Micro Motion Fork Density Meter Manuals & Guides

Configuration and Use Manual
MMI-20020959, Rev AC
Micro Motion® Fork Density Meters (FDM)
Configuration and Use Manual
April 2016

Safety messages

Emerson Flow customer service

Email:
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 704 136 666 New Zealand 099 128 804
Mexico +41 (0) 41 7686 111 France 0800 917 901 India 800 440 1468
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy 8008 77334 China +86 21 2892 9000
Venezuela +58 26 1731 3446 Central & Eastern +41 (0) 41 7686 111 Japan +81 3 5769 6803
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684

Contents

Contents
Part I Getting Started
Chapter 1 Before you begin ............................................................................................................3
1.1 About this manual ....................................................................................................................... 3
1.2 Model codes and device types ..................................................................................................... 3
1.3 Communications tools and protocols .......................................................................................... 4
1.4 Additional documentation and resources .................................................................................... 5
Chapter 2 Quick start .....................................................................................................................7
2.1 Power up the transmitter .............................................................................................................7
2.2 Check meter status ......................................................................................................................7
2.3 Make a startup connection to the transmitter ..............................................................................8
Part II Configuration and commissioning
Chapter 3 Introduction to configuration and commissioning ....................................................... 13
3.1 Default values ............................................................................................................................13
3.1.1 FDM default values ..................................................................................................... 13
3.2 Enable access to the off-line menu of the display ....................................................................... 14
3.3 Disable HART security ................................................................................................................15
3.4 Set the HART lock ...................................................................................................................... 17
3.5 Restore the factory configuration .............................................................................................. 17
Chapter 4 Configure process measurement ..................................................................................19
4.1 Verify the calibration factors ......................................................................................................19
4.1.1 Calibration factors ...................................................................................................... 20
4.2 Configure line density measurement ........................................................................................ 20
4.2.1 Configure Density Measurement Unit ................................................................................20
4.2.2 Configure Density Damping ........................................................................................... 22
4.2.3 Configure Density Cutoff ................................................................................................ 23
4.2.4 Configure two-phase flow parameters ........................................................................24
4.3 Configure temperature measurement .......................................................................................25
4.3.1 Configure Temperature Measurement Unit ........................................................................ 25
4.3.2 Configure Temperature Damping .....................................................................................26
4.3.3 Configure Temperature Input .......................................................................................... 27
4.4 Configure the pressure input ..................................................................................................... 30
4.4.1 Configure the pressure input using ProLink III ............................................................. 30
4.4.2 Configure the pressure input using the Field Communicator .......................................32
4.4.3 Options for Pressure Measurement Unit ........................................................................... 33
4.5 Set up the API referral application ..............................................................................................34
4.5.1 Set up the API referral application using ProLink III ...................................................... 34
4.5.2 Set up the API referral application using the Field Communicator ............................... 40
4.6 Set up concentration measurement .......................................................................................... 47
4.6.1 Preparing to set up concentration measurement ........................................................47
4.6.2 Set up concentration measurement using ProLink III ...................................................48
Configuration and Use Manual i
Contents
4.6.3 Set up concentration measurement using the Field Communicator ............................55
4.6.4 Using equations to calculate specific gravity, °Baumé, °Brix, °Plato, and °Twaddell ......60
4.6.5 Matrix switching ......................................................................................................... 61
4.6.6 Measuring Net Mass Flow Rate and Net Volume Flow Rate ..........................................63
4.7 Set up flow rate measurement ...................................................................................................63
4.7.1 Set up flow rate measurement using ProLink III ...........................................................63
4.7.2 Set up flow rate measurement using the Field Communicator .................................... 65
Chapter 5 Configure device options and preferences ....................................................................67
5.1 Configure the transmitter display .............................................................................................. 67
5.1.1 Configure the language used for the display ............................................................... 67
5.1.2 Configure the process variables and diagnostic variables shown on the display ...........68
5.1.3 Configure the number of decimal places (precision) shown on the display ..................68
5.1.4 Configure the refresh rate of data shown on the display ..............................................69
5.1.5 Enable or disable automatic scrolling through the display variables ............................ 69
5.2 Enable or disable the Acknowledge All Alerts display command ......................................................70
5.3 Configure security for the display menus .................................................................................. 71
5.4 Configure alert handling ............................................................................................................72
5.4.1 Configure Fault Timeout .................................................................................................72
5.4.2 Configure Alert Severity ................................................................................................. 73
5.5 Configure informational parameters ......................................................................................... 75
Chapter 6 Integrate the meter with the control system ................................................................77
6.1 Configure Channel B ..................................................................................................................77
6.2 Configure the mA output .......................................................................................................... 78
6.2.1 Configure mA Output Process Variable ............................................................................. 78
6.2.2 Configure Lower Range Value (LRV) and Upper Range Value (URV) ....................................80
6.2.3 Configure Added Damping ............................................................................................. 81
6.2.4 Configure mA Output Fault Action and mA Output Fault Level ...............................................82
6.3 Configure the discrete output ................................................................................................... 84
6.3.1 Configure Discrete Output Source ....................................................................................84
6.3.2 Configure Discrete Output Polarity ................................................................................... 85
6.3.3 Configure Discrete Output Fault Action ............................................................................. 86
6.4 Configure an enhanced event ....................................................................................................87
6.5 Configure HART/Bell 202 communications ............................................................................... 88
6.5.1 Configure basic HART parameters ...............................................................................88
6.5.2 Configure HART variables (PV, SV, TV, QV) ..................................................................89
6.5.3 Configure burst communications ............................................................................... 91
6.6 Configure Modbus communications ..........................................................................................95
6.7 Configure Digital Communications Fault Action ................................................................................. 97
6.7.1 Options for Digital Communications Fault Action .................................................................97
Chapter 7 Complete the configuration ......................................................................................... 99
7.1 Test or tune the system using sensor simulation ........................................................................99
7.2 Back up transmitter configuration ............................................................................................. 99
7.3 Enable HART security ...............................................................................................................100
Part III Operations, maintenance, and troubleshooting
Chapter 8 Transmitter operation ................................................................................................105
8.1 Record the process variables ................................................................................................... 105
8.2 View process variables and diagnostic variables ...................................................................... 105
ii Micro Motion® Fork Density Meters (FDM)
Contents
8.2.1 View process variables using the display ................................................................... 105
8.2.2 View process variables and other data using ProLink III ............................................. 106
8.2.3 View process variables using the Field Communicator .............................................. 106
8.3 View and acknowledge status alerts ........................................................................................ 107
8.3.1 View and acknowledge alerts using the display ........................................................ 107
8.3.2 View and acknowledge alerts using ProLink III ...........................................................109
8.3.3 View alerts using the Field Communicator ................................................................ 110
8.3.4 Alert data in transmitter memory ............................................................................. 110
Chapter 9 Measurement support ............................................................................................... 113
9.1 Perform the inline calibration check procedure ....................................................................... 113
9.1.1 Perform the inline calibration check using ProLink III .................................................114
9.1.2 Perform the inline calibration check using the Field Communicator ..........................115
9.2 Perform the Known Density Verification procedure .................................................................115
9.2.1 Perform the Known Density Verification procedure using the display ....................... 116
9.2.2 Perform the Known Density Verification procedure using ProLink III ......................... 117
9.2.3 Perform the Known Density Verification procedure using the
Field Communicator ................................................................................................. 117
9.3 Adjust density measurement with Density Offset or Density Meter Factor ....................................... 118
9.4 Perform density offset calibration ............................................................................................119
9.4.1 Perform density offset calibration using the display .................................................. 120
9.4.2 Perform density offset calibration using ProLink III ....................................................121
9.4.3 Perform density offset calibration using the Field Communicator ............................. 121
9.5 Perform temperature calibration .............................................................................................122
9.5.1 Perform temperature calibration using the display ................................................... 122
9.5.2 Perform temperature calibration using ProLink III .................................................... 123
9.5.3 Perform temperature calibration using the Field Communicator ..............................123
9.6 Adjust concentration measurement with Trim Offset .................................................................124
9.7 Adjust concentration measurement with Trim Slope and Trim Offset ........................................... 125
9.8 Set up user-defined calculations .............................................................................................. 127
9.8.1 Equations used in user-defined calculations ..............................................................129
9.8.2 Measurement units used in user-defined calculations ............................................... 129
Chapter 10 Troubleshooting ........................................................................................................ 131
10.1 Quick guide to troubleshooting ...............................................................................................131
10.2 Check power supply wiring ......................................................................................................132
10.3 Check grounding .....................................................................................................................133
10.4 Perform loop tests ...................................................................................................................133
10.4.1 Perform loop tests using the display ......................................................................... 134
10.4.2 Perform loop tests using ProLink III ........................................................................... 135
10.4.3 Perform loop tests using the Field Communicator .................................................... 136
10.5 Status LED states ..................................................................................................................... 137
10.6 Status alerts, causes, and recommendations ........................................................................... 138
10.7 Density measurement problems ............................................................................................. 143
10.8 Temperature measurement problems .....................................................................................144
10.9 API referral problems ...............................................................................................................145
10.10 Concentration measurement problems ...................................................................................145
10.11 Milliamp output problems ....................................................................................................... 146
10.12 Discrete output problems ........................................................................................................148
10.13 Time Period Signal (TPS) output problems ...............................................................................148
10.14 Using sensor simulation for troubleshooting ........................................................................... 149
10.15 Trim mA outputs ..................................................................................................................... 149
Configuration and Use Manual iii
Contents
10.15.1 Trim mA outputs using ProLink III ..............................................................................149
10.15.2 Trim mA outputs using the Field Communicator .......................................................150
10.16 Check HART communications ................................................................................................. 151
10.17 Check Lower Range Value and Upper Range Value ......................................................................... 152
10.18 Check mA Output Fault Action ...................................................................................................... 153
10.19 Check for radio frequency interference (RFI) ............................................................................153
10.20 Check the cutoffs .................................................................................................................... 153
10.21 Check for two-phase flow (slug flow) .......................................................................................154
10.22 Check the drive gain ................................................................................................................ 154
10.22.1 Collect drive gain data .............................................................................................. 155
10.23 Check for internal electrical problems ..................................................................................... 155
10.24 Locate a device using the HART 7 Squawk feature ................................................................... 156
Appendices and reference
Appendix A Calibration certificate ................................................................................................ 157
A.1 Sample calibration certificate ................................................................................................. 157
Appendix B Using the transmitter display ..................................................................................... 159
B.1 Components of the transmitter interface ................................................................................ 159
B.2 Use the optical switches .......................................................................................................... 159
B.3 Access and use the display menu system .................................................................................160
B.3.1 Enter a floating-point value using the display ............................................................161
B.4 Display codes for process variables ..........................................................................................164
B.5 Codes and abbreviations used in display menus ...................................................................... 164
Appendix C Using ProLink III with the transmitter .........................................................................177
C.1 Basic information about ProLink III ...........................................................................................177
C.2 Connect with ProLink III ........................................................................................................... 178
C.2.1 Connection types supported by ProLink III ................................................................ 178
C.2.2 Connect with ProLink III over Modbus/RS-485 ...........................................................179
C.2.3 Connect with ProLink III over HART/Bell 202 ............................................................. 182
Appendix D Using the Field Communicator with the transmitter ................................................... 191
D.1 Basic information about the Field Communicator ....................................................................191
D.2 Connect with the Field Communicator .................................................................................... 192
Appendix E Concentration measurement matrices, derived variables, and process variables ........ 195
E.1 Standard matrices for the concentration measurement application ........................................ 195
E.2 Concentration measurement matrices available by order ........................................................196
E.3 Derived variables and calculated process variables .................................................................. 198
iv Micro Motion® Fork Density Meters (FDM)
Part I
Getting Started
Chapters covered in this part:
Before you begin
Quick start
Getting Started
Configuration and Use Manual 1
Getting Started
2 Micro Motion® Fork Density Meters (FDM)
1 Before you begin
Topics covered in this chapter:

About this manual

Model codes and device types
Communications tools and protocols
Additional documentation and resources
1.1 About this manual
This manual provides information to help you configure, commission, use, maintain, and troubleshoot the Micro Motion Fork Density Meter (FDM).
The following versions of the FDM are documented in this manual:
Fork Density Meter with Analog Outputs
Fork Density Meter with Analog Output and Discrete Output
Fork Density Meter with Time Period Signal Output

Before you begin

For the Fork Density Meter with FOUNDATION™ Fieldbus, see Micro Motion® Fork Density Meters with FOUNDATION™ Fieldbus: Configuration and Use Manual.
Important
This manual assumes that the following conditions apply:
The meter has been installed correctly and completely, according to the instructions in the
installation manual.
The installation complies with all applicable safety requirements.
The user is trained in all government and corporate safety standards.
1.2 Model codes and device types
Your device can be identified by the model code on the device tag.

Model codes and device typesTable 1-1:

Model code Device nickname I/O
FDM*****C FDM mA Two mA outputs
RS-485 terminals
FDM*****D FDM DO One mA output
One discrete output
RS-485 terminals
Electronics mount­ing
Integral
Integral
Configuration and Use Manual 3
Before you begin
Model codes and device types (continued)Table 1-1:
Model code Device nickname I/O
FDM*****B FDM TPS One mA output
One Time Period Sig-
nal output
FDM*****A FDM FF FOUNDATION™ field-
bus
Restriction
The FDM and FDM FF support a complete set of application and configuration options. The FDM DO and FDM TPS support a subset of configuration options. Refer to the product data sheet for details.

1.3 Communications tools and protocols

You can use several different communications tools and protocols to interface with the device. You may use different tools in different locations or for different tasks.
Electronics mount­ing
Integral
4-wire remote transmitter
Communications tools, protocols, and related informationTable 1-2:
Communica­tions tool Supported protocols Scope In this manual For more information
Display Not applicable Basic configuration and
commissioning
ProLink III Modbus/RS-485
HART/Bell 202
Service port
Field Commu­nicator
HART/Bell 202 Complete configuration
Tip
You may be able to use other communications tools from Emerson Process Management, such as AMS Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this manual. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at www.micromotion.com.
Complete configuration and commissioning
and commissioning
Complete user informa­tion. See Appendix B.
Basic user information. See Appendix C.
Basic user information. See Appendix D.
Not applicable
User manual
Installed with soft-
ware
On Micro Motion
user documentation CD
On Micro Motion
web site (www.micromo‐
tion.com)
User manual on Micro Motion web site (www.micromo‐
tion.com )
4 Micro Motion® Fork Density Meters (FDM)
1.4 Additional documentation and resources
Micro Motion provides additional documentation to support the installation and operation of the device.

Additional documentation and resourcesTable 1-3:

Topic Document
Device installation Micro Motion Fork Density Meters (FDM): Installation Manual
Product data sheet Micro Motion Fork Density Meters: Product Data Sheet
All documentation resources are available on the Micro Motion web site at
www.micromotion.com or on the Micro Motion user documentation DVD.
Before you begin
Configuration and Use Manual 5
Before you begin
6 Micro Motion® Fork Density Meters (FDM)
2 Quick start
Topics covered in this chapter:

Power up the transmitter

Check meter status

Make a startup connection to the transmitter
2.1 Power up the transmitter
The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement.
1. Ensure that all transmitter and sensor covers and seals are closed.
WARNING!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are tightly closed. For hazardous area installations, applying power while housing covers are removed or loose can cause an explosion.

