Microlife BP A90 Navigation Manual

Europe / Middle-East / Africa
Microlife AG Espenstrasse 139 9443 Widnau / Switzerland Tel. +41 / 71 727 70 30 Fax +41 / 71 727 70 39 Email admin@microlife.ch www.microlife.com
Asia
Microlife Corporation. 9F, 431, RuiGang Road, NeiHu Taipei, 114, Taiwan, R.O.C. Tel. 886 2 8797-1288 Fax.886 2 8797-1283 Email service@microlife.com.tw www.microlife.com
North / Central / South America
Microlife USA, Inc.
1617 Gulf to Bay Blvd., 2nd Floor Ste A Clearwater, FL 33755 / USA
Tel. +1 727 442 5353 Fax +1 727 442 5377 Email msa@microlifeusa.com www.microlife.com
IB BP A90 Var 1 2713
Microlife BP A90
EN 1 FR 9 ES 17 PT 25
Microlife BP A90
Guarantee Card BP A90
Name of Purchaser / Nom de l'acheteur / Nombre del comprador / Nome do comprador
Serial Number / Numéro de série / Número de serie / Número de série
Date of Purchase / Date d'achat / Fecha de compra / Data da compra
Specialist Dealer / Revendeur / Vendedor especializado / Revendedor autorizado
Microlife BP A90
1
ON/OFF button
2
Display
3
Cuff Socket
4
Battery Compartment
5
Cuff
6
Cuff Connector
Display
7
Systolic Value
8
Diastolic Value
9
Pulse
AT
Pulse Rate
AK
Battery Display
AL
Stored Value
Read the instructions carefully before using this device.
Type BF applied part
1
EN
Dear Customer, Your new Microlife blood pressure monitor is a reliable medical device for taking measurements on the upper arm. It is simple to use, accurate and comes highly recommended for blood pressure moni­toring in your home. This device was developed in collaboration with physicians and clinical tests carried out prove its measurement accu­racy to be of a very high standard.* Please read through these instructions carefully so that you under­stand all functions and safety information. We want you to be happy with your Microlife product. If you have any questions, problems or want to order spare parts please contact Microlife-Customer Service. Your dealer or pharmacy will be able to give you the address of the Microlife dealer in your country. Alternatively, visit the internet at www.microlife.com where you will find a wealth of invaluable informa­tion on our products. Stay healthy – Microlife AG!
* This device uses the same measuring technology as the award winning «BP 3BTO-A» model tested according to the British Hyper­tension Society (BHS) protocol.
Table of Contents

1. Important Facts about Blood Pressure and Self-Measurement

• How do I evaluate my blood pressure?
2. Using the Device for the First Time
• Inserting the batteries
• Selecting the correct cuff
3. Taking a Blood Pressure Measurement using this Device
4. Data Memory
5. Battery Indicator and Battery change
• Low battery
• Flat battery – replacement
• Which batteries and which procedure?
• Using rechargeable batteries
6. Error Messages
7. Safety, Care, Accuracy Test and Disposal
• Safety and protection
• Device care
• Accuracy test
•Disposal
8. Guarantee
9. Technical Specifications Guarantee card (see Back Cover)
1. Important Facts about Blood Pressure and Self­Measurement
Blood pressure
generated by the pumping of the heart. Two values, the (upper) value and the
The device indicates the pulse rate (the number of times the heart
beats in a minute).
Permanently high blood pressure values can damage your
health and must be treated by your doctor!
Always discuss your values with your doctor and tell him/her if you
have noticed anything unusual or feel unsure. Never rely on
single blood pressure readings.
Enter your readings in the enclosed blood pressure diary. This
will give your doctor a quick overview.
There are many causes of excessively
values
ment where appropriate. Besides medication, relaxation techniques, weight loss and exercise can also lower your blood pressure.
Under no circumstances should you alter the dosages of any
drugs prescribed by your doctor!
Depending on physical exertion and condition, blood pressure is
subject to wide fluctuations as the day progresses. You should
therefore take your measurements in the same quiet condi-
is the pressure of the blood flowing in the arteries
diastolic
(lower) value, are always measured.
. Your doctor will explain them in more detail and offer treat-
high blood pressure
systolic
2BP A90
EN
tions and when you feel relaxed! Take at least two measure-
ments per day, one in the morning and one in the evening.
It is quite normal for two measurements taken in quick succession
to produce significantly different results.
Deviations between measurements taken by your doctor or in the
pharmacy and those taken at home are quite normal, as these situations are completely different.
Several measurements provide a much clearer picture than just
one single measurement.
Leave a small break of at least 15 seconds between two
measurements.
If you are pregnant, you should monitor your blood pressure very
closely as it can change drastically during this time!
If you suffer from an irregular heartbeat, measurements taken
with this device should only be evaluated after consultation with your doctor.
The pulse display is not suitable for checking the frequency
of heart pacemakers!
3

How do I evaluate my blood pressure?

