Microlife BP 3BU1-5 Instruction Manual

Page 1
microlife BP 3BU1-5
Wrist Watch Blood Pressure Monitor
Instruction Manual (1-31)
øøËË··ÓÓ ÎΡˇ ËËÁÁÏÏÂÂÂÂÌÌËˡˇ ÍÓ‚ˇÌÓ„Ó ‰‡‚ÎÂÌˡ ̇ Á‡ÔˇÒڸ (32-63)
Automatyczny nadgarstkowy ciÊnieniomierz krwi
Instrukcja obs∏ugi (64-93)
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 2
Page 2
Armenia
Alfa Pharmmedimpex 8-33 Komitas ave. AM-375033 Yerevan / Armenia Tel. +374 1 569 698 Fax +374 1 544 772
Belarus
Degreemed Av. Fabrichnaia 22, 2 floor BY-220033 Minsk / Belarus Tel. +375 172 06 5349 Fax +375 172 06 5349
Bulgaria
BULOPTICA 29 “Iskar” Str. BG-Sofia 1000 / Bulgaria Tel. +359 298 08 498 Fax +359 298 11 709
Croatia
MEDICUS MATEA Goleska 20 HR-10 020 Zagreb / Croatia Tel. +385 1 654 25 55 Fax +385 1 654 25 60
Estonia
Oriola AS Estonia Saku tn. 8 EE-11314 Tallinn / Estonia Tel. +372 65 15 100 Fax +372 65 15 111
Greece
Karabinis ALFA – GAUZE Ltd. 151 Lauriou str. Peania GR-19002 Athens Tel. +30 1 66 45 751 Fax +30 1 66 46 000
Georgia
Babutsa Ltd. 13 Gamsakhurdia av. GE-380015 Tbilisi / Georgia Tel. +995 (32) 379 873 Fax +995 (32) 530 532
Hungary
MEDTRON Ltd. Budapest 62 Pf. 920 HU-1386 Budapest / Hungary Tel. +361 412 8004 Fax +361 412 8004
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 3
Page 3
Latvia
Oriola-Riga SIA Dzelzavas cela 120M LV-1082 Riga / Latvia Tel. +371 780 2461 Fax +371 780 2460
Lithuania
UAB "ORIOLA VILNIUS" Laisves pr. 75 LT-2022 Vilnius / Lithuania Tel. +370 2 688 401 Fax +370 2 688 400
Poland
ECO CLIMA Mc Gregor Ul. 11-go Listopada 97 PL-41800 Zabrze / Poland Tel. +48 32 274 14 26 Fax +48 32 274 14 25
Romania
Algomed Str. V.Poloni nr. 9 Sector 5 RO-Bucuresti / Romania Tel. +40 1 411 89 08
Slovakia
Interpharm Slovakia a.s. Podzahradna ul. C. 47 SI-821 06 Bratislava / Slovakia Tel. +421 245 24 4962 Fax +421 245 25 7961
Slovenia
Simps'S d.o.o. IOC Trzin Motnica 3 SI-1236 Trzin / Slovenia Tel. + 386 1 5621 338 Fax + 386 1 5621 339
Ukraine
Microlife Ukraine ul. Serova-Naberesznaya 5a UA-49000 Dniepropetrovsk / Ukraine Tel. +38 (0562) 342 206 Fax +38 (056) 778 57 63 UA-03187 Kiev / Ukraine Tel. +38 (044) 568 20 92 Fax +38 (044) 267 73 84
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 4
Page 4
Wrist Watch Blood Pressure Monitor
Instruction Manual
Table of contents
1. Introduction
1.1. Features of the BP 3BU1-5
1.2. Important information about self-measurement
2. Important information on the subject of blood-pressure and its measurement
2.1. How does high/low blood-pressure arise?
2.2. Which values are normal?
2.3. What can be done, if regular high/low values are obtained?
2.4 MAM technology (Microlife Average Mode)
3. The various components of the blood-pressure monitor
4. Putting the blood-pressure monitor into operation
4.1. Inserting the batteries
4.2. Reading the set date
4.3. Setting the time and date and user selection
5. Carrying out a measurement
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 4
Page 5
6. Printer functions
6.1. How to print the last stored measurement (single data print)
6.2. How to print all measurements stored in memory (data print out with chart)
6.3. Stop printing
7. Error messages/malfunctions
8. Care and maintenance, recalibration
9. Guarantee
10. Reference to standards
11. Technical specifications
12. www.microlife.com
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 5
Page 6
1. Introduction
1.1. Features of the BP 3BU1-5
The blood-pressure monitor BP 3BU1-5 (with integrated time/date display) is a fully automatic, digital blood-pressure measuring device for use on the wrist, which enables very fast and reliable measure­ment of the systolic and diastolic blood-pressure as well as the pulse frequency by way of the oscillometric method of measuring.The de­vice offers a very high and clinical tested measurement accuracy and has been designed to provide a maximum of user-friendliness. Before using, please read through this instruction manual carefully and then keep it in a safe place. For further questions on the subject of blood-pressure and its measurement, please contact your doctor.