Quick start

2. Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. During this period, Alert 009 is active. The diagnostic routines should complete in approximately 30 seconds.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power has been off long enough to allow components to reach ambient temperature, allow the electronics to warm up for approximately 10 minutes before relying on process measurements. During this warm-up period, you may observe minor measurement instability or inaccuracy.
2.2 Check meter status
Check the meter for any error conditions that require user action or that affect measurement accuracy.
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alert A009 is active. This alert should clear automatically when the power-up sequence is complete.
2. Check the status LED on the transmitter.
Configuration and Use Manual 7
Quick start
Transmitter status reported by status LEDTable 2-1:
LED state Description Recommendation
Green No alerts are active. Continue with configuration or process meas-
urement.
Yellow One or more low-severity alerts are active. A low-severity alert condition does not affect
measurement accuracy or output behavior. You can continue with configuration or proc­ess measurement. If you choose, you can iden­tify and resolve the alert condition.
Flashing yellow Calibration in progress, or Known Density Veri-
fication in progress.
Red One or more high-severity alerts are active. A high-severity alert condition affects meas-
The measurement can fluctuate during the calibration process or change as a result of the calibration process. The alert will clear when the calibration is complete. Check the calibra­tion results before continuing.
urement accuracy and output behavior. Re­solve the alert condition before continuing.
View and acknowledge status alerts (Section 8.3)
Status alerts, causes, and recommendations (Section 10.6)

2.3 Make a startup connection to the transmitter

For all configuration tools except the display, you must have an active connection to the transmitter to configure the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in the appropriate appendix. Use the default communications parameters shown in the appendix.
Communications tool Connection type to use Instructions
ProLink III Modbus/RS-485
HART/Bell 202
Field Communicator HART/Bell 202 Appendix D
Postrequisites
(Optional) Change the communications parameters to site-specific values.
Appendix C
To change the communications parameters using ProLink III, choose Device Tools >
Configuration > Communications.
To change the communications parameters using the Field Communicator, choose
Configure > Manual Setup > HART > Communications.
8 Micro Motion® Fork Density Meters (FDM)
Quick start
Important
If you are changing communications parameters for the connection type that you are using, you will lose the connection when you write the parameters to the transmitter. Reconnect using the new parameters.
Configuration and Use Manual 9
Quick start
10 Micro Motion® Fork Density Meters (FDM)
Configuration and commissioning
Part II
Configuration and commissioning
Chapters covered in this part:
Introduction to configuration and commissioning
Configure process measurement
Configure device options and preferences
Integrate the meter with the control system
Complete the configuration
Configuration and Use Manual 11
Configuration and commissioning
12 Micro Motion® Fork Density Meters (FDM)

Introduction to configuration and commissioning

3 Introduction to configuration and
commissioning
Topics covered in this chapter:

Default values

Enable access to the off‐line menu of the display
Disable HART security
Set the HART lock
Restore the factory configuration
3.1 Default values
Default values for your meter are configured at the factory.
Important
Default values are based on your purchase order options. Therefore, the default values described in the following tables may not be the factory default values configured for your system. For absolute accuracy, refer to the configuration sheet that was shipped with your meter.
3.1.1 FDM default values
FDM default mA scaling valuesTable 3-1:
Variable Default 4 mA Default 20 mA
Density 0.500 g/cc 1.500 g/cc
Temperature -50.000°C
Drive gain 0.000 % 100.000 %
External temperature -50.000°C
External pressure 0.000 PSIg 1450.377 PSIg
Sensor time period 400 us 2900 us
Special equation output 0 100
API Referral option enabled
API density 0.500 g/cc 1.500 g/cc
-58.000°F
-58.000°F
200.000°C
392.000°F
200.000°C
392.000°F
Concentration Measurement option enabled
CM density @ ref 0.500 g/cc 1.500 g/cc
CM specific gravity 0.500 1.500
Configuration and Use Manual 13
Introduction to configuration and commissioning
FDM default mA scaling values (continued)Table 3-1:
Variable Default 4 mA Default 20 mA
CM concentration 0.000 % 100.000 %
External volume flow rate input enabled
Mass flow rate (calculated) -0.2 kg/s 0.2 kg/s
Volume flow rate (external) -0.2 l/s 0.2 l/s
Net mass flow -0.2 kg/s 0.2 kg/s
Net volume flow -0.2 l/s 0.2 l/s
Mass flow (mag input) -0.2 kg/s 0.2 kg/s
FDM default variablesTable 3-2:
Default variable Output option A Output options B and C
Primary Variable (PV), mA1 Sample Temperature Density
Secondary Variable (SV), mA2
Tertiary Variable (TV) Density Sensor Time Period
Quaternary Variable (QV) Drive Gain Drive Gain
Sensor Time Period Temperature

3.2 Enable access to the off-line menu of the display

Display Not available
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Display Menus > Offline Menu
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must enable it if you want to use the display to configure the transmitter.
Restriction
You cannot use the display to enable access to the off-line menu. You must make a connection from another tool.
14 Micro Motion® Fork Density Meters (FDM)

3.3 Disable HART security

A
If you plan to use HART protocol to configure the device, HART security must be disabled. HART security is disabled by default, so you may not need to do this.
Prerequisites
Strap wrench
3 mm hex key
Procedure
1. Power down the meter.
2. Using the strap wrench, loosen the grub screws and remove the transmitter end-
cap.
Transmitter with end-cap removedFigure 3-1:
Introduction to configuration and commissioning
A. Transmitter end‐cap
3. Using the hex key, remove the safety spacer.
Configuration and Use Manual 15
A
B
A
B
Introduction to configuration and commissioning
Transmitter with end-cap and safety spacer removedFigure 3-2:
A. Transmitter end‐cap B. Safety spacer
4. Move the HART security switch to the OFF position (up).
The HART security switch is the switch on the left.
HART security switchFigure 3-3:
A. HART security switch B. Unused
16 Micro Motion® Fork Density Meters (FDM)
5. Replace the safety spacer and end-cap.
6. Power up the meter.

3.4 Set the HART lock

If you plan to use a HART connection to configure the device, you can lock out all other HART masters. If you do this, other HART masters will be able to read data from the device but will not be able to write data to the device.
Restrictions
This feature is available only when you are using the Field Communicator or AMS.
This feature is available only with a HART 7 host.
Procedure
1. Choose Configure > Manual Setup > Security > Lock/Unlock Device.
2. If you are locking the meter, set Lock Option as desired.
Introduction to configuration and commissioning
Option Description
Permanent Only the current HART master can make changes to the device. The device will
remain locked until manually unlocked by a HART master. The HART master can also change Lock Option to Temporary.
Temporary Only the current HART master can make changes to the device. The device will
remain locked until manually unlocked by a HART master, or a power-cycle or device reset is performed. The HART master can also change Lock Option to Perma-
nent.
Lock All No HART masters are allowed to make changes to the configuration. Before
changing Lock Option to Permanent or Temporary, the device must be unlocked. Any HART master can be used to unlock the device.
Postrequisites
To avoid future confusion or difficulties, ensure that the device is unlocked after you have completed your tasks.

3.5 Restore the factory configuration

Display Not available
ProLink III Device Tools > Configuration Transfer > Restore Factory Configuration
Field Communicator Service Tools > Maintenance > Reset/Restore > Restore Factory Configuration
Configuration and Use Manual 17
Introduction to configuration and commissioning
Overview
Restoring the factory configuration returns the transmitter to a known operational configuration. This may be useful if you experience problems during configuration.
Tip
Restoring the factory configuration is not a common action. You may want to contact Micro Motion to see if there is a preferred method to resolve any issues.
18 Micro Motion® Fork Density Meters (FDM)

Configure process measurement

4 Configure process measurement
Topics covered in this chapter:

Verify the calibration factors

Configure line density measurement
Configure temperature measurement
Configure the pressure input
Set up the API referral application
Set up concentration measurement
Set up flow rate measurement
4.1 Verify the calibration factors
Display Not available
ProLink III Device Tools > Calibration Data
Field Communicator Configure > Manual Setup > Calibration Factors
Overview
The calibration factors are used to adjust measurement for the unique traits of the sensor. Your device was calibrated at the factory. However, you should verify that the calibration factors that are configured in your device match the factory values.
Prerequisites
You will need the factory values for the calibration factors. These are provided in two locations:
The calibration certificate shipped with your meter
The label inside the transmitter's end-cap
Important
If the transmitter is not the original component, do not use the values from the transmitter label.
Procedure
1. View the calibration factors that are stored in the device.
2. Compare them to the factory values.
If the values match, no action is required.
If the values do not match, contact Micro Motion customer service.
Configuration and Use Manual 19
Configure process measurement
Related information
Sample calibration certificate
4.1.1 Calibration factors
The original calibration factors are obtained from factory calibration, and are unique to each device. They are used to adjust measurements for the specific physical properties of the device.
The calibration certificate contains two sets of factors:
Density calibration coefficients
Temperature compensation coefficients
The calibration certificate also provides the results of the Known Density Verification procedure that was performed at the factory.
For each calibration performed at the factory, the calibration certificate contains the data used to calculate the calibration coefficients.
Related information
Sample calibration certificate
Define the relationship between density and the response of your sensor
Adjust density measurement for the effect of temperature on sensor response

4.2 Configure line density measurement

The density measurement parameters control how density is measured and reported.
Configure Density Measurement Unit (Section 4.2.1)
Configure Density Damping (Section 4.2.2)
Configure Density Cutoff (Section 4.2.3)
Configure two‐phase flow parameters (Section 4.2.4)
4.2.1 Configure Density Measurement Unit
Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > DENS
ProLink III Device Tools > Configuration > Process Measurement > Line Density > Density Unit
Field Communicator Configure > Manual Setup > Measurements > Density > Density Unit
Overview
Density Measurement Unit controls the measurement units that will be used in density calculations and reporting.
20 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
Restriction
If the API referral application is enabled, you cannot change the density measurement unit here. The density measurement unit is controlled by the API table selection.
Procedure
Set Density Measurement Unit to the option you want to use.
The default setting for Density Measurement Unit is g/cm3 (grams per cubic centimeter).
Related information
Set up the API referral application
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement Unit. Different communications tools may use different labels.
Options for Density Measurement UnitTable 4-1:
Label
Unit description
Specific gravity SGU SGU SGU
Grams per cubic centimeter G/CM3 g/cm3 g/Cucm
Grams per liter G/L g/l g/L
Grams per milliliter G/mL g/ml g/mL
Kilograms per liter KG/L kg/l kg/L
Kilograms per cubic meter KG/M3 kg/m3 kg/Cum
Pounds per U.S. gallon LB/GAL lbs/Usgal lb/gal
Pounds per cubic foot LB/CUF lbs/ft3 lb/Cuft
Pounds per cubic inch LB/CUI lbs/in3 lb/CuIn
Short ton per cubic yard ST/CUY sT/yd3 STon/Cuyd
Degrees API D API degAPI degAPI
Special unit SPECL special Spcl
Display (standard) ProLink III Field Communicator
Define a special measurement unit for density
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Line Density > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units
Configuration and Use Manual 21
Configure process measurement
Overview
A special measurement unit is a user-defined unit of measure that allows you to report process data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Procedure
1. Specify Density Special Unit Base.
2. Calculate Density Special Unit Conversion Factor as follows:
3. Enter Density Special Unit Conversion Factor.
4. Set User-Defined Label to the name you want to use for the density unit.
Density Special Unit Base is the existing density unit that the special unit will be based on.
a. x base units = y special units
b. Density Special Unit Conversion Factor = x÷y
The original density value is divided by this conversion factor.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for density
You want to measure density in ounces per cubic inch.
1. Set Density Special Unit Base to g/cm3.
2. Calculate Density Special Unit Conversion Factor:
a. 1 g/cm3 = 0.578 oz/in3
b. 1÷0.578 = 1.73
3. Set Density Special Unit Conversion Factor to 1.73.
4. Set User-Defined Label to oz/in3.
4.2.2 Configure Density Damping
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Line Density > Density Damping
Field Communicator Configure > Manual Setup > Measurements > Density > Density Damping
Overview
Density Damping controls the amount of damping that will be applied to the line density value.
22 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Tip
Density damping affects all process variables that are calculated from line density.
Procedure
Set Density Damping to the value you want to use.
The default value is 1.6 seconds. The range is 0 to 60 seconds.
Interaction between Density Damping and Added Damping
When the mA output is configured to report density, both Density Damping and Added Damping are applied to the reported density value.
Density Damping controls the rate of change in the value of the process variable in
transmitter memory. Added Damping controls the rate of change reported via the mA output.
If mA Output Process Variable is set to Density, and both Density Damping and Added Damping are set to non-zero values, density damping is applied first, and the added damping calculation is applied to the result of the first calculation. This value is reported over the mA output.
Related information
Interaction between mA Output Damping and process variable damping
4.2.3 Configure Density Cutoff
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Line Density > Density Cutoff Low
Field Communicator Configure > Manual Setup > Measurements > Density > Density Cutoff
Overview
Density Cutoff Low specifies the lowest density value that will be reported as measured. All density values below this cutoff will be reported as 0.
Procedure
Set Density Cutoff Low to the value you want to use.
The default value is 0.2 g/cm³. The range is 0.0 g/cm³ to 0.5 g/cm³.
Configuration and Use Manual 23
Configure process measurement
4.2.4 Configure two-phase flow parameters
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Line Density
Field Communicator Configure > Manual Setup > Measurements > Density
Overview
The two-phase flow parameters control how the transmitter detects and reports two­phase flow (gas in a liquid process or liquid in a gas process).
Note
Two-phase flow is sometimes referred to as slug flow.
Procedure
1. Set Two-Phase Flow Low Limit to the lowest density value that is considered normal in
your process.
Values below this will cause the transmitter to post Alert A105 (Two-Phase Flow).
Tip
Gas entrainment can cause your process density to drop temporarily. To reduce the occurrence of two-phase flow alerts that are not significant to your process, set Two-Phase Flow Low Limit slightly below your expected lowest process density.
You must enter Two-Phase Flow Low Limit in g/cm³, even if you configured another unit for density measurement.
2. Set Two-Phase Flow High Limit to the highest density value that is considered normal in
your process.
Values above this will cause the transmitter to post Alert A105 (Two-Phase Flow).
Tip
To reduce the occurrence of two-phase flow alerts that are not significant to your process, set Two-Phase Flow High Limit slightly above your expected highest process density.
You must enter Two-Phase Flow High Limit in g/cm³, even if you configured another unit for density measurement.
3. Set Two-Phase Flow Timeout to the number of seconds that the transmitter will wait for
a two-phase flow condition to clear before posting the alert.
The default value for Two-Phase Flow Timeout is 0.0 seconds, meaning that the alert will be posted immediately. The range is 0.0 to 60.0 seconds.
24 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
Detecting and reporting two-phase flow
Two-phase flow (gas in a liquid process or liquid in a gas process) can cause a variety of process control issues. By configuring the two-phase flow parameters appropriately for your application, you can detect process conditions that require correction.
Tip
To decrease the occurrence of two-phase flow alerts, lower Two-Phase Flow Low Limit or raise Two-Phase Flow High Limit.
A two-phase flow condition occurs whenever the measured density goes below Two-Phase Flow Low Limit or above Two-Phase Flow High Limit. If this occurs:
A two-phase flow alert is posted to the active alert log.
Line density is held at its last pre‐alert value for the number of seconds configured in
Two-Phase Flow Timeout.
If the two-phase flow condition clears before Two-Phase Flow Timeout expires:
Line density reverts to actual process density.
The two-phase flow alert is deactivated, but remains in the active alert log until it is
acknowledged.
If the two-phase flow condition does not clear before Two-Phase Flow Timeout expires, line density reverts to actual process density, but the two-phase flow alert remains active.
If Two-Phase Flow Timeout is set to 0.0 seconds, two-phase flow will cause a two-phase flow alert but will have no effect on how the meter measures or reports line density.

