Micro Motion Installation Instructions: Model D DL DT Installation Instruction Multi-Language | Micro Motion Manuals & Guides

P/ N 20000174 Rev. A March 2003
Micro Motion® Model D, DT, and DL Sensor
Multi Language Installation Instructions
Micro Motion
TM
Contents
English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Français . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Deutsch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Nederlands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Español . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Italiano . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Português . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Suomen kieli . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Norsk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Svensk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Dansk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
i
ii
English March 2003
Micro Motion® Model D, DT, and DL Sensor Installation Instructions
For online technical support, use the EXPERT2™ system at www.expert2.com. To speak to a customer service
representative, call the support center nearest you:
In the U.K., phone 0800 - 966 180 (toll-free)
Outside the U.K., phone +31 (0) 318 495 670
In the U.S.A., phone 1-800-522-MASS (1-800-522-6277)
In Canada and Latin America, phone (303) 530-8400
In Asia, phone (65) 6770-8155
Micro Motion
TM
1
BEFORE YOU BEGIN
About these instructions
This quick reference guide explains basic installation guidelines for all Micro Motion
®
Model D (except D600), DL, and DT sensors. For more information about sensors, refer to the instruction manual that was shipped with the sensor.
European installations
Micro Motion products comply with all applicable European directives when properly installed in accordance with the instructions in this quick reference guide. Refer to the EC declaration of conformity for directives that apply to a specific product.
The EC declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions are available on the internet at www.micromotion.com/atex or through your local Micro Motion support center.
INTRODUCTION
The sensor makes up one part of a Coriolis flowmeter. The other part is a transmitter.
Installation options
Model D, DT, and DL sensors may be connected to the Micro Motion transmitters listed in Table 1.
©2003, Micro Motion, Inc. All rights reserved. Micro Motion is a registered trademark of Micro Motion, Inc. The Micro Motion and Emerson logos are trademarks of Emerson Electric Co. All other trademarks are property of their respective owners.
2
Table 1. Transmitter and sensor compatibility guide
Transmitters
Model 1700/2700 (9-wire) X X Model 3500/3700 (9-wire) X X RFT9739 (7- or 9-wire) X X IFT9701 (9-wire) X RFT9712 (7- or 9-wire) X X
Model D and DL
Sensors Model DT Sensors
Model D, DT, and DL sensors are available with a 9-wire connection to a remote transmitter. Refer to Figures 1, 2, 3, and 4 for components of the sensor. (Note: Figures are not to scale.)
Figure 1. DS025, DH025, DH038, and DS040 sensors
Junction box
Purge connections
(optional)
Calibration tag
Flow direction arrow
Process connection
Sensor housing
Approvals tag
Ground screw
3
Figure 2. DS065, DS100, DH100, DS150, DH150, DS300, and DH300
sensors
Junction box
Flow direction arrow
Purge connection
(optional)
Calibration tag
Approvals tag
Ground screw
Figure 3. DT065, DT100, and DT150 sensors
Flow direction arrow
Lifting handle
Process connection
Sensor housing
Purge connection (optional)
Process connection
Calibration tag
Ground screw
Approvals tag
Sensor housing
Sensor cable
4
Figure 4. DL065, DL100, and DL200 sensors
Ground screw
Purge connection
(optional)
Sensor housing
Calibration tag
Approvals tag
Purge connection (optional)
Flow direction arrow
Process connection
Process connection
Junction box
STEP 1. Determining a location
The sensor may be located anywhere in the process line, as long as the following conditions are met:
Before operation, you must be able to stop flow through the sensor.
(During the zeroing procedure, flow must be stopped completely, and the sensor must be full of process fluid.)
During operation, the sensor must remain full of process fluid.
The sensor must be installed in an area that is compatible with the
classification specified on the sensor approvals tag (see Figures 1- 4).
Choose a location for the sensor based on the requirements described in this section.
5
Temperature limits
Process fluid temperature limits vary by sensor; refer to Table 2.
.
Table 2. Temperature specifications
Sensor type °F °C
DS025 DS040 DS065 DS100 DS150 DS150Z DS300 DS300Z DH025 DH038 DH100 DH150 DH300 DT065 DT100 DT150 DL065 DL100 DL200
–400 to +350 –240 to +177 –400 to +350 –240 to +177 –400 to +350 –240 to +177 –400 to +400 –240 to +204 –400 to +400 –240 to +204
+32 to +250 0 to +121
–400 to +400 –240 to +204
+32 to +250 0 to +121 –400 to +350 –240 to +177 –400 to +350 –240 to +177 –400 to +400 –240 to +204 –400 to +400 –240 to +204 –400 to +400 –240 to +204
+32 to +800 0 to +426
+32 to +800 0 to +426
+32 to +800 0 to +426 –400 to +350 –240 to +177 –400 to +350 –240 to +177 –400 to +400 –240 to +204
Process fluid temperature can be further restricted by ambient temperatures for ATEX approvals. For guidelines, go to www.micromotion.com/atex.
