Mettler Toledo MS, MS105, MS105DU, MS205DU, MS204 Reference Manual

C
Reference Manual
Semi-Micro Balances
MS
Table of Contents
1 Introduction 3
2 Safety Information 4
2.2 Product specific safety notes ......................................................................................... 4
3 Design and Function 6
3.1 Overview components .................................................................................................. 6
3.2 Overview type plate...................................................................................................... 7
3.3 Overview operation keys............................................................................................... 7
3.4 Display....................................................................................................................... 9
4 Installation and Putting into Operation 10
4.1 Selecting the location ................................................................................................... 10
4.2 Scope of delivery ......................................................................................................... 10
4.3 Unpacking .................................................................................................................. 10
4.4 Installing components .................................................................................................. 11
4.5 Connecting the balance ................................................................................................ 11
4.6 Setting up the balance.................................................................................................. 12
4.7 Adjustment (calibration) ............................................................................................... 13
4.8 Performing a simple weighing....................................................................................... 17
4.9 Transporting the balance .............................................................................................. 19
4.10 Weighing below the balance ......................................................................................... 20
4.6.1 Switching on the balance ............................................................................... 12
4.6.2 Leveling the balance...................................................................................... 12
4.6.3 Left/right operating of the glass draft shield ...................................................... 13
4.6.4 Adjusting the balance .................................................................................... 13
4.7.1 Fully automatic adjustment FACT .................................................................... 14
4.7.2 Manual adjustment with internal weight ........................................................... 14
4.7.3 Manual adjustment with external weight .......................................................... 14
4.7.4 Customer fine adjustment............................................................................... 15
4.9.1 Transporting over short distances.................................................................... 19
4.9.2 Transporting over long distances .................................................................... 20
4.9.3 Packaging and storage.................................................................................. 20
5 The Menu 22
5.1 What is in the menu ? .................................................................................................. 22
5.2 Menu operation ........................................................................................................... 23
5.3 Description of menu topic ............................................................................................. 24
5.3.1 Main menu................................................................................................... 24
5.3.2 Basic menu.................................................................................................. 25
5.3.3 Advanced menu............................................................................................ 26
5.3.4 Interface menu.............................................................................................. 31
6 Applications 39
6.1 Application piece counting ............................................................................................ 39
6.2 Application percent weighing......................................................................................... 42
6.3 Application statistics..................................................................................................... 44
6.4 Application formulation (Net total) ................................................................................. 46
6.5 Application totaling ...................................................................................................... 49
6.6 Application multiplication factor weighing ....................................................................... 51
6.7 Application division factor weighing ............................................................................... 53
6.8 Application density....................................................................................................... 55
6.8.1 Density determination of solids ....................................................................... 56
6.8.2 Density determination of liquids ...................................................................... 57
6.8.3 Formulae used to calculate density ................................................................. 58
Table of Contents 1Semi-Micro Balances
6.9 Application pipette check .............................................................................................. 60
6.10 Application routine test ................................................................................................. 64
6.11 Application diagnostics................................................................................................. 66
6.11.1 Repeatability test........................................................................................... 66
6.11.2 Display test .................................................................................................. 66
6.11.3 Key test........................................................................................................ 67
6.11.4 Motor test..................................................................................................... 67
6.11.5 Balance history............................................................................................. 67
6.11.6 Calibration history ......................................................................................... 68
6.11.7 Balance information ...................................................................................... 68
6.11.8 Service provider information ........................................................................... 69
7 Communication with Peripheral Devices 70
7.1 USB - interface and installation...................................................................................... 70
7.2 Send weight value via USB or RS232 to a PC using PC-Direct........................................... 71
7.2.1 PC-Direct via USB ......................................................................................... 71
7.2.2 PC-Direct via RS232 ..................................................................................... 72
7.2.2.1 Installing SerialPortToKeyboard software .................................................... 72
7.2.2.2 Settings on the balance............................................................................ 72
7.3 Send weight value via USB to PC using LabX Direct Balance............................................. 73
8 Maintenance 74
8.1 Maintenance table........................................................................................................ 74
8.2 Performing routine tests................................................................................................ 74
8.3 Cleaning..................................................................................................................... 74
8.3.1 Cleaning agents............................................................................................ 74
8.3.2 Cleaning the glass draft shield........................................................................ 76
8.3.3 Cleaning the balance..................................................................................... 77
8.3.4 Putting into operation after cleaning ................................................................ 77
9 Troubleshooting 78
9.1 Error messages ........................................................................................................... 78
9.2 Error symptoms........................................................................................................... 80
9.3 Status messages/Status icons ....................................................................................... 82
9.4 Putting into operation after correction of fault................................................................... 82
10 Technical Data 83
10.1 General data ............................................................................................................... 83
10.2 Model-specific data...................................................................................................... 84
10.3 Dimensions................................................................................................................. 86
10.4 Interface specification ................................................................................................... 87
10.4.1 RS232C interface .......................................................................................... 87
10.4.2 USB device................................................................................................... 88
10.4.3 Aux connection ............................................................................................. 88
10.4.4 MT-SICS interface commands and functions ..................................................... 88
11 Disposal 89
12 Accessories and Spare Parts 90
12.1 Accessories................................................................................................................. 90
12.2 Spare parts ................................................................................................................. 95
Index 97
Table of Contents2 Semi-Micro Balances
1 Introduction
Thank you for choosing a METTLER TOLEDO balance. The balance combines high performance with ease of use.
This document refers to the initially installed software version V 2.20.
EULA
The software in this product is licensed under the METTLER TOLEDO End User License Agreement (EULA) for Software.
When using this product you agree to the terms of the EULA.
Finding more information
www.mt.com/semimicro-analytical
u
Search for software downloads u www.mt.com/labweighing-software-download
Search for documents u www.mt.com/library
For further questions, please contact your authorized METTLER TOLEDO dealer or service representative.
Conventions and symbols
Key and/or button designations and display texts are shown in graphic or bold text, e.g.,
Refers to an external document.
www.mt.com/EULA
u
u
www.mt.com/contact
, DATE.
Note
For useful information about the product.
This symbol indicates press key briefly (less than 1.5 s).
This symbol indicates press and hold key down (longer than 1.5 s).
This symbol indicates a flashing display.
Elements of instructions
Prerequisites
§
1 Steps 2 ...
ð Intermediate results
ð Results
Introduction 3Semi-Micro Balances
2 Safety Information
Two documents named "User Manual" and "Reference Manual" are available for this instrument.
The User Manual is printed and delivered with the instrument.
The electronic Reference Manual contains a full description of the instrument and its use.
Keep both documents for future reference.
Include both documents if you transfer the instrument to other parties.
Only use the instrument according to the User Manual and the Reference Manual. If you do not use the instrument according to these documents or if the instrument is modified, the safety of the instrument may be impaired and Mettler-Toledo GmbH assumes no liability.
2.1 Definitions of signal warnings and warning symbols
Safety notes contain important information on safety issues. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results. Safety notes are marked with the following signal words and warning symbols:
Signal words
WARNING
A hazardous situation with medium risk, possibly resulting in death or severe injury if not avoided.
CAUTION
NOTICE
A hazardous situation with low risk, resulting in minor or moderate injury if not avoided.
A hazardous situation with low risk, resulting in damage to the instrument, other material damage, malfunctions and erroneous results, or loss of data.
Warning symbols
Electrical shock
General hazard: read the User Manual or the Reference Manual for information about the hazards and the resulting measures.
Personal protective equipment
Chemical resistant safety gloves are intended to protect hands against aggressive chemicals.
The protective goggles protect the eyes from flying parts and liquid splashes.
2.2 Product specific safety notes
Intended use
This instrument is designed to be used in laboratories by trained staff. The instrument is intended for weighing purposes.
Any other type of use and operation beyond the limits of technical specifications without written consent from Mettler-Toledo GmbH is considered as not intended.
Responsibilities of the instrument owner
The instrument owner is the person holding the legal title to the instrument and who uses the instrument or authorizes any person to use it, or the person who is deemed by law to be the operator of the instrument. The instrument owner is responsible for the safety of all users of the instrument and third parties.
Safety Information4 Semi-Micro Balances
METTLER TOLEDO assumes that the instrument owner trains users to safely use the instrument in their workplace and deal with potential hazards. METTLER TOLEDO assumes that the instrument owner provides the necessary protective gear.
Trained personnel
Persons performing weighing processes must fulfill the following basic knowledge requirements regarding the handling of METTLER TOLEDO instruments and associated software:
Are able to complete the tasks entrusted to them and independently detect and avoid any possible dangers.
Have expertise and experience as well as their familiarity with all applicable regulations.
Able to prove that they have undergone training.
Safety notes
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to death or injury. 1 Only use the approved METTLER TOLEDO power supply cable and AC/DC adapter with a
current-limited SELV output. 2 Connect the power cable to a grounded power outlet, ensure correct polarity. 3 Keep all electrical cables and connections away from liquids and moisture. 4 Check the cables and power plug for damage and replace damaged cables and power
plugs.
NOTICE
Risk of damage to the accessory or the main instrument due to the use of unsuitable parts!
Using unsuitable parts with the accessory or the main instrument can damage the accessory or the main instrument or cause them to malfunction.
− Only use parts from METTLER TOLEDO that are intended to be used with your accessory.
A listing of all parts can be found in the Reference Manual (RM).
Safety Information 5Semi-Micro Balances
3 Design and Function
C
1
2
3
4
5
6
7
8
9
10
11
12
7
13 14
15
16
17
18
3.1 Overview components
Components overview legend
Display
1
Model sticker (with approved models only)
3
Draft ring
5
Handle/Coupling element for the operation of
7
the draft shield doors
Handle for operation of the draft shield top
9
door
Drip tray
11
Socket for AC/DC Adapter
13
USB device interface
15
Kensington slot for anti-theft purposes
17
Operation keys
2
Level indicator
4
Weighing pan
6
Glass draft shield
8
Top cover
10
Leveling feet
12
Aux (connection for "ErgoSens" or foot-switch)
14
RS232C serial interface
16
Product label
18
Design and Function6 Semi-Micro Balances
3.2 Overview type plate
1 52 3 4
67
Weigh-Track
C
1 2 4 53 6
7
8
9
10
The balance type plate is located at the side of the balance, see example illustration and contains the following information:
Model designation
1
Readability
3
Manufacturer
5
Power supply
7
3.3 Overview operation keys
Year of manufacture
2
Maximum capacity
4
Serial number (SNR)
6
Terminal keys legend
No. Key Press briefly (less than 1.5 s) Press and hold (longer than 1.5 s)
1
To navigate back (scroll up) within menu
topics or menu selections
Decrease (numerical) parameters within
menu and in applications
To select the weighing application
Decrease (numerical) parameters quickly
within menu and in applications
Design and Function 7Semi-Micro Balances
No. Key Press briefly (less than 1.5 s) Press and hold (longer than 1.5 s)
On/Off
2
To navigate forward (scroll down) within
menu topics or menu selections
Increase (numerical) parameters within
menu and in applications
To select assigned F1 application and
entering the parameter settings of appli­cation.
Default F1 application assignment: Piece
counting
Increase (numerical) parameters quickly
within menu and in applications
3
4
Switch on
Zero/Tare
With entries: scroll down
To navigate through menu topics or menu
selections
To toggle between unit 1, recall value (if
selected), unit 2 (if different from unit 1)
Switch off
To select assigned F2 application and
entering the parameter settings of appli­cation.
Default F2 application assignment:
Percent weighing
and the application unit (if any)
5
To change display resolution (1/10d
no function
function) while application is running
Note: not available with approved
models in selected countries.
6
To enter or leave menu selection (from /
to menu topic)
To enter application parameter or switch
to next parameter
To confirm parameter
7
8
Printout display value
Printout active user menu settings
Transfer data
Execute predefined adjusting (calibration)
To select assigned F3 application and
entering the parameter settings of appli­cation.
Default F3 application assignment:
Statistics
no function
no function
procedure
9
10
Enter or leave menu (Parameter settings)
Save parameters
Cancel and to leave menu without saving
no function
no function
(one step back in the menu).
Design and Function8 Semi-Micro Balances
3.4 Display
Application icons
Menu locked Application formulation / Net total
Menu setting activated Application totaling
Application weighing Application multiplication factor
Application piece counting Application division factor
Application percent weighing Application density
Application statistics Application pipette check
While an application is running, the corresponding application icon appears at the top of the display.
Status icons
Indicates stored value (Memory) Applications diagnostics and routine test
Indicates net weight values Acoustic feedback for pressed keys activated
Adjustments (calibration) started Weighing range 1 (Dual Range models only)
FACT activated Weighing range 2 (Dual Range models only)
Service reminder Not used
Weight value field and weighing-in aid
Indicates negative values Brackets to indicate uncertified digits
(approved models only)
Indicates unstable values Marking of nominal or target weight
Indicates calculated values Not used
Not used
Unit field
g
kg
mg
ct
lb
oz
gram
kilogram
milligram
carat
pound
ounce
ozt
GN
dwt
mom
msg
tlh
troy ounce
grain
pennyweight
momme
mesghal
Hong Kong taels
tls
tlt
tola
baht
Singapore taels
Taiwan taels
tola
baht
Design and Function 9Semi-Micro Balances
4 Installation and Putting into Operation
4.1 Selecting the location
A balance is a sensitive precision instrument. The location where it is placed will have a profound effect on the accuracy of the weighing results.
Requirements of the location
According to the environmental condition, see chapter "General data".
Ensure there is:
indoor on
stable table
Avoid:
direct sunlight vibrations strong drafts temperature
with sufficient
distance (> 15 cm)
in level adequate lit
fluctuations
4.2 Scope of delivery
Balance with draft shield
Weighing pan with pan support
Draft ring
Drip tray
Protective cover
AC/DC adapter
Power cable (country specific)
1 User Manual
Declaration of conformity
4.3 Unpacking
Open the balance packaging. Check the balance for transport damage. Immediately inform a METTLER TOLEDO representative in the event of complaints or missing accessories.
Retain all parts of the packaging. This packaging offers the best possible protection for transporting the balance.
Installation and Putting into Operation10 Semi-Micro Balances
4.4 Installing components
C
1
2
3
CAUTION
Injury due to breaking glass
Careless handling with the glass components can lead to breakage off glass and damage cuttings.
− Always proceed focused and with care.
Push the side glass doors back as far as will go and place the following components on the balance in the specified order:
1 Place the drip tray (1) into the correct position. 2 Place the weighing pan (3). 3 Place the draft ring (2).
4.5 Connecting the balance
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to death or injury. 1 Only use the approved METTLER TOLEDO power supply cable and AC/DC adapter with a
current-limited SELV output. 2 Connect the power cable to a grounded power outlet, ensure correct polarity. 3 Keep all electrical cables and connections away from liquids and moisture. 4 Check the cables and power plug for damage and replace damaged cables and power
plugs.
NOTICE
Damage to the AC/DC adapter due to overheating
If the AC/DC adapter is covered or in a container, it is not sufficiently cooled and will overheat.
1 Do not cover the AC/DC adapter. 2 Do not put the AC/DC adapter in a container.
The balance is supplied with an AC/DC adapter and a country-specific power cable.
Install the cables so that they cannot be damaged or interfere with operation.
Insert the power cable in a grounded power outlet that is easily accessible.
Installation and Putting into Operation 11Semi-Micro Balances
1 Connect the AC/DC adapter to the connection socket on
On
Off
0.00
g
the back of your balance (see figure) and to the power
line. 2 Screw the plug tight to the balance. ð The balance is ready for use.
4.6 Setting up the balance
4.6.1 Switching on the balance
Before using the balance, it must be warmed up in order to obtain accurate weighing results. To reach operating temperature, the balance must be acclimatized and connected to the power supply for at least 60 minutes.
Switching on
− Press On.
ð The balance performs a display test. All segments
in the display light up briefly, WELCOME and software version. Maximum load and readability appears briefly. (Startup FULL mode only).
ð The balance is ready for weighing or for operation with
the last active application.
Legal-for-trade
Approved balances will execute an initial zero.
4.6.2 Leveling the balance
The balance has a level indicator and two adjustable leveling feet to compensate for slight irregularities in the surface of the weighing bench. The balance is exactly horizontal when the air bubble is in the middle of the level glass.
The balance must be leveled and adjusted each time it is moved to a new location. To level it, proceed as follows: 1 Position your balance at the selected location. 2 Align the balance horizontally.
