Microlife BP3SK1 3B User Manual

Page 1
WatchBP
Office
Professional
Automated Of ce
Blood Pressure
Monitor
BP 3SK1-3B
EN
3
Page 2
FCC Requirement
changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy, and if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: – Reorient or relocate the receiving antenna. – Increase the separation between the equipment and receiver. – Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. – Consult the dealer or an experienced radio/TV technician for help.
Preface
Microlife WatchBP Office (BP3SK1-3B) is an Automated Office Blood Pressure (AOBP) Monitor.
The device is a non-invasive digital blood pressure device using oscillometric technique and an upper-arm blood pressure cuff to measure systolic and diastolic blood pressures, pulse rate, mean arterial pressure (MAP) for use in pediatric and adult populations with arm cuff circumference sizes ranging from 14 -52 cm.
The device can be connected to a computer (PC) running the WatchBP Analyzer software. The measured patient data can be transferred from the blood pressure monitor to the PC by means of a USB cable connection or Bluetooth connection.The Bluetooth 4.2 connectivity allows data transferring from the device to PC, smartphones or tablet.
The device detects the appearance of atrial fibrillation during measurement and gives a warning signal together with the measured blood pressure value if atrial fibrillation is detected (optional).
The device provides aortic blood pressure parameters, includes central systolic blood pressure (cSYS), central pulse pressure (cPP) and central diastolic pressure (cDIA), non-invasively through the use of a brachial cuff (optional).
WatchBP
https://www.microlife.com/professional-products
product support:
WatchBP Software support:
https://www.microlife.com/support/software-professional-products
Developers support:
https://www.microlife.com/developers1
4
Page 3
WatchBP Office
Table of Contents
Product description
Contents ............................................................................................................................... 6
Device Version ....................................................................................................................... 6
Upgrading the version of the device ......................................................................................... 6
Product Overview .................................................................................................................. 7
Display .................................................................................................................................. 7
Initial set up
Attaching the power plug to the power adapter ........................................................................ 8
Power ON/OFF ...................................................................................................................... 8
Set the date, time and the safeguard pressure ......................................................................... 8
Before using the device
Selecting the correct cuff ........................................................................................................ 9
Fitting the cuff properly ............................................................................................................ 9
Taking measurements in MANUAL and AUTO mode
Turn on the power ................................................................................................................. 10
Connect the cuff to the device ............................................................................................... 10
Select an operation mode ..................................................................................................... 10
Settings of AUTO mode ................................................................................................... 10-11
Taking measurement in AUTO mode ..................................................................................... 12
Viewing stored values ........................................................................................................... 12
MANUAL mode Settings ....................................................................................................... 12
Taking measurement in MANUAL mode ................................................................................ 13
Special Functions
Screening for atrial brillation during blood pressure measurement ......................................... 14
About Atrial Fibrillation .......................................................................................................... 14
Central blood pressure parameters ....................................................................................... 14
How is central blood pressure measured? ............................................................................. 14
Accuracy of the central blood pressure parameters ................................................................ 15
MAP (Mean Arterial Pressure) ................................................................................................ 15
Using WatchBP Analyzer
Installing the Software Program ............................................................................................. 16
Connecting the Device to a Computer ................................................................................... 16
Start the Software Program ................................................................................................... 16
Transferring and deleting measurement data ......................................................................... 16
Bluetooth connectivity .......................................................................................................... 17
Rechargeable battery and power adapter
Rechargeable Battery ........................................................................................................... 17
Using a power adapter .......................................................................................................... 18
Safety, care, accuracy test and disposal
Device care ..................................................................................................................... 18-19
Cleaning the cuff ................................................................................................................... 19
Accuracy test ....................................................................................................................... 19
Disposal .............................................................................................................................. 19
Error messages and Troubleshooting ............................................................................ 20-21
Technical specifications ....................................................................................................... 22
5
EN
Page 4
Product description
The WatchBP Office consists of two major parts
• The device, cuffs and accessories.
• The WatchBP Analyzer Software.
With the WatchBP Analyzer Software
1) The device can be programmed for the blood pressure measurement procedure.
2) Measured blood pressure values can be downloaded to the PC.
3) A PDF report and Microsoft Excel spreadsheet for data analysis can be generated.
Download the latest WatchBP Analyzer Software from the Microlife website.
