Microlife TWIN200 1 User Manual

Simultaneous dual-arm measurement and auscultatory function for accurate office blood pressure measurement.
Instruction Manual
EN
1
sphygmomanometer and is designed to provide convenient, accurate and reliable in-office
blood pressure measurement following recommended guidelines of the European Society
of Hypertension (ESH) 1, American Heart Association (AHA), and World Health Organization
(WHO).
1
O’Brien E, Asmar R, Beilin L, Imai Y, et al. European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure
measurement. European Society of Hypertension Working Group on Blood Pressure Deviceing. J Hypertens 2003;21:821-848.
Table of Contents
Product description
Name of parts and display ................................. 3–4
Components and accessories ............................. 5–6
Before using WatchBP Office
Selecting and fitting the correct cuff .................. 7–8
Three operation modes
«AUSCULTATION» Mode .......................................... 9
«1st VISIT» Mode ............................................ 10–11
«FOLLOW-UP» Mode .............................................12
Taking measurements using WatchBP Office
«AUSCULTATION» Mode .................................. 13–14
«1st VISIT» Mode ............................................ 15–17
«FOLLOW-UP» Mode ....................................... 18–20
Special Functions
Hiding measurement values ................................21
Mean arterial pressure and pulse pressure ......... 22
Taking fewer than three measurements ............ 23
Skipping the 60-second countdown .................... 23
Setting inflation pressure ..................................... 24
Setting measurement intervals ........................... 24
Viewing Measurements
«AUSCULTATION» Mode ....................................... 25
«1st VISIT» and «FOLLOW-UP» Mode ............. 25–26
Controlling the WatchBP Office Wirelessly
Setting up the Bluetooth receiver .................. 27–28
Installation of the software program ................... 29
Controlling the device wirelessly .................. 29–30
Appendix
Rechargeable battery and power adaptor .......... 31
Troubleshooting ................................................... 32
Error messages ............................................... 33–34
Safety, care, accuracy test and disposal ......... 35–36
Technical specifications ....................................... 37
EN
L
R
mmHg
mmHg
SYS mmHg
DIA
SYS
DIA
mmHg
PUL /min
AUSCULT 1st FOLLOW UP
L/R
3 sec.
INTERVAL
3 sec.
7 sec.
HIDE
7 sec.
DC 6V 600mADC 6V 600mA
Product description
Name of parts
Cuff Socket (L)
3
Cuff Socket (R)
I/O switch Bluetooth Switch Main Adaptor Socket Battery Compartment
Recharge Indicator
M Button (Memory)
Start/Stop Button
Display
Mode Switch
Display
Mean Arterial Pressure (MAP)
Pulse Pressure (PP)
Mode Indication
External Power
Bluetooth
Rest and Arm Position Reminder
Battery Symbol
Hide Display
Memory Mode
Pulse Rate
Heart Beat
L/R Cuff
Value for Left Arm
Systolic Value
Diastolic Value
Value for Right Arm
Systolic Value
Diastolic Value
4
EN
System Requirements: 550MHz CPU. 256MB Memory, 1024x768 pixel resolution, 256 color, CD-ROM drive, 1 free USB port, 40MB free hard disk space, Microsoft Windows 2000 / XP / Vista.
Install CD
©
2
0
0
8
M
i
c
r
o
l
i
f
e
C
o
r
p
o
r
a
t
i
o
n
.
A
l
l
R
i
g
h
t
s
R
e
s
e
r
v
e
d
.
WatchBP Analyzer Office
Version 1.0.2.4
WatchBP Office and components
mmHg
mmHg
SYS mmHg
DIA
SYS
DIA
mmHg
PUL /min
AUSCULT 1st FOLLOW UP
How to take blood presssure measurement in three different modes
Quick start guide
Flip over the quick start guide to see ‘Product description’.
1
Plug cuff connector in the cuff socket
Turn on the device
Switch the mode
«1st VISIT» Mode (Double arm)
«FOLLOW-UP» Mode (Single arm)
«AUSCULTATION» Mode (Single arm)
Press Start/Stop button to begin a measurement
Note blood presure using auscultatory method
Place the stethoscope on the hollow of the elbow, over the brachial artery
Press Start/Stop button to begin a measurement
An averaged result will be displayed after three consecutive measurments
2
3
EN
1
Simultaneous dual-arm measurement and auscultatory function for accura te in- offi ce blood pressure measurement.
