microlife BP3GY1-2N Instruction Manual

Important Product and Safety Information
Follow Instructions for Use. This document provides important product operation and safety information regard­ing this Blood Pressure Monitor. Please read this document thoroughly before using the device and keep for future reference.
This blood pressure monitor is an automatic digital blood pressure measuring device for use by adults, including those with Diabetes mellitus, on the upper arm at home or in the doctor’s office. It enables a very fast and reliable measurement of the systolic and diastolic blood pressure, as well as the pulse rate, by way of the oscillometric method during cuff inflation. This device detects the appearance of irregular heartbeats during measurement and provides a warning signal when the irregular heartbeat is detected. The device can be connected via
®
Bluetooth
to a smart mobile device running the Microlife Connected
Health App.
This blood pressure monitor IS intended to be used:
• For self-measurement/monitoring of blood pressure and pulse in adults, including those with Diabetes mellitus; whereas the person being measured may be the user/operator of the device.
• Within a home healthcare environment.
• With a cuff located upon the user’s upper arm; ½ inch above the elbow over the artery as indicated on the cuff.
This blood pressure monitor IS NOT intended for use with:
• Pregnancy
• Preeclampsia
• Children under 12
• Neonatal patients
WARNING – Potentially hazardous situation that if not avoided may result in serious injury or death.
• Self-measuring means monitoring, not diagnosis or treatment. Unusual values must always be discussed with your doctor. Under no circum­stances should you independently alter the dosages of any drugs prescribed by your physician.
• Consult your physician before using this device if any of the following or similar conditions are present: arrhythmias such as atrial or ventric-
1
ular premature beats or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age, pregnancy, preeclampsia, renal diseases. Motion during
measurement, including trembling or shivering may affect
the measurement.
• Though not for use with children under 12, ensure that any children around this device are supervised; some parts are small enough to be swallowed and any present tubes or cables may provide a risk of strangulation.
• Ensure the cuff tubing is not kinked during use as harmful injury may occur due to
the effect of blood flow interference caused by high
pressure in the cuff not releasing.
DO NOT:
Use this device if you think it is damaged or anything appears unusual
(e.g. sporadic operation, open packaging upon purchase).
Use the displayed pulse for checking the frequency of heart pacemak-
ers as this device is not suitable for this action.
Open/modify this device; inaccuracy and/or harmful injury may result.Conduct many frequent measurements as this may result in harmful
injury due to blood flow interference.
Place the Cuff over a wound as this may cause further injury.Place and pressurize the Cuff over/near any present intravascular
access or therapy, or arteriovenous shunt, as this may cause blood flow interference and result in harmful injury.
Place and pressurize the Cuff over a limb near the side of a mastecto-
my as this may cause harmful injury.
Use the Blood Pressure Monitor on a limb simultaneously with other medical equipment on the same limb due to possible interference with such medical equipment.
Maintain pressure in the Cuff applied to the limb for a prolonged
amount of time. Ensure that circulation in the limb is not impaired by checking circulation if prolonged/repeated exposure to pressure occurs.
Service the device when being used or when power is supplied. When
the device is not in use and power is removed, cuff assembly, batter­ies and ac adapter (if applicable) may be replaced by the user with Microlife supplied replacements. No other parts/components are accessible.
2
CAUTION – Potentially hazardous situation that if not avoided may result in minor/moderate injury, property damage, and/or damage to the device
• This device contains sensitive electronics components. Avoid strong electrical or electromagnetic fields in the direct vicinity of the device (e.g., mobile telephones, microwave ovens). These can lead to temporary impairment of the measuring accuracy. Move the device to another location if interference is determined.
• When not using the Blood Pressure Monitor for extended lengths of time, remove the batteries to avoid potential battery leakage and damage to the monitor.
• When replacing the batteries, ensure all the batteries are replaced simultaneously to avoid battery damage and potential damage to the monitor. Microlife USA does not recommend using rechargeable batteries.
• To avoid inaccurate measurements and to lessen any discomfort from Cuff pressure, ensure the Cuff is placed correctly on the limb and fits correctly when snug (not tight), as indicated by markings with the Cuff.
• Consult your physician in cases of frequent irregular heartbeat detections.
• This Risk Indicator feature is provided in order to help you understand your potential blood pressure risk. However, this feature is neither a diagnosis nor a substitute for a medical examination. It is important to consult with your physician to determine your risk.
DO NOT:
Drop this device or expose it to strong vibrations; sensitive
components may be affected resulting in inaccuracies and/or operational issues.