Maximum wiring distances
The maximum cable length between the sensor and transmitter is located in Table 3.
Table 3. Maximum cable lengths
Cable to flowmeter component Maximum length
9-wire to a Model 1700/2700 transmitter 60 feet (20 meters) 9-wire to all other transmitters 1000 feet (300 meters)
6
Hazardous area installations
For installation in an area that requires intrinsic safety, refer to Micro Motion UL, CSA, SAA, or ATEX documentation, shipped with the sensor or available from the Micro Motion web site.
For a complete list of hazardous area classifications for Micro Motion sensors, refer to the Expert
system at www.expert2.com.
2
STEP 2. Orienting the sensor
The sensor will function properly in any orientation if the sensor tubes remain filled with process fluid. Micro Motion recommends installing D, DT, and DL sensors as shown in Figure 5.
Flow direction arrow
The sensor has a flow direction arrow (see Figures 1-4) to help you configure the transmitter for flow direction. Process fluid flowing in the direction opposite to the flow direction arrow may cause unexpected transmitter output unless the transmitter is configured appropriately. For instructions on configuring the transmitter’s flow direction parameter, refer to the transmitter instruction manual.
7
Liq
Figure 5. Recommended orientations for sensors
Orientations for liquids: tubes down, horizontal pipeline
DS025, DH038, DS040 DS065, DS100, DS150,
DS300
Orientations for gases: tubes up, horizontal pipeline, self-draining
DS025, DH038, DS040 DS065, DS100, DS150 DT065, DT100, DT150
Orientations for slurries: flag mount, vertical pipeline
DS025, DH038, DS040 DS065, DS100, DS150,
DS300
DT065, DT100, DT150
DT065, DT100, DT150
Flow
Liquids, gases, slurries Self-draining Approved by 3A for sanitary applications
DL065, DL100, DL200
Flow
Flow
8
STEP 3. Mounting the sensor
D and DT sensors
Use your common practices to minimize torque and bending load on process connections.
Figure 6 illustrates how to mount a D and DT
sensor. If possible, install wiring with the conduit opening pointed down to reduce the risk of condensation or excessive moisture in the junction box.
Figure 6. Mounting a D and DT sensor
CAUTION
Using the sensor to support piping can damage the sensor or cause measurement error.
Do not use the sensor to support pipe.
9
DL sensors
To ensure optimal performance of the DL sensor, follow these guidelines:
Attach piping to a stable structure in at least one location downstream and at least one location upstream from the sensor. Put supports as close as possible to the process connections. Supports should be attached to the same structure. See Figure 7.
Figure 7. DL sensor installation
Approved for 3A sanitary applications
Mounting leg
Mounting leg - bolted to sensor base plate
Min. 2 foot (0,6 m) flex cable
Pipe clamps and supports
Drip leg
Pipe clamps and supports
Sanitary flow valve
If multiple sensors are installed in series or in parallel, piping to each sensor must have separate supports. Clamp at least two fixed pipe supports on piping between sensors installed in series.
Flexible piping may be installed to minimize piping vibration transmitted to the sensor. Piping immediately adjacent to the sensor's process fittings must be rigid.
10
After process piping has been installed and properly supported, refer to Figure 8 and follow these guidelines when installing the sensor:
Install the sensor with user-supplied mounting legs, bolts, and lock washers (six 1/4 inch-20 UNC for a DL65 or DL100 sensor; two 1/4 inch-20 UNC and four 3/8 inch-16 UNC for a DL200 sensor).
Use appropriate user-supplied gaskets between process fittings on the sensor and process piping.
Figure 8. Self-draining DL sensor installations for sanitary
applications
Mounted to floor
4" (101mm)
minimum
Level
Level
4" (101mm)
minimum
Mounted to
ceiling
Level
11
Mounted to
wall
Level
4" (101mm)
minimum
STEP 4. Wiring the sensor to the transmitter
WARNING
Failure to comply with requirements for intrinsic safety in a hazardous area could result in an explosion.
• Make sure the hazardous area specified on the sensor approvals tag is suitable for the environment in which the sensor is installed. See Figures 1-4.
