Conventions and symbols used in these operating instructions1.17
Definition of signal warnings and symbols2.18
Product safety information2.28
Components3.110
Terminal keys3.211
User Interface3.311
Application home screen3.3.112
Input Dialogs3.3.213
Entering Characters and Numbers3.3.2.113
Changing the time and date3.3.2.214
Lists and tables3.3.315
Scope of delivery4.116
Installing the components4.217
Selecting the location4.318
Connecting the instrument4.419
Leveling the balance4.520
Leveling the balance with the level indicator4.5.120
Leveling the balance using the leveling assistant4.5.221
Transporting the balance4.622
Weighing below the balance4.722
General requirements4.823
Warming up the balance4.8.123
Adjusting the balance4.8.223
Switching on the balance5.124
Switching off the balance5.224
Main settings and activities at a glance5.325
Navigation on the touch screen5.426
Opening an application5.4.126
Scrolling5.4.226
Using shortcuts5.4.327
Performing a simple weighing5.527
Switching the weighing unit5.628
Changing the resolution5.731
Zeroing5.832
Taring5.932
Working with IDs5.1033
Defining an ID5.10.133
Workflow handling options5.10.234
Table of Contents3
Statistics5.1135
Activities636
Activities - Weighing and other applications6.136
Weighing6.1.137
Counting6.1.238
Percent Weighing6.1.340
Formulation6.1.442
Dynamic Weighing6.1.544
Check Weighing6.1.646
Factor Weighing6.1.750
Totaling6.1.851
Quick settings/Preferences7.157
Date and time7.1.157
Brightness7.1.257
Sound7.1.358
Language7.1.458
Glove mode7.1.558
Quick Adjustment7.1.658
Leveling assistant7.1.758
General configuration and data7.259
Weighing options7.2.159
Publishing7.2.260
Devices and connectivity7.2.362
System settings7.2.463
ISO-log7.2.563
Access protection7.2.664
Communication with Peripheral Devices865
PC-Direct Function8.165
USB Device Interface8.266
Cleaning the glass draft shield (0.1 mg and 1 mg models)11.171
Disposal11.273
Interface Specification1274
RS232C interface12.174
USB host12.275
USB device12.375
Table of Contents4
MT-SICS interface commands and functions12.476
Technical Data1377
General data13.177
Model-specific data13.278
Balances with readability of 0.1 mg, S platform with draft shield13.2.178
Balances with readability of 1 mg, S platform with draft shield13.2.280
Balances with readability of 0.01 g, S platform13.2.382
Balances with readability of 0.1 g to 1 g, S platform13.2.485
Dimensions13.386
Balances with readability of 0.1 mg and draft shield13.3.186
Balances with readability of 1 mg and draft shield13.3.287
Balances with readability of 0.01 g13.3.388
Balances with readability of 0.1 g13.3.489
Accessories and Spare Parts1490
Accessories14.190
Spare parts14.295
Index96
Table of Contents5
1Introduction
1
2
…
Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS line
combine a large number of weighing possibilities with easy operation.
These operating instructions apply to all balance models of the MS-TS line and are based on the initially
installed firmware (software) version V 3.30.
1.1Conventions and symbols used in these operating instructions
Key and button designations are indicated by a picture or text in square brackets (e.g. [ ]).
These symbols indicate an instruction:
Requirements
Steps
Results
u www.mt.com/balances
7Introduction
2Safety Information
2.1Definition of signal warnings and symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and warnings.
Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results.
WARNINGfor a hazardous situation with medium risk, possibly resulting in severe
injuries or death if not avoided.
CAUTIONfor a hazardous situation with low risk, resulting in damage to the device or
the property or in loss of data, or minor or medium injuries if not avoided.
Attention(no symbol)
for important information about the product.
Note(no symbol)
for useful information about the product.
General hazardElectrical shock
2.2Product safety information
Intended use
Your balance is used for weighing. Use the balance exclusively for this purpose. Any other type of use and
operation beyond the limits of technical specifications without written consent from Mettler-Toledo AG, is con
sidered as not intended.
It is not permitted to use the instrument in explosive atmosphere of gases, steam, fog, dust
and flammable dust (hazardous environments).
General safety information
This balance complies with current industry standards and the recognized safety regulations; however, it can
constitute a hazard in use. Do not open the balance housing: The balance contains no user-serviceable parts.
In the event of problems, please contact a METTLER TOLEDO representative.
Always operate and use your instrument only in accordance with the instructions contained in this manual. The
instructions for setting up your new instrument must be strictly observed.
If the instrument is not used according to these Operating Instructions, protection of the instrument may
be impaired and METTLER TOLEDO assumes no liability.
Staff safety
These operating instructions must be read and understood before using the balance. These operating instruc
tions must be retained for future reference.
The balance must not be altered or modified in any way. Only use METTLER TOLEDO original spare parts and
accessories.
8Introduction
Safety notes
CAUTION
Damage to the balance
a) Only use indoors in dry locations.
b) Do not use pointed objects to operate the touch screen!
The balance is of a very sturdy design, but is still a precision instrument. It must be han
dled with care.
c) Do not open the balance:
The balance contains no user-serviceable parts. In the event of problems, please contact
a METTLER TOLEDO representative.
d) Only use METTLER TOLEDO original accessories and peripheral devices for the balance.
These are specifically designed for the balance.
WARNING
Risk of electric shock
Use only the original universal AC/DC adapter delivered with your balance, and check that
the voltage printed on it is the same as your local power supply voltage. Only plug the
adapter into a socket which is grounded.
9Safety Information
3Design and Function
C
0.1 g
10
9
7
1
2514
0.01 g
9
5
10
1
2
7
14
6
MS
MS
3
1012
5
7
4
3
14
MS
MS
0.1 mg
10
1
2
5
7
4
14
6
11
12
8
1
3
15
14
3.1Components
10Safety Information
1Capacitive color TFT touch screen
2Operation keys
3Glass draft shield
4Handle for operation of the draft shield door
5Weighing pan
6Draft ring element
7Level indicator
8Kensington slot for anti-theft purposes
9Safety feet (0.01 g and 0.1 g models)
10Leveling foot
11USB Device connection
12RS232C serial interface
13Socket for AC/DC adapter
14USB Host connection
15Product label
3.2Terminal keys
534
1
2
No.KeyNameExplanation
1ON/OFFTo switch the instrument on or off.
2Capacitive color TFT touch screenGeneral navigation
3TareTo tare the balance.
4ZeroTo zero the balance.
5HomeTo return from any menu level, or other window
3.3User Interface
The screen is a capacitive color TFT touch screen. The screen not only displays information, it also allows the
user to enter commands by tapping on certain areas on its surface. You can choose the information displayed
on the screen, change balance settings or perform certain operations on the instrument.
Only those elements which are available for the current dialog appear on the display.
to the application home screen.
CAUTION
Do not touch the touch screen with pointed or sharp objects!
This may damage the touch screen.
11Design and Function
3.3.1Application home screen
The application home screen appears after the startup of the instrument. It always shows the last application
that was being used before the balance was switched off. The application home screen is the main screen of
the balance, from where every function can be accessed. You can always return to the application home screen
by pressing on the home button [ ] in the lower right corner of the screen.
Information and work bars
NameExplanation
1Weighing information barShows the weighing-in aid and general balance information.
2Work title barShows information about the current activity.
3Value barShows information about the current weighing process.
4Main navigationWork-related functions.
Information fields
NameExplanation
5Weighing-in aidA dynamic graphic indicator shows the used amount of the total
weighing range.
6Short balance informationReadability and capacity of the balance.*
7Weighing value fieldShows the value of the current weighing process.
8Coach text fieldShows instructions for the current weighing process.
* For approved balances: Min (minimum capacity) and e (Verification scale interval) are shown in the left
upper corner.
Action buttons
NameExplanation
9Main activity configurationTo configure the current application (e.g. Weighing).
10 Detailed balance InformationShows detailed technical data about the balance.
11 Weighing unitShows the unit of the current weighing process.
12 ActivitiesTo open the activities selection.
13 PrintTo produce a print.
14 Settings/PreferencesTo configure settings/preferences.
15 Status information fieldShows information about the system status.
12Design and Function
3.3.2Input Dialogs
3.3.2.1Entering Characters and Numbers
The keyboard dialog allows the user to enter characters including letters, numbers and a variety of special char
acters.
NameExplanation
1Input fieldShows the characters that have been entered.
2SwitchIf a switch appears on the right side in the content title bar, it must be switched
on in order to enter a value.
3DiscardTo discard the entered data and to exit the dialog.
4Delete allTo delete all entered characters.
5Page indicatorTo navigate when the dialog consists of several steps.
6DeleteTo delete the last entered character.
7ConfirmTo confirm the data entered.
8ShiftTo switch between lower or upper case letters.
9Specialized tabsTo switch the keyboard mode for entering letters, numbers or special characters.
10Explanation fieldExtra information about the value to enter (e.g. the maximum number of charac
ters available).
13Design and Function
3.3.2.2Changing the time and date
The dialog (Picker view) allows the user to set the time and date.
NameExplanation
1Change time/date formatVarious time/date formats can be chosen.
2Pick buttonIncrement
3Picker fieldShows the defined time/date.
4Pick buttonDecrement
5Selection tabsTabs of the selectable sub-categories
14Design and Function
3.3.3Lists and tables
Basic elements of a simple list include a content title and a list of sub-elements. Tapping on an element opens
a list of sub-elements or an input dialog.
NameExplanation
1List titleTitle of the current list
2Back buttonTo go one step back.
3List element titleTitle of the list element
4List element descriptionShows the sub-elements.
5Scroll positionThe list can be scrolled.
6Selection tabsTabs of the selectable sub-categories.
15Design and Function
4Installation and Putting into Operation
This section describes how to put the new instrument into operation.
