Mettler Toledo AB-S/FACT, PB-S/FACT, AB-S/PH, PB-S/PH, PB-L/PH Operating Instructions Manual

Operating instructions

METTLER TOLEDO B-S/FACT line of balances
• AB-S/FACT
• PB-S/FACT
off
O/T
On
off
O/T
METTLER TOLEDO
C
CLASSIC Plus
F
Cal/Menu
On
off
O/T
METTLER TOLEDO
C
Plus
CLASSIC
F
Cal/Menu
METTLER TOLE
DO
us
C
Pl
CLASS IC
F
Cal/Menu
On
off
O/T
METTLER TOLEDO
C
Plus
CLASSIC
F
Cal/Menu
www.mt.com/classic

Operating instructions in a nutshell

Press key briefly
Press and hold key
down until the desired display appears
automatic sequence
Switching on
On
88888888
0.00 g
Switching off
Off
OFF
Simple weighing
0.00 g
On
off
O/T
METTLER TOLEDO
C
• 1182.03 g
F
Cal/Menu
1250.00 g
Adjustment options
Cal/Menu
±
±
Taring
METTLER T
OLEDO
C
F
Cal/Menu
METTLER TOLEDO
C
F
Cal/Menu
O/T
METTLER TOLEDO
C
F
Cal/Menu
0.00 g
121.01 g
0.00 g
95.97 g
-121.01 g
O n
off
O/T
#
On
off
O/T
On
off
ŸÖEMU
CAL int
CAL E
AutoCal
FACt on
FACt off
Piece counting*
On
off
O/T
M
ETTLER
TO LED
O
C
F
Cal /Menu
#
Cal
On
o ff
O/T
METTLER TOLEDO
C
F
Cal
Cal
/Menu
F
±
On
o ff
O/T
METTLER TOLEDO
C
F
Cal /M
enu
Percent weighing*
On
o ff
O/T
METTLER TOLEDO
C
F
Cal /M enu
F
±
On
off
O/T
M ETTLER
T
OLEDO
C
F
Cal/Menu
Dynamic weighing*
Automatic start (Dyn A)
0.00 g
0.00 g
12.07 g
SEt 10
SEt 20
20
244
147.25 g
0.00 g
PCS
PCS
PCS
PCS
0.00 g
• 257.35 g
O n
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
‹ 260.33 g
Manual start (Dyn M)
0.00 g
0.00 g
• 257.35 g
O n
off
O/T
METTLER TO
LEDO
C
F
Cal/Menu
±
<
<
<
-- 3 --
<
-- 1 --
<
<
<
<
-- 3 --
<
-- 1 --
12.07 g
SEt100 %
100.0 %
Unit switching*
101.6 %
*These functions must be activated in the menu (section 4.3.4)
<
‹ 260.33 g
22.00 g
0.78
Free factor / selectable display increment*
O n
off
O/T
METTLER TOLEDO
C
F
C a l /M e
n u
F
±
±
** 12, 3 ...9, 0
±
±
On
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
oz
19.50 g
FAC ŸÖ
1. 0000
0. 0000
**
0. 0000
(2x)
0 9000
StEP
0 0900
0. 1000
StorEd
19.50 g
17.5
19.50 g
17.5
19.50 g
#
#

Contents

1 Getting to know B-S/FACT balances ...............................................4
1.1 Introducing the B-S/FACT line of balances .........................................4
1.2 Layout of B-S/FACT balances ..........................................................5
1.3 Overview of key functions ...............................................................6
2 Startup ........................................................................................7
2.1 Unpacking / standard equipment .....................................................7
2.2 Cautionary notes ...........................................................................7
2.3 Installing the draft shield ................................................................8
2.4 Setting up, leveling, preparations for weighing below the balance,
connecting to power supply ............................................................9
2.5 Adjusting (calibration) .................................................................11
3 Weighing ...................................................................................13
3.1 On/off switching .........................................................................13
3.2 Simple weighing ..........................................................................13
3.3 Taring ........................................................................................14
3.4 METTLER TOLEDO DeltaRange balances ........................................15
4 Menu .........................................................................................16
4.1 Overview ....................................................................................16
4.2 Menu operation ...........................................................................17
4.3 Description of menu options .........................................................18
5 Functions ...................................................................................25
5.1 Piece counting ............................................................................25
5.2 Percent weighing .........................................................................26
5.3 Dynamic weighing .......................................................................27
5.4 Weighing with free factor and/or selectable display increments .........29
5.5 Switching weight units .................................................................31
6 Technical data, optional equipment .............................................32
6.1 Technical data ............................................................................32
6.2 Interface .....................................................................................34
6.3 MT-SICS Interface commands and functions ...................................35
6.4 Optional equipment .....................................................................37
6.5 Dimensional drawings .................................................................38
7 Appendix ...................................................................................40
7.1 Typical printouts from METTLER TOLEDO
RS-P26 and LC-P45 printers ........................................................40
7.2 What if ...? .................................................................................41
7.3 Maintenance and cleaning ............................................................42
3

1 Getting to know B-S/FACT balances

1.1 Introducing the B-S/FACT line of balances

4
AB-S/FACT
(0.1 mg / 0.01mg)
PB-S/FACT
On
off
O/T
METTLER TOLEDO
Plus
C
CLASSIC
F
Cal/Menu
On
off
O/T
METTLER TOLEDO
Plus
C
CLASSIC
F
Cal /Menu
PB-S/FACT (0.1/1 g)
Balance features
• The B-S/FACT balance line ranges from high-resolution analytical balances (AB-S/FACT) with a readability of 0.1 mg / 0.01 mg through to precision balances (PB-S/FACT) with a readability of 0.001g to 1g. The weighing ranges extend from 51g to 8.1kg.
FACT (Fully Automatic Calibration Technology), fully automatic, motorized adjustment (calibration) with internal weight.
• In addition to basic weighing operations such as weighing, taring and adjusting (calibration) the functions
Piece counting”, “Percent weighing”, “Free Factor” or “Dynamic weighing” (automatic or manual start)
On
off
O/T
METTLER TOLEDO
Plus
C
CLASSIC
F
Cal/Menu
can be activated.
• METTLER TOLEDO DualRange balances have two ranges. These models also have a fine (semimicro) ran- ge from 0 to 31 g respectively 61 g. In this fine range the balance shows the result with a higher resolution, i.e. with one decimal place more.
• METTLER TOLEDO DeltaRange balances also have a movable fine range, with 10 times smaller display increments, over the the entire weighing range.
PB-S/FACT (0.01g)
• Several B-S/FACT balances are fitted with a glass draft shield in the factory; with other models a draft shield is available as an optional extra.
• All balances are fitted with an RS232C interface as standard.
• All models have an internal adjustment weight.
Note
All B-S/FACT balances are available as certified versions. Please ask your METTLER TOLEDO dealer for de-
On
off
O/T
METTLER TOLEDO
Plus
C
CLASSIC
F
Cal/Menu
PB-S/FACT (0.001g)
tails.

