Mettler Toledo XPR15002L/A, XPR16001L/A, XPR15002L, XPR20002LDR, XPR20002LDR/A Data Sheet

...
Excellence Level
XPR Large Precision Balances
Accurate, Flexible, Compliant
Outstanding Performance
The new Monobloc™ weighing cell delivers fast and precise results. The SmartPan weighing pan, on the XPR15002L and XPR20002LDR, minimizes air drafts, providing exceptional stability in busy areas. The LevelMatic weighing pan offers the highest weigh ing accuracy.
Extremely Durable
Integrated overload protection prevents damage to the weighing cell during the loading and unloading of heavy items. The full metal housing is chemically resistant, and protects against dust and splashing liquids to ensure a long balance lifetime.
Streamlined Processes
Intelligent features simplify and support your processes at every step. The integrated method library saves time and provides higher process security. The built-in notepad automatically records all task parameters and results, eliminating manual transcription.
Efcient Data Handling
No more manual transcription! Transfer of results and task parameters to a PC or software program is easy via USB or Ethernet. Peripheral devices, such as printers and barcode readers, simply plug in and you’re ready to go.
XPR Precision Balance Solutions
Go Beyond Weighing
The robust XPR large platform precision balances deliver outstanding weighing performance and support the highest requirements for data integrity and compliance.
These flexible balances offer wide-ranging solutions to meet the demands of your applications. E.g. high capacity balances with readability down to 1 mg enable you to weigh big and small quantities directly into the same container.
Customizable, integrated weighing methods guide you through each weighing process and all task parameters are automatically recorded on the built-in results notepad. With intelligent features automatically monitoring the status of your balance, you can rest assured that your results are valid.
SmartPan models
SmartPan weighing pans minimize the effects of air drafts on the weighing cell significantly improving stability and measurement accuracy.
LevelMatic weighing
1
The LevelMatic weighing pan minimizes corner load effects enabling you to weigh with an extreme level of accuracy.
XPR-L Precision Balances
XPR26003LC
Material number 303176 42 3 0 317644 303176 4 3
Limit values
Maximum capacity 26.1 kg 64.1 kg 32 .1 k g Repeatability 1 mg 5 mg 5 mg Repeatability 6 mg 15 m g 15 m g Linearity deviation 20 mg 50 mg 20 mg
Typical values
Repeatability 4 mg 10 mg 6.5 mg Linearity deviation 11. 4 m g 28 mg 18 mg Sensitivity offset (test weight) 140 mg (20 kg) 650 mg (50 kg) 140 mg (20 kg) Minimum weight (5% load, k=2, U=1.0%) 800 mg 2 g 1.3 g USP minimum weight (5% load, k=2, U=0.10%) 8 g 20 g 13 g Settling time 8 ... 12 s 8 ... 12 s 8 ... 12 s
Dimensions
Weighing pan size (mm) Weighing pan LevelMatic LevelMatic Standard pan Draft shield Draft shield W64 Draft shield W64 Optional
220 220
1
XPR64003LD5C
1
XPR32003LD5C
280 x 360
Material number 30 317522 30 317526 303176 4 5
Limit values
Maximum capacity (full range/fine range) 15.1 kg 20.1 kg / 4.2 kg 64.1 kg Readability (full range/fine range) 0.01 g 0.1 g / 0.01 g 0.01 g Repeatability (full range/fine range) 15 m g 80 mg / 30 mg 35 mg Linearity deviation 20 mg 60 mg 50 mg
Typical values
Repeatability (full range/fine range) 10 mg 60 mg / 15 mg 18 mg Linearity deviation 6 mg 20 mg 28 mg Sensitivity offset (test weight) 12 mg (15 kg) 24 mg (20 kg) 650 mg (50 kg) Minimum weight (5% load, k=2, U=1.0%) 2 g 3 g 3.6 g USP minimum weight (5% load, k=2, U=0.10%) 20 g 30 g 36 g Settling time 1.5 s 1.5 s 8...12s
Dimensions
Weighing pan size (mm) 172 x 205 172 x 205 280 x 360 Weighing pan Standard pan, SmartPan Standard pan, SmartPan Standard pan Draft shield Optional Optional Optional
2
METTLER TOLEDO XPR-L Precision Balances
XP R15 002 L XPR20002LDR XPR64002LC
Loading...