Mettler Toledo XPR1202S, XPR1203S/A, XPR1202S/A, XPR1203S, XPR10002S Data Sheet

...
Excellence Level
XPR Precision Balances
Accurate, Flexible, Compliant
Outstanding Performance
The SmartPan™ weighing pan minimizes the effects of air currents on the weighing cell to deliver faster and more accurate results. In most conditions a draf t shield is no longer needed, significantly reducing process steps.
Efcient Processes
Customizable, integrated weighing methods guide you through each weighing process, saving time and providing higher process security. The built-in notepad automatically records all task parameters and results, eliminating manual transcription.
Effortless Compliance
QA functions automatically monitor the balance level, routine tests and the minimum weight limit. The integrated StatusLight™ makes it easy to see at a glance if it’s okay to start your weighing task. Compliance management has never been this easy.
Error-free Data Handling
Data transfer to a PC or soft ware program is easy and error-free via USB or Ethernet. Integrated LabX functionalities assist you in ensuring data integrity. Just plug in a printer and barcode reader to eliminate manual transcription entirely.
XPR Precision Balance Solutions
Go Beyond Weighing
XPR small platform precision balances deliver outstanding weighing performance and support the highest requirements for data integrity and compliance.
Thanks to the extraordinary stability provided by the unique SmartPan weighing pan, results are delivered up to twice as fast even under tough weighing environments. Repeatability is improved up to two-fold.
The state-of-the-art user interface offers intelligent functions to simplify and support your processes at every step. With smart quality assurance features automatically monitoring the status of your balance, you can rest assured that your results are valid.
XPR-S Precision Balances
SmartPan models
SmartPan weighing pans minimize the effects of air drafts on the weighing cell significantly improving stability and measurement accuracy.
Material number 30 315631 30 315 628 30315 239
Limit values
Maximum capacity (full range/fine range) 210 g 410 g 510 g / 101 g Readability (full range/fine range) 0.1 m g 0.1 m g 1 mg / 0.1 mg Repeatability (full range/fine range) 0.2 mg 0.1 m g 0.5 mg / 0.1 mg Linearity deviation 0.2 mg 0.2 mg 0.4 mg
Typical values
Repeatability (full range/fine range) 0.08 mg 0.08 mg 0.4 mg / 0.08 mg Linearity deviation 0.07 mg 0.07 mg 0.7 m g Sensitivity of fset (test weight) 0.4 mg (200 g) 0.48 mg (400 g) 6 mg (500 g) Minimum weight (5% load, k=2, U=1.0%) 16 mg 16 m g 16 mg USP minimum weight (5% load, k=2, U=0.10%) 160 mg 160 mg 160 mg Settling time 2 s 2 s 1.5 s
Dimensions
Weighing pan size (mm) Weighing pan SmartPan SmartPan SmartPan Draft shield Pro draft shield Pro draft shield Pro draft shield
LevelMatic models
1
The LevelMatic weighing pan minimizes corner
load effects enabling you to weigh with an extreme level of accuracy.