1.2 Layout of B-S/FACT balances

6
5 4
3
2
On
off
O/T
1
METTLER TOLE
DO
us
C
Pl
CLASS IC
F
Cal/Menu
8
5
12
RS232C
1 Keys 2 Display with backlight (exept ABxx5-S/FACT) 3 Model plate with the following data:
“Max”: maximum capacity “d”: readability “Min”: minimum capacity (recommended minimum load; only relevant for certified balances) “e”: verification scale interval (smallest display increment tested during certification; only relevant for
certified balances) 4 Draft shield element 5 Weighing pan 6 Draft shield (standard supply with models AB-S/FACT and PBxx3-S/FACT) 7 Leveling feet 8 Hanger opening for weighing below the balance (underside of balance) 9 AC adapter socket
7
10 RS232C interface 11 Lug for optional antitheft device 12 Leveling control
7
11
10
9
7
5

1.3 Overview of key functions

C
O/ T
On
Off
Cal/ Menu
C
O/ T
On
Off
Cal/ Menu
6
The balances have two operator control levels: the weighing mode and the menu. The function of each indi- vidual key depends on the operator control level and how long the key is pressed.
Weighing mode
0.01 g
Menu
ŸÖENV
Key functions in weighing mode
Press briefly
OnSwitch on
Press and hold down
Off • Switch off balance (standby mode)
# • Zero/tare
F
CCancel function
Switch
Change settings
FCall function
A function must be activated in the menu, otherwise “F nonE” appears in the dis­play
± • Transfer weighing data via
interface
Confirm settings
Cal/MenuAdjust (calibrate)
• Show menu (hold key down until MENU appears)
Key functions in menu mode
Press briefly
C • Close menu (without
Press and hold down
----------
saving changes)
F
Change settings
± • Select menu options
----------
Cal/Menu • Save changes and close menu

2 Startup

2.1 Unpacking / standard equipment

On
off
O/T
METTLER TOLE
DO
C
F
C al
/M enu

2.2 Cautionary notes

All B-S/FACT balances are supplied in environmentally compatible packaging.
The standard equipment for every balance comprises:
AC adapter, to national codes
AC adapter with country-specific power cable (Models AB135-S/FACT / AB265-S/FACT)
Transparent plastic protective cover, fitted, to keep your balance clean at all times
Operating instructions, to ensure optimum utilization of your balance›s capabilities
CE declaration of conformity (in separate brochure 11780294)
AB-S/FACT with readability of 0.01/ 0.1 mg: Weighing pan ø 80 mm with draft shield element
• Draft shield “0.1 mg” (237 mm) PB-S/FACT with readability of 0.001 g: Weighing pan ø 100 mm with draft shield element
Draft shield “mg” (165 mm) PB-S/FACT with readability of 0.01 g: Weighing pan ø 180 mm with pan support
Retaining ring for protective cover, fitted PB-S/FACT with readability of 0.1/1 g: Weighing pan 180 x 180 mm with pan support
• Base plate and intermediate plate for protective cover,fitted
• B-S/FACT balances must not be operated in hazardous areas with the standard-supply AC adapter.
• Before connecting the AC adapter, verify that the voltage printed on it corresponds to the local mains voltage. If this is not the case, please contact your local METTLER TOLEDO dealer.
• B-S/FACT balances may only be used indoors in a dry environment.
• For use with an certified (CSA or equivalent) power supply, which must have a limited and SELV circuit output.
7

2.3 Installing the draft shield

8
Draft shield “0.1 mg” (237 mm):
ABxx5-S/FACT
1 2
1
2
On
off
O/T
METTLER TOLEDO
C
F
Cal
/Menu
On
off
O/T
1
On
off
O/T
METTLER TOLEDO
C
F
Cal
/Menu
2
METTLER TOLEDO
C
F
Cal/Menu

2.4 Setting up, leveling, preparations for weighing below the balance, connecting to power supply

The optimum location
The correct location makes an important contribution to the accuracy of the weighing results of high-resolution analytical and precision balances.
Hence, ensure a
• stable, vibration-free position as horizontal as possible
Avoid
• direct sunlight
• excessive temperature fluctuations
• drafts
The best location is on a stable bench in a corner protected against drafts, as far away as possible from doors, windows, radiators or the louvers of air conditioners.
Antitheft device
B-S/FACT balances are equipped with a lug for the optional antitheft device.
The antitheft device (cable with lock) is suitable for all models. It is available from METTLER TOLEDO unter the order number 590101.
9
10
Leveling
B-S/FACT balances have a spirit level and two adjustable leveling feet to compensate for slight irregularities in the surface of the weighing bench. The balance is exactly horizontal when the air bubble is in the middle of the level glass.
Procedure
Adjust the two leveling feet appropriately until the air bubble comes to rest exactly in the middle of the glass: Air bubble at “12 o›clock” turn both feet counterclockwise
Air bubble at “3 o›clock” turn left foot clockwise, right foot counterclockwise Air bubble at “6 o›clock” turn both feet clockwise Air bubble at “9 o›clock” turn left foot counterclockwise, right foot clockwise
Note
The balance should be leveled and adjusted (Section 2.5) each time it is moved to a new location.
Preparations for weighing below the balance
To carry out weighing operations below the balance, the special cover on the underside of the balance must be slackened (care: when doing this the balance may only be laid on its side, not turned upside down!), turned through 180° and retightened. This exposes the opening for the hanger, making weighing below the balance possible.
Connecting the power supply
Before connecting the AC adapter, check that the voltage printed on it corresponds to the local mains
voltage. If this is not the case, please contact your local METTLER TOLEDO dealer.
Plug the AC adapter into the AC adapter socket on the balance, and connect to the power supply. The balance performs a self-test. This test is finished when “OFF” appears. Press the «On» key briefly: the balance is in operational readiness. Before any work is performed with the
balance, it must be adjusted (Section 2.5).
Notes
To achieve accurate results with analytical balances (AB-S/FACT), these must be connected to the power supply for at least 60 minutes to warm up to operating temperature before the first weighing operation is carried out. An optional AccuPac B-S (rechargeable external battery) can be used to operate all B-S/FACT balances inde­pendently of the mains power supply

2.5 Adjusting (calibration)

To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its location. Adjusting is necessary:
• before the balance is used for the first time
• at regular intervals during weighing service
• after a change of location

2.5.1 Fully automatic adjustment (calibration) FACT

The factory setting is fully automatic adjustment FACT (Fully Automatic Calibration Technologiy) with the internal weight (see also Section 4.3.3). In this setting, you have no need worry about adjusting your balance. The balance adjusts itself automatically
• after the warm-up phase on connection to the power supply,
• when a change in the ambient conditions, e.g. the temperature could lead to a noticeable deviation in the
measurement.

