Mettler Toledo XPR1202S, XPR15002L/A, XPR1203S, XPR1203S/A, XPR10002S Instruction Manual

...
Precision Balances and Comparators
User Manual
XPR models

Overview balances with S weighing platform

9
1
2
3
8
10
4
5 mg / 10 mg
55
2
6
7
9
10
8
1

Legend balances with S weighing platform

DripTray
1
SmartPan weighing pan
2
Pro Draft Shield
3
Pro Draft Shield door handler
4
Weighing pan with protective cover
5
6
7
8
9
10
Safety foot
Weighing pan support cap
Weighing platform with protective cover
Leveling foot
Terminal with protective cover
Overview balances with S weighing platform 3Precision Balances and Comparators

Overview comparators with S weighing platform

10
0.1 mg / 1 mg
1
3
8
9
11
2
4
8
6
7
10 9
11
5
7
11
2
3
6 7
5
4

Legend comparators with S weighing platform

Bottom plate
1
DripTray
2
LevelMatic weighing pan holder
3
SmartPan weighing pan
4
LevelMatic weighing pan
5
Weighing pan
6
Pro Draft Shield door handler
7
Pro Draft Shield
8
Weighing platform with protective cover
9
Leveling foot
10
Terminal with protective cover
11
Overview comparators with S weighing platform4 Precision Balances and Comparators

Overview interface board S weighing platform

Legend interface board S weighing platform

Ethernet port
1
USB device ports
2
Fixations for optional terminal stand
3
Service seal
4
USB host port
5
Socket for terminal connection cable
6
Socket for AC/DC adapter
7

Overview balances with L weighing platform

10 mg
0.1 g / 1 g
7
1
2
3
6
8
1
4
5
9789

Legend balances with L weighing platform

Weighing pan support cap
1
Adapter plate
2
Draft shield element
3
SmartPan weighing pan
4
Weighing pan with protective cover
5
6
7
8
9
Weighing pan
Weighing platform
Leveling foot
Terminal with protective cover
Overview balances with L weighing platform 5Precision Balances and Comparators

Overview comparators with L weighing platform

1 mg / 5 mg / 10 mg
8
9
1
4
8
9
7
2
7
3
5
6
11 22 4
5
6
3

Legend comparators with L weighing platform

Adapter plate with LevelMatic base plate
1
Weighing pan support cap
2
Draft shield element
3
Weighing pan
4
LevelMatic weighing pan holder
5
LevelMatic weighing pan
6
Weighing platform
7
Leveling foot
8
Terminal with protective cover
9

Overview interface board L weighing platform

Overview comparators with L weighing platform6 Precision Balances and Comparators

Legend interface board L weighing platform

Socket for AC/DC adapter
1
Socket for terminal connection cable
2
Service seal
3
USB host port
4
USB device ports
5
Ethernet port
6

Overview terminal

11
22 33
44
55

Legend terminal

No. Key Name Explanation
1
2
3
4
5
ON/OFF Switches the balance on/off.
- 7 inch capacitive TFT touchscreen
Tare Tares the balance.
Zero Zeroes the balance.
Home To return from any menu level to the home screen.
By tapping goes into standby mode. To switch the balance completely off, it must be unplugged from the power supply.
Note
Do not disconnect the balance from the power supply unless the balance is not used for an extended period of time.
The touchscreen shows information about the current weighing process.
This function is used when the weighing process involves containers. After taring the balance, the screen shows Net which indicates that all displayed weights are net.
The balance must always be zeroed before starting the weighing process. After zeroing, the balance sets a new zero point.
, the balance is not completely switched off but
Overview terminal 7Precision Balances and Comparators

1 Safety Information

Read and understand the instructions in this manual before using the balance.
Keep this manual for future reference.
Include this manual if you pass on the balance to other parties.
If the balance is not used according to the instructions in this manual or if it is modified, the safety of the balance may be impaired and Mettler-Toledo GmbH assumes no liability.
Finding more information
Additional information about this balance can be found in the Operating Instructions on the CD­ROM or online.
u www.mt.com/XPR-precision-OI
u www.mt.com/library