XPR204S XPR404S XPR504SDR
68 68 68
2
METTLER TOLEDO XPR-S Precision Balances
XPR504S XPR2004SC
Material number 30 315 652 30 317 628
Limit values
Maximum capacity 510 g 2.3 kg Readability 0.1 m g 0.1 mg Repeatability 0.1 m g 0.6 mg Linearity deviation 0.3 mg 1 mg
Typical values
Repeatability 0.08 mg 0.3 mg Linearity deviation 0.2 mg 0.5 mg Sensitivity of fset (test weight) 0.3 mg (500 g) 6.8 mg (2 kg) Minimum weight (5% load, k=2, U=1.0%) 16 mg 60 mg USP minimum weight (5% load, k=2, U=0.10%) 160 mg 600 mg Settling time 2 s 3.5 s
Dimensions
Weighing pan size (mm) Weighing pan SmartPan LevelMatic Draft shield Pro draft shield Pro draft shield
68 130
1
XPR303SN
2
XPR303S XPR603SNDR
Material number 30316446 30 315753 3 0316529
Limit values
Maximum capacity (full range/fine range) 310 g 310 g 610 g / 120 g Readability (full range/fine range) 1 mg 1 mg 10 mg / 1 mg Repeatability (full range/fine range) 0.9 mg 0.9 mg 6 mg / 1 mg Linearity deviation 2 mg 2 mg 6 mg
Typical values
Repeatability (full range/fine range) 0.45 mg 0.45 mg 4 mg / 0.45 mg Linearity deviation 0.6 mg 0.6 mg 0.6 mg Sensitivity of fset (test weight) 1.2 mg (300 g) 1.2 mg (300 g) 1.6 mg (600 g) Minimum weight (5% load, k=2, U=1.0%) 90 mg 90 mg 90 mg
USP minimum weight (5% load, k=2, U=0.10%) 900 mg 900 mg 900 mg
Settling time 1.5 s 1.5 s 1.5 s
Dimensions
Weighing pan size (mm) 127 x 12 7 127 x 127 127 x 127 Weighing pan SmartPan Standard pan
SmartPan
SmartPan
Draft shield Optional Pro draft shield Optional
2
Supplied without draft shield (available as an accessory)
2
XPR603SDR XPR603SN
2
XPR603S
Material number 30316 525 30 31654 0 30316504
Limit values
Maximum capacity (full range/fine range) 610 g / 120 g 610 g 610 g Readability (full range/fine range) 10 mg / 1 mg 1 mg 1 mg Repeatability (full range/fine range) 6 mg / 1 mg 0.9 mg 0.9 mg Linearity deviation 6 mg 2 mg 2 mg
Typical values
Repeatability (full range/fine range) 4 mg / 0.45 mg 0.45 mg 0.45 mg Linearity deviation 0.6 mg 0.6 mg 0.6 mg Sensitivity of fset (test weight) 1.6 mg (600 g) 1.2 mg (600 g) 1.2 mg (600 g) Minimum weight (5% load, k=2, U=1.0%) 90 mg 90 mg 90 mg USP minimum weight (5% load, k=2, U=0.10%) 900 mg 900 mg 900 mg Settling time 1.5 s 1.5 s 1. 5 s
Dimensions
Weighing pan size (mm) 127 x 127 127 x 127 12 7 x 12 7 Weighing pan Standard pan,
SmartPan
SmartPan, Standard pan,
SmartPan
Draft shield Pro draft shield Optional Pro draf t shield
2
Supplied without draft shield (available as an accessory)
XP R1203 S XPR3003S XPR5003S
Material number 30316563 30 316701 30 316990
Limit values
Maximum capacity 1210 g 3.1 kg 5.1 k g Readability 1 mg 1 mg 1 mg Repeatability 0.8 mg 1 mg 1. 5 mg Linearity deviation 2 mg 6 mg 5 mg
Typical values
Repeatability 0.4 mg 0.6 mg 1 mg Linearity deviation 0.6 mg 2 mg 2 mg Sensitivity of fset (test weight) 1.5 mg (1.2 kg) 1.2 mg (3 kg) 4 mg (5 kg) Minimum weight (5% load, k=2, U=1.0%) 82 mg 120 m g 200 mg USP minimum weight (5% load, k=2, U=0.10%) 820 mg 1.2 g 2 g Settling time 1.5 s 2 s 2 s
Dimensions
Weighing pan size (mm) 127 x 127 127 x 127 127 x 127 Weighing pan Standard pan,
SmartPan
Standard pan, SmartPan
Standard pan, SmartPan
Draft shield Pro draft shield Pro draft shield Pro draft shield
METTLER TOLEDO XPR-S Precision Balances
3
XPR10003SC1XPR3003SD5 XPR6003SD5
Material number 30 317641 3031670 5 30316548
Limit values
Maximum capacity 10.1 k g 3 .1 k g 6.1 kg Readability 1 mg 5 mg 5 mg Repeatability 3.5 mg 6 mg 6 mg Linearity deviation 7 mg 6 mg 7 mg