Installation and Putting into Operation12 Semi-Micro Balances
C
È
3 Turning the two front leveling screws of the housing
until the air bubble is in the inner circle of the level indicator.
Example
Air bubble at 12 o'clock: turn both feet clockwise.
Air bubble at 3 o'clock: turn left foot clockwise, right foot counterclockwise.
Air bubble at 6 o'clock: turn both feet counterclockwise.
Air bubble at 9 o'clock: turn left foot counterclockwise, right foot clockwise.
4.6.3 Left/right operating of the glass draft shield
The glass draft shield of your balance can be adapted to the environ-
mental conditions and your weighing style, as well as to the type of
weighing and loading.
The position of the handles determines which door(s) of the draft
shield (left, right, or both) is/are opened.
Try various different combinations by moving the external handles into
the upper or lower position. We recommend you to set up the glass
draft shield so that it only opens on the side where the balance is
loaded. Your balance then works faster, because there are fewer
troublesome currents of air than when both doors of the draft shield are
opened together.
4.6.4 Adjusting the balance
To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its
location. This is also dependent on the ambient conditions. After reaching the operating temperature, it is
important to adjust the balance in the following cases:
Before the balance is used for the first time.
If the balance has been disconnected from the power supply or in the event of power failure.
After significant environmental changes (e.g. temperature, humidity, air draft or vibrations).
At regular intervals during weighing service.
4.7 Adjustment (calibration)
NOTICE
Before adjusting the balance, it must be warmed up.
Installation and Putting into Operation 13Semi-Micro Balances
4.7.1 Fully automatic adjustment FACT
------
ADJ DONE
0.00
g
ADJ.INT
------
2000.00
g
ADJ DONE
0.00
g
C
0.00
g
------
C
ADJ.EXT
The factory setting is fully automatic adjustment FACT (Fully Automatic Calibration Technology) with the internal weight.
More information can be found in the section [The Menu The balance adjusts itself automatically:
after the warm-up phase on connection to the power supply.
when a change in the ambient conditions, e.g. the temperature, could lead to a noticeable deviation in the
measurement.
on a predefined time, see menu topic FACT.
time interval. (with OIML accuracy class II approved models).
4.7.2 Manual adjustment with internal weight
In the menu topic CAL (Adjustment) of advanced menu ADJ.INT must be selected.
§
Weighing pan is unloaded.
§
− Press to execute the internal adjustment.
ð The balance adjusts itself automatically. ð The adjusting is finished when the message ADJ DONE
appears briefly on the display. The balance returns to
the last active application and is ready for operation.
}Page22].
4.7.3 Manual adjustment with external weight
Note
We recommend to disable FACT.
In the menu topic CAL (Adjustment) of advanced menu ADJ.EXT must be selected.
§
Required adjustment weight is ready.
§
Weighing pan is unloaded.
§
1 Press briefly to execute the external adjustment.
ð The required (predefined) adjustment weight value
flashes on the display.
2 Place adjustment weight in center of pan.
ð The balance adjusts itself automatically.
3 Remove adjustment weight, when 0.00 g flashes.
ð The adjusting is finished when the message ADJ DONE
appears briefly on the display. The balance returns to
the last active application and is ready for operation.
Installation and Putting into Operation14 Semi-Micro Balances
4.7.4 Customer fine adjustment
NOTICE
This function should be executed only by trained personnel.
The function customer fine adjustment ADJ.CUST.F allows you to adjust the value of the internal adjustment
weight with your own adjustment weight. The adjustable range of the adjustment weight is possible only in a
very small range. Customer fine adjustment impacts the function of internal adjustment. The customer fine
adjustment can be deactivated at any time.
Note
This feature is available on models with internal weight only.
Because of certification legislation, approved models cannot be adjusted with customer fine adjustment
(depending on selected countries' certification legislation).
Use certificated weights.
Balance and test weight have to be on operating temperature.
Observe the correct environmental conditions.
Installation and Putting into Operation 15Semi-Micro Balances
Execute customer fine adjustment
200.0000
g
ADJ.DONE
0.00
g
------
------
g
g
200.0005
200.0005
C
C
ADJ.CUST.F
EXECUTE
SET REF.
ADVANCED
------
0.00
g
-
+
The balance is under measuring condition.
§
Required adjustment weight is ready.
§
Weighing pan is unloaded.
§
1 Select in the menu ADVANCED: ADJ.CUST.F
2 Confirm ADJ.CUST.F with
3 To carry out this operation select EXECUTE.
4 Start Adjustment with
ð SET REF. appears briefly.
ð The last saved value flashes on the display.
5 Select the target adjustment weight.
• For coarse setting, press
• For fine setting, press + to increase the value or
press - to decrease the value.
6 Press and hold
ADJ.CUST.F.
ð The required adjustment weight value flashes in the
7 Place required adjustment weight in center of pan. 8 Remove adjustment weight when zero is flashing.
9 Wait until ADJ DONE briefly appears.
ð The adjusting is finished when the message ADJ DONE
appears briefly on the display. The balance returns to
the last active application and is ready for operation ð If the error message WRONG ADJUSTMENT WEIGHT
appears, the weight is not within the allowed value
range and could not be accepted. ADJ.CUST.F could
not be executed.
.
.
to change the value.
to confirm and execute
display. This could take some time.
Note
Storing the adjustment is not required.
Deactivate customer fine adjustment
1 Select in the menu ADVANCE.: ADJ.CUST.F.
2 Confirm ADJ.CUST.F with
3 To carry out this operation select RESET.
4 Start RESET by pressing
ð NO? appears.
5 Select YES? and confirm with
ð The adjusting is finished when the message ADJ DONE appears briefly on the display. The balance returns
to the last active application and is ready for operation with initial adjustment.
.
.
.
Installation and Putting into Operation16 Semi-Micro Balances
4.8 Performing a simple weighing
1250.38
0.00
1182.03
C
¢¢¢¢¢¢¢¢
¢¢¢¢¢¢
#
0.00
121.01
g
¢¢¢¢
#
0.01
g
0.00
g
C
C
121.01
g
¢¢¢¢
0.00
g
0.00
g
¢¢¢¢
95.97
g
¢¢¢¢¢¢¢
121.01
g
¢¢¢¢
-
N
et
Net
Net
#
C
C
The weighing application allows you to perform simple weighings and how you can accelerate the weighing process.
If your balance is not in the weighing mode, press and hold the
key down until WEIGHING appears in the
display. Release the key. Your balance is in the weighing mode and set to zero.
1 Press to tare the balance.
2 Place the sample on the weighing pan.
3 Wait until the instability detector
disappears and the
stability beep sounds.
4 Read the result.
Zeroing
Use the
zeroing key before you start with a weighing. 1 Unload the balance.
2 Press
to zero the balance.
ð All weight values are measured in relation to this
zero point.
Taring
If you are working with a weighing container, first set the balance to zero. 1 Place empty container on the weighing pan.
2 Press
3 Place weighing sample into the weighing container. ð The result appears in the display.
If the container is removed from the balance, the tare weight will be shown as a negative value.
ð The weight is displayed.
to set the balance to zero.
ð 0.00 g and Net appears in the display. Net
indicates that all weight values displayed are net values.
Note
Installation and Putting into Operation 17Semi-Micro Balances
The tare weight remains stored until the
0.78
oz
22.00
g
¢¢¢¢
*
22.00
g
¢¢¢¢
M
121.01
g
¢¢¢¢
0.00
g
0.00
g
M
121.01
g
0.00
g
C
C
¢¢¢¢¢¢¢¢
0%
100%
With METTLER TOLEDO DeltaRange balances, the fine range with its 10 times smaller display increments
key is pressed again or the balance is switched off.
(depending on the model) is available again after every taring operation.
METTLER TOLEDO DualRange balances
METTLER TOLEDO DualRange balances have two ranges. These models have a fix fine (semi-micro) range between 0 g and maximum capacity, fine range. In this fine range the balance shows the result with a higher resolution, i.e. with one decimal place more.
Switching weight units
key can be used at any time to toggle between weight unit UNIT 1, RECALL value (if selected) and
The weight unit UNIT 2 (if different from weight unit 1) and the application unit (if any).
− Press to set weight unit or recall value.
Recall / recall weight value
Recall stores stable weights with an absolute display value bigger than 10d.
Function RECALL is in the menu activated.
§
1 Load weighing sample.
ð The display shows weight value and stores stable
value.
2 Remove weighing sample.
ð The display shows zero.
3 Press
.
ð The display shows last stored stable weight value
for 5 seconds together with asterisk (*) and memory (M) symbols. After 5 seconds the display goes back to zero. This can be repeated unlimited times.
Delete last weight value
As soon a new stable weight value is displayed, the old recall value becomes replaced by the new weight value.
− Press
.
ð The recall value is set to 0.
If the power is switched off, the recall value is lost. The recall value can not be printed.
Weighing with the weighing-in aid
The weighing-in aid is a dynamic graphic indicator which shows the used amount of the total weighing range. You can thus recognize at a glance when the load on the balance approaches the maximum load.
Installation and Putting into Operation18 Semi-Micro Balances
Print / transmit data
SHUTOFF
On
Off
Press the
key to transmit the weighing results over the interface e.g. to a printer or a PC.
Switching off
− Press and hold the Off key until SHUTOFF appears on
the display. Release the key.
ð Balances switch into standby mode.
Note
After switching on from standby mode, your balance needs no warm-up time and is immediately ready for
weighing.
Standby mode is not possible with approved balances (only available in selected countries).
If your balance has been switched off after a preselected time, the display is dimly lit and shows date, time,
maximum load and readability.
If your balance has been switched off manually, the display is off.
To completely switch off mains operated balances, they must be disconnected from the power supply.
4.9 Transporting the balance
CAUTION
Injury due to breaking glass
Careless handling with the glass components can lead to breakage off glass and damage cuttings.
1 Do not lift the instrument by the glass draft shield. 2 Always proceed focused and with care.
1 Press and hold the Off key.
2 Disconnect the balance from the power supply. 3 Disconnect all interface cables.
4.9.1 Transporting over short distances
To move the balance over a short distance to a new location, follow the instructions below.
Installation and Putting into Operation 19Semi-Micro Balances
C
1 Hold the balance with both hands as shown. 2 Carefully lift the balance and carry it to its new
location.
If you want the balance put into operation, proceed as follows:
1 Connect in reverse order. 2 Level the balance. 3 Perform an internal adjustment.
4.9.2 Transporting over long distances
For transporting the balance over long distances, always using the original packaging.
4.9.3 Packaging and storage
Packaging
Store all parts of packaging in a save place. The elements of the original packaging are developed specifically for the balance and its components to ensure maximum protection during transportation or storing.
Storage
Store the balance under following conditions:
Indoor and in the original packaging.
According to the environmental condition, see chapter "General data".
When storing for longer than six months, the rechargeable battery maybe down (date and time get lost).
4.10 Weighing below the balance
Your balance is equipped with a hanger for performing weighing operations below the work surface (weighing below the balance).
1 Press and hold the Off key.
2 Disconnect the balance from the power supply. 3 Disconnect all interface cables.
4 Push the side doors and the top door of the glass draft shield completely to the back. Important: top cover
must be closed.
5 Remove weighing pan, draft ring and drip tray. 6 Carefully tip the balance over backwards, until it is lying on its back.
Installation and Putting into Operation20 Semi-Micro Balances
1
2
7 Remove the cap (1) and retain it. The hanger (2) for
weighing below the balance is easily accessible now.
8 Carefully turn the balance to its normal position and
reinstall all components in the reverse order.
Installation and Putting into Operation 21Semi-Micro Balances
5 The Menu
5.1 What is in the menu ?
The menu allows you to match your balance to your specific weighing needs. In the menu you can change the settings of your balance and activate functions. The main menu has 4 different menus and these contains different topics, each of which allows you various selection possibilities. Menu item PROTECT see chapter Description of menu topics > Main
menu.
Menu BASIC
Topic Description
DATE
TIME
UNIT 1
UNIT 2
KEY BEEP
STAB.BEEP
RESET
Menu ADVANCED
Topic Description
WEIG.MODE
ENVIRON.
CAL
ADJ.CUST.F
FACT
FACT PRT.
DATE.FORM
TIME.FORM
RECALL
SHUTOFF
BCKLIGHT
DISPLAY
AUTOZERO
ZERO RNG
LANGUAGE
ASSIGN:F1
ASSIGN:F2
ASSIGN:F3
DIAGNOSE
SERV.ICON
SRV.D.RST
Setting the current date.
Setting the current time.
Specification of the 1st weight unit in which the balance should show the result.
Specification of the 2nd weight unit in which the balance should show the result.
Setting the key beep level.
Setting the stability beep level.
Call up of the factory settings.
Adapting the balance to the weighing mode.
Matching the balance to the ambient conditions.
Settings for the type of adjustment (calibration).
Executing customer fine adjustment.
Settings for fully automatic balance adjustment based on a selected time.
Switching the automatic FACT printout on or off.
Setting the date format.
Preselection of the time format.
Switching the application recall for storing stable weights on or off.
Setting the time after which the balance should be switched off automatically.
Setting the time after which the display backlight should be switched off automatically.
Adjusting the brightness and contrast of the display.
Switching the automatic zero correction (Autozero) on or off.
Setting the zero limit of the zero/tare key.
Setting the preferred language.
Selection of assigned F1 key application and entering their parameter settings.
Selection of assigned F2 key application and entering their parameter settings.
Selection of assigned F3 key application and entering their parameter settings.
Starting a diagnostic application.
Switching the service icon (service reminder) on or off.
Reset service date and hours (service reminder).
The Menu22 Semi-Micro Balances
Menu INT.FACE
406.2
g
¢¢¢¢
ADVANCED
BASIC
TIME
DATE
Topic Description
RS232
HEADER
SINGLE
SIGN.L
LINE.FEED
ZERO PRT.
COM.SET
BAUDRATE
BIT/PAR.
STOPBIT
HD.SHAKE
RS.TX.E.O.L.
RS CHAR
USB
USB COM.S.
USB E.O.L.
USB CHAR
INTERVAL
ERGOSENS
Matching the serial interface RS232C to a peripheral unit.
Setting the header for printout of individual values.
Setting the information for printout of individual values.
Setting the footer for printout of individual values.
Setting line feed for printout of individual values.
Setting the auto print function for printing zero.
Setting the data communication format of the serial interface RS232C.
Setting the transfer speed of the serial interface RS232C.
Setting the character format (Bit/Parity) of the serial interface RS232C.
Setting the character format (stop bit) of the serial interface RS232C.
Setting the transfer protocol (Handshake) of the serial interface RS232C.
Setting the end of line format of the serial interface RS232C (outgoing data).
Setting the char set of the serial interface RS232C.
Matching the USB interface to a peripheral unit.
Setting the data communication format of the USB interface.
Setting the end of line format of the USB interface.
Setting the char set of the USB interface.
Selection of the time interval for the simulated print key press.
Settings for external key e.g. METTLER TOLEDO ErgoSens
5.2 Menu operation
In this section you will learn how to work with the menu.
Select menu
1 Press to activate main menu.
ð The first menu BASIC is displayed (except menu
protection is active).
2 Press
up + / – keys).
3 Press
Note
The menu selection BASIC, ADVANCED or INT.FACE can not be saved. The selection PROTECT must be saved.
Select menu topic
1 Press .
ð The next menu topic appears in the display.
2 Press
menu topic.
3 Press – key to return to the previous menu topic.
repeatedly to change menu (Scrolling down/
to confirm the selection.
or the + key, the balance switches to the next
Change settings in a selected menu topic
The >> flashing symbol in the display indicates selectable options available.
The Menu 23Semi-Micro Balances
ENVIRON.
STABLE
ENVIRON.
1 Press .
+ 01.2500
+ 01.6500
+ 01.6500
SAVE:YES
SAVE:NO
ð The display shows the current setting in the
selected menu topic.
2 Press
selection.
3 Press – key to return to the previous selection.
ð After the last selection, the first is shown again.
4 Press
For store the setting, see section Saving settings and
closing the menu.
Change settings in a submenu selection
The same procedure as for menu topics.
or the + key, the balance switches to the next
to confirm the setting.
Input principle of numerical values
1 Press for input of numerical values. 2 Press
application). ð The selected digit or the selected value is blinking.
3 Press + to scroll up or – to scroll down for changing
digits or values.
4 Press
to select a digit or a value (depending on the
to confirm the input.
Saving settings and closing the menu
1 Press briefly to leave menu topic.
2 Press
ð Changes are saved.