4.3 Configure temperature measurement

The temperature measurement parameters control how temperature data from the sensor is reported.
Configure Temperature Measurement Unit (Section 4.3.1)
Configure Temperature Damping (Section 4.3.2)
Configure Temperature Input (Section 4.3.3)
4.3.1 Configure Temperature Measurement Unit
Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > TEMP
ProLink III Device Tools > Configuration > Process Measurement > Line Temperature > Temperature Unit
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temperature Unit
Configuration and Use Manual 25
Configure process measurement
Overview
Temperature Measurement Unit specifies the unit that will be used for temperature measurement.
Restriction
If the API referral application is enabled, the API table selection automatically sets the temperature measurement unit. Configure the API referral application first, then change the temperature measurement unit if desired.
Procedure
Set Temperature Measurement Unit to the option you want to use.
The default setting is Degrees Celsius.
Related information
Options for Temperature Measurement Unit
Set up the API referral application
The transmitter provides a standard set of units for Temperature Measurement Unit. Different communications tools may use different labels for the units.
Options for Temperature Measurement UnitTable 4-2:
Unit description
Degrees Celsius °C °C degC
Degrees Fahrenheit °F °F degF
Degrees Rankine °R °R degR
Kelvin °K °K Kelvin
Display ProLink III
4.3.2 Configure Temperature Damping
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Line Temperature > Temperature Damping
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temp Damping
Label
Field Communica­tor
Overview
Temperature Damping controls the amount of damping that will be applied to the line temperature value, when the on-board temperature data is used (RTD).
26 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Tip
Temperature Damping affects all process variables, compensations, and corrections that use temperature data from the sensor.
Procedure
Enter the value you want to use for Temperature Damping.
Default: 4.8 seconds
Tips
A high damping value makes the process variable appear smoother because the reported value
changes slowly.
A low damping value makes the process variable appear more erratic because the reported value
changes more quickly.
Whenever the damping value is non-zero, the reported measurement will lag the actual
measurement because the reported value is being averaged over time.
In general, lower damping values are preferable because there is less chance of data loss, and less
lag time between the actual measurement and the reported value.
The value you enter is automatically rounded down to the nearest valid value.
4.3.3 Configure Temperature Input
Temperature data from the on-board temperature sensor (RTD) is always available. Optionally, you can set up an external temperature device and use external temperature data.
Tip
Use an external device only if it is more accurate than the internal RTD.
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the internal RTD temperature in some areas and an external temperature in others. The transmitter stores the internal RTD temperature and the external temperature separately. However, the transmitter stores only one alternate temperature value, which may be either an external temperature or the configured fixed value. Accordingly, if you set up polling for temperature in one area, and digital communications in another, and configure a fixed temperature value in a third, the fixed value will be overwritten by polling and digital communications, and polling and digital communications will overwrite each other.
Configuration and Use Manual 27
Configure process measurement
Configure Temperature Input using ProLink III
Configure Temperature Input using the Field Communicator
Configure Temperature Input using ProLink III
ProLink III Device Tools > Configuration > Process Measurement > Line Temperature > Line Temperature Source
Procedure
Choose the method to be used to supply temperature data, and perform the required setup.
Option Description Setup
Internal RTD tem­perature data
Polling The meter polls an external de-
Temperature data from the on­board temperature sensor (RTD) is used.
vice for temperature data. This data will be available in addi­tion to the internal RTD tem­perature data.
a. Set Line Temperature Source to Internal RTD. b. Click Apply.
a. Set Line Temperature Source to Poll for External Value. b. Set Polling Slot to an available slot. c. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
Poll as Secondary Other HART masters will be on the net-
network. The Field Communicator is not a HART master.
work. The Field Communicator is not a HART master.
d. Set External Device Tag to the HART tag of the temperature
device.
e. Click Apply.
Digital communica­tions
A host writes temperature data to the meter at appropriate in­tervals. This data will be availa­ble in addition to the internal RTD temperature data.
a. Set Line Temperature Source to Fixed Value or Digital Communica-
tions. b. Click Apply. c. Perform the necessary host programming and communica-
tions setup to write temperature data to the meter at appro-
priate intervals.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs group on the ProLink III main window.
Need help? If the value is not correct:
28 Micro Motion® Fork Density Meters (FDM)
For polling:
- Verify the wiring between the meter and the external device.
- Verify the HART tag of the external device.
For digital communications:
- Verify that the host has access to the required data.
- Verify that the host is writing to the correct register in memory, using the correct data
type.
Configure Temperature Input using the Field Communicator
Choose the method to be used to supply temperature data, and perform the required setup.
Method Description Setup
Internal RTD tem­perature data
Polling The meter polls an external de-
Temperature data from the on­board temperature sensor (RTD) is used.
vice for temperature data. This data will be available in addi­tion to the internal RTD tem­perature data.
a. Choose Configure > Manual Setup > Measurements > External Inputs
> Temperature.
b. Set External Temperature to Disable.
a. Choose Configure > Manual Setup > Measurements > External Inputs
> Temperature. b. Set External Temperature to Enable. c. Choose Configure > Manual Setup > Inputs/Outputs > External Device
Polling. d. Choose an unused polling slot. e. Set Poll Control to Poll as Primary or Poll as Secondary.
Configure process measurement
Digital communica­tions
A host writes temperature data to the meter at appropriate in­tervals. This data will be availa­ble in addition to the internal RTD temperature data.
Option Description
Poll as Primary No other HART masters will be on the
network. The Field Communicator is not a HART master.
Poll as Secondary Other HART masters will be on the net-
work. The Field Communicator is not a HART master.
f. Set External Device Tag to the HART tag of the external tem-
perature device. g. Set Polled Variable to Temperature.
a. Choose Configure > Manual Setup > Measurements > External Inputs
> Temperature. b. Set External Temperature to Enable. c. Perform the necessary host programming and communica-
tions setup to write temperature data to the meter at appro-
priate intervals.
Configuration and Use Manual 29
Configure process measurement
Postrequisites
Choose Service Tools > Variables > External Variables and verify the value for External Temperature.
Need help? If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
- Verify the wiring between the meter and the external device.
- Verify the HART tag of the external device.
For digital communications:
- Verify that the host has access to the required data.
- Verify that the host is writing to the correct register in memory, using the correct data
type.
If necessary, apply an offset.

4.4 Configure the pressure input

Pressure data is required to calculate base density from line density.
The meter does not measure pressure, so you must provide an external pressure
input.
You must use absolute pressure.
Pressure data is required for several different measurements.
There are several different methods to obtain pressure data.
Tip
A fixed pressure value is not recommended as it can produce inaccurate process data.
Configure the pressure input using ProLink III (Section 4.4.1)
Configure the pressure input using the Field Communicator (Section 4.4.2)
4.4.1 Configure the pressure input using ProLink III
1. Choose Device Tools > Configuration > Process Measurement > Line Pressure.
2. Set Pressure Type to match the pressure measurement from the external pressure device.
Option Description
Absolute The external pressure device reports absolute pressure.
Gauge The external pressure device reports gauge pressure.
30 Micro Motion® Fork Density Meters (FDM)
Restriction
If Line Pressure Source is set to Fixed, you cannot configure Pressure Type. You must enter the pressure value in the required form. To set Pressure Type, you may need to change the setting of Line Pressure Source.
The meter requires gauge pressure. If you select Absolute, the device will convert the input pressure value to the equivalent gauge pressure.
3. Set Pressure Unit to the unit used by the external pressure device.
Restriction
If the API referral application is enabled, the API table selection automatically sets the pressure measurement unit. Configure the API referral application first, then change the pressure measurement unit if necessary.
4. Choose the method used to supply pressure data and perform the required setup.
Option Description Setup
Polling The meter polls an external de-
vice for pressure data.
a. Set Pressure Source to Poll for External Value. b. Set Polling Slot to an available slot. c. Set Polling Control to Poll as Primary or Poll as Secondary.
Configure process measurement
Digital communica­tions
Option Description
Poll as Primary No other HART masters will be on the
network. The Field Communicator is not a HART master.
Poll as Secondary Other HART masters will be on the net-
work. The Field Communicator is not a HART master.
d. Set External Device Tag to the HART tag of the temperature
device.
A host writes pressure data to the meter at appropriate inter­vals.
a. Set Pressure Source to Fixed Value or Digital Communications. b. Perform the necessary host programming and communica-
tions setup to write pressure data to the meter at appropri­ate intervals.
Postrequisites
The current pressure value is displayed in the External Pressure field. Verify that the value is correct.
Need help? If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
- Verify the wiring between the meter and the external device.
Configuration and Use Manual 31
Configure process measurement
- Verify the HART tag of the external device.
For digital communications:
- Verify that the host has access to the required data.
- Verify that the host is writing to the correct register in memory, using the correct data
type.
If necessary, apply an offset.
Note
Do not use the offset in conjunction with the fixed pressure value. Enter the adjusted value.
Related information
Set up the API referral application
4.4.2 Configure the pressure input using the Field Communicator
1. Choose Configure > Manual Setup > Measurements > External Inputs > Pressure.
2. Set Pressure Input to Enable.
3. Set Pressure Type to match the pressure measurement from the external pressure
device.
Option Description
Absolute The external pressure device reports absolute pressure.
Gauge The external pressure device reports gauge pressure.
Restriction
If Line Pressure Source is set to Fixed, you cannot configure Pressure Type. You must enter the pressure value in the required form. To set Pressure Type, you may need to change the setting of Line Pressure Source.
The meter requires gauge pressure. If you select Absolute, the device will convert the input pressure value to the equivalent gauge pressure.
4. Set Pressure Unit to the unit used by the external pressure device.
Restriction
If the API referral application is enabled, the API table selection automatically sets the pressure measurement unit. Configure the API referral application first, then change the pressure measurement unit if necessary.
5. Set up the pressure input.
a. Choose Configure > Manual Setup > Inputs/Outputs > External Device Polling.
b. Choose an unused polling slot.
32 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
c. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the network. The
Field Communicator is not a HART master.
Poll as Secondary Other HART masters will be on the network. The
Field Communicator is not a HART master.
d. Set External Device Tag to the HART tag of the external pressure device.
e. Set Polled Variable to Pressure.
Postrequisites
Choose Service Tools > Variables > External Variables and verify the value for External Pressure.
Need help? If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
- Verify the wiring between the meter and the external device.
- Verify the HART tag of the external device.
For digital communications:
- Verify that the host has access to the required data.
- Verify that the host is writing to the correct register in memory, using the correct data
type.
If necessary, apply an offset.
Related information
Set up the API referral application
4.4.3 Options for Pressure Measurement Unit
The transmitter provides a standard set of measurement units for Pressure Measurement Unit. Different communications tools may use different labels for the units. In most applications, Pressure Measurement Unit should be set to match the pressure measurement unit used by the remote device.
Options for Pressure Measurement UnitTable 4-3:
Label
Unit description
Feet water @ 68 °F FTH2O Ft Water @ 68°F ftH2O
Inches water @ 4 °C INW4C In Water @ 4°C inH2O @4DegC
Inches water @ 60 °F INW60 In Water @ 60°F inH2O @60DegF
Configuration and Use Manual 33
Display ProLink III Field Communicator
Configure process measurement
Options for Pressure Measurement Unit (continued)Table 4-3:
Label
Unit description
Inches water @ 68 °F INH2O In Water @ 68°F inH2O
Millimeters water @ 4 °C mmW4C mm Water @ 4°C mmH2O @4DegC
Millimeters water @ 68 °F mmH2O mm Water @ 68°F mmH2O
Millimeters mercury @ 0 °C mmHG mm Mercury @ 0°C mmHg
Inches mercury @ 0 °C INHG In Mercury @ 0°C inHG
Pounds per square inch PSI PSI psi
Bar BAR bar bar
Millibar mBAR millibar mbar
Grams per square centimeter G/SCM g/cm2 g/Sqcm
Kilograms per square centimeter KG/SCM kg/cm2 kg/Sqcm
Pascals PA pascals Pa
Kilopascals KPA Kilopascals kPa
Megapascals MPA Megapascals MPa
Torr @ 0 °C TORR Torr @ 0°C torr
Atmospheres ATM atms atms
Display ProLink III Field Communicator

