program starts, you will be asked to supply your name and organization
name for registration purposes, and supply a directory path into which ADView’s files can be
loaded (a default directory path will be suggested).
Installation & Configuration Manual71
Using ADView and ProLink II
5. Follow the installation instructions until installation is complete. It will normally only take a
few minutes. You can abandon the installation if you need to do so.
8.1.3Starting ADView
Start the ADView software by navigating through the Start Menu to the program entry of ADView 6.
Left-click on it once and the window shown below will then appear.
Note: Developments in ADView may mean that the screen shots differ slightly from the ones you will see on
your PC screen.
Each of the six icons gives you access to the various facilities of ADView. You can choose to connect
a Modbus device to one of the PC’s serial ports, or you can use ADView’s built-in simulation of the
meter.
To run the simulation, choose
the appropriate density meter option. Then, click on the
Options > Simulate board response from the menu bar and choose
OK buttons, as necessary, to return to the
main ADView screen. When simulation is chosen, ADView ignores the serial port and supplies
simulated data. However, you do still need to click on the
by the Connect button. Then, click on the
OK buttons, as necessary, to return to the main ADView
Communications Setup button followed
screen.
Setting up serial communications
To operate with a real Modbus device, you will need to connect it to a suitable power supply (see the
technical manual for the device) and need a connection to a serial port on the PC. Full details for
connecting to the Modbus (RS-485) link on the meter are in Chapter 4.
72Micro Motion 7835/45/47 Liquid Density Meters
Using ADView and ProLink II
ADView automatically configures the selected port with the correct settings for the device. For the
meter, this is 9600 baud rate, 8 data bits, no parity, 1 stop bit, and Xon/Xoff (software) flow control.
8.1.4Understanding ADView Features
ADView facilities
The main ADView window gives access to the various facilities available. A brief description of each
is listed below. Using the facilities is largely intuitive so that you can quickly learn the system.
Communications Setup
Sets up and checks RS-232/RS-485 communications.
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
Board Configuration
Enables you to select the measured parameter and range for the analog output,
and to configure density referral by entering matrix values or K factors, as well as
special calculations, line pressure and averaging time.
Displays instantaneous values of a selectable output parameter and the analog
output.
Data logging
Provides tabular data from transmitters of line and base density, temperature and
special function. One parameter can be displayed as a graph.
Data can also be logged to a file in either Excel (tab delimited) or Notepad (space
delimited) formats.
The frequency at which results are logged can be set, and logging can be started
and stopped.
Register dump/load
With this facility you can dump the contents of all (or selected) Modbus registers
from the device, or alternatively transmit values to them. File format is selectable
(Excel/tab delimited, or Notepad/Space delimited).
Transducer details
Shows a list of meter details such as type, serial number, calibration dates,
software version, etc.
Diagnostics
Enables you to view:
- live sensor readings
- the status of the meter
- values of working coefficients
You can also verify calculations.
Installation & Configuration Manual73
Using ADView and ProLink II
Menu bar
File
Tools
Options
Window
Help
ExitExit ADView program.
Health CheckDetermines whether the system is functioning correctly.
Register Read/WriteA facility for reading or writing to any of the Modbus
registers (see Section F.7)
Direct Comms.Enables you to specify exactly what will be transmitted on
the Serial link (see Appendix F).
Engineer StatusOnly used by Micro Motion service engineers.
Simulate board response/
Actual Board
Enable / disable screensaverAllows you to select between these two options. When
About ADViewDisplays software version number.
Allows you to select between these two options
enabled, the screensaver operates as configured by the
Windows system settings.
Provides a means of opening or selecting ADView’s
facilities.
Configuring a slave address
The factory configuration sets the slave address to 1. However, in many applications it will be
necessary to allocate another address. In a multi-drop application, where several Modbus devices are
connected on the same network, it is essential to configure unique slave addresses for each device.
To do this, you will need to run ADView and use the Register Read/Write facility. Check the value in
Register 30 (Modbus Slave Address). If it is not the required value, enter the desired value and click
on the write button. The meter will now be configured with the new slave address.
