Mettler Toledo RM50, RM40 Instruction Manual

Refractometers
RM40 / RM50
Operating Instructions

Table of Contents

Introduction1 7
Description of the Refractometer2 8
Principles of Refractometry3 10
Description of Functions4 12
Setup5 15
Definition of the Refractive Index3.1 10 Method of Measurement3.2 10 Design of Measuring Cell3.3 10
Layout of the Terminal4.1 12 Operating the Touchscreen4.2 12 Homescreen4.3 12 The User Interface4.4 13
Entering Data in the User Interface.4.4.1 13 Shortcuts4.4.2 14
Adjustments & Tests5.1 15 Adjustment Sets5.1.1 16 Test Sets5.1.2 16
Hardware5.2 16 Cell5.2.1 17 Automation5.2.2 17 External Instruments5.2.3 19 Peripherals5.2.4 20 Sensors5.2.5 23 Auxiliary instruments5.2.6 24
User settings5.3 25 Language5.3.1 25 Screen5.3.2 25 Beep5.3.3 25 Shortcuts5.3.4 26 Keyboards5.3.5 26
Global Settings5.4 26 System5.4.1 26 User Management5.4.2 26 Users5.4.2.1 27 Account Policy5.4.2.2 27 User Groups5.4.2.3 27 Analysis and Resources Behavior5.4.3 28 Physical Properties5.4.4 29
Tables & Values5.5 29 Tables5.5.1 30 Auxiliary Values5.5.2 31
Maintenance & Service5.6 31 MT Service5.6.1 31 Import / Export5.6.2 32 Add External Cell5.6.3 33 Reset to Factory Settings5.6.4 33 Firmware5.6.5 33 Update5.6.6 33 Hardware / Firmware summary5.6.7 34 Board Tests5.6.8 34
Table of Contents 3
Cell5.6.9 34 Export Adjustments / Tests / Measurements5.6.10 35
Methods6.1 36 Establishing Methods6.1.1 36 Creating a Method Copy6.1.2 38 Modifying or Deleting Methods6.1.3 38 Method Syntax6.1.4 38 Standard Data6.1.5 40 Condition6.1.6 40 Method Function6.1.7 41 Title6.1.7.1 42 Configuration6.1.7.2 42 Sample6.1.7.3 42 Fill6.1.7.4 44 Measure6.1.7.5 45 Calculation6.1.7.6 47 Clean6.1.7.7 48 Online Display6.1.7.8 50 Report6.1.7.9 50 Cell Test6.1.7.10 51 Temperature Compensation6.1.7.11 51 Adjustment6.1.7.12 52 Test6.1.7.13 52 Instruction6.1.7.14 52 Auxiliary Value6.1.7.15 52 Wait6.1.7.16 53 Auxiliary instrument6.1.7.17 53 PowerShower6.1.7.18 54 Stir6.1.7.19 55 Line rinse6.1.7.20 56 Park6.1.7.21 56
Products6.2 56 Create Products6.2.1 57 Linking Methods with Products6.2.2 57 Parameters for Products6.2.3 57 Calculations in Products6.2.4 59
Delete Series7.1 61 Creating a series copy7.2 61
Statistics8.1 63 Information8.2 63 Data8.3 63
Automation9.1 65 Action: Dry9.1.1 65 Action: Rinse9.1.2 65 Action: Pump sample9.1.3 66 Action: Move to position9.1.4 66
Table of Contents4
Cell9.2 66 Action: Cell conditioning9.2.1 67 Cell Test9.2.2 67
Analysis Sequence10 68
Starting an Analysis10.1 68 Start from Editor (Methods/Products/Series)10.2 69 Start from Homescreen10.3 69 Start via Shortcuts10.4 69 Start with Barcode Reader10.5 69
Start with Handheld Reader10.5.1 70 Start with Built-in Barcode Reader10.5.2 70 Start with ErgoSens10.5.3 70
Continuous run10.6 71 Analysis Termination10.7 71 Errors in the Analysis Sequences10.8 71
Malfunction Types: Error10.8.1 71 Malfunction Types: Terminate error10.8.2 71 Malfunction Types: Critical error10.8.3 72
Tasks and Online Screen11 73
Tasks11.1 73 "Tasks" Button11.1.1 73
Online Screen11.2 73 Method type: Measurement11.2.1 73 Method Type: Adjustment11.2.2 74 Method Type: Test11.2.3 74 Method Type: Clean11.2.4 74
Appendix12 75
Raw Data12.1 75 Result proposals12.2 76 Refractive Index for Water12.3 78 Density12.4 78 Critical errors12.5 80
Index 82
Table of Contents 5
1Introduction
Simple and compact
The METTLER TOLEDO RM40/RM50 refractometers are modern, compact instruments suitable for use in a vast diversity of application areas. They can be used, for example, in quality control as well as in research and development and meet the most demanding requirements.
These compact refractometers perfectly combine simple, easy-to-understand operation with an extremely high level of measuring accuracy and outstanding reliability. With their plug & play capability, they automatically detect external devices and sensors.
These refractometers can be operated as standalone instruments or run from a computer using the LabX PC software. Straightforward user guidance on the large color touchscreen enables intuitive operation. User-defin able shortcuts allow all functions to be activated directly from the home screen with a single click.
Touchscreen control of the instrument and the method function parameters are described in the Operating Instructions. The Installation Instructions explain all the necessary steps for setting up your instrument. You are then guided through the first refractive index measuring process with the aid of a practical example. If you have any additional questions, METTLER TOLEDO is always available to assist you.

7Introduction

2Description of the Refractometer
The RM40 / RM50 refractometers measure the refractive index nD of liquids. The two instruments differ in mea suring precision. Both instruments:
measure liquids whose refractive indices range respectively from 1.3200 to 1.7000 (RM40) and to
1.58000 (RM50).
require only minimal sample quantities for measurements (0.5 mL),
maintain the temperature of the sample constant with the aid of a built-in Peltier thermostat in a range from 5 to 100°C (RM40) and 5 to 75°C (RM40).
directly indicate the sugar content of samples in various sugar units (Brix, invert sugar, HFCS42 or HFCS55)
calculate the concentration of a solution by means of data from measured standard solutions, literature tables or formulas
are equipped with an integrated test function that enables regular testing of the accuracy of the measure ment,
enable checking of the product-related specification limits and generate rolling statistics.
The samples can be introduced manually with a syringe. For the optimal use of the RM refractometers, the fol lowing pumps and sample changers are available:
A peristaltic pump (FillPal): For the automatic filling, emptying and rinsing of the cell.
The autosampler InMotion™ The autosampler for a completely automatic measurement of up to 303 aqueous, low viscous samples in series.
The automatic sample and cleaning unit (SC1): for fully automated measurements of free-flowing and viscous samples. The system automatically cleans and dries itself on completion of a measurement so that it is ready for the next measurement.
The sample changer (SC30): The SC30 sample changer for the completely automatic measurement of up to 30 samples in series.
The following devices can be connected:
Computer for operation under LabX
External measuring cells:
METTLER TOLEDO DX40, DX45 or DX50 density - modules
METTLER TOLEDO RX40 or RX50 refractive index - modules
Barcode reader for scanning sample data, set values of certified standards and for starting measurements
finger print reader for user identification
Compact printer (USB-P25), to print out results
External instruments:
METTLER TOLEDO S20 – SevenEasy™ pH
METTLER TOLEDO S30 – SevenEasy™ Conductivity
METTLER TOLEDO S220 - SevenCompact™ pH/Ion
METTLER TOLEDO S230 - SevenCompact™ Conductivity
Lovibond colorimeters PFX880/ PFXi880, PFX950/ PFXi950 and PFX995/ PFXi995 series, Tintometer
Minolta colorimeters CM-5 / CR-5
USB stick
External sensors:
ErgoSens – Infrared movement sensor for the automatic start of measurements
LevelSens – Level sensor for waste bottle
AtmoSens – Air pressure sensor for measuring air pressure and taking the latter into account in adjust ments and tests with air.
Special features of the refractometer:

8 Description of the Refractometer

A maintenance-free LED is used as a light source for the measurements.
The measuring prism is made of sapphire. It is therefore extremely corrosion resistance and very robust. It also has a high thermal conductivity.
For temperature measurement, NTC thermistors with extreme long-term stability are used. This means that there is no need to regularly readjust the measuring cell.
9Description of the Refractometer
3Principles of Refractometry
n
1
n
2
n
1
β
n
2
α
sin sin
=
1
2
β
β
1
2
3
4
3.1Definition of the Refractive Index
The refractive index of a substance is the ratio of the speed of a light beam in a vacuum to its speed in the sub stance (dimensionless).
If a light beam crosses at a certain angle from an optically less dense to an optically denser medium (e.g. from air to water), it is deflected; when the beam crosses from one medium to another perpendicularly no directional change takes place. According to Snell's law, the ratio of the refractive index of the two media is proportional to the ratio of the incident and refrac tive angle of the light beam:
If a light beam passes from an optically denser to a less dense medium, it is also deflected. If the angle of incidence α is increased, then it reaches a critical value, at which the light beam no longer enters the less dense medium (refractive angle
= 90°). When this "critical angle" is exceeded, total internal reflection occurs. The refractive index is calculated from this critical angle α :
As refraction depends on the wavelength of the incident light, the refractive index n is measured as standard with the D line of sodium (wavelength 589.3 nm) and designated nD.
The refractive index not only depends on the wavelength of the light but also on the temperature of the mea sured sample. For this reason the temperature must always be stated along with the result, e.g. n The standard temperature is 20 °C.
3.2Method of Measurement
= 90o —> sinβ =1
1: Critical angle 2: Total internal reflection
25
.
D
The light emitted from the light source passes through the prism and hits the sample. In the process it is partly refracted (angle of incidence < critical angle) and partly reflected (angle of incidence > critical angle).
The reflected light is measured by means of an optical sensor (CCD). The boundary between the dark and light area corre sponds to the critical angle required for calculation of the refractive index.
1: Light source 2: Prism 3: Sample 4: Optical sensor (CCD)
3.3Design of Measuring Cell
The light source is a light-emitting diode (LED), whose beam passes through a polarization filter, an interfer ence filter (589.3 nm) and various lenses before it reaches the sample via the sapphire prism. The reflected light (angle of incidence > critical angle) is deflected via a lens to the optical sensor that determines the critical angle. The temperature in the prism/sample boundary is measured with a built-in sensor.

10 Principles of Refractometry

1: Light source 2: Polarization filter 3: Interference filter 4: Lens 5: Lens 6: Prism 7: Lens 8: Measuring cell 9: Optical sensor (CCD)
11Principles of Refractometry
4Description of Functions
4.1Layout of the Terminal
The control panel of the terminal consists of an integrated touchscreen and the following buttons, located next to the touch-sensitive surface of the display: You can press these buttons any time, regardless of which dialog you are currently using.

Two Home buttons that bring you back to the Homescreen.

Info, opens the info screen that shows the data related to the specific device.
Reset that interrupts the ongoing task and processing of the pending tasks. To continue the waiting tasks, open the task list (with the Tasks button) and click Resume.
4.2Operating the Touchscreen
The touchscreen is automatically activated when the instrument is switched on.
To select a button or an input element in the dialog window, you simply touch the screen using a soft blunt object or a fingertip.
It is also possible to select input elements using a USB mouse. To do this, simply connect the mouse to a suit able USB port on the instrument.
Never touch the surface of the touchscreen with pointed or sharp objects! This may damage the screen!
4.3Homescreen
Home is the first screen that is displayed when you start up the instrument or when you log in. Home is the
main screen. On the left-hand side of the screen you will see five buttons that lead to the following dialog win dows:
Methods / Products: The button leads you to the method or product editor, in which you can create and administer the methods or products (see "Methods and Products").
Series: In this dialog, you can create and manage series of individual samples, e.g. for using a sample changer (see "Series").
Results: Here you administer the results of your analyses (see "Results and Statistics").
Setup: You can administer the following points here:
Adjustment and test sets
The hardware and all resources used by the instrument
User and global settings
Tables (internal and user defined) and auxiliary values
Maintenance and service of the instrument
Manual: This button takes you to manual operations.
In addition, there is another area that can be configured individually by each user (with the necessary autho rization). Each user can store up to eight shortcuts here. With these shortcuts, defined methods, products, series and manual operations can be started directly from the homescreen (see "Functional Description"): The user interface > Shortcuts").
Standby display: The standby display continuously shows the current cell temperature (Tcell) and set tem perature (Tset), even if no task is running. Via the standby display in the Homescreen you open the Cell data dialog.
By pressing the home button on the control panel of the terminal, you return to Home.
See also
Methods and Products (page36)
Series (page60)
Results and Statistics (page62)
Shortcuts (page14)

12 Description of Functions

4.4The User Interface
A
B
C
T
1
2
3
T
T
T
T
T
i
The graphical user interface consists of the following five basic elements:
The title bar at the top of the display specifies the name of the current dialog.
In the top right-hand corner you will see Tasks button, which signals the presence of ongoing processes (see "Tasks and online screen: Tasks (page73)").
The navigation bar, located below the title bar, specifies the path to the current dialog.
The scroll bar on the right-hand side of the screen becomes visible if the content of the screen extends beyond the viewable area. If this occurs, use either the arrows or the area in between them to move the viewable area of the screen up or down.
Five buttons are located at the bottom of the screen. The function of these buttons varies and depends on the context of the current dialog.
4.4.1Entering Data in the User Interface.
There are different types of input fields in the user interface. They allow you to enter data or select data from a list. Input fields can also be deactivated and their contents are then displayed as information only and cannot be changed in the corresponding dialog.
The various types of input fields are identified by an icon to the right of the screen:
Text input fields
Number input fields
Drop-down lists Selecting these fields opens a drop-down list from which you can
List fields Selecting these fields opens a menu list in a new window.
Parameter field Selecting these fields opens a new dialog with various additional para
Formula fields Freely definable formulas can be entered in these fields
In these fields text can be freely entered.
Numbers can be entered in these fields.
select an entry.
meters.
Info field
In addition to the input fields there are checkboxes that can be checked in order to select certain functionalities. Checkboxes can affect the scope of the corresponding dialog, i.e. input fields can be hidden or visible depend ing on whether the checkbox is checked.
The values in deactivated input fields are displayed as information only and cannot be edited in the corresponding dialog.
13Description of Functions
Sorting lists
There are lists that can be sorted alphabetically or numerically by column in ascending or descending order. To do this, simply touch the parameter in the header row by which you would like to sort the list. A small arrow in the header row indicates the parameter by which the list is sorted and whether it is sorted in ascending or descending order.
4.4.2Shortcuts
Shortcuts allow you to start methods, series, and manual operations directly from the homepage. You can locate shortcuts by means of the button AddToHome on the Homescreen. AddToHome is found in the respec tive Start analysis dialog of the methods, products, series or of the manual operation.
Two types of shortcuts are supported. Direct shortcuts which, when selected, start the task immediately without warning (only if the other settings allow this), and normal shortcuts which take you to the corresponding Start analysis start dialog from which you can start the task.
Shortcuts are user-specific, i.e. each individual user can create shortcuts for the tasks that they would personal ly like to conduct. Shortcuts are managed in Setup, under the subcategory "User Settings". Here you can delete or modify shortcuts, or change their position on the Homescreen.
for methods
for products
for series

Shortcuts

(lead you initially to the
Start analysis dialog)
Direct shortcuts
(task starts immediately, directly and with out advance warning)
for manual opera tions
A maximum of 8 shortcuts can be saved on the Homescreen. As soon as this maximum is reached, "AddTo Home" in the start dialogs of methods, products, series and manual operations is deactivated.
14 Description of Functions
5Setup
This section tells you how to set up the refractometer acccordance with your requirements so that you can carry out the measurements.
The following summary shows the buttons available in Setup for the various setting options:

Adjustment setsAdjustments & Tests Test sets

Hardware
User settings
Global settings
Maintenance & Service
Cell Automation External instruments Peripherals Sensors Auxiliary instruments
Language Screen Beep Shortcuts Keyboards
System User management Analysis and resources behavior Physical Properties
TablesTables & Values Auxiliary values
MT service Import / Export Add external cell Reset to factory settings Firmware Update Hardware / Firmware summary Cell Touch screen adjustment Export of adjustments / Tests / Measurements
The "Expired Resources" button
The Expired Resources button is located on the setup overview screen. It provides you with a summary of all expired resources, stating the: type, name and expiration date of the respective resource.
Expired resources are entered if the setting "Monitoring adjustment set/test set/auxiliary value" is activated dur ing setup.
Below you will find a detailed description of the setting options available in setup:
5.1Adjustments & Tests
Navigation: Home > Setup > Adjustments & Tests
Adjustment and test sets can be administered as described below. You can create new sets and delete existing ones (when deleting you receive a prompt with the option to cancel). Before an adjustment or test can be per formed, an adjustment or test set must be defined.
A maximum of six different adjustment sets or test sets respectively can be entered in the set list.
Before an adjustment or test can be performed, an adjustment or test set must be defined.

15Setup

5.1.1Adjustment Sets

Via the button Adjustment sets you will obtain a list of the defined sets. The default set "Air&Water20.00C" is always available and cannot be deleted. When you click the sets, you obtain more detailed information about the individual sets. The parameters "Adjustment mode", "Temperature" and "Set name" are displayed.

Creating adjustment sets You can create your own adjustment sets via the button New. The Adjustment set parameters dialog opens.
Parameters Description Displayed if
Adjustment mode Defines with which standard the adjustment will be performed.
Temperature Defines at which temperature the adjustment will be performed. ­Set name The set is uniquely identified in the system via this name. -
Delete adjustment sets
1 To delete a created set, click the desired entry in the Adjustment sets dialog.
2 In the Adjustment set parameters dialog, click the button Delete.
When a set is deleted, the set history will also be deleted. Methods that refer to the deleted set are no longer executable.
5.1.2Test Sets
-
Set "Standards" requires a valid adjustment with "Air&Water" at this temperature. (set "Air&Standard" and set "Water&Standard" are not available for refractometers).

Via the button Test sets you receive a list of the defined sets. The default set "Water20.00C" is always available and cannot be deleted. When you click the test sets, you obtain more detailed information about the individual sets. The parameters "Test mode", "Temperature", "Set name" and depending on the mode "Standard name" are displayed.

Creating test sets You can create your own test sets via the button New. The Test set parameters dialog is opened.
Parameters Description Displayed if
Test mode Defines with which standard the test will be performed. ­Temperature Defines at which temperature the test will be performed. ­Standard name Description of the standards "Test mode" = "Stan
Set name The set is uniquely identified in the system via this name. -
Delete test sets
1 To delete a created set, click the desired entry in the Test sets dialog.
2 In the Test set parameters dialog click the button Delete.
When a set is deleted, the set history will also be deleted. Methods that relate to the deleted set are no longer executable.
5.2Hardware
Navigation: Setup>Hardware
In this dialog window you can configure all the hardware components connected to the meter. These include:
Cell
Automation
External instruments
Peripherals
Sensors
Auxiliary instruments
dard" or "Brix standard"
16 Setup
5.2.1Cell
Navigation: Home > Setup > Hardware > Cell
The measuring cell is connected via an internal interface in the compact device. The instrument can be extend ed to create a 2–cell instrument (see also "Setup: Maintenance & Service > Add external cell (page33)").

– Touch the button Cell in the Hardware dialog.

