1) Introduction
The SlopeView TLL-90 is a low cost and high accurate device for measuring surface
and single axis inclinations. It is based on the revolutionary technology called MicroElectro-Mechanical Systems (MEMS), in conjunction with sophisticated software
techniques to accomplish a guaranteed accuracy of better than ±0.05º and with
measurement resolution of 0.01º in the favorable measurement range.
SlopeView TLL-90 provides not only ultra high precision measurement of angle but
also bundles with a Class2 tile laser for extending the angle projection and provides
alignment aid for the user’s work. Together with other distinguished features such as
dot matrix LCD, beeper alert and precision settings, SlopeView TLL-90 is the best
tool to make your work better and easier.
Features:
1) High accuracy ± (0.05° + 1%) , high resolution 0.01°
2) Laser Alignment
3) Dual and Single axis with user friendly LCD display angle
4) Rechargeable
5) V Shape metal case for easy to fit at the corner or pipe.
6) Magnetic base
7) Audible alarming at settable angle range
MIB Instruments Co. ,Ltd
3

2) Specifications
Accuracy: ± (0.05° + 1%)
Measuring range: Single axis: 360°, Dual axis: ±40°
Resolution: 0.01°
Response time: <0.4 second
Laser Beam: Class 2, 635nm red laser, < 1mW
Audio sound: 60dB @ 30cm
Operating temperature: 0 to 50°C
Storage temperature: -10 to 60°C
User Interface: Mono-color LCD with backlight
Supply Power: Rechargeable Li-Ion Battery, 3.7Volt.
Charger port: 5V 500mA Mini type-B USB port
Power Consumption: Standby: 200uA, Operation: 20mA.
Standby Battery Life: 2000 hours
Operating Battery Life: 40 hours
Dimensions (in mm): 93(L) x 66(W) x 28(H)
Magnetic Base: affix at the bottom and back.
MIB Instruments Co. ,Ltd
4

3) Button Functions
Function Descriptions
Button
Normal Mode Set Mode
In normal operation, this button turns the
device ON/OFF.
Press and hold for 3 seconds to enable or
disable backlight
During set function,
this button serves as
an escape key for
each selection menu.
When pressed, the curren t reading is set to
zero; subsequent measurem ents ar e
relative to this reading. The LCD will show
the icon to indicate the device is in
zero mode.
Press and hold for 3 seconds to enable or
disable sound. The icon on LCD will
be displaced accordingly. The buzzer beep
could be set at different accuracy level.
Refer to section “Angle Alarming”.
Serves as the upward
key for option
selection.
MIB Instruments Co. ,Ltd
5

When pressed , the current value will be
freezed, the unit icon will be flash to
indicate the reading is on hold.
Press and hold for 3 seconds to change
the measurement unit to either degree/mm
per meter. The measurement unit will be
displayed on the LCD accordingly.
Serves as the
downward key for
option selection.
La
Se/
ser
t
Press and hold for 3 seconds to enter Set
mode, for set mode options.
Press to turn laser ON/OFF. The laser line
is aligned in parallel with the Top and
Bottom side edges, and it is in
perpendicular with the Back side. To switch
on/off the laser beam, simply press the
Laser Button again.
Serves as Set key for
selected option.
MIB Instruments Co. ,Ltd
6

LCD Icons Representations
Battery status indication icons
These icons indicate the battery level. There are 3
levels representing empty, half and full. When the
In dual-axis mode, the LCD displays the direction
of tilt graphically. The unit of scale is displayed
below the graph with different scales. The scale is
adjusted automatically depending on the
measured angle. These icons show the available
scales.
These 2 icons show the unit of the measurement.
User can switch the measured unit to either
mm/m (mm per meter) or in degree mode. The
mm/m unit indicates how many mm/m to level. It
is calculate by:
mm/m = sin (Angle) * 1000 mm
MIB Instruments Co. ,Ltd
7

Sound notification on.
Blank as off
Show this logo as relative value is showing. When
the Zero button is pressed, the unit reset current
angle to zero.
Direction of tilt icons, show the tilt angle direction
with respect to the unit bottom right hand corner.
Dual-Axis Mode. Both the X and Y axis angle will
be showed. Dual axis mode measures inclination
upto +-30 degree for each axis before it
automatically switch to single axis mode.
MIB Instruments Co. ,Ltd
8

Single Axis Mode. Measure slope upto +-90
degree.
The triangle icon indicates the direction of tilt with
respect to the bottom right corner of the unit.
MIB Instruments Co. ,Ltd
9