Attention!
1.2. Important information about self-measurement
Do not forget: self-measurement means control, not diagnosis or treatment. Unusual values must always be discussed with your doctor. Under no circumstances should you alter the dosages of any drugs prescribed by your doctor.
The pulse display is not suitable for checking the frequency of heart pacemakers!
In cases of cardiac irregularity (Arrhythmia), measurements made with this instrument should only be evaluated after con­sultation with the doctor.
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 6
Page 7
2. Important information on the subject of blood-pressure and its measurement
2.1. How does high/low blood-pressure arise?
The level of blood-pressure is determined in a part of the brain, the so-called circulatory centre, and adapted to the respective situation by way of feedback via the nervous system. To adjust the blood-pressure, the strength and frequency of the heart (Pulse), as well as the width of circulatory blood vessels is altered. The latter is effected by way of fine muscles in the blood-vessel walls. The level of arterial blood-pressure changes periodically during the heart activity: During the «blood ejection» (Systole) the value is maxi­mal (systolic blood-pressure value), at the end of the heart’s «rest pe­riod» (Diastole) minimal (diastolic blood-pressure value). The blood-pressure values must lie within certain normal ranges in order to prevent particular diseases.
2.2. Which values are normal?
Blood pressure is too high if at rest, the diastolic pressure is above 90mmHg and/or the systolic blood-pressure is over 160mmHg. In this case, please consult your doctor immediately. Long-term values at this level endanger your health due to the associated advancing damage to the blood vessels in your body. Should the systolic blood-pressure values lie between 140mmHg and
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 7
Page 8
Even with normal blood-pressure values, a regular self-check with your blood-pressure monitor is recommended. In this way you can detect possible changes in your values early and react appropriately. If you are undergoing medical treatment to control your blood pres­sure, please keep a record of the level of your blood pressure by car­rying out regular self-measurements at specific times of the day. Show these values to your doctor. Never use the results of your
measurements to alter independently the drug doses pre­scribed by your doctor.
Table for classifying blood-pressure values (units mmHG) according to World Health Organization:
Range Systolic Diastolic Measures
Blood-pressure Blood-pressure
Hypotension lower than 100 lower than 60 Check with
your doctor
Normal range between 100 between 60 Self-check
and 140 and 90 Mild between 140 between 90 Consult your hypertension and 160 and 100 doctor Moderately between 160 between 100 Consult your serious and 180 and 110 doctor hypertension Serious higher than 180 higher than 110 Consult your
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 8
Page 9
hypertension». Please consult your doctor if you suspect that this might be the case.
Correctly measured diastolic blood-pressure values above
120mmHg require immediate medical treatment.
2.3. What can be done, if regular increased/low values are
obtained?
a) Please consult your doctor. b) Increased blood-pressure values (various forms of hypertension)
are associated long- and medium term with considerable risks to health. This concerns the arterial blood vessels of your body, which are endangered due to constriction caused by deposits in the vessel walls (Arteriosclerosis). A deficient supply of blood to important organs (heart, brain, muscles) can be the result. Furthermore, with long-term continuously increased blood-pres­sure values, the heart will become structurally damaged.
c) There are many different causes of the appearance of high
blood pressure. We differentiate between the common primary (essential) hypertension, and secondary hypertension. The latter group can be ascribed to specific organic malfunctions. Please consult your doctor for information about the possible origins of your own increased blood pressure values.
d) There are measures which you can take, not only for reducing a
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 9
Page 10
A) Eating habits
Strive for a normal weight corresponding to your age. Reduce
overweight!
•Avoid excessive consumption of common salt.
•Avoid fatty foods.
B) Previous illnesses
Follow consistently any medical instructions for treating previ-
ous illness such as:
Diabetes (Diabetes mellitus)
Fat metabolism disorder
Gout
C) Habits
Give up smoking completely
Drink only moderate amounts of alcohol
Restrict your caffeine consumption (Coffee)
D) Physical constitution:
After a preliminary medical examination, do regular sport.