Board configuration
The board configuration controls the way in which the meter will process and present data, user
settings, calibration constants and other factors. This data is stored in non-volatile memory known as
registers; a full list of the registers used in the meter is given in Appendix F.
To configure the meter, it is necessary to write data into the configuration registers using the
RS-485/Modbus link. ADView provides a convenient and graphical way of doing this without you
needing to know about register addresses and data formats.
Certain parameters are not available for configuration by ADView, including the Density Offset value
which may be required to fine tune the calibration of the meter. However, ADView does have tools for
reading and writing to individual Modbus registers (using the
and for direct communication on the Modbus (using
Tools > Direct Comms). More details and
Tools > Register Read/Write facility),
examples are given in Appendix F, but for the significant majority of applications these tools will not
be required.
74Micro Motion 7835/45/47 Liquid Density Meters
Using ADView and ProLink II
There is no facility within ADView or the meter to ‘reset’ to a default configuration. Therefore,
before attempting any alterations to the configuration, you are strongly advised to use the
Register Dump/Load facility in ADView to store the existing configuration. Then, if any mishap
occurs, you will be able to restore the configuration from the saved file.
ADView’s Board Configuration window is shown below:
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
To exit from any of the configuration windows without making any changes, press the
Esc key on
your computer keyboard.
Density referral (Configure… button)
To configure the density referral calculation, you will need to enter the relevant information.
•For matrix referral, this is a set of four values of density for each of up to five different
temperatures; Appendix D gives more details on this.
•For API referral, you can select the product type, which automatically adjusts the coefficients
of the General Density Equation (see Chapter 6), or enter your own values.
Installation & Configuration Manual75
Using ADView and ProLink II
The Log Setup button – which is activated
when logging has been stopped – enables
you to configure the frequency of logging,
where the logged data will be filed, and the
format of the data.
Display Selection dropdown
list to select the transmitter
and parameter to be displayed
on the graph
For selecting the parameter to
be logged.
Click Show Graph
to configure and
display graph
For multi-drop
configurations, the
output of up to three
transmitters can be
displayed
simultaneously.
Tabular display of
instantaneous
output of
transmitter.
Select analog
output of another
transmitter.
Click OK to close
Data Logging window
Graphical
representation of
analog output.
Click Start to
start logging.
Click Stop to stop
logging
Special function (Configure… button)
The range of special functions (calculated parameters) that are available depends on the referral type
selected.
Special FunctionAPI referralMatrix referral
Specific Gravity✓✓
API°✓
% mass✓
% volume✓
° Baumé✓
° Brix✓
User defined quartic✓
None✓✓
When you select the Special Function you require, the configuration window will alter to allow you to
input the relevant parameters, if applicable. Note that you can only select one Special Function to be
available at any one time.
When you are satisfied with the configuration, you should save it to a file, using the
Dump/Load
facility, as a safeguard against subsequent loss or alteration.
Register
Data logging
ADView’s Data Logging function is a useful tool for checking setups and performing experimental
data capture. The diagram below explains some of the features.
76Micro Motion 7835/45/47 Liquid Density Meters
Using ADView and ProLink II
Address of unit
being accessed
Enter desired filename
for Dump, or required
filename for Load.
Choose which
sets of registers to
save to file, or
simply save all of
them.
You can also
specify individual
registers.
Choose data
delimiter
(Dump only)
Restore a previously
saved set of register
data from file.
Store the selected
register data to a file.
Register DUMP/LOAD
This facility is essential for saving the configuration of your meter. You should use it to save the
current configuration before you start to alter it, in order to restore it if things go wrong for any
reason. Also, if you send the transmitter away for servicing or re-calibration, you should save the
current configuration. Details are given below.
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
Register Read/write
In a few cases, it may be useful to write directly to a single Modbus register. Two likely occasions for
using this feature are to set the Slave Address of the unit and to configure a density offset. Appendix F
has a complete list of the registers.
Before making any changes to individual registers, you should save the current configuration to
a file (see Register DUMP/LOAD) to safeguard your configuration if anything goes wrong.
From ADView’s menu bar, select
Tools > Register Read/Write.