• For installation in an area that requires intrinsic safety, refer to Micro Motion UL, CSA, SAA, or ATEX documentation, shipped with the sensor or available from the Micro Motion web site.
• For hazardous area installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
CAUTION
Failure to seal the sensor junction box housing and transmitter housing could cause a short circuit, which would result in measurement error or flowmeter failure.
• Ensure integrity of gaskets and o-rings.
• Install drip legs in conduit or cable.
• Seal all conduit openings.
12
Model D sensor junction box
Most Model D sensors are shipped with a junction box for wiring. For Model DT sensors, see Model DT conduit and junction box below. For Model D and DL sensors:
If not already installed, install the junction box on the sensor, following the wiring instructions on the junction box.
If possible, install wiring with the junction box opening pointed down, or with a drip leg in the conduit or cable, to reduce the risk of condensation or excessive moisture in the junction box. See Figure 9.
Next, follow the guidelines in 9-wire cable wiring to wire the sensor to the transmitter.
Figure 9. Model D sensor junction box
3/4" NPT female conduit opening
Drip leg in conduit or cable
Model DT conduit and junction box
DT sensors come with a 3-foot (1 meter) pigtail of pre-installed cable and a 3-foot (1 meter) piece of conduit that needs to be fitted over the pre­installed cable. Refer to Figure 10.
Slide the conduit over the pre-installed cable.
Screw the conduit fitting end into the sensor.
13
The other end of the conduit can be connected to a user-supplied junction box or directly to a transmitter.
If the conduit is connected to a user-supplied junction box, connect the wires to the terminals on the junction box. If possible, install wiring with junction box openings pointed down, or with drip legs in the conduit, to reduce the risk of condensation or excessive moisture in the junction box. Next, connect the 9-wire cable from the junction box to the transmitter by following the instructions in 9-wire cable wiring.
If the conduit is connected directly to a transmitter, refer to the wiring instructions in the transmitter Quick Reference Guide.
Figure 10. Model DT sensor wiring
3 ft. (1 m) factory-supplied flex conduit
Liquid tight to meet CE requirements
for European installations
1/2" NPT conduit
fitting
User-supplied junction box
Grounding screw
Drip leg
14
9-wire cable wiring
Follow the steps below to connect the 9-wire cable between the sensor and the transmitter.
1. Prepare the cable according to the instructions in Micro Motion’s
9-Wire Flowmeter Cable Preparation and Installation Guide.
2. Insert the stripped ends of the individual wires into the terminal
blocks. No bare wires should remain exposed.
For D and DL sensors, match the wires color for color. For wiring
at the transmitter, refer to the transmitter Quick Reference Guide.
For DT sensors, connect the sensor cable wires to the junction
box terminals. Then, connect the 9-wire cable to the junction box terminals, matching the colors of the wires to the numbers of the DT sensor terminals, as indicated in Table 4. For wiring at the transmitter, refer to the transmitter Quick Reference Guide.
3. Tighten the screws to hold the wires in place.
4. Ensure integrity of gaskets, then tightly close and seal the
junction box cover and all housing covers on the transmitter.
Table 4. DT sensor wiring to transmitter
DT sensor terminal number 9-wire cable colors
1Brown 2Red 3 Orange 4 Yellow 5 Green 6Blue 7 Violet 8Gray 9White
15
STEP 5. Grounding the sensor
CAUTION
Improper grounding could cause measurement error.
To reduce the risk of measurement error:
• Ground the flowmeter to earth, or follow ground network requirements for the facility.
• For installation in an area that requires intrinsic safety, refer to Micro Motion UL, CSA, SAA, or ATEX documentation, shipped with the sensor or available from the Micro Motion web site.
• For hazardous area installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
The sensor can be grounded via the piping if the joints in the pipeline are ground-bonded. If the sensor is not grounded via the piping, connect a ground wire to the sensor grounding screw (see Figures 1-4) or the junction box grounding screw.
If national standards are not in effect, follow these guidelines:
Use copper wire, 14 AWG (2,5 mm²) or larger wire size for grounding.
Keep all ground leads as short as possible, less than 1 ohm impedance.
Connect ground leads directly to earth, or follow plant standards.
Refer to the transmitter Quick Reference Guide for instructions on grounding the transmitter.