*
https://www.microlife.com/support/software-professional-products
Contents
1. WatchBP Office AOBP Monitor (dependent on purchase version*)
2. WatchBP Office Cuff – Size M (22- 32cm)
3. WatchBP Office Cuff – Size L (32- 42cm)
4. Data Cable
5. Mains adapter
6. Instruction manual
7. Quick start guide
Quick start guide
Product description
Cuff Socket
Display
M Button (Memory)
To enter memory record and
Down function in setting
Up Button
To enter SETTING and scroll up/switch
Start/Stop Button
Start measurement(s) base on the
AUTO or AUS mode in standby
Mode Button
To switch the mode
WatchBP Office
Power Adapter Socket
USB port
DC 7.5V 1.5A
Battery Compartment
BP 3SK1-3B
MODE
Instruction Manual
1
EN
WatchBP
Office
Blood Pressure
Monitoring System
Professional
MODE
Device Version
* The device can be upgraded with special features. There are three different versions of the device:
Standard version: standard AOBP Monitor
AFIB version: standard AOBP Monitor with Microlife Atrial Fibrillation Detector
CBP version: standard AOBP Monitor with Microlife Atrial Fibrillation Detector and
Central Blood Pressure measurement
Upgrading the device
The Atrial Fibrillation Detector and Central Blood Pressure measurement of the device can be activated through the WatchBP Analyzer. An activation key is needed for activation, the activation key is specific for the device as it matches the ID. Please contact Microlife or the local distributor for additional information.
6
Page 5
Product Overview
WatchBP Office
Cuff Socket
Display
button (Memory)
To enter memory record and
Down function in setting
button (Up)
To enter SETTING and scroll up/switch
button (Start/Stop)
Start measurement(s) based on the
AUTO or MANUAL mode in standby
button (Mode)
To switch the mode
To conrm selection in setting
Display
External power
Auto mode
# Measurements Auto mode
CBP indicator
AFIB indicator
Bluetooth indicator
USB indicator
Interval time
Battery indicator
Rest indicator
Unit
Aus mode
Memory
HIDE
Power Adapter Socket
USB port
DC 7.5V 1.5A
MODE
Battery Compartment
Date / Time
SYS / cSYS
DIA / cDIA
MAP / cPP
Pulse
7
EN
Page 6
Initial set up
Attaching the power plug to the power adapter
Select a suitable plug attachment and attach to the power adapter as shown here.
Charge the battery completely
When using the device for the first time, charge the battery until the recharge indicator on the device turns green.
Power ON/OFF
Press button to switch on the device.
Press and hold button for 3 seconds to switch off the device and turn off the LCD screen. The device displays ‘oFF’ before turning off.
Set the date, time and the safeguard pressure Set the year - Press and hold the button for 3 seconds to enter setting mode.
The year number flashes in the display. Use the or button to select the year. Use the button to confirm your selection and move on to month setting.
Set the month – Use the or b utt on to se lec t th e m on th. U se th e b utt o n to confirm your selection and move on to day setting.
Set the day – Pre s s th e or but t on to se le c t th e da y. Us e t he b ut t on to confirm your selection and move on to time setting.
Set the time
the date and time are set, and the current time is displayed.
Set the safeguard pressure - Use the or button to select the highest inflation pressure or AUTO mode. Use the button to confirm and finish the settings. Once you have finished the setting mode the current time is displayed.
– Once you have set the hour and minutes and pressed the button,
The “highest inflation pressure” can be programmed to the device. The suggested Inflation Pressure is 30 to 40 mmHg
*
above the expected systolic value of the patient. You can select 160, 180, 200, 220 or, 240mmHg or use the default (device Displays “- - -’) then the device will automatically inflate the cuff to the optimal cuff pressure. If the selected Highest Inflation Pressure selected is too low to measure a patient’s blood pressure it may result in re-pumping or an error (“Err “) will be shown.
The date and time on the device automatically synchronizes with the date and time on the computer when connected
*
with the WatchBP Analyzer.
8
Page 7
WatchBP Office
Before using the device
Selecting the correct cuff
A variety of different cuff sizes are available. M and L size cuffs are provided with the device. Use the cuff marker to select the cuff size that best matches the circumference of the patient’s upper arm.
Cuff Size Circumference (cm) Circumference (inch)
S
M
L
L-XL
Each cuff is provided with 130 cm air tube.
*
Use only cuffs provided by Microlife!