Installation & Configuration Manual77
Using ADView and ProLink II
To see a complete list of
Modbus register numbers
and descriptors, click
here.
Choose the one you want
to access.
For non-numerical values,
click here to see complete
list of possible entries and
select one to write into the
register.
Enter numerical values
directly.
The Write button
causes the current
value to be written to
the selected register.
You can read and write to any
number of registers. When you
have done all you want to, click
OK.
The Read button
causes the current
value of the chosen
register to be
displayed.
The current
register number
appears here.
8.2Using ProLink II Software
8.2.1Overview
ProLink II is a Windows-based configuration and management tool for Micro Motion transmitters. It
provides complete access to transmitter functions and data.
This chapter provides basic information for connecting ProLink II to your transmitter. The following
topics and procedures are discussed:
•Requirements (see Section 8.2.2)
•Configuration upload/download (see Section 8.2.4)
The instructions in this manual assume that users are already familiar with ProLink II software. For
more information on using ProLink II, see the ProLink II manual.
8.2.2Requirements
To use ProLink II with a 7835/45/47 liquid density meter, the following are required:
•ProLink II v2.9 or later
•Signal converter(s), to convert the PC port’s signal to the signal used by the transmitter
-For RS-485 connections, an RS-485 to RS-232 signal converter. See Section 4.5 for
information.
-25-pin to 9-pin adapter (if required by your PC)
8.2.3Connecting from a PC to a transmitter
Table 8-1 describes the options for connecting ProLink II to your transmitter. See Section 4.4 and
Section 4.5 for more information.
78Micro Motion 7835/45/47 Liquid Density Meters
Using ADView and ProLink II
Table 8- 1Connection options for 7835/45/47 liquid density meter
ConnectionPhysical layerProtocol
RS-485 terminals or RS-485 network
(see Section 4.5)
RS-485Modbus
8.2.4ProLink II configuration upload/download
ProLink II provides a configuration upload/download function which allows you to save configuration
sets to your PC. This allows:
•Easy backup and restore of transmitter configuration
•Easy replication of configuration sets
Micro Motion recommends that all transmitter configurations be downloaded to a PC as soon as the
configuration is complete.
To access the configuration upload/download function:
1. Connect ProLink II to your transmitter as described in Section 4.4 and Section 4.5.
2. In the ProLink II software application, open the
•To save a configuration file to a PC, use the
File menu.
Load from Xmtr to File option.
Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)Calibration and PerformanceUsing Adview and ProLink IIRemote Display and Digital Comm.Electrical Connections (EG Option)
•To restore or load a configuration file to a transmitter, use the
option.
8.2.5ProLink II language
ProLink II can be configured for the following languages:
•English
•French
•German
To configure the ProLink II language, choose
Tools > Options.
In this manual, English is used as the ProLink II language.
Send to Xmtr from File
Installation & Configuration Manual79
Using ADView and ProLink II
80Micro Motion 7835/45/47 Liquid Density Meters
Chapter 9
General Maintenance
9.1General
The 7835/45/47 liquid density meters have no moving parts, which reduces the maintenance
requirement to simple visual checks for leaks and physical damage.
Check calibrations should be carried out at specified intervals in order to highlight any malfunction or
deterioration in meter performance. If a fault or a drop in meter performance is discovered, further
tests are required to identify the cause of the fault. Remedial action is limited to cleaning the tube,
making good any poor connections and replacing the maintaining amplifier or, in extreme cases, the
entire instrument.
Extreme care is required in the handling of the meter during transit, its installation into the
pipeline and its removal from the pipeline.
Faults generally fall into two main categories: erratic readings or readings outside limits.
•Erratic Readings
Normally caused by the presence of gas bubbles in the flowing liquid. Severe electrical
interference or severe pipeline vibrations can also cause this effect.
•Readings Outside Limits
Normally caused by deposition and/or corrosion on the resonating tube.
Since an electrical fault could also cause either of the two faults, and since examination for deposition
or corrosion requires the removal off-line of the meter, it is recommended that the electrical system be
checked first.