Choose sports which require stamina and avoid those which re-
quire strength.
•Avoid reaching the limit of your performance.
With previous illnesses and/or an age of over 40 years, please
consult your doctor before beginning your sporting activities. He
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 10
Page 11
An advanced measurement accuracy is achieved by the auto-
matic analysis of three successive measurements.
The new system provides reliable values for the doctor and can
be used as the basis for reliable diagnostics and medication therapy for high blood pressure.
A) Why MAM?
Human blood pressure is not stable
Scattering of devices
B) Key advantages
Reduction of:
Device scattering
Insufficient rest prior to measurement
Movement artefacts
Cuff positioning influences
C) Medical benefits
Improved accuracy
Reliable patient self-measurement data for the doctor
Safe hypertension diagnostic
Reliable therapy control
D) Measurement sequence
•With the full measurement cycles, the total measurement time
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 11
Page 12
3. The various components of the blood-pressure monitor
The illustration shows the blood-pressure monitor BP 3BU1-5, con­sisting of:
Wrist cuff Type WC1 13.5–19.5 cm, for wrist circumference of 13.5 to 19.5 cm
4. Putting the blood-pressure monitor into operation
4.1. Inserting the batteries
After you have unpacked your device, first insert the batteries. The battery compartment is located on the left lateral side of the device (see illustration).
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 12
Page 13
b) Insert the batteries (2 x size AAA 1.5V), thereby observing the
indicated polarity.
c) If the battery warning appears in the display, the batteries are
empty and must be replaced by new ones.
Attention!
After the battery warning appears, the device is blocked until
the batteries have been replaced.
Please use «AAA» Long-Life or Alkaline 1.5V Batteries. The use
of 1.2V Accumulators is not recommended.
If the blood-pressure monitor is left unused for long periods,
please remove the batteries from the device.
Functional check: Press and hold the 0/I button to test all the dis­play elements. When functioning correctly all segments appear.
4.2. Reading the set date
Please press the «Time» key.
4.3. Setting the time and date and user
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 13
Page 14
You must then re-enter the date and current time. For this, please proceed as follows (Example: Entering 2001-06-20 Time 09:30 o’­clock):
1) Press the TIME button
for at least 3 seconds. The display now indi­cates the set year, dur­ing which the last two characters blink.
2) The correct year can be
entered by pressing the MEMORY button. (Example: 1 x press)
3) Press the TIME button
again. The display now switches to the current date, during which the first character (month)
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 14
Page 15
4) The corresponding
month can now be en­tered by pressing the MEMORY button. Example: 6 x presses)
5) Press the TIME button
again. The last two characters (day) are now blinking
6) The corresponding day
can now be entered by pressing the MEMORY button. (Example: 15 x presses)
7) Press the TIME button
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 15
Page 16
8) The corresponding hour
can now be entered by pressing the MEMORY button. (Example: 9 x presses)
9) Press the TIME button
again. The last two characters (Minutes) now blink.
10) The exact time can now
be entered by pressing the MEMORY button. (Example: 30 x presses)
11) Now after all settings
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 16
Page 17
User selection:
This advanced blood pressure monitor allows you to track blood pressure readings for 2 individuals independently. a) Before measurement
make sure you set the unit for the intended user. The unit can track results for 2 individuals. (User 1, User 2)
b) The unit is set to User 1.
Click the TIME button to change to User 2.
c) click the MEMORY but-
ton to change into User 1
d) We suggest the first
person to take their pressure to be User 1.
Further Information
With each press of the button
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 17
Page 18
5. Carrying out a measurement
5.1. Before the measurement:
•Avoid eating, smoking as well as all forms of exertion directly
before the measurement. All these factors influence the mea­surement result. Try and find time to relax by sitting in an arm­chair in a quite atmosphere for about ten minutes before the measurement.
Measure always on the same wrist (normally left).
Attempt to carry out the measurements regularly at the same
time of day, since the blood-pressure changes during the course of the day.
5.2. Common sources of error:
Note: Comparable blood-pressure measurements always re­quire the same conditions! These are normally always quiet conditions.
All efforts by the patient to support the arm can increase the
blood-pressure. Make sure you are in a comfortable, relaxed po­sition and do not activate any of the muscles in the measure­ment arm during the measurement. Use a cushion for support if necessary.