4.1Scope of delivery
Components0.1 mg1 mg0.01 g0.1 g
236 mm–––Draft shield
168 mm–––
Weighing pan
Draft shield element––
Pan support
Bottom plate––
ESD plate––
Protective cover
Universal AC/DC adapter (country-specific)---
AC/DC adapter (without power cable)-
Operating instructions printed or on CD-ROM (countryspecific)
Quick guide
EC declaration of conformity
Ø 90 mm–––
127 x 127 mm–––
170 x 200 mm–––
190 x 226 mm–––
16Design and Function
4.2Installing the components
3
4
5
6
1
2
2
M
S
3
4
5
1
2
MS
Balances with readability of 0.1 mg and draft shield
(236mm)
Place the following components on the balance in the
specified order:
Note
Push the side glass back as far as it will go and hold
the draft shield (1) with both hands on the bars at the
top.
1 Turn draft shield lock (2) to position (unlock).
2 Place draft shield on the balance.
3 Turn draft shield lock to (lock) and place bottom
plate (3).
4 Place draft ring element (4) and weighing pan (6)
with pan support (5).
Note
For cleaning the draft shield see section Cleaning the
glass draft shield (0.1 mg and 1 mg models)
(page71).
Balances with readability of 1 mg and draft shield
(168mm)
Place the following components on the balance in the
specified order:
Note: Push the side glass back as far as it will go
and hold the draft shield (1) with both hands on the
bars at the top.
1 Turn draft shield lock (2) to position (unlock).
2 Place draft shield on the balance.
3 Turn draft shield lock to (lock) and place bottom
plate (3).
4 Place weighing pan (5) with pan support (4).
Note
For cleaning the draft shield see section Cleaning the
glass draft shield (0.1 mg and 1 mg models)
(page71).
17Installation and Putting into Operation
Balances with readability of 0.01 g
MS
2
3
4
1
MS
3
2
1
– Place the following components on the balance in
the specified order:
●
ESD plate (1)
●
Pan support (2)
●
Weighing pan (3)
●
Draft ring element (4)
Balances with readability of 0.1 g
– Place the following components on the balance in
the specified order:
●
ESD plate (1)
●
Pan support (2)
●
Weighing pan (3)
4.3Selecting the location
Your balance is a precision instrument and will reward you for finding its optimum location with higher accura
cy and dependability. Select a stable surface that is as horizontal as possible. The surface must be able to
safety carry the weight of a fully loaded balance.
Observe ambient conditions (see Technical Data).
Avoid the following:
●
Direct sunlight
●
Powerful drafts (e.g. from fans or air conditioners)
●
Excessive temperature fluctuations
●
Vibrations
18Installation and Putting into Operation
4.4Connecting the instrument
The balance is supplied with a country-specific AC/DC adapter or with a country-specific power cable. The pow
er supply is suitable for all line voltages in the range: 100 - 240 VAC, 50/60 Hz. For detailed specifications,
see the section Technical Data.
CAUTION
Allow the balance to warm up for 30 minutes (0.1 mg models need 60 minutes) to adapt itself to the ambient
conditions.
Connect the AC/DC adapter to the connection socket on the back of your balance
and to the power line.
Attention
●
Check if the local power supply falls within this range. If this is not the case, under no circumstances con
nect the AC/DC adapter to the power supply, but contact a METTLER TOLEDO representative.
●
The power plug must be accessible at all times.
●
Prior to use, check the power cable for damage.
●
Route the cable in such a way that it cannot be damaged or cause a hindrance when working.
●
Ensure that no liquid ever comes into contact with the AC/DC adapter.
19Installation and Putting into Operation
4.5Leveling the balance
B
B
A
A
Exact horizontal positioning and stable installation are prerequisites for repeatable and accurate weighing
results.
Note
To compensate for small irregularities or inclinations (±2 %) at the location, the instrument must be leveled
and adjusted each time it is moved to a new location.
The balance can be levelled with the leveling assistant and/or with the level indicator on top of the terminal.
There are two adjustable leveling feet to compensate for slight irregularities in the surface of the weighing
bench.
4.5.1Leveling the balance with the level indicator
All balances have a level indicator with an air bubble to level the balance manually.
Leveling balances with readability of 0.1 mg and 1 mg
– Adjust the two leveling feet until the bubble comes to rest
exactly in the middle of the glass:
Air bubble at"12 o'clock"turn both feet clockwise.
Air bubble at"3 o'clock"turn left foot clockwise, right
foot counterclockwise.
Air bubble at"6 o'clock"turn both feet counterclockwise.
Air bubble at"9 o'clock"turn left foot counterclockwise,
right foot clockwise.
Leveling balances with readability of 0.01 g and 0.1 g
1 Remove the clamps (A) for the safety feet by turning them out
wards.
2 Turn the clamps (A) outwards as far as they will go (~ 90°), so
that the safety feet can move freely.
3 Now level the balance by turning both leveling screws (B) until the
bubble is in the inner circle of the level indicator (see procedure
above).
4 Secure the safety feet by turning the clamps (A) inwards as far as
they will go.
20Installation and Putting into Operation
4.5.2Leveling the balance using the leveling assistant
When the balance is switched on at its new location, the symbol The instrument is out of level appears in
the status information field at the left side of the screen.
1 Tap on [ ].
The screen Notifications opens.
2 Select The instrument is out of level.
The function Leveling assistant opens.
The function Leveling assistant is a step-by-step guidance that helps in leveling the balance.
After following the instructions, the leveling assistant will show the next steps. Follow the steps until the balance
is levelled.
Note
Always use the physical air bubble of the level indicator as a reference.
21Installation and Putting into Operation
4.6Transporting the balance
B
A
Switch off the balance and remove the power cable and any interface cable from the balance. Refer to the notes
in section Selecting the location (page18) regarding the choice of an optimal location.
Transporting over short distances
CAUTION
For balances with a draft shield:
Never lift the balance by its glass draft shield. The draft shield is not sufficiently fastened to
the balance.
Transporting over long distances
CAUTION
If you would like to transport or send your balance over long distances, use the complete
original packaging.
4.7Weighing below the balance
The balances are equipped with a hanger for carrying out weighings below the work surface.
CAUTION
●
Do not place the balance on the pan support location bolt (0.1 mg and 1 mg models).
●
Models with a glass draft shield: Carefully lift the draft shield from the weighing platform and put it aside.
1 Switch off the balance and remove the power cable and any inter
face cable from the balance.
2 Remove the draft ring element (10 mg models).
3 Remove the weighing pan and pan support.
4 Remove the bottom plate and unlock the draft shield (models with
draft shield). Carefully lift the draft shield from the weighing plat
form and put it aside.
5 Turn the balance carefully on its side.
6 Remove and retain the 2 screws (A) and the cover plate (B). The
hanger is now accessible.
7 Then return the balance to its normal position and simply reinstall
all components in the reverse order.
22Installation and Putting into Operation
4.8General requirements
4.8.1Warming up the balance
Before working with the balance, it must be warmed up in order to obtain accurate weighing results. To reach
operating temperature, the balance must be connected to the power supply for at least:
●
30 minutes for balances with a readability of 1mg (0.01ct) to 5g.
●
60 minutes for balances with a readability of 0.1mg (0.001ct) and higher.
4.8.2Adjusting the balance
To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its
location and depending on the ambient conditions. After reaching the operation temperature, an adjusting is
necessary in the following cases:
●
Before the balance is used for the first time.
●
When the balance (readability of 0.1 mg) was disconnected from the power or in case of power failure.
●
After a change of location and after the warming-up phase.
●
At regular intervals during weighing service.
23Installation and Putting into Operation
5Weighing Made Simple
This section describes how to perform a simple weighing. In addition, the basic concept of the navigation and
the basic functions of the balance are explained.
5.1Switching on the balance
Switching on the balance for the first time
1 Remove any load from the weighing pan.
2 Connect the balance using the AC/DC adapter to the mains.
After the start screen has disappeared, the balance starts with the application home screen.
After the balance has switched on for the first time, it can be switched on by pressing long on [ ].
Note
When the balance is switched on for the first time, the home screen of the application Weighing opens. If the
balance is switched on again, it always starts with the home screen of the application that was used before
switching off.
5.2Switching off the balance
1 Press on [ ] and hold until the dialog Switch-off appears.
2 Tap on [ ].
the balance switches off and goes into standby mode.
Note
●
After switching on from standby mode, the balance needs no warm-up time and is immediately ready for
weighing.
●
If the balance has been switched off after a preselected time, the display is dimly lit and shows the time and
the date.
●
If the balance has been switched off manually, the display is off.
To completely switch off the balance, it must be disconnected from the power supply.
24Installation and Putting into Operation
5.3Main settings and activities at a glance
The diagram below provides an overview of the main settings of an application (e.g. Weighing). Depending on
the application, the selectable options and their content can differ. Every application is based on this concept.
25Weighing Made Simple
5.4Navigation on the touch screen
To interact with the balance, use the screen and the operating keys at the bottom of the screen. The navigation
on the screen is similar to the navigation of a smart phone or tablet PC.
5.4.1Opening an application
To open settings or applications, tap with your finger on the symbol of the application (e.g. [ ] Weighing).
5.4.2Scrolling
When the options are too numerous to be shown on one screen, a blue bar on the right side appears. This
means that the user can scroll up and down. For scrolling, simply place your finger somewhere in the list and
drag it up or down.
26Weighing Made Simple
5.4.3Using shortcuts
To simplify navigation on the capacitive color TFT touch screen, there are a few shortcuts that allow quick
access to the most important areas of the balance. The weighing value field on the application home screen for
example works as a shortcut (see diagram below), so does the weighing unit next to the weighing value field.
Depending on the application, there may be other shortcuts that can be used.
Note
Every setting that can be changed directly via shortcut, can also be changed in the main configuration settings
of the application.
5.5Performing a simple weighing
When you switch on the balance for the first time, the home screen of the application Weighing opens auto
matically. If the balance was used already, the last used application before the balance was switched off
opens. If another application is running, switch to the application Weighing.