9.3General maintenance procedure
This procedure is recommended for any periodic maintenance carried out on the system and forms the
basis of any faultfinding task.
9.4Physical checks
Physical checks are as follows:
•Examine the meter and its mounting bracket, pipe couplings and electrical cables for signs of
damage and corrosion.
•Check the meter for signs of fluid leakage and the state of the rupture plate.
Installation & Configuration Manual81
General Maintenance
Notes:
•Any physical damage to the meter case or mounting brackets may have adverse effects on the
meter performance and a full calibration would be advisable to verify its accuracy.
•Any oil leakage can generally be remedied by servicing.
9.4.1Check calibration
Checking the calibration is as follows:
•Carry out a check calibration using methods detailed in Chapter 6.
•Compare the results obtained with the current calibration certificate figures to identify any
substantial deterioration in the meter's performance or any malfunction.
Notes:
•A substantial drop in meter performance is likely due to a build-up of deposition on the
vibrating tube, which can be removed by the application of a suitable solvent. See
Section 9.4.2 below.
•Malfunctions may be the result of electrical/electronic faults in either the meter circuit or the
readout equipment. The readout equipment should be proved before attention is directed to the
meter as detailed under Section 9.4.2.
9.4.2Remedial servicing
The required servicing falls into two categories – electrical and mechanical.
Electrical servicing
1. Follow the steps below.
Units with Advanced Electronics:
a.Check the voltage between terminals 7 and 8 is between 9.5 V and 28 V.
Units with Standard Electronics:
a.Carry out power supply and current consumption tests at the meter terminals. These
should give: 17 mA
± 1 mA at 15.5 V to 30 V.
b.Remove the power supply to the meter. If current consumption is suspect, replace the
meter amplifier.
c.Identify the drive coils (terminals 7 and 8) and disconnect the drive coil wires from the
amplifier. Measure the resistance of the drive coils. This should be: 95
± 5 ohms at 20 °C
(68 °F).
d.Reconnect the drive coil wires to the amplifier.
2. Identify the pick-up coils (terminals 9 and 10) and disconnect the pick-up coil wires from the
amplifier. Measure the resistance of the pick-up coils. This should be: 95
± 5 ohms at 20 °C
(68 °F).
Reconnect the pick-up coil wires to the amplifier.
3. Follow the steps below.
Units with Advanced Electronics:
a.With no power to the unit, check the 100 Ω RTD element across terminals 4 and 7. The
value of the element resistance is temperature dependent. For this data, see Appendix F.
82Micro Motion 7835/45/47 Liquid Density Meters
General Maintenance
Units with Standard Electronics:
a.Check the 100 Ω RTD element across the terminals 11 and 12 (ensure terminals 3 to 6 are
b.Check for continuity between terminals 11 and 3, and terminals 11 and 4, also from
4. Carry out an insulation test by removing all the input connections to the amplifier terminals
(1 to 7 inclusive) and short-circuit the terminals together. Test their insulation resistance to the
metal case using a 500 V dc insulation tester (current limited to 5 mA maximum). This
resistance must be greater than 2 MΩ.
Remove the short-circuit, and reconnect the input leads if required.
Mechanical servicing
Mechanical servicing comprises mainly of keeping the inner surface of the vibrating tube clear of
deposition and corrosion. Deposition may be removed by the use of a suitable solvent. Alternatively,
the instrument can be removed from the pipeline and cleaned mechanically. Care is required to
prevent damage to the inner surface of the tube during the cleaning.
disconnected). The value of the element resistance is temperature dependent. For this data,
see the product specifications appendix.
Great care is essential in handling the meter during transit, installation into the pipeline and
removal from the pipeline.
Ensure that the meter is not transported when it contains hazardous fluids. This includes fluids which
may have leaked into, and are still contained, within the case.
Installation & Configuration Manual83
General Maintenance
84Micro Motion 7835/45/47 Liquid Density Meters
Appendix A
7835 Specifications
A.1Density performance
Accuracy
Operating Range
Repeatability
Stability
Process Temperature Effect
(Corrected)
Process Pressure Effect
(Corrected)
(1) Accuracy is dependent upon the calibraton option chosen. Density range for which this accuracy applies depends on the
calibration option chosen.
