Safety Precaution 2
Important safeguards 2
Looking after your Micro CO meter 2
Introduction 2
% C O H b
CO-
OFF
Before you begin 2
Micro CO system overview 3
CO sensor 5
Micro CO meter exploded view 6
Disassembling the Micro CO
meter for servicing 7
Replacing the internal battery 7
Replacing the fuel cell 7
Reassembling the Micro CO meter 8
Calibrating the Micro CO meter 9
Circuit description 11
Power supply 11
micro
Fuel cell amplifier circuit 12
Reset circuit 12
Display 12
Sounder 12
Indicator lights 13
RS232 interface 13
Specifications 14
Technical support 15
Parts List 16
Circuit Diagram 18
Micro CO Meter
Service Manual
073-12 Iss. 1.1 Sept 2005
Page 2
Micro CO Meter
Service Manual
Information in this document is subject to change without notice and does not represent a commitment
on the part of Micro Medical Limited. Only the parts supplied by Micro Medical Limited should be used
to complete the service operation described in this manual. If in any way you feel unsure about the
successful completion of the service operation you should contact Micro Medical Limited or its
appointed agent in your country or region and arrange the despatch of the product to a Micro Medical
Limited Service Centre.
Copyright 2005 by Micro Medical Limited All rights reserved
Drawing no. 073-12
Version 1.1
September 2005
All other products are trademarks or registered trademarks of their respective owners.
1
Page 3
Safety Precaution
The servicing of this device is intended to be carried out by a properly trained and
competent electronics engineer, or experienced in the maintenance and servicing
of medical devices. Read this manual thoroughly before proceeding with the
service. If in any doubt please contact the service centre at Micro Medical Limited
or their accredited agent in your country or region.
Important Safeguards
ο Read all of the instructions.
ο Keep the instructions in a safe place for later use.
ο Follow all warnings and instructions marked on the product.
ο When replacement parts are required, be sure to use replacement parts
specified by Micro Medical that have the same characteristics as the original
parts. Unauthorised substitutions may result in fire, electric or other hazards.
ο Do not place on an unstable table.
ο The product should be operated only from the type of power source indicated
on the label.
Looking after your Micro CO Meter
ο Avoid exposing the Micro CO Meter to solvents including alcohol and
disinfectants.
ο Avoid operating the Micro CO Meter in dusty conditions or near to heating
appliances or radiators.
ο Do not keep the Micro CO Meter in a damp place or expose it to extreme
temperatures.
Introduction
This service manual provides you with information to carry out the servicing of the
Micro CO Meter. It is a process, which is relatively straightforward but must be
carried out in a logical sequence. Our advice is to familiarise yourself with the
contents of this manual before attempting to carry out the procedure of replacing
the parts supplied in the sensor replacement kit for the Micro CO Meter.
Before You Begin
Before you begin the servicing operation, please read the section on Circuit
description very carefully:
2
Page 4
Micro CO Meter system overview.
The Micro Medical Micro CO meter consists of a hand held microcomputer
unit (1) incorporating a CO sensor and is supplied with a mouthpiece
adapter with integral one way valve (2) together with disposable cardboard
mouthpieces (5).
The microcomputer unit is powered by a single alkaline PP3 battery (4)
and is supplied with a reducing connector for calibration (3).
1
2
%COHb
CO-PPM
OFF
Micro CO
3
4
5
3
Page 5
The CO sensor is an electrochemical fuel cell, and works through the
reaction of carbon monoxide at one electrode and oxygen (from ambient
air) at the other. This reaction generates an electrical current proportional
to the concentration of CO exposed to the sensing surface of the fuel cell.
The current output signal from the sensor is conditioned using a current to
voltage converter and is applied to an analogue to digital (A/D) converter.
When the unit is first switched on the microprocessor records the baseline
reading on the A/D input and uses this value to auto zero the instrument.
The subject is requested to breathe in maximally, hold the breath for 10
seconds, and then to expire fully through the mouthpiece connected to the
microcomputer unit with the mouthpiece adapter. As an aid to timing the
breath holding period a countdown from 10 to 0 is displayed after the unit
is turned on.
