Guidelines for the Selection and Operation
of Provers with Micro Motion ELITE
®
Coriolis
Flow Meters
Micro Motion® ELITE® ow meters are high-
precision Coriolis ow meters that are often used
in the oil and gas industry in conjunction with
volume provers. These guidelines are designed to
aid in the selection of a prover size that will result
in consistent proving repeatability, while taking
into consideration the balance between:
• Maximum proving efciency
• Minimum prover size and cost
• Minimum prover wear and maintenance
Substantial experience from laboratory testing and
eld proving forms the basis of the prover size
recommendations and the Total Prove Time (TPT)
predictions in these guidelines. However, results
may vary if unstable process conditions exist
during proving.
It should be noted that these guidelines are based
on conservative estimates from the available data.
If repeatability requirements are already consistently being met, there is no need to change the
process or the prover size.
Proving Methods and Proving Data Evaluation
The American Petroleum Industry (API) Manual
of Petroleum Measurement Standards (MPMS)
Chapter 4.8, Second Edition, Operation of Proving
Systems, Annex A, Evaluating Meter Proving
Data explains the relationship between the number
of proving runs, the observed repeatability, and the
random uncertainty of the resulting meter factor.
One important principal is that a lower meter factor
uncertainty will always result as more runs are
collected and averaged.
Prover Sizing and Selection for FixedVolume Provers (Does Not Apply to Master
Meter Provers)
Important Note: The prover size should never
result in a pass time of less than 0.5 seconds or
a pre-run time of less than 0.25 seconds.
Total Prove Time (TPT) is dened as the total
accumulated amount of time during which the
prover displacer was travelling between the detector
switches. The minimum TPT that is needed to
achieve the target meter factor random uncertainty
can be used to size the prover.
Estimated minimum TPT values that may be
expected to pass repeatability requirements for
different meter sizes and ow rates are shown in
Table 1. The velocity of the uid as it travels through
the meter ow tubes is also shown in Table 1 in
units of feet per second (fps). Velocities above
60 fps are not recommended when proving with
a xed-volume prover and may result in excessive
TPT to pass repeatability. For applications above
60 fps, master meter proving is recommended
instead of xed-volume provers.
Equation 1 describes TPT:
www.Emerson.com/MicroMotion
Base Prover Volume(BPV)
𝑇𝑃𝑇
Flow rate
×(# 𝑜 𝑟𝑢𝑛𝑠 × (# 𝑜 𝑝𝑎𝑠𝑠𝑒𝑠 𝑝𝑒𝑟 𝑟𝑢𝑛
Reference Guide
MC-001597 Rev E
1/2018
ELITE® Coriolis Flow Meters
To use Table 1 to size a prover, there are two
methods to select from:
Method 1 (Determine the BPV): Find the minimum
TPT value from Table 1. Multiply the ow rate by
0.0117 to convert from BPH to gallons per second.
Then enter the TPT, ow rate, the number of runs,
and the passes per run (if averaging multiple passes
per run) into Equation 2 to nd the minimum BPV
needed in gallons.
Equation 2:
𝐵𝑃
TPT x Flow Rate
(# 𝑜 R𝑢𝑛𝑠×(Passes pe Run
Example: CMFHC4 meter at 6500 BPH
• From Table 1: Velocity ≈ 30 fps. TPT =
20 seconds.
• Convert: 6500 BPH X 0.0117 = 76 gallons
per second
• If 5 single-pass runs are required, BPV =
(20 seconds X 76 gallons per second)
÷ (5 runs X 1 pass per run) = 304 gallons
or 7.2 BBLS.
• If 10 runs and 3 passes per run are acceptable,
BPV = (20 seconds X 76 gallons per second) ÷
(10 runs X 3 pass per run) = 50 gallons or
1.2 BBLS.
Method 2 (Determine the Number of Passes
Needed): Find the minimum TPT value from
Table 1. Multiply the ow rate by 0.0117 to convert
from BPH to gallons per second. Then, insert
the ow rate and a BPV in gallons into Equation
3 to estimate the total number of passes that
will be needed for a prover size with that BPV. If
averaging multiple passes per run, divide the total
number of passes needed by the number of runs
and round up to determine the minimum number
of passes per run needed. For single-pass runs,
the number of runs needed will equal the total
number of passes needed.
Equation 3:
𝑇𝑜𝑡𝑎 # 𝑜 𝑃𝑎𝑠𝑠𝑒𝑠 𝑇𝑃𝑇×
Flow rate
Base Prover Volume(BPV)
Example: CMF400 meter at 2300 BPH
• From Table 1: Velocity ≈ 40 fps. TPT ≈
30 seconds.
• Convert: 2300 BPH X 0.0117 = 27 gallons
per second
• If the BPV is 170 gallons (4 BBLS), Total #
of passes = 30 seconds X 27 gallons per
second ÷ 170 gallons = 5 total passes
(5 total passes ÷ 5 runs = 1 pass per run
for 5 runs).
Page 2
• If the BPV is 65 gallons (1.55 BBLS), Total # of
passes = 30 seconds X 27 gallons per second
÷ 65 gallons = 12 total passes
(12 total passes ÷ 5 runs = 3 passes per
run for 5 runs).
Reference Guide
MC-001597 Rev E
1/2018
ELITE® Coriolis Flow Meters
Table 1. Estimated Minimum Total Prove Time (TPT)
Values shown are the velocity inside the meter in feet per second (fps)
followed by the Total Prove Time (TPT) in seconds.
