Micro Motion, Inc. pursues a policy of continuous development and product improvement. The specification in this document
may therefore be changed without notice. To the best of our knowledge, the information contained in this document is
accurate and Micro Motion, Inc. cannot be held responsible for any errors, omissions or other misinformation contained
herein. No part of this document may be photocopied or reproduced without prior written consent of Micro Motion, Inc.
Cont-2
Exd 7835 Technical Manual Contents
IMPORTANT NOTICE
DO NOT drop the transducer. HANDLE WITH CARE
DO NOT use liquids incompatible with MATERIALS OF CONSTRUCTION
DO NOT allow axial loading from PIPEWORK STRESSES to exceed ½ TONNE
DO NOT operate the transducer above its RATED PRESSURE
DO NOTPRESSURE TEST above the specified TEST PRESSURE
DO NOT expose the transducer to excessive vibration (>0.5g continuous)
DO NOT rotate the electrical housing because this may invalidate the IP rating.
ENSURE all ELECTRICAL SAFETY requirements are applied
ENSURE transducer and associated pipework are PRESSURE TESTED to 1½
times the maximum operating pressure after installation
ENSUREtransducer is not TRANSPORTED when it contains hazardous fluids.
This includes fluids that may have leaked into, and are still contained,
within the case. The Returns Forms in Appendix F MUST be copied, completed and returned to the factory with the returned instrument
5.3 General Maintenance Procedure .........................................................................................5-1
Appendix A 7835 Specification ............................................................. A-1
Cont-5
Contents Exd 7835 Technical Manual
Appendix B Electronics Specifications ................................................ B-1
Appendix C Specimen Calibration Certificates ................................... C-1
Appendix D Conversion Tables and Product Data .............................. D-1
Appendix E Returns Forms .................................................................... E-1
Cont-6
Exd 7835 Technical Manual Introduction
Chapter 1
Introduction
1.1 Product Overview
This meter will provide a continuous on-line measurement of density and temperature of the process fluid being measured.
The construction of the meter is to explosion proof standards, allowing installation in hazardous areas. Operational
parameters can be found in the specification in appendix A.
Figure 1.1: Sideways view of meter
Liquid density is determined from the resonant frequency of a vibrating tube containing the liquid, and liquid temperature is
determined from a 100: Platinum Resistance Thermometer (PRT). For information on the calculation of density and
temperature, please refer to chapter 4.
1.2 Product Range
The product range is summarised in Table 1.1 (below). The meters are identical mechanically, except for the material
used in the wetted parts, and the flanges/couplings. A fully welded design is utilised to ensure maximum reliability in the
most severe environments. In the unlikely event of a leak occurring in the centre tube assembly, the outer casing will
withstand a line pressure rating of up to 1450psi (100 Bar).
For further details of the product range, please refer to appendix A.
Table 1.1: Explosion proof meter range
Meter
7835
Tube material
(wetted parts)
NI-SPAN-C
®
Low temperature coefficient and long term stability, appropriate for fiscal
Features
applications.
Page 1-1
Introduction Exd 7835 Technical Manual
SUPPLY
PRT
1.3 Electronics Product Range
The meters described above may be operated with frequency output electronics only. The electronics are not directly
interchangeable due to the complex internal wiring. For details of the performance of the electronics, refer to Appendix B.
Table 1.2
Standard
Electronics
o Basic amplifier circuit providing a frequency signal (indicating liquid
density) and PRT resistance (indicating liquid temperature).
o Features 6 screw-terminals for power in, and outputs (See Figure 1.2).
o Interfaces with a Signal Converter or Flow Computer (see Section 3).
Figure 1.2: Standard Electronics
Page 1-2
Exd 7835 Technical Manual Mechanical Installation
Chapter 2
Mechanical Installation
2.1 General
This chapter describes the mechanical installation of the 7835 liquid density meter.
Remember! Always handle the meters with care
When handling...
x Don’t drop the meter or subject it to severe mechanical shock.
x Don’t expose the meter to excessive vibration.
When installing…
x Ensure axial loading from pipework does not exceed ½ tonne.
x Ensure all electrical safety requirements are met (see safety instruction booklet 78355061/SI).
x Ensure that the meter and associated pipework are pressure tested to 1½ times the maximum operating pressure.
x Ensure that the electronics housing is not rotated. Rotating the housing may invalidate its’ IP rating.
When operating...
x Don’t use liquids incompatible with the construction.
x Don’t operate the meter above its rated pressure.
When transporting...
x Ensure that the meter does not contain hazardous fluids, including those that may have leaked into the case.
Page 2-1
Mechanical Installation Exd 7835 Technical Manual
2.2 Installation Planning
When planning the installation of a meter, it is important to consider the following factors:
Safety
When installing in a process line, it is important that the construction material of the wetted parts (tube)
is matched to the non-corrosive performance of the liquid passing through the instrument. Failure to
observe this requirement can cause deterioration of the central tube (the bellows) and loss in
measurement accuracy, or even a failure if leaking occurs. For advice on which meter in the range is
appropriate, please contact the factory.
Note: The NI-SPAN-C
®
material of the central tube is not rated for ‘sour’ service as defined in NACE
specification MR071-2000. For advice in this application, please contact the factory.
Serviceability
Performance
Pipe stresses
and vibration
Gas bubbles
Meter
orientation
Installing the meter in a by-pass configuration allows it to be removed for servicing, or calibration,
without affecting the main pipeline. Possible by-pass configurations are shown in Figure 2.2c.
Whilst the electronics may be accessed via the electronics housing end-caps, ensure that the
electronics housing is not rotated. Rotating the housing may invalidate its’ IP rating. If the
housing is rotated, contact the factory (see back page).
Axial load should not exceed ½ tonne, so pipe-work should have a degree of flexibility. Excessive pipe
vibration should be avoided. See Figure 2.2b for preferable mounting positions.
The presence of gas bubbles can seriously affect the meter performance and so the following points
should be considered:
x The liquid must always be at a pressure substantially above its vapour pressure.
x All pipe-work couplings and joints must be airtight.
x No vortex should be present at the inlet to the meter.
x Cavitations, caused by pumping, should not generate bubbles from dissolved gases.
x If a pump is used it should ‘push’ rather than ‘pull’ the product through the meter.
For low flow rates, e.g. 750 litres/hour (2.7 gal/min.), the meter should preferably be mounted vertically
or at an incline, with the flow in an upwards direction.
If the liquid contains solid particles, the direction of flow should be upwards unless the particles are
large enough not to be carried with the flow, in which case the direction of flow should be reversed.
The meter should be mounted with the electric cable running downwards thereby minimising the
ingress of water should a cable gland become defective.
Flow rate
A fast flow rate, e.g. 3000 litres/hour (11 gal/min.), will help to achieve good temperature equilibrium
and have a self-cleaning action.
