Follow Instructions for Use. This document provides
important product operation and safety information regarding
this Blood Pressure Monitor. Please read this document
thoroughly before using the device and keep for future
reference.
This blood pressure monitor is an automatic digital blood pressure measuring device for use by adults, on the upper arm at home or in the doctor’s
office. It enables a very fast and reliable measurement of the systolic and
diastolic blood pressure, as well as the pulse rate, by way of the oscillometric method during cuff inflation. This device detects the appearance of
irregular heartbeats during measurement and provides a warning signal
when the irregular heartbeat is detected. The device can be connected via
Bluetooth to a smart mobile device running the App.
This blood pressure monitor IS intended to be used:
•
For self-measurement/monitoring of blood pressure and pulse in adults;
whereas the person being measured may be the user/operator of the
device.
•
Within a home healthcare environment.
•
With a cuff located upon the user’s upper arm; ½ inch above the elbow
over the artery as indicated on the cuff.
This blood pressure monitor IS NOT intended for use with:
•
Pregnancy
•
Preeclampsia
•
Children under 12
•
Neonatal patients
WARNING – Potentially hazardous situation that if not
avoided may result in serious injury or death.
•
Self-measuring means monitoring, not diagnosis or treatment. Unusual
values must always be discussed with your doctor. Under no circumstances should you independently alter the dosages of any drugs
prescribed by your physician.
•
Consult your physician before using this device if any of the following or
similar conditions are present: arrhythmias such as atrial or ventricular
premature beats or atrial fibrillation, arterial sclerosis, poor perfusion,
diabetes, age, pregnancy, preeclampsia, renal diseases. Motion during
Page 3
2
measurement, including trembling or shivering may affect the measurement.
•
Though not for use with children under 12, ensure that any children
around this device are supervised; some parts are small enough to be
swallowed and any present tubes or cables may provide a risk of
strangulation.
•
Ensure the cuff tubing is not kinked during use as harmful injury may
occur due to the effect of blood flow interference caused by high pressure
in the cuff not releasing.
DO NOT:
Use this device if you think it is damaged or anything appears unusual
(e.g. sporadic operation, open packaging upon purchase).
Use the displayed pulse for checking the frequency of heart pacemakers
as this device is not suitable for this action.
Open/modify this device; inaccuracy and/or harmful injury may result.
Conduct many frequent measurements as this may result in harmful
injury due to blood flow interference.
Place the Cuff over a wound as this may cause further injury.
Place and pressurize the Cuff over/near any present intravascular access
or therapy, or arteriovenous shunt, as this may cause blood flow interference and result in harmful injury.
Place and pressurize the Cuff over a limb near the side of a mastectomy
as this may cause harmful injury.
Use the Blood Pressure Monitor on a limb simultaneously with other medical
equipment on the same limb due to possible interference with such medical
equipment.
Maintain pressure in the Cuff applied to the limb for a prolonged amount
of time. Ensure that circulation in the limb is not impaired by checking
circulation if prolonged/repeated exposure to pressure occurs.
Service the device when being used or when power is supplied. When
the device is not in use and power is removed, cuff assembly, batteries
and ac adapter (if applicable) may be replaced by the user with Microlife
supplied replacements. No other parts/components are accessible.
CAUTION – Potentially hazardous situation that if not avoided
may result in minor/moderate injury, property damage, and/or
damage to the device
•
This device contains sensitive electronics components. Avoid strong
Page 4
3
electrical or electromagnetic fields in the direct vicinity of the device
(e.g., mobile telephones, microwave ovens). These can lead to temporary
impairment of the measuring accuracy. Move the device to another
location if interference is determined.
• When not using the Blood Pressure Monitor for extended lengths of
time, remove the batteries to avoid potential battery leakage and
damage to the monitor.
• When replacing the batteries, ensure all the batteries are replaced
simultaneously to avoid battery damage and potential damage to the
monitor. Microlife USA does not recommend using rechargeable batteries.
• To avoid inaccurate measurements and to lessen any discomfort from
Cuff pressure, ensure the Cuff is placed correctly on the limb and fits
correctly when snug (not tight), as indicated by markings with the Cuff.
• Consult your physician in cases of frequent irregular heartbeat
detections.
• This Risk Indicator feature is provided in order to help you understand
your potential blood pressure risk. However, this feature is neither a
diagnosis nor a substitute for a medical examination. It is important to
consult with your physician to determine your risk.
DO NOT:
Drop this device or expose it to strong vibrations; sensitive components
may be affected resulting in inaccuracies and/or operational issues.
Use the Blood Pressure Monitor outside of its specified operation
temperature and humidity rating, or if stored outside of its specified
storage temperature and humidity rating. Avoid storage in direct
sunlight.