If the wrist artery lies considerably lower (higher) than the
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 18
Page 19
peated after a 5 minute pause or after the arm has been held up in order to allow the accumulated blood to flow away (after at least 3 minutes).
5.3. Fitting the cuff
a) Remove all eventual objects
and jewellery (e.g. wristwatch) from the wrist in question. Draw the cuff over the wrist.
b) The distance between the
cuff and the hand should be approx. 10 mm.
c) Secure the cuff with the
Velcro fastener, so that it lies comfortably and not too tight, whereby no space should remain between the cuff and the wrist.
d) Lay the arm on a table, with
the palm upwards.Support the
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 19
Page 20
5.4. Measuring procedure
After the cuff has been appropriately positioned, the measurement can begin:
a) Press the 0/I-button, the
pump begins to inflate the cuff. In the display, the in­creasing cuff-pressure is continually displayed.
b) After reaching the inflation
pressure, the pump stops and the pressure slowly falls away. The cuff-pressure (large characters) is dis­played during the measure­ment. When the device has detected the pulse, the heart symbol in the display begins to blink and a beep tone is audible for every pulse beat.
c) When the measurement has
been concluded, a long beep tone sounds. The measured
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 20
Page 21
The measurement results are displayed, until you switch the device off. If no button is pressed for 5 minutes, the device switches auto­matically off, to save the batteries.
d) When the unit is set to Average Mode
setting generally 3 separate measure­ments will take place in succession and calculates your detected blood pressure value. There will be 35 seconds resting time in-between each measurement. A count down indicates the remaining time and a beep will sound 5 seconds before the 2nd and 3rd readings will begin. In case that the single data of each cycle differ too much from each other, a fourth measurement is performed before the re­sult will be displayed. In rare cases the blood pressure is such unstable that even after four measurements the data vary too much. In this case „ERR 6“ is shown and no result can be given. If one of measurement causes an error message it is repeated.
5.5. Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for any reason (e.g. the patient feels unwell), the 0/I
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 21
Page 22
5.6. Memory – storage and recall of the measurements
The blood-pressure monitor automatically stores each of the last 30 measurement val­ues. By pressing the MEMORY button, the last measurement (MR1) as well as the fur­ther last 29 measurements (MR2, MR3, ..., MR30) can be displayed one after the other.
(MR1: Values of the (MR2-MR30: Values of the last measurement) measurement before MR1)
Further information
Measurements should not occur soon after each other, since other­wise, the results will be falsified. Wait therefore for several minutes in a relaxed position, sitting or lying, before you repeat a measure­ment.
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 22
Page 23
In order to delete all stored readings, depress the MEMORY button for at least 7 seconds, the display will show the symbol " CL” and 3 short beep sounds will be heard to indicate dele­tion of stored readings.
Note:
If your last stored measure­ment was taken in the Average Mode, you will see the „MAM“ symbol in approximately 3 seconds during the deletion procedure. (The display will show these symbols during deletion of all stored measure­ments if the last measurement is taken from the Average Mode)
6. Printer Funtions
This unit can be used in connection with the Microlife diagnostic printer. The unit can print out measurements stored in the memory by pressing the PRINTER button. There are 2 kinds of formats available.
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 23
Page 24
b) Connect the printer with the monitor and switch the
printer on (for details please refer to the manual of the printer).
c) Click the PRINTER button of the monitor once.
The display will show "Pr1” and the printer will print the last stored measurement for User 1 or 2 based on your selection.
6.2. How to print all measurements stored in memory and a
data chart, (data print out with chart).
a) Select correct printer objective first before you
depress the PRINTER button.
b) Connect the printer with the monitor and switch the printer on
(for details please refer to the manual of the printer).
INPUT Socket
Printer Port Socket
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 24
Page 25
6.3. Stop printing
If you would like to stop printing, click the PRINTER button during the printing process.
NOTE:
Further details regarding the printer can be found in the printer instruction manual.
7. Error messages/malfunctions
If an error occurs during a measurement, the mea­surement is discontinued and a corresponding error code is displayed (Example: Error No. 2).
Error No. Possible cause(s) ERR 1 No pulse has been detected. ERR 2 Unnatural pressure impulses influence the measurement
result. Reason: The arm was moved during the Measurement (Artefact).
ERR 3 The inflation of the cuff takes too long. The cuff is not
correctly seated.