Typical values
Repeatability 2 mg 3 mg 3 mg Linearity deviation 3.2 mg 2 mg 2 mg Sensitivity of fset (test weight) 34 mg (10 k g) 4 mg (3 kg) 7.2 mg (6 kg) Minimum weight (5% load, k=2, U=1.0%) 400 mg 600 mg 600 mg USP minimum weight (5% load, k=2, U=0.10%) 4 g 6 g 6 g
Settling time 3.5 s 2 s 2 s
Dimensions
Weighing pan size (mm)
130
Weighing pan LevelMatic Standard pan
172 x 205 172 x 205
Standard pan
SmartPan
SmartPan
Draft shield Draft Shield W12 Optional Optional
XP R1202S XPR2002S XPR4002S
Material number 30 317123 30 316996 30 317 269
Limit values
Maximum capacity 1210 g 2.1 k g 4.1 k g Readability 0.01 g 0.01 g 0.01 g Repeatability 8 mg 8 mg 8 mg Linearity deviation 20 mg 20 mg 20 mg
Typical values
Repeatability 4 mg 4 mg 4 mg Linearity deviation 6 mg 6 mg 6 mg Sensitivity of fset (test weight) 15 mg (1.2 kg) 25 mg (2 kg) 25 mg (4 kg) Minimum weight (5% load, k=2, U=1.0%) 820 mg 820 mg 820 mg USP minimum weight (5% load, k=2, U=0.10%) 8.2 g 8.2 g 8.2 g Settling time 1.2 s 1.2 s 1.2 s
Dimensions
Weighing pan size (mm) 172 x 205 172 x 205 172 x 205 Weighing pan
Standard pan SmartPan
Standard pan SmartPan
Standard pan SmartPan
Draft shield Optional Optional Optional
4
METTLER TOLEDO XPR-S Precision Balances
XPR6002SDR XPR6002S XPR8002S XPR10002S
Material number 30 317458 3 0 31742 8 30317466 30317433
Limit values
Maximum capacity (full range/fine range) 6.1 kg / 1.2 kg 6.1 k g 8.1 kg 10.1 kg Readability (full range/fine range) 0.1 g / 0.01 g 0.01 g 0.01 g 0.01 g Repeatability (full range/fine range) 60 mg / 8 mg 8 mg 8 mg 8 mg Linearity deviation 60 mg 20 mg 20 mg 20 mg
Typical values
Repeatability (full range/fine range) 40 mg / 4 mg 4 mg 4 mg 4 mg Linearity deviation 6 mg 6 mg 6 mg 6 mg Sensitivity of fset (test weight) 36 mg (6 kg) 12 mg (6 kg) 12 mg (8 kg) 12 mg (10 kg) Minimum weight (5% load, k=2, U=1.0%) 820 mg 820 mg 820 mg 820 mg USP minimum weight (5% load, k=2, U=0.10%) 8.2 g 8.2 g 8.2 g 8.2 g Settling time 1.2 s 1.2 s 1.5 s 1.5 s
Dimensions
Weighing pan size (mm) 172 x 205 172 x 205 172 x 205 172 x 205 Weighing pan Standard pan
SmartPan
Standard pan SmartPan
Standard pan SmartPan
Standard pan SmartPan
Draft shield Optional Optional Optional Optional
XPR4001S X PR 60 01S XPR8001S XPR10001S
Material number 30 3174 47 30 3174 91 30 31 7511 3031748 3
Limit values
Maximum capacity 4.1 kg 6.1 k g 8.1 k g 10.1 k g Readability 0.1 g 0.1 g 0.1 g 0.1 g Repeatability 80 mg 80 mg 80 mg 80 mg Linearity deviation 60 mg 60 mg 10 0 m g 10 0 mg
Typical values
Repeatability 40 mg 40 mg 40 mg 40 mg Linearity deviation 20 mg 20 mg 30 mg 30 mg Sensitivity of fset (test weight) 50 mg (4 kg) 50 mg (6 kg) 120 mg (8 kg) 120 mg (10 kg) Minimum weight (5% load, k=2, U=1.0%) 8.2 g 8.2 g 8.2 g 8.2 g USP minimum weight (5% load, k=2, U=0.10%) 82 g 82 g 82 g 82 g Settling time 0.8 s 0.8 s 1 s 1 s
Dimensions
Weighing pan size (mm) 190 x 223 190 x 223 190 x 223 190 x 223 Weighing pan Standard pan Standard pan Standard pan Standard pan Draft shield Optional Optional Optional Optional
Features
Outstanding Performance
• Max capacity from 200 g to 10 kg
• Readability from 0.1 mg to 0.1 g
• High performance models from 2 kg (0.1 mg) to 10 kg (1 mg)
• SmartPan and LevelMatic weighing pans
Quality Assurance
• GWP Approved built-in qualit y assurance monitoring
• Customizable tolerance profiles
• MinWeigh warning function
• Adjustment and routine test history
• Balance ready StatusLight™