1.1 Definition of signal warnings and symbols

Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results.
WARNING
CAUTION
NOTICE
for a hazardous situation with medium risk, possibly resulting in death or severe injury if not avoided.
for a hazardous situation with low risk, resulting in minor or moderate injury if not avoided.
for a hazardous situation with low risk, resulting in damage to the instrument, other material damage, malfunctions and erroneous results, or loss of data.
General hazard Electrical shock

1.2 Product safety information

Intended use
This balance is designed to be used in laboratories by trained staff. The balance is intended for weighing purposes.
Any other type of use and operation beyond the limits of technical specifications without written consent from Mettler-Toledo GmbH is considered as not intended.
General safety information
This balance complies with current industry standards and the recognized safety regulations; however, it can constitute a hazard in use. Do not open the balance housing: The balance contains no user-serviceable parts. In the event of problems, please contact a METTLER TOLEDO representative.
Always operate and use your instrument only in accordance with the instructions contained in this document. The instructions for setting up your new instrument must be strictly observed.
If the instrument is not used according to the Operating Instructions, protection of the instrument may be impaired and METTLER TOLEDO assumes no liability.
Staff safety
This printed document must be read and understood before using the balance. This printed document must be retained for future reference.
The balance must not be altered or modified in any way. Only use METTLER TOLEDO original spare parts and accessories.
Safety Information8 Precision Balances and Comparators
Safety notes
WARNING
Danger of death or serious injury due to electric shock!
Contact with parts that contain a live current can lead to injury and death. If the balance cannot be shut down in an emergency situation, people can be injured or the balance can be damaged.
1 Only use the original universal AC/DC adapter delivered with your balance. 2 Check that the voltage printed on the balance is the same as your local power supply
voltage. ð If this is not the case, under no circumstances connect the AC/DC adapter to the power
supply, but contact a METTLER TOLEDO representative.
3 Only plug the adapter into a socket which is grounded.
NOTICE
Damage to the balance due to wrong usage.
1 Only use indoors in dry locations. 2 Do not use pointed objects to operate the touchscreen. 3 Only use original METTLER TOLEDO accessories and peripheral devices. 4 Do not open the balance housing. It contains no user-serviceable parts. Contact a
METTLER TOLEDO representative in case of any issue.
Safety Information 9Precision Balances and Comparators

2 Basic Navigation

2.1 Main sections at a glance

The method work screen is the central navigation point where all the menus and settings can be reached (the picture shows the method work screen of the method General Weighing). The sections Balance menu,
Methods and Protocol can be opened by tapping the drawers along the method work screen.

2.2 Method work screen

The method work screen is the basic screen that appears when working with the balance. Depending on the method, the method work screen can slightly differ, but the basic elements appear for all methods.
No. Name Explanation
User name Shows the name of the current user.
1
Level indicator Indicates if the balance is leveled.
2
Weighing value field Shows the current weighing value.
3
Info weight Shows the current weighing value in another unit.
4
Basic Navigation10 Precision Balances and Comparators
No. Name Explanation
5 Drawer Protocol
6 Sample status OK
7 Sample status Not
OK
8 Button Add to
protocol
Weighing action
9
field
Method information
10
area
Weighing-in aid Defines a target weight with upper und lower tolerances.
11
Weighing value
12
area
13 Method name
In this section the weighing results are being shown.
The weighing result was within the defined tolerances.
The weighing result was out of the defined tolerances.
Adds the result to the protocol. Depending on the selected method the button can have different functions.
This field contains actions referring to the current task.
Contains information about sample IDs, method IDs or task IDs.
In this area the result of the current weighing process is shown.
Shows the name of the current method.

2.3 Opening a section/function

A menu item or a function can be selected or activated by tapping it.

2.4 Scrolling

If a list of available options or results is longer than the size of the screen, a scroll bar appears on the right side of the list. Place the finger on the list and move the finger up and down to scroll through the list.
Basic Navigation 11Precision Balances and Comparators

3 Installation and Putting into Operation

Disclaimer for comparators
In this document "balance" is a terminology preamble and stands for comparators.
Comparators are characterized by their higher resolution compared to balances and are mainly used for differential weighing application, such as the calibration of standard weights. Beside standard balance tests, comparators have also been tested with differential repeatability (ABA repeatability) during production.
This information is valid for models of the line/s:
XPR2004SC XPR26003LC XPR5004SC XPR32003LD5C XPR2003SC XPR64003LD5C XPR5003SC XPR64002LC XPR10003SC
Finding more information
Additional information about this balance can be found in the Operating Instructions on the CD­ROM or online.
u www.mt.com/XPR-precision-OI
u www.mt.com/library
WARNING
Risk of electric shock
The balance must be disconnected from the power supply before performing all setup and assembly work.