*
Contact Microlife or its authorized distributor to purchase cuffs.
*
M and L size cuffs are included as standard accessories.
*
Fitting the cuff properly
1 Place the cuff over the upper arm so that the air
tube and artery mark arrow point towards the lower arm. The artery mark on the cuff must be placed over the brachial artery.
2 Lay the cuff on the arm. Make sure that the lower
edge of the cuff lies approximately 2 to 3 cm ( ¾ to 1 inch) above the elbow.
3 Wrap and tighten the cuff around the arm.
4 Leave free space with the size of 2 fingers between
the arm of the patient and the cuff. Excessive tightness may cause venous congestion and discoloration of the limb. If the cuff is wrapped too loosely, it cannot be inflated properly, and the measured values may be inaccurate. Remove all clothing covering or constricting the measurement arm. Clothing may interfere with measurement accuracy.
5 Cuffs that do not fit properly may lead to inaccurate
readings. Use a different size cuff if the range index at the end of the cuff does not fall into the range specified by the range stripes.
14-22 5.5-8.7
22-32 8.7-12.6
32-42 12.6-16.5
32-52 12.6-20.5
2–3cm
(¾ to 1 inch)
9
EN
Page 8
Taking measurements in MANUAL and AUTO Mode
Turn on the power
Turn on the device by pressing the button of the device.
Connect the cuff to the device
Connect the cuff to the device by inserting the cuff connector into the cuff connector socket.
Select an operation mode
There are two measurement modes that can be used. Press the button to switch between
AUTO or MANUAL Mode.
Settings of AUTO Mode The measurement program in AUTO Mode of the device can be set, includes Number of
Measurements, Resting Time (Countdown time), Interval Time, AFIB detector, CBP measurement, HIDE and Average calculation (Discard 1st measurement).
1 Set the Number of Measurements –
Press the button when the device is in AUTO mode to first enter setting of Number of Measurements. Use the button to scroll up and use button to scroll down among one to six measurements. Press button to confirm the number
of measurements and enter Resting Time
setting.
2 Set the Resting Time – Pre s s th e
button to scroll up and use button to scroll down among 15, 30, 60, 120, 180, 240, 300 seconds of Resting Time. P r e s s button to confirm and enter Interval
Time setting.
3 Set the Interval Time – between
measurements - Press the button to scroll up and use button to scroll down among 15, 30, 60, 120, 180, 240, 300 seconds of Interval Time. P r e ss button to confirm and enter AFIB detector setting.
10
Page 9
4 Set the Atrial fibrillation (AFIB) detector
– Pre s s th e or but t on to sw it c h the AFIB detector ON or OFF. Pr e ss button to confirm.
Set the AFIB detector option appears only for the device
*
version with AFIB detector. If the device has an activated AFIB detector then it is witched on in default.
5 Set the Central blood pressure (CBP)
measurement – P r e ss o r b ut t on
to switch ON or OFF CBP measurement. Press button to confirm. If the device is upgraded with CBP then it is switched on in default.
6 Set the Hide function – The device
features a Hide function in order to prevent influence on blood pressure in patients due to nervousness triggered by visible blood pressure measurements. Press or button to switch ON or OFF the Hide Function.
WatchBP Office
7 Average calculation – The device features
allow you to discard 1st measurement from averaging while the selected number of measurement is 3, 4, 5, or 6 measurements. Press or button to switch ON or OFF and press button to confirm the setting of Discarding 1
st
measurement
(d-1) feature and finish the setting of AUTO mode . Once you go through the settings, the device returns to standby.
The last settings programmed to the device are the default of AUTO mode until you set the program again.
*
11
EN
Page 10
Taking measurement in AUTO Mode
Select AUTO mode. Press the button to perform automatic measurements based on the settings of AUTO mode. The device shows all the settings and then starts counting down the Resting Time before the first measurement. The average measurement reading is displayed and saved after the measurements are complete.
The device switches the display for the average Systolic Blood Pressure (SYS), Diastolic Blood Pressure, Mean Arterial Pressure (MAP), central Systolic Blood Pressure (cSYS), central Diastolic Blood Pressure, and central Pulse Pressure (cPP) automatically if central
blood pressure measurement is enabled.
If CBP measurement is enabled, the cuff pressure is held at
*
around 60 mmHg for around 10 seconds to collect sufficient pulse waves.
Press button during countdown to skip the countdown.
