Microlife AG1-30 User Manual [ru]

Read the instructions carefully before using this device.
шВВ‰ ‚ЛНУЛТЪ‡ММˇП ФЛО‡‰Ы Ы‚‡КМУ ФУ˜ЛЪ‡ИЪВ ˆ˛ iМТЪЫНˆi˛. шВВ‰ ЛТФУО¸БУ‚‡МЛВП ФЛ·У‡ ‚МЛП‡ЪВО¸МУ ФУ˜ЪЛЪВ ‰‡ММУВ ЫНУ‚У‰ТЪ‚У.
Europe / Middle-East / Africa
Microlife AG Espenstrasse 139 9443 Widnau / Switzerland Tel. +41 / 71 727 70 30 Fax +41 / 71 727 70 39 Email admin@microlife.ch www.microlife.com
Asia
Microlife Corporation. 9F, 431, RuiGang Road, NeiHu Taipei, 11492, Taiwan, R.O.C. Tel. 886 2 8797-1288 Fax.886 2 8797-1283 Email service@microlife.com.tw www.microlife.com
North / Central / South America
Microlife USA, Inc. 1617 Gulf to Bay Blvd., 2nd Floor Ste A Clearwater, FL 33755 / USA Tel. +1 727 442 5353 Fax +1 727 442 5377 Email msa@microlifeusa.com www.microlife.com
IB AG1-30 V3-UA 2212
Microlife BP AG1-30
EN
Aneroid Blood Pressure Kit
Instruction Manual (1-13)
RU
Механический прибор для измерения артериального давления
Pуководство по пользованию (14-27)
UA
Прилад для вимірювання артеріального тиску
Інструкція для користувачів (28-41)
Aneroid Blood Pressure Kit
Instruction Manual
Table of contents
1. Introduction
1.1. Features
1.2. Important information about self-measurement
2. Important information on the subject of blood pressure and its measurement
2.1. How does high/low blood pressure arise?
2.2. Which values are normal?
2.3. What can be done, if regular high/low values are obtained?
3. The various components of the blood pressure kit
4. Carrying out a measurement
4.1. Before the measurement
4.2. Common sources of error
4.3. Fitting the cuff
4.4. Measuring procedure
4.4.1. Connect stethoscope tubing to the chestpiece
4.4.2. Inflating the cuff
4.4.3. Systolic blood pressure reading
4.4.4. Diastolic blood pressure reading
4.4.5. Recording your readings
5. Malfunction / Troubleshooting
6. Care and maintenance, recalibration
7. Guarantee
8. Reference to standards
9. www.microlife.com
10. Technical specifications
EN
1
1. Introduction
1.1. Features
The aneroid blood pressure kit is a non-automated, mechanical blood pressure measuring device for use on the upper arm. It offers proven reliability and superior performance at an economical price. With the advanced non-stop pin mechanism and ergonomic bulb with complete valves, the BP AG1-30 ensure you a precise and consistent measurement. With stethoscope attached to the D-ring cuff, BP AG1-30 can be easily operated. Nevertheless, its durable nylon cuff, high-grade bearing and aneroid bellow provide consistent operation. The entire unit stores in a zippered bag for easy portability.
Before using, please read through this instruction manual carefully and then keep it in a safe place. For further questions on the subject of blood pressure and its measurement, please contact your doctor.
Attention!
1.2. Important information about self-measurement
• Donotforget:self-measurement means control, not diagnosis or treatment. Unusual values must always be discussed with your doctor. Under no circumstances should you alter the dosages of any drugs prescribed by your doctor.
2. Important information on the subject of blood pressure and its measurement
2.1. How does high/low blood pressure arise?
The level of blood pressure is determined in a part of the brain, the so-called circulatory centre, and adapted to the respective situation by way of feedback via the nervous system. To adjust the blood pressure, the strength and frequency of the heart (Pulse), as well as the width of circulatory blood vessels is altered. The latter is effected by way of fine muscles in the blood-vessel walls.
Thelevelofarterialbloodpressurechangesperiodicallyduringtheheartactivity:Duringthe
«blood ejection» (Systole) the value is maximal (systolic blood pressure value), at the end of 2
the heart’s «rest period» (Diastole) minimal (diastolic blood pressure value). The blood pressure values must lie within certain normal ranges in order to prevent particular diseases.