Navigation:
[ ] > Activities - Weighing and other applications > [ ] Weighing
1 Press [] to zero the balance or [] to tare the balance.
The initial screen looks like this:
2 Place a sample on the weighing pan.
The unstability symbol appears and the value in the weighing value field becomes light blue.
27Weighing Made Simple
After a short period of time, the weighing pan is stable. The unstability symbol disappears and the
value in the weighing value field becomes dark blue again.
The weighing process is finished. The result appears on the screen.
5.6Switching the weighing unit
There are several weighing units available. The default value is country-specific.
The weighing unit can be chosen via the main configuration of the current application or via shortcut. This
example describes how to change the weighing unit via shortcut.
1 Tap on the weighing process unit (shortcut) gram (g).
28Weighing Made Simple
the screen Main unit opens.
2 Put your finger somewhere in the list and drag up to scroll down.
3 Choose another weighing unit (e.g. ounce (oz)) by tapping on it.
4 Tap on [ ] to confirm the chosen weighing unit.
29Weighing Made Simple
The weighing unit gram (g) has been changed to ounce (oz).
Note
With approved balances, this menu topic has a fixed setting and cannot be changed.
30Weighing Made Simple
5.7Changing the resolution
There are several resolutions available. The default resolution is instrument-specific.
The resolution can be changed as follows:
1 Tap on the weighing value field.
2 Tap on 1d - 0.0001 g (Default).
3 Confirm the chosen resolution by tapping on [ ].
31Weighing Made Simple
5.8Zeroing
1 Unload the balance.
2 Press [] to set the balance to zero.
All weight values are measured in relation to the zero point.
Note
Press [] before starting the weighing process.
5.9Taring
1 Tap on [] to set the balance to zero.
2 Place the empty container on the balance.
3 Press [] to tare the balance.
The resolution has been changed.
The weight is displayed.
The status information field shows Net and the weighing value field shows 0.000 g.
32Weighing Made Simple
5.10Working with IDs
Identifications (IDs) contain descriptive text for individual measurements, which enables samples to be easily
allocated to specific tasks or customers. This feature defines identifications in order to comment measurements,
such as company ID, badge ID or sample ID.
Identifications must be defined in the contextual settings of an application. Usage and definition of the ID differ
depending on the application the ID is used for.
Identification dialog screen
The Identification dialog screen can slightly differ depending on the application in which the ID is used. The dia
log screen always consists of two parts:
●
The table with the ID definitions on the upper part of the screen.
●
The Workflow handling options on the lower part of the dialog screen.
5.10.1Defining an ID
A new ID is defined as follows:
1 Open an application (e. g. Weighing).
2 Tap on the Main configuration symbol on the upper left corner.
3 Tap on [ID] Identifications
4 Tap on ID 1.
The input dialog ID 1 opens. The input dialog is inactive.
5 Activate ID 1 with the switch in the title bar.
The input dialog ID 1 is activated.
The title bar shows ID 1 - Label.
6 Define ID 1 - Label.
7 Tap on [ ] to confirm the definition.
The input dialog ID 1 - Value opens.
8 Define ID 1 - Value.
9 Tap on [ ] to confirm the definition.
The screen Identifications opens showing the defined ID.
10 Tap on [ ] to accept the defined ID.
Repeat the steps for the other IDs.
33Weighing Made Simple
Note
The maximum length of an ID is 16 characters.
The picture below shows an example of defined IDs. IDs 1-3 are defined, ID 4 is still inactive.
5.10.2Workflow handling options
The workflow handling options differ depending on the application in which they are used. The following func
tions are available:
●
Input prompt
●
Autoincrement
●
Assign to sample
Input prompt
The function Input prompt can be used for every ID. If the function Input prompt is activated, the ID will be
prompted on the display before it is used. The user can decide whether to use the default value that has been
defined with the ID or to define an individual value.
Autoincrement
The function Autoincrement specifies that the last part of the ID is incremented with each use of that ID. There
are two basic functions depending on how the ID is defined:
●
If there is no counter in the ID, the system automatically adds a counter to the ID starting with 1 (e.g. the ID
Process will be Process1 in the next use).
●
If a counter is part of the ID, the system automatically increments the ID starting at that counter (e.g. the ID
Process 1 will be Process 2 in the next use).
Note
●
The counter must be set at the end of the ID, otherwise the system doesn't recognize the number as a
counter (e.g. in 567Apple the system doesn't recognize 567 as a counter).
●
If the ID has no counter and a maximum length of 16 characters, the last few characters will be overwritten
by the counter.
Assign to sample
If the option Assign to sample is activated, the ID will be used for each sample.
34Weighing Made Simple
5.11Statistics
The function Statistics generates statistics for a series of values. The option is available for every application.
Defining Statistics
1 Open an application (e.g. Weighing).
2 Tap on the main configuration symbol of the application.
The main configuration screen opens.
3 Tap on [ ].
The screen Statistics opens.
4 Activate the function Statistics.
5 Define the available options.
6 Tap on [ ] to confirm the adjustments.
Statistics configuration
The following options can be defined:
ParameterExplanationValues
Acceptance
range
Accepting mode To define if a weight sample is added automatically to the result.
* Factory setting
Note
The function Statistics is not available for the applications Totaling and Formulation.
To define the acceptable deviation in relation to the average val
ue.
The sample is automatically accepted if: the sample is at least
10d and a deflection according to the SNR command described
in the MT-SICS manual MT-SICS interface commands and func
tions (page76) was detected.
1%...100% (10%*)
Manual* | Automatic
35Weighing Made Simple
6Activities
The section Activities can be opened by tapping on [ ].
The section Activities includes the following two sub-sections:
●
Activities - Weighing and other applications
●
Activities - Adjustments and tests
6.1Activities - Weighing and other applications
Activities - Weighing and other applications includes the following applications:
●
Weighing see at section Weighing (page37) and Weighing Made Simple (page24)
●
Counting see section Counting (page38)
●
Percent Weighing see section Percent Weighing (page40)
●
Formulation see section Formulation (page42)
●
Dynamic Weighing see section Dynamic Weighing (page44)
●
Check Weighing see section Check Weighing (page46)
●
Factor Weighing see section Factor Weighing (page50)
●
Totaling see section Totaling (page51)
36Weighing Made Simple
6.1.1Weighing
The application Weighing allows the user to perform simple weighings. For more information about the basic
weighing functions see the section Weighing Made Simple (page24).
Navigation
[ ] Activities > Activities - Weighing and other applications > [ ] Weighing
Weighing – Main configuration
The following options can be defined:
ParameterExplanationValues
Main unitTo define the main unit of the weighing process.Available values are
Additional unitTo define an additional weighing unit.Available values are
ResolutionTo define the resolution of the weighing process.Available values are
model-specific.
model-specific.
model-specific
37Activities
6.1.2Counting
The application Counting determines a specific number of pieces, based on pre-determined reference piece
weight.
Navigation
[ ] Activities > Activities - Weighing and other applications > [ ] Counting
Defining Reference piece weight
To define the Reference piece weight, Reference piece number and Reference weight must successively be
defined. The system will automatically navigate from one option to the other.
Defining Reference piece number
1 Tap on [].
the screen Counting – Main configuration opens.
2 Tap on [Reference piece weight].
The dialog screen Reference piece number opens.
3 Delete the value by tapping on [ ].
4 Enter the number of reference pieces.
5 Tap on [ ] to confirm the defined reference piece number.
Note
The reference piece number must be a natural number between 1 to 999.
Defining Reference weight
Reference weight can be defined manually by entering the weight of the reference pieces or by weighing the
reference pieces.
Defining Reference weight manually
1 Delete the reference weight by tapping on [ ].
2 Enter the new reference weight.
3 The reference unit can be changed by tapping in the sensitive area at the right side of the input field.
4 Confirm the new value by tapping on [ ].
The reference piece weight has been defined.
5 Confirm the new reference piece weight by tapping on [ ].
Defining Reference weight by weighing
1 Tap on [ ].
2 Place reference weight on the weighing pan.
3 Tap on [ ] to confirm the weight.
The screen Counting – Main configuration opens.
4 Tap on [ ] to confirm the configurations.
38Activities
The following picture shows the home screen of the application. The work title bar displays the defined Refer
ence piece number and Reference piece weight that can be used as shortcuts.
Note
A fixed minimum reference piece number of 10 and inactive reference weight options for approved balances
(except for some selected countries).
Counting – Main configuration
The following options can be defined:
ParameterExplanationValues
Reference piece
weight
Additional unitTo activate/deactivate and define the additional unit of the weigh
* Factory setting
To define the number and the weight of the reference piece/s.1...999
(10*)
Available values are
ing process.
model-specific.
39Activities
6.1.3Percent Weighing
With the application Percent Weighing a sample weight can be checked as percentage to a reference target
weight.
Navigation
[ ] Activities > Activities - Weighing and other applications > [ ] Percent Weighing
Defining Reference weight
There are two possible ways to define the reference weight: The reference weight can be defined manually by
entering the value or by weighing the reference weight.
Defining the reference weight by weighing
1 Tap on [].
The screen Percent weighing – Main configuration opens.
2 Tap on Reference weight.
3 Tap on [ ].
4 Press [] to zero the balance.
5 Place the reference weight on the weighing pan.
6 Tap on [ ] to accept the value of the reference weight.
The screen Reference weight reappears.
7 Tap on [ ] to accept the reference weight.
8 Tap on [ ] to accept the reference weight and to return to the application home screen.
Defining the Reference Weight manually
1 Tap on [].
2 Tap on Reference weight.
3 Delete the default value by tapping on [ ].
4 Enter the value of the reference weight.
5 Confirm the reference weight by tapping on [ ].
The reference weight has now been defined and will be shown in the work title bar.
40Activities
The picture shows the home screen of the application Percent Weighing. The work title bar displays a defined
reference weight of 200g that can be used as shortcut.