3.1 Unpacking the balance

Open the balance packaging and check for transportation damage or missing parts. Please inform a METTLER TOLEDO representative in the event of missing or defective parts.
We recommend retaining the original box with its packaging elements. The packaging elements are designed to protect the balance. Use the packaging elements and the original box to store and transport the balance.

3.2 Scope of delivery

NOTICE
Risk of damage to the instrument due to the use of unsuitable parts!
Using unsuitable parts with the instrument can damage the Instrument or cause it to malfunction.
− Only use parts from METTLER TOLEDO that are intended to be used with your instrument.

3.2.1 Balances with S weighing platform

Components 0.1mg
Shield
Weighing platform with protective cover
Terminal with protective cover
Installation and Putting into Operation12 Precision Balances and Comparators
with
Pro
Draft
1mg
with
Pro
Draft
Shield
1 mg
without
Pro
Draft
Shield
5mg / 10 mg
0.1g
Components 0.1mg
Shield
Terminal support
Terminal connection cable (pre-assembled)
Pro Draft Shield
Weighing pan 127 x 127 mm
Weighing pan 172 × 205mm
Weighing pan 193 × 223mm
SmartPan weighing pan
Weighing pan support
DripTray
Ring seal
Weighing hook for below-the-balance weighing
AC/DC adapter
Power cable (country-specific)
Operating instructions and user manuals printed or on CD-ROM (country-specific)
Production certificate
CE declaration of conformity
with
Pro
Draft
1mg
with
Pro
Draft
Shield
1 mg
without
Pro
Draft
Shield

3.2.2 Comparators with S weighing platform

Components 0.1mg 1mg
Weighing platform with protective cover
Terminal with protective cover
Terminal support
Terminal connection cable (pre-assembled)
Pro Draft Shield (not for XPR10003SC)
Draft shield XP W12 (only for XPR10003SC)
SmartPan weighing pan
LevelMatic weighing pan Ø 130 mm (not for XPR2003SC)
Bottom plate
Ring seal
Weighing hook for below-the-balance weighing
AC/DC adapter
Power cable (country-specific)
Operating instructions and user manuals printed or on CD-ROM (country-specific)
5mg / 10 mg
0.1g
Installation and Putting into Operation 13Precision Balances and Comparators
Components 0.1mg 1mg
Mass calibration software MC Link
Production certificate
CE declaration of conformity

3.2.3 Balances with L weighing platform

Components 10mg 0.1g / 1g
Weighing platform
Terminal with protective cover
Terminal support
Terminal connection cable
Draft shield element
Weighing pan 172 × 205mm
Weighing pan 280 × 360mm
SmartPan weighing pan
Power cable (country-specific)
Operating instructions and user manuals printed or on CD-ROM (country-specific)
Production certificate
CE declaration of conformity

3.2.4 Comparators with L weighing platform

Components 1mg 5 mg 10mg
Weighing platform
Terminal with protective cover
Terminal support
Terminal connection cable
Draft shield element XP W64 (only for XPR26003LC and XPR64003LD5C)
Weighing pan 280 × 360mm (not for XPR64003LD5C and XPR64002LC)
LevelMatic weighing pan Ø 220 mm with draft shield element and weighing pan holder (not for XPR32003LD5C)
Weighing pan Ø 220 mm with draft shield cover (only for XPR64002LC)
Power cable (country-specific)
Transport case (only for XPR64002LC)
Operating instructions and user manuals printed or on CD-ROM (country-specific)
Mass calibration software MC Link
Production certificate
Installation and Putting into Operation14 Precision Balances and Comparators
Components 1mg 5 mg 10mg
11
2
CE declaration of conformity