Use the Blood Pressure Monitor outside of its specified operation temper-
ature and humidity rating, or if stored outside of its specified storage temperature and humidity rating. Avoid storage in direct sunlight.
Use this device in a moving vehicle; inaccurate measurements may result.Use third party accessories. Only use Microlife authorized accessories,
such as cuffs or AC adapters, as those not approved for use with the device may provide inaccurate measurements, injury, and/or damage the device.
3
STANDARDS
In addition to the standards stated in the Instruction Manual:
• This medical device is compliant with medical device and non-invasive blood pressure monitor standards IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and AAMI/ANSI/IEC 80601-2-30, and
• Electromagnetic standards IEC 60601—1-2 along with FCC Part 15, and
• Clinical Testing per standard ISO 81060-2:2013 was conducted on blood pressure device using the same measurement technology.
Please note: According to international standards, your monitor should be checked for accuracy every 2 years.
TYPE
IP20
Protected against solid foreign particles with a diameter of more than 12.5 mm,
no protection against water. Keep Dry. Type BF Applied Part Batteries and electronic instruments must be disposed of in accordance
with the locally applicable regulations, not with domestic waste.
Expected Life
Monitor 5 Years Cuff 2 Years
BHS
The B.H.S. (British Hypertension Society) clinical protocol was used to measure the accuracy of this product. Blood pressure units using the same measurement technology are graded “AA” for systolic/diastolic accuracy by independent investigators using the BHS protocol. This is the highest grading available for blood pressure monitors. Please see bhsoc.org for more information. (Uses the same algorithm as B.H.S. graded “AA” model number 3BT0-1)
FCC
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Changes or modifications to the product are not approved by Microlife USA and could void the user's authority to operate the equipment under FCC jurisdiction.
4
This equipment has been verified to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not used in accordance with the instruc­tions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment on and off, the user is encouraged to try to correct the interference by increasing the distance between the product and the affected device; or if applicable, reorientation the receiving antenna, or moving the product’s power plug to another receptacle.
Trademark Usage:
Apple, the Apple logo, iPad, and iPhone are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc.
Android and Google Play are both trademarks of Google Inc.
®
The Bluetooth by
Bluetooth SIG, Inc. and any use of such marks in this Blood Pressure
word mark and logos are registered trademarks owned
Monitor is under license. Other trademarks and trade names are those of their respective owners.
5
Bluetooth
®
Instruction M
Automatic
anual
Blood P
ressure Monitor
Table of Contents
1. Introduction
our autom
1.1. Y iabetes m
D
1.2.
omponents of your blood pressure monitor
C
2.
Blood pressure m
2.1.
Blood pressure cuff
2.2.
How
3.
atic blood pressure m ellitus validation
onitor
do I get started?
onitor
3.1. Protective screen film
3.2. Inserting the batteries
/ screen lock
itch
er sw
3.3. Pow sing the interactive touch screen
3.4. U
3.5. Setting the date and tim
e
3.6. Select the user
andar
and monit
er
m
re
su
ement
ur
eadings
st
or
?
t
n
e
t
elect
3.7. S
rolif
ic
3.8. M
onnecting t
3.9. C
3.10.Using t
o I
d
ow
H
4.
eparing
4.1. Pr ommon err
4.2. C
4.3. Using
t
4.4. Taking a
4.5. Memor
opping
t
4.6. S
he meas
aging Mode
er
e Av
he cuf
he AC pow
a
ke
a
t
ake a meas
to t
s
or
f
he cuf
ur
meas
iewing r
Rev
:
y
a meas
ing mode:
ur
f
er adapt
ea
m
ement
ement
ur
4.7. Hypertension risk indicator
4.8. Irregular heartbeat detector
4.9. Battery change indicator
Bluetooth® functions & application (app)
5.
5.1. Activating your Bluetooth®
5.2. Pairing your device
5.3. Unpairing your device
Error messages/troubleshooting
6.
d
Microlif
or
e Av
eraging
Mode
6
7. How is blood pressure measured?
7.1. What is blood pressure, and why does it fluctuate?
7.2. Measurement Guidelines
7.3. What can I do to change my blood pressure?
8. Care and maintenance
9. Limited warranty
10. Technical specifications
11. How to contact us
7
1. Introduction
1.1. Your Automatic Blood Pressure Monitor
Thank you for purchasing a fully automatic blood pressure monitor. Your monitor is designed to provide fast and reliable digital readings of your pulse, and systolic / diastolic blood pressure using the oscillometric method on your upper arm. It offers clinically proven accuracy and has been designed to be user friendly.