16
17
Visit us on the Internet at www.micromotion.com
Micro Motion United Kingdom
Emerson Process Management Limited Horsfield Way Bredbury Industrial Estate Stockport SK6 2SU U.K. T 0800 966 180 F 0800 966 181 www.emersonprocess.co.uk
Micro Motion Europe
Emerson Process Management Wiltonstraat 30 3905 KW Veenendaal The Netherlands T +31 (0) 318 495 670 F +31 (0) 318 495 689
Micro Motion Inc. USA
Worldwide Headquarters
7070 Winchester Circle Boulder, Colorado 80301 T (303) 530-8400
(800) 522-6277
F (303) 530-8459
Micro Motion
TM
Micro Motion Asia
Emerson Process Management 1 Pandan Crescent Singapore 128461 Republic of Singapore T (65) 6777-8211 F (65) 6770-8003
Micro Motion Japan
Emerson Process Management Shinagawa NF Bldg. 5F 1-2-5, Higashi Shinagawa Shinagawa-ku Tokyo 140-0002 Japan T (81) 3 5769-6803 F (81) 3 5769-6843
18
Guide condensé
Français Mars 2003
Notice d’installation des capteurs Micro Motion®
modèles D, DT et DL
Pour une aide technique en ligne, consultez notre système EXPERT parler à un technicien, appelez votre centre de service le plus proche :
En France, appelez le 01 49 79 74 96 ou, gratuitement, le 0800 917 901
En Suisse, appelez le 041-768 6111
™ sur internet à www.expert2.com. Si vous désirez
2
En Belgique, appelez le 02-716 77 11
Pour de l'assistance technique, appeler le +31 (0) 318 495 630
Aux Etats-Unis, appelez le 1-800-522-MASS (1-800-522-6277)
Au Canada et en Amérique Latine, appelez le (303) 530-8400
En Asie, appelez le (65) 6770-8155
Micro Motion
TM
19
AVANT DE COMMENCER
A propos de ce document
Ce guide condensé explique les principes de base d’installation des capteurs Micro Motion
®
modèles D (sauf D600), DL et DT. Pour des informations plus détaillées, consulter le manuel d’instructions qui a été livré avec le capteur.
Installations au sein de l’Union Européenne
Ce produit Micro Motion est conforme à toutes les directives européennes en vigueur s’il est installé conformément aux instructions de cette notice. Pour connaître la liste des directives qui s’appliquent à ce produit, consulter la déclaration de conformité CE.
La déclaration de conformité CE et le manuel contenant les instructions et schémas d’installation ATEX sont disponibles sur internet à www.micromotion.com/atex ou en contactant votre centre de service Micro Motion.
INTRODUCTION
Le capteur constitue l'un des deux éléments du débitmètre à effet Coriolis ; l'autre élément est le transmetteur.
Options d’installation
Les capteurs modèles D, DT et DL peuvent être raccordés aux transmetteurs Micro Motion indiqués au tableau 1.
©2003, Micro Motion, Inc. Tous droits réservés. Micro Motion est une marque déposée de Micro Motion, Inc. Les logos Micro Motion et Emerson sont des marques de Emerson Electric Co. Toutes les autres marques appartiennent à leurs propriétaires respectifs.
20
Tableau 1. Compatibilité des transmetteurs et des capteurs
Transmetteur Capteurs D et DL Capteurs DT
Modèle 1700/2700 (9 fils) X X Modèle 3500/3700 (9 fils) X X RFT9739 (7 ou 9 fils) X X IFT9701 (9 fils) X RFT9712 (7 ou 9 fils) X X
Les capteurs modèles D, DT et DL doivent être raccordés à un transmetteur déporté par l’intermédiaire d’une liaison 9 fils. Les éléments constitutifs du capteur sont indiqués aux figures 1, 2, 3 et 4 (remarque : ces figures ne sont pas à l’échelle).
Figure 1. Capteurs DS025, DH025, DH038 et DS040
Boîte de jonction
Connexions de purge
Vis de mise à la terre
(en option)
Plaque signalétique
d’étalonnage
Flèche indiquant le sens d’écoulement du fluide
Raccord
Boîtier
Plaque signalétique de certification
21
Figure 2. Capteurs DS065, DS100, DH100, DS150, DH150, DS300 et
DH300
Boîte de jonction
Flèche indiquant le sens d’écoulement du fluide
Connexion de purge
(en option)
Plaque signalétique
d’étalonnage
Plaque signalétique
de certification
Vis de mise à la terre
Figure 3. Capteurs DT065, DT100 et DT150
Flèche indiquant le sens
d’écoulement du fluide
Poignée de levage
Raccord
Boîtier
Connexion de purge (en option)
Raccord
Plaque signalétique
d’étalonnage
Vis de mise à la terre
22
Plaque signalétique de certification
Boîtier
Câble de liaison au transmetteur
Figure 4. Capteurs DL065, DL100 et DL200
Vis de mise à la
Connexions de purge
(en option)
Plaque signalétique de
Flèche indiquant le sens
d’écoulement du fluide
certification
Raccord
terre
Boîtier
Plaque signalétique d’étalonnage
Connexion de purge (en option)
Raccord
Boîte de jonction
Etape 1. Choix de l'emplacement
Le capteur peut être installé n’importe où sur la ligne du process, à condition que les points suivants soient respectés :
Une vanne d’arrêt parfaitement étanche doit être installée en aval du
capteur (au cours de la procédure d’auto-réglage du zéro, le débit doit être complètement arrêté et le capteur doit demeurer rempli de fluide).