Use this device in a moving vehicle; inaccurate measurements may
result.
Use third party accessories. Only use Microlife authorized accessories,
such as cuffs or AC adapters, as those not approved for use with the
device may provide inaccurate measurements, injury, and/or damage the
device.
Page 5
4
STANDARDS
In addition to the standards stated in the Instruction Manual:
•
This medical device is compliant with medical device and non-invasive
blood pressure monitor standards IEC 60601-1, IEC 60601-1-2, IEC
60601-1-11, and AAMI/ANSI/IEC 80601-2-30, and
•
Electromagnetic standards IEC 60601—1-2 along with FCC Part 15, and
•
Clinical Testing per standard ISO 81060-2:2013 was conducted on blood
pressure device using the same measurement technology.
Please note: According to international standards, your monitor should be
checked for accuracy every 2 years.
TYPE
IP20
Protected against solid foreign particles with a diameter of more
than 12.5 mm, no protection against water. Keep Dry.
Type BF Applied Part
Batteries and electronic instruments must be disposed of in accordance
with the locally applicable regulations, not with domestic waste.
Expected Life
Monitor
5 Years
Cuff
2 Years
Trademark Usage:
Apple, the Apple logo, iPad, and iPhone are trademarks of Apple Inc.,
registered in the U.S. and other countries. App Store is a service mark of
Apple Inc.
Android and Google Play are both trademarks of Google Inc.
The Bluetooth® Smart word mark and logos are registered trademarks
owned by Bluetooth SIG, Inc. and any use of such marks in this Blood
Pressure Monitor is under license. Other trademarks and trade names are
those of their respective owners.
Page 6
5
Federal Communications Commission (FCC) Statement
15.21
You are cautioned that changes or modifications not expressly approved by
the part responsible for compliance could void the user’s authority to
operate the equipment.
15.105(b)
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation.
This equipment generates uses and can radiate radio frequency energy and,
if not installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this
equipment does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the
following measures:
-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
-Consult the dealer or an experienced radio/TV technician for help.
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two
conditions:
1) this device may not cause interference and
2) this device must accept any interference, including interference that may
cause undesired operation of the device.
Page 7
6
Premium Automatic Blood Pressure Monitor
Instruction Manual
Table of Contents
1.
Introduction
1.1. Your automatic blood pressure monitor
1.2. Diabetes validation
2.
Components of your blood pressure monitor
2.1. Blood pressure monitor
2.2. Blood pressure cuff
3.
How do I get started?
3.1. Screen film
3.2. Inserting the batteries
3.3. Power switch / screen lock
3.4. Using the interactive touch screen
3.5. Setting the date and time
3.6. Select the user
3.7. Select the measuring mode: standard or Microlife Averaging Mode
3.8. Microlife Averaging Mode
3.9. Connecting the cuff and monitor
3.10.Using the AC power adapter
4.
How do I take a measurement?
4.1. Preparing to take a measurement
4.2. Common errors
4.3. Using the cuff
4.4. Taking a measurement
4.5. Memory: Reviewing readings
4.6. Stopping a measurement
4.7. Hypertension risk indicator
4.8. Irregular heartbeat detector
4.9. Battery change indicator
5.
Bluetooth functions and application (App)
5.1. Bluetooth® icon indicator (on monitor)
5.2. iOS devices
5.3. Android OS Phones
6.
Error messages/troubleshooting
7. How is blood pressure measured?
7.1. What is blood pressure, and why does it fluctuate?
Page 8
7
7.2. Measurement Guidelines
7.3. What can I do to change my blood pressure?
8. Care and maintenance
9. Limited warranty
10. Technical specifications
11. How to contact us
Page 9
8
1. Introduction
1.1. Your Automatic Blood Pressure Monitor
Thank you for purchasing a fully automatic blood pressure monitor. Your
monitor is designed to provide fast and reliable digital readings of your
pulse, and systolic / diastolic blood pressure using the oscillometric method
on your upper arm. It offers clinically proven accuracy and has been
designed to be user friendly.
If you have additional questions regarding blood pressure measurements
please contact Microlife Customer Service at 1-800-568-4147 or contact
your doctor.
1.2. Diabetes validation
Microlife’s commitment to disease-state management has led to our
clinical validation for use with diabetes. This is a worldwide first that
reinforces our commitment to quality, accuracy and meaningful innovation.
Page 10
9
2.
