
Reference Guide
MC-002152
2/2018
ELITE® Coriolis Flow Meters
Guidelines for Troubleshooting and Operation
of Provers with Micro Motion ELITE
®
Coriolis
Flow Meters
Micro Motion® ELITE® ow meters are high-
precision Coriolis ow meters that are often used
in the oil and gas industry in conjunction with
volume provers. These guidelines are designed
to aid in troubleshooting and operation of provers
that are in use together with Micro Motion Coriolis
meters. Proving results may vary if unstable
process conditions exist during proving.
Proving Methods and Proving Data Evaluation
The American Petroleum Industry (API) Manual
of Petroleum Measurement Standards (MPMS)
Chapter 4.8, Second Edition, Operation of Proving
Systems, Annex A, Evaluating Meter Proving
Data explains the relationship between the number
of proving runs, the observed repeatability, and the
random uncertainty of the resulting meter factor.
One important principal is that a lower meter factor
uncertainty will always result as more runs are
collected and averaged.
Prover Size and Selection for Fixed-Volume
Provers (Does Not Apply to Master Meter
Provers)
Important Note: The prover size should never
result in a pass time of less than 0.5 seconds or
a pre-run time of less than 0.25 seconds.
Prover Conditions
It is important to prove at conditions that are as
similar as possible to the expected operating
conditions. There are many conditions and factors
that can inuence the success of proving systems.
• Prover equipment and all supporting reference
measurement devices must be well-maintained
and veried to ensure measurement traceability,
reproducibility, and repeatability (API MPMS
Ch. 4 and Ch. 21.2, paragraph 2.11).
• Stability of ow rate, density, temperature, and
pressure is critical during proving. System design,
prover settings, and maintenance can all impact
ow rate stability during proving.
• The liquid inside the piping connecting the meter
to the prover should remain stable.
o
Minimize piping between the meter and
prover.
o
Avoid dead-end branches between
meter and prover that may act as a volume
“spring” with compressible uids.
Increasing the Base Prover Volume (BPV) will result in
passing repeatability requirements with fewer prover
passes. Increasing the number of passes will allow
for passing repeatability requirements with a smaller
prover. A decision may be made either to size the
prover with a smaller BPV (lower capital investment)
with a longer overall proving time, or with a larger BPV
with reduced runs (less long-term wear and tear) and
shorter overall prove time.
www.Emerson.com/MicroMotion

Reference Guide
MC-002152
2/2018
ELITE® Coriolis Flow Meters
• Flow pulsation from PD pumps, including lack
of back-pressure regulation, may inuence
repeatability and additional passes may be
needed to meet random uncertainty requirements.
• Enabling compensation for the effect of pressure
on the meter (consult the Transmitter Conguration and Use Manual) can improve repeatability
in applications where line pressure varies by more
than 30 psig during proving runs.
Meter Operation
When using smaller provers, it is important to
ensure that the meter is congured for optimum
ltering and speed of response.
1. Select the fastest speed of response available:
o
5700 transmitter: select “Low Filtering”
response mode
• Sufcient back pressure must be maintained on
both the prover and the meter to avoid vapor
breakout and to maintain a stable ow rate during
displacer launch and travel. Minimum recom-
mended back pressure is shown by Equation 4
(from API MPMS Ch. 5.6).
Equation 1: ρb ≥ 2 Δρ + 1.25 ρ
e
Where: ρb = Minimum back pressure (psig)
Δρ = Pressure drop across meter at max.
ow rate
ρe = Equilibrium vapor pressure at
operating temperature (psia)
• Accurate prover density measurement is crucial
when mass proving with a volumetric prover. The
following tolerances are advised when using a
pycnometer (API MPMS Ch. 14.6).
o Max. temperature difference = 0.2 °F
o Max. pressure difference = 1 psi
o Density Meter Factor (DMF) repeatability
should be 0.05% or better between consecutive
pycnometer tests
o
2700 transmitter: select “Special” for Update
Rate and either “Special” or “Low Filtering”
for Calculation Speed
2. Set ow damping to a value between 0.0 and
0.08 seconds.
3. Set density damping to 0.16 seconds.
4. Proving Wizard software is available from
Emerson to aid in preparing Micro Motion
Coriolis meters for proving
When the meter factor remains stable between
proving events, this indicates that the meter zero
setting value is good. A change in the meter factor
may or may not be related to the meter zero, so it is
important to always perform a Zero Verication Test
(consult Transmitter Conguration and Use manual)
before making any zero adjustments. Only adjust
the meter zero if advised to by the Zero Verication
Test. If a meter zero is adjusted, reprove the meter.
Coriolis ow meters do not have a xed K-factor
(number of pulses output per unit of ow). The
number of pulses output per unit of ow (e.g. pulses
per barrel) from a Coriolis meter is an adjustable
parameter that can be set to any desired value.
However, the frequency of pulses during the highest
ow rate must not exceed the pulse input capacity
of the prover pulse counting device.
Page 2