Lorsque le débitmètre est en exploitation, les tubes du capteur doivent
toujours demeurer remplis du fluide du process.
Le capteur doit être installé dans une zone qui est compatible avec le
certificat de conformité indiqué sur la plaque signalétique de certification (voir les figures 1 à 4).
De plus, le choix de l’emplacement du capteur doit répondre aux contraintes suivantes :
Limites de température
Les limites de température du fluide dépendent du type de capteur. Voir le tableau 2.
23
.
Tableau 2. Limites de température du fluide
Capteur °F °C
DS025 DS040 DS065 DS100 DS150 DS150Z DS300 DS300Z DH025 DH038 DH100 DH150 DH300 DT065 DT100 DT150 DL065 DL100 DL200
–400 à +350 –240 à +177 –400 à +350 –240 à +177 –400 à +350 –240 à +177 –400 à +400 –240 à +204 –400 à +400 –240 à +204
+32 à +250 0 à +121
–400 à +400 –240 à +204
+32 à +250 0 à +121 –400 à +350 –240 à +177 –400 à +350 –240 à +177 –400 à +400 –240 à +204 –400 à +400 –240 à +204 –400 à +400 –240 à +204
+32 à +800 0 à +426
+32 à +800 0 à +426
+32 à +800 0 à +426 –400 à +350 –240 à +177 –400 à +350 –240 à +177 –400 à +400 –240 à +204
Pour le certificat de conformité ATEX, les limites de température du fluide dépendent aussi de la température ambiante. Pour plus d’informations, consulter le manuel d’installation ATEX disponible à www.micromotion.com/atex.
Longueur maximum du câble de liaison au transmetteur
La longueur totale du câble reliant le capteur au transmetteur ne doit pas excéder les valeurs indiquées au tableau 3.
Tableau 3. Longueur maximum du câble de liaison au transmetteur
Câble de liaison Longueur maximum
9 fils vers transmetteur modèle 1700/2700 20 mètres 9 fils vers tout autre transmetteur 300 mètres
24
Installations en atmosphère explosive
Pour les installations devant être conformes aux règles de sécurité intrinsèque, consulter le guide d’installation Micro Motion ATEX, UL, CSA ou SAA, livré avec le capteur ou disponible sur le site internet de Micro Motion.
Pour la liste détaillée des certificats de conformité des capteurs Micro Motion, consulter le système Expert
sur internet à www.expert2.com.
2
ETAPE 2. Orientation du capteur
Le capteur fonctionnera correctement dans n'importe quelle orientation à condition que les tubes de mesure restent constamment remplis du fluide à mesurer. Micro Motion recommande d’installer le capteur comme illustré à la figure 5.
Sens d'écoulement
Une flèche est gravée sur le corps du capteur pour permettre la configuration du paramètre « sens d’écoulement » du transmetteur (voir les figures 1-4). Si le fluide s'écoule dans la direction opposée à la flèche, les sorties du débitmètre risquent de ne pas réagir comme prévu si le débitmètre n'est pas configuré correctement. Pour configurer le paramètre « sens d’écoulement » du transmetteur, consulter le manuel d'instructions du transmetteur.
25
Liq
Figure 5. Orientations conseillées
Orientation conseillée pour les liquides propres: tubes vers le bas, ligne horizontale
DS025, DH038, DS040 DS065, DS100, DS150,
DS300
Orientation conseillée pour les gaz: tubes vers le haut, ligne horizontale, auto­vidangeable par gravité
DS025, DH038, DS040 DS065, DS100, DS150 DT065, DT100, DT150
Orientation conseillée pour les liquides chargés : montage en drapeau, ligne verticale
DS025, DH038, DS040 DS065, DS100, DS150,
DS300
DT065, DT100, DT150
DT065, DT100, DT150
Circulation ascendante
Liquides propres, gaz, liquides chargés Auto-vidangeable par gravité Autorisé 3A pour les applications sanitaires
DL065, DL100, DL200
26
Circulation ascendante
Circulation ascendante