Components of your blood pressure monitor
2.1. Measuring unit
Cuff Socket
(yellow)
Hypertension
Risk Indicator
Memory
Date/Time
User Switch
Start, Stop,
Enter
AC Adapter Port
MAM mode
switch
Select 1 or 3
Measur ements
Power, Screen Lock
Page 11
10
START
STOP
User 1 & 2
Guest
Bluetooth
Systolic
Battery Indicator
Forward
Backwar d
Memory
Diastolic
Pulse
MAM
Enter
Irregular Heartbeat
Pause
Heartbeat
Error 2
Error 3
Page 12
2.2. Wide Range Cuff:
For arm circumference 8.7” - 16.5” (22 cm - 42 cm)
If you ever need to buy a replacement cuff, call
us toll-free at 1-800-568-4147.
Anti-Slip Cuff:
Your wide range cuff is equipped anti-slip grips
that ease the process of correctly positioning the
cuff on your arm.
Cuff Sizing:
The proper fit is essential for accurate blood
pressure readings. The average arm circumference
for a woman is 11", and the average for a man is
13". This cuff can accommodate arm circumferences
from 8.7" to 16.5" and fits most adults.
Arm circumference should be measured with a
measuring tape in the middle of the relaxed upper
arm. If the cuff is too small, call 1-800-568-4147 for
additional information.
Note:
• Do not force cuff connection into the opening.
• Make sure the cuff connection is not pushed into
the AC adapter port.
18”
16”
13”
12”
9”
medium M
L
large
sure-fit SF
S
F
+
su
re-f it
+
Page 13
12
3.
Low Battery Indicator
unlocked
locked
How do I get started?
3.1. Screen film
There is a protective piece of static film covering the
monitor’s screen. Please remove it before use.
3.2. Inserting the batteries
After you have unpacked your device, insert the batteries. The battery compartment is located on the back of
the device.
a) Remove the battery cover.
b)
Insert the batteries (4 x size AAA 1.5 V), observ-
ing the indicated polarity.
c)
If a battery warning appears in the display, the
batteries are discharged and must be replaced.
Attention!
•
After the low battery indicator appears, the
device won’t function until the batteries have
been replaced.
•
Please use AAA 1.5 V batteries and replace them
all at the same time.
•
If the blood pressure monitor is not used for long
periods, remove the batteries from the device.
3.3. Power switch / screen lock
To activate your monitor, push the switch on
the right hand side of the monitor to the
unlocked position.
Page 14
13
3.4. Using the interactive touch screen
Touch zones
Enter Enter
TIM E P M
This monitor features an interactive screen
with 2 touch zones which operate similarly to
traditional buttons, but require only a light
touch of the finger to operate.
The following functions are made easier
through the interactive touch screen:
•
Setting time and date
•
Scrolling through memories
•
Selecting the user
•
Starting/stopping the blood pressure
measurement
3.5. Setting the date and time
1.
After the new batteries are activated and
the monitor is unlocked, the year number
flashes in the display. You can advance
the year by pressing the plus or minus
icon. To confirm and then set the month,
press enter.
2.
You can now set the month using the
minus and plus. To confirm and then set
the day, press enter.
3.
Please follow the instructions above to
set the day, hour and minutes.
4.
Once you have set the last minute and
pressed enter, the date and time are set
and the time is displayed.
5.
If you want to change the date and time,
press and hold the time zone down for
approximately 3 seconds until the year
number starts to flash. Now you can enter
the new values as described above.
6.
To bypass setting date and time, press the
time.
Page 15
14
GuestMod
3.6. Select the user
This blood pressure monitor is designed to
store 120 measurements for each of two users
Memories are not stored in guest mode.
e
Before taking a measurement, be certain that
the correct user has been selected.
a) With the monitor unlocked, press and hold the user icon, then release.
b) The user icon will begin to blink. Tap again to change users.
c) To confirm your selection wait 3 seconds, or press STOP.
3.7. Select measuring mode: standard or
Microlife Averaging Mode (MAM)
This instrument enables you to select either standard
(single measurement) or measurement averaging mode
(automatic triple measurement).
a) To select Standard mode, push the switch on the side
of the monitor to Position "1."
b) To select Averaging mode, push the switch to Position
"3." Note the "MAM" icon will illuminate on the screen.
If you select 1, then only one measurement will be taken.
If you select 3, the unit will inflate and deflate three times
resulting in one final averaged measurement.
3.8. Microlife Averaging Mode (MAM)
• In Microlife Averaging Mode (MAM), three measure-
ments are automatically taken in succession and the
result is then automatically analyzed and displayed.
Because your blood pressure constantly fluctuates, a
result determined in this way is more reliable than
one produced by a single measurement.
• After pressing the STOP/START button the selected
Microlife Averaging Mode appears in the display as
the MAM symbol.
• The bottom, right-hand section of the display shows
a 1, 2 or 3 to indicate which of the 3 measurements
is currently being taken.
Page 16
15
3.9. Connecting the cuff and monitor
Insert the yellow cuff tube connector into
the yellow socket on the left side of the
instrument.