Percent weighing – Main configuration
The following options can be defined:
ParameterExplanationValues
Reference
weight
Main unitTo define the main unit of the weighing process.Available values are
To define the reference weight manually or by weighing.-220.0000 ... -
0.0010 | 0.0010 ...
220.0000
model-specific.
41Activities
6.1.4Formulation
The application Formulation allows the user to:
●
Weigh-in (add and store) up to 999 individual component weights and display the total.
●
Tare/pre-tare and store up to 999 container weights and display the total.
●
fill up the sum of all component net weight values by adding a further component to a higher value.
Note
Total number of containers + component weights ≤ 999.
Navigation
[ ] Activities > Activities - Weighing and other applications > [ ] Formulation
Performing Formulation
1 Press [] to zero the balance
2 If a container is used: Place container on the weighing pan and press [] to tare the balance.
If the balance is tared, the status information field shows Net.
3 Place the first component weight.
The weighing value field shows the value of the first component weight.
4 Tap on [ ] to add the first component weight.
5 Place the second component weight.
The weighing value field shows the value of the second component weight.
6 Tap on [ ] to add the second component weight.
7 Continue adding components untill all the components are weighed.
Defining Fill up sample
With the function Fill up sample an additional component weight can be added to the total weight of all com
ponents to reach a desired target weight.
The weighing value field shows the total net weight.
1 Tap on [ ].
2 Place the fill up sample.
3 Tap on [ ] to confirm.
Discard value
The last weighing result can be discarded as follows:
1 Tap on [ ].
The dialog screen Confirm discard opens. It is an overview of all the weighing results within the current
weighing process.
2 Tap on [ ] to discard the last weighing result.
CAUTION
Discard is only possible until the last time the balance has been tared.
42Activities
Display results
When the weighing process is finished, the results can be shown as follows:
1 Tap on [ ].
2 Tap on [ ] View result.
Note
The results can also be displayed by using the shortcut in the work title bar.
Note
Depending on the procedure, several containers can be used (number of used containers ≤ 999).
Pause the application
1 Tap on [ ].
2 Tap on [ ] Pause.
The application Formulation is paused and another application can be used in the meantime.
3 Open the application Formulation again.
4 Tap on [ ].
The process can be continued.
Formulation – Main configuration
The following options can be defined:
ParameterExplanationValues
Main unitTo define the main unit.Available values are
model-specific.
ResolutionTo define the resolution of the weighing process.Available values are
model-specific.
Terminate the application
1 Tap on [ ].
2 Tap on [ ] Finish and discard data.
The home screen of the application Formulation opens.
43Activities
6.1.5Dynamic Weighing
The application Dynamic Weighing determines the weights of unstable samples or when the weighing process
is being executed under unstable ambient conditions. The balance calculates the weight as an average of a
number of weighing operations over a defined time.
Navigation
Activities > Activities - Weighing and other applications > [ ] Dynamic Weighing
Defining Measuring time
1 Tap on [].
The screen Dynamic weighing – Main configuration opens.
2 Tap on [Measuring time].
The dialog screen Measuring time in seconds opens.
3 Tap on [ ] to delete the value in the input field.
4 Enter a value between 3 and 120 seconds.
5 Tap on [ ] to confirm the entered value.
The screen Dynamic weighing – Main configuration opens.
6 Tap on [ ] to confirm the defined measuring time.
Note
The displayed measuring time on the application home screen can be used as shortcut to define the measuring
time.
Defining Start mode
1 Tap on [].
2 Tap on Start mode.
3 Chose Automatic or Manual.
4 Tap on [ ] to confirm the chosen option.
5 Tap on [ ] again.
The home screen of the application Dynamic Weighing opens again.
Performing Dynamic Weighing
When the measuring time and the start mode have been defined, the dynamic weighing process can be started.
1 Press [] to zero the balance.
2 If using a container: place container on the weighing pan and press [] to tare the balance.
3 Place sample weight.
If the option Start mode is set to Automatic, the weighing process start automatically with relative sta
bility.
It the option Start mode is set to Manual, Tap on [ ] to start the weighing process.
The weighing process starts. The defined measuring time in the work title bar is counting down.
44Activities
When the weighing process has finished, the screen looks as follows:
Note
The weighing process will automatically be aborted when an overload or underload is detected.
Dynamic weighing – Main configuration
The following options can be defined:
ParameterExplanationValues
Measuring timeTo define the measuring time in seconds.3...120
(3 seconds*)
Start modeTo define the start mode.Automatic*
Manual
Main unitTo define the main weighing unit.Available values are
model-specific.
Additional unitTo define the additional weighing unit.Available values are
model-specific.
ResolutionTo define the resolution of the weighing process.Available values are
model-specific.
* Factory setting
45Activities
6.1.6Check Weighing
The application Check Weighing allows the user to check the deviation of a sample weight within a tolerance
limit to a reference target weight. The target weight can be determined manually or by weighing; the tolerance
limit must be defined manually.
Navigation
[ ] Activities > Activities - Weighing and other applications > [ ] Check Weighing
Before performing Check Weighing
Before performing a check weighing, the following options can be defined:
●
Target weight
●
± Tolerance limits
●
Tolerance threshold
Defining the target weight manually by entering the nominal weight
1 Tap on [] .
The screen Check weighing - Main configuration opens.
2 Tap on [Target, tolerance].
3 Tap on [ ] Target weight.
4 Delete the default value by tapping on [ ].
5 Enter the nominal weight.
Tap on [ ] to confirm the defined nominal weight.
46Activities
Defining the target weight by weighing
1 Tap on [].
The screen Check weighing - Main configuration opens.
2 Tap on [Target, tolerance].
3 Tap on [ ] Target weight.
4 Tap on [ ] to open the weighing dialog.
5 Place the reference weight on the weighing pan.
6 Tap on [ ] to confirm the defined target weight.
Defining the lower and upper limits manually by entering a percentage value or weight
1 Tap on [].
The screen Check weighing - Main configuration opens.
2 Tap on [Target, tolerance].
3 Tap on [] ± Tolerance limits.
4 Delete the default value by tapping on [ ].
5 Enter the tolerance limit.
6 Tap on [ ] to accept the new tolerance limit.
Defining Tolerance threshold
With the option Tolerance threshold a value limit can be set. Is the value of the check weight below the defined
threshold, it won't be checked.
1 Tap on []. .
The screen Check weighing - Main configuration opens.
2 Tap on Tolerance threshold.
3 Define Tolerance threshold.
4 Tap on [ ].
5 Tap on [ ] to confirm.
Note
The option Tolerance threshold always refers to the lower tolerance limit.
47Activities
Performing Check Weighing
After defining the target weight and the tolerance limits, the application Check Weighing can be performed. The
weighing-in aid at the top bar provides a visual confirmation whether the sample falls within the defined toler
ances.
In the example below, the defined target weight is 100.000g and the tolerance limit is ±2.5%. The sample
weight is 97.0000g.
– Place the sample weight on the weighing pan.
The weight is stable and the unstability symbol disappears.
The value is out of tolerance, the weighing-in aid bar and the weighing value field are red.
In the example below, the defined target weight is still 100.000g and the tolerance limit is ±2.5%. The sam
ple weight is 99.0000g.
– Place the sample weight on the weighing pan.
The weight is stable and the unstability symbol disappears.
The value is within tolerances, the weighing-in aid bar and the weighing value field are green.
Note
If the weight is below a defined tolerance threshold, the background color of the screen doesn't change.
48Activities
Check weighing - Main configuration
The following options can be defined:
ParameterExplanationValues
Target, toler
ance
To define the target weight in the defined unit and its tolerance
limits in %.
Target weight: 100g*
± Tolerance limits:
0.10...100 (2.5%*)
Tolerance
threshold
Within tolerance
To define the threshold. Values below the defined threshold are
1%...100% (1%*)
not being checked.
To create an acoustic signal when the result is within tolerance.OFF* | ON
beeps
Main unitTo define the main unit of the weighing process.Available values are
model-specific.
ResolutionTo define the resolution of the weighing process.Available values are
model-specific.
* Factory setting
Note
The selectable weighing units and the resolution can differ depending on the balance model.
●
Only those weight units allowed by the appropriate national legislation are selectable.
●
With approved balances, this menu topic has fixed settings and cannot be changed.
49Activities
6.1.7Factor Weighing
The application Factor Weighing multiplies or divides a pre-defined factor by the measured weight value (in
grams) and calculates it to a predefined number of decimal places.
Navigation
[ ] Activities > Activities - Weighing and other applications > [ ] Factor Weighing
Defining Factor, step
1 Tap on [].
The screen Factor weighing – Main configuration opens.
2 Tap on [Factor, step].
The dialog screen Factor – Multiplication opens.
3 Tap on [ ] to delete the defined value.
4 Define Factor.
5 Tap on [ ] to change the operation from Multiplication to Division or vice versa.
6 Tap on [ ] to confirm.
7 Tap on [ ].
The dialog screen Step opens.
8 Define Step.
9 Tap on [ ] to confirm.
10 Tap on [ ] again to confirm the defined adjustments.
Note
The allowed range for the steps depends on the defined factor and the resolution of the balance.
Factor weighing – Main configuration
The following options can be defined:
ParameterExplanationValues
Factor, stepTo define the factor and the step.Factor | Step
Additional unitTo define an additional weighing unit.Available values are
model-specific.
50Activities
6.1.8Totaling
The application Totaling allows the user to weigh different samples, adding their weight values and to totalize
them.
Navigation
[ ] Activities > Activities - Weighing and other applications > [ ] Totaling
Performing Totaling
1 Press [] to zero the balance.
2 If using a container: Place empty container on the weighing pan and press [] to tare the balance.
3 Place the first weight.
4 Wait until the unstability symbol disappears.
5 Tap on [ ] to accept the weight and to start the procedure.
6 Place next sample weight.
7 Tap on [ ] to accept the second sample weight.
When the balance is stable, the weighing value becomes dark blue.