Parameters Description Displayed if
Cell Name of the measuring cell ­Type Measuring cell type ­Status Shows whether the cell is connected. ­Serial number Serial number of the cell -
If an adjustment has been performed, an entry in the adjustment set appears in the setup of the corresponding cell. If you touch the entry, the data of the most recent adjustment that was executed with the set will be dis played.
Parameters Description Displayed if
Adjustment set By touching the set entry, you obtain the values for the most
Monitoring adjustment set
Adjustment inter val
Reminder Before the set expires, a message appears indicating that the
Days before expi ration
If a test has been performed, an entry in the test set appears in the setup of the corresponding cell. If you touch the entry, the data of the most recent adjustment that were executed with the set will be displayed.
Parameters Description Displayed if
Test set By touching the set you go to the values for the most recent test
Monitoring test set
Test interval Service life of test "Monitoring test set"
Reminder Before the set expires, a message appears indicating that the
Days before expi ration
Via the button History you go to a list that contains a maximum of ten adjustment or test entries for the selected set. In the process the measured refractive index values are displayed (in the adjustment mode air and water only the value for water). The history can be displayed in graphic form via the button Graph. You can view the history data by touching an entry.
With single cell instruments you go directly to the list with the cell parameters
In the case of two cell instruments the connected cells are listed. You have to touch one of the displayed cells so that the Cell parameters dialog opens.
The Cell parameters dialog is opened.
-
recent adjustment of this set. Monitoring of the service life of the adjustment can be activated.
The process, if the service life of a set has expired, is defined in the method function "Measure".
Validity of adjustment "Monitoring adjustment
set" activated "Monitoring adjustment
adjustment or test has expired. Specifies the number of days after which a warning is triggered. "Reminder" activated
for this set. Monitoring of the service life of the test can be activated. The
process, if the service life of a set has expired, is defined in the method function "Measure".
adjustment or test has expired. Specifies the number of days after which a warning is triggered. "Reminder" activated
set / test set" activated
activated "Monitoring adjustment
set / test set" activated
-
-
-
5.2.2Automation
Navigation: Home > Setup > Hardware > Automation
The automation units that can be installed are listed below:
17Setup
DryPal (drying pump)
FillPal (sample pump)
SC1 (automation unit for a sample)
SC30 (automation unit for 30 samples)
InMotion (automation unit for up to 303 samples)
The peripheral devices have an automatic PnP (Plug and Play) – identification. They can also be manually cre ated in the setup.
Via the button Automation in the Hardware dialog, you access a list with defined automation units. By touch ing a list entry, you obtain more detailed information about the corresponding automation unit.
Parameters Description Displayed if
Type Instrument Type ­Power purge unit The external diaphragm (Optional) has a much higher output
than the diaphragm pump integrated in the SC1/SC30. The pump
"Type" = "SC1"/ "SC30"
is actually adjusted at the SC1/SC30.
Heating option This is where the heating option is activated. The temperature is
directly set on the heating unit.
Limit sensor Optical sensor that detects the sample and prevents it leaving the
heated area of the heating option (info field).
External drain valve
Located in the heating unit. Prevents the samples flowing back after filling process (info field).
"Type" = "SC1"/ "SC30"
"Type" = "SC1"/ "SC30"
"Type" = "SC1"/"SC30"
Solvent 1 Specifies the solvent at the connection Rinse 1 "Type" = "SC1"/
"SC30"
Solvent 2 Specifies the solvent at the connection Rinse 2 "Type" = "SC1"/
"SC30"
Speed "Low" Value for speed level "Low" of the FillPal or InMotion autosampler
(in % of the maximum speed).
Speed "Medium" Value for speed level "Medium" of the FillPal or InMotion
autosampler (in % of the maximum speed).
Speed "High" Value for speed level "High" of the FillPal or InMotion autosam
pler (in % of the maximum speed).
Rate at 100% Value for max. possible pump output (depending on the installed
"Type" = "FillPal" / "InMotion"
"Type" = "FillPal" / "InMotion"
"Type" = "FillPal" / "InMotion"
"Type" = "InMotion"
pump which is defined as sampling pump).
Stirrer output Specifies the pump/stirrer port on the InMotion where your stirrer
"Type" = "InMotion"
is connected.
Sampling pump output
PowerShower output
Specifies the pump/stirrer port on the InMotion where your sam pling pump is connected.
Specifies the pump/stirrer port on the InMotion where your pump for PowerShower is connected.
"Type" = "InMotion"
"Type" = "InMotion"
Rate Value for required pump output for PowerShower. "Type" = "InMotion" Beaker height Specifies the height of the used beakers. "Type" = "InMotion"
A distinction should be made between the following two cases in the installation of instruments:
1.
There is still no setup entry available in the unit (delivery state). A new entry has been generated and the parameters automatically generated.
2.
A setup entry was previously manually created in the instrument: The PnP parameters are automatically entered, the remaining parameters, previously edited by the user, remain unchanged.
18 Setup
When a PnP instrument is unplugged, the status changes to "not installed".
Below is described how you can administer the various devices in setup. This includes the creation of devices or changing the parameters in the setup.
Create automation units
So that methods for the application of automation units can be created and configured without the automation unit being connected, these units must be created in the setup via the button New.
Delete automation unit
It is not possible to delete the entry of a connected device.
If you want to delete an entry for a unit that is not connected from the list, a message with the termination option appears indicating that after deletion the methods that use the external instruments will no longer be exe cutable.
5.2.3External Instruments
Navigation: Home > Setup > Hardware > External instruments
The external instruments that can be installed are listed below. They all have automatic PnP – identification:
METTLER TOLEDO S20 – SevenEasy™ pH
METTLER TOLEDO S30 – SevenEasy™ Conductivity
METTLER TOLEDO S220 - SevenCompact™ pH/Ion
METTLER TOLEDO S220 - SevenCompact™ Conductivity
Lovibond colorimeters PFX880/ PFXi880, PFX950/ PFXi950 and PFX995/ PFXi995 series, Tintometer
Minolta colorimeters CM-5/ CR-5
The Lovibond colorimeters can be connected to the USB interfaces. For this purpose the USB – RS232 adapter is required (contained in the connection kit).
For each device type only one entry is possible.
Install devices
A distinction should be made between the following two cases in the installation of instruments:
1.
There is still no setup entry available in the unit (delivery state). A new entry has been generated and the parameters automatically generated.
2.
A setup entry was previously manually created in the instrument: The PnP parameters are automatically entered, the remaining parameters, previously edited by the user, remain unchanged.
When a PnP instrument is unplugged, the status changes to "not installed".

Via the button External instruments in the Hardware dialog, you can obtain the list of instruments. By touching a list entry, you obtain more detailed information about the corresponding instrument.

Create device
So that methods can be created and configured with external instruments, without having to connect the instru ment, the setup entries of the external instruments can be manually created via the button New.
The pH and conductivity meters can be assigned names individually using the "Instrument" parameter.
Delete devices
It is not possible to delete the entry of a connected device.
If you want to delete an entry for a unit that is not connected from the list, a message with the termination option appears indicating that after deletion the methods that use the external instruments will no longer be exe cutable.
SevenEasy™ / SevenCompact™ pH/ion and conductivity meters
1 Select the following for the SevenEasy™ and SevenCompact™ measuring instruments (also refer to the
operating instructions for the measuring instrument concerned):
Manual end point measured value acquisition, set the unit [pH]" ([mV] not supported).
2 For the SevenCompact pH and conductivity meters, the following must also be selected (also refer to the rel
evant operating instructions):
an interval time of 1s (in menu>Interval measurements), interface=printer (in menu>Data transmission), "Send data to interface" (in "Type of data transmission").
19Setup
3 For temperature compensation, set the unit [°C] at the measuring instruments. Two types of temperature
compensation exist:
- ATC (automatic temperature compensation) with a temperature sensor connected.
- MTC (manual temperature compensation) with no temperature sensor connected. Note: The pH or conductivity meters switch automatically, depending on the arrangement. If the temperature unit at the pH meter or conductivity meter is set to [°C], the results "TpH" and "TCond" will be converted to the temperature unit set in the instrument, i.e. either [°C] or [°F] . By contrast, if [°F] is set at the pH or conductivity meter, no result will be calculated. "--" will be output in the results field.
The conductivity units are [mS/cm]/ [μS/cm] (depending on the range) or [mS/m][μS/m] (depending on the range). However, the results are expressed in [μS/cm]. If you wish to have the result displayed in [mS/cm] or [mS/m][μS/m], it will be necessary to recalculate the result using the "Calculation" method function.
Colorimeter
- Lovibond
The Lovibond colorimeter can be used for extinction or transmission measurements. The wavelengths are out put in 5nm (in the range 420 – 710nm).
A color measurement may take longer than 25s.
- Minolta colorimeters CM-5 / CR-5
The Minolta colorimeters can be used for transmission measurements (liquids). Transmission data are supplied in 10nm increments within the spectral range 360 – 740nm. A color measurement takes approximately 1s. The measurements involve high-energy light flashes. For this reason, color scales are not offered for reflection measurements in the crude oil industry.
The CR-5 colorimeter does not supply any spectral data.
Calibration is possible directly at the colorimeter only. The USB link must be disconnected.
"Illuminant" and "Observer" can be set in the "Configuration" method function. For more information, see "Methods and products: Method functions> Configuration (page42)".
- Color scales Navigation: Home > Setup > Hardware > External instruments > Parameters
– Click on "Colorimeter".
The "External instrument parameters" dialog appears.
If a device has been connected, a list of color scales is displayed via the Color scales button, that were avail able in the most recently connected device. If the device has been manually configured and no device has yet been connected, then the list is empty.
See also
Continuous run (page71)
5.2.4Peripherals
Navigation: Home > Setup > Hardware > Peripherals

In the dialog Peripherals, the following devices and settings can be configured:

Barcode reader
USB stick
Fingerprint reader
Printer
Personal Computer (PC) – settings
Network settings
Network storage
20 Setup
Below is described how you can administer the various devices in setup. This includes the creation of devices or changing the parameters in the setup.
Barcode reader
Barcode readers have Plug&Play (PnP) recognition and can be installed via the USB interface.
The following barcode readers can be installed:
Handheld readers
Built-in readers
You can create the barcode readers via New. A maximum of one entry can be created for both types.
Parameters Description Displayed if
Type Instrument Type ­Serial number Serial number of the corresponding device ­Barcode Informa
tion
Sample ID: on the barcode there is only the sample ID. Method ID: on the barcode there is the method ID (with this a
-
saved method can be selected during scanning). Product ID: on the barcode there is the product ID (with this a saved product can be selected during scanning). Sample ID/Method ID: on the barcode there are the sample ID and method ID (with this a saved method can be selected during scanning). Sample ID/Product ID: on the barcode there are the sample ID and product ID (with this a saved product can be selected during scanning).
Start pos. sample ID
Number of char acters
Start pos. method ID
Number of char acters
Start pos. prod uct ID
Number of char acters
Immediate start If this parameter has been activated, when a task is started with
Start position of the sample ID on the barcode "Barcode information"
with sample ID
Length of the sample ID on the barcode "Barcode information"
with sample ID
Start position of the method ID on the barcode "Barcode information"
with method ID
Length of the method ID on the barcode "Barcode information"
with method ID
Start position of the product ID on the barcode "Barcode information"
with product ID
Length of the product ID on the barcode "Barcode information"
with product ID
"Type" = "Handheld the barcode reader, the Start analysis dialog is skipped and the task is started immediately.
reader",
"Barcode information"
with method ID, product
ID or sample ID
Example of barcode with sample ID and method ID (161218522). (Sample ID=1612 and method ID 18522).
Start pos. sample ID 1
Number of characters 4
Start pos. method ID 5
Number of characters 5
USB stick
Commercially available USB sticks from USB Version 1.1 are supported.
Fingerprint reader
You can use a fingerprint reader to authenticate users on the titrator. In order to do this, the fingerprint reader must be activated on the titrator. The following parameters are available for this:
Parameters Description Displayed if
Activate finger
Activates the connected fingerprint reader. -
print reader
21Setup
Printer (and USB-RS232 data export)
Date can be printed with the USB-P25 (strip printer) or with a network printer.
Data can be exported using the USB data export box. The data are exported in the XML (UTF-8) formats.
For data export to an RS interface, you need the USB RS232 adapter (USB data export box).
As desired, data can be printed either as a summary or in a user defined format (via "Print Mode" parameter "Report" method function; also see "Methods and products: Methods> Method functions> Report")
For printing/USB data export, the following should be taken into consideration:
In the "Report" method function, the "Print / USB-RS232 data export" parameter must be selected.
You can define the following parameters via the Printer button (Home>Setup>Hardware>Peripherals):
Parameters Description Displayed if
Printer type USB compact printer
­The USB compact printer does not support all languages. This printer can only print out a limited quantity of analysis data and results.
USB-RS232 data export
For the RS-232 data export, the data is transmitted regardless of the selected language. Only a limited quantity of data and results can be exported.
Network printer
Every connected printer in your local network that supports HP PLC 3 or Epson ESC/P 2 can be used.
Status Indicates whether the selected printer type is installed (info field). Only for printer type
"USB compact printer" and "USB-RS232 data export"
Baud rate The baud rate for data transmission via the USB-RS232 interface. "Printer type"= "USB-
RS232 data export"
Data bit The number of data bits is displayed (info field). "Printer type"= "USB-
RS232 data export"
Stop bit The number of stop bits is displayed (info field). "Printer type"= "USB-
RS232 data export"
Parity Defines the parity protocol. ("Even", "Odd", "None") . "Printer type"= "USB-
RS232 data export"
Handshake Data transmission via the RS-232 interface ("Xon-Xoff", "None"). "Printer type"= "USB-
RS232 data export"
Type Choose the printer protocol. "Printer type"= "Net
work printer"
Network name Define your local network name here. "Printer type"= "Net
work printer"
Port number Define your local port number here. "Printer type"= "Net
work printer"
Paper size Choose the print-format of the report (A4 / Letter). "Printer type"= "Net
work printer"
22 Setup
PC settings
Only one PC connected per measuring instrument can be present at one time. You can select if you wish to set up a connection to the laboratory program "LabX".
Parameters Description Displayed if
Set up connec tion to LabX at
If this parameter is activated, a connection to LabX will be estab lished on startup.
-
start up
Changes to the PC settings are not implemented until after the measuring instrument is rebooted.
Network settings
You can define the following parameters for the network – settings:
Parameters Description Displayed if
Type Type of network connection ­Obtain IP
address auto matically
IP Address IP address of the instrument ­Subnet mask Subnet mask of the device ­Standard gate
way
The device automatically reboots after any change to the network settings.
Network storage
You can define the following parameters for a network – storage:
Transfer via Method for transferring data
Server PC or server name. Users should have read-write access. Maxi
Share name Name of the share folder ­User name Define your own user name. ­Domain Domain name for the user. ­Password Password for network share. ­Target folder Name of target folder for PDFs or results.
First folder level Can optional be used to sort data. (None | User name | Instru
Second folder level
If this device has been activated, the device automatically obtains an IP address.
Standard gateway of the device "Type" = "Ethernet"
Default: Network share
mum 60 alphanumeric characters.
Arbitrary IP address or name of server (for network share, only server name allowed).
A pdf_export folder and a results folder are automatically creat ed in the target folder when specified. If no target folder is speci fied the folders are created in the root folder.
ment ID | Date | Method ID | Analysis comment)
Default: None
Can optional be used to sort data. (None | User name | Instru ment ID | Date | Method ID | Analysis comment)
Default: None
"Type" = "Ethernet"
-
-
-
-
"First level folder" = activated
5.2.5Sensors
Navigation: Home > Setup > Hardware > Sensors
Sensors can be activated in the setup. The following sensors can be connected:
ErgoSens: Infrared sensor for contactless start of measurement (see "Analysis sequence: Start of analyses").
Before you can use ErgoSens, it has to be activated:
Parameters Description Displayed if
Activate ErgoSens
WasteSens or LevelSens: Level sensors for waste bottle.
It is determined whether the maximum filling level of the waste bottle is reached. If the maximum filling level has been reached, a message appears prompting the operator to empty the waste bottle. The task is then inter rupted.
Before you can use WasteSens or LevelSens, it has to be activated:
Activates the ErgoSens -
23Setup
Parameters Description Displayed if
Activate WasteSens/Lev elSens
If you connect the density - module, you can use an air pressure sensor:
AtmoSens: Atmospheric pressure sensor for measuring the absolute air pressure.
If an AtmoSens is connected, the atmospheric pressure (if required) is measured with the AtmoSens. If no AtmoSens is connected, the air pressure is read out from the current value from Home > Setup > Global
settings> Physical properties.
Parameters Description Displayed if
Verify AtmoSens availability
activates the WasteSens or LevelSens -
If this parameter is activated, the use of the AtmoSens is enforced for tasks which require the pressure.
5.2.6Auxiliary instruments
Navigation: Setup>Hardware>Auxiliary Instruments
Auxiliary instruments can be any instruments that access the 24V output or USB-RS232 adapter of the measur ing instrument and that are to be used in a method.
An auxiliary instrument that accesses the 24V output is switched on for a predefined period and then switched off again via the corresponding command. The instruments are controlled via the "Auxiliary instrument" method function.
Auxiliary instruments form part of a method, while peripheral devices are classified as input/output devices (printers, barcode readers etc.), which do not have direct access to methods.
Starting from the auxiliary instrument list, you can add new auxiliary instruments or select existing ones or modify their parameters. In addition, the selected auxiliary instruments can be deleted.
A maximum of 30 auxiliary instruments can be entered.

Choose the New button in the Auxiliary Instruments dialog window to open the Auxiliary Instrument Para meters dialog.

-
1 Before a new auxiliary instrument can be added, you must first use the "Control type" parameter to select
the manner in which the auxiliary instrument is to be controlled:. The following values are available for "Control type":
24V output
USB-RS-232
2 You can assign a name of your choice to the auxiliary instrument.
The parameters for the control types are listed below:
24V output
Parameters Description Displayed if
Control type Indicates which port on the measuring instrument is to be used
for the auxiliary instrument.
Name A name of your choice. -
USB-RS-232 interface
Parameters Description Displayed if
Control type Indicates which port on the measuring instrument is to be used
for the auxiliary instrument.
Name A name of your choice. ­Adapter Defines which adapter is used. Maximum 2 auxiliary instruments
of type USB-RS-232 can be used in the same method (by using adapter 1 and 2)
Baud rate The baud rate for data transmission via the RS-232 interface of
the adapter.
Data bit Defines the number of data bits. "Control type" = "USB-
"Control type" = "USB­RS232"
"Control type" = "USB­RS232"
RS232"
-
-
24 Setup
Stop bit Defines the number of stop bits. (2 stop bits can only be selected
Parity Defines the parity protocol. "Control type" = "USB-
Handshake Data transmission via the RS-232 interface of the adapter. "Control type" = "USB-
A suitable adapter (USB data export box) is required for the USB-RS-232 connection.
5.3User settings
Navigation: > >
These settings contains the options that can be made specifically for each currently logged in user.
You can configure the language, the screen settings (for the touchscreen), the layout of the alphanumeric and numeric keyboard, the use of beeps, and shortcuts for each user.
5.3.1Language
Navigation: Home > Setup > User settings > Language
The following languages are available:
German
English
French
Italian
Polish
Portuguese
Spanish
Chinese
Russian
if 7 data bits are also selected at the same time.)
"Control type" = "USB­RS232"
RS232"
RS232"
The language can be defined both for the operation of the terminal as well as for the protocols that are to be printed out from a printer.
Parameters Description Displayed if
Touchscreen Language of the user interface ­Report Language of the printout* -
*
Chinese can not be selected as language for reports
5.3.2Screen
Navigation: Home > Setup > User – settings > Screen
The following parameters can be set in the screen settings:
Parameters Description Displayed if
Primary color Color of the user interface ­Brightness Brightness of the display ­Button shape Shape of buttons on the touchscreen ­Screen saver Activates screen saver ­Wait time Time for the display of the screen saver Activates screen saver
5.3.3Beep
Navigation: Home > Setup > User settings> Beep
The audio signal is set specifically for each user. You can activate the audio signal in the Audio-Signal set tings dialog.
Parameters Description Displayed if
25Setup
5.3.4Shortcuts
Navigation: Home > Setup > User – settings > Shortcuts
In this dialog, every user has the opportunity to administer his or her own selected shortcuts. The list of all shortcuts for the logged in user can be viewed. Individual shortcuts can be selected and deleted. You can determine the following parameters:
Parameters Description Displayed if
Description This text is the name of the shortcut that is displayed in the
Immediate start When the shortcut is pressed, the display switches directly over
Homescreen Position
5.3.5Keyboards
Navigation: Home > Setup > User – settings > Keyboards
In the Keyboard settings dialog, you can define the layout for the alphanumeric and the numeric input fields. The following settings are available:
Parameters Description Displayed if
ABC keyboard Defines the layout of the alphanumeric input field. ­123 keyboard Defines the arrangement of the keys for the numeric input field. -
­Homescreen.
­to the online screen without opening the Start analysis dialog.
Selection of the position on the Homescreen -
5.4Global Settings
Navigation: Home > Setup > Global setting

In the Global Settings, you can make general settings to the measuring instrument that apply to all users of the instrument. The settings in this dialog can only be changed by users with the appropriate authorizations.