*
Press button to cancel remaining measurements at anytime during the measurement sequence. Display the results
*
(average) if available.
Viewing stored values
The device stores blood pressure values of the last measurement procedure in AUTO mode.
Press the button to reveal the average of the measurements of AUTO mode. Continue pressing the button to review individual measurements.
The device switches the display for values of the individual measurement (including SYS, DIA, MAP, cSYS, cDIA and cPP
*
values) if central blood pressure measurement is enabled.
MANUAL Mode Settings
The measurement program in MANUAL mode can be set to preferences. The program includes setting the Highest Cuff Pressure and Hide Cuff Pressure during deflation.
Set Highest Inflation Pressure – pr e ss button when the device is in MANUAL mode to enter setting of Highest Cuff Pressure. U s e button or button to scroll among 160, 180, 200, 220, 240 mmHg and auto ‘- - -‘. Use the button to confirm and move to HIDE Pressure setting.
When auto ‘ - - - ‘ is selected, the device automatically inflates
*
the cuff to the correct cuff pressure.
The Highest Inflation Pressure is considered as a safeguard
*
pressure. The device automatically inflates the cuff to the optimal cuff pressure but not higher than the selected Highest Cuff Pressure.
12
Page 11
WatchBP Office
Hide Cuff Pressure during deflation
This option helps you to determine Korotkoff K1 and K5 sound readings without digit preference. To use the HIDE function in MANUAL mode, push button to select the setting of HIDE function and confirm the selection by button and finish the setting of MANUAL mode.
Taking measurement in MANUAL mode
Select the MANUAL mode if auscultatory blood pressure measurement is preferred above oscillometric blood pressure measurement. In MANUAL mode, the device serves as a pressure gauge. No oscillometric measurements will be taken. Systolic and diastolic Korotkoff sounds are determined by the physician using a stethoscope placed over the brachial artery.
Start inflation – Press the button to start cuff inflation. When the maximum inflation pressure is reached, the device will automatically begin a linear deflation at a rate of 3 mmHg/sec as is recommended by the guidelines.
Assess the Korotkoff K1 and K5 sound
– when pushing the button during cuff deflation the cuff pressure at the time of pushing is temporarily stored so that the systolic (K1) and diastolic (K5) pressures can be seen afterwards. The device can store up to 4 pressures.
Push the button at any time to start fast deation and nish the measurement of MANUAL mode and show the pressure you have marked. Alternatively, the device quickly deates and shows the pressures after having pushed the button 4 times, or the cuff pressure has reached 20mmHg during the deation cycle.
After the measurement in MANUAL mode, the device displays all marked cuff pressures for one minute.
Re-inflate
– Push and Hold the Up button during deflation to re-inflate for as long as the button is held up to a max of 299mmHg. Release the button to continue deflation. Exceeding 299 mmHg will result in an immediate release of cuff pressure and a ‘HI’ Error message.
Deflate faster
– Push and hold the button during deflation to release the pressure in
the cuff faster, at around 8 -12 mmHg/second.
The recommended deflation rate for auscultation is 2-3 mmHg per second. Do not assess K1 or K5 sound while
*
holding the button.
If HIDE in MANUAL mode is selected, the cuff pressure during deflation will not be displayed. The display shows "---"
*
during cuff deflation.
13
EN
Page 12
Special Functions
Screening for atrial fibrillation during blood pressure measurement
The device is designed to screen for atrial fibrillation during blood pressure measurements (optional) with high accuracy: a sensitivity of 98% and a specificity value of 92%*. If atrial fibrillation is detected this will be shown in the report.
Verberk et al. Screening for atrial fibrillation with automated blood pressure measurement: Research evidence and
practice recommendations. Int J Cardiol 2016: 465–473.
About Atrial Fibrillation
Atrial fibrillation is a common heart rhythm problem and a common cause of major strokes. It affects 8% of those 65 years and older and about 20% of all strokes are caused by atrial fibrillation. Atrial fibrillation is a rhythm problem that can last from a few minutes, to days or weeks and even years. Atrial fibrillation can lead to the formation of blood clots in the heart. These clots can break off and flow to the brain causing stroke. One sign of atrial fibrillation is palpitations. However, many people have no symptoms and therefore may remain undetected whereas diagnosing atrial fibrillation early followed by adequate treatment can largely reduce the chance of getting a stroke.
Central blood pressure parameters
The device is designed to assess central blood pressure parameters (optional).