A low flow rate, e.g. 1000 litres/hour (3.7 gal/min.), is recommended if the product contains particles
which may cause erosion.
The meters exhibit a small flow dependent density reading. For flow rates up to 15000 litres per hour
Temperature
(55 gal/min) and assuming no consequent line pressure or product changes, the maximum density
offset will be less than 0.2kg/m
The inlet pipework should be thermally lagged to ensure good temperature stabilisation.
3
.
Stability
Page 2-2
Exd 7835 Technical Manual Mechanical Installation
2.3 Meter Mounting and Pipework
This section considers in more detail the mounting of the meters and the design of the associated pipework, including the
calculation of pressure drop in the meter.
Installation drawings for all the meter versions are reproduced in the Appendices, along with detailed drawings of the
flanges/couplings. The preferred methods of supporting the meter are shown in Figure 2.2a.
Figure 2.2a: Preferred Methods of Mounting Meter
For continuously high flow rates, the mounting position can be selected to simplify the associated pipework and help
minimise the pressure and temperature losses (see Figure 2.2b below.)
The meter should be considered as a straight pipe of 23.6mm (0.929”) internal diameter and 1.03m (40.551”) in length.
The following formula has been proven to apply to the meter by measurements at 12000 litres per hour (44 gal/min).
V
), and
)
X
(
V m/s)
X = 2cS X =10cS
h =
u
Where : h = Pressure drop (bars)
f = Friction coefficient
L = Pipe length (m) = 1.03
D = Internal pipe diameter (mm) = 23.6
V
= Mean fluid velocity (m/s)
U = Fluid density (g/cc)
g = 9.81 (m/s
For viscous or laminar flow (Reynolds Number
Frictional Coefficient (
For turbulent flow (
R greater than 2500):
e
Frictional Coefficient (
Where: Pipe
f ) =
f ) =
R =
e
R
16
R
e
064.0
23.0
e
X
2
)
R less than 2000):
e
uuDV1000
X = Kinematic viscosity (cS)
2
VLf200
Uuuuu
Dg
In addition to the pressure drop caused by the liquid flow through the instrument, it will be necessary to calculate the
pressure drop in any associated sample pipework before concluding the system design requirements.
Page 2-6
Exd 7835 Technical Manual Mechanical Installation
2.5 Post-Installation Checks
After installation, the meter should be pressure tested to 1.5 times the maximum working pressure of the system but NOT
to a value exceeding the meter test figure shown on the meter label.
Caution:
If the pressure test figure is exceeded, the meter may be irrevocably damaged.
Page 2-7
Mechanical Installation Exd 7835 Technical Manual
Page 2-8
Exd 7835 Technical Manual Electrical Installation
Chapter 3
Electrical Installation
3.1 General
This chapter describes the electrical installation of a Exd 7835 liquid density meter with Standard Electronics fitted.
3.1.1 Power supply
The power supply to the meter must have the following specification:
Voltage: Nominally 24V dc, but in the range 18 to 30V dc.
Current: >30mA.
If several meters are to be used within a local area, one power supply can be used to power them all; where the meters
are distributed over a wide area and cabling costs are high, it may be more cost effective to use several smaller, local
power supplies.
3.1.2 Ground connections
It is not necessary to earth the meter through a separate connection; this is usually achieved directly through the
metalwork of the installation.
3.1.3 Cabling
Cables should conform to BS2538. In the USA, use Belden 9402 (two-pair) or Beldon 85220 (single-pair). Other cables
that are suitable are those that meet BS5308 Multi-pair Instrumentation Types 1 and 2, Belden Types 9500, 9873, 9874,
9773, 9774 etc.
The typical maximum recommended cable length for the above cable types is 1000m (3200ft.), but care must be taken to
ensure that the power supply at the meter is at least 20V. Thus, for 24V power supply, the overall resistance for the power
supply connections (both wires in series) must be less than 100 ohms.
In order to complete the wiring, you will need the following parts:
x ¾” NPT to M20 gland adapter.
x ¾” NPT blanking plug.
x M20 x 1 cable gland (not supplied).
The gland adapter and blanking plug are supplied with each meter. These two parts are “EExd” rated. However, you will
need to get a suitably rated cable gland. Alternative parts may be required in order to meet local electrical installation
regulations.
In hazardous areas, all parts must be explosion-proof.
Page 3-1
Electrical Installation Exd 7835 Technical Manual
3.1.4 Installation in explosive areas
The meter is an explosion-proof and flameproof device. Therefore, the connections shown in the wiring diagrams later are
applicable. However, it is essential to observe the rules of compliance with current standards concerning flameproof equipment:
1. Electronics housing caps should be tightened securely and locked in position by their locking screws.
2. The electrical cable or conduit should have an appropriate explosion-proof cable gland fitted.
3. If any electrical conduit entry port is not used, it should be blanked off using the appropriate explosion-proof
blanking plug, with the plug entered to a depth of at least five threads.
4. The spigot must be locked in place.
In addition, the electrical installation must strictly adhere to the safety information given in safety instruction booklet
78355061/SI, which will have accompanied this manual.
Page 3-2
Exd 7835 Technical Manual Electrical Installation
3.2 Use with Flow Computers and Signal Converters
When operated in conjunction with a flow computer or signal converter, only the meter can be operated in the
hazardous area. The flow computer / signal converter must be sited in a non-hazardous (safe) area only.
Cable screens should only be earthed at one end of the cable to prevent “ground loops”. This is normally done at the end
of the cable which is in a control room panel. At the end of the cable which is connected to the meter, the screen is
normally cut back and insulated. If the cable passes through a junction box, the screen continuity is assured by screen
earth terminals on the junction box.
Cable braid is not the same as cable screen, but is another form of armouring and will be earthed at both ends through
conductive glands
3.2.1 System Connections
The density system connections are illustrated in Figure 3.1 (below).
Note: The PRT within the Exd 7835 is designed to only operate in 4-wire mode. If the PRT is connected in
2-wire mode, an additional resistance will be seen that is a result of the EMC protection circuitry. When operating in 4-wire
mode, this additional resistance is not seen and the operation is unaffected.
Standard Electronics
1
2
3
4
PRT
5
6
330:
SUPPLY(+24V)
SIGNAL +VE
SUPPLY(0V)
SIGNAL -VE
PRT PWR +VE
PRT SIG +VE
PRT SIG -VE
PRT PW R -VE
7950/7951/7955
Figure 3.1 – Electrical connection diagram
7835 with Standard Electronics to Flow Computers / Signal Converters.
SEE FLOW COMPUTER/
SIGNAL CONVERTER
HANDBOOK FOR
CONNECTION DETAILS
Page 3-3
Electrical Installation Exd 7835 Technical Manual
3.3 Use with Customer’s Own Equipment
3.3.1 System Connections
Power supply to Density Meter: 15.5V to 33V d.c., 25mA min.
Power supply to PRT: 5mA max.