Global settings include:
The system settings

User management for creating user accounts and assigning rights.

The settings for Analysis and resources behavior.
Physical properties for defining the "Temperature unit" and the air pressure.
5.4.1System
Navigation: Home > Setup > Global settings > System
In the System settings dialog the following buttons are available:
Identification:In this menu you can give the measuring instrument a freely definable ID consisting of at least four characters.
In addition the following information is displayed: Device, serial number and firmware version.
Date / time:You can define the format used to display the date and time and set the device date and time.
Data storage: If this parameter is activated, all results that are saved under Home > Results are deleted (only when device is shut down).
5.4.2User Management
Navigation: Home > Setup > Global settings > User management
By means of the User management dialog you can administer users and account policies for the instrument. A maximum of 30 different users can be defined, from which only maximum one may be logged in with the instrument (1 user operation).
User accounts can be deleted and edited.
There is a default user with the user name "Administrator" (user group: Administrators). This cannot be deleted.
26 Setup
5.4.2.1Users
Navigation: Home > Setup > Global settings > User management > User
Via the button New in the User dialog you can open the User parameters of the dialog. It is possible to define new users here.
"User management" is the only area in which users with administrator rights are able to edit settings and, there fore, create or make changes to users.
Create a new user in the User parameters dialog as follows:
1 Define a user name.
2 Assign the user to a group. The following user groups are available:
Administrator Expert Technician Operator
For information on the rights of these user groups, see "Setup (Setup: Global settings> User management> User groups (page27)".
The following parameters are available in the User parameters dialog:
Parameters Description Displayed if
User name The user is uniquely identified in the system via the name. ­Full Name Complete name of the user. ­User Group Selection of the user group for the user. Depending on the user
Reset password Resets the password for the user to "123456". Activates "Enforce pass
Lock user Locks the user. Activates "Enforce pass
Enforce pass word change
Created by The administrator logged in at time of creation ­Created on Date of creation and time (info field) ­Modified by The administrator logged in at time of creation ­Modified at Change date and time (info field) -
-
group the user has various rights.
word/fingerprint"
word/fingerprint"
With the next login the entry of a new password will be enforced. Activates "Enforce pass
word/fingerprint"
5.4.2.2Account Policy
Navigation: Home > Setup > Global setting > User management > Account policies
The following parameters can be edited in the dialog:
Parameters Description Displayed if
Enforce pass word/fingerprint
Min. password length
If the parameter "Enforce password" is deactivated, then the instrument starts directly, i.e. not via the login screen (only if one user is defined – corresponds to the factory settings). If several users are defined, the user name can be selected in the login screen from a list.
5.4.2.3User Groups
Each user is assigned to a user group. The following four user groups (with decreasing rights going down ward) should be distinguished:
Administrator
Expert
If this parameter has been activated, you can only log in by entering the password (or via the fingerprint reader, if the para meter "Activate fingerprint reader" is activated, in Home >
Setup > Hardware > Peripherals > Fingerprint reader).
The minimum length of the user passwords Activates "Enforce pass
word/fingerprint"
-
27Setup
Technician
Operator
The following table presents the user rights that are assigned to the corresponding user group:
List of user rights
Right Minimum required user group
Start analysis via Shortcuts. Operator Start analysis via Start button in the Homescreen Operator Start analysis with ErgoSens Operator Start analysis from method editor (for method type:
Technician measurement, cleaning and test), product and series editor (for method type: measurement, cleaning, test and adjustment)
Adding/deleting items for series in the Start analysis
Technician dialog (does not apply for adjustment-type methods)
Executing manual operations Technician Editing shortcuts
Technician (Setup>User settings)
Including and excluding results Technician Deleting results Expert Adding/deleting items for series in the Start analysis
Expert dialog (also applies for adjustment-type methods)
Setting the language for touchscreen and reports Expert Start analysis from method editor (for method type:
Expert measurement, cleaning, test and adjustment)
Editing methods, products, series Expert Editing "Adjustments & Tests"
Expert (Home > Setup )
Editing "Hardware"
Expert (Home > Setup)
Editing "Analysis and resources behavior"
Expert (Home > Setup > Global setting)
Editing "Tables & Values"
Expert (Home > Setup)
Setting the language for reports Expert Editing "Physical properties"
Expert (Home > Setup > Global setting)
Running "Maintenance & Service" in Home >
Setup, except:
Importing/exporting user management and memo
Expert
ry copy
Running "Maintenance & Service" (Home > Setup), including:
Importing/exporting user management and memo ry copy
Editing "User management" (Home > Setup > Global setting)
5.4.3Analysis and Resources Behavior
Navigation: Home > Setup > Global settings > Analysis and resources behavior.
The following settings can be defined:
28 Setup
Administrator
Administrator
Parameters Description Displayed if
Show required resources at start
Confirm end of the analysis
Reset statistics if sample ID changes
Verify USB stick availability
Verify printer availability
Information on identification of PnP resources
Verify network storage availabil ity
If this parameter is activated, all resources are shown during startup that are required for performance of the analysis. Note: If a required resource is not available, the "Needed resources" dialog is also shown without this parameter having being activated.
If this parameter is activated, the OK button is shown at the end of the task. The task is not ended until you confirm with OK. This parameter is especially used if the result should remain on the online screen at the end of the measurement so that, for exam ple, it can be copied.
If this parameter is activated, the rolling statistics will be interrupt ed, if the sample ID differs from the previous analysis. If it is not activated, the rolling statistics will only be interrupted if the method ID (or product ID) differs from the previous analysis.
If this parameter is activated and a USB stick is used in the method, the task will be checked when the start is started to see if the USB stick is available. If none is available the task cannot be started.
If this parameter is activated and a printer is used in the method, the task will be checked when the task is started to see if the printer is available. If the printer is not available the task cannot be started (does not work for network printers).
If this parameter is activated, a message is shown when a PnP resource is detected.
If this parameter is activated and network storage is used in the method, the task will be checked when the task is started to see if a network storage is available. If no network storage is available, the task cannot be started.
-
-
-
-
-
-
-
5.4.4Physical Properties
Navigation: Home > Setup > Global setting > Physical properties

In the Physical properties dialog you can defined the parameters "Temperature unit" [oC] or [oF] and "Atmospheric pressure" in [hPa].

Parameters Description Displayed if
Temperature unit Temperature unit applicable for all ranges of the instrument
(global setting). Either Celsius or Fahrenheit can be selected. All inputs or outputs are in the temperature unit selected here. The setting is saved after a restart.
Atmospheric pressure
The atmospheric pressure is only required when using the density - module.
Input of atmospheric pressure that is used for calculating the nominal value of the air density. The atmospheric pressure is required for an adjustment or a test with air or a cell test. If the AtmoSens is connected, the reading will be displayed here.
5.5Tables & Values
Navigation: Home > Setup > Tables & values
The instrument has tables and auxiliary values that can be used for the calculation of results (see also "Results (page62)").
With measurements the raw data (e.g. temperature value and refractive index) are delivered. These results are fed into the relevant tables so that corresponding results can be displayed in the respective units.
Tables are always assigned to an application. The list can be sorted according to application.
There are two types of tables:
-
-
29Setup
METTLER TOLEDO tables: These tables are included in the factory settings, they are only listed and can be neither edited, viewed nor deleted.
User-defined tables: Tables are laid out in the form of a value table (X-Y). They can be created, edited and deleted.
Below is a description of how to create your own tables.
In addition to the tables, you can also administer the auxiliary values in setup. The auxiliary values can thereby be used in formulas. By means of the method function "Auxiliary value", an auxiliary value from any desired formula can be assigned raw data or calculated results.
See also
Results and Statistics (page62)
5.5.1Tables
Navigation: Home > Setup > Tables & values > Tables

By opening the Tables dialog you can see a list of available tables (METTLER TOLEDO and user-defined).

How to create tables in the setup is described below.
You can create a table with the New button.
Parameters Description Displayed if
Name Table name: The name uniquely identifies the table in the system.
Application Application area of table. Facilitates the sorting of the table list. ­Input value Table heading for the input value. ­Output value Table heading for the output value. ­Fit type Definition of the curve type for the calculation.
1 Save the new table.
2 Enter value pairs via New.
Parameters Description Displayed if
"Input value" Input value of the value pair (corresponding to the measuring cell
"Output value" Output value of the value pair (e.g. Brix) -
At least four value pairs must be entered and then saved via the Save button.
– Touch the button Graph. The fit function is graphically presented (the button Graph is only displayed if the
table has been saved). Note: The axis labeling corresponds to the specified "Input" – (x-axis) and "Output" parameters (y-axis).
The table has been successfully entered. The coefficients are calculated according to the selected function. There is a "delta" column that shows the deviation from the calculated value at the effective table point (only for the "Fit type" = "Polynomial"). This column is not shown when the table is being edited.
­The name is entered into the formula in the method function "Cal culation".
-
• Linear interpolation: Corresponds to a segmented curve.
• Lagrange interpolation: Lagrange fit via four points
• 1 order polynom
• 2 order polynom
• 3 order polynom
A dialog opens with the name of the new table.
­used, e.g. density or refractive index)
30 Setup
In the Table – parameters dialog, you can again remove the table via the Delete button.
A table must contain at least 4 and contain a maximum of 200 value pairs.
A maximum of 30 user-defined tables may be defined.
Tables cannot be deleted or modified when they are currently in use.
The table name must be unique.
5.5.2Auxiliary Values
Navigation: Home > Setup > Tables & Values> Auxiliary values
Via the New button, you open the Auxiliary values parameters dialog. In it you can use the following parame ters to define the auxiliary value:
Parameters Description Displayed if
Name Unique designation of the auxiliary value: The name is entered
Comment Short comment on the auxiliary value (e.g. about the unit) ­Value Numerical value ­Determination
method Determination
date Performed by Person who performed determination ­Monitoring auxil
iary value Interval Specifies the time period for monitoring of the auxiliary value. "Monitoring auxiliary
Expiry date Expiry date of auxiliary value (info field) "Monitoring auxiliary
Reminder Defines whether a reminder should be issued before expiry of the
Days before expi ration
A maximum of 100 auxiliary values can be saved in the instrument.
Auxiliary values cannot be deleted or modified when they are currently in use.
When an auxiliary value is assigned with the "Auxiliary value" method function, this is updated in the setup immediately after completion of the method function.
-
into the formula in the method function "Calculation".
Type of determination method (info field) -
Date on which determination was performed (info field) -
Specifies whether the auxiliary value is to be monitored. -
value" activated
value" activated "Monitoring auxiliary
auxiliary value.
value" activated
Specifies the number of days after which a warning is triggered. "Reminder" activated
5.6Maintenance & Service
Navigation: Home > Setup > Maintenance & Service
The following functions are available to you:

MT service

Import / Export
Add external cell
Reset to factory settings
Firmware
Update
Hardware / Firmware summary
Cell
One-point temperature alignment (only for density - module)
Whole-range temperature alignment
Fan check
Add manual adjustment
Export of adjustments / tests / measurements
5.6.1MT Service
Navigation: Home > Setup > Maintenance & Service> MT Service
31Setup
Via the button MT Service you can open the Last MT services dialog window.
In this dialog, you can view and print out a list of the most recent (max. 10) METTLER TOLEDO services. Under each date, the user name of the METTLER TOLEDO service technicians and the date and time of the service appointment are displayed. The most recently performed service always appears at the top of the list.
Via the Settings button in the Last MT services dialog window you can open the Service data dialog window, via which you can edit the service life of the last service due date. You can define whether a warning should be issued before the service life expires (requires administrator rights). You can define the following parameters:
"Service life" (in days) of the most recently performed service.
"Reminder": Defines whether a reminder should be issued before expiry of the service life of the service.
"Days before expiration" Number of days before expiration of the service life on which the instrument should issue a warning. The entered value must be smaller than the value for the service life (only appears if "Reminder" has been activated).
5.6.2Import / Export
Navigation: Home > Setup > Maintenance & Service> Import / Export
With the aid of this feature you can save the data on a UBS stick (export) or import data from a stick. The following can be imported or exported:
Single method
All methods
Single product
All products
Single series
All series
Single table (only affects user-specific tables)
All tables (only affects user-specific tables)
User management
Memory copy
The following rules apply for Import / Export:
1.
Import/Export is only possible when a USB stick is connected.
2.
Import/Export is only possible when no task is running.
3.
Products can only be imported if referenced methods are already available on the device required for the import.
4.
Series can only be imported if referenced methods or products of the series are already available on the device required for the import.

– In the Maintenance & Service dialog, open the Import / Export dialog window. In this dialog, you can

define the following parameters:
"Action": Data can be exported to a USB stick or imported back from a USB stick.
"Data": Data that is to be exported or imported can be selected.
Memory copy
You can create a memory copy from your data. This includes adjustment and test data, auxiliary values, short cuts and fingerprints, data relating to automation, peripherals and external instruments. When a firmware upgrade of the instrument has to be done, the instrument can be restored from the memory copy. The memory copy can also be imported to another instrument. In this case, the Test and Adjustment history are not imported and an adjustment of the instrument is required.
Uploading data from a backup copy results in the existing data being overwritten by the data in the memory copy. In this way you can restore the initial status.
32 Setup
Not contained in a memory copy are any saved results, default parameters for manual operations, internal tables, hardware for cell creation or parameters for added external cells.
When you import / export user management settings the entire user management settings with all users and their properties are exported or imported.
If the import is canceled, all previously imported users are deleted and only a default user (administrator) is created.
You must have administrator rights before you can create and reimport a backup copy.
5.6.3Add External Cell
Navigation: Home > Setup > Mainten. & Service > Add external cell
Perform the action "Add external cell". In the Cell dialog box, the cell type must be selected (Home > Setup
> Hardware > Cell).
When you connect the external cell (module), the status changes to "installed" and the effective serial number is entered.
If the external cell is removed, the entry in the cell list is retained. The status changes to "not installed". The seri al number and the cell type however continue to be displayed.
5.6.4Reset to Factory Settings
Navigation: Home > Setup > Maintenance & Service > Reset to factory settings

With the button Reset to factory settings you can reset your settings.

If you do this all created data, amendments, settings, setup entries or results are lost. Before you activate "Reset to factory settings" you should create a memory copy.
5.6.5Firmware
Navigation: Home > Setup > Maintenance & Service > Firmware
With the aid of this feature you can display a list showing the most recent firmware updates. The first entry in the list corresponds to initial operation. All list entries are saved with a date and FW version.
5.6.6Update
Navigation: Home > Setup > Maintenance & Service > Update

With the aid of this feature a firmware update of the instrument, device board and cells can be performed via a USB stick. After the button update is touched, all available and associated components that contain firmware are presented in a list (Instrument FW, cell FW, Device Board FW).

You must have administrator access rights.
- The instrument is in idle mode (no task is active).
- To update a cell, the cell must be connected. In the process each update must be performed separately for each connected cell that is to be updated.
1 Via the button update you can open the Firmware – List dialog. Select the relevant components from this
list.
2 Plug in the USB stick with the firmware update files into the instrument.
The instrument recognizes the USB stick and enables the update.
3 Start the update.
The instrument reboots and starts the update program via the USB stick. The update can be performed in the update program.
4 When the update program is terminated, the instrument reboots with the normal application.
Instrument update:
33Setup
All results, settings, setup entries, methods, products and individual modifications are lost during the instrument firmware update. On the other hand, the service history, the instrument firmware history and serial number are all retained.
Cell update:
Only updates from connected and recognized cells are offered, whose update file is on the USB stick.
With the cell update, no changes are made to the application program and to the individual settings and methods.
Device board update:
With the device board update, no changes are made to the application program and to the individual settings and methods.
5.6.7Hardware / Firmware summary
Navigation: Home > Setup > Maintenance & Service > Hardware / Firmware summary
You can view a list with all connected components along with the corresponding information on hardware or firmware. The list can be saved as a file on a USB stick. The file name contains the serial number of the device and the current date and time. Example: HW_SW_Info_5124560983_23_03_2010_08_23.cvs
5.6.8Board Tests
Navigation: Home > Setup > Maintenance & Service > Board tests
With this function, the interfaces of various boards can be tested:
Main board
Device board
External cell board, if an external cell is connected
The sequence for testing the boards is described below:

1 Via the button Board tests you can open the Boards dialog. In this dialog, buttons are provided, which

2 Touch the button for the desired board and you will receive an information list with the connections on the
3 Touch the Start button, to go to the list of test functions.
4 You can select another interface and proceed as described above or you can cancel the test with the Stop
5.6.9Cell
Navigation: Home > Setup > Maintenance & Service > Cell

Via the Cell you get to the service function of the cells.