3 Press
ð Changes are not saved.
4 Press
to execute SAVE:YES.
to execute SAVE:NO.
to toggle between SAVE:YES and SAVE:NO.
Cancel
− Press C for leaving menu topic or menu selection without saving.
Note
If no entry is made within 30 seconds, the balance reverts to last active application mode. Changes are not saved. If changes are made, the balance asks SAVE:NO.
5.3 Description of menu topic
In this section you will find information regarding the individual menu topics and the available selections.
5.3.1 Main menu
Selecting the menu.
BASIC The small BASIC menu for simple weighing is displayed.
ADVANCED The extended ADVANCED menu for further weighing settings is
displayed.
The Menu24 Semi-Micro Balances
Note
The menu selection BASIC, ADVANCED or INT.FACE can not be saved.
To activate PROTECT ON or OFF, this selection must be saved.
5.3.2 Basic menu
DATE – Date
Setting the current date according to date format.
Note
A reset of the balance will not change this setting.
TIME – Time
Setting the current time according to time format
INT.FACE The menu INT.FACE for all interface parameter settings for
peripheral devices e.g. printer is displayed.
PROTECT
Menu protection. protection of balance configurations against unmeant manipulation.
OFF Menu protection is off. (Factory setting)
ON Menu protection is on. The menu BASIC, ADVANCED and
INT.FACE are not displayed. This is indicated with
in the
display.
+1H
Set the current time forwards by 1 hour (to adjust summer or winter time). (Factory setting)
-1H
Set the current time backwards by 1 hour (to adjust summer or winter time).
SET TIME
Enter the current time.
Note
A reset of the balance will not change this setting.
UNIT 1 – Weight unit 1
Depending on requirements, the balance can operate with the following units (depending on the model)
Legal-for-trade
Only those weight units allowed by the appropriate national legislation are selectable.
With approved balances, this menu topic has a fixed setting and cannot be changed.
Units:
1)
g
2)
kg
3)
mg
ct
lb
oz
ozt
GN
1)
factory setting
2)
not with 0.01 mg, 0.1 mg and 1 mg balances
3)
with 0.01 mg, 0.1 mg and 1 mg balances
4)
the Malaysian tael has the same value
Gram
Kilogram
Milligram
Carat
Pound
Ounce (avdp)
Ounce (troy)
Grain
dwt
mom
msg
tlh
tls
tlt
tola
baht
Pennyweight
Momme
Mesghal
Tael Hong Kong
4)
Tael Singapore
Tael Taiwan
Tola
Baht
The Menu 25Semi-Micro Balances
UNIT 2 – Weight unit 2
If it is required to show the weighing results in weighing mode in an additional unit, the desired second weight unit can be selected in this menu topic (depending on the model). Units see UNIT 1. Select NO, if you do not want to use UNIT 2.
Legal-for-trade
Only those weight units allowed by the appropriate national legislation are selectable.
KEY BEEP – Key beep
This menu topic allows you to select the volume of the key beep. The according key beep is emitted during the setting.
MED Medium level (Factory setting)
HIGH
OFF
LOW
High level
Beep switched off
Low level
STAB.BEEP – Stability beep
If the unstable symbol disappears, the stability beep becomes active. This menu topic allows you to preselect the volume of the stability beep.
MED Medium level (Factory setting)
HIGH
OFF
LOW
High level
Beep switched off
Low level
RESET – Reset balance settings
This menu topic allows you to call-up the factory settings.
To toggle between YES? and NO? press
Note
A reset of the balance will not change the DATE, TIME and ZERO RNG settings.
5.3.3 Advanced menu
WEIG.MODE – Weighing mode settings
This setting can be used to to adapt the balance to the weighing mode.
ENVIRON. – Environment settings
This setting can be used to match your balance to the ambient conditions.
(or + or –).
UNIVERS. For all standard weighing applications. (Factory setting)
DOSING
STANDARD
UNSTABLE
STABLE
For dosing liquid or powdery products. With this setting, the balance responds very quickly to the smallest changes in weight.
Setting for an average working environment subject to moderate variations in the ambient conditions. (Factory setting)
Setting for a working environment where the conditions are continuously changing.
Setting for a working environment which is practically free from drafts and vibrations.
CAL – Adjustment (calibration)
In this menu topic you can preselect the function of the external weights by pressing the adjustment (calibration) are printed out.
ADJ.OFF The adjustment is switched off. The key has no function.
The Menu26 Semi-Micro Balances
key. If you have attached a printer to your balance, the data of the
key. Your balance can be adjusted with internal or
ADJ.INT Internal adjustment: adjustment is performed at a keystroke with
the built-in weight (depending on the model, see technical data).
ADJ.EXT External adjustment: adjustment is performed at a keystroke with
a selectable external weight.
100.00 g Defining the external adjustment weight: define the weight of the external adjustment weight (in grams).
Factory setting: depends on the model.
ADJ.CUST.F – Customer fine adjustment
At this menu topic you can fine-adjust the internal weights. Further information refer to chapter Customer fine adjustment.
EXECUTE Start customer fine adjustment ADJ.CUST.F.
RESET Deactivate customer fine adjustment after confirming with YES?.
NO?
YES?
No deactivation.
Confirm to deactivation.
FACT – Fully automatic adjustment
Fully automatic internal adjustment (calibration) FACT (Fully Automatic Calibration Technology) provides fully automatic balance adjustment based on temperature criteria and on preselected time.
TIME
12:00
OFF FACT is switched off.
Execute FACT (with selected time).
Specify the time for a fully automatic adjustment to take place every day.
Factory setting: 12:00 (according to time format)
FACT PRT. – Protocol trigger for FACT
This setting specifies whether an adjustment report should be printed automatically.
Note: this menu topic does not affect the printing of adjustments with an internal or external adjustment
weight.
OFF Protocol switched off: if the balance adjusts automatically
(FACT), a protocol is not printed out.
ON
Protocol switched on: a record is printed out after every automatic adjustment of the balance (FACT).
Note: the protocol is printed out without a line for signatures.
DATE.FORM – Date format
This menu topic allows you to preselect the date format. The following date formats are available:
Display examples Printing examples
DD.MM.Y
MM/DD/Y
Y-MM-DD
D.MMM Y
MMM D Y
Factory setting: DD.MM.Y
01.02.2009 01.02.2009
02/01/09 02/01/2009
09-02-01 2009-02-01
1.FEB.09 1.FEB 2009
FEB.1.09 FEB 1 2009
TIME.FORM – Time format
This menu topic allows you to preselect the time format. The following date formats are available:
Display examples
The Menu 27Semi-Micro Balances
24:MM
12:MM
24.MM
12.MM
Factory setting: 24:MM
15:04
3:04 PM
15.04
3.04 PM
RECALL – Recall
This menu topic allows you to switch the RECALL function on or off. When it is switched on recall stores the last stable weight if the absolute display value was bigger than 10d.
OFF
ON
Note: the recall value is displayed with an asterisk and cannot be printed.
RECALL switched off (Factory setting)
RECALL switched on
SHUTOFF – Automatic shutoff
If the automatic shutoff function is activated, the balance automatically switches itself off after a preselected time of inactivity (i.e. with no key being pressed or changes of weight occurring etc.) and is switched to the standby mode.
A.OFF 10 min Automatic shutoff after 10 minutes of inactivity. (Factory setting)
A.OFF – Automatic shutoff not activated.
A.OFF 2 min Automatic shutoff after 2 minutes of inactivity.
A.OFF 5 min Automatic shutoff after 5 minutes of inactivity.
BCKLIGHT – Backlight
Under this menu topic, the display backlight can be switched off automatically. If the automatic switch-off is activated, the backlight will turn off automatically after the selected period of inactivity has elapsed. The backlight is reactivated when a key is pressed or the weight is changed.
B.L. ON Backlight is always on. (Factory setting)
B.L. 30 s Automatic switch-off after 30 seconds inactivity.
B.L. 1 min Automatic switch-off after 1 minute inactivity.
B.L. 2 min Automatic switch-off after 2 minutes inactivity.
B.L. 5 min Automatic switch-off after 5 minutes inactivity.
DISPLAY – Display settings
This menu topic allows you to adjust brightness and contrast of the display.
BRIGHTN
50% Factory setting: 50%
CONTRAST
75% Factory setting: 75%
To set the brightness in 1% steps.
To set the contrast in 1% steps.
AUTOZERO – Automatic zero setting
This menu topic allows you to switch the automatic zero setting on or off.
ON AUTOZERO switched on (Factory setting). The automatic zero
setting continuously corrects possible variations in the zero point that might be caused through small amounts of contamination on the weighing pan.
OFF AUTOZERO switched off. The zero point is not automatically
corrected. This setting is advantageous for special applications (e.g. evaporation measurements).
The Menu28 Semi-Micro Balances
Legal-for-trade
With approved balances, this setting is not available in selected countries.
ZERO RNG – Zero range
This menu topic allows you to set a zero limit for the
key will execute a zero. Above this limit the key will execute a tare.
1.2 g
To set the upper limit of the zero setting range as weight in the definition unit of the balance. (Factory setting: 0.5% of weighing range)
Note: with approved balances, this setting is not available
and fixed to 3e (only available in selected countries).
key. Up to and including this limit the
Note
A reset of the balance will not change this setting.
LANGUAGE – Language
Factory setting: generally, the language of the destination country (if available) or English is set.
The following languages are available:
ENGLISH
DEUTSCH
FRANCAIS
ESPANOL
ITALIANO
RUSSIAN
РУССКИИ
English
German
French
Spanish
Italian
Russian
POLSKI
CESKY
MAGYAR
NEDERL.
BR.PORTUG.
Polish
Czech
Hungarian
Dutch
Brazil Portuguese
ASSIGN:F1 – Assign application key F1
At this menu topic you can assign an application to the F1 key. The following applications are available (depending on the model):
COUNTING Piece counting (Factory setting)
PERCENT
STAT
FORMULA
TOTALING
FACTOR M
FACTOR D
DENSITY
PIPETTE
Percent weighing
Statistics
Formulation / Net total
Totaling
Multiplication factor
Division factor
Density
Pipette check
ASSIGN:F2 – Assign application key F2
At this menu topic you can assign an application to the F2 key. The following applications are available (depending on the model):
PERCENT Percent weighing (Factory setting)
STAT
FORMULA
TOTALING
FACTOR M
FACTOR D
DENSITY
PIPETTE
Statistics
Formulation / Net total
Totaling
Multiplication factor
Division factor
Density
Pipette check
The Menu 29Semi-Micro Balances
COUNTING
Piece counting
ASSIGN:F3 – Assign application key F3
At this menu topic you can assign an application to the F3 key. The following applications are available (depending on the model):
STAT Statistics (Factory setting)
FORMULA
TOTALING
FACTOR M
FACTOR D
DENSITY
PIPETTE
R. TEST
COUNTING
PERCENT
Formulation / Net total
Totaling
Multiplication factor
Division factor
Density
Pipette check
Routine test
Piece counting
Percent weighing
DIAGNOSE – Diagnostics application
At this menu topic you can start a diagnostic application. For more information see chapter Application diagnostics.
The following diagnostics are available:
REPEAT.T
DISPLAY
KEYPAD T
CAL.MOT. T
BAL.HIST
CAL.HIST
BAL.INFO
PROVIDER
Repeatability test (models with internal weights only)
Display test
Key test
Motor test (models with internal weights only)
Balance history
Calibration history
Balance information
Service provider information
SERV.ICON – Service reminder
This menu topic allows you to switch the service reminder
ON Service reminder
on or off.
switched on (factory setting).You will be informed after a preset time (e.g. one year or 8000 operating hours) to call service for recalibration. This will be indicated by the flashing service icon:
OFF Service reminder
switched off.
. (Factory setting)
SRV.D.RST – Service date reset
This menu topic allows you to reset service date and hours.
Note
This menu topic is only available if SERV.ICON setting ON was selected.
To toggle between YES? and NO? press
RS CHAR – Char set RS232C
1)
(or + or -).
At this menu topic you can set the character set of the transmitted data to different RS232C serial receivers.
IBM/DOS Char set IBM/DOS (Factory setting)
ANSI/WIN
Char set ANSI/WINDOWS
The Menu30 Semi-Micro Balances
Note
Not visible for 2nd display.
Each device has separate settings.
5.3.4 Interface menu
RS232 – RS232C interface
At this menu topic you can select the peripheral device connected to the RS232C interface and specify how the data is transmitted.
1)
PRINTER Connection to a printer. (Factory setting)
Only one printer possible.
Refer to your printer documentation for recommended printer settings.
PRT.STAB
PRT.AUTO
PRT.ALL
PC-DIR.
PRT.STAB
PRT.AUTO
PRT.ALL
HOST
SEND.OFF Send mode switched off. (Factory setting)
SEND.STB
SEND.CONT
SEND.AUTO
SEND.ALL
2.DISPLAY
If the printed. (Factory setting)
Every stable weight value will be printed, without pressing the key.
If the regardless of stability.
Connection to a PC: the balance can send data (as a keyboard) to the PC used for PC applications e.g. Excel. The balance sends the weight value without the unit to the PC.
If the followed by an enter. (Factory setting)
Every stable weight value will be sent followed by an enter, without pressing the
If the an enter regardless of stability.
Connection to a PC, barcode reader etc.: the balance can send data to the PC and receive commands or data from the PC).
If the
All weight value updates will be sent regardless of stability, without pressing the
Every stable weight value will be sent, without pressing the key.
If the of stability.
Connection of an optional auxiliary display unit. The trans­mission parameters cannot be selected. Settings are automat­ically set.
key is pressed, the next stable weight value will be
key is pressed, the weight value will be printed
key is pressed, the next stable weight value will be sent
key.
key is pressed, the weight value will be sent followed by
key is pressed, the next stable weight value will be sent.
key.
key is pressed, the weight value will be sent regardless
NOTICE
Risk of damage to the device due voltage on connector
Devices could be damaged because of the voltage on connector Pin 9.
− Make sure that no other device is connected at COM1 as a 2nd display.
The Menu 31Semi-Micro Balances
Important
If you select 2nd display 2.DISPLAY, first make sure that no
other device is connected at COM1 as a 2nd display. Necessary for powering the 2nd display, see chapter
Interface specification.
HEADER – Options for the printout header of individual values
This menu topic allows you to specify the information that is to be printed at the top of the printout for every individual weighing results (after pressing
).
Note
This menu topic is only available if PRINTER setting was selected.
NO The header is not be printed. (Factory setting)
DAT / TIM
D / T / BAL
Date and time are printed.
Date, time and balance information (Balance type, SNR, Balance ID) are printed.
Balance ID only if set.
SINGLE – Options for printing out the result of individual values
This menu topic allows you to specify the information that is to be printed for every individual weighing result (after pressing
).
Note: this menu topic is only available if PRINTER setting was selected.
NET
The value of the net weight from the current weighing is printed. (Factory setting)
G / T / N
The values of the gross weight, the tare weight and the net weight are printed.
SIGN.L – Options for the printout footer for signature line of individual values
This menu topic allows you to set a footer for signature at the bottom of the printout for every individual weighing result (after pressing
).
Note: this menu topic is only available if PRINTER setting was selected.
OFF The signature footer is not be printed. (Factory setting)
ON
The signature footer is printed.
LINE.FEED – Options for complete the printout of individual values
This menu topic allows you to specify the number of blank lines to complete the printout (line feed) for every individual weighing result (after pressing
).
Note: this menu topic is only available if PRINTER setting was selected.
0 Possible numbers of blank lines: 0 to 99. (Factory setting = 0)
ZERO PRT. – Options for PRT.AUTO
1)
This menu topic allows you to specify the auto print function PRT.AUTO for printing zero YES or NO.
OFF Zero is not be printed (Zero +/- 3d). (Factory setting)
ON
Zero is always printed.
Note: this menu topic is only available if PRT.AUTO function of the PRINTER or PC-DIR. was selected.
The Menu32 Semi-Micro Balances
COM.SET – Options for the data communication format (RS232C) (HOST)
1)
This menu topic allows you to set the data format depending on which peripheral device is connected.
Note: this menu topic is only available if HOST setting was selected.
MT-SICS The MT-SICS data transfer formats is used. (Factory setting)
For more information see chapter MT-SICS Interface Commands and Functions.