2.5.2 Manual adjustment (calibration)

To obtain accurate results, the balance must be connected to the power supply for 30 minutes (AB-S/FACT analytical balances 60 minutes) in order to reach operating temperature before adjusting.
Cal/Menu
CAL int
à
------
à
CAL donE
à
0.0000 g
Adjusting with internal weight
To carry out this operation, in the second menu option (Adjustment) select “CAL int” (=factory setting)
(Section 4.1).
Cal
Unload weighing pan Press and hold the «Cal/Menu» key down until “CAL” appears in the display, then release key. The balance adjusts itself automatically.
The adjusting is finished when the message “CAL done” appears briefly in the display, followed by
“0.0000 g”. The balance is again in weighing mode and ready for operation.
11
Cal/Menu
Cal/Menu
On
off
O/T
C
On
off
O/T
C
12
Adusting with external weight
CAL E
à
To carry out this operation, in the second menu option (Adjustment) select “CAL E” Have required adjustment weight ready Unload weighing pan Press and hold the «Cal/Menu» key down until “CAL” appears in the display, then release key. The re-
CAL
à
quired adjustment weight value flashes in the display.
Place adjustment weight in center of pan. The balance adjusts itself automatically. When “0.00 g” flashes, remove adjustment weight.
2000.00
Cal
à
METTLER TOLEDO
F
Cal/Menu
------
0.00
Cal
à
Cal
à
METTLER TOLEDO
F
Cal/Menu
------
Cal
à
The adjusting is finished when the message “CAL done” appears briefly in the display, followed by
“0.00 g”. The balance is again in weighing mode and ready for operation.
Notes
• Because of certification legislation, the certified models cannot be adjusted with an external weight.
• The adjustment procedure can be terminated at any time with the «C» key. The message “Abort” appears
briefly to confirm that adjustment has been canceled, and the balance reverts to weighing mode.
CAL donE
à
0.00 g

3 Weighing

3.1 On/off switching

On
0.00 g
Off

3.2 Simple weighing

0.00 g
88888888
à
OFF
à
Switching on
Remove any load from weighing pan and press «On» key briefly. The balance performs a display test (all segments in the display light up briefly). When zero is displayed, the balance is ready for operation.
Switching off
Press and hold the «Off» key down until “OFF” appears in the display. Release the key.
Place weighing sample on the weighing pan.
On
off
O/T
METTLER TOLEDO
C
F
Cal /Menu
• 1182.03 g
Read the result.
Wait until the stability detector “
” disappears.
1250.00 g
13

3.3 Taring

On
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
#
0.00 g
121.01 g
0.00 g
14
Place empty container on the balance.
The weight is displayed.
Tare: press the «#» key briefly.
95.97 g
On
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
-121.01 g
On
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
Add weighing sample to container. The net weight is now displayed.
If the container is removed from the balance, the tare weight will be shown as a negative value. The tare weight remains stored until the «#» key is pressed again or the balance is switched off.
Note
With METTLER TOLEDO DeltaRange balances (next Section), the fine range with its 10 times smaller display increments is available again after every taring operation.

3.4 METTLER TOLEDO DeltaRange balances

#
812.7 g
0.00 g
METTLER TOLEDO DeltaRange balances have a movable fine range with 10 times smaller display incre­ments over their entire weighing range. In this fine range an additional decimal place always appears in the display.
The balance operates in the fine range
• after switching on
• after every taring operation
If the fine range is exceeded, the balance display automatically switches to coarser display increments.
15
16

4 Menu

4.1 Overview

In the menu you can change the weighing unit (for certified balances, only if the national certification law allows), select additional functions and carry out various settings. A description of the individual menu options is given in Section 4.3.
Notes
1)
This menu option is only available
on AB135-S/FACT and AB265-S/ FACT models.
2)
With certified balances, this menu
option has a fixed setting and can­not be changed.
3)
These menu option is only shown
if “FACT on” (Factory setting) has been selected in menu option 3 (FACT).
4)
With certified balances, only tho-
se weighing units allowed by the appropriate national weights and measures legislation may be selected.
5)
This menu option is only shown if
“Host” has been selected in menu option 12 (Peripheral unit).
6)
These menu options are only shown
if “Host” or “Printer” has been selec­ted in menu option 12 (Peripheral unit).
Menu option
Factory setting
1 Reset
rESEt
List CAL E
17 Handshake
HS OFF
HS HArd HS SoFt
2 Adjustment
CAL int
6)
Cal
Cal
16 Bit/Parity
7b-odd 7b-no
3 FACT
7b-E
8b-no
2)
FACt on
FACT off
6)
4 Protocol
AutoCal
15 Baud rate
bd 2400
bd 1200
bd 600 bd 9600
bd 19200
Prot on
Prot off
6)
bd 4800
3)
AutoCal
5 Function
F count
F nonE
F 100%
#
FAC d
F
F
#
dyn ÑÖ
F FAC ÑÖ
F
14 Send mode
S. oFF
S. Cont S. Stb
PCS
dyn A
5)
6 Weighing mode
Std
robuSt doS
2)
SEnSor
13 Send format
SICS
PÑÖ
7 Vibration adapter
8 Repeatability
UniuErS FASt -rEL
StAbLE unStAbLE FASt rELiAbLE
12 Peripheral unit
5)
5)
PrintEr
11 Autozero
HoSt
2.diSPLAY
A.ZEro
no A2ZEro
5)
1)
9 Weighing
2)
unit 1
Unit 1
cl kg
Unit 1 Unit 1
t tl mg
Unit 1 Unit 1
S tl ct
Unit 1 Unit 1
H tl lb
Unit 1 Unit 1
moz
Unit 1 Unit 1
mo ozt
Unit 1 Unit 1
dwt GN
Unit 1 Unit 1
10 Weighing unit 2
Unit 2
cl kg
Unit 2 Unit 2
mmg
Unit 2 Unit 2
mo ct
Unit 2 Unit 2
dwt lb
Unit 2 Unit 2
GN oz
Unit 2 Unit 2
ozt
Unit 2
g
4)
g

4.2 Menu operation

Cal/Menu
±
Cal/Menu
C
0.01 g
ŸÖEMU
à
rESEt
rESEt
Unit 1 g
Unit 1 g
Unit 1 kg
STorED
0.01 g
0.01 g
Opening the menu
In weighing mode, press and hold down the «Cal/Menu» key until “MENU” appears in the display. Release the key: the 1st menu option is displayed.
Select menu options
The «±» key is used to select individual menu options with their current settings one after the other.
Change settings
The «» key is used to change the setting at the selected menu option. Every time the key is pressed, the next setting is displayed. Once the desired setting appears in the display, the next menu option can be selected (see above) or you can close the menu (see following sections).
Saving settings and closing the menu
Hold the «Cal/Menu» key down until “StorEd” appears in the display. Release the key and the balance reverts to weighing mode. All changes are saved.
Abort Press the «C» key briefly. The balance reverts to weighing mode. Changes are not saved.
Note If no entry is made within 45 seconds, the balance reverts to weighing mode. Changes are not saved.
17

4.3 Description of menu options

18

4.3.1 Reset or recording of balance settings (1st menu option “RESET”)

Reset balance settings
Cal/Menu
Cal/Menu
rESEt
r donE
0.00 g
rESEt
LISt
StorEd
0.00 g
Select “Reset”, press and hold down the «Cal/Menu» key until the message “r donE” confirms that all
menu settings have been reset. The balance then reverts to weighing mode and works with the factory
settings (Section 4.1).
Recording balance settings
Select “List” and hold down the «Cal/Menu» key until the message “StorEd” is displayed. The current balance settings are transmitted to the peripheral device connected to the interface. To do
this the setting “Printer” must always be selected at the 11th menu option (Peripheral unit). The current balance settings are saved at the same time.
CAL int
CAL E
Cal
Cal