4.5 Set up the API referral application

The API referral application corrects line density to reference temperature and reference pressure according to American Petroleum Institute (API) standards. The resulting process variable is referred density.
Set up the API referral application using ProLink III (Section 4.5.1)
Set up the API referral application using the Field Communicator (Section 4.5.2)
4.5.1 Set up the API referral application using ProLink III
This section guides you through the tasks required to set up and implement the API referral application.
1. Enable the API referral application using ProLink III
2. Configure API referral using ProLink III
3. Set up temperature and pressure data for API referral using ProLink III
Enable the API referral application using ProLink III
The API referral application must be enabled before you can perform any setup. If the API referral application was enabled at the factory, you do not need to enable it now.
34 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
1. Choose Device Tools > Configuration > Transmitter Options.
2. If the concentration measurement application is enabled, disable it and click Apply.
The concentration measurement application and the API referral application cannot be enabled simultaneously.
3. Enable API Referral and click Apply.
Configure API referral using ProLink III
The API referral parameters specify the API table, measurement units, and reference values to be used in referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process fluid.
You must know the reference temperature and reference pressure that you want to use.
Procedure
1. Choose Device Tools > Configuration > Process Measurement > API Referral.
2. Specify the API table to use.
Each API table is associated with a specific set of equations.
a. Set Process Fluid to the API table group that your process fluid belongs to.
API table group Process fluids
A tables Generalized crude and JP4
B tables Generalized products: Gasoline, jet fuel, aviation fuel, kerosene,
heating oils, fuel oils, diesel, gas oil
C tables Liquids with a constant base density or known thermal expansion
coefficient (TEC). You will be required to enter the TEC for your process fluid.
D tables Lubricating oils
E tables NGL (Natural Gas Liquids) and LPG (Liquid Petroleum Gas)
Restriction
The API referral application is not appropriate for the following process fluids: propane and propane mixes, butane and butane mixes, butadiene and butadiene mixes, isopentane, LNG, ethylene, propylene, cyclohexane, aeromatics, asphalts, and road tars.
b. Set Referred Density Measurement Unit to the measurement units that you want to
use for referred density.
c. Click Apply.
Configuration and Use Manual 35
Configure process measurement
3. Refer to the API documentation and confirm your table selection.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
5. Set Reference Temperature to the temperature to which density will be corrected in
6. Set Reference Pressure to the pressure to which density will be corrected in referred
These parameters uniquely identify the API table. The selected API table is displayed, and the meter automatically changes the density unit, temperature unit, pressure unit, reference temperature, and reference pressure to match the API table.
Restriction
Not all combinations are supported by the API referral application. See the list of API tables in this manual.
a. Verify that your process fluid falls within range for line density, line temperature,
and line pressure.
If your process fluid goes outside any of these limits, the meter will post a status alert and will report line density instead of referred density until the process fluid goes back within range.
b. Verify that the referred density range of the selected table is adequate for your
application.
referred density calculations. If you choose Other, select the temperature measurement unit and enter the reference temperature.
density calculations.
API tables supported by the API referral application
The API tables listed here are supported by the API referral application.
API tables, process fluids, measurement units, and default reference valuesTable 4-4:
Process fluid API table Referred density (API)
Generalized crude and JP4 5A Unit: °API
Range: 0 to 100 °API
23A Unit: SGU
Range: 0.6110 to 1.0760 SGU
3
Generalized products (gaso­line, jet fuel, aviation fuel, kerosene, heating oils, fuel oils, diesel, gas oil)
53A Unit: kg/m
Range: 610 to 1075 kg/m³
5B Unit: °API
Range: 0 to 85 °API
23B Unit: SGU
Range: 0.6535 to 1.0760 SGU
Default reference temperature
60 °F 0 psi (g)
60 °F 0 psi (g)
15 °C 0 kPa (g)
60 °F 0 psi (g)
60 °F 0 psi (g)
Default reference pressure
36 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
API tables, process fluids, measurement units, and default reference values (continued)Table 4-4:
Default reference
Process fluid API table Referred density (API)
53B Unit: kg/m
Range: 653 to 1075 kg/m³
Liquids with a constant den­sity base or known thermal expansion coefficient
Lubricating oils 5D Unit: °API
NGL (natural gas liquids) 23E Unit: SGU 60 °F 0 psi (g)
LPG (liquid petroleum gas) 24E Unit: SGU 60 °F 0 psi (g)
Restriction
These tables are not appropriate for the following process fluids: propane and propane mixes, butane and butane mixes, butadiene and butadiene mixes, isopentane, LNG, ethylene, propylene, cyclohexane, aeromatics, asphalts, and road tars.
6C Unit: °API 60 °F 0 psi (g)
24C Unit: SGU 60 °F 0 psi (g)
54C Unit: kg/m³ 15 °C 0 kPa (g)
Range: 10 to +40 °API
23D Unit: SGU
Range: 0.8520 to 1.1640 SGU
53D Unit: kg/m³
Range: 825 to 1164 kg/m³
3
temperature
15 °C 0 kPa (g)
60 °F 0 psi (g)
60 °F 0 psi (g)
15 °C 0 kPa (g)
Default reference pressure
Set up temperature and pressure data for API referral using ProLink III
The API referral application uses temperature and, optionally, pressure data in its calculations. You must decide how to provide this data, then perform the required configuration and setup.
Tip
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value may produce inaccurate process data.
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the internal RTD temperature in some areas and an external temperature in others. The transmitter stores the internal RTD temperature and the external temperature separately. However, the transmitter stores only one alternate temperature value, which may be either an external temperature or the configured fixed value. Accordingly, if you set up polling for temperature in one area, and digital communications in another, and configure a fixed temperature value in a third, the fixed value will be overwritten by polling and digital communications, and polling and digital communications will overwrite each other.
Configuration and Use Manual 37
Configure process measurement
Important
Line pressure data is used in several different measurements and calculations. The transmitter stores only one pressure value, which may be either the external pressure or the configured fixed value. Accordingly, if you choose a fixed pressure for some uses, and an external pressure for others, the external pressure will overwrite the fixed value.
Prerequisites
If you plan to poll an external device, the primary mA output (Channel A) must be
The pressure measurement must be gauge pressure, not atmospheric pressure. If
The pressure device must use the pressure unit that is configured in the transmitter.
If you are using an external temperature device, it must use the temperature unit
Procedure
1. Choose Device Tools > Configuration > Process Measurement > API Referral.
2. Choose the method to be used to supply temperature data, and perform the
wired to support HART communications.
your pressure input is absolute, you can set the pressure type to absolute and it will be converted to gauge for the API calculations.
that is configured in the transmitter.
required setup.
Option Description Setup
Internal RTD tem­perature data
Polling The meter polls an external de-
Temperature data from the on­board temperature sensor (RTD) is used.
vice for temperature data. This data will be available in addi­tion to the internal RTD tem­perature data.
a. Set Line Temperature Source to Internal RTD. b. Click Apply.
a. Set Line Temperature Source to Poll for External Value. b. Set Polling Slot to an available slot. c. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
Poll as Secondary Other HART masters will be on the net-
d. Set External Device Tag to the HART tag of the temperature
device.
e. Click Apply.
network. The Field Communicator is not a HART master.
work. The Field Communicator is not a HART master.
38 Micro Motion® Fork Density Meters (FDM)
Option Description Setup
Digital communica­tions
A host writes temperature data to the meter at appropriate in­tervals. This data will be availa­ble in addition to the internal RTD temperature data.
a. Set Line Temperature Source to Fixed Value or Digital Communica-
tions. b. Click Apply. c. Perform the necessary host programming and communica-
tions setup to write temperature data to the meter at appro-
priate intervals.
3. Choose the method used to supply pressure data and perform the required setup.
Option Description Setup
Polling The meter polls an external de-
vice for pressure data.
a. Set Pressure Source to Poll for External Value. b. Set Polling Slot to an available slot. c. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
Poll as Secondary Other HART masters will be on the net-
Configure process measurement
network. The Field Communicator is not a HART master.
work. The Field Communicator is not a HART master.
Digital communica­tions
d. Set External Device Tag to the HART tag of the temperature
device.
A host writes pressure data to the meter at appropriate inter­vals.
a. Set Pressure Source to Fixed Value or Digital Communications. b. Perform the necessary host programming and communica-
tions setup to write pressure data to the meter at appropri-
ate intervals.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs group on the ProLink III main window.
The current pressure value is displayed in the External Pressure field. Verify that the value is correct.
Need help? If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
- Verify the wiring between the meter and the external device.
- Verify the HART tag of the external device.
For digital communications:
- Verify that the host has access to the required data.
Configuration and Use Manual 39
Configure process measurement
- Verify that the host is writing to the correct register in memory, using the correct data
type.
If necessary, apply an offset.
4.5.2 Set up the API referral application using the Field Communicator
This section guides you through the tasks required to set up and implement the API referral application.
1. Enable the API referral application using the Field Communicator
2. Configure API referral using the Field Communicator
3. Set up temperature and pressure data for API referral using the Field Communicator
Enable the API referral application using the Field Communicator
The API referral application must be enabled before you can perform any setup. If the API referral application was enabled at the factory, you do not need to enable it now.
1. Choose Overview > Device Information > Applications > Enable/Disable Applications.
2. If the concentration measurement application is enabled, disable it.
The concentration measurement application and the API referral application cannot be enabled simultaneously.
3. Enable the API referral application.
Configure API referral using the Field Communicator
The API referral parameters specify the API table, measurement units, and reference values to be used in referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process fluid.
You must know the reference temperature and reference pressure that you want to use.
Procedure
1. Choose Configure > Manual Setup > Measurements > API Referral.
2. Choose API Referral Setup.
3. Specify the API table that you want to use for measurement.
Each API table is associated with a specific set of equations.
a. Set API Table Number to the number that matches the API table units that you
want to use for referred density.
40 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
Your choice also determines the measurement units to be used for temperature and pressure, and the default values for reference temperature and reference pressure.
API Table Number
Measurement unit for referred density
Temperature measurement unit
Pressure meas­urement unit
Default refer­ence tempera­ture
Default refer­ence pressure
5 °API °F psi (g) 60 °F 0 psi (g)
(1)
6
°API °F psi (g) 60 °F 0 psi (g)
23 SGU °F psi (g) 60 °F 0 psi (g)
(1)
24
SGU °F psi (g) 60 °F 0 psi (g)
53 kg/m³ °C kPa (g) 15 °C 0 kPa (g)
(1)
54
(1) Used only with API Table Letter = C.
kg/m³ °C kPa (g) 15 °C 0 kPa (g)
b. Set API Table Letter to the letter of the API table group that is appropriate for your
process fluid.
API Table Letter Process fluids
A Generalized crude and JP4
B Generalized products: Gasoline, jet fuel, aviation fuel, kerosene,
heating oils, fuel oils, diesel, gas oil
(1)
C
D Lubricating oils
E NGL (Natural Gas Liquids) and LPG (Liquid Petroleum Gas)
Liquids with a constant base density or known thermal expansion coefficient (TEC). You will be required to enter the TEC for your process fluid.
(1) Used only with API Table Number = 6, 24, or 54.
Restriction
The API referral application is not appropriate for the following process fluids: propane and propane mixes, butane and butane mixes, butadiene and butadiene mixes, isopentane, LNG, ethylene, propylene, cyclohexane, aeromatics, asphalts, and road tars.
API Table Number and API Table Letter uniquely identify the API table. The selected API table is displayed, and the meter automatically changes the density unit, temperature unit, pressure unit, reference temperature, and reference pressure to match the API table.
Restriction
Not all combinations are supported by the API referral application. See the list of API tables in this manual.
Configuration and Use Manual 41
Configure process measurement
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
5. Refer to the API documentation and confirm your table selection.
6. If required, set Ref Temperature to the temperature to which density will be corrected
7. If required, set Ref Pressure (Gauge) to the pressure to which density will be corrected
a. Verify that your process fluid falls within range for line density, line temperature,
and line pressure.
If your process fluid goes outside any of these limits, the meter will post a status alert and will report line density instead of referred density until the process fluid goes back within range.
b. Verify that the referred density range of the selected table is adequate for your
application.
in referred density calculations.
The default reference temperature is determined by the selected API table.
a. Choose Service Tools > Maintenance > Modbus Data > Write Modbus Data.
b. Write the desired reference temperature to Registers 319–320, in the
measurement unit required by the selected API table. Use 32-bit IEEE floating­point format.
in referred density calculations.
The default reference pressure is determined by the selected API table. API referral requires gauge pressure.
a. Choose Service Tools > Maintenance > Modbus Data > Write Modbus Data.
b. Write the desired reference pressure to Registers 4601–4602, in the
measurement unit required by the selected API table. Use 32-bit IEEE floating­point format.
API tables supported by the API referral application
The API tables listed here are supported by the API referral application.
API tables, process fluids, measurement units, and default reference valuesTable 4-5:
Process fluid API table Referred density (API)
Generalized crude and JP4 5A Unit: °API
Range: 0 to 100 °API
23A Unit: SGU
Range: 0.6110 to 1.0760 SGU
53A Unit: kg/m
Range: 610 to 1075 kg/m³
3
Default reference temperature
60 °F 0 psi (g)
60 °F 0 psi (g)
15 °C 0 kPa (g)
Default reference pressure
42 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
API tables, process fluids, measurement units, and default reference values (continued)Table 4-5:
Default reference
Process fluid API table Referred density (API)
Generalized products (gaso­line, jet fuel, aviation fuel, kerosene, heating oils, fuel oils, diesel, gas oil)
Liquids with a constant den­sity base or known thermal expansion coefficient
Lubricating oils 5D Unit: °API
NGL (natural gas liquids) 23E Unit: SGU 60 °F 0 psi (g)
LPG (liquid petroleum gas) 24E Unit: SGU 60 °F 0 psi (g)
5B Unit: °API
Range: 0 to 85 °API
23B Unit: SGU
Range: 0.6535 to 1.0760 SGU
53B Unit: kg/m
Range: 653 to 1075 kg/m³
6C Unit: °API 60 °F 0 psi (g)
24C Unit: SGU 60 °F 0 psi (g)
54C Unit: kg/m³ 15 °C 0 kPa (g)
Range: 10 to +40 °API
23D Unit: SGU
Range: 0.8520 to 1.1640 SGU
53D Unit: kg/m³
Range: 825 to 1164 kg/m³
3
temperature
60 °F 0 psi (g)
60 °F 0 psi (g)
15 °C 0 kPa (g)
60 °F 0 psi (g)
60 °F 0 psi (g)
15 °C 0 kPa (g)
Default reference pressure
Restriction
These tables are not appropriate for the following process fluids: propane and propane mixes, butane and butane mixes, butadiene and butadiene mixes, isopentane, LNG, ethylene, propylene, cyclohexane, aeromatics, asphalts, and road tars.
Set up temperature and pressure data for API referral using the Field Communicator
The API referral application uses temperature and, optionally, pressure data in its calculations. You must decide how to provide this data, then perform the required configuration and setup.
Tip
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value may produce inaccurate process data.
Configuration and Use Manual 43
Configure process measurement
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the internal RTD temperature in some areas and an external temperature in others. The transmitter stores the internal RTD temperature and the external temperature separately. However, the transmitter stores only one alternate temperature value, which may be either an external temperature or the configured fixed value. Accordingly, if you set up polling for temperature in one area, and digital communications in another, and configure a fixed temperature value in a third, the fixed value will be overwritten by polling and digital communications, and polling and digital communications will overwrite each other.
Important
Line pressure data is used in several different measurements and calculations. The transmitter stores only one pressure value, which may be either the external pressure or the configured fixed value. Accordingly, if you choose a fixed pressure for some uses, and an external pressure for others, the external pressure will overwrite the fixed value.
Prerequisites
If you plan to poll an external device, the primary mA output (Channel A) must be wired to support HART communications.
The pressure measurement must be gauge pressure, not atmospheric pressure. If your pressure input is absolute, you can set the pressure type to absolute and it will be converted to gauge for the API calculations.
The pressure device must use the pressure unit that is configured in the transmitter.
If you are using an external temperature device, it must use the temperature unit that is configured in the transmitter.
Procedure
1. Choose the method to be used to supply temperature data, and perform the
required setup.
Method Description Setup
Internal tempera­ture
Temperature data from the on­board temperature sensor (RTD) will be used for all meas­urements and calculations. No external temperature data will be available.
a. Choose Configure > Manual Setup > Measurements > External Inputs
> Temperature.
b. Set External Temperature to Disable.
44 Micro Motion® Fork Density Meters (FDM)
Method Description Setup
Polling The meter polls an external de-
vice for temperature data. This data will be available in addi­tion to the internal tempera­ture data.
Digital communica­tions
A host writes temperature data to the meter at appropriate in­tervals. This data will be availa­ble in addition to the internal temperature data.
a. Choose Configure > Manual Setup > Measurements > External Inputs
> Temperature. b. Set External Temperature to Enable. c. Choose Configure > Manual Setup > Inputs/Outputs > External Device
Polling. d. Choose an unused polling slot. e. Set Poll Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
Poll as Secondary Other HART masters will be on the net-
f. Set External Device Tag to the HART tag of the external tem-
perature device. g. Set Polled Variable to Temperature.
a. Choose Configure > Manual Setup > Measurements > External Inputs
> Temperature. b. Set External Temperature to Enable. c. Perform the necessary host programming and communica-
tions setup to write temperature data to the transmitter at
appropriate intervals.
Configure process measurement
network. The Field Communicator is not a HART master.
work. The Field Communicator is not a HART master.
2. Choose the method to be used to supply pressure data, and perform the required setup.
Configuration and Use Manual 45
Configure process measurement
Method Description Setup
Polling The meter polls an external de-
vice for pressure data.
Digital communica­tions
A host writes pressure data to the meter at appropriate inter­vals.
a. Choose Configure > Manual Setup > Measurements > External Inputs
b. Set Pressure Input to Enable. c. Choose Configure > Manual Setup > Inputs/Outputs > External Device
d. Choose an unused polling slot. e. Set Poll Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
Poll as Secondary Other HART masters will be on the net-
f. Set External Device Tag to the HART tag of the external pres-
g. Set Polled Variable to Pressure.
a. Choose Configure > Manual Setup > Measurements > External Inputs
b. Set Pressure Input to Enable. c. Perform the necessary host programming and communica-
> Pressure.
Polling.
network. The Field Communicator is not a HART master.
work. The Field Communicator is not a HART master.
sure device.
> Pressure.
tions setup to write pressure data to the transmitter at ap­propriate intervals.
Postrequisites
Choose Service Tools > Variables > External Variables and verify the values for External Temperature and External Pressure.
Need help? If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
- Verify the wiring between the meter and the external device.
- Verify the HART tag of the external device.
For digital communications:
- Verify that the host has access to the required data.
- Verify that the host is writing to the correct register in memory, using the correct data
type.
If necessary, apply an offset.
46 Micro Motion® Fork Density Meters (FDM)
Configure process measurement