Instruction Manual
AC Adaptor X1 (Input: 100-240V~50/60Hz 0.7A Output: +7.5V 2A)
M size (22cm~32cm) X2 L size (32cm~42cm) X2
5
WatchBP Office blood pressure device (1 pcs.)
WatchBP Office Software CD X1
Bluetooth Receiver Driver
Install CD
Bluetooth receiver X1 Install CD X1
Manual Instruction X1 Quick Start Guide X1
Selective Cuffs
Please contact Microlife or its distributor to purchase cuffs.
*
M (Medium size)
22 - 32 cm (8.75 - 12.5 inches)
With air tube 130 cm
L (Large size)
32-42 cm (12.5-16.5 inches)
With air tube 130 cm
6
EN
Before using WatchBP Office
R
Selecting the correct cuff
Two different cuff sizes are provided with the WatchBP Office: Medium and Large. Select the cuff size that best matches the circumference of the patient’s upper arm.
M (Medium size)
22 - 32 cm (8.75 - 12.5 inches)
M is the correct size for most people.
L (Large size)
32-42 cm (12.5-16.5 inches)
Connect the cuff to the device by inserting the cuff connector into the cuff connector socket.
7
Fitting the cuff properly
1) Place the cuff over the left (right) upper arm so that the air tube and artery mark arrow point toward the lower arm.
2) Lay the cuff on the arm. Make sure that the lower edge of the cuff lies approximately ¾” to 1”(2 to 3cm) above the elbow.
2–3cm
3) Wrap and tighten the cuff around the arm.
4) Leave a little free space between the arm and the cuff. The patient should be able to fit 2 fingers between the arm and the cuff. Clothing must not restrict the arm. Remove all clothing covering or constricting the measurement arm.
5) Cuffs that do not fit properly can result in incorrect blood pressure measurements. Use a different size cuff if the range Index at the end of the cuff does not fall into the range specified by the range stripes.
6) Apply the same steps to the other arm if a double arm measurement is required.
The range index of cuff should
*
be fallen into this range.
8
EN
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
Three operation modes
mmHg
mmHg
SYS mmHg
DIA
SYS
DIA
mmHg
PUL
/min
AUSCULT 1st FOLLOW UP
L/R
3 sec.
INTERVAL
3 sec.
7 sec.
HIDE
7 sec.
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
Choose a desired operation mode
The WatchBP Office device includes three convenient operation modes: «AUSCULTATION», «1st VISIT» and «FOLLOW-UP» Mode.Use the Mode Switch on the side of the device to select the desired mode.
9
«AUSCULTATION» Mode
Select «AUSCULTATION» Mode for blood pressure measurement of patients with arrhythmia, and, if needed, to confirm if a patient is suitable for the oscillometric method using a digital blood pressure monitor.
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
«1st VISIT» Mode
Select «1st VISIT» Mode to complete a fully- automated triple measurement on both arms according to recommended ESH/AHA blood pressure measurement protocols for a patient’s first office visit.
Simultaneous dual-arm measurements
In «1st VISIT» Mode, the WatchBP Office device measures the patient’s blood pressure on both arms simultaneously, helping to determine the preferred measurement arm and reveal other potential cardiovascular risks.
10
EN
Three operation modes (cont.)
Automated triple measurement
In «1st VISIT» Mode, the WatchBP Office device automatically takes three consecutive measurements at one minute intervals by default. The result of these three measurements are averaged to produce the first visit blood pressure measurement.
1
2
3
60 sec. 60 sec.
The user can manually set up the measurement
*
interval of 15, 30, 45 or 60 seconds in 1st Visit and Follow-up Mode. (Please refer to special functions section page.24 “Set up measurement interval”).
11
Determine proper measurement arm
If the measurement result between the two arm differs more than 20mmHg for Systolic or 10mmHg for Diastolic for all three measurements, the device will indicate the higher measurement arm by a flashing L or R.
The higher arm should be used as the preferred measurement arm for future measurements.
«FOLLOW-UP» Mode
Select «FOLLOW-UP» Mode to complete a fully- automated triple measurement on the preferred measurement arm. According to recommended ESH/AHA blood pressure measurement protocols for follow-up blood pressure measurement.
Automated triple measurement
In «FOLLOW-UP» Mode, the WatchBP Office device automatically takes three consecutive measurements at one minute intervals by default. The result of these three measurements are averaged to produce the follow-up visit blood pressure measurement.
1
2
3
60 sec. 60 sec.
The user can manually set up the measurement
*
interval of 15, 30, 45 or 60 seconds in 1st Visit and Follow-up Mode. (Please refer to special functions section page.24 “Set up measurement interval”).
12