The microprocessor then records the peak value obtained and displays
this on a 3½ digit LCD display.
The value can be displayed either as parts per million (ppm) concentration
in the expired air or as the equivalent percentage carboxyhaemoglobin
(%COHb) using the mathematical relationships described by Jarvis et al,
for concentrations below 90ppm and by Stewart et al for higher
concentrations.
Rapid estimation of carboxyhaemoglobin levels in fire fighters
JAMA 1976; 235:390-392
As a quick guide to the estimated smoking level, red, amber and green
indicator lights are provided. The following levels of CO activate these
lights:
CO(ppm) %COHb Cigarette consumption Indicator
0 - 5 0 - 0.8 Non smoker Green
6 - 10 1 - 1.6 Light smoker Amber
11 - 72 1.8 - 12 Heavy smoker Red
>72 >12 Suspected poisoning Red + alarm
4
Page 6
CO Sensor
The sensor is an electrochemical micro fuel cell using gaseous diffusion
barrier technology resulting in a direct response to volume concentration
rather than partial pressure.
The cell consists of a working electrode and a counter electrode separated
by a thin layer of electrolyte. The gaseous diffusion barrier limits the flow
of gas to the sensing electrode and ensures the electrochemical activity of
the electrode is far in excess of the amount of gas with which it has to
deal.
Gas diffusing onto the working electrode reacts at the surface of the
electrode by oxidation.
CO reacts at the working electrode according to the equation:
CO + H2O → CO2 + 2H+ + 2e
The counter electrode acts to balance out the reaction at the sensing
electrode by reducing oxygen in air to water:
½O2 + 2H+ + 2e- → 2 H2O
-
5
Page 7
Micro CO meter exploded view
Item 6
Item 5
Item 4Item 3Item 1Item 2
Item 10
Item 9
Item 7 Item 8
6
Page 8
Disassembling the Micro CO meter for servicing.
The Micro CO microcontroller unit comprises of a solid state electronic circuit
in a robust ABS housing and does not require any preventative maintenance.
Routine maintenance consists of replacing the fuel cell and internal lithium
battery when they are exhausted. The 3.6 volt lithium battery is continuously
monitored by the microcontroller and the message bt2 will be displayed when
the voltage falls below 3 volts. The fuel cell exhibits a gradual loss of
sensitivity with time and has effectively expired when the unit can no longer
be calibrated.
When either the fuel cell or the lithium battery has expired, replace by
following the procedure below.
1. Turn the unit face down and slide back the battery compartment.
2. Remove the PP3 battery.
3. Remove the 2 self tapping screws (Item 1) and put to one side.
4. Lift the top moulding (Item 6) from the bottom moulding (Item 2).
5. Remove the PCB (Item 3) from the bottom moulding.
6. If the fuel cell (Item 4) is exhausted carefully remove by gently pulling
away from the PCB.
7. If the battery (Item 8) has expired, then note the orientation and remove
from the PCB-mounted clips by pulling on the battery strap (Item 7) directly
away from the PCB. Put the battery strap to one side and discard the
battery.
Replacing internal battery of the Micro CO.
1 Place the battery strap over the new battery (Cat No: BAT5900) and push
into the PCB-mounted clips ensuring the correct orientation. (Take great
care not to short circuit the terminations even momentarily as the
low internal impedance of lithium batteries will result in a high
current consumption and greatly reduced life).
Replacing fuel cell of the Micro CO.
1 Remove the CO sensor (Cat No: CEL7300) from the plastic container.
2 Carefully insert the CO cell into the PCB with the correct orientation.
7
Page 9
Reassembling the Micro CO meter
1. Place the PCB into the bottom moulding.
2. Tuck the battery strap between the battery and the side wall of the bottom
moulding, away from the slide switch.
3. Ensure that the slide switch (Item 9) and the switch plate on the top
moulding (item10) are both positioned at the bottom of their travel.
4. Ensure that the rubber gasket (Item 5) is in place in the top moulding.
5. Place the top moulding on top of the bottom moulding ensuring the rubber
gasket fits over the cell correctly and secure using the two self-tapping
screws.
