Micro Motion® Model 3500 Transmitter
(MVD) or Model 3300 Peripheral
Installation manual for rack mount
Safety and approval information
This Micro Motion product complies with all applicable European directives when properly installed in accordance with the
instructions in this manual. Refer to the EC declaration of conformity for directives that apply to this product. The EC declaration of
conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions are available on
the internet at www.micromotion.com or through your local Micro Motion support center.
Information affixed to equipment that complies with the Pressure Equipment Directive can be found on the internet at
www.micromotion.com/documentation.
For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the transmitter
configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.micromotion.com/
documentation.
Return policy
Micro Motion procedures must be followed when returning equipment. These procedures ensure legal compliance with government
transportation agencies and help provide a safe working environment for Micro Motion employees. Failure to follow Micro Motion
procedures will result in your equipment being refused delivery.
Information on return procedures and forms is available on our web support system at www.micromotion.com, or by phoning the
Micro Motion Customer Service department.
Micro Motion customer service
Email:
• Worldwide: flow.support@emerson.com
• Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South AmericaEurope and Middle EastAsia Pacific
United States800-522-6277U.K.0870 240 1978Australia800 158 727
3.1Connect input and output wiring ........................................................................................... 13
3.2Connect the Model 3500 to the sensor .................................................................................. 13
3.3Wire the sensor to the remote core processor........................................................................ 19
3.4Connect the power supply wiring........................................................................................... 21
Installation Manual3
1Planning
Planning
This installation manual explains basic installation guidelines for mounting the Micro
Motion
rack.
For information on I.S. applications, refer to Micro Motion approval documentation.
For complete instructions about configuration, maintenance, and service, refer to the
instruction manual shipped with the transmitter.
Improper installation in a hazardous area can cause an explosion.
For information about hazardous applications, refer to the appropriate Micro Motion approval
documentation, shipped with the meter or available from the Micro Motion web site.
Hazardous voltage can cause severe injury or death.
Install transmitter and complete all wiring before supplying power.
Improper installation can cause measurement error or meter failure.
Follow all instructions.
®
Model 3300 or Model 3500 MVD applications platform in a 19-inch (486,2 mm)
WARNING!
WARNING!
CAUTION!
1.1Installation kit
For mounting in a rack, the Model 3300 or Model 3500 installation kit includes the
following parts:
•One DIN 41612/IEC 60603-2, Type D connector for input/output wiring, with solder
tails (Model 3300 only) or screw terminals
•(Model 3500 only) One DIN 41612/IEC 60603-2, keyed Type D connector for sensor
wiring with screw terminals
•One plug-in connector for power supply wiring
•Four Model 3300 or six Model 3500 slotted cheese-head machine screws, size M2.5x8,
for securing wiring connectors to the rack
Installation Manual5
Planning
1 U = 1 HE = 1.750 inches (44,5 mm)
3 U (3 HE)
3 U (3 HE)
3 U (3 HE)
1 U (1 HE)
1 U (1 HE)
1.2Choose a location
Choose a location for the transmitter based on the requirements described below.
1.2.1Environmental requirements
Install the transmitter where ambient temperature is between –4 and +140 °F (–20 and
+60 °C).
If multiple applications platforms are installed, provide at least 1 U (1 HE) of vertical space
between racks to ensure proper ventilation. See Figure 1-1.
Figure 1-1: Space requirements for proper ventilation
1.2.2Dimensions
6Model 3500 Transmitters or Model 3300 Peripherals - Rack Mount
The Model 3300 or Model 3500 has the following dimensions. See Figure 1-2:
•Height: 128 mm (3 U or 3 HE)
•Width: 142 mm (28 HP or 28 TE)
•Depth: 160 mm
The Model 3300 or Model 3500 conforms to DIN standard EN 60297-3-101 (IEC
60297-3-101) for 19-inch (486,2 mm) racks. Up to three enclosures fit into one rack. See
Figure 1-1.
Figure 1-2: Dimensions for mounting in rack
25 HP
(25 TE)
3 U
(3 HE)
4 13/16
(122,5)
5 1/16
(128,5)
4 x M2.5x11
1 U = 1 HE = 1.750 inches (44,5 mm)
1 HP = 1 TE = 0.200 inch (5,1 mm)
5 9/16
(142,2)
28 HP (28 TE)
Optional screw terminal
7 11/32
(186,7)
8 33/64
(216,2)
Approvals tag
Intrinsic safety shield
(Model 3500 only)
6 11/16
(169,9)
7 29/32
(200,6)
1
(25,4)
inches
(mm)
inches
(mm)
Rear rail for mounting connectors
19-inch (486,2 mm) rack
Planning
1.3Cable lengths
Maximum cable length from the sensor to the Model 3500 transmitter depends on the
installation type and cable type.
Installation typeMaximum cable length
4-wire remote transmitterFigure 1-3, and Table 1-1 for maximum length of the 4-wire
Remote core processor with remote
transmitter
cable
Figure 1-4, and Table 1-1 for maximum length of the 4-wire
cable and the 9-wire cable
Installation Manual7