3.3 Selecting the location

A balance is a sensitive precision instrument. The location where it is placed will have a profound effect on the accuracy of the weighing results.
Choose a solid table that is as horizontal as possible.
Choose the table according to the maximum capacity of the balance.
The balance must only be used indoors and up to a
maximum altitude of 4,000m above sea level.
Before switching on the balance, wait until all parts are at
room temperature (+10 °C to +30°C). The humidity must be between 10% and 80% non­condensing.
The power plug must be accessible.
Vibration-free location
No direct sunlight
No excessive temperature fluctuations
No strong drafts

3.4 Assembling balances with S weighing platform

3.4.1 Attaching the terminal to the weighing platform

The terminal is usually placed in front of the weighing platform on the terminal support. The terminal can also be placed individually, e.g. beside the weighing platform or it can be fixed on an additional terminal stand.
NOTICE
Damage to the balance
The weighing platform and the terminal are not safely fixed by the terminal support and may fall off when carrying.
− Remove the terminal from the weighing platform and place it on the weighing pan when carrying the balance.
1 Place the weighing platform on a flat surface. 2 Position the terminal support (1) in front of the
weighing platform. The plug of the pre-mounted terminal connection cable (2) must lie between the terminal support (1) and the weighing platform.
Installation and Putting into Operation 15Precision Balances and Comparators
3 4
3 Push the terminal support towards the weighing
1
2
43
platform. The far end of the terminal support (3) must be pushed into the lock element (4) at the bottom of the weighing platform.
4 Use the terminal connection cable to connect the
terminal with the weighing platform.
5 Place the terminal on top of the terminal support. 6 Push the terminal towards the weighing platform until
the terminal locks into the terminal support.
ð The terminal is mounted and connected to the
weighing platform.

3.4.2 Balances 0.1 mg with Pro Draft Shield and SmartPan weighing pan

Note
The 0.1 mg balances with Pro Draft Shield are equipped with a ring seal. The aim of the ring seal is to isolate the weighing chamber from draft and must always be correctly installed at the bottom of the Pro Draft Shield.
1 Turn the Pro Draft Shield carefully sideways into
horizontal position.
2 Push the ring seal (2) through the opening (1) on the
bottom of the Pro Draft Shield.
3 Fix the ring seal in the opening of the Pro Draft Shield.
The ring seal (3) must be properly fixed into the opening. The upper edge (4) of the ring seal must lay all around the bottom of the Pro Draft Shield.
4 Turn the Pro Draft Shield carefully back into vertical
position.
Installation and Putting into Operation16 Precision Balances and Comparators
5
5 Open the Pro Draft Shield with the door handlers on
6
7
1
both sides.
6 Hold the Pro Draft Shield on the top bars (5) on both
sides and place it on top of the weighing platform.
7 Place the DripTray (6) into the Pro Draft Shield. 8 Place the SmartPan weighing pan (7) into the Pro
Draft Shield on top of the DripTray (6).

3.4.3 Balances 1 mg with Pro Draft Shield

1 Open both side doors of the Pro Draft Shield. 2 Hold the Pro Draft Shield on the top bars (1) on both
sides.
3 Place the Pro Draft Shield on top of the weighing
platform.
Installation and Putting into Operation 17Precision Balances and Comparators
2
3
4
4 Place the DripTray (2) into the Pro Draft Shield.
1
2
3
4
5 Place the SmartPan weighing pan (3) into the Pro
Draft Shield on top of the DripTray (2).
6 The weighing pan (4) is optional and can be placed
into the Pro Draft Shield on top of the SmartPan weighing pan (3).

3.4.4 Comparators 0.1 mg and 1 mg with LevelMatic weighing pan

1 Open both side doors of the Pro Draft Shield. 2 Lift the Pro Draft Shield by holding on the top bars (1)
on both sides.
3 Place the Pro Draft Shield on top of the weighing
platform.
4 Place the bottom plate (2) into the Pro Draft Shield. 5 Place the LevelMatic holder (3) into the Pro Draft
Shield on top of the bottom plate (2).
6 Place the LevelMatic weighing pan (4) into the Pro
Draft Shield on top of the LevelMatic weighing pan holder (3).
Installation and Putting into Operation18 Precision Balances and Comparators

3.4.5 Balances 1 mg with SmartPan weighing pan

11
22
11
22
33
4
2
3
11
1 Place the DripTray (1) on top of the weighing
platform.
2 Place the SmartPan weighing pan (2) on top of the
DripTray (1).
3 The weighing pan can be placed on top of the
SmartPan weighing pan if needed.