The frequency at which the meter is operating can be detected by using a series resistor in the +VE power line. The value
of resistance to be used for a given supply voltage must not exceed the value obtained from the LOAD NOMOGRAM
(Figure 3.3). The electrical connections to be made are shown in Figure 3.2.
Note: The PRT within the Exd-densitometer is designed to only operate in 4-wire mode. If the PRT is connected in
2-wire mode, an additional resistance will be seen that is a result of the EMC protection circuitry. When operating in 4-wire
mode, this additional resistance is not seen and the operation is unaffected.
Standard Electronics
1
POS +
SIG
2
NEG -
See Note
1PF
R
1PF
3
4
PRT
5
6
Note: See “Load Resistance” (Figure 4.3) to determine R value.
Figure 3.2 – Electrical connection diagram
7835 with Standard Electronics to Customer’s Own Equipment
Note: It is recommended that the actual load resistor should
be 50 ohms less than that given by the Nomogram.
35
(33)
Maximum Supply Voltage ‘E’
Supply Voltage
SIGNAL +VE
POWER +VE
POWER ?VE
SIGNAL ?VE
PRT SUPPLY +VE
PRT SIGNAL
PRT SUPPLY ?VE
Page 3-4
Supply Voltage (volts d.c.)
30
25
20
15
e
n
i
L
e
c
n
a
t
is
s
e
R
d
a
o
L
for Given Supply Voltage
Maximum Load Resistance
0
100
200
Maximum Load Resistance (ohms)
300
400500600
700
Figure 3.3 – Load Resistance
Exd 7835 Technical Manual Electrical Installation
q
3.4 Post-Installation Checks
After installation, the following procedure will indicate to a high degree of confidence that the meter is operating correctly.
Measure the current consumption and the supply voltage at the meter amplifier. This should be within the limits:
15.5V to 30Vdc, 17mA ±1mA
With the meter empty, clean and dry, measure the periodic time of the output signal and check that it is as specified on the
meter calibration certificate (air check), to within the limits given in the table below.
Table 3.1
Meter type
7835 ±60ns
Air check limit at 20
C
Added temperature effect
±10ns / qC
Page 3-5
Electrical Installation Exd 7835 Technical Manual
Page 3-6
Exd 7835 Technical Manual Calibration and Performance
W
W
Chapter 4
Calibration and Performance
4.1 General
The Exd 7835 meters are calibrated at the factory, and are shipped with test and calibration certificates.
The calibration certificate specifies various calibration constants that allow the user to convert the output periodic time
signal from the meter into a density value. (See Appendix C for specimen calibration certificates.)
For units with Standard Electronics, the calibration constants will need to be programmed into a signal processing
instrument such as a signal converter. Density calculations are performed on the signal processing instrument.
4.2 Interpretation of Calibration Certificate
4.2.1 General Density Equation
The basic meter constants, K0, K1 and K2 are computed from the factory calibration on three fluids. Using these
constants and the general density equation, the density of the liquid within the meter can be calculated.
The general density equation is:
Where: D = Uncorrected density of liquid (kg/m
K0, K1, and K2 = Constants from the Calibration Certificate.
D = K0 + K1.
W = Periodic time (Ps) of vibration = 1/f where ‘f’’ is the frequency of vibration.
On the calibration certificate, you can see that the basic meter constants (K0, K1, and K2) are determined from a
calibration at a temperature of 20qC (68qF) and at a pressure of 1 bar (14.5psi):
x On a metric certificate: DENSITY CALIBRATION AT 20 DEG. C AND AT 1 BARA
x On imperial certificate: DENSITY CALIBRATION AT 68 DEG. F AND AT 14.5 PSIG
If the operating conditions of the meter differ from that of the calibration conditions, the density calculated using
equation (1) must be corrected.
4.2.2 Temperature Correction
If the meter operates at temperatures other than 20qC (68qF), a correction to the density calculated using equation (1)
must be made using the temperature coefficient constants from your calibration certificate.
The equation used to apply temperature correction is:
K18 and K19 = Constants from the Calibration Certificate
Note:
1. K18 and K19 are the temperature coefficient constants on the calibration certificate.
= Temperature corrected density (kg/m3)
t
D = Density calculated using equation (1).
t = Temperature (degrees C)
Page 4-1
Calibration and Performance Exd 7835 Technical Manual
4.2.3 Pressure Correction
The meter design has a unique facility to reduce the influence of the line pressure on the density measurement, but a
correction may be required for a residual pressure effect.
This residual pressure effect before a pressure correction is shown schematically for the Exd 7835 in Figure 4.1.
During the calibration of the meter, which is normally performed at a pressure of 1 bar (14.5psi), the pressure influence
is also measured. This data is also shown on the calibration certificate.
The equation used to apply pressure correction is:
This residual pressure effect after a pressure correction is shown schematically for the Exd 7835 in Figure 4.2.
Notes:
= Temperature and pressure corrected density (kg/m3).
p
= Temperature corrected density (kg/m3) calculated using equation (2).
D
t
P = Pressure in bar absolute.
K21 = K21A + K21B (P – 1)
1. K20A, K20B, K21A and K21B are the pressure coefficient constants on the calibration certificate.
2. The pressure correction is further enhanced on units that operate above 41 Bar by having sets of pressure
coefficient constants covering subsets of the full operating pressure range.
Note that only one set of pressure coefficient constants is selected from your calibration certificate according to your
operating pressure range. If your operating pressure range falls within the range of two sets of pressure coefficient
constants, contact the factory for a new calibration certificate.
3. If it is required to apply temperature and pressure corrections, the temperature correction is applied first.
Figure 4.1: Pressure effect on 7835 before pressure correction
50
)
3
the meter fall within these bands
Uncorrected pressure effects on
40
30
20
10
Density Offset (kg/m
0
Page 4-2
-10
020406080100120140
Pressure (Bar Absolute)
Exd 7835 Technical Manual Calibration and Performance
Figure 4.2: Residual pressure effect after pressure correction – 7835 (100Bar) units
This figure shows the typical residual error curves after pressure corrections for
7835 (100Bar) units using three sets of pressure coefficient constants. Each set
covers a sub-set of the 100Bar range.
Note that only one set of pressure coefficient constants is selected from your
calibration certificate according to your operating pressure range.
For specimen calibration certificates, see Appendix C.
1.00
0.80
0.60
0.40
0.20
0.00
0 102030405060 708090100110
-0.2 0
-0.4 0
Magnitude (kg/m3)
-0.6 0
-0.8 0
-1.0 0
The uncertainty specification for a 7835 is indicated by the upper and lower limit
lines. The uncertainty for the 7835 pressure coefficients is ±0.003 kg/m
addition to the instrument calibration uncertainty of +/-0.15kg/m
New K2 0 K 21
Uppe r limit
Low e r limit
Pressure (BarG)
3
3
. This is in
.