Table for classifying blood pressure values in adults in accordance with the World Health Organisation (WHO) in 2003. Data in mmHg.
Range Systolic Diastolic Recommendation
blood pressure too low
1. blood pressure optimum
2. blood pressure normal
3. blood pressure slightly high
4. blood pressure too high
5. blood pressure far too high
6. blood pressure dangerously high
The higher value is the one that determines the evaluation. Example: a readout value between 150/85 or 120/98 mmHg indicates «blood pressure too high».
100 
100 - 120 60 - 80
120 - 130 80 - 85
130 - 140 85 - 90
140 - 160 90 - 100
160 - 180 100 - 110
180
60
110
Consult your doctor
Self-check
Self-check
Consult your doctor
Seek medical advice
Seek medical advice
Urgently seek medical advice!

2. Using the Device for the First Time

Inserting the batteries

After you have unpacked your device, first insert the batteries. The battery compartment (4 x 1.5 V, size AA), thereby observing the indicated polarity.

Selecting the correct cuff

Microlife offers 3 different cuff sizes: S, M and L. Select the cuff size to match the circumference of your upper arms (measured by close fitting in the centre of the upper arm). M is the correct size for most people.
Cuff size for circumference of upper arm
S 17 - 22 cm (6.75 - 8.75 inches) M 22 - 32 cm (8.75 - 12.5 inches) L 32 - 42 cm (12.5 - 16.5 inches)
Only use Microlife cuffs.
Contact Microlife Service if the enclosed cuff 5 does not fit.
Connect the cuff to the device by inserting the cuff connector 6 into the cuff socket 3 as far as it will go.
4
is at the back of the device. Insert the batteries

3. Taking a Blood Pressure Measurement using this Device

Checklist for taking a reliable measurement
1. Avoid activity, eating or smoking immediately before the measurement.
2. Sit down for at least 5 minutes before the measurement and relax.
3. Always measure on the same arm (normally left).
4. Remove close-fitting garments from the upper arm. To avoid constriction, shirt sleeves should not be rolled up - they do not interfere with the cuff if they are laid flat.
5. Always ensure that the correct cuff size is used (marking on the cuff).
Fit the cuff closely, but not too tight.Make sure that the cuff is positioned 2 cm above the elbow.The artery mark located on the cuff (ca. 3 cm long bar) must
lie over the artery which runs down the inner side of the arm.
Support your arm so it is relaxed.Ensure that the cuff is at the same height as your heart.
6. Press the ON/OFF button 1 to start the measurement.
7. The cuff will now pump up automatically. Relax, do not move and do not tense your arm muscles until the measurement result is displayed. Breathe normally and do not talk.
8. When the correct pressure is reached, the pumping stops and the pressure falls gradually. If the required pressure was not reached, the device will automatically pump some more air into the cuff.
9. During the measurement, the heart symbol AT flashes in the display and a beep sounds every time a heartbeat is detected.
10.The result, comprising the systolic 7 and the diastolic 8 blood pressure and the pulse 9 is displayed and longer beep is heard. Note also the explanations on further displays in this booklet.
11.When the device has finished measuring, remove the cuff.
EN
4BP A90
12.Switch off the device. (The monitor does switch off automatically after approx. 1 min.).
You can stop the measurement at any time by pressing the
ON/OFF button (e.g. if you feel uneasy or an unpleasant pres­sure sensation).

4. Data Memory

This device always stores the last result at the end of the measurement. To recall the reading, press and hold the ON/OFF button device is switched off. All display elements are shown now. Release the button when the stored result with the letter «

5. Battery Indicator and Battery change

Low battery

When the batteries are approximately ¾ empty the battery symbol AK will flash as soon as the device is switched on (partly filled battery displayed). Although the device will continue to measure reliably, you should obtain replacement batteries.

Flat battery – replacement

When the batteries are flat, the battery symbol AK will flash as soon as the device is switched on (flat battery displayed). You cannot take any further measurements and must replace the batteries.
1. Open the battery compartment 4 at the back of the device.
5
M» AL
1
while the
is displayed.
2. Replace the batteries – ensure correct polarity as shown by the symbols in the compartment.

Which batteries and which procedure?

Use 4 new, long-life 1.5 V, size AA batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.

Using rechargeable batteries

You can also operate this device using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the device as they may become damaged (total discharge as a result of low use of the device, even when switched off). Always remove the rechargeable batteries if you do not intend
to use the device for a week or more. Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the information regarding charging, care and durability.

