Mettler Toledo MP220, MP225 Installation Manual

Page 1
1
Installation
1. Unpack the meter, power supply, electrode, electrode arm, buffer sachets, Guide Book and calibration certificate. Keep the calibration certificate in a safe place.
2. Check the power supply unit is the correct voltage for your local supply. If it is not, contact your distributor.
3. Attach the electrode arm to the left or righthand side of the meter:
a. Turn the meter upside down and seat the
dome of the electrode arm base into the left or righthand meter recess.
b. Insert and tighten the fixing screw. c. Turn the meter and base the right way up
and locate the electrode arm onto the base fixing post. Press down to snap into place.
Input and Output Connections
MP220 Rear Panel
power supply unit socket
combination electrode socket
automatic temperature compensation probe socket
Page 2
2
Display
stability/auto endpoint indicator
electrode condition shown here
prog. menu icons shown here (and data transfer icon - MP225 only)
temperature, electrode offset/slope shown here
memory, cal and error number shown here
automatic or manual temperature compensation indicator
power supply unit socket
pH or combination electrode socket
data output socket
recorder output socket
reference electrode socket
Input and Output Connections (continued)
MP225 Rear Panel
1. Disconnect the shorting clip from the pH socket and retain it by clipping it over the socket.
2. Connect the electrode to the pH socket. If you are using an electrode incorporating ATC connect the other lead to the ATC socket.
3. If you are using a separate ATC probe connect it to the ATC socket.
4. Connect the power supply unit to the DC socket.
automatic temperature compensation probe socket
Page 3
3
3. Repeat steps 1 and 2 using other buffers for a 2 point (or 3 point - MP225 only) calibration.
Electrode Condition Indicator
95.0 - 105.0% 90.0 - 94.9% 85.0 - 89.9% electrode in electrode electrode good needs needs condition cleaning conditioning
n When you press Cal the pH buffer you selected for
cal 1 (Program Menu) is displayed for 2 seconds, and then the meter starts measuring. If you want to use another pH buffer (choice of 3) press Cal again. Press Cal repeatedly to exit the calibration routine.
n The decimal point flashes during calibration
measurement. When the electrode output has stabilized the stability indicator appears.
Manual endpoint - press Read to endpoint. Auto endpoint - the meter automatically endpoints.
n When a calibration has endpointed you can press
Mode to display the absolute mV value and tempera­ture of the buffer. The display reverts to the buffer and E0 values, or buffer and slope values for a 2 point (or 3 point - MP225 only) calibration after 2 seconds.
buffer
Calibrating
1. 1-point calibration
Example
Use auto endpoint or to freeze the reading.
2. Rinse the electrode and blot dry
to return to sample measurement, or continue with step 3 for further cal points
2 seconds
Page 4
4
n If you press Mode to select pH or mV mode the meter
begins measuring immediately. You will not need to press Read to start the reading.
n The decimal point flashes during sample
measurement. When the electrode output has stabilized the stability indicator appears.
Manual endpoint - press Read to endpoint. Auto endpoint - the meter automatically endpoints.
n If you have the RS232 cable connected to the Data
socket (MP225 only), sample data is transmitted at endpoint.
Measuring Samples
1. Select pH or mV mode
2. Measure sample
Use auto endpoint or to freeze the reading.
3. Rinse, blot and store the electrode.
Page 5
5
Operating Hints
n Remove the wetting cap from the end of the
electrode and the rubber cap from the fill hole (if fitted) before using the electrode.
n Calibrate using buffers with values that bracket
that of the sample. With a new pH electrode, or after maintenance, we recommend you use a buffer close to pH 7 for the first calibration point.
n For greatest accuracy, buffers and samples
should be at the same temperature.
n Do not use solutions after the expiry date. n When transferring the electrode from one
solution to another, rinse it with distilled water and blot dry with tissue paper - do not wipe the electrode as this may cause polarization and slow response.
n When you select the calibration buffers
(Program Menu), set the buffer you will use most frequently for 1-point calibrations as cal 1 and the one you will use most frequently for the second calibration point as cal 2, and for the MP225, the one you will use most frequently for the third calibration point as cal 3.
Page 6
6
A
A
R
R
R
Using Auto Endpoint
The auto read feature can be used in pH and mV mode to automatically determine a stable endpoint during calibrations and sample measurements.
Press . The display shows the auto endpoint indicator . The display will freeze automatically when a stable endpoint is reached. To manually freeze the display press Read. To turn auto endpoint off press again.
Using the Memory
Entering a Reading into Memory
The MP220 can store 10 endpointed results and the MP225 can store 20.
Press when the measurement has endpointed. M 1 is displayed (or M 2 to M 20 if readings have already been stored).
M 10 (MP220) or M 20 (MP225) indicates the memory is full.
Recalling Memory
You can only recall stored memories if the current measurement has endpointed.
Press - the last stored memory is displayed. Press or to scroll through memories. RM 1 to
RM 20 indicates which memory is being displayed. M 0 indicates no memories stored.