4.3.2 Adjustment (2nd menu option)

In this menu option you can select whether you wish to adjust the balance using the internal or the external adjustment weight. Because of certification legislation, the certified models cannot be adjusted with an ex­ternal weight.
Adjusting with internal adjustment weight
Adjusting with external adjustment weight
4.3.3 FACT (3rd menu option)
In this menu option you can switch the fully automatic internal adjustment (calibration) FACT on or off.
FACt on
FACt off
Prot on
Prot off
AutoCal
AutoCal
AutoCal
FACT switched on This is the factory setting. The balance adjusts (calibrates) itself fully automatically.
FACT switched off Note: With the certified versions of balances, this option can not be selected, i.e. FACT is always active

4.3.4 Protocol (4th menu option)

This menu item can be used to switch the printing of the FACT fully automatic adjustment (calibration) on or off. Note: This menu item can only be selected if FACT is switched on and does not affect the printing of adjust­ments with an internal or external adjustment weight (Section 4.3.3).
Protocol switched on
A record is printed out after every automatic adjustment of the balance (FACT).
Note: The protocol is printed out without a line for signatures.
Protocol switched off
This is the factory setting. If the balance adjusts automatically (FACT), a protocol is not printed out.
19
20
F count
F 100 %
F dyn A
F dyn ŸÖ
F FAC ŸÖ
F FAC d
F nonE
Std
doS
robuSt
SEnSor
PCS
4.3.5 Functions (5th menu option / Operating Section 5)
In addition to simple weighing, the following functions can be selected: F count Piece counting F 100 % Percent weighing F dYn A Dynamic weighing with automatic start F dYn M Dynamic weighing with manual start F FAC M Multiply free factor value by weight, change size of display increment F FAC d Divide free factor value by weight, change size of display increment F nonE No function, simple weighing (factory setting)

4.3.6 Weighing mode (6th menu option)

This setting allows you to adapt the balance to the weighing mode. Select “Std” (standard) for all normal weighing processes or “doS” (dispensing) for dispensing liquid or powdery weighing samples. In this setting, the balance responds very quickly to smallest weight changes.
When set to “robuSt” (absolute weighing), the balance responds only to greater weight changes, and the weighing result is very stable.
When set to “SEnSor”, the balance delivers a raw, unfiltered weighing signal, and is only suitable for special applications. (Only available on AB135-S/FACT / AB265-S/FACT models).
UniuErS
unStAbLE
StAbLE
FASt -rEL
rELiAbLE
FASt
Unit 1 g
Unit 1 mg
Unit 1 m
……
Unit 1 t tl

4.3.7 Vibration adapter (7th menu option)

The vibration adapter can be used to match your balance to the ambient conditions (vibrations, drafts at location). If you work in surroundings which are practically free from temperature fluctuations, drafts, and vibrations, select “StabLE”. On the other hand, if you work in surroundings where the conditions are constantly changing, select “unStAbLE”. For normal ambient conditions, select “UnivErS” (Standard). This is the factory setting.
4.3.8 Measurement release (Repeatability) (8th menu option)
(Only on AB135-S/FACT / AB265-S/FACTmodels)
You can use this setting to specify how rapidly the balance considers the measurement value to be stable and releases it. The “FASt” setting is recommended if you require rapid results and their repeatability is of secondary importance. The “rELiAbLE” setting gives very good repeatibility of the measurement results, but lengthens the stabilization time. When the balance is set to “FASt -rELiAbLE” (fast and reliable), the weight display is quickly released as stable, and gives good repeatability. This is the factory setting.

4.3.9 Weighing unit 1 (9th menu option “UNIT 1”)

Depending on requirements, the balance can operate with the following units (possible with certified balances only if permitted by national weights and measures legislation):
Unit Conversion factor Comments
g gram factory setting, 0.1 mg kg kilogram 1 kg = 1000 g not with 0.01 mg and 1 mg balances mg milligram 1 mg = 0.001 g with 0.01 mg, 0.1 mg and 1 mg balances ct carat 1 ct = 0.2 g lb Pfund 1 lb oz Unze 1 oz ozt Troy Unze 1 ozt GN Grain 1 GN dwt Pennyweight 1 dwt ≈ 1,555173843 g mo Momme 1 m Mesghal 1 H tl Hong Kong Taels 1 S tl Singapur Taels 1 tls 37,799366256 g the Malaysian tael has the same value t tl Taiwan Taels 1 tlt 37,499995313 g
cl Tical 1
= 453,59237 g = 28,349523125 g = 31,1034768 g = 0,06479891 g
mom 3,749999953 g msg 4,6083162 g
tlh 37,42900 g
tical 16,3293 g
21
A.zEro
no A.zEro
22

4.3.10 Weighing unit 2 (10th menu option “UNIT 2”)

If it is required to show the weighing results in weighing mode in an additional unit by pressing the «» key, the desired second weighing unit can be selected in this menu option. The same weighing units are available as under “UNIT 1”, with the exception of the tael units (“H tl”, “S tl” und “t tl”).

4.3.11 Autozero (11th menu option)

This menu option allows you to switch the automatic zero correction on or off. When it is switched on, the zero point is automatically corrected for drift or contamination of the weighing pan.
The following settings are available:
Autozero switched on
The zero point is automatically corrected.
Autozero switched off
The zero point is not automatically corrected. This setting is advantageous for special applications (e.g. eva­poration measurements).
Note
With certified balances, this setting is possible only with a resolution of e = 10d.
PrintEr
HoSt
2.dISPLAY

4.3.12 Peripheral unit (12th menu option)

At this menu option you can select the peripheral device connected to the optional RS232C interface. The balance automatically saves the appropriate settings (Section 4.3.13 – 4.3.17) for every peripheral device.
Printer Connected to a printer. Host Connection to any desired peripheral device. Aux. display Connection of an optional auxiliary display unit (communications parameters cannot be selec-
ted).
SICS
PŸÖ

4.3.13 Send format (13th menu option)

Note: This menu option is only available if you have selected the setting “Host” in the 12th menu option
(Peripheral unit)! At this menu option you specify the data transfer format. “SICS”: The MT-SICS data transfer formats are used. Please refer to the “MT-SICS Reference Manual
Basic-S balances 11780447” available from your METTLER TOLEDO dealer or download from the Internet (www.mt.com/
sics-classic). More Information please find in the Section 6.3.
“PM”*: The following PM balance data transfer formats are used: S. Stb: /////1.67890/g S. Cont: S////1.67890/g
SD///1.39110/g
Additional information can be found on the Internet at www.mt.com/
classic.
* unidirectional, no MT-SICS commands are accepted.
S. oFF
S. Stb
S. Cont

4.3.14 Send mode (14th menu option)

Note: This menu option is only available if you have selected the setting “Host” in the 12th menu option
(Peripheral unit)! At this menu option you specify how data is transmitted to a peripheral device. S. oFF Send mode switched off S. Stb The next possible stable value will be transferred after the «±» key has been triggered. S. Cont All values are transferred automatically.
23
24
bd 2400
bd 4800
bd 19200
7b-E
7b-no
8b-no
7b-odd
HS oFF
HS Soft
HS HArd

4.3.15 Baud rate (15th menu option)

Note: This menu option is only available if you have selected the setting “Host” in the 12th menu option
(Peripheral unit)! The baud rate (data transfer rate) determines the speed of transmission via the serial interface. The unit is the
baud (bd) = 1 bit/second. The following settings are available: 600 bd, 1200 bd, 2400 bd, 4800 bd, 9600 bd and 19200 bd. For problem-free data transmission the sending and receiving devices must be set at the same value.