4.6 Set up concentration measurement

The concentration measurement application calculates concentration from line density and line temperature.
Related information
Preparing to set up concentration measurement Set up concentration measurement using ProLink III Set up concentration measurement using the Field Communicator
4.6.1 Preparing to set up concentration measurement
The procedure for setting up concentration measurement application depends on how your device was ordered and how you want to use the application. Review this information before you begin.
Requirements for concentration measurement
To use the concentration measurement application, the following conditions must be met:
The concentration measurement application must be enabled.
One or more concentration matrices must be stored in your transmitter.
Tip
In most cases, the concentration matrix that you ordered was loaded at the factory. If you need to add concentration matrices and you are using ProLink III, you can load concentration matrices from a file or you can build a custom matrix. If you are using the Field Communicator, you can build a custom matrix but you cannot load a matrix from a file. This manual does not discuss building a custom matrix. For information on building a custom matrix, see Micro Motion Enhanced Density Application: Theory, Configuration, and Use.
If your concentration matrices use Specific Gravity as the derived variable, the
reference temperature values must be set.
Temperature Source must be configured and set up.
One matrix must be selected as the active matrix (the matrix used for
measurement).
Optional tasks in setting up concentration measurement
The following tasks are optional:
Modifying names and labels
Configuring operational parameters
- Extrapolation alerts
- Calculation method (matrix or equation)
- Matrix switching
Configuration and Use Manual 47
Configure process measurement
4.6.2 Set up concentration measurement using ProLink III
This section guides you through the tasks required to set up, configure, and implement concentration measurement.
Restriction
This section does not cover building a concentration matrix. See Micro Motion Enhanced Density Application: Theory, Configuration, and Use for detailed information on building a matrix.
1. Enable the concentration measurement application using ProLink III
2. Load a concentration matrix using ProLink III
3. Set reference temperature values for specific gravity using ProLink III
4. Set up temperature data for concentration measurement using ProLink III
5. Modify matrix names and labels using ProLink III
6. Modify operational parameters for concentration measurement using ProLink III
7. Select the active concentration matrix using ProLink III
Enable the concentration measurement application using ProLink III
The concentration measurement application must be enabled before you can perform any setup. If the concentration measurement application was enabled at the factory, you do not need to enable it now.
1. Choose Device Tools > Configuration > Transmitter Options.
2. If the API referral application is enabled, disable it and click Apply.
The concentration measurement application and the API referral application cannot be enabled simultaneously.
3. Set Concentration Measurement to Enabled and click Apply.
Load a concentration matrix using ProLink III
At least one concentration matrix must be loaded onto your transmitter. You can load up to six.
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not need to load any matrices.
Prerequisites
Standard matrices for the concentration measurement application Concentration measurement matrices available by order
The concentration measurement application must be enabled on your device.
For each concentration matrix that you want to load, you need a file containing the matrix data. The ProLink III installation includes a set of standard concentration matrices. Other matrices are available from Micro Motion.
48 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
Tips
If you have a custom matrix on another device, you can save it to a file, then load it to the
current device.
If you have a matrix file in ProLink II format, you can load it using ProLink III.
You must know the following information for your matrix:
The derived variable that the matrix is designed to calculate
The density unit that the matrix was built with
The temperature unit that the matrix was built with
Important
All concentration matrices on your transmitter must use the same derived variable.
If you change the setting of Derived Variable, all existing concentration matrices will be deleted
from transmitter memory. Set Derived Variable before loading concentration matrices.
If you want the meter to calculate Net Mass Flow Rate, the derived variable must be set to
Mass Concentration (Density). If your matrix is not set up for Mass Concentration (Density), contact Micro Motion for assistance or for a custom matrix.
If you want the meter to calculate Net Volume Flow Rate, the derived variable must be set to
Volume Concentration (Density). If your matrix is not set up for Volume Concentration (Density), contact Micro Motion for assistance or for a custom matrix.
If you plan to use matrix switching, you must identify the two matrices to be used for switching and load them into Slot 1 and Slot 2.
Procedure
1. Choose Device Tools > Configuration > Process Measurement > Line Density and set Density Unit to the density unit used by your matrix.
Important
When you load a matrix, if the density unit is not correct, concentration data will be incorrect. The density units must match at the time of loading. You can change the density unit after the matrix is loaded.
2. Choose Device Tools > Configuration > Process Measurement > Line Temperature and set Temperature Unit to the temperature unit used by your matrix.
Important
When you load a matrix, if the temperature unit is not correct, concentration data will be incorrect. The temperature units must match at the time of loading. You can change the temperature unit after the matrix is loaded.
3. Choose Device Tools > Configuration > Process Measurement > Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
Configuration and Use Manual 49
Configure process measurement
4. In Step 1, ensure that the setting of Derived Variable matches the derived variable
5. Load one or more matrices.
6. (Optional) Set the density and temperature units to the units you want to use for
Related information
used by your matrix. If it does not, change it as required and click Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from transmitter memory. Verify the setting of Derived Variable before continuing.
a. In Step 2, set Matrix Being Configured to the location (slot) to which the matrix will
be loaded.
b. Click Load Matrix from a File, navigate to the matrix file on your computer, and load
it.
c. Repeat until all required matrices are loaded.
measurement.
Matrix switching Derived variables and calculated process variables Configure Temperature Measurement Unit Configure Density Measurement Unit Measuring Net Mass Flow Rate and Net Volume Flow Rate
Set reference temperature values for specific gravity using ProLink III
When Derived Variable is set to Specific Gravity, you must set the reference temperature to be used for density measurement and the reference temperature of water, and then verify the density of water at the configured reference temperature. The two reference temperature values affect specific gravity measurement.
Additionally, the two reference temperature values affect any concentration process variable that is calculated by equation, rather than by matrix, because the equations are based on specific gravity.
Typically, the two reference temperature values are the same, but this is not required.
Restriction
If Derived Variable is not set to Specific Gravity, do not change any of these values. These are set by the active concentration matrix.
Procedure
1. Choose Device Tools > Configuration > Process Measurement > Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
50 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and click Change Matrix.
3. Scroll to Step 3, then perform the following actions:
a. Set Reference Temperature for Referred Density to the temperature to which line
density will be corrected for use in the specific gravity calculation.
b. Set Reference Temperature for Water to the water temperature that will be used in
the specific gravity calculation.
c. Set Water Density at Reference Temperature to the density of water at the specified
reference temperature.
The transmitter automatically calculates the density of water at the specified temperature. The new value will be displayed the next time that transmitter memory is read. You can enter a different value if you want to.
4. Click the Apply button at the bottom of Step 3.
Related information
Using equations to calculate specific gravity, °Baumé, °Brix, °Plato, and °Twaddell
Set up temperature data for concentration measurement using ProLink III
The concentration measurement application uses line temperature data in its calculations. You must decide how to provide this data, then perform the required configuration and setup. Temperature data from the on-board temperature sensor (RTD) is always available. You can set up an external temperature device and use external temperature data if you want to.
The temperature setup that you establish here will be used for all concentration measurement matrices on this meter.
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the internal RTD temperature in some areas and an external temperature in others. The transmitter stores the internal RTD temperature and the external temperature separately. However, the transmitter stores only one alternate temperature value, which may be either an external temperature or the configured fixed value. Accordingly, if you set up polling for temperature in one area, and digital communications in another, and configure a fixed temperature value in a third, the fixed value will be overwritten by polling and digital communications, and polling and digital communications will overwrite each other.
Prerequisites
If you plan to poll an external device, the primary mA output (Channel A) must be wired to support HART communications.
Procedure
1. Choose Device Tools > Configuration > Process Measurement > Concentration Measurement.
Configuration and Use Manual 51
Configure process measurement
The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 4.
3. Choose the method to be used to supply temperature data, and perform the required setup.
Option Description Setup
Internal RTD tem­perature data
Polling The meter polls an external de-
Temperature data from the on­board temperature sensor (RTD) is used.
vice for temperature data. This data will be available in addi­tion to the internal RTD tem­perature data.
a. Set Line Temperature Source to Internal RTD. b. Click Apply.
a. Set Line Temperature Source to Poll for External Value. b. Set Polling Slot to an available slot. c. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
Poll as Secondary Other HART masters will be on the net-
network. The Field Communicator is not a HART master.
work. The Field Communicator is not a HART master.
Digital communica­tions
d. Set External Device Tag to the HART tag of the temperature
device.
e. Click Apply.
A host writes temperature data to the meter at appropriate in­tervals. This data will be availa­ble in addition to the internal RTD temperature data.
a. Set Line Temperature Source to Fixed Value or Digital Communica-
tions. b. Click Apply. c. Perform the necessary host programming and communica-
tions setup to write temperature data to the meter at appro-
priate intervals.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs group on the ProLink III main window.
Need help? If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
- Verify the wiring between the meter and the external device.
- Verify the HART tag of the external device.
For digital communications:
52 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
- Verify that the host has access to the required data.
- Verify that the host is writing to the correct register in memory, using the correct data
type.
If necessary, apply an offset.
Modify matrix names and labels using ProLink III
For convenience, you can change the name of a concentration matrix and the label used for its measurement unit. This does not affect measurement.
1. Choose Device Tools > Configuration > Process Measurement > Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and click Change Matrix.
3. Scroll to Step 3, then perform the following actions:
a. Set Concentration Units Label to the label that will be used for the concentration
unit.
b. If you set Concentration Units Label to Special, enter the custom label in User-Defined
Label.
c. In Matrix Name, enter the name to be used for the matrix.
4. Click the Apply button at the bottom of Step 3.
Modify operational parameters for concentration measurement using ProLink III
You can enable and disable extrapolation alerts, set extrapolation alert limits, and control matrix switching. These parameters control the behavior of the concentration measurement application but do not affect measurement directly. Additionally, for certain types of concentration measurement, you can select the calculation method to be used.
1. Choose Device Tools > Configuration > Process Measurement > Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and click Change Matrix.
3. Scroll to Step 4.
4. Set up extrapolation alerts.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density or line temperature goes outside the range, the transmitter will extrapolate concentration values. However, extrapolation may
Configuration and Use Manual 53
Configure process measurement
5. If applicable, use the Equation Selection parameter to specify the type of calculation to
affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is occurring, and can also be used to initiate matrix switching. Each concentration matrix has its own extrapolation alert settings.
a. Set Extrapolation Alert Limit to the point, in percent, at which an extrapolation alert
will be posted.
b. Enable or disable the high and low limit alerts for temperature and density, as
desired, and click Apply.
Important
If you plan to use matrix switching, you must enable the appropriate extrapolation alerts.
Example: If Extrapolation Alert Limit is set to 5%, High Extrapolation Limit (Temperature) is enabled, and the matrix is built for a temperature range of 40 °F to 80 °F, an extrapolation alert will be posted if line temperature goes above 82 °F.
be used, and click Apply.
This option is available only when Derived Variable is set to Specific Gravity.
Option Description
Specific Gravity
Baume Specific gravity is calculated as described above, and the result is used in the
Brix Specific gravity is calculated as described above and the result is used in the
Plato Specific gravity is calculated as described above, and the result is used n the
Twaddell Specific gravity is calculated as described above, and the result is used in the
Referred density is calculated from the matrix. The result is used in the specific gravity equation. The output is specific gravity.
°Baumé equation. The output is specific gravity and °Baumé (light or heavy).
°Brix equation. The output is specific gravity and °Brix.
°Plato equation. The output is specific gravity and °Plato.
°Twaddell equation. The output is specific gravity and °Twaddell.
6. Enable or disable Matrix Switching as desired, and click Apply.
When matrix switching is enabled and an extrapolation alert occurs, the transmitter automatically switches from the matrix in Slot 1 to the matrix in Slot 2, or vice versa. This occurs only if no extrapolation alert would be generated by the other matrix. Matrix switching is not applicable to any other slots.
Related information
Using equations to calculate specific gravity, °Baumé, °Brix, °Plato, and °Twaddell Matrix switching
54 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
Select the active concentration matrix using ProLink III
You must select the concentration matrix to be used for measurement. Although the transmitter can store up to six concentration matrices, only one matrix can be used for measurement at any one time.
1. Choose Device Tools > Configuration > Process Measurement > Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Active Matrix to the matrix you want to use and click Change Matrix.
Note
To support matrix switching, you must select the matrix in Slot 1 or the matrix in Slot 2.
Related information
Matrix switching
4.6.3 Set up concentration measurement using the Field Communicator
This section guides you through most of the tasks related to setting up and implementing the concentration measurement application.
Restrictions
This section does not cover building a concentration matrix. See Micro Motion Enhanced
Density Application: Theory, Configuration, and Use for detailed information on building a
matrix.
You cannot load a concentration matrix using the Field Communicator. If you need to load a
matrix, you must use ProLink III.
1. Enable the concentration measurement application using the Field Communicator
2. Set reference temperature values for specific gravity using the Field Communicator
3. Provide temperature data for concentration measurement using the Field Communicator
4. Modify matrix names and labels using the Field Communicator
5. Modify concentration measurement operational parameters using the
Field Communicator
6. Select the active concentration matrix using the Field Communicator
Enable the concentration measurement application using the Field Communicator
The concentration measurement application must be enabled before you can perform any setup. If the concentration measurement application was enabled at the factory, you do not need to enable it now.
1. Choose Overview > Device Information > Applications > Enable/Disable Applications.
Configuration and Use Manual 55
Configure process measurement
2. If the API referral application is enabled, disable it.
3. Enable the concentration measurement application.
Set reference temperature values for specific gravity using the Field Communicator
When Derived Variable is set to Specific Gravity, you must set the reference temperature to be used for density measurement and the reference temperature of water, and then verify the density of water at the configured reference temperature. The two reference temperature values affect specific gravity measurement.
Additionally, the two reference temperature values affect any concentration process variable that is calculated by equation, rather than by matrix, because the equations are based on specific gravity.
Typically, the two reference temperature values are the same, but this is not required.
The concentration measurement application and the API referral application cannot be enabled simultaneously.
Restriction
If Derived Variable is not set to Specific Gravity, do not change any of these values. These are set by the active concentration matrix. To check the setting of Derived Variable, choose Configure > Manual Setup > Measurements > Conc Measure (CM) > CM Configuration.
Important
Do not change the setting of Derived Variable. If you change the setting of Derived Variable, all existing concentration matrices will be deleted from transmitter memory.
Procedure
1. Choose Configure > Manual Setup > Measurements > Conc Measure (CM) > Configure Matrix.
2. Set Matrix Being Configured to the matrix you want to modify.
3. Choose Reference Conditions, then perform the following actions:
a. Set Reference Temp to the temperature to which line density will be corrected for
use in the specific gravity calculation.
b. Set Water Ref Temp to the water temperature that will be used in the specific
gravity calculation.
c. Set Water Ref Density to the density of water at the specified reference
temperature.
The transmitter automatically calculates the density of water at the specified temperature. The new value will be displayed the next time that transmitter memory is read. You can enter a different value if you want to.
Related information
Using equations to calculate specific gravity, °Baumé, °Brix, °Plato, and °Twaddell
56 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
Provide temperature data for concentration measurement using the Field Communicator
The concentration measurement application uses line temperature data in its calculations. You must decide how to provide this data, then perform the required configuration and setup. Temperature data from the on-board temperature sensor (RTD) is always available. You can set up an external temperature device and use external temperature data if you want to.
The temperature setup that you establish here will be used for all concentration measurement matrices on this meter.
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the internal RTD temperature in some areas and an external temperature in others. The transmitter stores the internal RTD temperature and the external temperature separately. However, the transmitter stores only one alternate temperature value, which may be either an external temperature or the configured fixed value. Accordingly, if you set up polling for temperature in one area, and digital communications in another, and configure a fixed temperature value in a third, the fixed value will be overwritten by polling and digital communications, and polling and digital communications will overwrite each other.
Prerequisites
If you plan to poll an external device, the primary mA output (Channel A) must be wired to support HART communications.
Procedure
Choose the method to be used to supply temperature data, and perform the required setup.
Method Description Setup
Internal RTD tem­perature data
Temperature data from the on­board temperature sensor (RTD) is used.
a. Choose Configure > Manual Setup > Measurements > External Inputs
> Temperature.
b. Set External Temperature to Disable.
Configuration and Use Manual 57
Configure process measurement
Method Description Setup
Polling The meter polls an external de-
vice for temperature data. This data will be available in addi­tion to the internal RTD tem­perature data.
Digital communica­tions
A host writes temperature data to the meter at appropriate in­tervals. This data will be availa­ble in addition to the internal RTD temperature data.
a. Choose Configure > Manual Setup > Measurements > External Inputs
b. Set External Temperature to Enable. c. Choose Configure > Manual Setup > Inputs/Outputs > External Device
d. Choose an unused polling slot. e. Set Poll Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
Poll as Secondary Other HART masters will be on the net-
f. Set External Device Tag to the HART tag of the external tem-
g. Set Polled Variable to Temperature.
a. Choose Configure > Manual Setup > Measurements > External Inputs
b. Set External Temperature to Enable. c. Perform the necessary host programming and communica-
> Temperature.
Polling.
network. The Field Communicator is not a HART master.
work. The Field Communicator is not a HART master.
perature device.
> Temperature.
tions setup to write temperature data to the meter at appro­priate intervals.
Postrequisites
Choose Service Tools > Variables > External Variables and verify the value for External Temperature.
Need help? If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
- Verify the wiring between the meter and the external device.
- Verify the HART tag of the external device.
For digital communications:
- Verify that the host has access to the required data.
- Verify that the host is writing to the correct register in memory, using the correct data
type.
If necessary, apply an offset.
58 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
Modify matrix names and labels using the Field Communicator
For convenience, you can change the name of a concentration matrix and the label used for its measurement unit. This does not affect measurement.
1. Choose Configure > Manual Setup > Measurements > Conc Measure (CM) > Configure Matrix.
2. Set Matrix Being Configured to the matrix you want to modify.
3. Set Matrix Name to the name to be used for the matrix.
4. Set Concentration Units to the label that will be used for the concentration unit.
5. If you set Concentration Units to Special, choose Concentration Label and enter the custom label.
Modify concentration measurement operational parameters using the Field Communicator
You can enable and disable extrapolation alerts, set extrapolation alert limits, and control matrix switching. These parameters control the behavior of the concentration measurement application but do not affect measurement directly. Additionally, for certain types of concentration measurement, you can select the calculation type to be used.
1. Choose Configure > Manual Setup > Measurements > Conc Measure (CM) > Configure Matrix.
2. Set Matrix Being Configured to the matrix you want to modify.
3. If applicable, set Equation Type to the type of calculation to be used.
Option Description
Specific Gravity
Baume Specific gravity is calculated as described above, and the result is used in the
Brix Specific gravity is calculated as described above and the result is used in the
Plato Specific gravity is calculated as described above, and the result is used n the
Twaddell Specific gravity is calculated as described above, and the result is used in the
This option is available only when Derived Variable is set to Specific Gravity.
4. Set up extrapolation alerts.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density or line temperature goes outside the range, the transmitter will extrapolate concentration values. However, extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is occurring, and can also be used to initiate matrix switching. Each concentration matrix has its own extrapolation alert settings.
Referred density is calculated from the matrix. The result is used in the specific gravity equation. The output is specific gravity.
°Baumé equation. The output is specific gravity and °Baumé (light or heavy).
°Brix equation. The output is specific gravity and °Brix.
°Plato equation. The output is specific gravity and °Plato.
°Twaddell equation. The output is specific gravity and °Twaddell.
Configuration and Use Manual 59
Configure process measurement
5. Choose Configure > Manual Setup > Measurements > Conc Measure (CM) > CM Configuration,
a. Set Extrapolation Alert Limit to the point, in percent, at which an extrapolation alert
will be posted.
b. Choose Online > Configure > Alert Setup > Concentration Measurement Alerts.
c. Enable or disable the high and low alerts for temperature and density, as desired.
Important
If you are using matrix switching, you must enable the appropriate extrapolation alerts.
Example: If Extrapolation Alert Limit is set to 5%, High Extrapolation Limit (Temperature) is enabled, and the matrix is built for a temperature range of 40 °F to 80 °F, an extrapolation alert will be posted if line temperature goes above 82 °F.
then enable or disable Matrix Switching as desired.
When matrix switching is enabled and an extrapolation alert occurs, the transmitter automatically switches from the matrix in Slot 1 to the matrix in Slot 2, or vice versa. This occurs only if no extrapolation alert would be generated by the other matrix. Matrix switching is not applicable to any other slots.
Related information
Using equations to calculate specific gravity, °Baumé, °Brix, °Plato, and °Twaddell Matrix switching
Select the active concentration matrix using the Field Communicator
You must select the concentration matrix to be used for measurement. Although the transmitter can store up to six concentration matrices, only one matrix can be used for measurement at any one time.
Set Active Matrix to the matrix you want to use.
Related information
Matrix switching
4.6.4 Using equations to calculate specific gravity, °Baumé, °Brix, °Plato, and °Twaddell
Whenever the derived variable is set to Specific Gravity, you have the option of using equations to calculate °Baumé, °Brix,°Plato, and °Twaddell, instead of matrix referral. Whenever the equation method is used, the active matrix is used to measure referred density. This value is used to calculate specific gravity. The result of the specific gravity calculation is then used in the equations used to calculate °Baumé, °Brix, °Plato, or °Twaddell.
Specific gravity is always calculated using the two reference temperatures that are specified during concentration measurement configuration.
60 Micro Motion® Fork Density Meters (FDM)
If you are measuring in °Baumé, the meter will automatically select the appropriate equation according to the specific gravity of the process fluid, and will switch equations when specific gravity crosses 1.0.
Equations used for specific gravity, °Baumé, °Brix, °Plato, and °TwaddellTable 4-6:
Current value of
Equation
Specific Gravity N/A
°Baumé 1.0 or greater
°Brix N/A
°Plato N/A
°Twaddell N/A
specific gravity Equation
(°Baumé heavy)
Less than 1.0 (°Baumé light)
ρ
SG =
°Baumé = 145
°Baumé =
°Brix = 318.906
RefTemp
ρ
WaterRefTemp
140
130
SG
145
SG
(
384.341 SG
66.1086
+
)
(
SG
)
2
°Plato =(668.72 × SG)− 463.37 −(205.347 × SG
°Twaddell = 200 ×(SG 1
)
Configure process measurement
2
)
SG
ρ
RefTemp
ρ
WaterRefTemp
Specific gravity of process fluid
Density of process fluid at Reference Temperature for Referred Density, as measured using the active concentration matrix
Density of water at Reference Temperature for Water
Related information
Set reference temperature values for specific gravity using ProLink III Set reference temperature values for specific gravity using the Field Communicator
4.6.5 Matrix switching
Matrix switching can be used to measure different process fluids without resetting the active matrix manually. Matrix switching can also be used to increase measurement accuracy.
When matrix switching is enabled, the meter automatically switches between the matrices in Slot 1 and Slot 2 whenever an extrapolation alert is present for the active matrix but would not be generated by the other matrix. For example:
The matrix in Slot 2 is active, the high-density extrapolation alert is enabled, and
matrix switching is enabled. Line density goes above the range of the matrix plus the extrapolation limit. The meter posts an alert, then checks the range of the matrix in Slot 1. No extrapolation alert would be posted, so the meter automatically switches to the matrix in Slot 1.
Configuration and Use Manual 61
Configure process measurement
The matrix in Slot 2 is active, the high-density extrapolation alert is enabled, and
You can control the conditions that trigger matrix switching by enabling or disabling specific extrapolation alerts. For example, if the low-density and high-density extrapolation alerts are enabled, but the low-temperature and high-temperature extrapolation alerts are disabled, matrix switching will be triggered only by changes in line density. Changes in line temperature will not trigger matrix switching.
Depending on your application, you may need to set up your matrices and extrapolation limits so that there is no overlap in density and/or temperature, or so that there is slight overlap.
On the display, the matrix number for the active matrix will alternately flash with the concentration and referred density units.
Example: Using matrix switching to measure different process fluids
The line may contain either of two process fluids, depending on the current product. The matrix in Slot 1 is appropriate for the first process fluid. The matrix in Slot 2 is appropriate for the second process fluid. Whenever the line is switched, an extrapolation alert is posted for the current matrix, and the meter automatically switches to use the other matrix.
matrix switching is enabled. Line density goes above the range of the matrix plus the extrapolation limit. The meter posts an alert, then checks the range of the matrix in Slot 1. The current line density would also generate an extrapolation alert for this matrix, so the meter does not switch.
To ensure that the correct matrix is used, there can be no overlap between the ranges of the two matrices. In other words:
If you are using density to trigger matrix switching, there can be no density values
that are within the ranges of both matrices, after the extrapolation limits are applied.
If you are using temperature to trigger matrix switching, there can be no
temperature values that are within the ranges of both matrices, after the extrapolation limits are applied.
If you are using both density and temperature to trigger matrix switching, there can
be no density or temperature values that are within the ranges of both matrices, after the extrapolation limits are applied.
Example: Using matrix switching to increase measurement accuracy
For some process fluids, measurement accuracy is increased when the matrix has a narrower temperature or density range. By using two matrices with adjacent or slightly overlapping ranges, increased accuracy is available over a wider range of process variation.
To ensure continuous process measurement, there should be no gap between the ranges, after the extrapolation limits are applied.
62 Micro Motion® Fork Density Meters (FDM)
Configure process measurement
4.6.6 Measuring Net Mass Flow Rate and Net Volume Flow Rate
Net Mass Flow Rate is calculated by multiplying concentration by the mass flow rate. Net Volume Flow Rate is calculated by multiplying concentration by the volume flow rate.
To measure Net Mass Flow Rate, the following are required:
A mass flow rate process variable, either measured or calculated, depending on
what is available on your meter
Concentration measurement configured and active, with Mass Concentration
(Density) set as the derived variable
To measure Net Volume Flow Rate, the following are required:
A volume flow rate process variable, either measured or calculated, depending on
what is available on your meter
Concentration measurement configured and active, with Volume Concentration
(Density) set as the derived variable

