Site preparation for the METTLER TOLEDO ParticleTrack G600Ex system is the end user's
responsibility. The following should be considered to ensure successful installation of the
system:
Power Requirements
A temporary power cord is provided with the ParticleTrack G600Ex system for
convenience of installation and commissioning. The final installation must comply
with EN60079-14 and use fittings and wiring rated appropriately for the location. The
equipment shall be permanently installed to the main power supply through metal conduit
with > 75 °C temperature-rated cable and connected to protective earthing. A 10-ampere
circuit breaker should be included in the building installation to provide overcurrent
protection. The circuit breaker should be within easy reach of the operator and marked as
the disconnecting device for the connection.
Air/Gas Requirements
A ParticleTrack G600Ex system requires a source of clean, dry, and pressure-regulated
instrument air or inert gas to drive the probe optical scanner. The quality of the air or
gas supply must meet the specifications of the American National Standards Institute/
Instrument Society of America (ANSI/ISA) S7.0.01-1996 Quality Standard for Instrument
Air. Air/gas must:
• Have a dew point at least 10 °C [50 °F] lower than the minimum temperature to
which any part of the system will be exposed.
• Contain less than 1 ppm total oil or hydrocarbons
• Contain less than 1 ppm particulates at a maximum size of 3microns
• Be free of any corrosive contaminants and flammable or toxic gases.
Air supply pressure 4.5 to 5.2 barg [65–75 psig]
Maximum ow rate 40 L/min [1.4 SCFM]
Table 5. Probe air/gas specication
Area of Intended Use
Ensure the area of intended use has adequate space to mount the base unit and a
support structure capable of supporting the system weight. The base unit has mounting
tabs for four (4) 8 mm [5/16 inch] minimum diameter stainless steel bolts (end user-
provided). The support structure must not be subject to excessive vibration. The probe
conduit should not bend beyond 20 centimeters (8 inches) and it should not be put
into an “S” shape. Include a suitable work location for the control computer that allows
connectivity through a USB cable.
The base unit must be convenient for access (eye level) and readily accessible for
service. It can also operate in a horizontal position—placed back side down on a
flat horizontal surface. Position the base unit to enable easy access to the purge and
pressurization system air supply. The ParticleTrack G600Ex base unit contains sensitive
electronic components that should be protected from severe environmental conditions.
If the area is outdoors in warmer climates, the base unit must be shaded from direct
sunlight.