4.3.16 Bit/Parity (16th menu option)

Note: This menu option is only available if you have selected the setting “Host” in the 12th menu option
(Peripheral unit)! At this menu option you can set the character format for the attached peripheral device. 7b–E 7 data bits/even parity 7b–no 7 data bits/no parity 8b–no 8 data bits/no parity 7b–odd 7 data bits/odd parity

4.3.17 Handshake (17th menu option)

Note: This menu option is only available if you have selected the setting “Host” in the 12th menu option
(Peripheral unit)! This menu option allows you to match the data transmission to different serial receivers. HS oFF No handshake HS SoFt Software handshake (XON/XOFF) HS HArd Hardware handshake (RTS/CTS)

5 Functions

5.1 Piece counting

Requirement
The function “F count” must be activated in the menu (Section 4).
#
On
o ff
O/T
M E TTL
E R T O L
E D O
C
F
C
a l /M e
n u
0.00 g
Place empty container on the balance and tare by briefly pressing the «#» key.
Setting the reference: a reference weight must first be entered for piece counting:
Add a number of reference pieces to container. Possible numbers* are 5, 10, 20, 50, 100 and “no” (this
setting deactivates the piece counting function). *approved balances min 10
On
o ff
O/T
M E T
TL ER
T O LE
D O
C
F
C
a l /M e
n u
F
SEt 10
PCS
Caution: Take into account minimum values: min. reference weight = 10d (10 digits), min. piece
weight* = 1d (1 digit)! *approved balances min 3e
Note: 1 digit corresponds to 1 display increment.
Hold the «F» key down until “SEt ... PCS” is displayed.
±
SEt 20
20
PCS
PCS
Repeatedly press the «» key until the display equals the number of reference pieces entered.
Confirm the number of reference pieces with the «±» key (or wait 7 seconds, in which case the number
is adopted automatically). The current number of pieces (PCS = pieces) is displayed.
Note: The current reference weight remains stored until the reference setting is changed or the power supply is interrupted.
On
240
o ff
O/T
M E TT
LE R
TO L
E D O
C
F
C
al /M enu
144.81 g
PCS
Switching between piece count and weight display
Add weighing sample to the container and read off number of pieces. Press the «» key. The weight is displayed. Return to the piece count display by pressing the «» key again.
25

5.2 Percent weighing

On
off
O/T
M
ETTLER TOLEDO
C
F
C a l /M en
u
F
±
On
off
O/T
M
ETTLER TOLEDO
C
F
C al /M en
u
0.00 g
148.50 g
SEt100 %
100.0 %
101.6 %
150.88 g
26
Requirement
The function “F 100 %” must be activated in the menu (Section 4).
Set target weight
Target weight (Reference weight, which corresponds to 100 %) in center of pan.
Hold the «F» key down until “SEt 100 %” is displayed.
Press the «» key to select “SEt 100 %” or “SEt no %” (Percent weighing deactivated).
The «±» key can be used briefly to confirm or automatic acceptance after 7 seconds. The target weight
is specified.
Note: The current target weight remains stored until a new target weight is set or the power supply is inter­rupted.
Percent weighing / switching
Place weighing sample in center of pan. The weight of the sample is displayed as a percentage of the target weight.
Press the «“» key, The weight is displayed. (Unit 1 and and Unit 2, if activated).
101.6 %
Return to display in percent: pressing the «“» key again.

5.3 Dynamic weighing

Dynamic weighing is suitable for the weighing of unstable weighing samples. The mean value of the weighing results is determined over a specified time period (weighing time). The more unstable the weighing sample, the longer the selected weighing time.
Requirement
“F dYn A” for automatic start or “F dYn M” for manual start must be activated in the menu (Section 4). Factory setting is a weighing time of 3 seconds (t = 3").
Tare container
Tare: Press the «#» key.
0.00 g
Dynamic weighing with automatic start (F dYn A)
On
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
<
0.00 g
<
• 257.35 g
<
-- 3 --
<
-- 1 --
The «» key can be used select the dynamic weighing. The display shows the symbol ,.
Load weighing sample. As soon as the balance is relatively stable, weighing starts automatically.
During the weighing time, a “count down” runs in the display.
<
‹ 260.33 g
Read off result. The result of the dynamic weighing is displayed with (= calculated value) and remains in the display
until the weighing sample is removed from the weighing pan or the container.
27
28
0.00 g
On
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
±
<
0.00 g
<
• 257.35 g
<
-- 3 --
<
-- 1 --
Dynamic weighing with manual start (F dYn M)
The «» key can be used select the dynamic weighing. The display shows the symbol ,.
Load weighing sample.
Start weighing with the «±» key.
During the weighing time, a “count down” runs in the display.
Read off result.
The result of the dynamic weighing is displayed with (= calculated value) and remains in the display
until the weighing sample is removed from the weighing pan or the container.
Notes
• The weighing cycle with the same weighing sample can be restarted with the «±» key.
• The «“» key can be used to switch between dynamic weighing and normal weighing.
• For weighing goods below 5 g the weighing must be started manually with the «±» key, even for dy-
namic weighing with automatic start.
Changing the weighing time
Press and hold the «F» key, until “t = 3"” appears in the display.
Repeatedly press the «» key, until the desired weighing time appears.
Possible values are 3", 5", 10", 20", 1", 2".
<
‹ 260.33 g
<
F
t á 3 ”
<
t á 5 ”
±
0.00 g
Selection with the «±» key briefly to confirm or by automatic acceptance after 3 seconds.
Note
The set weighing time remains stored until it is reset or the power supply fails.