The work title bar shows the number of samples (2 samples) and the total weight of the samples ( =
30.0000g).
Discard value
If a weighing value was incorrect, it can be discarded from the result as follows:
1 Tap on [ ].
The screen Confirm discard opens.
2 Tap on [ ].
The incorrect value has been deleted. The weighing process can be continued.
Display results
1 Tap on [ ].
2 Tap on [ ] View result.
Note
The results can also be displayed by using the shortcut in the work title bar.
51Activities
Pause the application
1 Tap on [ ].
2 Tap on [ ] Pause.
The application Totaling is paused and another application can be used in the meantime.
3 Open the application Totaling again.
4 Tap on [ ].
The process can be continued.
Totaling – Main configuration
The following options can be defined:
ParameterExplanationValues
Accepting mode To define if a weight sample is added automatically to the result.
The sample is automatically accepted if: the sample is at least
10d and a deflection according to the SNR command described
in the MT-SICS manual MT-SICS interface commands and func
tions (page76) was detected.
Main unitTo define the main unit of the weighing process.Available values are
ResolutionTo define the resolution of the weighing process.Available values are
Terminate the application
1 Tap on [ ].
2 Tap on [ ] Finish and discard data.
The home screen of the application Totaling opens.
Note
If the option Workflow reports is set to Automatic, a printer symbol appears in the list element. By tapping on
[ ] Discard data the process can be aborted without printing the reuslts.
Manual* | Automatic
model-specific.
model-specific.
52Activities
6.2Activities - Adjustments and tests
Activities - Adjustments and tests contains the following elements:
●
FACT, see section FACT Fully automatic internal adjustment (page54)
●
Internal Adjustment, see section Internal Adjustment (page55)
●
External Adjustment, see section External Adjustment (page56)
53Activities
6.2.1FACT Fully automatic internal adjustment
FACT stands for Fully Automatic Calibration Technology. It is activated as default value.
FACT means that the balance adjusts itself based on the following criteria:
●
in case the conditions change (temperature difference > 2°C), which could lead to a noticeable deviation in
the measurement.
●
on a predefined day time programmed by the user.
Setting FACT
You can define the day time of FACT as follows:
1 Tap on [ ].
2 Tap on [ ]. .
The screen Activities - Adjustments and tests opens.
3 Tap on [FACT].
4 Activate Fully automatic adjustment.
The Fully automatic adjustment dialog opens.
5 Select the time (hours : minutes) with the pick buttons.
6 Tap on [ ] to confirm the defined time.
The time underneath FACT has been updated and shows the time of the daily adjustment.
CAUTION
If the function FACT is not activated, all the FACT functionalities, such as temperature FACT and time FACT are
inactive.
Note
When defining the time, hold the pick button to scroll faster.
54Activities
6.2.2Internal Adjustment
Note
The function Internal Adjustment is available for models with internal weight only (see technical data).
Perform an internal adjustment manually
1 Tap on [Internal Adjustment].
The internal adjustment procedure starts. The screen shows Adjustment ongoing, please wait....
When the internal adjustment procedure is successfully completed, the results of the internal adjustment
appear.
2 Tap on [ ] to confirm the results.
Note
If the balance is configured and connected to a printer, the results of the adjustment process are printed.
55Activities
6.2.3External Adjustment
Note
Because of certification legislation, approved balances cannot be adjusted with an external weight (depending
on the certification legislation of the selected country).
Performing an external adjustment
1 Open Activities - Adjustments and tests.
2 Tap on External Adjustment.
The External Adjustment screen opens.
3 Tap on [] to define the adjustment weight according to the weight certificate.
4 Confirm the adjustment weight by tapping on [ ].
5 Prepare the adjustment weight and tap on [ ] to start the adjustment process.
6 Load the adjustment weight in the center of the weighing pan.
7 Unload the adjustment weight from the weighing pan.
When the external adjustment procedure is successfully completed, the result of the external adjustment
appear.
56Activities
7General Settings
The section General settings is divided into two sub-sections:
●
Quick settings/Preferences
●
General configuration and data
7.1Quick settings/Preferences
The following options are available:
●
Date and Time
●
Brightness
●
Sound
●
Language
●
Glove mode
●
Quick Adjustment
●
Leveling assistant
7.1.1Date and time
In this menu topic, the date and time can be defined. Tap on [ ] for Time and tap on [25] for Date.
The following parameters can be defined:
ParameterExplanationValues
TimeTo define the time format.24:MM* | 12:MM |
The date can be defined with the pick buttons.Hours | Minutes
DateTo define the date format.DD.MM.YYYY* |
The date can be defined with the pick buttons.Day | Month | Year
Notifies when zero/tare bar actions are not available, provides
additional feedback in case of input errors, messages and status
notifications.
zero/tare bar.
High
OFF | Low* | Medium |
High
OFF* | Low | Medium |
High
7.1.5Glove mode
With the function Glove mode activated, the touch-screen becomes more sensitive and easier to navigate when
wearing gloves.
ParameterExplanationValues
Glove modeTo activate or deactivate the function Glove mode.ON | OFF*
* Factory setting
7.1.6Quick Adjustment
ParameterExplanationValues
Quick Adjust
ment
* Factory setting
Note
If the option Quick Adjustment is activated, the symbol appears in the main navigation of an application.
7.1.7Leveling assistant
The function Leveling assistant helps to adjust the balance (e.g. when the balance has changed location).
For more information, see section Leveling the balance (page20).
To activate/deactivate and define the function Quick Adjustment. Internal Adjustment |
External Adjustment
58General Settings
7.2General configuration and data
The following options are available:
●
Weighing options
●
Publishing
●
Devices and connectivity
●
System settings
●
ISO-Log
●
Access protection
7.2.1Weighing options
The following options can be defined:
ParameterExplanationValues
Weighing modeTo define Weighing mode.Universal* = for stan
EnvironmentTo adapt the balance to the environmental circumstances.Standard* = for stan
AutozeroTo activate or deactivate AutozeroON* | OFF
AutoTareTo activate or deactivate AutoTareON | OFF*
MinWeighTo define MinWeighNot defined* | Cus
RecallTo retain and recall the last stable weight.ON | OFF*
* Factory setting
Note
The function Autozero can not be deactivated for approved balances (except for some selected countries).
dard weighing applica
tions
Dosing = for dosing liq
uid or powdery samples
dard environments
Unstable = for unstable
environments
Stable = for stable
environments
tomized | Certificate
59General Settings
MinWeigh method
The option MinWeigh method notifies when a measurement value is not fulfilling the required weighing accura
cy. The usage of MinWeigh is application-specific (contextual settings).
Defining MinWeigh method
1 Tap on MinWeigh.
The dialog MinWeigh configuration opens.
2 Tap on MinWeigh method.
The dialog MinWeigh method opens.
3 Select method and confirm by tapping on [ ].
Depending on the chosen method, the selectable options in the dialog MinWeigh configuration vary.
The following methods are available:
●
Customized
●
Certificate
●
OIML
Note
The method OIML is only available for approved balances.
7.2.2Publishing
In this section, the print and output options can be defined.
Note
Depending on the connected peripheral device (see section Devices and connectivity (page62)) the available
options in this section can differ. It is possible that not every options described in this section is available for
the used peripheral device.
Print
This option can be activated or deactivated.
The following options can be defined:
ParameterValues
Single valuesManual, stable* | Manual, all values | Automatic, stable
Workflow
reports
Automatic* |
Manual
FACT reportON* | OFF
* Factory setting
60General Settings
Printouts data (Weighing applications only)
The following options can be defined:
ParameterExplanationValues
Date, timeTo add date and time.ON | OFF*
Balance infor
mation
MinWeigh Infor
mation
Leveling infor
mation
Gross/TareTo add gross/tare to single values.ON | OFF*
Signature lineTo add the signature line.ON | OFF*
Empty linesTo add empty lines.ON | OFF*
* Factory setting
Send weight value
This option can be activated or deactivated.
The following options can be chosen:
ParameterExplanation
Manual, stable* Send next stable weight at will
Manual, all val
ues
Automatic, sta
ble
Automatic, con
tinuous
* Factory setting
To add information about the balance.ON | OFF*
To activate/deactivate MinWeigh InformationON | OFF*
To activate/deactivate Leveling informationON | OFF*
Send any stable or unstable weight at will
Send next stable weight automatically
Send any stable or unstable weight automatically
Autopublish
The option Autopublish can be activated or deactivated with the switch in the work title bar.
A numerical value for the function Autopublish can be defined.
61General Settings
7.2.3Devices and connectivity
This section describes how the peripheral devices can be configured.
cation
Sleep modeTo activate/deactivate and define the function Sleep mode.After 30 seconds
Backlight OFFTo define the functionBacklight OFF.After 30 seconds
Out of level noti
fication
Service due
notification
* Factory setting
System and data management
The following options can be defined:
ParameterExplanation
Touch screen
adjustment…
Level indicator
center
adjustment…
Reset balance… To reset the balance to default.
Note
Run the Level indicator center adjustment only when the electronic in/out level notification (status icon) and the
visual bubble position don't match.
CAUTION
By resetting the balance, any changes to general settings and contextual settings that have been made as well
as any temporary collected data (e.g. paused applications or statistics) will be lost.
A balance identification that can be defined by the user.Values can be defined
individually.
After 1 minute
After 2 minutes
After 5 minutes
After 10 minutes*
After 1 minute
After 2 minutes
After 5 minutes
After 10 minutes
To activate/deactivate the function Out of level notification.ON* | OFF
To activate/deactivate the function Service due notification.ON* | OFF
The balance is executing a screen adjustment.
The adjustment sets the exact center position of the level indicator and corrects any possi
ble shifts.
7.2.5ISO-log
In this section, detailed history information about performed adjustments, intensity of use and settings can be
displayed.
The following options can be selected:
ParameterExplanation
ISO-Log –
Adjustments
ISO-Log – Bal
ance
ISO-Log – Set
tings and status
Shows detailed information about the performed adjustments.