This dialog window contains the following buttons:
Button Parameters
One-point temperature alignment Adjustment of the cell temperature to the set temperature. The
guide you to the corresponding board information for the connections.
board and the required resources that must be plugged into the corresponding connection.
- By touching a test, you change over to the corresponding test and this can then be started. Pop-ups guide you through the test process and show you any errors identified.
- The test result is also displayed in the interface list.
button.
When all interfaces for the selected board have been tested and all have the status "successfully passed", the results can be written onto the USB stick.
block thermistor is adjusted to the certified cell thermistor at the current set temperature.
The One-point temperature alignment dialog shows the selected cell, the set (Tset), the cell (Tcell) and the cell block temperature (Tblock).
34 Setup
Button Parameters
Whole-range temperature alignment Adjustment of the cell temperature to the set temperature across
the entire temperature range.
Note: The alignment takes about two hours.
The Whole range temperature alignment dialog shows the selected cell, the current values of the set (Tset), cell (Tcell), cell block- (Tblock) and the ambient temperature (Tambient).
The alignment can be started when the temperature in the cell is stable and the cell is dry and clean.
Note:
If there is an error message, the current values are saved on the display so that it can be seen at which temperature the error occurred.
With Stop the alignment can be interrupted at any time.
During alignment, the temperature per interval is saved in a file. This file can be saved on the USB stick.
Fan check The functional efficiency of the fans is tested.
The display is periodically updated. Note: The instrument is not switched off when there is an error display.
Add manual adjustment This function simulates an adjustment. It is used in order to per
form a measurement without previously adjusting the instrument.
Note: Only use for demonstration or method development processes!
5.6.10Export Adjustments / Tests / Measurements
Navigation: Home > Setup > Maintenance & Service > Export of adjustments /
Tests / Measurements
With the aid of this function, the adjustment, test and measurement history can be saved on the USB stick so that it can then be imported into an EXCEL file.
The histories can be viewed and exported.
The measurement history is only exported if the measured data agree with the current cell configuration (if the external cell type is changed, the measured data cannot be exported from the previous cell).
35Setup
6Methods and Products
Before you can perform a measurement you need a method. A method is an analysis program and consists of a sequence of method functions (some with method subfunctions), which are processed by the instrument in sequence.
Products are always linked to a method, and multiple products can be linked to the same method. It is therefore possible to work with only a few methods.
Using the Print button, you can generate a PDF file containing the methods or product parameters and product calculations.
The Print button is active only when a USB stick is connected.
How to create methods and products, how to access them and apply them is described below.
6.1Methods
A method basically consists of the configuration, sample feed, the actual measurement, cleaning, result calcu lation and the creation of a report. These steps are defined as functions. The individual functions in turn consist of parameters, whose values or names can be changed.
Method types
The instrument distinguishes between the following method types with different objectives:
MS (Measurement), for the performance of a measurement
ADJ (Adjustment), for an adjustment of the measuring cell(s)
TE (Test), for testing the measuring cell(s)
CL (Cleaning), for cleaning the measuring cell(s)

6.1.1Establishing Methods

When creating methods (max. number of methods: 30) you can make use of method templates with prede fined method types and hence structures. The corresponding parameters thereby contain suitable default val ues.
Select: Home > Methods / Products > Methods
1 Choose New to create a new method on the basis of a template.
2 Select the desired method type from the available templates. Now you can adjust this method to your own
specifications by adding or deleting method functions, or by adjusting the settings.
After a template has been chosen, the Configuration dialog opens. Here, according to the defined setup, you can select the cell mode, the cell, automation unit and the external devices (pH, conductivity and colorimeter) with which you wish to carry out the method.
Parameters Description Displayed if
Cell mode Defines the cell(s) ­Automation Defines the automation unit used by the method. The choice
available depends on the automation units defined in Home >
Setup > Hardware > Automation.
Note
-
If the filling and rinsing is performed with "FillPal" and drying with "DryPal", then "FillPal+DryPal" must be selected.
If the InMotion autosampler is used together with a DryPal in method type "Clean", both (InMotion and DryPal) must be selected.
Lid handling Defines whether CoverUp (optional lid handling system on InMo