2.2 Which values are normal?
Blood pressure is too high if at rest, the diastolic pressure is above 90 mmHg and/or the systolic blood pressure is over 140 mmHg. In this case, please consult your doctor immediately. Long­term values at this level endanger your health due to the associated advancing damage to the blood vessels in your body.
With blood pressure values that are too low, i.e. systolic values under 100 mmHg and/or diastolic values under 60 mmHg, likewise, please consult your doctor.
Even with normal blood pressure values, a regular self-check with your blood pressure monitor is re­commended. In this way you can detect possible changes in your values early and react appropriately.
If you are undergoing medical treatment to control your blood pressure, please keep a record of the level of your blood pressure by carrying out regular self-measurements at specific times of the day. Show these values to your doctor. Never use the results of your measurements to
alter independently the drug doses prescribed by your doctor.
Tableforclassifyingbloodpressurevalues(unitsmmHg)accordingtoWorldHealthOrganization:
Range Systolic Diastolic Recommendation
blood pressure too low
1. blood pressure optimum
2. blood pressure normal
3. blood pressure slightly high
4. blood pressure too high
5. blood pressure far too high
6. blood pressure dangerously high
< 100 < 60 100 - 120 60 - 80 120 - 130 80 - 85 130 - 140 85 - 90 140 - 160 90 - 100 160 - 180 100 - 110
180 110
Consult your doctor Self-check Self-check Consult your doctor Seek medical advice Seek medical advice Urgently seek medical advice!
3
Further information
• Ifyourvaluesaremostlystandardunderrestingconditionsbutexceptionallyhighundercon
ditions of physical or psychological stress, it is possible that you are suffering from so-called «labile hypertension». Please consult your doctor if you suspect that this might be the case.
• Correctlymeasureddiastolicbloodpressurevaluesabove120mmHgrequireimmediate medical treatment.
2.3. What can be done, if regular increased/low values are obtained?
a) Please consult your doctor. b) Increased blood pressure values (various forms of hypertension) are associated long- and
medium term with considerable risks to health. This concerns the arterial blood vessels of your body, which are endangered due to constriction caused by deposits in the vessel walls (Arteriosclerosis). A deficient supply of blood to important organs (heart, brain, muscles) can be the result. Furthermore, with long-term continuously increased blood pressure values, the heart will become structurally damaged.
c) There are many different causes of the appearance of high blood pressure. We differentiate
between the common primary (essential) hypertension, and secondary hypertension. The latter group can be ascribed to specific organic malfunctions. Please consult your doctor for information about the possible origins of your own increased blood pressure values.
d) There are measures which you can take, not only for reducing a medically established high
bloodpressure,butalsoforprevention.Thesemeasuresarepartofyourgeneralwayoflife:
A) Eating habits
• Striveforanormalweightcorrespondingtoyourage.Reduceoverweight!
• Avoidexcessiveconsumptionofcommonsalt.
• Avoidfattyfoods.
B) Previous illnesses
Followconsistentlyanymedicalinstructionsfortreatingpreviousillnesssuchas:
• Diabetes(Diabetesmellitus)
• Fatmetabolismdisorder
• Gout
4
C) Habits
• Giveupsmokingcompletely
• Drinkonlymoderateamountsofalcohol
• Restrictyourcaffeineconsumption(Coffee)
D) Physical constitution
• Afterapreliminarymedicalexamination,doregularsport.
• Choosesportswhichrequirestaminaandavoidthosewhichrequirestrength.
• Avoidreachingthelimitofyourperformance.
• Withpreviousillnessesand/oranageofover40years,pleaseconsultyourdoctorbefore
beginning your sporting activities. He will advise you regarding the type and extent of types of sport that are possible for you.
3. The various components of the Aneroid Blood Pressure Kit
TheillustrationshowstheBPAG130,consistingof:
a) Measuringunit:
Manometer
Soft bag
Ear piece
Stethoscope
b) Cuff:
Type AC-1M, for arm circumference 22-32 cm or Type AC-1L, for arm circumference 32-42 cm (available as special accessory)
Bulb & Valves
Cuff
5
4. Carrying out a measurement
4.1. Before the measurement
• Avoideating,smokingaswellasallformsofexertiondirectlybeforethemeasurement.All
these factors influence the measurement result. Try and find time to relax by sitting in an armchair in a quite atmosphere for about ten minutes before the measurement.