Shows detailed information about the balance history.
Shows detailed information about changes of the settings.
63General Settings
7.2.6Access protection
With the function Access protection, certain functionalities of the balance can be protected by a numerical
passcode.
The following options can be defined:
ParameterExplanationValues
ApplicationsTo protect the application.ON | OFF*
Adjustments and
tests
SettingsTo protect all settings.ON | OFF*
PasscodeTo define the passcode by the user.1....9 (1-16 digits)
* Factory setting
User passcode
The user passcode can be defined by the user. The default value is 12345678. The length of the freely change
able user passcode is restricted to 16 digits.
Note
Passcode and access options are not affected by a balance reset.
What if you forgot the passcode?
When you forgot or lost the passcode, please contact a Mettler Toledo representative or visit the Mettler Toledo
balance support site.
To protect adjustments and tests.ON | OFF*
64General Settings
8Communication with Peripheral Devices
8.1PC-Direct Function
The numerical value displayed at the balance can be transferred to the cursor position in Windows Applications
(e.g. Excel, Word), by typing with the keyboard.
Note: The units will not be transferred.
Requirements
●
PC with one of the following Microsoft Windows® operating systems 32bit/64bit: XP (SP3), Vista (SP2),
Win 7 (SP1) or Win 8
●
Serial interface RS232, USB or BT option
●
Administrator rights for installing software (for USB not required)
●
Windows Application (e.g. Excel)
●
Balance to PC connection with cable RS232 or USB
Settings on the balance:
CAUTION
Disconnect the USB connection from the balance prior to changing settings.
Balance Interface Settings (see Interface Menu):
●
Topic RS232 or USB Device: set PC-Direct and select the most appropriate option for the desired weighing
result.
●
Topic End of line (transmitted end of line character):
– set <TAB> to write into the same row (e.g. in Excel).
– set <CR><LF> to write into the same column (e.g. in Excel).
●
Save changes.
Settings on the PC:
Installing SerialPortToKeyboard
Operation of PC-Direct via serial port RS232 requires the installation of SerialPortToKeyboard on your host
computer.
Using CD-ROM
1 Insert the product CD in the CD/DVD drive of the host computer.
2 Double click the folder SerialPortToKeyboard.
Using internet
1 Go to the site http://www.mettler-toledo-support.com.
2 Log in to the METTLER TOLEDO Balance Support Site (registration with the serial number of a METTLER
TOLEDO instrument is required).
3 Click Customer Support
4 Click appropriate product folder and save the program file SerialPortToKeyboard.exe on your specified
storage location.
Installation procedure
1 Right-click on SerialPortToKeyboard.exe and select Run as Administrator from the menu.
2 Follow the installer's instructions.
65Communication with Peripheral Devices
Settings for SerialPortToKeyboard
➀
➁
➂
1 Select the serial port (COM) to be used for connec
tion with the balance.
2 Set the baud rate to 9600.
3 Activate "Connect"
Note
Note
●
The window can be minimized.
●
Closing the window terminates the session.
Checking operation
1 Start SerialPortToKeyboard (RS232)
2 Start Excel (or another application) on the PC.
3 Activate a cell in Excel.
According to your selected PC-Direct option, the displayed values will appear e.g. in the column one after the
other one in the different rows.
8.2USB Device Interface
To perform the functionality USB DEVICE with a PC equipped only with a USB Interface, you have to install an
appropriate USB Driver on the PC first.
Requirements
●
Balance with USB Device Interface.
●
PC with one of the Microsoft Windows® operating system 32bit/64bit: XP (SP3), Vista (SP2), Win 7 (SP1)
or Win 8.
●
Administrator rights for installing software.
●
PC to balance USB connection cable.
Installing USB driver on the PC:
Using CD-ROM
1 Insert the product CD in the CD/DVD drive of the host computer.
2 Double click the folder USB Driver.
3 Click USBDriverInstaller.exe.
Using internet
1 Connect to the Internet
2 Go to the site http://www.mettler-toledo-support.com.
3 Log in to the METTLER TOLEDO Balance Support Site (registration with the serial number of a METTLER
TOLEDO instrument required).
4 Click Customer Support.
5 Click appropriate product folder.
6 Click USB Driver.
66Communication with Peripheral Devices
7 Click USBDriverInstaller.exe.
Installing procedure
1 Click Save to download to your specified location.
2 Right-click on the downloaded install program: USBDriverIn
staller.exe
3 If a safety warning appears, allow Windows to install.
4 Click Next and follow the installer's instructions.
Installing instrument
1 Switch the balance off.
2 Connect the balance to the preferred USB Port on the PC.
3 Switch the balance on.
4 Follow the instructions of the Wizard and install the software auto
matically (recommended)
Note: The wizard appears again for each USB port, either on your PC
or if another balance is connected.
Warning: Do not click Cancel as for the connected USB port, it might
not be possible anymore to perform the installation process.
67Communication with Peripheral Devices
9Firmware (Software) Updates
METTLER TOLEDO is continuously improving its balance firmware (software) for the benefit of customers. In
order that the customer can benefit quickly and easily from further developments, METTLER TOLEDO makes the
latest firmware versions available on the Internet. The firmware made available on the Internet has been devel
oped and tested by Mettler-Toledo AG using processes that meet the guidelines of ISO 9001. Mettler-Toledo AG
does not, however, accept liability for consequences that might arise from updating the firmware.
9.1Operating principle
You will find all the relevant information and updates for your balance on the METTLER TOLEDO website at the
following address:
www.mettler-toledo-support.com
A program known as "e-Loader II" is loaded onto your computer together with the firmware update. You can
use this program to download the firmware to the balance. The "e-Loader II" can also save the settings in your
balance before the new firmware is downloaded to it. You can reload the saved settings into the balance manu
ally or automatically after the software is downloaded.
If the selected update includes an application that is not described in these instructions (or that has been updat
ed in the meantime) you can download the corresponding instructions in Adobe Acrobat® PDF format.
Note
New applications might not be visible unless the type data are updated by a service technician.
Requirements
The minimum requirements for obtaining applications from the Internet and downloading them into your bal
ance are as follows:
●
PC with one of the following Microsoft Windows® operating systems:
•
Microsoft® Windows® XP Home or Professional with Service Pack 3 (32 bit)
•
Microsoft® Windows Vista® Home Premium, Business, Ultimate, or Enterprise with Service Pack 2 (32 bit
and 64 bit)
•
Microsoft® Windows 7 with Service Pack 1 Home Premium, Professional, Ultimate, or Enterprise (32 bit
and 64 bit)
•
Microsoft ® Windows ® 8 , Professional or Enterprise
●
Administrator rights for installing software.
●
PC to balance connection cable (e.g. No. 11101051 see chapter accessories)
68Communication with Peripheral Devices
9.2Update procedure
Installing "e-Loader II" software from the Internet onto the PC.
1 Connect to the Internet.
2 Go to the site http://www.mettler-toledo-support.com.
3 Log in to the METTLER TOLEDO Balance Support Site (registration with the serial number of a METTLER
TOLEDO instrument required).
4 Click on Customer Support.
5 Click the appropriate product folder.
6 Click the firmware version (e-Loader II) you need and save it on your specified storage location.
7 Right-click on the firmware SNxxx.exe.
8 Follow the installer's instructions.
Loading the new firmware into the balance.
1 Right-click on METTLER TOLEDO e-Loader II and select Run as Administrator from the menu.
2 Follow the instructions, which will take you step-by-step through the installation.
69Firmware (Software) Updates
10Error and Status Messages
Error messages in the display draw your attention to incorrect operation or that the balance could not execute a
procedure properly.
Error messageCauseRectification
NO STABILTYNo stabilityEnsure more stable ambient condi
Adjustment aborted - Weight out
of range
EEPROM error - Please contact
your MT-Support representative.
Wrong adjustment weight on the
weighing pan or none at all.
●
EEPROM (memory) error.
●
Excessive mains voltage fluctu
ation or strong glitches
occurred.
tions. If not possible, check settings
for environment.
Place required adjustment weight in
center of pan.
Please contact your MT-Support
representative.
Wrong cell data - Please contact
your MT-Support representative.
No standard adjustment - Please
contact your MT-Support repre
sentative.
Program memory defect - Please
contact your MT-Support repre
sentative.
Temperature sensor defect Please contact your MT-Support
representative.
Wrong type data set - Please con
tact your MT-Support representa
tive.
Battery backup lost - Please
check date and time settings.
Weight is above initial zero
range.
Weight is below initial zero
range.
Memory fullMemory full.Clear the memory and start a new
Weight out of rangeSample weight is outside the
Wrong cell data.Please contact your MT-Support
representative.
No standard calibration.Please contact your MT-Support
representative.
Program memory defect.Please contact your MT-Support
representative.
Temperature sensor defect.Please contact your MT-Support
representative.
Wrong load cell brand.Please contact your MT-Support
representative.
Wrong type data set.Please contact your MT-Support
representative.
Backup battery is empty. This bat
tery ensures that the date and time
are not lost when the balance is
disconnected from power.
Overload - The weight on the pan
exceeds the weighing capacity of
the balance.
UnderloadCheck that the weighing pan is
Wrong weighing pan or pan is not
empty.
Wrong weighing pan or pan is
missing.
allowed range.
Connect the balance to the power
supply for charging the battery
(e.g. during the night) or contact
METTLER TOLEDO customer ser
vice.
Reduce the weight on the weighing
pan.
positioned correctly.
Mount correct weighing pan or
unload weighing pan.
Mount correct weighing pan.
evaluation.
Unload the pan and load a new
sample weight.
70Firmware (Software) Updates
11Maintenance
A
B
C
WARNING
Risk of electric shock
– The instrument must be disconnected from the power supply, before cleaning or other
maintenance work to be performed.
– Use only the power cord from METTLER TOLEDO, if it needs replacing.