3.4.6 Balances 5 mg / 10 mg with SmartPan weighing pan

1 Place the weighing pan support caps (1) on top of the
weighing platform.
2 Place the DripTray (2) on top of the weighing
platform.
3 Place the SmartPan weighing pan (3) on top of the 4
weighing pan support caps (1).
4 Place the weighing pan with the protective cover (4)
on top of the SmartPan weighing pan (3).

3.4.7 Balances 0.1 g

1 Place the 4 pan support caps (1) on top of the
weighing platform.
2 Place the weighing pan support (2) on top of the 4
weighing pan support caps (1).
3 Place the weighing pan with the protective cover (3)
on top of the weighing pan support (2).
Installation and Putting into Operation 19Precision Balances and Comparators

3.5 Assembling balances with L weighing platform

1
2
3
2
11

3.5.1 Attaching the terminal to the weighing platform

The terminal can be attached to the long side or to the short side of the L weighing platform.
1 Turn the weighing platform upside down. 2 Dismantle the screws (1) on the long side or the
screws (2) on the short side of the weighing platform.
3 Connect the terminal to the weighing platform with the
terminal connection cable.
4 Attach the terminal support to the long side or to the
short side of the weighing platform. Fix the terminal support with the screws from the weighing platform.
5 Insert the terminal connection cable (3) into the cable
channel.
Note
When inserting the terminal connection cable into the cable channel, the terminal connection cable must be inserted simultaneously from both directions. The terminal connection cable must not have any play between the plug and the cable channel (see picture).
6 Turn the weighing platform.

3.5.2 Balances 0.1 g and 1 g

1 Place the weighing pan support caps (1) on top of the
weighing platform.
2 Place the weighing pan (2) on top of the weighing
pan support caps (1).
Installation and Putting into Operation20 Precision Balances and Comparators

3.5.3 Balances 10 mg with SmartPan weighing pan

2
3
11
4
5
1
1 Place the adapter plate (1) on top of the weighing
platform.
2 Place the weighing pan support caps (2) on top of the
adapter plate.
3 Place the draft shield element (3) on top of the
adapter plate (1).
4 Place the SmartPan weighing pan (4) on top of the
weighing pan support caps (2).
5 Place the weighing pan (5) on top of the SmartPan
weighing pan (4) if needed.

3.5.4 Removing the transport arrestments (only for comparators)

Note
Before installing the weighing pan, the transport arrestments on top of the weighing platform must be removed. Keep the transport arrestments for transporting the comparator.
1 Remove the transport arrestments (1) at the top of the
weighing platform.
2 Close the openings with the enclosed plastic covers.
Installation and Putting into Operation 21Precision Balances and Comparators

3.5.5 Comparators 1 mg / 5 mg with LevelMatic weighing pan

2
11
3
4
1 Place the adapter plate with the LevelMatic base plate
(1) on top of the weighing platform.
2 Place the draft shield element (2) on top of the
adapter plate (1).
3 Place the LevelMatic weighing pan holder (3) on top
of the LevelMatic base plate (1).
4 Place the LevelMatic weighing pan (4) on top of the
LevelMatic weighing pan holder (3).