• Removeanygarmentthatfitscloselytoyourupperarm.
• Measurealwaysonthesamewrist(normallyleft).
• Attempttocarryoutthemeasurementsregularlyatthesametimeofday,sincetheblood
pressure changes during the course of the day.
4.2. Common sources of error
Note: Comparable blood pressure measurements always require the same conditions! These are
normally always quiet conditions.
• Alleffortsbythepatienttosupportthearmcanincreasethebloodpressure.Makesureyou
are in a comfortable, relaxed position and do not activate any of the muscles in the measu­rement arm during the measurement. Use a cushion for support if necessary.
• Ifthearmarteryliesconsiderablylower(higher)thantheheart,anerroneouslyhigher(lower)
blood pressure will be measured! (Each 15 cm difference in height results in a measurement error of 10 mmHg!)
• Cuffsthataretoonarrowortooshortresultinfalsemeasurementvalues.Selectingthecorrect
cuff is of extraordinary importance. The cuff size is dependent upon the circumference of the arm (measured in the centre). The permissible range is printed on the cuff. If this is not suitable for your use, please contact your dealer. Note: Only use clinically approved Original-Cuffs!
• Aloosecufforasidewaysprotrudingairpocketcausesfalsemeasurementvalues.
4.3. Fitting the cuff
a) Push the cuff over the left upper arm so that the tube
points in the direction of the lower arm.
6
b) Lay the cuff on the arm as illustrated. Make certain
that the lower edge of the cuff lies approximately 2 to 3 cm above the elbow and that the tube leaves the cuff on the inner side of the arm.
Important! The mark (ca. 3 cm long bar) must lie
exactly over the artery which runs down the inner side of the arm.
c) Tighten the free end of the cuff and close the cuff
with the closer.
d) There must be no free space between the arm and
the cuff as this would influence the result. Clothing must not restrict the arm. Any piece of clothing which does (e.g. a pullover) must be taken off.
e) Secure the cuff with the closer in such a way that it
lies comfortably and is not too tight. Lay the arm on the table (palm upwards) so that the cuff is at the same height as the heart. Make sure that the tube is not kinked.
f) Remain seated quietly for two minutes before you
begin the measurement.
Comment:
If it is not possible to fit the cuff to the left arm, it can also be placed on the right one. However all measurements should be made using the same arm.
2–3 cm
tube
7
4.4. Measuring procedure
4.4.1. Connect stethoscope tubing to the chestpiece
After the cuff is properly worn on the arm, connect stethoscope tubing of to the chestpiece attached on the cuff. Make sure the chestpiece is in contact with skin and above the brachial artery. Wear the binaural (earpieces) properly to check the Korotkoff’s sound during measurement. Before using the stethoscope, be sure there is no crack on the diaphragm, earpieces, and tubing. Any improper setup or damage of the stethoscope will cause distorted sound or poor sound transmission to make inaccurate reading.
4.4.2. Inflating the cuff
Close the air valve on the bulb by turning the screw clock wise. Do not over-tighten. Squeeze the inflation bulb with the hand at a steady rate until the pointer on the gauge is 30 mmHg above your normal systolic pressure value. If you are not sure the value, inflate to 200 mmHg first.
4.4.3. Systolic blood pressure reading
Slowly open air valve by turning screw counter clockwise and hold stethoscope chestpiece over brachial artery. Proper de­flation rate is essential for an accurate reading, so you should practice and master a recommended defla tion rate of 2-3mmHg per second or a drop of 1-2 marks on the pressure gauge each heartbeat. You should not keep the cuff inflated any longer than necessary. As the cuff begins to deflate, you must listen care­fully with the stethoscope. Note the reading on the gauge as soon as you hear a faint, rhythmic tapping or thumping sounds. This is the systolic blood pressure reading. Listen care fully and familiarize yourself with pulse (Korotkoff’s) sound.