If you have additional questions regarding blood pressure measurements please contact Microlife Customer Service at 1-800-568-4147 or contact your doctor.
1.2. Diabetes mellitus validation
Diabetes, and stiff arteries, can make accurate blood pressure readings a challenge. Microlife's commitment to health has led to a breakthrough clinical validation for accurate readings in the presence of diabetes. This innovation reinforces our commitment to quality, accuracy, and helping you take control of your health.
Diabetes mellitus (or diabetes) is a chronic, lifelong condition that affects your body's ability to use the energy found in food. There are three major types of diabetes: type 1 diabetes, type 2 diabetes, and gestational diabetes.
Note: this monitor does not measure blood glucose.
8
2. Components of your blood pressure monitor
2.1. Measuring unit
Cuff Socket
(yellow)
Hypertension Risk Indicator
Memory
Date/Time
User Switch
Start, Stop, Enter
AC Adapter Port
MAM mode switch
Select 1 or 3 Measurements
Power, Screen Lock
9
STOP
START
Systolic
Diastolic
Pulse
User 1 & 2 Guest Bluetooth® Battery Indicator
Forward Backward
Mem
or
y
Enter MAM
Irregular Heartbeat
Pause Heartbeat Error 2 Error 3
10
2.2. Wide Range Cuff:
sure-tSFsure-t+
9”
12”
13”
16”
18”
M
L
large
medium
SF
+
For arm circumference 8.7” - 16.5 ” (22 cm - 42 cm) If you ever need to buy a replacement cuff, call
us toll-free at 1-800-568-4147.
Anti-Slip Cuff:
Your wide range cuff is equipped with anti-slip grips that ease the process of correctly position­ing the cuff on your arm.
Cuff Sizing:
The proper fit is essential for accurate blood pressure readings. The average arm circumference for a woman is 11", and the average for a man is 13". This cuff can accommodate arm circumfer­ences from 8.7" to 16.5" and fits most adults.
Arm circumference should be measured with a measuring tape in the middle of the relaxed upper arm. If the cuff is too small, call 1-800-568-4147 for additional information.
Note:
• Do not force cuff connection into the port opening.
• Make sure the cuff connection is not pushed into the AC adapter port.
11
3. How do I get started?
3.1. Protective screen film
There is a protective piece of static film covering the monitor’s screen. Please remove it before use.
3.2. Inserting the batteries
After you have unpacked your device, insert the batteries. The battery compartment is located on the back of the device.
a) Remove the battery cover. b) Insert the batteries (4 x size AAA 1.5 V),
observing the indicated polarity.
c) If a battery warning appears in the display, the
batteries are discharged and must be replaced.
Attention!
• After the low battery indicator appears, the device won’t function until the batteries have been replaced.
Please use AAA 1.5 V
batteries and replace
them all at the same time.
• If the blood pressure monitor is not used for long periods, remove the batteries from the device.
• We do not recommend using rechargeable batteries.
3.3. Power switch / screen lock
To activate your monitor, push the switch
unlocked
on the right hand side of the monitor to the unlocked position.
Low Battery Indicator
locked
12
3.4. Using the interactive touch screen
TIME PM
This monitor features an interactive screen with 2 touch zones similarly to traditional
buttons, but require
only a light touch of the finger to operate. The following functions are made easier
through the interactive touch screen:
• Setting time and date
• Scrolling through memories
• Selecting the user
• Starting/stopping the blood pressure measurement
3.5. Setting the date and time
1. After the new batteries are inserted and the monitor is unlocked, the year number flashes in the display. You can advance the year by pressing the plus or minus icon. To confirm and then set the month, press enter.
2. You can now set the month using the minus and plus. To confirm and then set the day, press enter.
3. Please follow the instructions above to set the day, hour and minutes.
4. Once you have set the last minute and pressed enter, the date and time are set and the time is displayed.
5. If you want to change the date and time, press and hold the time zone down for approximately 3 seconds until the year number starts to flash. Now you can enter the new values as described above.
6. To bypass setting date and time, press the time area of the touch screen.
which operate
Touch
(Enter)
(Enter)
ones
z
13
3.6. Select the user
This blood pressure monitor is designed to store 120 measurements for each of two users. Memories are not stored in guest mode.
