ROUTINE CLEANING/DECONTAMINATION
Dried blood, saline, and other deposits inside the handpiece are a major cause of equipment malfunction.
Proper cleaning and inspection prior to sterilization will avoid delays during the surgical procedure.
CAUTION: The handpiece is sensitive to moisture. DO NOT immerse the handpiece in saline,
disinfectant, or any other liquid. DO NOT use an ultrasonic cleaner or any other automated cleaning equipment, such as washer/decontaminators or washer/sterilizers. Ultrasonic cleaning, in particular, can damage
the bearings in the handpieces, potentially resulting in overheating or seizure.
1. Disassemble and transport to the decontamination area.
a. If using the throttle lever handpiece, slide the safety lock to theL
position. If using the throttleless handpiece, unplug the foot control.
b. Remove and discard used saw blades, burs, and drills after use,
handling them as any contaminated sharp instrument is handled.
Reuse of the saw blades and burs is not recommended.
2. Clean the device immediately, using an appropriate detergent solution.
a. Immediate decontamination protects personnel and prevents transmission of pathogens.
b. Make sure the air hose is connected to the handpiece to keep detergent solution and water
out of the motor.
c. Clean the air hose thoroughly with warm water, mild detergent, and a soft brush. Pay
vigorous attention to crevices.
3. Use a cannulation brush on cannulated handpieces.
a. Clean the cannulated shaft of wire drivers with the small cylindrical
wire driver cannulation brush (9600-063).
b. Remove bur guard from drills.
c. Clean inside drill collets and bur guards with the drill cannulation
brush (9600-064).
4. Flush the noses of drills and wire drivers with a Water-Pik. Flushing removes blood, debris, and saline
deposits.
5. Rinse air hose. With the air hose connected, rinse components under running water to remove all traces
of detergent solution. If possible, use distilled water for the final rinse.
6. Dry. Disconnect the air hose from the handpiece. Dry all the components with a lint-free towel.
7. Lubrication of head modules. DO NOT lubricate or oil the handpieces. Lubrication will clog the motor
and prevent it from starting. Also take special precautions to avoid the use of cleaners which contain
lubrication.
8. Accidental immersion. If a handpiece is accidentally immersed in saline, disinfectant, cleaning fluid, or
any other corrosive substance, take the following steps to save the handpiece.
a. Totally immerse the handpiece in distilled water for 1 minute to dilute the corrosive fluid.
DO NOT allow the water to dry in the handpiece.
b.Immediately after soaking, steam sterilize in a prevacuum sterilizer at 270°F(132°C) for
4 minutes followed by an 8 minute drying time. Sterilizing will dry out the handpiece,
avoid rusting, and prevent contamination from collecting in the motor.
24
L
MicroAire Instructional Manual