
1
Installation
1. Unpack the meter, power supply, electrode,
electrode arm, buffer sachets, Guide Book and
calibration certificate. Keep the calibration
certificate in a safe place.
2. Check the power supply unit is the correct
voltage for your local supply. If it is not, contact
your distributor.
3. Attach the electrode arm to the left or righthand
side of the meter:
a. Turn the meter upside down and seat the
dome of the electrode arm base into the left
or righthand meter recess.
b. Insert and tighten the fixing screw.
c. Turn the meter and base the right way up
and locate the electrode arm onto the base
fixing post. Press down to snap into place.
Input and Output Connections
MP220 Rear Panel
power supply
unit socket
combination
electrode socket
automatic temperature
compensation probe socket

2
Display
stability/auto endpoint
indicator
electrode
condition
shown
here
prog. menu icons shown
here (and data transfer
icon - MP225 only)
temperature,
electrode
offset/slope
shown here
memory,
cal and
error
number
shown here
automatic or
manual
temperature
compensation
indicator
power supply
unit socket
pH or
combination
electrode socket
data output
socket
recorder output
socket
reference
electrode
socket
Input and Output Connections (continued)
MP225 Rear Panel
1. Disconnect the shorting clip from the pH socket
and retain it by clipping it over the socket.
2. Connect the electrode to the pH socket. If you
are using an electrode incorporating ATC
connect the other lead to the ATC socket.
3. If you are using a separate ATC probe connect it
to the ATC socket.
4. Connect the power supply unit to the DC socket.
automatic temperature
compensation probe
socket

3
3. Repeat steps 1 and 2 using other buffers for a
2 point (or 3 point - MP225 only) calibration.
Electrode Condition Indicator
95.0 - 105.0% 90.0 - 94.9% 85.0 - 89.9%
electrode in electrode electrode
good needs needs
condition cleaning conditioning
n When you press Cal the pH buffer you selected for
cal 1 (Program Menu) is displayed for 2 seconds, and
then the meter starts measuring. If you want to use
another pH buffer (choice of 3) press Cal again. Press
Cal repeatedly to exit the calibration routine.
n The decimal point flashes during calibration
measurement. When the electrode output
has stabilized the stability indicator appears.
Manual endpoint - press Read to endpoint.
Auto endpoint - the meter automatically endpoints.
n When a calibration has endpointed you can press
Mode to display the absolute mV value and temperature of the buffer. The display reverts to the buffer and
E0 values, or buffer and slope values for a 2 point (or 3
point - MP225 only) calibration after 2 seconds.
buffer
Calibrating
1. 1-point calibration
Example
Use auto endpoint or
to freeze the reading.
2. Rinse the electrode and blot dry
to return to sample measurement, or
continue with step 3 for further cal points
↓ 2 seconds

4
n If you press Mode to select pH or mV mode the meter
begins measuring immediately. You will not need to
press Read to start the reading.
n The decimal point flashes during sample
measurement. When the electrode output
has stabilized the stability indicator appears.
Manual endpoint - press Read to endpoint.
Auto endpoint - the meter automatically endpoints.
n If you have the RS232 cable connected to the Data
socket (MP225 only), sample data is transmitted at
endpoint.
Measuring Samples
1. Select pH or mV mode
2. Measure sample
Use auto endpoint or
to freeze the reading.
3. Rinse, blot and store the electrode.

5
Operating Hints
n Remove the wetting cap from the end of the
electrode and the rubber cap from the fill hole (if
fitted) before using the electrode.
n Calibrate using buffers with values that bracket
that of the sample. With a new pH electrode, or
after maintenance, we recommend you use a
buffer close to pH 7 for the first calibration point.
n For greatest accuracy, buffers and samples
should be at the same temperature.
n Do not use solutions after the expiry date.
n When transferring the electrode from one
solution to another, rinse it with distilled water
and blot dry with tissue paper - do not wipe the
electrode as this may cause polarization and
slow response.
n When you select the calibration buffers
(Program Menu), set the buffer you will use
most frequently for 1-point calibrations as cal 1
and the one you will use most frequently for the
second calibration point as cal 2, and for the
MP225, the one you will use most frequently for
the third calibration point as cal 3.