4.7 Set up flow rate measurement

The meter does not measure flow rate directly. However, you can provide volume flow rate data to the meter and use this data to calculate mass flow rate.
4.7.1 Set up flow rate measurement using ProLink III
Prerequisites
To calculate mass flow rate, you must be able to supply volume flow rate data to the
meter.
If you plan to poll an external device, the primary mA output must be wired to
support HART communications.
Procedure
1. Choose Device Tools > Configuration > I/O > Inputs > External Inputs.
2. Set Mass Flow (Calculated) to Enabled and click Apply.
3. Set Mass Flow Rate (Calculated) Unit to the unit in which the mass flow rate will be
reported.
4. Set Line Volume Flow Rate Unit to the units used by the external volume measurement
device
5. Set Line Volume Flow Source to the method to be used to retrieve volume flow data
and perform the required setup.
Configuration and Use Manual 63
Configure process measurement
Option Description Setup
Polling The meter polls an external de-
vice for volume flow data and calculates the equivalent mass flow rate.
Digital communica­tions
A host writes volume flow data to the meter at appropriate in­tervals, and the meter calcu­lates the equivalent mass flow rate.
Tip
A fixed value is not recommended. A fixed value may produce inaccurate process data.
a. Set Line Volume Flow Source to Poll for External Value. b. Set Polling Slot to an available slot. c. Set Polling Control to Poll as Primary or Poll as Secondary. d. Set External Device Tag to the HART tag of the volume flow
a. Set Line Volume Flow Source to Fixed Value or Digital Communica-
b. Perform the necessary host programming and communica-
The following process variables are now available:
Line Volume Flow Rate
Mass Flow Rate (Calculated)
measurement device.
tions.
tions setup to write volume flow data to the meter at appro­priate intervals.
If the concentration measurement application is enabled and configuration requirements are met, one of the following process variables is also available:
Net Volume Flow Rate
Net Mass Flow Rate
Postrequisites
To verify the volume flow rate, view the value displayed on the ProLink III main window, in the Inputs group.
To verify the calculated mass flow rate, set up one of the ProLink III gauges to display it.
Need help? If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
- Verify the wiring between the meter and the external device.
- Verify the HART tag of the external device.
For digital communications:
- Verify that the host has access to the required data.
- Verify that the host is writing to the correct register in memory, using the correct data
type.
If necessary, apply an offset.
Related information
Measuring Net Mass Flow Rate and Net Volume Flow Rate
64 Micro Motion® Fork Density Meters (FDM)
4.7.2 Set up flow rate measurement using the Field Communicator
Prerequisites
To calculate mass flow rate, you must be able to supply volume flow rate data to the
meter.
If you plan to poll an external device, the primary mA output must be wired to
support HART communications.
Procedure
1. Choose Configure > Manual Setup > Measurements > Mass (Calculated) and set Mass Flow
Rate Unit to the unit in which the mass flow rate will be reported.
2. Choose Configure > Manual Setup > Measurements > Volume and set Volume Flow Rate Unit
to the unit used by the external volume measurement device.
3. Choose Configure > Manual Setup > Measurements > External Inputs > Volumeand set Volume
Flow Source to Enabled.
4. Choose the method to be used to supply volume flow rate data, and perform the
required setup.
Configure process measurement
Method Description Setup
Polling The meter polls an external de-
vice for volume flow rate data.
Digital communica­tions
A host writes volume flow rate data to the meter at appropri­ate intervals.
a. Choose Configure > Manual Setup > Inputs/Outputs > External Device
Polling. b. Choose an unused polling slot. c. Set Poll Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
Poll as Secondary Other HART masters will be on the net-
d. Set External Device Tag to the HART tag of the external pres-
sure device. e. Set Polled Variable to Volume from Mag/Vortex Meter.
a. Perform the necessary host programming and communica-
tions setup to write volume flow rate data to the transmitter
at appropriate intervals.
network. The Field Communicator is not a HART master.
work. The Field Communicator is not a HART master.
Tip
A fixed value is not recommended. A fixed value may produce inaccurate process data.
Configuration and Use Manual 65
Configure process measurement
The following process variables are now available:
Line Volume Flow Rate
Mass Flow Rate (Calculated)
If the concentration measurement application is enabled and configuration requirements are met, one of the following process variables is also available:
Net Volume Flow Rate
Net Mass Flow Rate
Postrequisites
Choose Service Tools > Variables > External Variables and verify the values for External Volume Flow Rate and Mass Flow Rate (Calculated).
Need help? If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
For digital communications:
If necessary, apply an offset.
- Verify the wiring between the meter and the external device.
- Verify the HART tag of the external device.
- Verify that the host has access to the required data.
- Verify that the host is writing to the correct register in memory, using the correct data
type.
Related information
Measuring Net Mass Flow Rate and Net Volume Flow Rate
66 Micro Motion® Fork Density Meters (FDM)

Configure device options and preferences

5 Configure device options and
preferences
Topics covered in this chapter:

Configure the transmitter display

Enable or disable the Acknowledge All Alerts display command
Configure security for the display menus
Configure alert handling
Configure informational parameters
5.1 Configure the transmitter display
You can control the process variables shown on the display and a variety of display behaviors.
Configure the language used for the display (Section 5.1.1)
Configure the process variables and diagnostic variables shown on the display
(Section 5.1.2)
Configure the number of decimal places (precision) shown on the display
(Section 5.1.3)
Configure the refresh rate of data shown on the display (Section 5.1.4)
Enable or disable automatic scrolling through the display variables (Section 5.1.5)
5.1.1 Configure the language used for the display
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > LANG
ProLink III Device Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Language
Overview
Display Language controls the language used for process data and menus on the display.
Procedure
Select the language you want to use.
The languages available depend on your transmitter model and version.
Configuration and Use Manual 67
Configure device options and preferences
5.1.2 Configure the process variables and diagnostic variables shown on the display
Display Not available
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Display Variables
Overview
You can control the process variables and diagnostic variables shown on the display, and the order in which they appear. The display can scroll through up to 15 variables in any order you choose. In addition, you can repeat variables or leave slots unassigned.
Restriction
You cannot set Display Variable 1 to None or to a diagnostic variable. Display Variable 1 must be set to a process variable.
Procedure
For each display variable you want to change, assign the process variable you want to use.
5.1.3 Configure the number of decimal places (precision) shown on the display
Display Not available
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Decimal Places
Overview
You can specify the number of decimal places (precision) that are shown on the display for each process variable or diagnostic variable. You can set the precision independently for each variable.
The display precision does not affect the actual value of the variable or the value used in calculations.
Procedure
1. Select a variable.
2. Set Number of Decimal Places to the number of decimal places you want shown when
the process variable or diagnostic variable appears on the display.
For temperature and density process variables, the default value is 2 decimal places. For all other variables, the default value is 4 decimal places. The range is 0 to 5.
68 Micro Motion® Fork Density Meters (FDM)
Configure device options and preferences
Tip
The lower the precision, the greater the change must be for it to be reflected on the display. Do not set the precision too low or too high to be useful.
5.1.4 Configure the refresh rate of data shown on the display
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > RATE
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Display Behavior > Refresh Rate
Overview
You can set Refresh Rate to control how frequently data is refreshed on the display.
Procedure
Set Refresh Rate to the desired value.
The default value is 1000 milliseconds. The range is 100 milliseconds to 10,000 milliseconds (10 seconds).
5.1.5 Enable or disable automatic scrolling through the display variables
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > AUTO SCRLL
ProLink III Device Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Display Behavior > Auto Scroll
Overview
You can configure the display to automatically scroll through the configured display variables or to show a single display variable until the operator activates Scroll. When you set automatic scrolling, you can also configure the length of time each display variable is displayed.
Procedure
1. Enable or disable Auto Scroll as desired.
Option Description
Enabled The display automatically scrolls through each display variable as specified
by Scroll Rate. The operator can move to the next display variable at any time using Scroll.
Configuration and Use Manual 69
Configure device options and preferences
Option Description
Disabled (de­fault)
The display shows Display Variable 1 and does not scroll automatically. The operator can move to the next display variable at any time using Scroll.
2. If you enabled Auto Scroll, set Scroll Rate as desired.
The default value is 10 seconds.
Tip
Scroll Rate may not be available until you apply Auto Scroll.
5.2 Enable or disable the Acknowledge All Alerts display command
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > ACK
ProLink III Device Tools > Configuration > Transmitter Display > Ack All
Field Communicator Configure > Manual Setup > Display > Display Menus > Acknowledge All
Overview
You can configure whether or not the operator can use a single command to acknowledge all alerts from the display.
Procedure
1. Ensure that the alert menu is accessible from the display.
To acknowledge alerts from the display, operators must have access to the alert menu.
2. Enable or disable Acknowledge All Alerts as desired.
Option Description
Enabled (default) Operators can use a single display command to acknowledge all alerts at
once.
Disabled Operators cannot acknowledge all alerts at once. Each alert must be ac-
knowledged separately.
70 Micro Motion® Fork Density Meters (FDM)
Configure device options and preferences

5.3 Configure security for the display menus

Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Display Menus
Overview
You can control operator access to different sections of the display off-line menu. You can also configure a passcode to control access.
Procedure
1. To control operator access to the maintenance section of the off-line menu, enable
or disable Off-Line Menu.
Option Description
Enabled (default) Operator can access the maintenance section of the off-line menu. This
access is required for configuration and calibration, including Known Density Verification.
Disabled Operator cannot access the maintenance section of the off-line menu.
2. To control operator access to the alert menu, enable or disable Alert Menu.
Option Description
Enabled (default) Operator can access the alert menu. This access is required to view and
acknowledge alerts, but is not required for Known Density Verification, configuration, or calibration.
Disabled Operator cannot access the alert menu.
Note
The transmitter status LED changes color to indicate that there are active alerts, but does not show specific alerts.
3. To require a passcode for access to the off-line menu, enable or disable Off-Line
Password.
Option Description
Enabled Operator is prompted for the off-line passcode at entry to the off-line
menu.
Disabled (default) No passcode is required for entry to the off-line menu.
4. Set Off-Line Password to the desired value.
Configuration and Use Manual 71
Configure device options and preferences
The default value is 1234. The range is 0000 to 9999.
Tip
Record your passcode for future reference.