(2) Stated accuracy is for operating density range of 0.3 to 1.1 g/cc (300 - 1100 kg/m
(3) Temperature effect is the maximum measurement offset due to process fluid temperature changing away from the
density calibration temperature.
(4) Pressure effect is defined as the change in sensor density sensitivity due to process pressure changing away from the
calibration pressure. To determine factory calibration pressure, refer to calibration document shipped with the 7835. If
data is unavailable, contact the factory.
Recognizing the increased emphasis on safety by chemical, hydrocarbon, and process markets alike,
these Micro Motion density meters have been enhanced by the introduction of an optional outer 725
psi (50 bar) or secondary 1450 psi (100 bar) pressure retaining capability. In the unlikely event of an
instrument failure, the meter safely contains any leakage. As a further safety feature, all welds are
qualified to
ASME 9/EN ISO 15614–1 standards and can undergo dye penetration testing to ASME standards, if
required. Furthermore, the flange welds may be x-rayed to most recognized international standards.
Case pressure
Case failure pressure
Standard Containment
Limited by the Case failure
pressure specification below
Fitted with burst disc which
will fail between 290–435 psi
(20–30 bar)
Optional Outer
Containment
725 psi (50 bar)
Standard engineering practice
2900 psi (200 bar)5727 psi (395 bar)
Optional Second
Containment
1450 psi (100 bar)
designed to B31.3
Glass to metal seal failure
Installation & Configuration Manual87
A.9Weight
Weight
Intrinsically safe (Ex ia)48 lb (22 kg)
Explosion proof (Ex d)77 lb (35 kg)
A.10Electrical
Power supply (Frequency Output
version)
Power supply (Transmitter version)
Outputs (Frequency Output version)
Outputs (Transmitter version)
16 to 28 VDC at 17 mA maximum
18 to 28 VDC at 80 mA
Current modulation on power supply line
Analog
Accuracy
Repeatability
Out-of-range
Pulse output
(Ex ia transmitter only)
Communications (Ex ia
transmitter only)
2 (+1 with HART option board)
0.1% of reading plus 0.5% of full scale
±0.025%
2 to 20 mA on 4–20 mA
(Programmable alarm state)
Open collector output. Alarm status or
frequency.
RS-485, Modbus (standard), HART
(optional)
88Micro Motion 7835/45/47 Liquid Density Meters
Appendix B
7845/7847 Specifications
B.1Density performance
Accuracy±0.0001 g/cc
±0.00035 g/cc
±0.005 g/cc
±0.1 kg/m
±0.35 kg/m3
±5.0 kg/m
3
3
(Enhanced calibration)
(Standard calibration)
(Entrained Gas Option)
(1)
(2)
(3)
Operating RangeUp to 3 g/cc Up to 3000 kg/m3
Repeatability±0.00005 g/cc
±0.001 g/cc
±0.05 kg/m
±1.0 kg/m
Stability±0.00035 g/cc0.35 kg/m
Process Temperature Effect
(Corrected)
Process Pressure Effect
(Corrected)
(1) Accuracy is dependent upon the calibraton option chosen. Density range for which this accuracy applies depends on the
calibration option chosen.
(2) Stated accuracy is for operating density range of 0.6 to 1.2 g/cc (600 – 1200 kg/m
1.6 g/cc (1600 kg/m3), contact the factory for accuracy specifications.
(3) Percentage of entrained gas range 0 to 100%.
(4) This is the maximum measurement offset due to process fluid temperature changing away from the density calibration
temperature.
(5) Pressure effect is defined as the change in sensor density sensitivity due to process pressure changing away from the
calibration pressure. To determine factory calibration pressure, refer to calibration document shipped with the 7845/47. If data
is unavailable, contact the factory.
(1) –58 °F to 320 °F (–50 °C to +160 °C) with high temperature kit option.
(1)
–58 °F to +230 °F (–50 °C to +110 °C)
B.2.1Integral temperature sensor
Technology100 Ohms RTD (4 wire)
AccuracyBS 1904 Class, DIN 43760 Class A.