5.4 Weighing with free factor and/or selectable display increments

In this menu option a custom “free factor” can be defined at will.
This value is then either multiplied (“F FAC M”) by the weighing result (in grams), i.e. reading = factor x weight, or it is divided (“F FAC d”) by the weight, i.e. reading = factor / weight. The range of possible factors depends on the resolution of the balance. Common values are: F FAC d 1 to 10'000 F FAC M < 1 to 10'000
The “free factor” (FAC M) function can, for example, be used to calculate the price of the material weighed directly or to calculate the weight per defined unit of surface area. It can also be used to convert the weight into any desired alternative unit. This facility for dividing the factor by the weight (FAC d) is required for instance in the textile industry to determine yarn count.
The ability to select the display increments makes it possible to specify how the weighing result is to be presented, the choice of display increments being limited by the set factor and the resolution of the balance model itself.
Requirement
The function “F FAC M” or “F FAC d” must be activated in the menu (Section 4).
Entering the free factor and/or the display increments
19.50 g
Hold the «F» key down until “FAC M” or “FAC d” appears in the display. The «» key can be used to switch between “FAC M” and “FAC d,” and “noFAC M” and “noFAC d”
On
off
O/T
METTLER TOLEDO
C
F
Cal
/Menu
(function deactivated) .
Release the key. Either the factor 1 appears as default value or the factor that
was saved most recently.
F
FAC ÑÖ
1. 0000
29
30
±
±
±
±
0. 0000
0. 0000
0. 9000
StEP
0. 0900
0. 1000
StorEd
19.50 g
17.55
#
This value can now be changed: Pressing the «“» key increases the value of the blinking numeral. The point appears after the numeral
9, and then the numeral 0 reappears.
Note: The point appears in the selection only if a point was not used in the preceding places. Pressing the key once changes the value by one increment. Pressing the «±» key moves the position of the numeral to be changed to the right. After the last place,
the position shifts back to the first place.
Hold the «±» key down until the selected factor is confirmed (no automatic acceptance). “StEP” ap-
pears in the display. The program automatically changes to the entry of the display increments. As stan­dard, the smallest possible display increment or the value last saved, if this is still valid, appears.
Note: “Error” is displayed instead of “StEP”: The factor entered lies outside the valid range. The program
then defaults to the value Factor Min. or Factor Max., which can be changed or accepted. This value can be changed in the same way as for the free factor (see above). Hold the «±» key down until the selected display increment is confirmed (no automatic acceptance).
“StorEd” appears in the display. The program automatically changes to the weight display. Note: Displays “Error” instead of “StorEd”. The display increment entered lies outside the valid range. The
program then defaults to the value Step Min. or Step Max., which can be changed or accepted. Press the «“» key. The weight of the object in the pan is calculated with the selected factor and dis-
played in the selected increment size. The symbol “#” appears instead of a weight unit. The weight is
calculated in grams.
Note
If you only want to change the display increments, set the free factor at exactly 1.
Toggling between displaying the calculated value and the measured weight
The «» key can be used to switch between Unit 1, Unit 2 (immediately activated) and the calculated value
On
19.50 g
off
O/T
METTLER TOLEDO
C
F
Cal /Menu
(#), provided that the “F FAC M” or “F FAC d” function is activated.
17.55
19.50 g

5.5 Switching weight units

#
64.71 g
64710 mg
Requirement
Different weight units must be activated in the menu for unit 1 and unit 2 (Section 4).
The «» key can be used at any time to toggle between the two weighing units selected in the menu
(“UNIT 1” and “UNIT 2”).
Notes:
• Switching between weight units may be blocked with certified balances, depending on national
weights and measures legislation.
• This function is not available with dynamic weighing.
31

6 Technical data, optional equipment

6.1 Technical data

Standard equipment of B-S/FACT balances
• Protective cover, transparent, made from Barex
• AC adapter to national codes according list on section 6.4 Balance power input: 8–14.5 VAC, 50/60Hz,
6 VA or 9.5–20 VDC, 6 W
• FACT, fully automatic adjustment with internal weight
• Built-in RS232C interface
• Draft shield with AB-S/FACT and PBxx3-S/FACT models
• All models can weigh below balance.
• Display with backlight (except ABxx5-S/FACT).
32
Materials
• Housing: die-cast aluminum, painted
• Weighing pan: Chromium-nickel steel, X2CrNiMo 17 13 2 (1.4404)
Protection
• Protected against dust and water
• Pollution degree: 2
• Overvoltage category: class II
• EMC: see declaration of conformity (separate brochure 11780294)
Ambient conditions The technical data are valid unter the following ambient conditions:
• Ambient temperature 10 °C … 30 °C
• Relative humidity 15 % … 80 % at 31 °C, linear decreasing to
50% at 40 °C
noncondensing Operability is assured at ambient temperatures bet-
ween 5 and 40 ºC.
Technical data
AB54-S/FACT AB104-S/FACT AB204-S/FACT AB304-S/FACT AB135-S/FACT
DualRange
AB265-S/FACT DualRange
Readability 0.1 mg 0.1 mg 0.1 mg 0.1 mg 0.01 mg**/0.1 mg 0.01 mg**/0.1 mg Max. capacity 51 g 110 g 220 g 320 g 31 g**/120 g 61 g**/220 g Repeatability (sd) 0.1 mg 0.1 mg 0.1 mg 0.1 mg 0.03 mg
a)
/0.1 mg 0.03 mga)/0.1 mg Linearity 0.2 mg 0.2 mg 0.2 mg 0.4 mg 0.2 mg 0.2 mg Sensitivity temperature drift
2.5 ppm/ °C 2.5 ppm/ °C 2.5 ppm/ °C 2.5 ppm/ °C 2.5 ppm/ °C 2.5 ppm/ °C (10 °C … 30 °C) Settling time, typical 3.5 s 3.5 s 4 s 5 s 4 s / 15 s** 4 s / 15 s** Adjustment weight built-in built-in built-in built-in built-in built-in Backlight yes yes yes yes no no External dimensions of balance
245x321x344 mm 245x321x344 mm 245x321x344 mm 245x321x344 mm 245x321x344 mm 245x321x344 mm
(W/D/H) External dimensions of packaging
(W/D/H)
381x436x495 mm (0.082 m
3
)
381x436x495 mm (0.082 m3)
419x494x521 mm (0.108 m3)
381x436x495 mm (0.082 m3)
419x494x521 mm (0.108 m3)
419x494x521 mm
(0.108 m3) Weighing pan ø 80 mm ø 80 mm ø 80 mm ø 80 mm ø 80 mm ø 80 mm Usable heigh of draft shield 237 mm 237 mm 237 mm 237 mm 237 mm 237 mm Net weight (with packaging) 5.8 kg (7.8 kg) 5.8 kg (7.8 kg) 6.4 kg (9.1 kg) 5.8 kg (7.8 kg) 6.4 kg (9.1 kg) 6.4 kg (9.1 kg)
** Fine range(DualRange)
a)
Fine range up to 10 g (Dual Range)
Technical data PB153-S/FACT PB303-S/FACT PB303-SDR/FACT
PB403-S/FACT PB503-S/FACT PB602-S/FACT PB1502-S/FACT PB3002-S/FACT
DeltaRange
Readability 0.001 g 0.001 g 0.001 g*/0.01 g 0.001 g 0.001 g 0.01 g 0.01 g 0.01 g Max. capacity 151 g 310 g 60 g*/310 g 410 g 510 g 610 g 1510 g 3100 g Repeatability (sd) 0.001 g 0.001 g 0.001 g*/0.008 g 0.001 g 0.001 g 0.01 g 0.01 g 0.01 g Linearity 0.002 g 0.002 g 0.01 g 0.002 g 0.002 g 0.02 g 0.02 g 0.02 g Sensitivity temperature drift
6 ppm/ °C 6 ppm/ °C 6 ppm/ °C 6 ppm/°C 6 ppm/°C 6 ppm/ °C 6 ppm/ °C 6 ppm/ °C
(10 °C … 30 °C) Settling time, typical 2 s 2 s 2 s 2 s 3 s 2 s 2 s 2 s Adjustment weight built-in built-in built-in built-in built-in built-in built-in built-in Backlight yes yes yes yes yes yes yes yes External dimensions of balance
(W/D/H)
External dimensions of packaging (W/D/H)
245x321x280 mm
381x436x495 mm (0.082 m3)
245x321x89 245x321x89 245x321x89
381x436x273 mm
(0.045 m3) Weighing pan ø 100 mm ø 180 mm Usable heigh of draft shield 165 mm — Net weight (with packaging) 4.9 kg (6.5 kg) 3.6 kg (4.7 kg)
* Fine range (DeltaRange)
33
33
34
Technical data PB3002-SDR/FACT
PB4002-S/FACT PB1501-S/FACT PB3001-S/FACT PB5001-S/FACT PB8001-S/FACT PB8000-S/FACT
DeltaRange
Readability 0.01 g*/0.1 g 0.01 g 0.1 g 0.1 g 0.1 g 0.1 g 1 g Max. capacity 600 g*/3100 g 4100 g 1510 g 3100 g 5100 g 8100 g 8100 g Repeatability (sd) 0.01 g*/0.08 g 0.01 g 0.08 g 0.08 g 0.08 g 0.08 g 0.8 g Linearity 0.1 g 0.02 g 0.1 g 0.1 g 0.1 g 0.1 g 1 g Sensitivity temperature drift
6 ppm/ °C 6 ppm/ °C 10 ppm/ °C 10 ppm/ °C 10 ppm/ °C 10 ppm/ °C 10 ppm/ °C
(10 °C … 30 °C) Settling time. typical 2 s 2 s 1.5 s 1.5 s 1.5 s 2 s 1 s Adjustment weight built-in built-in built-in built-in built-in built-in built-in Backlight yes yes yes yes yes yes yes External dimensions of balance
245x321x89 mm 245x321x89 mm
(W/D/H) External dimensions of packaging
(W/D/H)
381x436x273 mm (0.045 m3)
381x436x273 mm
(0.045 m3) Weighing pan ø 180 mm 180 x 180 mm Net weight (with packaging) 3.6 kg (4.7 kg) 3.6 kg (4.7 kg)
* Fine range (DeltaRange)