36 Methods and Products

tion) is used in the method.
Automation unit"= "InMotion"
pH Defines whether the pH meter is used in the method. "Method type"= "Mea
surement" pH meter is defined in "Home > Setup >
Hardware > External instruments".
Conductivity Defines whether the conductivity meter is used in the method. "Method type"= "Mea
surement" Conductivity meter is defined in "Home >
Setup > Hardware > External instruments"
Color Defines whether the colorimeter is used in the method. "Method type"= "Mea
surement" Colorimeter is defined in "Home > Setup >
Hardware > External instruments".
Up to the automation unit, these settings cannot be retrospectively modified.
Color settings (Minolta only)
Illuminant The following illuminants can be selected:
Observer Viewing angle:
Activates "Illuminant" so that the illuminant type can be defined and "Observer" so that the angle can be entered.
A, C, D65, D50, ID65, ID50, F2 to F12
2°C or 10°C; default 2°C
-
"Color settings" (Minol ta only)" activated
"Color settings" (Minol ta only)" activated
The parameters "Color settings" (Minolta only), "Illuminant" and "Observer" are used only with Minolta col orimeters.
1 Confirm with OK.
2 In the "Title" method function, enter a new method ID. Afterwards, the new method will be stored under this
method ID.
3 Save the method.
You have now created a new method with its own ID. This method already has predefined method func tions. To change the method functions proceed as follows:
4 Select available method functions to modify their parameters in line with your requirements.
5 Choose Insert to add additional method functions to the method.
6 Now use the arrow-shaped "Insert" button to select the required position for the new method function in the
method. (You will only be able to insert the method functions that are allowed in the corresponding location based on the method syntax.) Note: It is a good idea to save at regular intervals.
7 From the list, select the method function that you want to insert.
8 Modify the individual parameters of the method function.
The new method function will be displayed in the method.
9 To delete a method function, select the function in question and then choose Delete.
The method function will disappear from the method. Note: Some essential method functions cannot be deleted. (The "Delete" button is not available in these cases.)
10 After inserting all desired method functions, you can store the new method in the device by choosing Save.
When creating a new method, follow the rules specified by the device. These are described in Chapter "Method Syntax – Rules for Establishing methods".
37Methods and Products
6.1.2Creating a Method Copy
You can copy an existing method by changing the ID of the method in the method function "Title". When the method is saved a new method is created.
6.1.3Modifying or Deleting Methods
Modifying a method
To modify a method, select: Home > Methods / Products > Methods
1 From the displayed list of methods, select the method that you want to modify.
2 As soon as the methods functions of the selected method appear on the screen, you can edit and save the
method.
Delete Method
It is simple to delete newly created methods. Select: Home > Methods.
1 Select the method that you want to delete.
2 Choose Delete method to delete the method from the memory.
Shortcuts, products and series that refer to the method will also be deleted.
6.1.4Method Syntax
The following rules describe in which point method functions can be inserted or deleted. Only the available method functions are displayed for insertion.
Rules for method type "Measurement" (MS)
The following method functions cannot be deleted: Title, configuration, sample, filling, measure and online display.
The method functions "Calculation" (R1, R2,….Rx) and "Temperature compensation" (TC1, TC2,…TCx) are indexed.
A maximum of 20 method functions "Calculation" can be inserted. The method function "Temperature compensation" can be inserted a maximum of five times, in each case immediately after the method function "Measure". Insertion of the "Report" method function is limited for the following method types:
- twice for "Measurement" and "Test"
- once for "Adjustment" and
- "Report" cannot be inserted for "Cleaning".
Method functions can be inserted in the method as follows:
38 Methods and Products
Rules for method type "Adjustment" (ADJ)
The following method functions cannot be deleted: Title, Configuration, sample, Filling, Measuring and adjustment.
The method functions are executed twice between "Sample" and "Adjustment".
The following method functions can be inserted in the method template:
Template Method function
Title
Instruction, Auxiliary instrument
Configuration
Instruction, Auxiliary instrument
Sample
Clean, Instruction, Auxiliary value, Wait, Auxiliary instrument, Line
rinse
Fill
Instruction, Auxiliary value, Instruction, Wait, Auxiliary instrument
Measure
Clean, Instruction, Auxiliary value, Wait, Auxiliary instrument, Line
rinse
Clean
Clean, Instruction, Auxiliary value, Wait, Auxiliary instrument, Line
rinse
Adjustment
Instruction, Auxiliary instrument
Report
Rules for method type "Test" (TE)
The following method functions cannot be deleted: Title, Configuration, Sample, Filling, Measure and Test.
The following method functions can be inserted in the method template:
Template Method function
Title
Instruction, Auxiliary instrument
Configuration
Instruction, Auxiliary instrument
Sample
Clean, Instruction, Auxiliary value, Wait, Auxiliary instrument, Line
rinse
Fill
Instruction, Auxiliary value, Wait, Auxiliary instrument
Measure
Report, Clean, Instruction, Auxiliary value, Wait, Auxiliary instrument,
PowerShower, Line rinse
Clean
Clean, Instruction, Auxiliary value, Wait, Auxiliary instrument
Test
39Methods and Products
Report, Clean, Instruction, Auxiliary value, Wait, Auxiliary instrument,
PowerShower, Line rinse
Clean
Report, Clean, Instruction, Auxiliary value, Wait, Auxiliary instrument,
PowerShower, Line rinse
Report
Report, Clean, Instruction, Auxiliary value, Wait, Auxiliary instrument,
PowerShower, Line rinse
The "Report" method function can be inserted a maximum of two times. There is no limit on the other method functions.
Rules for method type "Clean" (CL)
The following method functions cannot be deleted: Title and Configuration.
For the method type "Clean" at least a method function "Clean" or "Cell test" must be available.
The following method functions can be inserted in the method template:
Template Method function
Title
Instruction, Auxiliary instrument
Configuration
Clean, Instruction, Wait, Auxiliary instrument, Stir, Park, PowerShower,
Line rinse
Clean
Clean, Instruction, Wait, Auxiliary instrument, Stir, Park, PowerShower,
Line rinse
6.1.5Standard Data
The standard data are editable according to mode and standard.
Parameters Description Displayed if
Standard name Description of the standards ­Lot number Lot number / batch number of the standard Standard not air Certification date Certification date of the standard Standard not air Expiry date Std Expiry date of the standard Standard not air nD nominal Nominal refractive index value of standard Cell = R d nominal Nominal density of standard Cell = D Uncertainty Limits of error of standard – belongs to the nominal value Standard not air
6.1.6Condition
A logical condition can be defined to determine whether a particular method function should be executed based on a result (true or false). The condition is in the format of a formula. An empty condition is interpreted as true.
Formulas can be used in various method functions. In formulas you can enter the calculated results Rx, tem perature-compensated values (TCx), tables, auxiliary values and raw data.
Formula Editor Navigation: Home > Methods / Products > Methods > Method > Calculation
With the formula editor both numerical values and symbols can be entered. The symbols H, T and C are made available when you select the 1 button in the formula editor. If you touch H, the list of auxiliary values opens; T opens the table list; C opens the list of color scales for the colorimeters. For colorimeters, it is possible to define whether measurements should be carried out in extinction or transmission mode. You can enter raw data manually or via insert via the button "Proposal". You will find the raw data in the "Annex: Raw data".
40 Methods and Products
Mathematical Functions and Operators
The following mathematical functions and operators can be used in formulas:
Functions Comparison operators
Logarithm to the base 10 lg(x) equal to = Logarithm to the base e Ln(x) larger than > Exponential to base 10 pw(x) or
larger than or equal to > =
scientific notation Exponential to base e ex(x) smaller than < Square sq(x) smaller than or equal to <= Square root sr(x) x in the range of ... < x < ...
not equal to < >
Mathematical operators Logical operators
Addition + and AND Subtraction - or OR Multiplication * Division /
Logical operators are only permitted within formulas of the subfunction "Condition".
The use of mathematical operators (+, -, * and /) within a parenthetical expression is not possible.
Formula Syntax
- Brackets
There are three types of brackets that can be used in formulas:
Round brackets "( )": e.g. T[Table_name(d)] and mathematical operators ln(d)
Square brackets "[ ]"
- Table name T[Table_name()] , e.g. T[Brix_d_NBS(d)];
- Auxiliary value H[Auxiliary_value], e.g. H[Auxiliary_Value_1]
Curly brackets "{ }": Relation to the cell, e.g. d{DX}. if the cell is not specified, then the internal cell is used as default.
- Fixed abbreviations
The following abbreviations are defined:
RM Internal refractive index cell
DX Density - module RX Refractive index module R1 ... Rx Results from the method function "Calculation" TC1 … TCx Results from the method function "temperature compensation" A, B, C, D, x, y Coefficients (A-D) and variables (x,y) can be used in formulas, e.g. Ax+B.
- Syntax examples
Simple raw result: R1 = nD
Temperature compensation: R1 = TC1
Tables: R1 = T[Brix_nD_20C (nD)]
The following values can be entered for tables:
- Internal tables: d, dRaw, TC
- User-defined tables: All generated analysis data
- Density modules: d, dRaw, TC
- Internal refractive index cell: nD, TC
6.1.7Method Function
All method functions for the instrument are listed below:
41Methods and Products
6.1.7.1Title
The method function "Title" contains general method data.
Parameters Description Displayed if
Type Method type (measurement, adjustment, test, cleaning): Defined
Method ID The method is uniquely identified in the system via the method
Title Title of the method. ­Author Name of creator (info field) ­Created on Date of creation and time (info field) ­Modified at Change date and time (info field) ­Modified by User who made change (info field) ­Protect When this is activated, only the author or an administrator can
SOP On activation a text is displayed before the start. ­SOP-Text Displayed SOP text "SOP" activated
6.1.7.2Configuration
Used for making settings for cells, automation and external instruments. The settings in the method function "Configuration" affect the parameters of the following method functions.
Parameters Description Displayed if
Cell mode Display of cells to be used in the method (info field) ­Automation Defines the automation unit used by the method. The choice
by the selection of the method template during creation of the method (info field).
ID. Once the method has been saved, the method ID can no longer be changed. A change of the method ID results in the cre ation of a copy of the method.
edit or delete the method.
available depends on the automation units defined in Home >
Setup > Hardware > Automation.
Note
If the filling and rinsing is performed with "FillPal" and drying with "DryPal", then "FillPal+DryPal" must be selected.
If the InMotion autosampler is used together with a DryPal in method type "Clean", both (InMotion and DryPal) must be selected.
-
-
-
-
Lid handling Defines whether CoverUp (optional lid handling system on InMo
External instru ments
Color settings (Minolta only)
Illuminant The following illuminants can be selected:
Observer Viewing angle:
The parameters "Color settings" (Minolta only), "Illuminant" and "Observer" are used only with Minolta col orimeters.
6.1.7.3Sample
By means of the method function "Sample", sample data for the following method types are defined:
Method type: Measurement
Parameters Description Displayed if
Viscosity correc tion
Automation unit"=
tion) is used in the method. Display of external instruments (pH, conductivity, color) to be
used in the method is to be measured (info field) Activates "Illuminant" so that the illuminant type can be defined
and "Observer" so that the angle can be entered.
A, C, D65, D50, ID65, ID50, F2 to F12
2°C or 10°C; default 2°C
Activation of the viscosity correction Density - Module
"InMotion"
-
-
"Color settings" (Minol ta only)" activated
"Color settings" (Minol ta only)" activated
defined
42 Methods and Products
Viscosity ≤2000:
For samples with viscosity less than 2000 mPa*s >2000: For samples with viscosity greater than 2000 mPa*s
- Set value:
Density - Module con nected
Viscosity correction activated
With this selection, a known viscosity can be entered Viscosity Value Viscosity of the sample "Viscosity" = "Set value" Sample ID Sample ID can be entered; you can still change this ID at the
analysis start. Correction factor This factor (f) can be included in the calculations. It can be
"Method type" = "Mea surement"
­changed before performance of the analysis in the Start analysis dialog.
Comment A comment was entered before performance of the analysis in the
­Start analysis dialog.
Method type "Adjustment" / "Test" = The standards will be shown additionally.
Method type: Adjustment
Parameters Description Displayed if
Viscosity correc tion
Activation of the viscosity correction Density - Module
defined
Viscosity ≤2000 (info field) Density - Module
defined
Viscosity correction activated
Adjustment mode Controls the selection of the adjustment sets and the standard
­specific data. The selection is filtered according to the modes of the available sets in the Method > Setup > Adjustment
sets. The following adjustment modes are available:
- Air&Water (recommend settings)
- Air&Standard (only for density - module)
- Water&Standard (only for density - module)
- Standards
Adjustment set The adjustment set defines the standards and the temperature at
­which the adjustment is performed. With Air&Water the instru ment accesses its internally saved nominal value tables, with Standard the nominal value must be entered by the user. The selection of the sets is filtered via the adjustment mode. The adjustment set must be entered for each cell.
Adj. temperature The adjustment temperature is displayed (info field). ­Standard 1 ... 2 Shows the parameters of the standard ­Comment A comment was entered before performance of the analysis in the
­Start analysis dialog.
Method type "Adjustment" / "Test" = The standards will be shown additionally.
Please note: For using an adjustment set "Standards" in an RM/RX, a valid adjustment based on Air&Water must be stored in the instrument for the same temperature.
Method type: Test
Parameters Description Displayed if
Viscosity correc tion
Viscosity ≤2000:
Activation of the viscosity correction Density - Module
defined
Density module speci For samples with viscosity less than 2000 mPa*s >2000: For samples with viscosity higher than 2000 mPa*s.
fied and
viscosity correction acti
vated
43Methods and Products
Test mode Controls the selection of the test sets and the standard specific
Test set The test set defines the standards and the temperature at which
Test temperature The test temperature is displayed (info field). ­Standard Shows the parameters of the standard ­Comment A comment was entered before performance of the analysis in the
You can find detailed information on the standards in "Methods and products: Methods > Standard data (page40)".
6.1.7.4Fill
­data. The selection is filtered according to the modes of the avail able sets in the Home > Setup > Test sets. The following test modes are available:
- Air (Internal standard value table) (only for density - module)
- Water (internal standard value table)
- Default
- Brix standard
­the test is performed. The selection of the sets is filtered via the test mode. The test set must be entered for each cell.
­Start analysis dialog.
Method type "Adjustment" / "Test" = The standards will be shown additionally.
Sample feed via the automation unit specified in the method function "Configuration".
Parameters Description Displayed if
Prompt for sam ple
Speed Pump speed
Max. speed after [s]
Filling mode Defines the way in which the sample feed is to be stopped.
Fill time Filling time in [s] "Automation" =
Fill ratio Defines the percentage by which the cell is "overfilled". For exam
When activated a confirmation appears on completion of the method function "Fill", stating that the task is interrupted until confirmed.
"Automation" =
"SC1"/ "SC30":
• "Maximum": Sample feed with maximum pump speed
• "Reduced": Sample feed with speed set at the automation unit
• "Automatic": The pump speed changes according to the defined time period from "Reduced" to "Maximum".
"FillPal" / "InMotion":
• "High"
• "Medium"
• "Low" The corresponding pump speed can be set under Home >
Setup > Hardware > Automation .
The pump switches over to the maximum speed after this time "Automation" = "SC1"/
"Fixed duration": The sample feed is stopped after a certain time. "Automatic": The measuring signal is used to detect when the sample fluid has reached the measuring cell (not for "InMotion")
ple: After 10s pumping, the sample is recognized in the cell, the fill ratio is 150%, i.e. the pump switches off after 15s (feed dura tion +feed duration x("fill rate"-100)/100).
"FillPal"/ "SC1"/ "SC30" / "InMotion"
"SC30"
"Speed" = "Automatic"
"Automation" = "FillPal"/ "SC1"/ "SC30"/ "InMotion"
"FillPal"/ "SC1"/ "SC30" / "InMotion"
"Filling mode" = "Fixed duration"
"Automation" = "FillPal" / "SC1" / "SC30"
"Filling mode" = "Auto matic"
-
44 Methods and Products
Sample detectionatSelection of cell at which the sample detection is performed. "Automation" =
Max. fill duration Sample feed is stopped at the latest after this time, if no new
Pressure release Defines if the pressure shall be released after filling the cell(s). "Automation" = "SC1"/
6.1.7.5Measure
Via the method function "Measure", the measurement, data capture and error detection can be configured. It contains the following subfunctions that are described below: "Cell" "Measured value acquisition" "Error detection"
Subfunction Cell – method type: Measurement
Parameters Description Displayed if
Measurement temperature
Adjustment set Defines which adjustment set is used. ­Action when
expired
Adj. temperature The adjustment temperature is displayed (info field). ­Verify test set Defines that a method can be executed for the valid test set (the
Test set Selection of test set to be verified "Verify test set" activat
Action This parameter defines the behavior with an expired test set and
sample is detected in the cell.
The task is continued to completion, however the method func tions "Fill" and "Measure" are skipped.
Input of meas. temperature; the temperature can be defined per cell.
This parameter defines the behavior if an adjustment set expired. In Home > Setup > Cell the monitoring of the adjustment set can be set. Action:
• "Block": The method is no longer executable.
• "Warn: A message is shown at the method start. In spite of this, the method is performed. The analysis is labeled with the status "OK*".
• "None": No action
last test was passed with the corresponding test set).
whether the last test was valid or not. (in Home > Setup > Cell it is possible to configure moni toring of the test set):
• "Block": The method can no longer be run.
• "Warn": A notification is displayed when the method is started. The method is still executed despite this notification. The analysis is tagged with the status "OK*".
"FillPal"/ "SC1"/ "SC30"
"Filling mode" = "Auto matic"; 2-cell method
"Automation" = "FillPal"/ "SC1"/ "SC30"
"Filling mode" = "Auto matic"
"SC30"
-
-
-
ed "Verify test set" is
enabled
Subfunction cell – method type: Adjustment
Parameters Description Displayed if
Adj. temperature The adjustment temperature is displayed (info field). -
Subfunction cell – method type: Test
Parameters Description Displayed if
Test temperature The test temperature is displayed (info field). ­Adjustment set Defines which adjustment set is used. -
45Methods and Products
Action when expired
This parameter defines the behavior if an adjustment set expired. In Home > Setup > Cell the monitoring of the adjustment
-
set can be set. Action:
• "Block": The method is no longer executable.
• "Warn: A message is shown at the method start. In spite of this, the method is performed. The analysis is labeled with the status "OK*".
• "None": No action
Adj. temperature The adjustment temperature is displayed (info field). -
Subfunction measurement acquisition – method types: Measurement, adjustment and test
Parameters Description Displayed if
Wait time R The refraction index measurement begins after this time has
­elapsed. This wait time ensures that the prism is completely wet ted with liquid. In the case of multiple measurements, each mea surement is started after this wait time has elapsed.
Wait time Col The colorimetric measurement begins after this time has elapsed.
In the case of multiple measurements, only the first measurement is started after this wait time has elapsed. All other measure ments begin with no wait time.
End point Fixed duration:
External instrument = Colorimeter ("Configu ration" method tem plate: "Color" activated)
­The measurement is stopped after a certain time. This function is used to shorten the measurement time if high precision is not required. The measurement is stopped on achievement of the maximum measurement reliability at the latest.
Automatic:
The measurement is ended if the defined stability criteria have been achieved. (For the method types "Adjustment" and "Test" the "End point"= "Automatic" (info field))
Meas. duration Duration of measurement in [s] "End point" = "Fixed
duration"
Meas. reliability Setting of stability criteria for meas. value acquisition.
- Maximum: Highest reliability of measured
"End point" = "Automat ic"
- High
- Medium
- Minimum: Fastest measurement
Max. meas. duration
Maximum duration of measurement. If the measured value has not stabilized by this time, the measurement will be canceled
"End point" = "Automat
ic" with the status "Error". (depending on the stability criteria of the parameter "Meas. reliability").
You record the following measuring errors with the bubble check or a multiple measurement:
air bubbles in the measuring cell (Bubble Check)
Solvent residues in the measuring cell
Solid particles in the sample
With the Bubble Check variations in the measuring signal are analyzed.
With Multiple measurement, n measurements are performed. Between measurements, the sample is subject to continual movement. The standard deviation of the n measurements is added, and checked to see whether it is smaller than the maximum standard deviation (Max. SD) defined in the method.
Parameters Description Displayed if
Bubble Check When this is activated, the system checks whether any bubbles
Multiple mea surement
No. of measure ments
46 Methods and Products
-
are present in the cell. On activation several measurements are performed on the same
-
sample Input of number of measurements "Multiple measurement"
activated
Refill ratio Measurement for further movement of the sample between two
measurements. The refill ratio is calculated using the duration of the sample addition. For example: The addition duration is 10s and the refill ratio is 50%, i.e. pumping continues for 5seconds.
Max. SD Maximum permitted absolute standard deviation of the measure
ments for density, refractive index or color. When the deviation is exceeded the measurement is interrupted with the status "Error".
The maximum standard deviation must be entered in the function of the measuring cell type. The default value in the method is provided for an R40 cell. This should be adjusted for an R50 cell, e.g. 0.00002.
For the calculation of the standard deviation SD of the color results, the first color result in the formula of the "Calculation" method function with the smallest Rx index is used. If no color result is included in the calculation, "–" will be displayed for the standard deviation, i.e. SD is within Max. SD.
Single values (d1 to d10) are available as raw results and can be used in a calculation (i.e. for a 3-fold mea surement all 3 single values can be reported).
Repeat if failed If this is activated, the measurement is repeated if a bubble check
has failed or if "Max. SD" or "Max. Deviation" are exceeded. Note: If "Multiple measurement" is activated, all measurements are repeated.
For "Repeat if failed", no wait time is included.
No. of meas. points
Max. deviation The max. permissible absolute deviation of the measured values
Defines the number of measured points used for determining the mean value. Note: If multiple measurement is activated, only the measured values from the last measurement repetition are used.
for the pH or conductivity meter. If this value is exceeded, the measurement is interrupted with an "Error".
"Automation" = "FillPal"/ "SC1"/ "SC30"/ "InMotion"
"Repeat if failed" acti vated
"Multiple measurement" activated
Multiple measurement activated
"Bubble Check" activat ed, or "Multiple measurement" activated, or External instrument defined (pH meter, con ductivity meter, col orimeter)
pH or conductivity meter defined in method
pH or conductivity meter defined in method
Subfunction Error detection – method type: Adjustment / Test
If you are using a density - module, the following additional functions are available:
Parameters Description Displayed if
Bubble Check When this is activated, the system checks whether any bubbles
6.1.7.6Calculation
A result can be defined by means of the method function "Calculation". For this purpose any desired formula with raw data, tables or other calculated results can be entered (see also "Formula syntax and tables" and "Results (page62)"). Results are displayed on the online screen. They are saved after the final "Calculation" method function has completed. Results receive an index Rx, by which they can be referenced in formulas of other method functions. The calcu lation number (Rx) is increased (from R1 to R20) when the "Calculation" method function is inserted, indepen dently of the sequence of the method functions. By means of the button "Results proposals" you can select the predefined results with unit, coefficients and formulas. You will find the list of predefined results in the Annex.
Parameters Description Displayed if
Name Name of the result generated in the calculation ­Unit Unit of result ­Formula Formula input for the calculation. -
-
are present in the cell.
47Methods and Products
Coefficients Specifies the coefficients that can be used in the formula.
Note: For powers to the base 10 the "e"-convention may be used (e.g. 1.2e-4 --> 0.00012).
Variable x/ Vari abley
A / B / C / D The coefficients that can be used in the formula "Coefficients" activated Decimal places Defines the number of decimal places in the calculated result. ­Result limits Limit values for going above and below a result ­Lower limit Lower limit of the result "Result limits" activated Upper limit Upper limit of the result "Result limits" activated Interrupt if limits
exceeded
Statistics For the calculation, rolling statistics are activated. ­Print statistics Defines whether the statistics should be printed out for each sam
Formula (Condi tion)
See also
Calculations in Products (page59)
Use of variables in the formula "Coefficients" activated
Defines that the measurement will be interrupted if the result is outside the limits. It then receives the status "Error". If this parameter is not activated, the measurement continues even if the limit value is exceeded. The analysis then receives the status "Limit exceeded"
ple. Define a specific condition for calculation. "Condition" activated
"Result limits" activated
"Statistics" activated
-
6.1.7.7Clean
Cleaning the measuring cell by means of the automation unit specified in the method function "Configuration".
Parameters Description Displayed if
Drain Draining the measuring cell ­Drain Direction Defines the valve setting for draining:
Drain Mode Fixed duration:
Drain duration Draining time in [s] "Drain mode" = "Fixed
Drain ratio Duration of draining (in percentage of filling duration) "Drain mode" = "Auto
Rinse Cycle (1) Activation of the cleaning cycle "Automation" = "Fillpal"
Waste: A waste bottle is drained. Vial: It is drained into a vial.
Draining is stopped after a defined time. Auto: The drain duration depends on the sample feed (total duration equals the time for which the pump was in operation during fill ing).
"Automation" = "SC1" /
"SC30"
"Automation" =
"FillPal"/ "InMotion"
(only for drain direction:
Waste)
"Drain" activated
"Automation" = "InMo
tion" (only drain mode:
Fixed duration)
"Drain" activated
duration"
matic"
/ "SC1" / "SC30" /
"InMotion"
"Automation" = "InMo
tion" (only method type
"clean", 2 rinse cycles
possible)
48 Methods and Products
Solvent Automation unit SC1 / SC30: Selecting the solvent. The selection
"Rinse cycle" activated
depends on the solvents entered in Home > Setup >
Hardware > Automation
The solvents are listed in the list field along with the connection. The name of the solvent is displayed during method execution in the notification that prompts the rinse process.
Rinse Mode "Fixed duration":
Rinsing is stopped after a defined rinse duration. Automatic: The rinse duration depends on the sample feed (Total duration is the time for which the pump was in operation for filling).
"Automation" = "FillPal"/ "SC1"/ "SC30" / "InMotion"
"Automation" = "InMo tion" (only rinse mode: Fixed duration)
"Rinse cycle 1" activat ed
Rinse duration Rinse time in [s] "Rinse mode" = "Fixed
duration"
Rinse rate Duration of rinsing as percentage of filling duration "Rinse mode" = "Auto
matic"
Air addition To increase the efficiency of rinsing, air is mixed with the solvent.
A higher air ratio causes a more turbulent flow, meaning more intensive cleaning and lower consumption of fluid.
"Automation" = "SC1"/ "SC30"
"Rinse cycle" activated If Power Purge Unit (PPU) is used and activated in the setup of "SC1" / "SC30", the parameter "normal" is not available (only "low" and "very low").
Rinse cycle 2 Activation of 2nd rinsing cycle; for the description of the parame
Rinse cycle 1 activated ter, see parameter "Rinse cycle 1".
Dry Activation of drying ­Dry mode Fixed duration:
Drying activated Drying is stopped after a defined time For the density - the "dry mode" can also be set to "Automatic": Automatic: Drying is interrupted, when the oscillation value becomes stable.
Dry duration Drying time in [s] "Dry mode" = "Fixed
duration"
Max. dry dura tion
Interrupt if failed If this parameter is activated, the measurement is given the status
Maximum drying time: If the oscillation value has not stabilized by the end of this peri od, drying is terminated.
"Error", if with dry mode "automatic" the oscillation value has not
Density - Module
defined
Dry mode = Automatic
"Dry mode" = "Auto
matic" stabilized within the defined "Max. dry duration". The task list is then interrupted. If the parameter has not been activated, the measurement proceeds as normal, even if drying failed. This parameter is necessary, for example, for when execution of the "Drying" method function is associated with a condition.
Prompt for DryPal
Choose if you want a prompt on the homescreen for connecting the DryPal to the aspiration tube for drying.
"Automation" = "InMo
tion"
"Dry" activated
Suppress prompts
Formula (Condi tion)
Choose if you want to suppress prompts on the homescreen for the cleaning process.
"Automation" = "FillPal"
/ "FillPal&DryPal"" Define a specific condition for cleaning. "Automation" = "InMo
tion"
Only if "Condition" acti
vated
49Methods and Products
6.1.7.8Online Display
You can define the number of displayed results for the online screen (two or four). You can also specify where the results should be placed.
Parameters Description Displayed if
Displayed results The number of results on the online screen can be selected. ­Top / bottom Place at which the calculated results can be displayed on the
Top left / Top right / Bottom left / Bottom right
6.1.7.9Report
This method function defines the type and scope of the data to be output for a report sent to network, a USB stick, printer (stripe printer or network printer) or via a USB-RS232 interface. The "Report" method function refers to all previous method functions. "Report" contains the following parameters:
Parameters Description Displayed if
Export to USB stick
Export to network Analysis data can be exported to a network folder in a XML, CSV
Print/ USB­RS232 data export
Print Mode The following values are available: "Summary" or "User defined".
"Print Mode" applies only to printing on the printer. For USB-RS232 data export, a fixed data record is exported. For information on the scope and syntax of the data export, see "RS232 Interface Description"
The following parameters are displayed for "Print Mode"="User defined":
Method/Product information
Calculated results
Data The raw results produced during the determination and used in
Sample informa tion
Resource infor mation
The "Condition" and "Formula" parameters can be found in "Methods and products: Methods: Conditions and formulas".
screen. Place at which the calculated results can be displayed on the
screen.
Analysis data can be exported to a USB stick in a CSV or/and PDF format. Multiple measurements are arranged in succession in the same file. Data export to the USB stick is only possible when the USB stick is available. This can be checked using the option "Verify USB stick availability" (Home>Setup>Global settings>Analysis and resources behavior).
or/and PDF format. Multiple measurements are arranged in suc cession in the same file. Data export to the network is only possible when the network connectivity is available. This can be checked using the option "Verify network storage availability" (Home>Setup>Global set tings>Analysis and resources behavior).
Data are either sent to a USB compact printer, to a network printer or exported via the USB-RS232 interface. You can define the type of data transfer using the "Printer type" parameter in the Setup menu (Home>Setup>Hardware>Peripherals>Printer).
If "Summary" has been selected, a brief summary of the most important data is printed. If "User defined" has been selected, you are presented with addi tional parameters enabling you to define which data should be printed.
Method information is printedor exported, e.g. method ID or product ID, method type. A complete method is not printed.
The results from the "Calculation" method functions can be print ed or exported.
the calculations are printed or exported. Information on the sample (e.g. sample ID, viscosity, standard
name) can be printed or exported. Information on the resource used (e.g. auxiliary value, adjust
ment set, cell type) can be printed or exported.
"Displayed results" = "2"
"Displayed results" = "4"
-
-
-
-
"Print Mode"= "User defined"
"Print Mode"= "User defined"
"Print Mode"= "User defined"
"Print Mode"= "User defined"
"Print Mode"= "User defined"
50 Methods and Products
The "Report" method function can be inserted a maximum of two times per method for the method types "Mea surement" and "Test" and only once for "Adjustment". See also the flow chart in "Methods and products: Meth ods: Method Syntax".
6.1.7.10Cell Test
The cell test can only be performed for the density - module.
The cell test is used for testing the cell in respect of dryness and cleanness.
Parameters Description Displayed if
Tolerance Maximum permissible deviation of the measured air density from
the theoretical value
Max. test dura tion
Maximum duration of the cell test in [s]. If the measured value has not stabilized by this time, the cell test will be interrupted and counts as having been failed.
Interrupt if failed If activated, the measurement receives the status "Error" if the cell
test is failed. The task list is interrupted. If the parameter has not been activated, the measurement proceeds as normal, even if the cell test has been failed.
6.1.7.11Temperature Compensation
Temperature compensation allows you to measure a sample at a temperature and then to output the measured value at another temperature. This temperature compensation is used in the following cases:
The measured value must for example be stated at 15°C, however the sample is too viscous at this tem perature. Therefore these samples must be measured at correspondingly higher temperatures so that they can be pumped to the measuring cell and remain fluid there.
For accelerated measurements. If the sample is delivered at 35°C for example and the measured value is required at 15°C, the sample can be measured at 35°C and the measured value can be compensated to 15°C.
If the required reference temperature is outside the range of the measuring instrument (e.g. 120°C), the sample can be measured at 75°C and the result compensated to 120°C.
-
-
-
To use the temperature compensation, the temperature dependence of the sample must be known.
If you have connected a density - module, you can also use the API table (ASTM-D 1250). The following tables are used (for crude oils, refined products, lubricants):
53A, 53B and 53D for 15°C
5A, 5B and 5D for 60°F
59A, 59B and 59D, for 20°C
The aforementioned ASTM or API tables take as input values the readings taken from a hydrometer, which do not however make allowance for the coefficient of expansion of the glass hydrometer. Therefore the measured density cannot be directly used as the input value for these tables.
The measured values are converted to the desired temperature. The output value (TC) of this method function can be used in the method function "Calculation".
Parameters Description Displayed if
Compensation type
API (only for external density measuring cell): Temperature compensation with API tables
-
Formula: Temperature compensation by entering any desired formula
API table Selection of the relevant API table: Crude oil, refined products,
lubricants
Density - Module
defined
Compensation type =
API
51Methods and Products
Compensation ­temp.
Output Selection of the unit in which the temperature compensated value
Formula Input of a formula for temperature compensation. "Compensation type" =
Coefficients Specifies the coefficients that can be used in the formula.
A / B / C / D The coefficients that can be used in the formula "Coefficients" activated
6.1.7.12Adjustment
You can use the "Adjustment" method function to store the adjustment data in the setup for the corresponding cell(s). Here you can also perform an adjustment analysis that should be used to verify the accuracy of the determined adjustment data. In this process, the system checks the deviation from the last adjustment. If the maximum deviation is exceeded, you can decide at the end of the adjustment whether the data is transferred.
Parameters Description Displayed if
Adjustment analysis
Standard 1/2 Shows the used standards (info field) "Adjustment analysis"
Max. deviation Input of the maximum permissible deviation. "Adjustment analysis"
Compensation type = API (only for density - module) You can choose from among three reference temperatures (15°C, 60°F, 20°C). The temperature is entered as the input value in the algorithm for calculating the API compensation.
Compensation type = Formula
The defined value is for information purposes only and is not included in the calculation.
"Compensation type" =
is to be output.
Note: For powers to the base 10 the "e"-convention may be used (e.g. 1.2e-4 --> 0.00012).
Evaluation of the adjustment -
"API"
"Formula"
activated
activated
-
-
6.1.7.13Test
This method compares the measured with the theoretical test value. The data from the test is saved in the setup of the corresponding cell(s).
Parameters Description Displayed if
Tolerance d/nD Allowed tolerance for the test -
6.1.7.14Instruction
You can determine the output of a message to be displayed during the analysis process. The analysis is inter rupted while the message is displayed.
Parameters Description Displayed if
Instruction Freely definable text. Via the input %x% reference can be made
Continue after Confirmation:
Time interval Maximum time interval for the display of the instruction (early
Print The specified text is printed out from the printer. -
to results or raw data, e.g. %R1%.
The analysis is continued as soon as the message has been confirmed. Time interval: The analysis is continued at the latest after the defined time peri od has elapsed.
confirmation continues the analysis)
-
-
"Continue after" = "Time interval"
6.1.7.15Auxiliary Value
By means of the method function "Auxiliary value" a result can be assigned an auxiliary value. These auxiliary values can be accessed in a formula (also in other methods).
52 Methods and Products
Parameters Description Displayed if
Name Selection of auxiliary value The selection relations to the list in the
setup (Home > Setup > Tables & Values >
Auxiliary values)
Formula H= Formula for the calculation of the auxiliary value. ­Limits By means of limits, the checking of the calculated value can be
defined. If the value is outside the limits, it will not be saved as an auxiliary value.
Lower limit Lower limit for the calculated value. "Limits" activated Upper limit Upper limit for the calculated values "Limits" activated
6.1.7.16Wait
By means of the method function "Wait" a method can be interrupted.
Parameters Description Displayed if
Wait time Wait time before execution of next method function -
6.1.7.17Auxiliary instrument
This method function activates external auxiliary instruments and enables the instrument to communicate with these auxiliary instruments. Auxiliary instruments can be connected either to the "Aux" socket on the instrument board, the 24V output or to a USB interface by means of the USB-RS adapter (RS232). The parameters described in the following apply to all control types. These are followed by parameters that can be defined explicitly for the relevant control type:
Parameters Description Displayed if
Control type Indicates which port on the measuring instrument is to be used
for the auxiliary instrument.
Name A name of your choice. ­Condition Logical condition for defining whether or not a method function is
executed based on a result (true or false).
Formula Formula input for the calculation. -
A method function can be executed using a specified calculation formula (see also "Methods and products: Methods> Calculations and formulas").
-
-
-
-
Control type: 24V output
Parameters Description Displayed if
Mode Defines the mode for controlling the control output.
- Fixed duration: The control output is switched on for the defined time period.
- On| Off: The control outlet is switched on or off. After a sample series the control outlet is automatically switched off.
Duration Here you can enter a time span, in [sec], for which the control
output should be switched on.
Control type: USB-RS232
Parameters Description Displayed if
Send output sequence
Output sequence The control sequence for the signal receiver – can also contain a
Wait for input sequence
Determines whether an output sequence should be sent. "Control type" = "USB-
formula embedded in % characters, e.g. "%R1%" or "%d%" or ASCII control characters in the format \xxx. Here, xxx is the deci mal code of the control character.
The method function waits for a response sequence from the external device.
"Control type" = "Output
24V"
"Control type" = "Output
24V",
"Mode" = "Fixed dura
tion"
RS232"
"Control type" = "USB-
RS232" and "Send out
put sequence" activated
"Control type" = "USB-
RS232"
53Methods and Products
Maximum time The maximum waiting time for the input sequence, in [sec]. After
it expires, the method is continued even if no sequence was received.
"Control type" = "USB­RS232" and "Wait for input sequence" activated
Input sequence The input sequence from the external device. The sequence can
contain ASCII control characters in the format \xxx. Here, xxx is the decimal code of the control character.
"Control type" = "USB­RS232" and "Wait for input sequence" activated
Input sequence with result
Defines whether the input sequence of the external device con tains results that should be imported.
"Control type" = "USB­RS232" and "Wait for input sequence" activated
Start sequence Start sequence of the incoming sequence from the external
device. This is the reference position for the subsequent results. The start sequence can contain ASCII control characters in the format \xxx. Here, xxx is the decimal code of the control charac ter.
"Control type" = "USB­RS232", "Wait for input sequence" activated and "Input sequence with result" activated
Total length Length from beginning of start sequence to end of final result.
It is only after this length has been received that the results are extracted from the input sequence.
"Control type" = "USB­RS232", "Wait for input sequence" activated and "Input sequence with result" activated
Number of results
The number of results that should be extracted from the sequence from the external device.
"Control type" = "USB­RS232", "Wait for input sequence" activated and "Input sequence with result" activated
Start position Start position (beginning) of the corresponding result; counted
from the beginning of the start sequence. (Leading blanks before the result are ignored.) Up to 10 results are supported.
"Control type" = "USB­RS232", "Wait for input sequence" activated and "Input sequence with result" activated
Max. length Maximum length of the result; beginning at the start position of
the result.
"Control type" = "USB­RS232", "Wait for input sequence" activated and "Input sequence with result" activated
The ASCII control characters can be found at: http://www.asciitable.com/
Results are stored under the variables "AuxInst x[y]" (x: result number; y: number of the "Auxiliary instru ment" method function). For example: AuxInst 3[2]; where "3" represents the third result in the input sequence, which was received in the second auxiliary instrument method function in the method.
6.1.7.18PowerShower
Rinsing with the defined number of rinse cycles in the setup.
54 Methods and Products
Parameters Description Displayed if
Beaker position Define the position of the rinse beaker.
Current position: Rinsing is performed in the same beaker.
Next beaker: Rinsing is performed in the beaker next to the tower (current posi tion + 1).
Absolute position: Rinsing is performed in a fix defined beaker on the sample rack.
Position (Beaker) Define the position of your rinse beaker on the sample rack. "Automation" = "InMo
Rinse cycles Define the rinse cycles. "Automation" = "InMo
Volume per cycle Define the solution volume which shall be used per rinse cycle. "Automation" = "InMo
Volume (Fill beaker)
Volume (Fill cell and tubes)
Formula (Condi tion)
Define the solution volume that shall be filled into the rinse beaker. The titration head will go into position "sample" for that.
Define the solution volume that shall be filled into the measuring cell and into the tubes. The titration head will go into position "sample" for that.
Define a specific condition for rinsing with PowerShower. "Automation" = "InMo
"Automation" = "InMo
tion"
tion"
Only if "Absolute posi
tion" is choosed
tion"
Only if "PowerShower
rinse" is activated
tion"
Only if "PowerShower
rinse" is activated
"Automation" = "InMo
tion"
Only if "Fill beaker" is
activated
"Automation" = "InMo
tion"
Only if "Fill beaker" is
activated
tion"
Only if "Condition" is
activated
6.1.7.19Stir
With this method function you can define a fixed pre-stirring duration and also, if the stirrer shall be activated after the defined pre-stirring duration.
Parameters Description Displayed if
Speed 1 Define the stirrer speed in percent depending on the maximum
Stir duration Define the stirring time for the pre-stirring. You can define two
Continue stirring Continued stirring after the defined stirring time in "Stir duration". "Automation" = "InMo
Speed 2 Define the stirrer speed after pre-stirring in percent depending on
speed of your stirrer (pre-stirring).
stirring sections (e.g. 10 s higher speeded for pre-stirring fol lowed lower speeded until the end of measurement).
the maximum speed of your stirrer.
"Automation" = "InMo
tion"
"Automation" = "InMo
tion"
tion"
"Automation" = "InMo
tion"
Only if "Continue stir
ring" is activated
55Methods and Products
Formula (Condi tion)
6.1.7.20Line rinse
With this method fuction you can reach a better degree of cleanliness of the lines. For this a mixture between air and cleanling solution will be pumped through the lines, so the cleaning effect is much better than with clean ing solution only. The pump is running intermittently.
Parameters Description Displayed if
Beaker position Absolute position
Duration Define the cleaning duration which depends on the length of the
Descent rate Define the descent rate of the titration head during the rinse dura
Formula (Condi tion)
Define a specific condition for stirring. "Automation" = "InMo
tion"
Only if "Condition" is activated
"Automation" = "InMo The system will move to the defined position on the turntable for the start of a sample.
Current position
The system will start looking for the first beaker to come.
Next beaker
The system will move to the beaker which is next to the tower (home position + 1). If no beaker can be detected, the sample changer moves to the next position (home position + 2) and repeats until a beaker is detected.
installed lines and on the wished degree of purity.
tion (very low, low, medium, high, very high). Define a specific condition for line rinse. "Automation" = "InMo
tion"
Only in method type
"clean"
"Automation" = "InMo
tion"
"Automation" = "InMo
tion"
tion"
"Condition" is activated
6.1.7.21Park
With this method function you can park the InMotion head in a beaker, e.g. parking the pH electrode in a beaker with buffer.
Park position Define the park position.
Formula (Condi tion)
6.2Products
By means of products, product-specific adjustments to methods can be made. Each product is respectively associated with one method. and several products can be association with the same method. The product parameters are a subset of the method parameters. Products are created and managed in the product editor.
When a method is saved, the products that access this method are updated.
"Automation" = "InMo
Absolute position
The system will move to the defined position on the turntable.
Current position
The system will start looking for the first beaker to come.
Next beaker
The system will move to the beaker which is next to the tower (home position + 1). If no beaker can be detected, the sample changer moves to the next position (home position + 2) and repeats until a beaker is detected.
Define a specific condition for method function "park". "Automation" = "InMo
tion"
Only in method type
"clean"
tion"
"Condition" is activated
56 Methods and Products
6.2.1Create Products
Select: Home > Methods / products > Products
You create a new product if you either select and configure defined products from the list and save them under a new ID or generate a completely new product. The following description shows how you can generate a new product:
Open the method selection window via the button New in the Products dialog. Here all the methods of the type Measurement are listed that can be assigned to a product.
1 Touch the corresponding method in the list that should be assigned to the product.
The Product dialog is opened. The parameter "Linked method" shows the ID of the selected methods. Note: You can retrospectively assign another method to the product (same configuration).
2 You can edit and configure product parameters, assign the product an ID and a title.
You can start the product immediately or return to the product editor.
6.2.2Linking Methods with Products
A distinction is made between two cases:
1.
A product is created and linked with a method. On linking of a product to a method (before saving the first time) the values are taken over from the linked
method.
2.
A product has been previously created, the linked method is changed (relinking). In this case only the methods for the relinking are suggested that agree with the currently linked method in
the configuration. After initial saving, the product parameters are not overwritten by the parameters of the method. For the measurement the product parameters are: such as e.g. Multiple measurements, Calculations (Coeffi cients and result limits) are used and not the parameters of the corresponding method. The entries in the method however remain unchanged.
When you start a product measurement, the parameters defined in the method are "overwritten" by the prod uct parameters.
A maximum of 100 products can be defined.
If no measurement methods have been defined on the device, no product can be created.
Product ID’s can be used for barcode recognition.
Several products can be linked to the same method.
A product is always assigned to exactly one method.
When a method is deleted, all linked products are also deleted.
If a product is saved under a new ID, the values are retained in the product and are not overwritten by stan dard values from the method.
6.2.3Parameters for Products
The following parameters can be defined:
Parameters Description Displayed if
Product ID The product is uniquely identified in the system via the product
ID. Product title Name of the product ­Linked method Selection of methods used in the product. With a new link, only
methods with the same configuration (cell mode and external
instruments) are shown. Cell mode Display of cells to be used in the method (info field) ­External instru
ments Color settings
(Minolta only)
Display of external instruments (pH, conductivity, color) to be
used in the method is to be measured (info field)
Activates "Illuminant" so that the illuminant type can be defined
and "Observer" so that the angle can be entered.
-
-
-
-
57Methods and Products
Illuminant The following illuminants can be selected:
A, C, D65, D50, ID65, ID50, F2 to F12
Observer Viewing angle:
2°C or 10°C; default 2°C
Speed Pump speed
"SC1"/ "SC30":
• "Maximum": Sample feed with maximum pump speed
"Color settings" (Minol ta only)" activated
"Color settings" (Minol ta only)" activated
"Automation" = "FillPal"/ "SC1"/ "SC30" / "InMotion"
• "Reduced": Sample feed with speed set at the automation unit
• "Automatic": The pump speed changes according to the defined time period from "Reduced" to "Maximum".
"FillPal" / "InMotion":
• "High"
• "Medium"
• "Low" The corresponding pump speed can be set under Home >
Setup > Hardware > Automation .
Max. speed after [s]
The pump switches over to the maximum speed after this time "Automation" = "SC1"/
"SC30"
"Speed" = "Automatic"
Multiple mea surement
No. of measure ments
Refill ratio Measurement for further movement of the sample between two
On activation several measurements are performed on the same sample
Input of number of measurements "Multiple measurement"
activated
"Automation" = measurements. The refill ratio is calculated using the duration of the sample addition. For example: The addition duration is 10s and the refill ratio is 50%, i.e. pumping continues for 5seconds.
"FillPal"/ "SC1"/
"SC30"/ "InMotion"
"Repeat if failed" acti
vated
"Multiple measurement"
activated
Max. SD Maximum permitted absolute standard deviation of the measure
ments for density, refractive index or color. When the deviation is
Multiple measurement
activated exceeded the measurement is interrupted with the status "Error".
The maximum standard deviation must be entered in the function of the measuring cell type. The default value in the method is provided for an R40 cell. This should be adjusted for an R50 cell, e.g. 0.00002.
No. of meas. points
Defines the number of measured points used for determining the mean value.
pH or conductivity
meter defined in method Note: If multiple measurement is activated, only the measured values from the last measurement repetition are used.
Max. deviation The max. permissible absolute deviation of the measured values
for the pH or conductivity meter. If this value is exceeded, the
pH or conductivity
meter defined in method measurement is interrupted with an "Error".
Repeat if failed If this is activated, the measurement is repeated if "Max. SD" or
"Max. Deviation" are exceeded. Note: If "Multiple measurement" is activated, all measurements are repeated.
"Multiple measurement"
activated, or
External instrument
defined (pH meter, con
ductivity meter, col
orimeter)
Author Name of creator (info field) ­Created on Date of creation and time (info field) ­Modified at Change date and time (info field) ­Modified by User who made change (info field) -
-
58 Methods and Products
Protect When this is activated, only the author or an administrator can
edit or delete the method.
For the density module the following parameters are available:
Bubble Check When this is activated, the system checks whether any bubbles
are present in the cell.
Viscosity correc tion
Viscosity ≤2000:
Viscosity Value Viscosity of the sample "Viscosity" = "Set value"
Activation of the viscosity correction Density - Module
For samples with viscosity less than 2000 mPa*s >2000: For samples with viscosity greater than 2000 mPa*s
- Set value: With this selection, a known viscosity can be entered
6.2.4Calculations in Products
The calculations defined in the method (Rx) are listed in the product. Press the button Calculations in the prod uct dialog (Navigation: Home > Methods / products > Products). For a description of the parame ters also see "Methods and products: Methods > Method functions > Calculation (page47)".
The following rules apply for calculations in products:
1.
When the methods are saved calculations in methods and products are compared on the basis of calcula tion indexes (R1 – R1, R2 – R2 etc.). If these indices from product and linked method do not coincide, they will be automatically aligned, i.e. product calculations that do not occur in the method will be automatically deleted and method calculations that do not occur in the product will be automatically inserted.
2.
If a parameter is changed in the method function "Calculation" that is only visible in the product (non­editable, e.g. formula), then the entire product calculation will be overwritten in the associated product.
-
-
defined Density - Module con
nected
Viscosity correction activated
59Methods and Products
7Series
Series act as a starting aid for several tasks. The tasks are individually lined up in the task list after the series start.
Series can be created, edited saved and deleted.
Create series Navigation: Home > Series
You can create shortcuts for series. A maximum of 5 series can be saved. The maximum number of series entries is 60. Series can consist of products and / or methods. Series can consist of the same or different entries (i.e. several entries (from the type method or product). All entries can be deleted. Series can also be saved without entries. If you are in the Start analysis dialog of the series, series entries can be deleted, inserted or selected.
Via the Insert button, you open the Series item dialog to define parameters; see also "Create series". If a change of series is made in the Start analysis dialog and a shortcut is created, this shortcut does not
point to the changed series but to the series created in the series editor. If no methods are defined in the instrument, no new series can be created.
Via the button Series in the Homescreen you open the Series list dialog. The available series are displayed.
1 Touch the button New.
A new series is generated. You can enter a series ID or use the default ID.
2 Save the series.
A dialog with the title of the newly created series opens. You can add new entries or edit existing ones. To edit an existing entry, touch the relevant entry. The Series item dialog opens.
3 For a new series entry, press the button Insert in the dialog of the selected series.
The Series item dialog is opened.
4 Enter the parameters and confirm with OK.
In the case of multiple entries, "Insert" markers appear behind the series entries.
5 Press the Insert marking area at the desired point.
The Series item dialog is opened.
6 Enter the parameters and confirm with OK.
7 Save the series.
The parameters for the Series item dialog are listed below:
Parameters Description Displayed if
Item type The series item can either be a product or a method (only if at
-
least one product is defined in the system).
Product ID Selection of product "Item type" = "Product" Method ID For "Item type"="Method", it is possible to select the relevant
­method here. For "Item type"="Product", the associated method is displayed here (Info Field).
Method type Displays the type of method, e.g. measurement ­Sample ID Input of the sample ID "Method type" = "Mea
surement"
Number of
To define the quantity of series entries. -
entries