If your cuff connector is another color, please
contact customer service as additional steps
are needed to obtain an accurate reading.
3.10. Using the AC power adapter
You may also operate this monitor using the
included AC adapter. Use only the included
AC adapter to avoid damaging the unit.
a) Ensure that the AC adapter and cable are
not damaged.
b) Plug the adapter cable into the AC
adapter port on the right side of monitor.
c) Plug the AC adapter into a 110 V power
socket (U.S. or Canada).
d) Test that power is available by pressing
the STOP/START button.
Note:
• No power is taken from the batteries
while the AC adapter is connected to the
instrument.
• If the power is interrupted during a
measurement (e.g., by removal of the
adapter from the wall socket), the instrument must be reset by removing the plug
from the instrument.
If you have any questions regarding the AC
adapter, call us at 1-800-568-4147.
Page 17
16
4. How do I take a measurement?
Please note: You should always be seated before and during
measurement.
4.1. Preparing to take a measurement
• Avoid eating and smoking as well as all forms of exertion directly
before measurement. These factors influence the measurement result.
Find time to relax by sitting in an armchair in a quiet atmosphere for
about 5 minutes before your measurement.
• Remove any garment that fits closely to your upper arm.
• Always measure on the same arm (normally left).
• Compare measurements at the same time of day, since blood pressure
changes during the course of the day (as much as 20–40 mmHg).
4.2. Common errors
Note: Comparable blood pressure measurements always require
the same conditions. Conditions should always be quiet.
• All efforts by the user to support the arm can increase blood pressure.
Make sure you are in a comfortable, relaxed position and do not flex
any of the muscles in the measurement arm during the measurement.
Use a cushion for support if necessary.
• If the arm artery lies considerably lower or higher than the heart, an
erroneously high or low blood pressure will be measured. Each 15 cm
(6”) difference in height between your heart and the cuff results in a
measurement error of 10 mmHg.
• Cuffs that are too narrow or too short result in false measurement
values. Selecting the correct cuff is extremely important. Cuff size is
dependent upon the circumference of the arm (measured in the center).
The permissible range is printed on the cuff. If this is not suitable for
your use, please call 1-800-568-4147.
• A loose cuff or a sideways protruding air pocket causes false measure-
ment values.
• With repeated measurements, blood accumulates in the arm, which
can lead to false results. Consecutive blood pressure measurements
should be repeated after a 1 minute pause or after your arm has been
held up in order to allow the accumulated blood to flow away.
Page 18
17
4.3. Using the cuff
a) Pass the end of the cuff through the flat
metal ring so that a loop is formed. The
hook and loop material must be facing
outward. (Ignore this step if the cuff has
already been prepared.)
b) Slide the cuff onto your upper left arm.
The rubber tube should be on the inside of
your arm extending downward to your
hand. Make certain the cuff lies approximately ½” (1 to 2 cm) above the elbow.
IMPORTANT:
The red strip on the edge
of the cuff (Artery Mark) must lie over the
artery which runs down the inner side of
the arm.
TIP:
Align red artery mark to pinky finger.
c) To secure the cuff, wrap it around your
arm and press the hook and loop material
together. Check the position of the green
SIZING SEAM on the cuff. Ensure green
sizing seam overlaps green size range bar.
d) There should be little free space between
the arm and the cuff. You should be able
to fit 2 fingers between your arm and the
cuff. Clothing must not restrict the arm.
Any piece of clothing which does must be
removed. Cuffs that don’t fit properly
result in false measurement values.
Measure your arm circumference if you
are not sure of proper fit.
8.7” - 16.5 ”
Brachia l Art ery
½
”
Page 19
18
e)
Lay your arm on a table so the cuff is at
the same height as your heart. Make sure
your arm is relaxed and the tube is not
kinked.
f)
Remain seated quietly for 5 minutes
before you begin the measurement.
Comment
If it is not possible to fit the cuff to your left arm, it
can also be placed on your right arm. However, all
measurements should be made using the same arm.
Comparable blood pressure measurements always
require the same conditions (relax for several
minutes before a reading).
Cuff on
right arm
Page 20
19
4.4. Taking a measurement
After the cuff has been appropriately positioned the
measurement can begin. Remain still during your
measurement, do not flex muscles.
a) Remain seated quietly for 5 minutes before you
begin the measurement.
b) Press the START/STOP button. After the system
check, the monitor displays OK and the pump
begins to inflate the cuff. On the display, the
increasing cuff pressure is continually shown.
c) After automatically reaching an individual
pressure, the pump stops and the pressure will
release. The cuff pressure is displayed during the
measurement.
d) When the device has detected your pulse, the
heart symbol in the display begins to blink.
e) When the measurement has been concluded, the
air will automatically release from the cuff. The
measured systolic and diastolic blood pressure
values, as well as the pulse, are now displayed.
f) The measurement results are displayed until you
switch the device off. If no button is pressed for 1
minute, the device switches off automatically.
g) When the unit is set to the MAM (Microlife
Averaging Mode) setting, 3 separate measurements will take place in succession, after which
your result is calculated and displayed as a single,
averaged measurement (individual results are not
displayed). There is a 15 second resting time
between each measurement. A count-down
indicates the remaining time between measurements.