TPT = prover pass time X passes per run X runs per proving*
(*total number of runs required to reach ±0.027% meter factor uncertainty may vary.)
Page 3
Reference Guide
MC-001597 Rev E
1/2018
ELITE® Coriolis Flow Meters
• The liquid inside the piping connecting the meter
to the prover should remain stable.
o
Minimize piping between the meter and
prover.
o
Avoid dead-end branches between
meter and prover that may act as a volume
“spring” with compressible uids.
• Sufcient back pressure must be maintained on
both the prover and the meter to avoid vapor
breakout and to maintain a stable ow rate during
displacer launch and travel. Minimum recommended back pressure is shown by Equation 4
(from API MPMS Ch. 5.6).
Table 1 (cont.) Estimated Minimum Total Prove Time (TPT)
Increasing the BPV will result in meeting the mini-
mum TPT with fewer passes. Increasing the number
of passes will allow reaching the minimum TPT with
a smaller prover. A decision may be made either to
size the prover with a smaller BPV (lower capital
investment) with a longer overall proving time, or
with a larger BPV with reduced runs (less long-term
wear and tear) and shorter overall prove time.
Prover Conditions
It is important to prove at conditions that are as
similar as possible to the expected operating conditions. There are many conditions and factors that
can inuence the success of proving systems.
• Prover equipment and all supporting reference
measurement devices must be well-maintained
and veried to ensure measurement traceability,
reproducibility, and repeatability (API MPMS Ch. 4
and Ch. 21.2, paragraph 2.11).
• Stability of ow rate, density, temperature, and
pressure is critical during proving. System design,
prover settings, and maintenance can all impact
ow rate stability during proving.
Equation 4:ρb ≥ 2 Δρ + 1.25 ρ
e
Where: ρb = Minimum back pressure (psig)
ρ = Pressure drop across meter at max.
Δ
ow rate
ρ
= Equilibrium vapor pressure at
e
operating temperature (psia)
• Accurate prover density measurement is crucial
when mass proving with a volumetric prover.
The following tolerances are advised when using
a pycnometer (API MPMS Ch. 14.6).
o
Max. temperature difference = 0.2 °F
o
Max. pressure difference = 1 psi
o
Density Meter Factor (DMF) repeatability
should be 0.05% or better between consecutive
pycnometer tests
• Flow pulsation from PD pumps, including lack
of back-pressure regulation, may inuence
repeatability and additional passes may be
needed to meet random uncertainty requirements.
• Enabling compensation for the effect of pressure
on the meter (consult the Transmitter Congura-
tion and Use Manual) can improve repeatability
in applications where line pressure varies by
more than 30 psig during proving runs.
Page 4
Reference Guide
MC-001597 Rev E
1/2018
ELITE® Coriolis Flow Meters
API MPMS Chapter 5.6 Measurement of Liquid
Hydrocarbons by Coriolis Meters denes a pulse
scaling factor (PSF) as the number of pulses output
by a Coriolis meter per unit of ow. In this way, a
Coriolis meter PSF is like a mechanical meter’s
K-factor, but adjustable.
Micro Motion transmitters typically provide a frequency output of up to 10,000 Hz (10,000 pulses per
second). Many eld device pulse input specications
have a similar maximum pulse input rate of 10,000
Hz. However, some eld devices have a maximum
pulse input rate that is less than 10,000 Hz. The
PSF of the Coriolis owmeter can be adjusted to
stay within the constraints of any eld device that is
counting the pulses from the meter per the equations
shown here:
Meter Operation
When using smaller provers, it is important to ensure
that the meter is congured for optimum ltering and
speed of response.
1. Select the fastest speed of response available:
o
5700 transmitter: select “Low Filtering”
response mode
o
2700 transmitter: select “Special” for Update
Rate and either “Special” or “Low Filtering”
for Calculation Speed
2. Set ow damping to a value between 0.0 and
0.08 seconds.
3. Set density damping to 0.16 seconds.
• Proving Wizard software is available from
Emerson to aid in preparing Micro Motion
Coriolis meters for proving
Coriolis owmeters do not have a xed K-factor
(number of pulses output per unit of ow). The
number of pulses output per unit of ow (e.g. pulses
per barrel) from a Coriolis meter is an adjustable
parameter that can be set to any desired value.
However, the frequency of pulses during the highest
ow rate must not exceed the pulse input capacity
of the prover pulse counting device.
When the meter factor remains stable between
proving events, this indicates that the meter zero
setting value is good. A change in the meter factor
may or may not be related to the meter zero, so it is
important to always perform a Zero Verication Test
(consult Transmitter Conguration and Use manual)
before making any zero adjustments. Only adjust the
meter zero if advised to by the Zero Verication Test.
If a meter zero is adjusted, reprove the meter.
Page 5
Reference Guide
MC-001597 Rev E
1/2018
Master Meter Proving
Micro Motion Coriolis meters can be used as master
meters per API MPMS Ch. 4.5 for proving with the
following advantages:
• A Coriolis master meter can be used to prove in
either volume and/or mass units.
• Pass duration can be lengthened to improve
repeatability.
• Maintaining stable process conditions is much
easier with no effects due to a displacer launch.
• Low maintenance and high reliability, with no
seals or moving parts.
ELITE® Coriolis Flow Meters
Need More Information?
Emerson has extensive eld experience in mass
and volume proving of our Micro Motion Coriolis meters.
Contact us at 1-800-522-6277 or visit our website at
www.MicroMotion.com
Emerson Automation Solutions
7070 Winchester Circle
Boulder, CO USA 80301
T +1 303 527 6277
F +1 303 530 8459