• LevelControl: level warning with graphical leveling guide
• Temperature and time programmable automatic adjustment (proFACT)
• User Management, individual profiles password protected
• Change history log
Efficient Operation
• 7 inch color capacitive touchscreen terminal, glove compatible
• Storage of weighing methods (including sample series and tolerances)
• Integrated results notepad
Seamless Process
• Easy connectivit y and data export: 4 x USB, 1 LAN
• Easy connection of peripheral devices: printers, PC, barcode readers, footswitches and sensors
• Integrated label editor, alphanumeric and barcode
• Export of process and results data to XML or Excel
• Alibi Memory (available with legally approved models)
Sustainable Value
• Integrated overload protection
• Chemically resistant materials
• Designed for easy cleaning
LabX Laboratory Software
Connect your balance to LabX software to benefit from a fully integrated solution which can satisfy all your data handling needs.
• SOP user guidance on the instrument
• Automatic data handling and calculations
• Easy data access and documentation
• Full process and results traceability
• Centralized management of multiple instruments
METTLER TOLEDO XPR-S Precision Balances
5
Selected Accessories
Printers
The P-50 family of printers generate fast, high­quality printouts on paper, continuous label and individual labels with IDs and barcodes.
Hand/foot Switches
Keep your hands free to handle your sample and equipment: use a remote switch to perform selected balance operations.
Barcode Reader
Read in sample and user IDs quickly and without error with a handheld barcode reader.
Protective Covers
Disposable terminal and platform covers protect your balance and help avoid cross-contamination. A dust cover shields your balance against soiling.
Accessory Material Numbers
Universal antistatic kit 1110 7 7 6 7
USB Antistatic kit 30215452
P-52 RUE dot matrix printer; RS232, USB, Ethernet, optional BT 30237290
P-56 RUE thermal direct printer; RS232, USB, Ethernet, optional BT 30094673
P-58 RUE thermal direct printer; RS232, USB, Ethernet, optional BT 30094674
USB barcode reader 30417466
Terminal stand for S-platform balances 30125 07 7
Bluetooth adapter, connects via USB 30416089
ErgoSens: infrared sensor for touch-free remote operation (USB connection) 30300915
Footswitch: switch for remote operation (USB connection) 30 3125 5 8
Protective cover for S-platform balances: 5 mg, 10 mg, 0.1 g 30215056
Protective cover for S-platform balances: 0.1 mg, 1 mg 30300967
Protective cover for XPR terminal 30125 37 7
Protective cover for standard weighing pan, S-platform: 5 mg, 10 mg (172 x 205mm) 3010 6207
Protective cover for standard weighing pan, S-platform: 0.1 g (190 x 223 mm) 3 01062 04
USB-RS232 converter: connects peripheral devices to the balance via RS232 64088427
Terminal extension cable: length 4.5 m 30300920
Terminal Stand
Place your terminal on a stand which is quickly mounted at the back of your balance. The display is easier to read and good posture is maintained.
Antistatic Kit
Eliminate electrostatic charges on samples and containers by passing them through the free­standing ionizer.
CarePac® certified weight
sets for routine testing
www.mt.com/carepacs
METTLER TOLEDO Group
Laboratory Weighing Local contact: www.mt.com/contacts
Subject to technical changes © 09/2017 METTLER TOLEDO. All rights reserved 30402312B Global MarCom 2232 RK
www.mt.com/xpr-precision
For more information
Loading...