If one of the measurements causes an error
message, it will be repeated one more time. If any
additional error occurs, the measurement will be
discontinued and error code displayed.
Pumping
Pressu re
Measuring
Systolic
Diastolic
Pulse
Measurement
Complete
Page 21
20
4.5. Memory: Reviewing readings
At the end of a measurement, this monitor automatically
stores each result with date and time. This unit stores
120 memories for each of 2 users.
Viewing the stored values
With the unit unlocked, press "M." The display first
shows "A," then shows an average of all measurements
stored in the unit for that user. Press the plus or minus
icon to scroll between stored readings.
Please note: Measurements for each user are averaged
and stored separately. Be certain that you are viewing
the measurements for the correct user.
To exit, press STOP.
Memory full
When the memory has stored 120 results, a new,
measured value is stored by overwriting the oldest value.
Clear all values
If you are sure that you want to permanently remove all
stored values, hold down "M" until "CL" appears and then
release "M". If you do not want to clear the values, press
STOP. To permanently clear the memory, press "M" while
"CL" is flashing.
Individual values cannot be cleared.
4.6. Stopping a measurement
If it is necessary to interrupt a blood pressure measurement for any reason (e.g., the patient feels unwell), the
STOP/START button can be pressed at any time. The
device then immediately lowers the cuff pressure
automatically and enters sleep mode.
Page 22
21
4.7. Hypertension risk indicator
The bars on the left-hand edge of the display show the possible range for
a blood pressure reading. The reading can fall within the normal (green),
borderline (yellow) or danger (orange, dark orange) range.
The classification is based on standards established by the 2017 AHA and ACC.
Refer to the chart in section 7.2 of this instruction manual for details of the
classifications.
A triangle is displayed to the right of the traffic light bar according to your
measurement.
• If your triangle is in the green zone, your measurement is "Normal,"
according to AHA/ACC 2017 standards.
• If your triangle is in the lower yellow zone, your measurement is "Elevated."
• If your triangle is in the upper yellow zone, your measurement is "Pre-
Hypertensive."
• If your triangle is in the orange zone, it is "Stage 1 Hypertensive."
• If your triangle is in the dark orange zone, it is "Stage 2 Hypertensive."
certain pulse irregularities were detected during the
measurement. In this case, the result may deviate
from your normal basal blood pressure – repeat the
measurement. In most cases, this is no cause for
concern. However, if the symbol appears on a
regular basis (e.g., several times a week with
measurements taken daily), we advise you to tell
your doctor.
Please show your doctor the following explanation:
Information on frequent appearance of the
irregular heartbeat symbol
This instrument is an oscillometric blood pressure
monitor device that also analyzes pulse frequency
during measurement. The instrument is clinically
tested.
If pulse irregularities occur during the measurement,
the irregular heartbeat symbol is displayed with the
measurement.
If the symbol appears frequently or if it suddenly
appears more often than usual, we recommend the
patient seek medical advice. The instrument does
not replace a cardiac examination but serves to
detect pulse irregularities at an early stage.
Page 24
23
4.9. Battery change indicator
Batteries almost discharged
When the batteries are approximately 75% used, the
battery symbol will flash a few times as soon as the
instrument is switched on. Although the
instrument will continue to measure reliably, you
should obtain replacement batteries.
Batteries discharged – replacements required
When the batteries are discharged, the battery
symbol will appear, unblinking, as soon as the instrument is switched on. You cannot take any additional
measurements and must replace the batteries.
1.
Open the battery compartment on the bottom of
the instrument.
2.
Replace all of the batteries at the same time,
ensure they are correctly connected, as shown on
the symbols in the compartment.
3.
The memory retains all values although date and
time (and possibly also set alarm times) must be
reset – the year number will flash automatically
after the batteries are replaced.
4.
To set date and time, follow the procedure
described in section 3.2.
Note:
Use four new AAA 1.5 V batteries. Do not use
batteries beyond their expiration date. If the monitor
is not going to be used for a prolonged period, the
batteries should be removed.
Page 25
24
5. Bluetooth functions and application (App)
Please download the Microlife Connected Health App
(referred to as App in this document) from Apple’s
App Store® or Google Play® before connecting your
devices. The App can be executed on a number of
mobile platforms with the following specifications:
Compatibility:
•
Apple® devices running iOS 8.0 or higher on
iPhone®, iPad®, and iPod touch®.