6. Reconnect the PP3 battery ensuring correct polarity.
8
Page 10
n
Calibrating the CO meter
Calibration will remain stable to within 2% over one month and typically to
within 10% over 6 months. Micro Medical supplies calibration gas (20ppm CO
in air) and recommends that the unit is recalibrated on a 6 monthly basis. See
page 13 for calibration accessories.
To carry out the calibration locate the calibration switch on the right hand side
of the instrument as shown below.
Serial
port
Calibratio
Switch
Push the slide switch to the CO - PPM position and wait for the unit to display
zero.
9
Page 11
Screw the control valve firmly onto the cylinder and connect the gas supply as
shown below:
Plastic Tubing
Control
valve
20 ppm carbon
monoxide in air
Flow
indicator
Reducing
Connector for
calibration
The plastic tubing supplied with the gas should be pushed firmly over the
reducing connector.
Slowly turn the control knob anti-clockwise until the ball in the flow indicator is
between the two marks. This will then supply a gas flow of approximately
0.25 l/min. Apply this flow for 25 seconds and if the meter does not read
20ppm use the calibration tool to push the calibration switch.
The unit will then beep 3 times, store the new calibration value, and display
the following:
The gas supply should then be turned off.
10
Page 12
Circuit description
(Refer to the parts list and circuit diagram)
The circuit is based on the Hitachi one time programmable (OTP)
microcontroller HD64F3687FP (U5) operating at a clock frequency of
14.745MHz.
This processor contains EPROM, RAM, and eight 10-bit analogue to digital
(A/D) converters. The current output signal from the sensor is conditioned
using a current to voltage converter and is applied to an A/D input of the
microcontroller. Voltages derived from the external PP3 battery and the
internal lithium battery are also connected to the A/D converter inputs.
When the unit is first switched on the microprocessor records the baseline
reading from the A/D and uses this value to auto-zero the instrument. The
signal from the CO sensor is continuously monitored and the peak of the
calculated carbon monoxide concentration is displayed and transmitted
through the RS232 driver (U7).
Power Supply
The unit has two separate supplies. The externally accessible alkaline 9 volt
PP3 battery (BAT 1) provides the main supply. The instrument may be
switched on and off with the slide switch and may also be switched off by a
signal from the processor. This is done if the unit is left on, without use, for a
period of 4 minutes in order to conserve battery power. The supply is
controlled by gates U2 and U3 arranged in a bi-stable configuration, and
powered continuously from BAT1. When the slide switch is moved to the
‘CO-PPM’ position one end of R7 is pulled low. This transition is differentiated
by the action of C15 and R6 so that a momentary pulse appears on pin 2 of
U3. This pulse will toggle the bi-stable circuit so that pin 4 of U2 will go low,
turning transistor TR2 on, and supplying 9 volts to the low drop-out regulator,
U4. When the slide switch is returned to the ‘off’ position pin 1 of U2 is pulled
low, the bi-stable action is reversed, and TR2 will be turned off. If the unit is
left on without use for approximately 4 minutes then pin 44 of U5 is driven
high, under software control, turning on TR1 which will also turn the unit off
via the bi-stable circuit. When this happens the slide switch must be pushed
to the ‘off’ position and then to the ‘CO-PPM’ position in order to initiate
another pulse through C15 to turn the unit back on again. The output of the 5
volt regulator supplies the processor and associated circuitry. C13 and C14
smooth the input and output of U4.
BAT1 is monitored by an A/D input of the microcontroller, pin 62 through the
potential divider, R13 and R14. When this battery falls below 6.6 volts a
battery low warning is temporarily indicated on the display upon switch on.
When the battery falls below 6.1 volts the message is displayed permanently
and the unit cannot be used.
11
Page 13
The secondary power supply is provided by a single 3.6 volt lithium cell,
BAT2, used to permanently power the fuel cell amplifier circuit consisting of
the op-amp U1 and associated passive circuitry. The amplifier circuit draws
less than 11uA from the 1000mA-Hr lithium battery giving a nominal
operational life of greater than 10 years.