Page 4-3
Calibration and Performance Exd 7835 Technical Manual
4.2.4 Velocity of Sound Correction
The Velocity of Sound (VOS) in the process liquid may have an effect on the accuracy of the indicated density. The
calibration of the Exd 7835 sensors has been optimised to a density/VOS relationship as indicated in Figure 4.3.
If the VOS of the process fluid deviates substantially from the relationship in Figure 4.5 it may be desirable to apply a
correction. This may be achieved by the simple introduction of a calibration offset using the data in Figure 4.3. Adjustment of
the value K0 in the basic equation will achieve this.
Alternatively, the following equations may be used:
Velocity of Sound (m/s)
2600
2400
2200
2000
1800
1600
1400
1200
1000
800
600
400
200
Where:
0
200
D
D
D
V
V
V
V
V
Dp
1
VOS
VOS
P
= Calibration VOS (m/s)
C
= Liquid VOS (m/s)
A
may be obtained direct from Figure 6.3 or may be calculated as follows:
C
=
C
=
C
«
«
P
¬
= Velocity of sound and temperature corrected density (kg/m3)
= Temperature and pressure corrected density (kg/m3)
Figure 4.3: Optimised Velocity of Sound Relationship for 7835
Exd 7835 Technical Manual Calibration and Performance
4.3 Calibration
4.3.1 Factory Calibration
The Exd 7835 Liquid Density Meters are calibrated prior to leaving the factory against Transfer Standard instruments,
traceable to National Standards.
Three fluids are used in the calibration - ambient air whose density is derived from tables, a hydrocarbon oil of about
815kg/m
are connected in parallel between two Transfer Standard Instruments on a special flow rig at the factory. During a
calibration and as the liquid flows through the instruments, readings are only taken when the indicated densities on the
two Transfer Standard Instruments agree. In this way, a high integrity of calibration is achieved.
Measurements are also made under conditions of changing temperature and pressure to establish the magnitude of these
effects on the instrument. From all this data, a calibration certificate is generated for each instrument.
Samples of the instruments are further tested by a quality assurance team at the factory to verify the calibration.
3
density and a high-density fluid in the range 1400 to 1500kg/m3 density. Several of the instruments-under-test
4.3.2 Calibration of Transfer Standards
The Transfer Standard instruments used in the calibration are selected instruments that are calibrated and certified by the
UKAS Certified Calibration Laboratory.
Transfer Standard calibration uses a number of ‘density certified’ liquids. The densities of these certified liquids are
obtained using the Primary measurement system, whereby glass sinkers of defined volumes are weighed in samples of
the liquids.
Calibration is performed by pumping each certified liquid through the Transfer Standard in a closely controlled manner
and recording the output signal in each case. A calibration certificate is issued for each Transfer Standard.
Calibrations are repeated, typically every six months, producing a well-documented density standard.
4.3.3 Instrument Calibration Certificate
Each instrument is issued with its own calibration certificate containing four important pieces of data:
(
a) The instrument serial number.
b) The output signal/density relationship. This is based on three calibration points - air, medium density and high-density
(
fluids. The air and high density fluid points are offset to achieve the product velocity of sound/density profile described
earlier, however, the signal value at Air Density is also given for check purposes.
(
c) Temperature coefficient data, describing the correction which should be applied to achieve the best accuracy if the
instrument is operating at product temperatures other than 20°C (68°F).
(
d) Pressure coefficient data, describing the correction that should be applied to achieve the best accuracy if the
instrument is operating at elevated pressures.
A second page of the calibration certificate is retained by the factory and contains all the calibration measurements.
See Appendix C for
specimen calibration certificates.
4.3.4 Pressure Test
A hydrostatic pressure test is carried out to a pressure value specified on the instrument label and on the instrument
calibration certificate. This test loads the instrument structure to a pressure which exceeds the maximum permitted
operating pressure of the instrument.
Note: During manufacture, the welded structure is pressure tested to conform to the requirements of EN50018-1997. The
outer case is able to withstand 100bar of internal pressure in the event of tube/bellows failure.
Page 4-5
Calibration and Performance Exd 7835 Technical Manual
4.3.5 Calibration Check Methods
There are two methods employed in calibration checks:
a) Air checkpoint, which is simple and convenient and highlights long term drift, corrosion and deposition.
(
b) Liquid calibration verification comprising two choices:
(
xDrawing off a sample of the liquid being measured and obtaining its density, using a hydrometer (for stable
liquids) or pyknometer (for unstable liquids).
xUsing a second density meter.
Ambient Air Check
(a) Isolate, drain and if necessary, disconnect the meter from the pipeline.
b) Clean and dry the wetted parts of the meter and leave them open to the ambient air.
(
c) Apply power to the instrument and check that the time period of the output signal agrees with the 'Air Check' figure
(
shown in the calibration certificate, to within acceptable limits (e.g. 60ns).
Some variation between the two figures is to be expected due to changes in ambient air conditions. The density
indication if using the K0, K1 and K2 factors will be about -0.9kg/m
3
because the basic density equation has been
optimised for best performance over the normal operating density range.
This test will indicate whether there has been a calibration offsets due to corrosion, deposition or long term drift. When
this test is applied to the 7845 instruments, their temperature coefficient has a significant effect and must be
considered (typically 0.3ms/°C).
Reconnect the meter to the pipeline if serviceable or remove it for further servicing.
Liquid Density Check
1. Sample Method
If it is necessary to verify the calibration using liquid at operating conditions, then the following sample methods are
recommended:
a) FOR STABLE LIQUIDS:
(
Draw off a sample of the liquid into a suitable container, at the same time noting the indicated density, temperature
and pressure of the liquid.
Measure the density of the sample under defined laboratory conditions, using a hydrometer or other suitable
instrument.
Refer the density measurement under laboratory conditions to that under the line operating conditions of temperature
and pressure.
Compare the referred density figure with that indicated by the density meter.
Note: It is essential that a good understanding of the physical properties (temperature coefficient, etc.) of the liquid is
acquired when using this method.
b) FOR UNSTABLE LIQUIDS:
(
Couple a pressure pyknometer and its associated pipework to the pipeline so that a sample of the liquid flows through it.
When equilibrium conditions are reached, the meter density reading is noted as the pyknometer is isolated from the
sample flow.
Remove the pyknometer for weighing to establish the product density.
Compare the pyknometer registered density with that obtained from the meter.
Page 4-6
Exd 7835 Technical Manual Calibration and Performance
Sampling Techniques
Sampling should comply with the international sampling standards (ISO 3171, ASTM D 4177, API 8.2 and IP 6.2).