•
Android® phones running Android® 5.0 or higher.
Questions?
•
To access the tutorial, go to the App home screen
(Microlife Connected Health) and press the
“Settings” button.
•
Press “Tutorial” and select your topic.
•
If you have any issues with your setup, please
contact customer service toll free: 1-866-464-6184
for additional support
5.1. Bluetooth® icon indicator (on monitor)
The Bluetooth® icon indicator on your blood
pressure monitor, located in the upper left portion of
the screen, is designed to provide information about
the connection between your device and the monitor.
Please note the following:
•
The bluetooth® icon on your blood pressure
monitor will flash initially when ready to connect
with a device, and continue flashing for 2 minutes
until a connection occurs,
•
If a connection does not occur within 2 minutes, the
Bluetooth® function on the monitor will automatically turn off.
•
If the bluetooth® icon flashes for 10 seconds and
then turns off, there is a problem with the connection.
•
The icon will remain unblinking on the monitor
when data is uploading to the App.
Bluetooth
Icon
Page 26
25
T
On your device:
Bluetooth® icon
roubleshooting: Bluetooth® connection:
•
Monitor: Turn off and wait for 1 minute.
•
Device (eg. phone): Return to the Main Menu and
reselect Download Data again.
If connection issues persist, please contact customer
support
toll free: 1-866-464-6184.
5.2. iOS devices:
Setting up your iOS device:
•
In the settings menu, turn Bluetooth® “ON”
•
Select the Microlife Connected Health App
•
Select “User Profiles”
•
Select “+” at the top of the screen to create a new
user profile.
•
Review the terms of service, and select “Accept” to
proceed.
•
Enter your information and select the check mark to
confirm.
Connecting the devices for the first time:
On your device:
•
Open the Microlife Connected Health App and select
“Settings.”
On your blood pressure monitor:
•
Press and hold the power button on your monitor
for 8 seconds. Once connected, a Bluetooth® icon
will appear on the blood pressure monitor’s
screen.
•
Select “Pair Device”
•
“Paired Successfully” will appear if connected.
Setting up your blood pressure monitor:
Confirm the correct user is selected on your blood pressure
monitor:
•
Confirm the user icon on the top right hand side of
Page 27
26
the monitor screen is set to user 1 or 2.
•
To change users on your monitor, see section 3.8.
Automatic measurement transfer (iOS):
•
Open the Microlife Connected Health App
•
Select “Download data.”
•
Take your blood pressure.
•
Once complete, the blood pressure monitor will
automatically activate a Bluetooth® connection
with your iOS device. A Bluetooth® icon will
appear on the monitor’s screen when connected.
(The devices must have been previously connected,
and be in range).
•
An image of a blood pressure monitor and a phone
connecting will appear, and the readings will
transfer.
•
To save your data and return to the main menu,
select the check mark in the upper right hand
corner.
Your measurements will now be available for review
through the “View Data” button on the home screen.
Manual measurement transfer (iOS):
1.
Open the Microlife Connected Health App
2.
Select “User Profiles,” and select the profile which
you would like to upload. Once selected, the
profile will be highlighted, and a check mark will
appear to its left.
3.
Use the home button to exit and return to the main menu.
4.
Select “Download data.” An image of a blood
pressure monitor and a phone connecting will
appear, and the readings will transfer.
5.
To save your data and return to the main menu,
select the check mark in the upper right hand corner.
Your measurements will now be available for review
through the “View Data” button on the home screen.
Download data
Download
data
User Profiles
Download
data
Page 28
27
Saving readings stored on your monitor (iOS):
• Open the Microlife Connected Health App
• Select the profile which you would like to upload.
Once selected, the profile will be highlighted, and
check mark will appear to its left.
• On your blood pressure monitor: Press and hold the
power button for 8 seconds. Once connected, a
Bluetooth® icon will appear on the blood pressure
monitor’s screen.
•
Follow steps 4 & 5 in “saving measurements
manually”
5.3 Android OS Phones
Setting up your phone:
• In the settings menu, turn Bluetooth® “ON”
• Select the Microlife Connected Health App
• Select “User Profile”
• Select “+” at the top of the screen to create a new
user profile.
• Review the terms of service, and select “Accept”
to proceed.
• Enter your information and select the check mark
to confirm.
On your blood pressure monitor:
• Confirm the correct user is selected on your blood
pressure monitor. The user icon on the top left
hand side of the monitor screen should be set to
your preference - 1 or 2).
• Press and hold the “START/STOP” button on the
monitor for 8 seconds until the Bluetooth® icon
flashes.
• To change users on your monitor, see section 3.8.