4) Battery Charging
The SlopeView TLL-90 has a built in Lithium Ion rechargeable battery. A standard
charger is supplied that the input voltage is 110V to 240V AC, 50/60Hz, and the
output is 5V DC, 500mA. The charger operating procedure is list below:
1) Plug the Charger into AC socket, the RED on the charger should turned ON,
2) Plug the USB charging cable to the Charger,
3) Insert the other end of the USB cable to the unit,
4) The battery icon on the LCD blinks to indicate charging in process. Upon
charging complete, the icon stops blinking.
5) The charging time is approx. 3 hours.
The unit could also be charged by connecting the USB cable to the unit and a
computer’s USB port. This has the same effect when charging the unit with the
provided adaptor.
Note: When the unit is turned OFF, and plug in the USB charge cable, the LCD will
no show anything, it is NORMAL. Once the unit is turned ON, the battery icon should
flash indicating the unit is in charging mode.
MIB Instruments Co. ,Ltd
10

5) Degree or mm Measurement Mode
Degree Mode:
LCD Icon:
Degree mode 360° for one circle, it will display the tilt angle relative gravity leveling.
mm/m Mode:
1 meter
LCD Icon:
Millimeter per meter mode. Place the tester in
a tilt plane.
LCD display Y (mm) for the plane is 1 M length.
For example, if you are setting a table to level.
- Table surface is 1.5M length
- Set to MM mode; place the device on the
table.
- The reading is 5mm/m on LCD
- That means Y= 5X1.5= 7.5mm difference
from leveling.
Degree and mm/m Mode Switching:
Press and hold the “HOLD” key for 3 seconds to switch between the Degree and mm
mode.
1.5M
MIB Instruments Co. ,Ltd
11

6) Relative/ Absolute Measurement
Relative Measurement
LCD Icon:
Absolute Measurement
LCD Icon: Blank
Relative and Absolute Mode Switching:
1 Press and release the “” key to set the relative measurement zero point
2 Press and release the “” key to cancel the relative zero point and back
to absolute measurem ent mo de.
7) Hold Function
Holding Mode:
LCD Icon blinking:
Holding function:
1. Press and release the “HOLD” key to activate holding function, digit will
freeze.
2. Press and release the “HOLD” key to cancel the holding function.
MIB Instruments Co. ,Ltd
12

8) Auto Power Off
For no movement in 30 minutes, the unit will power off.
Or we can set to never sleep mode at below instruction.
Power auto off setting:
1. Press and hold “Laser/Set” key and enter MENU mode
2. Select “POWER” by “Zero” and “Hold” key, press “Laser/Set” key to enter
Power mode
3. Scroll “NEVER” or “30MIN” by “Zero” and “Hold” key
4. Press “Laser/Set” key to confirm NEVER or 30MIN (30 minutes) sleep
9) Restore Factory Setting
When you find that the u nit is abnormal, you can restore the unit to factory setti ng.
All calibration setting will be restored to factory setting.
*Not suggest restore to factory setting in normal status, if for accuracy drift, please
follow Calibration. After factory set, please redo calibration to ensure the accuracy.
Restore to factory setting:
1. Press and hold “Laser/Set” key and enter MENU mode
2. Select “FACTORY SET” by “Zero” and “Hold” key, press “Laser/Set” key to
enter FACTORY SET mode
3. Scroll “YES” or “NO” by “Zero” and “Hold” key
4. Press “Laser/Set” key to confirm
MIB Instruments Co. ,Ltd
13

10) Angle Alarming
Alarming Mode:
LCD Icon:
Alarming Angle setting:
1. Press and hold “Laser/Set” key and enter
MENU mode
BUZZER
>SINGLE : 0 0 . 0 0
DUAL. X: 0 0 . 0 0
DUAL. Y: 0 0 . 0 0
RANGE : 0 0 . 5 0
2. Select “BUZZER” by “Zero” and “Hold” key,
press “Laser/Set” key to enter BUZZER setting
Press and hold “Up” and “Down” key for fast
scrolling the digit.
3. Press “Laser/Set” to enter setting value
SINGLE Vertical / Single axis mode alarming angle (Degree)
DUAL.X Horizontal/ Dual axis mode X axis alarming angle (Degree)
DUAL.Y Horizontal/ Dual axis mode Y axis alarming angle (Degree)
RANGE The range (Degree)that will trigger the audible alarming
For example: SINGLE. set to 20.00
RANGE set to 01.00
While the unit is in +19 ْ to +21 ْ or -19 ْ to -21 ْ, the unit will
alarm
MIB Instruments Co. ,Ltd
14