Guest Mode
Before taking a measurement, be certain that the correct user has been selected.
a) With the monitor unlocked, press and hold the user icon, then release. b) The user icon will begin to blink. Tap again to change users. c) To confirm your selection wait 3 seconds, or press STOP.
3.7. Select measuring mode: standard or Microlife Averaging Mode (MAM)
This instrument enables you to select either standard (single measurement) or measurement averaging mode (automatic triple measurement).
a) To select Standard mode, push the switch on the side
of the monitor to Position "1."
b) To select Averaging mode, push the switch to Position
"3." Note the "MAM" icon will illuminate on the screen.
If you select 1, then only one measurement will be taken. If you select 3, the unit will inflate and deflate three times resulting in one final averaged measurement.
3.8. Microlife Averaging Mode (MAM)
• In Microlife Averaging Mode (MAM), three measure­ments are automatically taken in succession and the result is then automatically analyzed and displayed. Because your blood pressure constantly fluctuates, a result determined in this way is more reliable than one produced by a single measurement.
• After pressing the START button the selected Microlife Averaging Mode appears in the display as the MAM symbol.
• The bottom, right-hand section of the display shows a 1, 2 or 3 to indicate which of the 3 measurements is currently being taken.
14
3.9. Connecting the cuff and monitor
Insert the yellow cuff tube connector into the yellow socket on the left side of the instrument.
If your cuff connector is another color, please contact customer service as additional steps are needed to obtain an accurate reading.
3.10. Using the AC power adapter
You may also operate this monitor using the included AC adapter.
Use only the included AC adapter to avoid damaging the unit.
a) Ensure that the AC adapter and cable
are not damaged.
b) Plug the adapter cable into the AC
adapter port on the right side of monitor.
c) Plug the AC adapter into a 110 V power
socket (U.S. or Canada).
d) Test that power is available by pressing
the START button.
Note:
• No power is taken from the batteries
while the AC adapter is connected to the instrument.
• If the power is interrupted during a
measurement (e.g., by removal of the adapter from the wall socket), the instrument must be reset by removing the plug from the instrument.
If you have any questions regarding the AC adapter, call us at 1-800-568-4147.
15
4. How do I take a measurement? Please note: You should always be seated before and during measurement.
4.1. Preparing to take a measurement
• Avoid eating and smoking as well as all forms of exertion directly before measurement. These factors influence the measurement result. Find time to relax by sitting in an armchair in a quiet atmosphere for about 5 minutes before your measurement.
• Remove any garment that fits closely to your upper arm.
• Always measure on the same arm (normally left).
• Compare measurements at the same time of day, since blood pressure changes during the course of the day (as much as 20–40 mm Hg).
4.2. Common errors
Note: Comparable blood pressure measurements always require the same conditions. Conditions should always be quiet.
• All efforts by the user to support the arm can increase blood pressure. Make sure you are in a comfortable, relaxed position and do not flex any of the muscles in the measurement arm during the measurement. Use a cushion for support if necessary.
• If the arm artery lies considerably lower or higher than the heart, an erroneously high or low blood pressure will be measured. Each 15 cm (6”) difference in height between your heart and the cuff results in a measurement error of 10 mm Hg.
• Cuffs that are too narrow or too short result in false measurement values. Selecting the correct cuff is extremely important. Cuff size is dependent upon the circumference of the arm (measured in the center). The permissible range is printed on the cuff. If this is not suitable for your use, please call 1-800-568-4147.
• A loose cuff or a sideways protruding air pocket causes false measurement values.
• With repeated measurements, blood accumulates in the arm, which can lead to false results. Consecutive blood pressure measurements should be repeated after a 1 minute pause or after your arm has been held up in order to allow the accumulated blood to flow away.
16
4.3. Using the cuff
a) Pass the end of the cuff through the flat
metal ring so that a loop is formed. The hook and loop material must be facing outward. (Ignore this step if the cuff has already been prepared.)
b) Slide the cuff onto your upper left arm.
The rubber tube should be on the inside of your arm extending downward to your hand. Make certain the cuff lies approximately ½” (1 to 2 cm) above the elbow.
IMPORTANT: The red strip on the edge of the cuff (Artery Mark) must lie over the artery which runs down the inner side of the arm.