60 Series

For the entry type "Method" of the type "Adjustment" and "Test", you can change the associated standard data using the parameter "Standard". For this, see also "Methods and products: Methods> Standard data (page40)".
Standard For "Method type"="Adjustment", it is possible to modify the stan
A series can also be saved without entries. All entries can be deleted.
In the dialog of the corresponding series the Start button is found with which you can open the Start analysis dialog. In this you have the option to change your items again before the actual analysis start. To do this press the button Items.
In addition to the parameters described above, the following parameters are displayed:
Correction factor Any correction factor that can be used in calculations. "Method type" = "Mea
For more information on the parameters in the Start analysis dialog, see also "Analysis sequence: Starting an analysis (page68)".
7.1Delete Series
1 Touch the series to be deleted in the Series list dialog.
A list with the corresponding items for the series appears.
2 In the dialog press the Parameters button.
In the dialog there is a Delete button with which you can delete the series.
A prompt with the interruption criterion appears before you finally delete the series.
dard data. For "Method type"="Test", the standard used is displayed (Info Field).
"Method type" = "Adjustment" "Method type" = "Test"
surement"
7.2Creating a series copy
Using the Parameters button, you can then copy an existing series by changing the "Series ID". When the series is saved a new series is created.
61Series
8Results and Statistics
Navigation:Home > Results
You can get directly from the Homescreen to your analysis data via the Results button. Here the most recent 60 analyses with analysis status, analysis data, raw data, user-defined results, information on the analysis and statistical data are stored. The performed analyses are listed, with the exception of manual operations and methods of the type "Clean".
The analysis list contains the following information:
with methods of the method type and with products "P"
Methods or product ID
Sample ID (or Adjusting/Test mode for the method types "Adjustment" and "Test")
Status, Result (Rx), Sample and Date/Time: Using the buttons Status, Rx, Sample and Date/Time, you can change between the various display modes. The Status and Rx buttons enable you to change between the analysis status and the result of the first cal culation of the method (for the method type "Measurement").2 The Date/Time and Sample buttons enable you to change between the date and time of the analysis and the sample ID. If Sample is selected, "Sample ID" appears (for the method type "Measurement"), "Air&Water" (for the method type "Adjustment") and, in the case of methods that use the SC30 sample changer, the sample rack position is displayed.
Executed analyses can have the following statuses:
OK
OK*
Corresponds to the status "OK" with one of the following restrictions:
One adjusting or test set has expired.
An error has occurred after the method function "Measurement".
Error
Before the end of the method function "Measurement" an error has occurred.
Failed (Methods of the type adjustment and Test):
"Adjustment": Adjustment analysis failed.
"Test": Tolerance exceeded.
Lim. exceeded (Measurement type methods): The limits defined in the method function "Calculation" have been exceed (for method type "Measurement").
Excluded:
The analysis has been manually excluded.
Analysis with the statuses "Error", "Failed" or "Lim. exceeded" are tagged in red.
Via the button Export analyses of the type "Measurement" can be exported.
By touching Delete all you can delete the analyses with all data.
To view the data of the performed analyses, proceed as follows.
– Touch the relevant analysis.
For the selected analysis a new dialog opens. This dialog window contains the following buttons:
- Calculated Results (only for the method type "Measurement")
- Statistics (only for the method types "Measurement" and "Statistics")
- Information
- Data
- Back
- Report
- Exclude
By touching the button Exclude individual analyses can be excluded from the statistics.
You can print or export a report manually by touching the Report button. The data included are the data used in the method. (See also the "Report (page50)" method function ("Methods and products: Methods>Meth ods>Method functions").

62 Results and Statistics

See also
Calculation (page47)
Results and Statistics (page62)
8.1Statistics

Via the button Statistics you can also display the statistics (the statistics are listed per calculation, if in the method function "Calculation" the parameter "Statistics" has been activated). As a component of the statistics (Rolling statistics) the following values are displayed:

Mean value x of a result Rx (Mean [Rx])
The standard deviation of a result Rx (SD)
The following criteria apply for the interruption of the rolling statistics:
1.
Change of method ID or product ID. The automatic sample ID extension which is added to the ID when working in continuous run (XY__1, XY__2) is ignored for the statistic. That means only the user edited ID XY is taken as sample ID.
2.
In the setup (Navigation: Home > Setup > Global settings > Analysis and resources
behavior), you can activate the setting "Reset statistics with other sample ID". In this case the statistics
are also reset if the sample ID changes.
The rolling statistics are not interrupted in measurements with the status "Error" and "Excluded". The measure ments are not however taken into account in the rolling statistics. On the other hand, analyses with the status "OK*" or "Lim. exceeded" are included in the statistics (the description of the analysis is found at the start of this section "Results and Statistics (page62)").
8.2Information
You received detailed information for the selected analysis result, e.g.:
Automation unit
Status
Cell
Method ID
Sample ID
Date / Time
Performed by
8.3Data
The raw data generated in the analysis is shown below Data, e.g.:
Set temperature (standard) and the cell temperature
nD
nD1... nD10 (single values from multiple measurements)
Meas. duration
External instruments
- pH meter
pH
Max. Deviation
Deviation
Temperature
- Conductivity meter
Conductivity
Max. Deviation
63Results and Statistics
Deviation
Temperature
- Colorimeter
Illuminant
Observer
For multiple measurements:
- maxSDCOL
- SDCol
Path length
64 Results and Statistics
9Manual Operation
With the help of the manual operations, you may call up various functions of the instrument that are indepen dent of the immediate execution of an analysis but which for example, may be helpful during the preparation of the samples.
Cell
Via the manual operation "Cell" a cell conditioning run can be started. If you have connected the density ­module, you can also start a cell test.
9.1Automation
Navigation: Home > Manual operations > Automation
You can perform the following manual operations with the automation units:
Action DryPal FillPal SC1 SC30 InMotion
Dry X X X Rinse X X X X Drain X Pump sample X X X X PowerShower X Stir X Move to position X X Move lift X
If no automation is defined in the setup, the manual operation "Automation" does not appear.