Page 29
28
Saving readings to your Android phone:
On your phone:
• Open the Microlife Connected Health App
• Select “User Profiles,” and then select the profile
which you would like to upload (the profile will
be highlighted in gray when selected).
•
Note: At least one blood pressure reading must
be taken and recorded on the monitor before
the data can be transferred
On your blood pressure monitor:
• Press and hold the power button on your
monitor for 8 seconds. Once connected, a
Bluetooth® icon will flash on the blood
pressure monitor’s screen.
On your phone:
• Select “Download data.” Your phone will
automatically search and list available
Bluetooth® devices.
• Select your monitor from the list (BP3GY1-2N),
and the readings will automatically transfer.
• After the readings have been transfered, a
summary of all your readings will be
displayed.
• To save your readings and return to the main
menu (home screen), select the check mark in
the upper right hand corner.
• To view saved readings, return to the home
screen and select “View Data”
Note: If no connection between the monitor and
the phone has been made within 10 seconds, a
“SCAN” button will appear in the top right
corner of the phone’s screen
• To search for your monitor again, press
“SCAN.”
• Once the phone has detected your monitor,
select it from the list (BP3GY1-2N).
Bluetooth® icon
Download
data
Page 30
29
6. Error messages/troubleshooting
If an error occurs during a measurement, the measurement
is discontinued and a corresponding error code is displayed
(example: Error no. 2).
Error No. Possible cause(s)/Solutions
ERR 1 Weak signal. The tube may have loosened, or no
pulse was detected.* Ensure cuff connections are tight
with proper cuff placement. See section 4.3.
ERR 2
Movement detected. Unnatural pressure impulses
influenced the measurement result. Reason: The arm
was moved during the measurement (artifact). Repeat
measurement, keeping still and quiet.
ERR 3
No pressure detected. Inflation of the cuff takes too
long. The cuff is not correctly seated or the hose
connection is not tight. Re-position cuff and repeat
the measurement.
ERR 5
Abnormal reading. The measured readings indicate an
unacceptable difference between systolic and
diastolic pressures. Take another reading following
directions carefully. Contact your doctor if you continue to get unusual readings.
HI
The cuff pressure is too high. Relax for 5 minutes and
repeat the measurement.*
LO
The pulse is too low (less than 40). Repeat the
measurement.*
*If this or any other problem occurs repeatedly, please consult your
doctor.
Page 31
30
Other possible errors and their solutions
If problems occur when using the device, the following points should be
checked and, if necessary, the corresponding measures are to be taken:
Malfunction Remedy
The display remains blank when the
instrument is switched on although
the batteries are in place.
The pressure does not rise although
the pump is running.
The device frequently fails to
measure blood pressure values or
the values measured are too low or
high.
Every measurement results in a
different value, although the device
functions normally and normal
values are displayed.
Blood pressure values differ from
those measured by my doctor.
After the instrument has inflated the
cuff the pressure falls very slowly,
or not at all. (No reasonable
measurement possible.)
1. Check batteries for the correct
polarity.
2. If the display is unusual, remove
the batteries and exchange them
for new ones.
Check the connection of the cuff
tube and connect properly.
1. Check the positioning of the cuff.
2. Measure blood pressure again in
peace and quiet, carefully following the details in section 4.
Please read the following information and points listed in section 4.2
"Common errors." Repeat the
measurement.
Please note: Blood pressure
fluctuates continually so successive measurements will show
some variability.
Record the daily development of the
measured values and consult your
doctor (see section 4.1)
Please note: Individuals visiting
their doctor frequently experience anxiety which can result
in a higher reading than at home
under resting conditions.
1. Check cuff connections.
2. Ensure the unit has not been
tampered with.
Page 32
31
7. How is blood pressure measured?
7.1. What is blood pressure, and why does it fluctuate?
Your level of blood pressure is determined in the circulatory center of the
brain and adjusts to a variety of situations through feedback from the
nervous system. To adjust blood pressure, the strength and frequency of
the heart (pulse), as well as the width of circulatory blood vessels is altered.
Blood vessel width is affected by fine muscles in the blood vessel walls.
Your level of arterial blood pressure changes periodically during heart
activity. During the "blood ejection" (Systole), the value is highest (systolic
blood pressure value). At the end of the heart’s "rest period" (Diastole),
pressure is lowest (diastolic blood pressure value).
Blood pressure values must lie within certain normal ranges in order to
prevent particular diseases.
Blood pressure is too high if your systolic blood pressure is over 160
mmHg and/or your diastolic pressure is above 100 mmHg, while at rest.
In this case, please consult your physician immediately. Long-term values
at this level endanger your health due to continual damage to the blood
vessels in your body.