BAT2 is buffered by U8B and monitored by an A/D input of the
microcontroller, pin 63. R27, between BAT2 and the inverting amplifier,
prevents the battery from discharging through the op-amp when the supply is
switched off. The battery has an end point of 3 volts and when this level is
reached the message bt2 will be displayed. When this happens follow the
procedure outlined in Servicing.
Fuel cell amplifier circuit.
The fuel cell provides an output current, from the counter electrode,
proportional to the concentration of target gas at the sensing surface. This
current is converted to a voltage by the action of U1 and the associated
passive components at a nominal transfer factor of 387mV/μA. This signal is
further amplified by U8A with a gain of 2.08 and applied the A/D input of the
microcontroller, pin 44, after being filtered by the action of R8 and C4. The
sensitivity of the CO fuel cell is 0.045 +/- 0.015μA giving a nominal amplified
output of 36mV/ppm CO.
Reset Circuit
This microcontroller has internal reset circuitry that requires C12 and D5 for
proper operation.
Display
The display is a custom 3½ digit low power LCD. The segments and
backplane are driven directly by microcontroller. The backplane is driven by a
square wave of nominally 60Hz. The individual segments are driven by a
similar square wave that is in phase with the backplane when the segment is
off and 180 degrees out of phase when the segment is on.
Sounder
The sounder is operated by a 1kHz square wave generated by pins 24 and 25
of U5.
12
Page 14
Indicator lights
The three indicator LED’s are driven from pins 51, 52 and 53 of the
microcontroller through transistors TR3, TR4 and TR5.
RS232 interface
The transmit and receive ports for RS232 transmission are on pins 46 and 45
respectively of the microcontroller. The transmission is converted from logic
to RS232 levels by the U7.
13
Page 15
Specifications
Type Electro - chemical fuel cell
Range 0 - 100ppm
Resolution 1ppm
Green indicator light 0 to 6ppm (0 to 1% COHb)
Amber indicator light 7 to 10ppm (1.1 to 1.6% COHb)
Red indicator light 11 to 72ppm (1.8 to 12% COHb)
Flashing red light + alarm >72ppm (>12% COHb)
Accuracy +/-5% of full scale of 1ppm whichever
is the greater
Sensitivity drift 0.5%/°C
Sensor Life >2 to 5 years
Response time <15 sec (to 90% of reading)
Hydrogen cross sensitivity <15%
Operating temperature 15 to 25°C
Operating pressure Atmospheric +/-10%
Pressure coefficient 0.02% signal per mBar
Relative humidity 15 – 90% continuous
(Non condensing) (0 – 99% intermittent)
Baseline drift 0ppm (auto zero)
Long term drift <2% signal loss per month
Power source Single Alkaline 9 volt PP3
Main battery life >30 hours of continuous use
Internal battery life 10 years
Weight 180g (Including battery)
Dimensions 170 x 60 x 26mm
Display 3½ digit LCD
Storage temperature -20 to +70 deg Celsius
Storage humidity 30% to 90%
14
Page 16
Technical Support
Great Britain and World Headquarters
Micro Medical Ltd
PO Box 6
Rochester
Kent ME1 2AZ
Telephone + 44 (0)1634 893500
Fax +44 (0)1634 893600
Web Site http://www.micromedical.co.uk
Email support@micromedical.co.uk
Contact Micro Medical Ltd for the local agent in your region or
country for local service:
15
Page 17
Parts List
Designation Description
U1 OP90GS. Precision micro power surface mount
U2 BU4S11. Single NAND gate
U3 BU4S11. Single NAND gate
U4 LM2931M5.0. Low drop-out 5 volt regulator
U5 HD64F3687FP. Hitachi microcontroller
U6 24LC00. 16 Byte surface mount serial EEPROM
U7 MAX3221CAE. Maxim RS232 transceiver
U8 MCP602-I/SN. Microchip dual rail-to-rail OP-AMP
D1 ZHCS750 Schottky diode
D2 3mm green LED
D3 3mm red LED
D4 3mm yellow LED
D5 BAS21. Small signal diode