Central blood pressure is the pressure in the ascending aorta, the largest artery that originates from the left ventricular of the heart and from where oxygen is distributed to all parts of the body through the systemic circulation. Central Systolic Blood Pressure and Central Pulse Pressure provided by this monitor are determined directly through pulse volume plethysmography (PVP) waveform analysis. Central Diastolic Blood Pressure by this monitor is calculated by subtraction of Central Systolic Blood Pressure and Central Pulse Pressure.
How is central blood pressure measured?
The device measures brachial systolic and diastolic blood pressure as usual. However, where the cuff normally totally deflates after the blood pressure measurement, the cuff now stops deflating at approximately 60 mmHg cuff pressure to keep a stable pressure on the brachial artery for approximately 10 seconds which is needed to acquire brachial pulse volume plethysmography (PVP) waveforms (pulse volume recording). During these 10 seconds approximately 10 PVP waveforms are recorded from which one average PVP waveform is determined and analyzed. From the average PVP waveform, some characteristic points (parameters) are identified that are directly related to arterial compliance (stiffness) and wave reflections. With these parameters and previously measured peripheral (regular) blood pressure the central systolic blood pressure value and the central pulse pressure value are then determined
14
1
.
Page 13
WatchBP Office
The time that is needed to determine the central blood pressure value may vary among patients; i.e. with faster heart rate, less time is required for collecting the number of required PVP waveforms. It is very important to keep the arm still during the time the PVP waveforms are collected.
Accuracy of the central blood pressure parameters
The accuracy of central blood pressure parameters performed with this device can only reliably be determined against intra-arterial blood pressure measurement. The device is a certified equivalence with the WatchBP Office Central that has been validated against simultaneous recorded intra-arterial blood pressure measurement in 85 subjects and showed high accuracy
Sung, S.H., et al., Measurement of central systolic blood pressure by pulse volume plethysmography with a
noninvasive blood pressure monitor. Am J Hypertens, 2012. 25: 542-8.
Cheng, H.M., et al., Measurement accuracy of a stand-alone oscillometric central blood pressure monitor: a
validation report for Microlife WatchBP Office Central. Am J Hypertens, 2013. 26: 42-50.
MAP (Mean Arterial Pressure)
The device measures the true mean arterial pressure (MAP) of the patient. Each measurement includes a single MAP value. The MAP value will always be displayed together with the systolic and diastolic blood pressure value.
2
.
15
EN
Page 14
Using WatchBP Analyzer
Installing the Software Program
The latest WatchBP Analyzer Software is available from the Microlife website.
https://www.microlife.com/support/software-professional-products
Double click the download installer and simply follow the instructions provided in the installation window on the computer screen.
Connecting the Device to a Computer
It is important to only use the USB cable provided.
Start the Software Program
Start the software program. The date and time on the device automatically synchronizes with the date and time on the computer when successfully connected with WatchBP Analyzer PC software.
If the device and WacthBP Analyzer software is connected successfully:
• <USb> is displayed on the LCD screen of the device.
• The device ID, model, version of the Device and batteries condition etc. are displayed on the WatchBP Analyzer software.
DC 7.5V 1.5A
Transferring measurement data
Connect the device to the PC. Start the WatchBP Analyzer software program.
Click <Download> button of the WatchBP Analyzer to transfer the measurement data on the device to a computer.
Deleting measurements
The measurement data on the device will be automatically deleted after clicking <Program Device> in the WatchBP Analyzer software to program a measurement schedule for the next patient.
Press and hold button of the device for 7 seconds displays CL, presses button again to clear the memory
*
See instruction manual of WatchBP Analyzer for details.
16
Page 15
WatchBP Office
Bluetooth connectivity
Pairing the device
Press and hold the button for around 7 seconds, untill the Bluetooth icon flashes and starts pairing mode. The unique 6-digit device ID of the unit is displayed. Connect the device and confirm pairing. The Bluetooth icon is displayed on the LCD screen of the device to show the presence of Bluetooth connection.
Press and hold the button for 5 seconds to clear the
connection.
Rechargeable battery and power adapter
Rechargeable Battery
The device has a built-in, rechargeable Ni­MH battery pack that can perform up to 400 measurement cycles on a full charge. The battery can be recharged with the power adapter provided with the device. The empty battery indicator is displayed when the battery is low.