11) Laser Alignment
Laser ON/OFF:
Press and release “Laser/Set” key to switch the tile laser on or off.
Draw a laser line on wall in different angle:
1) Install the metal plate with tripod and place near
the wall
2) Tune the metal plane at vertical angle.
3) Place the TLL90 on the metal plate, the ma gnetic
base will stick the TLL90 on the metal plate.
4) Laser on; tune the position of the TLL90 to set to
your specific angle.
Draw a vertical line or horizontal line without
tripod:
1) Place the metal plate on a table
2) Place TLL90 on the metal plate with dual axis
mode
3) Tune the 2 screw and let the metal plane become
level
4) You can then laser on the TLL90 and it will project
a vertical or horizontal line to the wall within 5m.
MIB Instruments Co. ,Ltd
15

12) Calibration
Calibrate the unit, once you found that there is accuracy drift on the unit. You can
verify the accuracy at below step:
At Step1, you measured X and Y value, X1 and Y1
At Step2, you measured X2 and Y2, in theory X1=-X2, Y1=-Y2.
If the error is too large, you can enter calibration mode to eliminate the error
Accuracy drift is causing by large ambient temperature change (5 to 10 Degree
Celsius) or the unit has been dropped.
MIB Instruments Co. ,Ltd
17

Calibration Procedure:
Step 1: Press and hold the “Laser/Set” key enter Menu mode. Select “Calibration”
mode, press “Set”. Place the unit on a flat table (no need perfect level table)
LCD display “CALIBRATE DUAL AXIS PRESS SET”, press set and buzzer will beep;
wait until the beep sound stop. While the buzzer is beeping keep the unit stable
Step 2: Then rotate the unit 180 degree with the other side against the at the same
place. Press the Set button again, and wait for the beep finished.
A
A
Step 1 Step 2
CALIBRATE
DUAL AXIS
ROTATE 180
DEGREE,
PRESS SET
MIB Instruments Co. ,Ltd
18

Step 3: LCD display “CALIBRATE HORIZONTAL PRESS SET”. Place the unit
horizontal like the picture “STEP 3” and then press Set, wait until the beep sound
stop.
Step 4: Then rotate the unit 180 degree at the same place. Press the Set button
again, and wait for the beep finished.
A
A
Step 3
Step 4
CALIBRATE
HORIZONTAL
ROTATE 180
DEGREE,
PRESS SET
MIB Instruments Co. ,Ltd
19

Step 5: LCD display “CALIBRATE VERTICAL PRESS SET”, mention that the LCD in
upper side, hold on a flat wall. And then press “Laser/Set” Key. Wait for the “Beep”
sound stop.
Step 6: Then rotate the unit 180 degree with the other side against the at the same
place wall (LCD in upper side). Press the Set button again, and wait for the beep
finished. Now, the LCD should go back to the selection menu. The calibration is done,
by selecting “Back” to go back for normal operation
ROTATE 180
DEGREE,
PRESS SET
CALIBRATE
VERTICAL
A
A
Step 5
Step 6
MIB Instruments Co. ,Ltd
20

13) PC Communication
Slopeview TLL-90 has a PC data logging function.
You can not directly plug the USB cable between PC to TLL-90 by a general USB
cable.
You need to optional purchase the PC adapter (model number: SV-RS232)
You can continue to data logging TLL-90 X and Y inclination data to PC.
Sampling time is around 2 Hz.
Specification:
1) RS232 Com Port 9600 baud rate
2) USB connection (SV-RS232 include a RS232 to USB adapter
Detail can visit our website or contact our sales person.
MIB Instruments Co. ,Ltd
21

15) Warranty
The SlopeView TLL-90 comes with 12 months warranty from the data of purchase.
The warranty ensures the SlopeView TLL-90 is free of defects in materials and
workmanship during manufacturing. In case of malfunction or defects due to
materials or workmanship, MIB Instruments Co. will repair or replace the device
provided that the original purchaser meets the following guidelines:
1) Notify the dealer from whom the unit was purchased about the failure,
2) Return the unit to the dealer,
3) Provide original receipt of purchase,
This warranty does not cover damage resulting from accident, misuse or abuse,
tampering, disassembled or attempted by unauthorized agencies. This also includes
units that have been modified in any fashion, or units that have not been maintained
in accordance with the instructions in this manual.
END.
MIB Instruments Co. ,Ltd
23