6. Error Messages

If an error occurs during the measurement, the measurement is inter­rupted and an error message, e.g. «ERR 3», is displayed.
Error Description Potential cause and remedy
«
ERR 1
» Signal too
weak
ERR 2
» Error signal During the measurement, error signals were
«
ERR 3
» No pressure
«
«
«
in the cuff
ERR 5
» Abnormal
result
HI
» Pulse or cuff
pressure too high
The pulse signals on the cuff are too weak. Re-position the cuff and repeat the measurement.*
detected by the cuff, caused for instance by movement or muscle tension. Repeat the measurement, keeping your arm still.
An adequate pressure cannot be generated in the cuff. A leak may have occurred. Check that the cuff is correctly connected and is not too loose. Replace the batteries if necessary. Repeat the measurement.
The measuring signals are inaccurate and no result can therefore be displayed. Read through the checklist for performing reliable measurements and then repeat the measurement.*
The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over 200 beats per minute). Relax for 5 minutes and repeat the measurement.*
Error Description Potential cause and remedy
«LO» Pulse too
* Please consult your doctor, if this or any other problem occurs repeatedly.
low
If you think the results are unusual, please read through the information in «Section 1.» carefully.
The pulse is too low (less than 40 beats per minute). Repeat the measurement.*

7. Safety, Care, Accuracy Test and Disposal

Safety and protection
This device may only be used for the purposes described in these
instructions. The manufacturer cannot be held liable for damage caused by incorrect application.
This device comprises sensitive components and must be treated
with caution. Observe the storage and operating conditions described in the «Technical Specifications» section.
Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
EN
6BP A90
The cuffs are sensitive and must be handled with care.Only pump up the cuff once fitted.Do not use this device close to strong electromagnetic fields such
as mobile telephones or radio installations.
Do not use this device if you think it is damaged or notice anything
unusual.
Never open this device.If the device is not going to be used for a prolonged period the
batteries should be removed.
Read the additional safety information in the individual sections of
this instructions.
Ensure that children do not use this device unsupervised; some parts are small enough to be swallowed.

Device care

Clean the device only with a soft, dry cloth.

Accuracy test

We recommend this device is tested for accuracy every 2 years or after mechanical impact (e.g. being dropped). Please contact Microlife­Service to arrange the test (see foreword).

Disposal

Batteries and electronic devices must be disposed of in accor­dance with the locally applicable regulations, not with domestic waste.
7

8. Guarantee

This device is covered by a purchase. The guarantee is valid only on presentation of the guarantee card completed by the dealer (see back) confirming date of purchase or the receipt.
Batteries, cuff and wearing parts are not included.Opening or altering the device invalidates the guarantee.The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance with the operating instructions.
Please contact Microlife-Service (see foreword).
3 year guarantee
from the date of

9. Technical Specifications

Operating temperature:
Storage temperature: -20 - +55 °C / -4 - +131 °F
Weight: 718 g (including batteries) Dimensions: 146 x 47 x 64 mm Measuring procedure:
10 - 40 °C / 50 - 104 °F 15 - 95 % relative maximum humidity
15 - 95 % relative maximum humidity
oscillometric, corresponding to Korotkoff method: Phase I systolic, Phase V diastolic
Measurement range: 20 - 280 mmHg – blood pressure
Cuff pressure display range: 0 - 299 mmHg
Resolution: 1 mmHg Static accuracy: pressure within ± 3 mmHg Pulse accuracy: ± 5 % of the readout value Voltage source: Reference to
standards:
This device complies with the requirements of the Medical Device Directive 93/42/EEC. Technical alterations reserved.
40 - 200 beats per minute – pulse
4 x 1.5 V Batteries; size AA EN 1060-1 /-3 /-4; IEC 60601-1; IEC 60601-1-2 (EMC)
EN
8BP A90
Microlife BP A90
1
Bouton ON/OFF (marche/arrêt)
2
Ecran
3
Prise pour brassard
4
Logement des piles
5
Brassard
6
Connecteur brassard
Ecran
7
Tension systolique
8
Tension diastolique
9
Pouls
AT
Fréquence des battements de coeur
AK
Indicateur d'état de charge des piles
AL
Valeur enregistrée
Veuillez lire attentivement les instructions avant d'utiliser ce produit.
Partie appliquée du type BF
9
FR
Cher client, Votre nouveau tensiomètre Microlife est un appareil médical fiable conçu pour prendre la tension sur le haut du bras. Il est facile d'emploi, précis et vivement recommandé pour surveiller la tension chez soi. Cet appareil a été développé en collaboration avec des médecins. Les tests cliniques dont il a fait l'objet ont montré que les résultats affichés sont caractérisés par une très grande précision.* Veuillez lire ces instructions attentivement pour comprendre toutes les fonctions et informations sur la sécurité. Nous souhaitons que cet appareil Microlife vous apporte la plus grande satisfaction possible. Si vous avez des questions, des problèmes ou désirez commander des pièces détachées, veuillez contacter le Service Clients Microlife. Le revendeur ou la pharmarcie chez qui vous avez acheté cet appareil sont en mesure de vous fournir l'adresse du représentant Microlife dans votre pays. Vous pouvez aussi visiter notre site Internet à l'adresse www.microlife.fr, où vous trouverez de nombreuses et précieuses informations sur nos produits. Restez en bonne santé avec Microlife AG!
* Cet appareil applique la même technologie de mesure que le modèle primé «BP 3BTO-A» testé conformément aux standards de la Société Britannique de l'Hypertension (BHS).