MT-PM
The following PM balance commands are supported: S Send value SI Send immediate value SIR Send immediate value and repeat SR Send value and repeat SNR Send next value and repeat T Tare TI Tare immediately B Base *) MI Modify ambient vibration MZ Modify auto zero M Modified settings reset ID Identify CA Calibrate D Display (only symbol N and G available) *) Limitation:
Negative values are limited up to the current tare value.
B command is additive.
The sum of the B values plus the previous tare value, before a
"TA", "T" or "Z" is sent, must be less than the total weighing range.
SART
The following Sartorius commands are supported: K Ambient conditions: very stable L Ambient conditions: stable M Ambient conditions: unstable N Ambient conditions: very unstable O Block keys P Print key (print, auto print; activate or block) Q Acoustic signal R Unblock keys S Restart/self-test T Tare key W Calibration/adjustment (depending on the menu setting)
Z Internal calibration/adjustment
*)
**)
f0_ Function key (F) f1_ Function key (CAL) s3_ C key x0_ Perform internal calibration
**)
x1_ Print balance/scale model
The Menu 33Semi-Micro Balances
x2_ Print weighing cell serial number x3_ Print software version
*)
may be inaccessible on verified balances/scales
**)
only on models with built-in motorized calibration weight
Functionality mapping
HOST settings: Sartorius printer settings:
SEND.OFF
SEND.STB
SEND.ALL
SEND.CONT
SEND.AUTO
not applicable
manually print with stability
manually print without stability
automatically print without stability
similar applicable to automatically print when load is changed
BAUDRATE – Baud rate RS232C
1)
This menu topic allows you to match the data transmission to different serial RS232C receivers. The baud rate (data transfer rate) determines the speed of transmission via the serial interface. For problem-free data trans­mission the sending and receiving devices must be set at the same value.
The following settings are available: 600 bd, 1200 bd, 2400 bd, 4800 bd, 9600 bd, 19200 and 38400 bd. (default: 9600 bd)
Note
Not visible for 2nd display.
Each device has separate settings.
BIT/PAR. – Bit/Parity RS232C
1)
At this menu topic you can set the character format for the attached RS232C serial peripheral device.
8/NO 8 data bits/no parity (Factory setting)
7/NO
7/MARK
7/SPACE
7/EVEN
7/ODD
7 data bits/no parity
7 data bits/mark parity
7 data bits/space parity
7 data bits/even parity
7 data bits/odd parity
Note
Not visible for 2nd display.
Each device has separate settings.
STOPBIT – Stop bits RS232C
1)
At this menu topic you can set the stop bits of the transmitted data to different RS232C serial receivers.
1 BIT 1 Stop bit (Factory setting)
2 BITS
2 Stop bits
HD.SHAKE – Handshake RS232C
1)
This menu topic allows you to match the data transmission to different RS232C serial receivers.
XON/XOFF Software handshake (XON/XOFF) (Factory setting)
RTS/CTS
OFF
Hardware handshake (RTS/CTS)
No handshake
The Menu34 Semi-Micro Balances
Note
Not visible for 2nd display.
Each device has separate settings.
RS.TX.E.O.L. – End of line RS232C
1)
At this menu topic you can set the end of line character of the transmitted outgoing data to different RS232C serial receivers.
(CR)(LF)
<CR><LF> Carriage Return followed by Line feed (ASCII-Codes 013+010) (Factory setting)
(CR)
(LF)
(TAB)
<CR> Carriage Return (ASCII-Code 013)
<LF> Line feed (ASCII-Code 010)
<TAB> Tabulator to the right (ASCII-Code 009), only settable if
PC-DIR. is selected.
Note
Not visible for 2nd display.
Each device has separate settings.
RS CHAR – Char set RS232C
1)
At this menu topic you can set the character set of the transmitted data to different RS232C serial receivers.
IBM/DOS Char set IBM/DOS (Factory setting)
ANSI/WIN
Char set ANSI/WINDOWS
Note
Not visible for 2nd display.
Each device has separate settings.
USB – USB interface
At this menu topic you can select the mode of the USB device interface and specify how the data is transmitted.
NOTICE
Disconnect the USB connection from the balance prior to changing settings.
Note
This port is not usable for printers or displays.
PC-DIR.
SEND.OFF Send mode switched off. (Factory setting)
SEND.STB
SEND.CONT
SEND.AUTO
SEND.ALL
HOST
Connection to a PC: the balance can send data (as a keyboard) to the PC used for PC applications e.g. Excel.
Note: the balance sends the weight value without the unit to
the PC.
If the
key is pressed, the next stable weight value will be sent.
All weight value updates will be sent regardless of stability, without pressing the
key.
Every stable weight value will be sent, without pressing the key.
If the
key is pressed, the weight value will be sent regardless
of stability. Connection to a PC, Barcode Reader etc.: the balance can send
data to the PC and receive commands or data from the PC).
The Menu 35Semi-Micro Balances
SEND.OFF Send mode switched off. (Factory setting)
SEND.STB
SEND.CONT
SEND.AUTO
If the
All weight value updates will be sent regardless of stability, without pressing the
Every stable weight value will be sent, without pressing the
key is pressed, the next stable weight value will be sent.
key.
key.
SEND.ALL
If the
key is pressed, the weight value will be sent regardless
of stability.
USB COM.S. – Options for the data communication format (USB)
This menu topic allows you to set the data format depending on which peripheral device is connected.
MT-SICS The MT-SICS data transfer formats is used. (Factory setting)
For more information see chapter MT-SICS Interface Commands
and Functions.
MT-PM
The following PM balance commands are supported: S Send value SI Send immediate value SIR Send immediate value and repeat SR Send value and repeat SNR Send next value and repeat T Tare TI Tare immediately B Base *) MI Modify ambient vibration MZ Modify auto zero M Modified settings reset ID Identify CA Calibrate D Display (only symbol N and G available) *) Limitation:
Negative values are limited up to the current tare value.
B command is additive.
The sum of the B values plus the previous tare value, before a
"TA", "T" or "Z" is sent, must be less than the total weighing range.
SART
The following Sartorius commands are supported: K Ambient conditions: very stable L Ambient conditions: stable M Ambient conditions: unstable N Ambient conditions: very unstable O Block keys P Print key (print, auto print; activate or block) Q Acoustic signal R Unblock keys S Restart/self-test T Tare key W Calibration/adjustment (depending on the menu setting)
*)
The Menu36 Semi-Micro Balances
Z Internal calibration/adjustment
**)
f0_ Function key (F) f1_ Function key (CAL) s3_ C key x0_ Perform internal calibration
**)
x1_ Print balance/scale model x2_ Print weighing cell serial number x3_ Print software version
*)
may be inaccessible on verified balances/scales
**)
only on models with built-in motorized calibration weight
Functionality mapping
HOST settings: Sartorius printer settings:
SEND.OFF
SEND.STB
SEND.ALL
SEND.CONT
SEND.AUTO
not applicable
manually print with stability
manually print without stability
automatically print without stability
similar applicable to automatically print when load is changed
USB E.O.L. – End of line USB
At this menu topic you can set the end of line character of the transmitted data to USB device.
(CR)(LF)
<CR><LF> Carriage Return followed by Line feed (ASCII-Codes 013+010) (Factory setting)
(CR)
(LF)
<CR> Carriage Return (ASCII-Code 013)
<LF> Line feed (ASCII-Code 010)
(TAB) <TAB> Horizontal tab (ASCII-Code 011), only settable if PC-DIR.
is selected.
USB CHAR – Char set USB
At this menu topic you can set the character set of the transmitted data to USB device.
ANSI/WIN Char set ANSI/WINDOWS (Factory setting)
IBM/DOS
Char set IBM/DOS
INTERVAL – Print key simulation
At this menu topic you can activate a simulation of the
key. INTERVAL simulates a print key press every x
seconds.
Range: 0 to 65535 seconds 0 sec: disables the print key simulation
Factory setting: 0 sec
Note: the executed action is according to the configuration of the print key, see interface setting.
ERGOSENS – Settings for external key
The METTLER TOLEDO ErgoSens or external contact switches (optional, see section accessories) can be connected to the "Aux" connection and these can be used to execute certain weighing functions.
OFF Deactivate (Factory setting)
->0<-
Zero setting
The Menu 37Semi-Micro Balances
->T<-
PRINT
Taring
Print
———————————————————————————
1)
Note for 2nd RS232C interface
If an optional 2nd interface is installed, the menu topic is
displayed for each interface, e.g
BAUDRATE.1 for standard interface BAUDRATE.2 for optional 2nd interface
Only one printer can be set if two RS232 interfaces are existing.
———————————————————————————
The Menu38 Semi-Micro Balances
6 Applications
COUNTNG
D
A
B C
A
B
C
D
FIX 10
PCS
C
10PCS
------
FIX 10
PCS
A
6.1 Application piece counting
The Piece counting application allows you to determine the number of pieces put on the weighing pan.
Requirement: the function COUNTING must be assigned to an Fx key. See advanced menu topic ASSIGN:Fx, factory setting: F1.
− Press and hold the appropriate assigned Fx key to activate the function COUNTING.
Piece counting first requires the setting of a reference weight, there are 4 possibilities
Setting the reference by multiple pieces with fix
reference values.
Setting the reference by multiple pieces with
variable reference values.
Setting the reference for 1 piece in weighing mode.
Setting the reference for 1 piece in manual mode.
Setting the reference by multiple pieces with fix reference values
1 Select a number of reference pieces by scrolling with
. Possible numbers* are 5, 10, 20 and 50.
2 Press
empty container on the weighing pan and press
3 Add the selected number of reference pieces to
container.
4 Press
to zero the balance. If using: place
to tare the balance.
to confirm.
Legal-for-trade
* with approved balances in selected countries: min 10.
Applications 39Semi-Micro Balances
Setting the reference by multiple pieces with variable reference values
25 PCS
C
---------
VAR.REF
B
15
PCS
REF.
25
PCS
REF.
0.00
g
1
PCS
2.74
g
1
C
---------
PCS.WGT
C
1 Select VAR.REF by scrolling with .
2 Press
3 Select a number of reference pieces by scrolling up +
key or down - key. Speed up by press and hold. Possible numbers* are 1 to 999.
4 Press
empty container on the weighing pan and press
5 Add the selected number of reference pieces to
container.
6 Press
to confirm.
to zero the balance. If using: place
to tare the balance.
to confirm.
Legal-for-trade
* with approved balances in selected countries: min 10.
Setting the reference for one piece in weighing mode
1 Select PCS.WGT by scrolling with .
2 Press
empty container on the weighing pan and press
3 Add one reference piece to container.
ð The weight of one piece is displayed.
4 Press
to zero the balance. If using: place
to tare the balance.
to confirm.
Legal-for-trade
With approved balances, this setting is not available in selected countries.
Applications40 Semi-Micro Balances
Setting the reference for one piece in manual mode
2.68
g
*
0PCS
2.74
g
*
PCS.WGT
D
2.74
g
2.74
g
,
812.7
g
75
PCS
1 Select PCS.WGT by scrolling with .
2 Press 3 Enter the final reference one piece weight by scrolling
up + key or down - key. Speed up by press and hold.
4 Press
to confirm.
to confirm.
Legal-for-trade
With approved balances, this setting is not available in selected countries.
Switching between manual mode and weighing mode
− Press to switch between manual and weighing mode.
ð By switching from weighing mode to manual mode the
weight value will be transferred and can be changed manually.
Note
if without any key press within 60 seconds or by pressing C, the balance returns to the previous active appli­cation.
On completion of the setting procedure, your balance is ready.
Switching between piece count and weight display
− Press key at any time to switch the display between piece display, weighing unit UNIT 1, RECALL value (if activated) and weighing unit UNIT 2 (if different from UNIT 1).
The RECALL value is displayed with an asterisk (*)
and icon M and can not be printed.
Take into account minimum values: min. reference weight = 10d (10 digits), min. piece weight* = 1d (1
digit)! * with approved balances in selected countries: min 3e
The current reference weight remains stored until the reference setting is changed.
Terminate the application
Press and hold
to terminate the application and to return to the weighing application.
Applications 41Semi-Micro Balances
6.2 Application percent weighing
PERCENT
A
B
A
B
0.0
%
175.28
g
*
A
SET 100%
I00.0
%
175.28
g
¢¢¢¢
SET 100%
C
0.00
g
B
175.28
g
175.28
g
,
The Percent weighing application allows you to check a sample weight as percentage to a reference target weight.
Requirement: the function PERCENT must be assigned to an Fx key. See advanced menu topic ASSIGN:Fx, factory setting: F2.
− Press and hold the appropriate assigned Fx key to activate the function PERCENT.
Percent weighing first requires the setting of a reference weight that should corresponds to 100%, there are 2 possibilities
Setting the reference in manual mode (enter
100%).
Setting the reference in weighing mode (weigh
100%).
Setting the reference by manual mode (enter 100%)
1 Press to activate manual mode. 2 Select the reference target weight (100%) by scrolling
up + key or down – key. Speed up by press and hold.
3 Press
to confirm.
Setting the reference by weighing mode (weigh 100%)
1 Press to zero the balance. If using: place
empty container on the weighing pan and press
to tare the balance.
2 Load the reference weight (100%). Reference weight
must be at least +/- 10d.
to confirm.
3 Press
Switching between manual mode and weighing mode
− Press to switch between manual and weighing mode.
ð By switching from weighing mode to manual mode the
Applications42 Semi-Micro Balances
weight value will be transferred and can be changed manually.
Note
,
128.25
g
73.17
%
128250
mg
128.25
g
M
if without any key press within 60 seconds or by pressing C, the balance returns to the previous active appli­cation.
On completion of the setting procedure, your balance is ready.
Switching between percent and weight display
− Press key at any time to switch the display between
percent display, weighing unit UNIT 1, RECALL value (if activated) and weighing unit UNIT 2 (if different from UNIT 1).
The RECALL value is displayed with an asterisk (*)
and icon M and can not be printed.
The current set weight remains stored until it is
redetermined.
Terminate the application
Press and hold
to terminate the application and to return to the weighing application.
Applications 43Semi-Micro Balances
6.3 Application statistics
CLEAR M?No
STAT
M .CLEARED
46.36
g
0.00
g
- 1 -
C
999
C
0.00
g
The Statistics application allows you to generate statistics of a series of weighing values. 1 to 999 values are possible.
Requirement: the function STAT must be assigned to an Fx key. See advanced menu topic ASSIGN:Fx. Connect a printer or a PC if present.
1 Press and hold the appropriate assigned Fx key to
activate the function STAT.
2 Press 3 Press
4 Press
If the memory is already cleared (at the first start of this application or sample counter is 0) the memory clear question will be not displayed.
Weighing the first sample weight
1 Press to zero the balance. 2 Load the first sample weight.
3 Press
ð The display shows the sample count - 1 - and the
4 When the sample counter is displayed you may press
and hold C to undo (drop) this sample.
5 Unload the first sample weight.
to continue the last statistics. to start a new statistical evaluation.
and select Yes to clear the memory.
.
current weight is stored as sample and the weight is printed out.
Weighing further sample weights
The same procedure as for the first sample weight.
1…999 samples are possible.
The next value will be accepted if the sample weight is in the range 70% –130% of the current average
value. OUT OF RANGE will be displayed if the sample is not accepted.
Results
− Press , if the numbers of sample are greater than or equal to 2.
ð The results are displayed and printed.
Displayed results
1 Press
to show the next statistical value.
2 Press C to cancel displaying results and to continue weighing next sample.
Applications44 Semi-Micro Balances
5
N
X
50
.
530
g
S.DEV
3.961
g
S.REL
7.84
%
MIN
46.36
g
MAX
55.81
g
DIFF
9.45
g
SUM
252.65
g
Terminate the application
Press and hold
to terminate the application and to return to the weighing application.
0.5 seconds
number of samples
average
standard deviation
relative standard deviation
lowest value (minimum)
highest value (maximum)
different between the minimum
and the maximum
sum of all values
Applications 45Semi-Micro Balances
6.4 Application formulation (Net total)
CLEAR M?
No
FORMULA
M .CLEARED
#
-T1-
0.00
g
Net
10.33
g
C
0.00
g
#
The Formulation (Net total) application allows you to
weigh in (add and store) up to 999 individual component weights and displays the total.
If a printer is connected, the component weights are printed individually and as a total.
tare/pre-tare and store up to 999 container weights and displays the total. If a printer is
connected, the tare weights are printed out individually and as a total.
fill up the sum of all component net weight values by adding a further component to a
higher value.