6.2 Interface

RS232C interface and interface accessories
GND
96
AB-S,PB-S RS232C
DATA
RxD
IN
TxD
Every B-S/FACT balance is fitted with an RS232C interface
OUT
for attachment to a peripheral device (e.g. printer or PC with a 9-pin male connector). Matching to a different device can
15
HAND SHAKE
CTS
RTS
be carried out in the menu (Sections 4.3.12 – 4.3.17). You will find a detailed description of the available inter-
face commands in the brochure “Reference Manual MT-
IN
OUT
SICS Basic-S balances 11780447” available from your METTLER TOLEDO dealer or download from the Internet (www.mt.com/
sics-classic).
The wide range of features of the B-S/FACT balances regarding documentation of the results can not be exploited to the full until a printer, e.g. the RS-P26 or LC-P45 from METTLER TOLEDO is attached. The printed results make a decisive contribution to a simp­le way of working in compliance with GLP/GMP.

6.3 MT-SICS Interface commands and functions

Many of the balances and scales used have to be capable of integration in a complex computer or data acquisition system.
To enable you to integrate balances in your system in a simple manner and utilize their capabilities to the full, most balance functions are also available as appropriate com­mands via the data interface.
All new METTLER TOLEDO balances launched on the market support the standardized command set “METTLER TOLEDO Standard Interface Command Set” (MT-SICS). The commands available depend on the functionality of the balance.
Basic information on data interchange with the balance
The balance receives commands from the system and acknowledges the command with an appropriate response.
Command formats
Commands sent to the balance comprise one or more characters of the ASCII character set. Here, the following must be noted:
Enter commands only in uppercase.
• The possible parameters of the command must be separated from one another and from the command name by a space (ASCII 32 dec., in this description represented
/).
as
• The possible input for “text” is a sequence of characters of the 8-bit ASCII character set from 32 dec to 255 dec.
• Each command must be closed by C
The characters CRLF, which can be inputted using the Enter or Return key of most entry keypads, are not listed in this description, but it is essential they be included for com­munication with the balance.
(ASCII 13 dec., 10 dec.).
RLF
Example S – Send stable weight value
Command S Send the current stable net weight value. Response S/S/WeightValue/Unit Current stable weight value in unit actually set
under unit 1. S/I Command not executable (balance is currently
executing another command, e.g. taring, or timeout as stability was not reached).
S/+ Balance in overload range. S/- Balance in underload range.
Example
Command S Send a stable weight value. Response S/S/////100.00/g The current, stable weight value is 100.00 g.
35
35
36
The MT-SICS commands listed below is a selected list of available commands. For additional commands and further information please refer to the Reference Manual “MT-SICS Basic-S balances 11780447” downloadable from the Internet under
www.mt.com/sics-classic.
S – Send stable weight value
Command S Send the current stable net weight value.
SI – Send value immediately
Command SI Send the current net weight value, irrespective of balance stability.
SIR – Send weight value immediately and repeat
Command SIR Send the net weight values repeatedly, irrespective of balance stability.
Z – Zero
Command Z Zero the balance.
@ – Reset
Command @ Resets the balance to the condition found after switching on, but without a zero setting being performed.
SR – Send weight value on weight change (Send and Repeat)
Command SR Send the current stable weight value and then send continuously the stable weight value after every weight change. The weight change must be at least 12.5 % of the last stable weight value, minimum = 30d.
ST – Send stable weight after pressing ± (transfer) key
Command ST Inquiry of actual status of the ST function.
SU – Send stable weight value with currently displayed unit
Command SU As the “S” command, but with the currently displayed unit.

6.4 Optional equipment

AC adapters
For all models except AB135-S/FACT / AB265-S/FACT:
• AC adapter universal (EU, USA, AU, UK) 11120270
100–240 VAC/50–60 Hz, 0.3 A 12 VDC, 0.84 A
For AB135-S/FACT / AB265-S/FACT (as well as for all B-S/FACT models):
• AC adapter universal (bench version)* 11107909
100-240 VAC/50–60 Hz, 12 V, 2.0 A *(appropriate cable for country also required)
AccuPac B-S
Rechargeable external power source for weighing operation with no mains connection 21254691
Adjustment weights
Available as OIML weights (E1, E2, F1, with calibration certificate) For further details see METTLER TOLEDO Weights brochure 11795461 or see www.mt.com/weights
Antitheft device
Cable with lock (for all models)
18 hours
00590101
Auxiliary display (RS/LC-BLD)
Auxiliary display including RS cable for connection to the RS232C interface and separate AC adapter
Auxiliary display (RS232)
Auxiliary display for connection to the RS232C interface 12120057
Density kits (for AB-S/FACT only)
• For determination of solids
• For determination of liquids with displacement body
Draft shields
• Glass zylinder draft shield 11137305 (for 0.1/1 mg balances)
• Draft shield with sliding doors
“mg” (165 mm) * 11137468
• Draft shield with sliding doors
“0.1 mg” (237 mm) * 11103682
• Draft shield
* Balances with a weighing pan Ø 180 mm require the weighing pan Ø 175 mm 11103680
“mg” (141 mm) * 11103683
00033360 + 00210260
00224200
00033360
Interface cable
• RS9–RS25: (m/f), length 2 m 11101052
• RS9–RS9: (m/f), length 1 m 11101051
• RS9–RS9: (m/m), length 1 m 21250066
• USB-RS232 converter cable 11103691
Printer, Application printer (LC-P45)
Plain-paper printer, 24 characters, with additional functions (time, date, statistic, multiplier etc.)
Printer, Report printer (RS-P26)
Plain-paper printer, 24 characters, with additional functions (time, date) 12120788
Protective cover
• For PB-S/FACT models (1 piece) 11103681
• For AB-S/FACT models (1 piece) 11135408
Transport case
• For all PB-S/FACT models, with room for balance, draft shield 165 mm / 141 mm and weight 11101050
• For all AB-S/FACT models, has space for balance and draft shield 237 mm 11103834
00229119
37
37