For further details of these procedures, reference should be made to:
Institute of Petroleum: Petroleum Measurement Manual
Part VII Section 1 - Method IP 160 (Hydrometer Method)
(BS2000-160, ISO3675, ASTM 1298)
Institute of Petroleum: Petroleum Measurement Manual
Part VII Section 2 - Continuous Density Measurement
American Petroleum Institute: Manual of Petroleum Measurement Standards
Chapter 14 - Natural Gas Fluids - Section 6:
Installing and proving density meters used to measure hydrocarbon liquid with
densities between 0.3 and 0.7gm/cc at 15.56°C (60°F) and saturation vapour
pressure, 1991.
2. Second Density Meter
(a) Connect the second density meter to the pipeline adjacent to meter being checked so that it receives the same
sample of fluid under the same conditions of temperature and pressure as the meter under test.
(
b) Connect the second meter to its readout equipment, switch on and allow both systems to reach equilibrium
conditions.
c) Compare the two readings, making any necessary corrections.
(
It is often the practice, especially in fiscal metering applications, to use two or more density meters in a continuous
measurement mode as a means of improving the integrity of the measurement system. Any unacceptable discrepancies
between the measurements can immediately raise the necessary alarm signals.
This method of automatic checking has proved to be a very successful technique and where there is a facility for two
instruments, the practice of exchanging one for a newly calibrated instrument is proving successful. This is sometimes
referred to as the "Substitution Method".
It is very important when using one instrument to verify the performance of a second and similar instrument, to ensure
there are no unaccounted for systematic errors which would are not highlighted.
Page 4-7
Calibration and Performance Exd 7835 Technical Manual
4.4 Performance
Meters are generally calibrated at the factory using specified fluids at 20qC and 1 bar absolute. When operating at other
conditions, it is necessary to increase the uncertainty of measurement by the magnitude of the offsets if no corrections are
applied or increase by a fraction of the offsets if corrections are applied.
As a general guide, Table 4.1 lists the sources and magnitudes of the offsets affecting the meters covered in this manual,
with an example given in Table 4.2.
Table 4.1: Source and Magnitude of Measurement Offsets
Error Source
A Primary Standard
B Transfer Standard
C Instrument Accuracy
7835
0.05 kg/m
0.1 kg/m
0.15 kg/m
3
3
3
(at calibration conditions)
D Temperature (uncorrected)
Temperature (corrected)
E Pressure (uncorr’d at 50bar)
Pressure (uncorr’d at 100bar)
Pressure (corrected)
F Velocity of Sound (uncorr’d)
Velocity of Sound (corrected)
G Long term stability
r 0.02 kg/m
r 0.005 kg/m
-1 to +2 kg/m
+7 to +15 kg/m
r 0.003 kg/m
See Section 4.2
20% of offset
0.15 kg/m
3
/deg C
3
/deg C
3
/bar
3
/year
3
3
For total operational accuracy, the square root of the sum of the squares of each error source (C to G) is recommended,
i.e.
Effective Total =
22222
GFEDC
For example, if we consider instruments operating at 50qC (122qF) and 50 bar, six months after calibration and with no
VOS offset, the total operational accuracy after corrections have been applied is derived as follows:
Table 4.2: Total Operational Accuracy for Example Quoted
Error Source
C
D
E
F
G
7835
0.15
0.15
0.15
-
0.07
Effective Total0.27
For better accuracy, it would be necessary to carry out an on-line calibration at the operating conditions.
Higher accuracy can be obtained, by request, for all instruments by the use of water calibration or by UKAS certified
laboratory calibration of selected fluids.
Note:
The tables above relate to the effect of uncertainties on the time period output of the meter, and do not take into account
any uncertainty in the measurement of the time period itself.
Page 4-8
Exd 7835 Technical Manual General Maintenance
Chapter 5
General Maintenance
5.1 General
The Exd 7835 liquid density meters have no moving parts, which reduces the maintenance requirement to simple visual
checks for leaks and physical damage.
Check calibrations should be carried out at specified intervals in order to highlight any malfunction or deterioration in meter
performance. If a fault or a drop in meter performance is discovered, further tests are required to identify the cause of the
fault. Remedial action is limited to cleaning the tube, making good any poor connections and replacing the maintaining
amplifier or, in extreme cases, the entire instrument.
Caution:
Extreme care is required in the handling of the meter during transit, its
installation into the pipeline and its removal from the pipeline.
5.2 Fault Analysis
Faults generally fall into two main categories: erratic readings or readings outside limits.
Erratic Readings: Normally caused by the presence of gas bubbles in the flowing liquid. Severe electrical
Readings Outside Limits: Normally caused by deposition and/or corrosion on the resonating tube.
Since an electrical fault could also cause either of the two faults, and since examination for deposition or corrosion
requires the removal off-line of the meter, it is recommended that the electrical system be checked first.
interference or severe pipeline vibrations can also cause this effect.
5.3 General Maintenance Procedure
This procedure is recommended for any periodic maintenance carried out on the system and forms the basis of any
faultfinding task.
5.3.1 Physical Checks
Physical checks are as follows:
a.Examine the meter and its mounting bracket, pipe couplings and electrical cables for signs of damage and corrosion.
b.Check the meter for signs of fluid leakage and the state of the rupture plate.
Notes:
1.Any physical damage to the meter case or mounting brackets may have adverse effects on the meter performance
and a full calibration would be advisable to verify its accuracy.
2.Any oil leakage can generally be remedied by servicing.
Page 5-1
General Maintenance Exd 7835 Technical Manual
5.3.2 Check Calibration
Checking the calibration is as follows:
a.Carry out a check calibration using methods detailed in Chapter 4.
b.Compare the results obtained with the current calibration certificate figures to identify any substantial deterioration in
the meter's performance or any malfunction.
Notes:
1. A substantial drop in meter performance is likely due to a build-up of deposition on the vibrating tube, which can be
removed by the application of a suitable solvent. See 'Remedial Servicing' section below.
2.Malfunctions may be the result of electrical/electronic faults in either the meter circuit or the readout equipment. The
readout equipment should be proved before attention is directed to the meter as detailed under 'remedial servicing'.
5.3.3 Remedial Servicing
The required servicing falls into two categories - electrical and mechanical.
1. Electrical Servicing
a) Carry out power supply and current consumption tests at the meter terminals. These should give:
17mA ±1mA at 15.5V to 30V
Remove the power supply to the meter. If current consumption is suspect, contact the factory.
2. Mechanical Servicing
Mechanical servicing comprises mainly of keeping the inner surface of the vibrating tube clear of deposition and corrosion.
Deposition may be removed by the use of a suitable solvent. Alternatively, the instrument can be removed from the pipeline
and cleaned mechanically. Care is required to prevent damage to the inner surface of the tube during the cleaning.
Caution: Great care is essential in handling the meter during transit,
installation into the pipeline and removal from the pipeline.
Ensure that the meter is not transported when it contains hazardous fluids. This includes fluids which may have leaked
into, and are still contained, within the case.
Page 5-2
Exd 7835 Technical Manual 7835 Specification
Appendix A
7835 Specification
This appendix describes the performance and the mechanical design of the various versions of 7835 liquid density
meters. The flange / coupling variations do not affect the meter performance.