Clearing the Memories
Press then press or until M C is displayed. Press Mode to clear the memories, M 0 indicates
the memories have been cleared. (Press Read to exit without clearing the memories).
b697
M
M
M
M
Page 7
7
Using Continuous Data Transfer Mode
(MP225 only)
In continuous data transfer mode readings are sent to the data output approximately every second. If the measurement endpoints (manually, or using auto endpoint) data transfer stops. If you recall memories all stored readings are output, (from the first to the last stored).
To select continuous data transfer mode press and hold for 2 seconds.
n To maintain continuity of sample readings
calibration data is only output at endpoint.
R
Resetting Sample ID (MP225 only)
The sample identity number is sent to the serial output and increments 1 every time a sample measurement endpoints.
To reset the sample ID number to 1, press and hold Read for 2 seconds. The sample ID icon flashes three times to show the number has been reset.
Recalling Calibration Data (MP225 only)
You can only recall calibration data if the current measurement has endpointed - press Read if necessary.
Press and hold Cal for 2 seconds. The display shows the date and time of the last calibration, the buffer value, the mV offset (E
0
) for 1-point calibrations, and the slope value (%) for 2 (3) point calibrations.
Page 8
8
MTC
The MP225 has a preset reminder interval of 1 hour. Change the interval using and
. Press Mode to enter the value and move
on. Time and Date (MP225 only)
Time and date are displayed during calibration recall, and will be sent to the serial output. Time is shown on the main display, the date is shown on the lower part of the display. Set the time and date using
and . Press Mode to enter the value
and move on.
Manual Temperature Compensation
(MP220 and MP225) You can enter temperature manually
between –5.0 and 105.0°C. (An ATC probe will override manual compensation.) The meters have a preset temperature of 25°C. Use and to change temperature. Press Mode to enter the value and move on.
Program Menu
The Program Menu allows you to set manual temperature compensation, pH calibration buffers and, for the MP225 only, calibration reminder and date and time.
You can only enter the Program Menu if the current measurement has endpointed - press Read if necessary.
Press and hold the Mode key for 2 seconds to access the Program Menu - appears.
Press Mode to scroll through the options and press and to change the value. Press Read to exit the Program Menu at any time. If you press Read when a value is flashing that value will not be entered.
Calibration Reminder (MP225 only) The calibration reminder interval can be
set, in hours, between 0 and 4 (where 0 = no reminder). When the selected time has elapsed the calibration reminder symbol will appear.
M
R
M
R
M
R
M
R
Page 9
9
cal 1
cal 2
cal 3
Program Menu (continued)
pH Buffer Selection (MP220 and MP225)
The buffers are grouped in sets (b = 1, b = 2, b = 3); select your required set first using and : set 1 = 7.00, 4.00, 10.01, 1.68 set 2 = 7.00, 4.01, 9.21, 2.00, 11.00 set 3 = 6.87, 4.01, 9.18, 1.68
Press Mode to enter the set and move on. NOTE: Check buffer set 2 is selected to use
the buffers supplied Choose three buffers (cal 1, cal 2, cal 3)
from the set for calibration. For ease of use, set the buffers in the order that you will use them.
For example: Set 2 (factory settings) cal 1 = 7.00, cal 2 = 4.01, cal 3 = 9.21
(2.00 and 11.00 not selected).
These can be rearranged in any order using
and . Press Mode to enter the value
and move on. Press Read to exit the Program Menu.
M
R
M
R
b697
Page 10
10
Interfacing (MP225 only)
Serial Output
The MP225 can interface with computers, printers (GA42), Acquire Data Acquisition software and other RS232 compatible devices via the RS232 cable. The maximum source or sink current available is ± 10mA.
Pin Signal Pin Signal
1 device enable (link to pin 5) 6 not used 2 RS232 Tx (signal transmit) 7 not used 3 not used 8 not used 4 not used 9 not used 5 protective ground (0V)
Data Socket
Type of communication - uni directional, baud rate 2400, data format - 7 data bits : 1 stop bit : even parity
n Data is sent to the Data socket at measurement
endpoint (manual or automatic), calibration recall and memory recall.
n To use the output to monitor continually, you need to
select continuous data transfer mode.
Recorder Output
An output is available for both modes. The polarity of the output is the same as the polarity of the input signal.
Approximate Recorder Outputs
pH - Output follows display and provides approxi­mately 60 mV output/pH unit. This value varies with changes in electrode slope and temperature.
mV - Output follows display, i.e. output varies from –1999 to 1999 mV.
Recorder output is not available for temperature.
Rec Socket Polarity
protective ground (0V)
signal transmit
0V
data output
51
96
Page 11
11
Problem Solving
Err 1 - offset value (E0) out of range
Check correct buffer is used. Check mV reading for pH 7 buffer is 0 ± 30 mV. If it
is not, clean or replace the electrode. Err 2 - slope out of range
Slope less than 85%, or not calculable. Electrode needs cleaning, conditioning or
replacing. Check correct buffers are used.
Slope more than 105.0%. Check calibration buffers.