Please observe the following notes:
Note
On no account use cleaning agents which contain solvents or abrasive ingredients, as this can result in dam
age to the touch screen.
11.1Cleaning the glass draft shield (0.1 mg and 1 mg models)
Remove the following parts:
1 Remove weighing pan, draft shield element and pan support.
2 Remove the bottom plate.
3 Unlock the draft shield, lift it off the balance and place it on a clean surface.
1 Push the glass doors (A) back.
2 Turn the two lock covers (B) on the front as
far as they will go.
1 Tilt the front glass (C) forward.
2 Remove the front glass.
71Maintenance
– Pull the top glass door (D) out from the
E
F
G
H
front.
– Push the lock button (G) to release the rear
glass.
– Lift the side glass doors (E) at (F) and pull
them out from the front.
– Remove the rear glass (H).
72Maintenance
1 Turn the draft shield lock to the (Service)
position.
2 Remove the draft shield lock.
After cleaning reinstall all components in the reverse order. For balance mounting see section Installing the
components (page17-18).
11.2Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies
to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point
specified for electrical and electronic equipment. If you have any questions, please contact
the responsible authority or the distributor from which you purchased this device. Should this
device be passed on to other parties (for private or professional use), the content of this regu
lation must also be related.
Thank you for your contribution to environmental protection.
73Maintenance
12Interface Specification
15
96
IN
GND
OUT
HAND
SHAKE
IN
DATA
OUT
RxD
TxD
CTS
RTS
OUT
POWER
SUPPLY
+12V
2nd di
splay mode only
12.1RS232C interface
Each balance is equipped with an RS232C Interface as standard for the attachment of a peripheral device (e.g.
printer or computer).
SchematicItemSpecification
Interface typeVoltage interface according to EIA
RS-232C/DIN66020 CCITT V24/V.28)
Max. cable length15 m
Signal levelOutputs:
Many of the instruments and balances used have to be capable of integration in a complex computer or data
acquisition system.
To enable you to integrate balances in your system in a simple manner and utilize their capabilities to the full,
most balance functions are also available as appropriate commands via the data interface.
All new METTLER TOLEDO balances launched on the market support the standardized command set "METTLER
TOLEDO Standard Interface Command Set" (MT-SICS). The commands available depend on the functionality of
the balance.
For further information please refer to the Reference Manual MT-SICS downloadable from the Internet under
u www.mt.com/sics-newclassic
76Interface Specification
13Technical Data
13.1General data
Power supply for 0.1 mg balances
●
AC adapterPrimary: 100–240VAC, -15%/+10%, 50/60Hz
●
Cable for AC adapter3-core, with country-specific plug
●
Balance power supply12VDC ±3%, 2.25A, maximum ripple: 80mVpp
13.2.1Balances with readability of 0.1 mg, S platform with draft shield
MS104TSMS204TS
Limit values
Maximum capacity120 g220 g
Readability0.1 mg0.1 mg
Repeatability (at nominal load) 0.1 mg0.1 mg
Linearity deviation0.2 mg 0.2 mg
Sensitivity temperature drift 1.5 ppm/°C1.5 ppm/°C
Typical values
Repeatability (at nominal load) 0.08 mg0.08 mg
Linearity deviation0.06 mg0.06 mg
Minimum sample weight (acc. to USP)0.16 g0.16 g
Minimum sample weight (U=1 %, k=2)0.016 g0.016 g
Minimum capacity OIML0.01 g0.01 g
Settling time2 s2 s
Adjustment Int.Cal / FACTInt.Cal / FACT
Balance dimensions (W x D x H)204x347x348 mm204x347x348 mm
Weighing pan dimensionsø 90 mmø 90 mm
Usable height of draft shield236 mm236 mm
Weight of balance 5.9 kg5.9 kg
Weights for routine testing
OIML CarePac
Weights
ASTM CarePac
Weights
#11123002
100g F2, 5g E2
#11123102
100g 1, 5g 1
#11123001
200g F2, 10g E2
#11123101
200g 1, 10g 1
78Technical Data
MS304TS
Limit values
Maximum capacity320 g
Readability0.1 mg
Repeatability (at nominal load) 0.1 mg
Linearity deviation0.2 mg
Sensitivity temperature drift 1.5 ppm/°C
Typical values
Repeatability (at nominal load) 0.08 mg
Linearity deviation0.06 mg
Minimum sample weight (acc. to USP)0.16 g
Minimum sample weight (U=1 %, k=2)0.016 g
Minimum capacity OIML0.01 g
Settling time2 s
Adjustment Int.Cal / FACT
Balance dimensions (W x D x H)204x347x348 mm
Weighing pan dimensionsø 90 mm
Usable height of draft shield236 mm
Weight of balance 5.9 kg
Weights for routine testing
OIML CarePac
Weights
ASTM CarePac
Weights
#11123001
200g F2, 10g E2
#11123101
200g 1, 10g 1
79Technical Data
13.2.2Balances with readability of 1 mg, S platform with draft shield
MS303TSMS403TS
Limit values
Maximum capacity320 g420 g
Readability1 mg1 mg
Repeatability (at nominal load) 1 mg1 mg
Linearity deviation2 mg 2 mg
Sensitivity temperature drift 3 ppm/°C3 ppm/°C
Typical values
Repeatability (at nominal load) 0.7 mg0.7 mg
Linearity deviation0.6 mg0.6 mg
Minimum sample weight (acc. to USP)1.4 g1.4 g
Minimum sample weight (U=1 %, k=2)0.14 g0.14 g
Minimum capacity OIML0.02 g0.02 g
Settling time1.5 s1.5 s
Adjustment Int. Cal / FACTInt. Cal / FACT
Balance dimensions (W x D x H)204x347x283 mm204x347x283 mm
Weighing pan dimensions127x127 mm127x127 mm
Usable height of draft shield168 mm168 mm
Weight of balance 5.6 kg5.6 kg
Weights for routine testing
OIML CarePac
Weights
ASTM CarePac
Weights
#11123001
200g F2, 10g F1
#11123103
200g 1, 10g 1
#11123000
200g F2, 20g F1
#11123100
200g 1, 20g 1
80Technical Data
MS603TSMS1003TS
Limit values
Maximum capacity620 g1020 g
Readability1 mg1 mg
Repeatability (at nominal load) 1 mg1 mg
Linearity deviation2 mg 2 mg
Sensitivity temperature drift 3 ppm/°C3 ppm/°C
Typical values
Repeatability (at nominal load) 0.7 mg0.7 mg
Linearity deviation0.6 mg0.6 mg
Minimum sample weight (acc. to USP)1.4 g1.4 g
Minimum sample weight (U=1 %, k=2)0.14 g0.14 g
Minimum capacity OIML0.02 g0.02 g
Settling time1.5 s1.5 s
Adjustment Int. Cal / FACTInt. Cal / FACT
Balance dimensions (W x D x H)204x347x283 mm204x347x283 mm
Weighing pan dimensions127x127 mm127x127 mm
Usable height of draft shield168 mm168 mm
Weight of balance 5.6 kg5.6 kg
Weights for routine testing
OIML CarePac
Weights
ASTM CarePac
Weights
#11123007
500g F2, 20g F1
#11123107
500g 1, 20g 1
#11123008
1000g F2, 50g F1
#11123108
1000g 1, 50g 1
81Technical Data
13.2.3Balances with readability of 0.01 g, S platform
MS1602TSMS3002TSMS4002TS
Limit values
Maximum capacity1620 g3200 g4200 g
Readability0.01 g0.01 g0.01 g
Repeatability (at nominal load) 0.01 g0.01 g0.01 g
Linearity deviation0.02 g 0.02 g 0.02 g
Sensitivity temperature drift 3 ppm/°C3 ppm/°C3 ppm/°C
Typical values
Repeatability (at nominal load) 0.007 g0.007 g0.007 g
Linearity deviation0.006 g0.006 g0.006 g
Minimum sample weight (acc. to USP)14 g14 g14 g
Minimum sample weight (U=1 %, k=2)1.4 g1.4 g1.4 g
Minimum capacity OIML0.5 g0.5 g0.5 g
Settling time1.5 s1.5 s1.5 s
Adjustment Int. Cal / FACTInt. Cal / FACTInt. Cal / FACT
Balance dimensions (W x D x H)194x347x99 mm194x347x99 mm194x347x99 mm
Weighing pan dimensions170x200 mm170x200 mm170x200 mm
Weight of balance 4.6 kg4.6 kg4.6 kg
Weights for routine testing
OIML CarePac
Weights
ASTM CarePac
Weights
#11123008
1000g F2, 50g F2
#11123106
1000g 1, 50g 1
#11123009
2000g F2, 100g
F2
#11123109
2000g 1, 100g 1
#11123010
2000g F2, 200g
F2
#11123110
2000g 4, 200g 4
82Technical Data
MS4002TSDRMS6002TSMS6002TSDR
Limit values
Maximum capacity4200 g6200 g6200 g
Maximum capacity, fine range820 g–1220 g
Readability0.1 g0.01 g0.1 g
Readability, fine range0.01 g–0.01 g
Repeatability (at nominal load) 0.1 g0.01 g0.01 g
Repeatability, at fine range (at nominal
load)
Linearity deviation0.2 g 0.02 g 0.2 g
Linearity deviation, fine range0.02 g –0.02 g
Sensitivity temperature drift 3 ppm/°C3 ppm/°C3 ppm/°C
Typical values
Repeatability (at nominal load) 0.05 g0.007 g0.05 g
Repeatability, at fine range (at nominal
load)
Linearity deviation0.02 g0.006 g0.02 g
Linearity deviation, at fine range0.006 g–0.006 g
Minimum sample weight (acc. to USP)14 g14 g14 g
Minimum sample weight (U=1 %, k=2)1.4 g1.4 g1.4 g
Minimum capacity OIML0.5 g0.5 g0.5 g
Settling time1.5 s1.5 s1.5 s
Adjustment Int. Cal / FACTInt. Cal / FACTInt. Cal / FACT
Balance dimensions (W x D x H)194x347x99 mm194x347x99 mm194x347x99 mm
Weighing pan dimensions170x200 mm170x200 mm170x200 mm
Weight of balance 4.6 kg5.1 kg5.1 kg