3.6 Connecting the balance to the mains

WARNING
Danger of death or serious injury due to electric shock.
Contact with parts that contain a live current can lead to injury and death. 1 To connect the balance, only use the supplied three-core power cable with equipment
grounding conductor. 2 Only use a three-pin power socket with earthing contact. 3 Only use standardized extension cable with equipment grounding conductor. 4 Do not disconnect the equipment grounding conductor.
NOTICE
Damage to the balance due to short circuit!
Damage to the insulation on the AC/DC adapter can result in a short circuit and damage the balance.
Installation and Putting into Operation22 Precision Balances and Comparators
The balance is supplied with an AC/DC adapter and a country-specific power cable. The AC/DC adapter is suitable for use with the following voltage range: 100 – 240 VAC, 50/60Hz.
Connect the balance to the mains as follows:
1 Check whether your local power supply falls within the range of the delivered AC/DC adapter. If this is
1 Only use the original AC/DC adapter supplied with your balance. 2 Route the cable so that it cannot be damaged by external influences. 3 Make sure that the power plug is always accessible.
Note
not the case, under no circumstances connect the AC/DC adapter to the power supply, but contact a METTLER TOLEDO representative.
11
22
2 Connect the AC/DC adapter (2) to the AC/DC socket
(1).
3 Use the power cable to connect the AC/DC adapter to
the mains.
ð The balance performs a self-test after connection to the
power supply and is then ready to use.

3.7 Setting up the Balance

3.7.1 Switching on the balance

When the balance is connected to the mains it starts automatically.
EULA (End User License Agreement)
When the balance is switched on for the first time, the EULA (End User License Agreement) appears on the screen. Read the conditions, tap I accept the terms in the license agreement. and tap
Note
Before the balance can be used, it must warm up first. The warm-up time is at least 30 minutes after connecting the balance to the power supply. When the balance was switched on from standby-mode, the balance is ready for operation immediately.

3.7.2 Leveling the balance

When the balance is switched on for the first time or when it is switched on after the location of the balance has been changed, the message Balance is out of level appears. By tapping opens. Follow the instructions on the screen to level the balance.
Navigation:
Note
A leveled balance and a stable installation are prerequisites for accurate weighing results.

3.7.3 Performing a simple weighing

When the balance is switched on for the first time, the method work screen of the method General Weighing opens (the method work screen is described in more detail in the section Method work screen).
3.7.3.1 Zeroing
1 Clear the weighing pan. 2 Close the weighing chamber if the balance has a draft shield. 3 Tap ð The balance is zeroed.
3.7.3.2 Taring
If a weighing container is used, the balance must be tared.
1 Clear the weighing pan. 2 Close the weighing chamber if a draft shield is being used. 3 Tap 4 Open the weighing chamber if a draft shield is being used. 5 Place the container on the weighing pan. 6 Close the weighing chamber if a draft shield is being used.
Balance menu > Leveling aid
to zero the balance.
to zero the balance.
OK.
the function Leveling aid
Installation and Putting into Operation 23Precision Balances and Comparators
7 Tap to tare the balance. ð The balance is tared.
3.7.3.3 Performing General Weighing
1 Place the weight on top of the weighing pan and wait until the weight value is stable. 2 Tap ð The weighing process was successfull. The weight value has been transferred to the section Protocol.
Add to protocol.

3.7.4 Setting the balance to standby mode

The balance can be set to standby mode by holding again.

3.7.5 Switching off the balance

To completely switch off the balance it must be disconnected from the mains. By holding goes only into standby mode.
Note
When the balance was completely switched off for some time, it must be warmed up for at least 30 minutes before it can be used.
. The standby mode can be finished by holding
the balance
Installation and Putting into Operation24 Precision Balances and Comparators

4 Maintenance

4.1 Cleaning

We recommend periodically cleaning the draft shield, the weighing platform, and the terminal. For cleaning the weighing chamber use the brush supplied with it. The maintenance interval depends on your standard operating procedure (SOP). The balance is made of high-quality, resistant materials and can therefore be cleaned with a commercially available, mild cleaning agent.
WARNING
Danger of death or serious injury due to electric shock!
Contact with parts that contain a live current can lead to injury and death. If the device cannot be shut down in an emergency situation, people can be injured or the device can be damaged.
1 Disconnect the device from the power supply prior to cleaning and maintenance. 2 Only use METTLER TOLEDO power cables, if these need to be replaced. 3 Make sure that no liquid enters the device or AC/DC adapter. 4 Do not open the device or AC/DC adapter. These do not contain any user-serviceable
parts.
NOTICE
Damage of the balance due to the use of inappropriate cleaning agents!
− On no account use cleaning agents which contain solvents or abrasive ingredients, as this can result in damage to the terminal overlay.
Note
Please contact your METTLER TOLEDO dealer for details of the available service options. Regular servicing by an authorized service engineer ensures constant accuracy for years to come and prolongs the life of your balance.