A.1 Performance
Density Range: 0 to 3000 kg/m
Accuracy: 0.15 kg/m3 (Over range 300 - 1100 kg/m3)
0.1 kg/m
Repeatability: 0.02 kg/m
Stability: 0.15 kg/m
Temperature Range: -50qC (122qF) to +110qC (230qF)
Temperature Coefficient: Uncompensated at 850kg/m
Compensated 0.005 kg/m
Pressure Range: 0 to 100 bar (1450psi) or as defined by flanges
Pressure Coefficient: Uncompensated (see Chapter 4)
(The temperature and pressure coefficients for each instrument are as specified by the
instrument calibration certificate.)
Compensated 0.003 kg/m
Test Pressure: 1.5 x maximum operating pressure.
Temperature Sensor
x Technology: 100 ohm PRT (4 wire)
x Range: -200qC (-328qF) to 200qC (572qF)
x Accuracy: BS 1904 Class, DIN 43760 Class A.
3
3
(with calibration in water)
3
3
per year
3
0.02 kg/m
3
/qC typical
3
/qC
3
/bar
Page A-1
7835 Specification Exd 7835 Technical Manual
A.2 Mechanical
Material: Wetted parts - Ni-Span-C
Run-out: Flange 1MM
Weight: 35 KG
Dimensions: See Figure A.1.
The 7835 is primarily intended for use with hydrocarbon products but may also be used with other process liquids, if they
are compatible with the NI-SPAN-C
There are various versions of the 7835 meter; each allocated an alphabetic suffix to identify the type of flange/coupling
fitted. The installation drawing gives details of the meter’s dimensions (see Figure A.1). Figure A.2 shows the general
outline of a flange with the differing dimensions for each flange type tabulated.
The meter variations available are:
Meter Version Flange/Coupling Type
7835A ASA 900 RF
7835B ASA 600 RF
7835D ASA 600 RJF
7835E ASA 900 RJF
7835F ASA 600 RF Smooth Face
7835H DIN 2635 RF DN25/PN40
7835J DIN 2635/2512 GVD DN25/PN40
7835K DIN 2637 RF DN25/PN100
A.4 Safety Approval
See safety instruction booklet 78355061/SI for safety approval information.
Page A-2
Exd 7835 Technical Manual 7835 Specification
Figure A.1: Installation Drawing for 7835 with Standard Electronics
Page A-3
7835 Specification Exd 7835 Technical Manual
Page A-4
Figure A.2: Flanges used on the 7835 Liquid Density Meter
where D = Density ( Uncorrected )
Dt = Density ( Temp Corrected )
DP = Density ( Pressure Corrected )
T = Periodic Time ( uS )
t = Temperature ( DEG.C )
P = Pressure (BarA)
------------- | FINAL TEST & |
| INSPECTION |
| |
| |
------------- Ref No:- LD7835/V5.0/FVA DATE : 15MAY07
Figure C.1: Example of certificate with 1 set of pressure coefficients (Metric Units)
where D = Density ( Uncorrected )
Dt = Density ( Temp Corrected )
DP = Density ( Pressure Corrected )
T = Periodic Time ( uS )
t = Temperature ( DEG.F )
P = Pressure (PSIG)
------------- | FINAL TEST & |
| INSPECTION |
| |
| |
| |
------------- Ref No:- LD7835/V5.0/FVA DATE : 15MAY07
Figure C.2: Example of certificate with 1 set of pressure coefficients (US Units)
where D = Density ( Uncorrected )
Dt = Density ( Temp Corrected )
DP = Density ( Pressure Corrected )
T = Periodic Time ( uS )
t = Temperature ( DEG.C )
P = Pressure (BarA)
------------- | FINAL TEST & |
| INSPECTION |
| |
| |
| |
------------- Ref No:- LD7835/V5.0/FVA DATE : 17MAR07
Figure C.3: Example of certificate with 3 sets of pressure coefficients (Metric Units)
where D = Density ( Uncorrected )
Dt = Density ( Temp Corrected )
DP = Density ( Pressure Corrected )
T = Periodic Time ( uS )
t = Temperature ( DEG.F )
P = Pressure (PSIG)
------------- | FINAL TEST & |
| INSPECTION |
| |
| |
| |
------------- Ref No:- LD7835/V5.0/FVA DATE : 17MAR07
Figure C.4: Example of certificate with 3 sets of pressure coefficients (US Units)
where D = Density ( Uncorrected )
Dt = Density ( Temp Corrected )
DP = Density ( Pressure Corrected )
T = Periodic Time ( uS )
t = Temperature ( DEG.C )
P = Pressure (BarA)
------------- | FINAL TEST & |
| INSPECTION |
| |
| |
| |
------------- Ref No:- LD7835/V5.0/FVA DATE : 03MAY07
Figure C.5: Example of certificate with 4 sets of pressure coefficients (Metric Units)
where D = Density ( Uncorrected )
Dt = Density ( Temp Corrected )
DP = Density ( Pressure Corrected )
T = Periodic Time ( uS )
t = Temperature ( DEG.F )
P = Pressure (PSIG)
------------- | FINAL TEST & |
| INSPECTION |
| |
| |
| |
------------- Ref No:- LD7835/V5.0/FVA DATE : 03MAY07
Figure C.6: Example of certificate with 4 sets of pressure coefficients (US Units)
Page C-6
Exd 7835 Technical Manual Conversion Tables and Product Data
Appendix D
Conversion Tables and Product Data
D.1 Conversion Tables
To convert the left-hand column of units into the top row of units, multiply by the factor in the box.
Exd 7835 Technical Manual Conversion Tables and Product Data
Velocity of sound in liquids (continued)
Oil (castor) 19 1500 -4.1
Oil (olive) 22 1440 -2.8
Octane 20 1197 ----
Oxygen -186 950 -6.9
n-Pentane 20 1044 -4.2
n-Propyl acetate 26 1182 ----
Toluene 20 1320 -4.3
Turpentine 25 1225 ----
Water (distilled) 10 1447.2 ----
20 1482.3 ----
30 1509.1 ----
50 1542.5 ----
70 1554.8 ----
Water (sea) -4 1430.2 ----
00 1449.5 ----
05 1471.1 ----
15 1507.1 ----
25 1534.7 ----
o-Xylene 22 1352 ----
Page D-7
Conversion Tables and Product Data Exd 7835 Technical Manual
Page D-8
Exd 7835 Technical Manual Return Forms
Appendix E
Return Forms
The Returns Forms, contained in this Appendix, must be copied and completed whenever a meter is to be returned to
the factory for servicing, calibration or repair. This must be done before the product is shipped.
There are separate Returns Forms for New/Unused equipment, and for Used equipment. Please select accordingly.