Weights for routine testing
OIML CarePac
Weights
ASTM CarePac
Weights
0.01 g–0.01 g
0.007 g–0.007 g
#11123010
2000g F2, 200g
F2
#11123110
2000g 4, 200g 4
#11123011
5000g F2, 200g
F2
#11123111
5000g 4, 200g 4
#11123011
5000g F2, 200g
F2
#11123111
5000g 4, 200g 4
83Technical Data
MS12002SMS12002TS
Limit values
Maximum capacity12200 g12200 g
Readability0.01 g0.01 g
Repeatability (at nominal load) 0.01 g0.01 g
Linearity deviation0.02 g0.02 g
Sensitivity temperature drift 3 ppm/°C3 ppm/°C
Typical values
Repeatability (at nominal load) 0.007 g0.007 g
Linearity deviation0.006 g0.006 g
Minimum sample weight (acc. to
USP)
Minimum sample weight (U=1 %,
k=2)
Minimum capacity OIML0.5 g0.5 g
Settling time1.5 s1.5 s
Adjustment Int. Cal / FACTInt. Cal / FACT
Balance dimensions (W x D x H)194x347x99 mm194x347x99 mm
Weighing pan dimensions170x200 mm170x200 mm
Weight of balance 5.2 kg5.2 kg
13.2.4Balances with readability of 0.1 g to 1 g, S platform
MS8001TS
Limit values
Maximum capacity8200 g
Readability0.1 g
Repeatability (at nominal load) 0.1 g
Linearity deviation0.2 g
Sensitivity temperature drift 3 ppm/°C
Typical values
Repeatability (at nominal load) 0.07 g
Linearity deviation0.06 g
Minimum sample weight (acc. to USP)140 g
Minimum sample weight (U=1 %, k=2)14 g
Minimum capacity OIML5 g
Settling time1 s
Adjustment Int. Cal / FACT
Balance dimensions (W x D x H)194x347x99 mm
Weighing pan dimensions170x200 mm
Weight of balance 4.6 kg
Weights for routine testing
OIML CarePac
Weights
ASTM CarePac
Weights
#11123011
5000g F2, 200g F2
#11123111
5000g 4, 200g 4
85Technical Data
13.3Dimensions
236
77
86
245
21
173
174
140
347
194
204
348
74
98
218
85
ø 90
115
140
99
13.3.1Balances with readability of 0.1 mg and draft shield
Models:
MS104TS
MS204TS
MS304TS
86Technical Data
13.3.2Balances with readability of 1 mg and draft shield
Sinker for density of liquids in conjunction with density kit00210260
Calibrated (sinker + certificate)00210672
Recalibrated (new certificate)00210674
Calibrated thermometer with certificate11132685
Weighing pans
Draft shields
Dynamic weighing pan MS-DWP-21 with 4 liter bowl (for bal
30006471
ances with readability of 0.01g and 0.1g )
Draft shield with sliding doors "mg" (usable height 168mm)12122405
Draft shield with sliding doors "0.1 mg" (usable height
12122404
236mm)
Draft shield MS-DS-21 for models with readability of 0.1g up
12121014
to 0.01g.
90Technical Data
Printers
RS-P25 printer with RS232C connection to instrument11124300
Paper roll, set of 5pcs00072456
Paper roll, self-adhesive, set of 3pcs11600388
Ribbon cartridge, black, set of 2pcs00065975
RS-P26 printer with RS232C connection to instrument (with
date and time)
Paper roll, set of 5pcs00072456
Paper roll, self-adhesive, set of 3pcs11600388
Ribbon cartridge, black, set of 2pcs00065975
RS-P28 printer with RS232C connection to instrument (with
date, time and applications
Paper roll, set of 5pcs00072456
Paper roll, self-adhesive, set of 3pcs11600388
Ribbon cartridge, black, set of 2pcs00065975
P-56RUE thermal printer with RS232C, USB and ethernet con
nections, simple printouts, date and time, label printing (limit
ed).
Paper roll, white, set of 10pcs30094723
Paper roll, white, self-adhesive, set of 10pcs30094724
Paper roll, white, self-adhesive labels, set of 6pcs30094725
P-58RUE Thermal printer with RS232C, USB and ethernet con
nections, simple printouts, date and time, label printing, bal
ance applications: statistics, formulation, totaling,
Paper roll, white, set of 10pcs30094723
Paper roll, white, self-adhesive, set of 10pcs30094724
Paper roll, white, self-adhesive labels, set of 6pcs30094725
RS232 - USB converter cable – Cable with converter to connect
a balance (RS232) to a USB port
64088427
91Accessories and Spare Parts
Cables for USB interface
METTLER TOLEDO
000000
USB (A –B) connection cable for connection to PC,
length=1m
Cable replacement (wireless)
Auxiliary displays
RS232 auxiliary display AD-RS-M712122381
Power supplies
AC/DC universal adapter (EU, USA, AU, UK) 100–240VAC,
50/60 Hz, 0.3 A, 12 VDC 0.84 A
30241476
11120270
AC/DC adapter (without power cable) 100–240VAC, 0.8A,
50/60Hz, 12VDC2.5A
11107909
92Accessories and Spare Parts
Protective covers
Protective cover30216667
Anti-theft devices
Steel cable11600361
Software
LabX direct balance (simple data transfer)11120340
Weighing below the balance
Transport cases
Transport case11124245
93Accessories and Spare Parts
Adjustment weights
OIML / ASTM Weights (with calibration certificate) see www.mt.com/weights
Density kit MS-DNY-43 for MS-TS balances
(d=0.1mg/1mg)
Bluetooth RS232 serial adapter ADP-BT-S for wireless connec
tion between printer and balance* or between balance and
PC*.
Fits printers P-56 / P-58.
The balance and PC* connection is supported by the following
balance models (SW V2.20 or higher required): MS, MS-S,
MS-L, ML, PHS, JP, JS.
* Bluetooth interface required
●
1 Bluetooth RS232 serial adapter (slave)
●
1 MT-DB9 male to female connector
●
1 MT-DB9 male to male connector
Bluetooth RS232 serial adapter ADP-BT-S for wireless connec
tion between printer and balance* or between balance and
PC*.
Fits printers P-56 / P-58.
The balance and PC* connection is supported by the following
balance models (SW V2.20 or higher required): MS, MS-S,
MS-L, ML, PHS, JP, JS.
* Bluetooth interface required
●
2 Bluetooth RS232 serial adapters paired (slave/master)
●
1 MT-DB9 male to female connector
●
1 MT-DB9 male to male connector
11142143
30086494
30086495
94Accessories and Spare Parts
14.2Spare parts
7
4
2
1
3
5
6
4
1
2
3
5
6
7
9
10
8
M
S
MS
Draft shield
DrawingPosDescriptionPart No.
Weighing pans / draft shield elements / support
DrawingPosDescriptionPart no.
5Draft shield lock12122013
6Bottom plate12122019
Draft shield "168 mm"
1Top glass with handle12121884
2Rear glass low12122015
3Side glass door left low with handle12121881
4Side glass door right low with handle12121883
7Front glass low12122014
Draft shield "236 mm"
1Top glass with handle12121884
2Rear glass high12122012
3Side glass door left high with handle12121880
4Side glass door right high with handle12121882
7Front glass high12122011
10.1 mgWeighing pan Ø 90 mm12122010
20.1 mgPan support Ø90mm11124249
30.1 mgDraft ring element12122008
41 mgWeighing pan 127x127mm12122009
51 mgPan support 127x127mm for
12122017
models up to 999g
51 mgPan support 127x127mm for
12122016
models from 1000g
60.01 gDraft ring element
12122018
170x200mm
70.01 gWeighing pan 170x200mm11124247
80.01 gPan support 170x200mm12121064
70.1 gWeighing pan 190x226mm11124248
80.1 gPan support 190x226mm12121066
9from
Pan support caps11131029
0.01g
10Leveling foot11106323
95Accessories and Spare Parts
Index
A
AC adapter77
Access protection64
Accessories90
Action buttons12
Activities12
Weighing37
Applications36
Adjusting23
Adjustments53
Adjustments and Tests53
air bubble20
Ambient conditions18
Application home screen12
Assign to sample34
Autoincrement34
Autopublish61
General8
Intended use8
Signal words8
Staff safety8
Warning Symbols8
Scope of delivery16
Selecting the location18
Send weight value61
Setting FACT54
Setting/Preferences12
Short balance information12
Signal words8
Simple weighing27
Sleep Mode63
Software update68
Sound58
Spare parts95
Staff safety8
Statistics35
Statistics configuration35
Status information field12
Switching
On23
Switching off24
Switching on24
Switching on the balance for the
first time
Switching the weighing unit28
Symbols7
System and Data Management63
System settings63
T
Taring32
Technical Data77
Technical data dimensions86
Technical data general77
Threshold49
Time57
Time format57
Totaling51
Touch beep58
Touch Screen Adjustment63
Transporting the balance22
U
USB Device62
USB device interface66, 75
USB driver installation66
USB Host75
User Interface11
V
Value bar12
W
Warm-up time23, 77
Warning Symbols8
Weighing37
Weighing below the balance22
Weighing information bar12
Weighing options59
Weighing unit12
Weighing value field12
Weighing-in aid12
Work title bar12
Workflow Feedback Beep58
Z
Zero setting32
Zeroing32
24
Index97
GWP®
is the global weighing standard, ensuring consistent accuracy
of weighing processes, applicable to all equipment from any manufacturer. It helps to:
• Choose the appropriate balance or scale
•
Calibrate and operate your weighing equipment with security
• Comply with quality and compliance standards in laboratory and