– Open the Automation dialog window via the button Automation in Manual operations.

In it you can define the following parameters to perform a manual operation with an automation unit:
Parameters Description Displayed if
Cell mode Defines the cell(s) ­Automation Definition of the automation unit for manual operation ­Action Specifies the action with which the manual operation is to be per
Below the parameters are listed that are to be entered for the selected action:
9.1.1Action: Dry
Parameters Description Displayed if
Dry mode Fixed duration:
Dry duration Drying time in [s] "Dry mode" = "Fixed
Max. dry dura tion
formed. The offered selection depends on the automation select ed.
Drying is stopped after a defined time For the density - the "dry mode" can also be set to "Automatic": Automatic: Drying is interrupted, when the oscillation value becomes stable.
Maximum drying time: If the oscillation value has not stabilized by the end of this peri od, drying is terminated.
-
Drying activated
duration" Density - Module
defined
Dry mode = Automatic
9.1.2Action: Rinse
Parameters Description Displayed if
Solvent Definition of the solvent at the connections "Rinse 1" and "Rinse2""Type" = "SC1"/
Rinse duration Duration of the rinse; Input from infinity "∞" is possible. -
"SC30"

65Manual Operation

Rinse position Absolute position
The system will move to the defined position on the turntable for the start of a sample.
Current position
The system will start looking for the first beaker to come.
Next beaker
The system will move to the beaker which is next to the tower (home position + 1). If no beaker can be detected, the sample changer moves to the next position (home position + 2) and repeats until a beaker is detected.
9.1.3Action: Pump sample
Parameters Description Displayed if
Direction Cell:
Pump in direction cell Vial: Pumpin direction vial
Note: FillPal only pumps in one direction. The automation units SC1 /SC30 can pump in both directions.
Speed Pump speed
"SC1"/ "SC30":
• "Maximum": Filling with maximum pump speed
• "Reduced": Filling with speed set at the automation unit
"FillPal" / "InMotion":
• "High"
• "Medium"
• "Low" The appropriate pump speed can be set in Home > Setup >
Hardware > Automation.
Duration Duration of pumping input from infinity is possible. -
"Type" = "InMotion"
-
Automation = "FillPal" / "SC1" / "SC30" / "InMo tion"
9.1.4Action: Move to position
Parameters Description Displayed if
Go to Home:
Sample changer (SC30 / InMotion) moves to position "Home"
Relative position:
Sample changer (SC30 / InMotion) moves a definable number of steps Absolute position: Sample changer (SC30 / InMotion) moves to an indefinite posi tion Next vial / beaker: Sample changer (SC30 / InMotion) moves to next vial / beaker
Position Absolute position, that is approached by the sample changer. "Go to" = "Absolute
Direction Definition of the moving direction of the sample rack. Type = "SC30" / "InMo
9.2Cell
Navigation: Home > Manual operations > Cell
Here you can perform cell conditioning.
Type = "SC30" / "InMo tion"
position"
tion"
66 Manual Operation
9.2.1Action: Cell conditioning
By means of Cell conditioning (standard setting) you can assign the cell a certain temperature, so that the cell is already at operating temperature before measurement takes place (e.g. evening before). The set temperature and the adjustment set are read out from the selected method and sent to the cell.
In the corresponding method the set temperature and the adjustment set have already been specified for the conditioning of the measuring cell(s).
You can define the corresponding parameters for the following actions:
Parameters Description Displayed if
Type Method or product (at least one product must be defined) ­Method Method, with which the cell conditioning is to be performed. "Type" = "Method" Product Product, with which the cell conditioning is to be performed. "Type" = "Product"
Methods of the type "CLEAN" are not available here for selection.
9.2.2Cell Test
If you are using a density - module, you can perform a cell test for it.

The Cell test is designed to check whether the cell is dry and clean. With this test a measurement is carried out with the air. The measured density of the air is compared with the theoretical value. If the difference of the measured results is within the specified tolerance, the cell test counts as having been passed.

Parameters Description Displayed if
Max. test dura tion
Tolerance Permitted tolerance for passing the cell test Action = cell test
Maximum duration of the cell test in [s]. If the measured value has not stabilized by this time, the cell test will be interrupted and counts as having been failed.
Action = cell test
67Manual Operation
10Analysis Sequence
10.1Starting an Analysis
An analysis can be started in various ways:
Start from the editor (Methods/Products/Series)
Start from Homescreen
Start via shortcuts
Start with barcode reader
Start with ErgoSens
Depending on the settings and the way in which the program is called up, you can enter or check analysis­related data in the Start analysis dialog. The following parameters may be displayed:
Parameters Description Displayed if
Method ID The method is uniquely identified in the system via the method
ID. Once the method has been saved, the method ID can no longer be changed. A change of the method ID results in the cre ation of a copy of the method.
Method type Displays the type of method, e.g. measurement ­No. of samples Enables you to enter the sample quantity (max. 120). "Method type" = "Mea
Sample ID Sample ID can be entered; you can still change this ID at the
analysis start.
Start position You can define different start positions.
Home
Position number 1 on the turntable. The changer will move to position 1 for the start of a sample.
Absolute position
The system will move to the defined position on the turntable for the start of a sample.
Current position
The system will start looking for the first beaker to come.
Next beaker (only InMotion) The system will move to the beaker which is next to the tower (home position + 1). If no beaker can be detected, the sample changer moves to the next position (home position + 2) and repeats until a beaker is detected.
Continuous run If "Continuous run" is activated, the same measurement will be
started automatically at the end of a measurement.
Continuous run remains in effect until it is terminated by means of the Stop button.
The sample vials are processed successively until an empty space is found on the sample changer.
-
surement" "Method type" = "Mea
surement" "Method type" =
"Adjustment" / "Mea surement" / "Test" / "Clean"(only InMotion)
"Method type" = "Mea surement"
Interval Defines the period in [min] between two successive analyses.
Correction factor Any correction factor that can be used in calculations. "Method type" = "Mea
Comment A comment was entered before performance of the analysis in the
For Series, the following parameters can be modified in the Series item dialog.
1 To do this, select a defined series (in the Series list dialog) (see: "Series (page60)").

68 Analysis Sequence

The next analysis does not begin until the interval duration has elapsed.
Start analysis dialog.
Method type "Adjustment" / "Test" = The standards will be shown additionally.
"Continuous run" acti vated
surement"
-
The corresponding dialog box containing the defined series items appears.
2 Touch Start.
The Start analysis dialog box appears.
3 Touch the button "Entries"
The Series item dialog of the selected series opens.
4 If you touch an entry of the type "Method" or "Product" or touch the button Insert, the following parameters
will be available:
Parameters Description Displayed if
Item type The series item can either be a product or a method (only if at
least one product is defined in the system). Product ID Selection of product "Item type" = "Product" Method ID For "Item type"="Method", it is possible to select the relevant
method here.
For "Item type"="Product", the associated method is displayed
here (Info Field). Method type Displays the type of method, e.g. measurement ­Sample ID Input of the sample ID "Method type" = "Mea
surement"
Correction factor Any correction factor that can be used in calculations. "Method type" = "Mea
surement"
Number of entries
For a "Method" of the type "Adjustment" and "Test", you can modify the associated standard data using the "Standard" parameter, see also "Methods and products: Methods> Standard data (page40)".
Standard For "Method type"="Adjustment", it is possible to modify the stan
To define the quantity of series entries. -
"Method type" = dard data. For "Method type"="Test", the standard used is displayed (Info Field).
"Adjustment"
"Method type" = "Test"
-
-
10.2Start from Editor (Methods/Products/Series)
If a method, a product or a series is selected, then these can be started via Start. Start from the editor always proceeds via the Start analysis dialog.
10.3Start from Homescreen
In the Homescreen the Start button is also found that leads you directly to the Start analysis dialog. In it the method, productor series is listed that you most recently started. By touching Start again the analysis is imme diately started.
10.4Start via Shortcuts
You can create shortcuts for methods/products/series and manual operations that enable an immediate analy sis start. You can create a shortcut in the Start analysis dialog via the button AddToHome. In the process you can define the position on the Homescreen and the title of the shortcut. You can also continue to define whether after touching the shortcut you are directed via the Start analysis dialog or whether the analysis is to be started immediately. After it is created, the shortcut will appear on the selected position on the home screen and can be selected there by touching it.
10.5Start with Barcode Reader
The following two barcode reader types can be connected to the device:
Handheld readers
Built-in reader, this can be integrated in the sample changer (SC30 or InMotion).
69Analysis Sequence
Under Home > Setup > Hardware > Peripherals > Barcode readers you can configure the barcode readers. Here the information contained in the barcode and the format can be specified. If the barcode contains the sample, method or product ID you can also activate the parameter "Immediate start". When you are in the Start analysis dialog and scan in the barcode of a sample with a new sample ID, the bar code is transferred to the parameter "Sample ID". The task starts immediately. The most recently completed method starts with the new sample ID.
The specified data from standards can be scanned in for adjustments and tests. To do this, start an adjusting and test method. In the Start analysis dialog touch the Default, whereupon the Default Data dialog opens. This displays the standard data that you scanned in with the barcode.
Test methods can be started via barcode.
10.5.1Start with Handheld Reader
If you are in the Homescreen, the online screen or in the task list, you can start an analysis by scanning in a barcode. Depending on the information in the Barcode reader parameters dialog (Home > Setup >
Hardware > Peripherals > Barcode readers), a distinction is made between the following sce
narios:
1.
Parameter "Barcode information" = "Sample ID": After the barcode has been scanned, the Start analysis dialog is displayed. Here you can select the method/ the product, with which the analysis is to be performed. By pressing Start the analysis is started.
2.
Parameter "Barcode information" = "Sample ID/Method ID" or "Sample ID/Product ID": After the barcode has been scanned, the Start analysis dialog is displayed. By pressing Start the analysis is started.
3.
Parameter "Barcode information" = "Method ID" or "Product ID": After the barcode has been scanned, the Start analysis dialog is displayed. Here you can enter the sample ID.
If "Immediate start" is activated in the setup of the barcode reader, then the analysis is immediately started, i.e. without the Start analysis dialog being displayed. The sample ID defined in the method is adopted.
10.5.2Start with Built-in Barcode Reader
If a built-In barcode reader is connected, the button BC Start appears in the Homescreen (this button is only activated, if no task is lined up). By pressing this button the Start analysis dialog is displayed. Depending on the information in the Barcode reader parameters dialog (Home > Setup > Hardware >
Peripherals > Barcode readers), a distinction is made between the following scenarios:
1.
Parameter "Barcode information" = "Sample ID": In the Start analysis dialog you can select the method/ product, with which the analysis is to be performed (all analyses are performed with the same method/ product). After Start is pressed, the barcode of the first sample vial is scanned in and the first analysis started. The vials are sequentially processed until an empty item appears on the sample changer (SC30 or InMotion).
2.
Parameter "Barcode information" = "Sample ID/Method ID" or "Sample ID/Product ID": After Start is pressed, the barcode of the first sample vial is scanned in and the first analysis started. The vials are sequentially processed according to the method/product information on the barcode. If there is an empty position on the sample changer (SC30 or InMotion), the analysis sequence is terminated.
3.
Parameter "Barcode information" = "Method ID" or "Product ID": After Start is pressed, the barcode of the first sample vial is scanned in and the first analysis started. The vials are sequentially processed according to the method/product information on the barcode. The sample ID is taken over from the method or from the Start analysis dialog. If there is an empty position on the sam ple changer (SC30 or InMotion), the analysis sequence is terminated. By touching Stop or Reset you can cancel the analysis sequence at any stage.
10.5.3Start with ErgoSens
The ErgoSens is a movement sensor. A hand movement in front of the sensor immediately starts the method, productor series most recently used by you.
70 Analysis Sequence
If you have activated the parameter "Confirm end of the analysis", at the end of the analysis a confirmation via the ErgoSens takes place; instead of confirmation by OK.
An analysis start by barcode with the ErgoSens can be initiated with the button Barcode start (for SC30) (if the last start was initiated with this button).
10.6Continuous run
At the start of a measuring method or product by the editor, homescreen or a shortcut, you can select "Continu ous run". At the end of an analysis, the same analysis will start again automatically after the defined wait time (interval). If a sample changer (SC30 or InMotion) is used, the sample vials are processed successively until an empty space is found on the sample changer. The sample ID receives the suffix "__01" and is incremented.
An analysis with the "Continuous run" parameter activated is terminated in the following ways:
- via Stop button,
- in the event of a fatal error,
- if no more sample vials are available (SC30 or InMotion).
10.7Analysis Termination
Ongoing analyses can be canceled in the online screen via the button Stop or Reset (detailed information on the online screen can be found in"Tasks and online screen (page73)"). The current analysis is stopped and the processing of the next task is interrupted. To continue the waiting tasks, open the task list (with the Tasks button) and press Continue.
10.8Errors in the Analysis Sequences
Analyses can exhibit faults. Three types of faults can be identified during an analysis sequence:
Error (e.g.: max. fill duration exceeded)
Error with termination (e.g.: sample needle on the SC30 block)
Critical error (e.g.: fan stopped)
10.8.1Malfunction Types: Error
Malfunctions of the type Error are for example:
Max. filling duration exceeded
Max. meas. duration exceeded
The malfunction Error triggers the following behavior:
A message appears with information and hints on the error.
The task is continued to completion, however the method functions "Fill" and "Measure" are skipped.
The processing of further tasks is interrupted. Exception: With the SC30 or InMotion autosampler the tasks list is only interrupted, if the error occurs in one of the method functions "Clean" or "Cell test" (only for the density - module).
The analysis with the status "Error" is listed in the analysis list under Home > Results.
Not all errors on the pH or conductivity meter can be displayed on the refractometer.
10.8.2Malfunction Types: Terminate error
Malfunctions of the type Error with termination are for example:
Errors with respect to temperature adjustment
Sample needles from sample changer (SC30 or InMotion) blocked.
The malfunction Error with termination triggers the following response:
An information message on the relevant error appears.
The task is terminated immediately.
The processing of further tasks is interrupted. No further actions are executed, including cleaning or move ment of the sample changer (SC30 or InMotion).
71Analysis Sequence
10.8.3Malfunction Types: Critical error
Malfunctions of the type Critical error are for example (also see "Annex: Critical error":
Fan stopped
Temperature sensor malfunction
Temperature too high
The malfunction Critical error triggers the following response:
A message appears with information on the error.
The device switches off automatically after 20 seconds.
Critical errors can also occur when no task is running.
72 Analysis Sequence
11Tasks and Online Screen
If an analysis is started, the online screen is displayed. If a task is already running and a new analysis is started, the task list with the waiting tasks is displayed.
11.1Tasks
Tasks are processed in sequence. No tasks can be performed in parallel. The task list shows tasks that are currently in the queue and tasks that are currently running.
The task list offers the following options:
Display of online screen: By touching the list entry of the ongoing task, the online screen is displayed.
Interrupt: By touching the button Interrupt, the processing of the tasks is interrupted. The ongoing task is then carried out to completion. By touching Resume you can continue the processing of the task.
Remove all: This button is only visible, if the task list is interrupted and no more tasks are running. By touching this but ton all lined up tasks are removed.
Move tasks: Touch a task. By changing the "number", you can move the task.
Delete individual tasks: Touch a task. By touching the button Remove the task is deleted from the list.
Editing the sample ID Entering a comment.
11.1.1"Tasks" Button

The Tasks button is located at the top right of the display. If it is activated, as soon as at least one task is lined up.

By touching the button Tasks you go to the online screen or the task list, if several tasks are lined up.
Status display of the button Tasks:
Blue: No task is lined up.
Yellow: A task is running right now.
Yellow / blue blinking: A task has been finished and is waiting for confirmation
Red: The task list is interrupted and no task is running.
11.2Online Screen
The online screen displays the status and the readings of the ongoing analysis.
11.2.1Method type: Measurement
The info boxes at the top of the online screen show either the method or product ID, method or product title and the sample ID as well as the current analysis sequence step. At the bottom of the screen the set and actual temperatures (Tset and Tcell) are displayed. The calculations (Rx) that are displayed can be defined using the method function "Online Display". When you touch the button for individual calculations, the data relating to the calculation concerned are dis played. From here you can select another calculation (specified in the method) via the button Select result that is to be displayed during the runtime (each result Rx can only be displayed once). In the online screen the button Calculated results is displayed that shows all calculations (Rx) for the method.