Page E-1
Return Forms Exd 7835 Technical Manual
Page E-2
Micro Motion Return Policy
For Use in the U.S.A.
With New and Unused Micro Motion Equipment
Definitions
New and unused equipment
Only equipment that has not been removed from the original shipping package will be considered new
and unused. New and unused equipment includes sensors, transmitters, or peripheral devices which:
• Were shipped as requested by the customer but are not needed, or
• Were shipped incorrectly by Micro Motion.
Used equipment
All other equipment is considered used. This equipment must be completely decontaminated and cleaned
before being returned. Document all foreign substances that have come in contact with the equipment.
Before you begin
This document is for returning new and unused equipment to Micro Motion in the United States.
• For instructions on returning used equipment, our used equipment return policy is available as a
• For instructions on returning equipment to Emerson offices around the world, our international return
To obtain any of our return policies, procedures, and forms, contact the Micro Motion Customer Service
Department during business hours:
• In the U.S.A., phone 1-800-522-6277 or 1-303-527-5200 between 6:00 a.m. and 5:30 p.m. (Mountain
• In Europe, phone +31 (0) 318 495 555, or contact your local sales representative.
• In Asia, phone (65) 6777-8211, or contact your local sales representative.
The latest return policies, procedures, and forms are also available from the Micro Motion web site:
www.micromotion.com.
Restock fees
Restock fees might apply, depending on the reason for return:
• If you ordered the wrong equipment, a restock fee will be charged.
• If you no longer require the equipment (for example, if your project has been cancelled), a restock fee
• If we shipped the wrong equipment, a restock fee will not be charged.
separate document.
policies are available as separate documents.
Standard Time), Monday through Friday, except holidays.
A Return Material Authorization (RMA) number must be obtained prior to returning any equipment
to Micro Motion for any reason.
To obtain an RMA number, contact the Micro Motion Customer Service Department at 1-800-522-6277 or
1-303-527-5200 between 6:00 a.m. and 5:30 p.m. (Mountain Standard Time), Monday through Friday,
except holidays.
• No product returns will be accepted without an RMA number.
• Each returned sensor must be issued a separate RMA number. A sensor and its associated
transmitter may be shipped in the same package with a single RMA number.
• If no sensor is being returned, all transmitters and peripheral devices being returned may be shipped
together, in one package, with a single RMA number.
Step 2 Preparing equipment for return
Only equipment that has not been removed from the original shipping package will be considered new
and unused. New and unused equipment must be returned in its original packaging.
Before returning new and unused equipment:
a. Clearly mark the RMA number on the outside of the original shipping package(s).
b. Clearly mark on the outside of each package: “NEW AND UNUSED”.
c. Complete and sign the “New and Unused Statement” on page 4.
d. Include one copy of the statement inside the original shipping package, and attach one copy to the
outside of each package.
e. Close and reseal all packages.
Step 3 Shipping instructions
Required shipping documents
The customer must provide a Packing List and Bill of Lading for each shipment. The Bill of Lading
contains information necessary for the carrier to ship the freight, such as consignee of shipment, payment
terms, number of pieces in shipment, weight, etc. The Bill of Lading should also contain the following
address:
Ship-to Party
Micro Motion Inc.
C/O Veolia Environmental Services
9131 East 96 Avenue
Henderson, CO 80640
Attn: RMA #__________
Micro Motion Return Policy for Use in the U.S.A. with New and Unused Micro Motion Equipment2
Document submittal
Submit the following shipping documents inside the shipping container:
• One (1) copy of the Packing List.
Submit the following shipping documents to your Micro Motion customer service representative:
• One (1) copy of the Packing List.
• One (1) copy of the Bill of Lading.
Shipping charges
The customer is responsible for all shipping charges.
Veolia has been instructed to refuse any collect shipments.
Micro Motion Return Policy for Use in the U.S.A. with New and Unused Micro Motion Equipment3
Statement of New and Unused Equipment
1)Return Material Authorization (RMA) Number:
Equipment Identification
2)For each instrument being returned, list a description or model number and its serial number.
Description or Model NumberSerial number
3)Reason for return:
Shipping Requirements
4)Clearly mark RMA number and “NEW AND UNUSED” on each shipping package.
5)Include one copy of this document inside the original shipping package, and attach one copy to the outside of each package in a
visible location.
6)Ship all equipment to:Address correspondence to:
Attn: RMA# _____________
Micro Motion Inc.
C/O Veolia Environmental Services
Sensor Department
9131 East 96 Avenue
Henderson CO 80640 USA
Micro Motion, Inc.
7070 Winchester Circle
Boulder CO 80301 USA
Attn: Repairs
Definition and Restock Fees
Only equipment that has not been removed from the original shipping package will be considered new and unused. New and
unused equipment includes sensors, transmitters, or peripheral devices which:
• Were shipped as requested by the customer but are not needed, or
• Were shipped incorrectly by Micro Motion.
Restock fees might apply, depending on the reason for return:
• If the customer ordered the wrong equipment, a restock fee will be charged.
• If the customer no longer requires the equipment (for example, if a project was cancelled), a restock fee will be
charged.
• If Micro Motion shipped the wrong equipment, a restock fee will not be charged.
THIS EQUIPMENT IS BEING RETURNED AS “NEW AND UNUSED,” PER THE DEFINITION STATED ABOVE.
I UNDERSTAND A RESTOCK FEE MIGHT BE CHARGED.
By:
(Signature)(Print name)
Title:Date:
Company:
Phone:Fax:
Micro Motion Return Policy
For Use in the U.S.A.
With Used Micro Motion Equipment
Definitions
New and unused equipment
Only equipment that has not been removed from the original shipping package will be considered new
and unused. New and unused equipment includes sensors, transmitters, or peripheral devices which:
• Were shipped as requested by the customer but are not needed, or
• Were shipped incorrectly by Micro Motion.
Used equipment
All other equipment is considered used. This equipment must be completely decontaminated and cleaned
before being returned. Document all foreign substances that have come in contact with the equipment.
Before you begin
This document is for returning used equipment to Micro Motion in the United States.
• For instructions on returning new and unused equipment, our new and unused equipment return
policy is available as a separate document.
• For instructions on returning equipment to Emerson offices around the world, our international return
policies are available as separate documents.
To obtain any of our return policies, procedures, and forms, contact the Micro Motion Customer Service
Department during business hours:
• In the U.S.A., phone 1-800-522-6277 or 1-303-527-5200 between 6:00 a.m. and 5:30 p.m. (Mountain
Standard Time), Monday through Friday, except holidays.
• In Europe, phone +31 (0) 318 495 555, or contact your local sales representative.
• In Asia, phone (65) 6777-8211, or contact your local sales representative.
The latest return policies, procedures, and forms are also available from the Micro Motion web site:
www.micromotion.com.
These procedures must be followed for you to meet governmental requirements. They also help us
provide a safe working environment for our employees. Failure to follow these requirements will result in
your equipment being refused delivery.