6.5 Dimensional drawings

All dimensions in millimeters (mm)
38
PB-S/FACT balance with readability of 0.01 g
10.5
213
266
184.5
1)
PB-S/FACT balance with readability of 0.1 g / 1 g
184.5
266
1)
180 x 180
321
10
89
190
245
Ø
181
89
10.5
321
190
10
245
213
1) Opening for hanger
All dimensions in millimeters (mm)
AB-S/FACT balance with readability of 0.01 mg / 0.1 mg and draft shield 0.1 mg (237 mm)
128.5
91.5
237
PB-S/FACT balance with readability of 1 mg and draft shield mg (165 mm)
128.5
91.5
344
165
280
10.5
184.5
213
141
1) Opening for hanger
92.5
184.5
266
321
163.5
73.5
10
1)
Ø
39
38
100
39
141
266
321
163.5
73.5
10
10.5
1)
38
Ø
84
245
190
193
213
92.5
245
190
193

7 Appendix

40
7.1 Typical printouts from METTLER TOLEDO
Function: Adjusting (FACT)
- BALANCE CALIBRATION -
12.02.2007 09:42:15
METTLER TOLEDO Type: PB3002-S/FACT SNR: 1118015657 SW: 1.20
Internal Cal. done
--------- END ----------
Function: Adjusting (external)
- BALANCE CALIBRATION -
12.02.2007 09:45:10
METTLER TOLEDO Type: PB3002-S/FACT SNR: 1118015657 SW: 1.20
Weight ID: .............
Weight: 2000.00 g
External Cal. done
Signature:
........................
--------- END ----------
1)
Only possible with LC-P45.
Function: Percent weighing
----- % - WEIGHING ----­Ref. 10.008 g
100.00 %
60.01 g
599.59 %
Function: List Printout of the current balan­ce settings
--- LIST OF SETTINGS ---
12.02.2007 09:48:16
METTLER TOLEDO Type: PB3002-S/FACT SNR: 1118015657 SW: 1.20
-----------------------­Application: Dynamic A
-----------------------­Weighing Parameters: Weighing Mode Standard Unit 1 g Unit 2 g A.Zero On
-----------------------­Peripheral Devices: P.Device Printer Baud 2400 Bit/Parity 7b-even Handshake Off
P.Device Host Sendmode Off Baud 9600 Bit/Parity 8b-no Handshake Soft
--------- END ----------
RS-P26 and LC-P45 printers
Function: Piece counting Printout with reference weight
---- PIECE COUNTING ---­APW: 0.99460 g Out of: 10 PCS
27.000 g 27 PCS
Function: Verification of the calibration (adjustment) with external weight.
Function is triggered via the
1)
printer.
----- BALANCE TEST -----
12.02.2007 09:52:12
METTLER TOLEDO Type: PB3002-S/FACT SNR: 1118015657 SW: 1.20
Weight ID: .............
Target : ...............
Actual : .......199.98 g
Diff : ...............
External test done
Signature:
........................
--------- END ----------
Function: Dynamic weighing
--- DYNAMIC WEIGHING --­Weigh Time: 2 s
DW 49.999 g
Function: Statistics Function is triggered via the
1)
printer.
12.02.2007 10:44:07 ID 666 SNR: 1118015657 1 1100.15 g 2 1600.10 g 3 1699.95 g n 3 x 1466.733 g s 321.372 g srel 21.91 % min. 1100.15 g max. 1699.95 g dif. 599.80 g
--------- END ----------
Function: Multiplier Function is triggered via the
1)
printer.
12.02.2007 08:23:22 ID 242 SNR: 1118015657
Factor 1.65
588.43 g * 970.9095
Notes
The operating instructions for the LC-P45 include a descrip­tion of the functions that are triggered via that printer.
The
RS-P26 prints all reports in
English. This applies also to the LC-P45 reports that originate
in the balance. In the case of reports triggered by the LC-P45, the following languages may be selected: German, English,
French, Spanish or Italian.

7.2 What if ...?

Error/Error message
неееем
î____ï
Error 1
Error 2
Error 3
Error 4
Error 6
0.000
g
Abort
Cause
Overload
Underload
No stability
• in taring or adjusting (calibration)
• when reference weight for piece counting is placed on pan
Wrong adjustment weight on pan or none at all
Reference number for piece counting too small
Internal fault Contact METTLER TOLEDO customer service.
No standard calibration Contact METTLER TOLEDO customer service.
Wrong weighing pan or pan missing
Adjustment aborted with the «C» key
Rectification
Remove sample from weighing pan, zero again (tare).
Check whether weighing pan is positioned properly.
Wait for stability before pressing key. Ensure more stable ambient conditions. Remove weighing pan and clean if necessary
Place required adjustment weight in center of pan.
Increase number of reference pieces.
Mount correct weighing pan.
41

7.3 Maintenance and cleaning

42
Service
Regular servicing of your balance by a service technician prolongs its working life. Ask your METTLER TOLEDO dealer for details of servicing options.
Cleaning
Every now and then, clean the weighing pan, draftshield element, draftshield (depending on the model) and housing of your balance using a damp cloth. Your balance is made of high-quality, durable materials and can therefore be cleaned with a standard, mild cleaning agent.
Please observe the following notes
• On no account use cleaning agents, which contain solvents or abrasive ingredients, as this can result in damage to the terminal overlay.
• After working with chemicals, it is advisable to wash or clean the weighing pan and the bottom plate (if draft shield fitted).
• Although all materials are of high quality, corrosion may occur if corrosive substances are deposited on chrome steel for an extended period of time (and if air is excluded, for example by a coating of grease).
• Ensure that no liquid comes into contact with the balance or the AC adapter!
• Never open the balance or AC adapter – they contain no compo­nents, which can be cleaned, repaired or replaced by the user.
• Soiled protective covers can be replaced on all balance types (see Optional equipment).
Disposal
In conformance with the European Directive 2002/96/EC on Wa­ste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste.
This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regula­tions at the collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible autho­rity or the distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
To protect your METTLER TOLEDO product’s future: METTLER TOLEDO Service assures the quality, measuring accuracy and preservation of value of all METTLER TOLEDO products for years to come. Please send for full details about our attractive terms of service. Thank you.
*11780762
Subject to technical changes and to the availability
of the accessories supplied with the instruments.
© Mettler-Toledo AG 2007 11780762C Printed in Switzerland 0710/2.12
Mettler-Toledo AG, Laboratory & Weighing Technologies, CH-8606 Greifensee, Switzerland Phone +41-44-944 22 11, Fax +41-44-944 30 60, Internet: http://www.mt.com
Loading...