73Tasks and Online Screen

11.2.2Method Type: Adjustment
The info fields at the top of online screen show the method ID, the adjustment standards and the current analy sis sequence step. At the bottom of the display the set and actual temperatures are shown. When the method is running the refractive index (incl. standard value) is displayed.
11.2.3Method Type: Test
The info fields at the top of online screen show the method ID, the test standards and the current analysis sequence step. At the bottom of the screen the set and actual temperatures are displayed. During the method the refractive index (incl. nominal value) and deviation from the nominal value (incl. tolerance) are displayed online.
11.2.4Method Type: Clean
The info fields at the top of online screen show the method ID and the current analysis sequence step. At the bottom of the screen the set and actual temperatures are displayed.
74 Tasks and Online Screen
12Appendix
12.1Raw Data
Refractometer measuring cell
nD Refractive index nD1... nD10 Single values from multiple measurement
Common raw data
Tset Set temperature Tcell Actual temperature at the time of measured value acquisition SD Standard deviation for multiple measurements. MaxSD Maximum permitted absolute standard deviation of measurements if
p Atmospheric pressure f Correction factor R1…R20 Results from "Calculation" method function TC1…TC5 Results from "Temperature compensation" method function H[ ] Auxiliary values (global variables) AuxInst1[1] ... AuxInst1[x],
AuxInst2[1] ... AuxInst2[x]
RLim[1] ... RLim[x] The result describes whether the limit is within or outside of the result
MeasOK Result for the measurement status (0: "error" - measurement failed, 1:
TIME Duration from the start of the method to the time of calculation of
DRY Result of the "Clean" method function for "Drying mode" = "Automatic"
TE Result of the test of the method of type "Test" (0:"not passed", 1:
ADJAN Result of the test of the method of type "Test" (0:"not passed", 1:
t Measurement duration (Method type "Measurement" or "Test"). t1 Measurement duration standard 1 (method type "Adjustment"). t2 Measurement duration standard 2 (method type "Adjustment").
multiple measurements are taken.
The result of the "Auxiliary instrument" method function from the sequence of the external auxiliary instrument (see also Methods and products: MethodsMethod functions> Auxiliary instrument (page53)).
limits (0: within limits, 1: outside limits).
"OK").
"TIME".
(0:"not passed", 1: "passed").
"passed").
"passed")
Density - module
OSC Oscillation: Actual measurement signal OSC1... OSC10 Single oscillation from multiple measurements. d Density
• During the measurement, the extrapolated density is displayed as soon as it is received.
• If viscosity correction is activated, the viscosity-corrected density is
transferred at the end of the measurement. d1...d10 Single values from multiple measurements. dRaw Uncorrected density. Differences to d:
• During the measurement, the non-extrapolated density is displayed.
• At the end of the measurement, the viscosity-corrected density is not
transferred, even if viscosity correction is activated. dA Apparent density SG Specific weight: Density of the sample divided by the density of water
at the measurement temperature. SGA Apparent specific weight

75Appendix

SG4 Specific weight 4oC: Density of the sample divided by the density of
water at 4oC. SG4A Apparent specific weight 4oC SG4 Specific weight 60oF: Density of the sample divided by the density of
water at 60oF SG60A Apparent specific weight 60oF F Adjustment factor CT Result of the "Cell test" method function (0: "not passed", 1:
"passed"). DevCT Deviation of the cell test from the nominal value.
External instruments
pH pH value DevpH Maximum deviation in the pH measurement. MaxDevpH The maximum permitted absolute deviation of measured values for pH. TpH Temperature pH COND Conductivity DevCOND Maximum deviation in the conductivity measurement. MaxDevCOND The maximum permitted absolute deviation of measured values for
conductivity. TCOND Temperature conductivity SDCOL Standard deviation of the measured values for color in the case of mul
tiple measurements. MaxSDCOL Maximum permitted absolute standard deviation of the measured val
ues for color, if multiple measurements are taken.
12.2Result proposals
List of the predefined results
Application d: Others
Twaddell number Based on the specific weight SG at meas. temperature.
Baumé degree Based on the specific weight SG at meas. temperature.
Application d: Alcohol
Jap. Sake deg. Sake Grade (Japan), based on the specific weight SG, measured at
Milk degrees Based on the specific weight SG at meas. temperature.
Alc. (%v/v) OIML @20°C Alc. (%v/v) OIML @60°F Alc. (%v/v) OIML @15°C
Alc. (%w/w) OIML % weight ethanol, as per the OIML R-22 (International Organization of
Alc. d @20°C Alc. d @60°F Alc. d @15°C
°Twad = 200 * (SG-1)
For samples heavier than water:
H. °Be = ((1/SG) - 1) * -144.3), at 15°C
H. Bé (US) = ((1/SG) - 1) * -145), at 60°F
For samples lighter than water:
Lt. °Bé = 10 + 144.3 * ((1/SG) - 1), at 15°C
Lt. °Bé (US) = 10 + 145 * ((1/SG) - 1), at 60°F
15°C
Jap. Sake deg. = ((1/SG) - 1) * 1443
Milk deg. =1000 * (SG-1)
% volume of ethanol at the stated temperature, as per the OIML R-22
(International Organization of Legal Metrology, Recommendation 22:
alcohol tables), temperature scale IPTS 68, based on the absolute
density (in vacuum)
Legal Metrology, Recommendation 22: alcohol tables), temperature
scale IPTS 68, based on the absolute density (in vacuum)
True density (in vacuo) of an ethanol sample at the stated tempera
ture, as per OIML R-22
76 Appendix
Alc. dA @20°C Alc. dA @60°F Alc. dA @15°C
Alc. SGA @20°C Alc. SGA @60°F Alc. SGA @15°C
Alc. (Proof) USA US proof degree at 15.56°C (60°F), based on the true density Alc. (%v/v) HM C&E
Alc. (%w/w) HM C&E Alc. (%v/v) Canada @20°C % volume ethanol at 20°C, as per the Canadian ethanol tables Alc. (proof) UK UK proof degree at 15.56°C (60°F) Alc. (%v/v) OIML ITS90 @20°C
Alc. (%v/v) OIML ITS90 @60°F Alc. (%v/v) OIML ITS90 @15°C
Alc. (%w/w) OIML ITS90 % weight ethanol, as per the OIML R-22 (International Organization of
Alc. (%v/v) AOAC % volume ethanol at 60°F, or % weight ethanol, as per AOAC tables
Alc. (%v/v) Gay-Lussac %-volume ethanol at 15°C, as per OIML R-22 (International Organiza
Apparent density (in Air) of an ethanol sample at the stated tempera ture
Apparent relative density (in air), SG(t/t), of an ethanol sample at the stated temperature
% volume ethanol as per H.M. C&E Table, at 20°C
% volume of ethanol at the stated temperature, as per the OIML R-22 (International Organization of Legal Metrology, Recommendation 22: alcohol tables), new temperature scale ITS 90, based on the true den sity (in vacuum)
Legal Metrology, Recommendation 22: alcohol tables), new tempera ture scale ITS 90, based on the true density (in vacuum)
(American Organization of Analytical Chemists), based at the true den sity at 20°C.
tion of Legal Metrology, Recommendation 22: (alcohol tables), new temperature scale IPTS 68, based on the true density (in vacuum)
Application d: Sugar
Plato d Extract-content in percentage weight (% w/w), Plato table, from true
density at 20°C
Brix d NBS 113 Saccharose content in percentage weight (% w/w), NBS table 113,
from true density at 20°C
Brix d Emmerich Saccharose content in percentage weight (% w/w), according to A.
Emmerich, Zuckerindustrie 119 (1994), from true density at 20°C
HFCS42 d %-weight HFCS syrup (High Fructose Corn Syrup) with 42 % fructose
fraction, based on true density at 20°C. Must be measured at 20°C.
HFCS55 d %-weight HFCS syrup (High Fructose Corn Syrup) with 55 % fructose
fraction, based on true density at 20°C. Must be measured at 20°C.
Invert sugar d %-weight invert sugar, based on true density at 20°C. Must be mea
sured at 20°C.
KMW d Klosterneuburg sugar grade (Austria). Precise sugar content in grape
juice. Based on true density at 20°C. Babo (KMW) d Sugar content in grape juice (Italy). Based on true density at 20°C. Oechsle d Oechsle degree in grape juice, based on the specific weight at 15°C.
°Oe = (SG-1) * 1000, with d in g/cm3
Application nD: Others
Zeiss (14.45) Zeiss number, based on nD(20°C). Pure water gives a value of
14.45.
Zeiss (15.0) Zeiss number, based on nD(20°C). Pure water gives a value of 15.0.
Application nD: Sugar
Brix nD @ Tx Saccharose content in %.weight as per ICUMSA, 20th session in Col
orado Springs (1990). Also corresponds to OIML R 108 (1993).
Result without temperature compensation to 20°C. Brix nD comp 20C Saccharose content in %.weight as per ICUMSA, 20th session in Col
orado Springs, 1990. Also corresponds to OIML R 108 (1993). Result
compensated to 20°C. Invert sugar nD % weight invert sugar, as per ICUMSA, 20th session in Colorado
Springs (1990). Result without temperature compensation to 20°C.
77Appendix
HFCS 42 nD % weight HFCS syrup (High Fructose Corn Syrup) with 42 % fructose
ρ
HFCS 55 nD % weight HFCS syrup (High Fructose Corn Syrup) with 55 % fructose
Oechsle nD Oechsle degree from grape juice, based on the refractive index at
12.3Refractive Index for Water
fraction. Based on the refractive index at 20°C. Must be measured at 20°C.
fraction. Based on the refractive index at 20°C. Must be measured at 20°C.
20°C.
Temp.
[°C] 5 1.33388 25 1.33250 45 1.32985 65 1.32628 6 1.33385 26 1.33240 46 1.32969 66 1.32609 7 1.33382 27 1.33229 47 1.32953 67 1.32589 8 1.33378 28 1.33217 48 1.32937 68 1.32568 9 1.33373 29 1.33206 49 1.32920 69 1.32548 10 1.33369 30 1.33194 50 1.32904 70 1.32527 11 1.33363 31 1.33182 51 1.32887 71 1.32507 12 1.33358 32 1.33170 52 1.32870 72 1.32486 13 1.33352 33 1.33157 53 1.32852 73 1.32464 14 1.33345 34 1.33144 54 1.32835 74 1.32443 15 1.33338 35 1.33131 55 1.32817 75 1.32422 16 1.33331 36 1.33117 56 1.32799  17 1.33324 37 1.33104 57 1.32781  18 1.33316 38 1.33090 58 1.32762  19 1.33307 39 1.33075 59 1.32744  20 1.33299 40 1.33061 60 1.32725  21 1.33290 41 1.33046 61 1.32706  22 1.33280 42 1.33031 62 1.32687  23 1.33271 43 1.33016 63 1.32668  24 1.33261 44 1.33001 64 1.32648
Refractive index (nD)
Temp.
[°C]
Refractive index (nD)
Temp.
[°C]
Refractive index (nD)
Temp.
[°C]
Refractive index (nD)
12.4Density
Density of air
= 0.0012930/(1+0.00367*t) * (p/1013.25) t = temperature in [°C] p = absolute atmospheric pressure in [mbar] or [hPa]
Temp. [°C]
0 0.00129 46 0.00111 1 0.00129 47 0.00110 2 0.00128 48 0.00110 3 0.00128 49 0.00110 4 0.00127 50 0.00109 5 0.00127 51 0.00109 6 0.00127 52 0.00109 7 0.00126 53 0.00108 8 0.00126 54 0.00108 9 0.00125 55 0.00108 10 0.00125 56 0.00107 11 0.00124 57 0.00107 12 0.00124 58 0.00107
Density [g/cm3]
Temp. [°C]
Density [g/cm3]
78 Appendix
Temp. [°C]
Density [g/cm3]
Temp. [°C]
Density [g/cm3]
13 0.00123 59 0.00106 14 0.00123 60 0.00106 15 0.00123 61 0.00106 16 0.00122 62 0.00105 17 0.00122 63 0.00105 18 0.00121 64 0.00105 19 0.00121 65 0.00104 20 0.00120 66 0.00104 21 0.00120 67 0.00104 22 0.00120 68 0.00103 23 0.00119 69 0.00103 24 0.00119 70 0.00103 25 0.00118 71 0.00103 26 0.00118 72 0.00102 27 0.00118 73 0.00102 28 0.00117 74 0.00102 29 0.00117 75 0.00101 30 0.00116 76 0.00101 31 0.00116 77 0.00101 32 0.00116 78 0.00101 33 0.00115 79 0.00100 34 0.00115 80 0.00100 35 0.00115 81 0.00100 36 0.00114 82 0.00099 37 0.00114 83 0.00099 38 0.00113 84 0.00099 39 0.00113 85 0.00099 40 0.00113 86 0.00098 41 0.00112 87 0.00098 42 0.00112 88 0.00098 43 0.00112 89 0.00097 44 0.00111 90 0.00097 45 0.00111
Density of water
According to F. Spieweck, H. Bettin; Review: Solid and liquid density determination Part 1: tm – Technisches Messen 59 (1992) 6, p. 237 – 244 Part 2: tm – Technisches Messen 59 (1992) 7/8, p. 285 – 292
Temp. [°C]
Density [g/cm3]
Temp. [°C]
Density [g/cm3]
0 0.99984 46 0.98979 1 0.99990 47 0.98936 2 0.99994 48 0.98892 3 0.99996 49 0.98848 4 0.99997 50 0.98803 5 0.99996 51 0.98757 6 0.99994 52 0.98711 7 0.99990 53 0.98664 8 0.99985 54 0.98617 9 0.99978 55 0.98569 10 0.99970 56 0.98520
79Appendix
Temp. [°C]
11 0.99960 57 0.98471 12 0.99950 58 0.98421 13 0.99938 59 0.98370 14 0.99924 60 0.98319 15 0.99910 61 0.98267 16 0.99894 62 0.98215 17 0.99877 63 0.98162 18 0.99859 64 0.98109 19 0.99840 65 0.98055 20 0.99820 66 0.98000 21 0.99799 67 0.97945 22 0.99777 68 0.97889 23 0.99754 69 0.97833 24 0.99730 70 0.97776 25 0.99704 71 0.97719 26 0.99678 72 0.97661 27 0.99651 73 0.97602 28 0.99623 74 0.97543 29 0.99594 75 0.97484 30 0.99565 76 0.97424 31 0.99534 77 0.97363 32 0.99502 78 0.97302 33 0.99470 79 0.97241 34 0.99437 80 0.97179 35 0.99403 81 0.97116 36 0.99368 82 0.97053 37 0.99333 83 0.96989 38 0.99296 84 0.96925 39 0.99259 85 0.96861 40 0.99221 86 0.96796 41 0.99183 87 0.96730 42 0.99143 88 0.96664 43 0.99103 89 0.96598 44 0.99062 90 0.96531 45 0.99021
Density [g/cm3]
Temp. [°C]
Density [g/cm3]
12.5Critical errors
Error code Error
E007 Fan 1 stopped
E030 Fan 2 stopped (only for the density - module)
E028 Temperature sensor malfunction E029 Thermomodule malfunction E031 Hardware error E033 Memory error E064 Maximum temperature exceeded E075 Electric current too high E076 Voltage too high
80 Appendix
If a critical error occurs, a message is displayed. The instrument then switches off automatically after 20 sec onds. If the problem is not resolved by a restart, please contact your authorized METTLER TOLEDO service agent.
81Appendix

Index

Numerics
24V Output 24
A
Adjustment
Method function 52 Method syntax 39 MF sample 43
Adjustment data
Export 35
Adjustment parameters
Setup, Cell 17
Adjustment sets
Create 16 Delete 16
Analyses
Stopping 71
Analysis
Start 68 Analysis status 62 AtmoSens
Verify availability 24 Audio-Signal
Setting 25 Automation
Manual operation 65 Automation units
>Delete 19
Create 18
Types 17 Auxiliary Instrument 53 Auxiliary Instruments
24V Output 24
USB-RS-232 24 Auxiliary values
Create 31
Method function 52 AuxInst 75
Variable auxiliary instru
ment
B
Barcode reader
Start analysis 70 Beep
Setting 25 Bubble Check 46 Buttons
Info 12
Reset 12
C
Calculation, method function 47 Calculations in products 59 Cell
Conditioning 67
Maintenance & Service 34
Manual operation 66 Cell parameters 17 Cell test 67
54
Cell test, method function 51 Cell update 34 Cell, method function "Measure"
Method type "Adjustment" 45 Method type "Measure" 45 Method type "Test" 45
Clean
Method syntax 40 Clean, method function 48 Color scales 20
Formula editor 40 Condition, method function 40 Conditioning
Cell 67 Configuration dialog 36 Configuration, method function 42 Configure barcode reader 21 Configure USB stick 21 Continuous run 71 Control type
Output 24V 53
USB-RS-232 53
D
Data
Analysis results 63 Data storage 26 Date and time 26, 26 Define account policies 27 Density of air
Temperature dependency 78 Density of water
Temperature dependency 79 Drying 49, 65
E
ErgoSens
Activate 23
Start analysis 70 Error 71 Error detection, method function "Measure"
Method type "Adjustment
/ Test" Errors, critical 80 Expired Resources 15 Export
Adjustment, test, mea
surement history External devices
Delete 19 External instruments
Colorimeter 20
Conductivity meter 19
Create 19
Edit 19
Install 19
pH meter 19
Types 19 Extinction
Formula editor 40
47
35
Index82
F
Filling, method function 44 Fingerprint reader 21 Firmware
List 34 Firmware history 33 Formula editor 40 Formula syntax 41 Formulas
Example 41
in method functions 40
Initials 41
G
Global settings
Analysis and resources
behavior
Physical properties 29
System 26
Types 26
User management 26
H
Handheld reader
Start analysis 70 Hardware
Auxiliary Instruments 24
List 34 History
Adjusting, test entries 17 Homescreen
Buttons 12
I
Identification
Define system settings 26 Import/Export of data 32 Information
Analysis results 63 Input fields
Types 13 Instruction, Method function 52 Instrument update 33 Instruments
Description 8
L
Language
Setting 25 Layout of Terminal 12 LevelSens
Activate 23 Line rinse 56 Linking of methods with prod ucts Lists
Sorting 14 Lovibond 20
M
Maintenance & Service
Firmware history 33
28
57
Functions 31 Import / Export of data 32 MT Service 32 Reset to factory settings 33 update 33
Malfunction
Critical error 72
Error with termination 71 Mathematical Functions and Operators Measure
Subfunction "Cell" 45
Subfunction "Error detec
tion"
Subfunction "Measure
ment acquisition" Measure, method function 45 Measured value acquisition, method function "Measure"
Method types "Measure",
"Adjustment", "Test" Measurement
Method syntax 38 Measurement history
Export 35 Memory copy 32 Menu structure 15 Method function
Wait 42 Method functions
Adjustment 52
Aux. value 52
Calculation 47
Cell test 51
Clean 48
Condition 40
Configuration 42
Filling 44
Instruction 52
Measure 45
Online display 50
Report 50
Temperature compensa
tion
Test 52
Title 42
Wait 53 Method Functions, Parameters
Auxiliary Instrument 53 Method quantity 36 Method syntax
Adjustment 39
Clean 40
Measurement 38
Test 39 Method type
Adjustment 43
Test 43 Method types 36 Methods
Create 36
Linking with products 57
41
47
46
46
51
Index 83
Minolta 20 MT Service, Maintenance & Ser vice
N
Network – settings 23 Network - storage 23
O
Online display, method function 50 Online Screen
Adjustment 74 Clean 74
Measurement 73 Online screen and tasks 73 Operation of the touchscreen 12 Output 24V 53
P
Parameters
Products 57 Park 56 PC – settings 22 Peripherals
Barcode reader 21
Device types 20
Fingerprint reader 21
Network – settings 23
PC – settings 22
Printer 22
USB data export 22
USB stick 21 PowerShower 54 Printer
Setup 22
USB data export 22 Printing a PDF file 36 Printing a report
Manually 62 Products
Calculations 59
Create 57
Linking with methods 57
Parameters 57 Pump sample 66
Q
Quantity of methods 36
R
Raw data 75 Report, method function 50 Reset to factory settings 33 Results 62 Rights for users 27 Rinse 65 Rotate turntable 66
S
Sample, method function 42 Screen
Setting 25 Sensors
32
Activate 23
Series
Create, delete 60
Tasks 60 Setup-Menu 15 Shortcuts
Create 26
Homescreen 14 Sorting lists 14 Standard data 40 Standby display 12 Statistics
Results 63
Rolling statistics 63 Status
Analyses 62 Stir 55 Syntax, methods 38 System settings
Date and time 26
Define 26
T
Tables
Create 30 Tasks and online screen 73 Temperature compensation, method function Test data
Export 35 Test parameters
Setup, Cell 17 Test sets
Create 16
Delete 16 Test, method function 52 Test, method syntax 39 Test, method type 43 Title, method function 42 Touchscreen 12 Transmission
Formula editor 40
U
Update, firmware 33 USB compact printer 22 USB data export 22 USB-RS-232 24, 53 USB-RS232 data export box 22 User – settings
Shortcuts 26 User groups 27 User Interface
Input Fields 13 User management
Administer users 27
Define account policies 27 User rights 27 User settings
Audio-Signal 25
Beep 25
Keyboards 26
51
Index84
Language 25 Screen 25
W
Wait, method function 53 WasteSens
Activate 23
Index 85
Mettler-Toledo AG, Analytical
CH-8603 Schwerzenbach, Switzerland Tel. +41 (0)44 806 77 11 Fax +41 (0)44 806 73 50 www.mt.com
www.refractometry.com
For more information
Subject to technical changes. © Mettler-Toledo AG 51710794B
*51710794*
Loading...