Requirement: the function FORMULA must be assigned to an Fx key. See advanced menu topic ASSIGN:Fx. Connect a printer or a PC if present.
Note
Connect a printer or a PC if present.
1 Press and hold the appropriate assigned Fx key to
activate the function FORMULA.
2 Press
3 Press
4 Press
If the memory is already cleared (sample and tare/pre-tare counter is zero) the memory clear question will be not displayed.
to continue formulation weighing.
or (+ or -) to start a new formulation.
and select Yes to clear the memory.
Tare container
If used.
1 Press to zero the balance. 2 Place the empty container on the weighing pan.
3 Press
to tare the balance.
ð The container is tared and the tare count - T1 - is
displayed and the tare weight is printed.
If you pre-tare via MT-SICS (e.g. bar code reader) -
PT1 - is displayed.
Zero range setting (menu topic ZERO RNG) has no
effect. The zero-limit is less than or equal 10d.
Applications46 Semi-Micro Balances
Weighing the first component weight
Net
Net
C
8
N
T.TOTAL
452
.
76
g
G.TOTAL
546.79
g
N.TOTAL
94.03
g
FILL UP
1 Load the first component weight.
2 Press
.
ð The display briefly shows the component count - 1
- , the current weight is stored as sample and the
component weight is printed. The display is set back to zero.
Weighing further component weights
The same procedure as for the first component weight with the same or new container).
1…999 sample values are possible.
max 999 tare values are possible.
max 999 pre-tare values are possible.
Results
− Press , if the numbers of sample are greater than or
equal to 2. ð The results are displayed and printed.
Displayed results
1 Press
to show the next statistical value.
2 Press C to cancel displaying results and to continue weighing next component.
0.5 seconds
number of samples
sum of all tare values (T and
PT)
sum of all component gross
weight values
sum of all component net
weight values
Function FILL UP
This function allows you to add an additional component weight to the total weight of all components to reach a desired target weight (Fill up).
Starting the fill up function
1 Press + to activate the function FILL UP.
2 Press – to deactivate the function FILL UP.
Applications 47Semi-Micro Balances
Filling up with an additional component weight
Net
Net
94.03
100.00
Net
0.00
F9
The last total of the component weights is displayed.
§
1 Add component weight until the desired target weight is
reached.
2 Press ð The display briefly shows the next component count
marked with F, the current weight is stored as sample and the component weight is printed. The display is set back to zero.
to confirm.
Filling up further additional component weights
The same procedure, beginning with starting up the FILL UP function.
Terminate the application
Press and hold
to terminate the application and to return to the weighing application.
Applications48 Semi-Micro Balances
6.5 Application totaling
CLEAR M?
No
TOTALING
M .CLEARED
0.00
g
46.36
g
¢¢¢¢¢¢
- 1 -
999
0.00
g
0.00
g
¢¢
C
C
879
N
TOTAL
8789
.
79
g
The TOTALING application allows you to weigh in different samples to add their weight values and to totalize them. 1 to 999 samples are possible.
Requirement: the function TOTALING must be assigned to an Fx key. See advanced menu topic ASSIGN:Fx.
1 Press and hold the appropriate assigned Fx key to
activate the function TOTALING.
2 Press
3 Press
4 Press If the memory is already cleared (sample counter is 0) the
memory clear question will be not displayed.
Weighing in the sample weight
1 Press to zero the balance. If using: place empty
container on the weighing pan an press
the balance. 2 Load the first sample weight.
3 Press
ð The display shows the sample count - 1 - and the
4 When the sample counter is displayed you may press
and hold C to undo (drop) this sample.
5 Unload the first sample weight.
ð The display shows zero.
or (+ or -) to start a new totaling evaluation.
and select Yes to clear the memory.
to zero the balance.
.
current weight is stored.
to tare
Weighing in further sample weights
The same procedure as for the first sample weight.
1…999 samples are possible.
Results
− Press , if the numbers of sample are greater than or
Displayed results
1 Press
2 Press C to cancel.
equal to 2.
ð The results are displayed and printed.
to show the totalized value.
0.5 seconds
number of samples
totalized value
Applications 49Semi-Micro Balances
Terminate the application
Press and hold
to terminate the application and to return to the weighing application.
Applications50 Semi-Micro Balances
6.6 Application multiplication factor weighing
FACTOR M
SET F.MUL
0001.0000
0009.8
Fac
100
Fac
SET STEP
001.0
Stp
005.0
Stp
The Multiplication factor weighing application allows you to multiply the weight value (in grams) by a predefined factor (result = factor * weight) and have it calculated to a predefined number of decimal places.
Requirement: the function FACTOR M must be assigned to an Fx key. See advanced menu topic ASSIGN:Fx.
− Press and hold the appropriate assigned Fx key to
activate the function FACTOR M.
Setting the factor value
Zero for multiplication factor value is outside the allowed range, the error message FACTOR OUT OF RANGE will be displayed.
1 Press to execute SET.F.MUL.
ð Either the factor 1 appears as default value or the
factor that was saved most recently.
2 Press
ð The selected digit is blinking.
3 Press + to scroll up or – to scroll down to changing
digits. 4 Press
to select a digit.
to confirm (no automatic acceptance).
Setting the step value
SET.STEP appears in the display, and the program changes automatically to allow the display increments to be
entered. The smallest possible display increment appears as default value, or the last value that was saved. The allowed range for the step depends on the factor and the resolution of the balance. If it is outside the
allowed range the error message STEP OUT OF RANGE will be displayed.
1 Press to execute SET.STEP.
2 Press
ð The selected digit is blinking.
3 Press + to scroll up or – to scroll down to changing
digits. 4 Press if without any key press within 60 seconds or by pressing
C, the balance returns to the previous active application.
On completion of the setting procedure, your balance is ready.
to select a digit.
to confirm (no automatic acceptance).
Applications 51Semi-Micro Balances
Weighing procedure
C
9810
1000.02
g
0
C
1 Press to zero the balance. 2 Load sample weight. 3 Read the result.
ð The appropriate calculation is then made using the
weight of sample and the selected factor, the result
being displayed with the selected display step. No
units are displayed.
4 Unload sample weight.
Toggling between displaying the calculated value and the measured weight
− Press
key at any time to switch the display between percent display, weighing unit UNIT 1, RECALL
value (if activated) and weighing unit UNIT 2 (if different from UNIT 1).
Terminate the application
Press and hold
to terminate the application and to return to the weighing application.
Applications52 Semi-Micro Balances
6.7 Application division factor weighing
FACTOR D
SET F.DIV
0001.0000
009.8
100
Fac
Fac
0
SET STEP
0.0001
0.0050
Stp
Stp
The Division factor weighing divide a predefined factor by the weight value (in grams) (result = factor / weight) and have it rounded to a predefined number of decimal places.
Requirement: the function FACTOR D must be assigned to an Fx key. See advanced menu topic ASSIGN:Fx.
− Press and hold the appropriate assigned Fx key to
activate the function FACTOR D.
Setting the factor value
Zero for division factor value is outside the allowed range, the error message FACTOR OUT OF RANGE will be displayed.
1 Press to execute SET.F.DIV.
ð Either the factor 1 appears as default value or the
factor that was saved most recently.
2 Press
ð The selected digit is blinking.
3 Press + to scroll up or – to scroll down to changing
digits. 4 Press to confirm (no automatic acceptance).
to select a digit.
Setting the step value
SET.STEP appears in the display, and the program changes automatically to allow the display increments to be
entered. The smallest possible display increment appears as default value, or the last value that was saved. The allowed range for the step depends on the factor and the resolution of the balance. If it is outside the
allowed range the error message STEP OUT OF RANGE will be displayed.
1 Press to execute SET.STEP.
2 Press
3 Press + to scroll up or – to scroll down to changing
4 Press if without any key press within 60 seconds or by pressing
C, the balance returns to the previous active application.
On completion of the setting procedure, your balance is ready.
to select a digit.
ð The selected digit is blinking.
digits.
to confirm (no automatic acceptance).
Applications 53Semi-Micro Balances
Weighing procedure
C
0.010
1000.02
g
0.000
C
1 Press to zero the balance. 2 Load sample weight. 3 Read the result.
ð The appropriate calculation is then made using the
weight of sample and the selected factor, the result
being displayed with the selected display step. No
units are displayed.
4 Unload sample weight.
Toggling between displaying the calculated value and the measured weight
− Press
key at any time to switch the display between percent display, weighing unit UNIT 1, RECALL
value (if activated) and weighing unit UNIT 2 (if different from UNIT 1).
Terminate the application
Press and hold
to terminate the application and to return to the weighing application.
Applications54 Semi-Micro Balances
6.8 Application density
DENSITY
SOLID
LIQUID
WEIGH IN
60.0020
g
¢¢¢¢
The Density application allows you to determine the density of solid bodies and liquids. Determination of the density uses Archimedes' principle according to which a body immersed in a fluid undergoes an apparent loss in weight which is equal to the weight of the fluid it displaces.
To determine the density of solid bodies, we recommend you to work with the optional density kit which contains all the attachments and aids needed for convenient and precise density determination. To determine the density of liquids, you additionally need a sinker which you can also obtain from your METTLER TOLEDO dealer.
Note for performing of density determinations
You can also use the hanger for weighing below the balance which belongs to your
balance.
If a METTLER TOLEDO printer is attached to your balance, the settings will be automat-
ically recorded.
We recommend you to consult the operating instructions enclosed with the density kit.
Requirement: the function DENSITY must be assigned to an Fx key. See advanced menu topic ASSIGN:Fx. Density kit is installed.
− Press and hold the appropriate assigned Fx key to
activate the function DENSITY.
Setting the method for density determination
1 Select:
SOLID, the function for the density determination of
solids, or
LIQUID, the function for the density determination of
liquids with a sinker. 2 Press
to confirm.
Switching the display between user guidance and weighing
− Press to toggle the display between user guidance
and weighing.
Terminate the application
Press and hold
to terminate the application and to return to the weighing application.
Applications 55Semi-Micro Balances
6.8.1 Density determination of solids
H-2-0
ETHANOL
FREE
°C
20.0
°C
21.8
g/cc
g/cc
01.00200
01.00000
PRESS EN
WEIGH IN AIR
WEIGH IN LIQUID
Setting the parameter of the auxiliary liquid
Method SOLID is set.
§
1 Press or (+ or -) to select the auxiliary liquid:
H-2-O for distilled water, ETHANOL or FREE for a freely
definable auxiliary liquid.
2 Press
If you have selected water or ethanol as the auxiliary liquid
1 Enter the current temperature of the auxiliary liquid
(read off on thermometer).
2 Change the value by scrolling up + or down -. The
temperature ranges from 10°C to 30.9°C.
3 Press The densities of distilled water and ethanol in the range
10°C to 30.9°C are stored in the balance.
If you have selected a freely definable auxiliary liquid
1 Enter the density of the auxiliary liquid at the current
temperature (read off on thermometer).
2 Press
ð The selected digit is blinking.
3 Press + to scroll up or – to scroll down to changing
digits.
4 Press if without any key press within 60 seconds or by pressing
C, the balance returns to the previous active application.
On completion of the setting procedure, your balance is ready.
Taring the balance is possible at any time.
The balance prompts you: PRESS ENTER TO START.
− Press to start. ð Tare/Zero is executed.
The balance prompts you to weigh the solid in air WEIGH IN AIR.
1 Load the solid. 2 Press The balance prompts you to weigh the solid in the
auxiliary liquid WEIGH IN LIQUID.
to confirm.
to confirm.
to select a digit.
to confirm.
to initiate the measurement.
1 Load the solid. 2 Press
ð The balance now shows the determined density of
Applications56 Semi-Micro Balances
to initiate the measurement.
the solid.
5.988
g/cc
*
This result has already been corrected for the air
010.0000
cm3
012.0000
cm3
PRESS EN
WEIGH IN AIR
WEIGH IN LIQUID
1.000
g/cc
*
buoyancy. The buoyancy caused by the two immersed
wires (ø0.6mm) can be neglected.
− Press C the balance returns to PRESS ENTER TO
START.
Result
− Press .
ð The result will be printed.
6.8.2 Density determination of liquids
Setting the displacement volume of your sinker
Method LIQUID is set.
§
1 Press 2 Press to select a digit.
ð The selected digit is blinking.
3 Press + to scroll up or – to scroll down to changing
digits. 4 Press if without any key press within 60 seconds or by pressing
C, the balance returns to the previous active application.
On completion of the setting procedure, your balance is ready.
Taring the balance is possible at any time.
The balance prompts you: PRESS ENTER TO START.
− Press to start.
The balance prompts you to weigh the sinker in air WEIGH IN AIR.
to confirm the default value of 10.0 cm3 or change it if needed.
to confirm.
1 Position the sinker. 2 Press The balance prompts you to weigh the sinker in the liquid
WEIGH IN LIQUID.
1 Pour the liquid into the beaker. Make sure that the
2 Press
This result has already been corrected for the air
− Press C the balance returns to PRESS ENTER TO
to initiate the measurement.
sinker is immersed by al least 1 cm in the liquid, and
that there are no air bubbles in the container.
to initiate the measurement.
ð The balance now shows the determined density of
the liquid at the current temperature (read off on the thermometer).
buoyancy. The buoyancy caused by the immersed wire
(ø 0.2 mm) of the sinker can be neglected.
START.
Applications 57Semi-Micro Balances
Result
− Press . ð The result will be printed.
6.8.3 Formulae used to calculate density
The DENSITY application is based on the formulae listed below.
Formulae for determining the density of solids with compensation for air density
= Density of the sample
A = Weight of the sample in air B = Weight of the sample in the auxiliary liquid V = Volume of the sample
= Density of the auxiliary liquid
= Density of air (0.0012 g/cm3)
= Weight correction factor (0.99985), to take the atmospheric buoyancy of the adjustment weight
into account
Formula for determining the density of liquids with compensation for air density
= Density of the liquid
P = Weight of the displaced liquid V = Volume of the sinker
= Density of air (0.0012 g/cm3)
= Weight correction factor (0.99985), to take the atmospheric buoyancy of the adjustment weight
into account
Applications58 Semi-Micro Balances
Density of H2O given in g/cm3, according to the "American Institute of Physics Handbook".
Density of C2H5OH given in g/cm3, according to the "American Institute of Physics Handbook".
Applications 59Semi-Micro Balances
6.9 Application pipette check
PIPETTE
H20 20.0 °C
H20 21.3 °C
1013 hPa
985 hPa
The PipetteCheck application allows you to check the volume of pipettes from any manufacturer, with the gravimetric method. For checking pipettes we recommend using the optional METTLER TOLEDO evaporation trap. This evaporation trap minimizes moisture evaporation for more accurate results.
Maximum 3 different test volumes are possible for checking the pipette. The test volumes recommended by the manufacturer are usually 10%, 50%, and 100% of the nominal volume of the pipette. The liquid to determine the volume of the pipette is water and the following conditions must be known:
Current temperature of the test liquid
Current barometric air pressure of the test environment
Current relative humidity of the test environment
Based on the test results on the printout of the statistic and your specifications, you can decided whether the pipette can be used for further applications (successfully or failed).
Requirement
A printer must be connected.
The evaporation trap is ready installed (recommended).
The function PIPETTE must be assigned to an Fx key. See advanced menu topic
ASSIGN:Fx. Density kit is installed.
− Press and hold the appropriate assigned Fx key to activate the function PIPETTE.
Setup
Setting the test liquid temperature
The setting range is 15.0°C up to 30.0°C.
1 Press + to scroll up or – to scroll down to changing
digits.
2 Press
to confirm (no automatic acceptance).
Setting the barometric air pressure of the test environment
The setting range is 850 hPa up to 1090 hPa.
1 Press + to scroll up or – to scroll down to changing
digits.
2 Press
to confirm (no automatic acceptance).
Setting the relative humidity of the test environment
The setting range is 20% up to 90%.
Applications60 Semi-Micro Balances
50 %
61 %
1 Press + to scroll up or – to scroll down to changing
S.SIZE 4
S.SIZE 10
00250
µl
µl
00100
digits.
2 Press
Setting the sample size
Set how many measurements should be performed for the selected test volume before the measurement cycle is complete. The setting range is from 4 up to 10.
1 Select the sample size. 2 Press 3 Press
to confirm (no automatic acceptance).
to toggle between the values 4 to 10. to confirm (no automatic acceptance).
Setting the test volume
The setting range is 1µl up to 20000µl. 1 Press to select a digit.
ð The selected digit is blinking.
