4Installation and Putting Into Operation
4.1Scope of delivery
1 Open the packaging and carefully remove all components.
2 Check the delivered items.
The standard scope of delivery contains the following items:
●
Balance
●
Weighing pan and weighing pan support
●
Protective cover for load cell cone (mounted)
●
Protective cover (mounted)
●
Stackable cover
●
Universal AC/DC adapter (country specific)
●
Operating instructions printed or on CD-Rom depending on the country.
●
Quick Guide (English) depending on the country.
●
EC declaration of conformity
4.2Selecting a location
The correct location makes an important contribution to the accuracy of the weighing results of preci
sion balances. Select a stable, vibration-free position that is as horizontal as possible. The surface
must be able to safely carry the weight of a fully loaded balance.
Observe ambient conditions see Technical Data (page59).
Avoid the following:
●
Vibrations
●
Excessive temperature fluctuations
●
Direct sunlight
●
Powerful drafts (e.g. from fans or air conditioners)
4.3Leveling the balace
The balances have a level indicator and two or four adjustable
leveling feet to compensate for slight irregularities in the surface
of the weighing bench. The balance is exactly horizontal when
the air bubble is in the middle of the level glass.
Align the balance horizontally by turning the leveling feet of the
balance housing until the air bubble is in the inner circle of the
level indicator.
Note: The balance should be leveled and adjusted each time it is
moved to a new location.
13Installation and Putting Into Operation