TIP: Align red artery mark to pinky finger.
c) To secure the cuff, wrap it around your
arm and press the hook and loop materi­al together. Check the position of the green SIZING SEAM on the cuff. Ensure green sizing seam overlaps green size range bar.
d) There should be little free space
between the arm and the cuff. You should be able to fit 2 fingers between your arm and the cuff. Clothing must not restrict the arm. Any piece of clothing which does must be removed. Cuffs that don’t fit properly result in false measurement values. Measure your arm circumference if you are not sure of proper fit.
8.7” - 16.5”
½
Brachial Artery
17
e) Lay your arm on a table so the cuff is at
the same height as your heart. Make sure your arm is relaxed and the tube is not kinked.
f) Remain seated quietly for 5 minutes
before you begin the measurement.
Comment If it is not possible to fit the cuff to your left arm, it
can also be placed on your right arm. However, all measurements should be made using the same arm. Comparable blood pressure measurements always require the same conditions (relax for several minutes before a reading).
Cuff on
right arm
18
4.4. Taking a measurement
After the cuff has been appropriately positioned the measurement can begin. Remain still during your measurement, do not flex muscles, and refrain from talking.
a) Remain seated quietly for 5 minutes before you
begin the measurement.
b) Press the START button. After the system
check, the monitor displays OK and the pump begins to inflate the cuff. On the display, the increasing cuff pressure is continually shown.
c) After automatically reaching an individual
pressure, the pump stops and the pressure will deflate slowly. The cuff pressure is displayed during the measurement.
d) When the device has detected your pulse, the
heart symbol in the display begins to blink. When the measurement has been concluded, the
e)
air will automatically release from the cuff. The measured systolic and diastolic blood pressure values, as well as the pulse, are now displayed.
f
) The measurement results are displayed until you
switch the device off. If no button is pressed for 1 minute, the device switches off automatically.
g) When the unit is set to the MAM (Microlife
Averaging Mode) setting, 3 separate measure­ments will take place in succession, after which your result is calculated and displayed as a single, averaged measurement (individual results are not displayed). There is a 15 second resting time between each measurement. A count-down indicates the remaining time between measurements.
If one of the measurements causes an error message, it will be repeated one more time. If any additional error occurs, the measurement will be discontinued and error code displayed.
Pumping Pressure
Measuring
Systolic
Diastolic
Pulse
Measurement
Complete
19
4.5. Memory: Reviewing readings
At the end of a measurement, this monitor automati­cally stores each result with date and time. This unit stores 120 memories for each of 2 users.
Viewing the stored values
• With the unit unlocked, press "M." The display first shows "A," then shows an average of all measurements stored in the unit for that user.
• Press the plus or minus icon to scroll between stored readings. Please note: Measurements for each user are averaged and stored
separately. Be certain that you are viewing the measurements for the correct user.
• To exit, press STOP.
Memory full
When the memory has stored 120 results, a new, measured value is stored by overwriting the oldest value.
Clear all values
• If you are sure that you want to permanently remove all stored values, hold down "M" until "CL" appears and then release "M".
• To continue and erase the values, press "M"again while "CL" is flashing.
• To discontinue with the clearing of memories, press “STOP.”
• Individual values cannot be cleared.
4.6. Stopping a measurement
If it is necessary to interrupt a blood pressure measurement for any reason (e.g., the patient feels unwell), the “STOP” button can be pressed at any time. The device then immediately lowers the cuff pressure automatically and enters sleep mode.
20
4.7. Hypertension risk indicator
Dk. Orange Dk. Orange Dk. Orange Dk. Orange Dk. Orange
The bars on the left-hand edge of the display show the possible range for a blood pressure reading. The reading can fall within the normal (green), borderline (yellow) or danger (orange, dark orange) range.
The classification is based on standards established by the American Heart Association (AHA) and American College of Cardiology (ACC) in 2017.
Refer to the chart in section 7.2 of this instruction manual for details of the classifications.
A triangle is displayed to the right of the traffic light bar according to your measurement. If the display shows a triangle in the:
green zone, your measurement is "Normal."
lower yellow zone, it is "Elevated."
upper yellow zone, it is "Stage 1 Hypertensive."
orange zone, it is "Stage 2 Hypertensive.”
lower dark orange zone, it is "Stage 2 Hypertensive.”
upper dark orange zone, it is "Hypertensive Crisis."
"Normal"
Blood Pressure
Reading
"Elevated"
Blood Pressure
Reading
"Stage 1 Hypertensive"
Blood Pressure
Reading
"Stage 2 Hypertensive"
Blood Pressure
Reading
"Hypertensive Crisis"
Blood Pressure
Reading
21
Loading...
+ 49 hidden pages