2 Press + to scroll up or – to scroll down to changing
digits.
3 Press
check. ð Application header and settings are printed.
if without any key press within 60 seconds or by pressing C, the balance returns to the previous active application.
to confirm the setting and to start the pipette
Applications 61Semi-Micro Balances
Pipette check start
11.34
µl
SAMPLE I
C
Sample 4 or 10
0.00
µl
SAMPLE 2
0.00
µl
SYS.ERR
1.24
µl
abs
SYS.ERR
10.91
%
rel
RAND.ERR
0.13
µl
abs
RAND.ERR
1.32
%
rel
NEW T.VOL
Yes
Sample1 appears briefly on the display (Sample counter).
§
The balance shows zero.
§
1 Weigh the first sample. 2 Press
to confirm.
ð The weight of sample1 is stored and printed.
3 Weigh next samples with the same procedure as for
the first sample, until the defined number of samples is reached.
ð The statistics is printed and shown on the display.
Displayed results
1 Press
(several times) to show all statistic values on the display first.
2 Press C to continue the pipette check.
systematic error, absolute
0.5 seconds
systematic error, relative
random error, absolute
random error, relative
Continue pipette check
Continue the check with another test volume. Three different test volumes are possible. After three test
volume, the balance terminates the check of the pipette automatically.
Terminate pipette check.
NEW T.VOL appears on the display.
§
1 Select YES to continue or select NO to terminate the
check of the pipette.
2 Press
3 Press
Continue next pipette check
1 Set the next test volume.
to toggle between YES or NO.
to confirm.
Applications62 Semi-Micro Balances
00250
µl
µl
00100
2 Press to select a digit.
ð The selected digit is blinking.
3 Press + to scroll up or – to scroll down to changing
digits.
4 Press
check. ð The sample header with the new test volume is
5 To continue, refer to section "Pipette check start". Same
procedure for further test volume.
to confirm the setting and to start the pipette
printed.
Terminate pipette check
The printout will be completed and the balance returns to the weighing application.
Applications 63Semi-Micro Balances
6.10 Application routine test
R.TEST
TARE
No
USE
WGT
TEST
Ref
5000.00
g
CTRL. LMT
Ref
2.50
g
The Routine test application allows you to determine the sensitivity of the balance. More about periodic sensitivity tests (routine tests), See GWP® (Good Weighing Practice) on
http://www.mt.com/gwp.
GWP gives clear recommendation for routine testing:
how should I test my balance?
how often?
where can I reduce efforts?
More about test weights, see
Requirement
The function R. TEST must be assigned to an F3 key. See advanced menu topic
ASSIGN:F3.
It is recommended to connect a printer or a PC to the balance for showing the results.
1 Press and hold the appropriate assigned Fx key to
activate the function R. TEST.
2 Press
use a tare weigh for test.
3 Press
4 Press
It is recommended to test the sensitivity without tare
load. (Factory setting: No).
If using tare: Make sure that tare weight plus test weight is not exceeding max. load.
or (+ or -) to toggle between No or Yes to
to confirm.
to zero or tare the balance.
http://www.mt.com/weights.
Setting the reference test weight value
The default value of the test weight: Next smaller OIML weight than the maximum load of your balance according to the GWP® recommendation.
1 For changing the value, press + to scroll up or – to
scroll down. Progressing speed by press and hold.
2 Press
to confirm.
Setting the control limit
The default value of the control limit: Test weight × weighing process tolerance / 2 Example: 5000 g × 0.1% / 2 = 2.50 g.
1 For changing the value, press + to scroll up or – to
scroll down. Progressing speed by press and hold.
2 Press
to confirm.
Setting the warning limit
The default value of the warning limit: Warning limit = control limit / safety factor Example: 2.5 g / 2 = 1.25 g.
Applications64 Semi-Micro Balances
LMT
WARN.
Ref
1.25
g
1 For changing the value, press + to scroll up or – to
PRESS EN
PRESS EN
5000.00
g
scroll down. Progressing speed by press and hold. 2 Press The default values of control limit and the warning limit are
evaluated according the GWP recommendation. These are based under the assumption that the weighing process tolerance is 0.1% and the safety factor is 2.
On completion of the setting procedure, your balance is ready.
The test weight must be acclimatized to the ambient temperature of the balance. 1 Press to start.
ð Follow the instructions on the display. 2 Load the test weight (displayed value), when the test
weight value is flashing. ð The printout starts after the weighing pan is unloaded.
to confirm.
Exit the current test procedure
− Press and hold
for executing a new application.
What if warning limit or control limit are FAILED?
The "SOP for Periodic Sensitivity Tests (Routine Tests)" provides information about measures when routine tests fail.
Find a download version of these SOPs on http://www.mt.com/gwp, link http://http://www.mt.com/
ch/en/home/library/operating-instructions/laboratory-weighing/ free_standard_operating_procedures_for_balance_straightforward_testing.html
.
Content of SOP
Preparation
Test procedure
Evaluation
Deviation
– If Warning Limit FAILED
– If Control Limit FAILED
Applications 65Semi-Micro Balances
6.11 Application diagnostics
The Diagnostics application allows you to carry out predefined diagnostics tests and to view or print predefined sets of balance information. This diagnostics tool helps you find errors faster and more efficiently. Requirement: A printer or a PC is connected to the balance for showing the results.
1 Activate ADVANCED menu.
2 Press
3 Press
6.11.1 Repeatability test
The repeatability test allows you to repeat tests with internal weight for a given number of times.
Note: on models with internal weights only.
1 Press
ð R. TST. 10 appears on the display.
2 Enter the number of times (blinking) by pressing + or –. Possible values are 5, 10 (factory setting), 20,
50, 100 times.
3 Press
ð The message RUNNING REPEAT TEST appears till the tests are completed.
4 Press 5 Press
6 Press C to cancel the test procedure.
ð The balance will return to the topic DIAGNOSE.
Sample information displayed
to activate the function DIAGNOSE.
to select appropriate tests.
to activate repeatability test REPEAT.T.
to start the test.
to print the test information.
to scroll forward through the displayed list.
Displayed for 0.5 s Display
S DEV
MAX. TEMP
MIN. TEMP
MEAN. TEMP
TOT.TIME
* 0.004 g
21.2 °C
21.0 °C
21.1 °C
00:01:26
Examples
Repeatability test is a tool to do functional check with the balance. It may be performed:
To check function of balance
– during installation to store print out with installation documents. – after preventative maintenance to store print out with installation maintenance report. – when remarkable decrease of weighing performance occurs, so that you can email/fax print out to
service support provider for diagnose purposes.
To develop the optimal environment settings, see menu topic ENVIRON..
Measure the time you need to perform repeatability test with each STABLE, STANDARD and UNSTABLE setting. The setting with the fastest total time suits best for the existing environmental conditions.
6.11.2 Display test
The display test allows you to test the display of the balance.
1 Press
ð All possible segments and icons on the display will illuminate.
2 Press
Applications66 Semi-Micro Balances
to start DISPLAY.
to print the test information.
3 Press C to cancel the test procedure.
ð The balance will return to the topic DIAGNOSE.
6.11.3 Key test
The key test allows you to test the keys of the balance.
1 Press
ð The message KEY TEST - PRESS KEY TO BE TESTED is displayed scrolling during the duration of the
key test.
2 Press every key briefly. Each press of a key beeps and echoes with OK on the display.
3 Second press C key to print the test information.
ð The test procedure will be cancelled and the balance will return to the topic DIAGNOSE. If a key has not
been tested before printing, then the test results will be indicated with a ---- line.
Sample information displayed
to start KEYPAD T.
Key Display
1/10 D OK
MENU OK
CAL OK
PRINT OK
6.11.4 Motor test
The motor test allows you to test the calibration motor of the balance.
Note: on models with internal weight only.
1 Press
ð RUNNING is displayed during the motor test. A motor test is deemed successful when all the motor
2 Press
3 Press C to cancel the test procedure.
ð The balance will return to the topic DIAGNOSE.
to start CAL.MOT. T.
positions have been successfully tested. At the end of the test, the test information will be printed.
for printout.
6.11.5 Balance history
The balance history function allows you to view and print the history of the balance.
1 Press
2 Press 3 Press
4 Press C to cancel the test procedure.
ð The balance will return to the topic DIAGNOSE.
to start BAL.HIST.
for printout. to scroll forward through the displayed list of balance history information.
MINUS OK
+ PLUS OK
TOGGLE OK
ENTER OK
C C OK
0/T OK
Applications 67Semi-Micro Balances
Sample information displayed
Operation time (year:day:hour) 00:018:04
Total load kg 115.7191 kg
Number of weighings 1255
Number of key pressed 4931
Number of motor movements 1012
Backlight time (year:day:hour) 00:018:04
Next service due date 01:01:2010
6.11.6 Calibration history
The calibration history function allows you to view and print information of the last 30 (thirty) balance adjustment. Adjustments made by a service technician and normal user are counted together.
1 Press
2 Press 3 Press
4 Press C to cancel the test procedure.
ð The balance will return to the topic DIAGNOSE.
Sample information displayed
S = External adjusted service 05:03:09S 01
F = FACT 05:03:09F 02
I = Internal adjusted 04:03:09I 28
E = External adjusted user 03:03:09E 29
F = FACT 02:03:09F 30
to start CAL.HIST.
for printout. key to scroll forward through the displayed list of adjustments history information.
Information Display
Note Display
-3 PPM
2 PPM
·
·
·
-1 PPM
4 PPM
1 PPM
·
·
·
The PPM value is the change since the last performed FACT or adjustment.
6.11.7 Balance information
The balance information function allows you to view and print information about your balance.
1 Press
2 Press 3 Press
4 Press C to cancel the test procedure.
ð The balance will return to the topic DIAGNOSE.
Sample information displayed
Balance type
Applications68 Semi-Micro Balances
to start BAL.INFO.
for printout. to scroll forward through the displayed list of Balance information.
Information Display
TYPE MS6002S
Information Display
Max. load
Software platform
Serial number
Type definition number
Software version
Cell ID
Cell type
Tolerance revision number
Language
6.11.8 Service provider information
The service provider Information function allows you to print information about your service provider.
1 Press
ð The service provider information appears. 2 Press ð The service provider information will be printed and the balance will return to the topic DIAGNOSE.
to start PROVIDER.
.
MAX 6200 g
PLATFORM RAINBOW
SNR 1234567890
TDNR 9.6.3.411
SOFTWARE V1.00
CELL ID 1172400044
CELL TYPE MMAI6000G2
TOLERANCE NO2
LANGUAGE ENGLISH
Applications 69Semi-Micro Balances
7 Communication with Peripheral Devices
This section lists some typical examples of where the balance can communicate with peripheral devices. In most cases, it is possible to do so via USB or RS232. The chapters are primarily restricted to a USB setup.
7.1 USB - interface and installation
Before connecting the balance via the USB device interface and using either the HOST or PC-Direct function, the appropriate METTLER TOLEDO USB driver has to be assigned to the PC first. The USB driver can be found on
www.mt.com/labweighing-software-download. If you have any questions, please contact a METTLER TOLEDO
representative.
Note
If you connect the balance via USB to the PC before installing the METTLER TOLEDO USB driver, Windows will automatically install the wrong driver.
Requirements
Balance with USB device interface
PC with one of the following Microsoft Windows
Win 10
Administrator rights for installing software
USB connection cable to link PC to balance
Download USB driver
1 Connect to the internet. 2 Go to the site
3 Click Download Driver in section USB Driver for Advanced and Standard level laboratory balances.
ð A pop-up window with interactions appears.
4 Click e.g. Open.
ð The extract screen appears.
5 Extract the file MT_Lab_Advanced-Standard_USB_Driver_Installer_v1.24.zip to your specified location.
6 Right-click on the downloaded installation program MT_Lab_Advanced-
Standard_USB_Driver_Installer_v1.24.exe and select Run as Administrator.
7 If a safety warning appears, confirm windows to perform the installation.
8 Click Next and follow the installer's instructions.
www.mt.com/labweighing-software-download.
®
32-bit/64-bit operating systems: Win 7 (SP1), Win 8 or
Installing the balance
1 Switch the balance off.
2 Connect the balance to the preferred USB port on the PC.
3 Switch the balance on.
Communication with Peripheral Devices70 Semi-Micro Balances
7.2 Send weight value via USB or RS232 to a PC using PC-Direct
The numerical value displayed at the balance can be transferred to the cursor position in windows applications (e.g. Excel, Word) in the same way as typing with the keyboard.
The data is transferred via USB or via the serial RS232 interface. The weight value without the unit will be transferred.
Requirements
PC with one of the following Microsoft Windows
Win 10
Serial interface RS232 or USB
Administrator rights for installing the SerialPortToKeyboard software (if data transfer is via RS232)
Windows application (e.g. Excel)
Connection between balance and PC via cable RS232 or USB
®
32-bit/64-bit operating systems: Win 7 (SP1), Win 8 or
7.2.1 PC-Direct via USB
The balance can send data (as a keyboard) to the PC used for PC applications, e.g. Excel. The balance sends the weight value without the unit to the PC.
Use the USB connection cable to connect the balance with the PC. Connect the USB cable to the USB device on the balance.
NOTICE
Disconnect the USB connection from the balance prior to changing settings.
Balance interface settings, see chapter "Interface menu".
Topic USB
− Set PC-DIR. and select the most appropriate option for the desired weighing result.
Topic USB E.O.L./USB E.O.L
1 Set TAB to write into the same row (e.g. in Excel).
2 Set CR LF to write into the same column (e.g. in Excel).
3 Save changes.
The balance must be disconnected from the PC.
§
1 Connect the balance to the PC.
2 Place the sample on the weighing pan. 3 Press
, the next stable weight will be send to the cursor position of your application.
Communication with Peripheral Devices 71Semi-Micro Balances
7.2.2 PC-Direct via RS232
Balance interface settings, see chapter "Interface menu".
7.2.2.1 Installing SerialPortToKeyboard software
The operation of PC-Direct via serial port RS232 requires the installation of SerialPortToKeyboard on your host computer. The file SerialPortToKeyboard can be found on you have any questions, please contact a METTLER TOLEDO representative.
Download SerialPortToKeyboard
1 Connect to the internet. 2 Go to the site
3 Click Download Software and Instructions in section SerialPortToKeyboard software for Advanced and
Standard level laboratory balances.
ð A pop-up window with interactions appears.
4 Click e.g. Open.
ð The extract screen appears.
5 Extract the file SerialPortToKeyboard_V_1.11_installer_and_instructions.zip to your specified location.
6 Right-click on the downloaded installation program SerialPortToKeyboard_V_1.11.exe and select Run as
Administrator.
7 If a safety warning appears, confirm windows to perform the installation.
8 Click Next and follow the installer's instructions.
www.mt.com/labweighing-software-download.
www.mt.com/labweighing-software-download. If
Checking operation
1 Start SerialPortToKeyboard (RS232)
2 Start Excel (or another application) at the PC. 3 Activate a cell in Excel.
Settings at the PC
Settings for SerialPortToKeyboard
1 Select the serial port COM for the connection with the balance.
2 Set the Baud Rate to 9600.
3 Activate Connect.
Closing the window terminates the session.
According to your selected PC-DIR. option, the displayed values will appear e.g. in the column one after the other one in the different rows.
7.2.2.2 Settings on the balance
Balance interface settings, see chapter "Interface menu".
Topic RS232
− Set PC-DIR. and select the most appropriate option for the desired weighing result.
Communication with Peripheral Devices72 Semi-Micro Balances
Topic RS.TX.E.O.L./RS E.O.L.
1 Set TAB to write into the same row (e.g. in Excel).
2 Set CR LF to write into the same column (e.g. in Excel).
3 Save changes.
7.3 Send weight value via USB to PC using LabX Direct Balance
The software LabX Direct Balance transfers weighing results and additional data such as the date and time or the measurement unit to an Excel file or any other application. The software can be ordered at METTLER TOLEDO and must be installed on a PC.
LabX Direct is installed on the PC according to the user requirements.
The METTLER TOLEDO USB* driver is installed on your PC.
The balance is connected to the PC via the USB* device interface.
Configure balance
Balance interface settings, see chapter "Interface menu".
Topic USB
1 Set HOST and select the most appropriate option for the desired weighing result (SEND.STB/SEND.CONT/
SEND.AUTO/SEND.ALL).
2 Save changes.
Transfer values
1 Start LabX Direct Balance on your PC. 2 Select the correct COM port used by the balance. 3 Configure all settings according to the user requirements.