5.4 Configure alert handling

The alert handling parameters control the transmitter’s response to process and device conditions.
Configure Fault Timeout (Section 5.4.1)
Configure Alert Severity (Section 5.4.2)
5.4.1 Configure Fault Timeout
Display Not available
ProLink III Device Tools > Configuration > Fault Processing
Field Communicator Configure > Alert Setup > Alert Severity > Fault Timeout
Overview
Fault Timeout controls the delay before fault actions are performed.
Restriction
Fault Timeout is applied only to the following alerts (listed by Status Alert Code): A003, A004, A008, A016, A033. For all other alerts, fault actions are performed as soon as the alert is detected.
Procedure
Set Fault Timeout as desired.
The default value is 0 seconds. The range is 0 to 60 seconds.
If you set Fault Timeout to 0, fault actions are performed as soon as the alert condition is detected.
The fault timeout period begins when the transmitter detects an alert condition. During the fault timeout period, the transmitter continues to report its last valid measurements.
If the fault timeout period expires while the alert is still active, the fault actions are performed. If the alert condition clears before the fault timeout expires, no fault actions are performed.
72 Micro Motion® Fork Density Meters (FDM)
5.4.2 Configure Alert Severity
Display Not available
ProLink III Device Tools > Configuration > Alert Severity
Field Communicator Configure > Alert Setup > Alert Severity > Change Alert Severity
Overview
Use Alert Severity to control the fault actions that the transmitter performs when it detects an alert condition.
Restrictions
For some alerts, Alert Severity is not configurable.
For some alerts, Alert Severity can be set only to two of the three options.
Tip
Micro Motion recommends using the default settings for Alert Severity unless you have a specific requirement to change them.
Configure device options and preferences
Procedure
1. Select a status alert.
2. For the selected status alert, set Alert Severity as desired.
Option Description
Fault Actions when fault is detected:
The alert is posted to the Alert List.
Outputs go to the configured fault action (after Fault Timeout has expired, if
applicable).
Digital communications go to the configured fault action (after Fault Timeout
has expired, if applicable).
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
Outputs return to normal behavior.
Digital communications return to normal behavior.
The status LED returns to green.
Informa­tional
Ignore No action
Actions when fault is detected:
The alert is posted to the Alert List.
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
The status LED returns to green.
Configuration and Use Manual 73
Configure device options and preferences
Status alerts and options for Status Alert Severity
Status alerts and Status Alert SeverityTable 5-1:
Alert number Alert title Default severity
A001 EEPROM Error Fault No
A002 RAM Error Fault No
A003 No Sensor Response Fault Yes
A004 Temperature Overrange Fault No
A006 Characterization Required Fault Yes
A008 Density Overrange Fault Yes
A009 Transmitter Initializing/Warming Up
A010 Calibration Failure Fault No
A014 Transmitter Failure Fault No
A016 Sensor Temperature (RTD) Failure Fault Yes
A020 Calibration Factors Missing Fault Yes
A021 Transmitter/Sensor/Software Mis-
A029 Internal Electronics Failure Fault No
A030 Incorrect Board Type Fault No
A033 Insufficient Pickoff Signal Fault Yes
A037 Sensor Check Failed Fault Yes
A038 Time Period Signal Out of Range Fault No
A100 mA Output 1 Saturated Informational To Informational or
A101 mA Output 1 Fixed Informational To Informational or
A102 Drive Overrange Informational Yes
A104 Calibration in Progress Informational To Informational or
A105 Two-Phase Flow Informational Yes
A106 Burst Mode Enabled Informational To Informational or
A107 Power Reset Occurred Informational Yes
A113 mA Output 2 Saturated Informational To Informational or
A114 mA Output 2 Fixed Informational To Informational or
User can reset severity
Ignore Yes
or Significant Process Instability
Fault No
match
Ignore only
Ignore only
Ignore only
Ignore only
Ignore only
Ignore only
74 Micro Motion® Fork Density Meters (FDM)
Configure device options and preferences
Status alerts and Status Alert Severity (continued)Table 5-1:
User can reset
Alert number Alert title Default severity
A115 No External Input or Polled Data Informational To Informational or
A116 Temperature Overrange (API Referral) Informational Yes
A117 Density Overrange (API Referral) Informational Yes
A118 Discrete Output 1 Fixed Informational To Informational or
A120 Curve Fit Failure (Concentration) Informational No
A121 Extrapolation Alert (Concentration) Informational Yes
A122 Pressure Overrange (API Referral) Informational Yes
A132 Sensor Simulation Active Informational Yes
A133 EEPROM Error (Display) Informational Yes
A136 Incorrect Display Type Informational Yes
severity
Ignore only
Ignore only

5.5 Configure informational parameters

Display Not available
ProLink III Device Tools > Configuration > Meter Information
Field Communicator Configure > Manual Setup > Info Parameters
Overview
The informational parameters can be used to identify or describe your meter. They are not used in process measurement and they are not required.
Procedure
Enter data as desired.
Parameter Description
Meter Serial Num­ber
Message A message to be stored in device memory. The message can contain up to
Descriptor A description of this device. The description can contain up to 16 characters.
Date A static date (not updated by the meter). Enter the date in the form mm/dd/yyyy.
Flange Type The sensor flange type for this device. Obtain the value from the documents
The serial number of the device. Enter the value from the device tag.
32 characters.
shipped with the device or from a code in the model number.
Configuration and Use Manual 75
Configure device options and preferences
Tips
The Field Communicator does not support all informational parameters. If you need to configure
all of the informational parameters, use ProLink III.
The Field Communicator allows you to configure HART Tag and HART Long Tag from this location.
These parameters are replicated from Configure > Manual Setup > HART > Communications. These parameters are used in HART communications.
76 Micro Motion® Fork Density Meters (FDM)

Integrate the meter with the control system

6 Integrate the meter with the control
system
Topics covered in this chapter:

Configure Channel B

Configure the mA output
Configure the discrete output
Configure an enhanced event
Configure HART/Bell 202 communications
Configure Modbus communications
Configure Digital Communications Fault Action
6.1 Configure Channel B
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CONFIG CH B
ProLink III Device Tools > Configuration > I/O > Channels
Field Communicator Configure > Manual Setup > Inputs/Outputs > Channels > Channel B
Overview
Depending on your device, you can configure Channel B to operate as either an mA output or a discrete output.
Restriction
You cannot configure Channel B on the FDM TPS device. On this device, Channel B always operates as a TPS output.
Prerequisites
The configuration of Channel B must match the wiring. See the installation manual for your device.
To avoid causing process errors:
Configure Channel B before configuring the mA output or discrete output.
Before changing the channel configuration, ensure that all control loops affected by
the channel are under manual control.
Procedure
Set Channel B as desired.
Configuration and Use Manual 77
Integrate the meter with the control system
Option Description
mA output Channel B will operate as the secondary mA output.
Discrete output Channel B will operate as a discrete output.

6.2 Configure the mA output

The mA output is used to report the configured process variable. The mA output parameters control how the process variable is reported.
The FDM mA device has two mA outputs: Channel A and Channel B. Both outputs are fully configurable.
The FDM DO device has one mA output: Channel A. The output is fully configurable.
The FDM TPS device has one mA output: Channel A. The output is fully configurable.
Important
Whenever you change an mA output parameter, verify all other mA output parameters before returning the meter to service. In some situations, the transmitter automatically loads a set of stored values, and these values may not be appropriate for your application.
Configure mA Output Process Variable (Section 6.2.1)
Configure Lower Range Value (LRV) and Upper Range Value (URV) (Section 6.2.2)
Configure Added Damping (Section 6.2.3)
Configure mA Output Fault Action and mA Output Fault Level (Section 6.2.4)
6.2.1 Configure mA Output Process Variable
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CONFIG MAO 1 > AO 1 SRC
OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > MAO 2 > CONFIG MAO 2 > AO 2 SRC
ProLink III Device Tools > Configuration > I/O > Outputs > mA Output > mA Output 1 > Source
Device Tools > Configuration > I/O > Outputs > mA Output > mA Output 2 > Source
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output 1 > Primary Variable
Configure > Manual Setup > Inputs/Outputs > mA Output 2 > Secondary Variable
Overview
Use mA Output Process Variable to select the variable that is reported over the mA output.
Prerequisites
If you are using the HART variables, be aware that changing the configuration of mA Output Process Variable will change the configuration of the HART Primary Variable (PV) and/or the
HART Secondary Variable (SV).
78 Micro Motion® Fork Density Meters (FDM)
Integrate the meter with the control system
Procedure
Set mA Output Process Variable as desired.
Default settings are shown in the following table.
Default settings for mA Output Process VariableTable 6-1:
Default process variable assign-
Device Channel mA output
FDM mA Channel A Primary mA output Density
Channel B Secondary mA output Temperature
FDM DO Channel A Primary mA output Density
FDM TPS Channel A Primary mA output Temperature
ment
Postrequisites
If you changed the setting of mA Output Process Variable, verify the settings of Lower Range Value (LRV) and Upper Range Value (URV).
Options for mA Output Process Variable
The transmitter provides a basic set of options for mA Output Process Variable, plus several application-specific options. Different communications tools may use different labels for the options.
Options for mA Output Process VariableTable 6-2:
Label
Process variable
Standard
Line Density DENS Line Density Density
Line Temperature TEMP Line Temperature Temperature
Line Temperature (External) EXT T Line Temperature (External
Line Pressure (External) EXT P Line Pressure (External or
Drive Gain DGAIN Drive Gain Drive Gain
Sensor Time Period TP B Sensor Time Period Sensor Time Period
User-Defined Calculation Output
Volume Flow Rate (External) MAG V Volume Flow Rate (External) Volume from Mag/Vortex
Mass Flow Rate (Calculated) MAG M Mass Flow Rate (Calculated) Calculated Mass Flow from
Display ProLink III Field Communicator
External Temperature
or Fixed)
External Pressure
Fixed)
UCALC User-Defined Calculation
Output
User-Defined Calculation Output
Meter
Mag Meter Input
Configuration and Use Manual 79
Integrate the meter with the control system
Options for mA Output Process Variable (continued)Table 6-2:
Label
Process variable
API referral
Referred Density (API) RDENS Referred Density (API) Density at Reference (API)
Concentration measurement
Specific Gravity SG Specific Gravity Specific Gravity (CM)
Concentration CONC Concentration Concentration (CM)
Referred Density (Concen­tration)
Net Mass Flow Rate NET M Net Mass Flow Rate Net Mass Flow Rate (CM)
Net Volume Flow Rate NET V Net Volume Flow Rate Net Volume Flow Rate (CM)
Display ProLink III Field Communicator
RDENS Referred Density (Concen-
tration)
Density at Reference (CM)
6.2.2 Configure Lower Range Value (LRV) and Upper Range Value (URV)
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CONFIG MAO 1 > 4 mA
OFF-LINE MAINT > OFF-LINE CONFG > IO > CONFIG MAO 1 > 20 mA
OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > MAO 2 > CONFIG MAO 2 > 4 mA
OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > MAO 2 > CONFIG MAO 2 > 20 mA
ProLink III Device Tools > Configuration > I/O > Outputs > mA Output > mA Output 1 > Lower Range Value
Device Tools > Configuration > I/O > Outputs > mA Output > mA Output 1 > Upper Range Vaue
Device Tools > Configuration > I/O > Outputs > mA Output > mA Output 2 > Lower Range Value
Device Tools > Configuration > I/O > Outputs > mA Output > mA Output 2 > Upper Range Vaue
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output 1 > mA Output Settings > PV LRV
Configure > Manual Setup > Inputs/Outputs > mA Output 1 > mA Output Settings > PV URV
Configure > Manual Setup > Inputs/Outputs > mA Output 2 > mA Output Settings > SV LRV
Configure > Manual Setup > Inputs/Outputs > mA Output 2 > mA Output Settings > SV URV
Overview
The Lower Range Value (LRV) and Upper Range Value (URV) are used to scale the mA output, that is, to define the relationship between mA Output Process Variable and the mA output level.
Prerequisites
Ensure that mA Output Process Variable is set to the desired process variable. Each process variable has its own set of LRV and URV values. When you change the values of LRV and URV, you are configuring values for the currently assigned mA output process variable.
80 Micro Motion® Fork Density Meters (FDM)
Integrate the meter with the control system
Ensure that the measurement unit for the configured process variable has been set as desired.
Procedure
Set LRV and URV as desired.
LRV is the value of mA Output Process Variable represented by an output of 4 mA. The
default value for LRV depends on the setting of mA Output Process Variable. Enter LRV in the measurement units that are configured for mA Output Process Variable.
URV is the value of mA Output Process Variable represented by an output of 20 mA. The
default value for URV depends on the setting of mA Output Process Variable. Enter URV in the measurement units that are configured for mA Output Process Variable.
Tip
For best performance:
Set LRVLSL (lower sensor limit).
Set URVUSL (upper sensor limit).
Set these values so that the difference between URV and LRV is ≥ Min Span (minimum span).
Defining URV and LRV within the recommended values for Min Span, LSL, and USL ensures that the resolution of the mA output signal is within the range of the bit precision of the D/A converter.
Note
You can set URV below LRV. For example, you can set URV to 50 and LRV to 100.
The mA output uses a range of 4–20 mA to represent mA Output Process Variable. Between LRV and URV, the mA output is linear with the process variable. If the process variable drops below LRV or rises above URV, the transmitter posts an output saturation alert.
6.2.3 Configure Added Damping
Display Not available
ProLink III Device Tools > Configuration > I/O > Outputs > mA Output > mA Output 1 > Added Damping
Device Tools > Configuration > I/O > Outputs > mA Output > mA Output 2 > Added Damping
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output 1 > mA Output Settings > PV Added Damping
Configure > Manual Setup > Inputs/Outputs > mA Output 2 > mA Output Settings > SV Added Damping
Overview
Added Damping controls the amount of damping that will be applied to the mA output.
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Configuration and Use Manual 81
Integrate the meter with the control system
Added Damping affects the reporting of mA Output Process Variable through the mA output only. It does not affect the reporting of that process variable via any other method (e.g., a frequency output or digital communications), or the value of the process variable used in calculations.
Note
Added Damping is not applied if the mA output is fixed (for example, during loop testing) or if the mA output is reporting a fault. Added Damping is applied while sensor simulation is active.
Procedure
Set Added Damping to the desired value.
The default value is 0.0 seconds. The range is 0.0 to 440 seconds.
When you specify a value for Added Damping, the transmitter automatically rounds the value down to the nearest valid value.
Interaction between mA Output Damping and process variable damping
When mA Output Process Variable is set to density or temperature, Added Damping interacts with Density Damping or Temperature Damping.
Example: Damping interaction
Configuration:
mA Output Process Variable = Density
Density Damping = 1 second
Added Damping = 2 seconds
Result: A change in density will be reflected in the mA output over a time period that is greater than 3 seconds. The exact time period is calculated by the transmitter according to internal algorithms which are not configurable.
Related information
Interaction between Density Damping and Added Damping
6.2.4 Configure mA Output Fault Action and mA Output Fault Level
Display Not available
ProLink III Device Tools > Configuration > I/O > Outputs > mA Output > mA Output 1 > Fault Action
Device Tools > Configuration > I/O > Outputs > mA Output > mA Output 2 > Fault Action
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output 1 > mAO1 Fault Settings > MAO1 Fault Action
Configure > Manual Setup > Inputs/Outputs > mA Output 2 > MAO2 Fault Settings > MAO2 Fault Action
82 Micro Motion® Fork Density Meters (FDM)
Integrate the meter with the control system
Overview
mA Output Fault Action controls the behavior of the mA output if the transmitter encounters an internal fault condition.
Note
For some faults only: If Fault Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed.
Procedure
1. Set mA Output Fault Action to the desired value.
The default setting is Downscale.
Restriction
If Digital Communications Fault Action is set to NAN (not a number), you cannot set mA Output Fault Action to None. If you try to do this, the device will not accept the configuration.
2. If you set mA Output Fault Action to Upscale or Downscale, set mA Output Fault Level as
desired.
Postrequisites
CAUTION!
If you set mA Output Fault Action to None, be sure to set Digital Communications Fault Action to None. If you do not, the output will not report actual process data, and this may result in measurement errors or unintended consequences for your process.
Options for mA Output Fault Action and mA Output Fault Level
Options for mA Output Fault Action and mA Output Fault LevelTable 6-3:
Option mA output behavior mA Output Fault Level
Upscale Goes to the configured fault level Default: 21.5 mA
Range: 21.0 to 21.5 mA
Downscale (default) Goes to the configured fault level Default: 3.2 mA
Range: 3.2 to 3.6 mA
Internal Zero Goes to the mA output level associated
with a process variable value of 0 (zero), as determined by Lower Range Value and
Upper Range Value settings
None Tracks data for the assigned process vari-
able; no fault action
Not applicable
Not applicable
Configuration and Use Manual 83
Integrate the meter with the control system