Installation & Configuration Manual89
7845/7847 Specifications
B.3Pressure ratings
Maximum operating
pressure
7845
7847
1450 psi (100 bar) or flange limit
290 psi (20 bar) or flange limit
Test pressureTested to 1.5 x the maximum operating pressure
PED complianceComplies with European directive 97/23/EC on Pressure Equipment.
B.4Hazardous area classifications
ATEX Intrinsically Safe
ATEX-approved I.S. 7845/47: Certification to EN 60079-0: 2006 and EN 60079-11: 2007 for use in Europe
(1)
7845/47 (Frequency Output)
7845/47 (Transmitter)
(1)
Remote Display (Optional)ATEX II 1 G, Ex ia IIC, T4 (Ta –40°C...+60°C)
(1) Entrained gas 7845/7847 (Frequency Output) approved for use in non-hazardous areas only.
(784x****AJ****)ATEX II1G, Ex ia IIC
T6 (Ta –40 °C...+40 °C)
T4 (Ta –40°C...+70°C)
(784x****(D/H)J****)
(784x****(B/F)J****)
ATEX II1G, Ex ia IIB T4 (Ta –40°C...+60°C)
ATEX II1G, Ex ia IIC T4 (Ta –40°C... +60°C)
CSA Intrinsically Safe
CSA-approved I.S. 7845/47: Certification to CSA C22-2 No 142, CSA C22-2 No 175, UL 508 and UL 913 for use in
Canada/USA
(1)
7845/47 (Frequency Output)
7845/47 (Transmitter) and
optional Remote Display
(1)
(784x****AL****)Class I, Division 1 Groups C & D, T3C
(784x****(B/F)L****)
(784x****(D/H)L****)
Class I, Division 1, Groups A, B, C & D, T4 (Single
instrument)
Class I, Division 1, Groups C & D, T4 (Hart Multi-drop)
(1) Entrained gas 7845/7847 (Frequency Output) approved for use in non-hazardous areas only.
B.5OIML R117-1 classifications (7845 Only)
Evaluation to OIML R117-1 Edition 2007 (E) and Measuring Instrument Directive (2004/22/EC) Annex MI-005
Viscosity range 0.75 cP to 50 cP (0.75 mPa
Density range0.7 g/cc to 1.2 g/cc (700 kg/m
Temperature range ambient–40 °F to +158 °F (–40 °C to +70 °C)
Environmental classMechanical: M2
Electromagnetic: E2
Maximum pressureFluid temperature rangeAccuracy Class
• 870.2 psi (60 bar)• +41 °F to +95 °F (+5 °C to +35 °C)• 0.3
• 870.2 psi (60 bar)• –14 °F to +131 °F (–10 °C to +55 °C)• 1.0
·s to 50 mPa·s)
3
to 1200 kg/m3)
90Micro Motion 7835/45/47 Liquid Density Meters
7845/7847 Specifications
B.6General classifications
B.6.1Electromagnetic compatibility
All versions conform to the latest international standards for EMC, and are certified compliant with:
•Emissions: EN 61326 – 1997 (Heavy Industrial Environment)
•Radiated emissions in the range 30 MHz to 100 MHz, and conducted emissions in the range
0.15 MHz to 30 MHz complying with standard EN 61000-4
•Immunity: BS EN 61000-6.2
B.7Materials of construction
Wetted parts316L Stainless steel
Case finish316L Stainless steel
Flange316L Stainless steel
B.8Fluid containment
Recognizing the increased emphasis on safety by chemical, hydrocarbon, and process markets alike,
these Micro Motion density meters have been enhanced by the introduction of an optional outer
725 psi (50 bar) or secondary 1450 psi (100 bar) pressure retaining capability. In the unlikely event of
an instrument failure, the meter safely contains any leakage. As a further safety feature, all welds are
qualified to ASME 9/BS/EN288 standards and can undergo dye penetration testing to ASME
standards, if required. Furthermore, the flange welds may be x-rayed to most recognized international
standards.