ð The weighing result (and additional data) is transferred to the PC program either automatically. * The required settings and workflow are very similar when using an RS232 connection. In this case, a
METTLER TOLEDO USB driver does not need to be installed on the PC as a prerequisite.
Communication with Peripheral Devices 73Semi-Micro Balances
8 Maintenance
To guarantee the functionality of the balance and the accuracy of the weighing results, a number of maintenance actions must be performed by the user.
8.1 Maintenance table
Maintenance action Recommended interval Remarks
Performing an internal adjustment
Performing routine tests (sensitivity test, repeata­bility test).
METTLER TOLEDO recommends to perform at least a sensitivity test.
Cleaning Depending on the degree of pollution or your
Daily
After cleaning
After leveling
After changing the location
After cleaning
internal regulations (SOP), clean the instrument:
After every use
After change of sample
see chapter "Fully automatic adjustment FACT"
see below
see chapter "Cleaning the
balance"
8.2 Performing routine tests
There are several routine tests. Depending on your internal regulations, specific routine test must be performed by the user.
METTLER TOLEDO recommend to perform an sensitivity test after cleaning and reassembling the balance. To perform a routine test, proceed how described in chapter "Application routine test".
8.3 Cleaning
WARNING
Death or serious injury due to electric shock
Contact with parts carrying a live current can lead to injury and death. 1 Disconnect the instrument from the power supply prior to cleaning and maintenance. 2 Prevent liquid from entering the instrument, terminal or AC/DC adapter.
8.3.1 Cleaning agents
The following table presents the cleaning tools and cleaning agents recommended by METTLER TOLEDO.
Maintenance74 Semi-Micro Balances
Tools Cleaning agents
Around the balance
Balance terminal
Balance draft shield
Weighing area
Paper tissue
Lab bench
Brush
Dishwasher
Water
Acetone
Ethanol (70%)
Isopropanol
(70%)
Hydrochloric acid
(3-10 %)
Sodiumhydroxide
(0.2-1.0 M)
(R) n.a. (R) (D) R (D) (D) (D) or weighing table
Balance
(R)
R
R R R R
housing
Top housing (R)
Bottom
(R)
R
R
R R R R
R R R R
housing
Back
(R)
R
R R (R) R
housing
Feet R (R)
Terminal
Screen /
R
R R R R
(R) (R) (R) R R R
(R) R R R R R
display
Glass pans R (R) R (R) R R R R R
Glass free
R (R)
R
R R R R
pans
Non-
(R)
R (R) (R) R R R removable handles and frames
Draft ring
R (R) R
(R) (R) R (R) R element / Draft shield element
Weighing
(R)
R (R) R R R R
pan
Drip tray R (R) R
(R) (R)
Peracetic acid
(2-3%)
R
Symbol descriptions:
: Best recommendation by METTLER TOLEDO, can be used without limitation. R: Recommended by METTLER TOLEDO, can be used without limitation. (D): Depending on the material used: individual durability and resistance to acid and alkali must be evaluated. (R): Partially recommended by METTLER TOLEDO (for selected balance line only, please check the respective
operating instructions, might also depend on your internal regulations).
Not recommend.
Maintenance 75Semi-Micro Balances
8.3.2 Cleaning the glass draft shield
1
2
3
C
CAUTION
Injury due to breaking glass
Careless handling with the glass components can lead to breakage off glass and damage cuttings.
− Always proceed focused and with care.
Removing and inserting the side door glass panels and top door glass panels.
1 Always hold the 2 parallel guided glass panels
together and parallel with one hand, see illustrations.
2 Push all the glass panels as far as they will go to the
back.
3 Turn the top cover (1) to the front. 4 Pull the side door glass panels and the top door glass
panels towards the back and off. (observe the important note above)
5 Turn the two lock covers (2) on the front as far as they
will go to unlock the front glass.
6 Tilt the front glass forward and pull it out. 7 Remove draft ring. 8 Remove weighing pan. 9 Remove drip tray. After cleaning reinstall all components in reverse order
through the observance of the important notes.
Note
Side door glass panels: The guide pin must be placed in the guide slot (3).
After inserting the glass doors (side and top), close the top cover so that they can not fall out.
Front glass: The writing "front" must be show forwards.
Maintenance76 Semi-Micro Balances
8.3.3 Cleaning the balance
On/Off
NOTICE
Damage due to improper cleaning
Improper cleaning can damage the weighing cell or other essential parts. 1 Do not use any cleaning agents other than the ones specified in the Reference Manual or
Cleaning Guide.
2 Do not spray or pour liquids on the instrument. Always use a moistened lint-free cloth or
a tissue.
3 Always wipe out from inside to outside of the instrument.
Protective equipment:
Gloves
Goggles
The following procedures describe the cleaning of the weighing pan and all components.
Around the balance
− Remove any dirt or dust around the balance and avoid further contaminations.
Cleaning the balance
1 Use a lint-free cloth moistened with a mild cleaning agent to clean the surface of the balance. 2 Remove powder or dust at first with a disposable tissue. 3 Remove sticky substances with a damp lint-free cloth and a mild solvent (e.g. isopropanol or ethanol
70%).
Cleaning all removable parts
− Clean all removed part with a damp cloth or a tissue and a mild cleaning agent or clean them in a
dishwasher up to 80°C.
Note
Useful details to avoid soiling are described in the METTLER TOLEDO "SOP for Cleaning a Balance".
8.3.4 Putting into operation after cleaning
1 Reassemble the balance. 2 Check the functionality of the draft shield. 3 Press 4 Warm up the balance. Wait 1h for the acclimatization, before starting the tests. 5 Check the level status, level the balance if necessary. 6 Perform an internal adjustment. 7 Perform a routine test due to the internal regulations of your company. METTLER TOLEDO recommends to
perform an repeatability test after cleaning the balance.
8 Press ð The balance has been putting into operation and is ready to use.
See also
2 Leveling the balance 2 Adjustment (calibration) 2 Application routine test 2 Repeatability test
to switch on the balance.
to zero the balance.
}Page12
}Page13
}Page64
}Page66
Maintenance 77Semi-Micro Balances
9 Troubleshooting
Possible causes for faults and fault correction tasks are described in the following chapters. If there are faults that cannot be corrected through the instructions below, contact the METTLER TOLEDO.
9.1 Error messages
Error message Possible cause Diagnostic Remedy
NO STABILTY
WRONG ADJUSTMENT WEIGHT
REFERENCE TOO SMALL
EEPROM ERROR - PLEASE CONTACT CUSTOMER SERVICE
WRONG CELL DATA ­PLEASE CONTACT CUSTOMER SERVICE
NO STANDARD ADJUSTMENT - PLEASE CONTACT CUSTOMER SERVICE
Vibrations at the workplace.
Draft due to untight draft shield and /or open window.
The location is not suitable for weighing.
Something is touching the weighing pan.
Wrong adjustment weight. Check weight. Place correct weight on the
Reference for piece counting too small.
Data in EEPROM damaged.
Defect load cell data. Please contact your
Please contact your
Place beaker with tap water on the weighing table. Vibrations cause ripples on the water surface.
Make sure draft shield or window is closed.
Check and observe the
Check for touching parts or dirts.
Increase reference weight.
Please contact your
Protect weighing
location against vibrations (vibration absorber, etc.).
Set weighing
parameters coarser (change ENVIRON. from STABLE to
STANDARD or even UNSTABLE.
Find a different
weighing location (by agreement with customer).
Close draft shield or
window.
Set weighing
parameters coarser (change ENVIRON. from STABLE to
STANDARD or even UNSTABLE.
requirements for the location, refer to the "Selecting the location" chapter.
Remove touching parts or clean the balance.
weighing pan.
METTLER TOLEDO-Support representative.
METTLER TOLEDO-Support representative.
METTLER TOLEDO-Support representative.
Troubleshooting78 Semi-Micro Balances
Error message Possible cause Diagnostic Remedy
PROGRAM MEMORY DEFECT - PLEASE CONTACT CUSTOMER SERVICE
TEMP SENSOR DEFECT ­PLEASE CONTACT CUSTOMER SERVICE
WRONG LOAD CELL BRAND - PLEASE CONTACT CUSTOMER SERVICE
WRONG TYPE DATA SET ­PLEASE CONTACT CUSTOMER SERVICE
Battery backup lost
INITIAL ZERO RANGE EXCEEDED
BELOW INITIAL ZERO RANGE
MEM FULL
FACTOR OUT OF RANGE
STEP OUT OF RANGE
OUT OF RANGE
Please contact your
METTLER TOLEDO-Support representative.
Temperature sensor of load cell defect.
Wrong load cell installed. – Please contact your
Wrong type data set. Please contact your
Backup battery is empty. This battery ensures that the date and time are not lost when the balance is disconnected from power.
Wrong weighing pan. Pan is missing. Pan is not empty.
Wrong weighing pan. Pan is missing. Pan is not empty.
Memory full. Clear the memory by
Factor is outside the allow range.
Step is outside the allow range.
Sample weight is outside the allow range.
Please contact your
METTLER TOLEDO-Support representative.
METTLER TOLEDO-Support representative.
METTLER TOLEDO-Support representative.
Connect the balance to the power supply for charging the battery (full capacity after 2 days of charging).
Check weighing pan. Mount correct weighing
Check weighing pan. Mount correct weighing
Select a new factor.
Select a new step.
Unload the pan and load a
Battery must be recharged. Please contact METTLER TOLEDO-Support represen­tative.
pan or unload weighing pan.
pan.
finishing all applications where a measurement is ongoing.
new sample weight.
Troubleshooting 79Semi-Micro Balances
9.2 Error symptoms
Error symptoms Possible cause Diagnostic Remedy
Display is dark Instrument is switched off. – Switch on the instrument.
Power plug not connected. Check Connect power cable to
power supply.
Membrane keypad does not function
Power supply not connected to balance.
Power supply is faulty. Check/test Replace power supply.
Wrong power supply. Check that input data on
Connector socket on balance is corroded or faulty.
Display is faulty. Replace display. Please contact your
Keypad faulty. Replace keypad. Please contact your
Check Connect power supply.
Use proper power supply. type plate match the power supply values.
Check Please contact your
METTLER TOLEDO-Support
representative.
METTLER TOLEDO-Support
representative.
METTLER TOLEDO-Support
representative.
Troubleshooting80 Semi-Micro Balances
Error symptoms Possible cause Diagnostic Remedy
The value drifts into plus or minus
Room, environment not suitable.
Environmental recom­mendations
Windowless, non air-
conditioned room, e.g. basement.
Only one person in the
weighing room.
Sliding doors.
Standard doors cause pressure changes.
No draft in weighing
room (check with suspended threads).
No air conditioning
(temperature oscillates, draft).
Acclimatize balance,
take dummy measurements.
Instrument uninter-
ruptedly connected to the power supply (24h per day).
Direct sunlight or other heat source.
Is any sun shade (blinds, curtains, etc.) available?
Select location according to chapter "Selecting the location" (customer responsibility).
Weighing sample absorbs moisture or evaporates moisture.
Is the weighing result
with a test weight stable?
Sensitive weighing
Use aids.
Cover weighing
sample.
samples e.g. paper, cardboard, wood, plastic, rubber, liquids.
Weighing sample is electrostatically charged.
Is the weighing result
with a test weight stable?
Sensitive weighing
Increase air humidity
in weighing chamber (45% - 50%).
Use ionizer.
samples e.g. plastic, powder, insulating materials.
Weighing sample is hotter or colder than the air in the weighing chamber.
Instrument has not yet reached thermal equilibrium.
Weighing operation with test weight does not show this effect.
Was there a power
outage?
Was power supply
disconnected?
Bring weighing sample to room temperature before weighing.
Acclimatize instrument
for at least 1 hour. Depending on climatic conditions, extend this period accordingly.
Instrument switched on
for at least 1 hour, refer to the "General data" chapter.
Troubleshooting 81Semi-Micro Balances
Error symptoms Possible cause Diagnostic Remedy
Display shows overload or underload
The weight on the weighing pan exceeds the
Check weight. Reduce the weight on the
weighing pan.
weighing capacity of the instrument.
Wrong weighing pan. Slightly lift or press
Use proper weighing pan. weighing pan. The weight display appears.
No weighing pan. Install weighing pan.
Incorrect zero point at switch-on.
Switch off balance.
Disconnect and
reconnect power cable.
Display flashes 0.00000 Loose cables. Check all cable
connections.
Connect all cables.
Please contact your
METTLER TOLEDO-Support
representative if the
problem persists.
Taring not possible Vibrations at the
workplace.
Display unstable. Press Tare again.
Place beaker with tap water on the weighing table. Vibrations cause ripples on the water surface.
Protect weighing
location against vibrations (vibration absorber, etc.).
Set weighing
parameters coarser (change ENVIRON. from STABLE to
STANDARD or even UNSTABLE.
Find a different
weighing location (by agreement with customer).
9.3 Status messages/Status icons
Status messages are displayed by means of small icons. The status icons indicate the following:
Icon Status description Diagnostic Remedy
Service due.
See menu topic SERV.ICON in chapter "Description of menu
Please contact your METTLER TOLEDO-Support representative.
topic" -> "Advanced menu".
9.4 Putting into operation after correction of fault
After correcting the fault, perform the following steps to put the balance into operation:
Ensure that the balance is completely reassemble and cleaned.
Reconnect the balance to the power supply.
Troubleshooting82 Semi-Micro Balances
10 Technical Data
10.1 General data
Standard power supply
AC/DC adapter: Primary: 100 - 240 V~ 0.8 A, 50 – 60 Hz 60 - 80 VA
Cable for AC/DC adapter: 3-core, with country-specific plug
Secondary: 12 V DC, 2.5 A LPS (with electronic overload protection)
Polarity:
output
Balance power consumption: 12 V DC, 0.3 A
with a current limited SELV (Safety Extra Low Voltage)
Protection and standards
Overvoltage category: II
Degree of pollution: 2
Protection: Protected against dust and water
Standards for safety and EMC: See Declaration of Conformity
Range of application: For use in closed interior rooms only
Environmental conditions
Height above mean sea level: Up to 4000 m
Ambient temperature: 5 - 40 °C
Relative air humidity: Max. 80% up to 31°C, linearly decreasing to 50% at 40°C,
noncondensing
Warm-up time:
At least 60 minutes after connecting the balance to the power supply. When switched on from standby, the instrument is ready for operation immediately.