6.3 Configure the discrete output

The discrete output is used to report specific meter or process conditions. The discrete output parameters control which condition is reported and how it is reported. Depending on your purchase option, your transmitter may have one discrete output or no discrete outputs.
Important
Whenever you change a discrete output parameter, verify all other discrete output parameters before returning the meter to service. In some situations, the transmitter automatically loads a set of stored values, and these values may not be appropriate for your application.
Configure Discrete Output Source (Section 6.3.1)
Configure Discrete Output Polarity (Section 6.3.2)
Configure Discrete Output Fault Action (Section 6.3.3)
6.3.1 Configure Discrete Output Source
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > DO > CONFIG DO > DO SRC
ProLink III Device Tools > Configuration > I/O > Outputs > Discrete Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Source
Overview
Discrete Output Source controls which device condition or process condition is reported via the discrete output.
Procedure
Set Discrete Output Source to the desired option.
The default setting for Discrete Output Source is Fault.
Options for Discrete Output Source
Options for Discrete Output SourceTable 6-4:
Label
Option
Enhanced Event 1–5 Enhanced Event 1
Calibration in Progress Calibration in Progress Calibration in Progress ON Site-specific
ProLink III Field Communicator
Enhanced Event 1
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
State
ON Site-specific
OFF 0 V
Discrete output volt­age
84 Micro Motion® Fork Density Meters (FDM)
Integrate the meter with the control system
Options for Discrete Output Source (continued)Table 6-4:
Label
Option
Fault (default) Fault Indicator Fault ON Site-specific
Important
This table assumes that Discrete Output Polarity is set to Active High. If Discrete Output Polarity is set to Active Low, reverse the voltage values.
State
OFF 0 V
OFF 0 V
Discrete output volt­ageProLink III Field Communicator
6.3.2 Configure Discrete Output Polarity
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > DO > CONFIG DO > DO POLAR
ProLink III Device Tools > Configuration > I/O > Outputs > Discrete Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Polarity
Overview
Discrete outputs have two states: ON (active) and OFF (inactive). Two different voltage levels are used to represent these states. Discrete Output Polarity controls which voltage level represents which state.
Procedure
Set Discrete Output Polarity as desired.
The default setting is Active High.
Options for Discrete Output Polarity
Options for Discrete Output PolarityTable 6-5:
Polarity Description
Active High When asserted (condition tied to DO is true), the cir-
cuit draws as much current as it can, up to a maximum of 10 mA.
When not asserted (condition tied to DO is false), the
circuit draws less than 1 mA.
Configuration and Use Manual 85
Integrate the meter with the control system
Options for Discrete Output Polarity (continued)Table 6-5:
Polarity Description
Active Low When asserted (condition tied to DO is true), the cir-
cuit draws less than 1 mA.
When not asserted (condition tied to DO is false), the
circuit draws as much current as it can, up to a maxi­mum of 10 mA.
6.3.3 Configure Discrete Output Fault Action
Display Not available
ProLink III Device Tools > Configuration > Fault Processing
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Fault Action
Overview
Discrete Output Fault Action controls the behavior of the discrete output if the transmitter encounters an internal fault condition.
Note
For some faults only: If Fault Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed.
CAUTION!
Do not use Discrete Output Fault Action as a fault indicator. If you do, you may not be able to distinguish a fault condition from a normal operating condition. If you want to use the discrete output as a fault indicator, set Discrete Output Source to Fault and set Discrete Output Fault Action to None.
Procedure
Set Discrete Output Fault Action as desired.
The default setting is None.
86 Micro Motion® Fork Density Meters (FDM)
Options for Discrete Output Fault Action
Options for Discrete Output Fault ActionTable 6-6:
Discrete output behavior
Integrate the meter with the control system
Label
Upscale Fault: discrete output is ON
Downscale Fault: discrete output is OFF
None (default) Discrete output is controlled by its assignment
Polarity=Active High Polarity=Active Low
(site-specific voltage)
No fault: discrete output is con-
trolled by its assignment
(0 V)
No fault: discrete output is con-
trolled by its assignment
Fault indication with the discrete output
To indicate faults via the discrete output, set Discrete Output Source to Fault. Then, if a fault occurs, the discrete output is always ON and the setting of Discrete Output Fault Action is ignored.

6.4 Configure an enhanced event

Display Not available
ProLink III Device Tools > Configuration > Events > Enhanced Events
Field Communicator Configure > Alert Setup > Enhanced Events
Fault: discrete output is OFF
(0 V)
No fault: discrete output is con-
trolled by its assignment
Fault: discrete output is ON
(site-specific voltage)
No fault: discrete output is con-
trolled by its assignment
Overview
An enhanced event is used to provide notification of process changes. An enhanced event occurs (is ON) if the real-time value of a user-specified process variable moves above (HI) or below (LO) a user-defined setpoint, or in range (IN) or out of range (OUT) with respect to two user-defined setpoints. You can define up to five enhanced events.
Procedure
1. Select the event that you want to configure.
2. Specify Event Type.
Configuration and Use Manual 87
Integrate the meter with the control system
Options Description
HI x > A
LO x < A
IN A ≤ x ≤ B
OUT x ≤ A or x ≥ B
3. Assign a process variable to the event.
4. Set values for the required setpoints.
The event occurs when the value of the assigned process variable (x) is greater than the setpoint (Setpoint A), endpoint not included.
The event occurs when the value of the assigned process variable (x) is less than the setpoint (Setpoint A), endpoint not included.
The event occurs when the value of the assigned process variable (x) is in range, that is, between Setpoint A and Setpoint B, endpoints included.
The event occurs when the value of the assigned process variable (x) is out of range, that is, less than Setpoint A or greater than Setpoint B, end- points included.
For HI and LO events, set Setpoint A.
For IN and OUT events, set Setpoint A and Setpoint B.
5. (Optional) Configure a discrete output to switch states in response to the event status.
Related information
Configure Discrete Output Source

6.5 Configure HART/Bell 202 communications

HART/Bell 202 communications parameters support HART communications with the transmitter's primary mA terminals over a HART/Bell 202 network.
Configure basic HART parameters (Section 6.5.1)
Configure HART variables (PV, SV, TV, QV) (Section 6.5.2)
Configure burst communications (Section 6.5.3)
6.5.1 Configure basic HART parameters
Display OFF-LINE MAINT > OFF-LINE CONFG > HART
ProLink III Device Tools > Configuration > Meter Information
Device Tools > Configuration > Communications > Communications (HART)
Field Communicator Configure > Manual Setup > HART > Communications
88 Micro Motion® Fork Density Meters (FDM)
Integrate the meter with the control system
Overview
Basic HART parameters include the HART address, HART tags, and the operation of the primary mA output.
Restrictions
Your device supports HART 7. If you are using HART 5, HART Long Tag is not available.
HART Tag, HART Long Tag, and mA Output Action are not configurable from the display.
Procedure
1. Set HART Address to a unique value on your network.
Valid address values are between 0 and 15. The default address (0) is typically used unless you are in a multidrop environment.
Tip
Devices using HART protocol to communicate with the transmitter may use either HART Address or HART Tag (Software Tag) to identify the transmitter. Configure either or both, as
required by your other HART devices.
2. Set HART Long Tag to a unique value on your network.
3. Ensure that mA Output Action is configured appropriately.
Options Description
Enabled (Live) The primary mA output reports process data as configured. This is
the appropriate setting for most applications.
Disabled (Fixed) The primary mA output is fixed at 4 mA and does not report process
data.
Important
If you use ProLink II or ProLink III to set HART Address to 0, the program automatically enables mA Output Action. If you use ProLink II or ProLink III to set HART Address to any other value, the
program automatically disables mA Output Action. This is designed to make it easier to configure the transmitter for legacy behavior. Always verify mA Output Action after setting HART Address.
6.5.2 Configure HART variables (PV, SV, TV, QV)
Display Not available
ProLink III Device Tools > Configuration > Communications > Communications (HART)
Field Communicator Configure > Manual Setup > Inputs/Outputs > Variable Mapping
Configuration and Use Manual 89
Integrate the meter with the control system
Overview
The HART variables are a set of four variables predefined for HART use. The HART variables include the Primary Variable (PV), Secondary Variable (SV), Tertiary Variable (TV), and Quaternary Variable (QV). You can assign specific process variables to the HART variables, and then use standard HART methods to read or broadcast the assigned process data.
Tip
The Tertiary Variable and Quaternary Variable are also called the Third Variable (TV) and Fourth Variable (FV).
Restriction
On some devices, the PV is fixed to a specific process variable and cannot be changed.
Related information
Set up the API referral application
Options for HART variables
Options for HART variablesTable 6-7:
Primary Varia-
Process variable
Standard
Line Density
Line Temperature
Line Temperature (External)
Line Pressure (External)
Drive Gain
Sensor Time Period
Volume Flow Rate (External)
Mass Flow Rate (Calculated)
User-Defined Calculation Output
Board Temperature
Input Voltage
API referral
Referred Density (API)
Concentration measurement
Specific Gravity
Concentration
Referred Density (Concentration)
ble (PV)
Secondary Variable (SV)
Third Variable (TV)
Fourth Varia­ble (QV )
90 Micro Motion® Fork Density Meters (FDM)
Integrate the meter with the control system
Options for HART variables (continued)Table 6-7:
Primary Varia-
Process variable
Net Mass Flow
Net Volume Flow
ble (PV)
Secondary Variable (SV)
Third Variable (TV)
Fourth Varia­ble (QV )
Interaction of HART variables and transmitter outputs
The HART variables are automatically reported through specific transmitter outputs. They may also be reported through HART burst mode, if enabled on your transmitter.
Restriction
One some devices, the PV and the primary mA output are fixed to a specific process variable and cannot be changed.
HART variables and transmitter outputsTable 6-8:
HART variable Reported via Comments
Primary Variable (PV) Primary mA output If one assignment is changed, the other is changed auto-
matically, and vice versa.
Secondary Variable (SV) Secondary mA output, if
present on your transmitter
Tertiary Variable (TV) Not associated with an out-
put
Quaternary Variable (QV) Not associated with an out-
put
If you have a secondary mA output: If one assignment is changed, the other is changed automatically.
If you do not have a secondary mA output: The SV must be configured directly, and the value of the SV is available on­ly via digital communications.
The TV must be configured directly, and the value of the TV is available only via digital communications.
The QV must be configured directly, and the value of the QV is available only via digital communications.
6.5.3 Configure burst communications
Burst mode is a mode of communication during which the transmitter regularly broadcasts HART digital information to the network via the primary mA output.
Restriction
Burst communications, including trigger mode and event notification, are not available on HART/ RS-485. These features are supported only on HART/Bell 202.
Configure HART burst messages
Configure HART trigger mode
Configure HART event notification
Configuration and Use Manual 91
Integrate the meter with the control system
Configure HART burst messages
Display Not available
ProLink III Device Tools > Configuration > Communications > Communications (HART)
Field Communicator Configure > Manual Setup > HART > Burst Mode
Overview
Burst messages contain information on process variables or transmitter status. You can configure up to three burst messages. Each message can contain different information. Burst messages also provide the mechanism for trigger mode and event notification.
Restriction
If you are using a HART 5 host, only one burst message is supported.
Procedure
1. Navigate to the burst message you want to configure.
2. Enable the burst message.
3. Set Burst Option to the desired content.
Options for burst message contentsTable 6-9:
HART command
1 Source (Primary Variable) Primary Variable The transmitter sends the primary variable (PV) in
2 Primary Variable (Percent
Range/Current)
3 Process Variables/Current Process Vars/Current The transmitter sends the PV’s actual milliamp read-
9 Read Device Variables with
Status
33 Transmitter Variables Field Device Vars The transmitter sends four user-specified process
48 Read Additional Transmitter
Status
Label
DescriptionProLink III Field Communicator
the configured measurement units in each burst message (e.g., 14.0 g/sec, 13.5 g/sec, 12.0 g/sec).
Pct Range/Current The transmitter sends the PV’s actual mA level and
the PV’s percent of range in each burst message (e.g.,11.0 mA 25%).
ing and the PV, SV, TV, and QV values in measure­ment units in each burst message (e.g.,11.8 mA, 50 g/sec, 23 °C, 50 g/sec, 0.0023 g/cm3).
Device Variables with Status The transmitter sends up to eight user-specified
process variables in each burst message.
variables in each burst message.
Read Additional Device Sta­tus
The transmitter sends expanded device status infor­mation in each burst message.
4. Depending on your choice, select the four or eight user-specified variables for the burst message, or set the HART variables as desired.
92 Micro Motion® Fork Density Meters (FDM)
Integrate the meter with the control system
Important
If you change the HART Primary Variable (PV) or Secondary Variable (SV), the process variables assigned to the primary mA output and the secondary mA output (if applicable) are automatically changed to match. The PV cannot be changed on devices with fixed mA output assignments.
Configure HART trigger mode
Display Not available
ProLink III Device Tools > Configuration > Communications > Communications (HART)
Field Communicator Configure > Manual Setup > HART > Burst Mode > Burst Message x > Configure Update Rate
Overview
Trigger mode uses the burst message mechanism to indicate that a process variable has changed. When trigger mode is implemented, the bursting interval (HART update rate) changes if Primary Variable or Burst Variable 0 moves above or below the user-specified trigger level. You can set up a different trigger on each burst message.
Restriction
This feature is available only with a HART 7 host.
Prerequisites
Before you can configure trigger mode, the corresponding HART burst message must be enabled.
Procedure
1. Select the burst message for which you will set up trigger mode.
2. Set Trigger Mode to the type of trigger you want to use.
Option Description
Continuous The burst message is sent at Default Update Rate. The burst interval is not affected
by changes in process variables.
Falling When the specified process variable is above Trigger Level, the burst message
is sent at Default Update Rate.
When the specified process variable is below Trigger Level, the burst message
is sent at Update Rate.
Rising When the specified process variable is below Trigger Level, the burst message
is sent at Default Update Rate.
When the specified process variable is above Trigger Level, the burst message
is sent at Update Rate.
Configuration and Use Manual 93
Integrate the meter with the control system
Option Description
Windowed This option is used to communicate that the process variable is changing rapid-
On Change If any value in the burst message changes, the burst message is sent at Up-
3. Ensure that Primary Variable or Burst Variable 0 is set to the variable that will activate the trigger. If it is not, reconfigure the burst message contents.
4. Set Trigger Level to the value of the process variable at which the trigger will be activated.
5. Set Default Update Rate to the burst interval to be used when the trigger is not active.
6. Set Update Rate to the burst interval to be used when the trigger is active.
ly. Trigger Level defines a deadband around the most recently broadcast value.
If the process variable stays within this deadband, the burst message is sent
at Default Update Rate.
If the process variable moves outside this deadband in either direction, the
burst message is sent at Update Rate.
date Rate.
If no values change, the burst message is sent at Default Update Rate.
Configure HART event notification
Display Not available
ProLink III Device Tools > Configuration > Communications > Communications (HART) > Event Notification
Field Communicator Configure > Manual Setup > HART > Event Notification
Overview
Event notification uses the burst message mechanism to indicate that an alert has occurred. When event notification is enabled and one or more of the selected alerts occurs, each active burst message will broadcast HART Command 119 until the condition is acknowledged by a HART master.
Tip
Event notification affects only HART burst messages. Whether an alert is selected for event notification or not, alert severity, alert status (active or inactive), fault timeout, and alert acknowledgment operate as normal.
Restriction
This feature is available only with a HART 7 host.
Prerequisites
If you are using the Field Communicator, you must enable a burst message before you can configure event notification.
94 Micro Motion® Fork Density Meters (FDM)
Loading...