NOTE: Where 0 mV/pH unit = 0%, and
59.16 mV/pH unit = 100% (at 25°C)
Err 3 - pH buffer outside temperature limits pH buffers must be between 5°C and 50°C for
accurate calibration.
. Displayed (out of range)
Check electrode is connected and immersed in sample. Check wetting cap is removed.
Data Entry Errors
Date flashes and reverts to previous setting - invalid date entered.
Entered temperature value changes to –5.0 or 105.0
- the meter will not accept values outside this range.
Dashes displayed instead of buffer value during calibration buffer set up - that buffer has already been selected for a cal point. Cal points must have different values.
Page 12
12
Maintenance
There are no user replaceable parts in the meters or power supply unit. Do not remove the covers.
The meters require very little maintenance. Occasionally wipe the meter with a damp cloth. The casework is made of ABS/PC which is known to be affected by some organic solvents, including toluene, xylene and methyl-ethyl-ketone. It is good practice to wipe away any spillages as soon as they occur.
Electrode Maintenance
Refer to the electrode product insert for full details on maintaining your electrode.
Spares and Accessories
52000100 InLab 413 ‘3 in 1’ pH Electrode 52000118 InLab 410 pH Electrode 51300164 30K NTC Probe 51302118 Electrode Arm and Base 51302152 Recorder Cable (MP225 only) 51302125 RS232 Cable (MP225 only) 00229170 GA42 Printer (MP225 only) 00229119 LC - P45 GLP Printer (MP225 only) 51300047 Guide to pH Measurement 51302069 pH 4.01 Buffer Sachets, 30/pack 51302047 pH 7.00 Buffer Sachets, 30/pack 51302070 pH 9.21 Buffer Sachets, 30/pack 51302068 Rainbow Pack Buffer Sachets
(10 each of pH 4.01, 7.00 and 9.21) 51340058 pH 4.01 Buffer Solution, 250 mL x 6 51340060 pH 7.00 Buffer Solution, 250 mL x 6 51300194 pH 9.21 Buffer Solution, 250 mL x 6
c897
Page 13
13
1.68 2.00 4.00 4.01 6.87 7.00 9.18 9.21 10.01 11.00
5°C 1.67 2.02 4.00 4.01 6.95 7.09 9.39 9.45 10.25 11.72 10°C 1.67 2.01 4.00 4.00 6.92 7.06 9.33 9.38 10.18 11.54 15°C 1.67 2.00 4.00 4.00 6.90 7.04 9.28 9.32 10.12 11.36 20°C 1.68 2.00 4.00 4.00 6.88 7.02 9.22 9.26 10.06 11.18
25°C 1.68 2.00 4.00 4.01 6.87 7.00 9.18 9.21 10.01 11.00
30°C 1.68 1.99 4.01 4.01 6.85 6.99 9.14 9.16 9.97 10.82 35°C 1.69 1.99 4.02 4.02 6.84 6.98 9.10 9.11 9.93 10.64 40°C 1.69 1.98 4.03 4.03 6.84 6.98 9.07 9.06 9.89 10.46 45°C 1.70 1.98 4.04 4.04 6.83 6.97 9.04 9.03 9.86 10.28 50°C 1.71 1.98 4.06 4.06 6.83 6.97 9.01 8.99 9.83 10.10
Buffer Tables
The MP220 and MP225 automatically correct pH buffers for temperature using the values shown in the table.
b697
Page 14
14
Specifications
MP220 MP225
Measurement pH 0.00 to 14.00 –2.00 to 16.00 Ranges mV ± 1999 mV ± 1999 mV
Temp. -5.0 to 105.0°C –5.0 to 105.0°C
Calibration 2 of 3 3 of 3 Points selectable selectable
Resolution pH 0.01 0.01
mV 1 1 Temp. 0.1 0.1
Relative pH ± 0.01 ± 0.01 Accuracy* mV ± 1 mV ± 1 mV
Temp. ± 0.5°C ± 0.2°C
Isopotential pH 7.00 7.00 Point
Outputs Recorder
Serial
* ± 1 least significant digit
MP220 MP225
Memory 10 memories 20 memories Temp. –5.0 to 105.0°C –5.0 to 105.0° C
Compensation auto/manual auto/manual Display LCD LCD Input Conditions Impedance Impedance
>10
12
ohms >1012 ohms
Operating Conditions
Temperature 5 to 40°C 5 to 40°C Humidity at 35°C max. 85% max. 85%
(non condensing) Installation Category 2 2 Pollution Category Degree 2 Degree 2
Size inches
101/2 x 71/2 x 21/2 101/2 x 71/2 x 21/2
mm 265 x 190 x 65 265 x 190 x 65
Weight lb 1
3/4 1 3/4
kg 0.8 0.8
b697
Page 15
15
Specifications (continued)
Power Requirements
The MP220 and MP225 are supplied with an appropriate power supply unit.
USA/Japan 100 - 120V, 50/60Hz, 0.85VA Europe 230V, 50Hz, 1.1VA
Output from PSU: 9V DC Meter Power Rating: 0.3VA
NOTE: The meters should only be used with the power supply unit supplied.
Loading...