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 25
Page 26
Other possible malfunctions and their elimination
If problems occur when using the device, the following points should be checked and if necessary, the corresponding measures are to be taken:
Malfunction Remedy
The display remains empty 1.Check batteries for correct polarity when the instrument and if necessary insert correctly. is switched on although 2.If the display is unusual, re-insert the batteries are in place. batteries or exchange them. The device frequently fails 1.Check the positioning of the cuff. to measure the blood 2.Measure the blood-pressure again in pressure values, or the peace and quiet under observance of values measured the details made under point 5. are too low (too high). Every measurement 1.Please read the following information produces a different and the points listed under «Common value although the sources of error». Repeat the instrument functions measurement. normally and the values Please note: Blood pressure displayed are normal fluctuates continually so s
uccessive measurements will show some variability.
Blood pressure 1.Record the daily development of the measured differ from values and consult your doctor.
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 26
Page 27
Further Information
The level of blood-pressure is subject to fluctuations even with healthy people. Important thereby is, that comparable measure-
ments always require the same conditions (Quiet conditions)!
If, in spite of observing all these factors, the fluctuations are larger than 15mmHg, and/or you hear irregular pulse tones on several occa­sions, please consult your doctor.
For licensing, the device has been subjected to strict clinical tests, by which the computer programme used to measure the blood-pressure values was tested by experienced specialist doctors in Germany. The same computer programme is used in every individual device, and has thus also been clinically tested. The manufacture of the devices takes place according to the terms of the European standard for blood-pressure measuring devices (see technical data) under the supervision of the Technical Monitoring Association Essen (RWTüV-Essen).
You must consult your specialist dealer or chemist if there are tech­nical problems with the blood-pressure instrument. Never attempt to repair the instrument yourself! Any unauthorised opening of the instrument invalidates all guaran­tee claims!
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 27
Page 28
8. Care and maintenance, recalivration
a) Do not expose the device to either
extreme temperatures, humidity, dust or direct sunlight.
b) The cuff contains a sensitive
air-tight bubble. Handle this carefully and avoid all types of straining through twisting or buckling.
c) Clean the device with a soft,
dry cloth. Do not use petrol, thinners or similar solvent. Spots on the cuff can be re­moved carefully with a damp cloth and soapsuds. The cuff must not be washed!
d) Do not drop the instrument or
treat it roughly in any way.
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 28
Page 29
Periodical recalibration
Sensitive measuring devices must from time to time be checked for accuracy. We therefore recommend a periodical inspection of the static pressure display every 2 years. Your specialist dealer would be pleased to provide more extensive information about this.
9. Guarantee
The blood-pressure monitor BP 3BU1-5 is guaranteed for 2 years from date of purchase. This guarantee includes the instrument and the cuff. The guarantee does not apply to damage caused by im­proper handling, accidents, not following the operating instructions or alterations made to the instrument by third parties. The guarantee is only valid upon presentation of the guarantee card filled out by the dealer.
Name and company address of the responsible dealer:
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 29
Page 30
10. Reference to standards Device standard: Device corresponds to the requirements of
the European standard for non-invasive blood- pressure monitor EN1060-1 / 12:95 EN1060-3 / 09:97
DIN 58130, NIBP – clinical investigation ANSI / AAMI SP10, NIBP – requirements
Electromagn. compatibility: Device fulfils the stipulations of the
European standard EN 60601-1-2
Clinical testing: The clinical performance test was carried
out in Germany according to the DIN 58130 / 1997 procedure N6 (sequential).
The stipulations of the EU-Guidelines 93/42/EWG for Medical Products Class IIa have been fulfilled.
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 30
Page 31
11. Technical specifications Weight: 148g (with batteries + cuff)
Size: 85 x 77 x 75mm (including cuff) Storage temperature: –5 to +50°C Humidity: 15 to 85% relative humidity
maximum
Operation temperature: 10 to 40°C Display: LCD-Display
(Liquid Crystal Display)
Measuring method: oscillometric Pressure sensor: capacitive Measuring range:
SYS/DIA: 30 to 280 mmHg
Pulse: 40 to 200 per minute Cuff pressure display range: 0-299 mmHg Memory: toring the last 2 x 30
measurements automatically
Measuring resolution: 1 mmHg Accuracy: Pressure within ± 3 mmHg
Pulse ± 5 % of the reading
Power source: 2 dry cells (Batteries) UM-4,
size AAA 1.5V
Accessories: storage case
IB BP 3BU1-5 VarB 0702.qxd 13.08.2002 8:41 Uhr Seite 31