4.2 Disposal

In conformance with the European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
Maintenance 25Precision Balances and Comparators

5 Technical Data

FCC Rules
This device complies with Industry Canada licence-exempt RSS standard(s) and part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful inter­ference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Technical Data26 Precision Balances and Comparators

5.1 Balances with S weighing platform

Power supply
AC adapter: Primary: 100 – 240VAC, 50/60 Hz
Secondary: 12VDC ±3%, min. 2.5A, max. 6.0 A (with
electronic overload protection) Cable for AC adapter: 3-core, with country-specific plug Balance power supply: 12 VDC ±3%, 2.25 A, maximum ripple: 80 mVpp
Protection and standards
Overvoltage category: II Degree of pollution: 2 Protection: Protected against dust and water Standards for safety and EMC: See Declaration of Conformity Range of application: For use in closed interior rooms only
Environmental conditions
Height above mean sea level: Up to 4000m Ambient temperature: 5–40°C Relative air humidity: Max. 80% up to 31°C, linearly decreasing to 50% at 40°C,
Warm-up time:
Environmental conditions Comparators
Height above mean sea level: Up to 4000m Ambient temperature: 10 – 30 °C ±0.5°C / 12 h Relative air humidity: 40% – 70% ±5%/4h Maximum air speed: 0.15 m/sec Warm-up time: At least 60 minutes after connecting the comparator to the
Materials
Weighing platform: Die-cast aluminum, lacquered, plastic and chrome steel Terminal: Die-cast zinc, chromed and plastics Weighing pan: Chrome-nickel steel X2CrNiMo-17-12-2 SmartPan weighing pan with drip
tray: LevelMatic weighing pan
(Comparators only): Draft shield: Aluminum, plastic, chrome steel and glass Draft shield element: Chrome steel X2CrNiMo-17-13-2
noncondensing
At least 30 minutes after connecting the balance to the power
supply; when switched on from standby-mode, the balance is
ready for operation immediately
power supply. When switched on from standby-mode, the
comparator is ready for operation immediately.
Die-cast zinc, chromed
Chrome steel, plastic
Technical Data 27Precision Balances and Comparators

5.2 Balances with L weighing platform

Power supply
AC adapter: Primary: 100 – 240VAC, 50/60 Hz
Secondary: 12VDC ±3%, min. 2.5A, max. 6.0 A (with
electronic overload protection) Cable for AC adapter: 3-core, with country-specific plug Balance power supply: 12 VDC ±3%, 2.25 A, maximum ripple: 80 mVpp
Protection and standards
Overvoltage category: II Degree of pollution: 2 Protection: Protected against dust and water Standards for safety and EMC: See Declaration of Conformity Range of application: For use in closed interior rooms only
Environmental conditions
Height above mean sea level: Up to 4000m Ambient temperature: 5–40°C Relative air humidity: Max. 80% up to 31°C, linearly decreasing to 50% at 40°C,
Warm-up time:
Environmental conditions Comparators
Height above mean sea level: Up to 4000m Ambient temperature: 10 – 30 °C ±0.5°C / 12 h Relative air humidity: 40% – 70% ±5%/4h Maximum air speed: 0.15 m/sec Warm-up time: At least 60 minutes after connecting the comparator to the
Materials
Weighing platform: Sheet aluminum, die cast, laquered, plastic and chrome steel Terminal: Die-cast zinc, chromed and plastics Weighing pan: Chrome-nickel steel X5CrNi18-10 SmartPan weighing pan with drip
tray: LevelMatic weighing pan
(Comparators only): Draft shield: Plastic, chrome-nickel steel, aluminium and glass
noncondensing
At least 30 minutes after connecting the balance to the power
supply; when switched on from standby-mode, the balance is
ready for operation immediately
power supply. When switched on from standby-mode, the
comparator is ready for operation immediately.
Die-cast zinc, chromed
Chrome-nickel steel and aluminium
Technical Data28 Precision Balances and Comparators
Mettler-Toledo GmbH
30357071
Im Langacher 44 8606 Greifensee, Switzerland www.mt.com/contact
Subject to technical changes. © Mettler-Toledo GmbH 09/2017 30357071Aen