If your systolic blood pressure values are between 140 mmHg and 159
mmHg and/or the diastolic blood pressure values are between 90 mmHg
and 99 mmHg, consult your physician. Regular self-checks are necessary.
If you have blood pressure values that are too low (i.e., systolic values
under 105 mmHg and/or diastolic values under 60 mmHg), consult your
physician.
Even with normal blood pressure values, a regular self-check with your
blood pressure monitor is recommended. You can detect possible changes
in your values early and react appropriately.
If you are undergoing medical treatment to control your blood pressure,
keep a record of values along with time of day and date. Show these
values to your physician. Never use the results of your measurements
to independently alter the medication prescribed by your
physician.
Page 33
32
7.2. Measurement Guidelines
Category
Systolic
Diastolic
(mmHg) (mmHg)
Normal
<120 and <80
Elevated
120-129 and <80
Hypertension
Stage 1 Hypertension
130-139 or 80-89
Stage 2 Hypertension
140-179 or 90-119
Hypertensive Crisis
≥
180 or ≥120
The following standards for assessing high blood pressure (in adults) have
been established by the AHA and ACC in 2017.
Additional information
•
This chart reflects a 2017 update to blood pressure standards.
•
If your values are mostly normal under resting conditions but exceptionally high under conditions of physical or psychological stress, it is
possible that you are suffering from so-called "labile hypertension."
Consult your doctor.
Increased blood pressure values (various forms of hypertension) are
associated with considerable health risks over time. Arterial blood
vessels in your body are endangered due to constriction caused by
deposits in the vessel walls (arteriosclerosis). A deficient supply of
blood to important organs (heart, brain, muscles) can result from
arteriosclerosis. Furthermore, the heart will become structurally
damaged with increased blood pressure values.
c)
There are many different causes of high blood pressure. We differentiate between the common primary (essential) hypertension and secondary hypertension. The latter group can be ascribed to specific organ
Page 34
33
malfunctions. Please consult your doctor for information about the
possible origins of your own increased blood pressure values.
d)
There are measures which you can take to reduce and even prevent high
blood pressure. These measures must be permanent lifestyle changes.
1)
Eating habits
•
Strive for a normal weight corresponding to your age. See your doctor for your
ideal weight.
•
Avoid excessive consumption of common salt.
•
Avoid fatty foods.
2)
Previous illnesses
•
Consistently follow all medical instructions for treating illness such as:
•
Diabetes (diabetes mellitus)
•
Fat metabolism disorder
•
Gout
3)
Habits
•
Give up smoking completely.
•
Drink only moderate amounts of alcohol.
•
Restrict your caffeine consumption (e.g., coffee).
4)
Physical constitution
•
After a preliminary medical examination, do regular exercise.
•
Choose sports which require stamina and avoid those which require
strength.
•
Avoid reaching the limit of your performance.
•
With previous illnesses and/or an age of over 40 years, please consult
your doctor before beginning your exercise routine. You must receive
advice regarding the type and extent of exercise that is appropriate for
you.
Page 35
34
8.
Gasoline
Care and maintenance
a) The cuff contains a sensitive airtight
bubble. Handle this cuff carefully and avoid
all types of stress through twisting or
buckling.
b)
Clean the device with a soft, dry cloth. Do
not use gasoline, thinners or similar
solvents. Spots on the cuff can be removed
carefully with a damp cloth and soapsuds.
Do not wash the cuff in a dishwasher,
clothes washer or submerge it in
water.
c)
Handle the tube carefully. Do not pull on it.
Do not allow the tubing to kink and keep it
away from sharp edges.
d)
Never open the monitor. This invalidates
the manufacturer’s warranty.
Page 36
35
9.
Limited Warranty
Your Automatic Blood Pressure Monitor is warranted for 5 years by
Microlife USA Inc, against manufacturer defects for the original purchaser
only, from date of purchase.
The 5 year warranty applies to the monitor only. The following
accessories are warranted for 1 year: cuff and AC adapter. Batteries
are not covered by this warranty.
The warranty does not apply to consequential and incidental damages, or
damage caused by batteries, improper handling, and accidents.
Professional use, not following the operating instructions, and alterations
made to the monitor or accessory by third parties, are also not included in
this warranty. Some states do not allow the exclusion or limitation of
incidental or consequential damages, so the above limitation or exclusion
may not apply to you.
Microlife USA Inc will investigate your concern. A monitor or accessory as
defined by this warranty, determined to be out of specification, will be
replaced and shipped to you at no cost. A monitor or accessory as defined
by this warranty, determined to be within specification, will be returned to
you with a report of findings, at no cost.
Please use the below customer service contact information to reach
Microlife USA Inc. regarding any warranty concerns. We ask that you
please contact us before sending any product back in order to better
identify, and more quickly process, your concern.