A Return Material Authorization (RMA) number must be obtained prior to returning any equipment
to Micro Motion for any reason.
To obtain an RMA number, contact the Micro Motion Customer Service Department at 1-800-522-6277 or
1-303-527-5200 between 6:00 a.m. and 5:30 p.m. (Mountain Standard Time), Monday through Friday,
except holidays.
• No product returns will be accepted without an RMA number.
• Each returned sensor must be issued a separate RMA number. A sensor and its associated
transmitter may be shipped in the same package with a single RMA number.
• If no sensor is being returned, all transmitters and peripheral devices being returned may be shipped
together, in one package, with a single RMA number.
Step 2 Cleaning and decontamination
All equipment being returned must be thoroughly cleaned and decontaminated of all foreign substances,
including all substances used for cleaning the equipment, prior to shipment. This requirement applies to
the sensor tubes, sensor case exterior, sensor case interior, electronics, and any part that might have
been exposed to process fluids or cleaning substances.
Shipping equipment that has not been decontaminated may cause a violation of U.S. Department of
Transportation (DOT) regulations. For your reference, the requirements for packaging and labeling
hazardous substances are listed in DOT regulations 49 CFR 172,178, and 179.
If you suspect that the sensor case interior may be contaminated, the case must be completely drained
and flushed to remove contaminants.
Contents of sensor case may be under pressure. Contents of sensor case may be hazardous. Take appropriate
measures to avoid the hazards associated with gaining access to a contaminated case interior. Avoid
exposure to hazardous materials.
Decontamination/Cleaning Statement
A blank Decontamination/Cleaning Statement is provided on the final page of this document. You may
copy and use this form to return any Micro Motion sensor.
• A Decontamination/Cleaning Statement is required for each sensor being returned.
• Each form must be fully completed and include a signature. If the statement is not completed, the
customer may be charged for decontamination and cleaning.
If the equipment has been exposed to a known hazardous substance with any characteristic that can be
identified in the Code of Federal Regulations, 40 CFR 261.20 through 261.24, the chemical abstracts
number and hazardous waste number/hazard code must be stated in the space provided on the form.
Two (2) copies of each Decontamination/Cleaning Statement must be provided:
• One (1) copy must be attached to the outside of the package.
• One (1) copy must be provided inside the package.
Micro Motion Return Policy for Use in the U.S.A. with Used Micro Motion Equipment2
Step 3 Material Safety Data Sheets (MSDS)
Included with the returned equipment, you must provide a Material Safety Data Sheet (MSDS) for each
substance that has come in contact with the equipment being returned, including substances used for
decontamination and cleaning.
An MSDS is required by law to be available to people exposed to specific hazardous substances,
with one exception: if the equipment has been exposed only to food-grade substances or potable water, or
other substances for which an MSDS is not applicable, the Decontamination/Cleaning Statement form
alone is acceptable.
Two (2) copies of each MSDS must be provided:
• One (1) copy must be attached to the outside of the package.
• One (1) copy must be provided inside the package.
Step 4 Packaging
Shipping a sensor and transmitter or sensor only
To meet DOT requirements for identifying hazardous substances, ship only one sensor per package. A
sensor and its associated transmitter may be shipped in the same package.
Shipping a transmitter or peripheral device without a sensor
If no sensor is being returned, all transmitters and peripheral devices being returned may be shipped
together, in one package.
Equipment installed on a portable cart, in a protective cabinet or with special wiring and process
connections
Micro Motion is equipped to repair sensors, transmitters and peripheral devices manufactured by Micro
Motion only. Our repair department cannot work on equipment installed in a customer-supplied cabinet,
on a portable cart as part of a system, or with any wiring or piping attached. Any returned equipment
other than Micro Motion sensors, transmitters and peripheral devices will be considered the responsibility
of the customer.
Step 5 Shipping
Required shipping documents
The customer must provide a Packing List and Bill of Lading for each shipment. The Bill of Lading
contains information necessary for the carrier to ship the freight, such as consignee of shipment, payment
terms, number of pieces in shipment, weight, etc. The Bill of Lading should also contain the following
address:
Ship-to Party
Micro Motion Inc.
C/O Veolia Environmental Services
9131 East 96 Avenue
Henderson, CO 80640
Attn: RMA #__________
Micro Motion Return Policy for Use in the U.S.A. with Used Micro Motion Equipment3
Document submittal
Submit the following shipping documents inside the shipping container:
• One (1) copy of the Packing List.
Submit the following shipping documents to your Micro Motion customer service representative:
• One (1) copy of the Packing List.
• One (1) copy of the Bill of Lading.
The address is listed as follows:
Micro Motion, Inc.
Attn: (Your customer service representative)
7070 Winchester Circle
Boulder, CO 80301 USA
RMA# _________________
Shipping charges
The customer is responsible for all shipping charges.
Veolia has been instructed to refuse any collect shipments.
Micro Motion Return Policy for Use in the U.S.A. with Used Micro Motion Equipment4
Sensor Decontamination/Cleaning Statement
Refer to Micro Motion Return Policy for Use in the U.S.A. with Used Micro Motion Equipment
1)Return Material Authorization (RMA) Number:
2)Equipment to be returnedModel Number: Serial Number:
3)Reason for return
Process and Decontamination/Cleaning Fluids
4)List each substance to which the equipment was exposed. Attach additional documents if necessary.
Used for
CAS#
Common name
5)Please circle any hazards and/or process fluid types that apply:
InfectiousRadioactiveExplosivePyrophoricPoison Gas
CyanidesSulfidesCorrosiveOxidizerFlammablePoison
CarcinogenPeroxideReactive – AirReactive – WaterReactive-Other (list)
Other hazard category (list)
if available
hazardous waste
(20 CFR 261)
[] Yes[] No
[] Yes[] No
[] Yes[] No
[] Yes[] No
[] Yes[] No
[] Yes[] No
EPA waste code
if used for hazardous
waste
6)Describe decontamination/cleaning process. Include MSDS description for substances used in decontamination and cleaning
processes. Attach additional documents if necessary.
Shipping Requirements
Failure to comply with this procedure will result in the shipment being refused
7)Ship only one sensor per box. RMA number must be noted on the shipping package.
8)Include inside the package: one copy of this document and all required Material Safety Data Sheets (MSDS).
9)Attach to the outside of the package: one copy of this document, and all required Material Safety Data Sheets (MSDS).
10) Ship equipment to:Address correspondence to:
Micro Motion Inc. Attn: RMA# _____________
C/O Veolia Environmental Services
Sensor Department
9131 East 96 Avenue
Henderson CO 80640 USA
Micro Motion Inc.
7070 Winchester Circle
Boulder CO 80301 USA
Attn: Repairs
EQUIPMENT HAS BEEN CLEANED AND DECONTAMINATED OF ANY HAZARDOUS SUBSTANCES