MFJ MFJ-269 Calibration Procedure Guide

2. - MFJ 269 TEST & CALIBRATION PROCEDURE
g
g
2.1. Introduction
The box can be tested completely after assembling with the battery holder initially unassembled. An initial PCB pre calibration procedure is assumed. The operations indicated in the procedure can be followed under normal conditions (getting oscillations impedance and SWR meter working), otherwise some troubleshooting may be required. A multimeter
2.2 Procedure
- Check appearance of box, if any defects inform to the assembly persons.
- Set the unit in test mode by pressing the mode and the gate time switches buttons in the order previously indicated.
a spectrum analyzer, a scope (optional), a calibrated frequency counter, a
2.2.1 Checkin
2.2.2 Band Overlappin
- Set the TEST mode (read 1.I if it is necessary to remind the procedure).
- Set the 4th TEST mode step where the internal oscillator frequency will be displayed and
The unit at the HF/VHF mode has to cover continuously from 170 to 1.8 MHz. Check that the band overlapping realized in the PCB precalibration is still correct.
the RF amplifier bias and the AGC
"
"
- Using the variable capacitor, slowly cover the band looking for a consistent frequency reading in the counter module. Notify, if erratic behavior is observed.
- Repeat the previous operation to cover all the bands. Readjust the setting where necessary.
2.2.3 Voltage calibration
.
p
It may me necessary to repeat the previous interactively to get the desired settings.
Adjust the trimmer R67 to read 50 ohms on the 7 needle meter. .,
-
- Set the unit around 430 Hz and verif an SWR reading close to one and a zero ohm im
edance needle meter reading. 145 MHz
2.24 Inductance and capacitance modes
- Set the unit to a frequency higher than 60 MHz and verify a close to
p
F reading.
4
-Press and hold the mode and the gate ti me switch b uttons. As soon the Advanced 2 message is di splayed, release b oth switches.
Loading...
+ 4 hidden pages