Materials
Housing/Terminal: Die-cast aluminum / Plastic (PA12)
Weighing pan: Stainless steel X2CrNiMo 17-12-2 (1.4404)
Draft shield element: Stainless steel X2CrNiMo 17-12-2 (1.4404)
Draft shield: Plastic (PBT), glass
Protective cover: Plastic (PET)
Technical Data 83Semi-Micro Balances
10.2 Model-specific data
Limit Values
Maximum capacity 120 g 120 g 220 g
Nominal load 100 g 100 g 200 g
Readability 0.01 mg 0.1 mg 0.1 mg
Maximum capacity in fine range 42 g 82 g
Readability in fine range 0.01 mg 0.01 mg
Repeatability (at nominal load) 0.04 mg 0.08 mg 0.08 mg
Repeatability (5% load) 0.02 mg 0.02 mg 0.02 mg
Linearity deviation 0.1 mg 0.15 mg 0.2 mg
Eccentricity deviation (test load) 0.3 mg (50 g) 0.3 mg (50 g) 0.3 mg (100 g)
Sensitivity offset (at nominal load) 1)0.4 mg 0.4 mg 0.8 mg
Sensitivity temperature drift
Typical values
Repeatability (5% load) 0.015 mg 0.015 mg 0.015 mg
Linearity deviation (sd) 0.03 mg 0.05 mg 0.06 mg
Eccentricity deviation (sd) (test load)
Sensitivity offset (at nominal load) 0.2 mg 0.2 mg 0.4 mg
Minimum weight (according to USP, 5% load)
Minimum weight (U=1%, k=2, 5% load)
Settling time 1.5 s 1.5 s 1.5 s
Settling time in fine range 3 s 3 s
Dimensions & other specifications
Balance dimensions (W × D × H) 247 × 358 × 331 mm 247 × 358 × 331 mm 247 × 358 × 331 mm
Weighing pan dimensions ø 80 mm ø 80 mm ø 80 mm
Usable height of draft shield 235 mm 235 mm 235 mm
Weight of the balance 6.8 kg 6.8 kg 6.8 kg
Weights for routine testing
Weights / OIML Class 5 g (E2) / 100 g (F2) 5 g (E2) / 100 g (F2) 10 g (E2) / 200 g
Weights / ASTM Class 5 g (ASTM 1)/ 100 g
sd = Standard deviation
2)
In the temperature range 10…30°C
MS105 MS105DU MS205DU
2)
0.0002%/°C 0.0002%/°C 0.0002%/°C
0.1 mg (50 g) 0.1 mg (50 g) 0.1 mg (100 g)
30 mg 30 mg 30 mg
3 mg 3 mg 3 mg
(F2)
(ASTM 1)
5 g (ASTM 1)/ 100 g (ASTM 1)
1)
After adjustment with internal weight
10 g (ASTM 1)/ 200 g (ASTM 1)
Technical Data84 Semi-Micro Balances
MS204
Limit Values
Maximum capacity 220 g
Nominal load 200 g
Readability 0.1 mg
Maximum capacity in fine range
Readability in fine range
Repeatability (at nominal load) 0.09 mg
Repeatability (5% load) 0.07 mg
Linearity deviation 0.2 mg
Eccentricity deviation (test load) 0.3 mg (100 g)
Sensitivity offset (at nominal load) 1)0.8 mg
Sensitivity temperature drift
2)
0.0002%/°C
Typical values
Repeatability (5% load) 0.05 mg
Linearity deviation (sd) 0.06 mg
Eccentricity deviation (sd) (test
0.1 mg (100 g)
load)
Sensitivity offset (at nominal load) 0.4 mg
Minimum weight (according to USP,
100 mg
5% load)
Minimum weight (U=1%, k=2, 5%
10 mg
load)
Settling time 1.5 s
Settling time in fine range
Dimensions & other specifications
Balance dimensions (W × D × H) 247 × 358 × 331 mm
Weighing pan dimensions ø 80 mm
Usable height of draft shield 235 mm
Weight of the balance 6.8 kg
Weights for routine testing
Weights / OIML Class 10 g (E2) / 200 g (F2)
Weights / ASTM Class 10 g (ASTM 1)/ 200 g (ASTM 1)
sd = Standard deviation
2)
In the temperature range 10…30°C
1)
After adjustment with internal weight
Technical Data 85Semi-Micro Balances
10.3 Dimensions
C
174
358
205
247
255
234
126
80.2
277
331
77
80.6
ø 80
134
12
14
157
147
140
224
Technical Data86 Semi-Micro Balances
10.4 Interface specification
15
9 6
IN
GND
OUT
HAND SHAKE
IN
DATA
OUT
RxD
TxD
CTS
RTS
OUT
POWER SUPPLY
+12V
2nd display mode only
10.4.1 RS232C interface
Each balance is equipped with an RS232C Interface as standard for the attachment of a peripheral device (e.g. printer or computer).
Schematic Item Specification
Interface type Voltage interface according to EIA RS-232C/
DIN66020 CCITT V24/V.28)
Max. cable length 15 m
Signal level Outputs:
+5 V … +15 V (RL = 3–7 kΩ) –5 V … –15 V (RL = 3–7 kΩ) Inputs: +3 V … +25 V –3 V … -25 V
Connector Sub-D, 9-pole, female
Operating mode Full duplex
Transmission mode Bit-serial, asynchronous
Transmission code ASCII
Baud rates 600, 1200, 2400, 4800, 9600, 19200,
38400 (software selectable)
Bits/parity 7-bit/none, 7-bit/even, 7-bit/odd, 8-bit/none
(software selectable)
Stop bits 1 stop bit
Handshake None, XON/XOFF, RTS/CTS (software selectable)
End-of-line <CR><LF>, <CR>, <LF> (software selectable)
Power supply for 2nd display
+ 12 V, max 40 mA (software selectable, 2nd display mode only)
Technical Data 87Semi-Micro Balances
10.4.2 USB device
1 VBUS (+5 VDC) 2 D– (Data –) 3 D+ (Data +) 4 GND (Ground)
Shield Shield
3.5 mm
GND
Do not connect!
Switching contact
Each balance is equipped with a USB device interface as standard for the attachment of a peripheral device (e.g. computer).
Note
This interface is not able to communicate with a printer.
Schematic Item Specification
10.4.3 Aux connection
Standard In conformity with USB specifications revision
1.1
Speed Full-speed 12 Mbps (requires shielded cable)
Function CDC (Communication Device Class) serial port
emulation
Power usage Suspended device: Max 10 mA
Connector Type B
You can connect the METTLER TOLEDO "ErgoSens" or an external switch to socket Aux. This allows you to start functions such as taring, zeroing or printing.
External connection
Connector:
3.5 mm stereo jack connector Electrical data:
Max. voltage 12 V Max. current 150 mA
10.4.4 MT-SICS interface commands and functions
Many of the instruments and balances used have to be able to integrate into a complex computer or data acquisition system.
To easily integrate a balance into a system and utilize its capacity to the full extent, most balance functions are also available as corresponding commands via the data interface.
All new METTLER TOLEDO balances launched on the market support "METTLER TOLEDO Standard Interface Command Set" (MT-SICS). The commands available depend on the functionality of the balance.
For further information, please contact your METTLER TOLEDO representative.
Refer to the MT-SICS Reference Manual which can be downloaded from the Internet at
Technical Data88 Semi-Micro Balances
www.mt.com/library
u
11 Disposal
In conformance with the European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties, the content of this regulation must also be related.
Disposal 89Semi-Micro Balances
12 Accessories and Spare Parts
12.1 Accessories
Description Part No.
Printers
RS-P25/01 (EMEA) printer with RS232 connection to instrument
11124300
Paper roll (length: 20 m), set of 5pcs 00072456 Paper roll (length: 13 m), self-adhesive, set of
3pcs Ribbon cartridge, black, set of 2pcs 00065975
11600388
RS-P25/02 (Asia-Pacific) printer with RS232C connection to instrument
Paper roll (length: 20 m), set of 5pcs 00072456 Paper roll (length: 13 m), self-adhesive, set of
3pcs Ribbon cartridge, black, set of 2pcs 00065975
RS-P25/03 (Northern America) printer with RS232 connection to instrument
Paper roll (length: 20 m), set of 5pcs 00072456 Paper roll (length: 13 m), self-adhesive, set of
3pcs Ribbon cartridge, black, set of 2pcs 00065975
RS-P26/01 (EMEA) printer with RS232 connection to instrument (with date and time)
Paper roll (length: 20 m), set of 5pcs 00072456 Paper roll, self-adhesive (length: 13 m), set of
3pcs Ribbon cartridge, black, set of 2pcs 00065975
RS-P26/02 (Asia-Pacific) printer with RS232C connection to instrument (with date and time)
Paper roll (length: 20 m), set of 5pcs 00072456 Paper roll (length: 13 m), self-adhesive, set of
3pcs Ribbon cartridge, black, set of 2pcs 00065975
11124310
11600388
11124320
11600388
11124303
11600388
11124313
11600388
RS-P26/03 (Northern America) printer with RS232 connection to instrument (with date and time)
Paper roll (length: 20 m), set of 5pcs 00072456 Paper roll, self-adhesive (length: 13 m), set of
3pcs Ribbon cartridge, black, set of 2pcs 00065975
Accessories and Spare Parts90 Semi-Micro Balances
11124323
11600388
RS-P28/01 (EMEA) printer with RS232 connection to instrument (with date, time and applications)
Paper roll (length: 20 m), set of 5pcs 00072456 Paper roll, self-adhesive (length: 13 m), set of
3pcs Ribbon cartridge, black, set of 2pcs 00065975
11124304
11600388
RS-P28/02 (Asia-Pacific) printer with RS232C connection to instrument (with date, time and applications)
Paper roll (length: 20 m), set of 5pcs 00072456 Paper roll (length: 13 m), self-adhesive, set of
3pcs Ribbon cartridge, black, set of 2pcs 00065975
RS-P28/03 (Northern America) printer with RS232 connection to instrument (with date, time and applications)
Paper roll (length: 20 m), set of 5pcs 00072456 Paper roll, self-adhesive (length: 13 m), set of
3pcs Ribbon cartridge, black, set of 2pcs 00065975
P-52RUE dot matrix printer RS232C, USB and Ethernet connections, simple print-outs
Paper roll (length: 20 m), set of 5pcs 00072456 Paper roll (length: 13 m), self-adhesive, set of
3pcs Ribbon cartridge, black, set of 2pcs Dimension of the label 56 × 18mm
P-56RUE thermal printer with RS232, USB and ethernet connections, simple print-outs, date and time
Paper roll, white (length: 27 m), set of 10pcs 30094723 Paper roll, white, self-adhesive (length: 13 m), set
of 10pcs
11124314
11600388
11124324
11600388
30237290
11600388
00065975
30094673
30094724
P-58RUE thermal printer with RS232, USB and ethernet connections, simple print-outs, date and time, label printing, balance applications: statistics, formulation, totaling,
Cables for RS232 interface
RS9 – RS9 (m/f): connection cable for PC, length = 1 m 11101051
30094674
Paper roll, white (length: 27 m), set of 10pcs 30094723 Paper roll, white, self-adhesive (length: 13 m), set
of 10pcs Paper roll, white, self-adhesive labels (550 labels),
set of 6pcs Dimension of the label 56 × 18mm
Accessories and Spare Parts 91Semi-Micro Balances
30094724
30094725
RS9 – RS25 (m/f): connection cable for PC, length = 2m 11101052
RS232 - USB converter cable – cable with converter to connect a balance (RS232) to a USB port
Cables for USB interface
USB (A –B) connection cable, length=1m 30241476
Cable replacement (wireless)
Bluetooth RS232 serial adapter ADP-BT-S for wireless connection between:
Balance and PC
Compatible with following balance lines: MS, MS-L, MS-TS, ML-T, ME-T, ME, PL-E, JP, JS, JE
Printer and balance
Compatible with P-50 printers and the following balance lines:
XPR, XSR (additional Bluetooth USB adapter 30416089 required for balance)
64088427
30086494
Density determination
Bluetooth RS232 serial adapter set ADP-BT-P for wireless
30086495 connection between printer and balance. Compatible with printers P-50 and the following balance models (SW V2.20 or higher required): JP, JS, MS, MS-S, MS-L, ML, MS-TS, ML-T, ME-T, PHS, PL-E.
2 Bluetooth RS232 serial adapter paired (slave/master)
1 MT-DB9 male to female connector
1 MT-DB9 male to male connector
Density kit MS-DNY-54 for Semi-Micro balances 30004077
Glass beaker, height 100 mm, ø 60 mm 00238166
Accessories and Spare Parts92 Semi-Micro Balances
Pipette check
METTLER TOLEDO
000000
Auxiliary displays
Sinker for density of liquids in conjunction with density kit 00210260 Calibrated (sinker + certificate) 00210672 Recalibrated (new certificate) 00210674
Calibrated thermometer with certificate 11132685
SmartCheck Trap 50 ml*, > 20 - 2000 µl
30215436
* 50 ml capacity with 0.01 mg readability only in combination with MS105. For other models either readability or capacity is limited.
RS232 auxiliary display AD-RS-M7 12122381
External switches
Protective covers
Anti-theft devices
ErgoSens, optical sensor for hands-free operation 11132601
Auxiliary Footswitch with selectable function for balances 11106741
Protective cover for semi micro balances 30006615
Anti-theft cable with lock 11600361
Accessories and Spare Parts 93Semi-Micro Balances
Software
Various
LabX direct balance (simple data transfer) 11120340
AC/DC adapter (without power cable) 100–240VAC, 0.8A, 50/60Hz, 12VDC2.5A
Country-specific 3-Pin power cable with grounding conductor.
Power cable AU 00088751 Power cable BR 30015268 Power cable CH 00087920 Power cable CN 30047293 Power cable DK 00087452 Power cable EU 00087925 Power cable GB 00089405 Power cable IL 00225297 Power cable IN 11600569 Power cable IT 00087457 Power cable JP 11107881 Power cable TH, PE 11107880 Power cable US 00088668 Power cable ZA 00089728
11107909
Adjustment weights
OIML / ASTM weights (with calibration certificate) see www.mt.com/weights
Accessories and Spare Parts94 Semi-Micro Balances
12.2 Spare parts
C
6
7
8
9
10
11
17
12
5
2
1
13
14
15
4
3
16
Order no. Designation Remarks
11142253 Level cover
1
30003679 Front glass panel
2
11142229 Front glass lock, right
3
11142228 Front glass lock, left
4
11142244 Top cover
5
30003678 Side door front, left Including: Handle
6
11133079 Door back left
7
11133082 Door top front Including: Handle
8
11133081 Door top back
9
11133077 Door back right
10
30003677 Side door front, right Including: Handle
11
30003777 Weighing pan
12
11142206 Draft ring
13
30003778 Drip tray
14
11122623 Plastic cap
15
12104936 Weighing below balance cap
16
30104835 Leveling foot
17
Accessories and Spare Parts 95Semi-Micro Balances
Accessories and Spare Parts96 Semi-Micro Balances
Index
A
AC/DC adapter 83 Accessories 90 Adjusting 13 Adjustment 15, 26, 27 Advanced menu 22 Application density 55 Application diagnostics 30, 66 Application formulation 46 Application icons 9 Application multiplication factor weighing 51 Application percent weighing 42 Application piece counting 39 Application pipette check 60 Application Routine test 64 Application statistics 44 Application totaling 49 Assign application 29, 30 Auto print 32 Automatic adjustment 14 Automatic shutoff 28 Automatic zero setting 28 Autozero 28 Aux connection 88 Average (Statistics) 44
B
Backlight 28 Balance history 67 Balance information 68 Basic menu 22, 25 Baudrate 34 Beep 26 Below-the-balance weighing 20 Bit/Parity 34
C
Calibration 26, 27 Calibration history 68 Cancel 24 Change settings 23, 24 Char set 30, 35, 37 Cleaning 74 Closing the menu 24 Control limit 65
Customer fine adjustment 15, 27
D
Data communication format 33, 36 Date 25 Date format 27 Density 55 Density kit 55 Density of distilled water 59 Density table for ethanol 59 Diagnose 30 Diagnostics 66 Diagnostics application 30 Dimensions 86 Display 28, 31 Display panel 9 Display test 66 Disposal 89 Distilled water 59 Dosing 26 Draft shield 11, 74, 76 DualRange 18
E
End of line 35, 37 Environment 26 Environmental conditions 83 ErgoSens 37, 88 Ethanol 59 External key 37
F
FACT 14, 27 Factor weighing 51 Fill up 47 Foot switch 88 Formulation 46 Fully automatic adjustment 14, 27 Function fill up 47 Function PC-Direct 71
G
Good weighing practice 64 GWP 64, 65
H
Handshake 34
Index 97Semi-Micro Balances
Header 32 Host 31, 35
I
N
Net 17 Numerical values 24
Icons 9 Input principle 24 Installing the components 11 Interface
MT-SICS 88
USB device 88 Interface menu 23, 31 Interface RS232C 31, 87 Interface USB device 35 Internal weight 14 Interval 37 Introduction 3
K
Key assign 29, 30 Key beep 26 Key functions 7 Key test 67
L
Language 29 Level indicator 13 Leveling 12 Leveling the balance 12 Line feed 32 Liquid 56 Liquids 57 Location 10
M
Main Menu 24 Manual adjustment with internal weight 14 Materials 83 Menu 22, 24 Menu advanced 22 Menu basic 22, 25 Menu interface 23, 31 Menu operation 23 Menu protection 25 Menu topic 23, 24 Motor test 67 MT-SICS 88 Multiplication factor weighing 51
O
Operating temperature 12 Operation keys 7
P
PC-DIR 31 PC-Direct 71 Percent weighing 42 Performing a simple weighing 17 Piece counting 39 Pipette check 60 Print 19 Printer 31 Protect 25 Protection and standards 83 Protocol trigger 27
R
Recall 18, 28 Repeatability test 66 Reset 26 Routine test 64 RS232C interface 31, 87
S
Safety information 4
Signal words 4
Warning symbols 4 Saving settings 24 Select menu 23 Select menu topic 23 Selecting the location 10 Service 30, 74 Service date reset 30 Service icon 30 Service provider information 69 Service reminder 30 Shutoff 19, 28 Signal words 4 Signature line 32 Single 32 Sinker 55 Solids 56 SOP 65
Index98 Semi-Micro Balances
Loading...