• When using the device for the rst time, charge the battery until the recharge indicator turns to green.
• The orange recharge indicator indicates that the recharge is in progress.
• A green recharge indicator indicates that the recharge is completed.
• A green and orange changing recharge indicator, means that there is a charging error. Make sure that the correct Mains Adapter is used. If the condition persists, contact Microlife or the local distributor.
17
EN
Page 16
DC 7.5V 1.5A
Using a power adapter
Only use the Mains Aadapter supplied with the device to recharge the device.
1) Plug the adapter cable into the power socket of the device.
2) Plug the adapter plug into the wall socket.
DC 7.5V 1.5A
The battery will be recharged if the device is attached to an AC power source. After the battery is fully recharged, the charging will stop. No battery power will be used if the adapter is plugged in. The battery must always remain within the device even when using AC power.
3) If the battery starts losing capacity, contact your local dealer for replacement battery. The battery can be replaced.
Safety, care, accuracy test and disposal
Safety and protection
This device may only be used for the purposes as described in these instructions. The device comprises of sensitive components and must be treated with caution. The manufacturer cannot be held liable for damage caused by incorrect application.
Follow the Instructions for Use. This document provides important product operation and safety information regarding this Blood Pressure Monitor. Please read this document thoroughly before using the device and keep for future reference.
• Only activate the pump when the cuff is connected to the device.
• Do not use the device if you think it is damaged or if anything appears unusual.
• Read the further safety instructions in the individual sections of the instruction manual.
18
Page 17
WatchBP Office
M (Medium size) for upper arm
22 - 32 cm (8.7 - 12.6 inches)
With air tube 130 cm
Observe the storage and operating conditions as described in the “Technical specications” section of this manual.
Protect the device from water and moisture
Protect the device from direct sunlight
Protect the device from extreme heat and cold
Device care
Use a soft cloth with one of the following recommended cleaning solutions to wipe the exterior of the device:
• Mild soap and water.
• Hydrogen peroxide solution (3% diluted with water).
• Sodium hypochlorite solution (1:10 dilution of household chloride bleach in water).
Cleaning the cuff
Remove the bladder. Fold and place the cuff cover inside a washing bag. Wash the cuff cover with warm water (43°C; 110°F) and a mild detergent in the washing machine.
Pasteurization: wash the cuff cover in 75°C(167°F) hot water for 30 minutes.
Avoid proximity to electromagnetic fields, such as those produced by mobile phones
Never open the device
Protect the device from impact and drops
MODE
Accuracy test
We recommend the device to be tested for accuracy every 2 years or after mechanical impact (e.g. Being dropped). Please contact Microlife to arrange an accuracy test.
Disposal
Batteries and electronic instruments must be disposed of in accordance with the locally applicable regulations, and not as domestic waste.
Do not iron the cuff!
19
EN
Page 18
Error messages and Troubleshooting
If an error occurs during measurement, the measurement is interrupted and an error message
«Er» is displayed.
Error Description Potential cause and remedy
"Er 1"
"Er 2"
“Er 3”
“Er 5”
"Er 11"
"Er 12"
Signal too
weak
Error signal
No pressure in
the cuff
No valid results
Signal too
weak during
central blood
pressure
measurement
Error signal
during central
blood pressure
measurement
The pulse signals on the cuff are too weak. Reposition the cuff
and repeat the measurement.
During the measurement, error signals were detected by the
cuff, caused for instance by movement or muscle tension.
Repeat the measurement, keeping your arm still.
An adequate pressure cannot be generated in the cuff. A
leak may have occurred. Replace the blood pressure cuff if
necessary. Repeat the measurement.
The measuring signals are inaccurate, and no result can
therefore be displayed. Read through the checklist for
performing reliable measurements and then repeat the
measurement.
The pulse signals on the cuff are too weak. Re-position the cuff
and repeat the measurement.
During the measurement, error signals were detected by the
cuff, caused, for instance, by movement or muscle tension.
Repeat the measurement, when keeping the arm still.
“Er 13”
20
Cuff pressure
errors during
central blood
pressure
measurement
An adequate pressure cannot be generated in the cuff. A leak
may have occurred. Check if the cuff is correctly connected
and is not too loose. Replace the blood pressure cuff if
necessary. Repeat the measurement.