8
4.4.4. Diastolic blood pressure reading
Allow the pressure to continue dropping at the same deflation rate. When your diastolic blood pressure value reached, the thumping sound stops. Deflate the cuff valve completely. Remove the cuff from arm and stethoscope from ears.
4.4.5. Record your readings
Repeat the measurement at least two times. Do not forget to record your readings and the time of the day measurement is made immediately after you finish measuring. A suitable time is first thing in the morning, or just before evening meals. Remember that your physician is the only person qualified to analyze your blood pressure.
Further information Measurements should not occur soon after each other, since otherwise, the results will be falsified. Wait therefore for several minutes in a relaxed position, sitting or lying, before you repeat a measurement.
5. Other possible malfunctions and their elimination
If problems occur when using the device, the following points should be checked and if
necessary,thecorrespondingmeasuresaretobetaken:
Malfunction
The sound transmission is poor, distorted or there is extraneous noise.
Remedy
1. Check the earpieces if they are plugged or cracked. If not, make sure they do not fit poorly as worn.
2. Check the tube if it is broken or twisted.
3. Check the bell and diaphragm of chestpiece if there is any crack.
4. Make sure the chestpiece is in proper contact with skin and over brachial artery during measuring.
Clean or replace any defective parts if found to
avoid inaccurate reading.
9
The pressure does not rise although the bulb is pumping.
The deflation rate can not be set to 2-3 mmHg/ sec. by adjusting the air release valve.
Pointer is not at 0 +/- 3 mmHg at rest.
FURTHER INFORMATION The level of blood pressure is subject to fluctuations even with healthy people. Important thereby is, that comparable measurements always require the same conditions (rest condition)! If, in spite of observing all these factors, the fluctuations are more than 15mmHg, and/or you hear irregular pulses on several occasions, please consult your doctor.
You must consult your specialist dealer or chemist if there are technical problems with the blood pressure instrument. Never attempt to repair the instrument yourself! Any unauthorised opening of the instrument invalidates all guarantee claims!
6. Care and maintenance, recalibration
With proper care and maintenance, this blood pressure measuring device will provide years of
satisfactoryservice.FollowthegeneralRulesbelow:
•Donotdrop.
•Neverinflatebeyond300mmHg.
•Donotexposethedevicetoeitherextremetemparatures,humidity,ordirectsunlight.
•Nevercontactthecufffabricwithasharpinstrument,sincethiscouldcausedamage.
10
1. Make sure that the valve is closed.
2. Make sure the cuff is properly connected to bulb and manometer
3. Check if the cuff, tube and bulb is leaky. Replace the defective parts if any
•Disassemblethevalvefrombulbtocheckifthere
is any blockage in the airway of the valve. Clean the blockage and try again. If it still does not work, replace it to avoid inaccurate reading.
1. Make sure that the valve is open for zero check.
2. If still more than 3mmHg deviation, contact your dealer to recalibrate the manometer.
•Alwaysdeflatecuffcompletelybeforestorage.
•Donotdismantlemanometerunderanycircumstance.
•Storethewholedeviceinstoragebagprovided,tokeepallthepartsclean.
•Storagetemperaturecondition:20°C to +70 °C at a relative air humidity of 85 % (non-
condensing).
•Wipeoffthemanometerandbulbwithadampcloth.Sterilizationisnotnecessary,sincethe
parts of manometer should not come into direct contact with the patient‘s body during measurement.
•Removethebladderfirst,andwipethecloser,bladderandtubeswithadampcloth.Thecuffcan
be washed with soap and cold water. But do rinse the cuff with clear water and keep it air dry.
Periodical recalibration
Sensitive measuring devices must from time to time be checked for accuracy. We therefore recommend a periodical inspection of the static pressure display every year. Your specialist dealer would be pleased to provide more extensive information about this.
7. Guarantee
This blood pressure monitor is guaranteed for 2 years from date of purchase. This guarantee includes the instrument and the cuff. The guarantee does not apply to damage caused by improper handling, accidents, not following the operating instructions or alterations made to the instrument by third parties. The guarantee is only valid upon presentation of the guarantee card filled out by the dealer.
Nameandcompanyaddressoftheresponsibledealer:
11
Loading...
+ 30 hidden pages