Page 19
WatchBP Office
"Er 15"
"Er F"
“Er A”
"HI"
"LO"
Abnormal
result of
central blood
pressure
reading
The device
has gone into
“single fault
condition”
Flash memory
error
Pulse or cuff
pressure too
high
Pulse too low
Troubleshooting
The measuring signals are inaccurate so that no result can be
displayed. Read through the checklist for performing reliable
measurements and then repeat the measurement.
Single fault condition means that the measurement is aborted
to protect the patient from being harmed or the device
from being damaged. Re-position the cuff and repeat the
measurement. Replace the batteries if necessary. If the error
persists, contact Microlife or the local distributor.
Possible hardware fault. Try again. If the error persists, contact
Microlife or the local distributor.
The pressure in the cuff is too high (over 299 mmHg) OR the
pulse is too high (over 239 beats per minute). Relax for 5
minutes and repeat the measurement.
The pulse is too low (less than 30 beats per minute). Repeat the
measurement.
Problem Possible cause Solutions
No power (No LCD display)
Cuff does not inflate properly
No result displayed after
Power supply is not properly plugged in
Battery is fully discharged
Loose connection of the tube
Leakage of the tube / bladder
Device is in MANUAL mode
Plug the power supply into the wall socket.
Recharge the rechargeable battery by plugging in
the power supply.
Make sure the tube of the cuff is securely connected
to the device.
Check for cracks on the tube or the bladder. Replace
the blood pressure cuff if necessary.
Switch to AUTO Mode and repeat the
measurements.
measurements
21
EN
Page 20
Technical specifications
Operation temperature/ humidity:
Storage temperature/ humidity:
Weight:
Dimensions:
Measuring method:
Measurement range:
Cuff pressure display:
Pulse accuracy:
Power source:
Expected service life:
• 10 to 40 °C (50 to 104 °F)/ 15 - 90 % relative maximum humidity
• -20 to 55 °C (-4 to 131 °F)/ 15 - 90 % relative maximum humidity
• 620g (including rechargeable battery pack)
• 220.4 x 121.7 x 63.3 mm
• Oscillometric, Systolic blood pressure = K1; Diastolic blood pressure = K5
• 60 - 255mmHg - systolic blood pressure; 30 - 200mmHg - diastolic blood pressure; 30 - 239 beats per minute - pulse
• Range: 0 - 299 mmHg; Resolution: 1 mmHg; Static accuracy: pressure within ± 3 mmHg;
• ±5 % of the readout value
• Rechargeable battery pack; 4.8V 2400 mAh; Mains power supply DC
7.5V, 1.5 A
• 2 years
Reference to Standards:
Electromagnetic Compatibility:
22
• Device corresponds to the requirements of the standard for non­invasive blood pressure monitor. IEC 60601-1: 2005+A1:2012 IEC 60601-1-2:2014 ANSI/AAMI/ISO 81060-2 ANSI/AAMI/IEC 80601-2-30
• Device fullls the stipulations of the standard IEC 60601-1-2.
The stipulations of the EU Directive 93/42/EEC for Medical Devices Class IIa have been fullled.
Type BF applied part
Microlife reserves the right to alter technical specications without prior written notice.
Page 21
Guarantee Card
WatchBP Office
This device is covered by a two-year guarantee from the date
of purchase. This guarantee is valid only on presentation of the
guarantee card completed by the owner conrming date of purchase
or purchase receipt. Batteries and wearing parts are not covered by
this guarantee.
Name:
Address:
Date:
Telephone:
Email:
Product: WatchBP Office
Product number: BP3SK1-3B
Date:
23
EN
Page 22
Europe / Middle-East / Africa
Microlife AG
Espenstrasse 139
9443 Widnau, Switzerland
Tel. +41 71 727 7000
Fax. +41 71 727 7011
Email: watchbp@microlife.ch
www.watchbp.com
Asia
Microlife Corporation
9F., No.431, Ruiguang Rd., Neihu Dist.,
Taipei City 114, Taiwan (R.O.C.)
Tel. +886 2 8797 1288
Fax +886 2 8797 1283
Email: watchbp@microlife.com.tw
www.watchbp.com
North / Central / South America
Microlife USA, Inc.
1617 Gulf to Bay Blvd
2nd Floor, Suite A
Clearwater, FL 33755, USA
Tel. +1 727 442 5353
Fax +1 727 442 5377
Email: msa@microlifeusa.com
www.watchbp.com
IB WatchBP Office EN 29
19
Loading...